ZKM 2 Manual
ZKM 2 Manual
ZIRCONIA OXYGEN
ANALYZER CONVERTER
Type: ZKMA, ZKMB
INZ-TN2ZKMA-E
PREFACE
We are grateful for your purchase of Fuji Direct Insertion Type Zirconia Oxygen Analyzer Converter (ZKM).
First read this instruction manual carefully until an adequate understanding is acquired, and
then proceed to installation, operation and maintenance of the converter. Improper handling
may result in accidents or injury.
The specifications of this converter will be changed without prior notice for further product
improvement.
Modification of this converter is strictly prohibited unless a written approval is obtained
from the manufacturer. Fuji will not bear any responsibility for a trouble caused by such a
modification.
This instruction manual shall be stored by the person who actually uses the converter.
After reading through the manual, be sure to keep it near at hand for future reference.
This product falls within category 9 (Monitoring and control instruments including indus-
trial monitoring and control instruments) of Annex 1 of EU RoHS directive (2011/65/EU).
Please do not use it for consumer use.
This instruction manual should be delivered to the end user without fail.
INZ-TN2ZKMA-E -i-
SAFETY PRECAUTIONS
First of all, read this “SAFETY PRECAUTIONS” carefully, and then use in the correct way.
Installation, transportation, wiring, use, maintenance of this product shall be carried out by suitably trained
personnel.
First-time users should operate the instrument under the supervision of a fully competent person in the
operation.
Be sure to observe the instructions shown below, because they describe important information on safety.
Those safety precautions are ranked in 3 levels, “DANGER”, “CAUTION” and “PROHIBITION”.
The items noted under “ CAUTION” may also result in serious trouble depending on circumstances.
All the items are important and must be fully observed.
- ii - INZ-TN2ZKMA-E
Cautions on wiring
Be sure to turn off all the power before performing wiring. Otherwise
CAUTION electric shock may result.
Be sure to perform class D grounding work.
Otherwise, electric shock or failure may result.
Select a proper wiring material that satisfies the ratings of the instru-
ment. Otherwise, electric shock or fire may result.
Connect power source of correct rating. Connection of a power source
of incorrect rating may lead to a risk of fire.
Cautions on use
Leaving the converter unused for a long time or restarting it after dis-
CAUTION use requires procedures different from normal operation or suspension
procedures. Be sure to follow the instructions in each instruction man-
ual. Otherwise, intended performance may not be achieved, or acci-
dents or injury may result.
Do not operate the converter for a long time with its door left open.
Otherwise, dust, foreign matter, etc. may stick on internal walls, there-
by causing failure.
Before maintenance and check, be sure to turn off the main power
CAUTION supply and wait until the detector is cooled adequately. Otherwise,
you may suffer a burn.
Before removing the detector from the flue for maintenance and
check, make sure the furnace is stopped. Otherwise, you may suffer a
burn.
Before working, take off a wrist watch, finger ring or the like metallic
accessories. And never touch the instrument with a wet hand. Other-
wise, you will have electric shocks.
If the fuse is blown, eliminate the cause, and then replace it with the
one of the same capacity and type as before. Otherwise, it may cause
electric shocks or failure.
Others
INZ-TN2ZKMA-E - iii -
WARRANTY AND MAINTENANCE
1. Scope of application
To use this equipment, the following conditions must be met:
the use of the equipment incurs no risk of a serious accident even if a failure or malfunction occurs on the
equipment, and
in case of product failure or malfunction, safety measures such as redundant design, prevention of mal-
function, fail safe system, foolproof mechanism are provided outside of the equipment.
Be sure to use this instrument under the conditions or environment mentioned in this instruction manual.
Please consult us for specifications for the following applications:
4. Warranty
4-1. Period of warranty
1) Warranty period for this product including accessories is one year after delivery.
2) Warranty period for the parts repaired by our service providers is six months after the completion of repair.
4-2. Scope of warranty
1) If any failure or malfunction attributable to Fuji Electric occurs in the period of warranty, we shall pro-
vide the product after repairing or replacing the faulty part for free of charge at the place of purchase or
delivery.
The warranty does not apply to failure or malfunctions resulting from:
a) inappropriate conditions, environment, handling or usage that is not instructed in a catalog, instruction
book or user's manual, or overuse of the product
b) other devices not manufactured by Fuji Electric
c) improper use, or an alteration or repair that is not performed by Fuji Electric
d) inappropriate maintenance or replacement of expendable parts listed in the instruction book or the
catalog
e) damages incurred during transportation or fall after purchase
f) any reason that Fuji Electric is not responsible for, including a disaster or natural disaster such as
earthquake, thunder, storm and flood damage, or inevitable accident such as abnormal voltage.
- iv - INZ-TN2ZKMA-E
2) Regardless of the time period of the occurrence, Fuji Electric is not liable for the damage caused by the
factors Fuji Electric is not responsible for, opportunity loss of the purchaser caused by malfunction of Fu-
ji Electric product, passive damages, damage caused due to special situations regardless of whether it was
foreseeable or not, and secondary damage, accident compensation, damage to products that were not
manufactured by Fuji Electric, and compensation towards other operations.
5. Failure diagnosis
Regardless of the time period of the occurrence, if any failure occurs, the purchaser shall perform a primary
failure diagnosis. However, at the purchaser's request, Fuji Electric shall provide the diagnosis service for a
fee. In such a case, the purchaser shall be charged for the service.
6. Service life
This product, excluding limited-life parts and consumable parts, is designed for a service life of 10 years un-
der a general condition (average ambient temperature of 30°C).
The service life may be shortened depending on operating conditions and environment. To ensure the service
life, it is important to perform planned maintenance of the product including limited-life parts and consuma-
ble parts.
7. Maintenance plan
Maintenance can be divided into “preventive maintenance” and “corrective maintenance”. Preventive
maintenance can further classified into “daily inspection” and “periodic inspection”. Preventive maintenance
is achieved through systematic implementation of "daily inspection" and "periodic inspection".
INZ-TN2ZKMA-E -v-
8. Limited-life parts and consumable parts
This product contains the following limited-life parts and consumable parts which may affect the service life
of the product itself.
(1) LCD
Design life: approx. five years for continuous use
Symptoms of LCD screen's end-of-life: deterioration of display, backlight failure, etc.
Factors which affect LCD's life: temperature. The life is shortened by half when the temperature rises
by 10°C. (Arrhenius' law)
Replacement: Estimate the lifetime of LCD according to your operating environment, and have the
LCD replaced at appropriate time.
(2) Aluminum electrolytic capacitors
Design life: 5 years under general working conditions (annual average of ambient temperature: 30°C)
Symptoms when a capacitor loses its capacity: deterioration of power quality, malfunction
Factors which affect capacitor life: temperature. The life is shortened by half when the temperature
rises by 10°C. (Arrhenius' law)
Replacement: Estimate the lifetime of capacitor according to your operating environment, and have
the capacitor replaced or overhauled at appropriate time, at least once in 10 years.
Do not use capacitors beyond its lifetime. Otherwise, electrolyte leakage or depletion may cause odor, smoke,
or fire. Please contact Fuji Electric or its service providers when an overhaul is required.
10. Period for repair and provision of spare parts after product discontinuation (maintenance period)
The discontinued models (products) can be repaired for five years from the date of discontinuation. Also,
most spare parts used for repair are provided for five years from the date of discontinuation. However, some
electric parts may not be obtained due to their short life cycle. In this case, repair or provision of spare parts
may be difficult even in the above period.
Please contact Fuji Electric or its service providers for further information.
- vi - INZ-TN2ZKMA-E
Contents
PREFACE ..........................................................................................................................................................i
SAFETY PRECAUTIONS ............................................................................................................................. ii
WARRANTY AND MAINTENANCE...........................................................................................................iv
CHECKING OF CONTENTS OF THE PACKAGE ....................................................................................x
STORAGE CONDITIONS ..............................................................................................................................x
1. GENERAL ....................................................................................................................................................1
1.1 Direct insertion type zirconia oxygen analyzer ........................................................................................1
2. OPERATING PARTS AND THEIR FUNCTIONS ...................................................................................2
2.1 Outline Drawing .......................................................................................................................................2
2.2 Terminal block ..........................................................................................................................................3
2.3 Didplay and operation panel .....................................................................................................................3
2.4 Internal constitution ..................................................................................................................................4
3. INSTALLATION ..........................................................................................................................................5
3.1 Installation site ..........................................................................................................................................5
3.2 How to install the converter ......................................................................................................................6
3.2.1 Mounting on panel (ZKMA).............................................................................................................6
3.2.2 Pipe mounting (ZKMA) .....................................................................................................................6
3.2.3 Mounting on panel (ZKMB) .............................................................................................................7
3.2.4 Pipe mounting (ZKMB) ....................................................................................................................7
4. WIRING AND PIPING ................................................................................................................................8
4.1 Before wiring ............................................................................................................................................8
4.2 Wiring to terminals ...................................................................................................................................9
4.2.1 Wiring to power supply and detector .................................................................................................9
4.2.2 Cable gland and input and output lines ............................................................................................ 10
4.2.3 Allocation of the terminal block ....................................................................................................... 11
4.3 Wiring and piping diagram ..................................................................................................................... 12
4.4 Handling of standard gas (An article on separate order) ........................................................................ 16
5. OPERATION .............................................................................................................................................. 17
5.1 Preparation for operation ........................................................................................................................ 17
5.2 Key operation flow diagram (outline) .................................................................................................... 19
5.3 Initial parameter value table ................................................................................................................... 23
5.3.1 Parameters related to measurement .................................................................................................. 23
5.3.2 Parameters related to calibration ...................................................................................................... 23
5.3.3 Parameters related to blowdown (displayed if the option is provided) ............................................ 24
5.3.4 Parameters related to maintenance ................................................................................................... 24
5.3.5 Other parameters .............................................................................................................................. 25
6. OPERATION START AND STOP .......................................................................................................... 27
6.1 Operation start ........................................................................................................................................ 27
6.2 Operation stop ......................................................................................................................................... 27
6.3 Actions during operation......................................................................................................................... 28
6.4 Check the contents of display ................................................................................................................. 29
6.4.1 Check of state information ............................................................................................................... 29
6.4.2 Checking the error information ........................................................................................................ 30
6.4.3 Checking the alarm information....................................................................................................... 30
6.5 Oxygen detector standard output voltage ............................................................................................... 31
INZ-TN2ZKMA-E - vii -
7. CALIBRATION ......................................................................................................................................... 32
7.1 Preparation ............................................................................................................................................. 32
7.2 Manual calibration ................................................................................................................................. 33
7.3 Auto calibration (option) ........................................................................................................................ 35
7.4 Remote calibration ................................................................................................................................. 37
7.5 All calibration (option) ........................................................................................................................... 39
8. BLOWDOWN (OPTION)......................................................................................................................... 40
8.1 Preparation for blowdown ...................................................................................................................... 40
8.2 Manual blowdown.................................................................................................................................. 40
8.3 Automatic blowdown ............................................................................................................................. 41
8.4 Remote blowdown ................................................................................................................................. 43
9. MAINTENANCE AND CHECK.............................................................................................................. 45
9.1 Checking ................................................................................................................................................ 45
9.2 Spare parts .............................................................................................................................................. 45
9.3 Replacement of fuse ............................................................................................................................... 46
9.4 Adjustment of monitor contrast.............................................................................................................. 47
9.5 Troubleshooting ..................................................................................................................................... 48
10. SETTING AND OPERATING OF PARAMETER ............................................................................... 50
10.1 Measured menu .................................................................................................................................... 50
10.1.1 Display range setting screen .......................................................................................................... 50
10.1.2 Decimal point position setting screen............................................................................................ 51
10.1.3 Full scale setting screen ................................................................................................................. 52
10.1.4 Setting screen for calculation time of maximum and minimum values appears ........................... 53
10.2 Calibration menu .................................................................................................................................. 54
10.2.1 Automatic calibration setting (option) ........................................................................................... 54
10.2.2 Date and time for starting automatic calibration (option) ............................................................. 55
10.2.3 Cycle time setting of automatic calibration (option) ..................................................................... 56
10.2.4 Performing all calibration (option) ................................................................................................ 57
10.2.5 Performing a manual span calibration ........................................................................................... 58
10.2.6 Performing a manual zero calibration ........................................................................................... 59
10.2.7 Calibration gas setting ................................................................................................................... 60
10.2.8 Calibration waiting setting (option)............................................................................................... 61
10.2.9 Calibration error clear ................................................................................................................... 62
10.2.10 Operation setting screen of calibration range .............................................................................. 63
10.3 Blowdown menu (option) .................................................................................................................... 64
10.3.1 Automatic blowdown setting ......................................................................................................... 64
10.3.2 Date and time setting of automatic blowdown .............................................................................. 65
10.3.3 Automatic blowdown cycle setting ............................................................................................... 66
10.3.4 Procedure for setting blowdown time............................................................................................ 67
10.3.5 Perfoming manual blowdown........................................................................................................ 68
10.4 Maintenance menu ............................................................................................................................... 69
10.4.1 Error log display ............................................................................................................................ 69
10.4.2 Clearing error logs ......................................................................................................................... 71
10.4.3 Alarm log display .......................................................................................................................... 72
10.4.4 Clearing alarm logs ....................................................................................................................... 73
10.4.5 Performing a manual sensor check ................................................................................................ 74
10.4.6 Maintenance mode setting ............................................................................................................ 75
10.4.7 Password setting ............................................................................................................................ 76
10.4.8 PID auto tuning ............................................................................................................................. 77
10.5 Parameter menu ................................................................................................................................... 78
10.5.1 Current date and time setting ......................................................................................................... 78
10.5.2 Contact input setting ...................................................................................................................... 79
10.5.3 Selection of alarm contact output .................................................................................................. 80
10.5.4 High limit setting of oxygen concentration ................................................................................... 81
- viii - INZ-TN2ZKMA-E
10.5.5 Lower limit setting of oxygen concentration ................................................................................. 82
10.5.6 HH limit setting of oxygen concentration ...................................................................................... 83
10.5.7 LL limit setting of oxygen concentration ....................................................................................... 84
10.5.8 Hysteresis Setting ........................................................................................................................... 85
10.5.9 Hold treatment setting (maintenance hold) ................................................................................... 87
10.5.10 Hold value setting (maintenance hold)........................................................................................ 88
10.5.11 Setting of hold setting value (maintenance hold) ........................................................................ 89
10.5.12 Setting of measurement recovery time (maintenance hold) ........................................................ 90
10.5.13 Hold treatment setting (error hold).............................................................................................. 91
10.5.14 Hold value setting (error hold) ..................................................................................................... 92
10.5.15 Setting of hold setting value (error hold) .................................................................................... 93
10.5.16 Setting of key lock ....................................................................................................................... 94
10.5.17 Setting of automatic OFF time ..................................................................................................... 95
10.5.18 Station number setting.................................................................................................................. 96
10.5.19 Adjustment screen for analog output 0% ..................................................................................... 97
10.5.20 Adjustment screen for analog output 100% ................................................................................. 98
10.5.21 Fuel coefficient setting (option) ................................................................................................... 99
10.6 Factory menu ...................................................................................................................................... 100
10.6.1 Password setting screen................................................................................................................ 100
11. SPECIFICATIONS ................................................................................................................................ 101
11.1 Specifications ...................................................................................................................................... 101
11.2 Code symbols ...................................................................................................................................... 104
INZ-TN2ZKMA-E - ix -
CHECKING OF CONTENTS OF THE PACKAGE
Check that all of the following are contained in the delivered package.
STORAGE CONDITIONS
-x- INZ-TN2ZKMA-E
1. GENERAL
This manual describes the installation, operation, and the maintenance of the zirconia oxygen analyzer con-
verter. Read it carefully before using the converter. For the detector, flow guide tube and ejector used with
the converter, refer to relevant instruction manuals.
Caution
Power voltage for the converter must conform to that for the detector to be connected. Don’t use any
power voltage different from the power specifications of the detector. Otherwise it may result in damage
to the detector.
100/120V AC50/60Hz for ZFK8R1
200/240V AC50/60Hz for ZFK8R3
Operating environment
1) Operating temperature: -20 to 55°C
2) Operating humidity: 95%RH or less, non condensing
3) Power voltage: 100 to 120V AC 50/60Hz or 200 to 240V AC 50/60Hz
4) Pollution degree: 2
5) Installation category: II
6) Altitude: up to 2000m
INZ-TN2ZKMA-E -1-
2. OPERATING PARTS AND THEIR FUNCTIONS
2.1 Outline Drawing
(1) ZKMA
(2) ZKMB
-2- INZ-TN2ZKMA-E
2.2 Terminal block
① LED
INZ-TN2ZKMA-E -3-
2.4 Internal constitution
(1) ZKMA
② CPU board ① I/O board
③ Communication
board (optional)
④ Fuse
(2) ZKMB
You can see the CPU board, the I/O board, and the communication board (optional) if you remove the M3
screw and open the inner cover.
M3 screw
④ Fuse
⑦ Packing
⑤ Terminal block ⑥ Earth terminal
-4- INZ-TN2ZKMA-E
3. INSTALLATION
DANGER
This unit is not explosion-proof type. Do not use it in an atmosphere of explosive gases. Otherwise
serious accidents such as explosion or fire may result.
CAUTION
For installation, observe the rule on it given in the instruction manual and select a place where the
weight of converter can be endured. Installation at an unsuited place may cause turnover or fall and
there is a risk of injury.
Before transport, fix the door so that it will not open. Otherwise, the casing may be separated and fall
to cause an injury.
During installation work, care should be taken to keep the unit free from entry of cable chips or other
foreign objects. Otherwise, it may cause fire, failure or malfunction.
Secure at least 100 mm of space between the converter and nearby wall. Also secure a space of opening the
front cover for maintenance.
Secure a cable wiring space under the case.
N
E
OP 100
or
147 147 more
400 or more
Unit: mm
E
OP 100
or
240 240 more
580 or more
Unit: mm
INZ-TN2ZKMA-E -5-
3.2 How to install the converter
Installation
-6- INZ-TN2ZKMA-E
3.2.3 Mounting on panel (ZKMB)
Unit: mm
Mounting dimensions Installation
Installation
2
5
3,4
INZ-TN2ZKMA-E -7-
4. WIRING AND PIPING
CAUTION
: Wiring work must be carried out with all power supplies turned off. Otherwise electric
shock may result.
: Be sure to ground the Converter. (Class D grounding)
(3) Provide adequate protection of the exclusive cable (6 cores in total), which connects the detector to
converter, using wire protection tube, etc. Separate these cables from the power cable (noise preven-
tion).
(4) Keep the wire for output signals as far as possible (more than 30cm) from the power line and heavy
current lines to prevent induced noise. Also, wherever possible use a shielded cable and earth one
point of the shield.
Note) For connection of the lines to the external terminals, use of ring crimp solderless terminal
with insulation sleeve is recommended.
-8- INZ-TN2ZKMA-E
4.2 Wiring to terminals
(2) <ZKMB>
Note 1: Fix the exclusive cable (O2 sensor input / O2 sensor thermocouple input) with the cable gland so
that its sheath is 10 mm or less.
When attaching the nut, turn it by hand until it does not move and then tighten it with a spanner
by about 1/4 turn.
Note 2: Wire a protective earth to earth terminal of above figure. (Class D, grounding resistance: 100Ω
or less)
Note 3: Wind the power cable around the ferrite core twice.
INZ-TN2ZKMA-E -9-
4.2.2 Cable gland and input and output lines
(1) <ZKMA>
(2) <ZKMB>
- 10 - INZ-TN2ZKMA-E
4.2.3 Allocation of the terminal block
INZ-TN2ZKMA-E - 11 -
4.3 Wiring and piping diagram
4.3.1.1 Flow guide tube system
- 12 - INZ-TN2ZKMA-E
4.3.1.2 Flow guide tube system (with valve)
INZ-TN2ZKMA-E - 13 -
4.3.1.3 Ejector system
- 14 - INZ-TN2ZKMA-E
4.3.1.4 Ejector system (with valve)
INZ-TN2ZKMA-E - 15 -
4.4 Handling of standard gas (An article on separate order)
4.4.1.1 Operation
(1) Make sure the handle is closed on the high pressure gas container, then detach the cap nut.
(2) Attach the high pressure gas container using the cap nut with packing of the pressure reduction valve.
(3) Make sure the secondary pressure adjusting valve is turned fully counterclockwise (pressure not ap-
plied) and the outlet needle is turned fully clockwise (closed), then open the handle.
(4) Turn the secondary pressure adjusting valve clockwise and set to the normal value of 20 to 30 kPa,
then open the outlet needle slowly to allow the gas to flow.
Cap nut
Packing
Secondary pressure
adjusting valve
Outlet needle
High pressure
gas container
4.4.1.2 Piping
(1) The gas outlet of the pressure reduction valve is of Rc 1/4 (internal thread).
Prepare the joint and tube (such as 6/ 4 teflon tube).
4.4.1.3 Caution
(1) Fasten securely, so there is no gas leakage from the pressure reduction valve connection or from
threaded part of the joint.
(2) Store high pressure gas containers in a place protected from direct sunlight and rain.
(3) After use, be sure to close the handle.
- 16 - INZ-TN2ZKMA-E
5. OPERATION
5.1 Preparation for operation
Preparation can be performed after installation or on the bench.
(2) Confirmation of the power supply specifications (Please check the main power supply and the
power supply voltage specification of the detector.)
OXYGEN ANALYZER
The message shown left appears on the LCD screen.
VER . YY/MM
(4) Warm-up
Warm-up takes 10 minutes or less. Then the analyzer starts temperature control for the detector
ZFK.
Note 1: If the temperature control completes within 10 minutes, the analyzer stops warm-up to
start operation. When you first use the analyzer, proceed to the step 5 “PID auto tuning”.
Note 2: Even if the temperature control does not complete within 10 minutes, the analyzer stops
warm-up to start operation, however, for 7 minutes after the warm-up stoped, the analyzer
cannot detect the error of the heater temperature nor display it. (Carry out the step 5 “PID
auto tuning” during this period.)
(7) Calibration
At the first operation, perform manual calibration after warm-up using a calibration gas.
Refer to “chapter 7” for calibration procedures.
INZ-TN2ZKMA-E - 17 -
(8) Auto calibration (option)
Automatic calibration may be performed at specified time intervals.
Refer to “7.3” for automatic calibration settings.
Operation
- 18 - INZ-TN2ZKMA-E
5.2 Key operation flow diagram (outline)
INZ-TN2ZKMA-E - 19 -
- 20 - INZ-TN2ZKMA-E
INZ-TN2ZKMA-E - 21 -
- 22 - INZ-TN2ZKMA-E
5.3 Initial parameter value table
Reference
Parameter setting Displayed message Range Initial value
paragraph
DISPLAY RANGE
Display range Range1 or Range2 Range-1 10.1.1
RANGE1 RANGE2
Reference
Parameter setting Displayed message Range Initial value
paragraph
Auto calibration function AUTO CALIBRATION Invalid
(Displayed if the option is YES or NO (Auto calibration 10.2.1
YES NO
provided.) function: Invalid)
Date and time for starting
automatic calibration START DATE Date and time in the fu-
99/01/01 00:00 10.2.2
(Displayed if the option is 99/01/01 00:00 ture in the calendar
provided.)
Automatic calibration
cycle time AUTO CAL. CYCLE 00d 00h to 99d23h
07d 00h 10.2.3
(Displayed if the option is 07d 00h (h: 00 to 23)
provided.)
Span:
Calibration gas concen- Span:
SPAN ZERO 00.010 to 50.000 vol%
tration-1 20.600 vol%
Zero: 10.2.7
calibration gas concentra- 20.600% 01.000% Zero:
00.010 to 25.000 vol%
tion-2 01.000 vol%
in 0.001 vol% steps
INZ-TN2ZKMA-E - 23 -
5.3.3 Parameters related to blowdown (displayed if the option
is provided)
Reference
Parameter setting Displayed message Range Initial value
paragraph
NO
Automatic blowdown BLOW DOWN (The automatic
YES or NO 10.3.1
function YES NO blowdown func-
tion is invalid.)
Date and time for starting START DATE Date and time in the fu-
99/01/01 00:00 10.3.2
automatic blowdown 99/01/01 00:00 ture in the calendar
Reference
Parameter setting Displayed message Range Initial value
paragraph
MAINTENANCE MODE
Maintenance mode YES or NO NO 10.4.6
YES NO
NEW PASSWORD
Password 0000 to 9999 0000 10.4.7
0123
- 24 - INZ-TN2ZKMA-E
5.3.5 Other parameters
Reference
Parameter setting Displayed message Range Initial value
paragraph
DI1 to DI3
[NONE]
DI1
[BLOW DOWN ON]
[NONE]
DI 1 [HEATER OFF]
DI2
Contact inputs 1 to 3 [PROHIBIT CAL.] 10.5.2
NONE [NONE]
[REMOTE CAL.]
DI3
[REMOTE HOLD]
[NONE]
[CALCULATE REST]
[OUTPUT RANGE]
[ALARM NONE]
[HIGH ALARM]
DO ALARM SET [LOW ALARM]
Alarm contact output [H-HIGH ALARM] [ALARM NONE] 10.5.3
ALARM NONE
[L-LOW ALARM]
[H/L ALARM]
[HH/LL ALARM]
Upper limit of oxygen
concentration HIGH ALARM 0.001 to 55.000 vol% in
50.000 vol% 10.5.4
Range-1 50.000 vol% 0.001 vol% steps
Range-2
Lower limit of oxygen
concentration LOW ALARM 0.001 to 55.000 vol% in
00.020 vol% 10.5.5
Range-1 00.020 vol% 0.001 vol% steps
Range-2
Upper 2 limit of oxygen
concentration H-HIGH ALARM 0.001 to 55.000 vol% in
55.000 vol% 10.5.6
Range-1 55.000 vol% 0.001 vol% steps
Range-2
Lower 2 limit of oxygen
concentration L-LOW ALARM 0.001 to 55.000 vol% in
00.010 vol% 10.5.7
Range-1 00.010 vol% 0.001 vol% steps
Range-2
Hysteresis
(Oxygen concentration HYSTERESIS
alarm) 0 to 20 % in 1 % steps 10 % 10.5.8
10 %
Range-1
Range-2
Analog output hold func- NO
tion OUTPUT HOLD (Analog output 10.5.9
YES or NO
Maintenance hold YES NO hold function is 10.5.13
Error hold invalid.)
Output value of analog [0 %] (4 mA/0V)
output hold OUTPUT SELECT [100 %] (20 mA/1V) 10.5.10
[0 %](4 mA/0V)
Maintenance hold 0% [Last output value] 10.5.14
Error hold [Setting value]
Setting the value of ana-
log output hold HOLD VALUE 10.5.11
0 to 100 % in 1 % steps 0%
Maintenance hold 000 % 10.5.15
Error hold
INZ-TN2ZKMA-E - 25 -
Reference
Parameter setting Displayed message Range Initial value
paragraph
KEY LOCK No
Key lock function YES or NO (Key lock func- 10.5.16
YES NO
tion is invalid.)
BACKLIGHT TIME 0 to 99
Automatic OFF time 10 minutes 10.5.17
10 m in 1-minute steps
STATION NO
Station No. 0 to 99 01 10.5.18
01
FUEL COEFF.
FUEL COEFFICIENT 0.00 to 1.99 0.70 10.5.21
0.70
- 26 - INZ-TN2ZKMA-E
6. OPERATION START AND STOP
6.1 Operation start
After correct wiring and piping has been completed, turn on the converter to start measurement.
Note: 10 min. of warm-up time is necessary after power ON.
INZ-TN2ZKMA-E - 27 -
6.3 Actions during operation
While the instrument is operating, the following displays can be changed.
- 28 - INZ-TN2ZKMA-E
6.4 Check the contents of display
The condition of the unit is displayed on the left of the LCD with three letters. The maximum of three items
are displayed on one display. If there are four or more items, “▼” is displayed at the bottom of the screen.
Scroll the screen with the key to display the fourth and subsequent items.
The unit displays the following three pieces of information:
(1) Condition information (“6.4.1”), (2) Error information (“6.4.2”), (3) Alarm information (“6.4.3”)
INZ-TN2ZKMA-E - 29 -
6.4.2 Checking the error information
Display
State Remarks
message
Appears when control temperature of the heater exceeds
Er1 Fault of heater temperature the set range.
The heater control is stopped.
Appears when disconnection is detected at the sensor, or
Er2 Disconnection detection thermocouples for temperature control.
The heater control is stopped.
Er3 Sensor error Appears when the A/D value is saturated.
Appears when the span calibration is abnormal.
Er4 Span calibration error (The calibration gas is unstable. / The calibration factor
setting is inappropriate.)
Appears when the zero calibration is abnormal.
Er5 Zero calibration error (The calibration gas is unstable. / The calibration factor
setting is inappropriate.)
You can select one of the following seven alarms to output to the alarm contact (Contact No. 21 and 22 of the
external terminal blocks) when an oxygen concentration error occurs.
(1) [Not used] : No alarm is output to the contact output.
(2) [High limit alarm] : Alarm contact is output when an high limit alarm occurs.
(3) [Lower limit alarm] : Alarm contact is output when a lower limit alarm occurs.
(4) [HH limit alarm] : Alarm contact is output when an HH limit alarm occurs.
(5) [LL limit alarm] : Alarm contact is output when a LL limit alarm occurs.
(6) [High/lower limit alarm] : Alarm contact is output when an high or lower limit alarm occurs.
(7) [HH / LL limit alarm] : Alarm contact is output when an HH or LL limit alarm occurs.
- 30 - INZ-TN2ZKMA-E
6.5 Oxygen detector standard output voltage
O2 concentration Output value O2 concentration Output value O2 concentration Output value
(%) (mV) (%) (mV) (%) (mV)
0.01 176.38 5.0 32.73 25.0 -4.475
0.1 123.15 10.0 16.71 30.0 -8.689
0.5 85.95 15.0 7.333 40.0 -15.34
1.0 69.93 20.0 0.683 50.0 -20.50
1.5 60.56 20.6 0 – –
2.0 53.91 21.0 -0.445 – –
INZ-TN2ZKMA-E - 31 -
7. CALIBRATION
In order to maintain good accuracy, proper calibration using standard gas is necessary. The following 4
methods of calibration are provided.
(1) Manual calibration (“7.2”), (2) Auto calibration (option) (“7.3”),
(3) Remote calibration (“7.4”), (4) All calibration (option) (“7.5”)
7.1 Preparation
Check of piping and wiring
Perform wiring and piping correctly referring to Item “4.3”. At this time, the main valve of standard
gas should be left open. Since high pressure is present at piping connections, use cap nut joints and take
special care with regard to air-tightness. Calibration gas flow should be 1.5 ± 0.5 L/min.
Setting of calibration gas concentration
Referring to “10.2.7 Calibration gas setting” set the oxygen concentration in standard gas cylinder to
be used.
Setting of calibration range
Set the range for calibration according to “10.2.10 Operation setting screen of calibration range.”
- 32 - INZ-TN2ZKMA-E
7.2 Manual calibration
Description
Span/zero is calibrated once by key operation.
Calibration must be made in the order of span and zero.
Perform calibration after a calibration gas is supplied to the detector and the output signal of the de-
tector becomes stable.
If the unit does not have an auto calibration function, the operator shall perform open and close opera-
tions, or adjust the flow rate of calibration gas.
During calibration, if the analog output hold function (maintenance hold) is enabled, the analog output
signal is held at the set value. Even after the calibration, the hold is maintained during the set time as a
measurement recovery time.
Operation
Procedure
(6) If the operator opened the span gas valve manually, close the
valve.
(7) Display the screen on the right in accordance with the key op-
CALIBRATION MENU
eration summary and press the key, the manual zero
calibration screen appears. MANUAL ZERO CAL.
INZ-TN2ZKMA-E - 33 -
(8) Press the key to perform manual zero calibration.
MANUAL ZERO CAL.
If supplying calibration gas manually (without the auto-
START
calibration function)
The operator shall open the span gas valve manually and adjust
the flow rate to 1.5 ± 0.5 L/min.
In the case of unit with the auto calibration function, an exter-
nal solenoid valve can be driven by using the contact output
signal from the terminal block.
(9) Oxygen concentration value and cell electromotive force are
displayed. MANUAL ZERO CAL.
Wait until the oxygen concentration is stabilized. 2.01 % 053.9 mV
(12) If the operator open the zero gas valve manually, close the
valve.
How to cancel
Press the key to cancel the operation.
After the cancellation, be sure to close the valves of span gas and zero gas.
- 34 - INZ-TN2ZKMA-E
7.3 Auto calibration (option)
Description
Calibration is performed at time intervals set in advance.
The solenoid valve is driven by contact signal to feed the standard gas for automatic calibration with
span gas and zero gas.
“CAL” is displayed on the left of the measurement screen during automatic calibration.
If the output signal hold is set, the output signal is held to the set value during calibration.
After the calibration, the hold is maintained until the time set in the measurement recovery time elaps-
es.
For automatic calibration, it is necessary to set “10.2.2 Date and time for starting automatic calibra-
tion (option)”, “10.2.3 Cycle time setting of automatic calibration (option)”, “10.2.7 Calibration
gas setting”, “10.2.8 Calibration waiting setting (option)”, and “10.5.12 Setting of measurement re-
covery time (maintenance hold)”.
Refer to Sections “4.2” and “4.3” for the wiring of solenoid valves.
Operation Setting the automatic calibration so that it is performed every four days from 13:00,
Procedure
(example) 2015/02/25
Displayed message
Key operation Description
(LCD)
(1) Display the screen on the right in accordance with the key op-
CALIBRATION MENU
eration summary and press the key.
SET AUTO CAL
(3) Use the key to select the auto calibration valid (YES).
AUTO CALIBRATION
Press the key to set the value.
YES NO
INZ-TN2ZKMA-E - 35 -
(5) The screen on the right appears.
SET AUTO CAL
AUTO CALIBRATION
(6) Press the key to display the screen on the right and press
SET AUTO CAL
the key.
START DATE
The date and time for starting automatic calibration screen ap-
pears.
(7) Use the and key to set the auto calibration starting
START DATE
date and time.
(Set the date and time of the future.) 15/02/25 13:00
Press the key to set the value.
(8) Press the key.
START DATE
15/02/25 13:00
(10) Press the key to display the screen on the right and press
SET AUTO CAL
the key.
AUTO CAL. CYCLE
The cycle time setting of automatic calibration screen appears.
(11) Use the and key to set the auto calibration cycle
AUTO CAL. CYCLE
time.
04d 00h
Press the key to set the value.
How to cancel
Press the key to cancel the operation.
Note
Automatic calibration is not performed under the following conditions.
Warm-up is being performed.
Contact of “Prohibition of calibration” is being input.
Contact of “Heater off” is being input.
- 36 - INZ-TN2ZKMA-E
7.4 Remote calibration
This function is available only for the version with auto-calibration.
You can perform all calibration by the contact input of the external terminal block.
To perform remote calibration, install piping and wiring for the standard gas cylinder and the solenoid valve
according to Section 4.
(1) Set one of the contact inputs DI 1 to 3 to “Remote calibration” in accordance with the following op-
eration procedure.
(2) Close the contact set to the “Remote contact” for one second or more (depending on the settings of
(17) to (19) and (20) of the terminal block).
(3) Remote calibration is started. “RCL” is displayed on the left of the display panel, which disappears
when the calibration is completed.
17 18 19 20
You can arbitrarily set the contact inputs (17), (18), (19) and (20) of the external terminal block (see “10.5.2
Contact input setting”).
Piping and wiring for the standard gas cylinder and the solenoid valve shall be installed.
Description
You can perform all calibration by the contact input using this function.
The solenoid valve is driven by contact signal from the terminal block to feed the standard gas for
automatic calibration with span gas and zero gas.
Refer to Sections “4.2” and “4.3” for the wiring of solenoid valves.
Operation
Procedure
(4) Press the key several times and select “REMOTE CAL.”.
DI 1
Press the key to set the value.
REMOTE CAL.
INZ-TN2ZKMA-E - 37 -
(5) Press the key.
DI 1
REMOTE CAL.
How to cancel
Press the key to cancel the operation.
Note
Automatic calibration is not performed under the following conditions.
Warm-up is being performed.
Contact of “Remote blow” is being input.
Contact of “Prohibition of calibration” is being input.
Contact of “Heater off” is being input.
- 38 - INZ-TN2ZKMA-E
7.5 All calibration (option)
Description
Perform sensor maintenance [sensor check (setting), sensor recovery (setting)], span and zero calibra-
tion once for each sequentially by key operation.
Actuate the solenoid valve attached to the exterior by the contact signal from the terminal block and
supply standard gases sequentially. Span and zero gas calibration are automatically performed.
If the output signal hold is set, the output signal is held to the set value during calibration. After the
calibration, the hold is maintained until the time set in the measurement waiting time elapses.
To perform sensor maintenance (sensor check, sensor recovery) during calibration, “10.4.5 Perform-
ing a manual sensor check” are required.
Note that the sensor recovery is performed if it is determined to be required by the sensor check.
Refer to Sections “4.2” and “4.3” for the wiring of solenoid valves.
Operation
Procedure
(3) The value of the concentration of oxygen and the cell electro-
motive force are displayed while executing the all calibration. ALL CALIBRATION
20.61 % 000.1 mV
(4) After the all calibration is completed, the display returns to the
screen on the right. CALIBRATION MENU
ALL CALIBRATION
How to cancle
Press the key to cancel the operation.
INZ-TN2ZKMA-E - 39 -
8. BLOWDOWN (OPTION)
In order to prevent the flow guide tube from clogging with dust contained in gas being measured, dust depos-
its in the flow guide tube is removed by blowing compressed air such as instrumentation air, etc. Use the
blowdown function by one of the following three methods.
(1) Manual blowdown (“8.2”), (2) Automatic blowdown (“8.3”),
(3) Remote blowdown (“8.4”)
Description
You can perform blowdown operation once by key operation using this function.
Operation
Procedure
How to cancel
Press the key to cancel the operation.
- 40 - INZ-TN2ZKMA-E
8.3 Automatic blowdown
Description
Blowdown operation is performed at time intervals set in advance.
Using contact signal from the terminal block, drive the solenoid valve and remove dust by blowing
instrumentation air, etc. into the flow guide tube with blowdown nozzle.
“BLW” is displayed on the left of the measurement screen during automatic blowdown.
If the output signal hold is set, the output signal is held to the set value before start of blowdown dur-
ing blowdown. After the calibration, the hold is maintained until the time set in the measurement wait-
ing time elapses.
To perform automatic blowdown, “10.3.2 Date and time setting of automatic blowdown” and
“10.3.3 Automatic blowdown cycle setting” and “10.3.4 Procedure for setting blowdown time” are
required.
Automatic blowdown
starting date
Automatic blowdown
interval
Blowdown contact Closed Closed
Operation Setting the blowdown so that it is performed for 30 seconds every 24 hours from 13:00,
Procedure
(example) 08/02/25
Displayed message
Key operation Description
(LCD)
(1) Display the screen on the right in accordance with the key op-
BLOW DOWN MENU
eration summary and press the key.
SET AUTO BLOW
(3) Use the key to select the auto blowdown valid (YES).
BLOW DOWN
Press the key to set the value.
YES NO
(6) Press the key to display the screen on the right and press
SET AUTO BLOW
the key.
START DATE
The date and time setting of automatic blowdown screen ap-
pears.
INZ-TN2ZKMA-E - 41 -
(7) Use the and key to set the auto blowdown starting
START DATE
date and time.
15/02/25 13:00
(Set the date and time of the future.)
Press the key to set the value.
(8) Press the key.
START DATE
15/02/25 13:00
(10) Press the key to display the screen on the right and press
SET AUTO BLOW
the key.
AUTO BLOW CYCLE
The auto setting blowdown interval screen appears.
(11) Use the and key to set the auto blowdown interval.
AUTO BLOW CYCLE
Press the key to set the value.
24h 00m
(14) Press the key to display the screen on the right and press
SET AUTO BLOW
the key.
BLOW DOWN TIME
The setting blowdown time screen appears.
(15) Use the and key to set the blowdown time.
BLOW DOWN TIME
(Common with the manual blowdown.)
030 S
Press the key to set the value.
How to cancel
Press the key to cancel the operation.
- 42 - INZ-TN2ZKMA-E
8.4 Remote blowdown
You can perform blowdown by the contact input of the external terminal block.
To perform remote blowdown, install piping and wiring for the supply air and the solenoid valve according
to Section 4.
(1) Set one of the contact inputs DI 1 to 3 to “Blowdown ON” in accordance with the following opera-
tion procedure.
(2) Close the contact set to the “Blowdown ON” for one second or more (depending on the settings of
(17) to (19) and (20) of the terminal block).
(3) Blowdown is started. “RBL” is displayed on the left of the display panel, which disappears when the
blowdown is completed. Blowdown continues for the time you set in “10.3.4 Procedure for setting
blowdown time”.
17 18 19 20
You can arbitrarily set the contact inputs (17) to (19) and (20) of the terminal block (see “10.5.2 Contact
input setting”).
Piping and wiring for the supply air and the solenoid valve shall be installed.
Description
You can perform blowdown by the contact input using this function.
Actuate the solenoid valve attached to the exterior by the contact signal from the terminal block and
flow supply air sequentially. Blowdown is automatically performed.
Refer to Sections “4.2” and “4.3” for the wiring of solenoid valves.
Operation
Procedure
(4) Press the key several times and select “BLOW DOWN
DI 1
ON”.
BLOW DOWN ON
Press the key to set the value.
INZ-TN2ZKMA-E - 43 -
(5) Press the key.
DI 1
BLOW DOWN ON
How to cancel
Press the key to cancel the operation.
- 44 - INZ-TN2ZKMA-E
9. MAINTENANCE AND CHECK
9.1 Checking
Please regularly check the instrument and carry out maintenance work to use it always in good condition.
Check especially the following items. Perform maintenance and check every year or two.
Calibration gas pressure Replace calibration gas cylinder when the primary
pressure becomes around 1 MPa.
Check for loose cable gland Check if the screw is tightened.
If the screw is loose, tighten it.
Case packing If there are any cracks on the surface of packing,
Periodic inspection
Packing of cable gland If there is any cracks in the packings inside cable
glands, replace the packing with new one. Pay
attention to the exposed surface of packing if
there are any cracks.
INZ-TN2ZKMA-E - 45 -
9.3 Replacement of fuse
If a fuse blows, turn off the power, and replace the fuse after investigating the cause.
Open the front door and you can see the a fuse on the case.
To replace the fuse, insert a flathead screwdriver or coin into the fuse cap and turn it to the left while pressing
it in order to remove the cap and replace the fuse.
Put the cap on the fuse and turn it to the right to fix it.
- 46 - INZ-TN2ZKMA-E
9.4 Adjustment of monitor contrast
You can adjust the monitor contrast to improve the readability.
VR1 Open the cover, and you can see the volume
knob (VR1) on the CPU board. Turn the
knob with a flathead screwdriver to adjust
the contrast.
ZKMA
VR1
ZKMB
INZ-TN2ZKMA-E - 47 -
9.5 Troubleshooting
Phenomena Probable causes Checking methods (normal value) Remedy
No display Converter fuse blown out Check the fuse and supply voltage speci- Replace fuse Check
fication. Power supply voltage
Indication does Filter and/or flow guide tube Visual check of filter and flow guide Clean or replace filter
not change or clogged tube for contamination or clogging. Tighten pipe connec-
slow response Check for loosen and gas leaks at piping tions
connections and mounting place of de-
tector.
Sensor deterioration Change over between zero and span gas Replace sensor
and check if 5 minutes or longer is need-
ed for 90% response.
Decrease in flow velocity of Check response to process gas after Increase process gas
exhaust gas shutting down calibration gas. Flow into the flow
Move the direction (mounting position) guide tube.
of “arrow” of the flow guide slightly.
Temperature Break of wiring Cable check of wiring Replacement
alarm contin- Wrong wiring Wiring check Correct wiring
ues for more Source voltage is too low. Check of supply voltage specification Check power supply
than 10 min. Break of thermocouples Break check Replace sensor
after power
switched ON Blown heater fuse Cable check of fuse Replace fuse
Break in detector heater Check heater resistance Replace sensor
50 to 55Ω for 115V, 200 to 250Ω for
220V (Excluding wiring resistance)
Automatic Difference between calibration Check the set value for calibration gas Set proper value (Re-
calibration is gas concentration and its setting concentration. fer to “10.2.7”)
not possible Wrong parameters setting Check automatic calibration intervals. Set proper parame-
ters
The calibration is prohibited in Check if the calibration is not prohibited Set proper parame-
the contact input of the external in the contact input of the external ter- ters
terminal block. minal block. Correct wiring
The heater is set to off at the Check if the heater is set to off at the Set proper parame-
contact input of the external contact input of the external terminal ters
terminal block. block. Correct wiring
Zero and/or Difference between calibration Check the set value for calibration gas Set proper value
span alarm gas concentration and its setting concentration.
or misconnection between zero Check piping. Correct wiring
and span gas
Indication too Loose flange and its surround- Check for gas leaks in detector and Tighten mounting
high or too low ings Deteriorated O-rings mounting part of flow guide tube flange. screws
Replace sensor
Check for leaks from the outside. Seal
Detector is faulty. Check for gas leaks at calibration gas Tighten connectors
inlet. Replace sensor
Check sensor voltage (mV) for higher or
lower than other detector when flowing
zero gas. (See “6.5 Oxygen detector
standard output voltage”)
Abnormal detector temperature Refer to check items for detector tem- Replace sensor
perature alarm described above.
Indication difference between Oxygen concentration is higher in dry Normal
dry and wet base measurement base.
Unstable read- The power supply voltage is too Check that the supplied voltage is as Auto-tuning
ings high or too low. specified.
- 48 - INZ-TN2ZKMA-E
Phenomena Probable causes Checking methods (normal value) Remedy
Disconnection Break of thermocouples Cable check of wiring Replace the defec-
detection error Break of sensor Wiring check tive parts.
Wrong wiring Correct wiring
Turn on/off the
power supply.
Range cannot “Range setting” is set in the Check if “Range setting” is set in the Cancel “Range set-
be switched. contact input setting. contact input setting. ting” in the contact
input setting.
INZ-TN2ZKMA-E - 49 -
10. SETTING AND OPERATING OF PARAMETER
10.1 Measured menu
Operation
Procedure
(4) When it is fixed, the display returns to the screen on the right.
MEASURE MENU
DISPLAY RANGE
Note
If “Range setting” is set in the contact input setting, you cannot change the display range on this
screen.
- 50 - INZ-TN2ZKMA-E
10.1.2 Decimal point position setting screen
Description
You can set the decimal point position of full scale for oxygen concentration display using this func-
tion.
Settable range: Select one of the following.
(1) “00.00”: Displayed with two-digit integer and two decimal places.
(2) “0.000”: Displayed with one-digit integer and three decimal places.
Operation
Procedure
Setting the display of two-digit integer and two decimal places (Range 1)
(example)
Displayed message
Key operation Description
(LCD)
(1) Display the screen on the right in accordance with the key op-
MEASURE MENU
eration summary and press the key.
OUTPUT RANGE
(3) Use the key to select the two-digit integer and two deci-
DECIMAL POINT
mal places.
00.00
Press the key to set the value.
(5) When it is fixed, the display returns to the screen on the right.
RANGE 1
DECIMAL POINT
Note
If changing “0.000” to “00.00,” “25.00” is set as the full scale value.
If changing “00.00” to “0.000,” “5.000” is set as the full scale value.
INZ-TN2ZKMA-E - 51 -
10.1.3 Full scale setting screen
Description
You can set the full scale value for display of oxygen concentration value using this function.
Settable range: If the decimal point position is set to “00.00”: 02.00 to 50.00 vol%
If the decimal point position is set to “0.000”: 2.000 to 9.000 vol%
Operation
Procedure
(2) Press the key to display the screen on the right and press
RANGE 1
the key.
FULL SCALE
The full scale setting screen appears.
(3) Use the and key to set the full scale value.
FULL SCALE
Press the key to set the value.
20.00
- 52 - INZ-TN2ZKMA-E
10.1.4 Setting screen for calculation time of maximum and min-
imum values appears
Description
You can set the calculation time of maximum and minimum values of oxygen concentration value
using this function.
Settable range: 0 to 240h
Operation
Procedure
(4) When it is fixed, the display returns to the screen on the right.
MEASURE MENU
CALCULATE TIME
INZ-TN2ZKMA-E - 53 -
10.2 Calibration menu
Operation
Procedure
(3) Use the key to select the auto calibration valid (YES).
AUTO CALIBRATION
Press the key to set the value.
YES NO
Note
If the time for automatic calibration comes during manual treatments (calibration, blowdown, or sen-
sor check) or remote treatments (calibration, blowdown, or heater off), the treatment being performed
is prioritized and the automatic calibration starts after the treatment is completed.
If the time for automatic calibration comes at the same time with automatic blowdown, the automatic
blowdown starts first and the automatic calibration starts after the automatic blowdown is completed.
“Prohibition of calibration” is set in the contact input setting and the contact input is on, automatic
If
calibration is not performed.
If disconnection is detected (O2 sensor input, O2 sensor thermocouple input, or thermocouple input
(combustion control: option)), or a heater temperature error or A/D saturation error occurs, automatic
calibration is not performed.
Automatic calibration is not available during warm-up operation.
- 54 - INZ-TN2ZKMA-E
10.2.2 Date and time for starting automatic calibration (option)
Description
You can set the date and time for starting automatic calibration using this function.
Automatic calibration is performed in a specified cycle from a specified date and time.
If it is invalid, the automatic calibration does not start at a specified date and time.
Settable range: date and time in the future in the calendar
Operation
Procedure
(2) Press the key to display the screen on the right and press
SET AUTO CAL
the key.
START DATE
The auto calibration starting date and time setting screen ap-
pears.
(3) Use the and key to set the auto calibration starting
START DATE
date and time.
15/01/01 00:00
Press the key to set the value.
Note
You cannot change the setting value during automatic calibration or remote calibration.
Check that “Current date and time setting” in the parameter menu is properly set.
INZ-TN2ZKMA-E - 55 -
10.2.3 Cycle time setting of automatic calibration (option)
Description
You can set the automatic calibration cycle using this function.
The cycle starts from a specified date and time for automatic calibration.
Settable range: 00d 00h to 99d 23h (h: 00 to 23)
Operation
Procedure
(2) Press the key to display the screen on the right and press
SET AUTO CAL
the key.
AUTO CAL. CYCLE
The cycle time setting of automatic calibration appears.
(3) Use the and key to set the auto calibration starting
AUTO CAL. CYCLE
date and time.
04 d 00 h
Press the key to set the value.
Note
You cannot change the setting value during automatic calibration or remote calibration.
- 56 - INZ-TN2ZKMA-E
10.2.4 Performing all calibration (option)
Description
You can perform all calibration on the screen using this function.
Zero calibration is automatically performed after the span calibration.
You cannot perform “All calibration” during warming-up operation.
Operation
Procedure
How to cancel
Press the key to cancel the operation.
INZ-TN2ZKMA-E - 57 -
10.2.5 Performing a manual span calibration
Description
Before starting span calibration, the operator shall supply span gas to the detector and check that the
display is stabilized.
You cannot perform manual span calibration during warming-up operation.
Operation
Procedure
(6) If the operator opened the span gas valve manually, close the
valve.
How to cancel
Press the key to cancel the operation.
After the cancellation, be sure to close the valves of span gas.
- 58 - INZ-TN2ZKMA-E
10.2.6 Performing a manual zero calibration
Description
Before starting zero calibration, the operator shall supply zero gas to the detector and check that the
display is stabilized.
You cannot perform manual zero calibration during warming-up operation.
Operation
Procedure
(6) The operator shall close the zero gas valve manually.
How to cancel
Press the key to cancel the operation.
After the cancellation, be sure to close the valves of zero gas.
INZ-TN2ZKMA-E - 59 -
10.2.7 Calibration gas setting
Description
Set calibration gas concentration (span/zero calibration gas concentrations).
Use the calibration gas concentration 1 for the range 1, and the calibration gas concentration 2 for the
range 2.
Use normal air (atmosphere) as a span calibration gas and set its concentration to 20.600% O2/N2.
Settable range: Span calibration gas 00.010 to 50.000 %O2/N2
Zero calibration gas 00.010 to 25.000 %O2/N2
Operation
Procedure
(3) Use the and key to change the calibration gas con-
SPAN ZERO
centrations.
20.600% 02.000%
Press the key to set the value.
Note
You cannot change the setting value during automatic calibration or remote calibration.
Set with span calibration gas concentrations ≥ zero calibration gas concentrations.
- 60 - INZ-TN2ZKMA-E
10.2.8 Calibration waiting setting (option)
Description
Set the waiting time from supply of calibration gas to start of calibration.
(Set the time so that the calibration gas becomes stable before the calibration.)
Settable range: 10 to 999sec.
Operation
Procedure
Note
You cannot change the setting value during automatic calibration or remote calibration.
INZ-TN2ZKMA-E - 61 -
10.2.9 Calibration error clear
Description
You can clear the errors occurred during calibration using this function.
If an error occurs during calibration, an error display (Er4, Er5) and abnormal contact output (close)
continues until the next calibration is properly completed.
Clear the error display on the measurement screen and open the abnormal contact output.
Error log information is not cleared.
Operation
Procedure
- 62 - INZ-TN2ZKMA-E
10.2.10 Operation setting screen of calibration range
Description
During calibration, you can select single or common range for the calibration factor using this func-
tion.
Settable range: Select one of the following.
(1) “Range interlock”: Performs calibration of the range that is currently displayed
and sets the calibration factors of the other ranges to the same
value as above.
(2) “Display range”: Performs calibration of the range that is currently displayed.
Operation
Procedure
(4) When it is fixed, the display returns to the screen on the right.
CALIBRATION MENU
CURRENT
INZ-TN2ZKMA-E - 63 -
10.3 Blowdown menu (option)
Operation
Procedure
(3) Use the key to select the auto blowdown valid (YES).
BLOW DOWN
Press the key to set the value.
YES NO
Note
If the time for automatic calibration comes during manual treatments (calibration, blowdown, sensor
check, or sensor recovery) or remote treatments (calibration, blowdown, or heater off), the treatment
being performed is prioritized and the automatic calibration starts after the treatment is completed.
If the time for automatic blowdown comes at the same time with automatic calibration, the automatic
blowdown starts first.
If disconnection is detected (O2 sensor input, O2 sensor thermocouple input, or thermocouple input
(combustion control: option)), or heater temperature error or A/D saturation error occurs, automatic
calibration is not performed.
- 64 - INZ-TN2ZKMA-E
10.3.2 Date and time setting of automatic blowdown
Description
You can set the date and time for starting automatic blowdown using this function.
Automatic blowdown is performed in a specified cycle from a specified date and time.
If it is invalid, automatic blowdown does not start at a specified date and time.
Settable range: date and time in the future in the calendar
Operation
Procedure
Setting the date and time for starting automatic blowdown to 13:00, 2015/02/25
(example)
Displayed message
Key operation Description
(LCD)
(1) Display the screen on the right in accordance with the key op-
BLOW DOWN MENU
eration summary and press the key.
SET AUTO BLOW
Note
You cannot change the setting value during automatic blowdown or remote blowdown.
Check that “Current date and time setting” in the parameter menu is properly set.
INZ-TN2ZKMA-E - 65 -
10.3.3 Automatic blowdown cycle setting
Description
You can set the automatic blowdown cycle using this function.
The cycle starts from a specified date and time for automatic blowdown.
Settable range: 00h 00m to 99h 59m (m: 00 to 59)
Operation
Procedure
(2) Press the key to display the screen on the right and press
SET AUTO BLOW
the key.
AUTO BLOW CYCLE
The auto blowdown cycle setting screen appears.
(3) Use the and key to select the auto blowdown cycle.
AUTO BLOW CYCLE
Press the key to set the value.
24 h 00 m
Note
You cannot change the setting value during automatic blowdown or remote blowdown.
Set the blowdown cycle value larger than the blowdown time.
- 66 - INZ-TN2ZKMA-E
10.3.4 Procedure for setting blowdown time
Description
You can set the blowdown time using this function (common with manual blow down).
Settable range: 0 to 999 sec.
Operation
Procedure
(2) Press the key to display the screen on the right and press
SET AUTO BLOW
the key.
BLOW DOWN TIME
The procedure for setting blowdown time screen appears.
(3) Use the and key to set the blowdown time.
BLOW DOWN TIME
Press the key to set the value.
030 s
Note
You cannot change the setting value during automatic blowdown or remote blowdown.
Set the blowdown cycle value smaller than the blowdown time.
INZ-TN2ZKMA-E - 67 -
10.3.5 Perfoming manual blowdown
Description
You can perform blowdown on the screen using this function.
If you perform remote control (calibration, blowdown, turning off the heater) during manual blow-
down, remote control is prioritized and manual blowdown is stopped.
Operation
Procedure
How to cancel
Press the key to cancel the operation.
- 68 - INZ-TN2ZKMA-E
10.4 Maintenance menu
Operation
Procedure
(4) Press the key to display the previous piece of error log
YY/MM/DD HH:MM
information.
(5) Press the key, the display returns to the screen on the
ERROR LOG
right.
DISP CLEAR
(6) Press the key again to return to the screen on the right.
MAINTE MENU
ERROR LOG
INZ-TN2ZKMA-E - 69 -
Error logs
Display message Status
Sensorline Error Sensor line disconnection of the zirconia oxygen sensor was detected.
TC-line Error Temperature control line disconnection of the zirconia oxygen sensor was detected.
Sub temp. Error Line disconnection of the thermocouple for combustion control was detected.
Warm-up was not completed within the warm-up monitoring time (45 minutes).
Warm-up Error Warm-up is properly completed if the heater temperature of the zirconia oxygen sen-
sor becomes the control temperature (800°C) ± 1°C and stable for one minute.
Cell temp. Error Heater temperature exceeds the specified range (800°C ± 70°C)
The concentration of the calibration span gas being supplied is not stable.
Span gas Error (In a discrimination treatment of stability, the error of ± 0.2% or more compared to the
value in the previous treatment continues.)
The concentration of the calibration zero gas being supplied is not stable.
Zero gas Error (In a discrimination treatment of stability, the error of ± 0.2% or more compared to the
value in the previous treatment continues.)
Span cal. Error Span calibration failed. (Calibration factor could not be determined.)
Zero cal. Error Zero calibration failed. (Calibration factor could not be determined.)
Sensor Error, An error was detected in the A/D conversion of oxygen concentration value of the zirco-
A/D data Error nia oxygen sensor. (260 mV or more, –50 mV or less)
- 70 - INZ-TN2ZKMA-E
10.4.2 Clearing error logs
Description
You can clear all error logs saved using this function.
Operation
Procedure
(2) Use the key to select the error log clear screen.
ERROR LOG
DISP CLEAR
(6) Press the key again to return to the screen on the right.
MAINTE MENU
ERROR LOG
INZ-TN2ZKMA-E - 71 -
10.4.3 Alarm log display
Description
You can display alarm logs on the screen using this function.
A latest piece of alarm information is displayed first.
The maximum of 12 pieces of alarm information are saved.
Press the key to display the older pieces of alarm information.
The latest piece of alarm information is displayed next to the oldest piece of alarm information.
The oldest piece of alarm information is overwritten by a new one.
Operation
Procedure
(2) Use the key to select the alarm log display screen.
ALARM LOG
DISP CLEAR
(4) Press the key to display the previous piece of alarm log
YY/MM/DD HH:MM
information.
(5) Press the key, the display returns to the screen on the
ALARM LOG
right.
DISP CLEAR
(6) Press the key again to return to the screen on the right.
MAINTE MENU
ALARM LOG
Alarm logs
Display message Status
High alarm Oxygen concentration value exceeded a specified upper limit.
Low alarm Oxygen concentration value exceeded a specified lower limit.
Hi-High alarm Oxygen concentration value exceeded a specified Hi-High limit.
Low-Low alarm Oxygen concentration value exceeded a specified Low-Low limit.
- 72 - INZ-TN2ZKMA-E
10.4.4 Clearing alarm logs
Description
You can clear all alarm logs using this function.
Operation
Procedure
(2) Use the key to select the alarm log clear screen.
ALARM LOG
DISP CLEAR
(6) Press the key again to return to the screen on the right.
MAINTE MENU
ALARM LOG
INZ-TN2ZKMA-E - 73 -
10.4.5 Performing a manual sensor check
Description
Manually supply atmospheric air or air from cylinder to the detector in order to measure the electro-
motive force of the sensor.
Operation
Proce-
(example)
Key operation Description Displayed message
(1) Display the screen on the right in accordance with the key op-
MAINTE MENU
eration summary and press the key, the manual sensor
check performing screen appears. SENSOR CHECK
Note
If the electromotive force of the sensor is out of the range from -5.0 mV through +5.0 mV, it is rec-
ommended to replace the sensor.
- 74 - INZ-TN2ZKMA-E
10.4.6 Maintenance mode setting
Description
You can set the maintenance mode to valid or invalid with this function.
If the maintenance mode is set to valid, the analog output signal is held at the set value (see “10.5.10
Hold value setting.”) and the contact output for maintenance of the external contact is on. The data
portion of the measurement screen flickers.
Operation
Procedure
(2) Use the key to select the maintenance mode valid (YES).
MAINTENANCE MODE
Press the key to set the value.
YES NO
Note
If an error occurs while the maintenance mode is enabled, error handling is prioritized.
If the analog output hold function (error hold) is enabled, the analog output signal is held at the value
set at the hold value setting (error hold).
The data portion of the measurement screen flickers and is highlighted.
INZ-TN2ZKMA-E - 75 -
10.4.7 Password setting
Description
You can set a password for switching the “Key lock function” valid /invalid, which is to prevent unau-
thorized people from making various setting or operating the unit manually (modification, etc.)
Note: Refer to “Setting of key lock” Paragraph (“10.5.16”).
When you set the “new password” you desire, the screen transits to the password authentication
screen automatically.
After you input the “old password" in the password authentication screen, the new password will be
registered.
The factory-set password is “0000”.
An authorized person should manage the set password for remembrance’ sake.
Settable value: 4 digits from 0 to 9
Operation
Procedure
(2) Use the key and the key to input the new
NEW PASSWORD
password.
0123
(4) Use the key and the key to input the old
OLD PASSWORD
password.
9999
- 76 - INZ-TN2ZKMA-E
10.4.8 PID auto tuning
Description
Heater temperature of the detector is PID controlled. This "PID auto-tuning" function optimizes each
value of P (proportion), I (integration) and D (derivation) for the environment where the unit is in-
stalled.
Note: Each value of P, I, and D has been set at factory. If temperature is not properly controlled, per-
form PID auto tuning.
Measured value and analog output become unstable during PID auto tuning because the controlled
temperature goes up and down.
You can hold the analog output value during PID auto tuning, because it is a part of maintenance (Re-
fer to 10.5.9 “Hold treatment setting (maintenance hold)”).
You cannot use PID auto tuning with auto calibration or automatic blowdown at the same time.
Operation
Procedure
How to cancel
Press the key to cancel PID auto tuning.
If you cancel PID auto tuning, each value of P, I and D is to be the value before tuning.
INZ-TN2ZKMA-E - 77 -
10.5 Parameter menu
Operation
Procedure
(2) Use the and key to set the date and time.
DATE SET
Press the key to set the value.
15/02/25 13:00
- 78 - INZ-TN2ZKMA-E
10.5.2 Contact input setting
Description
You can set the functions for the contact inputs 1 to 3 using this function.
Settable range: Select one of the following
(1) [NONE] : Performs no treatment by contact input.
(2) [BLOW DOWN ON] : Performs blowdown by contact input.
(Switch OFF to ON to perform blowdown.)
(3) [HEATER OFF] : Turn off the heater by contact input.
(OFF/ON:Heater ON/Heater OFF)
(4) [PROHIBIT CAL.] : Sets if calibration is prohibited or valid by contact in-
put.
(OFF/ON: Calibration is valid/prohibited.)
(5) [REMOTE CAL.] : Performs all calibration by contact input.
(Switch OFF to ON to perform calibration.)
(6) [REMOTE HOLD] : Holds the AO by contact input.
(OFF/ON: not held/held)
(7) [CALCULATE REST] : Resets maximum and minimum calculations of O2 by
contact input.
(Switch OFF to ON to perform calibration.)
(8) [OUTPUT RANGE] : Switches the range by contact input.
(OFF/ON: Range-1/Range-2)
Note) The functions other than “NONE” cannot be set for multiple contacts.
Operation
Procedure
(2) Use the key to select the contact input 1 setting screen.
DIGITAL INPUT
Press the key to set the value.
DI 1
(Also follow this procedure for the contact inputs 2 and 3.)
(3) Use the key to select the function for contact input 1.
DI 1
NONE
INZ-TN2ZKMA-E - 79 -
10.5.3 Selection of alarm contact output
Description
You can set the alarm conditions for alarm contact output using this function.
Settable range: Select one of the following.
(1) [ALARM NONE] : Alarm contact output is not performed.
(2) [HIGH ALARM] : Alarm contact output is performed when an high limit
alarm occurs.
(3) [LOW ALARM] : Alarm contact output is performed when an lower limit
alarm occurs.
(4) [HH ALARM] : Alarm contact output is performed when an HH limit alarm
occurs.
(5) [LL ALARM] : Alarm contact output is performed when an LL limit alarm
occurs.
(6) [H/L ALARM] : Alarm contact output is performed when an high or lower
limit alarm occurs.
(7) [HH/LL ALARM] : Alarm contact output is performed when an HH or LL limit
alarm occurs.
Operation
Procedure
Setting the lower limit alarm function for alarm contact output
(example)
Displayed message
Key operation Description
(LCD)
(1) Display the screen on the right in accordance with the key op-
PARAMETER MENU
eration summary and press the key.
DO ALARM SET
- 80 - INZ-TN2ZKMA-E
10.5.4 High limit setting of oxygen concentration
Description
You can set the high limit of oxygen concentration using this function.
Use the oxygen concentration alarm 1 for the range 1, and the oxygen concentration alarm 2 for the
range 2.
Settable range: 0.001 to 55.000 vol%
Operation
Procedure
(2) Use the key to select the oxygen concentration high limit
OXYGEN ALARM 1
value setting screen, and press the key.
HIGH ALARM
Note
A setting error occurs if the following condition is not satisfied:
“HH limit of oxygen concentration” ≥ “high limit of oxygen concentration” ≥ “Lower limit of oxygen
concentration” ≥ “LL limit of oxygen concentration”
INZ-TN2ZKMA-E - 81 -
10.5.5 Lower limit setting of oxygen concentration
Description
You can set the lower limit of oxygen concentration using this function.
Use the oxygen concentration alarm 1 for the range 1, and the oxygen concentration alarm 2 for the
range 2.
Settable range: 0.001 to 55.000 vol%
Operation
Procedure
Note
A setting error occurs if the following condition is not satisfied:
“HH limit of oxygen concentration” ≥ “high limit of oxygen concentration” ≥ “Lower limit of oxygen
concentration” ≥ “LL limit of oxygen concentration”
- 82 - INZ-TN2ZKMA-E
10.5.6 HH limit setting of oxygen concentration
Description
You can set the HH limit of oxygen concentration using this function.
Use the oxygen concentration alarm 1 for the range 1, and the oxygen concentration alarm 2 for the
range 2.
Settable range: 0.001 to 55.000 vol%
Operation
Procedure
Note
A setting error occurs if the following condition is not satisfied:
“HH limit of oxygen concentration” ≥ “High limit of oxygen concentration” ≥ “Lower limit of oxygen
concentration” ≥ “Low-Low limit of oxygen concentration”
INZ-TN2ZKMA-E - 83 -
10.5.7 LL limit setting of oxygen concentration
Description
You can set the LL limit of oxygen concentration using this function.
Use the oxygen concentration alarm 1 for the range 1, and the oxygen concentration alarm 2 for the
range 2.
Settable range: 0.001 to 55.000 vol%
Operation
Procedure
Note
A setting error occurs if the following condition is not satisfied:
“HH limit of oxygen concentration” ≥ “High limit of oxygen concentration” ≥ “Lower limit of oxygen
concentration” ≥ “LL limit of oxygen concentration”
- 84 - INZ-TN2ZKMA-E
10.5.8 Hysteresis Setting
Description
You can set the hysteresis for alarm condition of oxygen concentration.
Use the oxygen concentration alarm 1 for the range 1, and the oxygen concentration alarm 2 for the
range 2.
Perform the setting using the percentage (%) of the range compared to the full scale.
Settable range: 0 to 20 %
Operation
Procedure
Setting the hysteresis for alarm condition of oxygen concentration to “20%” (Range 1)
(example)
Displayed message
Key operation Description
(LCD)
(1) Display the screen on the right in accordance with the key op-
PARAMETER MENU
eration summary and press the key.
OXYGEN ALARM 1
(2) Use the key to select the hysteresis setting screen, and
OXYGEN ALARM 1
press the key.
HYSTERESIS
INZ-TN2ZKMA-E - 85 -
Hysteresis:
If the value fluctuates around the condition value, there is a possibility that alarms occur frequent-
ly. When determining alarms, set a hysteresis width for the condition in order to prevent chattering.
For alarm check, set the percentage (%) of the range compared to the full scale as hysteresis width
(see the figure below).
This is common among “HH limit value,” “High limit value,” “Lower limit value,” and “LL limit
value.”
HH limit value
Hysteresis
Oxygen concentration value
LL limit value
HH limit judgment
LL limit judgment
HH / LL limit judgment
- 86 - INZ-TN2ZKMA-E
10.5.9 Hold treatment setting (maintenance hold)
Description
You can set if the analog output hold function is valid or invalid using this function.
If the analog output hold function is valid, the value set for the analog output (see “10.5.10 Hold
value setting (maintenance hold)”) is held at the value set for analog output when the following
treatment is performed.
・ Calibration (Auto, All, Manual, Remote)
・ Blowdown (Auto, Manual, Remote)
・ Sensor diagnosis, PID auto tuning
・ While the maintenance mode is set to “Valid.”
Operation
Procedure
(3) Use the key to select the output hold valid (YES).
OUTPUT HOLD
Press the key to set the value.
YES NO
Note
If an error occurs while the analog output hold function (error hold) is set to “Valid,” error hold pro-
cessing is prioritized.
Analog output signal during warm-up is held at 0% (4 mA/0 V).
INZ-TN2ZKMA-E - 87 -
10.5.10 Hold value setting (maintenance hold)
Description
Using this function, you can set (select) the output value of analog output signal when the analog out-
put hold function (maintenance hold) is enabled.
If the maintenance mode is set to “Valid,” analog output signal is held at the value set in this proce-
dure.
Settable range: Select one of the following.
(1) [0%] : Held at 0% (4 mA/0 V)..
(2) [100%] : Held at 100% (20 mA/1 V).
(3) [Last value] : Held at the value immediately before the value for analog hold.
(4) [Setting value] : Held at the value set as the “10.5.11 Setting of hold setting val-
ue (maintenance hold)”.
Operation
Procedure
(2) Press the key to display the screen on the right and press
AO HOLD (MAINTE)
the key.
OUTPUT SELECT
The analog output hold value setting screen appears.
(3) Use the key to select the hold value.
OUTPUT SELECT
Press the key to set the value.
0%
- 88 - INZ-TN2ZKMA-E
10.5.11 Setting of hold setting value (maintenance hold)
Description
Using this function, you can set the output value of analog output signal to an arbitrary value when the
analog output hold function (maintenance hold) is enabled.
This function is to set the output value of an analog output at a percentage (%) of the full-scale value,
when “setting value” is selected and specified on the “10.5.10 Hold value setting (maintenance
hold).”
Settable range: 0 to 100 %
Operation
Procedure
(2) Press the key to display the screen on the right and press
AO HOLD (MAINTE)
the key.
HOLD VALUE
The hold setting value setting screen appears.
(3) Use the and key to set the hold value.
HOLD VALUE
Press the key to set the value.
000 %
INZ-TN2ZKMA-E - 89 -
10.5.12 Setting of measurement recovery time (maintenance
hold)
Description
This function is to set the recovery time (hold of extension) from the analog output hold function
(maintenance hold).
Settable range: 0 to 300 sec.
Operation
Procedure
(2) Press the key to display the screen on the right and press
AO HOLD (MAINTE)
the key.
MEAS. WAIT TIME
The measurement recovery time setting screen appears.
(3) Use the and key to set the measurement recovery
MEAS. WAIT TIME
time.
010 S
Press the key to set the value.
- 90 - INZ-TN2ZKMA-E
10.5.13 Hold treatment setting (error hold)
Description
Using this function, you can set whether the analog output hold function is valid or invalid when an
error occurs.
If the analog output hold function (error hold) is set to valid, analog output signal is held at the set
value (see “10.5.14 Hold value setting (error hold)”) if an error occurs.
Operation
Procedure
(3) Use the key to select the output hold valid (YES).
OUTPUT HOLD
Press the key to set the value.
YES NO
Note
If an error occurs while the analog output hold function (error hold) is set to “Valid,” error hold pro-
cessing is prioritized.
Analog output signal during warm-up is held at 0% (4 mA/0 V).
INZ-TN2ZKMA-E - 91 -
10.5.14 Hold value setting (error hold)
Description
Using this function, you can set (select) the output value of analog output signal when the analog out-
put hold function (error hold) is enabled.
If the maintenance mode is set to “Valid,” analog output signal is held at the value set in this proce-
dure.
Settable range: Select one of the following.
(1) [0%] : Held at 0% (4 mA/0 V).
(2) [100%] : Held at 100% (20 mA/1 V).
(3) [Last value] : Held at the value immediately before the value for analog hold.
(4) [Setting value] : Held at the value set as the “10.5.15 Setting of hold setting val-
ue (error hold)”.
Operation
Procedure
(2) Press the key to display the screen on the right and press
AO HOLD (ERROR)
the key.
OUTPUT SELECT
The analog output hold value setting screen appears.
(3) Use the key to select the hold value.
OUTPUT SELECT
Press the key to set the value.
0%
- 92 - INZ-TN2ZKMA-E
10.5.15 Setting of hold setting value (error hold)
Description
Using this function, you can set the output value of analog output signal to an arbitrary value when the
analog output hold function (error hold) is enabled.
This function is enabled if “Setting value” is set at “10.5.14 Hold value setting (error hold).”
Set the output value of analog output signal as a percentage (%) of the full-scale value of the display
range. 0% is equivalent to 0 vol% (4 mA/0 V) and 100 % is to the full-scale value (20 mA/1 V).
Settable range: 0 to 100 %
Operation
Procedure
(2) Press the key to display the screen on the right and press
AO HOLD (ERROR)
the key.
HOLD VALUE
The hold setting value setting screen appears.
(3) Use the and key to set the hold value.
HOLD VALUE
Press the key to set the value.
000 %
INZ-TN2ZKMA-E - 93 -
10.5.16 Setting of key lock
Description
Authorized person can set if the key lock is valid or invalid using this function.
You need a “password” to make a setting if the key lock is valid or invalid.
Note: Refer to “Password setting” (Paragraph “10.4.7”).
If the key lock is valid, you cannot make settings and manual operation (manual calibration, manual
browdown, etc.). However, you can see the screen transition and set values.
Operation
Procedure
(2) Use the key to select the key lock valid (YES) or invalid
KEY LOCK
(No).
YES NO
(4) Use the key and the key to input the password.
INPUT PASSWORD
0123
- 94 - INZ-TN2ZKMA-E
10.5.17 Setting of automatic OFF time
Description
You can set the time for automatically turning off the backlight of the LCD (screen) using this func-
tion.
When the time set for turning off the backlight elapses after the last operation, the backlight is turned
off.
(Press any key to turn on the backlight.)
If 00 minutes is set, the backlight is not turned off.
Settable range: 0 to 99 min.
Operation
Procedure
Setting the time for automatically turning off the backlight to 10 minutes
(example)
Displayed message
Key operation Description
(LCD)
(1) Display the screen on the right in accordance with the key op-
PARAMETER MENU
eration summary and press the key.
BACKLIGHT TIME
(2) Use the and key to set the automatic OFF time.
BACKLIGHT TIME
Press the key to set the value.
10 m
INZ-TN2ZKMA-E - 95 -
10.5.18 Station number setting
Description
You can set the station number of the unit for MODBUS communication using this function.
Settable range: 0 to 99
Operation
Procedure
- 96 - INZ-TN2ZKMA-E
10.5.19 Adjustment screen for analog output 0%
Description
You can adjust the analog output 0% using this function.
Operation
Procedure
INZ-TN2ZKMA-E - 97 -
10.5.20 Adjustment screen for analog output 100%
Description
You can adjust the analog output 100% using this function.
Operation
Procedure
- 98 - INZ-TN2ZKMA-E
10.5.21 Fuel coefficient setting (option)
Description
You can set the fuel coefficient [k], which is for calculation of combustion efficiency, using this func-
tion.
Settable range:0.00 to 1.99
Operation
Procedure
(2) Use the key and the key to input the fuel coefficient.
FUEL COEFF.
0.73
INZ-TN2ZKMA-E - 99 -
10.6 Factory menu
- 100 - INZ-TN2ZKMA-E
11. SPECIFICATIONS
11.1 Specifications
INZ-TN2ZKMA-E - 101 -
- 102 - INZ-TN2ZKMA-E
INZ-TN2ZKMA-E - 103 -
11.2 Code symbols
<Converter>
- 104 - INZ-TN2ZKMA-E
<Detector>
INZ-TN2ZKMA-E - 105 -
<Ejector>
<Exclusive cable>
- 106 - INZ-TN2ZKMA-E