Olympus GF Type UCT180 User Manual
Olympus GF Type UCT180 User Manual
CAUTION : Balloons Used with This Product Contain Natural Rubber Latex, Which May Cause
Allergic Reactions.
USA: CAUTION: Federal law restricts this device to sale by or on the order of a
physician.
Contents
Contents
Symbols......................................................................................... 1
Chapter 4 Operation.................................................................. 57
4.1 Insertion .......................................................................................... 59
4.2 Observation of the ultrasound image .............................................. 64
4.3 Using EndoTherapy accessories .................................................... 66
4.4 Withdrawal of the endoscope.......................................................... 73
4.5 Removal of the balloon ................................................................... 74
4.6 Transportation of the endoscope .................................................... 75
Symbols
The meaning(s) of the symbol(s) shown on the package, the back cover of this
instruction manual and/or this instrument are as follows:
Refer to instructions.
Caution
Serial number
Ultrasonic endoscope
Do not resterilize
Lot number
Nonsterile
Keep dry
Date of manufacture
Manufacturer
Intended use
This instrument has been designed to be used with an Olympus universal
endoscopic ultrasound center or a diagnostic ultrasound system (ALOKA CO.,
LTD), video system center, light source, documentation equipment, monitor,
EndoTherapy accessories and other ancillary equipment.
Instruction manual
This instruction manual contains essential information for using this instrument
safely and effectively. Before use, thoroughly review this manual and the
manuals of all equipment that will be used during the procedure. Then use the
equipment as instructed.
Keep this and all related instruction manuals in a safe and accessible location. If
you have any questions or comments about any information in this manual,
please contact Olympus.
Elastography:
Mode for displaying the relative elasticity information of a tissue using
color images.
For more details, refer to the instruction manual for the ultrasound
instrument for which elastography is available.
User qualifications
If there are official standards for user qualifications for performing endoscopy
and endoscopic treatment that are defined by the hospital’s medical
administrators or other official institutions, such as academic societies on
endoscopy, follow those standards. If there are no official qualification standards,
the operator of this instrument must be a physician approved by the medical
safety manager of the hospital or person in charge of the department
(department of internal medicine, etc.).
The physician should be capable of safely performing the planned endoscopy
and endoscopic treatment following guidelines set by the academic societies on
endoscopy, etc., and considering the difficulty of endoscopy and endoscopic
treatment. This manual does not explain or discuss endoscopic procedures.
Instrument compatibility
Refer to the “System chart” in the Appendix to confirm that this instrument is
compatible with the ancillary equipment being used. Using incompatible
equipment can result in patient or operator injury and/or equipment damage.
This instrument complies with EMC standard for medical electrical equipment,
edition 3 (IEC 60601-1-2: 2007). However, when connected with an instrument
that complies with EMC standard for medical electrical equipment, edition 1
(IEC 60601-1-2: 1993), the whole system complies with edition 1.
After using this instrument, reprocess and store it according to the instructions
given in the endoscope’s companion reprocessing manual. Improper and/or
incomplete reprocessing or storage can present an infection control risk, cause
equipment damage, or reduce performance.
The balloons are disposable, and are intended for a single use only; a new one
must be used for each patient. Do not attempt to reuse or resterilize a balloon.
Spare equipment
Be sure to prepare another endoscope to avoid interruption of the examination
due to equipment failure or malfunction.
Maintenance management
The probability of failure of the endoscope and ancillary equipment increases as
the number of procedures performed and/or the total operating hours increase.
In addition to the inspection before each procedure, the person in charge of
medical equipment maintenance in each hospital should inspect the items
specified in this manual periodically. An endoscope with an observed irregularity
should not be used, but should be inspected by following Section 10.1,
“Troubleshooting guide” on page 154. If the irregularity is still observed after
inspection, contact Olympus.
Signal words
The following signal words are used throughout this manual:
• For reasons described below, do not rely on the NBI1 observation mode
alone for primary detection of lesions or to make a decision regarding
any potential diagnostic or therapeutic intervention.
NBI has not been demonstrated to increase the yield or
sensitivity of finding any specific mucosal lesion including
colonic polyps or Barrett’s esophagus.
NBI has not been demonstrated to aid in differentiating
establishing the presence or absence of dysplasia or
neoplastic changes within mucosa or mucosal lesions.
Ultrasound endoscope
Mouthpiece
(MB-142, 2 pcs.)
Balloon 3 for the USA
(MAJ-249 sterile, 20 pcs.)
Balloon for countries other
than the USA (MAJ-213, Cleaning brush
nonsterile, 20 pcs.) (MAJ-1534)
Injection tube
Channel plug (MH-946)
(MAJ-621)
Balloon applicator Water-resistant cap Suction cleaning adapter Air/water channel cleaning adapter
(MAJ-675) (MH-553, 2 pcs.) (MH-856) (MAJ-629)
Endoscope-side
connector
Ultrasound
connector
Ultrasonic cable
(MAJ-1597)
Balloon 3
Balloon 3 is shipped sterile in sets of 20 pieces, enclosed in a resealable
package. The correct model to be used with this endoscope is listed in Table 1.1
below.
Endoscope Balloon 3
GF-UCT180 MAJ-249
Table 1.1
2.1 Nomenclature
Universal cord
2. Suction connector
4. Endoscope connector
Water-resistant cap
(MH-553) Product number and serial number
8. UP/DOWN angulation
control knob
Suction cylinder
7. Elevator control lever
10. Suction valve (MAJ-1443)
20. RIGHT/LEFT
angulation
control knob
2.3 Specifications
Environment
Specifications
Endoscope functions
Model GF-UCT180
Optical system Field of view 100
Direction of view 55 forward oblique
Depth of field 3 – 100 mm
Insertion tube Distal end outer
ø 14.6 mm
diameter
Distal end enlarged 1. Air/water nozzle
2. Objective lens
3. Light guide lens
4. Forceps elevator
5. Instrument channel outlet
6. Ultrasonic transducer
7. Balloon water supply port
8. Balloon suction port
9. Balloon attachment groove
6. 1.
7. 8.
9. 2.
3.
5. 4.
1 For more details, refer to the instruction manual for the CV-180.
CISPR 11 of emission:
Group 1, Class B
Notch
Fitting part
Chain part
Connecting plate
Hole
Figure 2.1
1. Confirm that the chain for water-resistant cap is free from cracks, flaws,
wear, deformation, or other damages (see Figure 2.1).
2. Align the notch on the connecting plate with the pin on the venting connector
of the water-resistant cap (MH-553, see Figure 2.2).
3. Place the connecting plate over the venting connector (see Figure 2.2).
4. Confirm that the connecting plate is securely attached to the foot of the
venting connector and can be smoothly rotated.
5. Place the hole on the fitting part over the endoscope’s S-cord connector
mount.
Pin
Venting connector
Water-resistant cap
Figure 2.2
Before each procedure, prepare and inspect this instrument as instructed below.
Inspect other equipment to be used with this instrument as described in their
respective instruction manuals. If this instrument malfunctions, do not use it.
Return it to Olympus for repair as described in Section 10.3, “Returning the
endoscope for repair” on page 161. If any irregularities are suspected after
inspection, follow the instructions given in Chapter 10, “Troubleshooting”.
Monitor
Ultrasonic cable
Light source
Water container
Suction pump
Ultrasound endoscope
Figure 3.1
2. Inspect the boot and the insertion section near the boot for bends, twists, or
other irregularities.
3. Inspect the external surface of the entire insertion section including the
bending section and the distal end for dents, bulges, swelling, scratches,
holes, sagging, transformation, bends, adhesion of foreign bodies, missing
parts, protruding objects, or other irregularities.
4. Gently holding the insertion section with one hand, carefully run your other
hand back and forth over the entire length of the insertion section (see
Figure 3.2). Confirm that no objects or metallic wire protrude from the
insertion section. Also, confirm that the insertion tube is not abnormally rigid.
Figure 3.2
5. Using both hands, bend the insertion tube of the endoscope into a
semicircle. Then, moving your hands as shown by the arrows in Figure 3.3,
confirm that the entire insertion tube can be smoothly bent to form a
semicircle and that the insertion tube is pliable.
Figure 3.3
6. Gently hold the midpoint of the bending section and a point 20 cm from the
distal end. Push and pull gently to confirm that the junction between the
bending section and the insertion tube is not loose.
7. Inspect the objective lens and light guide lens at the distal end of the
endoscope’s insertion section for scratches, cracks, stains, or other
irregularities.
8. Inspect the air/water nozzle at the distal end of the endoscope’s insertion
section for abnormal swelling, bulges, dents, or other irregularities.
1. Confirm that both the UP/DOWN and RIGHT/LEFT angulation locks move
all the way in the “F ” direction.
3. When the UP/DOWN and RIGHT/LEFT angulation control knobs are turned
to their respective neutral positions as shown in Figure 3.4, confirm that the
bending section returns smoothly to an approximately straight condition.
Figure 3.4
1. Move the UP/DOWN angulation lock all the way in the opposite direction of
the “F ” mark. Then turn the UP/DOWN angulation control knob in the
“ U” or the “D ” direction until it stops.
2. Confirm that the angle of the bending section is roughly stabilized when the
UP/DOWN angulation control knob is released.
3. Confirm that the bending section straightens out when the UP/DOWN
angulation lock is moved all the way in the “F ” direction and the
UP/DOWN angulation control knob is released.
1. Turn the RIGHT/LEFT angulation lock all the way in the opposite direction of
the “F ” mark. Then turn the RIGHT/LEFT angulation control knob in the
“R ” or the “ L” direction until it stops.
2. Confirm that the angle of the bending section is roughly stabilized when the
RIGHT/LEFT angulation control knob is released.
3. Confirm that the bending section straightens out when the RIGHT/LEFT
angulation lock is turned in the “F ” direction and the RIGHT/LEFT
angulation control knob is released.
1. Move the elevator control lever slowly all the way in the opposite direction of
the “ U” direction. Visually confirm that the portion of the elevator wire
extending from the distal end of the insertion section is not broken or bent
(see Figure 3.5).
2. While observing the forceps elevator at the distal end of the insertion
section, slowly move the elevator control lever all the way in the “ U”
direction. Confirm that the lever can be operated smoothly and that the
forceps elevator is raised smoothly. Also confirm that the forceps elevator
remains stationary when pushed from behind while holding the elevator
control lever stationary (see Figure 3.5).
3. Move the elevator control lever slowly all the way in the opposite direction of
the “ U” direction. Confirm that the lever can be operated smoothly and
that the forceps elevator is lowered smoothly (see Figure 3.5).
Forceps elevator
Figure 3.5
Confirm that the top hole of the air/water valve is not blocked
(see Figure 3.6). If the hole is blocked, air is fed continuously
and patient pain, bleeding, and/or perforation can result.
1. Confirm that the holes of the valves are not blocked (see Figures 3.6 and
3.7).
2. Confirm that the valves are not deformed or cracked (see Figures 3.6 and
3.7).
3. Check for excessive scratching or tears in the air/water valve’s seals (see
Figure 3.6).
Hole
Cap
Spring Skirt
Top view
Packings
Seal
Figure 3.6
Cap
Spring
Skirt
Packing
Hole
Suction valve (MAJ-1443)
Figure 3.7
1. Confirm that the slit and hole on the biopsy valve have no splits, cracks,
deformations, discoloration, or other damage (see Figure 3.8).
Normal Abnormal
Slit Discoloration
Hole
Cap Main body Splits, cracks
Figure 3.8
Slit
Figure 3.9
2. Using your fingers, check all surfaces of the mouthpiece for scratches,
cracks, or other irregularities (see Figure 3.10).
Opening
Outer flange
Main body
Figure 3.10
Suction valve
Skirt
Figure 3.11
Air/water valve
Skirt
Air/water cylinder
Figure 3.12
Attach the biopsy valve to the instrument channel port of the endoscope (see
Figure 3.13). Confirm that the biopsy valve fits properly.
Biopsy valve
Instrument
channel port
Figure 3.13
1. Prepare and inspect the light source, video system center, ultrasound
diagnostic equipment, monitor, water container, suction pump, and
EndoTherapy accessories as described in their respective instruction
manuals.
Figure 3.14
SSD-10
Transducer port Ultrasound connector
Lock handle
ProSound 71
Ultrasonic connector
Lock handle
ProSound F751
PROBE 2 PROBE 1
PROBE 3
PROBE 4
Lock handle
Ultrasound connector
Figure 3.15
Firmly connect the suction tube from the suction pump to the
suction connector on the endoscope connector. If the suction
tube is not attached properly, debris may drip from the tube
and can pose an infection control risk, cause equipment
damage, and/or reduce suction capability.
2. Insert the endoscope connector completely into the output socket of the light
source.
3. Make sure that the inside of the videoscope cable connector is dry and free
of debris.
4. Place the water container’s water supply channel onto the water supply
connector on the endoscope connector at an angle of 90 and push it in until
it stops (see Figure 3.16 (1)).
5. Turn the water container’s connection adapter 90 clockwise to align the air
supply channel with the air supply connector of the endoscope connector
(see Figure 3.16 (2)).
6. Push the water container’s connection adapter again until it stops (see
Figure 3.16 (3)).
7. Confirm that the water container’s connection adapter fits properly and that
it cannot be rotated (see Figure 3.16 (4)).
Endoscope connector
Air supply channel
Metal tip
Water supply channel
Figure 3.16
8. Align the mark on the videoscope cable with mark 1 on the endoscope
connector and push it in until it stops (see Figure 3.17).
Mark (gray)
Figure 3.17
9. Turn the connector of the videoscope cable towards mark 2 until it stops
(see Figure 3.17).
10. Confirm that the mark on the videoscope cable is aligned with mark 2 on the
endoscope connector.
11. Connect the suction tube from the suction pump to the suction connector on
the endoscope connector (see Figure 3.18).
Suction pump
Suction connector
Suction tube
Figure 3.18
1. Make sure that the inside of the ultrasonic cable connector is dry and free of
debris.
2. Align the mark on the ultrasonic cable with mark A on the ultrasound
connector and push it until it stops (see Figure 3.19).
4. Confirm that the mark on the ultrasonic cable is aligned with mark B on the
endoscope connector (see Figure 3.19).
Mark A (red)
Mark A (red)
Mark (red)
Mark (red)
Figure 3.19
1. Turn the video system center, light source, and monitor ON and inspect the
WLI and NBI endoscopic image as described in their respective instruction
manuals.
3. While observing the palm of your hand, confirm that the WLI and NBI
endoscopic image is free from noise, blur, fog, or other irregularities.
4. Angulate the endoscope and confirm that the WLI and NBI endoscopic
image does not momentarily disappear or display any other irregularities.
Depress every remote switch and confirm that the specified functions work
normally.
2. Immerse the distal end of the insertion section in sterile water to a depth of
10 cm and confirm that no air bubbles are emitted when the air/water valve
is not operated.
3. Cover the hole in the air/water valve with your finger and confirm that air
bubbles are continuously emitted from the air/water nozzle.
4. Uncover the hole in the air/water valve and confirm that no air bubbles are
emitted from the air/water nozzle.
1. Keep the air/water valve’s hole covered with your finger and depress the
valve to the first stage. Observe the endoscopic image and confirm that
water flows on the entire objective lens.
2. Release the air/water valve. While observing the endoscopic image, confirm
that the emission of water stops and that the valve returns smoothly to its
original position.
3. While observing the endoscopic image, feed air after feeding water by
covering the hole in the air/water valve with your finger. Confirm that the
emitted air removes the remaining water from the objective lens and clears
the endoscopic image.
1. Place the container of sterile water and the endoscope at the same height.
For the inspection, adjust the suction pressure to the same level as it will be
during the procedure.
2. Immerse the distal end of the insertion section in sterile water with the
endoscope’s instrument channel port at the same height as the water level
in the water container. Press the suction valve and confirm that water is
continuously aspirated into the suction bottle of the suction pump.
3. Release the suction valve. Confirm that suction stops and the valve returns
to its original position.
4. Depress the suction valve to the first stage and aspirate water for one
second. Then, release the suction valve for one second. Repeat this several
times and confirm that no water leaks from the biopsy valve.
5. Remove the distal end of the endoscope from the water. Depress the
suction valve and aspirate air for a few seconds to remove any water from
the instrument and suction channels.
2. Release the suction valve. Confirm that suction stops and the suction valve
returns to its original position.
Keep your eyes away from the distal end when inserting
EndoTherapy accessories. Extending the EndoTherapy
accessory from the distal end could cause eye injury.
1. Confirm that the forceps elevator is lowered, then insert the EndoTherapy
accessory through the biopsy valve. Confirm that the EndoTherapy
accessory extends smoothly from the distal end, and that a foreign object
does not come out.
Figure 3.20
3. When setting the frequency, choose the setting value CLA2, and then the
frequencies available with the keyboard (see Figure 3.21). Refer to
EU-ME1’s instruction manual for instructions on how to set the frequency.
Figure 3.21
Figure 3.22
6. Press the “FREEZE” switch to change the ultrasound image to the FREEZE
mode.
Figure 3.23
4. Press the “FREEZE” switch to change the ultrasound image to the FREEZE
mode.
2. Insert the front end of the balloon into the installation side of the balloon
applicator. Slide the rear band of the balloon into the groove of the balloon
applicator (see Figure 3.24).
Rear end
Figure 3.24
3. Insert the distal end of the endoscope into the balloon applicator until it
contacts the front end of the balloon.
4. Remove the rear band from the balloon applicator and attach it to the
endoscope’s balloon attachment groove (see Figure 3.25).
Balloon attachment
groove
Distal end of
the endoscope
Figure 3.25
• Do not feed water into the balloon when the water level in the
water container is too low. The balloon may not be inflated
properly. If this occurs, refer to Section 10.1,
“Troubleshooting guide” on page 154.
• Disconnecting the water container from the endoscope just
after removing the air bubbles from the balloon may cause
them to reenter the balloon when inflating the balloon again.
Therefore, confirm that there are no air bubbles in the balloon
according to the following procedure.
1. Check the water level in the water container. If the water level is low, fill the
water container with sterile deaerated water. After adding water to the water
container or removing the water container from the endoscope, cover the
hole in the air/water valve and depress it to the first stage for about
15 seconds. Turn the light source and suction pump ON.
2. Cover the hole in the air/water valve and completely depress the valve to fill
the balloon with deaerated water.
5. If the balloon has no air bubbles in it, completely depress the suction valve
until all the water has been aspirated from the balloon.
6. If air bubbles are observed in the balloon, lower the distal end of the
endoscope and repeat Steps 2 through 5 until they are completely removed
from the balloon. After confirming that all air bubbles have been removed,
repeat Steps 2 through 5 two times to make sure they are completely
removed.
7. Completely depress the suction valve until all the water has been aspirated
from the balloon.
8. Tie the rear end of the balloon tightly with a sterile cotton thread.
Do not obstruct the balloon water supply port with the cotton
thread. If the balloon water supply port is obstructed, the
balloon cannot be inflated or deflated (see Figure 3.26).
Figure 3.26
Chapter 4 Operation
This manual does not explain or discuss clinical endoscopic procedures. It only
describes basic operation and precautions related to the operation of this
instrument. Therefore, the operator of this instrument must be a physician or
medical personnel under the supervision of a physician and must have received
sufficient training in clinical endoscopic technique.
4.1 Insertion
Figure 4.1
Keep the elevator control lever moved all the way in the
opposite direction of the “ U” direction while inserting or
withdrawing the endoscope into or from the patient. If the
elevator control lever is moved all the way in the “ U”
direction and the forceps elevator is raised while inserting or
withdrawing the endoscope into or from the patient, this may
cause patient injury.
> 10 cm
Figure 4.2
1. Move the elevator control lever in the opposite direction of the “ U” until it
stops.
3. Place the mouthpiece between the patient’s teeth or gums, with the outer
flange on the outside of the patient’s mouth.
1. Operate the angulation control knobs as necessary to guide the distal end
for insertion and observation.
2. The endoscope’s angulation locks are used to hold the angulated distal end
in position.
• If the sterile water level in the water container is too low, then
air, not water, will be supplied. In this case, turn the airflow
regulator on the light source OFF and add sterile water to the
water container until it reaches the specified water level.
• If air/water feeding does not stop, turn the airflow regulator
on the light source OFF and replace the air/water valve with a
new one.
Air/water feeding
1. Cover the air/water valve’s hole to feed air from the air/water nozzle at the
distal end (see Figure 4.3).
2. Depress the air/water valve to the first stage to feed water onto the objective
lens (see Figure 4.3).
Suction valve
Air/water valve
Figure 4.3
Suction
During the procedure, make sure that the suction bottle does
not fill completely. Aspirating fluids into a full bottle may
cause the suction pump to malfunction.
Depress the suction valve to the first stage to aspirate excess fluid or other
debris obscuring the endoscopic image (see Figure 4.3).
Refer to the light source’s instruction manual for instructions on how to adjust the
brightness.
1. Position the distal end of the endoscope near the target site and push the
“FREEZE” switch on the ultrasound system to change the ultrasound image
to the REAL-TIME mode.
2. Connect the water supply unit’s water-feeding valve to the biopsy valve of
the endoscope.
3. Depress the water supply unit’s water feeding valve as described in its
instruction manual and supply the necessary amount of sterile deaerated
water.
1. Position the distal end of the endoscope near the target site and press the
“FREEZE” switch on the ultrasound system to change the ultrasound image
to the REAL-TIME mode.
2. Cover the small hole in the air/water valve of the endoscope and completely
depress the valve to inflate the balloon.
3. While viewing the ultrasound image, inflate the balloon to the desired size.
Do not inflate the balloon to a diameter greater than 3 cm.
4. Use the endoscope’s angulation mechanism to bring the balloon into full
contact with the intestinal wall. Move the distal end to the target site.
Observation
When the ultrasound image of the target appears, adjust the ultrasound system
as described in its instruction manual to obtain a suitable image.
2. Move the elevator control lever all the way in the “ U” direction.
4. Confirm that the tip of the EndoTherapy accessory is closed or retracted into
its sheath and insert the EndoTherapy accessory slowly and straight into the
slit of the biopsy valve.
8. Manipulate the elevator control lever to adjust the height of the elevator (see
Figures 4.4 and 4.5).
Figure 4.4
Figure 4.5
Withdraw the EndoTherapy accessory slowly while the tip of the EndoTherapy
accessory is closed and/or retracted into its sheath.
1. Close the tip of the EndoTherapy accessory and/or retract it into its sheath.
1. When using the balloon method, aspirate water from the balloon by
depressing the suction valve while covering the hole in the valve.
1. Use a clean, lint-free cloth to gently wipe and dry the balloon surface.
2. Roll up the rear end of the balloon with your fingers (see Figure 4.6).
3. After the removal of the balloon, confirm that the surface of the ultrasound
transducer is free from scratches. In case the ultrasound transducer surface
is scratched, stop using the endoscope and contact Olympus.
Figure 4.6
Figure 4.7
5.1 Instructions
• Chapters 5, “Reprocessing: General Policy” through 8, “Cleaning and
Disinfection Equipment” describe recommended procedures and
equipment for cleaning and disinfecting or sterilizing this instrument.
• This instruction manual contains essential information on reprocessing
this instrument safely and effectively.
• Before reprocessing, thoroughly review the manuals of the reprocessing
chemicals and all equipment that will be used, and reprocess the
equipment as instructed.
• Keep this and all related instruction manuals in a safe, accessible
location.
• If you have any questions or comments about any information in this
manual, or if a problem that cannot be solved occurs while reprocessing,
contact Olympus.
For the types and conditions of the means of cleaning, disinfection, and
sterilization to be adopted, please make judgments from your professional
viewpoints.
5.3 Precautions
Balloon (MAJ-213)4
Cleaning brush (MAJ-1534)
Balloon 3 (MAJ-249)
Single use channel
cleaning brush (BW-201T)
Single use channel-opening
cleaning brush (MAJ-1339)
Ultrasonic cable (MAJ-1597)
Table 6.1
When nonsterile water is used after disinfection, wipe the endoscope and flush
the channels with 70% ethyl or isopropyl alcohol, then air-dry all internal
channels to inhibit the growth of residual bacteria. Do not reuse rinse water.
Table 6.2
Table 6.3
Process Parameters
Prevacuum Temperature 132 – 134C
(270 – 274F)
Exposure time 5 minutes
Channel-opening cleaning
brush (MH-507) or Cleaning brush Water-resistant cap
single use channel-opening Biopsy valve (MAJ-1534) (MH-553, 2 pcs)
cleaning brush (MAJ-853)
(MAJ-13391)
• Detergent solution • Large basin with a • Large basins for • Clean, lint-free cloths
tight-fitting lid for rinsing
detergent and
disinfectant solution
• Clean water • Large basin for rinsing • 500 cm3 (500 ml) • Sterile, cotton swabs
and leakage testing container
• 70% ethyl or • Disinfectant solution • 30 cm3 (30 ml) syringe • 5 cm3 (5 ml) syringe
isopropyl alcohol
Figure 7.1
The two water-resistant caps are attached to the videoscope cable connector
and the ultrasonic cable connector on the endoscope to protect the connectors
from water penetration during reprocessing. For leakage testing, the venting
connector on the water-resistant cap must be connected to the leakage tester
(MB-155) (see Figure 7.2).
Groove
Venting connector
Seal
Figure 7.2
The channel cleaning brush is used to brush the inside of the instrument
channel, suction channel, and the interior and/or openings of the suction valve,
air/water valve, AW channel cleaning adapter, and biopsy valve (see Figure 7.3).
Brush head
Figure 7.3
The single use channel cleaning brush is used to brush the inside of the
instrument channel, suction channel, and the interior and/or openings of the
suction valve, AW channel cleaning adapter, and biopsy valve (see Figure 7.4).
Tip
Figure 7.4
The channel cleaning brush or the single use single-ended cleaning brush is
used to clean the balloon channel (see Figure 7.5).
Metal tip
Shaft
Figure 7.5
The channel-opening cleaning brush is used to brush the external surface of the
distal end of the endoscope, the suction cylinder, the irrigation port, and the
instrument channel port (see Figure 7.6).
Brush handle
Brush head
Shaft
Figure 7.6
The single use channel-opening cleaning brush is used to brush the external
surface of the distal end of the endoscope, the suction cylinder, the irrigation
port, and the instrument channel port (see Figure 7.7).
Brush head
Handle
Shaft
Figure 7.7
The cleaning brush is used to brush the external surface of the distal end of the
endoscope (see Figure 7.8).
Grip section
Figure 7.8
The suction cleaning adapter is used to aspirate reprocessing fluids from the
distal end of the endoscope through the instrument channel (see Figure 7.9).
Weighted end
Connecting end
Figure 7.9
Notice card
Pakings
Piston
Figure 7.10
The injection tube is used to inject detergent solution, disinfectant solution, water
and alcohol into the air/water channel, instrument channel and suction channel
and to flush air through the channels to expel fluids (see Figure 7.11).
Connector plug
Suction channel tube
Notice card
Figure 7.11
The channel plug is used to plug the openings of the instrument channel port,
air/water and suction cylinders during cleaning (see Figure 7.12).
Cylinder plug
Figure 7.12
Luer port
Connecting end
Figure 7.13
The cleaning adapter for instrument channel port is attached to the instrument
channel port during reprocessing so that the endoscope may be connected to
reprocessing equipment (see Figure 7.14).
Notice card
Figure 7.14
The channel cleaning brush part of the single use combination cleaning brush is
used to brush the inside of the instrument channel, suction channel, and the
interior and/or openings of the suction valve, AW channel cleaning adapter, and
biopsy valve. The channel-opening cleaning brush part of the single use
combination cleaning brush is used to brush the external surface of the distal
end of the endoscope, the suction cylinder, the irrigation port, and the instrument
channel port (see Figure 7.15).
Figure 7.15
1. Confirm that the inside of the water-resistant caps is dry and free from
debris (see Figure 7.2 on page 90). Wipe with a dry cloth when the inside of
the water-resistant cap is wet or debris is detected.
2. Confirm that the seal inside the water-resistant cap is free from scratches,
flaws, and debris.
Check for cracks, scratches, flaws, debris and other damage (see Figure 7.9 on
page 93).
Check for cracks, scratches, flaws, debris and other damage (see Figure 7.10 on
page 94).
1. Confirm that all components of the injection tube are free from cracks,
scratches, flaws and debris (see Figure 7.11 on page 94).
3. Attach the 30 cm3 (30 ml) syringe to the air/water channel port. With the
filter mesh immersed in rinse water, withdraw the syringe plunger and
confirm that rinse water is drawn into the syringe. Depress the plunger and
confirm that rinse water is emitted from the air pipe port. Confirm that water
is not emitted from the suction port.
4. Attach the 30 cm3 (30 ml) syringe to the suction channel port. With the filter
mesh immersed in rinse water, withdraw the syringe plunger and confirm
that rinse water is drawn into the syringe. Depress the plunger and confirm
that rinse water is emitted from the distal end of the suction channel tube.
Confirm that water is not emitted from the suction port.
Confirm that the cylinder plug and biopsy valve cap are free from cracks,
scratches, flaws and debris (see Figure 7.12 on page 95).
Check for cracks, scratches, flaws, debris and other damage (see Figure 7.14 on
page 96).
Check for cracks, scratches, flaws, debris and other damage (see Figure 7.13 on
page 95).
1. Confirm that the brush section and the metal tip at the distal end are
securely in place. Check for loose or missing bristles (see Figure 7.4 on
page 91).
3. Check for debris on the shaft and/or in the bristles of the brush head.
1. Check the brush head for loose or missing bristles (see Figure 7.6 on
page 92).
3. Check for debris on the shaft and/or in the bristles of the brush head.
1. Check the brush head for loose or missing bristles (see Figure 7.8 on
page 93).
2. Confirm that the tip and bristles at the distal end are securely in place.
Check for loose or missing bristles.
4. Check for debris on the shaft and/or in the bristles of the brush.
Tip
Packaging
Figure 7.16
4. Check for debris on the shaft and/or in the bristles of the brush head.
Packaging Shaft
Figure 7.17
1. Remove the brush from its packaging just prior to use (see Figure 7.18).
2. Confirm that the channel cleaning brush part and the tip at the distal end are
securely attached.
3. Check the channel cleaning brush and the channel-opening cleaning brush
parts for loose or missing bristles.
4. Check the bristles of the channel cleaning brush and the channel-opening
cleaning brush parts for any damage. If the bristles are crushed, gently
straighten them with your fingertips.
5. Check the shaft for bends, scratches, and other damage. Do not reprocess
the single use combination cleaning brush prior to use. The brush may be
damaged.
Brush head
Handle
Tip
Shaft
Packaging
Figure 7.18
7.3 Precleaning
Preclean the endoscope at the bedside in the procedure room immediately after
each procedure. These steps are to be performed when the suction pump is still
connected to the endoscope. During precleaning, wear appropriate personal
protective equipment.
Equipment needed
Prepare the following equipment, and wear appropriate protective equipment.
Preparation
1. Turn the ultrasound center, video system center and light source OFF.
2. Wipe the entire insertion section with a clean, lint-free cloth soaked in
detergent solution. Wipe from the boot at the control section toward the
distal end.
4. Remove the distal end of the insertion section from the detergent solution.
Depress the suction valve to the first stage and aspirate air for 10 seconds.
Then depress the suction valve completely and aspirate air for 10 seconds.
Suction valve
Detergent solution
Figure 7.19
3. Depress the air/water valve completely to feed water into the balloon
channel and wash away any liquid that remained in the balloon channel.
5. Remove the air/water valve from the endoscope and place it in a container
of detergent solution (see Figure 7.20).
Water may drip from the air/water valve or cylinder when the
air/water valve is detached. The water dripping from the
air/water valve or cylinder is clean (i.e., sterile water in the
water container). If water is dripping from the air/water valve
or cylinder, hold the control section higher than the water
container during water feeding.
6. Attach the air/water channel cleaning adapter to the air/water cylinder of the
endoscope (see Figure 7.20).
8. Immerse the distal end of the insertion tube in the water (see Figure 7.19).
9. Depress the air/water channel cleaning adapter to feed water through the
channels for about 30 seconds. Release the adapter to feed air through the
channels for 10 seconds or more.
10. Turn the light source, the EVIS video system center and the ultrasound
imaging system OFF.
Figure 7.20
Flush detergent solution and air into the elevator wire channel
1. Connect the washing tube to the elevator channel plug (see Figure 7.21).
Washing tube
Figure 7.21
2. Immerse the distal end of the insertion section in the clean water.
3. Use the 5 cm3 (5 ml) syringe, to slowly flush detergent solution through the
elevator wire channel several times until no air bubbles exit the distal end
(see Figure 7.22).
4. Use the 5 cm3 (5 ml) syringe to slowly flush clean water through the elevator
wire channel several times (see Figure 7.22).
5. Use the 5 cm3 (5 ml) syringe to slowly flush air through the elevator wire
channel several times until a steady stream of air bubbles exit the distal end
(see Figure 7.22).
Elevator channel
Elevator wire channel plug
Endoscope connector
Figure 7.22
3. Disconnect the suction tube from the suction connector of the endoscope
connector.
4. Disconnect the metal tip of the water container from the air/water supply
connector of the endoscope connector. Then, attach the metal tip to the tip
receptacle on the lid of the water container as described in the water
container’s instruction manual.
7. Detach the air/water channel cleaning adapter, suction valve, and biopsy
valve from the endoscope and place them in a container of detergent
solution.
Equipment needed
Prepare the following equipment, and wear appropriate personal protective
equipment.
1. Confirm that the inside of the water-resistant cap is dry and free from debris.
Wipe with a dry, lint-free clean cloth if the inside of the water-resistant cap is
wet or there is debris.
2. Align the EW character (see Figures 7.23 (a) and 7.24 (a)) on the
water-resistant caps with mark 2 on the electrical connector housing and the
ultrasonic cable connector housing. When using a manual disinfector, align
the KC/TD character (see Figures 7.23 (b) and 7.24 (b)) on the
water-resistant caps with mark 2 on the electrical connector housing and the
ultrasonic cable connector housing.
3. Align the pin on the electrical connector and the ultrasonic cable connector
with the grooves on the water-resistant caps.
4. Push the water-resistant caps into position and rotate each of them
clockwise until they stop (approximately 45).
Water-resistant
cap
2. 2.
Rotate Rotate
3. 3.
4. 4.
Figure 7.23
Water-resistant
cap
2. 2.
Rotate Rotate
3. 3.
4. 4.
Figure 7.24
1. Fill a basin with clean water. Use a basin that is at least 40 cm by 40 cm (16”
by 16”) in size and deep enough to allow the endoscope to be completely
immersed.
2. Insert the leakage tester connector into the output socket of the
maintenance unit or the light source (see Figure 7.25) and turn the
maintenance unit or the light source ON. Set the light source’s airflow
regulator switch to its maximum level.
Maintenance unit 2.
or light source
Water-resistant
4.
cap (MH-553)
Leakage tester
connector cap
Venting connector
Figure 7.25
3. Confirm that the leakage tester is emitting air by gently depressing the pin
located inside the leakage tester’s connector cap.
4. Connect the leakage tester’s connector cap to either one of the two venting
connectors that are in the water-resistant caps of the video cable connector
and the ultrasonic cable connector (see Figure 7.25).
5. With the leakage tester connected, immerse the entire endoscope in water
and observe for approximately 30 seconds while angulating the bending
section. Confirm that there is no location on the endoscope from that a
continuous series of bubbles emerges.
6. Remove the endoscope from the basin with the leakage tester attached.
8. Disconnect the leakage tester from the maintenance unit or the light source.
9. Wait 30 seconds, or until the covering of the bending section contracts to its
pre-expansion size. Disconnect the leakage tester’s connector cap from the
venting connector.
After completing the leakage test, perform manual cleaning according to the
procedures described below.
Equipment needed
Prepare the following equipment and wear appropriate personal protective
equipment.
Preparation
1. Fill a basin with fresh detergent solution at the temperature and
concentration recommended by the detergent manufacturer. Use a basin
that is at least 40 cm by 40 cm (16” by 16”) in size and deep enough to
completely immerse the endoscope.
2. Fill a basin with clean water. Use a basin that is at least 40 cm by 40 cm (16”
by 16”) in size and deep enough to completely immerse the endoscope.
Ultrasound transducer
Lint-free cloth
Air/water nozzle
opening
Objective lens
Figure 7.26
1. Turn the elevator control lever all the way in the opposite direction of the
“ U” direction. perform the following brushing in the detergent solution.
2. While holding the distal end, brush the groove of the interior of the forceps
elevator with the cleaning brush (MAJ-1534) until all debris is removed (see
Figure 7.27).
Forceps elevator
Cleaning brush
(MAJ-1534)
Figure 7.27
3. Look at the left side of the instrument channel opening from the distal end,
brush the interior of this part with the cleaning brush (MAJ-1534) until all
debris is removed (see Figure 7.28).
Cleaning brush
(MAJ-1534)
Figure 7.28
4. Look at the right side of the instrument channel opening from the distal end,
brush the interior of this part with the cleaning brush (MAJ-1534) until all
debris is removed (see Figure 7.29).
Cleaning brush
(MAJ-1534)
Figure 7.29
5. Move the elevator control lever in the “ U” direction to raise the forceps
elevator.
6. While holding the distal end, brush both sides of the forceps elevator and
the opposite side of the groove of the forceps elevator with the cleaning
brush (MAJ-1534) until all debris is removed (see Figure 7.30).
Forceps elevator
Cleaning brush
(MAJ-1534)
Figure 7.30
7. Brush the distal end of the endoscope, using the cleaning brush
(MAJ-1534).
8. Move the elevator control lever all the way in the opposite direction of the
“ U” direction to lower the forceps elevator.
2. With the forceps elevator raised, flush the interior of the forceps elevator
with detergent solution using the 30 cm3 (30 ml) syringe (see Figure 7.31).
Forceps elevator
Figure 7.31
While the endoscope is submerged, brush the instrument and the suction
channel, suction cylinder, instrument channel port, balloon channel and irrigation
port according to the following procedures (see Figure 7.32).
Suction cylinder
A
B
C
E
B
B
Instrument
channel port
D
C
A A
Suction
connector
Figure 7.32
3. Insert the channel cleaning brush (BW-20T), the single use channel
cleaning brush (BW-201T), or the single use combination cleaning brush
(BW-412T) into the center opening of the suction cylinder as illustrated by A
in Figure 7.32. Using short strokes, feed the brush through the insertion tube
until it emerges from the distal end of the endoscope.
4. Clean the bristles with your fingertips in the detergent solution. Carefully pull
the brush out through the channel and back out of the suction cylinder.
1. Grip the channel cleaning brush (BW-20T), the single use channel cleaning
brush (BW-201T), or the single use combination cleaning brush (BW-412T)
at a point 3 cm from the bristles.
2. Insert the channel cleaning brush (BW-20T), the single use channel
cleaning brush (BW-201T), or the single use combination cleaning brush
(BW-412T) straight into the opening of the suction cylinder as illustrated by
B in Figure 7.32. Using short strokes, feed the brush through the universal
cord until it emerges from the suction connector on the endoscope
connector.
3. Clean the bristles with your fingertips in the detergent solution. Carefully pull
the brush out through the channel and back out of the suction cylinder.
1. Grip the channel cleaning brush (BW-7L) or the single use single-ended
cleaning brush (BW-400L) at a point 3 cm from the bristles.
2. Insert the channel cleaning brush (BW-7L) or the single use single-ended
cleaning brush (BW-400L) straight into the opening of the suction cylinder
as illustrated by C in Figure 7.32. Using short strokes, feed the brush until it
reaches the distal end near the balloon attachment groove.
It can be confirmed whether the brush has reached the distal end by looking
at the tip of brush through the groove of the balloon water suction port
(Rough standard of brush insertion length: 150 cm from point).
3. Pull the brush out through the channel and clean the bristles with your
fingertips in the detergent solution.
3. Pull the brush out and clean the bristles with your fingertips in the detergent
solution.
3. Pull the brush out and clean the bristles with your fingertips in the detergent
solution.
5. Insert the channel-opening cleaning brush into the other instrument channel
port and repeat Steps 1 through 3 until all debris is removed.
2. Attach the cleaning adapter for instrument channel port to the instrument
channel port.
3. Attach the suction cleaning adapter to the cleaning adapter for instrument
channel port (see Figures 7.33 and 7.34).
4. Connect the suction tube from the suction pump to the suction connector on
the endoscope. Turn the suction pump ON.
5. Immerse both the endoscope’s distal end and the weighted end of the
suction cleaning adapter in the detergent solution.
6. Depress the suction valve to the first stage and aspirate detergent solution
for approximately 30 seconds.
7. Remove both the endoscope’s distal end and the weighted end of the
suction cleaning adapter from the detergent solution.
8. Depress the suction valve to the first stage and aspirate air for
approximately 30 seconds.
10. Depress the suction valve completely and aspirate detergent solution for
approximately 30 seconds.
11. Remove the endoscope’s distal end from the detergent solution.
12. Depress the suction valve completely and aspirate air for approximately
30 seconds.
14. Disconnect the suction tube and remove the suction valve and the suction
cleaning adapter.
15. Reprocess the suction tube, the suction cleaning adapter and suction valve
as described in Section 7.9, “Cleaning, disinfection, and sterilization
procedures for reusable parts and reprocessing equipment” on page 139.
Suction cleaning
adapter
Weighted end
Figure 7.33
Suction Instrument
port channel port
Suction
Balloon channel Instrument channel valve
Suction
cylinder
Endoscope connector Universal cord
Suction
Suction connector channel
Suction tube
Suction pump
Suction bottle
Figure 7.34
2. Lower the channel plug’s cylinder plug onto the air/water and suction
cylinders and slide the plug until it stops (see Figure 7.35).
Suction cylinder
Biopsy valve cap
Instrument
channel port
Figure 7.35
3. Attach the injection tube’s connector plug to the air and water supply
connectors on the endoscope connector (see Figures 7.36 and 7.37).
4. Attach the injection tube’s air pipe port to the air pipe on the endoscope
connector (see Figures 7.36 and 7.37).
5. Attach the injection tube’s suction channel tube to the suction connector on
the endoscope connector (see Figures 7.36 and 7.37).
6. Immerse the suction port of the injection tube in the detergent solution.
Syringe
Suction channel tube Air/water
channel tube
Suction port
Air pipe port
Figure 7.36
7. Attach the 30 cm3 (30 ml) syringe to the injection tube’s air/water channel
port (see Figure 7.37).
8. Inject 150 cm3 (150 ml) of detergent solution into the air/water channel and
the balloon channel.
9. Detach the channel plug, the cleaning adapter for instrument channel port
and injection tube from the endoscope, and leave all items immersed.
Channel plug
Air channel Air/water cylinder
Water channel
Suction
cylinder
Connector plug
Water supply
connector
Universal cord
Air pipe port
Injection tube
Suction port
Air/water
channel port
Suction channel port
Syringe
Figure 7.37
Washing tube
Figure 7.38
2. Use the 5 cm3 (5 ml) syringe to inject 15 cm3 (15 ml) of detergent solution
into the elevator wire channel via the washing tube (see Figure 7.39).
Elevator channel
plug
Elevator wire channel
Endoscope connector
Figure 7.39
2. Cover the basin with a tight-fitting lid to minimize the diffusion of detergent
vapors.
3. Soak the endoscope and all reprocessing equipment for the amount of time
and at the temperature and concentration recommended by the detergent
manufacturer.
4. Remove the endoscope and all reprocessing equipment from the detergent
solution.
6. Place the endoscope and all reprocessing equipment in clean water and
gently agitate them to thoroughly rinse.
2. Attach the 30 cm3 (30 ml) syringe to the suction channel port and inject
150 cm3 (150 ml) of clean water into the suction channel and the balloon
channel (see Figures 7.37 and 7.40).
3. Attach the 30 cm3 (30 ml) syringe to the injection tube’s air/water channel
port and inject 150 cm3 (150 ml) of clean water into the air/water channel
and the balloon channel (see Figures 7.37 and 7.40).
4. Connect the washing tube to the elevator channel plug (see Figure 7.38).
5. Attach the 5 cm3 (5 ml) syringe to the washing tube and flush the elevator
wire channel with 5 cm3 (5 ml) of clean water (see Figure 7.39).
6. Remove the endoscope, together with all equipment, from the water.
7. Cover the distal end and control section of the endoscope with a clean,
lint-free cloth.
8. Using the 30 cm3 (30 ml) syringe, flush the air/water, balloon, and suction
channels with 150 cm3 (150 ml) of air, respectively, via the injection tube
(see Figures 7.37 and 7.40).
9. Use the 5 cm3 (5 ml) syringe, to flush the elevator wire channel with 10 cm3
(10 ml) of air via the washing tube (see Figure 7.39).
10. Remove the clean, lint-free cloth from the distal end and the control section
of the endoscope.
11. Detach the channel plug, the cleaning adapter for instrument channel port
and injection tube from the endoscope.
Instrument channel
Balloon channel
Suction channel
Universal cord
Suction
channel tube
Suction connector
Suction port
Injection tube
Air/water
channel port
Syringe
Figure 7.40
2. Inspect the endoscope and all equipment for residual debris. Should debris
remain, repeat the procedures given in this section.
2. Carefully coil the endoscope’s insertion tube and universal cord, and
completely immerse the endoscope in the detergent solution.
3. Fill all channels with the detergent solution following the procedures
described in this section.
5. Cover the basin with a tight-fitting lid to minimize the diffusion of detergent
vapors.
Equipment needed
Prepare the following equipment and wear appropriate personal protective
equipment.
Preparation
1. Fill a basin with disinfectant solution at the temperature and concentration
recommended by the disinfectant manufacturer. Use a basin that is at least
40 cm by 40 cm (16” by 16”) in size and deep enough to completely
immerse the endoscope.
2. Connect the cleaning adapter for the instrument channel port to the
endoscope.
3. Attach the channel plug and the injection tube to the endoscope. Attach the
channel plug’s biopsy valve cap to the cleaning adapter for instrument
channel port (see Figures 7.35 and 7.36).
4. Connect the washing tube to the elevator channel plug (see Figure 7.38).
2. Confirm that the suction port of the injection tube is immersed in disinfectant
solution.
3. Using the 30 cm3 (30 ml) syringe, flush at least 150 cm3 (150 ml) of
disinfectant solution through the air/water and suction channels respectively.
Confirm that no bubbles exit the distal end of the endoscope.
4. Using the washing tube and the 5 cm3 (5 ml) syringe, flush 15 cm3 (15 ml)
of disinfectant solution through the elevator wire channel. Confirm that no
bubbles exit the distal end of the endoscope.
5. While immersed in disinfectant solution, raise the forceps elevator and flush
the vicinity of the forceps elevator with disinfectant solution using the 30 cm3
(30 ml) syringe.
6. With the endoscope, channel plug, injection tube, cleaning adapter for the
instrument channel port and washing tube completely immersed in
disinfectant solution, disconnect all equipment from the endoscope. Leave
the endoscope and all equipment immersed in the disinfectant solution.
2. Immerse the distal end in disinfectant solution, and move the elevator
control lever in the “U” direction to slope the forceps elevator.
4. Cover the basin with a tight-fitting lid to minimize the diffusion of disinfectant
vapors.
5. Soak the endoscope and all equipment in disinfectant solution for the
amount of time and at the temperature and concentration recommended by
the disinfectant manufacturer.
2. Remove the injection tube’s suction port from the disinfectant solution.
3. Attach the 30 cm3 (30 ml) syringe to the injection tube’s air/water channel
port and flush the air/water channel with 150 cm3 (150 ml) of air.
4. Attach the 30 cm3 (30 ml) syringe to the injection tube’s suction channel port
and flush the suction channel with 150 cm3 (150 ml) of air.
5. Attach the 5 cm3 (5 ml) syringe to the washing tube and flush the elevator
wire channel with 15 cm3 (15 ml) of air.
6. Remove the endoscope and all equipment from the disinfectant solution.
Equipment needed
Prepare the following equipment, and wear appropriate personal protective
equipment.
2. Immerse the endoscope, channel plug, injection tube and cleaning adapter
for the instrument channel port in the sterile water. Using a sterile, lint-free
cloth, thoroughly rinse and wipe all external surfaces.
3. Connect the cleaning adapter for the instrument channel port, the channel
plug, and the injection tube to the endoscope. Place the suction port in
sterile water.
4. Attach the 30 cm3 (30 ml) syringe to the injection tube’s air/water channel
port and inject 150 cm3 (150 ml) of sterile water into the air/water channel
and the balloon channel.
5. Attach the 30 cm3 (30 ml) syringe to the suction channel port and inject
150 cm3 (150 ml) of sterile water into the suction channel and the balloon
channel.
7. Using the 5 cm3 (5 ml) syringe, inject 15 cm3 (15 ml) of sterile water into the
elevator wire channel.
8. Remove the endoscope, together with all equipment, from the water.
9. Using the 30 cm3 (30 ml) syringe, flush the air/water and suction channels
with 150 cm3 (150 ml) of air.
10. Using the 5 cm3 (5 ml) syringe, flush the elevator wire channel with 15 cm3
(15 ml) of air.
11. Disconnect the injection tube only. Connect a sterile suction tube from the
suction pump to the suction connector on the endoscope. Turn the suction
pump ON and aspirate air for at least 15 seconds.
12. Turn the suction pump OFF and disconnect all equipment from the
endoscope.
13. Use a sterile, lint-free cloth to thoroughly wipe and dry the external surfaces
of the endoscope and all equipment.
2. Inject nonsterile water and air following the procedure given in “Sterile water
rinse” on page 136.
3. Immerse the suction port of the injection tube in the alcohol. Using the
30 cm3 (30 ml) syringe, flush the air/water and suction channels with
150 cm3 (150 ml) of the alcohol, respectively.
4. Remove the suction port of the injection tube from the alcohol. Flush the
air/water and suction channels with 150 cm3 (150 ml) of air.
5. Using the 5 cm3 (5 ml) syringe, flush the washing tube with 15 cm3 (15 ml)
of the alcohol through the elevator wire channel.
6. Using the 5 cm3 (5 ml) syringe, flush the washing tube with 15 cm3 (15 ml)
of air through the elevator wire channel.
9. Use a sterile, lint-free cloth to thoroughly wipe and dry the external surfaces
of the endoscope and all equipment.
10. Use a sterile, lint-free cloth moistened with alcohol to thoroughly wipe the
external surfaces of the endoscope and all equipment.
11. Connect the cleaning adapter for instrument channel port, channel plug and
injection tube to the endoscope. Using the 30 cm3 (30 ml) syringe, flush the
air/water channel with 150 cm3 (150 ml) of air.
12. Using the washing tube and the 5 cm3 (5 ml) syringe, flush 15 cm3 (15 ml)
of air through the elevator wire channel.
13. Disconnect the injection tube only. Connect a clean suction tube from the
suction pump to the suction connector on the endoscope and aspirate air for
15 seconds.
15. Use sterile cotton swabs to dry the inside of the air/water cylinder, suction
cylinder, and instrument channel port.
7.8 Sterilization
Figure 7.41
This section includes cleaning, disinfection, and sterilization procedures for the
reusable parts and reprocessing equipment listed below.
Equipment needed
Prepare the following equipment, and wear appropriate personal protective
equipment.
Manual cleaning
1. Fill a basin with clean water. Use a basin that is deep enough to completely
immerse all equipment.
2. Fill a basin with the detergent solution at the temperature and concentration
recommended by the detergent manufacturer. Use a basin that is deep
enough to completely immerse all equipment.
3. Detach the biopsy valve’s cap from the main body, before immersing in the
detergent solution (see Figure 7.42).
Slit
Figure 7.42
4. Immerse all equipment in the detergent solution. Using a clean, soft brush or
lint-free cloth, meticulously clean all external surfaces in detergent solution.
5. Clean the bristles of the cleaning brushes (BW-20T, BW-7L and MH-507) or
the single use single-ended cleaning brush (BW-400L) thoroughly while the
brushes are immersed.
6. While immersed, depress and release the pistons of the air/water valve,
suction valve and air/water channel cleaning adapter.
7. Using the channel cleaning brush (BW-20T, BW-7L), the single use channel
cleaning brush (BW-201T), the single use combination cleaning brush
(BW-412T), or the single use single-ended cleaning brush (BW-400L)
thoroughly brush the hole of the suction valve and air/water valve, the
openings of biopsy valve and balloon applicator until no debris can be seen
(see Figures 7.43 and 7.44). Clean the bristles of the brush while it is
immersed in detergent solution.
Hole
Figure 7.43
Channel cleaning
brush (BW-7L),
single use Hole
single-ended cleaning
brush (BW-400L)
Air/water valve
Figure 7.44
8. To clean the suction cleaning adapter, attach the 30 cm3 (30 ml) syringe and
flush the tube thoroughly with detergent solution.
9. Soak all equipment for the amount of time and at the temperature
recommended by the detergent manufacturer.
10. Remove all equipment from the detergent solution and place it in clean
water.
12. While immersed, depress and release the pistons of the air/water valve,
suction valve and air/water channel cleaning adapter.
13. Attach the 30 cm3 (30 ml) syringe to the suction cleaning adapter and flush
the tube with the clean water.
15. Attach the 30 cm3 (30 ml) syringe to the suction cleaning adapter and flush
air to remove the clean water.
16. Use a clean, lint-free cloth to thoroughly wipe and dry the external surfaces
of all equipment.
High-level disinfection
3. Use a lint-free cloth and/or 30 cm3 (30 ml) syringe, to wipe and/or flush all
surfaces with the disinfectant solution to remove all air bubbles.
4. Use the 30 cm3 (30 ml) syringe, to flush the interior and recessed parts of all
equipment while immersed. Ensure that all air bubbles are expelled.
5. While immersed, depress and release the pistons of the valves and
air/water channel cleaning adapter. Ensure that all air bubbles are expelled.
6. Use the 30 cm3 (30 ml) syringe, to flush the recessed parts of the biopsy
valve with disinfectant solution while the valve is immersed. Ensure that all
air bubbles are expelled.
8. Use the 30 cm3 (30 ml) syringe to flush the inside of the suction cleaning
adapter thoroughly with disinfectant solution while immersed in the
disinfectant solution. Ensure that all air bubbles are expelled.
9. Cover the basin with a tight-fitting lid to minimize the diffusion of disinfectant
vapor. Soak all equipment in disinfectant solution for the amount of time and
at the temperature and concentration recommended by the disinfectant
manufacturer.
10. Remove the suction cleaning adapter from the disinfectant solution. Hold
the adapter and tilt it to expel the disinfectant solution remaining in the
adapter.
1. Fill a small basin with sterile water. Use a basin that is deep enough to
completely immerse all equipment.
4. Use the 30 cm3 (30 ml) syringe to flush the interiors and recessed parts of
all equipment with sterile water while immersed. Ensure that all air bubbles
are expelled.
5. While immersed, depress and release the pistons of the air/water valve,
suction valve, and air/water channel cleaning adapter. Ensure that all air
bubbles are expelled.
7. Attach the 30 cm3 (30 ml) syringe to the suction cleaning adapter and flush
the adapter with 30 cm3 (30 ml) of sterile water while immersed. Ensure that
all air bubbles are expelled.
8. Remove the suction cleaning adapter from the sterile water. Hold the
adapter and tilt it to expel the sterile water remaining in the adapter.
9. Use the 30 cm3 (30 ml) syringe, to flush air to dry the inside of the suction
cleaning adapter.
11. Use a sterile, lint-free cloth to thoroughly wipe and dry all external surfaces.
13. Store the components following the instructions given in Chapter 9, “Storage
and Disposal”.
1. Fill a small basin with clean water. Use a basin that is deep enough to
completely immerse all equipment.
6. Use the 30 cm3 (30 ml) syringe, to flush the interiors and recessed parts of
all equipment with alcohol to remove all air bubbles.
7. While immersed, depress and release the pistons of the air/water valve,
suction valve, and air/water channel cleaning adapter. Ensure that all air
bubbles are expelled.
9. Attach the 30 cm3 (30 ml) syringe to the suction cleaning adapter and flush
the adapter with 30 cm3 (30 ml) of alcohol to remove all air bubbles.
10. Remove the suction cleaning adapter from the alcohol. Hold the adapter
and tilt it to expel the alcohol remaining in the adapter.
11. Use the 30 cm3 (30 ml) syringe to flush air to dry the inside of the suction
cleaning adapter.
13. Use a sterile, lint-free cloth, to thoroughly wipe and dry all external surfaces.
Sterilization
In this section, the sterilization methods for the equipment compatible with
ethylene oxide gas sterilization or steam sterilization (autoclaving) listed in Table
6.1 on page 82, are described below.
The channel plug (MAJ-621), the injection tube (MH-946), the cleaning adapter
for instrument channel port (MAJ-350) and washing tube (MH-974) that have
been cleaned and disinfected with the endoscope can also be sterilized by the
procedures described below.
Equipment needed
Care
1. If the cable is soiled with blood or other potentially infectious materials, wipe
off all gross debris using a soft, lint-free cloth moistened with neutral
detergent.
2. To remove dust, dirt and other non-patient debris, wipe the cable’s surface
with a soft, lint-free cloth moistened with 70% ethyl or isopropyl alcohol.
OER 1
MAJ-33, MAJ-41, MAJ-43 –
EW-30 1
Table 8.1
OER-Pro OER-AW
GF-UCT180 Group31 Group31
Table 8.2
Be sure to attach all required connectors to the endoscope and accessories. For
details concerning appropriate connectors, refer to the instructions of the AER
manufacturer.
Manually clean and disinfect or sterilize any endoscopes and accessories which
are not compatible withe the AER.
1. Detach all equipment (the air/water valve, suction valve, biopsy valve, and
water-resistant cap) from the endoscope.
2. Confirm that all surfaces of the endoscope (especially the interior of the
channels, the distal end, lens, and electrical contacts) are completely dry.
4. Hang the endoscope in the storage cabinet with the distal end hanging
freely. Make sure that the insertion section hangs vertically and as straight
as possible.
Store the balloon at room temperature, in a clean and dry environment. Do not
store it in direct sunlight. If you have any questions about any information in this
manual, contact Olympus.
2. The ultrasonic cable should be stored without being tightly coiled or twisted.
9.5 Disposal
When disposing of this endoscope or any of its components, follow all applicable
national and local laws and guidelines.
Chapter 10 Troubleshooting
Endoscope functions
Angulation
Irregularity
Possible cause Solution
description
Resistance is The angulation lock(s) is Rotate angulation lock(s) in the
encountered when engaged. “F ” direction.
rotating angulation
control knob(s).
Air/water feeding
Irregularity
Possible cause Solution
description
No air feeding. The air pump is not Press the “LOW”, “MED”, “HIGH”
operating. button on the light source as
described in the light source’s
instruction manual.
The air/water valve is Replace it with a new one.
damaged.
No water feeding. The air pump is not Press the “LOW”, “MED”, “HIGH”
operating. button on the light source as
described in the light source’s
instruction manual.
There is no sterile water in Add sterile water to fill the container
the water container. to the specified water level.
The air/water valve is Replace it with a new one.
damaged.
The air/water valve The air/water valve is dirty. Remove the air/water valve.
is sticky. Reprocess the air/water valve and
then attach it again.
The air/water valve is Replace it with a new one.
damaged.
The air/water valve An incorrect air/water valve Use a correct air/water valve.
cannot be is used.
attached. The air/water valve is Replace it with a new one.
damaged.
Suction
Irregularity
Possible cause Solution
description
The suction is The biopsy valve is not Attach it correctly.
absent or attached properly.
insufficient. The biopsy valve is Replace it with a new one.
damaged.
The suction pump is not set Adjust the suction pump’s setting as
properly. described in its instruction manual.
The suction valve is Replace it with a new one.
damaged.
The suction valve The suction valve is dirty. Remove the suction valve.
is sticky. Reprocess the suction valve and
attach it again.
The suction valve is Replace it with a new one.
damaged.
The suction valve The suction valve is Replace it with a new one.
cannot be damaged.
attached. An incorrect suction valve Use a correct suction valve.
is used.
Liquid leaks out of The biopsy valve is Replace it with a new one.
the biopsy valve. damaged.
The biopsy valve is not Attach it correctly.
attached properly.
Irregularity
Possible cause Solution
description
An image is An incompatible video Use a compatible video system
abnormal. system center is being center.
used.
An incompatible light Use a compatible light source.
source is being used.
There is no video Not all power switches are Turn all power switches ON.
image. ON.
An image is not The objective lens is dirty. Feed water to remove mucus, etc.
clear.
An image is The light source is not set Adjust the light source’s setting as
excessively dark or properly. described in its instruction manual.
bright.
EndoTherapy accessories
Irregularity
Possible cause Solution
description
An EndoTherapy An incompatible Refer to the “System chart” in the
accessory does not EndoTherapy accessory is Appendix and select a compatible
pass through the being used. EndoTherapy accessory. Confirm
instrument channel that the color code on the
smoothly. EndoTherapy accessory matches
that on the endoscope. Although
the color code of the GF-UCT180 is
orange, do not use EndoTherapy
accessories designed for a
Ø 4.2 mm or Ø 6.0 mm channel,
use only EndoTherapy accessories
designed for a o 3.7 mm channel.
The elevator The elevator wire or the Clean and disinfect or sterilize the
control lever does elevator wire channel is elevator wire channel as described
not operate dirty. in Chapter 7, “Cleaning,
smoothly. Disinfection, and Sterilization
Procedures” that has the model
number of your endoscope on the
cover.
Others
Irregularity
Possible cause Solution
description
The remote switch does The wrong remote Operate the correct remote
not work. switch is operated. switch.
The remote switch Set the remote switch function
function has been set correctly as described in the video
incorrectly. system center’s instruction
manual.
Irregularity
Possible cause Solution
description
Echo appears where Multiple echoes Move the endoscope’s distal end
there is no object. generated from an to another position and check to
object with strong see if the echo in question moves
reflection. to a greater degree than other
echoes. If this happens, multiple
echoes have coalesced and
disturbed the ultrasound image.
Keep the endoscope’s distal end
away from any object with strong
reflection.
Irregularity
Possible cause Solution
description
No water is fed into the The air/water and/or Reprocess the valve.
balloon and/or no suction valve is dirty.
aspiration.
The balloon water Reprocess the channel.
channel is dirty.
2. When the NBI endoscopic image is displayed, switch to the WLI endoscopic
image by operating the video system center and light source.
2. Operate the video system center and the light source to switch to the
endoscopic image that is still displayed.
3. Follow the procedure of Step 3 above in “Withdrawal when the WLI and NBI
endoscopic images appear on the monitor”. Carefully withdraw the
endoscope under the visible observation mode when the WLI endoscopic
image is not displayed.
2. Turn the video system center and light source OFF and then ON again. If
the WLI or NBI endoscopic image appears or the frozen image is restored,
follow the procedure given in “Withdrawal when EITHER the WLI or the NBI
endoscopic image does not appear on the monitor”, beginning from Step 2.
If all endoscopic images still do not appear or the frozen image cannot be
restored, perform the following steps.
3. Turn the video system center, light source, monitor, suction pump OFF.
Before returning the endoscope for repair, contact Olympus. With the
endoscope, include a description of the malfunction or damage and the name
and telephone number of the individual at your location who is most familiar with
the problem. Also include a repair purchase order.
When returning the endoscope for repair, follow the instructions given in
“Transporting outside the hospital” on page 75.
Suction pumps
Biopsy valve
(MAJ-853)
Ultrasonic cable
(MAJ-1597)
ALOKA
Diagnostic ultrasound
system Balloon
SSD-10 (MAJ-213)
Universal endoscopic
ProSound 71 Balloon 3
ultrasound center
ProSound F751 (MAJ-249)
(EU-ME1)
Balloon applicator
Mouthpiece (MAJ-675)
(MB-142)
Videoscope cable
EXERA II (MAJ-1430) Videoscope cable
EXERA (MAJ-843)
Electrosurgical units
EndoTherapy accessories
UES-20, UES-30
Electrosurgical accessories
PSD-30, PSD-60
ESG-100
Leakage tester
(MB-155)
Water-resistant cap
(MH-553)
Maintenance unit
(MU-1)
Air/water valve Suction valve
(MAJ-1444) (MAJ-1443)
Air/water channel
cleaning adapter
(MAJ-629)
EndoTherapy accessories
Please not that some of the accessories may not be available in some areas.
Biopsy forceps
Alligator type with
With needle Rat tooth Alligator type
rat tooth
Endoscope
GF-UCT180 FB-13Q-1 FB-39Q-1, FB-40Q-1 FB-45Q-1 FB-46Q-1
Endoscope
FB-19N-1, FB-26N-1,
GF-UCT180 FB-24Q-1 FB-50Q-1 FB-37U-1
FB-28R-1
Disposable biopsy
Biopsy forceps (fenestrated)
forceps
Alligator jaws and Alligator jaws and
Alligator jaws and
rat tooth (swinging rat tooth with needle
rat tooth Alligator jaws-step
type/elongated (swinging type/
(swinging type)
cups) elongated cups)
Endoscope
GF-UCT180 FB-53Q-1 FB-54Q-1 FB-55Q-1 FB-212U
Endoscope
GF-UCT180 FB-222U FB-232U FB-242U FG-14P-1
Grasping forceps
Rubber tips
Basket type Flower basket type Tripod type
(non-latex)
Endoscope
FG-16U, FG-18Q-1,
GF-UCT180 FG-301Q FG-20P-1 FG-45U-1
FG-22Q-1, FG-23Q-1
Rotatable
Grasping forceps Disposable grasping forceps
grasping forceps
Rat tooth with
Pentapod type Flower basket type Basket type
alligator type
Endoscope
GF-UCT180 FG-46U-1 FG-401Q FG-402Q, FG-403Q FG-44NR-1
Endoscope
GF-UCT180 PW-1V-1 FD-5U FD-210U FD-230U
Disposable
Mechanical lithotriptor mechanical
Heat probe
lithotriptor
Basket type Slide type Slide type
Endoscope
BML-202Q,
GF-UCT180 CD-21Z, CD-120U BML-2Q-1 BML-4Q-1 BML-203Q,
BML-204Q
Endoscope
GF-UCT180 M1-2U NA-200H-8022 NA-11J-KB
Electrosurgical accessories
Electrosurgical snare
Crescent Hexagonal Oval Mini oval
Endoscope
GF-UCT180 SD-5U-1, SD-7P-1 SD-6U-1, SD-8P-1 SD-9U-1, SD-11U-1 SD-12U-1, SD-13U-1
Endoscope
SD-210U-10, SD-210U-15,
GF-UCT180 SD-16U-1 SD-17U-1
SD-240U-10 SD-240U-15
Endoscope
SD-210U-25,
GF-UCT180 SD-221U-25 SD-230U-20
SD-240U-25
Immunity test IEC 60601 test level Compliance level Electromagnetic environment — Guidance
Electrostatic Contact: Same as left Floors should be made of wood, concrete, or
discharge (ESD) 2, 4, 6 kV ceramic tile that hardly produces static. If
IEC 61000-4-2 Air: floors are covered with synthetic material that
2, 4, 8 kV tends to produce static, the relative humidity
should be at least 30%.
Electrical fast 2 kV Same as left Mains power quality should be that of a typical
transient/burst for power supply lines commercial (original condition feeding the
IEC 61000-4-4 1 kV facilities) or hospital environment.
for input/output lines
Surge Differential mode: Same as left Mains power quality should be that of a typical
IEC 61000-4-5 0.5, 1 kV commercial or hospital environment.
Common mode:
0.5, 1, 2 kV
Voltage dips, short < 5% UT Same as left Mains power quality should be that of a typical
interruptions, and (> 95% dip in UT) commercial or hospital environment. If the
voltage variations for 0.5 cycle user of this instrument requires continued
on power supply operation during power mains interruptions, it
40% UT
input lines is recommended that this instrument be
(60% dip in UT)
IEC 61000-4-11 powered from an uninterruptible power supply
for 5 cycle or a battery.
70% UT
(30% dip in UT)
for 25 cycle
< 5% UT
(> 95% dip in UT)
for 5 seconds
Power frequency 3 A/m Same as left It is recommended to use this instrument by
(50/60 Hz) maintaining enough distance from any
magnetic field equipment that operates with high current.
IEC 61000-4-8
Definition: UT is the a.c. mains voltage prior to application of the test level.
Compliance
Immunity test IEC 60601 test level Electromagnetic environment — Guidance
level
Recommended separation distance
Conducted RF 3 Vrms 3 V (V1) 3.5
(150 kHz – 80 MHz) d = --------- P
IEC 61000-4-6 V1
Recommended separation distance
Radiated RF 3 V/m 3 V/m (E1) 3.5
(80 MHz – 2.5 GHz) d = --------- P
IEC 61000-4-3 E1
80 MHz – 800 MHz
7
d = ----- P
E1
800 MHz – 2.5 GHz
Definition: Where “P” is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer and “d” is the recommended separation distance in meters (m).
Symbol key
MI Mechanical Index
TISscan Soft Tissue Thermal Index in an auto-scanning mode
ITA.3 (Z1) Derated spatial peak, temporal average intensity at axial distance z1
(milliwatts per square centimeter)
z1 Axial distance corresponding to the location of max [min (W.3 (Z), ITA.3
(Z) × 1 cm2)] where Z Zbp (centimeters)
zsp For MI, the axial distance at which Pr.3 is measured for TIB, the axial
distance at which TIB is a global maximum (i.e., zsp = zB.3)
(centimeters)
[email protected] Axial distance corresponding to the maximum of the derated
spatial-peak pulse intensity integral (megapascals)
deq (Zsp) Equivalent beam diameter as a function of axial distance z, and equal
to [(4/) (W0/ITA(z))] 0.5, where ITA(z) is the temporal-average intensity
as a function of z (centimeters)
fc Center frequency (MHz). For MI, fc is the center frequency associated
with the transmit pattern giving rise to the global maximum reported
value of MI. For TI, for combined modes involving transmit patterns of
unequal center frequency, fc is defined as the overall ranges of center
frequencies of the respective transmit patterns
Dim. of Aaprt Active aperture dimensions for azimuthal (x) and elevational (y)
planes (centimeters)
PD Pulse duration (microseconds) associated with the transmit pattern
giving rise to the reported value of MI
PRF Pulse repetition frequency associated with the transmit pattern giving
rise to the reported value of MI (Hz)
pr@PIImax Peak rarefactional pressure at the point where the free-field,
spatial-peak pulse intensely integral is a maximum (megapascals)
deq@PIImax Equivalent beam diameter at the point where the free-field,
spatial-peak pulse intensity integral is a maximum (centimeters)
FL Focal length, or azimuthal (x) and elevational (y) lengths, if different
(centimeters)
IPA.3@MImax Derated pulse-average intensity at the point of global maximum
reported MI (watts per square centimeter)
ISPTA.3 Derated spatial peak temporal average intensity (milliwatts per square
centimeter)
Derated ISPTA Derated spatial peak temporal average intensity (milliwatts per square
centimeter)
PII Pulse intensely integral
TIS TIB
Index label MI TISnon-scan Non- TIC
TISscan
Aaprt<=1 Aaprt>1 scan
Maximum Index Value 0.88 – 0.71 – 1.7 0.96
p r. (MPa) 2.0
W0 (mW) – 20 24 26
min of [W(z 1), I TA.(z 1)] (mW) –
z1 (cm) –
Assoc. z bp (cm) –
Acoustic z sp (cm) 1.4
Param. [email protected] (cm) 1.4
d eq(z sp) (cm) 0.19
ƒc (MHz) 5.3 – 7.4 – 5.3 5.3
Dim of A aprt X (cm) – 0.74 – 0.74 0.74
Y (cm) – 0.50 – 0.50 0.50
PD (µsec) 0.47
PRF (Hz) 5210
p r @PII max (MPa) 2.5
Other Info. d eq @ PII max (cm) 0.18
Focal Length FL x (cm) – 4.3 – 4.3
FL y (cm) – 1.0 – 1.0
I pa. @MImax (W/cm2) 209
DVA% (%) 100 – 89 – 97 100
Function CHE B – eF ExPHD – eF ExPHD eF ExPHD
B/W Frequency Select (MHz) n/a n/a n/a n/a n/a n/a
B/W Focus n/a n/a n/a n/a n/a n/a
B/W-Range (cm) n/a n/a n/a n/a n/a n/a
B/W Line Density n/a n/a n/a n/a n/a n/a
B/W Scan Area (%) n/a n/a n/a n/a n/a n/a
Doppler Image Select PENET – PENET – PENET PENET
Doppler Frequency Select (MHz) 5 MHz – 7.5 MHz – 5 MHz 5 MHz
Doppler Focus F5 – F15 – F11 F16
Operator
Doppler Velocity Range (cm/sec) 19.9 – 36.2 – 76.0 39.8
Controls
Color Image Select n/a n/a n/a n/a n/a n/a
Color Doppler Frequency
(MHz) n/a n/a n/a n/a n/a n/a
Select
Flow Focus n/a n/a n/a n/a n/a n/a
Color Velocity Range (m/sec) n/a n/a n/a n/a n/a n/a
Color Line Density n/a n/a n/a n/a n/a n/a
Color Average n/a n/a n/a n/a n/a n/a
Flow Filter n/a n/a n/a n/a n/a n/a
Flow Scan Area (%) n/a n/a n/a n/a n/a n/a
Triplex Velocity Range n/a n/a n/a n/a n/a n/a
Display TISmax 0.71
Display TIBmax 1.7
TIS TIB
Index label MI TISnon-scan Non- TIC
TISscan
Aaprt<=1 Aaprt>1 scan
Maximum Index Value 0.88 – 0.76 – 1.7 0.89
p r. (MPa) 2.0
W (mW) – 22 22 24
min of [W(z 1), I TA.(z 1)] (mW) –
z1 (cm) –
Assoc. z bp (cm) –
Acoustic z sp (cm) 1.2
param. [email protected] (cm) 1.4
d eq(z sp) (cm) 0.19
ƒc (MHz) 5.3 – 7.4 – 5.1 5.3
Dim of A aprt X (cm) – 0.74 – 0.74 0.74
Y (cm) – 0.50 – 0.50 0.50
PD (µsec) 0.47
PRF (Hz) 5210
p r @PII max (MPa) 2.5
Other Info. d eq @ PII max (cm) 0.18
Focal Length FL x (cm) – 4.6 – 4.3
FL y (cm) – 1.0 – 1.0
I pa. @MImax (W/cm2) 209
DVA% (%) 100 – 100 – 100 100
Function CHE B – eF ExPHD – Off eF ExPHD
B/W Frequency Select (MHz) n/a n/a n/a n/a n/a n/a
B/W Focus n/a n/a n/a n/a n/a n/a
B/W-Range (cm) 4.0 – 2.0 – 2.0 2.0
B/W Line Density n/a n/a n/a n/a n/a n/a
B/W Scan Area (%) n/a n/a n/a n/a n/a n/a
Doppler Image Select n/a n/a n/a n/a n/a n/a
Doppler Frequency Select (MHz) n/a n/a n/a n/a n/a n/a
Doppler Focus n/a n/a n/a n/a n/a n/a
Operator
Doppler Velocity Range (cm/sec) n/a n/a n/a n/a n/a n/a
Controls
Color Image Select PENET – PENET – PENET PENET
Color Doppler Frequency
(MHz) 5 MHz – 7.5 MHz – 5 MHz 5 MHz
Select
Flow Focus F5 – F16 – F9 F16
Color Velocity Range (m/sec) 0.20 – 0.80 – 0.61 0.94
Color Line Density n/a n/a n/a n/a n/a n/a
Color Average n/a n/a n/a n/a n/a n/a
Flow Filter n/a n/a n/a n/a n/a n/a
Flow Scan Area (%) n/a n/a n/a n/a n/a n/a
Triplex Velocity Range n/a n/a n/a n/a n/a n/a
Display TISmax 0.76
Display TIBmax 1.7
Uncertainty
Quantity
(95% confidence)
W0 12.22%
PII 25.62%
Pr 13.19%
Fc 2.39%
ISPTA.3 62
TI type TIS, TIB
TI value 0.83
MI 0.99
IPA.3 @MImax 377
MI Mechanical index
P Output power
TI Thermal index
td Pulse duration
Uncertainty
Quantity
(95% confidence)
P 12.22%
PII 25.62%
Pr 13.19%
Fc 2.39%
TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt1cm2 Aaprt>1cm2 scan
Maximum Index Value 0.84 0.69 – – – 0.76
p r, (MPa) 2.0
P (mW) 27 – – 20
Min of [P(z s), I ta,(z s)] (mW) –
zs (cm) –
Associated z bp (cm) –
acoustic zb (cm) –
parameters z at max. Ipi, (cm) 1.2
d eq(z b) (cm) –
ƒ awf (MHz) 5.4 5.3 – – – 4.4
Dim of A aprt X (cm) 1.2 – – – 0.72
Y (cm) 0.50 – – – 0.50
td (µsec) 0.29
prr (Hz) 6940
p r at max. I pi (MPa) 2.3
Other
d eq at max. I pi (cm) –
information
I pa, at max.MI (W/cm ) 2 313
Focal Length FL x (cm) 2.8 – – 0.56
FL y (cm) 1.0 – – 1.2
DVA% (%) 100 100 – – – 100
Function CHE A Penet CHE B – – – CHE A Penet
B/W Frequency Select (MHz) 5.4 MHz 5.4 MHz – – – 4.3 MHz
B/W Focus F4 F9 – – – F2
B/W-Range (cm) 8.0 4.0 – – – 4.0
B/W Line Density HIGH HIGH – – – HIGH
B/W Scan Area (%) 25 25 – – – 25
Doppler Image Select n/a n/a n/a n/a n/a n/a
Doppler Frequency Select (MHz) n/a n/a n/a n/a n/a n/a
Operating
Doppler Focus n/a n/a n/a n/a n/a n/a
control
Doppler Velocity Range (cm/sec) n/a n/a n/a n/a n/a n/a
conditions
Color Image Select n/a n/a n/a n/a n/a n/a
Color Doppler Frequency Select (MHz) n/a n/a n/a n/a n/a n/a
Flow Focus n/a n/a n/a n/a n/a n/a
Color Velocity Range (m/sec) n/a n/a n/a n/a n/a n/a
Color Line Density n/a n/a n/a n/a n/a n/a
Color Average n/a n/a n/a n/a n/a n/a
Flow Filter n/a n/a n/a n/a n/a n/a
Flow Scan Area (%) n/a n/a n/a n/a n/a n/a
Triplex Velocity Range n/a n/a n/a n/a n/a n/a
NOTE1 Data should only be entered in one of the columns related to TIS.
NOTE2 Information need not be provided regarding TIC for any TRANSDUCER ASSEMBLY not intended for transcranial or
neonatal cephalic uses.
NOTE3 If the requirements of 201.12.4.2a are met, it is not required to enter any data in the columns related to TIS, TIB or TIC.
NOTE4 If the requirements of 201.12.4.2b are met, it is not required to enter any data in the columns related to MI.
TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt1cm2 Aaprt >1cm2 scan
Maximum Index Value 0.84 0.45 – – 0.096 0.56
p r, (MPa) 2.0
P (mW) 22 – 0.91 18
Min of [P(z s), I ta,(z s)] (mW) –
zs (cm) –
Associated z bp (cm) –
acoustic zb (cm) 0.51
parameters z at max. I (cm) 1.2
pi,
d eq(z b) (cm) 0.18
ƒ awf (MHz) 5.4 4.4 – – 4.4 4.3
Dim of A aprt X (cm) 1.1 – – 0.31 1.2
Y (cm) 0.50 – – 0.50 0.50
td (µsec) 0.29
prr (Hz) 5510
p r at max. I pi (MPa) 2.3
Other
d eq at max. I pi (cm) 0.17
information
I pa, at max.MI (W/cm 2) 313
Focal Length FL x (cm) 2.3 – – 4.3
FL y (cm) 1.2 – – 1.2
DVA% (%) 100 100 – – 99 99
Function CHE A Penet CHE A Penet – – CHE A Penet CHE A Penet
B/W Frequency Select (MHz) 5.4 MHz 4.3 MHz – – 4.3 MHz 4.3 MHz
B/W Focus F4 F7 – – F12 F12
B/W-Range (cm) 4.0 4.0 – – 4.0 4.0
B/W Line Density HIGH HIGH – – HIGH HIGH
B/W Scan Area (%) 25 25 – – 25 25
Doppler Image Select n/a n/a n/a n/a n/a n/a
Doppler Frequency Select (MHz) n/a n/a n/a n/a n/a n/a
Operating
Doppler Focus n/a n/a n/a n/a n/a n/a
control
Doppler Velocity Range (cm/sec) n/a n/a n/a n/a n/a n/a
conditions
Color Image Select n/a n/a n/a n/a n/a n/a
Color Doppler Frequency Select (MHz) n/a n/a n/a n/a n/a n/a
Flow Focus n/a n/a n/a n/a n/a n/a
Color Velocity Range (m/sec) n/a n/a n/a n/a n/a n/a
Color Line Density n/a n/a n/a n/a n/a n/a
Color Average n/a n/a n/a n/a n/a n/a
Flow Filter n/a n/a n/a n/a n/a n/a
Flow Scan Area (%) n/a n/a n/a n/a n/a n/a
Triplex Velocity Range n/a n/a n/a n/a n/a n/a
NOTE1 Data should only be entered in one of the columns related to TIS.
NOTE2 Information need not be provided regarding TIC for any TRANSDUCER ASSEMBLY not intended for transcranial or
neonatal cephalic uses.
NOTE3 If the requirements of 201.12.4.2a are met, it is not required to enter any data in the columns related to TIS, TIB or TIC.
NOTE4 If the requirements of 201.12.4.2b are met, it is not required to enter any data in the columns related to MI.
TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt 1cm2 Aaprt>1cm2 scan
Maximum Index Value 0.84 – 0.074 – 0.30 0.13
p r, (MPa) 2.0
P (mW) – 3.6 3.0 3.6
Min of [P(z s), I ta,(z s)] (mW) –
zs (cm) –
Associated z bp (cm) –
acoustic zb (cm) 1.2
parameters z at max. I (cm) 1.2
pi,
d eq(z b) (cm) 0.16
ƒ awf (MHz) 5.4 – 4.3 – 4.4 4.3
Dim of A aprt X (cm) – 0.74 – 0.55 0.74
Y (cm) – 0.50 – 0.50 0.50
td (µsec) 0.29
prr (Hz) 1000
p r at max. I pi (MPa) 2.3
Other
d eq at max. I pi (cm) 0.15
information
I pa, at max.MI (W/cm 2) 313
Focal Length FL x (cm) – 4.8 – 4.8
FL y (cm) – 1.2 – 1.2
DVA% (%) 100 – 100 – 100 100
Function CHE A Penet – CHE A Penet – CHE A Penet CHE A Penet
B/W Frequency Select (MHz) 5.4 MHz – 4.3 MHz – 4.3 MHz 4.3 MHz
B/W Focus F4 – F15 – F5 F15
B/W-Range (cm) 4.0 – 4.0 – 4.0 4.0
B/W Line Density n/a n/a n/a n/a n/a n/a
B/W Scan Area (%) n/a n/a n/a n/a n/a n/a
Doppler Image Select n/a n/a n/a n/a n/a n/a
Doppler Frequency Select (MHz) n/a n/a n/a n/a n/a n/a
Operating
Doppler Focus n/a n/a n/a n/a n/a n/a
control
Doppler Velocity Range (cm/sec) n/a n/a n/a n/a n/a n/a
conditions
Color Image Select n/a n/a n/a n/a n/a n/a
Color Doppler Frequency Select (MHz) n/a n/a n/a n/a n/a n/a
Flow Focus n/a n/a n/a n/a n/a n/a
Color Velocity Range (m/sec) n/a n/a n/a n/a n/a n/a
Color Line Density n/a n/a n/a n/a n/a n/a
Color Average n/a n/a n/a n/a n/a n/a
Flow Filter n/a n/a n/a n/a n/a n/a
Flow Scan Area (%) n/a n/a n/a n/a n/a n/a
Triplex Velocity Range n/a n/a n/a n/a n/a n/a
NOTE1 Data should only be entered in one of the columns related to TIS.
NOTE2 Information need not be provided regarding TIC for any TRANSDUCER ASSEMBLY not intended for transcranial or
neonatal cephalic uses.
NOTE3 If the requirements of 201.12.4.2a are met, it is not required to enter any data in the columns related to TIS, TIB or TIC.
NOTE4 If the requirements of 201.12.4.2b are met, it is not required to enter any data in the columns related to MI.
TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt1cm2 Aaprt >1cm2 scan
Maximum Index Value 0.88 – 0.62 – 1.7 0.93
p r, (MPa) 2.0
P (mW) – 17 22 14
Min of [P(z s), I ta,(z s)] (mW) –
zs (cm) –
Associated z bp (cm) –
acoustic zb (cm) 1.2
parameters z at max. I (cm) 1.4
pi,
d eq(z b) (cm) 0.19
ƒ awf (MHz) 5.3 – 7.5 – 5.1 5.2
Dim of A aprt X (cm) – 0.74 – 0.74 0.22
Y (cm) – 0.50 – 0.50 0.50
td (µsec) 0.47
prr (Hz) 5210
p r at max. I pi (MPa) 2.5
Other
d eq at max. I pi (cm) 0.18
information
I pa, at max.MI (W/cm 2) 209
Focal Length FL x (cm) – 3.4 – 0.27
FL y (cm) – 1.0 – 0.84
DVA% (%) 100 – 100 – 100 97
Function CHE B – Off – Off CHE A Std
B/W Frequency Select (MHz) n/a n/a n/a n/a n/a n/a
B/W Focus n/a n/a n/a n/a n/a n/a
B/W-Range (cm) n/a n/a n/a n/a n/a n/a
B/W Line Density n/a n/a n/a n/a n/a n/a
B/W Scan Area (%) n/a n/a n/a n/a n/a n/a
Doppler Image Select PENET – PENET – PENET STD
Doppler Frequency Select (MHz) 5 MHz – 7.5 MHz – 5 MHz 5 MHz
Operating
Doppler Focus F5 – F11 – F9 F1
control
Doppler Velocity Range (cm/sec) 39.8 – 53.1 – 61.3 114
conditions
Color Image Select n/a n/a n/a n/a n/a n/a
Color Doppler Frequency Select (MHz) n/a n/a n/a n/a n/a n/a
Flow Focus n/a n/a n/a n/a n/a n/a
Color Velocity Range (m/sec) n/a n/a n/a n/a n/a n/a
Color Line Density n/a n/a n/a n/a n/a n/a
Color Average n/a n/a n/a n/a n/a n/a
Flow Filter n/a n/a n/a n/a n/a n/a
Flow Scan Area (%) n/a n/a n/a n/a n/a n/a
Triplex Velocity Range n/a n/a n/a n/a n/a n/a
NOTE1 Data should only be entered in one of the columns related to TIS.
NOTE2 Information need not be provided regarding TIC for any TRANSDUCER ASSEMBLY not intended for transcranial or
neonatal cephalic uses.
NOTE3 If the requirements of 201.12.4.2a are met, it is not required to enter any data in the columns related to TIS, TIB or TIC.
NOTE4 If the requirements of 201.12.4.2b are met, it is not required to enter any data in the columns related to MI.
TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt 1cm2 Aaprt>1cm2 scan
Maximum Index Value 0.88 – 0.76 – 1.7 0.89
p r, (MPa) 2.0
P (mW) – 22 22 24
Min of [P(z s), I ta,(z s)] (mW) –
zs (cm) –
Associated z bp (cm) –
acoustic zb (cm) 1.2
parameters z at max. I (cm) 1.4
pi,
d eq(z b) (cm) 0.19
ƒ awf (MHz) 5.3 – 7.4 – 5.1 5.3
Dim of A aprt X (cm) – 0.74 – 0.74 0.74
Y (cm) – 0.50 – 0.50 0.50
td (µsec) 0.47
prr (Hz) 5210
p r at max. I pi (MPa) 2.5
Other
d eq at max. I pi (cm) 0.18
information
I pa, at max.MI (W/cm 2) 209
Focal Length FL x (cm) – 4.6 – 4.3
FL y (cm) – 1.0 – 1.0
DVA% (%) 100 – 100 – 100 100
Function CHE B – eF EPHD – Off eF EPHD
B/W Frequency Select (MHz) n/a n/a n/a n/a n/a n/a
B/W Focus n/a n/a n/a n/a n/a n/a
B/W-Range (cm) 4.0 – 2.0 – 2.0 2.0
B/W Line Density n/a n/a n/a n/a n/a n/a
B/W Scan Area (%) n/a n/a n/a n/a n/a n/a
Doppler Image Select n/a n/a n/a n/a n/a n/a
Doppler Frequency Select (MHz) n/a n/a n/a n/a n/a n/a
Operating
Doppler Focus n/a n/a n/a n/a n/a n/a
control
Doppler Velocity Range (cm/sec) n/a n/a n/a n/a n/a n/a
conditions
Color Image Select PENET – PENET – PENET PENET
Color Doppler Frequency Select (MHz) 5 MHz – 7.5 MHz – 5 MHz 5 MHz
Flow Focus F5 – F16 – F9 F16
Color Velocity Range (m/sec) 0.40 – 0.80 – 0.61 0.94
Color Line Density n/a n/a n/a n/a n/a n/a
Color Average n/a n/a n/a n/a n/a n/a
Flow Filter n/a n/a n/a n/a n/a n/a
Flow Scan Area (%) n/a n/a n/a n/a n/a n/a
Triplex Velocity Range n/a n/a n/a n/a n/a n/a
NOTE1 Data should only be entered in one of the columns related to TIS.
NOTE2 Information need not be provided regarding TIC for any TRANSDUCER ASSEMBLY not intended for transcranial or
neonatal cephalic uses.
NOTE3 If the requirements of 201.12.4.2a are met, it is not required to enter any data in the columns related to TIS, TIB or TIC.
NOTE4 If the requirements of 201.12.4.2b are met, it is not required to enter any data in the columns related to MI.
TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt1cm2 Aaprt >1cm2 scan
Maximum Index Value 0.88 – 0.50 – 1.6 0.95
p r, (MPa) 2.0
P (mW) – 14 15 16
Min of [P(z s), I ta,(z s)] (mW) –
zs (cm) –
Associated z bp (cm) –
acoustic zb (cm) 0.37
parameters z at max. I (cm) 1.4
pi,
d eq(z b) (cm) 0.18
ƒ awf (MHz) 5.3 – 7.5 – 5.3 4.3, 5.2
Dim of A aprt X (cm) – 0.74 – 0.31 0.22
Y (cm) – 0.50 – 0.50 0.50
td (µsec) 0.47
prr (Hz) 6640
p r at max. I pi (MPa) 2.5
Other
d eq at max. I pi (cm) 0.16
information
I pa, at max.MI (W/cm 2) 209
Focal Length FL x (cm) – 2.7 – 0.27
FL y (cm) – 1.0 – 0.84
DVA% (%) 100 – 100 – 92 100
Function CHE B – Off – CHE A Std CHE A Std
B/W Frequency Select (MHz) 4.7 MHz – 5 MHz – 4.7 MHz 4.3 MHz
B/W Focus F5 – F16 – F2 F12
B/W-Range (cm) 4.0 – 4.0 – 12.0 4.0
B/W Line Density HIGH – HIGH – HIGH HIGH
B/W Scan Area (%) 25 – 25 – 25 25
Doppler Image Select PENET – PENET – STD STD
Doppler Frequency Select (MHz) 5 MHz – 7.5 MHz – 5 MHz 5 MHz
Operating
Doppler Focus F5 – F8 – F2 F1
control
Doppler Velocity Range (cm/sec) 53.1 – 39.8 – 79.7 133
conditions
Color Image Select n/a n/a n/a n/a n/a n/a
Color Doppler Frequency Select (MHz) n/a n/a n/a n/a n/a n/a
Flow Focus n/a n/a n/a n/a n/a n/a
Color Velocity Range (m/sec) n/a n/a n/a n/a n/a n/a
Color Line Density n/a n/a n/a n/a n/a n/a
Color Average n/a n/a n/a n/a n/a n/a
Flow Filter n/a n/a n/a n/a n/a n/a
Flow Scan Area (%) n/a n/a n/a n/a n/a n/a
Triplex Velocity Range n/a n/a n/a n/a n/a n/a
NOTE1 Data should only be entered in one of the columns related to TIS.
NOTE2 Information need not be provided regarding TIC for any TRANSDUCER ASSEMBLY not intended for transcranial or
neonatal cephalic uses.
NOTE3 If the requirements of 201.12.4.2a are met, it is not required to enter any data in the columns related to TIS, TIB or TIC.
NOTE4 If the requirements of 201.12.4.2b are met, it is not required to enter any data in the columns related to MI.
TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt 1cm2 Aaprt>1cm2 scan
Maximum Index Value 0.88 0.91 – – – 0.82
p r, (MPa) 2.0
P (mW) 26 – – 27
Min of [P(z s), I ta,(z s)] (mW) –
zs (cm) –
Associated z bp (cm) –
acoustic zb (cm) –
parameters z at max. I (cm) 1.4
pi,
d eq(z b) (cm) –
ƒ awf (MHz) 5.3 5.5, 7.5 – – – 5.1, 5.5
Dim of A aprt X (cm) 1.1 – – – 1.1
Y (cm) 0.50 – – – 0.50
td (µsec) 0.47
prr (Hz) 8730
p r at max. I pi (MPa) 2.5
Other
d eq at max. I pi (cm) –
information
I pa, at max.MI (W/cm 2) 209
Focal Length FL x (cm) 2.7 – – 3.8
FL y (cm) 1.0 – – 0.84
DVA% (%) 100 100 – – – 100
Function CHE B Off – – – Off
B/W Frequency Select (MHz) 4.7 MHz 5 MHz – – – 5 MHz
B/W Focus F5 F15 – – – F14
B/W-Range (cm) 4.0 4.0 – – – 4.0
B/W Line Density HIGH HIGH – – – HIGH
B/W Scan Area (%) 25 25 – – – 25
Doppler Image Select n/a n/a n/a n/a n/a n/a
Doppler Frequency Select (MHz) n/a n/a n/a n/a n/a n/a
Operating
Doppler Focus n/a n/a n/a n/a n/a n/a
control
Doppler Velocity Range (cm/sec) n/a n/a n/a n/a n/a n/a
conditions
Color Image Select PENET PENET – – – PENET
Color Doppler Frequency Select (MHz) 5 MHz 7.5 MHz – – – 5 MHz
Flow Focus F5 F8 – – – F16
Color Velocity Range (m/sec) 0.76 0.50 – – – 0.61
Color Line Density LOW LOW – – – LOW
Color Average HIGH HIGH – – – HIGH
Flow Filter #4 #4 – – – #4
Flow Scan Area (%) 20 20 – – – 20
Triplex Velocity Range n/a n/a n/a n/a n/a n/a
NOTE1 Data should only be entered in one of the columns related to TIS.
NOTE2 Information need not be provided regarding TIC for any TRANSDUCER ASSEMBLY not intended for transcranial or
neonatal cephalic uses.
NOTE3 If the requirements of 201.12.4.2a are met, it is not required to enter any data in the columns related to TIS, TIB or TIC.
NOTE4 If the requirements of 201.12.4.2b are met, it is not required to enter any data in the columns related to MI.
TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt1cm2 Aaprt >1cm2 scan
Maximum Index Value 0.88 0.91 – – – 0.82
p r, (MPa) 2.0
P (mW) 26 – – 27
Min of [P(z s), I ta,(z s)] (mW) –
zs (cm) –
Associated z bp (cm) –
acoustic zb (cm) –
parameters z at max. I (cm) 1.4
pi,
d eq(z b) (cm) –
ƒ awf (MHz) 5.3 5.5, 7.5 – – – 5.1, 5.5
Dim of A aprt X (cm) 1.1 – – – 1.1
Y (cm) 0.50 – – – 0.50
td (µsec) 0.47
prr (Hz) 8730
p r at max. I pi (MPa) 2.5
Other
d eq at max. I pi (cm) –
information
I pa, at max.MI (W/cm 2) 209
Focal Length FL x (cm) 2.7 – – 3.8
FL y (cm) 1.0 – – 0.84
DVA% (%) 100 100 – – – 100
Function CHE B Off – – – Off
B/W Frequency Select (MHz) 4.7 MHz 5 MHz – – – 5 MHz
B/W Focus F5 F15 – – – F14
B/W-Range (cm) 4.0 4.0 – – – 4.0
B/W Line Density HIGH HIGH – – – HIGH
B/W Scan Area (%) 25 25 – – – 25
Doppler Image Select n/a n/a n/a n/a n/a n/a
Doppler Frequency Select (MHz) n/a n/a n/a n/a n/a n/a
Operating
Doppler Focus n/a n/a n/a n/a n/a n/a
control
Doppler Velocity Range (cm/sec) n/a n/a n/a n/a n/a n/a
conditions
Color Image Select PENET PENET – – – PENET
Color Doppler Frequency Select (MHz) 5 MHz 7.5 MHz – – – 5 MHz
Flow Focus F5 F8 – – – F16
Color Velocity Range (m/sec) 0.76 0.50 – – – 0.61
Color Line Density LOW LOW – – – LOW
Color Average HIGH HIGH – – – HIGH
Flow Filter #4 #4 – – – #4
Flow Scan Area (%) 20 20 – – – 20
Triplex Velocity Range n/a n/a n/a n/a n/a n/a
NOTE1 Data should only be entered in one of the columns related to TIS.
NOTE2 Information need not be provided regarding TIC for any TRANSDUCER ASSEMBLY not intended for transcranial or
neonatal cephalic uses.
NOTE3 If the requirements of 201.12.4.2a are met, it is not required to enter any data in the columns related to TIS, TIB or TIC.
NOTE4 If the requirements of 201.12.4.2b are met, it is not required to enter any data in the columns related to MI.
TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt 1cm2 Aaprt>1cm2 scan
Maximum Index Value 0.88 – 0.50 – 1.6 0.90
p r, (MPa) 2.0
P (mW) – 14 15 14
Min of [P(z s), I ta,(z s)] (mW) –
zs (cm) –
Associated z bp (cm) –
acoustic zb (cm) 0.37
parameters z at max. I (cm) 1.4
pi,
d eq(z b) (cm) 0.18
ƒ awf (MHz) 5.3 – 7.5 – 5.3 5.2, 5.4
Dim of A aprt X (cm) – 0.74 – 0.31 0.22
Y (cm) – 0.50 – 0.50 0.50
td (µsec) 0.47
prr (Hz) 11800
p r at max. I pi (MPa) 2.5
Other
d eq at max. I pi (cm) 0.16
information
I pa, at max.MI (W/cm 2) 209
Focal Length FL x (cm) – 2.7 – 0.27
FL y (cm) – 1.0 – 0.84
DVA% (%) 100 – 100 – 92 100
Function CHE B – Off – CHE A Std CHE A Std
B/W Frequency Select (MHz) 4.7 MHz – 5 MHz – 5.4 MHz 4.3 MHz
B/W Focus F5 – F16 – F1 F15
B/W-Range (cm) 4.0 – 4.0 – 10.0 4.0
B/W Line Density HIGH – HIGH – HIGH HIGH
B/W Scan Area (%) 25 – 25 – 25 25
Doppler Image Select PENET – PENET – STD STD
Doppler Frequency Select (MHz) 5 MHz – 7.5 MHz – 5 MHz 5 MHz
Operating
Doppler Focus F2 – F8 – F2 F1
control
Doppler Velocity Range (cm/sec) 106 – 53.1 – 93.8 133
conditions
Color Image Select PENET – PENET – STD STD
Color Doppler Frequency Select (MHz) 5 MHz – 5 MHz – 3.75 MHz 5 MHz
Flow Focus F5 – F16 – F16 F15
Color Velocity Range (m/sec) 0.33 – 0.25 – 0.33 0.33
Color Line Density LOW – LOW – LOW LOW
Color Average HIGH – HIGH – HIGH HIGH
Flow Filter #4 – #4 – #4 #4
Flow Scan Area (%) 20 – 20 – 20 20
Triplex Velocity Range HIGH – HIGH – HIGH HIGH
NOTE1 Data should only be entered in one of the columns related to TIS.
NOTE2 Information need not be provided regarding TIC for any TRANSDUCER ASSEMBLY not intended for transcranial or
neonatal cephalic uses.
NOTE3 If the requirements of 201.12.4.2a are met, it is not required to enter any data in the columns related to TIS, TIB or TIC.
NOTE4 If the requirements of 201.12.4.2b are met, it is not required to enter any data in the columns related to MI.
Distributed by
KeyMed House, Stock Road, Southend-on-Sea, Essex SS2 5QH, United Kingdom
Fax: (01702)465677 Telephone: (01702)616333
A8F, Ping An International Financial Center, No. 1-3, Xinyuan South Road,
Chaoyang District, Beijing, 100027 P.R.C.
Fax: (86)10-5976-1299 Telephone: (86)10-5819-9000
491B, River Valley Road #12-01/04, Valley Point Office Tower, Singapore 248373
Fax: 6834-2438 Telephone: 6834-0010