UV PhotoX Fluorometer Operations Manual Rev 1 - 1 February 2007
UV PhotoX Fluorometer Operations Manual Rev 1 - 1 February 2007
:
Effective Date: 03/05/07 Revision Date: 09/28/07 Revision No.: 1.1
Operation Manual:
UV PhotoX Fluorescence System
Model 53F
Notice of Confidentiality
The information contained within this document is confidential and proprietary to Custom Sensors &
Technology (CST) and may be covered under existing US or CST patents pending. This information shall
not be reproduced or further disclosed, in whole or in part, to anyone other than employees of the
company purchasing the product without prior written consent from Custom Sensors & Technology.
References
Document Name: Operation Manual: UV Photo-X Fluorescence System Model 53F
Document Number: 53F
Disclaimer
The information in this document is subject to change without notice and should not be construed as a
commitment by Custom Sensors & Technology. CST assumes no responsibility for any errors that may
appear in this document. This manual is believed to be complete and accurate at the time of publication.
In no event shall CST be liable for incidental or consequential damages in connection with or arising from
the use of this manual.
Safety Guidelines
PLEASE READ THE INSTRUCTIONS CAREFULLY BEFORE WORKING WITH THIS PRODUCT!
This device has left our facility after careful testing of all the transmitters functions and safety features.
The functioning and operational safety of the product can only be ensured if the user observes the usual
safety precautions as well as the specific safety guidelines stated in these operating guidelines:
• Before connecting the device to the electrical supply, please ensure that the operating voltage
stated on the power supply corresponds to the voltage supplied to the unit.
• The functioning and operational safety of the instrument can only be maintained under the
climatic conditions specified in the specification section of this manual
• If the instrument is moved from warm surroundings, condensate may form and interfere with the
functioning of the instrument. In this event, wait until the temperature of the transmitter
equilibrates to the new temperature before putting the transmitter back into operation.
• Transmitter controls, maintenance and repair work must only be carried out by a suitable qualified
technician.
• If there is any reason to assume that the product can no longer be employed without a risk, it
must be set aside and appropriately marked to prevent further use.
• The safety of the user may be endangered if the instrument:
o is visibly damaged;
o no longer operates as specified
o has been damaged in transport
• If you are in doubt, the product should be sent back to the factory
• The operator of this product must ensure that the following laws and guidelines are observed
when using this product around dangerous substances:
o EEC directives
o National Fire Protection Association
o Safety data-sheets of the chemical manufacturer
PhotoX Fluorometer
Installation and Operating Instructions
Table of Contents
Section 1: Unpacking ............................................................................................................ 5
Section 2: Hardware Installation...................................................................................... 7
2.1 Physical ................................................................................................................ 7
2.2 Optical................................................................................................................... 8
2.3 Electrical ............................................................................................................ 10
Section 3: Introduction to Fluorescence Monitoring ............................................... 12
3.1 Theory of Operation ...................................................................................... 12
3.2 Filter Based Fluorimetry .............................................................................. 12
Section 4: Glossary, Hardware Description, and Modes of Operation............. 14
4.1 Definition of Terms ........................................................................................ 14
4.2 Hardware Description ................................................................................... 15
4.3 Modes of Operation ....................................................................................... 16
4.3.1 Mode 1 Initialization.................................................................................. 16
4.3.2 Mode 2 Initialization.................................................................................. 16
Section 5: Instrument Controls....................................................................................... 18
5.1 Active or Operational Controls and Indicators.................................... 19
5.2 Factory and Troubleshooting Controls and Indicators..................... 20
5.3 Inactive Controls and Indicators .............................................................. 20
Section 6: Fluorometer Operations and Diagnostics .............................................. 21
6.1 Quick Start Operations Guide.................................................................... 21
6.2 Detailed Operations Guide.......................................................................... 25
6.2.1 First Power-Up Following Installation..................................................... 25
6.2.2 Mode 1 or Calibrated Operation ............................................................... 29
6.2.3 Mode 2 or Fixed Range Operation ........................................................... 32
6.2.4 Tips for Easy Operation ............................................................................... 34
6.3 Diagnostics and Troubleshooting ............................................................. 35
6.3.1 Diagnostic Indicators .................................................................................... 35
6.3.2 Basic Troubleshooting .................................................................................. 36
Section 7: System Specifications ................................................................................... 39
7.1 Optical Configuration .................................................................................... 39
7.2 Transmitter Specifications .......................................................................... 40
7.3 Light Source Specifications ........................................................................ 41
7.3.1 CST UVF-100 Xenon Flashlamp ................................................................ 41
7.4 Fluorescence Accessory Specifications .................................................. 42
7.4.1 Integrated Front Surface Probe Specifications ................................... 42
7.4.2 Stand-Alone Front Surface Probe Specifications ............................... 42
7.4.3 90º Fluorescence Flow Cell Specifications............................................ 43
Section 1: Unpacking
CST provides the 53F system in three ways: 1) Laboratory system mounted on a back
panel, 2) Process equipment mounted in an appropriate enclosure, and 3) OEM component
hardware. In all of these configurations it is critical to proper installation and setup that the
following parts be identified.
ii) Stand alone: the probe has no fiber optic cables attached
f) Calibration bottle
g) Adjustment/Initialization tool
3) Insure that required resources are available (power, purge air, etc.)
In addition to the physical, optical, and electrical installation requirements of the fluorometer it is
recommended that the transmitter be placed in a clean, dry area. If the area surrounding the
analyzer system exceeds our upper temperature limit of 115º F, it is recommended that the
transmitter be moved. For systems mounted in an enclosure, the enclosure may be purged with
clean, dry, oil free air (or nitrogen) to help dissipate heat within the enclosure.
Please follow the outlined sections below as a guideline to installing the 53F system for your
application or in the case of OEM components onto a suitable panel.
DO NOT supply power to the unit until specifically told to do so in the Instrument Setup
and Initialization procedure given in Section 6.
2.1 Physical
1. For permanent installations, bolt the fluorometer, back panel, or enclosure (as
appropriate) in place using mounting screws. The location should be secure, rigid, and
strong enough to support the weight of the installed system.
2. For laboratory use, the system is supplied on either a back panel with rubber feet or
mounting rack. Make sure there is sufficient space on a bench/table top to
accommodate the size of the hardware supplied.
3. Run required electrical wiring for instrument power and 4-20mA output signals. Review
Section 2.3 Electrical below prior to making electrical connections.
4. Run required dry and oil free purge air for the unit. (Only necessary with systems
mounted in an enclosure).
5. Run the fiber optic cable(s) in conduit or other appropriate protective measure as
desired. Refer to Section 2.2 Optical below for information on connecting the optical
components.
NOTE: Secure the fiber optic cables so that they are not free to move around, be
accidentally stepped on, leaned against, or otherwise damaged during normal
operations. The allowed bend radius for fibers supplied with the 53F system is eight
(8) inches.
6. Install probe in the sensing location, making sure there are no fluid leaks and the probe
has been fully tested under process conditions for temperature and pressure safety.
The sensing location could be the CST supplied calibration bottle for OEM and
laboratory based systems.
2.2 Optical
1. Remove the protective boots (typically black) from the SMA 905 connectors on the
fiber optic cables supplied for use with the stand alone probe or located on the
integrated probe.
2. Clean the fiber ends using a lint free cotton swab dipped in either spectroscopic
grade isopropyl alcohol (IPA, also referred to as 2-propanol) or methanol.
Spectroscopic grade acetone can also be used, but may de-laminate the cotton from
the swab by dissolving the binding agent.
3. Remove the red protective boots from the SMA 905 connections on the transmitter
and light source. For installations with the stand alone probe, there are also red
protective boots on the two probe SMA 905 connections.
4. Connect the probe to the transmitter and light source.
a. Integrated probe.
i. The fiber bundle is connected to the light source.
ii. The single fiber is connected to the transmitter.
NOTE: Pointing the probe at the over-head light will illuminate the fiber. The
single fiber will appear as a single bright spot, whereas the bundle will clearly
show several small bright spots. The fiber bundle also appears larger than the
single fiber. CST will usually label the fiber bundle as either “Bundle” or “Source”
to facilitate proper connection.
Input ( I )
(bundle, lamp
excitation)
Port
Output ( O )
(Return, emission,
transmitter)
Port
NOTE: Secure the fiber optic cables so that they are not free to move around, be
accidentally stepped on, leaned against, or otherwise damaged during normal
operations. The allowed bend radius for fibers supplied with the 53F system is eight (8)
inches.
NOTE: NEVER use a wrench, pliers, or other tool to tighten the SMA connections. The
connectors are designed to operate after being finger tightened. Over tightening the
SMA connections will result in damage to the connector and the fiber optic cable limiting
the light transmission capabilities of the system and typically requiring replacement of
the fiber optic cables for normal analyzer operation.
2.3 Electrical
Consult Appendix 3 for electrical connectivity drawings to assist with the installation.
1. Verify electrical connection between the light source and the transmitters
a) The 53F transmitter controls the light source flash rate.
b) Verify that there is a cable connection between the transmitter and light source module.
2. Connect power and ground to the system. Make sure that this step does NOT power up
the 53F transmitter. It is recommended that an ON/OFF switch be employed.
a) For panel mounted systems CST provides a terminal strip for electrical connectivity,
either 120 VAC (for systems equipped with an AC to 24 VDC power supply) or 24 VDC
per the specifications of the customer.
b) For laboratory systems, CST supplies the system with an attached 3-pronged 120V AC
power cord. CST typically installs an ON/OFF switch on these systems; however, the
customer may request this option be eliminated.
3. Connect the 4-20 mA signal output line.
a) For panel mounted systems CST provides a terminal strip for electrical connectivity.
b) For OEM systems, connectivity must be made to terminals on the TB1 connector
(reference the appropriate schematic in Appendix 3)
4. Span Check Filter Control
a) For panel mounted systems, verify that there is a connection on the terminal strip to
either an external normally closed relay or that a jumper wire is connected between the
terminal points (reference Appendix 3 for the electrical connectivity schematic). Failure
to have the connection in place will prevent the instrument from initializing.
b) Insure that there is either a jumper or connections to a terminal strip or the TB2
connector (top of the transmitter) between the COM and SPAN N.C. positions. Failure
to have the connection in place will prevent the instrument from initializing.
5. Insure that both the 24V Power and the Loop Output are both properly grounded through
Terminal 3 of the TB1 connector. Improper grounding could result in erroneous
measurement results and/or measurement instability.
NOTE: DO NOT power up the system at this time. Please reference Section 6 for
instructions on powering the unit and initializing the system.
Operational Tip: CST recommends that data logging of the 4-20 mA output from the 53F
system be done with a minimum logging interval of 4 seconds. Logging on faster time scales is
not necessary since the update period for the transmitter is 4 seconds.
CST fluorometer the filters are located internal to the transmitter and are specific to the
application. Therefore, the CST fluorometer is a dedicated instrument for monitoring only one
specific analyte of interest. In order to monitor a new analyte of interest, the user must return
the fluorometer to CST for factory reconfiguration.
In brief, the CST fluorometer works as follows: The light source launches excitation light
through the excitation wavelength selection filter and then into a fiber optic cable. The fiber
optic cable transfers the excitation light to the CST fluorescence probe mounted in the process.
The fluorescence probe launches light into the process sample, and collects the molecular
fluorescence (emission) from the sample. The emission light is then transferred from the probe
through fiber optic cable to the analyzer. The emission light passes through an emission filter in
order to remove any residual excitation energy collected by the fluorescence probe. The
emission light then impinges a detector and the fluorescence intensity is displayed on the
analyzer.
Unlike many fluorescence units on the market, the CST fluorometer utilizes a xenon
flash lamp to provide excitation energy. This lamp allows the CST fluorometer to be easily
configured to meet any excitation wavelength requirement within the UV and visible spectral
range with the appropriate filter selection. The xenon flash lamp also has an extended lifetime
compared to other common UV sources (deuterium, mercury vapor, etc), which reduces the
cost of ownership of the analyzer.
The remainder of this manual provides the user with the necessary tools to operate the
fluorometer and meet the measurement requirements. In addition to standard operation
methods and procedures, a section detailing some application development objectives
(Appendix 1) has been provided to aid the user in defining the parameters required to realize
optimal process monitoring.
Excitation Filter: The excitation filter is used to select the range of wavelengths, commonly
referred to as the pass band, used to cause or excite the molecular fluorescence. Wavelengths
not in the pass band are rejected and ideally never reach the sample. Removal of wavelengths
outside the pass band minimizes the possibility of false fluorescence readings due to detection
of light from the lamp which mimics the fluorescence signal.
Emission Filter: The emission filter is used to select the range of wavelengths, the pass band,
to be passed to the measure detector. Wavelengths not in the pass band are rejected and
ideally never reach the sample. It is critical that the excitation wavelengths never reach the
detector, since it will respond to excitation light. Excitation light impinging the measure detector
results in increased residual background levels, which reduces the dynamic range, the signal-to-
noise ratio and the signal-to-blank ratio.
Measure Detector: The light detector is most often a photomultiplier tube, though photodiodes
are increasingly being used. The light passing through the emission filter is detected by the
photomultiplier or photodiode. The light intensity, which is proportional to the analyte
concentration, is registered as a digital readout.
Span Filter: The span filter is used to check instrument operation. The span check filter used
by CST is a neutral density filter that attenuates the fluorescence signal. The degree of
attenuation is called the attenuation factor and this value is recorded in Section 7.1.
Sensitivity: The ability of the analyzer to detect a given level of analyte based on the molecular
fluorescence from the analyte. The actual limits of detection depend on the properties of the
analyte measured and the process conditions. Parameters such as pH, temperature, oxygen
content, and background solvent, to name but a few, may dramatically alter the fluorescence
intensity measured. Typically, detection of parts-per-million (PPM) and parts-per-billion (PPB)
analyte levels can be detected. In general, fluorescent measurements are 1,000 to 500,000
times more sensitive than absorbance based photometric measurements. Practically, sensitivity
means the minimum analyte concentration that can be measured above background
fluorescence in the process.
Specificity: The ability of the analyzer to monitor one specific analyte in a mixture of
background materials without interference from the background materials. In absorbance based
photometric measurements, interference problems are common since many materials absorb
light, making it difficult to isolate the targeted analyte in a complex mixture. However,
fluorometers are highly specific and less susceptible to interferences because fewer materials
exhibit molecular fluorescence. Furthermore, if background materials do absorb and emit light,
it is rare that they will emit the same wavelength of light as the analyte of interest.
Signal-to-noise ratio: Signal refers to the emission collected by the fluorescence probe and
monitored by the analyzer. Noise refers to the output from the instrument’s electronics, which is
present whether or not sample is being read and any collection of errant wavelengths not
removed by the optical filters. Noise is measured by placing the fluorescence probe in air and in
complete darkness (no stray light). CST determines the signal-to-noise as a quality assurance
tool to insure that the analyzer system meets design criteria. For process monitoring, the
signal-to-noise ratio is not as important as the signal-over-background ratio.
Signal-over-background ratio: Signal refers to the emission collected from a sample with
known analyte concentration by the fluorescence probe and monitored by the analyzer.
Background refers to the process liquid containing no analyte of interest and any stray light
present in the system. The signal-over-background ratio should be calculated during the
application engineering phase of the project. Knowing this ratio will help determine when the
stray light level changes and/or the background material fluorescence properties change. Refer
to Appendix 1 for additional details.
Dynamic Range: Dynamic range refers to the range of concentrations an instrument can read,
from the minimum to the maximum detectable concentration. The minimum detectable
concentration is determined by signal-to-noise and signal-over-background ratios. The
maximum detectable concentration is determined by the compound’s chemistry and by factors
such as instrument sensitivity ranges, optical pathlength, specificity of optical filters, etc.
Process Background: The liquid solution used to transport or sustain the analyte of interest in
the process. This solution has all the chemical constituents found in the process except the
analyte of interest.
Stand Alone Probe: CST front surface fluorescence probe, in which the fiber optic cable
connecting the probe to the transmitter is NOT connected to the probe. i.e. connecting the probe
to the transmitter requires four connections: 1) Two (2) on the probe, 2) One (1) to the light
source, and 3) One (1) to the 53F transmitter. This probe has SMA 905 connections on the
probe marked with an “I” for input (from the light source) and “O” for output (return to the 53F
transmitter).
Integrated Probe: CST front surface probe, in which the fiber optic cables are integrated into
the probe assembly, i.e. the only connections required for this probe are at the light source and
the transmitter, not on the probe itself. Typically, the light source connection will be marked with
a label reading “Bundle” or “Source”.
Light Source: CST typically uses a Xenon flash lamp for this system, however any light source
(deuterium lamp, tungsten halogen lamp, LED, laser diode, etc.) could be used with the 53F
transmitter.
In Mode 1, the background fluid will typically exhibit a non-zero display value. This is
why the CST fluorescence analyzer is a signal-over-background measurement. In Mode 1
operation, the 4-20 mA output response is scaled from 4 mA at 000 counts and 20 mA at 1999
counts, with the 1000 count calibration point equaling 12 mA.
NOTE: If the fluorescence of the process has not been characterized, Mode 2
initialization must be performed first followed by an application engineering study, refer
to Appendix 1 for guidance. This process is an advanced use of the 53F system, and
should only be performed in consultation with CST technical staff.
simplicity CST recommends using the process background material without any analyte of
interest (commonly referred to as the zero or blank sample). Mode 2 initialization sets the
absolute zero fluorescence level to a display value of 000 counts and the upper fluorescence
limit to a display value of 1999. In general, the background fluorescence level will not equal 000
counts. The background level is a combination of fluid native fluorescence and the native
fluorescence of the optical materials in the probe assembly. Probe fluorescence is most
prominent when the excitation wavelength is below 280 nm. Section 6 details the procedure for
Mode 2 initialization.
A description of the numbered controls and indicators are listed below and broken into
three categories: 1) Active (used during normal operations), 2) Troubleshooting (only used by
CST personnel, CST certified technicians, or at the request of a CST technical staff member),
and 3) Inactive controls (these controls have no relevance or function during both normal
operations and troubleshooting).
(2)
(1) (3)
(4) (5)
(6)
(7)
(12) (13)
Monitoring States:
Measure Detector
4 RED = too much light intensity contact CST
Status Indicator
Solid Green – acceptable
Solid Yellow – acceptable
Flashing Green – acceptable
Flashing Yellow – acceptable
Indicates the amount of energy impinging the reference
(excitation) detector
Monitoring States:
Reference Detector
6 RED = too much light intensity contact CST
Status Indicator
Solid Green – acceptable
Solid Yellow – low light intensity contact CST
Flashing Green – low light intensity contact CST
Flashing Yellow – low light intensity contact CST
Displays the fluorescence level in counts (arbitrary units)
Format XXXX (for a range of 000 to 1999)
7 Digital Display
Values outside the format range indicates an operational
failure (contact CST)
Span Check Filter Press, hold 1-2 seconds, and then release to initiate span
12
Motion Control check filter motion, either in or out.
This section is broken up into two parts: 1) The Quick Start Operations Guide and 2)
The Detailed Operations Guide. The Quick Start section is sufficient for all applications where
Mode 1 or calibrated operation is desired (90% of all applications). For users performing
application development work, it may be necessary to operate in both Mode 1 and Mode 2, so
detailed procedures are supplied for working in both Modes and switching between Modes.
Incorrectly switching between operational Modes can lead to significant operational difficulties
and instrument response instability.
Operation Note: The digital display and 4-20 mA output are correlated to fluorescence
intensity in counts (arbitrary units). To convert fluorescence intensity to relevant engineering
units the 4-20mA output must be scaled using an external device. The transmitter display
cannot be set to relevant engineering units, and has a fixed range of 000 to 1999 counts.
DO NOT power-up the unit until directed to do so in this procedure. Powering the
unit incorrectly will result in unstable measurements and errant readings.
CST shipped the unit in Mode 1 or the calibrated mode, after completion of QA/QC
testing.
TB 2 Connector TB 1 Connector
1) Insure that there is either a jumper or connections to a terminal strip on the TB2 connector
(top of the transmitter) between the COM and SPAN N.C. positions. Failure to have the
connection in place will prevent the instrument from initializing.
2) Insure that both the 24V Power and the Loop Output are both properly grounded through
Terminal 3 of the TB1 connector. Improper grounding could result in erroneous
measurement results and/or measurement instability.
3) Connect the probe to the transmitter and light source.
a) Integrated probe.
i) The fiber bundle is connected to the light source.
ii) The single fiber is connected to the transmitter.
NOTE: Pointing the probe at the over-head light will illuminate the fiber. The single fiber
will appear as a single bright spot, whereas the bundle will clearly show several small
bright spots. The fiber bundle also appears larger than the single fiber.
NOTE: Secure the fiber optic cables so that they are not free to move around, be
accidentally stepped on, leaned against, or otherwise damaged during normal
operations. The allowed bend radius for fibers supplied with the 53F system is eight (8)
inches.
4) Place the probe in a dark environment, containing a non-fluorescing sample (air, water,
etc.). Examples of appropriate environments are:
a) The calibration bottle (with the probe approximately ¼ - ½” from the bottom of the bottle),
or
b) A dedicated sampling point (slip stream, side vessel, etc.), or
c) The process pipe, vessel, etc.
5) Power up the analyzer and wait a minimum of 15 minutes before proceeding with the
procedure.
NOTE: The Measure Detector Status LED has four (4) allowed states for proper
fluorescence measurement. Flashing Yellow, Flashing Green, Solid Yellow and Solid
Green. If any of these four states occur, the unit should be functioning normally.
NOTE: NEVER allow the probe tip to be exposed to ambient light for longer than 1
minute. If exposure occurs, place the probe back in a dark environment for 2 hours and
suspend all work until the time has passed. Excessive light will result in RED indicator
LEDs on both the Measure and Reference Detector Status.
7) Add calibration fluid to the probe insertion region. Wait 2 minutes, the following should be
observed:
a) The Span Filter Indicator should be green (filter is out)
b) The Detector Status LEDs (located to the left of the display) should be:
i) Measure: Solid Yellow or Solid Green (typically)
ii) Reference: Solid Green
c) Assuming that the CST QA/QC calibration fluid is a relatively close match to your
calibration fluid, the display should be reading between 900 and 1100 counts (typically)
with a stability of +/- 20 counts. If you did not specify a calibration fluid, then agreement
between the factory calibration and your calibration sample is unknown. Consult the
CST QA/QC documentation for a description of the calibration solution and the actual
test results (designated as Mode 1).
d) If either Detector Status LED is RED, contact CST immediately for assistance.
NOTE: If the display does not fall within the range from 980 to 1020 counts repeat the
calibration step above. Contact CST for technical assistance if three (3) attempts at
calibration are not successful.
9) You are now ready to proceed with either lab work or process monitoring.
Operation Tip 1: The 53F system is designed for continuous 24/7 operation, so leave the
system running whenever possible. Cycling of power frequently will require calibration steps on
a shorter time interval.
Operational Tip 2: The calibration bottle supplied by CST is a common Nalgene™ bottle
supplied by almost every laboratory supply company. If you are using the bottle for calibration
and system validation, purchase a set of bottles from Fisher to insure that cross contamination
between samples does not occur.
Operational Tip 3: The span check filter can be used to provide additional points during
instrument testing and verification. The filter utilized is a neutral density filter and attenuates the
fluorescence response from any sample. Consult the CST QA/QC documentation for the exact
attenuation level in your instrument.
This procedure assumes that the hardware has been installed correctly and is
ready for operation. For OEM components, wiring and mounting will be required before
these procedures are applicable.
CST shipped the unit in Mode 1, or the calibrated mode, after completion of
QA/QC testing. Whenever possible, please leave the unit in Mode 1 when powering the
unit off. Mode 1 operation is the standard or normal mode of operation.
TB 2 Connector TB 1 Connector
1) Insure that there is either a jumper or connections to a terminal strip on the TB2 connector
(top of the transmitter) between the COM and SPAN N.C. positions. Failure to have the
connection in place will prevent the instrument from initializing.
2) Insure that both the 24V Power and the Loop Output are both properly grounded through
Terminal 3 of the TB1 connector. Improper grounding could result in erroneous
measurement results and/or measurement instability.
3) Verify the connection between the probe, transmitter, and light source.
a) Integrated probe.
i) The fiber bundle is connected to the light source.
ii) The single fiber is connected to the transmitter.
NOTE: Pointing the probe at the over-head light will illuminate the fiber. The single fiber
will appear as a single bright spot, whereas the bundle will clearly show several small
bright spots. The fiber bundle also appears larger than the single fiber.
i) Connect one fiber optic cable between the light source and the probe. The probe
connection is marked with an “I”.
ii) Connect one fiber optic cable between the transmitter and the probe. The probe
connection is marked with an “O”.
NOTE: Secure the fiber optic cables so that they are not free to move around, be
accidentally stepped on, leaned against, or otherwise damaged during normal
operations. The allowed bend radius for fibers supplied with the 53F system is eight (8)
inches.
4) Place the probe in a dark environment, containing a non-fluorescing sample (air, water,
etc.). Examples of appropriate environments are:
a) The calibration bottle (with the probe approximately ¼ - ½” from the bottom of the bottle),
or
b) A dedicated sampling point (slip stream, side vessel, etc.), or
c) The process pipe, vessel, etc.
5) Power up the analyzer and wait a minimum of 15 minutes before proceeding with the
procedure.
NOTE: CST recommends at least one (1) hour of warm-up when using the analyzer
system after an extended (>1 hour) power down period.
NOTE: The Measure Detector Status LED has four (4) allowed states for proper
fluorescence measurement. Flashing Yellow, Flashing Green, Solid Yellow and Solid
Green. If any of these four states occur, the unit should be functioning normally.
NOTE: NEVER allow the probe tip to be exposed to ambient light for longer than 1
minute. If exposure occurs, place the probe back in a dark environment for 2 hours and
suspend all work until the time has passed. Excessive light will result in RED indicator
LEDs on both the Measure and Reference Detector Status.
7) Add calibration fluid to the probe insertion region. Wait 2 minutes, the following should be
observed:
a) The Span Filter Indicator should be green (filter is out)
b) The Detector Status LEDs (located to the left of the display) should be:
i) Measure: Solid Yellow or Solid Green (typically)
ii) Reference: Solid Green
c) Assuming that the CST QA/QC calibration fluid is a relatively close match to your
calibration fluid, the display should be reading between 900 and 1100 counts (typically)
with a stability of +/- 20 counts. If you did not specify a calibration fluid, then agreement
between the factory calibration and your calibration sample is unknown. Consult the
CST QA/QC documentation for a description of the calibration solution and the actual
test results (designated as Mode 1).
d) If either Detector Status LED is RED, contact CST immediately for assistance.
NOTE: The Measure Detector Status LED has four (4) allowed states for proper
fluorescence measurement. Flashing Yellow, Flashing Green, Solid Yellow and Solid
Green. If any of these four states occur, the unit should be functioning normally.
NOTE: NEVER allow the probe tip to be exposed to ambient light for longer than 1
minute. If exposure occurs, place the probe back in a dark environment for 2 hours and
suspend all work until the time has passed. Excessive light will result in RED indicator
LEDs on both the Measure and Reference Detector Status.
NOTE: If the display does not fall within the range from 980 to 1020 counts repeat the
calibration step above. Contact CST for technical assistance if three (3) attempts at
calibration are not successful.
9) Checking unit operation with the span check neutral density filter.
a) Perform this step with the calibration fluid in place
b) Press and hold the SPAN CHK switch (top of the transmitter) for 1-2 seconds to actuate
the span check assembly. This will move the span check filter into the optical path and
attenuate the fluorescence intensity.
c) The Span Filter Indicator should be red (filter is in).
d) Wait 2 minutes for the signal level to stabilize.
e) The degree of signal attenuation is given in Section 7: Instrument Specifications -
Optical and is listed as the Attenuation Factor (a decimal). The expected signal level is
calculated by taking the Attenuation Factor * Display (or 4-20) value. For example, when
the attenuation factor is 0.407 and the displayed fluorescence level is 997, the expected
display value after inserting the span check filter would be 406 (always round up to the
next whole digit).
f) Press and hold the SPAN CHK switch (top of the transmitter) for 1-2 seconds to actuate
the span check assembly. This will move the span check filter out of the optical path.
g) The Span Filter Indicator should be green (filter is out).
h) Wait 2 minutes for the signal level to stabilize.
NOTE: During span check filter actuation, the displayed values will range from very low
to very high as the assembly moves through the optical path. This is normal operation
for the unit when the span check filter is used.
10) You are now ready to proceed with either lab work, generation of a complete calibration
curve, or process monitoring.
1) Verify that the probe is mounted in a dark environment (the CST calibration bottle, the
process monitoring point, etc.) where ambient light cannot reach the probe tip.
2) Verify that the probe tip is located in a process/test sample. The fluorescence
characteristics of this sample maybe unknown, but CST recommends that the sample be
non-fluorescing. Examples of non-fluorescing samples are:
a) Air,
b) Distilled water, or
c) Non-fluorescing solvents (methanol, cyclohexane, iso-octane, etc.).
ii) The Detector Status LEDs (located to the left of the display) should be:
(1) Measure: Solid Yellow or Solid Green (typically)
(2) Reference: Solid Green
iii) The display should read 1000 +/- 20 counts and be stable within this allowed range.
iv) If either Detector Status LED is RED, contact CST immediately for assistance.
NOTE: If the display does not fall within the range from 980 to 1020 counts repeat the
calibration step above. Contact CST for technical assistance if three (3) attempts at
calibration are not successful.
5) Checking unit operation with the span check neutral density filter.
a) Perform this step with the calibration fluid in place
b) Press and hold the SPAN CHK switch (top of the transmitter) for 1-2 seconds to actuate
the span check assembly. This will move the span check filter into the optical path and
attenuate the fluorescence intensity.
c) The Span Filter Indicator should be red (filter is in).
d) Wait 2 minutes for the signal level to stabilize.
e) The degree of signal attenuation is given in Section 7: Instrument Specifications -
Optical and is listed as the Attenuation Factor (a decimal). The expected signal level is
calculated by taking the Attenuation Factor * Display (or 4-20) value. For example, when
the attenuation factor is 0.407 and the displayed fluorescence level is 997, the expected
display value after inserting the span check filter would be 406 (always round up to the
next whole digit).
f) Press and hold the SPAN CHK switch (top of the transmitter) for 1-2 seconds to actuate
the span check assembly. This will move the span check filter out of the optical path.
g) The Span Filter Indicator should be green (filter is out).
h) Wait 2 minutes for the signal level to stabilize.
NOTE: During span check filter actuation, the displayed values will range from very low
to very high as the assembly moves through the optical path. This is normal operation
for the unit when the span check filter is used.
6) Generate a minimum three (3) point calibration curve for the analyzer. CST recommends a
five (5) point curve. This calibration will allow the operator to generate a predictive equation
that converts the display/4-20 fluorescence intensity (in counts) to the relevant
engineering/application unit. Typically, this will be the fluorophore concentration. The
procedure below can be used as a guide for generating the calibration curve.
a) CST recommends the following four solutions plus the calibration solution for generating
the curve
i) Zero fluorescence – typically the background material
ii) Tenth-span – solution has one-tenth the fluorophore concentration of the calibration
solution
iii) Half-span – solution has one-half the fluorophore concentration of the calibration
solution
iv) One and a Quarter Span – solution has 1.25 times the fluorophore concentration of
the calibration solution.
b) Prepare calibration curve solutions in advance in quantities appropriate for procedure
(this will depend on the size of volume of the probe insertion point).
c) The unit is calibrated and operating in Mode 1.
d) Admit the calibration curve solutions individually and record the fluorescence response
for each sample. Replicate this sequence three times.
e) Take the response data, plot, and then fit the data using either a linear calibration or
second order polynomial equation. CST typically uses the second order polynomial fit,
which provides a better predictor of the fluorescence response for fluorophore
concentrations higher than that of the calibration solution.
f) The analyzer is now fully calibrated and ready for use.
NOTE: The span check filter can be used to provide additional points during instrument
testing and verification. The filter utilized is a neutral density filter and attenuates the
fluorescence response from any sample. Consult the CST QA/QC documentation for the
exact attenuation level in your instrument.
7) You are now ready to proceed with either lab work or process monitoring.
1) Verify that the analyzer is operating in Mode 1. If necessary follow the Mode 1 calibration
procedure given in Section 6.2.3.
2) Make sure that the probe tip is immersed in a sample. The fluorescence level of the test
sample need not be known with regard to the Mode 1 calibration curve. Wait 2 minutes, the
following should be observed:
a) The Span Filter Indicator should be green (filter is out)
b) The Detector Status LEDs (located to the left of the display) should be:
i) Measure: Flashing Yellow, Flashing Green, Solid Yellow or Solid Green (dependent
on the fluorophore concentration)
ii) Reference: Solid Green
c) If either Detector Status LED is RED, contact CST immediately for assistance.
3) Initialize the analyzer to the Fixed Range (Mode 2 Operation)
a) Insert the Adjustment/Initialization tool into the hole marked INT PRESET and depress
the button for at least 8 seconds.
b) Wait 2 minutes, the following should be observed:
i) The Span Filter Indicator should be green (filter is out)
ii) The Detector Status LEDs (located to the left of the display) should be:
(1) Measure: Flashing Yellow, Flashing Green, Solid Yellow or Solid Green
(dependent on the fluorophore concentration)
(2) Reference: Solid Green
(3) If either Detector Status LED is RED, contact CST immediately for assistance.
c) The display should read a value between 000 and 1999 counts. If a 1, -1, or -500 is
shown, the unit is off-scale and only Mode 1 may be used to evaluate the fluorophore of
interest in the concentration range desired.
NOTE: The span check filter can be used to provide additional points during instrument
testing and verification. The filter utilized is a neutral density filter and attenuates the
fluorescence response from any sample. Consult the CST QA/QC documentation for the
exact attenuation level in your instrument.
4) Checking unit operation with the span check neutral density filter.
a) Perform this step with the calibration fluid in place
b) Press and hold the SPAN CHK switch (top of the transmitter) for 1-2 seconds to actuate
the span check assembly. This will move the span check filter into the optical path and
attenuate the fluorescence intensity.
c) The Span Filter Indicator should be red (filter is in).
d) Wait 2 minutes for the signal level to stabilize.
e) The degree of signal attenuation is given in Section 7: Instrument Specifications -
Optical and is listed as the Attenuation Factor (a decimal). The expected signal level is
calculated by taking the Attenuation Factor * Display (or 4-20) value. For example, when
the attenuation factor is 0.407 and the displayed fluorescence level is 997, the expected
display value after inserting the span check filter would be 406 (always round up to the
next whole digit).
f) Press and hold the SPAN CHK switch (top of the transmitter) for 1-2 seconds to actuate
the span check assembly. This will move the span check filter out of the optical path.
g) The Span Filter Indicator should be green (filter is out).
h) Wait 2 minutes for the signal level to stabilize.
NOTE: During span check filter actuation, the displayed values will range from very low
to very high as the assembly moves through the optical path. This is normal operation
for the unit when the span check filter is used.
Operation Tip 1: The 53F system is designed for continuous 24/7 operation, so leave the
system running whenever possible. Cycling of power frequently will require calibration steps on
a shorter time interval.
Operational Tip 2: The calibration bottle supplied by CST is a common Nalgene™ bottle
supplied by almost every laboratory supply company. If you are using the bottle for calibration
and system validation, purchase a set of bottles from Fisher to insure that cross contamination
between samples does not occur.
Operational Tip 3: The span check filter can be used to provide additional points during
instrument testing and verification. The filter utilized is a neutral density filter and attenuates the
fluorescence response from any sample. Consult the CST QA/QC documentation for the exact
attenuation level in your instrument.
Operational Tip 4: The First Power-Up Following Installation sequence should be completed
anytime the unit has been powered down for an extended time period (> 7 days).
Operational Tip 5: When powering the unit down, always leave the analyzer in Mode 1.
Restarting the analyzer and verifying operation is much easier with Mode 1.
Operational Tip 6: The 53F analyzer will restart after power down or power failures using the
operating parameters stored prior to the power loss. If power failure occurred in Mode 2, it is
likely that upon the restart analyzer operation will be erratic and unstable. Simply follow the
Mode 2 operation setup (Section 6.2.3) before continuing work.
Operational Tip 7: CST recommends that any sample to be used for Mode 1 calibration have
at least 100 counts of fluorescence intensity in Mode 2.
Monitoring States:
Measure Detector RED = too much light contact CST
Status Indicator Solid Green – acceptable
Solid Yellow – acceptable
Flashing Green – acceptable
Flashing Yellow – acceptable
Indicates the amount of energy impinging the reference
(excitation) detector
Monitoring States:
Reference Detector
RED = too much light intensity contact CST
Status Indicator
Solid Green – acceptable
Solid Yellow – low light intensity contact CST
Flashing Green – low light intensity contact CST
Flashing Yellow – low light intensity contact CST
Normal Range: 000 to 1999
TB 2 Connector TB 1 Connector
Solution: Power unit off and call CST for a Return Material Authorization (RMA) number.
This unit has too much light impinging the detector or detectors and needs to be
evaluated at the factory.
=====================================================================
Problem: The Span Filter Indicator is neither RED nor GREEN, the unit will not initialize
and there is a humming/buzzing sound.
Cause: There is not a jumper or connection between the COM and SPAN N.C. positions
on the TB2 connector or the common terminal. Therefore, the span assembly is
continually operating and the instrument will not initialize. Failure to have the
connection in place will prevent the instrument from initializing.
Solution: Make sure there is a jumper between the COM and SPAN N.C. positions on the
TB2 connector or terminal strip. For panelized systems with a terminal strip,
there must be a jumper or connectivity to an external/remote Form B contact
(normally closed) switch or relay.
=====================================================================
Problem: The Span Filter Indicator is neither RED nor GREEN and the unit will not
initialize. No sound is heard other than flashing of the lamp.
Cause: The transmitter optical assembly has been shoved up into the transmitter and the
span assembly cannot actuate.
Solution: Call CST for an RMA number. The unit must be repaired by CST.
=====================================================================
Problem: There is a solid yellow indicator on the REF Status LED
Cause: The light passing to the reference PMT does match the factory levels.
Solution: Step 1 - Make sure that the fiber connections to the light source and transmitter
are tight (finger tight). If no change, then:
Step 2 – (Integrated Probe) Make sure that the fiber bundle is connected to the
light source and that the single fiber is connected to the transmitter. (Stand Alone
Probe) Make sure that the port marked “I” on the probe has a fiber running to the
light source, and the port marked “O” has a fiber run to the transmitter. If no
change, then:
Step 3 – Power off the analyzer. Remove the probe from the insertion location
and disconnect the integrated probe (or the stand alone probe run fibers) from
the light source and transmitter.
Step 4 – (Integrated Probe) Hold a flashlight at the end of the probe, and
examine the light transmitted through the fibers to the end of the assembly.
There should be a very bright light emitting from the SMA connections. If the
output is not bright, then call CST for an RMA number, the probe must be
returned for evaluation. (Stand Alone Probe) Disconnect the run fibers from the
probe. Hold a flashlight at the end of the probe, and examine the light
transmitted through the SMA ports at the top of the probe. If the output is not
bright, then call CST for an RMA number, the probe must be returned for
evaluation. Next, place the flashlight at one end of a fiber optic cable and
examine the light transmitted through the fiber. Repeat procedure for the other
fiber. If either fiber has low transmission, contact CST for the purchase of a new
fiber.
Step 5 – if the probe (or probe + run fibers) all have acceptable transmission,
then reconnect to the analyzer and power the system. Wait 15 minutes. If there
is still no change to the LED color, contact CST for an RMA.
=====================================================================
Cut-on Wavelength =
Tmax =
Detectors: PMT
Silicon Photodiode
InGaS Photodiode
Mechanical
Width: 4.4” (112 mm)
Height: 9.5” (242 mm)
Depth: 3.5” (89 mm)
Weight: <3 lbs (1.4 kg)
Display 3½ Digits
Physical Dimensions
Probe Insertion Length:
Probe Overall Length:
Run Length (Fiber Portion):
Total Assembly Length:
Probe Diameter: 12.0 mm (0.472”)
12.7 mm (0.500”)
Physical Dimensions
Probe Insertion Length:
Probe Overall Length:
Run Fiber Length:
Run Fiber Core Diameter: 600 micrometer
800 micrometer
Probe Diameter: 12.0 mm (0.472”)
12.7 mm (0.500”)
Physical Dimensions:
Length (defined as the optical axis): 5.375” (137 mm)
Width (defined as the flow axis): 4.0” (102 mm)
Height (fluorescence axis): 3.625” (92 mm)
Enclosure Type:
Enclosure Material:
Area Classification:
Please reference the Purge Kit Manual and Specifications documentation supplied with
the analyzer system.
This section is provided to enable operators to use the system in the laboratory or
development environment to determine the best practices for using the analyzer system. CST
terms this activity as application development, and offers this service to customers on a contract
basis. However, CST will consult with customers on their internal application development
activities. Please contact either Mike Ponstingl ([email protected]) or Jess V Ford
([email protected]) for assistance. The phone number is 636-305-0666 and the fax
number is 636-305-1096.
The system was supplied with an optical configuration specific to a given fluorescence
excitation and emission pair, which is defined by the fluorophore of interest. However, the
configuration may be applicable to additional fluorophores as well.. Therefore, the information
provided below is only valid when the fluorophore of interest has excitation and emission
characteristics that correspond to the installed optical filters.
not relative to an arbitrary calibration level as in Mode 1. Finally, Series 3 is not useful in Mode
2, unless multiple fluorophore comparison is necessary.
It is possible to determine a range factor that will take the values recorded in Mode 2 and
convert them into approximate Mode 1 values, relative to a specific calibration sample (i.e. you
must choose one sample as the “calibration” sample). Outlined below are the steps involved:
1000
Range Multiplier =
Mode 2 Measured Value
Step 3: Convert the Mode 2 sample value into an approximate Mode 1 value by
CST recommends that only samples with at least 100 counts of fluorescence in Mode 2
be used as calibration fluids in Mode 1. If this guideline is not followed, measurement instability
may result.
⎛ x−4⎞
⎜ ⎟ *1999 = Counts
⎝ 16 ⎠
⎛ x−4⎞
⎜ ⎟ * y = Engineering Value
⎝ 16 ⎠
y
Attenuating Factor =
x
Operational Mode:
See attached.
See attached.