TI500 Globe-Trotter Neonatal Transport System: Instructions For Use
TI500 Globe-Trotter Neonatal Transport System: Instructions For Use
WARNING
To properly use this medical device, read and
                                                Model GT500
comply with these instructions for use.
This page intentionally left blank.
Trademarks
  The list of countries in which Dräger trademarks are registered can be found on the Dräger website. See
  www.draeger.com/trademarks.
This page intentionally left blank.
Table of Contents
Section 1: Symbol Definition and Intended Use
   Symbol Definition . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
       Typefaces and Icons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
       Technical Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
   Intended Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
   Indications for Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
Section 2: Introduction, Features, and Specifications
   Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
       Pneumatic System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
           Pneumatic Module with Air Cylinder - Gas Flow Sequence. . . . . . . . . . . . . . . . . . . . . . . . . 2-3
           Pneumatic Module with Compressor - Gas Flow Sequence . . . . . . . . . . . . . . . . . . . . . . . . . 2-5
       Transport Incubator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-7
       Modular Frame . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-7
       Variable Height Adjustable (VHA) Cart . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-7
   Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-8
       Standard Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-8
       Optional Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-8
       Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-8
   Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-9
       Standard Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-9
       Cart Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-10
       Optional Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-10
       Device Classification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-10
Section 3: Precautions and Safety Tips
   General Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
   Electrical Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
   Explosion Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
   Oxygen Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
   Electromagnetic Compatibility Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
   System Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
   Ventilator Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
   Pneumatic Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8
Section 4: Installation and Functional Checkout
   Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
       Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
           Scissor mount dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
                                                                                                                                                              i
       Cylinder Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
           Pin-Index Style Cylinders . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
   Functional Checkout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
       Electrical Checkout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
       Mechanical Checkout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
       Pneumatic Checkout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-6
       System Checkout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-9
Section 5: Instructions for Use
   Controls, Indicators and Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
   Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
       Blender Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
       Compressor Low Battery Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
   Instructions for Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
       General Tips and Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
       System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
           To Use an External AC Power Source. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-5
           To Use the Auxiliary Power Strip . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-6
           To Prepare the System for Transport . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-6
       Transport incubator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-6
       Pneumatic Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-7
           To Activate the Pneumatic Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-8
           To Operate the System when Air Compressor Battery is Low . . . . . . . . . . . . . . . . . . . . . . . 5-9
           To Assess Each Gas Cylinder . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-9
           To Monitor Gas Reserve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-9
           To Administer Oxygen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-10
           To Charge the Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-11
           To Check the Battery Condition After Charging the System . . . . . . . . . . . . . . . . . . . . . . . 5-11
           To Use the Mechanical Ventilation Circuits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-12
           To Use the Oxygen Analyzer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-13
       Infusion Devices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-14
       Suction System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-14
       Patient Monitor Securing Site . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-14
       Variable Height Adjustable Cart . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-15
                To Remove the Frame from the VHA Cart . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-15
                To Return the Frame to its Original Position on the VHA Cart . . . . . . . . . . . . . . . . . . . . . 5-15
ii
Section 6: Cleaning, Maintenance, and Replacement Parts
   Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
       General Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
       Steam Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
       Cleaning Difficult to Access Areas . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2
       Disinfecting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2
       Cleaning the System Frame and the Pneumatic Module Surfaces . . . . . . . . . . . . . . . . . . . . . . . 6-2
       Cleaning the Tie-Down Straps Used to Secure a Monitor or the Gas Hoses and
       Power Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2
       Cleaning and Sterilizing the Mechanical Ventilation Circuit . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2
       Disassembly for Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-3
       Assembly After Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-3
       Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-3
   Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-4
       Frame and Electrical Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-4
       Pneumatic Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-4
   Replacement Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-5
Section 7: Troubleshooting
   Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 1
                                                                                                                                                              iii
     This page intentionally left blank.
iv
                                                              Section 1
                                                      Symbol Definition
                                                      and Intended Use
Symbol Definition
  Below is a description of the typefaces, symbols and technical terms used in this manual.
     –   A WARNING identifies situations or actions that may affect patient or user safety. Disregarding a
         warning could result in patient or user injury.
     –   A CAUTION points out special procedures or precautions that personnel must follow to avoid
         equipment damage.
  • The symbol below indicates “Attention: Consult accompanying documents:”
                           Attention: Consult Accompanying Documents
     –   The instrument provides a specified degree of protection against electric shock, particularly the
         leakage current and reliability of the protective ground connection with a BF-type applied part.
                                                                                                        1-1
         –   A BF-type applied part indicates an applied part isolated from all other parts of the instrument to
             such a degree that the patient leakage current allowable in a single-fault condition is not
             exceeded.
      • The symbol below indicates “Caution: Electric shock hazard.”
                                        Caution: Electric Shock Hazard
      • The symbol below indicates the On and Off position of a power switch:
                                                    On and Off
• This symbol indicates: CAUTION: for incubator stability, shelf loading must not exceed 15 lb (6.8 kg).
• This symbol indicates that if a single O2 cylinder is used, place it in the front gas cylinder support.
1-2
  • The symbol below indicates “Phthalate present”:
                                           Phthalate Present
Technical Definitions
  • Air or O2 cylinder—A gas cylinder holding compressed gas, most frequently constructed of
    aluminum or steel. The outlet has a pin and hole or thread pattern that is unique to the cylinder
    contents.
  • Alternating Current (AC) power—Hospital, line, or wall electrical power that characteristically
    reverses its polarity at regular, recurring intervals.
  • Canadian Standards Association (CSA)—An organization that sets Canadian equipment standards.
  • Diameter Indexed Safety System (DISS)—A system of threaded pneumatic, fittings that are each
    unique to a given gas.
  • Direct Current (DC) power—Internal or battery electrical power that is unidirectional.
  • Fraction of Inspired Oxygen (FiO2) concentration—The value often used to specify a physician-
    ordered concentration of oxygen for delivery.
  • Gas blender—A device used to mix different gases to a desired proportion of each gas. The blender
    described in this manual mixes a combination of compressed breathing air (21% oxygen) and pure
    oxygen (100% oxygen).
  • Hand ventilator—A hand-operated device used for artificial ventilation.
  • International Electrotechnical Commission (IEC)—An organization that sets international
    equipment standards.
  • Mechanical ventilator—A mechanically operated device used for artificial ventilation.
  • Non-Interchangeable Screw Thread (NIST)—A method that specifies uniquely threaded pneumatic
    fittings for gases.
  • Pin Index Safety System—A system of matching high-pressure gas cylinders to their contents by
    following a uniquely patterned pin and hole key pattern. For example, the combined sites of 1 and 5 are
    for air, and the combined sites 2 and 5 are for oxygen gas.
  • Pneumatic—Operation of a device by compressed gas.
                                                                                                        1-3
Intended Use
      United States federal law restricts this device to sale by or on an order of a physician.
      The TI500 Globe-Trotter Neonatal Transport System is intended for the transport of high risk premature,
      low birth weight, or critically ill newborns between healthcare facilities. The means of transportation can
      be either ground or airborne (rotary or fixed wing). The system is also suitable for either transport or
      stationary care within the hospital. It includes up to three system components. These components include
      a frame which houses the incubator and the pneumatic module assembly, and a facility for gas cylinder
      storage. The system is arranged to serve as a complete unit for the care of an infant in need of
      transportation.
1-4
                                                      Section 2
                                        Introduction, Features,
                                             and Specifications
Introduction
  SHOCK HAZARD:
  Where the integrity of the external protective earth conductor arrangement is in doubt, operate
  the equipment from its internal electrical power source. Failure to do so could result in personal
  injury or equipment damage.
  The TI500 Globe-Trotter Neonatal Transport System is a low profile neonatal transport system. The basic
  configuration is comprised of:
  • Pneumatic module
  • Transport incubator (Model TI500)
  • Modular frame
  • Suction pump
  • Various Variable Height Adjustable (VHA) carts
  The incubator and pneumatic module are assembled on the modular frame, which mounts on a VHA cart.
  The VHA cart only provides portability and variable height for the transport system; therefore the
  modular frame can be lifted off the cart. All the transport components remain operational.
  The incubator is located on the left side of the frame, and the pneumatic module is located on the right
  side. The accessory shelf attaches to the rear of the frame and is located above the incubator. The
  ratcheting strap secures the optional monitor to the shelf.
  The suction pump is mounted on a holder and is secured to the rear of the pneumatic module. The
  disposable suction bottle is secured to the rear of the pump.
  An IV pump mounting frame is located at the rear of the pneumatic module. It is used to secure infusion
  pumps with pole-style clamps.
  The system accepts the UK or USA style, E- or D-size, aluminum or steel gas cylinders.
Pneumatic System
  The pneumatic module is an integrated system that takes air and oxygen from gas cylinders or wall
  sources. It provides ventilation and controls the flow of gas. The pneumatic module is available in two
  versions, a compressor version and an Air cylinder version. In the compressor version, the air is cooled,
  dried, and regulated as it exits the compressor. In both versions, 2-stage regulation reduces the cylinder
  gas pressure. In both versions, the user selects the source (wall or gas cylinder) for each gas using check
  valves. The module uses the source with the highest pressure. The blender mixes air and oxygen. It
  supplies blended gas to the flow-control valve and to the patient gas inlet of the ventilator. In the Air
                                                                                                          2-1
      cylinder version, blended gas is also used to power the ventilator. In the compressor version, pure oxygen
      is used to power the ventilator.
      Test ports are integrated in the system to verify the air and O2 cylinder pressures and the external gas
      input and output pressures. Check valves allow gas sources to be changed without loss of operating
      pressure. Additional check valves ensure that gas follows a forward path to the blender in the pneumatic
      module.
      The pneumatic module also allows analysis of O2 using an oxygen analyzer.
      The pneumatic module with Air cylinder includes two gas cylinder pressure gauges (for O2 and air). It
      supports use of one Air cylinder and one O2 cylinder. The pneumatic module with compressor includes a
      12 V air compressor with a battery to run the compressor and a power supply to charge the battery. It
      includes a gas cylinder pressure gauge for O2 and an illuminated On/Off switch for the compressor. It
      also supports use of two O2 cylinders.
      Both versions can also be used with external 345 kPa (50 psi) ± 34 kPa (5 psi) sources of O2 or air. They
      can be used with a combination of on-board, internal and external sources. Gauges are mounted on the
      front of the pneumatic module and gas fittings on the rear. Fittings conform to NIST or DISS standards.
2-2
Pneumatic Module with Air Cylinder—Gas Flow Sequence
                                                       2-3
                                   Reference                       Description
                               A                   Air cylinder
                               B                   Air cylinder pressure
                               C                   Regulator, high pressure
                               D                   Air gauge
                               E                   Test port
                               F                   Regulator, internal with gauge
                               G                   Gas blender
                               H                   Air (DISS or NIST) inlet
                               I                   O2 cylinder
                               J                   O2 cylinder pressure
                               K                   Oxygen gauge
                               L                   Oxygen (DISS or NIST) inlet
                               M                   Blender pressure
                               N                   Flow control valve
                               O                   Ventilator gas (on/off)
                               P                   Mechanical ventilator
                               Q                   Ventilator pop-off
                               R                   Gauge port tube
                               S                   Exhalation valve tube
                               T                   Gauge port
                               U                   Exhalation valve
                               V                   Patient gas
                               W                   Oxygen sensor tee
                               X                   Oxygen analyzer
      Oxygen Flow Sequence
      Oxygen is fed through a check valve from the front gas cylinder, into the pneumatic module, and to an
      adjustable 2-stage regulator. The O2 inlet of the blender is fed from the output of the O2 regulator or from
      externally supplied O2 at the O2 fittings on the rear of the pneumatic module.
      Air Flow Sequence
      Air is fed through a check valve from the rear gas cylinder, into the pneumatic module, and to an
      adjustable 2-stage air regulator. The air inlet of the blender is fed from the output of the air regulator or
      from externally supplied air at the air fittings on the rear of the pneumatic module.
      Blended Gas Sequence
      For hand ventilation, the blender output is supplied to an external calibrated flow control valve that is
      gravity-independent. For mechanical ventilation, a 2-way valve serves as the On/Off control for the
      mechanical ventilator. Low pressure blended gas powers the mechanical ventilator, and uses that same
      gas for patient gas. The output of the ventilator feeds the patient gas circuit. The patient circuit includes
      an adjustable pop-off valve and an O2 sensor. The O2 sensor feeds O2 percentage information to the
      oxygen analyzer.
2-4
Pneumatic Module with Compressor—Gas Flow Sequence
                                                     2-5
                                   Reference                       Description
                              A                   Inlet filter
                              B                   Heat exchanger
                              C                   Air compressor
                              D                   Air Compressor Pressure
                              E                   Accumulator tank (60 psi +/- 2 psi)
                              F                   Pressure switch
                              G                   Test port
                              H                   Filter with auto-relief water drain
                              I                   Regulator, high pressure
                              J                   Air gauge (internal)
                              K                   Regulator, internal with gauge
                              L                   Air (DISS or NIST) inlet
                              M                   O2 cylinder
                              N                   O2 cylinder pressure
                              O                   O2 cylinder pressure gauge
                              Q                   Oxygen (DISS or NIST) inlet
                              R                   Gas blender
                              S                   Blended gas (50 psi)
                              T                   Flow control valve
                              U                   Valve for ventilator logic gas (on/off) & patient
                                                  gas
                              V                   Ventilator logic gas (on/off)
                              W                   Mechanical ventilator
                              X                   Ventilator pop-off
                              Y                   Gauge port
                              Z                   Exhalation valve tube
                              AA                  Gauge port
                              BB                  Exhalation valve
                              CC                  Patient gas
                              DD                  Oxygen sensor tee
                              EE                  Oxygen analyzer
      Oxygen Flow Sequence
      Oxygen is fed through check valves from either front or rear gas cylinders, into the pneumatic module,
      and to an adjustable primary regulator. The secondary O2 regulator is fed from the output of the primary
      regulator or by externally supplied gas at the rear gas fittings. The same circuit also feeds the logic gas to
      the mechanical ventilator. The secondary O2 regulator is adjustable, has a built-in gauge, and includes a
      5-micron inlet filter.
      Air Flow Sequence
      Room air is drawn into the compressor through the inlet filter. The compressor produces low-pressure air.
      The air, cooled by a fan-equipped radiator, is then stored in an accumulator tank. The On/Off switch, and
      a pressure switch that senses the accumulator pressure, control the compressor. An adjustable air
      regulator with a built-in gauge feeds compressed air to the blender. The air regulator includes a 5-micron
      port filter, a water trap, and an automatic drain valve with a discharge tube. The blender can also receive
      air through the external fitting. In the blender, it is combined with O2 from the secondary O2 regulator.
2-6
  Blended Gas Sequence
  For hand ventilation, the blender output supplies an external calibrated flow control valve that is gravity-
  independent. For mechanical ventilation, a 4-way valve serves as the On/Off control for the mechanical
  ventilator. One circuit of the 4-way valve provides the mechanical ventilator with low-pressure O2 as
  logic gas. The other circuit of the 4-way valve provides blended gas as patient gas. The patient gas
  includes an adjustable pop-off valve and an O2 sensor. The O2 sensor feeds O2 percentage information to
  the oxygen analyzer. The patient circuit includes a gauge tube that feeds back to the mechanical
  ventilator.
Transport Incubator
  NOTE:
  All models contain the Transport Incubator, Model TI500.
  The Transport Incubator, Model TI500, permits transport of high-risk, premature, low birth-weight, or
  critically ill newborns. It provides a means to control air temperature and to change O2 concentrations
  and relative humidity. For additional information refer to the Transport Incubator, Model TI500,
  Instructions for Use.
Modular Frame
  The modular frame system includes the following:
      •   Overhead accessory shelf with tie-down strap
      •   Gas cylinder/yoke wrench
      •   Articulating wheel/handle for the gas cylinder valves
      •   Mounting bracket for mechanical ventilator valve
      •   End handle grip bars for lifting and maneuvering
      •   High-pressure tubes for delivering Air (pneumatic modules with Air cylinder only) and O2 to the
          pneumatic module
      •   Strap to secure the power cable during transit or contain excess power cable while charging
      •   Strap to secure the O2 and air hoses when external gases are not in use
                                                                                                          2-7
Features
      The Features subsection provides a list of the standard and optional features and available accessories for
      the TI500 Globe-Trotter Neonatal Transport System.
Standard Features
      The standard features for the TI500 Globe-Trotter Neonatal Transport System include:
      • Mechanical ventilator (with blender)
      • Transport incubator (Model TI500)
      • Modular frame
      • Suction Pump
      • Accessory shelf
      • Oxygen analyzer
      • High hood
Optional Features
      The optional features for the TI500 Globe-Trotter Neonatal Transport System are:
      • Air compressor
      • Variable height cart
Accessories
      The accessories used on the TI500 Globe-Trotter Neonatal Transport System are:
      • I V pole assembly
      • Skin probes
      • Patient ventilation circuits
      • Mattress
      • Mattress restraint straps
      • Humidity pad
2-8
Specifications
Standard Features
                       Feature                                          Dimension
   120 V AC model external power requirements 110/120 V, 50/60 Hz
   120 V AC outlet specifications                  Six, 10A breakers
   120 V AC consumption                            < 480-W worst case (incubator, compressor char-
                                                   ger)
   230 V AC model external power requirements 230 V, 50/60 Hz
   230 V AC outlet specifications                  Six, 5A breakers
   230-V consumption                               < 400-W worst case (incubator, compressor char-
                                                   ger)
   Chassis leakage current limit (120 V AC         300 A
   model)
   Chassis leakage current limit (230 V AC         500 A
   model)
   Pneumatic module inlet pressure                 345 kPa (50 psi) ± 34 kPa (5 psi)
   Ventilator input pressure (pneumatic module     345 kPa (50 psi) ± 34 kPa (5 psi)
   with Air cylinder version)
   Ventilator input pressure (pneumatic module     345 kPa (50 psi) ± 34 kPa (5 psi)
   with compressor version)
   Pneumatic ventilator output                     0 liters per minute (lpm) to 20 lpm
   Operating temperature range (normal             10 °C (50 °F) to 30 °C (86 °F)
   functioning)
   Operating temperature range (limited use)       10 °C (50 °F) to 40 °C (104 °F)
   Storage temperature range                       -20 °C (68 °F) to 60 °C (140 °F) ambient
   Operating humidity range                        0% to 95% relative humidity (RH) non-condensing
   TI500 Globe-Trotter Neonatal Transport Sys- 129.5 cm (51")
   tem length without VHA cart
   TI500 Globe-Trotter Neonatal Transport Sys- 56.5 cm (22.3")
   tem width without VHA cart
   TI500 Globe-Trotter Neonatal Transport Sys- 58.4 cm (23")
   tem height without VHA cart
   TI500 Globe-Trotter Neonatal Transport Sys- 104.5 kg (230.2 lb)
   tem weight (incubator and pneumatic module
   with compressor) without gas cylinders, I.V.
   poles, patient monitor, or VHA cart
   TI500 Globe-Trotter Neonatal Transport Sys- 92.2 kg (203.2 lb)
   tem weight (incubator and pneumatic module
   without compressor) without gas cylinders, I.V.
   poles, patient monitor, or VHA cart
   Noise level inside the incubator without alarms <60 dBa
   Noise level inside the incubator with alarms    <80 dBa
                                                                                                     2-9
Cart Features
                            Feature                                         Dimension
         VHA cart length                                  Model 147A -101.6 cm (40")
                                                          Model 35IT - 200.6 cm (79")
         VHA cart width                                   Model 147A - 55.9 cm (22")
                                                          Model 35IT - 61 cm (24")
         VHA cart height, low position                    Model 147A - 88.3 cm (32.8")
                                                          Model 35IT - 25.4 cm (10")
         VHA cart height, high position                   Model 147A - 118.8 cm (46.8")
                                                          Model 35IT - 107 cm (42")
         VHA cart weight                                  Model 147A - 22.3 kg (51 lb)
                                                          Model 35IT - 36.3 kg (80 lb)
Optional Features
                            Feature                                          Dimension
         Air compressor DC power requirements             12 V DC, 11 A
         Compressor output from regulator                 20 lpm at 207 kPa (30 psi) ± 14 kPa (2 psi)
         Compressor battery quantity and type             One sealed, gel-type, and rechargeable
         Compressor battery voltage                       12 V DC (nominal)
         Compressor battery charging time (from full      20 hours or less
         discharge)
         Compressor battery life expectancy               Minimum 200 full charge/discharge cycles
         Compressor battery size rating                   20 A hours
         Compressor battery run time (with full battery   Minimum of 3 hours with the blender set at 21%
         charge)                                          oxygen at 10 L/min
         Flow control valve output (pneumatic module      0 lpm to 15 lpm
         with Air cylinder)
         Flow control valve output (pneumatic module      0 lpm to 10 lpm
         with compressor)
         Service life                                     7 years
Device Classification
         Classification according to EU Directive         Class IIb
         93/42/EEC
         Protection class according to IEC 60601-1        I
         Applied part                                     Type BF
         Continuous operation
         Universal Medical Device Nomenclature Sys-       12-114
         tem (UMDNS) code
         Global Medical Device Nomenclature               35121
         (GMDN) code
2 - 10
                                       Section 3
                      Precautions and Safety Tips
General Precautions
  WARNING:
  Federal Law (U.S.) restricts this device to sale by or on the order of a physician.
  WARNING:
  Only properly trained personnel, directed by an appropriately qualified physician aware of the
  current known risks and benefits of infant transportation and use of the associated equipment,
  should use the TI500 Globe-Trotter Neonatal Transport System. Use by untrained personnel or
  unqualified physicians could result in personal injury or equipment damage.
  WARNING:
  Follow each product manufacturer instructions. Failure to do so could result in personal injury or
  equipment damage.
  WARNING:
  This product has been validated with the accessories and options listed in this manual. It
  complies with all relevant safety and performance requirements applicable to the device. Any
  person or organization who makes an unauthorized modification, or incorporates an
  unapproved attachment to the device, is responsible to ensure that the system still complies
  with those requirements.
  CAUTION:
  Take care not to scratch or otherwise damage unprotected surfaces.
  CAUTION:
  Do not use harsh cleansers, solvents, or detergents. Equipment damage could occur.
  CAUTION:
  For incubator stability, shelf loading must not exceed 15 lb (6.8 kg) - if single O2 cylinder is
  used, place in the front gas cylinder holder.
                                                                                                     3-1
Electrical Precautions
      SHOCK HAZARD:
      The circuit breaker ensures that the maximum current capacity of the 230 V AC powered
      system is 5 A; and the 120 V AC powered unit has a capacity of 10 A. To ensure grounding
      reliability, plug the AC power cable into only a properly grounded, three-wire, hospital-grade, or
      hospital use outlet. Failure to do so could result in personal injury or equipment damage.
      SHOCK HAZARD:
      When any of the electronics covers are removed, an electrical shock hazard exists. Only
      qualified personnel should perform service. Servicing by unqualified personnel could cause
      personal injury or equipment damage.
      SHOCK HAZARD:
      Do not reset a circuit breaker or replace a fuse without assessing and correcting what caused
      the breaker or fuse to activate. Personal injury or equipment damage could occur.
      SHOCK HAZARD:
      When plugging in items other than items supplied with the system, check the total current
      leakage to ensure that it does not exceed the acceptable limits. Failure to do so could result in
      personal injury or equipment damage.
      SHOCK HAZARD:
      Before use, check ground continuity between the systems and the AC plug ground pin. To
      ensure grounding reliability, connect the system power cable only to a properly grounded, three-
      wire, hospital-grade, or hospital-use outlet of the proper voltage and frequency. Do not use
      extension cords. Personal injury or equipment damage could occur.
      SHOCK HAZARD:
      If the system circuit breaker is activated, all non-battery-operated equipment ceases to operate.
      The plugged-in components must not exceed the maximum total current draw of the system to
      avoid activating the circuit breaker. Check the respective current draw of each system
      component, and verify that the total system current draw does not exceed the specified
      maximum draw. Failure to do so could result in personal injury or equipment damage.
      SHOCK HAZARD:
      Ensure that the AC output is of the appropriate voltage and ampere capacity. Some AC-
      powered equipment is dependent on a sine-wave AC source. Inverters provide either a sine-
      wave, a quasi-sinusoidal wave, or a square wave pattern. The incubator, suction pump,
      compressor charger, and monitor require at least a quasi-sinusoidal wave pattern. Other
      equipment could be sensitive to a non-sine wave AC source. Equipment damage could occur.
      SHOCK HAZARD:
      Unplug the power cable from its power source before performing any service or cleaning the
      system. Failure to do so could result in personal injury or equipment damage.
3-2
  SHOCK HAZARD:
  Do not expose the unit to excessive moisture. Personal injury or equipment damage could
  occur.
Explosion Precautions
  WARNING:
  Do not use the TI500 Globe-Trotter Neonatal Transport System in the presence of flammable
  anesthetics. Personal injury or equipment damage could occur.
  WARNING:
  Before performing electrical service procedures or changing any battery, shut off the O2
  cylinders. Failure to do so could result in personal injury or equipment damage.
  WARNING:
  Switch off all gas sources, and bleed off remaining pressure. When performing cleaning or
  maintenance procedures in an oxygen-enriched environment, a fire and explosion hazard
  exists. Personal injury or equipment damage could occur.
Oxygen Precautions
  WARNING:
  Switch off the cylinder gas when it is not in use or when the system is connected to an external
  gas source. Failure to do so could result in personal injury or equipment damage.
  WARNING:
  Improper use of supplemental oxygen can cause serious side effects, including blindness, brain
  damage, and death. The risks vary with each infant. The attending physician prescribes the
  method, concentration, and duration of oxygen administration. Failure to do so could result in
  serious side effects to the infant.
  WARNING:
  If it is necessary to administer oxygen in an emergency, notify the attending physician
  immediately. Failure to do so could result in infant injury or equipment damage.
  WARNING:
  Measure oxygen concentrations with a calibrated oxygen analyzer at intervals directed by the
  attending physician. Failure to do so could result in infant injury or equipment damage.
  WARNING:
  Oxygen therapy requires that special care is taken to prevent fire. Any materials that burn in air,
  and some that do not, are easily ignited, and burn rapidly in oxygen-enriched air. Accordingly,
  for safety, keep all sources of ignition away from the incubator and out of the area where it is
  being used. Failure to do so could result in personal injury or equipment damage.
                                                                                                  3-3
      WARNING:
      On high-pressure O2 cylinders, use only approved reducing or regulating valves marked for
      oxygen service. Do not use these valves for air or gases other than oxygen since they could be
      hazardous when returned to oxygen service. Operate such equipment strictly according to the
      directions of the valve manufacturer. Failure to do so could result in personal injury or
      equipment damage.
      WARNING:
      Although oxygen-compatible materials are used in the oxygen delivery system, take special
      care when high-pressure oxygen, such as found in a medical O2 cylinder, is used. If their
      ignition temperature is reached, violent ignition of oil, grease, greasy substances, small
      particles or dust, dirt, or other particulate contaminants, even small particles of metal, could
      occur in the presence of high-pressure oxygen. If the O2 cylinder valve is opened too rapidly, an
      instantaneous increase in temperature could occur due to friction, particle acceleration, or
      adiabatic compression. Serious personal injury or equipment damage could occur.
      WARNING:
      Keep oil, grease, greasy substances, dust, dirt, and other particulate contaminants away from
      oxygen regulators, cylinder valves, tubing, and all other oxygen equipment. Failure to do so
      could result in personal injury or equipment damage.
      WARNING:
      Always open the cylinder shut-off valves slowly and carefully. Failure to do so could result in
      personal injury or equipment damage.
      WARNING:
      When installing new or replacement O2 cylinders, clear their outlet ports by cracking the
      cylinder valve momentarily before attaching it to the equipment. Failure to do so could result in
      personal injury or equipment damage.
      WARNING:
      Antistatic or electrically conductive hoses and tubing must not be used.
3-4
Electromagnetic Compatibility Precautions
  General information on electromagnetic compatibility (EMC) pursuant to international EMC standard
  IEC 60601-1-2:
  Electromedical devices are subject to special precautionary measures concerning electromagnetic
  compatibility (EMC) and must be installed and put into operation in accordance with the EMC
  information provided in the technical documentation available from DrägerService.
          Connector pins with an electrostatic discharge (ESD) warning sign should not be
          touched and no connections should be made between these connectors without
          implementing ESD protective measures. Such precautionary procedures may include
          antistatic clothing and shoes, the touch of a ground stud before and during connecting
          the pins or the use of electrically isolating and antistatic gloves. All staff involved in the
          above shall receive instruction in these ESD precautionary procedures.
  CAUTION:
  Portable and mobile RF communications equipment can affect medical electrical equipment.
System Precautions
  WARNING:
  Compressed gas cylinders, such as air and oxygen, can become hazardous projectiles when
  the gas releases rapidly due to damage or other causes. To prevent movement or damage from
  shock or impact to the cart or incubator, securely fasten the cylinders. To prevent cylinder
  movement, tighten the hold-down mechanism well. Failure to do so could result in personal
  injury or equipment damage.
  WARNING:
  When the system is secured in a land ambulance, airplane, or helicopter, do not place anything
  on the right-hand side of the frame. Doing so could prevent access to the on-board cylinders.
  The cylinders are removed and installed from the right side of the frame. Allow for the needed
  space. Failure to do so could result in personal injury or equipment damage.
  WARNING:
  Do not attach the self-inflating bag to the incubator wall. An increase in carbon dioxide (CO2)
  and breathing work for the infant could result. Personal injury or equipment damage could
  occur.
  WARNING:
  For safety, due to the weight and complexity of the system, two people of sufficient strength
  must assemble and move the unit. Failure to use two people could result in personal injury or
  equipment damage.
  WARNING:
  The battery pack weighs 18.1 kg (40 lb). To avoid personal injury, exercise care when
  removing the battery tray.
                                                                                                     3-5
      WARNING:
      Do not place the TI500 Globe-Trotter Neonatal Transport System in service without each
      component passing the respective manufacturer checkout procedure. Otherwise, personal
      injury or equipment damage could occur.
      WARNING:
      Before clinical use, correct all sources of gas leakages. Failure to do so could result in personal
      injury or equipment damage.
      WARNING:
      Use clean, dry gas. Failure to do so could result in personal injury or equipment damage.
      WARNING:
      Do not use a cylinder without first removing the vinyl wrapper. Never use an E- or D-size
      cylinder missing its gas outlet wrapper without first “cracking” the gas cylinder valve to remove
      any debris. For the proper method, consult the facility gas cylinder safety procedures. Failure to
      do so could result in personal injury or equipment damage.
      WARNING:
      The cylinder gasket is the source of the seal. Use fresh cylinder gaskets. Failure to do so could
      result in personal injury or equipment damage.
      WARNING:
      Protect the webbing of the patient monitor securing strap from sharp edges and heat. If the
      webbing is cut or worn, replace it. Failure to do so could result in personal injury or equipment
      damage.
      WARNING:
      Only facility-authorized personnel should troubleshoot and perform preventive maintenance on
      the TI500 Globe-Trotter Neonatal Transport System. Troubleshooting and preventive
      maintenance performed by unauthorized personnel could result in personal injury or equipment
      damage.
      WARNING:
      Routine use of the TI500 Globe-Trotter Neonatal Transport System can result in the gradual
      loosening of the screw fasteners. This loosening is important when used in a helicopter
      environment where intense vibration is common. Assess the system components for signs of
      loosened fasteners, and secure when necessary. Failure to do so could result in personal injury
      or equipment damage.
      WARNING:
      Do not block the air vents on the sides of unit.
      CAUTION:
      Before transport, ensure that the frame is securely seated on the four alignment pins and that
      the two locking blocks are secured over the frame. Failure to do so could result in equipment
      damage.
3-6
  CAUTION:
  Tightening the securing bolt on a misaligned cylinder permanently damages the yoke inlet and
  keyed pins. Before securing the cylinder, align it correctly. Failure to do so could result in
  equipment damage.
  CAUTION:
  Before placing the TI500 Globe-Trotter Neonatal Transport System in clinical use, ensure that it
  is in optimal condition. The use of partially charged batteries and partially full gas cylinders
  provides less operating time from the on-board battery and gas supplies for the system. Failure
  to heed this caution could lead directly to equipment damage and indirectly to personal injury.
Ventilator Precautions
  WARNING:
  If the gas cylinder pressure display falls below 2068 kPa (300 psi), discontinue use of the
  ventilator, and change to ventilation with a hand ventilator. Failure to do so could result in
  personal injury or equipment damage.
  WARNING:
  During hand ventilation, the delivered oxygen is not being measured or displayed. Infant injury
  or equipment damage could occur.
  WARNING:
  Manometers are a common means of assessing the amount of Peak Inspiratory Pressure (PIP)
  and Positive End Expiratory Pressure (PEEP)/Continuous Positive Alveolar (Airway) Pressure
  (CPAP) delivered. However, there can be great differences between the readings of different
  manufacturer gauges. Pay particular attention when changing the sources of ventilation or
  pressure gauges. The user is the final judge in assessing the adequacy of ventilation. Failure to
  check the infant could result in personal injury or equipment damage.
  WARNING:
  When making oxygen adjustments, maintain the same total flow rate. Failure to do so could
  result in inadvertent ventilator setting changes, namely PIP and PEEP/CPAP. Personal injury or
  equipment damage could occur.
                                                                                                   3-7
      WARNING:
      The mechanical ventilation circuits and exhalation valve are specifically for use with the TI500
      Globe-Trotter Neonatal Transport System. Do not substitute other products without first
      ensuring proper operation. Personal injury or equipment damage could occur.
      WARNING:
      If the mechanical ventilator is used and the oxygen sensor is taken out of the sensor tee, close
      the sensor port with the cap. Failure to do so could result in a large leakage in the ventilator
      circuitry and improper ventilation. Patient injury or equipment damage could occur.
      CAUTION:
      Store the excess sensor cable. Failure to do so could result in equipment damage.
      WARNING:
      At cleaning and installation, inspect the ventilation circuit for damage. Discard damaged parts.
      Failure to do so could result in adverse performance. Personal injury or equipment damage
      could occur.
      WARNING:
      The ventilator included in the system is customized. It may not function in the same manner as
      a standard mechanical ventilator. They are not interchangeable. Refer to the separate
      instructions for use from the manufacturer. Failure to do so could result in personal injury or
      equipment damage.
      CAUTION:
      Do not autoclave or pasteurize the mechanical ventilation block. Equipment damage could
      occur.
Pneumatic Precautions
      WARNING:
      To continue pneumatic module operation when the compressor Low Battery indicator
      activates, connect the system to AC power, or provide compressed air to the pneumatic module
      using the external inlet fitting. If AC power or an external compressed air source is not available,
      discontinue use of the ventilator, and change to ventilation with a hand ventilator. Failure to do
      so could result in personal injury or equipment damage.
      WARNING:
      The gas provided from the pneumatic module is not heated or humidified. For humidification of
      ventilation gas during transport, follow hospital practice. Failure to do so could result in infant
      injury.
      WARNING:
      The pneumatic module acoustic alarm signals can be difficult to detect in a noisy environment,
      such as during air transport. Frequently assess the pneumatic module to anticipate and correct
      conditions resulting from a mechanical ventilator, air compressor, or gas blender condition.
      Failure to do so could result in personal injury or equipment damage.
3-8
WARNING:
When conditions exist that result in a Blender alarm, the mechanical ventilator of the pneumatic
module can malfunction. Change the form of ventilation to a hand ventilator. Then replenish the
exhausted gas source by replacing the empty on-board cylinder or by providing gas to the
respective pneumatic module gas inlet fitting. Failure to do so could result in personal injury or
equipment damage.
CAUTION:
Use a filter and water trap on the system air input hose. The filter and water trap minimize the
possibility of particulate matter or condensed water entering the pneumatic module. Both debris
and water adversely affect operation of the pneumatic module.
CAUTION:
Store the air and oxygen hoses of the pneumatic module in a clean environment. Introduction of
fluid or debris into the pneumatic module gas inlets can result in equipment damage.
CAUTION:
Continued operation of the pneumatic module during the Gas Supply Failure alarm, of the
pneumatic module blender, consumes approximately 10 liters per minute (lpm). Continued
operation of the blender during this condition substantially reduces the operation time from a
cylinder supply source. Failure to heed this caution could lead directly to equipment damage
and indirectly to personal injury.
                                                                                               3-9
         This page intentionally left blank.
3 - 10
                                                         Section 4
                                                  Installation and
                                              Functional Checkout
Installation
  The TI500 Globe-Trotter Neonatal Transport System is shipped in two cartons. One carton contains the
  incubator and pneumatic module assembled on a modular frame. The variable height adjustable (VHA)
  cart is in a separate carton.
  CAUTION:
  Take care not to scratch or otherwise damage unprotected surfaces.
  1. Remove the equipment from the two cartons. Take care not to scratch or otherwise damage
     unprotected surfaces.
  2. Discard the packing material.
  3. Assemble the unit (see “Assembly” on page 4-1).
  4. Install the gas cylinders (see “Cylinder Installation” on page 4-3).
  5. Perform the functional checks. Refer to “Functional Checkout” on page 4-5.
Assembly
  WARNING:
  For safety, due to the weight and complexity of the system, two people of sufficient strength
  must assemble and move the unit. Failure to use two people could result in personal injury or
  equipment damage.
  1. Open the front access panel of the incubator hood, and remove the components.
  2. Loosen the retention screw that secures the battery pack.
  WARNING:
  The battery pack weighs 18.1 kg (40 lb). To avoid personal injury, use care when removing the
  battery tray.
  3. Remove the battery pack, and connect the battery terminal leads.
  NOTE:
  The battery pack can only be removed and installed with the frame separated from the VHA cart.
  4. Reinstall the battery pack, and tighten the retention screw.
                                                                                                    4-1
      5. Secure the modular frame to the VHA cart.
         For all VHA carts, except for Model 147A, refer to the
         separate instructions for use from the manufacturer.
4-2
  Scissor mount dimensions
Cylinder Installation
  WARNING:
  Do not use a cylinder without first removing the vinyl wrapper. Never use an E- or D-size
  cylinder missing its gas outlet wrapper without first “cracking” the gas cylinder valve to remove
  any debris. For the proper method, consult the facility gas cylinder safety procedures. Failure to
  do so could result in personal injury or equipment damage.
                                                                                                 4-3
      WARNING:
      When the system is secured in a land ambulance, airplane, or helicopter, do not place anything
      on the right-hand side of the frame. Doing so could prevent access to the on-board cylinders.
      The cylinders are removed and installed from the right side of the frame. Allow for the needed
      space. Failure to do so could result in personal injury or equipment damage.
      Pin-Index Style Cylinders
      CAUTION:
      Take care when placing the cylinders in their respective cylinder holders. Failure to do so could
      result in equipment damage.
      1. To install a cylinder:
         a. Loosen the securing bolt on the cylinder yoke to accommodate the cylinder post valve.
          b. Align each cylinder post valve so that the cylinder gas outlet is facing inward.
      NOTE:
      The gas yokes can be aligned to accept the varying diameters of the UK-style steel cylinders or the US-
      style aluminum and steel cylinders.
         c. Grasp the yoke assembly firmly.
         d. Align the yoke to match the position of the respective style or composition of cylinder.
      NOTE:
      When installing the valve into the yoke, twist slightly at the base of the cylinder while gently pushing the
      post valve forward into the yoke.
      CAUTION:
      Tightening the securing bolt to misaligned cylinder permanently damages the yoke inlet and
      keyed pins. Before securing the cylinder, align it correctly. The cylinder gasket is the source of
      the seal. Always use new cylinder gaskets. Failure to do so could result in equipment damage.
      NOTE:
      To tighten the securing bolt on the cylinder yoke, use the side of the ratchet wrench labeled On. To
      loosen, use the Off side.
         e. Using the On side of the ratchet wrench provided, tighten the securing bolt on the cylinder yoke
            for the valve outlet.
         f. Secure the cylinder with the straps.
      2. To remove a cylinder:
         a. Close the valve.
         b. Remove the straps from the cylinder.
         c. Using the Off side of the ratchet wrench, loosen the securing bolt on the cylinder yoke.
         d. Perform the installation procedure in reverse order.
4-4
Functional Checkout
  WARNING:
  Do not place the TI500 Globe-Trotter Neonatal Transport System in service without each
  component passing the respective manufacturer checkout procedure. Personal injury or
  equipment damage could occur.
  Before use, fully charge the electronic equipment found in the system, such as the air compressor (if
  present), pumps, and monitors.
  Before first placing the TI500 Globe-Trotter Neonatal Transport System into use and after disassembling
  the unit for cleaning and maintenance, perform the functional checkout. To operate the components not
  manufactured by Draeger Medical Systems, Inc., refer to their respective manufacturer user manuals.
Electrical Checkout
  SHOCK HAZARD:
  Ensure that the power source is compatible with the electrical specifications shown on each
  component. For proper grounding reliability, plug the TI500 Globe-Trotter Neonatal Transport
  System power cable only into a properly marked, hospital-grade receptacle. Do not use
  extension cords. If any doubt exists as to the grounding connection, do not operate the
  equipment. Personal injury or equipment damage could occur.
  1. Plug the TI500 Globe-Trotter Neonatal Transport System power cable into an appropriate power
     source, and verify that the AC indicator illuminates.
  2. Ensure that components with this function indicate AC Charging mode, such as the incubator,
     infusion pump, suction pump, or patient monitor.
  3. Ensure that the retainer clip is engaged on the power cable.
  4. Coil up the excess power cable and secure it in the associated strap.
Mechanical Checkout
  CAUTION:
  Ensure that the frame is firmly seated on the four alignment pins and that the two locking blocks
  are secured before system use. Failure to do so could result in equipment damage.
  1. Ensure that the patient monitors are well secured.
  2. Verify that the system is clean and stocked with the necessary supplies.
  3. Verify that the incubator locking handle, on the cart, operates freely so that the modular frame system
     can be removed from the cart if necessary.
  4. Inspect the tie-down straps, the cord, and the hose straps for wear. Replace, if necessary.
  5. Visually inspect the system for worn, loose, or damaged components. Before use, replace or repair
     damaged components.
                                                                                                          4-5
Pneumatic Checkout
      WARNING:
      Before clinical use, correct all sources of gas leakages. Failure to do so could result in personal
      injury or equipment damage.
      WARNING:
      Use clean, dry gas. Failure to do so could result in personal injury or equipment damage.
      WARNING:
      E- and D-size cylinders come standard with a vinyl wrapper over the gas outlet port to keep the
      port clean. Do not use a cylinder without first removing the wrapper. Personal injury or
      equipment damage could occur.
      WARNING:
      Never use an E- or D-size cylinder missing its gas outlet wrapper without first “cracking” the
      cylinder valve to remove any debris. For a method, consult the facility gas cylinder safety
      procedures. Failure to do so could result in personal injury or equipment damage.
      CAUTION:
      Store the air and oxygen hoses of the pneumatic module in a clean environment. Introduction of
      fluid or debris into the pneumatic module gas inlets can result in equipment damage.
      CAUTION:
      A filter/water trap is recommended for use on the DISS or NIST gas-input hose to isolate
      contaminants before they enter into the pneumatic module. Failure to do so could result in
      equipment damage.
      Checking on-board gas sources
      NOTE:
      The blender alarm sometimes activates during this procedure. This alarm occurs when only one gas
      cylinder is turned on at a time. Disregard this alarm except during step 13.
      1. Switch off the Ventilator Gas Source control.
      2. Close the Flow Controller valve.
      3. Ensure that the gas cylinders are properly installed and connected into their respective holders (see
         “Cylinder Installation” on page 4-3).
      4. For a system with an Air cylinder pneumatic module, confirm that each cylinder is full. For
         compressor versions, go to step 5.
      NOTE:
      1200 psi is the recommended minimum for beginning transport. 300 psi is sufficient for testing.
         a. Open the Air cylinder valve.
         b. Confirm the cylinder pressure shown on the gauge of the pneumatic module.
         c. Open the O2 cylinder.
         d. Confirm the cylinder pressure shown on the gauge of the pneumatic module.
4-6
5. For a system with a compressor pneumatic module, confirm that each cylinder is full.
NOTE:
1200 psi is the recommended minimum for beginning transport. 300 psi is sufficient for testing.
   a. Open one O2 cylinder.
   b. Confirm the cylinder pressure shown on the gauge of the pneumatic module.
   c. Close the open O2 cylinder.
   d. Vent the system by opening the Ventilator Gas Source control and the Flow Controller valve,
      until the pressure shown on the gauge drops to 0.
   e. Switch off the Ventilator Gas Source control and close the Flow Controller valve.
   f. Open the second O2 cylinder (compressor version only).
   g. Confirm the cylinder pressure shown on the gauge of the pneumatic module.
NOTE:
1200 PSI is the recommended minimum for beginning transport. 300 PSI is sufficient for testing.
6. For a system with a compressor pneumatic module, confirm that the compressor operates.
   a. Ensure that the Globetrotter is plugged into the AC mains.
     b. Switch on the compressor. Then verify that the compressor cycles on, and remains on for at least
         2 min. The compressor shuts off within 5 minutes.
NOTE:
If the compressor does not remain on for at least 2 minutes or does not shut off within 5 minutes, there is
a fault in the compressor system. Discontinue use of the Globe-trotter and contact DrägerService.
7. Switch on the Ventilator Gas Source control.
8. Set the Flow Controller valve to 3 lpm.
9. Switch on the oxygen analyzer.
10. Ensure that the oxygen sensor is in the sensor tee at the rear of the pneumatic module.
11. Switch on the ventilator, and adjust settings as directed by qualified personnel.
12. Verify the ventilator output according to the instructions for use for the ventilator.
13. Confirm that the blender and oxygen analyzer function.
   a. Set the blender to 21%. Confirm that the oxygen analyzer shows 21% +/-3% within 2 minutes.
   b. Set the blender to 100%. Confirm that the oxygen analyzer shows 100% +/-3% within 2 minutes.
   c. Close the O2 cylinder valve. Confirm that the blender alarm activates within 2 minutes. Open the
      O2 cylinder valve.
   d. Switch off the compressor, if pneumatic module is equipped with a compressor. Confirm that the
      blender alarm activates within 2 minutes. Switch on the compressor.
   e. Close the Air cylinder valve, if pneumatic module is equipped with an Air cylinder. Confirm that
      the blender alarm activates within 2 minutes. Open the Air cylinder valve.
14. Switch off the ventilator and oxygen analyzer.
                                                                                                        4-7
      15. Confirm that the Flow Controller valve functions.
         a. Close the Flow Controller valve. Manually verify that gas flow stops.
         b. Open the Flow Controller valve to 3 lpm. Manually verify that gas flows.
         c. Open the Flow Controller valve to 10 lpm. Manually confirm that gas flow is greater than 3 lpm.
      16. Close all cylinder valves.
      17. Switch off the compressor, if equipped.
      18. After the gas stops flowing through the Flow Controller valve, switch off the Ventilator Gas Source
          control and Flow Controller valve.
      Checking external gas sources
      NOTE:
      The blender alarm activates during this procedure. This alarm occurs when only one gas cylinder is
      turned on at a time. Disregard this alarm except during step 12.
      1. Switch off the Ventilator Gas Source control.
      2. Close the Flow Controller valve.
      3. Clear the hoses of any debris.
         a. Disconnect the air hose from the pneumatic module.
         b. Attach the air hose to a source of medical-grade, compressed air (70 psi maximum).
         c. Briefly purge the air hose while pointing the open end away from any personnel.
         d. Disconnect the air hose from the air source.
         e. For the oxygen hose, repeat the procedure using medical-grade oxygen.
      4. Connect each hose from the appropriate central gas to the pneumatic module. Finger-tighten the
         fitting of each hose.
      5. Switch on the central gas supplies. Confirm the input pressures.
      6. Switch on the Ventilator Gas Source control.
      7. Set the Flow Controller valve to 3 lpm.
      8. Switch on the oxygen analyzer.
      9. Ensure that the oxygen sensor is in the sensor tee at the rear of the pneumatic module.
      10. Switch on the ventilator, and adjust settings as directed by qualified personnel.
      11. Verify the ventilator output according to the instructions for use for the ventilator.
      12. Confirm that the blender and oxygen analyzer function.
         a. Set the blender to 21%. Confirm that the oxygen analyzer shows 21% +/-3% within 2 minutes.
         b. Set the blender to 100%. Confirm that the oxygen analyzer shows 100% +/-3% within 2 minutes.
      13. Switch off the ventilator and oxygen analyzer.
      14. Close all wall terminal units.
      15. Switch off the compressor, if equipped.
4-8
  16. After the gas stops flowing through the Flow Controller valve, switch off the Ventilator Gas Source
      control and Flow Controller valve.
  17. Disconnect the hoses, and return them to their associated straps.
System Checkout
  The TI500 Globe-Trotter Neonatal Transport System is a system of integrated components commonly
  used to transport the acutely ill newborn. However, the pre-configured system does not replace the
  responsibility of the user to include any additional equipment necessary for the care of the infant.
  To complete the system checkout, each product used with the transport system must satisfy the
  manufacturer functional checks.
  Upon successful completion of each component checkout procedure, perform the following:
  1. Plug the TI500 Globe-Trotter Neonatal Transport System into an appropriate AC power source.
  2. Set the On/Off switch of the transport incubator to the On position so the incubator batteries can
     maintain a full charge.
  NOTE:
  On/Off switch of the transport incubator is located on the power chassis on the left side.
  3. Perform the Operational Checkout for the transport incubator (TI500). Refer to the separate
     instructions for use from the manufacturer.
                                                                                                          4-9
         This page intentionally left blank.
4 - 10
                                                         Section 5
                                              Instructions for Use
Controls, Indicators, and Connectors
  NOTE:
  This view is typical of the TI500 Globe-Trotter Neonatal Transport System.
                                                                               5-1
      The major system components are listed in this table:
                                 Reference                      Description
                                     A          TI500 Incubator
                                     B          Overhead accessory shelf
                                     C          I V pump mounting frame
                                     D          Pneumatic module
                                    E, I        Locking block
                                     F          Variable height cart (Model 147A shown)
                                     G          Recessed wheels
                                     H          Retaining latch
                                     J          Modular frame system
                                     K          Mechanical ventilation circuit couplings
                                     L          Exhalation valve holder
                                    M           Cable/hose strap
5-2
Alarms
  An alarm is signaled whenever a condition, which could be potentially hazardous, is detected. This
  section lists the TI500 Globe-Trotter Neonatal Transport System alarms.
  Refer to the separate instructions for use from the manufacturer for alarms associated with the ventilator
  and the accessories. For the incubator, refer to the Transport Incubator, Model TI500, Instructions for
  Use.
Blender Alarms
  When the blender detects the absence of a supply gas or a significant difference in the pressure in the gas
  supplies, it produces the Blender alarm or Gas Supply Failure alarm.
  CAUTION:
  Continued operation of the pneumatic module during the Gas Supply Failure alarm, of the
  pneumatic module blender, consumes approximately 10 litres per minute (lpm). Continued
  operation under this condition substantially reduces the operation time from a gas cylinder
  supply source. Failure to heed this caution could lead directly to equipment damage and
  indirectly to personal injury.
  During a Gas Supply Failure alarm, stop operation of the pneumatic module.
  WARNING:
  When conditions exist that result in a Blender alarm, the mechanical ventilator of the pneumatic
  module can malfunction. Change the form of ventilation to a hand ventilator. Then replenish the
  depleted gas source by replacing the on-board empty gas cylinder or by providing gas to the
  respective pneumatic module gas inlet fitting. Failure to do so could result in patient injury or
  equipment damage.
  During a Blender alarm:
  • Change the form of ventilation to a hand ventilator.
  • Replace the empty gas cylinder, or provide gas to the respective external gas inlet fitting.
                                                                                                          5-3
      When less than 30 min of battery operation remains, the Low Battery alarm is generated: an LED
      flashes, and an acoustic alarm signal activates.
System
      CAUTION:
      Before placing the TI500 Globe-Trotter Neonatal Transport System in clinical use, ensure that it
      is in optimal condition. The use of partially charged batteries and partially full gas cylinders
      provides less operating time from the on-board battery and gas supplies on the system. Failure
      to heed this caution could lead directly to equipment damage and indirectly to personal injury.
      To minimize the time needed before placing the system back into clinical use, follow the sequence for
      recharging the temporary supplies for the system. The supplies include batteries and gas cylinders.
      Before clinical use, ensure that each device/component passes its recommended checkout procedure.
5-4
To Use an External AC Power Source
SHOCK HAZARD:
Before use, check ground continuity between the systems and the AC plug ground pin. To
ensure grounding reliability, connect the system power cable only to a properly grounded, three-
wire, hospital-grade, or hospital-use outlet of the proper voltage and frequency. Do not use
extension cords. Personal injury or equipment damage could occur.
SHOCK HAZARD:
If the system circuit breaker is activated, all non-battery-operated equipment ceases to operate.
To avoid activating the circuit breaker, the plugged-in components must not exceed the
maximum total current draw of the system. Check the current draw of each system component,
and verify that the total system current draw does not exceed the specified maximum draw.
Failure to do so could result in personal injury or equipment damage.
SHOCK HAZARD:
Ensure that the AC output is of the appropriate voltage and ampere capacity. Some AC-
powered equipment is dependent on a sine-wave AC source. Inverters provide either a sine-
wave, a quasi-sinusoidal wave, or a square wave pattern. The incubator, suction pump,
compressor charger, and monitor require at least a quasi-sinusoidal wave pattern. Other
equipment can be sensitive to a non-sine wave AC source. Equipment damage could occur.
NOTE:
Ambulances are often equipped with an inverter to provide AC power.
Compressor version pneumatic system
When the system is plugged into an appropriate AC power source, the On AC LED on the rear of the
pneumatic module illuminates. If the battery requires a charge and the system is plugged into an AC
power source, both the On AC LED and the In Use LED illuminate. The On AC LED always
illuminates when the system is plugged into an appropriate AC power source. The In Use LED
illuminates if the system is plugged into an AC power source and the battery requires charging. It also
illuminates when the air compressor is in use.
Air cylinder version pneumatic system
When the system is plugged into an appropriate AC power source, the AC indicator on the front of the
pneumatic module illuminates.
                                                                                                      5-5
      To Use the Auxiliary Power Strip
      WARNING:
      Do not reset a circuit breaker or replace a fuse without assessing and correcting what caused
      the breaker or fuse to activate. Personal injury or equipment damage could occur.
      The AC power strip is on the rear of the pneumatic module. A 10 A circuit breaker is integral with the
      120-V system power strip, and a 5 A circuit breaker is integral to the 230-V systems.
      1. Plug the incubator, suction pump, and any optional equipment, such as patient monitor or infusion
         pump, into these outlets for AC power in the hospital or ambulance.
      SHOCK HAZARD:
      When plugging in items other than items supplied with the system, check the total leakage
      current. Ascertain that it does not exceed the acceptable limits. Failure to do so could result in
      personal injury or equipment damage.
      2. For devices not supplied with the system, ensure that the total leakage current does not exceed the
         acceptable limits.
      3. Ensure that the AC indicator is illuminated.
      NOTE:
      To charge the devices plugged into the outlets, the AC indicator must be illuminated.
      To Prepare the System for Transport
      Temporary, on-board, electric and pneumatic supplies, such as batteries and gas cylinders, can power the
      pneumatic system. Additionally, external electric and pneumatic supplies can power the pneumatic
      system during transport. External electric and pneumatic supplies include AC power and independent gas
      sources.
      For the individual battery recharge and use time of the transport system components, refer to the separate
      instructions for use from the manufacturer.
      To prepare the system for transport:
      SHOCK HAZARD:
      Unplug the power cable from its power source. Failure to do so could result in personal injury or
      equipment damage.
      1. Keep the incubator pre-warmed to 35 °C (95 °F) while on AC power.
      2. Unplug the power cable from its power source.
      3. Coil the power cable, and secure it in the power cable storage strap.
      4. Detach the hoses from the external gas sources and gas inlet fittings.
      5. Coil the hoses, and secure them in the hose storage strap.
      6. Secure the system into the ambulance in accordance with the method defined for each type of vehicle
         or mode of patient transport, such as air or land ambulance.
Transport incubator
      Refer to the Transport Incubator, Model TI500, Instructions for Use to operate the incubator.
5-6
Pneumatic Module
  WARNING:
  The ventilator included in the system is customized. It may not function in the same manner as
  a standard mechanical ventilator. They are not interchangeable. Refer to the separate
  instructions for use from the manufacturer. Failure to do so could result in personal injury or
  equipment damage.
  WARNING:
  The gas provided from the pneumatic module is not heated or humidified. For humidification of
  ventilation gas during transport, follow the facility practice. Failure to do so could result in infant
  injury or equipment damage.
  CAUTION:
  The gas output from the pneumatic module is dry. Evaluate the need for humidifying hand and
  mechanical ventilators, especially in situations of extended duration. Adapt your methods to the
  system. Failure to do so could result in equipment damage.
  WARNING:
  The acoustic alarm signals of the pneumatic module can be difficult to detect in a noisy
  environment, such as during air transportation. Frequently assess the pneumatic module to
  anticipate and correct conditions resulting from a mechanical ventilator, air compressor, or gas
  blender condition. Failure to do so could result in injury or equipment damage.
  NOTE:
  It is normal for water to exit the condensate drain tube below the pneumatic module. This phenomenon
  increases in humid climates and where extended use of the air compressor occurs, such as long transport,
  low FiO2, and high gas flow.
  To operate properly, the pneumatic module requires supplies of both oxygen and air at equal pressures.
  Oxygen and air can be supplied from either on-board cylinders or from external sources, such as central
  source medical oxygen and air or external cylinders. Gas supplied by on-board cylinders is regulated
  within the pneumatic module. Gas supplied by external sources must be regulated externally in
  accordance with the specifications table (“Specifications” on page 2-9). DISS and NIST fittings are
  available for use of external sources.
  Check-valves in the gas cylinder plumbing prevent gas supplied from external sources from depleting
  through the cylinder fittings, even when no cylinder is connected. On modules with a compressor, they
  also prevent the O2 cylinder with the higher pressure from cross-feeding into the cylinder with the lower
  pressure.
  Frequently assess the pneumatic module to anticipate and correct conditions resulting from a mechanical
  ventilator, air compressor, or gas blender condition.
  During a short or extreme current draw in the DC circuit, a 10 A circuit breaker, that is integral to the
  compressor On/Off switch, interrupts 12-V compressor operation.
  During a DC short, a 10 A fuse in the DC output line of the charger interrupts output.
  During a short or extreme current draw in the AC charger circuit, a 2 A circuit breaker in the 120 V AC
  charger or a 1 A circuit breaker in the 230 V AC system interrupts operation of the charger.
                                                                                                              5-7
      To Activate the Pneumatic Module
      When on-board cylinder gas supplies are in use, store the oxygen and air hoses for the pneumatic module
      in a clean environment.
      Before introducing an external gas source to the pneumatic module, prepare the hoses.
      To prepare hoses for use with external gas sources
      1. Unsnap the hose storage strap, and remove the air or oxygen hoses.
      2. Uncoil each hose, and purge it with gas.
         a. Disconnect the air hose from the pneumatic module.
         b. Attach the air hose to a source of 345 kPa (50 psi) ± 34 kPa (5 psi) medical-grade, compressed air.
         c. Briefly purge the air hose while pointing the open end away from any personnel.
         d. Disconnect the air hose from the air source.
         e. Repeat steps a through d using medical-grade oxygen in step b for the oxygen hose.
      WARNING:
      Use a filter and water trap on the system air input hose. The filter and water trap minimize the
      possibility of particulate matter or condensed water entering the pneumatic module. Both debris
      and water can adversely affect operation of the pneumatic module. Failure to do so could result
      in equipment damage and patient injury.
      3. Use a filter and water trap on the air input hose.
      4. Attach the hoses to their associated inlet fitting.
      5. Switch on the external gas supply, and switch off the corresponding on-board gas source.
      To Activate the Pneumatic Module
      1. Install and open the air and/or O2 cylinders, as required.
      2. Ensure that the power cable for the pneumatic module is connected to a power source.
      3. Switch on the compressor, if equipped, using the On/Off switch. If not using a compressor, Go to step
         5.
      4. Allow a few seconds for the system to develop enough pressure to charge the accumulator tank.
      NOTE:
      A pressure switch controls the compressor to maintain accumulator pressure at 414 kPa (60 psi) ± 34 kPa
      (5 psi). The compressor motor runs intermittently, even though the On/Off switch and circuit breaker are
      continuously illuminated.
      5. Switch on the On/Off switch on the Ventilator Gas Source on the pneumatic module.
      6. Switch on the oxygen analyzer.
      NOTE:
      The Blender alarm of the pneumatic module remains active until there are appropriate supplies of both
      gases.
5-8
To Operate the System When the Air Compressor Battery is Low
WARNING:
To continue operation of the pneumatic module when the compressor Low Battery indicator
activates, attach the system to AC power, or provide compressed air into the pneumatic
module. Use the NIST or DISS inlet fitting. If AC power or an external compressed air source is
not available, discontinue use of the ventilator, and change to ventilation with a hand ventilator.
Failure to do so could result in personal injury or equipment damage.
1. Attach the system to AC power, or provide compressed air into the pneumatic module by way of the
   NIST or DISS inlet fitting.
2. If AC power or an external compressed air source is not available, discontinue use of the ventilator,
   and change to ventilation with a hand ventilator.
To Assess Each Gas Cylinder
WARNING:
If the gas cylinder pressure display falls below 2068 kPa (300 psi), discontinue use of the
ventilator, and change to ventilation with a hand ventilator. Failure to do so could result in
personal injury or equipment damage.
1. Switch on the gas cylinder.
2. If the cylinder contents are 2068 kPa (300 psi) or less, replace the cylinder.
3. If the cylinder contents are greater than 2068 kPa (300 psi), anticipate the duration of future
   operation, and act accordingly.
4. Switch off the gas cylinder.
5. To bleed off any remaining gas, open the flow control valve, and wait until the flow of gas stops.
6. Restock the other necessary supplies expected to be used during the transport.
To Monitor Gas Reserve
WARNING:
The ventilation equipment relies on a pressured gas source for operation. Do not allow the
cylinder pressure to fall below 2068 kPa (300 psi). Regulator output could become
unpredictable, and could adversely affect operation of the pneumatic module. For ease in
monitoring the cylinder reserve, use the air, when present, and oxygen gauges. If the pressure
falls below 2068 kPa (300 psi), discontinue use of the ventilator, and change to ventilation with
a hand ventilator. Failure to do so could result in personal injury or equipment damage.
NOTE:
Ambulances are often equipped with additional gas cylinders that contain their own regulator. The
ambulance has a hospital-like central gas outlet configuration. However, if these sources are considered
to be part of the system gas supply, the transport operator retains the responsibility for assessing the
volume of gas reserves. Gas inlet fittings, specified by regulatory agencies, exist for air and oxygen. For
inlet gas pressure requirements, refer to “Specifications” on page 2-9.
1. Use the air, when present, and oxygen gauges to monitor the gas cylinder reserve.
2. If the pressure falls below 2068 kPa (300 psi), discontinue use of the ventilator, and change to
   ventilation with a hand ventilator.
                                                                                                        5-9
     To Administer Oxygen
     For stable oxygen delivery, attach a ventilation circuit to the patient.
    WARNING:
    Improper use of supplemental oxygen can cause serious side effects including blindness, brain
    damage, and death. The risks vary with each infant. The attending physician prescribes the
    method, concentration, and the duration of oxygen administration. Failure to do so could result
    in patient injury or equipment damage.
    WARNING:
    Measure oxygen concentrations with a calibrated oxygen analyzer at intervals directed by the
    attending physician. Failure to do so could result in patient injury or equipment damage.
    WARNING:
    Oxygen therapy requires special care to prevent fire. Any materials which burn in air, and some
    that do not, are easily ignited, and burn rapidly in oxygen-enriched air. Accordingly for safety,
    keep all sources of ignition away from the incubator and out of the area in which it is being used.
    Failure to do so could result in personal injury or equipment damage.
    WARNING:
    If oil, grease, or greasy substances come in contact with oxygen under pressure, a
    spontaneous, and violent ignition could occur. Keep these substances away from the incubator
    and out of the area in which it is being used. Failure to do so could result in personal injury or
    equipment damage.
    WARNING:
    On high-pressure O2 cylinders, use only approved reducing or regulating valves marked for
    oxygen service. Do not use these valves for air or gases other than oxygen since they can be
    hazardous when returned to oxygen service. Operate such equipment strictly in accordance
    with the directions of the valve manufacturer. Failure to do so could result in personal injury or
    equipment damage.
     To administer oxygen, use either the flow control valve or mechanical ventilator fresh gas flow as the
     source.
     The flow control valve has detent positions on the knob control. The valve has detents at 1/4, 1/2, 1, 2, 3,
     4, 5, 6, 8, 10, and 15. Combined ventilator and auxiliary flow control rates are limited to 30 lpm.
     NOTE:
     Specifications are ± 10% due to the manufacturer variance.
                                   Specifications of Typical-Size Gas Cylinders
                                       Dimensions
          Cylinder Size                                                  Contents          Weight, empty
                                   (Diameter x Length)
         D (aluminum)      11.4 cm (4.5") x 28 cm (11")              360 l             2 kg (4.5 lb)
         UK-type E         9.7 cm (3.8") x 80.8 cm (31.8")           580 l             5.2 kg (11.5 lb)
         E (aluminum)      10.8 cm (4.25") x 73.7 cm (29")           640 l             3.8 kg (8.5 lb)
         E (steel)         10.2 cm (4") x 73.7 cm (29")              640 l             5.6 kg (12.5 lb)
5 - 10
To Charge the Battery
1. Unsnap the AC power cable from the power cable strap.
NOTE:
Use only an AC outlet capable of sustaining the system operation. It must also be certified for the
environment of system use, such as hospital-grade or Federal Aviation Administration (FAA)-approved.
2. Uncoil the AC power cable to the length required to reach an appropriate AC outlet.
3. Plug the AC power cable into an appropriate AC outlet. Verify that the system AC indicator, located
   on the front of the pneumatic module, illuminates. For compressor versions of the pneumatic module,
   verify that the air compressor On AC indicator, located on the rear of the pneumatic module,
   illuminates.
To Check the Battery Condition after Charging the System
1. Check the condition of the battery before the first use of the TI500 Globe-Trotter Neonatal Transport
   System, after running beyond a Low Battery alarm, and at least every three months.
   a. Set the blender to 21%.
   b. Turn the ventilator gas source to ON.
   c. Turn the flow control valve clockwise to full on.
   d. Observe that the compressor runs for 3 hours.
2. If the Low Battery indicator, or the Low Battery alarm comes on before completing the check of the
   battery condition:
   a. Switch off the compressor.
   b. Place the unit on charge for 20 hours.
   c. After the unit is charged, check the condition of the battery again (refer to step 1).
   d. If it fails after the second charge cycle, replace the battery before putting the system back into
      service.
                                                                                                       5 - 11
     To Use the Mechanical Ventilation Circuits
    WARNING:
    The mechanical ventilation circuits and exhalation valve are specifically for use with the TI500
    Globe-Trotter Neonatal Transport System. Do not substitute other products without first
    ensuring proper operation. Patient injury or equipment damage could occur.
     The mechanical ventilator circuit exhausts gas outside the incubator environment to help maintain the
     incubator heat. The mechanical ventilation circuit is attached to the incubator wall to limit movement of
     the ET tube and to help prevent accidental disconnection. Mechanical ventilation tubing is routed inside
     the incubator wall to protect it from disconnection or kinking.
     A disposable ventilation circuit is included with each TI500 Globe-Trotter Neonatal Transport System.
     Refer to separate instructions for use.
                                            Mechanical
            To                              ventilation                                      Airway
            gauge                                                                            suction
                                            (MV) block
            barb                                                                             port
         To
         patient
         port
                                                                                     Airway
                                                                                     connector
                                             Incubator
                                                                                         Cover
                                             walls
                                                                                         plug
                                                           Valve
                                                           drive
                                                           tube
                          Exhalation
                          valve                           Use the cover plug for testing the MV function only.
5 - 12
To Use the Oxygen Analyzer
An oxygen analyzer is mounted in the upper, left-hand corner of the pneumatic module. The
manufacturer design and function of the oxygen analyzer is unaltered.
To use and care for the oxygen analyzer, refer to separate instructions for use from the manufacturer.
Sensor
WARNING:
If the oxygen sensor is removed from the sensor tee, cap the sensor port. Failure to do so could
result in a large leakage in the ventilator circuitry and improper ventilation. Patient injury or
equipment damage could occur.
WARNING:
If the mechanical ventilator is used, the oxygen sensor must be in the sensor tee, or the sensor
port cap must be securely installed. Failure to do so could result in patient injury or equipment
damage.
The sensor port, mounted above the mains power sockets on the rear of the pneumatic module, contains
the sensor tee.
To measure oxygen concentration in the ventilation circuit:
1. Install the sensor in the sensor port on the rear of the
   pneumatic module.
2. Connect the O2 analyzer cable to the sensor.
3. Tighten the capture nut.
CAUTION:
Store the excess sensor cable. Failure to do so could
result in equipment damage.
4. To protect the sensor cable, secure the excess cable.
Battery
For easy access, the battery is located on the rear of the
oxygen analyzer. For replacement and maintenance
instructions, refer to the separate instructions for use from
the manufacturer.
Calibration
Refer to the separate instructions for use from the manufacturer.
                                                                                                         5 - 13
Infusion Devices
     To use infusion pumps, refer to separate instructions for use from the manufacturer.
     To install infusion pumps:
     1. Using the infusion pump mounting frame at the rear of the pneumatic module, secure the infusion
        pumps with pole clamps. Orient the infusion pumps so they remain inside the rear of the frame.
     2. Route the IV tubing between the incubator hood and the pneumatic module. Then route the IV tubing
         into the right-hand, tubing grommet located on the front access panel.
     NOTE:
     Three additional open grommets exist at the head-end access panel. Using these grommets requires
     disconnecting the IV tubing before moving the infant through the front access panel.
     3. To protect the tubing to the infant, and to avoid hampering the mechanical ventilation block, coil any
        excess IV tubing inside the incubator.
Suction System
     The suction pump is mounted on a holder at the rear of the pneumatic module. The disposable suction
     bottle is secured to the rear of the pump.
     For device use, cleaning, and maintenance, refer to separate instructions for use from the manufacturer.
     A fully charged internal battery sustains the unit for 90 minutes. When the system is connected to an AC
     power source, a charger provides continuous operation. The pump On/Off switch is a rocker switch at the
     top of the pump. Adjust the pressure using the vacuum control knob located to the right of the pump
     On/Off switch. To increase the pressure, turn the knob clockwise. To decrease the pressure, turn the knob
     counterclockwise. A vacuum gauge built into the pump and located left of the On/Off switch displays the
     pressure in cm Hg and inch (") Hg increments.
     The suction tubing enters any open incubator grommet. However, use of the head-end access panel
     grommets requires disconnecting the tubing before removing or inserting the infant through the front
     access panel.
5 - 14
Variable Height Adjustable Cart
  To secure the Variable Height Adjustable (VHA)
  cart into a FERNO Model 175 ambulance
  fastener system, refer to “Assembly” on page
  4-1, step 5.
  To Remove the Frame from the VHA Cart
  1. Pull back the spring-loaded locking handle.
  2. Hook the retaining latch over the top tube of
     the cart.
  3. With an attendant at each end, firmly grasp
     the end grip bars of the frame.
  4. Lift the modular frame straight up and off the VHA cart.
  5. Place the frame on a level, clean surface capable of supporting its weight.
  6. Secure the frame to the new surface.
  To Return the Frame to its Original Position on the VHA Cart
  1. Position the four holes in the frame over the four pins on the VHA cart.
  2. To secure the frame to the VHA cart, release the locking handle.
  3. Visually confirm that the frame is firmly seated on the four alignment pins and that the two locking
     blocks are secured over the frame.
                                                                                                      5 - 15
         This page intentionally left blank.
5 - 16
                                                   Section 6
                                      Cleaning, Maintenance,
                                      and Replacement Parts
Cleaning
  WARNING:
  Follow the product manufacturer instructions. Failure to do so could result in personal injury or
  equipment damage.
  WARNING:
  Switch off all gas sources, and bleed off remaining pressure. When performing cleaning or
  maintenance procedures in an oxygen-enriched environment, a fire and explosion hazard
  exists. Personal injury or equipment damage could occur.
  SHOCK HAZARD:
  Before cleaning, unplug the unit from its power source. Failure to do so could result in personal
  injury or equipment damage.
  SHOCK HAZARD:
  Do not expose the unit to excessive moisture that would allow for liquid pooling. Personal injury
  or equipment damage could occur.
  CAUTION:
  Do not use harsh cleansers/detergents such as scouring pads or heavy-duty grease removers.
  Equipment damage could occur.
  CAUTION:
  Do not use alcohol to clean the incubator hood. Crazing could occur.
General Cleaning
  If there is no visible soilage with possible body fluids, we recommend that you clean the unit with a mild
  detergent and warm water. If disinfection is desired, use a combination cleanser/disinfectant as explained
  in “Disinfecting” on page 6-2.
  Thoroughly clean the system after each transportation of an infant. Before cleaning, remove all solid
  wastes and contaminants from each item. When using any cleaning agent, follow the directions of the
  manufacturer.
Steam Cleaning
  Do not use any steam cleaning device on the unit. Excessive moisture can damage mechanisms in this
  unit.
                                                                                                          6-1
Cleaning Difficult to Access Areas
      To remove difficult spots or stains, we recommend that you use standard household cleansers and a soft
      bristle brush. To loosen heavy, dried-on soil, saturate the spot first.
Disinfecting
      • Use an intermediate level tuberculocidal cleanser/disinfectant, such as CaviCide Surface
        Disinfectant/Decontaminant Cleaner to disinfect. Disinfect only after the unit is empty and
        disassembled (see “Disassembly for Cleaning” on page 6-3).
      NOTE:
      For the US market only, use only an intermediate level tuberculocidal cleanser/disinfectant registered by
      the Environmental Protection Agency (EPA).
      • Dilute the disinfectant as specified on the manufacturer label.
      • Thoroughly dry the unit before putting it back into use.
      • After removing all solid wastes and contaminants from the disassembled parts, refer to the respective
        cleaning procedures for the individual parts.
Cleaning the Tie-Down Straps Used to Secure a Monitor or the Gas Hoses and
Power Cable
      The tie-down straps are constructed of nylon.
      1. Use a mild detergent and warm water to release soil. If necessary, soak the straps.
      2. Follow the cleaning with a surface disinfectant.
      3. Before use, allow the tie-down straps to dry completely.
6-2
  WARNING:
  The ventilation circuit is disposable and is intended for single patient use only. Do not reuse.
  WARNING:
  At cleaning and installation, inspect the ventilation circuit for damage. Discard damaged parts.
  Failure to do so could result in adverse performance. Personal injury or equipment damage
  could occur.
  Clean the mechanical ventilation block with a mild detergent and warm water. Inspect for cracks.
  CAUTION:
  Do not autoclave or pasteurize the mechanical ventilation block. Equipment damage could
  occur.
  Use a cold (liquid) sterilization solution to sterilize the mechanical ventilation block.
                                                                                                         6-3
Maintenance
      WARNING:
      Only facility-authorized personnel should troubleshoot and perform preventive maintenance on
      the TI500 Globe-Trotter Neonatal Transport System. Troubleshooting and preventive
      maintenance performed by unauthorized personnel could result in personal injury or equipment
      damage.
      WARNING:
      Routine use of the TI500 Globe-Trotter Neonatal Transport System can result in the gradual
      loosening of the screw fasteners. This loosening is especially important when used in a
      helicopter where intense vibration is common. Assess the system components for signs of
      loosened fasteners, and secure when necessary. Failure to do so could result in personal injury
      or equipment damage.
      WARNING:
      At the end of the life of the TI500 Globe-Trotter Neonatal Transport System, safely dispose of
      the unit per local requirements. Failure to do so could result in personal injury.
      For disposal of consumable or defective parts, refer to “Disposal (For countries subject to the EU
      directive 2002/96/EC)” on page 6-3.
      To minimize the time needed before placing the system back into clinical use, follow the sequence for
      recharging the temporary supplies for the system. The supplies include batteries and gas cylinders.
Pneumatic Module
      Perform the following pneumatic module maintenance at the specified intervals:
      • Evaluate the high-pressure inlet gas hoses at least every 12 months for signs of fatigue. Also ensure
        that each hose is leak-free when connected to an external gas source.
      • For the oxygen analyzer, replace the 3 AA batteries every 12 months or as recommended by the
        manufacturer. Refer to the separate instructions for use from the manufacturer.
      For compressor-based systems, perform the following:
      • Replace the compressor inlet filter annually.
      • Replace the 12V, rechargeable battery every 2 years.
6-4
Replacement Parts
  The serial number for the TI500 Globe-Trotter Neonatal Transport System is located in the back of the
  pneumatic module. All pieces of equipment received from Dräger at the time of purchase are keyed to
  this serial number. When corresponding about the system or its components, refer to the serial number.
  For a list of parts and accessories for the various system models, refer to the table on page 6-5. Qualified
  service personnel should install or replace items not shown here.
                                                 Accessories
    Part Number                                           Description
   MU20508             TI500 Globe-Trotter Neonatal Transport System Model GT500
   MU04999             Replacement strap assembly, power cable
   MU09042             Handle, valve, removable, with chain
   MU04792             Wrench, ratchet, modification, 9/16" hex
   MU04852             Replacement block assembly, ventilation, mechanical
   MU16753             Cart, height adjustable
   MU04501             I.V. pole assembly, transport system
                                                Consumables
    Part Number                                           Description
   MU04991             Filter, compressor inlet
   MU04771             Ventilation circuit, mechanical, disposable, case of 20
   MP01442             Oxygen sensor, MX300
   13 358 04           Battery, AA, for MX300
   MU16052             Battery, 12V, rechargeable
   MU10918             Suction bottle, 800cc, box of 20
                                            Optional components
    Part Number                                           Description
   MU24474             MVP-10 kit
  For a list of parts and accessories not manufactured by Dräger, refer to the separate instructions for use
  from the manufacturer. For more information about the MX300, refer to the instructions for use from the
  manufacturer. For more information about the MVP-10k or MVP-10k1, refer to the instructions for use
  from the manufacturer.
                                                                                                           6-5
      This page intentionally left blank.
6-6
                                                                  Section 7
                                                           Troubleshooting
Troubleshooting
  WARNING:
  Only facility-authorized personnel should troubleshoot the TI500 Globe-Trotter Neonatal
  Transport System. Troubleshooting by unauthorized personnel could result in personal injury or
  equipment damage.
  For optimal performance, only qualified service personnel should service the items manufactured by
  Dräger Medical in the TI500 Globe-Trotter Neonatal Transport System. For customers requiring service
  outside of the US, contact your local, factory-authorized Dräger Medical distributor or subsidiary.
  To confirm the manufacturer current service and warranty agreement, refer to the product literature
  accompanying the item.
  For troubleshooting of the TI500 Globe-Trotter Neonatal Transport System, refer to the table on page 7-1.
  If you cannot localize the fault, remove the unit from service, and refer it to qualified service personnel.
                                               Troubleshooting
            Symptom                          Possible Cause                       Remedy
   With one gas source active, the  Residual opposing gas is still   Turn the blender to 60%, and
   Blender alarm does not acti-     present in the pneumatic module. vent the residual gas with the
   vate.                                                             flow control valve.
   With both gas sources on, the    The source gas is exhausted, or Replace the source gas with an
   Blender alarm activates.         the output pressure is too low.  adequate source.
                                    The gas regulators are not       Sample the outlet pressures, and
                                    matched.                         adjust accordingly.
                                    The air compressor DC circuit    Assess the air compressor DC
                                    breaker is tripped.              circuit, and reset it.
   The cylinder pressure falls at a The seal is damaged.             Replace the seal.
   rate of >1379 kPa (200 psi)
   over 5 min with the leakage
   test.
                                    The securing bolt and coupling   Tighten the securing bolt, but do
                                    are loose.                       not over-tighten.
                                                                                                           7-1
                Symptom                        Possible Cause                            Remedy
      The air compressor restarts     The water-relieving diaphragm        Clean the bowl and diaphragm
      within 2 min during the leak-   and the seal in the air filter/regu- with mild soap and water, dry,
      age test.                       lator assembly do not seat.          coat with a drop of oxygen-safe
                                                                           lubricant, and assemble.
                                      The internal check valve is dam- Refer the system to qualified ser-
                                      aged on the Diameter Indexed         vice personnel.
                                      Safety System (DISS) or Non-
                                      Interchangeable Safety Thread
                                      (NIST) inlet
                                      The air compressor valves or the Refer the system to qualified ser-
                                      pressure switch is damaged.          vice personnel.
      Gas is felt or heard escaping   The internal check valve is dam- Refer the system to qualified ser-
      from the DISS or NIST fittings aged.                                 vice personnel.
      when the on-board cylinders
      or the air compressor are in
      use.
      Gas is felt or heard escaping   The check valve is damaged.          Refer the system to qualified ser-
      from a cylinder yoke when the                                        vice personnel.
      DISS or NIST gas input is in
      use.
      The blender setting is not      The oxygen analyzer is out of        Calibrate the oxygen analyzer.
      within 5% of the oxygen ana- calibration.
      lyzer reading.
                                      The blender is out of calibration. Contact the manufacturer.
      The air compressor does not     No breathing air is in use, and the This condition is normal. Ensure
      activate with the On/Off        accumulator tank is fully            that the air compressor activates
      switch turned on and illumi-    charged.                             by allowing the consumption of
      nated.                                                               air.
      The air compressor does not     The circuit breaker switch is        Assess, and reset the circuit
      activate with the On/Off        tripped.                             breaker switch.
      switch turned on but not illu-
      minated.
                                      The battery is fully discharged.     Recharge the unit.
                                      The external DC source is not        Validate the presence and polar-
                                      present, or the polarity is          ity of the external DC source.
                                      reversed.
      The On AC LED on the com- The AC circuit breaker is                  Assess, and reset the AC circuit
      pressor charger is not lit when tripped.                             breaker.
      the system is plugged into AC
      power.
                                      AC power is not present.             Ensure that the plug is secured
                                                                           and fully seated in the receptacle.
7-2
          Symptom                         Possible Cause                        Remedy
The In Use LED on the battery     The internal battery charger fuse Refer the system to qualified ser-
charger does not illuminate       is blown.                         vice personnel.
when the system is plugged
into AC power and the air
compressor is active.
The gas cylinder post valve       The yoke is not aligned with the   Firmly grip the yoke assembly,
does not readily fit in the gas   cylinder.                          and maneuver it into position.
yoke.
A loud compressor noise is        The inlet filter is loose or       Assess, reset, or replace the inlet
heard.                            cracked.                           filter.
The variable height adjustable    The orientation of the VHA cart    Verify that the recessed wheel
(VHA) cart does not attach in     is wrong.                          end is positioned on the left side.
the FERNO 175 Ambulance
Fastening System.
                                  The spacing of the FERNO 175       Refer to the separate instructions
                                  Ambulance Fastening System is      for use from the manufacturer for
                                  wrong.                             adjustment.
                                                                                                           7-3
      This page intentionally left blank.
7-4
This page intentionally left blank.
                                      7-5
These instructions for use only apply to
TI500 Globe-Trotter Neonatal Transport
System Model GT500
with the Serial No.:
If no Serial No. has been filled in by Dräger,
these instructions for use are provided for gen-
eral information only and are not intended for
use with any specific machine or device.
This document is provided for customer informa-
tion only, and will not be updated or exchanged
without customer request.
Manufacturer Distributed in US by