Installation Manual ZVIEW - For HUMAN (ENG)

This document provides instructions for installing and removing the ZView digital radiography console software. It discusses system requirements, the installation process including installing drivers and setting configurations, and removing the software. The installation involves selecting setup options, installing the .NET framework, installing drivers, and configuring detector, generator and system settings. Removal involves backing up data and uninstalling the software. It also provides instructions for important operating system setup tasks to ensure proper functionality.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
100% found this document useful (1 vote)
809 views52 pages

Installation Manual ZVIEW - For HUMAN (ENG)

This document provides instructions for installing and removing the ZView digital radiography console software. It discusses system requirements, the installation process including installing drivers and setting configurations, and removing the software. The installation involves selecting setup options, installing the .NET framework, installing drivers, and configuring detector, generator and system settings. Removal involves backing up data and uninstalling the software. It also provides instructions for important operating system setup tasks to ensure proper functionality.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 52

ZView

DR Console Software

Install Manual
(for HUMAN)

Manual Version : 2.1


Revision : 003
Digital Radiography Console Software ZView

Table of Contents
1. Introduction ................................................................................ 4
2. Purpose ......................................................................................... 4
3. System requirements ............................................................... 5

3.1 Workstation ................................................................................................. 5

3.2 Monitor ......................................................................................................... 5

4.Installation & Delete software ............................................... 7

4.1 Installation Preparation ............................................................................ 7

4.2 Installation of the Net Framework ......................................................... 8

4.3 Installation Screen ...................................................................................... 8

4.4 Selecting the Setup route ......................................................................... 9

4.5 Selecting the route of Data Storage..................................................... 10

4.6 Selecting the type of the Detector ....................................................... 11

4.7 Selecting the type of the Generator .................................................... 12

4.8 Starting the Installation .......................................................................... 13

4.9 Installation of the various Drivers ........................................................ 14

4.10 Completing the Installation ................................................................. 15

4.11 Setting the Detector & Generator ..................................................... 16


4.11.1.Detector .............................................................................. 16
4.11.2.Generator............................................................................ 17
4.11.3.System................................................................................. 22

2
Digital Radiography Console Software ZView

4.12 Removing software .................................................................................. 24


4.12.1 Data Backup.......................................................................... 24
4.12.2 Software Removal ................................................................ 24

4.13 Must-Do Setup of OS ............................................................................... 25


4.13.1 Set up of NIC ......................................................................... 25
4.13.2 Window Patch for DVD ......................................................... 31
4.13.3 Virtual Memory Setup(Essential Setup) ................................ 32
4.13.4 DEP ....................................................................................... 39
4.13.5 Removal of Sleep Mode ........................................................ 47

5.Appendix ........................................................................................ 50

5.1. DICOM MWL interface by Text file ....................................................... 50


5.1.1. Principles of interfacing ......................................................... 50
5.1.2. Define order arument ............................................................ 51
5.1.3. Method of Interfacing ........................................................... 51
5.1.4. Sample ................................................................................... 52

3
Digital Radiography Console Software ZView

1. Introduction
Zview is the name of the digital radiography of Atlaim Co. Ltd.
Zview is the DR control software that acquires image from the DR
detector like flat panel, processes rapidly for clinical application, and
provides DICOM services for the integration with PACS system. This
manual explains how to install and operate Zview software.
(Manufacturer: ATLAIM Co., LTD.)

2. Purpose
ZView is the operation console system software integrated with
ATLAIM detectors or X-ray generator, based on FS-MLW (Faster
Specialized Multi Layered Wavelet) technology.
ZView is the product developed for the operator, designed at the
clinical hospital through many years of clinical tests, and finished by
service engineers.
ZView is the embedded software based on feel-TK enable to remove
noise and various artifacts and much clearer body tissue. The
embedded calibration function makes it possible to set the basic
calibration data (offset image and gain image).
It can also process progressive defect by providing defect view.
ZView is distributed and being used at the clinical fields after verifying

software on distributed worldwide. It is easy to use console system’s

software just by changing the type of the detector which DR makers

want. Integrating the X-ray’s generator operation panel, the total DR

system inside the software makes it possible to control APR, AEC,


Ready/Exposure with ease.

4
Digital Radiography Console Software ZView

ZView is integrated with various X-ray generators, supports APR


function, which enables to work as existing operation panel with just
one console software.

3. System requirements
3.1. Workstation
 Processor : Intel Pentium Intel Duo Core/Core 2 Duo or
compatible AMD Dual Core Processor or later, Win 7 : Intel i5
Second generation 2.0Ghz or above
 RAM : 2GB RAM
 HDD: At least 40GB hard disk (if possible, over 7,200 RPM). For
data backup and reliability, we recommend two separate
HDDs physically-1st HDD: System and Software, 2 HDD: Image
Data Storage, not separating the partition logically with just
one HDD.
 Network Card : 1Gbps Ethernet Network Card
 Video graphic adapter: at least over 1280 X 900 pixels, color
resolution-True Color mode(no shard memory), DVI interface
connector
 Operating System : Windows 7 32/64 bit(at least SP1),
Windows XP
prof.(at least SP2), Windows 2000 prof.(at least SP4)
 Need I/O : 1 Parallel port(synchronize Detector & X-ray
Generator & Software), 1 Serial port(in case Control of X-ray
generator), 1 USB port(Software Lock-Key)

3.2. Monitor
Over 19 inch (23 inch recommended)
 Brightness : over 300cd/m2
 Contrast Ratio : over DC 8000:1
 Response time : over 5 ms
 1600X1200 or 1920X1280 recommended (at least over 1280 X 900)

5
Digital Radiography Console Software ZView

4:3 , 16:9. 16: 10 ratio supported.


 High fidelity of grayscale and optimized luminance distribution
If you want to use high-resolution monitor, we recommend medical
LCD monitor (B/W or color). There is no problem because the present
software was made based on gray color when you want to use black
and white monitor.
For image quality control, we recommend the below among medical
monitors. It is because it needs to adjust the image at the same
brightness for diagnosis after taking X-ray. Most of monitors are made
according to each country’s medical standard so that there is no
problem to use them.

6
Digital Radiography Console Software ZView

4. Installation & Delete software

If you want to install on your new PC or reinstall after removing the old
version, please follow the below procedures.

4.1. Installation Preparation


Install detector or generator related drives after installing this software.
Insert the installation CD into CD-ROM drive or copy after making a
folder in your PC hard disk. The automatic installation program works
when using CD-ROM. If you do manually, go to the below folder and
double-click “Setup”.

Do not create a folder with other languages except English or long route
when using after copying the installation CD on your hard disk. It can make
problems when installing the driver.

7
Digital Radiography Console Software ZView

4.2. Installation of the Net Framework

If the .Net Framework 3.5 SP1 is not installed, install the .Net
framework first. If it is installed already, go to the next screen.
After installing the .Net Framework, make sure to reboot your PC. The
installation process continues after rebooting.

When you have installed the .Net Framework, you should reboot PC.
Afterward, the installation is progressing continuously.

4.3. Installation Screen


When you get the following screen, click “Next”.

8
Digital Radiography Console Software ZView

4.4. Selecting the Setup route


Specify the route of the software to be installed. Select the drive to
install from the below list and click “Next”.

The list is displayed the fixed drive, not are displayed CD-ROM or
Floppy disk. We recommend Zview to set up to Drive C:

“When you get the above message meaning the memory space is
less than 10GB, you can click “Yes” to continue. You need at least
10GB to install this software. This memory space includes created
data to be added after installation.

“When you get the above message meaning the same folder name
exists in the drive to install, you can not continue. For continuation,
you need to change the folder name in the drive as “_bak” into a
different name.

This is for the case that other software with the same name was
installed and the case that the backup was not done after removing
the previous installation.”

9
Digital Radiography Console Software ZView

4.5. Selecting the route of Data Storage

Select the route of data to be installed. From the below list, select the
drive and click “Next”.

You can also get the message like this.


When you get this message meaning the memory space is less
than 40GB, you can click “Yes” to continue. You need at least
40GB for software storage space. You can save 1,000 image
data there
When you get the above message meaning the same
folder name exists in the drive to install, you cannot
continue. For continuation, you need to change the
folder name in the drive as “_bak” into a different name.

This is for the case that other software with the same
name was installed and the case that the backup was
not done after removing the previous installation

10
Digital Radiography Console Software ZView

4.6. Selecting the type of the Detector


Select the type of the detector to use. From the below list, select the
detector and click “Next”.

Note

- If you have any other detector to add, please contact our sales staff.
-The available installed detectors are different by DR System Company.

11
Digital Radiography Console Software ZView

4.7. Selecting the type of the Generator

Select the type of the generator to use. From the below list, select the
generator and click “Next”. If you do not connect a generator, select
“DO NOT CONNECT GENERATOR” at the bottom of the list.

Note

“If you have any generator to add or connect, please contact our
sales staff.”
“ The contents of installation version is different according to
customers that’s because this version is fixed to applicable model. “
“If the model of generator is empty, please contact service staff”

12
Digital Radiography Console Software ZView

4.8. Starting the Installation


It displays all settings until now. If there is no problem, click “Next” for
installation.

13
Digital Radiography Console Software ZView

4.9. Installation of the various Drivers

You install the drivers to use in this process. It works automatically and
never stops compulsively.

If driver is failed to set up, the message will be displayed.

In this case, firstly press the OK button to continue setting up the


rest part after remembering the path of message.

The next step, press NO button in completed step of installation


and don’t reboot after finishing the installation.

By manual, reboot the system after looking for the path of drive
and finishing the installation

14
Digital Radiography Console Software ZView

4.10. Completing the Installation


After you finish installing the software, restart your computer to display
the installed software on the desktop. Please refer to how to use setup
for software setup.

15
Digital Radiography Console Software ZView

4.11. Setting the Detector & Generator


For detailed setup for each detector/generator, do on the below screen
after rebooting or starting setup.exe in the installation folder for re
modification.

4.11.1. Detector
Refer to the detector instructor of the respective manufacturer.

16
Digital Radiography Console Software ZView

4.11.2. Generator

(1) GENERATOR

Enter the workstation No. of the applicable bucky when setting up


the generator.
If it apply to connect with AUX ,Enter the applicable workstation No.
Unless it apply, enter the the workstation No. of STAND or TABLE .
Comport : Select the using serial communication port that is
communicated with generator.

17
Digital Radiography Console Software ZView

(2) IO Device for Trigger


Sometime, several detectors don’t support hardware trigger with X-
ray generator. In this case, you should have to use software trigger(if
the detector support it) with Digital IO Device. And you should have
to check a check box of Enable
i. IO Device Model
Select IO device model of PCI/LPCI/LPCIe-7250 or USB-7250.
ii. Ready Delay
This time mean that the check delay time is a starting time of
checking EXPOSUE button after pushing PREP(READY) button.
So, if you know a time of READY UP of generator, you type-in that
time value.

(3) ETC
i. 12 bit send
: To send the 12 bit dicom file must be set.
The rest part does not check this Because the raw data value is up to
14 bit. But PACS system is applies to process only 12 bit , always
check it .

ii. Interpolation
: This option is that you can choose ON/OFF of interpolation at the
time of image loading in review mode. When the images load in the

18
Digital Radiography Console Software ZView

review mode, the Zview apply interpolation for all images. So the
loading time needs more time according to image count.
But if you turn off this option, the Zview does interpolation when you
work as using windowing, panning, zooming.

iii. AutoCrop
: This function is available in the only exposure mode.
The image size is automatically modified by adjusting the collimator
size of x-ray tube.
This function is operated only one time automatically.
Also it is possible to use the manual mode after releasing the this
one.

iv. Virtual Keyboard


: It is applied to write the text and adjust the window width/level
without keyboard.

19
Digital Radiography Console Software ZView

v. Reject
: This function applies to cancel the examination or reject the
acquired image after exposure and “Reject” mark is displayed on the
image. The image data is not able to use for diagnostic purpose and
does not delete the image from the saved drive. If the check is
released, the image will be deleted with cancellation.

vi.Analysis
: If you want to know analysis or statistics for the used DR console
station, you can use this function. This function informs you of
Exposure counts, Dicom sending and rejected image counts. And you
can check the rejected analysis that is done by operator.

vii. Stitching Processed Image


: Check this function when you want to stitch the processed image at the
time. Unless you don’t want this, release the this function.

vii. Sensitivity
: Check this function when you want to display the Exposure Index or
Exposure Indicator on the image.

viii. Use Ralco


: Check this function when using the Ralco Auto collimator.
Please, refer to the additional installation manual if you need it in
detail.

20
Digital Radiography Console Software ZView

x. Show Wifi UI
When you use wireless detector, do checked this check box.
You can see the status of WIFI sensitivity and Battery.

xi. No Use Predefine shutter


Config ON/OFF about Pre-Define shutter size.

(4) DAP
i. Use DAP
Check this check box when you use DAP.

ii. Type
Select DAP Type

iii. Comport
When you connect Vacutec and iba directly, setup COM port No.

iv. Unit of DAP


Select Unit of DAP Value.

21
Digital Radiography Console Software ZView

4.11.3. System

① You can set up whether enable or disable the function of using


DICOM Worklist. This default value is using wokrlist.

② The viewsub can be used only in the state of integrated generator. If


you use one monitor, you cannot this function and you don’t need to
set the resolution of main cons
[2 monitors and integrated generator]
The main console view always use number “1” monitor. And the
viewsub always use number “2” monitor.
- PosX : Position X-coordinate of number 2 monitor or viewsub
console
- PosY : Position X-coordinate of number 2 monitor or viewsub
console

22
Digital Radiography Console Software ZView

Note

“You can find system width and height in “display set up of


Windows system”. “Start > Control panel > Display > Setup
TAB”
This subView is optional function. If you want these
functions, you call sale man.”

③ The feelVIEW is special viewer for detail seeing image using Medical
or high resolution monitor.
Pos X : X coordinates for opening feelVIEW windows
Pos Y : Y coordinates for opening feelVIEW windows
Width : Width of feelVIEW Monitor’ resolution
Height : Height of feelVIEW Monitor’ resolution

④ Rotation is set up according to detector position when image is


acquired. Rotation value must be entered to clock-direction and
choose one of the rotation value among 90,180 and270.

23
Digital Radiography Console Software ZView

4.12. Removing software


Please follow the below steps to remove any pre-installed software.

4.12.1. Data Backup


Back up the below files and folders at the route of the installed
software.
 AceView.ini
 AceView.mdb
 Cor folder
 DefaultCurve folder
Parameter Files
Other files or folders which other operators added

4.12.2. Software Removal

. From Add or Remove Programs of the Control Panel, click “Remove”


to uninstall the unnecessary software.

24
Digital Radiography Console Software ZView

4.13. Must-Do Setup of OS


4.13.1. Set up of NIC
(1) Internet Connection Sharing
The install engineer must setup for using two LAN card. This
configuration needs for priority of PACS network and Detector network.

A. Check items for setup


i. The Windows XP was installed.
ii. If it installed two network interface card in the system, the one is
interfaced or setup into PACS network or hospital network, another is
interfaced into X-ray Detector.
iii. If the IP address of interfaced the PACS or hospital network is
192.168.0.XXX, you should change IP address and default gateway IP
address (192.168.XXX.XXX)
iv. All of NIC(LAN Card) must be interface with network cable.

25
Digital Radiography Console Software ZView

B. Setup
i. Start -> Control Panel - > Network

ii. After check NIC(detector NIC or PACS NIC), you have to change

network name, ‘PACS’ and ‘DETECTOR’. After change network name,

select the PACS network and click the right button of mouse after

selecting‘properties‘.

26
Digital Radiography Console Software ZView

iii. Click the advanced.

iv. Check the box in the red window and click the “OK” to finish.

27
Digital Radiography Console Software ZView

v. After rebooting, IP is changed to 192.168.0.xxx.


It needs to change original IP address.
At the first time, IP address is 192.168.0.xxx that is connected with
hospital network or PACS .
It should change it to the assigned IP from hospital.

28
Digital Radiography Console Software ZView

(2) Adapters and Binding (Window XP, 7, 8 are common)


Select the Advanced  Advanced Settings.. in Menu
(If the Menu is hidden, it will be appeared by pressing ‘Alt+N’ Key)

Select ‘pacs’ and click the button, and then confirm the changing
connection order and click the ‘OK’ button

29
Digital Radiography Console Software ZView

30
Digital Radiography Console Software ZView

4.13.2. Window Patch for DVD


This update is a assistance for DVD burning in WindowXP operating
system. You should update patch Microsoft patch files for using DVD
burning.

* You should select a “hotfix” folder in installCD.

* First step, Install the “IMAPI_XP_SRV2003_x86.exe” file in the hotfix


folder.
* Second step, Install the “WindowsXP-KB932716-v2-x86-XXX” file in
the hotfix folder. You must check a file name that mated Window XP
Language version. For example, if your system is WindowXP English
version, you must install “WindowsXP-KB932716-v2-x86-ENU” file.

Note
“You don’t need this step on the Windows 7 Operation system”

31
Digital Radiography Console Software ZView

4.13.3. Virtual Memory Setup(Essential Setup)

 Window XP
This setup is for the stability of management of memory and store
image.

A. Click right button at icon of My Computer.

B. Click the Advanced tap.

C. Click the Settings in Performance

32
Digital Radiography Console Software ZView

D. Click advanced tap

E. Click change button in Virtual Memory

33
Digital Radiography Console Software ZView

34
Digital Radiography Console Software ZView

F. Change or setup.

i. Select HDD.(You must setup all HDD)


ii. Check in Custom size and type in , Initial size : up to 3072, Maximum
size : up to 4092.
iii. Click Setup button in paging file size for selected drive.
* Please set up the all of the used drive.
If you installed the program and DCM in C drive, set up only one drive.
But if you installed program and DCM separately, set up all drive at
the same way .

35
Digital Radiography Console Software ZView

 Window 7

36
Digital Radiography Console Software ZView

37
Digital Radiography Console Software ZView

38
Digital Radiography Console Software ZView

4.13.4. DEP
 DEP Setup in WindowsXP

39
Digital Radiography Console Software ZView

40
Digital Radiography Console Software ZView

 DEP Setup in Windows 7

41
Digital Radiography Console Software ZView

42
Digital Radiography Console Software ZView

 DEP Setup in Windows 8

- This function may not work exactly.


- There are CMOS disturbs this function to use, at that case you need to
set DEP release in OS.

i. System
Control Panel  System and Security  System

43
Digital Radiography Console Software ZView

ii. Click the Advanced system Setting and then the System Properties
window will be shown. And then move the Advanced Tab and click the
Settings button of Performance

iii. When the Performance Options window show, move the Data
Execution Prevention Tab and then select ‘Turn on DEP for all programs
and services except those I select’

iv. Click the ‘Add…’ button and select two files as like below
 Zview.exe
 WBMaker.exe

44
Digital Radiography Console Software ZView

45
Digital Radiography Console Software ZView

v. Confirm the ‘Zview’ and ‘WBMaker’ was included in List and then
click ‘Apply’ button

46
Digital Radiography Console Software ZView

4.13.5. Removal of Sleep Mode


If you are using the detector connecting by Ethernet,
To prevent, Ethernet connection is disconnected by Sleep mode. Don’t
use the Sleep mode
i. Open the ‘Control Panel’ and select ‘System and Security’.

47
Digital Radiography Console Software ZView

ii. Select ‘Power Option’

iii. Check ‘Balance(recommended)’ and click ‘Change plan settings’

48
Digital Radiography Console Software ZView

iv. Change ‘Never’ in ‘Put the computer to sleep’

49
Digital Radiography Console Software ZView

5. Appendix
5.1. DICOM MWL interface by Text file
5.1.1. Principles of interfacing

50
Digital Radiography Console Software ZView

5.1.2. Define order arument

5.1.3. Method of Interfacing


(1) Filename
- Order_Date.mwl (ex. Order_20140101123456.mwl)
(2) Store destination folder
- C:\Zview\OCS\
(3) Argument delimiter

51
Digital Radiography Console Software ZView

- “,” : You can use this character in all of Text file.


(4) Study delimiter
- CR : Carriage return
(5) Unknown value
- Blank
(6) Mandatory Argument
- Patient ID/Name/Sex/Modality/Date of Study Scheduled

5.1.4. Sample

52