Announcement of Readiness for Installation
Announcement of Readiness for
Installation
PlasmaQuant ICP-OES product family
Analytik Jena GmbH Phone +49 3641 77 70 Announcement of Readiness for Installation PlasmaQuant ICP-OES product 1/10
Konrad-Zuse-Strasse 1 [email protected] family
07745 Jena ⋅ Germany www.analytik-jena.com Version: C (11/2021)
Announcement of Readiness for Installation
Table of contents
1 Explanatory note............................................................................................................................................................. 3
2 Installation conditions .................................................................................................................................................... 4
2.1 Delivery and installation................................................................................................................................... 4
2.2 Installation and connection requirements ...................................................................................................... 4
2.3 Water cooling.................................................................................................................................................... 6
2.4 Control computer.............................................................................................................................................. 7
2.5 Installation preparations and required reagents ............................................................................................ 8
3 Device layout and space requirements .......................................................................................................................... 9
Analytik Jena GmbH Phone +49 3641 77 70 Announcement of Readiness for Installation PlasmaQuant ICP-OES product 2/10
Konrad-Zuse-Strasse 1 [email protected] family
07745 Jena ⋅ Germany www.analytik-jena.de Version: C (11/2021)
Announcement of Readiness for Installation
1 Explanatory note
The device may only be installed by our customer service or by persons authorized by Analytik Jena.
It is imperative that at least one person designated to operate the device is present during the briefing given by the ser-
vice technician.
Please provide us with confirmation that all the installation requirements have been met. We will be happy to send you
more detailed information if required.
Analytik Jena GmbH
Konrad-Zuse-Straße 1
D-07745 Jena
Phone: +49 3641/77 7407, Fax: +49 3641/77 7449
E-mail: [email protected]
With your signature you confirm that all the prerequisites for the complete installation of the device will have been ful-
filled before the arrival of our service technician.
If this is not the case and as a result additional costs are incurred (for example, it is necessary to arrange a second visit),
we reserve the right to charge for this according to our price list.
Please complete in block capitals:
Company
Department
Address
(street, number,
zip code, city)
Phone
E-mail
Customer number
Order number
Please confirm the details on this page by entering the date and your signature:
Date
Signature
Analytik Jena GmbH Phone +49 3641 77 70 Announcement of Readiness for Installation PlasmaQuant ICP-OES product 3/10
Konrad-Zuse-Strasse 1 [email protected] family
07745 Jena ⋅ Germany www.analytik-jena.com Version: C (11/2021)
Announcement of Readiness for Installation
2 Installation conditions
2.1 Delivery and installation
Guaranteed Not guaranteed
The installation location is on the first floor or accessible by a
suitable elevator.
If "no", please describe the conditions:
It is possible to reach the installation building with a trans-
porter.
If "no", please describe the conditions:
The device will be delivered directly to the final device location
by a transportation company.
The delivery by this company requires the presence of a com-
petent person.
The device must be positioned in such a way that allows easy
access from all sides. Alternatively, a heavy duty roller table
can be used.
2.2 Installation and connection requirements
Installation conditions
Complies Does not com-
ply
Table surface:
¡ 1800 mm x 750 mm (basic device + autosampler)
¡ Wipe, scratch and corrosion resistant, may not absorb moisture
¡ Load capacity min. 200 kg
¡ The rear of the device is accessible. A distance of 300 mm is maintained to the next
wall.
The installation location is free of:
¡ Direct sunlight and radiation from heaters
¡ High dust contamination
¡ Heavy mechanical oscillations and vibrations
¡ Magnetic fields
¡ Drafts
¡ Corrosive gases
Analytik Jena GmbH Phone +49 3641 77 70 Announcement of Readiness for Installation PlasmaQuant ICP-OES product 4/10
Konrad-Zuse-Strasse 1 [email protected] family
07745 Jena ⋅ Germany www.analytik-jena.com Version: C (11/2021)
Announcement of Readiness for Installation
Complies Does not com-
ply
The installation location meets the following environmental conditions:
¡ +15 °C to +35 °C, optimum +20 °C to +25 °C (air conditioning unit recommended)
¡ 20 to 90 % relative humidity at 20 °C
¡ Atmospheric pressure 0.7 bar to 1.06 bar
Electrical connection requirements
The emission spectrometer is operated on single-phase alternating current.
Complies Does not comply
Voltage 230 V ±10%
Frequency 50/60 Hz
Mains fuse (installation in the Safety fuse 32 A, slow blow, or cir-
building) cuit breaker, type C
The CEE surface box is installed.
Gas supply
"Not applicable" if the corresponding gas is not used.
Complies Does not Not
comply applica-
ble
Argon Quality ≥ 4.6
Inlet pressure 600 kPa (6 bar)
Connection Connection for hose inner diameter 4 mm
or Swagelok 6 mm with reinforcing sleeve
Oxygen (additive gas, Quality ≥ 4.5
optional)
Inlet pressure 600 kPa (6 bar)
Connection Connection for hose inner diameter 4 mm
or Swagelok 6 mm with reinforcing sleeve
The total consumption of argon is 13 to 21 l/min. The consumption of oxygen, as an optional additive gas,
is ≤ 0.04 l/min.
Analytik Jena GmbH Phone +49 3641 77 70 Announcement of Readiness for Installation PlasmaQuant ICP-OES product 5/10
Konrad-Zuse-Strasse 1 [email protected] family
07745 Jena ⋅ Germany www.analytik-jena.com Version: C (11/2021)
Announcement of Readiness for Installation
Exhaust unit
Complies Does not comply
Material Heat and corrosion resistant (rec-
ommended: V2A steel)
External pipe diameter ⌀ 125 mm
Adapter using flexible aluminum Length: approx. 1000 mm
pipe
Extraction output 3.5 m³/min … 5.5 m³/min
(optimum 4.0 to 4.5 m³/min)
The exhaust unit removes gases that are harmful to health, such as ozone or nitrous gases, which can be generated
during plasma operation. Correct extraction requires the form-fitting connection of an extraction hose to the smoke-
stack of the device.
2.3 Water cooling
A cooling system is required to operate the emission spectrometer.
¡ The cooling system can be supplied by Analytik Jena, but it must be ordered as a separate item.
¡ Alternatively, a cooling system that meets the minimum requirements can be used.
Please select an option in the table below.
Option Type/manufacturer Guaranteed
Option 1: Water-air cooler
The recirculating cooler is supplied by Analytik Water-water cooler
Jena
Option 2:
The customer provides the recirculating cooler.
The cooling system meets the minimum
requirements (see below).
The cooling system must meet the following minimum requirements:
Water flow in the cooling water circuit 1.5 to 2.0 l/min
Cooling water temperature range at device cooling water inlet 17 to 24 °C
Cooler target temperature 18 °C
Temperature stability ± 0.1 °C
Conductivity of the cooling water 50 to 200 µS/cm
Cooling performance 3000 VA
Pressure setting (max.) 600 kPa (6 bar)
Analytik Jena GmbH Phone +49 3641 77 70 Announcement of Readiness for Installation PlasmaQuant ICP-OES product 6/10
Konrad-Zuse-Strasse 1 [email protected] family
07745 Jena ⋅ Germany www.analytik-jena.com Version: C (11/2021)
Announcement of Readiness for Installation
For a water-water cooler, the customer must also comply with the following requirements for the water flow (on the
primary side):
Water volume in circulation 610 l/h (at 15 °C water temperature at the inlet side, 20 °C at
the outlet side and ∆p = 40 kPa)
If the recirculating cooler is included in the delivery from Analytik Jena, the following components are supplied:
Component Water-air cooler Water-water cooler
Water hose (on the secondary side) 3.5 m 3.5 m
Water hose (on the primary side) / 3.5 m
Connections / 0.5"
Power cable 2.7 m 2.7 m
Interface cable 3.5 m 3.5 m
(to the ICP-OES)
Minimum distance to adjacent devices 60 cm 30 cm
(please observe)
2.4 Control computer
The control computer can be supplied by Analytik Jena, but it must be ordered as a separate item. The control com-
puter must meet the following minimum requirements:
Minimum requirements for PC with Windows 8.1 or Windows 10 (32/64 Bit)
the control computer Graphics resolution 1280 x 1024
(1024 x 768 possibly with restrictions), Direct X 9. WDDM 2.0
Processor: 1.6 GHz Dual Core CPU
Working memory: 2 GB RAM (32 Bit). 4 GB RAM (64 Bit)
Free disk space: 4 GB (SSD recommended)
4 x USB 2.0 interfaces
Mouse / trackball, keyboard
A CD/DVD drive is required for installation.
Complies Does not com-
ply
Option 1:
The control computer is supplied by Analytik Jena.
Option 2:
The control computer is provided by the customer.
The control computer meets the minimum requirements.
Internet connection during installation is recommended. When connected to the Internet, updates to the operating sys-
tem as well as to additional software (Microsoft Office) are ensured. Remote access to the device is then also possible
for remote diagnosis. Customer consent is always required for remote diagnosis.
Analytik Jena GmbH Phone +49 3641 77 70 Announcement of Readiness for Installation PlasmaQuant ICP-OES product 7/10
Konrad-Zuse-Strasse 1 [email protected] family
07745 Jena ⋅ Germany www.analytik-jena.com Version: C (11/2021)
Announcement of Readiness for Installation
Operation of the analysis device without an Internet connection is ensured!
For the installation, administrator rights are required on the control computer.
If the control computer is to be integrated into a network, this must take place before installing the software. Installa-
tion within a virtual environment constitutes an exceptional case. Consultation with the Service department of Analytik
Jena or the providing partner organization is required for this.
Guaranteed Not guaranteed
Internet access is available during installation of the control software.
NET Framework 3.5 can be activated for installation.
Administrator rights are ensured during installation. Alternatively, the customer's IT de-
partment can provide support.
Network connection LAN
WLAN
Optionally, a Microsoft Office package can be included in the order. The customer must have a Microsoft ID for licens-
ing.
Yes No
The installation of an Office package is included in the order.
A Microsoft ID for the Microsoft Office license is provided.
2.5 Installation preparations and required reagents
The customer must provide the following reagents for installation and commissioning:
Complies Does not com-
ply
¡ Nitric acid HNO3 p.a. ≥ 65 %
¡ Deionized water
¡ ICP multi-element standard solution with the following elements:
Tl, Zn, Pb, Cd, Mn, Cu, Li, Mg, Ba
e.g., 111355, MERCK IV (23 elements in HNO, 3 %)
Guaranteed Not guaranteed
All the reagents required for installation are provided by the customer.
Analytik Jena GmbH Phone +49 3641 77 70 Announcement of Readiness for Installation PlasmaQuant ICP-OES product 8/10
Konrad-Zuse-Strasse 1 [email protected] family
07745 Jena ⋅ Germany www.analytik-jena.com Version: C (11/2021)
Announcement of Readiness for Installation
3 Device layout and space requirements
The emission spectrometer is a compact device designed for table-top operation. The required space depends on all the
components that make up the measuring station.
Components of the measuring station:
¡ Autosampler
¡ Cooling system (outside the laboratory or next to the workbench)
Please note! The cooler must be located on the same floor as the basic device. Place the cooler next to the work-
bench, not under it!
¡ Waste bottle (under the bench)
¡ The PC and printer may be placed on a separate side table.
Further optional components of the measuring station:
¡ Dilution system
¡ Switch valve with control unit (accessory for fast sample feeding)
Component Width x height x depth [mm] Mass [kg]
On the workbench
Basic device 990 mm x 940 mm x 855 mm 170 kg
ASPQ 3300 sampler 285 mm x 510 mm x 490 mm 15 kg
Teledyne Cetac ASX-560 sampler 580 mm x 620 mm x 550 mm 12 kg
Cetac Oils 7400 sampler 570 mm x 490 mm x 540 mm 23 kg
Teledyne Cetac SDX(HPLD) dilution system 132 mm x 254 mm x 117 mm 4.4 kg
Switch valve Cetac ASXPress Plus 58 mm x 128 mm x 217 mm 1.3 kg
with control unit 83 mm x 254 mm x 200 mm 1.4 kg
Outside the laboratory or next to the workbench
Water-air cooler 460 mm x 703 mm x 735 mm 92 kg
Water-water cooler 360 mm x 590 mm x 470 mm 33 kg (empty)
Under the workbench
Waste bottle (⌀ x height) 120 mm x 250 mm
To ensure unimpeded intake and exhaust of cooling air and effective cooling, the surfaces of the housing sides of the
water-air cooler require a minimum distance of 60 cm from adjacent objects.
Due to the waste heat generated and the possible noise pollution, it is recommended to place the water-air cooler out-
side the laboratory. An extension of the cooling water hoses is permitted if minimum pressure and flow volume are
maintained. The cooler must be located on the same floor as the basic device. If this is not the case, additional check
valves must be installed in the water circuit. Otherwise the water tank can run empty when the device is at a standstill.
These adjustments will not be carried out by Analytik Jena.
Analytik Jena GmbH Phone +49 3641 77 70 Announcement of Readiness for Installation PlasmaQuant ICP-OES product 9/10
Konrad-Zuse-Strasse 1 [email protected] family
07745 Jena ⋅ Germany www.analytik-jena.com Version: C (11/2021)
Announcement of Readiness for Installation
Figure 1 Space requirements (viewed from the front)
Figure 2 Space requirements (view from above)
Analytik Jena GmbH Phone +49 3641 77 70 Announcement of Readiness for Installation PlasmaQuant ICP-OES product 10/10
Konrad-Zuse-Strasse 1 [email protected] family
07745 Jena ⋅ Germany www.analytik-jena.com Version: C (11/2021)