MGIT 960 User Manual
MGIT 960 User Manual
System
User’s Manual
May, 1999
Document Number: MA-0117-D(prelim) Y E A R
Revision: C(prelim)
Catalog Number: 4405876
BACTEC® MGIT™ 960 System User’s Manual
Change History
Revision Date Pages Reason
B 8/98 Preface (i-iv), Sect. 5 (11-12, 17-18) Changes for software vers. 2.00
BACTEC and BBL are registered trademarks, and MGIT is a trademark of Becton Dickinson and
Company.
© Copyright Becton Dickinson and Company, 1997 – 1999. All rights reserved. No part of this
publication may be reproduced, transmitted, transcribed, stored in retrieval systems, or translated
into any language or computer language, in any form or by any means, electronic, mechanical,
magnetic, optical, chemical, manual, or otherwise, without the prior written permission of Becton
Dickinson Microbiology Systems, 7 Loveton Circle, Sparks, Maryland, 21152, United States of
America.
ii MA-0117-D(Prelim)
u Preface
Contents
1 – Introduction
1.1 System Overview .....................................................................................................................1-1
1.2 Instrument Overview ..............................................................................................................1-3
1.3 Manual Structure ....................................................................................................................1-7
1.4 Use of this Manual ..................................................................................................................1-7
1.5 Conventions ............................................................................................................................1-8
1.5.1 General ..........................................................................................................................1-8
1.5.2 Symbols Used on the Equipment ...................................................................................1-8
1.5.3 Notes, Cautions, and Warnings .....................................................................................1-9
1.6 Summary of Cautions and Warnings ......................................................................................1-9
2 – Installation
2.1 General ...................................................................................................................................2-1
2.2 Installation Policy ....................................................................................................................2-2
2.3 Instrument Specifications .....................................................................................................2-2
2.4 System Setup .........................................................................................................................2-4
2.4.1 General ..........................................................................................................................2-4
2.4.2 Setup Parameters ..........................................................................................................2-4
2.4.3 Setup for Use with the EpiCenter™ System ...................................................................2-8
2.4.4 External Connections .....................................................................................................2-8
2.5 System Startup .......................................................................................................................2-8
2.6 Software Installation ...............................................................................................................2-9
2.7 CLIA Compliance ....................................................................................................................2-9
MA-0117-D(Prelim) iii
BACTEC® MGIT™ 960 System User’s Manual
Contents
3.5 Brightness Dial ........................................................................................................................3-6
3.6 Barcode Scanner .....................................................................................................................3-6
3.7 Drawer Controls and Indicators ..............................................................................................3-7
3.7.1 Exterior Drawer Controls and Indicators ........................................................................3-7
3.7.2 Interior Drawer Indicators ..............................................................................................3-9
3.8 Circuit Breakers CB2 – CB7 ....................................................................................................3-10
3.9 Audible Tones and Alarms ....................................................................................................3-11
3.10 Printer .................................................................................................................................3-11
4 – Operation
4.1 General ...................................................................................................................................4-1
4.2 Using the Instrument Interface ...............................................................................................4-2
4.3 Handling and Preparing MGIT QC Tubes and Specimens .......................................................4-2
4.3.1 Media Quality Control ...................................................................................................4-3
4.3.2 Specimen Preparation ....................................................................................................4-3
4.4 Daily Maintenance ..................................................................................................................4-4
4.5 Entering New Tubes ..............................................................................................................4-6
4.6 Positive and Negative Specimens ............................................................................................4-8
4.6.1 Notification of Positive Tubes .......................................................................................4-8
4.6.2 Removing Positive Tubes ..............................................................................................4-8
4.6.3 Returning Positive Tubes to the Instrument for Further Testing ....................................4-9
4.6.4 Notification of Negative Tubes ......................................................................................4-9
4.6.5 Removing Negative Tubes ..........................................................................................4-10
4.7 Removing Ongoing Tubes .....................................................................................................4-11
4.8 Identifying Anonymous Tubes ..............................................................................................4-12
4.9 Resolving Station Errors ........................................................................................................4-14
4.10 Printing Reports ..................................................................................................................4-16
4.11 LIS Communications ...........................................................................................................4-18
4.12 Use with the EpiCenter System ...........................................................................................4-19
4.13 Power Failures .....................................................................................................................4-20
5 – Reference
5.1 General ...................................................................................................................................5-1
5.2 Software Menu Tree ...............................................................................................................5-2
iv MA-0117-D(Prelim)
u Preface
Contents
5.3 Display Types ........................................................................................................................5-3
5.3.1 Main Status Screen ........................................................................................................5-3
5.3.2 Main Activity Screen ......................................................................................................5-4
5.4 Icon Charts ...........................................................................................................................5-6
5.5 Reports .................................................................................................................................5-11
5.5.1 Unloaded Positives Report ...........................................................................................5-11
5.5.2 Unloaded Negatives Report .........................................................................................5-13
5.5.3 Unloaded Ongoings Report .........................................................................................5-15
5.5.4 Instrument Inventory Report .......................................................................................5-17
5.5.5 Quality Control Report .................................................................................................5-19
6 – Maintenance
6.1 General ...................................................................................................................................6-1
6.2 Routine Maintenance ..............................................................................................................6-2
6.2.1 Daily Maintenance .........................................................................................................6-2
6.2.2 Periodic Maintenance ....................................................................................................6-6
6.2.2.1 Air Filter Replacement ......................................................................................6-6
6.2.2.2 Calibrator Replacement ....................................................................................6-7
6.2.3 “As Needed” Maintenance ..........................................................................................6-10
6.2.3.1 Cleaning the Barcode Scanner Window ..........................................................6-10
6.2.3.2 Blocking a Station ...........................................................................................6-11
6.2.3.3 Unblocking a Station .....................................................................................6-12
6.2.3.4 Replacing a Barcode Label .............................................................................6-14
6.2.3.5 Decontamination .........................................................................................6-14
6.2.3.6 Reuniting Separated Mercury Columns on Thermometers ..............................6-15
6.3 Module Replacement ............................................................................................................6-16
6.3.1 General ........................................................................................................................6-16
6.3.2 Top Panel Removal ......................................................................................................6-16
6.3.3 Top Faceplate Removal ................................................................................................6-18
6.3.4 Bottom Faceplate Removal ..........................................................................................6-20
6.3.5 Barcode Scanner Replacement .....................................................................................6-20
6.3.6 Floppy Disk Drive Replacement ....................................................................................6-22
6.3.7 Liquid Crystal Display (LCD) Assembly Replacement ....................................................6-23
6.3.8 Speaker Replacement ..................................................................................................6-24
MA-0117-D(Prelim) v
BACTEC® MGIT™ 960 System User’s Manual
Contents
7 – Troubleshooting
7.1 General ...................................................................................................................................7-1
7.1.1 Instrument Service .........................................................................................................7-1
7.2 Error/Alert Messages ............................................................................................................7-2
7.3 Write Data to Disk Function ................................................................................................7-12
Glossary and Abbreviations
Appendix A – Limited Warranty
Appendix B – Replacement Parts
Appendix C – Software Update Log
Appendix D – International Contacts
Appendix E – Supplemental Procedures
E.1 Reading BBL® MGIT™ 7 ml Tubes Manually
E.2 Reprocessing Contaminated BBL® MGIT™ 7 ml Tubes
Notes
Index
Reader Comment Card
vi MA-0117-D(Prelim)
u Preface
Figures
1-1 BACTEC® MGIT™ 960 Test Technology .................................................................................1-2
1-2 BACTEC® MGIT™ 960 Instrument .........................................................................................1-4
1-3 Main Status Screen ...............................................................................................................1-5
1-4 Symbols Used on the BACTEC® MGIT™ 960 Instrument .......................................................1-8
MA-0117-D(Prelim) vii
BACTEC® MGIT™ 960 System User’s Manual
viii MA-0117-D(Prelim)
Introduction
1.1 System Overview
The BACTEC® MGIT™ 960 instrument is an in vitro diagnostic instrument designed and optimized
for the rapid detection of mycobacteria from clinical specimens (except blood and urine). Samples
are collected from patients, processed, and inoculated into BBL® MGIT™ 7 ml tubes (Mycobacteria
Growth Indicator Tube).
Microorganisms present in these specimens metabolize nutrients and oxygen in the culture. The
culture vials contain a fluorescent sensor that responds to the concentration of oxygen in the culture
medium. The instrument’s photo detectors measure the level of fluorescence, which corresponds to
the amount of oxygen consumed by organisms. Instrument detection of the presence of
microorganisms growing in the culture medium results from these fluorescence measurements.
(See Figure 1-1.)
The instrument automatically tests the tubes continuously. A row of Light Emitting Diodes (LEDs)
below the tubes illuminates, activating their fluorescent sensors. Then the instrument’s photo
detectors take the readings. A test cycle of all drawers is completed every 60 minutes. Positive
cultures are immediately flagged by an indicator light on the front of the drawer, an optional
audible alarm, and are displayed on the LCD screen.
When positive tubes are identified, the lab technologist removes them from the instrument for
confirmation of results, and for isolation and identification of the organism.
A single BACTEC® MGIT™ 960 instrument is capable of monitoring a total of 960 BBL® MGIT™ 7 ml
tubes. The tubes are arranged in three drawers, each of which holds up to 320 tubes, and are
continuously incubated. The practical capacity is typically 154 samples per week with a 6-week
protocol (115 samples per week with an 8-week protocol).
MA-0117-D(Prelim) 1–1
BACTEC® MGIT™ 960 System User’s Manual
➊
O2
➋ +
Test
Results ➐
LED Photo
Detector
➌ ➍
Computer
Raw Data
➎ Positivity
Analysis
1–2 MA-0117-D(Prelim)
!
Introduction
Each of the three drawers holds up to 320 BBL® MGIT™ 7 ml tubes. The drawers are designated
A, B, and C, from top to bottom. Each drawer contains a sample measurement module that
consists of a number of components:
The tube rack – The wells in the rack, into which tubes are inserted, are called “stations.”
The detector assembly – Tube testing is performed by a moving detector assembly that sits
below the rack. The assembly has 16 detectors, one for each row of stations. The assembly
moves from left to right and back, taking test readings for each of the 20 station columns and
the calibrator tube column at the far left.
Drawer status indicators – Three lamps are located on the front of each drawer. One indicator
lights any time a positive tube exists in the drawer; one indicator lights any time an out of
protocol negative exists in the drawer; and one indicator lights to inform you that there is an
error station in the drawer.
Station status LEDs – Indicator LEDs are located at each station. The color (red, green, or orange)
and state (on, off, or flashing) indicate the various station statuses (such as positive, negative,
error, etc.).
Barcode Scanner
A barcode scanner is located on the front of the instrument to provide the ability to scan tube
labels for specimen identification. The scanner turns on automatically whenever the system is
ready (and expecting) to scan a barcode.
Floppy Disk
The floppy disk drive is provided to enable you to update system software and to save data to
disk for troubleshooting and system diagnostic purposes.
MA-0117-D(Prelim) 1–3
BACTEC® MGIT™ 960 System User’s Manual
Floppy
Disk Barcode Scanner
Drive
On/Off Switch
1–4 MA-0117-D(Prelim)
!
Introduction
External Ports
A port on the rear of the instrument is used to connect a printer. The other ports are for Field
Service diagnostic use. A network port is for future connections to other systems. See Figure 1-4.
System Software
The system software presents a simplified user interface on the LCD Display, with picture icons to
represent all the functions, operations, setup parameters, and status conditions (see Section 5 –
Reference for charts of all icons). Routine system operations are performed by pressing the teal
soft key that corresponds to the key definition icon shown on the screen.
There are three basic types of displays:
Main Status Screen – When all the instrument drawers are closed, this screen appears. A
summary area shows the number of tubes that are positive, negative, ongoing, available, and
stations that are in error or anonymous. Also shown are the current date and time. Software
keys allow you to configure the setup parameters, perform routine daily maintenance, review
system errors, print reports, and check the instrument temperature. See Figure 1-3.
Configuration/Maintenance Screens – Accessible from the Main Status Screen, the
Configuration and Maintenance Screens allow you to set the protocol length, time and date and
their formats, audible alarm volume, instrument identification number, to select the desired
language for reports, verify the operation of all indicator lamps, block and unblock stations,
write data to a floppy disk, and to update system software.
Activity Screens – When an instrument drawer is opened, software key definitions appear that
enable you to enter new tubes; remove positive, negative, and ongoing tubes; identify
anonymous tubes; and resolve station error conditions.
MA-0117-D(Prelim) 1–5
BACTEC® MGIT™ 960 System User’s Manual
Components in the BACTEC® MGIT™ 960 instrument are selected and designed to maintain
electrical and optical integrity throughout the product’s life. All BACTEC instruments are
calibrated at the factory prior to shipment. In addition, once per hour, each detector in the
detector assembly reads the calibrator tube present in its row. After the calibrator tube is read,
the readings are used by the Built-In-Test to verify that the detection system is in calibration. As
described above, any readings for a given calibrator that exceed the expected range cause the
system to automatically block the row from use. Detector failures are reported in the
instrument’s Quality Control Report.
Calibration verification can be provided by printing the Quality Control Report
(see Section 5.5.5 – Quality Control Report). CLIA regulations state that calibration must be
performed and documented a minimum of once every six months.
1–6 MA-0117-D(Prelim)
!
Introduction
Section 1 – Introduction – provides an overview of the BACTEC® MGIT™ 960 instrument and its
uses in the mycobacteriology laboratory, its major hardware and software components. An
overview of this manual’s structure and conventions is also included.
Section 2 – Installation – gives specifications for installation of the BACTEC® MGIT™ 960
instrument and instructions for instrument setup.
Section 3 – Controls and Indicators – explains the use and meaning of all controls and
indicators of the system.
Section 4 – Operation – provides instructions for routine daily activities.
Section 5 – Reference – provides reference material on the user interface.
Section 6 – Maintenance – explains all user system maintenance, including parts replacements.
Section 7 – Troubleshooting – provides a convenient guide for identifying and correcting
system malfunctions.
The Glossary explains several instrument and computer terms used in this manual, as well as
abbreviations.
The Appendices contain warranty information, supplemental procedures, replacement parts list,
a software update form, and a listing of Becton Dickinson international contacts.
The Index provides a listing of major topics and associated page numbers.
BBL® MGIT™ 7 ml Media Package Insert – This document contains important information on the
use, storage, inoculation, performance, and limitations of barcoded BBL® MGIT™ 7 ml tubes. It
is included with each carton of tubes, and is available upon request from the Technical Services
Department.
MA-0117-D(Prelim) 1–7
BACTEC® MGIT™ 960 System User’s Manual
1.5 Conventions
1.5.1 General
Keys
The three keys that have fixed functions are the UP ARROW (Increase) key, the DOWN ARROW
(Decrease) key, and the SILENCE ALARM key. These keys are marked with symbols representing
their functions, and operate identically regardless of the active display or operation. Eight other
keys, the software (soft) keys are teal in color, and have functions that vary depending on the
active display. Each display shows icons representing the current soft key assignments at the
bottom of the screen. To perform the function represented by the icon, press the corresponding
soft key, located just below the icon defining it.
The three fixed function keys are always identified in your BACTEC® MGIT™ 960 manual by
CAPITAL LETTERS (e.g., SILENCE ALARM key). The software keys are always identified by
lowercase letters in quotes, and the words soft key (e.g., “tube entry” soft key).
!
Figure 1-4 – Symbols Used on the BACTEC® MGIT™ 960 Instrument
Top figure: Symbols for Serial Port, Remote Alarm Port, Printer Port, Keyboard Port, and Network
Port; Center figure: Symbol for electrical hazard; Bottom figure: Symbol for “refer to
accompanying documentation” (specifically, the user’s manual) for instructions
1–8 MA-0117-D(Prelim)
!
Introduction
NOTE
CAUTION
WARNING
The BACTEC® MGIT™ 960 system has been designed for the non-invasive detection of mycobacteria
so as to minimize risks associated with mycobacterial testing. However, to further reduce the risks of
accidental exposure to infectious agents, additional precautions should be taken.
It is strongly recommended that the BACTEC® MGIT™ 960 instrument be placed in the laboratory
used for routine culture of M. tuberculosis. For activities involving the propagation and manipulation
of M. tuberculosis or Mycobacterium species grown in culture, Biosafety Level 3 practice,
containment equipment, and facilities are required as recommended by CDC and NIH guidelines.
At a minimum, the instrument should be placed in a contained laboratory environment with
controlled access which has a tuberculosis exposure control plan.
The location should have surfaces which can be easily decontaminated using an appropriate topical
disinfectant.
MA-0117-D(Prelim) 1–9
BACTEC® MGIT™ 960 System User’s Manual
The instrument must not be placed in an open corridor or hallway that is accessible to the general
public or to the patient population.
It is mandatory that all system users become thoroughly familiar with all controls and indicators
before attempting to operate the instrument.
To avoid injury to fingers or hands, make sure that no one’s hands are holding a drawer handle
when another drawer is to be opened.
Specimen preparation must be done in a biological safety cabinet. Laboratory procedures involving
mycobacteria require special equipment and techniques to minimize biohazards.1
Pathogenic microorganisms including Hepatitis B Virus and Human Immunodeficiency Virus may be
present in specimens. “UNIVERSAL PRECAUTIONS”2,3 and local laboratory guidelines should be
followed in handling all items contaminated with blood or other body fluids.
In the event of tube leakage or breakage: 1) Close the instrument drawers; 2) Turn off the
instrument; 3) Vacate the area immediately; 4) Consult your facility/CDC guidelines. An inoculated
leaking or broken tube may produce an aerosol of mycobacteria, appropriate handling should be
observed.
If an inoculated tube is found to be leaking or is accidentally broken during collection or transport,
use the established procedure in your facility for dealing with mycobacterial spills. As a minimum,
“Universal Precautions” should be employed. Tube should be discarded in an appropriate manner.
An inoculated leaking or broken tube may produce an aerosol of mycobacteria; appropriate
handling should be observed.
When the system notifies you of alerts and errors, you should immediately respond to the condition.
Instrument power should be turned off, and the power cord should be disconnected, before
beginning any module replacement procedure.
One person should stand at the front of the instrument and hold the top faceplate as the four
screws are removed. If left unsupported, the faceplate can fall when the fourth screw is loosened.
If any error sub-codes other than those listed here appear, note the sub-code and contact Becton
Dickinson for assistance.
If the recommended corrective actions do not solve the problem, contact Becton Dickinson.
1 Kent, P.T., and G.P. Kubica. 1985. Public health mycobacteriology: a guide for the level III laboratory. USDHHS. Centers for
1–10 MA-0117-D(Prelim)
Installation
2.1 General
The BACTEC® MGIT™ 960 system should be installed in an area that is free from undue vibration,
direct sunlight, high humidity, dust, temperature extremes, and corrosive or explosive vapors or
gases. The system will operate within specifications in room temperatures between 19° – 30° C
(66.2° – 86° F). Relative humidity should be between 30% and 80% (non-condensing). The left, rear,
and right sides of the instrument should be placed at least four inches from any wall. Environments
which exceed these limits could adversely effect the performance of the system components.
The drawers should maintain their temperatures to within plus 1.0° or minus 2.0° C of the
temperature controller’s setting (37° C). This accuracy can be assured only if the room temperature
meets the requirements given above.
WARNINGS
CAUTION
MA-0117-D(Prelim) 2–1
BACTEC® MGIT™ 960 System User’s Manual
WARNINGS
THE BACTEC® MGIT™ 960 SYSTEM HAS BEEN DESIGNED FOR THE
NON-INVASIVE DETECTION OF MYCOBACTERIA SO AS TO MINIMIZE
RISKS ASSOCIATED WITH MYCOBACTERIAL TESTING. HOWEVER, TO
FURTHER REDUCE THE RISKS OF ACCIDENTAL EXPOSURE TO
INFECTIOUS AGENTS, ADDITIONAL PRECAUTIONS SHOULD BE
TAKEN:
• IT IS STRONGLY RECOMMENDED THAT THE BACTEC® MGIT™ 960
INSTRUMENT BE PLACED IN THE LABORATORY USED FOR ROUTINE
CULTURE OF M. TUBERCULOSIS. FOR ACTIVITIES INVOLVING THE
PROPAGATION AND MANIPULATION OF M. TUBERCULOSIS OR
MYCOBACTERIUM SPECIES GROWN IN CULTURE, BIOSAFETY LEVEL 3
PRACTICE, CONTAINMENT EQUIPMENT, AND FACILITIES ARE
REQUIRED AS RECOMMENDED BY CDC AND NIH GUIDELINES.
• AT A MINIMUM, THE INSTRUMENT SHOULD BE PLACED IN A
CONTAINED LABORATORY ENVIRONMENT WITH CONTROLLED
ACCESS WHICH HAS A TUBERCULOSIS EXPOSURE CONTROL PLAN.
• THE LOCATION SHOULD HAVE SURFACES WHICH CAN BE EASILY
DECONTAMINATED USING AN APPROPRIATE TOPICAL DISINFECTANT.
• THE INSTRUMENT MUST NOT BE PLACED IN AN OPEN CORRIDOR
OR HALLWAY THAT IS ACCESSIBLE TO THE GENERAL PUBLIC OR TO
THE PATIENT POPULATION.
2–2 MA-0117-D(Prelim)
#
Installation
Electrical Requirements
Environmental Requirements
Non-Operating Storage
Operating Conditions
MA-0117-D(Prelim) 2–3
BACTEC® MGIT™ 960 System User’s Manual
2–4 MA-0117-D(Prelim)
#
Installation
To enter configuration mode, press the soft key corresponding to the icon shown below (the
“configuration” soft key). (For a complete listing of system icons, see Section 5 – Reference.)
When you enter configuration mode, the first of several setup displays (Test Protocol Duration)
appears. Press the “configuration” soft key to advance through the setup displays, or press the “exit”
soft key to exit configuration mode.
Any changes to configuration parameters are in effect from the time of the change forward. Also
note that any changes you make cannot be “cancelled” per se – if you change a value, you must
manually change the new value back to its previous state.
Test Protocol Duration
Select the length, in days, of the automated testing protocol. The default setting is
42 days. To increase or decrease the number of days, use the UP ARROW or DOWN
ARROW key. You can choose from 1 to 56 days.
MA-0117-D(Prelim) 2–5
BACTEC® MGIT™ 960 System User’s Manual
2–6 MA-0117-D(Prelim)
#
Installation
LIS Communications
Set up parameters for communications with a Laboratory Information System (LIS).
Note that LIS Communications AND network (EpiCenter system) communications
CANNOT be enabled simultaneously. Changes to any LIS parameters require the
instrument to be rebooted to take effect. Use the “move to other field” soft key to
highlight successive fields (shown in reverse video).
Enable/Disable LIS Communications: Press UP or DOWN ARROW until | is displayed. To disable
LIS Communications, press UP or DOWN ARROW until O is displayed. When enabled, the
parameters discussed below appear on the screen.
Upload Results: Press UP or DOWN ARROW to select SOLICITED if you want the
BACTEC® MGIT™ 960 instrument to upload results to the LIS ONLY when requested by the LIS.
Select UNSOLICITED for the instrument to upload results automatically.
Consumable Tracking: Press UP or DOWN ARROW to select ENABLED if you want the
instrument to upload tube statuses of initial entry into and removal from the instrument. Select
DISABLED to withhold reporting of tube entry and removal.
Orphan Results: Press UP or DOWN ARROW to select ENABLED if you want to upload tube
results for “orphan” tubes (tubes without an associated Accession number). Select DISABLED to
withhold reporting of orphan tube statuses.
Frame Packing: Press UP or DOWN ARROW to select PACKED to upload information in multiple
records per frame. Select UNPACKED to upload one record per frame.
Baud Rate: Press UP or DOWN ARROW to select the desired baud rate. Select from 1200, 2400,
4800, 9600 (default), 19200.
Data Bits: Press UP or DOWN ARROW to select the number of data bits used in serial
communications with the LIS. Select from 7 or 8 (default).
Stop Bits: Press UP or DOWN ARROW to select the number of stop bits used in serial
communications with the LIS. Select from 1 (default) or 2.
Parity:Press UP or DOWN ARROW to select the method of parity check used in serial
communications with the LIS. Select from NONE (default), ODD, or EVEN.
MA-0117-D(Prelim) 2–7
BACTEC® MGIT™ 960 System User’s Manual
AST Setup
Enables setup of the MGIT 960 AST function. Refer to the BACTEC® MGIT™ 960 AST
Instructions for additional information. AST is not available for use in the USA.
The BACTEC® MGIT™ 960 instrument must be set up by a Becton Dickinson representative for use
with the EpiCenter system.
2–8 MA-0117-D(Prelim)
#
Installation
3 The first of five maintenance function displays appears. Press the “maintenance” soft key
three more times, until the “update software” icon appears (display 4 of 5).
The system reboots and then immediately begins to update the system software (provided the disk is
formatted, is not write-protected, and contains either the same or a later version of system
software). The names of any files being updated appear on the LCD Display, as well as a progress
indicator. When the update is complete, the user interface loads and you may proceed with normal
system operation.
MA-0117-D(Prelim) 2–9
BACTEC® MGIT™ 960 System User’s Manual
2–10 MA-0117-D(Prelim)
Controls and Indicators
3.1 General
This section describes the meaning and use of the controls and indicators of the
BACTEC® MGIT™ 960 instrument. The overall layout of the instrument cabinet is shown in
Figure 3-1. Individual components are illustrated in figures accompanying the related text.
WARNING
MA-0117-D(Prelim) 3–1
BACTEC® MGIT™ 960 System User’s Manual
Floppy
Disk Barcode Scanner
Drive
On/Off Switch
3–2 MA-0117-D(Prelim)
$
Controls and Indicators
On/Off Switch
MA-0117-D(Prelim) 3–3
BACTEC® MGIT™ 960 System User’s Manual
Display Area
Alarm Indicator
Soft Keys
3–4 MA-0117-D(Prelim)
$
Controls and Indicators
MA-0117-D(Prelim) 3–5
BACTEC® MGIT™ 960 System User’s Manual
Barcode Scanner
Brightness Dial
3–6 MA-0117-D(Prelim)
$
Controls and Indicators
WARNING
The drawer release latch is located on the rear of the teal drawer handle. The latch keeps the
drawer closed for normal operation. To open the drawer, depress the latch (toward you) and
slowly pull the drawer open. When closing a drawer, be sure you push it completely closed and
the latch locks the drawer into place. (This may be checked by gently pulling the drawer handle
toward you without depressing the release latch.)
MA-0117-D(Prelim) 3–7
BACTEC® MGIT™ 960 System User’s Manual
Negative Indicator
3–8 MA-0117-D(Prelim)
$
Controls and Indicators
Station Indicators
Station Indicators
MA-0117-D(Prelim) 3–9
BACTEC® MGIT™ 960 System User’s Manual
3–10 MA-0117-D(Prelim)
$
Controls and Indicators
Informational
Activity complete All positive tubes are removed Three short fast high beeps
Drawer closed Drawer was closed Two short fast high beeps
Alarm
3.10 Printer
For an explanation of controls and indicators on the printer, refer to the manufacturer’s operating
instructions furnished separately.
MA-0117-D(Prelim) 3–11
BACTEC® MGIT™ 960 System User’s Manual
3–12 MA-0117-D(Prelim)
Operation
4.1 General
This section describes the normal operation of the BACTEC® MGIT™ 960 system. The following
major topics are discussed:
• Preparing Specimens
• Using the Instrument Interface
• Daily Maintenance
• Entering New Tubes
• Positive and Negative Specimens
• Removing Ongoing Tubes
• Identifying Anonymous Tubes
• Responding to Alarms and Errors
• Printing Reports
• LIS Communications
• Use with the EpiCenter™ System
• Power Failures
These topics are offered in a general logical order which might fit the workflow of the average
laboratory. Some of the operations (such as printing reports and performing maintenance), may be
done at your convenience. Other operations, like monitoring the indicator lamps for new positives
and alarm conditions, should be ongoing throughout the day.
MA-0117-D(Prelim) 4–1
BACTEC® MGIT™ 960 System User’s Manual
WARNINGS
SPECIMEN PREPARATION MUST BE DONE IN A BIOLOGICAL SAFETY CABINET. LABORATORY
PROCEDURES INVOLVING MYCOBACTERIA REQUIRE SPECIAL EQUIPMENT AND TECHNIQUES
TO MINIMIZE BIOHAZARDS.1
PATHOGENIC MICROORGANISMS INCLUDING HEPATITIS B VIRUS AND HUMAN
IMMUNIDEFICIENCY VIRUS MAY BE PRESENT IN SPECIMENS. “UNIVERSAL PRECAUTIONS”2,3
AND LOCAL LABORATORY GUIDELINES SHOULD BE FOLLOWED IN HANDLING ALL ITEMS
CONTAMINATED WITH BLOOD OR OTHER BODY FLUIDS.
IN THE EVENT OF TUBE LEAKAGE OR BREAKAGE: 1) CLOSE THE INSTRUMENT DRAWERS; 2)
TURN OFF THE INSTRUMENT; 3) VACATE THE AREA IMMEDIATELY; 4) CONSULT YOUR
FACILITY/CDC GUIDELINES.
IF AN INOCULATED TUBE IS FOUND TO BE LEAKING OR IS ACCIDENTALLY BROKEN DURING
COLLECTION OR TRANSPORT, USE THE ESTABLISHED PROCEDURE IN YOUR FACILITY FOR
DEALING WITH MYCOBACTERIAL SPILLS. AS A MINIMUM, “UNIVERSAL PRECAUTIONS”
SHOULD BE EMPLOYED. TUBE SHOULD BE DISCARDED IN AN APPROPRIATE MANNER.
AN INOCULATED LEAKING OR BROKEN TUBE MAY PRODUCE AN AEROSOL OF
MYCOBACTERIA; APPROPRIATE HANDLING SHOULD BE OBSERVED.
1, 2, 3 Ibid.
4–2 MA-0117-D(Prelim)
%
Operation
MA-0117-D(Prelim) 4–3
BACTEC® MGIT™ 960 System User’s Manual
4–4 MA-0117-D(Prelim)
%
Operation
All three external indicator lamps on all three drawers should light, as well as the instrument Alarm
indicator. If any does not, contact Becton Dickinson to arrange replacement of the indicator.
Open drawer A. The soft key definitions now change to allow tests of station status LEDs. Press
the “test green LEDs” soft key:
All the green LEDs at all the stations should light. If any does not, you should block the station(s)
as described in Section 6.2.3.2 – Blocking a Station. Press the “test green LEDs” soft key again
to extinguish the green LEDs.
Now press the “test red LEDs” soft key:
All the red LEDs at all the stations should light. If any does not, you should block the station(s) as
described in Section 6.2.3.2 – Blocking a Station. Press the “test red LEDs” soft key again to
extinguish the red LEDs.
Repeat the tests of red and green station status LEDs for drawers B and C.
3 Check the thermometer readings of the Temperature QC tubes. If your manual readings are
within +1.0°/–2.0° C of 37° C, the controller and heaters are operating within the expected
range.
4 From the main status screen, press the “temperature” soft key to access drawer temperature
readings. Verify that the temperature is currently within 1.5° C of the manual reading for
each of the drawers. (More information on performing temperature verification is provided
in Section 6.2.)
MA-0117-D(Prelim) 4–5
BACTEC® MGIT™ 960 System User’s Manual
These instructions should be used to enter new BBL® MGIT™ 7 ml tubes into the instrument if the
Accession Barcoding feature (see Section 2.4.2) is disabled. If Accession Barcoding is enabled, see
the alternate instructions below.
1 Take the new cultures to the instrument. Open the desired drawer.
2 Press the “tube entry” soft key.
3 The barcode scanner turns on and the barcode icon appears in the main body of the display,
signaling that the instrument is ready to read a tube barcode sequence number. Scan the
tube’s barcode label. (To scan a tube barcode, place the tube in the alignment block in front
of the scanner with the barcode label facing the scanner. If necessary, rotate the tube
slightly so the scanner can read the label. The system beeps once to indicate a good scan.)
4 The assigned station is shown in the main body of the display (along with the scanned
sequence number). In addition, the station LEDs of the assigned station illuminate GREEN.
5 Carefully and completely insert the tube into the station.
6 Repeat Steps 3 – 5 for each of the new cultures you want to enter.
NOTE
Once the tubes are placed in stations, they should not be twisted or
turned. Tubes should not be removed except in the following conditions:
Removal of positive
Removal of negative
4–6 MA-0117-D(Prelim)
%
Operation
These instructions should be used to enter new BBL® MGIT™ 7 ml tubes into the instrument if the
Accession Barcoding feature (see Section 2.4.2) is enabled.
1 Take the new cultures to the instrument. Open the desired drawer.
2 Press the “tube entry” soft key.
3 The icon in the main body of the display shows an arrow pointing to the upper barcode (the
tube sequence number), and the barcode scanner turns on. This signals that the instrument
is ready to read a tube barcode sequence number. Scan the tube’s barcode label. (To scan a
tube barcode, place the tube in the alignment block in front of the scanner with the barcode
label facing the scanner. If necessary, rotate the tube slightly so the scanner can read the
label. The system beeps once to indicate a good scan.)
4 The icon in the main body of the display now shows an arrow pointing to the lower barcode
(the accession number), and the barcode scanner remains on. This signals that the
instrument is ready to read an accession barcode. Either scan the accession barcode label, or
press the “no accession barcode available” soft key.
5 The assigned station is shown in the main body of the display (along with the scanned tube
sequence and accession numbers). In addition, the station LEDs of the assigned station
illuminate GREEN.
6 Carefully and completely insert the tube into the station.
7 Repeat Steps 3 – 6 for each of the new cultures you want to enter.
NOTE
Once the tubes are placed in stations, they should not be twisted or
turned. Tubes should not be removed except in the following conditions:
Removal of positive
Removal of negative
MA-0117-D(Prelim) 4–7
BACTEC® MGIT™ 960 System User’s Manual
• When the drawer is opened, the “remove positive tubes” soft key appears in the soft key
definition area of the screen (see below)
• Until the condition is acknowledged, the audible alert sounds (if enabled)
3 All the positive stations illuminate with FLASHING GREEN, FLASHING RED indicators. Remove
one of the positive tubes.
4 The barcode scanner turns on and the barcode icon appears in the main body of the display,
signaling that the instrument is ready to read a tube barcode sequence number. Scan the
positive tube’s barcode label. The LEDs at this station extinguish.
5 Repeat Steps 4 – 5 to remove additional positive tubes. The POSITIVE indicator on the front
of the drawer (and at the top of the instrument) does not extinguish until all positive vials
are removed.
4–8 MA-0117-D(Prelim)
%
Operation
6 When all positive tubes are removed, the instrument beeps three times, the barcode scanner
turns off, and the “ok” icon appears in the main body of the display.
7 All instrument positive tubes should be stained for AFB and subcultured upon removal from the
instrument. Tubes should remain at room temperature while they are out of the instrument.
• Out-of-Protocol Negatives – In the summary region of the display, the tube count for each
drawer appears next to the filled circle with a minus sign icon
MA-0117-D(Prelim) 4–9
BACTEC® MGIT™ 960 System User’s Manual
3 All the final negative stations illuminate with FLASHING GREEN indicators.
4 The barcode scanner turns on and the barcode icon appears in the main body of the display,
signaling that the instrument is ready to read a tube barcode sequence number.
5 To remove all negative tubes (batch removal), press the “remove negatives – batch” soft
key. Re-move all the tubes in the indicated stations. The barcode scanner turns off, so you
cannotscan any of the tube barcode labels. Do not close the drawer until you have
removed all the tubes in the FLASHING GREEN stations. When all negative tubes are
removed, press the “ok” soft key. Be sure to remove all the tubes in the FLASHING GREEN
stations; any tubes you leave in the drawer will become anonymous at the next drawer test,
and will have to be identified before you can remove them. Once identified, they are treated
as newly entered tubes.
To remove negative tubes one at a time, remove the desired tubes and scan their barcode
labels. Continue to scan and remove all the desired individual negative tubes.
6 When all negative tubes are removed, the instrument beeps three times, and the “ok” icon
appears in the main body of the display.
4–10 MA-0117-D(Prelim)
%
Operation
3 The oldest ongoing station illuminates with FLASHING ORANGE indicators. In addition, a
display similar to the following appears (showing the same station as is flashing):
The information appearing in the main body of the display includes the station number,
barcode sequence number, accession number (if applicable), time in protocol, and current
growth unit.
4 The barcode scanner turns on and the barcode icon appears in the main body of the display,
MA-0117-D(Prelim) 4–11
BACTEC® MGIT™ 960 System User’s Manual
signaling that the instrument is ready to read a tube barcode sequence number.
5 If the station from which you want to remove the ongoing tube is the one displayed,
remove this tube and scan its barcode sequence number. If this is not the correct station,
use the UP or DOWN ARROW key to scroll through the list of ongoing stations until the
desired station appears. Then remove the desired tube and scan its barcode sequence
number.
6 Repeat Step 5 for any additional ongoing tubes you want to remove.
7 Ongoing tubes that are removed should either be subcultured to solid media (and the
BBL® MGIT™ 7 ml tube discarded according to facility guidelines), or should be incubated
offline at 37° C and manually read each day until the end of the testing protocol. (See
Appendix E – Supplemental Procedures, for information on reading tubes manually.)
These instructions should be used to identify anonymous BBL® MGIT™ 7 ml tubes in the instrument
if the Accession Barcoding feature (described in Section 2.4.2) is disabled. If Accession Barcoding is
enabled, see the alternate instructions below.
1 Open the desired drawer.
2 Press the “identify anonymous tubes” soft key.
3 The first anonymous station is shown in the main body of the display, and the ORANGE
station indicator at this station illuminates. In addition, the barcode scanner turns on and
the barcode icon appears in the main body of the display, signaling that the instrument is
ready to read a tube barcode sequence number.
4 Scan the tube’s barcode label. The scanned sequence number is now assigned to this
station. In addition, the station LEDs of the assigned station illuminate GREEN.
4–12 MA-0117-D(Prelim)
%
Operation
5 The next anonymous station is now shown in the main body of the display. Repeat Step 4 for
each of the anonymous tubes in the drawer. For anonymous tubes in other drawers, repeat
the procedure from Step 1 – 4.
To identify anonymous tubes – Accession Barcoding Enabled
These instructions should be used to identify anonymous BBL® MGIT™ 7 ml tubes in the instrument
if the Accession Barcoding feature (described in Section 2.4.2) is enabled.
1 Open the desired drawer.
2 Press the “identify anonymous tubes” soft key.
3 The first anonymous station is shown in the main body of the display, and the ORANGE
station indicator at this station illuminates.The icon in the main body of the display shows
an arrow pointing to the upper barcode (the tube sequence number), and the barcode
scanner turns on. This signals that the instrument is ready to read a tube barcode sequence
number.
4 Scan the tube’s barcode label.
5 The icon in the main body of the display now shows an arrow pointing to the lower barcode
(the accession number), and the barcode scanner remains on. This signals that the
instrument is ready to read an accession barcode. Either scan the accession barcode label, or
press the “no accession barcode available” soft key.
6 The scanned sequence number is now assigned to this station. In addition, the station LEDs
of the assigned station illuminate GREEN.
7 The next anonymous station is now shown in the main body of the display. Repeat Step 4 –
5 for each of the anonymous tubes in the drawer. For anonymous tubes in other drawers,
repeat the procedure from Step 1 – 6.
MA-0117-D(Prelim) 4–13
BACTEC® MGIT™ 960 System User’s Manual
CAUTION
When the system notifies you of alerts and errors, you should
immediately respond to the condition.
System alerts, which comprise all “E” type error codes except those numbered in the 30’s, are
reported in the system alert log. These errors cause the system alert icon to appear on the Main
Status screen and the System Alert indicator to illuminate, and can be reviewed by pressing the
“system alert” soft key (see below). The errors must be reviewed to clear the system alert condition.
Activity errors represent “E” type error codes numbered in the 30’s. These errors (such as scanning
an unexpected barcode) cause the activity error icon to appear on the activity screen (e.g., remove
positives, remove negatives, etc.). They do not put the system into an alert condition (though they
are reported in the system alert log). These errors can frequently be cleared by simply performing
the activity correctly (such as scanning the correct barcode).
Station errors (type E12) can occur from a number of causes. These errors are reported in the system
alert log, the Station Alert indicator on the drawer illuminates, and the resolve errors icon on the
Main Activity screen appears (if there are no anonymous vials). The general operation of resolving
error stations is shown in Figure 4-1. Specific suggestions for resolving error stations are provided in
Section 7.2 – Error/Alert Messages, under error code E12.
All the “E” type error codes are discussed in detail in Section 7.2 – Error/Alert Messages. The
audible tones are discussed in Section 3.9 – Audible Tones and Alarms.
4–14 MA-0117-D(Prelim)
➊
Open drawer and
MA-0117-D(Prelim)
press “resolve
station errors”
soft key
Press UP ARROW or
➌ Will the DOWN ARROW to resolve
Is the tube Can the tube
NO NO tube be
in the be returned YES next station error. When
returned
station? now? tube is ready to be
later?
returned, repeat
procedure from first step.
YES YES NO
4–15
Operation
%
BACTEC® MGIT™ 960 System User’s Manual
4–16 MA-0117-D(Prelim)
%
Operation
4 For the “unloaded tubes” reports, you are asked to confirm that the selected report printed.
or
(Note that the specific report icon appears in the blank area at the top left of the icon.)
You must press one of these two confirmation icons to print any other “unloaded tubes”
reports. If you press the “ok” soft key, the information contained in the report is
removed from the database.
The specific reports are discussed in greater detail in Section 5 – Reference.
MA-0117-D(Prelim) 4–17
BACTEC® MGIT™ 960 System User’s Manual
Enable instrument to
LIS Communications Enable or Disable
communicate with LIS
A number of technical parameters relating to communications between the two systems can
also be set in LIS Configuration. It is recommended that these parameters not be set without
consulting your local Becton Dickinson representative and your LIS vendor.
4–18 MA-0117-D(Prelim)
%
Operation
NOTE
• The BACTEC® MGIT™ 960 instrument obtains its date and time settings from the EpiCenter
system. If you need to correct the system date or time, perform the correction at the
EpiCenter™ system. The instrument will synchronize time within 10 minutes.
• Do not include any of the following characters in accession numbers entered at the
instrument:
* ? [ ] ! #
• Immediately after installation of the EpiCenter™ system, the Entry Date and Time (Start of
Protocol) of all tubes in the instrument will shift due to clock/time synchronization with the
EpiCenter system. This occurs only after new installations and only for tubes currently in the
instrument(s).
• The state of the link with the EpiCenter™ system is indicated on the Main Status screen by
one of two icons: network link okay, shown below left; and network link not okay, shown
below right.
• If the BACTEC® MGIT™ 960 instrument is connected to a LIS, it cannot be connected to the
EpiCenter™ system. However, if the BACTEC® MGIT™ 960 instrument is connected to
EpiCenter™, then a LIS connection can be established via EpiCenter.™
MA-0117-D(Prelim) 4–19
BACTEC® MGIT™ 960 System User’s Manual
If power is lost to the BACTEC® MGIT™ 960 instrument for more than 24 hours, it is strongly
recommended that all BBL® MGIT™ 7 ml tubes in the instrument be removed and incubated offline
at 37° C. These tubes must be read manually throughout the eight week protocol according to the
supplemental procedure in Appendix E.
You may, at your option, connect the system to an Uninterruptible Power Supply (UPS). Use of a UPS
may help prevent interruptions to testing.
4–20 MA-0117-D(Prelim)
Reference
5.1 General
This section presents reference material on the BACTEC® MGIT™ 960 instrument user interface. The
following information is presented:
• Software menu tree
• Display types and regions
• Icon Charts
• Reports
MA-0117-D(Prelim) 5–1
BACTEC® MGIT™ 960 System User’s Manual
5–2 MA-0117-D(Prelim)
&
Reference
MA-0117-D(Prelim) 5–3
BACTEC® MGIT™ 960 System User’s Manual
Instrument
Number
Current
Time
Summary
LIS
Enabled
Main Body
• The “tube entry” icon, which allows you to enter new BBL® MGIT™7 ml tubes for testing.
• The “remove positive tubes” icon. This icon appears only if there are positive tubes in the
instrument. It allows you to remove these positive tubes.
• The “remove negative tubes” icon. This icon appears only if there are final (out-of-protocol)
negative tubes in the instrument. It allows you to remove these negative tubes.
• The “identify anonymous tubes” icon. This icon appears only if there are anonymous tubes
in the instrument. It allows you to identify these anonymous tubes so that the system can
apply the correct positivity criteria and display any underlying statuses.
• The “resolve station errors” icon. This icon appears only if there are error stations in the
instrument and there are no anonymous tubes. It allows you to resolve the error stations.
• The “remove ongoing tubes” icon. This icon appears only if there are ongoing negative
tubes in the instrument. It allows you to remove these ongoing tubes.
In addition, system and station status information is always in view at the top of the screen. The
following information is presented:
• Instrument number
• Current date
5–4 MA-0117-D(Prelim)
&
Reference
• Current time
• LIS Indicator (if LIS Communications is enabled)
• Last station assigned during “tube entry” operation
• Summary region, showing status of drawers, number of tubes that are positive, negative,
ongoing, available, and stations that are in error, anonymous, or blocked (note that
summary counters may total greater than 960 because tubes can have multiple statuses)
The main body of the display presents icons that guide you through the operation or setup you are
performing.
The soft key assignment area (very bottom of the screen) shows the current soft key definitions.
These definitions change as you access different functions and screens. The Main Activity screen
icons are described in the first set of bullets above.
Main Body
MA-0117-D(Prelim) 5–5
BACTEC® MGIT™ 960 System User’s Manual
LIS active
Blocked station
5–6 MA-0117-D(Prelim)
&
Reference
Temperature display
Configuration Icons
Accession barcoding
Japanese language
enable/disable
LIS enable/disable
MA-0117-D(Prelim) 5–7
BACTEC® MGIT™ 960 System User’s Manual
Power was removed at this time Power was restored at this time
Maintenance Icons
5–8 MA-0117-D(Prelim)
&
Reference
Action/Navigation Icons
MA-0117-D(Prelim) 5–9
BACTEC® MGIT™ 960 System User’s Manual
Activity Icons
5–10 MA-0117-D(Prelim)
&
Reference
5.5 Reports
Five instrument reports are available for printing. The reports are as follows:
• Unloaded Positives – a list of all positive tubes removed from the system since the last time
the report was printed and confirmed (up to a maximum of 500 unloaded tubes)
• Unloaded Negatives – a list of all negative tubes removed from the system since the last time
the report was printed and confirmed (up to a maximum of 500 unloaded tubes)
• Unloaded Ongoings – a list of all ongoing (in protocol) tubes removed from the system since
the last time the report was printed and confirmed (up to a maximum of 500 unloaded
tubes)
• Instrument Inventory – a list of all current station assignments (excluding Available stations),
either for the whole instrument or for individual drawers
• Quality Control – a list of the status of all the detectors in the instrument with the date and
time of their last verification, as well as a list of blocked stations
Each of the reports is discussed in greater detail in the sections that follow.
MA-0117-D(Prelim) 5–11
BACTEC® MGIT™ 960 System User’s Manual
5–12 MA-0117-D(Prelim)
&
Reference
MA-0117-D(Prelim) 5–13
BACTEC® MGIT™ 960 System User’s Manual
5–14 MA-0117-D(Prelim)
&
Reference
MA-0117-D(Prelim) 5–15
BACTEC® MGIT™ 960 System User’s Manual
5–16 MA-0117-D(Prelim)
&
Reference
From this point, additional soft keys allow you to select which drawer(s) to report.
A sample Instrument Inventory Report is shown in Figure 5-6.
The Instrument Inventory Report provides the following information:
Report Header
Title (Instrument Inventory Report)
Instrument number
Current date and time
Current drawer temperatures
Current system software version
Page number
Main Body of Report
Tube Position – in the form D/RCC, where D is the drawer letter (A, B, or C), R is the Row
(A through S, omitting I, O, and Q),and CC is the column (01 to 20). This is the sort order
of the report.
Accession Number – the scanned accession barcode number, if one exists
Sequence Number – the scanned tube barcode sequence number
Growth Unit – the last growth unit reading
Tube Status – = for ongoing, + for positive, +T for threshold positive, – for negative, ?
for anonymous, ! for error
TIP/TTD – the time in protocol or time to detection, in the form D;HH, where D
represents the number of days and HH the number of hours from the start of protocol
(when the first instrument test occurred) to the time the tube was removed from the
instrument or declared positive. If asterisks appear in this field, it indicates that the value
is not available.
Protocol Length – the number of days of the assigned testing protocol
Start of Protocol – the date and time of the first instrument test
MA-0117-D(Prelim) 5–17
BACTEC® MGIT™ 960 System User’s Manual
5–18 MA-0117-D(Prelim)
&
Reference
MA-0117-D(Prelim) 5–19
BACTEC® MGIT™ 960 System User’s Manual
Detector Status
Row Sensor / Row Test
Designation Status Date/Time
A/A ' 10/17/96 12:00
A/B X 10/17/96 12:00
A/C ?
. . .
. . .
. . .
C/S ' 10/17/96 12:00
5–20 MA-0117-D(Prelim)
Maintenance
6.1 General
The BACTEC® MGIT™ 960 instrument requires minimal maintenance from the user to provide
reliable performance. Daily verifications include: checking the incubator temperature, printer paper
supply, and verifying indicator lamp operation. Routine preventive maintenance consists only of a
monthly changing or cleaning of the cabinet air filters and replacement of calibrator vials when they
expire. All other procedures are on an “as needed” basis. Any maintenance or repair not described in
this section should be performed by Becton Dickinson personnel only.
WARNING
MA-0117-D(Prelim) 6–1
BACTEC® MGIT™ 960 System User’s Manual
6–2 MA-0117-D(Prelim)
(
Maintenance
All three external indicator lamps on all three drawers should light, as well as the instrument
Alarm indicator. If any does not, contact Becton Dickinson to arrange replacement of the
indicator.
Open drawer A. The soft key definitions now change to allow tests of station status LEDs.
Press the “test green LEDs” soft key:
All the green LEDs at all the stations should light. If any does not, you should block the sta-
tion(s) as described in Section 6.2.3.2 – Blocking a Station. Press the “test green LEDs” soft
key again to extinguish the green LEDs.
Now press the “test red LEDs” soft key:
MA-0117-D(Prelim) 6–3
BACTEC® MGIT™ 960 System User’s Manual
All the red LEDs at all the stations should light. If any does not, you should block the sta-
tion(s) as described in Section 6.2.3.2 – Blocking a Station. Press the “test red LEDs” soft
key again to extinguish the red LEDs.
Repeat the tests of red and green station status LEDs for drawers B and C.
3 Three thermometers in special tubes are provided for temperature verification (QC) of
instrument incubation. It is at the discretion of each laboratory to determine if they wish to
perform temperature QC, at what frequency, and which stations. Three thermometers/tubes
enable you to place one in each drawer, if desired.
DO NOT FILL ANY OF THE TUBES WITH LIQUID. The most accurate readings of drawer incuba-
tion are obtained without liquid in the special tube.
It is recommended that the accuracy of the thermometer be verified against a calibrated lab
thermometer to insure the validity of the temperature verification.
Note that drawer openings and closings can affect the temperature of tubes, including the
temperature QC tube. It is strongly recommended that temperature verification be per-
formed first thing in the morning. If not, select a time when the drawers have remained
closed for a few hours. This will provide the most accurate readings of tube temperatures.
Select the drawer to be verified and the station into which you want to place the tempera-
ture QC tube. To prevent the system from trying to test the temperature QC tube in this sta-
tion, you should block it from use. See Section 6.2.3.2 – Blocking a Station, for
instructions on how to perform this activity.
After the desired station is blocked, place the tube/thermometer into the station. The ther-
mometer is already inserted in the tube and should not be removed – temperatures can be
read through the tube.
If your manual reading is within +1.0°/–2.0° C of 37° C, the controller and heaters are oper-
ating within their specifications. The instrument’s temperature controller is designed to
maintain cabinet temperatures within this range.
If you move the tube/thermometer to a different station, remember to unblock the current
station (Section 6.2.3.3 – Unblocking a Station) and to block the new station you select.
4 From the Main Status screen, press the “temperature” soft key to view drawer temperature
readings. Verify that the drawer temperatures are currently within 1.5° C of the manual
readings for each of the drawers. If the readings are not within this range, refer to the
instructions in Section 7 – Troubleshooting.
6–4 MA-0117-D(Prelim)
(
Maintenance
MA-0117-D(Prelim) 6–5
BACTEC® MGIT™ 960 System User’s Manual
WARNING
Change or clean the air filters behind the bottom faceplate monthly. If the instrument’s environment
is especially dusty, check the filters more frequently. These filters must remain clean and
unobstructed; restricted air flow may cause the instrument interior to reach excessive temperatures,
which can affect organism recovery and possibly cause hardware malfunctions or failures. The filters
can be cleaned and reused. See Figure 6-2.
Required Material
New or clean air filters
To Replace the Air Filters
1 The air filters are located behind the instrument’s bottom faceplate. To remove the
faceplate, grasp the faceplate along its bottom edge at the finger holes. Gently but firmly
pull straight out. The faceplate will pop free with medium force.
2 To remove a filter, lift it up slightly, then pivot the bottom outward. Lower the filter out of
the filter housing.
3 Wash the filter in a mycobactericidal disinfectant. Place it on a paper towel and dry it
thoroughly (if you are going to reuse it immediately).
4 Replace the filter by sliding the top up into the filter housing. Pivot the bottom in toward
the instrument, and lower the filter into place.
5 To replace the faceplace, hold it in place (the cutout should surround the On/Off Switch) and
firmly press in toward the instrument. The faceplate will snap into place.
6–6 MA-0117-D(Prelim)
(
Maintenance
Press the “maintenance” soft key four more times, until the maintenance display numbered
5/5 (replace calibrator tubes) appears. The following icon is shown in the main body of the
display:
5 Open the desired drawer. If you are replacing all the calibrators in the drawer, make sure the
display reads X/*00 (where X is the drawer letter) and press the “ok” soft key. If you are
replacing only one (or more than one, but less than all) calibrator tube, use the UP ARROW
or DOWN ARROW key to change the row value that is displayed. When the correct row
appears, press the “ok” soft key:
or
Repeat the selection process if more than one (but less than all) calibrator tube is replaced.
This step informs the system software that one or more calibrators have been replaced.
6 Close the drawer. Repeat Step 5 for the other drawers if necessary.
7 Press the “exit” soft key when complete.
MA-0117-D(Prelim) 6–7
BACTEC® MGIT™ 960 System User’s Manual
6–8 MA-0117-D(Prelim)
(
Maintenance
Lift 4 black
push-pull
fasteners
MA-0117-D(Prelim) 6–9
BACTEC® MGIT™ 960 System User’s Manual
6–10 MA-0117-D(Prelim)
(
Maintenance
3 Press the “maintenance” soft key again, until the maintenance display numbered 2/5 (block/
unblock stations) appears. The following icon is shown in the main body of the display:
MA-0117-D(Prelim) 6–11
BACTEC® MGIT™ 960 System User’s Manual
4 Open the desired drawer. The “block station” soft key now appears in the soft key definition
area.
5 Press the “block station” soft key. The first unblocked station in the drawer now appears in
the main body of the display. Initially, the column value of the station is highlighted. If the
column value is not correct, press the UP ARROW or DOWN ARROW key until the correct
column number appears. If the row value is not correct, press the “move to other field” soft
key to highlight the row value. Then press the UP ARROW or DOWN ARROW key to change
the row value until the correct row number appears.
6 When the correct station number is displayed, press the “block station” soft key to block the
station. Repeat Step 5 if neccessary to block additional stations.
7 If necessary, insert a bad station plug into any stations you block to prevent their inadvertent
use.
8 When all the desired stations are blocked, press the “exit” soft key.
6.2.3.3 Unblocking a Station
Unblocking a station returns a blocked station (which was unavailable for testing) to service. Note
that stations that were blocked automatically by the system (e.g., when a detector fails) cannot be
unblocked unless the detector or calibrator is replaced. Stations that you blocked manually (such as
to place a temperature QC tube in it) can be unblocked through the Unblock Station function.
To Unblock a Station
1 Make sure all the drawers are closed. When the drawers are closed, the Main Status screen
appears:
6–12 MA-0117-D(Prelim)
(
Maintenance
3 Press the “maintenance” soft key again, until the maintenance display numbered 2/5 (block/
unblock stations) appears. The following icon is shown in the main body of the display:
4 Open the desired drawer. The “unblock station” soft key now appears in the soft key
definition area.
5 Press the “unblock station” soft key. The first blocked station in the drawer now appears in
the main body of the display. Initially, the column value of the station is highlighted. If the
column value is not correct, press the UP ARROW or DOWN ARROW key until the correct
column number appears. If the row value is not correct, press the “move to other field” soft
key to highlight the row value. Then press the UP ARROW or DOWN ARROW key to change
the row value until the correct row number appears.
6 When the correct station number is displayed, press the “unblock station” soft key to
unblock the station. Repeat Step 5 if neccessary to unblock additional stations.
7 Be sure to remove any bad station plugs you may have previously inserted for stations you
are unblocking.
8 When all the desired stations are unblocked, press the “exit” soft key.
MA-0117-D(Prelim) 6–13
BACTEC® MGIT™ 960 System User’s Manual
WARNING
If a leakage or spill is limited to the surface of the drawer tube holder, the top surface can be cleaned
with mycobactericidal disinfectant. It is recommended that the instrument be turned off during any
cleaning activity to reduce the potential of aerosol production.
If a leakage or spill is discovered in any drawer and the instrument has not yet been turned off,
immediately turn off the instrument power switch. Close the affected drawer, follow the previous
recommendations as stated in the WARNING box, and contact Becton Dickinson.
Documentation of decontamination must be presented to a Becton Dickinson representative upon
request prior to any service.
6–14 MA-0117-D(Prelim)
(
Maintenance
MA-0117-D(Prelim) 6–15
BACTEC® MGIT™ 960 System User’s Manual
WARNING
6–16 MA-0117-D(Prelim)
(
Maintenance
• Unscrew 4 screws
• Pull panel rearward and lift off
(Rear of Instrument)
MA-0117-D(Prelim) 6–17
BACTEC® MGIT™ 960 System User’s Manual
CAUTION
One person should stand at the front of the instrument and hold the
top faceplate as the four screws are removed. If left unsupported, the
faceplate can fall when the fourth screw is loosened.
5 Replace the top faceplate by reversing the above steps. When plugging cables back into the
Control Electronics Module, be sure you carefully align the cable connectors with the
connectors on the board, and that you plug the correct cables into the correct connectors.
6–18 MA-0117-D(Prelim)
(
Maintenance
Unplug 6 cables
(Toward front
of instrument)
MA-0117-D(Prelim) 6–19
BACTEC® MGIT™ 960 System User’s Manual
6–20 MA-0117-D(Prelim)
(
Maintenance
Unplug cable
(Toward front
of instrument)
Unscrew 2 screws
MA-0117-D(Prelim) 6–21
BACTEC® MGIT™ 960 System User’s Manual
Loosen 3 screws
Unplug 2 cables
6–22 MA-0117-D(Prelim)
(
Maintenance
Remove 4 screws
MA-0117-D(Prelim) 6–23
BACTEC® MGIT™ 960 System User’s Manual
Unplug
cable
(Toward front
of instrument)
6–24 MA-0117-D(Prelim)
(
Maintenance
MA-0117-D(Prelim) 6–25
BACTEC® MGIT™ 960 System User’s Manual
6–26 MA-0117-D(Prelim)
Troubleshooting
7.1 General
7.1.1 Instrument Service
If your BACTEC® MGIT™ 960 instrument malfunctions or operates unusually in any way, you may
initially attempt to solve the problem by following the recommendations in this section. All other
servicing attempts will terminate the responsibility of the manufacturer under the terms of the
warranty.
If you cannot repair a system malfunction, contact your local Becton Dickinson representative
(contact numbers are listed in Appendix D).
This section discusses error messages, which appear when the system has encountered a known
problem. These messages are listed in numerical order, along with possible causes of the message
and corrective actions.
MA-0117-D(Prelim) 7–1
BACTEC® MGIT™ 960 System User’s Manual
CAUTION
When the system notifies you of alerts and errors, you should
immediately respond to the condition.
When the system encounters an alert or error condition, the error code (EXX, where XX is a two-digit
number) is either displayed on the screen or written into the system alert log. The error code is an
abbreviation for the conditions described in the listing below.
Codes in the E30 series (E30, E31, etc.) are displayed on the screen when they occur. (They also cause
the Activity Error tone to sound [sequence of short high beep and short low beep repeated four
times].) These are activity (or “workflow”) types of errors. In most cases, this means that some action
you have performed was not what the system expected, but you can usually perform the correct
action, as recommended below, without exiting the current operation. These activity errors are
flagged by the Activity Error icon:
System alerts, which comprise all error codes except those in the E30 series, are reported in the
system alert log. These errors cause the Alert tone (medium beep on for one second, off for 3
seconds, repeating) to sound (if it is enabled). Also the System Alert icon appears on the Main Status
screen. The errors must be reviewed to clear the system alert condition. The system alert log can be
viewed from the Main Status screen by pressing the “system alert” soft key:
The error messages are listed in numerical order. Error sub-codes are 8-digit numeric codes that
appear below the EXX readout on the system alert log. The sub-codes indicate specific conditions
detected. Many sub-codes are listed in the “Possible Causes” and “Corrective Actions” sections
below.
CAUTION
If any error sub-codes other than those listed here appear, note the
sub-code and contact Becton Dickinson for assistance.
If the recommended corrective actions do not solve the problem,
contact Becton Dickinson.
7–2 MA-0117-D(Prelim)
)
Troubleshooting
The software algorithms that determine tube positivity have been reset. This message is
informational.
POSSIBLE CAUSE(S)
• Another E type of error code (such as E02, E05, etc.) precedes the occurrence of E01.
CORRECTIVE ACTION(S)
* E01 is an informational message, and requires no corrective action per se. It is important
that whenever an error condition occurs, you acknowledge the error by reviewing the
system alert log, and take the recommended corrective action (if any).
An incubator or drawer temperature is too high (02, 08, 20, or 40) or too low (01, 04, 10). Any of
these numbers can appear in the 8-digit subcode in any of the following positions: zzyyxxww, where
zz represents a condition for Drawer C, yy a condition for Drawer B, xx a condition for Drawer A, and
ww a condition for the cabinet air. (For example, code E01 might show subcode 01202000, meaning
there is a high temperature condition in Drawers C, B, and A.)
POSSIBLE CAUSE(S)
• Drawer was kept open too long.
• Room temperature is not within recommended range, or other environmental specification
is not being met (such as instrument sitting in direct sunlight).
• Air filters are dirty, restricting fresh air intake.
CORRECTIVE ACTION(S)
* Check current air and plate temperatures on LCD display (Configuration display #2) to see if
temperature is still too high or too low.
* Minimize number and duration of drawer openings.
* Make sure environmental specifications are met (see Section 2 – Installation).
* Clean or replace air filters.
* Check temperature QC tubes to see if manual readings agree with displayed readings.
POSSIBLE CAUSE(S)
• System encountered a software error.
CORRECTIVE ACTION(S)
* Write data to disk (see Section 7.3) and call Becton Dickinson.
MA-0117-D(Prelim) 7–3
BACTEC® MGIT™ 960 System User’s Manual
A calibrator tube reading is out of tolerance (see chart to determine which calibrator). Below the E05
readout, three sub-codes are shown. The first row of sub-codes represents Drawer A, the second
row of sub-codes represents Drawer B, and the third row of sub-codes represents Drawer C. If the
sub-code 00000000 appears in the row, there is no problem in that drawer. If any numbers appear
in the row, refer to the chart below to determine which row within the drawer has caused the
problem. If a number other than those in the chart appears, contact Becton Dickinson.
SAMPLE CALIBRATION ERROR READOUT
E05 (Calibration error code)
00000001 (Sub-code for Drawer A; per chart, indicates Row A failure)
00000000 (Sub-code for Drawer B; indicates no calibrator failure)
00000040 (Sub-code for Drawer C; per chart, indicates Row G failure)
POSSIBLE CAUSE(S)
• Calibrator tube failure.
• Debris or spilled media inside instrument is affecting calibration readings.
CORRECTIVE ACTION(S)
* Check that there is no debris or spilled media inside instrument. Clean if necessary.
* If error recurs, replace calibrator tube at the row as indicated by the sub-code in the chart
below.
POSSIBLE CAUSE(S)
• Detector assembly motion was impeded.
• One of the station, home, or end flags was not read by the detector.
CORRECTIVE ACTION(S)
* Look for and if necessary remove any object impeding motion of detector assembly.
7–4 MA-0117-D(Prelim)
)
Troubleshooting
POSSIBLE CAUSE(S)
• Temporary electrical anomaly.
CORRECTIVE ACTION(S)
* Reboot instrument.
POSSIBLE CAUSE(S)
• Temporary electrical anomaly.
• Line voltage is out of spec.
• Air filters are dirty, restricting fresh air intake.
CORRECTIVE ACTION(S)
* Reboot instrument.
* Check/correct line voltage to instrument.
* Clean or replace air filters.
POSSIBLE CAUSE(S)
• Instrument has been off for longer than 4 hours.
• Four consecutive test cycles were missed (e.g., due to drawer openings).
• System clock was set more than 4 hours ahead.
CORRECTIVE ACTION(S)
* 20000000 – Investigate why no tests have occurred in 4 hours. If this error occurs and none
of the above events happened, use the “write data to disk function” (see Section 7.3) and
contact Becton Dickinson.
POSSIBLE CAUSE(S)
• Database checksum test failed.
CORRECTIVE ACTION(S)
* Write data to disk (see Section 7.3) and call Becton Dickinson.
* 00000002, 00000008 – All tubes should be incubated offline and read manually according
to the procedure in Appendix E.
MA-0117-D(Prelim) 7–5
BACTEC® MGIT™ 960 System User’s Manual
POSSIBLE CAUSE(S)
• Printer paper is jammed or exhausted.
• Printer cable is disconnected during printing.
• Printer power is turned off during printing.
• Printer was taken offline during printing.
CORRECTIVE ACTION(S)
20000000
* Check paper and clear jam or add paper if necessary.
* Check printer cables (power and communications), reattach if necessary.
* Turn printer power on.
* Place printer online.
* Request report again. It will not resume printing automatically when error condition is
corrected. .
NOTE
POSSIBLE CAUSE(S)
00000001, 00000002, 00000004, 00000008, 00000020
• System cannot detect a tube in a station where one should be. The tube may not be seated
in the station completely, or may have been pulled out of the instrument without being
scanned out.
• Tube is in a row in which a calibrator tube failed.
CORRECTIVE ACTION(S)
* If tube is in station, or was removed but you have now located it, use the “resolve station
errors” activity (Section 4.9) to scan the tube barcode label. If the station was in a
“hardware” error condition (such as with errors E05 and E06), the current station will be
blocked from use and the tube will be relocated to a new station (if one is available). If the
station was in a “software” error condition (error E12 with any of the sub-codes listed
above), the error condition is cleared after you scan the barcode label and confirm the error
resolution.
* If the tube cannot be located, press the “force station available” soft key. This forces the
tube’s protocol to be terminated and clears the error condition. Tubes that are cleared in
this manner are given the following status on reports: tubes with a status of positive are
reported as positives; tubes with a status of negative are reported as negatives; tubes with a
status of ongoing are reported as errors.
7–6 MA-0117-D(Prelim)
)
Troubleshooting
* Always be sure tubes are fully seated when placing them into stations.
* To avoid station errors, always use one of the tube removal activities (such as remove
positive tubes or remove negative tubes) to scan the tube out of the station when you
remove it. .
POSSIBLE CAUSE(S)
• Power was removed from instrument.
CORRECTIVE ACTION(S)
* Message is informational. If multiple power failures have occurred, only the latest one is
reported in the alert list. Note the power failure and restore times in your instrument log.
* If power is lost to the BACTEC® MGIT™ 960 instrument for more than 24 hours, it is strongly
recommended that all BBL® MGIT™ 7 ml tubes in the instrument be removed and incubated
offline at 37° C. These tubes must be read manually throughout the eight week protocol
according to the supplemental procedure in Appendix E.
POSSIBLE CAUSE(S)
• During the remove positive tubes, remove negative tubes, or resolve station errors activities,
the tube barcode you scanned is not the one the system expected for the station. Either you
pulled a tube from a different station than the one specified; or more than one tube has
been placed in the wrong station.
CORRECTIVE ACTION(S)
* If the station is still displayed on the activity screen, verify that you pulled the tube from the
specified station. If you did not (you will know because a tube is still in that station), press
the “exit” soft key, then press the “tube entry” soft key. Scan the tube barcode label and
place the tube where the system indicates. If there is a tube in the indicated station, do
not remove that tube to “swap” locations. Go to + below.
* If you pulled the tube from the specified station, then multiple tubes may be misplaced. You
must determine how many tubes are misplaced, and where they are located. It may be
helpful to print an Instrument Inventory Report for the drawer, add a column labeled “actual
station,” and relabel the existing station column as “assigned station.” If you prefer, you can
draw a chart, with one column labeled “tube sequence number,” a second column labeled
“assigned station,” and a third column labeled “actual station.”
Write down the tube sequence number of the tube that was just removed. Write the station
specified on the activity screen in the “actual station” column.
Press the “exit” soft key, then press the “tube entry” soft key. Scan the tube barcode label,
and write the station in the “assigned station” column of your chart. There will be a tube in
this station.
+ Place the first misplaced tube aside – its results must be confirmed manually (either
through AFB stain or reading manually as described in Appendix E). Tube test results cannot
be transferred reliably among multi-station misplacements.
MA-0117-D(Prelim) 7–7
BACTEC® MGIT™ 960 System User’s Manual
Remove the tube from the station currently displayed on the tube entry screen. Write this
tube sequence number on the chart, as well as the actual station. Now scan the current tube
barcode label. Write the station now displayed in the “assigned station” column. Place this
tube aside for manual confirmation/reading.
Go to the station now displayed. If there is a tube in this station, repeat the steps in the
previous paragraph. Continue to do so until the actual station matches the assigned station,
or until there is no tube in the assigned station. When one of these conditions occurs, you
have probably reached the end of the tube misplacements.
* All tubes involved in misplacement scenarios must be confirmed manually (either stained for
AFB or incubated offline and read manually as described in Appendix E). Keep the tubes at
room temperature until the next test cycle, when the station errors are flagged.
* When the next test cycle or drawer scan occurs, any stations from which tubes were
removed will go into error. Resolve all the station errors as described above under error E12
and in Section 4.9 after the misplaced tubes have been tested manually.
POSSIBLE CAUSE(S)
• Floppy disk is not inserted.
• Floppy disk is not formatted.
• Floppy disk is write-protected.
• Floppy disk is full.
CORRECTIVE ACTION(S)
* Insert formatted floppy disk.
* Move write-protect tab toward center of floppy disk.
* Use a new floppy disk.
POSSIBLE CAUSE(S)
• During Tube Entry or Resolve Station Errors, the system tried to allocate a station in the
current drawer but found that none are available.
CORRECTIVE ACTION(S)
* Try entering the tube in another drawer (with available stations). If there are no available
stations in other drawers, remove final negative tubes (or optionally, ongoing tubes near the
end of protocol) if any exist. If room cannot be made, the tube should be tested manually.
7–8 MA-0117-D(Prelim)
)
Troubleshooting
POSSIBLE CAUSE(S)
• The system has detected a tube in a station that does not have a tube assigned to it (i.e., an
anonymous tube). When you are identifying anonymous tubes, the system does not expect
a known barcode to be scanned (if it knows the barcode, that means the tube was
previously scanned in and assigned to a different station). Scanning a known barcode during
the identify anonymous activity causes this error to occur. The two main causes of the error
are: 1) you pulled a tube from a different station than the one specified in the identify
anonymous screen; or 2) more than one tube has been placed in the wrong station.
CORRECTIVE ACTION(S)
* To correct the error condition, you have to determine: 1) whether you pulled the wrong tube
or tubes have been misplaced; 2) if tubes are misplaced, where does the instrument think
they should be; and 3) where tubes are actually located.
* If the station is still displayed on the identify anonymous screen, verify that you pulled the
tube from the specified station. If you did not (you will know because a tube is still in that
station), press the “exit” soft key, then press the “tube entry” soft key. Scan the tube
barocde label and place the tube where the system indicates. If there is a tube in the
indicated station, do not remove that tube to “swap” locations. Go to + below.
* If you pulled the tube from the specified station, then multiple tubes may be misplaced. You
must determine how many tubes are misplaced, and where they are located. It may be
helpful to print an Instrument Inventory Report, add a column labeled “actual station,” and
relabel the existing station column as “assigned station.” If you prefer, you can draw a chart,
with one column labeled “tube sequence number,” a second column labeled “assigned
station,” and a third column labeled “actual station.”
Write down the tube sequence number of the tube that was just removed. Write the station
specified on the identify anonymous screen in the “actual station” column.
Press the “exit” soft key, then press the “tube entry” soft key. Scan the tube barcode label,
and write the station in the “assigned station” column of your chart. There will be a tube in
this station.
+ Place the first misplaced tube aside – its results must be confirmed manually (either
through AFB stain or reading manually as described in Appendix E). Tube test results cannot
be transferred reliably among multi-station misplacements.
Remove the tube from the station currently displayed on the tube entry screen. Write this
tube sequence number on the chart, as well as the actual station. Now scan the current tube
label. Write the station now displayed in the “assigned station” column. Place this tube aside
for manual confirmation/reading.
Go to the station now displayed. If there is a tube in this station, repeat the steps in the pre-
vious paragraph. Continue to do so until the actual station matches the assigned station, or
until there is no tube in the assigned station. When one of these conditions occurs, you have
probably reached the end of the tube misplacements.
* All tubes involved in misplacement scenarios must be confirmed manually (either stained for
AFB or incubated offline and read manually as described in Appendix E). Keep the tubes at
room temperature until the next test cycle, when the station errors are flagged.
MA-0117-D(Prelim) 7–9
BACTEC® MGIT™ 960 System User’s Manual
* When the next test cycle or drawer scan occurs, any stations from which tubes were
removed will go into error. In addition, the anonymous station that began the problem is
still anonymous. Go to the identify anonymous activity, and recall the anonymous vial that
originally generated this error. Scan one of your spare barcode labels, then discard it. Allow
this station to go into error too, then resolve all the station errors as described above under
error E12 and in Section 4.9 after the misplaced tubes have been tested manually.
POSSIBLE CAUSE(S)
• During a software update, an error occurred.
CORRECTIVE ACTION(S)
00000000, 00000002, 00000003
* Return to the “update software” operation and repeat the software update.
00000001
* You cannot install an older version of software than is currently on the instrument.
POSSIBLE CAUSE(S)
• The tube you just scanned belongs in a drawer other than the one currently open.
CORRECTIVE ACTION(S)
* Place the problem tube aside and complete the operation you are currently performing.
Then open the next drawer, press the “tube entry” soft key, and scan the barcode label of
the problem tube. If the error recurs, repeat this step for the third drawer, which is where
the tube should reside.
POSSIBLE CAUSE(S)
• You just scanned a tube barcode sequence number when the system expected a barcode
accession number, or vice versa. (The accession barcode feature must be enabled for this
error to occur.)
CORRECTIVE ACTION(S)
* The icon in the main body of the display shows which barcode the system expects you to
scan. The arrow pointing to the upper barcode indicates that the tube barcode sequence
number (preprinted on the BBL® MGIT™ tube label) is expected:
7–10 MA-0117-D(Prelim)
)
Troubleshooting
The arrow pointing to the lower barcode indicates that the accession barcode sequence
number is expected:
To prevent the error from occurring, always scan the type of barcode the system is prompt-
ing you to scan.
POSSIBLE CAUSE(S)
• During the write data to disk function, the system could not detect a floppy disk in the disk
drive.
CORRECTIVE ACTION(S)
* If a disk is in the disk drive, eject it and insert a blank, formatted, write-enabled disk in the
disk drive. Repeat the write data to disk operation.
POSSIBLE CAUSE(S)
• During the write data to disk function, the system filled the current floppy disk but has more
data left to write.
CORRECTIVE ACTION(S)
* Eject and label the current disk and insert another blank, formatted, write-enabled disk in
the disk drive.
MA-0117-D(Prelim) 7–11
BACTEC® MGIT™ 960 System User’s Manual
POSSIBLE CAUSE(S)
• Change was made to LIS enable/disable status or to one of the LIS parameters. Instrument
must be rebooted for changes to take effect.
CORRECTIVE ACTION(S)
* Press “OK” soft key to reboot instrument and implement changes, or press “not okay” soft
key to cancel changes.
POSSIBLE CAUSE(S)
00000002 – One of the drawers is not completely closed.
CORRECTIVE ACTION(S)
* Push the drawer closed.
POSSIBLE CAUSE(S)
• System encountered a software error.
CORRECTIVE ACTION(S)
* Write data to disk (see Section 7.3) and call Becton Dickinson.
7–12 MA-0117-D(Prelim)
)
Troubleshooting
4 An hourglass icon appears while the information is being written to the disk.
5 When the activity is complete, the system beeps three times. Remove the floppy disk by
pressing the disk eject button.
MA-0117-D(Prelim) 7–13
BACTEC® MGIT™ 960 System User’s Manual
7–14 MA-0117-D(Prelim)
l
Glossary and
Abbreviations
Below are some terms used within this manual which may be unfamiliar to the casual computer
user. Most terms are described within the context of the BACTEC® MGIT™ 960 system, rather than
in the strictly technical sense.
anonymous tube Anonymous tubes are those which have been placed in the
instrument without being assigned their location through Tube
Entry (have not been scanned into the instrument). Tests are
ongoing, but no other status (such as positive, negative, error)
can be indicated for anonymous tubes until they are identified to
the system.
MA-0117-D(Prelim) G–1
BACTEC® MGIT™ 960 System User’s Manual
D/RCC Designation within the manual for the station. “D” represents the
drawer letter (A, B, or C), “R” represents the row (A – S, omitting
I, O, and Q), and “CC” represents the column (01 – 20).
G–2 MA-0117-D(Prelim)
l
Glossary and Abbreviations
foreground task The currently active display or operation, with which the user is
interacting.
interface A go-between. The user interface for the BACTEC® MGIT™ 960
system is the displays, icons, and soft keys through which you
view information and perform activities.
kilobyte 1,024 bytes. Abbreviated “kb” or sometimes “k.” See also “byte.”
positive A tube is deemed positive if, during its test, it meets the
predefined criteria of the system software. These criteria relate to
the measurement of fluorescence in the tube sensor over one or
more test readings.
MA-0117-D(Prelim) G–3
BACTEC® MGIT™ 960 System User’s Manual
prompt A message from the system relating to the current activity. For
many of the activities you perform, the LCD displays offer
prompts in the form of icons in the main body of the screen.
Pulled Positive Any positive tube which has been removed through the Remove
Positive tubes operation can be placed back in the instrument for
further testing for up to 5 hours after removal (or until it goes
out of protocol, whichever comes first). During this re-entry
window of time, the tube is referred to as a “Pulled Positive.”
soft keys The eight teal keys below the LCD display whose functions vary
with each display. The functions are defined by the software, thus
the name.
G–4 MA-0117-D(Prelim)
Limited Warranty
f
This warranty gives you specific legal rights. Additionally, you may have other rights that vary from
region to region.
The BACTEC® MGIT™ 960 system is warranted to be free from defects in materials and workmanship
for a period of one year following delivery. Becton Dickinson’s sole responsibility under this warranty
shall be to repair or replace any instrument or its components (except for expendable supplies such
as printer cartridges, paper, or filters) which under normal operating conditions, prove to be
defective within one year of delivery.
Becton Dickinson Microbiology Systems will furnish new or remanufactured components upon its
option. All replacements shall meet new part specifications. Replaced components become the
property of Becton Dickinson.
It is understood that the equipment covered by this Agreement has been installed in accordance
with the recommendations and instructions in the BACTEC® MGIT™ 960 System User’s Manual.
Any damage to a BACTEC® MGIT™ 960 system resulting from the insertion or removal of cables that
connect this instrument to systems other than those approved or supplied by Becton Dickinson or
the failure of the owner to maintain reasonable care and precautions in the operation and
maintenance of the system will void this warranty and terminate the obligations of the
manufacturer as stated herein.
This warranty is in lieu of all other warranties, whether express or implied, including but not limited
to, warranties of merchantability, or fitness for a particular use. In no event will Becton Dickinson
Microbiology Systems be liable for indirect, incidental or consequential damages.
MA-0117-D(Prelim) A–1
BACTEC® MGIT™ 960 System User’s Manual
A–2 MA-0117-D(Prelim)
Replacement Parts
g
The following items may be ordered by contacting your local Becton Dickinson representative at the
numbers in Appendix D.
Item
4405874
MA-0117-D(Prelim) B–1
BACTEC® MGIT™ 960 System User’s Manual
B–2 MA-0117-D(Prelim)
Software Update Log
Received Version
h
Whenever you receive a software update, please take a moment to log it below. This can assist you
and Becton Dickinson personnel in identifying software revision levels, potential software problems,
etc.
MA-0117-D(Prelim) C–1
BACTEC® MGIT™ 960 System User’s Manual
C–2 MA-0117-D(Prelim)
International Contacts
Becton Dickinson de
Mexico, S.A. de C.V.
Monte Pelvoux 111 • 9th Floor
Col. Lomas de Chapultepec
i Becton Dickinson
Becton Dickinson and Company
2464 South Sheridan Way
Mississauga, Ontario
11000 Mexico D.F. Canada L5J 2M8
Voice: 52 5 237 1200 • Fax: 52 5 237 1287 Voice: (905) 855-5663 • Fax: (905) 855-6583
MA-0117-D(Prelim) D–1
BACTEC® MGIT™ 960 System User’s Manual
D–2 MA-0117-D(Prelim)
E.1
Supplemental Procedures
Reading BBL®
jMGIT™ 7 ml Tubes Manually
A situation may arise where it is necessary to remove one or more ongoing negative BBL® MGIT™
7 ml tubes from the instrument and to incubate them offline. These tubes should then be read
manually using an ultraviolet (UV) transilluminator (365 nm) or a Wood’s lamp with a longwave
bulb or blacklight.
NOTE
WARNING
Use of Positive and Negative Control tubes is only for the interpretation of fluorescence and is not
intended as a control for the performance of the media. The procedure that follows is for the
manual reading of tubes only, not for entry of controls into the instrument.
MA-0117-D(Prelim) E–1
BACTEC® MGIT™ 960 System User’s Manual
E–2 MA-0117-D(Prelim)
j
Supplemental Procedures
1 Add the contents of the contaminated BBL® MGIT™ 7 ml tube to a 50 ml plastic centrifuge
tube.
2 Add 8 ml NALC-NaOH solution to the centrifuge tube. With the cap tightened, vortex the
tube for 5 to 20 seconds.
3 Allow the tube to stand for 15 to 20 minutes. Do not treat for more than 20 minutes.
4 Add 35 ml sterile phosphate buffer pH 6.8. Replace the cap and mix the contents.
5 Concentrate the specimen in a centrifuge at a speed of 3,000 x g for 15 minutes.
6 Carefully decant the supernatant fluid from the pellet. Resuspend the pellet using a sterile
Pasteur pipette with sterile phosphate buffer pH 6.8.
7 Inoculate 0.5 ml of the suspension to a new BBL® MGIT™ 7 ml tube. Also add a drop (0.1 ml)
of the suspension to a 7H10 agar plate or other mycobacterial solid agar or egg-based
medium.
The routine inoculation of solid media is especially important for optimal recovery of mycobacteria
from tissue specimens as these specimen types are particularly susceptible to sporadic organism
recovery.
MA-0117-D(Prelim) E–3
BACTEC® MGIT™ 960 System User’s Manual
E–4 MA-0117-D(Prelim)
Notes
MA-0117-D(Prelim) Notes–1
BACTEC® MGIT™ 960 System User’s Manual
Notes–2 MA-0117-D(Prelim)
A
accession barcoding setup, 2-6
alarm volume setup, 2-6
alarms, 3-11
alerts, 3-11, 7-2
n
Index
D
database
save to disk, 7-12
date format setup, 2-5–2-6
date setup, 2-5
decontamination, 6-14
anonymous tube displays
definition, G-1 hierarchy, 5-2
identifying, 4-13 types, 5-3–5-5
arrow keys, 3-3 drawers
error station indicator, 3-8
B negative tube indicator, 3-8
overview, 1-3
barcode scanner positive tube indicator, 3-7
how to scan, 3-6 release latch, 3-7
overview, 1-3
Becton Dickinson E
how to contact, D-1 E01 error, 7-3
blocked station E02 error, 7-3
E04 error, 7-3
definition, G-1
E05 error, 7-4
C E06 error, 7-4
E07 error, 7-5
calibration E08 error, 7-5
calibrator replacement, 6-9 E09 error, 7-5
overview, 1-6 E10 error, 7-5
E11 error, 7-6
cautions, 1-10
E12 error, 7-6
circuit breakers E13 error, 7-7
assignments, 3-10 E30 error, 7-7
CLIA compliance, 2-9 E31 error, 7-8
MA-0117-D(Prelim) I–1
BACTEC® MGIT™ 960 System User’s Manual
K removing, 4-10
keypad O
alarm indicator, 3-4
On/Off switch, 3-3
arrow keys, 3-3
naming conventions, 1-8
overview, 1-4
P
SILENCE ALARM key, 3-4 positive tube indicator, 3-7
soft(ware) keys, 3-4 positives
indication, 3-7
L notification, 4-8
language setup, 2-6 removing, 4-8–4-9
LCD display returning for further testing, 4-9
automatic dimming, 3-5 power failures, 4-1
brightness dial, 3-6 power switch, 3-3
display types, 5-3–5-5 protocol setup, 2-5
I–2 MA-0117-D(Prelim)
n Index
MA-0117-D(Prelim) I–3
BACTEC® MGIT™ 960 System User’s Manual
I–4 MA-0117-D(Prelim)
Reader Comment Card
BACTEC® MGIT™ 960 System User’s Manual (MA-0117)
After you have become familiar with your instrument and its documentation, please take a moment to evaluate the
manual. Your comments are a valuable tool for us to improve the usefulness of the manuals. NOTE: For any evaluation of
“average” or below, please add a specific comment on the blank lines at the end.
Thank you!
Optional Name
Hospital
City, State
Fold Under & Tape
Fold Here
NO POSTAGE
NECESSARY IF
MAILED IN
THE UNITED
BUSINESS REPLY MAIL
PERMIT NO. 25 SPARKS, MD
BECTON DICKINSON
MICROBIOLOGY SYSTEMS
Technical Writing
7 Loveton Circle / P.O. Box 999
Sparks, MD 21152-0999