Biological Safety Cabinet Class II A2: BSC-1100IIA2-X BSC-1300IIA2-X BSC-1500IIA2-X
Biological Safety Cabinet Class II A2: BSC-1100IIA2-X BSC-1300IIA2-X BSC-1500IIA2-X
Class II A2
BSC-1100IIA2-X
BSC-1300IIA2-X
BSC-1500IIA2-X
USER MANUAL
Version 2019.05
Preface
Thank you very much for purchasing our class II A2 biological safety cabinet.
Please read the “Operating Instructions” and “Warranty” before operating this unit to assure proper
operation. After reading these documents, be sure to store them securely together with the
“Warranty” within touch for future reference.
Warning: Before operating the unit, be sure to read carefully and fully understand important
warnings in the operating instructions.
Disclaimer
Biobase shall not be liable for any equipment failure or damage, or for any direct or indirect damage
that may occur during the use of the equipment.
1.Malfunction or damage due to violation of the instructions, precautions, and intended use of this
manual.
2.Malfunction or damage caused by repair or alteration of the other company.
3.Malfunction or damage caused by use instruments of other company at the same time .
4.Malfunction or damage caused by operating environment not corresponding to the specified
operating environment (power conditions, installation environment, etc).
5.Malfunction or damage caused by natural disasters such as earthquakes and floods.
6.Malfunction or damage caused by the company unaware of the movement or transfer (transport)
after installation.
Contact information
Company: Jinan Biobase Biotech Co., Ltd
Company Address: NO. 51 South Gongye Road, Jinan, China
Service Email: [email protected]
Fax: +86-531-81219804
Tel: +86-531-81219801/03
Zip code: 250000
2
CONTENT
2. User Instructions............................................................................................................................ 14
2.1 Functions.....................................................................................................................................14
2.1.1 Product Concept................................................................................................................... 14
2.1.2 Application Range................................................................................................................ 14
2.1.3 Working theory/Air flow pattern and protected area............................................................14
2.1.4 Protected objects...................................................................................................................15
2.1.5 TECHNICAL PARAMETERS............................................................................................ 15
2.1.6 Performance Index............................................................................................................... 16
2.2 Product structure......................................................................................................................... 17
2.2.1 Structural composition of BSC-1100IIA2-X........................................................................17
2.2.2 Structural composition of BSC-1300IIA2-X....................................................................... 18
2.2.3 Structural composition of BSC-1500IIA2-X....................................................................... 19
2.2.4 Structure introduction........................................................................................................20
2.3 Control panel...............................................................................................................................22
2.4 Remote Control &Foot switch....................................................................................................23
2.4.1 Remote control..................................................................................................................... 23
2.4.2 The use of foot switch.......................................................................................................... 25
2.5 Instructions for Operation...........................................................................................................25
2.5.1 Normal Operation Notice..................................................................................................... 25
2.5.2 Operation Process.................................................................................................................26
2.6. Daily maintenance..................................................................................................................... 27
2.6.1 Preparations before maintenance..........................................................................................27
3
2.6.2 Clean the cabinet surface......................................................................................................27
2.6.3 Overall maintenance period...............................................................................................27
2.6.4 Maintenance methods...........................................................................................................27
2.6.5 Storage conditions................................................................................................................ 28
2.7 Methods and procedures for disinfection....................................................................................29
2.8 Replacement parts list.................................................................................................................31
2.9 Wiring diagram........................................................................................................................... 34
4. Warranty......................................................................................................................................... 43
4
1. Unpacking, Installation, Debugging
Please firstly check if packing box is in good condition. If the packing box is damaged, please take
photos.
1.1 Unpacking
Picture 1
5
Picture 2
Rapid unpacking diagram (Picture 3). Disassemble the screws shown in the below Picture, then
move the wooden pieces to right and left.
Picture 3
6
Using ordinary scissors to cut packing tape, take off the package cover, then move up the paper box
body.
7
1.3 Installation conditions and using environment
To avoid disturbances to the safety cabinet and its operator, follow the following guidelines, while
determining a suitable location for the cabinet:
a. The distance from the plane of the aperture to any circulation space should be at least 1000 mm,
so as to preserve a zone undisturbed by anyone other than the operator.
b. Biological safety cabinets should be placed in a position where there should be no opposing wall
(or other obstruction likely to affect the airflow) within 2000 mm of the front aperture.
c. Safety cabinets should not be installed in positions where they are likely to be affected by other
items or equipment. In particular the distance to the aperture of an opposing safety cabinet, fume
cupboard, or the edge of a local exhaust ventilation outlet should not be less than 3000 mm.
d. Any room air supply diffuser should not be within 1500 mm of the front aperture.
e. Doorways should not be within 1500 mm of the aperture or within 1000 mm of the side of the
safety cabinet.
f. The position of a safety cabinet should satisfy the spatial requirements (e.g. vision, lighting and
convenience of access) of the operator and personnel working nearby. When a cabinet is installed
on a bench top, the leading edge should be flush with or slightly overhanging the edge of the
bench top.
Working environment:
(1) Only is suitable for indoor;
(2) Ambient temperature: 15℃~35℃;
(3) Relative Humidity: ≤75%;
(4) Atmospheric pressure range: 70 kPa~106 kPa;
(5) Electrical parameters: Consistent with the rated voltage of the biosafety cabinet (See 2.1.5
technical parameter performance index);
(6) Power supply need to be grounded; (Judging method: testing the fire wire and the zero line of the
power supply with multimeter, the fire wire to ground voltage should be grid voltage and the zero
line to ground voltage should be 0, otherwise the power supply ground is bad).
(7) Test the voltage stability before using, if the voltage is unstable, should use the voltage regulator,
otherwise the control panel and transformer may be easily damaged.
1.4 Installation
8
c. Move the whole device to the final installation location;
The base stand will be packed at back of main body, please take it out before installation.
Picture 4
Remove the M10×55 Hexagon socket head screws, Buckle plug, M10 Stainless steel flat washer,
M10 Stainless steel spring washer, M10 Stainless steel cap nuts, referring to Picture 4 assembled
base, firmly fastening requirements.
9
e. Connect base stand and main body
Refer to Picture 5, 6 to connect base stand and main body.
Picture 5
Align the mounting holes on the bottom of the cabinet with the mounting bolts, and the cabinet
slowly descending the mounting base
10
Picture 6
Take out the M10*55 Hexagon socket head screws, Flat washer 10, Spring washer 10 from the
accessory box, and fasten tightly according to Picture 6.
Picture 7
11
1. Drain valve connect
2. Shim (Inner diameter*outer diameter*thicknessΦ20*Φ28*2mm)
3. Safety cabinet bottom installation holes
4. Ball coupling fastening nut
5. Rubber gasket (Inner diameter*outer diameter*thicknessΦ13*Φ19*2mm)
6. Drain valve
Take out drain valve coupling, shim, Ball nut, Rubber gasket, Drain valve, assembling from up to
down as Picture 7 illustrated.
g. Adjustment of Footmaster Caster
Picture 8
Clockwise rotate caster’ red part to low down the base feet and the height of the cabinet. Low down
all four casters can move the cabinet position. Counterclockwise rotate casters’ red part can rise the
base leg and height of cabinet. Raise all four casters can at same time can fix the cabinet. Adjust the
four Foot -masters makes the cabinet stable.
h. Foot Switch
Picture 9
Install Foot Switch as Picture 9. It’s socket is at the left top, connect the plug.
i. Installation of Water and Gas Tap(Optional)
12
Picture 10
1. Fastening Nut
2. Stainless Steel Water and Gas Taps
Take out fastening nuts, water and gas taps, installing as Picture 10.
First, make sure the Voltage and frequency to be same as logo showing, and then check the follows
items with power on:
Checking Items Normal situation
Inflow 0.53±0.025m/s , downflow
Wind speed display
0.33±0.025m/s
13
2. User Instructions
2.1 Functions
Picture 11
14
2.1.4 Protected objects
Biological safety cabinets (BSCs) are designed to protect the operator, the laboratory environment
and work materials from exposure to infectious aerosols and splashes that may be generated when
manipulating materials containing infectious agents, such as primary cultures, stocks and diagnostic
specimens.
Frequency 50 Hz 60Hz
External
1100×750×2250 mm 1300×750×2250 mm 1500×750×2250 mm
Size(W*D*H)
Working Zone
940×600×660 mm 1150×600×660 mm 1350×600×660 mm
Size(W*D*H)
Total Airflow
360m3/h 440m3/h 500m3/h
Volume
UV Lamp
30W 30W 40W
Consumption
LED lamp
12W*2 12W*2 16W*2
Consumption
Downflow
0.33±0.025m/s
Velocity
Notes: (1) Electric consumption power including power which operation area needs to load (Loading
no more than 500W)
(2) Our company has right for changing the products, if we need to change and re-design, please
forgive us for not notifying you.
15
2.1.6 Performance Index
1) Biological safety functions
Personnel protection, microbial colony count ≤5CFU;
Sample protection, microbial colony count ≤5CFU;
Cross contamination protection, microbial colony count ≤2CFU.
2) Leak-proof Cabinet
If cabinet pressurized to 500Pa, the pressure should be no less than 450 Pa after 30 min.
3) Integrity of HEPA Filter
Scan and detect the HEPA filter, the leakage rate at any point should not be >0.01%.
4) Vibration amplitude
The net vibration amplitude between frequency 10Hz and 10KHz is no more than 5μm(rms).
5) Illumination
The average illumination is no less than 650 lux, actual illumination is no less that 450 lux.
6) Mechanical performance
Structure design is reasonable, high quality materials are adopted for the cabinet.
It can resist shape global deformation caused by external force.
The working surface will not occur permanent deformation when weight put reaching 23kg.
7) Electrical properties
The voltage increases to 1390V(AC) in 5s and keep for another 5s without breakdown.
Grounding resistance ≤0.1Ω
16
2.2 Product structure
Picture 12
17
2.2.2 Structural composition of BSC-1300IIA2-X
Picture 13
18
2.2.3 Structural composition of BSC-1500IIA2-X
Picture 14
1. Footmaster Caster 8. UV lamp
2. Nameplate 9. Power supply lock
3. Power socket 10. LED Lamp
4. General power fuse socket 11. Water-proof socket
5. Fuse socket 12. Control panel
6. Blower fuse tube socket 13. Front window
7. Tube motor 14. Inflow grid
19
2.2.4 Structure introduction
1) Driving System of Front Window
2) Driving system consists of tube motor, front window, hauling sash and position switch.
3) Air Filtration System
4) Air Filtration System is the most important system of BSC. It consists of blower, supply filter
and exhaust filter. The function of Air Filtration System is transferring filtered air to work
area, ensure the down flow velocity, and keep Class 100 cleanness of work area.
5) UV Light
6) UV lamp is inside work area. So UV lamp can well sterilize all space of work area.
7) LED Light
8) The BSC is equipped with LED lamp. It can make sure average illumination inside work area
which meets standard requirements.
9) Air pipe
10) Air pipe is the ventage of differential pressure sensor.
Picture 15
The air pipe should not be blocked and please do not hang anything on the pipes,
20
Waterproof Sockets are located on the right side of the work area, which can be controlled by
SOCKET button.
(1) Please make sure the total load of sockets should be ≤ 500W;
(2) The waterproof socket can only be waterproof when its front cover is lowered, and the
socket can not be regarded as a waterproof socket when the front cover is opened.
Picture 16
Large LCD display indicates detailed key parameters, it is real-time display to reflect the equipment
working condition, such as effective working state of the filter, which is more intuitive. (Please refer
to 2.3)
15) Control of Front Window
Front window is motorized. It could be controlled by control panel. Glass door control motor for a
dedicated reversible speed motor, the mo tor has a large starting torque, smooth rotation and so on.
16) Structure
a) Biological Safety Cabinet’s both sides and back area are negative pressure air channel. And
the negative pressure keeps work area away from contamination.
b) Cabinet body is built of 1.2mm cold-rolled steel with anti-powder coating. Strong and steady.
c) Work area is fully made of stainless steel which looks beautiful and with corrosion resistance
performance.
21
d) Base stand is made of cold-rolled steel with anti-powder coating.
e) Soft touch type control panel, easy to handle and beautiful appearance.
Picture 17
1.UV Lamp 2. LED Lamp
3.Fan 4.Socket
5.Glass Window Up 5.Glass Window Down
7.Mute 8.Power
9.Alarm Status 10.Mute Status
11.Filter Changing Status 12.Temperature
13.System Working Time 14.Humidity
15.Glass Window Down Status 16.Glass Window Up Status
17.Foot Switch Status 18.UV Status
19.Socket Status 20.LED Lamp Status
21.Fan Status 22.Exhaust Filter Differential Pressure
23.Inflow velocity 24.Supply Filter Differential Pressure
25. Down-flow velocity 26.Filter Working Time
27.UV Lamp Working Time 28.Reservation Timing
a) LCD Screen
22
The working status of the equipment and operation can be seen on the LCD screen.
: To control blower working status. (It will not work when front window is fully
closed.)
Clock Adjustment:
Turn the power key, so machine is in standby state.
Press the light button, and then press the power button for 5 seconds. Then you see the state of clock
adjustment after a buzzer alarm.
Firstly, minute position is flashing, press UP and DOWN to adjust to present time. Then press the
MUTE button switching to hour position and adjust to present time. After that, press the light button
first, and press the power button for about 5 seconds. Data will be saved after a buzzer alarm.
23
even carry it with themselves during experiment for convenience.
This remote control adopt specific chip which is featured with good anti-jamming performance,
longer control distance and high control precision.
1. Power (POWER)
2. Reservation Time (SUB)
3. Timer (INSTALL TIMER)
4. Confirm (CONFIRM)
5. Cancel (CANCEL)
6. Turn up (+)
7. Turn down (-)
8. Fan (FAN)
9. UV (UV)
10. Illumination (LIGHT)
11. Socket (SOCKET)
12. Mute (MUTE)
13. Front window up (UP)
14. Front window down (DOWN)
Picture 18
A. Reservation Time (SUB)
a. Connect power, open power lock, and press the reservation timing button (SUB);
b. Adjust the time (minutes) by “+” or “-” button. Press the confirmation button (CONFIRM) to
confirm; and then adjust other minutes and hours position data in the same way;
c. After the time is confirmed, the corresponding display lamp lights by selecting the function
buttons (such as UV);
d. Press the POWER button again, the reservation function starts. Reserved time starts count
down. The corresponding setting function starts when the time counts down to zero.
B. Timer (INSTALL TIMER)
a. Connect power, open power lock, press button (POWER), the corresponding display lamp
lights by selecting the function buttons (such as UV);
b. Press button (INSTALL TIMER), adjust the time (minutes) by “+” or “-” button. Press button
(CONFIRM) to confirm; and then adjust other minutes and hours position data in the same
way;
c. After the time is confirmed, the Timer function starts. When the time counts down to zero, all
the functions will be off, the cabinet will be in standby mode.
C. Application of Reservation Time
24
Biological safety cabinet is equipped with special UV lamp. When turning on or turning off the
cabinet, sterilization time of UV lamp should be at least 30 minutes. In order to save the waiting time
of turning on or turning off the cabinet, we develop reservation time function. It realizes function of
automatic turning on or turning off the cabinet after the sterilization finished. Reservation time
setting range is from 0 to 99 hours and 59 minutes. This function helps operators to save time and
improve efficiency.
Picture 20
Press the left red switch by foot, front window goes up, press the right black switch, front window
goes down.
25
(8) No flame: No flame is allowed inside the cabinet. Using of fire will lead to airflow disorder,
and filter damage. If sterilization is required during the experiment, infrared sterilizer is
highly recommended.
(9) HEPA filter life: With the usage time increasing, dust and bacteria accumulate inside HEPA
filter. Filter Resistance is getting bigger, when it reaches the maximum point, there will be
audible and visual alarm. Please replace new HEPA filter, otherwise it will affect the safety
performance of the equipment. The used filter should be processed as medical waste.
(10)There is a negative passage surrounding the work area, which is sealed strictly in the
factory. The operator is not allowed to remove or loose screws of those parts. If necessary,
please contact service personal.
(11)Front Grille is used for air intake and drain. Do not block it, otherwise it will affect airflow.
Armrest is recommended to solve this problem and reducing the operator's wrist fatigue.
(12)Long-term use of biological safety cabinets will inevitably cause pollution (e.g. HEPA filters,
corner cabinets, etc.). In order to sterilize thoroughly every 500 hours, formalin
(formaldehyde) fumigation sterilizer is recommended. After sterilization, neutralize
formaldehyde gas with ammonium hydrogen carbonate. Make sure no sterilization gas
escapes during the whole process.
(13)The maximum storage period is one year. If the period is more than one year, performance
test should be done .
Serious declaration: we will take no responsibility for risks caused by improper operation and
man-made damages!
When front window is open or other button is pressed, the UV lamp can not be work.
d. Before doing experiment, please sterilize the cabinet for more than 30 minutes by UV lamp( the
front window must be closed)
(1) For safety of eyes and skin, people should leave room during the UV sterilization.
(2)UV lamp intensity should be tested regularly. If there is no test conditions, it should be
replaced, recommend once for three month.
26
e. Please move the front window at 200mm height from the work table, turn on the fan, make sure
the experiment should be started after fan working for at least half an hour.
For operating safety, please put testing materials inside the cabinet in advance, and keep
the front window at 200mm height from the work table during operation.
After finishing the experiment, please move the front window down to the bottom, and make sure to
sterilize the cabinet by UV lamp for 30 minutes before turning off the cabinet.
Because the operating time will directly affect the judgment of maintenance needs, we recommend
the user keep a detailed record of operating time for reference.
When doing maintenance, please pay attention to cut off the power, so as to avoid electric
shock
Disinfectants used for wiping should not damage 304 stainless steel.
27
a. Disinfect and clean operating area;
b. Clean the external surface and front window around the operating area;
c. Check the various functions of equipment;
d. Record this maintenance result
2) Monthly maintenance
a. Clean the the external surface and front window.
b. Wipe the working table, inner wall surface of operating area (excluding the wind distributing grid
of operating area) and the inner surface of glass door with 70﹪ medical alcohol or household
bleach diluted 1:100 (i.e, 0.05% sodium hypochlorite). Then wipe again with sterile water in
order to eliminate the rest chlorine.
c. Check the various functions of equipment;
d. Record this maintenance result;
3) Annual maintenance
a. Check the two conveyor belts of front window drive unit, and ensure that their tightness is
coincident.
b. Check the UV lamp and LED lamps.
c. Apply for testing the overall performance of cabinet on an annual basis to ensure the
performance safety. User is responsible for testing costs.
d. Record this maintenance result.
28
2.7 Methods and procedures for disinfection
Before sterilization, all the hydrogen chloride should be moved away from the cabinet.
There will result in carcinogenic substance-BCME when it meet with formaldehyde under the
ambient air.
a. Figure out the total volume by multiplying the height, width and depth.
b. The required weight of the triformol will be worked out through the total volume multiply
11g/m³, figuring out the weight of ammonium bicarbonate or its alternative by
chemometry.
c. It must be bubble-tight if there are exhaust pipe with the biosafety cabinet, which can be
realized at the end of the pipe or sealing at the control valve if there is valve nearby the
cabinet, it need more triformol to compensate the added volume if the length of exhaust
pipe is more than 3 meters. If the exhaust recirculation of the cabinet connect with the
exhaust system of the construction, interrupting the connection system of the biosafety
cabinet with construction and sealing (by using plastic film and plastic strips)
d. Sealing by plastic film as the waste gas is exhausted into the room from biosafety cabinet.
e. In order to urgently eliminate formaldehyde, sterilize and eliminate formaldehyde after
neutralizing, putting a tube nearby the biosafety cabinet in advance which is required to
connect with chemical smoke hood or other exhaust devices that is suitable for releasing
the harmful gas.
29
f. Put the heater, such as electric-heating frying pan which can be bought from market,
formaldehyde generator or neutralizer on the working table. Set up the temperature at
232℃~246℃, the triformol is spraying over the surface of the heater devices equably.
g. The neutralizer with the heating device is placed on the workbench. Neutralizers
(Ammonium bicarbonate or equivalent) should be isolated from the air in the cabinet
before use. The following two examples illustrate how to achieve air separation.
Eg.1
Ammonium bicarbonate or substitute is uniformly sprayed on the heated surface of the
heating device and covered with aluminum foil to prevent the carbonic acid hydrogen
ammonia or its substitute in disinfection reaction with formaldehyde. The aluminum foil
should be placed so that ammonia can be released during heating or ready to remove
aluminum foil at the beginning of the neutralization phase. Removal of aluminum foil is
not allowed when the formaldehyde leakage out of the safety cabinet of;
Eg.2
The safety cabinet with the gloves for a whole plastic film sealing. Ammonium
bicarbonate or equivalent substitutes are placed in a container in a safety cabinet. During
the neutralization phase, the disinfected personnel into the safety cabinet through the
gloves do not break the sealing system. The ammonium bicarbonate or equivalent
substitutes are removed from the sealed container and uniformly sprinkled on the heating
surface of the heating device, then the heating device is energized, the ammonium
bicarbonate or the substitute is heated to release the ammonia.
h. Putting heating plate, water beaker and hygrothermograph on the work table, do not
connect the wires to the power supply in the cabinet.
i. Sealing the front operating window with thick plastic film and plastic strips. Seal all the
area where possible to leak. Such as wire outlet, the surrounding of front operating
window and joint between plastic film and biosafety cabinet.
j. Measure the temperature and humidity inner the cabinet.
k. The humidity is at 60%~85% when temperature is above 21℃, heating the water in the
beak up to the expected temperature and humidity.
l. Before depolymerizing the formaldehyde, strictly restrict entry to the area around
biosafety cabinet or room in accordance with related regulation and safety measures.
m. Connect the wire of the heater with socket outside the cabinet.
n. 25% formaldehyde is depolymerized open the safety cabinet fan 10 s ~ l 5 s. after
the para-formaldehyde is depolymerized by 50%, 75% and 100%, repeating the
30
above steps. If the safety cabinet fan does not work, Use auxiliary fans to promote
air circulation in the cabinet, or prolong the disinfection time over Step P .
o. Disconnect the power supply of the heating plate and heater for the use of triformol
p. Keep the biosafety cabinet isolation at least 6 hours, it is better to leave it for a night.
q. Prepare the neutralizer, power off the heating device and the fan of biosafety cabinet until
ammonium bicarbonate wiped away, the operation of triformol is also the same, after the
decomposition of 20% ammonium bicarbonate, open the fan for 10 s ~ l 5 s, using
assisted fan or electric fan to promote the air circulation inner biosafety cabinet or
prolong the neutralization time for at least 6 h if the fan of the biosafety cabinet does not
work.
r. Keep the biosafety cabinet isolation at least 1 hour before open the sealing membrane;
s. If exhausting the formaldehyde caused by neutralization via tube, tear out the plastic
cover in the exhaust pipe, connect it with tube and seal. The plastic cover in the front of
the biosafety cabinet will be suck in and cut one or two openings make the fresh air be
into the cabinet if the tube work well, then the formaldehyde is exhausted via tube.
31
AAQ13 Key selection button LA42YZP-20B
AAQ14 Glass 1034*730*6.38mm
AAQ15 Foot switch Dual pedal (line length 4 meters)
32
AAJ06 LED Lamp T5 16W LED
AAJ07 UV lamp ballast 1*TL8-36W
AAJ09 Upper filter (Exhaust filter) 920*420*69mm
AAJ10 Downflow filter (Supply filter) 1380*470*69mm
AAJ11 Fan FH320
AAJ12 LCD control board (strong circuit
Control panel board, weak circuit board, display
screen)
AAJ13 Remote control (with battery)
AAJ14 Key selection button LA42Y2P-20B
AAJ15 Glass 1419*720*6.38mm
AAJ16 Foot switch Dual pedal (line length 4 meters)
33
1 2 3 4 5 6 7 8
BSC-1100IIA2-X Wiring diagram/BSC-1300IIA2-X Wiring diagram
D T&H SENSOR D
WINDOW H
WINDOW L
BLACK(Φ1)
T SA LL AB V U
BLACK(Φ0.5)
YELLOW(Φ0.5)
BLACK(Φ0.5)
GREEN(Φ0.5)
WHITE(Φ0.5)
RED(Φ0.5)
UV LAMP WHITE(Φ0.5)
BLUE(Φ1)
)5.0Φ(DER
)5.0Φ(DER
)5.0Φ(ETIHW
)5.0Φ(KCALB
)5.0Φ(EULB
Window
H
L
GND
VCC
H T
UV Ctrl
DNG
GREEN(Φ1) up
Foot
LAMP
BLACK(Φ1 ) down
C BLUE(Φ1) C
BLACK(Φ1 )
NAIJNA
MCU EIGHT CORE CABLE
LCD
TRANS-IN
GREEN(Φ1) L
LAMP 1 2V/0 .5A
34
BLUE(Φ1) N EX-FAN
GREEN(Φ1 )
L 1 8B2-EF GREEN(Φ1 )
N TRANS-OUT SPEAK
L
VU
N
SIXTEEN CORE CABLE
L
L TEKCOS THGIL
SOCKET
FUSE3 N
RED(Φ1) RED(Φ1) SPEAKER
L
5A N
BLUE(Φ1)
YELLOW GREEN(Φ1.0) BROWN(Φ1.0)
)1Φ(DER
DOWN
ROTOM ROOD
BLUE(Φ1.0) DOWN-FAN N UP
N
L N
B SOCKET B
YELLOW GREEN(Φ1.0)
BLUE(Φ1) BLACK(Φ1) BROWN(Φ1)
YELLOW(Φ1) BLUE(Φ1)
FUSE2 DOOR MOTOR
BROWN(Φ1.0)
L
WAVE FILTER
10A DOWN FAN
2.9 Wiring diagram
FOOT SWITCH UP
WINDOW H
WINDOW L
BLACK(Φ1)
T SA LL AB V U
D
BLACK(Φ0.5) D
WHITE(Φ0.5)
YELLOW(Φ0.5)
BLACK(Φ0.5)
GREEN(Φ0.5)
UV LAMP WHITE(Φ0.5)
RED(Φ0.5)
BLUE(Φ1)
)5.0Φ(DER
)5.0Φ(DER
)5.0Φ(ETIHW
)5.0Φ(KCALB
)5.0Φ(EULB
1 3
2
Window
H
L
GND
VCC
H T
UV Ctrl
GREEN(Φ1)
LAMP
DNG
BLUE(Φ0.5)
BLUE(Φ1)
RED(Φ0.5)
up
Foot
BSC-1500IIA2-X Wiring diagram
BLACK(Φ1 )
down
BLACK(Φ0.5)
NAIJNA
BLACK(Φ1 )
MCU EIGHT CORE CABLE
LCD
GREEN(Φ1)
LAMP
C BLUE(Φ1) TRANS-IN C
L
EX-FAN 1 2V/0 .5A
N
GREEN(Φ1 )
L GREEN(Φ1 )
1 8B2-EF
N TRANS-OUT SPEAK
L
VU
N
35
SIXTEEN CORE CABLE
L
L TEKCOS THGIL
SOCKET
N
FUSE3 RED(Φ1)
RED(Φ1) SPEAKER
L
5A N
BLUE(Φ1)
YELLOW GREEN(Φ1.0) BROWN(Φ1.0)
)1Φ(DER
DOWN
ROTOM ROOD
BLUE(Φ1.0) DOWN-FAN N UP
N
L N
SOCKET
YELLOW(Φ1)
YELLOW GREEN(Φ1.0)
B B
FUSE4
10A BLUE(Φ1) BLACK(Φ1) BROWN(Φ1)
YELLOW(Φ1)
BLUE(Φ1)
FUSE2 DOOR MOTOR
BROWN(Φ1.0) BROWN(Φ1.0)
L
WAVE FILTER
10A DOWN FAN
POWER POWER LOCK
YELLOW GREEN(Φ1.0)
BLUE(Φ1.0) BLUE(Φ1.0)
N
A A
1 2 3 4 5 6 7 8
3. Trouble shooting and Labels
Please confirm whether the power is connected or not, whether the power cord is obvious damaged
or not, whether the fuse is good or not, and whether the power locks are in the open state or not
before the fault diagnosis.
Faults Check parts Measures
UV lamp Change it
36
Micro Switch Check if Micro Switch is broken
Pressure or air speed Gas circuit Check whether gas circuit has dropped, is broken, or
displayed incorrectly jammed
Control panel Change it
37
Power key Check if power key is open, is broken or not
Micro switch Check whether the micro switch is good, and it works
normally or not.
No alarm Circuit Check whether connection circuit of micro switch is
good.
Control panel Change it
NOTES
(1) The above electrical parts must be operated by a qualified electrician in safety conditions
(cutting off power supply). The other parts are not allowed to remove; otherwise the user
should take responsibility by them;
(2) When failures are not occur, and the operator can’t solve, please notify our maintenance
department immediately. For your safety, please do not maintain equipment by yourself;
(3) The maintenance of this equipment is undertaken by trained and recognized technicians;
(4) If you need to order parts, contact the agent or our technical service department, and please
indicate the model and serial number of the cabinet purchased.
38
Picture 20
Picture 21
39
Picture 22
Picture 23
40
3.2 Label Description
Picture 24
2)Fuse label (Picture 25)
Picture 25
Note:
3)Ground label
Picture 26
Picture 27
41
5)UV lamp alarm label
Picture 28
Picture 29
Picture 30
Picture 31
42
4. Warranty
1) Warranty is 1 years from EX-factory date (excluding consumable accessories, UV and LED
lamp, fuse).
2) We will take no responsibility for risks caused by improper operation and man-made damages.
3) After the expiration of warranty, our company is also responsible for repairs, but the
corresponding maintenance cost should be charged.
4) Life time of biosafety cabinet is 8 years from production date on the label.
5) We can provide equipment drawings and necessary technical data for maintenance companies or
personnel trained by our company.
Warranty declaration:
One-year Warranty, Life-long Maintenance.
BIOBASE GROUP
Add:No.51 South Gongye Road,Jinan,China
Tel:0086-531-81219801
Fax:0086-531-81219804
E-mail:[email protected]
Web:www.biobase.cc
43