0% found this document useful (0 votes)
414 views36 pages

3 Osteosys DT Service Training English

The document provides training information for service engineers on the Osteosys DEXXUM T bone densitometry device, including an overview of device parts and modules, installation and operation procedures, troubleshooting assistance indicators, and methods for repairs, firmware updates, and realignment calibration. Trainees will learn how to install and operate the system through demonstrations, as well as how to diagnose and resolve common errors through the device's indicator lights.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
414 views36 pages

3 Osteosys DT Service Training English

The document provides training information for service engineers on the Osteosys DEXXUM T bone densitometry device, including an overview of device parts and modules, installation and operation procedures, troubleshooting assistance indicators, and methods for repairs, firmware updates, and realignment calibration. Trainees will learn how to install and operate the system through demonstrations, as well as how to diagnose and resolve common errors through the device's indicator lights.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 36

Service Engineer

training

• Osteosys Co., Ltd.


Index
1. Introduction
2. General information of device
3. Interior / exterior part’s names and interior module
4. Installation and operation (DEMO)
1) User program’s installation (DEMO)
2) Test program (DEMO)
3) PACS, OCS (DEMO)
4) Daily test (DEMO)
5) Measurement and analysis (DEMO)
6) Patient’s data back up and import (DEMO)
Index
5. A/S repairing
1) Indicator’s explanation of module
2) Power error (when power is not on)
3) Movement error
4) X-RAY exposure error

6. Firmware Download method


7. Module exchanging method
8. Alignment method
9. Calibration method
1. Introduction
Q: Osteoporosis?

A: Is a disease of bone that leads to increased risk of fracture.

In osteoporosis the bone mineral density (BMD) is reduced, bone microar


chitecture is disrupted, and the amount and variety of proteins in bone is
altered.

Normal Osteoporosis
1. Introduction
Q: Patient’s mortality risk in with femoral fractures?

A: 12~20% of femoral fractures’ patient died within 6 months and 20~25%


of patient have a possibility to pass way with a complication.

Spine fracture Femur fracture


1. Introduction
Q: The fragile parts of body to fracture
A:
1. Forearm
2. Spine
3. Femur

DEXXUM T
is the best system to measure
the Bone Mineral Density (BMD)
for forearm, spine and femur
with low dose exposure of X-ray.
2. General information of device

<DEXXUM T>
- X-ray system : DEXA (Dual Energy X-ray Absorptiometry)
- Scanning method : Pencil beam

Detector

BED
2. General information (Exterior)

Control panel Emergency SW

Curtain Film
Wing Bed
Main SW

Bed

AC Inlet
Ethernet port

Caster
2. General information (Exterior)

Carriage
Top Cover
Column Rear
Cover
Carriage
Bottom Cover

Right wing
Left wing Cover
Column Front Cover
Cover

Back
Cover

Left Side Right Side


Cover Front Cover Cover
2. Interior / exterior part’s names and interior module

5 1
1. Carriage 6. Motor driver for X axis
2. Generator (HFG) 7. MDF
3 4 6
3. Main board 8. Motor Y axis
4. Power board 9. DEXXUM T
2
5. Motor driver for Y axis

7
9
8
DEXXUM T Block sketch
4. Installation and operation (DEMO)

1) User program installation


2) Data base installation (Daily test)
3) Data base setting
DEXXUM T user program scanning process
Program start
Measure

Start to analysis

No Yes
Daily test Daily test
error issue? process/ Finish Finished
scanning?

Yes No
Register Patient and select Yes
measurement No Selected scan
Start to scan
area?

Insert (register)
correctly
No Yes Can select
Laser pointer ON reposition
Patient
User panel using selection? No
information?
Position select

Yes Yes
Standard axis moving and Moved to
No
wait the instruction standard axis?
4. Test program

1) Network connection
2) Modification kV, mA
3) Modification for sampling time
4) Get count
5) Get spectrum
6) Get temperature
7) Using scan mode
8) Using user operation switch
9) Shutter test
5. PACS interlock

1) PC setting to connect PACS (LAN CARD)


2) Method to connect PACS
3) OSC connection
6. A/S repairing

1) Instruction of Indicator
 Power Module Indicator
 SMPS Indicator
 Power Board Indicator
 Main Board Indicator
 X-ray Generator Indicator

2) Repair power error

3) Repair device movement’s error

4) Repair X-ray Generator’s error


6. A/S repairing
1) Instruction of indicator – Power Module

Power module is composed Power board(DC/DC) and SMPS(AC/DC).

Main
Emergency Board
Switch Power board
Insert AC
Motor
Driver

SMPS
6. A/S repairing
2) Instruction of indicator – SMPS

When is insert AC power, LED1’s ONSMPS must be always ON.

(
(
(
(
(
AC N (
AC L (
6. A/S repairing
3) Instruction of indicator – Power board

No 1 2 3 4 5 6 7 8

description DC Relay Generator Motor(A) Motor(T) -15V +15V +5V SMPS


ON ON ON
No 9 10 11 12 13 14 15 16

description 24V Generator Motor(A) Motor(T) _15V +15V +5V SMPS ON


Relay In ON (FUSE) (FUSE) (FUSE) (FUSE) (FUSE)

X X X X X X X X
If is ON, only Main Switch
O X X X X X X X

O X O O O O O O If is ON, Main Switch and Emergency


O X O O O O O O Switch

O O O O O O O O
If is ON, Main Switch and Emergency
O O O O O O O O Switch, also Main board ON
6. A/S repairing
4) Instruction of indicator – Main board
Main board has 7 segment indicators for Main board’s movement
and indicator’s Ethernet network connection.

Main board 7-segment indicator (reverse)


7S Description 7S Description
0 Done for initial 6 Saving location’ information

1 1 Laser pointer ON 7 Movement to TOP direction


2 Collect Spectrum data 8 Movement Bottom
2 direction
3 Collect Count data 9 Movement Left direction
4 Scan moving a Movement Right direction
5 Initialing b Automatic scan mode

Right
Ethernet Indicator
WD1 When Link with PC
Bottom Top
WD2 When connect by 100M bps

Left
6. A/S repairing
5) Instruction of indicator – X-ray Generator
X-ray Generator Indicator
ON/OFF Ready state
ON X-ray Shot state

Selectable Switch selection use


115 88VAC ~ 130VAC, 45Hz ~65Hz
230 176VAC ~ 264VAC , 45Hz ~ 65Hz

X-ray Generator’s indicator is OFF even power is connected,


check inserted AC and then open cover, check Drive board is moving.
6. A/S repairing
6) Instruction of indicator – Power error
 Power error’s reason

1. Main switch or emergency switch is OFF

2. Fuse destruction’s case, when input the excess current into device

3. Power supply error (Wall power error)

4. Destruction of main switch or destruction of emergency switch

5. Output of terminal in AC power path or disconnection of cable

6. SMPS of power board’s error

7. Main board’s error


6. A/S repairing
6) Instruction of indicator – Power error

1. Check the main board and emergency switch is ON.

2. Check Fuse state with the naked eyes and multi meter.

Ω
6. A/S repairing
6) Instruction of indicator – Power error

3. Check the Wall Power.

4. Remove the outlet from device, and switch’ state.

Main switch
1pole 2pole
Ω ON

Back View

Switch ON OFF
1 pole 0Ω -Ω Switch ON OFF

2 pole 0Ω -Ω 1 pole 0Ω -Ω
6. A/S repairing
6) Instruction of indicator – Power error
5. Check AC power path.

2 3 NO Descreption

1 Ielet(N,L)
5
1 2 Noise Filter
4
3 Main Switch

4 Inner Inlet

4 4
5 Terminal Block
6. A/S repairing
6) Instruction of indicator – Power error
6. Check the SMPS

Disconnect the main board and HFG’s power board and power ON.
If it approved till connector’s block 220V, but not to SMPS’ indicator and then exchange SMPS.

6. Check the main board and HFG’s state.

When Power enter, do connect the separated main board.


When is no error, do connect HFG.
6. A/S repairing
7) Movement’s error

 The kind of movement’s error

1. No image from selected area and scanning other area.

2. Do not stop with noise when device goes to initial position.

3. The device do not move.

4. It’s very noisy when the device moves.


6. A/S repairing
7) Movement’s error
1. Do not appear the selected scan area and appear unselected area image.

1) Check the User manual.


2) Check the alignment’s Laser pointer.
3) Check the alignment’s between X-ray Generator and Detector.

2. Do not stop with noise when the device is initialing.

1) Check the Limit Switch damage and Cable


2) Check the Limit Block’s location
6. A/S repairing
7) Movement’s error
3. The device does not move.

1) Check the power state of Motor Driver


2) When does not supply power, check the Power Board Indicator and power.
3) Check the Cable’s disconnection state.
4) Check the main module’s movement state and network’s connection state.

2. The device’s movement is very noisy.

1) Check the interior of device if is other strange substance.


2) Check the Cover’s fixed state.
3) Check the Main-board ~ Motor ~ Motor Driver Cable’s state.
6. A/S repairing
7) Movement’s error- X-ray’s error code

 The kind of X-ray exposure’s error

1. Generator power’s error issue(B-031, B-041 )

2. Generator movement’s error issue (B-011,B-012, B-018)

3. No find the X-ray Indicators error, but there is no detection from


Detector.
6. A/S repairing
7) Movement’s error- X-ray’s error code

1. Generator’s power Error.

1) Check with naked eyes the Generator’s Indicator, to estimate the


power’s approve.
2) Check the AC power is insert to Generator.
3) Check the Fuse state, opening the Cover.
4) Measure the CON3 to DC, and then check the results is 350~400VDC,
when is no output, exchange the main power board.

AC Power CON3
6. A/S repairing
7) Movement’s error- X-ray’s error code
2. Generator error code

1) Exchange the Generator for errors of KV, mA, filament Feedback.


2) B-011 error code, check the real exposure doing scintilllator on X-ray
shutter in Get count state of TEST Program. Exchange the detector
the count’s results even the exposure has no problem..

Sintillator shows the green colour when X-ray Energy touch.

3. The is no Indicator’s error, but detection in Detector

1) Check the real exposure doing scintillator on X-ray shutter in Get


count state of TEST Program. Exchange the detector the count’s results
even the exposure has no problem.
7. Firmware Download method

<DEMO>
8. Module exchanging method

<DEMO>

1) Exchange Main board


2) Exchange Power board
3) Exchange Film
4) Exchange Detector
5) Exchange X-Ray Generator
9. Alignment method

<DEMO >

1) Camera-Jig’s installation method

2) Exchange X-Ray Generator’s method

3) Alignment method

4) Alignment method using Scintillator


10. Calibration method

<DEMO >

1) Phantom’s name

2) Calibration method

3) Correction using Reference Phantom

4) Daily TEST

You might also like