Especificaciones Monitor N15 Mindray

This document provides specifications for the BeneVision N17/N15/N12 patient monitors. It details physical characteristics, display features, monitoring capabilities including ECG, respiration, pulse oximetry, non-invasive blood pressure, and temperature. Measurement ranges, accuracy, and technical standards are specified for each monitoring modality.

Uploaded by

Aniela
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
227 views

Especificaciones Monitor N15 Mindray

This document provides specifications for the BeneVision N17/N15/N12 patient monitors. It details physical characteristics, display features, monitoring capabilities including ECG, respiration, pulse oximetry, non-invasive blood pressure, and temperature. Measurement ranges, accuracy, and technical standards are specified for each monitoring modality.

Uploaded by

Aniela
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 4

BeneVision N17/N15/N12

Patient monitor

Physical Specifications QT Analysis


Weight Standard configuration, excluding modules, Patient Adult/Pediatric/Neonate.
recorder, battery and accessories. Parameters QT, QTc, ∆QTc
N17: 7.3 kg (16.1 lbs) QTc Formula Bazett, Fridericia, Framingham, or Hodges
N15: 5.4 kg (11.9 lbs) Range
N12: 4.1 kg (9.1 lbs) QT/QTc: 200 to 800 ms
Size QT-HR: Adult: 15 to 150 bpm
N17: 466 x 355 x 210 mm Pediatric/Neonate: 15 to 180 bpm
N15: 396 x 313 x 193 mm QT Accuracy ± 30 ms
N12: 313 x 290 x 161 mm Resolution QT 4 ms; QTc 1 ms
Display Respiration
Type Medical-grade color TFT LCD, capacitive touch Range 0 to 200 bpm
screen, support multi-touch operation. Resolution 1 rpm
178 ° viewing angle Apnea Alarm Time 10, 15, 20, 25, 30, 35, 40 sec
Screen & Resolution Accuracy
N17: 18.5-inch, 1920 x 1080 pixels (FHD) 0 - 120 rpm: ± 1 rpm
N15: 15.6-inch, 1920 x 1080 pixels (FHD) 121 - 200 rpm: ± 2 rpm
N12: 12.1-inch, 1280 x 800 pixels (WXGA) Lead I, II, or auto (default: lead II)
Waveforms N17: Up to 12 waveforms Pulse Oximetry
N15: Up to 10 waveforms Meet standards of ISO 80601-2-61.
N12: Up to 8 waveforms Module Mindray, Masimo, Nellcor
ECG Range 0 to 100 %
Meet standards of IEC 60601-2-27 and IEC 60601-2-25. Resolution 1%
Lead Sets Automatic 3/5/6/12 - lead recognition Accuracy
3-lead: I, II, III Mindray/Nellcor: ± 2 % (70 to 100%, Adult/Pediatric:)
5-lead: I, II, III, aVR, aVL, aVF, V ± 3 % (70 to 100%, Neonate)
6-lead: I, II, III, aVR, aVL, aVF, Va, Vb Unspecified (0 to 69%)
12-lead: I, II, III, aVR, aVL, aVF, V1 to V6 Masimo: ± 2 % (70 to 100%, Adult/Pediatric, non-motion)
Sweep Speed 6.25 mm/s, 12.5 mm/s, 25 mm/s, 50 mm/s ± 3 % (70 to 100%, Neonate, non-motion)
Gain Selection x 0.125, x 0.25, x 0.5, x 1, x 2, x 4, auto ± 3 % (70 to 100%, motion)
Waveform format Standard, Cabrera Unspecified (0 to 69%)
Input Signal Range ± 8 mV (p-p) Perfusion indicator (PI) Yes, for Mindray/Masimo SpO2
Electrode Offset Potential Tolerance ± 500 mV Pitch Tone Yes
Bandwidth Dual-SpO2 Yes, SpO2, SpO2b, ∆SpO2
Diagnostic Mode: 0.05 to 150 Hz Pulse Rate Range
Monitor Mode: 0.5 to 40 Hz Mindray/Nellcor: 20 to 300 bpm
Surgical Mode: 1 to 20 Hz Masimo: 25 to 240 bpm
ST Mode: 0.05 to 40 Hz Pulse Rate Accuracy
High Freq Cut-off (for 12-lead ECG analysis): Mindray: ± 3 bpm (20 - 300 bpm)
350 Hz, 150 Hz, 35 Hz, 20 Hz selectable Nellcor: ± 3 bpm (20 - 250 bpm)
CMRR Masimo: ± 3 bpm (non-motion)
Diagnostic: > 90 dB ± 5 bpm (motion)
Monitor, Surgical, ST mode: PR Refresh Rate 1 sec
> 105 dB (with notch filter on) Temperature
Pace detection Meet standard of ISO 80601-2-56.
Amplitude: ± 2 mV to ± 700 mV Method Thermal resistance
Width: 0.1 to 2 ms Channels Up to 8 channels
Rise time: 10 to 100 μs (without overshoot) Units of Measure Selectable °C or °F
Defibrillator Protection Withstand 5000VAC (360J) defibrillation Range 0 to 50 °C / 32 to 122 °F
Defib. Recovery Time ≤ 5 seconds Resolution 0.1 °C, 0.1°F
ESU recovery time ≤ 10 s Accuracy ± 0.1 °C or ± 0.2 °F (without probe)
Provides Glasgow resting 12-lead ECG algorithm. Refresh Rate 1 sec
Provides Mindray Multi(4)-lead ECG monitoring analysis algorithm. Genius TM 2 Tympanic Thermometer
(* These ECG specifications are from MPM Platinum module.) Measurement Range 33 to 42 °C / 91.4 to 107.6 °F
Heart Rate Calibrated Accuracy ± 0.1 °C (environment temperature 25 °C,
Measurement Range target temperature 36.7 to 38.9 °C)
Adult: 15 to 300 bpm ± 0.2 °C (environment temperature 16 °C,
Pediatric/Neonate: 15 to 350 bpm target temperature 33 to 42 °C)
Accuracy ± 1 bpm or ± 1%, whichever is greater. Resolution 0.1 °C, 0.1°F
Resolution 1 bpm Response Time < 2 sec
Arrhythmia Analysis Non-Invasive Blood Pressure
Patient Adult/Pediatric/Neonate. Meet standards of ISO 80601-2-30.
Monitored Arrhythmias Asystole, VFib/VTac, VTac, Vent. Brady, Extreme Method Oscillometry
Tachy, Extreme Brady, Vrhythm, PVCs/min, Modes Manual, Auto, STAT, Sequence
Pauses/min, Couplet, Bigeminy, Trigeminy, R on T, Units of Measure mmHg, kPa (user-selectable)
Run PVCs, PVC, Tachy, Brady, Missed Beats, PNP, Resolution 1 mmHg
PNC, Multif. PVC, Nonsus. VTac, Pause, Irr. Rhythm, Systolic range
AFib. Adult: 25 to 290 mmHg
ST Segment Analysis Pediatric: 25 to 240 mmHg
Patient Adult/Pediatric. Neonate: 25 to 140 mmHg
Range - 2.0 to + 2.0 mV (RTI) Diastolic range
Accuracy ± 0.02 mV or ± 10%, whichever is greater Adult: 10 to 250 mmHg
(- 0.8 to + 0.8 mV) Pediatric: 10 to 200 mmHg
Resolution 0.01 mV Neonate: 10 to 115 mmHg
Mean range
Adult: 15 to 260 mmHg 77 to 99 mmHg: ± 10% of reading
Pediatric: 15 to 215 mmHg 100 to 150 mmHg: ± (3 mmHg+8% of reading)
Neonate: 15 to 125 mmHg O2 Accuracy
Accuracy 0 to 25 %: ±1 %
Max Mean Error: ± 5 mmHg 25.1 to 80 %: ±2 %
Max Standard Deviation: 8 mmHg 80.1 to 100 %±3 %
Cuff Deflation Technique Step bleed Resolution
Initial Cuff Inflation etCO2: 1 mmHg
Adult: 80 to 280 mmHg (default: 160 mmHg) O2 (optional) : 1%
Pediatric: 80 to 210 mmHg (default: 140 mmHg) Sample Flow Rate
Neonate: 60 to 140 mmHg (default: 90 mmHg) Adult/Pediatric: 120 ml/min (with or without O2 monitoring)
Over Pressure Protection Neonate: 70 ml/min or 90 ml/min, selectable
Adult/ Pediatric: 297 ± 3 mmHg 90 ml/min (with O2 monitoring)
Neonate: 147 ± 3 mmHg Sample Flow Rate Tolerance
Max Measurement time ±15 ml/min or ±15 %, whichever is greater.
Adult/Pediatric: 180 sec Warm-up Time 90 sec (maximum), 20 sec (typically)
Neonate: 90 sec Measured with a neonatal watertrap and 2.5-meter neonatal sampling line, or
Assisting Venous Puncture Yes an adult watertrap and a 2.5-meter adult sampling line:
Pulse Rate Range 30 to 300 bpm Rise Time
Pulse Rate Accuracy ± 3 bpm or ± 3 %, whichever is greater etCO2: ≤ 250 ms @ 70 ml/min (Neonate watertrap)
IBP ≤ 250 ms @ 90 ml/min (Neonate watertrap)
Meet standard of IEC 60601-2-34. ≤ 300 ms @ 120 ml/min (Adult watertrap)
Number Up to 8 channels O2 (optional) : ≤ 800 ms @ 90 ml/min (Neonate watertrap)
Measurement Range -50 to 360 mmHg ≤ 750 ms @ 120 ml/min (Adult watertrap)
Resolution 1 mmHg Sampling Delay Time
Accuracy ± 1 mmHg or ±2 %, whichever is greater (excluding etCO2: ≤ 5.0 sec @ 70 ml/min (Neonate watertrap)
sensor error) ≤ 4.5 sec @ 90 ml/min (Neonate watertrap)
Sensitivity 5 μV/V/mmHg ≤ 5.0 sec @ 120 ml/min (Adult watertrap)
Impedance Range 300 to 3000 Ω O2 (optional) : ≤ 4.5 sec @ 90 ml/min (Neonate watertrap)
PPV Range 0 to 50 % ≤ 5.0 sec @ 120 ml/min (Adult watertrap)
PAWP Yes awRR Range 0 to 150 rpm
ICP measurement Support awRR Accuracy
Support waveforms overlapping. 0 to 60 rpm: ± 1 rpm
Pulse Rate Range 25 to 350 bpm 61 to 150 rpm: ± 2 rpm
Pulse Rate Accuracy ±1 bpm or ±1 %, whichever is greater Apnea Time 10, 15, 20, 25, 30, 35, 40 sec
Cardiac Output Provide VCO2, VO2, MVCO2, MVO2, EE, RQ parameters, when monitoring with RM
Method Thermodilution module.
Measurement Range 0.1 - 20 L/min Oridion Microstream CO2
Resolution 0.1 L/min Measurement Range 0 to 99 mmHg
Accuracy ±0.1 L/min or ±5%, whichever is greater Resolution 1 mmHg
TB Range 23 to 43 °C / 73.4 to 109.4 °F Accuracy
TB, TI Accuracy ± 0.1 °C (without sensor) 0 to 38 mmHg: ±2 mmHg
TB, TI Resolution 0.1 °C 39 to 99 mmHg: ±5 % + 0.08 % of the reading – 38 mmHg
PiCCO Sample Flow Rate 50 -7.5+15 ml/min
Parameters Measurement Range Coefficient of Variation Start-up Time 30 sec (typical)
CCO 0.25 to 25.0 L/min ≤ 2% Response Time 2.9 s (typical)
C.O. 0.25 to 25.0 L/min ≤ 2% awRR Range 0 to 150 rpm
GEDV 40 to 4800 ml ≤ 3% awRR Accuracy
SV 1 to 250 ml ≤ 2% 0 to 70 rpm: ±1 rpm
EVLW 10 to 5000 ml ≤ 6% 71 to 120 rpm: ±2 rpm
ITBV 50 to 6000 ml ≤ 3% 121 to 150 rpm: ±3 rpm
(Coefficient of variation is measured using synthetic and/or database wave Apnea time 10, 15, 20, 25, 30, 35, 40 sec
forms (laboratory testing.) Coefficient of variation= SD/mean error.) Capnostat Mainstream CO2
TB Range 23 to 43 °C / 73.4 to 109.4 °F Measurement Range 0 to 150 mmHg
TB, TI Accuracy ± 0.1 °C (without sensor) Resolution 1 mmHg
TB, TI Resolution 0.1 °C Accuracy
pArt/pCVP Range -50 to 300 mmHg 0 to 40 mmHg: ± 2mmHg
pArt/pCVP Accuracy ± 1 mmHg or ± 2 %, whichever is greater 41 to 70 mmHg: ± 5% of reading
ScvO2 71 to 100 mmHg: ± 8% of reading
Range 0 to 99 % 101 to 150 mmHg: ± 10% of reading
Accuracy ± 3% (50 to 80 %) Rise time < 60 msec
ICG awRR Range 0 to 150 rpm
Method Thoracic electrical bioimpediance (TEB) awRR Accuracy ±1 rpm
HR Range 40 to 200 bpm (ICG), accuracy ±2 bpm Provide VCO2, MVCO2, FeCO2, SlopeCO2, Vtalv, MValv, Vdaw, Vdaw/Vt, Vdalv,
C.O. Range 1.0 to 15 L/min Vdalv/Vt, Vdphy, Vd/Vt, when monitoring with RM module.
SV Range 5 to 250 ml Anesthesia Gases
Provides Monitoring Parameters ACI, VI, PEP, LVET, TFI, TFC, HR, C.O., C.I., SV, SVI, Meet standard of ISO 80601-2-55.
SVR, SVRI, PVR, PVRI, LCW, LCWI, LVSW, LVSWI, STR, VEPT Sampling Rate
Continuous Cardiac Output Interface Adult/pediatric: 200 ml/min
Measured Parameter Consistent with CCO-related parameters outputted Neonate: 120 ml/min
by Vigilance II®, Vigileo™, or EV1000 Sampling Rate Tolerance ±10 ml/min or ±10%, whichever is greater.
Vigilance II: CCO, CCI, C.O., C.I., SV, SVI, SVR, SVRI, RVEF, EDV, EDVI, Sampling Delay Time < 4 sec
ESV, ESVI, TB, SaO2, VO2, O2EI, DO2, ScvO2, SvO2, SQI Refresh Rate 1 sec
Vigileo: CCO, CCI, SV, SVI, SVR, SVRI, ScvO2, SvO2 Warm-up Time 45 sec to warm-up status
EV1000: CCO, CCI, CO, CI, SV, SVI, SVV, SVR, SVRI, GEF, CFI, 10 min to ready-to-measure status
GEDV, ITBV, ITBI, EVLW, EVWI, PVPI Measurement Range
Artema Sidestream CO2 CO2: 0 to 30 %
Meet standard of ISO 80601-2-55. N2O: 0 to 100 %
Measurement Range Des/Sev/Enf/Iso/Hal:
etCO2: 0 to 150 mmHg 0 to 30 %
O2 (optional) : 0 to 100 % O2: 0 to 100 %
CO2 Accuracy awRR: 2 to 100 rpm
0 to 40 mmHg: ± 2mmHg Resolution
41 to 76 mmHg: ± 5% of reading CO2: 0.1 %
N2O: 1% Paw ± 3% of reading
Des/Sev/Enf/Iso/Hal: MVe/MVi ± 10% of reading
0.1 % TVe/TVi Adult/Pediatric: ±10% or ±15 ml, whichever is
O2: 1% greater.
awRR: 1 rpm Infant: ±10% or ±6 ml, whichever is greater.
Full Accuracy awRR: ±1 rpm (4 to 99 rpm)
Gases Range (%REL) Accuracy (%ABS) ±2 rpm (100 to 120 rpm)
CO2: 0 to 1 % ± 0.1 % Provide loops display.
1 to 5 % ± 0.2 % Monitoring parameters include PEEP, Pmean, PIP, Pplat, PEF, PIF, MVe, MVi, TVe,
5 to 7 % ± 0.3 % TVi, RR, I:E, FEV1.0, Compl, RSBI, NIF, WOB, RAW.
7 to 10 % ± 0.5 % rSO2
> 10 % Not specified Patient Adult/Pediatric/Neonate.
N2O: 0 to 20 % ±2% Method INVOS, NIRS (Near Infrared Spectroscopy)
20 to 100 % ±3% Number Up to 4 channels
Des: 0 to 1 % ± 0.15 % Measurement Range 15 to 95 %
1 to 5 % ± 0.2 % NMT
5 to 10 % ± 0.4 % Meet the standard of IEC 60601-2-10
10 to 15 % ± 0.6 % Sensor Type Acceleromyography sensor
15 to 18 % ±1% Stimulation Modes ST, TOF, PTC, DBS3.2, DBS3.3
> 18 % Not specified Stimulation Current Range
Sev: 0 to 1 % ± 0.15 % 0 to 60 mA
1 to 5 % ± 0.2 % Stimulation Current Accuracy
5 to 8 % ± 0.4 % ± 5% or ±2 mA, whichever is greater.
>8% Not specified Stimulation Pulse Width 100,200 or 300μs,monophasic rectangle pulse
Enf/Iso/Hal: 0 to 1 % ± 0.15 % Stimulation Pulse Width Accuracy
1 to 5 % ± 0.2 % ± 10 %
>5% Not specified Max. Output Voltage 300 V
O2: 0 to 25 % ±1% BISx/BISx4
25 to 80 % ±2% Meet standard of IEC 60601-2-26.
80 to 100 % ±3% Method Bispectral Index
awRR: 2 to 60 rpm ± 1 rpm Impedance Range 0 to 999 kΩ
> 60 rpm Not specified EEG Bandwidth 0.25 to 100 Hz
Rise Time BIS Range 0 to 100 (BIS, BIS L, BIS R)
Sampling flow 120 ml/min, using the DRYLINE II TM watertrap and a SQI Range 0 to 100 % (SQI, SQI L, SQI R)
neonatal 2.5m sampling line, ASYM 0 to 100%
CO2/ N2O: ≤ 250 ms DSA Trend Yes
Iso/Hal/Sev/Des: ≤ 300 ms EEG
Enf: ≤ 350 ms Meet standard of IEC 60601-2-26.
O2: ≤ 600 ms EEG Channels Up to 4 channels
Sampling flow 200ml/min, using DRYLINE II TM watertrap and an adult Montage Mode Biopolar mode, referential mode
2.5m sampling line: Input Signal Range - 2 mVp-p to + 2mVp-p
CO2/ N2O: ≤ 250 ms Max. Input DC Offset ± 500 mV
Iso/Hal/Sev/Des: ≤ 300 ms CMRR ≥ 100 dB @51 kΩ imbalance and 60 Hz
Enf: ≤ 350 ms Noise Level ≤ 0.5 μV rms (1 Hz to 30 Hz)
O2: ≤ 500 ms Differential Input Impedance
Sampling Delay Time > 15 MΩ @10 Hz
Sampling flow 120 ml/min, using the DRYLINE II TM watertrap and a Electrode Impedance
neonatal 2.5m sampling line, Range 0 to 90 kΩ
CO2: ≤ 4 sec Accurancy ± 1 kΩ or ± 10%, whichever is greater
N2O: ≤ 4.2 sec Sampling Frequency 1024 Hz
O2: ≤ 4 sec Analog bandwidth 0.5 to 110 Hz
Enf /Iso/Hal/Sev/Des: ≤ 4.4 sec Spectrum analysis SEF, MF, PPF, TF, Delta, Theta, Alpha, and Beda
Sampling flow 200ml/min, using DRYLINE II TM watertrap and an adult Trend DSA, CSA
2.5m sampling line: tcGas
CO2: ≤ 4.2 sec Interfaces with TCM CombiM, TCM TOSCA or SenTec SDM monitor.
N2O: ≤ 4.3 sec Measurement Range
O2: ≤ 4 sec tcpCO2 5 to 200 mmHg
Enf/Iso/Hal/Sev/Des: ≤ 4.5 sec tcpO2 0 to 800 mmHg
Apnea time 10,15,20,25,30,35,40 sec SpO2 0 to 100 %
Provide MAC value (support calibrated by age). PR 25 to 240 bpm
Support two mixed gas identify and monitoring. Power 0 to 1000 mW
RM Accuracy
Method Diff-Pressure flow tcpCO2 TOSCA Sensor 92, tc Sensor 54:
Measurement Range Better than 1 mmHg (1 % or 10 % CO2)
Flow Adult/Pediatric: ± (2 to 120) L/min Better than 3 mmHg (33 % CO2)
Neonate: ± (0.5 to 30) L/min tc Sensor 84:
Paw -20 to 120 cmH2O Better than 1 mmHg (1 % or 10 % CO2)
MVe/MVi Adult/Pediatric: 2 to 60 L/min Better than 5 mmHg (33 % CO2)
Infant: 0.5 to 15 L/min tcpO2 tc Sensor 84:
TVe/TVi Adult/Pediatric: 100 to 1500 ml Better than 1 mmHg (0 % O2)
Infant: 20 to 500 ml Better than 3 mmHg (21 % O2)
awRR range 4 to 120 rpm Better than 5 mmHg (50 % O2)
Resolution Better than 25 mmHg (90 % O2)
Flow 0.1 L/min SpO2 ±3 % (70 to 100 %)
Paw 0.1 cmH2O PR ±3 bpm
MVe/MVi 0.01 L/min (MVe/MVi < 10 L/min) Power ±20 % of reading
0.1 L/min (MVe/MVi ≥ 10 L/min) iView (for N17 only)
TVe/TVi 1 ml CPU Intel Pentium N4200 2.5GHz
awRR: 1 rpm Memory 8 GB
Accuracy Hard-disk mSATA SSD 128GB
Flow Adult/Pediatric: ± 1.2 L/min or ± 10% of the OS Windows 10
reading, whichever is greater. Recorder
Neonate: ± 0.5 L/min or ± 10%, whichever is Type Thermal array
greater. Speed 25 mm/sec, 50 mm/sec
Trace Up to 3 (paper 50 mm width, 20 m length) ST Mode: 0.05 to 40 Hz
Supports integrated recorder module. QRS Delay ≤ 25 ms (in diagnostic mode, and non-paced)
Alarms Sensitivity 1 V/mV, ± 5 %
Audible indicator Yes, 3 different alarm tones, and prompt tone Pace Enhancement
Visible indicator Red/yellow/cyan LED, and alarm message Signal Amplitude: Voh ≥ 2.5 V
Provide AlarmSight infographic alarm indicator. Pulse Width: 10 ms ± 5 %
Data Storage Signal Rising and Falling Time:
Trends Data > 120 hrs @ 1min, 4 hrs @ 5 sec. ≤ 100 μs
Events 1000 events, including parameter alarms, IBP Analog Output
arrhythmia events, technical alarms, and so on. Bandwidth (- 3 dB; reference frequency: 10 Hz)
NIBP 1000 sets 0 to 40 Hz
Interpretation of resting 12-lead ECG results Max. Transmission Delay 30 ms
20 sets Sensitivity 1 V/100 mmHg, ± 5 %
Full disclosure 48 hours for all parameters and waveforms Interfacing
(8G storage card) AC Power Connector 1
48 hours at maximum. The specific storage time RJ45 Network Connector, 100 Base-TX, IEEE 802.3
depends on the waveforms stored and the number N17: 2 (1 for iView)
of stored waveforms. (2G storage card) N15/N12: 1
OxyCRG 48 hrs USB 2.0 Connector
ST review 120 hrs @1 min N17: 8 (4 for iView)
Minitrend Yes N15/N12: 4
Special Functions Nonstandard USB SMR Connector
Clinical Assistive Application (CAA): N17/N5: 1 to connect SMR, N1/T1 docking station
HemoSightTM, ST GraphicTM, SepsisSightTM, BoA N12: 1 to connect N1/T1 docking station
DashboardTM, EWS, GCS, 24hrs ECG Summary, Pace Standard DVI-D Video Interface Connector
View N17: 2 (1 for iView)
Support calculations (drug, hemodynamic, Oxygenation, Ventilation, Renal), N15/N12: 1
and Titration table. BNC Connector 1
Support wireless connection with BeneVision TM80 and BP10. Equipotential Grounding Terminal
Support nView remote display tool 1
Wi-Fi Communications Multifunction Connector for Defib Sync and Analog Output
Protocol IEEE 802.11a/b/g/n 1 on multi-parameter module
Modulation Mode DSSS and OFDM Module Slot
Operating Frequency N17/N15: 6 slots
IEEE 802.11b/g/n (2.4G): N12: 4 slots
ETSI/FCC/KC: 2.4 to 2.483 GHz Barcode Scanner Support 1D and 2D barcode
MIC: 2.4 to 2.495 GHz Keyboard & Mouse Support wire and wireless type via USB
IEEE 802.11a/n (5G): Remote Control Support
ETSI: 5.15 to 5.35 GHz, 5.47 to 5.725 GHz Network Printer Support
FCC: 5.15 to 5.35 GHz, 5.725 to 5.82 GHz Battery
MIC: 5.15 to 5.35 GHz Type Rechargeable lithium-ion
KC: 5.15 to 5.35 GHz, 5.47 to 5.725 GHz, Number of Battery 1
5.725 to 5.82 GHz Capacity 4500mAh, 11.1 VDC
Channel Spacing 5 MHz @ 2.4 GHz (802.11 b/g/n) Run Time when powered by a new fully-charged battery
20 MHz @ 5 GHz (802.11 a/n) at 25 ºC±5 ºC with 5-lead ECG , SpO2, and auto
Wireless Baud Rate IEEE 802.11a: 6 to 54 Mbps NIBP measurements every 15 min, and screen
IEEE 802.11b: 1 to 11 Mbps brightness set to 1.
IEEE 802.11g: 6 to 54 Mbps N17/N15: > 2 hrs
IEEE 802.11n: 6.5 to 72.2 Mbps N12: > 4 hrs
Output Power < 20dBm (CE requirement: detection Recharge Time 4.5 hrs to 90% when the monitor is off.
mode- RMS) Power Requirements
< 30dBm (FCC requirement, detection AC Voltage 100 to 240 VAC (±10 %)
mode- peak power) Current 2.0 to 0.9 A
Operating Mode Infrastructure Frequency 50 Hz/60 Hz (±3 Hz)
Data Security WPA-PSK, WPA2-PSK, WPA-Enterprise, Environmental requirements
WPA2-Enterprise (EAP-FAST, EAP-TLS, EAP-TTLS, Temperature Operating: 0 to 40 °C (32 to 104 °F)
PEAP-GTC, PEAP-MSCHAPv2, PEAP-TLS, LEAP)
Encryption: TKIP and AES Storage: -20 to 60 °C (-4 to 140 °F )
MPAN Communications Humidity Operating: 15 to 95 % (non condensing)
Modulation Mode GFSK Storage: 10 to 95 % (non condensing)
Operating Frequency 2402 to 2480 MHz Barometric Operating: 427.5 to 805.5 mmHg (57.0 to 107.4
Channel Spacing 2 MHz
kPa)
Wireless Baud Rate 1 Mbps
Output Power ≤ 2.5 mW Storage: 120 to 805.5 mmHg (16.0 to 107.4 kPa)
Data Security Private protocol Safety
MPAN is used in device pairing for BeneVision TM80, BP10 NIBP module and Type of Protection Class I
BeneVision N series patient monitor. Degree of Protection MPM/IBP/C.O./NMT/EEG module: CF
Output ScvO2/CO2/AG/BIS/rSO2 module: BF
Auxiliary Output
Standard Meets the requirements of ANSI/AAMI/IEC Protection Against Ingress of Fluids
60601-1 for short-circuit protection and leakage IPX1
current --------------------------------------------------------------
ECG Analog Output Some of functions marked with an asterisk may not be available. Please
Bandwidth (- 3 dB; reference frequency: 10 Hz)
contact your local Mindray sales representative for the most current
Diagnostic Mode: 0.05 to 150 Hz
Monitor Mode: 0.5 to 40 Hz information.
Surgical Mode: 1 to 20 Hz
| Healthcare with reach are registered trademarks or trademarks owned by Shenzhen Mindray Bio-medical Electronics Co., LTD
Mindray Building, Keji 12th Road South, © 2018 Shenzhen Mindray Bio-medical Electronics Co., Ltd. All rights reserved. Specifications subject to changes without prior notice.
High-tech Industrial Park, Nanshan, Shenzhen 518057, P.R. China P/N: ENG-BeneVision N17/N15/N12 Datasheet-210285x4P-20190610
Tel: +86 755 8188 8998 Fax: +86 755 26582680
E-mail: [email protected] www.mindray.com

You might also like