                  Dear customer
                  You would like to have and will have many years of satisfaction
                  with your páêçå~ X-ray unit.
                  Safety and reliability are necessary to ensure this.
                  Your dental dealership offers you service by specially trained
                  engineers for this purpose.
                  The maintenance should ensure that your product is
                  permanently safe and operational. All components subject to
                  normal wear and tear are checked and, if necessary, replaced.
                  Maintenance work may be performed by the operator only if
                  this is described on the part of Sirona Dental Systems GmbH,
                  otherwise only authorized service engineers of Sirona Dental
                  Systems GmbH or its authorized dealers may be entrusted
                  with the work.
                  In case you have not concluded a maintenance contract,
                  please contact the customer service department of your dental
                  dealership.
                  The performed maintenance must be documented in this
                  document, which must be kept with the unit.
                  We wish you much success and pleasure with your quality
                  product
                  from pfolk^=aÉåí~ä=póëíÉãëK
60 82 346 D 3285
D 3285.105.01.02.02
Sirona Dental Systems GmbH
List of Contents
                                                                                                          ÇÉìíëÅÜ
                             1   Yearly inspection by the system
                                 owner....................................................5
                                 1.1    Technical documents ............................... 5
                                 1.2    System accessories ................................. 6
                                 1.3    Surfaces of the unit .................................. 7
                                 1.4    X-ray unit .................................................. 8
                                 1.5    System safety .......................................... 8
                                 1.6    Exposure safety ....................................... 9
                                 1.7    Key functionality ....................................... 9
                                 1.8    Light localizer functionality (if present) ... 10
                                 1.9    Film cassette holder functionality ........... 10
                                                                                                          ÉåÖäáëÜ
                                 1.10   Image receptor functionality ................... 11
                                 1.11   Image quality .......................................... 11
                                 1.12   Height adjustment functionality .............. 12
                                 1.13   Documenting your yearly inspection ...... 12
                             2   Maintenance by the service
                                 engineer .............................................13
                                 2.1    Customer data .......................................        13
                                 2.2    Checking the records .............................           14
                                 2.3    Labels on the system .............................           14
                                 2.4    Serial numbers of the system ................                15
                                 2.5    Functional test ........................................     16
                                 2.6    Assessing X-ray exposures ...................                17
                                 2.7    Checking the tube data ..........................            17
                                 2.8    Diaphragm check ...................................          18
                                 2.9    Checking the cables for damage ...........                   18
                                 2.10   Safety checks .........................................      19
                                 2.11   Final work ...............................................   20
                                 2.12   Documenting maintenance work ............                    20
60 82 346 D 3285
D 3285.105.01.02.02                                                                                  3
    Sirona Dental Systems GmbH
              60 82 346 D 3285
4          D 3285.105.01.02.02
Sirona Dental Systems GmbH                                                            1 Yearly inspection by the system owner
1                 Yearly inspection by the system
                  owner
Inspection and maintenance must be performed at predetermined intervals to guarantee the health and safety
of all patients, users and third parties.
In order to guarantee the operational safety and reliability of your product, you as the system owner should
check your unit through at least once a year or commission your dental depot to do so.
In addition, your dental depot offers you maintenance of the system by specially trained engineers; see
Chapter 2 “Maintenance by the service engineer”.
All inspection and maintenance work performed by the system owner and/or service engineer must be
recorded in this document and kept near the unit!
                                                                                                                                ÉåÖäáëÜ
If one or more checks to be performed do not lead to satisfactory results, please contact your dental depot.
1.1 Technical documents
 Technical documents                                            Answer questions with yes (9) or no (−)
 Date (please enter month/year)
 Operating Instructions available?
 Installation Instructions available?
 On-site installation, dimensions,
 technical data available?
 Installation Report / Warranty
 Passport available?
 Declaration of Conformity by sys-
 tem integrator available?
 X-ray System Logbook available
 (for Germany only)?
                                        If no Ö Order missing documents from your dental depot.
60 82 346 D 3285
D 3285.105.01.02.02                                                                                                        5
1 Yearly inspection by the system owner                                                              Sirona Dental Systems GmbH
1.2 System accessories
    System accessories (see Operating Instructions)                   Answer questions with yes (9) or no (−)
    Not all accessories listed here may be included in
    the scope of supply.
    Date (please enter month/year)
                    Needle phantom
                    available?
                    Test phantom avai-
                    lable?
                    Special wrench
                    available?
                    Adjusting tool avai-
                    lable?
                    Bite block, yellow
                    available?
                    Contact segment,
                    yellow available?
                    Localizer avai-
                    lable?
                    Ear holders avai-
                    lable?
                    FH-positioner avai-
                    lable?
                    Contact support
                    piece available?
                                             If no Ö Order missing accessories from your dental depot.
                                                                                                               60 82 346 D 3285
6                                                                                                           D 3285.105.01.02.02
Sirona Dental Systems GmbH                                                                1 Yearly inspection by the system owner
1.3 Surfaces of the unit
 Cover parts                                                        Answer questions with yes (9) or no (−)
 Date (please enter month/year)
 All cover parts attached at X-ray
 unit (Fig. 1)?
 All screws on unit housing avai-
 lable?
 No cover parts damaged (no sur-
 face damages, e.g. scratches)?
 All cover parts in clean condition?
 Clean X-ray unit with the recommen-
 ded cleaning agents
 (see Operating Instructions).
                                                                                                                                    ÉåÖäáëÜ
                                          If no Ö Inform your dental depot.
                                    Fig. 1:
                                                                              Only with ORTHOPHOS 3 DS
60 82 346 D 3285
D 3285.105.01.02.02                                                                                                            7
1 Yearly inspection by the system owner                                                                           Sirona Dental Systems GmbH
1.4 X-ray unit
    X-ray tube assembly and stand condition                                  Answer questions with yes (9) or no (−)
    Date (please enter month/year)
    Tube assembly housing palpably/
    visibly oil-free (Fig. 1)?
    Drive spindle free from foreign
    objects and lubricated (Fig 1)?
    Foam shock absorber available
    (Fig 1)?
                                             If no Ö Inform your dental depot.
                                                                          Tube assembly housing
                                                                        Spindle
                                                       Rear side
                                                                                  Foam shock absorber
                               Fig. 1:
1.5 System safety
    Safety                                                                   Answer questions with yes (9) or no (−)
    Date (please enter month/year)
    Coiled cable with strain relief and
    anti-kink sleeve OK (Fig. 1)?
    Power cable free of damage
    (Fig. 2)?
    Control cable free of damage
    (Fig. 3)?
    All screws used for attaching the
    X-ray unit to the wall and floor
    tight (Fig. 4)?
    For Japan only: Is the unit imme-
    diately switched off when the
    emergency shutdown switch is
    activated?
                                             If no Ö Immediately order repair by a service engineer.
                          Anti-kink sleeve
    Fig. 1:                                  Fig. 2:                              Fig. 3:               Fig. 4:
                                                         PC        PE
                                                                                                                            60 82 346 D 3285
8                                                                                                                        D 3285.105.01.02.02
Sirona Dental Systems GmbH                                                                         1 Yearly inspection by the system owner
1.6 Exposure safety
 Exposure safety                                                      Answer questions with yes (9) or no (−)
 Date (please enter month/year)
 No repeated occurrence of a cer-
 tain error message (Fig. 1)?
 Optical and acoustic signal pre-
 sent when radiation is released
 (Fig. 2)?
 Exposure is immediately interrup-
 ted when radiation button is
 released prematurely, X-RAY dis-
 play no longer lights up(Fig. 3)?
                                            If no Ö Inform your dental depot.
                                                                                                                                                 ÉåÖäáëÜ
                                                                                                                          X-RAY no
                                                                     X-RAY lights up                                      longer lights up
                                                                                  Exposure time is                        Acoustic signal
                                                                                                                          no longer
     Error                                          P I                           indicated             II.3              sounding
     message     EI                                  60                                                   60
                                                                                  kV/mA lights up
                 0I                                   I0                                                   I0
                                                                                                                          LED above
                                                                                                                          Return key
                                                                                                                          lights up.
  Fig. 1:                               Fig. 2:                                          Fig. 3:
1.7 Key functionality
 Key functionality                                                    Answer questions with yes (9) or no (−)
 Date (please enter month/year)
 Membrane foil of the height adjus-
 tment buttons and multitimer keys
 without cracks or holes (Fig. 1)?
 All numbers and letters on the
 multitimer show up completely?
                                            If no Ö Inform your dental depot.
                                                                                       P I
                                                                                        60
                                                                                         I0
                                  Fig. 1:
60 82 346 D 3285
D 3285.105.01.02.02                                                                                                                          9
1 Yearly inspection by the system owner                                                                        Sirona Dental Systems GmbH
1.8 Light localizer functionality (if present)
 Light localizer                                                        Answer questions with yes (9) or no (−)
 Date (please enter month/year)
 Vertical and horizontal light beam
 visible (Fig. 1)?
 Position of the horizontal light
 beam can be adjusted?
                                            If no Ö Inform your dental depot.
                                    Vertical and
                                    horizontal light beam
                               Fig. 1:
1.9 Film cassette holder functionality
 Film cassette holder                                                   Answer questions with yes (9) or no (−)
 Date (please enter month/year)
 Lever locks when film cassette is
 inserted (Fig. 1)?
 Cassette holder driving cable free
 of damage and not spliced
 (Fig. 2)?
                                            If no Ö Inform your dental depot.
                                                             Lever for locking
                                                             the film cassette
                                                                                              Tension spring
     Fig. 1:                                                               Fig. 2: Driving cable    F
                                                                                                                         60 82 346 D 3285
10                                                                                                                    D 3285.105.01.02.02
Sirona Dental Systems GmbH                                                            1 Yearly inspection by the system owner
1.10 Image receptor functionality
 Image receptor                                                 Answer questions with yes (9) or no (−)
 Date (please enter month/year)
 Image receptor can be removed
 and turned in easily (Fig. 1)?
 Image receptor fits snugly in the
 guide?
                                      If no Ö Inform your dental depot.
                                                                            Rotary knob
                                                                                                                                ÉåÖäáëÜ
                      Fig. 1:
1.11 Image quality
 Image quality                                                  Answer questions with yes (9) or no (−)
 Date (please enter month/year)
 Typical exposure parameters are
 not adjusted in order to achieve a
 constant image quality?
                                      If no Ö Inform your dental depot.
60 82 346 D 3285
D 3285.105.01.02.02                                                                                                       11
1 Yearly inspection by the system owner                                                          Sirona Dental Systems GmbH
1.12 Height adjustment functionality
 Height adjustment                                                  Answer questions with yes (9) or no (−)
 Date (please enter month/year)
 Acoustic signal sounds when
 height is adjusted?
 Height adjustment works without
 jolting and without atypical run-
 ning noises (e.g. clacking, scuf-
 fing)?
                                          If no Ö Inform your dental depot.
1.13 Documenting your yearly inspection
The undersigned confirms that he/she has checked the unit for the above criteria and that he/she has informed
the competent dental depot in case of any defects.
Year          Inspection date:                Name:                           Signature:
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
13.
14.
                                                                                                           60 82 346 D 3285
12                                                                                                      D 3285.105.01.02.02
Sirona Dental Systems GmbH                                                       2 Maintenance by the service engineer
2                 Maintenance by the service engineer
General information for the service engineer
For performing maintenance work, the ORTHOPHOS 3 / 3 Ceph / DS Service Manual
(Order No. 58 35 744) is indispensable. All measures which are required for maintenance are described there
in Chapter “Maintenance” on page 1.
Furthermore, the Operating Instructions which have been supplied together with the unit are required.
NOTE                                                                                                            i
In addition to the yearly inspection by the system owner or persons authorized by him/her, preventive
maintenance has to be performed after 4, 7, 10 years and afterwards every two years.
                                                                                                                         ÉåÖäáëÜ
Should it be necessary to replace defective components along with parts subject to wear, this must be agreed
previously with the customer or dental practice staff.
If components bearing a serial number are replaced, the new serial number must be recorded in the table in
Chapter 2.4‚ “Serial numbers of the system”.
Discontinuation of spare part deliveries:
For each system, the spare part deliveries are discontinued after a certain period of time. The system owner will
be responsible for safety-relevant failures of systems which continue in operation after that time and can no lon-
ger be serviced due to missing spare parts.
2.1 Customer data
Customer
Last name:
First name:
Street:
Post code / city:
Phone:
60 82 346 D 3285
D 3285.105.01.02.02                                                                                                 13
2 Maintenance by the service engineer                                                             Sirona Dental Systems GmbH
2.2 Checking the records
 Checking the records of the                                      Answer questions with yes (9) or no (−)
 yearly inspection by the system owner
 Maintenance interval
 (after years)                            4        7       10       12       14
 Date (enter month/year)
 Yearly inspection by the system
 owner has been performed
 (see Chapter 1)?
 Records are complete?
 All questions answered with yes?
 All work which has led to a nega-
 tive result during the checks have
 been performed?
                                        If no Ö Perform yearly inspection prior to maintenance.
2.3 Labels on the system
 Comparing labels on the system with
 Installation Report / Warranty Passport                          Answer questions with yes (9) or no (−)
 Maintenance interval
 (after years)                            4        7       10       12       14
 Date (enter month/year)
 All labels according to Installation
 Report / Warranty Passport
 affixed (Fig. 1)?
 All labels legible?
                                        If no Ö Replace missing and/or illegible labels.
                           Fig. 1:
                                                                                                            60 82 346 D 3285
14                                                                                                       D 3285.105.01.02.02
Sirona Dental Systems GmbH                                                         2 Maintenance by the service engineer
2.4 Serial numbers of the system
                              Please enter new Serial No. here when replacing a component:
                              X-ray unit                               Date                     Signature
Serial No. (at initial start-up)
Serial No.
Serial No.
                       X-ray tube assembly                             Date                     Signature
Serial No. (at initial start-up)
Serial No.
Serial No.
Serial No.
                                                                                                                           ÉåÖäáëÜ
Serial No.
                                   Tube                                Date                     Signature
Serial No. (at initial start-up)
Serial No.
Serial No.
Serial No.
Serial No.
                              Multitimer                               Date                     Signature
Serial No. (at initial start-up)
Serial No.
Serial No.
Serial No.
                                   Ceph                                Date                     Signature
Serial No. (at initial start-up)
Serial No.
Serial No.
                           Image receptor                              Date                     Signature
Serial No. (at initial start-up)
Serial No.
Serial No.
                           Test phantom                                Date                     Signature
Serial No. (at initial start-up)
Serial No.
Serial No.
60 82 346 D 3285
D 3285.105.01.02.02                                                                                                  15
2 Maintenance by the service engineer                                                          Sirona Dental Systems GmbH
2.5 Functional test
Prior to starting with maintenance work, a functional test must always be performed.
In doing so, it is helpful to read out the error messages, if present.
Refer to Chapter 5.1‚ “Selecting Service routines”.
Inform the customer or dental practice staff about any defects you have detected.
 Checking the system functions
 (observe Operating Instructions)                                 Answer questions with yes (9) or no (−)
 Maintenance interval
 (after years)                             4        7        10    12      14
 Date (enter month/year)
 No atypical running noises
 audible?
 See chapter 7.1
 Drive spindle free from foreign
 objects, drive spindle lubricated,
 foam shock absorber attached?
 See chapter 7.1
 Jolt-free fine positioning possible?
 Correction switch for height adjus-
 tment functioning?
 Acoustic signals audible?
 Forehead support mounted verti-
 cally? See chapter 7.2
 Temple supports can be adjusted
 easily and symmetrically?
 See chapter 7.2
 With Ceph only
 Diaphragm wheel can be rotated
 and engages perceptibly?
 See chapter 7.3
 Film cassette
 Cassette holder engages securely
 in both end positions?
 See chapter 7.4
 Cassette holder driving cable is
 free of damage?
 Image receptor
 Image receptor can be screwed in
 and out easily and fits snugly in its
 guide?
 See chapter 7.5
 Light localizer adjustable?
 See chapter 7.6
 Cephalometer, conventional
 Head support can be rotated?
 See chapter 7.7
                                         If no Ö Correct fault.
                                                                                                         60 82 346 D 3285
16                                                                                                    D 3285.105.01.02.02
Sirona Dental Systems GmbH                                                         2 Maintenance by the service engineer
2.6 Assessing X-ray exposures
Ask to be shown the most recent X-ray exposures taken by the dentist (film or digital).
 Assessing X-ray exposures                                    Answer questions with yes (9) or no (−)
 Maintenance interval
 (after years)                         4        7        10    12      14
 Date (enter month/year)
 Unexposed surrounding border
 present?
 See chapter 7.8
 Exposure sharp (in case of films,
 observe processing conditions)?
 Density of X-ray image OK
 (observe position of density
                                                                                                                           ÉåÖäáëÜ
 switch; in case of film, observe
 processing conditions)?
                                     If no Ö Correct fault.
2.7 Checking the tube data
 Checking the tube data                                       Answer questions with yes (9) or no (−)
 Maintenance interval
 (after years)                         4        7        10    12      14
 Date (enter month/year)
 Actual kV/mA values and prehea-
 ting OK?
 See chapter 7.9
                                     If no Ö Correct fault.
60 82 346 D 3285
D 3285.105.01.02.02                                                                                                  17
2 Maintenance by the service engineer                                                         Sirona Dental Systems GmbH
2.8 Diaphragm check
Perform diaphragm check also with Cephalometer, if present.
 Performing diaphragm check                                      Answer questions with yes (9) or no (−)
 Maintenance interval
 (after years)                            4        7        10    12      14
 Date (enter month/year)
 ORTHOPHOS 3/3 Ceph
 Test phantom exposure/needle
 phantom exposure correct?
 See chapter 7.10
 ORTHOPHOS 3 DS/ 3 Ceph with
 upgrade Kit
 Test phantom exposure/needle
 phantom exposure correct?
 See chapter 7.11
                                        If no Ö Correct fault.
2.9 Checking the cables for damage
 Checking the cables                                             Answer questions with yes (9) or no (−)
 Maintenance interval
 (after years)                            4        7        10    12      14
 Date (enter month/year)
 Power cable OK?
 See chapter 7.12
 Protective ground wire OK?
 Control cable OK?
 Data transfer cable OK?
                                        If no Ö Correct fault.
                                                                                                        60 82 346 D 3285
18                                                                                                   D 3285.105.01.02.02
Sirona Dental Systems GmbH                                                          2 Maintenance by the service engineer
2.10 Safety checks
Move unit to working height. Switch power OFF at the main switch for the building installation or disconnect the
power plug. Remove covers from the unit.
 Performing safety checks                                      Answer questions with yes (9) or no (−)
 Maintenance interval
 (after years)                          4        7        10    12      14
 Date (enter month/year)
 Grounding straps have complete
 and firm contact?
 See chapter 7.13
 Cable shielding OK?
 See chapter 7.14
 Light barrier housing properly
                                                                                                                            ÉåÖäáëÜ
 attached and free of damage?
 See chapter 7.15
 Ring cable free of damage or
 signs of wear?
 See chapter 7.15
 Prestress of flat belt on rotation
 motor M1 sufficient?
 See chapter 7.16
 Counterbalancing disk on rotation
 motor M1 properly attached?
 Protective ground wire resistance
 complies with specifications?
 See chapter 7.17
 Protective ground wire connec-
 tions properly attached?
 Unit’s leakage current complies
 with specifications?
 See chapter 7.17
                                      If no Ö Correct fault.
60 82 346 D 3285
D 3285.105.01.02.02                                                                                                   19
2 Maintenance by the service engineer                                                        Sirona Dental Systems GmbH
2.11 Final work
 Final work
 (observe Operating Instructions)                               Answer questions with yes (9) or no (−)
 Maintenance interval
 (after years)                            4       7       10     12      14
 Date (enter month/year)
 System cleaned?
                                        If no Ö Clean system.
2.12 Documenting maintenance work
The undersigned confirms that he/she has checked the unit for the above criteria and that he/she has handed
over the unit in fully functional condition.
Date of maintenance              Name of engineer                         Signature of engineer
                                                                                                       60 82 346 D 3285
20                                                                                                  D 3285.105.01.02.02
Sirona Dental Systems GmbH   2 Maintenance by the service engineer
Space for remarks
                                                                     ÉåÖäáëÜ
60 82 346 D 3285
D 3285.105.01.02.02                                            21
tÉ=êÉëÉêîÉ=íÜÉ=êáÖÜí=íç=ã~âÉ=~åó=~äíÉê~íáçåë=ïÜáÅÜ=ã~ó=ÄÉ=êÉèìáêÉÇ=ÇìÉ=íç=íÉÅÜåáÅ~ä=áãéêçîÉãÉåíëK
«=páêçå~=aÉåí~ä=póëíÉãë=dãÄe=OMMM                             péê~ÅÜÉW=ÉåÖäáëÅÜ=                         mêáåíÉÇ=áå=dÉêã~åó
a=POURKNMRKMNKMOKMO===NNKOMMR                                 ûKJkêKW=NMS=UVV=                           fãéêáã¨=Éå=^ääÉã~ÖåÉ
páêçå~=aÉåí~ä=póëíÉãë=dãÄe
                      áå=íÜÉ=rp^W                             áå=`~å~Ç~W
c~Äêáâëíê~≈É=PN              páêçå~=aÉåí~ä=póëíÉãë=ii`        páêçå~=`~å~Ç~                         lêÇÉê=kç      SM=UO=PQS=a=POUR
SQSOR=_ÉåëÜÉáã               QUPR=páêçå~=aêáîÉI=pìáíÉ=NMM     PORM=oáÇÖÉï~ó=aêáîÉ=J=råáí=R
dÉêã~åó                      `Ü~êäçííÉI=k`=OUOTP              jáëëáëë~ìÖ~I=låí~êáç=iRi=RvS
ïïïKëáêçå~KÅçã               rp^                              `~å~Ç~