Service Manual Blood Collection Mixer ...
Service Manual Blood Collection Mixer ...
Model CM735
No. CAT.CM735ASM4.2Ee
ISSUE 7 DEC. 2005
Power Source
The Blood Collection Mixer CM735 (A) is operated by either a battery pack or an AC
power adaptor.
Caution
Do not put excessive weight on or push the tray downward in a forceful manner.
(This avoids damage to the weight-sensitive device under the tray.)
Do not place weights over 2 Kg on the tray.
Battery Capacity
Turn power switch off to save the battery when the machine is not to be used for hours.
If a machine, with a fully charged battery pack is left in but unused, the battery will be
exhausted in about 15-20 hours. A machine with a fully charged battery can operate
continuously for approximately 8 hours.
Battery Recharging
A low-voltage symbol flashes on the display panel and an alarm will sound when the
battery voltage gets low. To recharge the battery, just plug the AC power adaptor in
the rear panel jack, or take out battery and put it on the battery charger unit.
Cleaning Instructions
Use a soft cloth and mild detergent to clean the equipment. Do not use paint thinner,
benzene, solvent or strong detergent.
Keeping
Avoid using or leaving the machine in a location where the temperature is higher than
122 °F (50 °C). Keep it away from any heating source or direct sunlight.
i
Table of Contents
-------------------------------------------------------------------------------------------------------------
Page
Important Notes i
Table of Contents ii
List of Figures iii
1. Introduction 1-1
1-1. General 1-1
1-2. Features 1-2
1-3. Specifications 1-5
List of Figures
Page
[Figure 2-1] Top panel 2-1
[Figure 2-2] Rear panel 2-3
[Figure 2-3] Control panel 2-5
[Figure 2-4] LCD display panel 2-7
[Figure 2-5] Tube clamp module 2-9
[Figure 2-6] Battery charger unit 2-11
[Figure 3-1] CPU board functional block diagram 3-2
[Figure 3-2] Power supply board functional block diagram 3-4
[Figure 4-1] Battery charger unit connection 4-6
[Figure 4-2] Tube clamp module 4-8
[Figure 5-1] LCD font layout 5-2
[Figure 6-1] Troubleshooting flow charts 6-3
[Figure 7-1] System wiring diagram 7-2
[Figure 8-1] Exploded view of CM735 (A) 8-2
[Figure 8-2] Power supply board layout 8-3
[Figure 8-3] CPU board layout 8-4
[Figure 8-4] Sensor board layout 8-5
[Figure 8-5] Battery contact board layout 8-5
[Figure 8-6] Battery charger board layout 8-5
iii
1. Introduction
1-1. General
This Service Manual describes technical information of the Blood Collection Mixer
CM735 (A) to aid service and maintenance of the equipment.
1. Introduction
2. Appearance and Functions
3. System Descriptions
4. Standard Operating Procedures
5. Checks and Adjustments
6. Troubleshooting
7. Schematic Diagrams
8. Technical Illustrations
9. Parts List
1-1
1-2. Features
• The Blood Collection Mixer CM735 (A) provides efficient mixing of the blood with
the anti-coagulant in the blood bags.
• It is a lightweight mobile unit with maximum portability.
• The panel display indicates the net volume of blood collected in the blood bag. It
automatically closes the tube clamp when the collected volume reaches the preset
selected figure. The display shows the total weight on the tray at the end of the
donation and elapsed collection time.
• Various audible and visual alarms are provided to ensure accurate and safe collection
and mixing.
The Blood Collection Mixer monitors blood flow rate during the collection and
donation draw time. When the draw is less than 30 ml/minutes, the machine will alarm.
The alarm is both audible and visual. The alarm repeats every 60 seconds if blood flow
remains low.
The battery can be recharged while the mixer is in use if the AC power adaptor is
attached at rear panel jack of main unit. The battery can also be recharged with the
separate battery charger unit.
(A large battery charger, model BC-162, which charges 6 battery packs same time, is
available for purchase from your distributor. With this charger, one pack among 6 is
recharged fast in 1 hour, and other 5 packs in 10 hours.)
1-2
It takes about 10 hours to recharge a pack from exhausted level to fully charged level.
A fully charged battery will operate the mixer continuously for approximately 8 hours.
The blood collection mixer unit can also be powered directly by using the AC power
adaptor. When both a battery pack and a power adaptor are connected at the same time,
the power adaptor supplies operating power and the battery is under recharging.
The tube clamp module can be positioned on either the left side or the right side of the
tray. It can easily be remounted on either side of the tray by users.
The equipment is stored in a soft carrying bag, and weighs less than 4 kilograms.
The Blood Collection Mixer model CM735/D version works either with or without the
Data Collection Unit.
1-3
Collection Mixer Base Collection/Weighing Tray
1-4
1-3. Specifications
1-5
2. Appearance and Functions
2-1
(1) Control Panel - has seven control switches and an LCD display panel.
(2) LCD Display Panel - is a custom-designed display panel on which all information
is shown.
(3) Power Switch - turns the power on and off.
(4) Tube Clamp Module - is detachable, it can be mounted at either the left or right
side of the tray. A screwdriver is necessary to move the module.
(5) Tray - is plastic molded on which blood collection bags are loaded. Tray should be
detached when mixer is stored in the carrying/transportation case.
(6) Handle - is retractable.
(7) Weight Calibration Hole - In this hole is an adjustment screw to calibrate weight
and volume as necessary.
(8) Tray Leveling Screw - makes the tray rotate.
(9) Clamp Module Connection Hole - is used when user wants to install the tube clamp
module at the right side.
(10) LED Lamp - is illuminated when the power switch is on. The LED lamp starts
flashing when collection volume reaches the preset volume at the end of normal
collection process.
2-2
2-2. Rear Panel
2-3
(1) Battery Pack - is rechargeable and replaceable Nickel-Cadmium 12 VDC/2 AH.
(2) Battery Compartment Door - is opened by placing a finger in the hole and pushing
the door plate down.
(3) Power Adaptor Inlet - takes power directly from the AC power adaptor. When the
adaptor is plugged in, it supplies power to the equipment and the battery pack will
be recharged.
(4) Tray - is detachable. At the bottom of the tray attached is a magnetic pad, which
sticks to the metal Tray Mount (5).
2-4
2-3. Control Panel
2-5
(1) [START] Switch - starts collection process. Tray starts rocking, clamp is opened,
main display shows collected blood volume in [ML].
(2) [END] Switch - completes collection process and returns machine to initial state. Tray stops
rocking, main display shows total weight on the tray in [GRAM].
(3) [TIME] Switch - When depressed, it displays elapsed time of the collection. Time
memory is cleared by pressing [START] switch to begin next collection.
(4) [CLAMP] Switch - opens and closes tube clamp. This switch does not work at the
end of a normal collection until [END] switch is pressed.
(5) LCD Display Panel - shows all figures and symbols. Refer to the Section 3-4 for
details.
(6) [UP] Switch - increases the preset target volume by 5 ml each time pressed.
(7) [DOWN] Switch - decreases the preset target volume by 5 ml each time pressed.
(8) [PRESET] Switch - recalls one of the numbers that were factory-set. If the preset
numbers are 450 and 500 ml, the target volume displayed toggles as 500 - 450 -
500 - 450 - ... each time this switch is pressed. Power-on default figure is 500.
These preset numbers and default figure may be changed upon user’s request.
2-6
2-4. LCD Display Panel
2-7
(1) Main Display - displays either total weight, blood volume or elapsed time. [ML],
[GRAM] or [TIME] is displayed to indicate what is being displayed.
(2) Preset Target Volume Display - has three digits. This figure is controlled by [UP],
[DOWN] or [PRESET] switch.
(3) Low Battery Indicator - is illuminated when battery requires recharging. Beep tone
sounds when this symbol is first activated.
(4) [ALARM] indicator - is illuminated when an abnormal condition occurs, such as a
slow draw. Refer to the Section 5-3 for details.
(5) [TIME] indicator - is illuminated only when the [TIME] switch is depressed. The
main display shows the elapsed time in [min : sec].
(6) [BLEED] indicator - is illuminated while blood collection is in progress. When the
[START] button is depressed, this indicator is on and [ML] is on too. This indicator
goes off and [GRAM] is on when the [END] button is depressed.
(7) [END] indicator - is illuminated when the collection is completed. When the
collected volume reaches the preset target volume, this indicator is illuminated.
The tube clamp closes and beeps sound. The front LED lamp also flashes.
(8) [ML] indicator - is illuminated when the [START] button is depressed. It means the
main display is now showing ‘net’ volume of blood collected in the blood bag in
milliliters.
(9) [GRAM] indicator - is illuminated when the machine is not in collection process. It
means the main display is now showing total weight loaded on the tray in grams.
2-8
2-5. Tube Clamp Module
2-9
(1) Tube Clamp - presses blood bag tubing to stop blood flow. This clamp is opened
and closed by [CLAMP] switch except at the end of a normal collection. The [END]
switch must be pressed to indicate completion of the collection cycle, then
[CLAMP] switch is enabled to work. The clamp is automatically opened when the
[START] switch is pressed to start collection. It is automatically closed at the end of
collection, or when the [END] switch is pressed.
(2) Guide Pins - keep the blood bag tubing in proper position.
(3) Connection Plug - is a circular connector.
(4) Screws - should be removed to move the module to the other side.
2 - 10
2-6. Battery Charger and Adaptor
2 - 11
(1) Battery Charger Unit - is a power supply source to recharge the battery pack. The
Output Plug (5) of the AC Power Adaptor (4) is plugged in the Adaptor Inlet (2).
(3) Charging Indicator Lamp - is on during recharging. If the battery is removed from
the charger unit, the lamp will be off.
(4) AC Power Adaptor - converts the AC main line voltage to 18 VDC output.
2 - 12
3. System Descriptions
3-1. General
Blood Collection Mixer CM735 (A) consists of a main unit, a battery pack, a battery
charger unit, an AC power adaptor and a carrying case.
Main unit comprises a tube clamp module, five circuit board assemblies, a drive motor
assembly, a load cell, a display panel and a control key pad. Battery pack is inserted in
the compartment at rear panel of main unit.
Battery charger consists of a circuit board and a power adaptor. The power adaptor is
used for both recharging the battery and for supplying power to the main unit directly.
. CPU Board
. Power Supply Board
. Sensor Board
. Battery Contact Board
. DCU Interface Board
. Weighing Circuit
. Tray Rocking Mechanism
. Tube Clamp Module
Following sections describe operation theory and function of each assembly in detail.
3-1
3-2. CPU Board
The CPU board assembly comprises load cell amplifiers, an A-D converter, memories,
microprocessor, LCD panel and display circuit.
3-2
(1) Load cell amplifiers ;
Load cell generates very weak signal that is proportional to weight applied on it.
OP amp stages U1 and U2 amplify the weak signal by 60 dB and feed it to A/D
converter. A potentiometer VR2 adjusts offset voltage, and VR1 adjusts signal gain.
3-3
3-3. Power Supply Board
Power supply board comprises three voltage regulators, power switching circuit,
battery charging circuit, fuse, battery level check circuit, clamp motor drive circuit,
tray motor drive circuit and a power switch.
[Figure 3-2] represents the functional block diagram of power supply board.
3-4
(1) Voltage regulators ;
U4 is +5 V regulator and U3 is -5 V regulator. Negative voltage is made at U2
and regulated at U3 for components in CPU board.
U1 is a +12 V regulator for incoming power from the power adaptor.
(4) Fuse ;
There used are two electronic fuses.
PTC1 and PTC2 are ceramic fuses, that have positive-thermal-coefficient.
When current flows above some limit due to any trouble in the equipment, fuse
makes open-circuit. When trouble is removed, fuse recovers to normal condition.
3-5
3-4. Sensor Board
The board has a magnetic sensor called Hall sensor. The component senses magnetic
field and generate a signal. As the tray motor assembly has magnets on its cam
structure, this sensor senses rotation angle of motor shaft and tray position.
So tray motion and position information is checked by the sensor and fed to CPU.
The mechanism consists of a geared motor, cam and bearing, tray mount, magnets and
a tray leveling screw.
A turn of motor shaft makes tray rock a full cycle. Numbers of tray rocking and tray
position are checked by sensor board and fed to microprocessor on CPU board.
When tray is biased, it can be corrected by turning the tray leveling screw as it changes
the sensing angle of magnets by Hall sensor on sensor board.
Tube clamp module has a geared motor in it. This module can be mounted either left or
right panel of the equipment. Users can move it by unscrewing three screws.
3-6
3-8. Battery and Battery Charger
3-7
4. Standard Operating Procedures
(1) Open carrying bag and unpack mixer, tray, battery and AC adaptor.
(2) Insert battery pack into the compartment and close the door.
To operate the equipment by AC power, plug the AC power adaptor in the inlet on
the rear panel.
[Note 1] When both the battery and the AC power adaptor are connected together,
the adaptor supplies operating power and the battery is under recharging.
[Note 2] If anything is put on the tray when power is turned on, it is tared at zero
count. Tared weights must remain on the tray throughout the procedures.
If any weight is removed, total weight will have an error of an equal
amount.
4-1
4-2. Operating Procedures
[Note 3] The tubing length from the clamping point to the blood bag should not be
shorter than 30 centimeters (1 ft) to ensure proper rocking movement.
* Tube clamp is opened and the blood begins to flow into blood bag.
* Tray starts rocking.
* [BLEED] indicator is on.
* [ML] appears to indicate the number displayed is ‘net’ volume of blood
collected in blood collection bag, represented in milliliters.
[Note 4] Weight of blood bags and anti-coagulant, and any weights put on tray
are tared when [START] is pressed. All tared weights must remain on the
tray until the collection cycle ends. If any weights are removed, blood
volume will have an error of an equal amount.
[Note 5] Do not touch the tray or blood collection bags during the collection.
If the tray is pushed down, sensed weight may exceed the preset target
volume, then the clamp will close. When this happens, press [CLAMP]
button to re-open the clamp and continue collection. DO NOT press
[END] button, as it will clear the volume previously collected.
[Note 6] When the collected blood volume reaches 20 ml below the preset target volume, the
tray stops rocking until the preset target volume is reached. This process assures accurate
collection with a 1-2 ml tolerance. For example, if the target volume is set at 500 ml,
the tray stops rocking at 480 ml and resumes rocking when the tubing is clamped at
500ml.
[Note 7] If blood flow rate is slow, lower than 30 ml per minute, and remains
low for 1 minute, audible beeps will sound and the [ALARM] indicator will
be activated. Check donor phlebotomy site to ensure blood flow is
adequate. The beep will repeat once every minute until the flow is higher
than 30 ml per minute.
4-2
When the collected volume reaches the preset target volume, the collection ends.
* The tube clamp closes.
* The tray resumes rocking.
* LED lamp starts flashing.
* The [END] indicator appears to show that the collection is finished.
* The [ML] indicator remains on to show the main display indicates the
net volume of blood in milliliters.
* Beeps sound. The beeps will repeat once every 30 seconds until
[END] button is pressed.
* The [CLAMP] button will not work until [END] button is pressed.
[Note 8] Depressing [TIME] button displays donation draw time on main display.
(8) Crimp blood bag tubing and remove phlebotomy needle from the donor.
(9) Press [CLAMP] button to open clamp and remove blood bags.
4-3
4-3. Audible Alarm Signals
Different audible alarms are generated by a buzzer to indicate that the attention of the
phlebotomist is required.
* Press [END] button. Beeps stop and LED lamp stops flashing.
* Crimp blood bag tubing and remove phlebotomy needle.
* Press [CLAMP] button to open clamp and remove blood bags.
B. Slow bleed:
C. Low battery:
4-4
D. Tray removal:
E. 10 minutes passed:
F. 20 minutes passed:
[Remark]
Certain version of the blood collection mixer doesn’t have 10 and 20 minutes alarms
and/or tray removal alarm.
Certain version may have 30 minutes alarm. If no button switch has been pressed for 30 minutes, an
audible alarm starts. The beeps are same as 20 minutes alarm. Pressing any button switch will stop
beeps. This alarm is designed for battery saving purpose.
4-5
4-4. Recharging the Battery
The battery pack can be recharged in two ways. It can be recharged in the mixer unit,
or with the battery charger unit included.
(1) Plug the AC power adaptor into the jack on the battery charger unit.
(2) Plug the AC cord plug into a wall outlet.
(3) Put the battery pack in the charger unit. The charging indicator lamp will be lit.
* Battery pack doesn’t have polarity. Any direction of installation will be good.
(4) Leave plugged in for several hours. Battery will take about 10 hours to fully
recharge from a low voltage indicator.
(5) Unplug the AC cord plug and disconnect battery from charger unit after 10
hours.
4-6
[Note 1] Recharging a battery for more than 10 hours is not harmful to the battery or
any circuit.
[Note 3] A multi-unit battery charger, model BC-162, is available for purchase. The
features of BC-162 charger are;
- Six battery packs are recharged same time.
- One bay gives fast charging, so one pack can be recharged in about 1 hour.
The other five bays give standard charging, which takes 10 hours.
- One bay of this battery charger can be converted to make discharging by
intentionally activating a push button switch. This feature is to be used to
refresh the battery pack by clearing built-up memory in the pack.
4-7
4-5. Moving the Clamp Module
The tube clamp module is detachable and users can change its mounting place either
to the left or to the right side of the mixer.
4-8
5. Checks and Adjustments
This Chapter 5 describes how to check, adjust and calibrate several important parameters
of the equipment.
All procedures described in this chapter can be performed without opening the upper
cover. The test mode subroutines in Section 5-1 are not presented in the User's Manual.
More professional fault isolation and troubleshooting procedures that should be done
with upper cover open are described in Chapter 6.
5-1
5-1. Test Mode Subroutines
For service personnel to check major key components quickly without using any tool,
some software subroutine programs are prepared in the equipment.
One can check LCD fonts and pin connections, button switch failures, the load cell and
its circuit performance with these special test mode subroutines.
If any failure or malfunction is detected at these test mode procedures, fault isolation
and troubleshooting shall be done in accordance with the procedures in Chapter 6.
5-2
B. Button switch check
5-3
5-2. Weight Check
Following procedures describe how to adjust the weighing circuits. This adjustment is
very sensitive and requires time and accuracy, so one should be careful and patient.
(1) Prepare a small size flat (-) tip screwdriver, 2-3 mm (1/16-1/8”) in diameter.
(2) Turn power off and remove the tray.
(3) Remove the hole plug and locate a screw in the top panel (# 7 in Section 2-1).
(4) If the measured figure was smaller than the known weight, turn the screw
clockwise (CW), by about 1/8 of a turn (45 degrees).
If the measured figure was bigger than the known weight, turn the screw counter-
clockwise (CCW), by about 1/8 of a turn (45 degrees).
* Turning it CW will increase displayed figure, and CCW will decrease display.
(5) Install the tray and follow the accuracy check procedures above.
(6) If the equipment is still out of tolerance, repeat from (2) to (5) again.
If the equipment is now in tolerance, replace the hole plug.
[Note 1] The calibration screw is a potentiometer working very sensitive. Do not turn the
screw more than a quarter turn at one cycle of calibration. The screw adjustment
should be done while the power switch is being turned off.
[Note 2] Designed tolerance of blood volume measurement of the net collected blood in
blood bag is - 0 to + 3 milliliters. For example, if the preset target volume is set at
‘500’, it’s normal that the net collected blood volume comes between 500-503 ml.
But there exist two factors that may change the result.
1) The equipment measures weight only. Blood volume is calculated from blood
weight, divided by specific gravity of blood. If the specific gravity of a donor’s
blood is different from the set figure, it will result some error (about 1-2 ml).
2) If the tubing length between clamping point and blood bag is changed, or if it is
not free, it may push or lift the tray more or less. To minimize this type of error,
the tube length should be longer than 30 cm. Refer to [Note 3] in Section 4-2.
5-4
5-4. Tray Leveling
Tray leveling is ensured by a magnetic position sensor and precision mechanism. After
long use for years, the tray might become unlevel. Leveling of the tray can be simply
adjusted by users.
5-5
6. Troubleshooting
6-1. Precautions
(1) Power should be turned off when subassemblies are plugged in or taken out.
(2) Use proper test equipment and tools.
(3) Attention should be taken when checking AC main voltage.
(4) Excessive force should not be applied on the tray and load cell structures.
(5) When assembling upper cover case after troubleshooting, be sure that wiring
harnesses would not touch load cell, motor or any moving parts.
Wiring harnesses should not be too tight.
6-3. Preparation
Before carrying out troubleshooting procedures in Section 6-4 to locate a faulty assembly
or components, follow the procedures below first.
6-1
6-4. Troubleshooting Procedures
The troubleshooting flow charts in the [Figure 6-1] will guide service personnel to a
faulty assembly or components that should be replaced or corrected.
When a faulty assembly is isolated, it should be replaced with a new or good assembly.
In case that service personnel have good knowledge in electronics, proper components
and test equipment, the personnel can troubleshoot to isolate a single faulty component.
Technical information is prepared in the Chapters 7, 8 and 9.
6-2
[Figure 6-1] Troubleshooting Flow Charts (Page 1 of 3)
6-3
[Figure 6-1] Troubleshooting Flow Charts (Page 2 of 3)
6-4
6-5
7. Schematic Diagrams
7-1
7-2
8. Technical Illustrations
8-1
[Figure 8-1] Exploded view of CM735 (A)
8-2
[Figure 8-2] Power Supply Board Layout
8-3
[Figure 8-3] CPU Board Layout
8-4
[Figure 8-4] Sensor Board Layout
8-5
9. Parts List
Each component shall be identified with manufacturer's part number in the Parts List.
Location of a part and component can be found by a Find Number in the [Figure 8-1]
Exploded View or in the printed circuit board layouts in the [Figure 8-2] ~ [Figure 8-6].
Prefix 'A' in a find number denotes that the part is not a single component but an assembly,
and a lower-level parts list of which is presented at another place.
An order for service spare parts to distributors or to manufacturer should be accompanied
with its part number, or with a designation number if it is an independent unit such as the
Battery pack BN-122C or the Battery charger unit BC-129.
9-1
9-2
9-3
9-4
9-5
9-6
9-7
9-8
9-9
9 - 10