Manual tm2101
Manual tm2101
Force:
Resolution 1/250000
Precision 0.1%
Range Free option from 1 N to 20000 kN
Transducer 1--40mV/V
sensitivity
Excitation voltage 9VDC
Suitable for a servo controller with pulse input or a stepper motor controller
with pulse input; and used for AC frequency variable controller.
3. Double click “setup” to start installation program. The welcome dialog will
appear:
4. Click “Next>>” button, read the License Agreement in the following dialog:
5. Check “I accept the License Agreement(s)” option and click “Next>>” to enter
the next step:
6. Click “Next>>” and start the installation, it may take several minutes, please
be patient until the following screen shows.
7. Click “Finish” to complete the installation. It will ask you whether or not to
restart the windows system immediately.
8. Click “Yes” to restart your computer.
9. Copy “TM2101V2.xx” folder to local disk and double click to open it.
10. Right-click the “TestMaster” program icon and choose “Send to...” ->
“Desktop shortcut” menu to create a shortcut on the desktop.
Click “Find target…”, it will open the folder where the old version is.
3. Right-click the “data” folder of the old version, choose “Copy” in the popup
menu:
4. Open the “TM2101V2.42” folder that contains the new version, Right-click to show
the popup-menu, choose “Paste”:
A Dialog will popup to ask you whether to overwrite the “data” folder:
Choose “All” to overwrite all the data.
5. Delete the old shortcut named “TestMater” on the Desktop. Right-click
“TestMaster” executable and choose “Send to..” “Desktop shortcut” in the popup
menu to create a new shortcut:
3.1 Quickstart
A complete testing process includes three steps: Setting before test; execute test
and get test results.
1. Setting before test:
① Select a sample or add a new sample, as Fig.1;
② Select a project or add a new project, as Fig.2;
Fig. 1
Fig. 2
2. Execute test: click ’Test’ button or press its shortcuts ‘F7’ (Refer to 4.1
shortcuts instruction) , when ’Test’ button is in green connected state , it
indicates that testing process is running and awaiting ending automatically or
manual stop the test.
3. Get test results: Testing results will return automatically in the end. You can
also get different test results by choosing other items in the dropdown list. as
Fig.3.
Fig. 3
Notice: Testing should be executed in online state.
3.2 Main Interface Introduction
Main interface show in Fig. 1 and Fig. 2:
Fig. 1
Fig. 2
Software include title bar, menu bar, current volume displaying bar, control panel, testing time
displaying, lifting speed of machine, manual check point, material selection and modification, testing
methods’ selection and modification, control parameters’ selection and modification, results’ selection and
modification, results displaying form, history record operation, report forms operation ,force -
displacement curve display and so on.
Title bar: software name, version information, etc.
Menu bar: Software management and function operation, etc. Details in 3.3 Menu Introduction
Current volume displaying bar: Displaying current value such as Power, displacement. You can choose
displaying items from dropdown list, showing in Fig3.
Fig. 3
Control Panel: Control the machine up and down and testing execution operation and so on.
Rising and falling speed of machine: Adjust lifting and falling speed of machine.
Snap manually: Snap feature points manually in the curve, details in Fig. 3.4 Function detailed
information-Snap the point manually
Material selection and modification: See details in Fig. 3.4 Function detailed information-Material editor
Testing method selection and modification: See details in Fig. 3.4 Function detailed information-Control
method
Select and modify the control parameters: See details in Fig. 3.4 Function detailed information-Control
parameters
Select and modify the results: See details in Fig. 3.4 Function detailed information-Testing Results
Displayed results form: Get required results of testing
History record operation: Open record means open a bypast testing record only one at a time, when
repeated open it would cover the last one; Delete record means delete one or more selected records; Add record
means new open one or more records based on last one.
Report form operation: print form means export report form to the printer for printing execution; Edit
report form means open the report form for editing; Word report form means generate a report form of Word
format according to template of Word report form.
Force-Displacement curve Display: Display test curve. Details in Fig. 3.4 Function detailed information-
Graph Operation
3.3 Menu Introduction
Menu List:
Main menu First level Illustration Second level Illustration
sub-menu sub-menu
Management Log in Details in 3.4 Detailed
(M) explanation of all functions
Log in
Authority Details in3.4 Detailed
Management explanation of all functions
Authority Management
Modify Details in 3.4 Detailed
password explanation of all functions
Modify password
Log out Exit software , short key
Ctrl+Q
Setting(S) Unit System Details in 3.4 Detailed
explanation of all functions
Unit System
Calibration Details in 3.4 Detailed
explanation of all functions
Calibration
Connect Details in 3.4 Detailed COM1- Details in 3.4 Detailed
explanation of all functions- COM6 explanation of all functions
Connect Connect
Other Invalid
Zero(Z) Zeroed Force Zeroed current force value
Zeroed Route Zeroed current route value
Zeroed Zeroed current displacement
Displacemen value
t
Zeroed All Zeroed all of the above
values
Vision(V) Language Simplified Switch to selected language
Chinese of the software
Traditional Details in 3.4 Detailed
Chinese explanation of all functions
English Language Switch
Option Details in 3.4 Detailed
explanation of all functions
Option
Help(H) Help Open help documents
Documents
About… Show information of the
testing system
Register Details in 3.4 Detailed
explanation of all functions
Register
Active Details in 3.4 Detailed
explanation of all functions
Active
3.4 Detailed Information of all functions
1、 Log in
2、 Authority Management
3、 Modify Password
4、 Unit System
5、 Calibration
6、 Online
7、 Swich Language
8、 Option
9、 Register
10、Active
11、Edit Material
12、Control Methods
13、Control Parameter
14、Testing Result
15、Edit Report Form
16、Word Report Form
17、Snap Point Manually
18、Graph Operation
1. Log in
Click Menu “Management”->”Log in”, or press composite key Ctrl+L(3.5 Shortcut Key
lists) to enter Login interface, as shown in Fig.1:
Fig. 1
Choose user, input correct password and click confirm button, then you can use the
software within corresponding user authority. Administrator owns the highest user
authority and Anonymous owns the lowest. When selected ’Always login as the user’, user
will log in automatically as the same user when software start next time.
2. Authority Management
Click menu “Management”->”Authority Management” to enter Authority Management
Interface, as shown in Fig. 2:
Fig. 2
Authority Management is in disabled status when log in as Anonymous, as anonymous has
the lowest administrative privileges which does not have right of authority management.
User permission of all functions can be set in the Authority Management, the selected
item shows that user owns permission of it, otherwise does not have. For example, in
"Authority Management" project of above picture, Administrator is selected while the
tester is not selected, it shows the administrator has the right of authority management
and the tester does not have the permission.
3. Modify Password
Click menu “Management”-> “Modify login password” to enter Modify Login Password
Interface, see Fig.3:
Fig. 3
Modified password is current login user's password. For example, when current login
user is administrator, and then modifies the administrator password, when current login
user is tester, and then modifies the tester’s password.
4. Unit System
Click menu “Setting”->” Unit System”, or press composite key Ctrl+U(3.5 Shortcut
Key lists) to enter Unit System Interface, see Fig.4:
Fig. 4
Choose a unit and accuracy from the drop list, the accuracy means the digits after
the decimal point. And the unit determines the unit of the whole system, the operation of
all variable must according to this unit system.
5. Calibration
Click menu “Setting”->” Calibration”, or press composite key Ctrl+Shift+C (3.5
Shortcut Key lists) to enter Calibration Password Input Interface, see Fig. 5:
Enter into the Calibration Interface after input the right password, see Fig. 6:
Fig. 5
Fig. 6
The Calibration Interface contains five tabs: Force Sensor、Displacement Sensor、Rubber
Extensometer、Metal Extensometer and System Setting.
1. Calibrate the force sensor
Choose the tab “Force Sensor”, see Fig. 6. In the left side, there are four force
sensor channels: “50kg”、“0.0kg”、“0.0kg”、“0.0kg”.
Calibrate process as below:
1). Plug in a force sensor (for example: range of the force sensor is 50kg) to be
calibrated, select a sensor channel on the left side;
2). Input the range of force sensor into “Range”, for example “50kg”;
3). Input zero value into “Zero”, the default value is “0”;
4). Click “Empty” to empty the calibration table;
5). Calibrate the first point: empty the sensor, and input “0” kg in “Standard
Weight Value”, then click button “Sample”, so the AD value and standard weight value
will be input into the calibration table;
6). Calibrate the second point: Add certain value of load (for example 5 kg) on the
sensor, and input “5” kg into the “Standard Weight Value”, then click button
“Sample”, so the AD value and standard weight value of the second point will be input
into the calibration table;
7). Calibrate the third point: The method is same with the second point, if continue
to calibrate the third point, the AD value and standard weight value of the third point
must bigger than the second point, or the calibration will be inaccurate;
8). Calibrate the forth point: The method is same with the second point, if continue
to calibrate the forth point, the AD value and standard weight value of the forth point
must bigger than the second point, or the calibration will be inaccurate;
9). Calibrate the fifth point: The method is same with the second point, if continue
to calibrate the fifth point, the AD value and standard weight value of the fifth point
must bigger than the second point, or the calibration will be inaccurate;
10). Calibrate the sixth point: The method is same with the second point, if continue
to calibrate the sixth point, the AD value and standard weight value of the sixth point
must bigger than the second point, or the calibration will be inaccurate;
11). Click the button “Write the Standard Data into the acquisition”, the
calibration of a sensor will be completed if prompts “write successful”, please retry
if prompts “write failed”;
12). Please redo 1~11 steps if you need to calibrate the second force sensor, or set
the range of other sensors to zero, then click button “Write the Standard Data into the
acquisition”.
Attention: At least two points should be calibrated, step 7~10 can be skipped.
According to the character of most domestic sensors, normally need 3 points to be
calibrated to get very good accuracy, the value of the second point can be set as 3~5% of
the range of sensor, and the value of the third point can be set as 30~50% of the range
of sensor. The AD value and standard weight value of the latter point must bigger than
the point in front if calibrate more than 3 points, or the calibration will be
inaccurate. If the AD value decreases when the “Standard Weight Value” increases, you
can exchange the two signal line of sensor. AD value is about 8400000 when empty, and the
AD value will be proportional change when with load on the sensor, if the AD value
doesn't change when with load on the sensor, there must be some errors on the hardware,
maybe the force sensor damaged or the acquisition card has malfunction, please contact
with manufacturer.
If there is a little difference between measure force value and the standard weight
after calibration, please re-calibrate or modify the value of standard weight in the
calibration table, if the measure force value is bigger than the value of standard
weight, modify the value of standard weight of the second point in the calibration table
smaller, then click button “Write the Standard Data into the acquisition”, then re-
clear and re-compare until the two value become equal.
Fig. 8
The calibration process of rubber extensometer is same with the force sensor.
5. System Setting
Choose “System Setting” tab, see Fig. 10:
Fig. 10
Velocity Limit:Setting the maximum operate velocity and minimum operate velocity of
machine.
Reverse the Direction of Tensile and Compress : Determine whether reverse the
direction of Tensile and Compress.
Reverse the Direction of Up and Down:Determine whether reverse the direction of up
and down.
Probation Period : Can modify the probation period of software after choose this
button, the software can be continue used only if you input the right license when the
probation period expired.
Modify the Calibration Password:Modify the password calibration interface.
Modify the License:Modify the software License.
Write the Standard Data into the Acquisition : Write the current standard data into
the acquisition card, the data will be valid only when write the data into the
acquisition.
Get Standard Data from acquisition card :Get Standard Data from acquisition card to
current display.
Get Standard Data from computer:Get Standard Data from computer to current display.
OK : Save all data, switch the force sensor to the current choose channel. It is
independent with “Write the Standard Data into the acquisition”, Click “OK” does not
mean the current data is valid, the data will be valid only when write the data into the
acquisition.
Cancel:Don’t save all data, so the setting is invalid.
Fig. 11
A fail prompt window will be ejected when connect failed (as shown in Fig. 12), if
connect success, no prompt window eject. If the connect success, the software will
connect one time when it startups, the software will connect via the current COM at the
next startup time, the test can be successful only if the connect successfully.
Fig. 12
7. Language Change
Click Menu “View”-> “Language” to change the language of software, as shown in
Fig.13:
Fig. 13
The general version has three types of language: Chinese Simple、Chinese Traditional and English. And the
advanced version has ten types of language, if you have any special requirement, please contact us.
8. Option
Click Menu “View”-> “Option” or press the combination keys Ctrl+Shift+O (See 3.5
Shortcuts), then you can enter into the option interface, as shown in Fig. 14:
Fig. 14
There are two tabs in the “Option”: “Color” and “Display”, which can set up the
color and display.
9. Registration
Click Menu “Help”-> “Registration” or press and hold on the combination keys
Ctrl+R (See 3.5 Shortcuts), then you can enter into the registration interface, as shown
in Fig. 15:
Fig. 15
The software contain a probation period, you can continue to use the software only if
you input the right license when the probation period expired.
10. Activation
Click Menu “Help”-> “Activation” or press and hold on the combination keys
Ctrl+Shift+A (See 3.5 Shortcuts), then you can enter into the activation interface, as
shown in Fig. 16:
Fig. 16
According to the license of the system, you can activate the software to the advanced version after input
the right activate license. If connect fail, the license will be shown as 0, the activation can be operated only if
the connect successfully. If the function of the current software could not meet your test requirement, please
contact with us to upgrade to the advanced version.
11. Material Edit
You can choose an existed sample through the drop list of “Sample Data Choose” (As
shown in Fig. 17) or click button “Edit” to add or edit the sample data (As shown in
Fig. 18):
Fig. 17
Fig. 18
Add:Add a new material
Delete:Delete the selected material
OK:Save all modifies
Cancel:Don’t save the modify
12. Test method edit
You can choose an existed Project via drop list of “Project Select” (As shown in
Fig. 19) or click button “Edit” to add or edit the control program, each control
program contains “Test Method” and “Control Parameter”, you can click tab
“TestMethod” to edit the test method(As shown in Fig. 20)
Fig. 19
Fig. 20
Control Mode Value:The 1~20 steps “Control Mode Value” in the “Results List”
Shift Value:The 1~20 steps “Shift Value” in the “Results List”
Switch Condition : The percentage of the value of the switching conditions when
transform into the next action
Name List:List all existed control programs
Add:Add a new test method
Delete:Delete the selected test method
OK:Save all modifies
Cancel:Don’t save the modify
Use Default:Use the default system test method
13. Control parameters
Click tab “Control method” in “Control program”, you can edit the control
parameter. As shown in Fig. 22:
Fig. 22
Fig. 24
Name List:List all existed test results
Add:Add a new group of test results
Delete:Delete a group of selected test results
OK:Save all the modifies
Cancel:Don’t save all the modifies
The content in the result table is all the needed result for a completed test, you
can add a parameter to the table from the parameter list, or do some operation such as
delete 、 left shift or right shift via click a table. The button “Add” 、 ” Edit” and
“Delete” on the right side can operate on the parameter list. For example, chose
“tensile strength” in the result parameter, click “Edit” or double click “tensile
strength”, then the parameter edit interface will be eject. As shown in Fig. 25:
Fig. 25
The name is tensile strength, tensile strength=maximum force/area, in the table
above, you can find out the alternative symbol of maximum force is “h1”, the
alternative symbol of area is “m5”, so the formula is “h1/m5”, and the unit type is
stress (strength), the edit will be finished when click button “OK”. The table above is
the explanation of basic parameter symbol.
You can fill in interval set、automatic sample set and other tab on demand.
15. Report Edit
Click button “Edit Report”, you can enter into the report edit interface. As
shown in Fig. 26:
Fig. 26
Input the needed information in header, then click button “Print” can print the
report directly, Click button “Save as” can save the current report as a Word format
file, click button “Close” can exit current file.
16. Word report
Click button “Word report” can generate a word report. User can create a new Word
template, and chose the template in the drop-down list in “Word report” in tab
“Others” of menu “Test Results”, then the Word report will be produced by this
template. And a bookmark “Graph” and a table with two lines and two columns must be
included in the template; the position of the bookmark is the position of the curve, and
the table will be replaced by the test results. As shown in Fig. 27:
Fig. 27
Click Menu “Insert”-> “Bookmark” to insert a Curve, as shown in Fig. 28 and Fig.
29:
Fig. 28 Fig. 29
Input bookmark “Graph”, then click 添 加 (add), the curve will be add to current
position of the cursor.
In addition, if there has a text format “?**?” in the template and the content
between two “?” is same with the character in the test formula table, it will be
replaced by the actual value of the test.
You can change the maximum accommodate column number of the report test result if you
modify “NumColums=10” in file “Config.ini” under folder “data”, the default value
is 10.
17. Manual Snap
Click Menu “View”-> “Option”, function manual snap can be activated if you select
“manual Snap” in the display tab. As shown in Fig. 30 and Fig. 31:
Fig. 30
Fig. 31
During the test proceed, click button “Snap” or shortcuts F10, will fill the
current force、displacement and time into the form. During the non-test proceed, click button “Snap”,
if the cursor is invisible, it will be visible; if you don’t snap at the first time and
the cursor is visible, then the position of the cursor will be set as the current Snap
position.
18. Graph operation
Click right button of the mouse in the curve display window, then the graph operation
will be popped, As shown in Fig. 32:
Fig. 32
Return : Return to the initial state of the mouse, the mouse must return to initial
state if you want to drag the cursor on the curve
move:Parallel move the curve
Zoom in:Zoom in the curve
Zoom out:zoom out the curve
Zoom area:Press and hold on the left button of the mouse, drag a rectangular box to
zoom in part of the curve
Zoom all:Display the whole curve in the picture
Cursor : Indicate the cursor, current position of the cursor will be display on the
upper right corner, As shown in the blue marking in Fig. 32
Graph Tools : Indicate the graphics tools, show as the red marking in Fig. 32. the
graphics tools can also operate the curve, such as parallel move and zoom
Snap:Has the same function with “Manual Snap”
3.5 Shortcuts
Shortcuts list:
Shortcuts Function Shortcuts Function
F2 Up Ctrl + L Log in
F3 Down Ctrl + Q Exit
F4 Stop Ctrl + U Unit System
F5 Clear Ctrl + Shift + C Calibration
F6 Back position Ctrl + F Zero Force
F7 Start Ctrl + S Zero Course
F8 Finish Ctrl + E Zero Displacement
F9 Pause/Continue Ctrl + Shift + Z Zero
F10 Manual Sampling Ctrl + Shift + O Option
Ctrl + H Help
Ctrl + R Registration
Ctrl + Shift + A Activation
3.6 Configure files:
Text “AdFilter=5” in file “Initial.ini” means the sliding filter number of AD Value, it will be
effected when write acquisition card in the calibration interface.
Text “NumColums=10” in file “Config.ini” use for control the maximum accommodate
column number of the report test result.
4 Q&A
Q: The software complains “connection failed” when starting the test, how to solve
the problem?
A: “Connection failed” means the software cannot connect to the machine. Please
check the serial port is correctly connected; the machine is already powered on. If these
are confirmed, click “Option” —> “Connection”, choose the expected COM port to
connect. If you don’t know which COM port to connect, just try each COM port one by one
until successful connected. If all the above are done and the problem remains. Quit the
software, power off the machine, restart both the machine the software and retry.
Q: What to do when the software popup the “Software expired, please contact the provider.”?
A: Please contact the software provider to get registration code. The prompt will disappear after register
the software using the new registration code. For more information, see section 3.4.
Q: When I click “Up” or “Down” the speed of the workbench does not match the configured
speed(faster than expected),how to solve the problem?
A: Please uncheck “User options” ”Project Edit” ”control parameters” “Close loop control”.
Please don’t check this option without displacement coder or the displacement has been calibrated.
Q: After click the “Test” button, the workbench works for a while but stops when the sample is just under
stress, how to solve the problem?
A: Please increase the value of “User options” ”Project Edit” “TestMethod” “Break judge *.**%
of measuring range”. The software will not check break when the test value is below this value. So that the
experiment will not stop at the beginning because of the misjudgment of break.
Q: When I click “Up” or “Down”, the move direction of the workbench is opposite to the configured
value, how to solve the problem?
A: Please check “Options” “Calibration” “System options” “reverse Up-Down direction” and
click “write to chip” button.
Q: When the value of “Test Method” is set to “Tensile” or “Compress”, the move direction of the
workbench is opposite to the configured value, how to solve the problem?
A: Please check “Options” “Calibration” “System options” “reverse dir. Tensile&Compress” and
click “write to chip” button.
5 Functional differences between software versions
The software of TM2101 has three versions: Version A (Starter Edition); Version B
(Basic Edition); Version C (Professional Edition). The performance differences between
versions are as follows:
Function Version A Version B Version C
Management for USER Permission Y Y Y
Management for Unit Y Y Y
Calibrate System Y Y Y
Sample Data Editor Y Y Y
One-dimension Tension or Compression Test Y Y Y
Fixed Displacement Test Y Y Y
Fixed Strain Test Y Y Y
Multi-interval Analysis Y Y Y
Fixed Format Reports for Test Results Y Y Y
Word Format Reports for Test Results Y Y Y
Force-Displacement Curve Y Y Y
Force-Time Curve Y Y Y
Displacement -Time Curve Y Y Y
Stress-Strain Curve Y Y Y
Setting of Software Probation Period Y Y Y
Over Range Protection for Force/Route/Displacement Y Y Y
No. of Force Sensor Channels 4 4 4
Rubber Extensometer(for Large Deformation) Y Y Y
Metal Extensometer(for Small Deformation) Y Y Y
Closed-loop Control for Velocity Y Y Y
Closed-loop Control for Displacement N/A Y Y
Closed-loop Control for Force N/A Y Y
Fixed Force Control N/A Y Y
Fixed Stress Control N/A Y Y
Holding Pressure Test with Carton N/A Y Y
Single-pass Go Back and Forth Test N/A Y Y
Repeated Go Back and Forth Cycle Test N/A N/A Y
Multi-layer Nested Reciprocate Cycle Test N/A N/A Y
Multi-step Test N/A N/A Y
Fixed Force & Velocity Control N/A N/A Y
Fixed Stress & Velocity Control N/A N/A Y
Fixed strain & Velocity Control N/A N/A Y
Multi-step Data Analysis Capabilities N/A N/A Y
No. of Automatic Sampling Point 3 6 20
No. of Manual Sampling Point 5 10 30
Language Chinese Chinese Chinese
Simple Simple/Chinese Simple/Chinese
Traditional/English Traditional/English
The control method of version A is 1 step, suit for simple one-dimension tension or compression test;
The control method of version B is 2 steps, suit for slightly complex test, such as single reciprocating test and
holding pressure test with carton;
The control method of version C is multi-step, suit for various complicated control test.
6 Technical Support
If you have any question during use process, please contact us, we will be happy to serve you.
Please submit the issue via writing or e-mail, so that we could solve the problem for you as soon as
possible. Although we make every effort to, there maybe still be some omissions and errors, you are
welcome to criticize correction.
South China
Tel:0769-23129446
Fax:0769-23129447
Address : 501, C blocks, Tangbeixinyuan, Technology Boulevard,South city
District, Dong guan, Guangdong Province.
Web Address:www.rixine.com
E-mail:[email protected]
QQ :12638579
7 Appendix 1: Explanation of some calculate formula for test result
1. Ring stiffness
Thermoplastic pipe Determination of ring stiffness calculates according to GB/T 9647-2003.
S=(0.0186+0.025Y/d)F/(LY)
F----Force @ 3.0% elongation percentage. The unit is kN.
L----Sample length.The unit is m.
Y----Elongation @ 3.0% elongation percentage.The unit is m,Em.Y/d=0.03.
Please fill the form input fields with the following values:
Gauge column of sample data--> inside diameter of the measured pipe
Width column--> the length of the tested tube
First point of "test results" "auto sample setting" -->3
2. Elastic modulus:
Calculate according to GB8653-1988
Elastic modulus=Stress difference/strain difference of any two points in elastic range
The two points for elastic modulus calculation are set up in Tab “Other” in “Test Result”.
3. Tensile strength
Tensile strength=the large force/area
4. Fracture gauge
The Distance after sample fracture, need manual input.
5. Elongation after sample fracture
Elongation=(Fracture Distance -Original Distance)/Original Distance
6. Upper yield strength
Upper yield=Force on upper yield point/area.
7. Low yield strength
Low yield=Force on low yield point/area
For the material with no significant yield, use Rp0.2 as yield strength.
8. Provision of non-proportional extension strength
Non-proportional extension strength=Force on non-proportional extension point/area.
9. Provision of the total extension strength
Total extension strength= Force on total extension point/area.
10. Shear strength
Shear strength= Large Force/ (Distance * Width)
11. Shear modulus
Shear modulus= (Young’s modulus * Area/ Distance)*Thickness/ (Distance * Width)
12. Maximum bending strength
Maximum bending strength=3 * Large Force * Distance / (2 * Width * Thickness * Thickness)
13. Bending modulus
Bending modulus= (Young’s modulus * Area/ Distance) * Distance * Distance * Distance/ (4 * Width
* Thickness * Thickness * Thickness)
14. SDEV (Standard deviation)
SDEV= Radical of [(x1-x)^2+(x2-x)^2+......(xn-x)^2]/(n-1). Xn is the value of each sample, x is average
of all samples, n is the number of samples.
15. CPK(Process capability index)
Cpk=Cp(1-|Ca|), 其 中 Ca=(X-U)/(T/2),Cp =T/(6σ),T = Upper specification limit – Low specification
limit ; U=( Upper specification limit + Low specification limit)/2; X is the average, σis standard
deviation.
8 Appendix 2: Wiring diagram of RX2101
9 Appendix 3: Reference of connection diagram for RX2101 with
common servo drive
1. Panasonic A4 series AC servo drive:
The connection diagram can see in the wiring diagram, in addition, some parameters should be
configured as below:
1. P02=0(Position control mode).
2. P42=3(Pulse & direction control mode), P41 is use for change the direction of motor rotation.
P48 、 P49 、 P4A 、 P4B use for joint velocity control, the initial value of them can be set as:
P48=10000, P49=1, P4A=0, P4B=10000. If you use the internal encode of servo motor, please set
P44<=2500/reduction ratio of reducer, P45=0.
3. The frequency range of output pulse is 0.1~50 kHz.
2. Panasonic A series AC servo drive:
P02=0(Position control mode), P42=3(Pulse & direction control mode), P41 use for change the
direction of motor rotation. P46、P4A and P4B use for joint velocity control, the initial value of them
can be set as: P46=10000, P4A=0, P4B=10000.
▲The force sensor must connect with the board via shield wire, or there may exist some interference.
10 Appendix 4: Reference of connection diagram for RX2101 with
inverter
J13 use for connect with inverter, the connection diagram as below:
Helips A series inverter:
Parameters configure:
CD033=1,
CD034=1,
CD068 Use for velocity calibration.
In addition, please pay attention to the configuration of
CD007(Maximum operate frequency)、CD006(Minimum velocity control,
prevent the motor hot because of low velocity)、CD012(Time of start
to accelerate)、CD013(Time of stop to slow down)、CD032(Stop mode)、CD066~CD070. For the other
parameters, use the initial value.
The wiring diagram of J13 with “Up” &”down” solid state relay or 24V relay:
Appendix 5: Example of membrane switch