Astral 100 150 Clinical Guide
Astral 100 150 Clinical Guide
English
Contents
Introduction .......................................................................................................................... 1
Indications for use ............................................................................................................ 1
Contraindications .............................................................................................................. 1
Adverse effects ................................................................................................................ 1
General warnings and cautions......................................................................................... 2
The Astral system ................................................................................................................ 4
The Astral device .............................................................................................................. 5
The Astral device interface ............................................................................................... 6
Touch screen .................................................................................................................... 7
Information bar ................................................................................................................. 8
Menu bar .......................................................................................................................... 9
Bottom bar ....................................................................................................................... 9
Main screen ..................................................................................................................... 9
Pressure bar ................................................................................................................... 10
Settings bar .................................................................................................................... 10
Circuits and patient interfaces ........................................................................................... 11
Patient circuit accessories ................................................................................................. 11
Power accessories.............................................................................................................. 11
Power supplies ............................................................................................................... 12
Astral External Battery .................................................................................................... 12
ResMed Power Station II (RPSII) .................................................................................... 12
Remote Alarm II ............................................................................................................. 12
Pulse oximeter ............................................................................................................... 12
Optional accessories .......................................................................................................... 13
Astral Mobility bag.......................................................................................................... 13
Astral SlimFit Mobility bag .............................................................................................. 13
ResMed Homecare Stand .............................................................................................. 13
Astral Table Stand .......................................................................................................... 13
Aerogen® nebuliser......................................................................................................... 13
ResMed Connectivity Module (RCM) ............................................................................. 13
ResMed Connectivity Module Hospital (RCMH) ............................................................. 13
Using the Astral device...................................................................................................... 14
Connecting to mains power............................................................................................ 14
Powering on the device .................................................................................................. 14
Powering off the device ................................................................................................. 15
Accessing Clinical mode ................................................................................................. 15
Using the Setup Assistant .............................................................................................. 16
Enhanced access feature ............................................................................................... 19
Starting and stopping ventilation .................................................................................... 20
Locking and unlocking the touch screen ......................................................................... 21
Navigating the menus .....................................................................................................21
Monitors menu ............................................................................................................................... 22
Setup menu .................................................................................................................................... 23
Alarms menu .................................................................................................................................. 23
Information menu ........................................................................................................................... 24
Programs ........................................................................................................................24
Working with programs .................................................................................................................... 24
Setting up a new program ................................................................................................................ 26
Adjusting patient settings ................................................................................................................. 28
Monitoring ventilation .....................................................................................................29
Device settings ...............................................................................................................31
Adjusting device settings ................................................................................................................. 32
Copying device settings ................................................................................................................... 32
Exporting device settings onto USB stick ........................................................................................ 33
Importing device settings from the USB stick.................................................................................. 33
Service reminders............................................................................................................................. 34
Using the Service reminders ............................................................................................................ 34
Circuit options.....................................................................................................................36
Assembling patient circuits .............................................................................................36
Fitting the circuit adapter .................................................................................................................. 38
Connecting a single limb circuit with intentional leak ....................................................................... 38
Connecting a single limb circuit for invasive use .............................................................................. 39
Connecting a single limb circuit with expiratory valve ...................................................................... 40
Connecting a double limb circuit (Astral 150 only)............................................................................ 42
Connecting a mouthpiece circuit ...................................................................................................... 43
Changing the circuit selection .......................................................................................................... 44
Learn Circuit ..................................................................................................................................... 44
Accessories .........................................................................................................................48
Attaching patient circuit accessories ...............................................................................48
Attaching a humidifier .....................................................................................................48
Attaching a Heat Moisture Exchange (HME) ...................................................................50
Attaching an antibacterial filter ........................................................................................50
Adding supplemental oxygen ..........................................................................................51
Monitoring delivered oxygen ...........................................................................................53
Attaching a nebuliser.......................................................................................................53
Attaching other accessories ............................................................................................54
Attaching a pulse oximeter ............................................................................................................... 54
Attaching a remote alarm ................................................................................................................. 56
Astral Carry Bag ..............................................................................................................57
Travelling with the Astral device......................................................................................58
Power management ...........................................................................................................59
Connecting to mains power ............................................................................................60
Connecting the Astral External Battery ...........................................................................61
Using the External Battery................................................................................................................ 61
Connecting to a ResMed Power Station (RPSII) ..............................................................62
Connecting to an external DC power source ...................................................................62
Using the internal battery ................................................................................................63
Battery run time................................................................................................................................ 63
Storing and recharging...................................................................................................................... 64
Device power source indicators ......................................................................................64
Using the Astral device for the first time .......................................................................... 66
Ventilation modes .............................................................................................................. 67
(A)CV mode - Assisted volume-controlled ventilation...................................................... 68
P(A)CV mode - Assisted pressure-controlled ventilation ................................................. 69
P-SIMV - Pressure Synchronised Intermittent Mandatory Ventilation ............................. 71
V-SIMV - Volume Synchronised Intermittent Mandatory Ventilation ............................... 72
PS mode - Pressure support ........................................................................................... 74
(S)T mode - Spontaneous Ventilation with backup rate ................................................... 76
P(A)C mode .................................................................................................................... 77
CPAP mode .................................................................................................................... 78
iVAPS (intelligent Volume Assured Pressure Support) mode .......................................... 79
Target alveolar ventilation ............................................................................................... 80
intelligent Backup Rate (iBR) .......................................................................................... 81
Configuring iVAPS .......................................................................................................... 81
Adopting recently learnt targets ........................................................................................................81
Adopting recently learnt targets ........................................................................................................82
Entering the target values manually ..................................................................................................84
AutoEPAP....................................................................................................................... 86
Flow shape settings ....................................................................................................... 87
Interdependence of controls........................................................................................... 88
Dynamic settings limits .....................................................................................................................88
Making use of Astral's High Pressure limit .......................................................................................88
Triggering and cycling ..................................................................................................... 89
Triggering with intentional leak circuits .............................................................................................90
Cycling with intentional leak circuits .................................................................................................90
Triggering with valve circuits .............................................................................................................91
Changing the Trigger Type ................................................................................................................92
Cycling with valve circuits .................................................................................................................93
Triggering with mouthpiece circuits ..................................................................................................93
Supplementary features .................................................................................................... 94
Manual breath settings ................................................................................................... 94
Sigh settings .................................................................................................................. 96
Apnoea Settings ............................................................................................................. 97
Control settings for Apnoea ventilation .............................................................................................97
(A)CV Breath pattern .........................................................................................................................98
P(A)CV breath pattern ........................................................................................................................99
Safety volume settings ................................................................................................. 100
Non-invasive ventilation (NIV) ........................................................................................ 101
Mask NIV...................................................................................................................... 101
Mouthpiece NIV ........................................................................................................... 101
Mouthpiece ventilation ................................................................................................. 101
Safety considerations for mouthpiece ventilation ......................................................... 102
Alarms .............................................................................................................................. 103
Alarm priority ................................................................................................................ 104
Viewing the active alarms ............................................................................................. 106
Muting alarms .............................................................................................................. 106
Resetting alarms .......................................................................................................... 107
Adjusting alarm settings................................................................................................108
Setting the alarm volume ..............................................................................................109
Testing the alarm sounders and indicators ....................................................................109
Testing the Remote Alarm ............................................................................................110
Testing the alarms ............................................................................................................111
Testing the power alarms .............................................................................................111
Testing the pressure alarms ..........................................................................................111
Testing the ventilation monitoring alarms......................................................................112
Testing the oximetry alarms ..........................................................................................112
Testing the oxygen alarms ............................................................................................113
Testing the breathing circuit alarms ..............................................................................113
Testing the disconnection alarm ...................................................................................113
Alarm settings and conditions .........................................................................................114
Tidal volume alarms ......................................................................................................114
Minute volume alarms ..................................................................................................114
Respiratory rate alarms .................................................................................................115
Pressure alarms/limits ...................................................................................................115
High pressure ................................................................................................................................. 115
Obstruction ..................................................................................................................................... 116
Low pressure.................................................................................................................................. 116
PEEP ............................................................................................................................................... 116
Breathing circuit alarms.................................................................................................117
High leak ......................................................................................................................................... 117
NV Mask/Rebreathing..................................................................................................................... 117
Flow sensor not calibrated ............................................................................................................. 117
Circuit configuration alarms ............................................................................................................ 118
Disconnection alarm ....................................................................................................................... 118
Oxygen alarms ..............................................................................................................118
Oximetry alarms............................................................................................................119
SpO2............................................................................................................................................... 119
Pulse rate........................................................................................................................................ 119
Apnoea alarm ................................................................................................................119
Ventilation stopped alarm ..............................................................................................119
Power alarms ................................................................................................................120
System alarms ..............................................................................................................120
Detecting circuit disconnection and de-cannulation ......................................................121
Astral Disconnection Alarm .............................................................................................122
Adjusting the Disconnection Alarm ...............................................................................122
Setting and testing Disconnection Tolerance ................................................................122
Setting Activation Time .................................................................................................................. 124
Disabling (or enabling) the Disconnection Alarm ...........................................................125
Data management process ..............................................................................................127
Data management summary .........................................................................................128
Deleting patient data........................................................................................................ 129
Cleaning and maintenance .............................................................................................. 130
Single patient use ......................................................................................................... 130
Weekly ............................................................................................................................................130
Monthly ...........................................................................................................................................130
Multi-patient use .......................................................................................................... 131
Replacing components ................................................................................................. 132
Replacing the air filter ......................................................................................................................132
Replacing the expiratory flow sensor and antibacterial filter (Astral 150 only) ................................132
Replacing the double limb adapter (expiratory valve) ......................................................................134
Replacing the Oxygen sensor .........................................................................................................134
Replacing the internal battery ..........................................................................................................135
Servicing....................................................................................................................... 136
Maintenance Timetable ...................................................................................................................136
Internal Battery ................................................................................................................................136
Device information ..........................................................................................................................137
Upgrading software .........................................................................................................................137
Additional considerations for hospital or healthcare facilities ........................................ 137
Technical specifications ................................................................................................... 138
Monitoring........................................................................................................................ 143
Time parameters .......................................................................................................... 143
Volume and flow parameters........................................................................................ 143
Pressure parameters .................................................................................................... 144
Other parameters ......................................................................................................... 144
Accuracy of controls ........................................................................................................ 145
Measurement System and Accuracy .............................................................................. 148
Functional variants........................................................................................................... 148
Guidance and Manufacturer’s Declaration Electromagnetic Emissions
& Immunity....................................................................................................................... 149
Guidance and manufacturer’s declaration—electromagnetic emissions ....................... 149
Guidance and manufacturer’s declaration – electromagnetic immunity ........................ 150
Recommended separation distances between portable and mobile RF
communications equipment and the life support device............................................... 152
Potential impact of electromagnetic disturbances ........................................................ 152
Symbols............................................................................................................................ 153
Standards compliance ..................................................................................................... 154
Training and support ....................................................................................................... 154
Troubleshooting............................................................................................................... 155
Alarm troubleshooting .................................................................................................. 155
Learn Circuit troubleshooting........................................................................................ 160
General troubleshooting ............................................................................................... 163
Recommendations for suctioning ...................................................................................164
Limited warranty ..............................................................................................................165
Appendix A: Definitions ...................................................................................................166
Ventilation settings definitions ......................................................................................166
Measured and calculated parameter definitions ............................................................168
Appendix B: Ventilation parameters................................................................................170
Ventilation Parameters summary table ..........................................................................170
Ventilation displayed parameters...................................................................................174
Supplementary Features ...............................................................................................174
Supplementary Features displayed parameters.............................................................177
Appendix C: Alarm parameters .......................................................................................179
Introduction
Introduction
The Astral device provides mechanical ventilation to both ventilation dependent and non-dependent
patients. It delivers pressure and volume ventilation through either a valve or leak circuit, and is
compatible with a range of accessories to support specific use cases.
The information in this guide applies to both the Astral 100 and the Astral 150 devices. Where
information applies to only one of these devices, that device will be specified.
Note: Some features may not be available on your device.
WARNING
• Read the entire manual before using the Astral device.
• Use the Astral device only as directed by a physician or healthcare provider.
• Use the Astral device only for the intended use as described in this manual. Advice contained
in this manual does not supersede instructions given by the prescribing physician.
• Install and configure the Astral device in accordance with the instructions provided in this
guide.
CAUTION
The Astral device is not intended for use as an emergency transport ventilator.
Contraindications
The Astral device is contraindicated in patients with the following pre-existing conditions:
• pneumothorax or pneumomediastinum
• pathologically low blood pressure, particularly if associated with intravascular volume depletion
• cerebrospinal fluid leak, recent cranial surgery or trauma
• severe bullous lung disease
• dehydration.
WARNING
AutoEPAP is contraindicated when using an invasive interface.
Adverse effects
Report unusual chest pain, severe headache or increased breathlessness to your physician. The following
side effects may arise during use of the device:
• drying of the nose, mouth or throat
• nosebleed
• bloating
• ear or sinus discomfort
• eye irritations
• skin rashes.
English 1
Introduction
WARNING
• If you notice any unexplained changes in the performance of the device, if it is making unusual
or harsh sounds, if the device or the power supply are dropped or mishandled discontinue use
and contact your healthcare provider.
• For ventilator-dependent patients, always have alternate ventilation equipment available, such
as a back-up ventilator, manual resuscitator or similar device. Failure to do so may result in
patient injury or death.
• The Astral device is a restricted medical device intended for use by qualified, trained personnel
under the direction of a physician. Clinical supervision is required in critical care/intensive care
unit environments.
• Ventilator-dependent patients should be continuously monitored by qualified personnel or
adequately trained carers. These personnel and carers must be capable of taking the necessary
corrective action in the event of a ventilator alarm or malfunction.
• The internal battery is not intended to serve as a primary power source. It should only be used
when other sources are not available or briefly when necessary; for example, when changing
power sources.
• The Astral device is not intended to be operated by persons (including children) with reduced
physical, sensory or mental capabilities without adequate supervision by a person responsible
for the patient's safety.
• The Astral device is not intended to be operated by patients unless they have been given
adequate instruction concerning the operation of the device by a person responsible for the
patient's safety.
• The Astral device must not be used in the vicinity of an MRI or diathermy device.
• The effectiveness of ventilation and alarms should be verified including after any ventilation or
alarm setting change, any change in circuit configuration, or after a change to co-therapy (eg,
nebulisation, oxygen flow).
• The Astral device and AC Power Supply can get hot during operation. To prevent possible skin
damage do not leave the Astral device or AC Power Supply in direct contact with the patient for
extended periods of time.
• The device can provide therapies typically associated with both ventilator-dependent and
non-dependent patients. The mode of ventilation, circuit type, and alarm strategies should be
chosen after a clinical evaluation of each patient’s needs.
• The device must not be used at an altitude above 3000m or outside the temperature range of
0–40°C. Using the device outside these conditions can affect device performance which can
result in patient injury or death.
2
Introduction
A caution explains special measures for the safe and effective use of the device.
CAUTION
• Repairs and servicing of the device should only be performed by an authorised ResMed service
representative.
• The temperature of the airflow for breathing produced by the device can be as much as 6ºC
higher than the temperature of the room. Caution should be exercised if the room temperature
is warmer than 35ºC.
• Do not expose the device to excessive force, dropping or shaking.
• Dusty environments may affect device performance.
• The Astral device may experience interference in the vicinity of electronic article surveillance
(EAS). Keep the Astral device at least 20 cm away from the EAS.
English 3
The Astral system
WARNING
Before using any accessory, always read the accompanying User Guide.
4
The Astral system
Description
1 Adapter port
Can be fitted with single limb adapter, single limb leak adapter or double limb adapter (Astral 150 only).
2 Handle
3 Inspiratory port (to patient)
Provides an outlet for pressurised air to be delivered to the patient via the patient circuit. Includes FiO2 sensor on the
Astral 150. The FiO2 sensor is an optional accessory on the Astral 100.
4 Ethernet connector (service use only)
5 USB connector (for download to ResScan and connection of approved accessories)
6 Mini USB connector (for connection to RCM or RCMH)
English 5
The Astral system
Description
7 DC power inlet
8 Device on/off push button
9 SpO2 Sensor connector
10 Remote alarm five pin connector
11 Low flow oxygen input (up to 30 L/min)
12 Air inlet (complete with hypoallergenic filter)
Description
1 Touch screen
2 Power source indicators
AC (mains power supply)
DC (external battery or car accessory adapter or RPSII)
Internal battery
6
The Astral system
Description
3 Therapy on/off indicator
Device ready
Constant green display when the device is turned on but not ventilating.
Device ventilating
Flashes blue when the device is ventilating and the Ventilation LED setting is 'ON'.
Otherwise is 'OFF'.
4 Alarm mute/reset button
Illuminates when an alarm is triggered and flashes when the sound is muted.
5 Alarm bar
Touch screen
The main method of interacting with the Astral device is via the touch screen. The display on the touch
screen changes according to the function being performed.
Description
Locked Unlocked
2 Manual breath button
English 7
The Astral system
Description
6 Menu bar
7 Bottom bar
8 Start/Stop ventilation button
9 Main screen
10 Sub-menus
11 Pressure bar
Information bar
The Information bar is displayed at the top of the touch screen. The Information bar displays the
operating status of the device, including patient type, current circuit configuration, programs, information
messages, ventilation status, alarms and power status.
Description
Patient type – Adult
Message Will display alarms or information. Image above shows device in Standby. (Displayed when the device is powered
window on but not ventilating). Date and time will be displayed when the device is ventilating and there are no active
alarms.
Information messages are displayed in blue text. If the device Alert tone setting is 'On', you will be alerted to new
information messages by a single beep.
8
The Astral system
Menu bar
The Menu bar provides access to the four main menus in the Astral device.
Monitors menu
View real-time patient data in either waveform or monitoring formats including pressure, flow, leak, tidal
volume, synchronisation and oximetry.
Setup menu
Configure and view ventilation therapy or device settings; and import/export data.
Alarms menu
Configure and view alarms including alarm volume.
Bottom bar
The Bottom bar changes with the function of the device.
It can display buttons to Stop or Start ventilation and Apply or Cancel functions. It can also display
real-time readings.
Main screen
The Main screen displays the monitoring data, ventilation and device controls. Each function is accessed
through the various menus and tabs.
English 9
The Astral system
Pressure bar
The Pressure bar displays real-time therapy data while the Astral device is ventilating.
Patient pressure is shown as a bar graph. Peak inspiratory pressure is shown as a numerical value and
watermark. Spontaneously triggering and cycling is indicated by and .
The example below displays the pressure bar when a patient is spontaneously breathing.
Description
1 Peak inspiratory pressure (PIP) value
2
Spontaneous cycled breath marker—indicates patient-cycled breath
3 Peak inspiratory pressure marker
4 Current pressure
5 Positive end expiratory pressure (PEEP) setting
6
Spontaneous triggered breath marker—indicates patient-triggered breath
Settings bar
The settings bar appears on the right hand side of the touch screen when a therapy or alarm setting has
been selected for adjustment. The settings bar allows the settings to be adjusted via the up and down
scroll arrows or the minimum and maximum buttons. The example below displays the settings bar when
a setting has been selected for adjustment.
Description
1 Maximum button—adjusts setting to maximum limit
2 Up scroll arrow—increases setting by single increment
3 Down scroll arrow—decreases setting by single increment
10
Circuits and patient interfaces
Invasive
• endotracheal tubes
• tracheostomy tubes
Non-invasive
• vented and non-vented masks
• mouthpiece.
WARNING
The measurement of patient exhaled gas volume may be affected by leak.
Power accessories
WARNING
• The Astral device should only be used with accessories recommended by ResMed. Connection
of other accessories could result in patient injury or damage to the device.
• Connecting the Astral device to the battery of a battery-powered wheelchair may affect the
device performance and may result in patient harm.
English 11
Power accessories
Power supplies
In addition to the internal battery and mains power sources, the Astral device can be powered from the
following sources:
• Astral External Battery
• ResMed Power Station II
• Astral DC adapter.
Remote Alarm II
The Remote Alarm II is self-powered (via battery) and connected to the Astral device via a cable. If
required, a second Remote Alarm can be attached to the first Remote Alarm. This enables Remote
Alarms to be placed in two separate locations. The Remote Alarm II can also be connected to a hospital
alarm management system. For more information, see the Remote Alarm II user guide.
Pulse oximeter
Attaching a pulse oximeter to the Astral device enables real-time SpO2 and pulse readings to be viewed
from the Monitoring menu. For further information, refer to Attaching a pulse oximeter (see page 54).
12
Optional accessories
Optional accessories
The Astral device can be used with a range of optional accessories as follows:
• Astral Mobility Bag
• Astral SlimFit Mobility Bag
• ResMed Homecare Stand
• Astral Table Stand
• Aerogen® nebuliser
• ResMed Connectivity Module (RCM)
• ResMed Connectivity Module Hospital (RCMH).
Note: Some accessories may not be available in all regions.
Aerogen® nebuliser
If required, a nebuliser can be used in conjunction with the Astral device. ResMed recommends
Aerogen® nebuliser products. For more information, refer to Attaching a nebuliser (see page 53).
English 13
Using the Astral device
CAUTION
• To prevent possible damage to the ventilator, always secure it to its stand or place it on a flat,
stable surface. For mobile situations, ensure the Astral device is contained within its
mobility bag.
• Ensure the device is protected against water if used outdoors.
14
Using the Astral device
The closed padlock indicates the device is in Patient mode. To access clinical functions, such as the
Setup Assistant and program configurations, you must enter Clinical mode.
Clinical mode can be accessed from any screen, whether or not the Astral device is ventilating.
1. From the patient home screen, press and hold for 3 seconds, then release.
2. Select:
20 minutes—the device will return to patient mode after 20 minutes of inactivity, or
Unlimited—the device will remain unlocked until powered off or manually locked.
CAUTION
Only select Unlimited if the device will remain under the continuous supervision of qualified
trained personnel under the direction of a physician. When clinical access is no longer required,
exit Clinical mode to return to Patient mode.
English 15
Using the Astral device
2. From the Main menu press Setup . The Setup menu is displayed.
3. Select Setup Assistant and press Start.
16
Using the Astral device
5. Select the Patient type. This will automatically configure default alarm and setting ranges.
6. The Select Circuit type screen is displayed. Select the circuit type to be connected to the device.
8. Follow the prompts to attach and test the circuit. For detailed instructions on how to attach the
circuit, refer to Assembling patient circuits (see page 36).
When the Learn Circuit is complete, you will be prompted to select your required Interface type. The
interface type selection is used to determine if the disconnection alarm can be turned off and also
sets a suggested setting for the disconnection alarm.
Note: If the circuit is single with leak and Mask is selected as the Patient interface, the Select Mask Type screen
will appear.
English 17
Using the Astral device
18
Using the Astral device
WARNING
To prevent inadvertent alarm mute or reset, do not leave the patient in contact with the device
screen.
1. From the Main menu press Setup . The Setup menu is displayed.
2. Select the Patient Access tab from the Device Config. menu.
English 19
Using the Astral device
With this feature enabled, it is possible to switch between ‘Big buttons’ mode and standard. Simply
select the Home button from left hand corner of the Bottom bar.
Your screen will return to standard button size and the Home icon will be replaced by the Big buttons
icon .
To return to 'Big buttons' mode, simply select the Big buttons icon from the bottom bar.
Note: With the enhanced access feature enabled, your screen will return to 'Big buttons' mode once the screen
locks (after two minutes of inactivity).
WARNING
Always ensure that the values in the ventilation therapy and alarm settings are appropriate before
starting therapy.
Note: If using the device for the first time, ResMed recommends performing a functional test before starting
ventilation. Refer to Using the Astral device for the first time (see page 66).
To start ventilation:
1. Press the green on/off button at the back of the device (if power is not already on).
20
Using the Astral device
To stop ventilation:
Ventilation can be stopped at any time and from any screen.
1. If oxygen is connected, turn off the oxygen.
To manually lock the touch screen, from the Information bar press . When the touch screen is
locked the button is highlighted orange.
English 21
Using the Astral device
Monitors menu
The Monitors menu allows you to view real-time ventilation data and is comprised of three sub-menus:
• Waveforms
• Monitoring
• Trends
22
Using the Astral device
Setup menu
The Setup menu displays six different sub-menus:
• Programs—to configure therapy programs
• Setup Assistant—guides you through device setup
• Circuit—to set up the circuit
• Settings—to change the ventilation mode and related settings
• USB—to save patient data and import/export settings
• Device config.—to change the device configuration.
Alarms menu
Display the Alarms menu to view/update the individual thresholds for each alarm to trigger. Real-time
values are displayed between the upper and lower thresholds. The options available for adjustment
change depending on the circuit configuration.
English 23
Using the Astral device
Information menu
The Information menu is comprised of three sub-menus:
• Events—all logged event activity that has taken place is displayed. A breakdown of specific alarms,
settings or system events can also be viewed.
• Device—information about the actual device is displayed, eg, Model and Serial numbers, Software
version, and Next service due date.
• Battery—information about the state of charge of the internal and external batteries when connected
including the combined total battery charge.
Programs
WARNING
The safety and effectiveness of ventilation therapy and alarms settings should be verified for each
enabled Program.
24
Using the Astral device
5. When selected, the program name is displayed on the program label and on the information bar
when that program is active.
English 25
Using the Astral device
To change programs:
1. Select P2: (A)CV. P2 becomes the active program.
2. Select and start the Setup assistant. Refer to Using the Setup Assistant (see page 16).
Once the Setup assistant is complete, P2 will now be configured to the settings you have chosen
and is ready to Start ventilation. In the example below, P2 has been changed from (A)CV mode to
P(A)CV.
Note: Each program retains its own circuit configuration. When switching between programs, ensure you use the
correct circuit as 'learned' for that program.
WARNING
Always review and adjust alarm Settings when setting up a new program. For information, refer to
Adjusting alarm settings (see page 108).
26
Using the Astral device
3. Press the Ventilation mode button. The Select Ventilation Mode screen will display, the current
active Ventilation mode will be highlighted orange. Select your required Ventilation mode.
4. You will be returned to the Main Settings page. All changes you have made will be outlined in
orange, indicating a pending status until you press Apply. If the mode change is cancelled you will be
returned to the current active mode's primary settings page. Any modifications will be disregarded.
5. Once settings have been applied the program is updated with the new Mode selection. Perform a
Learn Circuit.
English 27
Using the Astral device
6. To make changes to the patient settings within this mode, refer to Adjusting patient settings (see
page 28).
Note: If the mode change is cancelled you will be returned to the current active mode's primary
settings page. All modifications will be disregarded.
2. Select the setting that you want to adjust. The selected setting is highlighted in orange and the
Settings bar appears on the right side of the screen.
28
Using the Astral device
Notes:
• The Apply button is disabled if one or more settings cannot be applied. Refer to Interdependence of controls
(see page 88). A message is displayed indicating the reason.
• If the mode change is cancelled you will be returned to the current active mode's primary settings page. All
modifications will be disregarded.
Monitoring ventilation
During ventilation, ventilation parameters are continuously monitored and displayed in real-time. The
measures are displayed as follows:
• Numerical values on the Monitoring screen and a summary of critical values in the bottom bar.
• Graphically on the Waveforms, Trends screen and Pressure bar.
Monitoring screen
The Monitoring screen displays all measured parameters in numerical form.
For details on the available monitoring parameters refer to the Ventilation Parameters Summary Table
(see page 170).
Waveforms
The Waveforms screen displays the last 15 seconds of patient airway pressure and flow in a graph. The
graph updates in real-time and when necessary the vertical axis will auto scale to accommodate changes
in amplitude.
Description
English 29
Using the Astral device
Trends screen
The Trends screen shows the 5th and 95th percentile values, as well as the median for the last 30 days
for each of the following parameters:
• Leak
• Minute ventilation
• Peak inspiratory pressure
• Tidal volume
• Respiratory rate
• Inspiratory time
• SpO2
• Pulse rate
• FiO2
• Alveolar ventilation.
30
Using the Astral device
Device settings
All device configuration settings are stored independently of patient programs. The configurable settings
are described in the following table.
English 31
Using the Astral device
32
Using the Astral device
2. From the Main menu, press Setup . The Setup menu is displayed.
3. Select the Settings tab from the USB sub-menu.
When the USB is ready, the Export > button will become active.
4. Press Export >.
Any Astral settings currently on the USB stick will be deleted and replaced with the settings from the current
device.
5. Press Confirm to proceed with the export.
When the export is complete, a status message is displayed.
3. From the Main menu, press Setup . The Setup menu is displayed.
4. Select the Settings tab from the USB sub-menu.
The Import > button will become active if there are settings from a compatible device available on the USB stick
and the device is not ventilating.
5. Press Import >.
A warning will be displayed to indicate that all device and patient program settings will be replaced.
6. Press Confirm to proceed with the import.
A progress bar will be displayed while the settings are being imported and applied.
When the import is complete, a status message is displayed as well as a reminder to check all settings have
been applied correctly.
Learn Circuit results will be reset to default parameters for each enabled program. Learn Circuit needs to be
performed with the appropriate patient breathing circuit for each enabled program.
All device configuration settings and all patient program settings should be verified to be appropriate for the
patient prior to commencing ventilation.
English 33
Using the Astral device
Service reminders
The Service reminders provide information that device servicing is required in the near future.
There are two types of Service reminders:
• Preventative maintenance reminder—the two year service interval is approaching.
Note: This reminder is linked to the Next service due date as displayed on the Device information page.
• Battery replacement reminder—the internal battery is approaching the end of service life.
Select the reminder button to display the reminder prompt. Acknowledgment of the prompt will clear the
reminder prompt and reminder button. The reminder button will remain if the user prompt has not been
acknowledged or the Next service due date has passed.
Note: More than one Service reminder can be present at a given time. In this case, the Service
reminders will be displayed sequentially after each confirmation.
34
Using the Astral device
2. From the Main menu, press Setup . The Setup menu is displayed.
3. Select the Reminders tab from the Device Config. sub-menu.
English 35
Circuit options
Circuit options
The Astral device supports a range of circuits (the device and accessories assembled together) to suit
individual patient needs. The device uses interchangeable circuit adapters.
The following table may assist in selecting suitable circuits and settings for different patient types:
Tidal volume range Recommended patient type setting Suitable circuit diameters
50 mL to 300 mL Paediatric 10 mm, 15 mm or 22 mm
> 300 mL Adult 15 mm or 22 mm
WARNING
• Use a double limb circuit for direct measurement of exhaled volumes. In this configuration, the
expired volume is returned to the ventilator for independent measurement. (Astral 150 only)
• The Astral device does not support monitoring of exhaled volumes when used with a single
limb circuit with expiratory valve.
• The patient circuit should be arranged so as not to restrict movement or pose a strangulation
risk.
• Only use circuit components that comply with the relevant safety standards including
ISO 5356-1 and ISO 5367.
CAUTION
For paediatric use, ensure that the patient circuit type fits and is suitable for use with a child. Use a
paediatric patient type for patients that weigh less than 23 kg and normally require less than
300 mL tidal volume.
36
Circuit options
1 Single limb leak Single limb circuit with intentional leak or mouthpiece circuit
2 Single limb Single limb circuit with expiratory valve (expiratory valve integrated into the circuit)
3 Double limb Double limb circuit (expiratory valve integrated into the adapter) OR single limb circuit with
(Astral 150 only) intentional leak or mouthpiece circuit
A Learn Circuit should be performed after any change of circuit. Astral will provide accurate therapy as long as the Learn Circuit
is completed. Refer to Learn Circuit (see page 44) for further information.
English 37
Circuit options
WARNING
• At low pressures, the flow through the mask vents may be inadequate to clear all exhaled
gases, and some rebreathing may occur when using a single limb circuit with intentional leak.
• Ensure the vent holes at the mask or at the ResMed Leak Valve are unobstructed. Ensure the
area around the vent holes is clear of bedding, clothes, or other objects and that the vents
holes are not directed towards the patient.
6. Attach the patient interface (eg, mask) to the Leak valve or the free end of the air tubing as
appropriate and adjust the mask type setting on the Astral device.
Note: If using the ResMed Leak Valve, select Leak Valve for the Mask type setting.
CAUTION
Always set up the ResMed Leak Valve in the breathing circuit with the arrows and the symbol
pointing in the direction of air flow from the Astral device to the patient.
For invasive ventilation, since the patient's upper respiratory system is bypassed by an artificial airway
device (for example endotracheal or tracheostomy tube) humidification of the inspired gas is required to
prevent lung injury.
English 39
Circuit options
To connect an Astral 'Quick Connect' Single Limb Circuit with expiratory valve:
1. Check the device is equipped with the single limb adapter (otherwise change the adapter).
2. Connect the air tubing to the inspiratory port on the device.
3. Attach the Astral Quick Connect circuit to the single limb adapter on the device (see diagram below).
4. Attach any required circuit accessories (eg, humidifier or filter).
5. Select the circuit type and perform a Learn Circuit.
6. Attach a patient interface (eg, mask) to the connector on the expiratory valve.
40
Circuit options
English 41
Circuit options
42
Circuit options
English 43
Circuit options
3. Select the circuit you want to change to which will be highlighted orange. You will be taken to the
Learn Circuit screen.
4. Press Start to run the Learn Circuit and follow the on-screen prompts.
Learn Circuit
In order to support a wide range of circuit configurations and accessories, the Astral device provides a
Learn Circuit function to determine the characteristics of the circuit. As part of the Learn Circuit
functionality the Astral performs a device self-test and a calibration of the FiO2 sensor (if installed).
CAUTION
To ensure optimum and accurate performance, it is recommended that the Learn Circuit function
be performed with every change of circuit configuration and at regular intervals not less than once
every three months.
Do not connect patient interfaces prior to performing the Learn Circuit. Patient interfaces include any
components placed after the single circuit's expiratory valve or exhalation port, or double limb circuit's 'Y'
piece (eg, HMEF, catheter mount, mask, tracheostomy tube).
The following table may assist in selecting suitable circuits and settings for different patient types:
Tidal volume range Recommended patient type setting Suitable circuit diameters
50 mL to 300 mL Paediatric 10 mm, 15 mm or 22 mm
> 300 mL Adult 15 mm or 22 mm
44
Circuit options
Note: Trigger type sets whether a pressure-based or flow-based trigger threshold is used when a Double circuit
is selected.
English 45
Circuit options
The following icons are used to report the Learn Circuit results:
46
Circuit options
English 47
Accessories
Accessories
For a full list of accessories, see Ventilation accessories on www.resmed.com under the Products page.
If you do not have internet access, please contact your ResMed representative.
WARNING
• Adding or removing circuit components can adversely affect ventilation performance. ResMed
recommends performing a Learn circuit every time an accessory or component is added to or
removed from the patient circuit. If the circuit configuration is changed, the Disconnection
Alarm needs to be checked for correct operation.
• Do not use electrically conductive or anti-static air tubing.
Attaching a humidifier
A humidifier or HME is recommended for use with the Astral device.
WARNING
• For invasive ventilation, since the patient's upper respiratory system is bypassed by an artificial
airway device (for example endotracheal or tracheostomy tube) humidification of the inspired
gas is required to prevent lung injury.
• Always place the humidifier on a level surface below the level of the device and the patient to
prevent the mask and tubing filling with water.
• Only use humidifiers that comply with the relevant safety standards, including ISO 8185 and
set up the humidifier according to the manufacturer's instructions.
• Monitor the air tubing for water condensation and / or spillage from the humidifier. Use
appropriate precautions to prevent water in the circuit transferring to the patient (eg, a water
trap).
For non-invasive ventilation, for patient experiencing dryness of the nose, throat or mouth, humidification
of the inspired gas will prevent subsequent irritation and discomfort.
CAUTION
Make sure that the water tub is empty and thoroughly dried before transporting the humidifier.
48
Accessories
When using heated humidification with a double limb circuit, condensation may form in the expiratory
flow sensor if the air is cooled to below its dew point. Condensation may also form in the patient circuit
and is most likely to form at high humidity settings and low ambient temperatures.
Condensation forming in the expiratory flow sensor may cause a loss of expiratory flow measurement
and compromised therapy (ie, auto-triggering, increased PEEP and activation of the leak alarm.
To prevent condensation at the Expiratory flow sensor, always follow the humidifier manufacturer's
instructions on how to prevent condensation and regularly check the patient circuit for condensation.
To ensure accurate therapy, Astral's Learn Circuit function should be performed prior to filling the water
tub.
English 49
Accessories
WARNING
Only use HMEs that comply with the relevant safety standards, including ISO 9360-1 and
ISO 9360-2.
Place the HME between the patient end of the circuit and the patient interface.
Do not connect patient interfaces prior to performing the Learn Circuit. Patient interfaces include any
components placed after the single circuit's expiratory valve or exhalation port, or double limb circuit's 'Y'
piece (eg, HMEF, catheter mount, mask, tracheostomy tube).
WARNING
• Regularly check the antibacterial filter and expiratory valve for signs of moisture or other
contaminants, particularly during nebulisation or humidification. Failure to do so could result
in increased breathing system resistance and/or inaccuracies in expired gas measurement.
• Only use antibacterial filters that comply with the relevant safety standards, including
ISO 23328-1 and ISO 23328-2.
CAUTION
The antibacterial filter must be used and replaced according to the manufacturer's specifications.
50
Accessories
WARNING
• To prevent the risk of cross-contamination, an antibacterial filter is mandatory if the device is to
be used on multiple patients.
• The expiratory module, internal antibacterial filter, expiratory flow sensor and blue membrane
come into contact with exhaled gases but do not form part of the inspiratory pathway.
English 51
Accessories
WARNING
• Use only medical grade oxygen sources.
• Always ensure that the device is ventilating before the oxygen supply is turned on.
• Oxygen flow must be turned off when the device is not ventilating so that oxygen does not
accumulate within the device enclosure. Explanation: Accumulation of oxygen presents a risk
of fire. This applies to most types of ventilators.
• Oxygen supports combustion. Oxygen must not be used while smoking or in the presence of
an open flame. Only use oxygen in well-ventilated rooms.
• Supplemental oxygen must be added into Astral’s oxygen inlet at the rear of the device.
Adding oxygen elsewhere, ie, into the breathing system via a side port or at the mask, has
potential to impair triggering and accuracy of therapy/monitoring and impair alarms (eg, High
Leak alarm, Non-vented mask alarm)
• The patient circuit and the oxygen source must be kept at a minimum distance of 2 m away
from any sources of ignition.
• Monitor supplemental oxygen using the integrated FiO2 sensor and alarms or use an external
O2 monitor compliant with ISO 80601-2-55.
• When operating Astral in its mobility bag do not add more than 6 L/min of supplemental
oxygen.
• Astral is not designed for use with heliox, nitric oxide or anaesthetic gases.
• Do not position the Astral device on its side as this may affect FiO2 monitoring accuracy.
Supplemental oxygen can also be added from an oxygen bottle however, a flow regulator must be fitted
to ensure the delivered oxygen remains at or below 30 L/min.
Before you remove supplemental oxygen from the device, ensure the Oxygen supply has been turned
off.
52
Accessories
WARNING
Do not position the Astral device on its side as this may affect FiO2 monitoring accuracy.
Attaching a nebuliser
If required, a nebuliser can be used in conjunction with the Astral device. ResMed recommends
Aerogen® nebuliser products—designed to operate in-line with standard ventilator circuits and
mechanical ventilators without changing ventilator parameters or interrupting ventilation.
WARNING
• Always connect antibacterial filters to both the inspiratory port and the expiratory inlet of the
Astral device to protect the device.
• Regularly check the antibacterial filter and expiratory valve for signs of moisture or other
contaminants, particularly during nebulisation or humidification. Failure to do so could result
in increased breathing system resistance and/or inaccuracies in expired gas measurement.
• Only operate the nebuliser when the device is ventilating. If ventilation is stopped, switch off
the nebuliser.
• Use of a gas jet nebuliser may affect ventilator accuracy. Monitor the patient and compensate
for the gas volume introduced by the gas jet nebuliser as appropriate.
• For full details on using a nebuliser, see the User Guide that comes with that device.
English 53
Accessories
Connect the nebuliser unit with a T-piece into the inspiratory limb of the breathing circuit before the
patient. If one of the Aerogen nebuliser models is being used (ie, Aeroneb Solo and Aeroneb Pro), it can
be powered via the USB connector at the rear of the Astral device, or the Aerogen USB AC/DC adapter.
WARNING
• Only use compatible NONIN™ finger pulse sensors*.
• Pulse oximeter sensors must not be used with excessive pressure for prolonged periods as this
can cause patient pressure injury.
• The pulse oximeter sensor and cable needs to be verified for compatibility with Astral,
otherwise patient injury can result.
CAUTION
Factors that may degrade pulse oximeter performance or affect the accuracy of the measurement
include the following: excessive ambient light, excessive motion, electromagnetic interference,
blood flow restrictors (arterial catheters, blood pressure cuffs, infusing lines, etc.), moisture in the
sensor, improperly applied sensor, incorrect sensor type, poor pulse quality, venous pulsations,
anaemia or low haemoglobin concentrations, cardiogreen or other intravascular dyes,
carboxyhaemoglobin, methaemoglobin, dysfunctional haemoglobin, artificial nails or fingernail
polish, or a sensor not at heart level.
54
Accessories
*Please refer to the Ventilation accessories on www.resmed.com under the Products page for part
numbers of oximeter accessories with confirmed compatibility. For information on how to use these
accessories, refer to the user guide that comes with these accessories.
Once you have attached the pulse oximeter, a message will briefly display in the information bar.
Real-time SpO2 and Pulse readings can be viewed from the Monitoring menu.
Notes:
• Values from the SpO2 sensor are averaged over 4 heartbeats.
• Included SpO2 sensor is calibrated for the display of functional oxygen saturation.
• The No SpO2 monitoring alarm will activate if the pulse oximeter has been disabled or has a degraded
signal for more than 10 seconds or has been disconnected.
English 55
Accessories
CAUTION
To remove the cable, pull firmly on the connector. Do not twist.
56
Accessories
WARNING
The Astral should not be operated while in the Carry Bag. To ventilate while travelling, use the
Astral Mobility bag or SlimFit mobility bag.
CAUTION
Do not place any heavy or bulky objects in the zippered pocket on the inside front of the bag. This
could result in damage to the LCD Touch screen.
English 57
Accessories
WARNING
The Astral device should not be operated while in the Carry Bag. To ventilate while travelling, use
the Mobility Bag or SlimFit Mobility bag.
58
Power management
Power management
Helpful hints!
• Connect the ventilator to the mains power whenever possible. In the event of battery failure, connect
to mains power immediately to resume ventilation.
• An external power source (Astral External Battery or RPSII) should always be available for
ventilator-dependent patients.
• An external power source (Astral External Battery or RPSII) should be in use in mobile situations,
including when mains power is unavailable or disrupted. Do not rely solely on the internal battery for
mobile use.
• Ensure the external battery is sufficiently charged before using in mobile situations.
WARNING
• Beware of electrocution. Do not immerse the device, power supply or power cord in water.
• Make sure the power cord and plug are in good condition and the equipment is not damaged.
• Keep the power cord away from hot surfaces.
• Explosion hazard—do not use in the vicinity of flammable anaesthetics.
English 59
Power management
WARNING
Ensure that the power cord does not pose a tripping or choking hazard.
Note: The power cord is equipped with a push-pull locking connector. To remove, grasp the power cord housing and
gently pull the connector from the device. Do not twist its outer housing or pull on the cord.
60
Power management
WARNING
• Do not attempt to connect more than two external batteries. Battery specific messages and
alarms on the Astral device will not operate for any additional units.
• In the unlikely event of an issue occurring with the external battery, Astral will sound an alarm
and notify the user indicating that the device is operating on internal battery power. Ventilation
will continue, however, users should connect to an alternative external power source (eg, AC
power or another external battery) as soon as possible.
Alarms and messages relating to the External Battery may occur from time to time. All message
information will be displayed on the Astral user interface, and will be accompanied by an audible signal.
Refer to the Alarms Troubleshooting (see page 155) for further information.
English 61
Power management
CAUTION
• When using the Astral device with an RPSII, the internal battery will not be charged.
• Do not use the RPSII and external battery together.
CAUTION
• When using a car auxiliary adapter, start the car before plugging in the device's DC adapter.
• If the external DC power source drops to below 11V, the Astral will switch to internal battery.
• When the device is turned off while connected to the DC adapter, it will continue to draw
power from the external DC power source.
To connect DC power:
1. Connect the DC plug of the external DC power supply unit to the rear of the device.
2. Plug the other end of the power cord into the power outlet.
62
Power management
WARNING
• When using the Astral device as a backup ventilator, ensure the internal battery level is
checked on a regular basis.
• As the battery ages, the available capacity decreases. When the remaining battery capacity is
low, do not rely on the internal battery as the primary power supply.
• The internal battery should be replaced every two years or sooner when there is a noticeable
reduction in usage time when fully charged.
• The internal battery is not intended to serve as a primary power source. It should only be used
when other sources are not available or briefly when necessary; for example, when changing
power sources.
CAUTION
• Revert to AC mains power when the remaining capacity of the battery is low.
• The internal battery may stop charging when ambient temperatures of 35°C or more are
reached. This will be indicated with a Power fault/No charging alarm message.
• The internal battery will be depleted if the device is left in storage for an extended period of
time. During storage, ensure the internal battery is recharged once every six months.
• Storing the Astral device at temperatures exceeding 50°C for extended periods will accelerate
battery ageing. This will not affect the safety of the battery or the device.
While connected to mains power, the internal battery continues to charge when the device is operating
or in standby.
For more information on the expected operating time of the internal battery, see the Technical
Specifications.
English 63
Power management
Display Description
When either the external or internal battery is in use, but the device is not ventilating, the battery charge
level is displayed. The battery percentage is an average of all batteries connected to the system. Full
details of individual battery capacities can be reviewed in the Battery information page.
When either the external or internal battery is in use during ventilation, the remaining usage is displayed as
estimated by current operating conditions. The total will be the sum of all batteries connected to the
system.
When either the external or internal battery is charging, the charge battery symbol and percentage charged
is displayed.
Note: Only the Astral external and internal batteries are included in battery indicator calculations. RPSII battery levels are
not displayed.
64
Power management
2. Battery information
The battery information can be accessed from the Battery sub-menu in the Information menu. This
menu has two tabs:
- Charge—displays the current charge level (0-100%) for any batteries currently detected by the
system, as well as the total system charge.
- Maintenance—displays the full charge capacity and the charge cycle count for any batteries
currently detected by the system.
Regularly check the charge level of the internal battery and any connected external batteries. It is
recommended to replace any batteries at 400 charge cycles.
English 65
Using the Astral device for the first time
CAUTION
If any of the following checks fail, contact your Healthcare provider or ResMed for assistance.
WARNING
If no alarm sounds, do not use the ventilator.
Press the power switch at the back of the device to turn on the device. Check that the alarm sounds
two test beeps and the LEDs for the alarm signal and the alarm mute/reset button flash. The device
is ready for use when the Patient Home screen is displayed.
5. Disconnect the device from the mains and external battery (if in use) so that the device is powered
by the internal battery. Check that the Battery Use alarm is displayed and the battery LED is on.
Note: If the charge state of the internal battery is too low an alarm occurs. Refer to Troubleshooting (see page
155).
6. Reconnect the external battery (if in use) and check that the LED for the DC power supply is lit. The
External DC Power Use alarm will be displayed and the Alarm LED will light.
7. Reconnect the device to mains power.
8. Check the pulse oximeter sensor (if in use).
Attach the accessories according to the set up descriptions. From the Monitoring menu, go to the
Monitoring screen. Check that the values for SpO2 and pulse are displayed.
9. Check the oxygen connection (if in use). Check for damage to hoses or leaks. Check remaining
capacity of oxygen cylinders.
10. Perform a Learn Circuit.
66
Ventilation modes
Ventilation modes
Astral supports a variety of ventilation modes. Available modes will vary depending on the circuit type
chosen. Mode settings can be accessed from Main Settings tab from the Settings menu.
Note: Some features may not be available on your device.
P(A)CV * * *
P-SIMV
V-SIMV
PS * *
CPAP *
(S)T
P(A)C
iVAPS
English 67
Ventilation modes
*The International ventilator standard indicates that Paediatric patient type is intended to be used for a patient receiving less
than 300 mL, however Astral permits adjustment of ‘Vt’ setting parameter up to 500 mL for cases where ‘Vt’ is set such that it
compensates for leak in the breathing circuit.
WARNING
ResMed does not recommend 500 mL as the upper limit for the pediatric tidal volume use;
however, clinicians may choose this upper limit based on their clinical determination.
Supplementary features:
• Apnoea response
• Sigh breath (Astral 150 only)
• Manual breath (Astral 150 only)
68
Ventilation modes
(A)CV breath pattern showing one patient triggered breath amongst time triggered breaths, with
inspiratory duration defined by Ti. The patient-triggered breath re-schedules the next time-triggered
breath.
English 69
Ventilation modes
Parameter Setting
Trigger When Trigger Type is set to Flow:
Adult: Off, 0.5 to 15 [1.0] (L/min)
Paed: Off, 0.5 to 15 [0.5] (L/min)
When Trigger Type is set to Pressure:
Off, Very Low to Very High [Medium]
Rise Time (msec) Min, 150 to 900 [200]
Note: Some default settings are different for Mouthpiece circuit.
Supplementary features:
• Safety Vt (Tidal Volume)
• Apnoea response
• Sigh breath (Astral 150 only)
• Manual breath (Astral 150 only)
The P(A)CV breath pattern showing one patient triggered breath amongst time-triggered breaths. The
Safety Vt feature is turned off.
70
Ventilation modes
Parameter Setting
Resp.rate (per min) Mandatory breaths:
Adult: 2 to 50 [10]
Paed: 5 to 80 [20]
PEEP (cmH2O) Off, 3.0 to 20.0 [5.0]
P control (cmH2O) Mandatory breaths:
Adult: 2 to 50 [7]
Paed: 2 to 50 [7]
PS (cmH2O) Spontaneous breaths:
Adult: 2 to 50 [7]
Paed: 2 to 50 [7]
Ti (sec) Mandatory breaths
Adult: 0.2 to 5.0 [1.0]
Paed: 0.2 to 5.0 [0.6]
Cycle (%) Spontaneous breaths:
5 to 90, Auto [Auto]
Trigger type Flow / Pressure
Trigger When Trigger Type is set to Flow:
(double limb circuit only)
Adult: 0.5 to 15 [1.0] (L/min)
Paed: 0.5 to 15 [0.5] (L/min)
When Trigger Type is set to Pressure:
(double and single limb circuit)
Very Low to Very High [Medium]
Rise Time (msec) Min, 150 to 900 [200]
English 71
Ventilation modes
Supplementary features:
• Apnoea response
• Manual breath
Spontaneous breaths are permitted between mandatory breaths as shown in the figure above. In order
to promote synchrony with the patient's spontaneous efforts, mandatory breaths may be patient
triggered. Such patient triggering will cause some variation in the mandatory ventilation respiratory rate.
72
Ventilation modes
Parameter Setting
Resp.rate (per min) Mandatory breaths:
Adult: 2 to 50 [15]
Paed: 5 to 80 [15]
PEEP (cmH2O) Off, 3.0 to 20.0 [5.0]
PS (cmH2O) Spontaneous breaths:
Adult: 2 to 50 [7]
Paed: 2 to 50 [7]
Vt (Tidal Volume) (mL) Mandatory breaths:
Adult: 100 to 2,500 [500]
Paed:50 to 300 [100]*
PIF (L/min) When Volume Breath option is set to PIF
Mandatory breaths:
Adult: 10 to 120 [50]
Paed: 5 to 60 [10]
Ti (Inspiratory time) (sec) When Volume Breath option is set to Ti
Mandatory breaths:
Adult: 0.3 to 3.0 [1.0]
Paed: 0.3 to 3.0 [0.6]
Flow shape (%) Mandatory breaths:
100 (Constant), 75, 50, 25 [100]
Cycle (%) Spontaneous breaths:
5 to 90, Auto [Auto]
Trigger type Flow/Pressure
Trigger When Trigger Type is set to Flow:
(double limb circuit only)
Adult: 0.5 to 15 [1.0] (L/min)
Paed: 0.5 to 15 [0.5] (L/min)
When Trigger Type is set to Pressure:
(double and single limb circuit)
Very Low to Very High [Medium]
WARNING
ResMed does not recommend 500 mL as the upper limit for the pediatric tidal volume use;
however, clinicians may choose this upper limit based on their clinical determination.
English 73
Ventilation modes
Supplementary features:
• Apnoea response
• Manual breath (Astral 150 only)
Spontaneous breaths are permitted between mandatory breaths as shown in the figure above. In order
to promote synchrony with the patient's spontaneous efforts, mandatory breaths may be patient
triggered. Such patient triggering will cause some variation in the mandatory ventilation respiratory rate.
74
Ventilation modes
Parameter Setting
Supplementary features:
• Apnoea response
• Safety Vt (Tidal Volume)
• Manual breath (Astral 150 only)
• This graph shows PS mode with Resp. rate enabled with a transition from spontaneous-triggered to
time-triggered breaths. Cycling is constrained within the limits of Ti Min and Ti Max.
• To allow the patient sufficient time to exhale, Ti cannot exceed two-thirds of the breath period.
(Breath period is equal to 60/Resp. rate).
• To allow sufficient time to reach the targeted inspiratory pressure, Rise time cannot exceed
two-thirds of Ti Max.
English 75
Ventilation modes
76
Ventilation modes
P(A)C mode
P(A)C is a bilevel ventilation mode delivering pressure-controlled mandatory breaths:
- Inspiration is either initiated by the ventilator at a set rate (time-triggered breath) or the patient
(spontaneous-triggered breath). Spontaneous-triggered breaths re-schedule the next
time-triggered breath. Resp. rate and Trigger can be set to Off, but not at the same time. When
Trigger is set to Off, the mode name will be shown on the Information bar as PC.
- End of inspiration is controlled by the ventilator (time-cycled breath).
Parameter Setting
Resp. rate (per min) Adult: Off, 2 to 50 [15]
Paed: Off, 5 to 80 [15]
EPAP (cmH2O) 2 to 25 [5]
IPAP (cmH2O) Adult: 4 to 50 [12]
Paed: 4 to 50 [12]
Ti (sec) Adult: 0.3 to 4.0 [1.0]
Paed: 0.3 to 4.0 [0.6]
Trigger Off, Very Low to Very High [Medium]
Rise Time (msec) Min, 150 to 900 [200]
• To allow the patient sufficient time to exhale, Ti cannot exceed two-thirds of 60/Resp. rate.
• To allow sufficient time to reach the targeted inspiratory pressure, Rise time cannot exceed
two-thirds of Ti.
English 77
Ventilation modes
CPAP mode
CPAP mode delivers a constant level of pressure during inspiration and expiration.
With CPAP administered via a valved breathing system, the inspiratory triggering level is settable to
optimise the control of the exhalation valve and minimise the patient's work of breathing. Adjust trigger
sensitivity to accurately report patient respiratory rate.
In a vented system, the inspiratory triggering level is settable to optimise monitoring and alarm
performance.
Parameter Setting
CPAP (cmH2O) All circuits:
3.0 to 20.0 [5.0]
Trigger type Double limb circuit:
Flow/Pressure
Trigger Double limb circuit:
When Trigger Type is set to Flow:
Adult: 0.5 to 15 [1.0] (L/min)
Paed: 0.5 to 15 [0.5] (L/min)
When Trigger Type is set to Pressure:
Very Low to Very High [Medium]
Single limb circuit:
Very Low to Very High [Medium]
Single with intentional leak:
Very Low to Very High [Medium]
Supplementary features:
• Apnoea response (Valve circuits only)
CPAP operation with a single limb circuit with intentional leak is displayed.
78
Ventilation modes
Parameter Setting
Target Va (L/min) 1.0 to 30.0 [5.2]
English 79
Ventilation modes
80
Ventilation modes
Configuring iVAPS
There are two ways in which you can configure iVAPS mode:
• Adopting the recently learnt targets, from any vented mode (CPAP, (S)T or PAC)—learns the patient's
breathing pattern and calculates the target values automatically, or
• Entering the target values manually.
English 81
Ventilation modes
3. From the on-screen selections, press Ventilation Mode, then select iVAPS.
Note: The Review option will only be available for selection if there is at least five minutes worth of patient data
available.
82
Ventilation modes
5. Review the Learnt Targets and make adjustment to Pt Height if required. Select Confirm.
6. Select Apply.
Note: With AutoEPAP switched on, the information window name is updated to iVAPS.AutoEPAP.
English 83
Ventilation modes
3. From the on-screen selections, press Ventilation Mode, then select iVAPS.
4. Select Skip.
84
Ventilation modes
English 85
Ventilation modes
AutoEPAP
iVAPS mode only.
Note: This feature may not be available on your device.
The purpose of EPAP is to maintain upper airway patency. AutoEPAP automatically adjusts pressure in
response to flow limitation or obstruction of the upper airway. EPAP is adjusted within Min EPAP and
Max EPAP settings with the response depending on the degree of the upper airway obstruction.
WARNING
AutoEPAP is contraindicated when using an invasive interface.
Pressure support is adjusted on top of the EPAP. The maximum delivered pressure, EPAP plus pressure
support, is limited to 2 cmH2O below the set High Pressure limit. If the sum of EPAP plus pressure
support exceeds the maximum pressure limit, pressure support is sacrificed to maintain airway patency
(ie, EPAP). However, pressure support will not drop below the set minimum pressure (Min PS).
(a) For flow limitation, EPAP will increase at a maximum rate of 0.5 cmH2O (0.5 hPa) per breath.
(b) For obstructive apnoea, EPAP will increase at a rate of approximately 1 cmH2O (1 hPa) per second
spent in inspiration on termination of the apnoea.
(c) EPAP will start to decrease from the first breath after upper airway obstruction has resolved and will
continue to decrease slowly until either another upper airway flow limitation/obstruction occurs or Min
EPAP is reached.
The AutoEPAP algorithm does not address any other titration target such as lung recruitment to improve
oxygenation or offset intrinsic PEEP. Min EPAP should be set to treat lower airway conditions. AutoEPAP
will increase airway pressure to treat upper airway conditions.
86
Ventilation modes
The figure illustrates how Flow Shape affects breath delivery for a fixed volume. With volume breath
option set to PIF (Peak Inspiratory Flow), adjusting the flow shape alters the inspiratory duration,
whereas with volume breath option set to Ti (Inspiratory time), adjusting the flow shape alters the Peak
Inspiratory Flow.
When the flow shape is set to 100%, the flow is generally constant during inspiration. For decreasing
percentages, the flow starts at the peak flow and declines to approximately the percentage setting of
this value at the end of inspiration.
English 87
Ventilation modes
Interdependence of controls
Dynamic settings limits
The adjustable range of one setting may be restricted by the value of another. When a dynamic limit of
this kind is reached, a message describing the limitation (interdependence) is displayed on the
Information bar and the Apply button is disabled.
To enable the Apply button, modify one of the conflicting settings. For example in this case, to continue
IPAP must be increased or EPAP decreased.
88
Ventilation modes
When making adjustments to the high pressure alarm limit, you are prompted to confirm the changes to
any affected settings. If these setting changes are not accepted, the change to the high pressure alarm
limit is not applied.
English 89
Ventilation modes
Note: Cycling is constrained within the limits of TiMin and TiMax. This means that the inspiratory period cannot be
less that TiMin or longer than TiMax.
90
Ventilation modes
English 91
Ventilation modes
2. From the Trigger type option, press Change. The current Trigger Type is highlighted.
3. Select Flow. You will be returned to the Circuit screen where the changed Trigger type will be
displayed.
92
Ventilation modes
Note: Cycling is constrained within the limits of TiMin and TiMax. This means that the inspiratory period
cannot be less that TiMin or longer than TiMax.
English 93
Supplementary features
Supplementary features
Manual breath settings
Manual breath feature allows the user to manually insert a breath within the currently delivered
breathing pattern.
The Manual breath feature is used to trigger manual breath(s) and can be pressed at any time.
• If pressed during expiration, the manual breath will be delivered immediately.
• If pressed during inspiration, the manual breath will be delivered 300 ms after the end of the current
inspiration.
The manual breath can be configured as a magnified version of the primary breath (magnification factor
set from 100 to 250%).
For pressure-target breaths, the inspiratory duration and pressure are magnified proportionally.
For volume-target breaths, the delivered volume is magnified proportionately.
Adjustable parameters:
Parameter Setting
Manual Breath Off, On [Off]
Magnitude (%) 100 to 250 [150]
Note: The Magnitude (%) value is clipped at 2500 mL and 500 mL for Adult and Paediatric patient types respectively.
For pressure-controlled mandatory breaths, the P control and breath duration are magnified by the
magnification factor. This applies to mandatory breaths in P(A)CV mode and in P-SIMV and V-SIMV
modes.
94
Supplementary features
For volume-controlled mandatory breaths, the volume is magnified by firstly flattening the flow shape,
then by extending the Ti and finally by increasing the PIF. This applies to mandatory breaths in (A)CV and
V-SIMV modes.
English 95
Supplementary features
Sigh settings
The Sigh breath feature allows the user to program the delivery of a 'sigh breath' at a regular interval
(sigh interval) within the currently delivered breathing pattern.
The sigh breath is a magnified version of the primary breath (Magnification factor set from 120 to 250%).
For pressure-target breaths, the inspiratory duration and P control are magnified proportionally.
For volume-target breaths, volume is magnified.
Adjustable parameters − (A)CV & P(A)CV Modes only:
Parameter Setting
Sigh Breath Off, On [Off]
Sigh Alert Off, On [Off]
Interval (min) 3 to 60 [10]
Magnitude (%) 120 to 250 [150]
Note: The Magnitude (%) value is clipped at 2500 mL and 500 mL for Adult and Paediatric patient types respectively.
96
Supplementary features
Apnoea Settings
The Astral device allows the clinician to define what should happen when the device detects an apnoea.
An apnoea refers to the absence of breath within a configurable period: T apnoea (Apnoea Interval).
Apnoea can be defined as an absence of patient-triggered breaths (No Spont Breath), or an absence of
any breaths (No Breath), whether they are patient, time, or manually-triggered breaths.
Three types of Apnoea response settings can be selected on the Astral device:
1. Alarm only
2. Alarm + apnoea ventilation ((A)CV breath pattern)
3. Alarm + apnoea ventilation (P(A)CV breath pattern)
4. Off
When Alarm + apnoea ventilation ((A)CV or P(A)CV) is selected, apnoea ventilation is automatically
triggered following the detection of an apnoea event. Apnoea ventilation is displayed on the Information
bar.
The Apnoea response will deactivate when the patient triggers three consecutive breaths.
ResMed recommends that Apnoea ventilation is enabled whenever the Resp. rate is set to Off.
WARNING
Setting Apnoea Detection to No breath and T apnoea to a value greater than 60s will make the
Apnoea Alarm and Response ineffective.
English 97
Supplementary features
WARNING
ResMed does not recommend 500 mL as the upper limit for the pediatric tidal volume use;
however, clinicians may choose this upper limit based on their clinical determination.
98
Supplementary features
English 99
Supplementary features
Parameter Setting
Safety Vt (mL) Adult: Off, 100 to 2500 [Off]
Paed: Off, 50 to 300 [Off]*
PS Max (PS mode) PS to 50 [PS + 5]
P control max (P(A)CV) P control to 50 [P control + 5]
IPAP Max (S(T), PAC modes) IPAP to 50 [IPAP + 5]
*The International ventilator standard indicates that Paediatric patient type is intended to be used for a patient receiving less
than 300 mL, however Astral permits adjustment of ‘Vt’ setting parameter up to 500 mL for cases where ‘Vt’ is set such that it
compensates for leak in the breathing circuit.
WARNING
ResMed does not recommend 500 mL as the upper limit for the pediatric tidal volume use;
however, clinicians may choose this upper limit based on their clinical determination.
100
Non-invasive ventilation (NIV)
Mouthpiece NIV
The decision to use mouthpiece ventilation is generally a consultative process between clinician and
patient after considering the risks and advantages of this form of therapy. For example, the patient must
possess adequate cognition, head/neck/mouth control, and oropharyngeal function, and significant
individualized training may be required compared to other forms of ventilation.
To independently assess the patient condition during mouthpiece ventilation, SpO2 can be monitored
using a pulse oximeter. Should the oximeter sensor become disconnected from the patient's finger, the
No SpO2 monitoring alarm will activate.
CAUTION
Mouthpiece ventilation may not be appropriate for particular patients and hence clinical discretion
is essential.
Mouthpiece ventilation
The settings below are suggested for ‘open’ or ‘sip’ mouthpiece ventilation where the patient exhales to
atmosphere frequently or continuously, eg, for on-demand daytime ventilation via a 15 mm non-vented
mouthpiece. Other circuit types, modes and settings are also available if required.
English 101
Non-invasive ventilation (NIV)
Low pressure alarms are sometimes used to imply circuit disconnection and are quick to activate. Should
this be an annoyance, for example when the patient is receiving a partial breath or missing a breath, or if
a false triggered breath occurs, it is at the discretion of the Clinician to turn OFF. Other mitigations may
need to be put in place to ensure that patient safety is not compromised. This may include external
monitoring, SpO2 alarm, or full-time supervision.
Mouthpiece Ventilation with Astral 100/150 mouthpiece circuit is not intended to support continuous
exhalation into the circuit. The non-user adjustable NV Mask/Rebreathing alarm will activate if the device
detects continuous exhalation into the circuit. For patients that may prefer continuous exhalation into the
circuit, a circuit with an expiratory valve or intentional leak should be considered.
102
Alarms
Alarms
The Astral device activates alarms to alert you to conditions that require attention to ensure patient
safety. When an alarm is activated, the Astral device provides both audible and visual alerts, and displays
an alarm message in the Alarm display on the Information bar.
As part of the alarm system (eg, the overpressure protection and system alarms), Astral may perform an
automatic restart. An automatic restart checks systems and ensures correct operation of the alarm.
Indicator Description
1 Alarm display Displays either the alarm message for the highest priority active alarm, or the last
alarm not yet reset.
Press the Alarm display for further alarm information.
Certain conditions may result in multiple alarms. indicates that there are
multiple active alarms. Press when displayed to view all alarms and
respond appropriately. Alarms are displayed in order of priority.
2 Active Alarms screen Displays the full set of active alarms. Will automatically display upon activation of
an alarm in Patient mode.
3 Information menu Some alarms clear automatically. To view a history of alarms, view the alarm log
through the Information menu.
English 103
Alarms
Indicator Description
Alarm priority
Alarms are classified into relative priority (high, medium and low) according to the severity and urgency
of the alarm condition. Respond to all alarms. An immediate response is required for high priority alarms.
104
Alarms
The following list of alarms is ordered by relative importance within priority. Adjustable alarms can be
changed to suit therapy and care requirements.
English 105
Alarms
in the Alarm display indicates that there are multiple active alarms. Although multiple alarms can be
active simultaneously, the Alarm display only shows the highest priority alarm. The full set of active
alarms is displayed in the Active alarms screen.
When the highest priority alarm is cleared, the next highest priority alarm displays in the Alarm display.
Muting alarms
You can temporarily mute the audible alert on the Astral device for a two minute period. The Alarm
display and Alarm bar continue to display the alarm as usual. If after two minutes the alarm condition is
still present, the audible alert will sound again.
You can also use the Alarm Mute in advance, to 'pre-silence' alarms that you expect to occur. This can
be helpful during suctioning procedures or when intending to disconnect the patient from the ventilator
for a short period.
If any of the following high priority alarms are triggered, the Alarm mute is automatically cancelled:
• Total power failure
• Critically low battery
• Incorrect circuit.
Press .
The alarm is silenced for two minutes. During that period, is displayed on the Information bar and
flashes.
Note: Pressing the Alarm mute/reset button again during the Alarm Mute period will reset the displayed alarm. Refer
to Resetting alarms (see page 107).
106
Alarms
1. Press . Alarm mute is active for two minutes. During that period, is displayed on the
Resetting alarms
The following alarms cannot be manually reset. For these alarms you must correct the cause of the
alarm. Resolving the alarm will automatically clear the display.
• Total power failure
• Critically low battery
• Low pressure
• Pressure line disconnected
• Battery inoperable
• Circuit disconnection
• Incorrect circuit adapter.
Resetting an alarm removes that alarm from the Alarm display and the Active alarms screen, and turns
off the visual and audible alerts. An active alarm should only be reset after the situation that caused the
alarm has been attended to. If the alarm condition has not been corrected, the alarm will activate again.
The Astral device may automatically clear an alarm when the condition that triggered the alarm is
corrected. When an alarm is cleared it no longer displays in the Active alarms screen and the audible and
visual alerts cease.
When an alarm is cleared or manually reset, the Alarm display then shows the next highest priority active
alarm.
To view a complete alarm history, see the Events log available from the Information menu.
2. Press again to reset the alarm. The alarm message is removed from the Alarm display. It
is also cleared from the Active alarms screen.
Note: You can carry out this procedure with the Active alarms screen open, if you want visibility of all the active
alarms as you perform the reset.
English 107
Alarms
2. Press Reset all to reset multiple alarms. Only those alarms that can be reset, will be reset. Any
remaining alarms will require user intervention and correction.
3. Complete any required action to resolve the remaining alarms.
4. Press OK to close the Active alarms screen and return to the previous screen.
CAUTION
Adjusting alarm thresholds to maximum or minimum values may render the alarm ineffective.
WARNING
The safety and effectiveness of alarms settings should be verified for each enabled Program.
Note: To adjust the volume of the audible alert, refer to Device settings (see page 31).
2. Press .
3. Press the alarm setting that you want to adjust. The selected setting is highlighted in orange and the
up and down scroll arrows appear. Minimum and maximum settable values appear with the scroll
arrows.
108
Alarms
WARNING
• Alarm volume cannot be adjusted separately for individual alarms. Adjusting the alarm volume
will change the volume of all alarms, independent of alarm priority level.
• When adjusting alarm volume, ensure that the alarm can be heard above the ambient noise
levels that the patient may experience in a variety of settings, including use in noisy
environments or inside mobility bags.
In the example below the current alarm volume is '5' however the minimum alarm volume has been set
at '3'. The '1' and '2' volume options are now disabled and not able to be selected by the patient or carer.
English 109
Alarms
3. Press Test to test the alarm. The alarm beeps twice and the LED flashes to indicate it is working
correctly. Confirm the alarm beeps twice. Confirm the Alarm bar flashes red, then yellow. Confirm
the mute button flashes.
WARNING
If no alarm sounds, do not use the ventilator.
CAUTION
If only one beep is heard, or the Alarm bar does not flash red, then yellow, return the device for
servicing.
CAUTION
A test of the Remote Alarm should be performed prior to initial use and every change of battery.
Test the alarm periodically as per the facility policy. For dependent patients perform a test on a
daily basis.
110
Testing the alarms
This section describes functional tests to confirm correct technical operation of Astral alarms. The
efficacy of therapy alarm limits should be assessed clinically.
It is recommended that alarms be tested individually. To do so, turn off all alarms except the alarm that is
going to be tested. If the ventilator fails any of the alarm tests, see Troubleshooting.
It is not necessary to test all alarms for every alarm condition. There is no specific sequence in which the
alarms must be tested.
Many of the following tests require use of a passive test lung.
CAUTION
When the alarm tests are completed ensure the Astral device is restored to its original state.
English 111
Testing the alarms
Low Vte Set the alarm limit above the current volume.
High Vte Set the alarm limit below the current volume.
Low Vti Set the alarm limit above the current volume.
High Vti Set the alarm limit below the current volume.
Low MVe Set the alarm limit above the current minute volume.
High MVe Set the alarm limit below the current minute volume.
Low MVi Set the alarm limit above the current minute volume.
High MVi Set the alarm limit below the current minute volume.
Low Resp Rate Set the alarm limit above the set respiratory rate.
High Resp Rate Set the alarm limit below the set respiratory rate.
Apnoea Set the respiratory rate to Off.
112
Testing the alarms
English 113
Alarm settings and conditions
Low Vti Inspiratory tidal volume is less than Low Vt for the duration Adult: Off, 50 to 2990 [100]
of three breaths at the current average respiratory rate. Paed: Off, 10 to 995 [25]
High Vti Inspiratory tidal volume is greater than High Vt for three Adult: Off, 60 to 3000 [2500]
consecutive breaths. Paed: Off, 25 to 1000 [500]
Low Vte Expiratory tidal volume is less than Low Vt for the duration Adult: Off, 50 to 2990 [100]
of three breaths at the current average respiratory rate. Paed: Off, 10 to 995 [25]
High Vte Expiratory tidal volume is greater than High Vt for three Adult: Off, 60 to 3000 [2500]
consecutive breaths Paed: Off, 25 to 1000 [500]
WARNING
The measurement of tidal volumes may be affected by leak.
Low MVi Inspiratory minute ventilation is less than Low MV Adult: Off, 0.5 to 59.9 [3]
Paed: Off, 0.2 to 59.9 [0.5]
High MVi Inspiratory minute ventilation is greater than High MV Adult: Off, 0.6 to 60 [20]
Paed: Off, 0.3 to 60 [10]
Low MVe Expiratory minute ventilation is less than Low MV Adult: Off, 0.5 to 59.9 [3]
Paed: Off, 0.2 to 59.9 [0.5]
High MVe Expiratory minute ventilation is greater than High MV Adult: Off, 0.6 to 60 [20]
Paed: Off, 0.3 to 60 [10]
114
Alarm settings and conditions
Low Resp Rate Respiratory rate is less than the Low Resp rate setting Adult: Off, 2 to 79 [4]
Paed: Off, 5 to 98 [12]
High Resp Rate Respiratory rate is greater than the High Resp rate setting Adult: Off, 3 to 80 [80]
Paed: Off, 6 to 99 [99]
Pressure alarms/limits
Pressure alarms include:
• High Pressure
• Obstruction
• Low Pressure
• PEEP.
High pressure
For pressure-target modes, the High Pressure setting acts as a pressure limit, holding pressure-target
therapies to 2 cmH2O less than the High Pressure setting. Refer to Making use of Astral's High Pressure
limit on page 88. This allows the clinician to globally limit pressure therapies if required (eg, in the context
of mask NIV).
The Astral High Pressure limit acts as a traditional High Pressure alarm for volume target therapies and
fault situations, as described below.
When a high inspiratory pressure is detected during volume target modes, the Astral device immediately
cycles to the expiratory phase and an information message is displayed. Sustained high pressure will
result in the following alarms.
High pressure may be caused by change in patient conditions.
English 115
Alarm settings and conditions
Obstruction
WARNING
When an obstruction is detected, ventilation is suspended. Ventilation can only resume when the
monitored pressure reduces to below 3 cmH2O and the monitored expiratory flow reduces below 1
L/min.
Obstruction may be caused by a circuit obstruction. Obstruction can typically be caused by a pinched
PEEP valve control tube or a blocked exhalation valve outlet.
Note: The Obstruction alarm will not be triggered if there is an obstruction after the vent, proximal expiratory valve,
or Y piece depending on which circuit is used. To detect such a condition, use the Apnoea, Low Vt and/or Low MV
alarms.
Low pressure
Low pressure may be caused by severe leak.
Note: To allow for mouthpiece ventilation and other exhale to atmosphere configurations, this alarm may be turned
off. For more information, refer to Mouthpiece Settings (see page 101).
PEEP
PEEP alarms activate when the monitored positive expiratory end pressure parameter (PEEP) exceeds
the corresponding alarm setting. When ventilation starts, or after a PEEP setting change, PEEP alarms
are temporarily disabled until:
• the monitored PEEP is within 2 cm H2O of the PEEP setting for three consecutive breaths, or
• 30 seconds has passed.
PEEP alarms may be caused by:
• circuit disconnection
• high leak
• device fault
• obstruction.
116
Alarm settings and conditions
Low PEEP Monitored PEEP is less than or equal to 2 cmH2O below the Low PEEP
PEEP ventilation setting for 10 consecutive breaths.
Off / On [On]
High PEEP Monitored PEEP is greater than 2 cmH2O above the PEEP N/A
setting for 10 consecutive breaths.
High leak
Alarm Activates when Alarm setting
High Leak The difference between MVe and MVi is greater than the Off, 20 to 80 [Off] (%)
(double limb) Leak alarm setting for a 10 second period.
High Leak Estimated unintentional leak is greater than the Leak alarm 5 to 80, Off [40] (L/min)
(single limb with setting for a 20 second period.
intentional leak)
NV Mask/Rebreathing
Alarm Activates when Alarm setting
NV mask/Rebreathing A non-vented mask is used in a vented circuit. Off / On [On]
An obstruction of the vents in a vented circuit leading to Note: Non-selectable in Mouthpiece
rebreathing. mode.
Rebreathing on a vented circuit using a non-vented mask or
on a mouthpiece circuit during continuous exhalation into
the circuit.
Rebreathing is greater than 20% for vented circuit or 50%
for mouthpiece circuit, for 10 consecutive breaths.
English 117
Alarm settings and conditions
Disconnection alarm
The disconnection alarm monitors multiple parameters to assess whether the ventilator breathing circuit
has become disconnected from the patient.
Note: This alarm should be configured in conjunction with other alarms that can be used to detect circuit
disconnection. Refer to the Detecting circuit disconnection and de-cannulation (see page 121) section.
CAUTION
Be certain that all forms of patient disconnection can be successfully detected, including the
patient interface (mask / cannula / endotracheal tube / mouthpiece) detaching from the patient.
Refer to Testing the disconnection alarm on page 113.
Oxygen alarms
High and low FiO2 alarms are only available when an oxygen sensor is installed. The absence at the start
of ventilation, or loss during ventilation, of oxygen monitoring is indicated by the No FiO2 alarm.
118
Alarm settings and conditions
Oximetry alarms
The oximetry alarms are only available when the pulse oximeter is connected to the Astral device. The
alarms are enabled even when the ventilator is in standby.
The SpO2 and Pulse rate alarms are automatically disabled when the pulse oximeter is disconnected
from the Astral device.
SpO2
Oxygen saturation alarms activate when the monitored SpO2 parameter exceeds the corresponding
alarm setting (no delay).
Pulse rate
Pulse rate alarms activate when the average pulse rate (Pulse) exceeds the corresponding alarm setting
(no delay).
Apnoea alarm
For details on how to configure the Apnoea alarm and activation conditions refer to Apnoea settings (see
page 97).
English 119
Alarm settings and conditions
Power alarms
Power alarms are not adjustable.
CAUTION
Data cannot be saved while there is a Critically low battery or Battery inoperable alarm. Program
selections made while these alarms are active may be lost if the device is restarted. Recording of
ventilation data and alarms is suspended.
System alarms
System alarms are not adjustable.
Safety reset complete the device activated a restart to resume normal operation.
Last device test failed ventilation is started after the device self-test fails.
Device overheating internal component/s of the device are becoming too hot.
System fault a technical fault is detected within the Astral device at initial power up or during
ventilation.
Safety system fault a technical fault is detected within the Astral device at initial power up or during
ventilation.
120
Detecting circuit disconnection and de-cannulation
CAUTION
Be certain that all forms of patient disconnection can be successfully detected, including the
patient interface (mask / tracheal tube / mouthpiece) detaching from the patient.
Refer to Testing the disconnection alarm on page 113.
Multiple alarms may be required. Independent monitoring can be used as an alternative.
WARNING
Alarm settings may be sensitive to any changes to the circuit, ventilation settings or co-therapy.
Test the effectiveness of the alarm after any of these changes are made.
The following table provides the most appropriate alarms for use in detecting circuit disconnection.
English 121
Astral Disconnection Alarm
Any active alarm will clear when the monitored value drops below the set Disconnection Tolerance.
CAUTION
Be certain that patient disconnection can be detected, including if the patient interface becomes
accidentally detached from the patient (eg, if the tracheostomy tube / endotracheal tube / mask /
mouthpiece remains attached to the circuit). For example, to check that accidental decannulation
of a tracheostomised patient can be detected, simulate disconnection using a tracheostomy tube
one size smaller than the patient’s tube.
Refer to Testing the disconnection alarm on page 113.
122
Astral Disconnection Alarm
English 123
Astral Disconnection Alarm
124
Astral Disconnection Alarm
5. Once confirmed, the Disconnect Alarm is disabled and the Disconnect Alarm OFF icon will be
displayed next to Standby on the Information bar.
English 125
Astral Disconnection Alarm
126
Data management process
To remove the USB stick, simply pull it out of the USB connector on completion of transfer. If data was
being transferred at the time, a message in the Information bar alerts you to a failed transfer.
WARNING
Only connect devices specially designed and recommended by ResMed to the data
communication ports. Connecting other devices could result in patient injury, or damage to the
Astral device.
To transfer data:
1. From the Settings menu select Patient Data from the USB sub-menu.
2. Press Save >. When the transfer is complete a status message is displayed.
English 127
Data management process
PEEP 1 Hz
Avg. P 1 Hz
Maximum Inspiratory flow 1 Hz
Leak 1 Hz
SpO2 1 Hz
Pulse 1 Hz
FiO2 1 Hz
Vt 1 Hz
MV 1 Hz
Va 1 Hz
Resp. rate 1 Hz
I:E ratio 1 Hz
Ti 1 Hz
Te 1 Hz
% Spont. trig
% Spont. cyc
RSBI 1 Hz
AHI2 Median
AI2 Median
1
One set of summary data is provided per day for each program used.
2
The Apnoea Index (AI) and Apnoea Hypopnoea Index (AHI) are only recorded when a single limb circuit with intentional leak is
used. AI indicates the number of times per hour that patient flow reduces by more than 75% of nominal flow for a period of
10 seconds or more. AHI indicates the number of times per hour that patient flow reduces by more than 50% of nominal flow for
a period of 10 seconds or more.
128
Deleting patient data
English 129
Cleaning and maintenance
WARNING
• A patient treated by mechanical ventilation is highly vulnerable to the risks of infection. Dirty or
contaminated equipment is a potential source of infection. Clean the Astral device and its
accessories regularly.
• Always turn off and unplug the device before cleaning and be sure it is dry before plugging
back in.
• Do not immerse the device, pulse oximeter or power cord in water.
The Astral device can be cleaned using an anti-bacterial solution on a clean, non-dyed disposable cloth.
After replacing any accessory in the patient circuit, ResMed recommends you perform a Learn Circuit.
CAUTION
Clean only exterior surfaces of the Astral device.
The following cleaning solutions are compatible for use on a weekly basis (except as noted) when
cleaning the external surfaces of the Astral device:
• Actichlor Plus
• Bleach (1:10) (May also be known as 'dilute hypochlorite').
• Isopropanol
• Cavicide*
• Mikrozid*
*Suitable for cleaning on a monthly basis only.
Weekly
1. Wipe the exterior of the device with a damp cloth using a mild cleaning solution.
2. Inspect the condition of the circuit adapter for entry of moisture or contaminants. Replace as
necessary, or at regular intervals not less than once every six months.
3. Test the alarm sounders, refer to Testing the alarm sounders (see page 109).
Monthly
1. Inspect the condition of the air filter and check whether it is blocked by dirt or dust. With normal use,
the air filter needs to be replaced every six months (or more often in a dusty environment).
2. Check the charge level of the internal battery by:
- removing external power and operating the device on internal battery for a minimum of
10 minutes.
- reviewing the remaining battery capacity, refer to Using the Internal battery (see page 63).
- restoring external power once the test is complete.
130
Cleaning and maintenance
Multi-patient use
WARNING
• To prevent the risk of cross-contamination, an antibacterial filter, placed on the inspiratory port
is mandatory if the device is to be used on multiple patients as under some fault conditions,
expired gas may be returned through the inspiratory port.
• The expiratory module, internal antibacterial filter, expiratory flow sensor and blue membrane
come into contact with exhaled gases but do not form part of the inspiratory pathway.
In addition to the cleaning and maintenance instructions for single patient use, you must perform the
following before the device is provided to a new patient.
Double limb adapter (used with Double limb circuits) For general hygiene the Double limb adapter should be replaced or
protected with an antibacterial filter.
Mask Masks should be reprocessed when used between patients. Refer
to the User guide provided with the mask in use.
Patient circuits Replace or sterilise. Refer to the manufacturer's recommended
cleaning instructions.
Humidifier Refer to the User Guide provided with the humidifier in use.
Internal battery Check the charge level by removing the external power and
operating the device on internal battery for a minimum of ten
minutes. Review the remaining battery capacity and restore
external power.
English 131
Cleaning and maintenance
Replacing components
Replacing the air filter
Inspect the condition of the air filter and check whether it is blocked by dirt or dust. With normal use, the
air filter needs to be replaced every six months (or more often in a dusty environment).
CAUTION
Do not wash the air filter. The air filter is not washable or reusable.
Replacing the expiratory flow sensor and antibacterial filter (Astral 150 only)
CAUTION
• Regularly check the double limb adapter (expiratory valve) for moisture and contaminants.
Particular care should be taken when using nebulisers or humidifiers.
• When replacing the expiratory flow sensor and antibacterial filter also replace the blue
membrane and adapter.
132
Cleaning and maintenance
To remove and replace the expiratory flow sensor and antibacterial filter:
Before replacing the expiratory flow sensor, turn off the device and remove mains power and/or external
battery.
1. Turn over the device and place on a soft surface (to protect the LCD screen).
2. Press and hold the eject button. Pull the cover out towards you.
3. Lift out the adapter and discard.
4. Remove the blue membrane (including the white antibacterial filter) and discard.
5. Remove and insert a new expiratory flow sensor.
6. Insert a new antibacterial filter.
7. Insert a new blue membrane ensuring the rear tab and surrounds sit flush in the enclosure.
8. Insert a new adapter, gently pushing down so it sits firmly in place.
9. Place the cover over the enclosure, ensuring the runners on the device and the cover are aligned.
Slide the cover back into place until the latch clicks.
CAUTION
Following the replacement of the Expiratory flow sensor, you must run a Learn Circuit to calibrate
the new sensor and ensure accurate therapy and monitoring.
To calibrate the Expiratory flow sensor, connect a double limb circuit. Prior to performing the Learn
Circuit, ensure double circuit is selected on the Circuit sub-menu.
English 133
Cleaning and maintenance
CAUTION
Following the replacement of the Oxygen sensor, you must run a Learn Circuit to calibrate the new
sensor and ensure accurate therapy and monitoring.
134
Cleaning and maintenance
WARNING
Turn off and disconnect oxygen before replacing the internal battery.
CAUTION
• Lithium-ion batteries have built-in safety protection circuits, but can still be dangerous if they
are not used correctly. Damaged batteries can fail or catch fire.
• As the battery ages, the available capacity decreases. When the remaining battery capacity is
low, do not rely on the internal battery as the primary power supply.
• Only recharge the internal battery inside the device or by using a ResMed approved charger.
Avoid hard, physical impact on the device.
• Use the internal battery in accordance with the intended use stated in this guide. Damage to
equipment or injury can result from modifying the equipment or its operation.
• The internal battery should be replaced every two years or sooner when there is a noticeable
reduction in usage time when fully charged.
English 135
Cleaning and maintenance
Servicing
WARNING
Inspection and repair should only be performed by an authorised agent. Under no circumstances
should you attempt to service or repair the device yourself. Failure to do so could void your Astral
device warranty, damage the Astral device or result in possible injury or death.
Note: Retain the original packaging of the Astral device for use when shipping to/from an authorised ResMed Service
Centre.
Maintenance Timetable
The Astral device should be serviced by an authorised ResMed Service Centre according to the following
schedule. The Astral device is intended to provide safe and reliable operation provided that it is operated
and maintained in accordance with the instructions provided by ResMed. As with all electrical devices, if
any irregularity becomes apparent, you should exercise caution and have the device inspected by an
authorised ResMed Service Centre.
With regular servicing, the expected service life of an Astral device is 8 years.
Servicing schedule from the date of first use:
Helpful hint!
If Service Reminders are enabled you will get a notification on the device when two year preventative
maintenance is nearly due.
Internal Battery
The expected life of the internal battery is two years. The internal battery should be replaced every two
years or sooner when there is a noticeable reduction in usage time when fully charged. During storage
ensure that internal battery is recharged once every six months.
Helpful hint!
If Service reminders are enabled you will get a notification on the device when there is a measurable
reduction in battery usage time.
136
Cleaning and maintenance
Device information
Device information, including the Next service due date indicating when the next preventative
Upgrading software
Software upgrades can only be performed by an authorised service representative. Astral has the ability
to be upgraded with a USB stick containing the appropriate software. Please contact your authorised
agent for further information.
Double limb adapters For general hygiene, the double limb adapter should be replaced between patients or protected with an
antibacterial filter.
Air filter Replace every six months (or more if necessary).
Mask Masks should be reprocessed or replaced between patients. Cleaning, disinfection and sterilisation
instructions are available from the ResMed website, www.resmed.com/downloads/masks. If you do
not have internet access, please contact your ResMed representative.
Patient circuits Follow the protocol of the hospital/healthcare facility and the recommendations of the manufacturer.
Humidifier Follow the protocol of the hospital/healthcare facility and the recommendations of the manufacturer.
Antibacterial filter Replace as required and between patients.
English 137
Technical specifications
Technical specifications
Operating pressure range Single limb with valve or double limb with valve: 3 to 50 hPa
Single limb with intentional leak: 2 to 50 hPa
CPAP: 3 to 20 hPa
Maximum working pressure limit: 10 to 80* hPa
*Not applicable on all device variants
Forced cycling occurs if the Pressure alarm limit is exceeded.
Operating tidal volume range (volume Adult patient type: 100 to 2500 mL
control modes) Paediatric patient type: 50 to 300 mL*
Circuit resistance and compliance range Paediatric patient setting:
for stated accuracy of monitoring and Circuit resistance range (circuit with intentional leak): 0 to 8 hPa at
control** 60 L/min
Circuit resistance range (circuit with valve): 0 to 20 hPa at 60 L/min
Circuit resistance range (mouthpiece circuit): 0 to 5 hPa at 60 L/min
Circuit compliance range: 0 to 4 mL / hPa
Adult patient setting:
Circuit resistance range (circuit with intentional leak): 0 to 20 hPa at
120 L/min
Circuit resistance range (circuit with valve): 0 to 35 hPa at 120 L/min
Circuit resistance range (mouthpiece circuit): 0 to 15 hPa at 120 L/min
Circuit compliance range: 0 to 4 mL / hPa
Breathing resistance under single Paediatric circuit
fault*** Inspiration: 2.2 hPa (at 15 L/min), 5.3 hPa (at 30 L/min)
Expiration: 2.4 hPa (at 15 L/min), 5.0 hPa (at 30 L/min)
Adult circuit
Inspiration: 5.7 hPa (at 30 L/min), 8.3 hPa (at 60 L/min)
Expiration: 4.2 hPa (at 30 L/min), 6.2 hPa (at 60 L/min)
Maximum flow 220 L/min
Inspiratory trigger (nominal) Inspiratory trigger occurs when patient flow exceeds trigger setting.
characteristics Double limb with valve (flow trigger): 0.5 to 15.0 L/min
Single limb with valve or double limb with valve: 1.6 to 10.0 L/min (in five
steps)****
Single limb with intentional leak: 2.5 to 15.0 L/min (in five steps)
Mouthpiece (tube only) circuit: 2.0 to 4.0 L/min (in four steps)
Expiratory cycle (nominal) Cycle occurs when inspiratory flow declines to the set percentage of peak
characteristics inspiratory flow.
5 to 90%
Sound pressure level 35 ± 3 dBA as measured according to ISO 80601-2-72:2015
Sound power level 43 ± 3 dBA as measured according to ISO 80601-2-72:2015
Alarm volume range 56–85 dBA (in five steps) as measured according to IEC 60601-1-8:2012
Data storage 7 days of high-resolution airway pressure, respiratory flow and delivered
volume (sampled at 25 Hz).
7 days of breath-related therapy data (sampled at 1 Hz).
365 days of statistical data per program.
138
Technical specifications
140
Technical specifications
WARNING
ResMed does not recommend 500 mL as the upper limit for the pediatric tidal volume use;
however, clinicians may choose this upper limit based on their clinical determination.
English 141
Technical specifications
WARNING
Under normal or single fault conditions the entire pneumatic flow path can become contaminated
with body fluids or expired gases if antibacterial filters are not installed on the ventilator outlet and
double limb adapter exhalation port.
142
Monitoring
Monitoring
This section specifies the monitored parameters of the Astral device. Monitored therapy parameters are
only displayed during ventilation. Other parameters (eg, battery charge level and oximetry data) are
monitored even when not ventilating.
The Astral device monitors the following parameters:
Time parameters
Parameter Updates Units Range Resolution Accuracy
Ti End of sec 0 to 10 0.01 ±(20 ms + 5%)
inhalation
Te End of sec 0 to 15 0.01 ±(20 ms + 5%)
exhalation
I:E End of breath N/A 1:9.9 to 9.9:1 0.1 ±10%
Resp. rate End of breath 1/min 0 to 99 1 ±1/min
English 143
Monitoring
Pressure parameters
Parameter Updates Units Range Resolution Accuracy
Pressure Continuous cmH2O or hPa 0 to 99 0.1 ±(0.5 hPa + 4% of
or mbar actual pressure)
PIP End of cmH2O or hPa 0 to 99 0.1 ±15%
inhalation or mbar
PEEP End of cmH2O or hPa 0 to 30 0.1 ±(0.5 hPa + 4% of
exhalation or mbar actual pressure)
Avg. P End of breath cmH2O or hPa 0 to 99 0.1 ±(0.5 hPa + 4% of
or mbar actual pressure)
Other parameters
Parameter Updates Units Range Resolution Accuracy
RSBI End of breath 1/min-L 0 to 999 1 N/A
% Spont. Trig. Start of % 0 to 100 1 N/A
inhalation
% Spont. Cyc. End of % 0 to 100 1 N/A
inhalation
Measured Battery Time Once per HH:MM N/A 1 min N/A
minute
Measured Battery Charge Once per % 0 to 100 1 N/A
Level minute
FiO2* End of % 18 to 100 1 ±(2.5% + 2.5% of
inhalation actual oxygen
concentration)**
SpO2 Once per % SpO2 0 to 100 1 Refer to Nonin
second Xpod 3012
technical
specifications at
www.nonin.com
Pulse rate Once per 1/min 18 to 321 1 Refer to Nonin
second Xpod 3012
technical
specifications at
www.nonin.com.
* FiO2 monitoring automatically compensates for atmospheric pressure variation.
** The measured FiO2 sampled at 25 Hz has a response time of <12s to 90% of the final value. The long term output drift is <1% volume oxygen
per month.
144
Accuracy of controls
Notes:
• All flows and volumes are measured at BTPS conditions.
• To achieve specified accuracies, a successful Learn Circuit must be performed prior to measurement testing.
Specified accuracies exclude mask compliance.
• Pressure may be displayed in cmH2O, hPa or mbar. The conversion factor of the pressure units is 1. This means
the displayed pressure will have the same values regardless of the unit selected.
Accuracy of controls
The following table shows the setting range and accuracy for the adjustable parameters.
Parameter Range Resolution Accuracy
Delivered Pressure 2 to 50 hPa ±(0.5 hPa + 5% of target)
IPAP 4 to 50 hPa 0.2 hPa Refer to delivered pressure
EPAP 2 to 25 hPa 0.2 hPa Refer to delivered pressure
CPAP 3 to 20 hPa 0.2 hPa Refer to delivered pressure
PEEP Off, 3 to 20 hPa 0.2 hPa Refer to delivered pressure
PS 2 to 50 hPa (Valved circuits) 0.2 hPa Refer to delivered pressure
0 to 50 hPa (iVAPS)
P control 2 to 50 hPa 0.2 hPa Refer to delivered pressure
Vt (Tidal volume) Adult: 100 to 2500 mL 10 mL Valved circuits:
Paed: 50 to 300 mL* 5 mL ±12 mL or 10%, whichever is greater
Safety Vt (Safety tidal Adult: 100 to 2500 mL 10 mL Vented circuits:
volume) Paed: 50 to 300 mL* 5 mL ±15 mL or 15%, whichever is greater
Valved circuits:
±12 mL or 10%, whichever is greater
Target Va 1 to 30 L/min 0.1 L/min ±(0.1 +30% of target) L/min;
Resp.rate < 12 bpm
±(0.1 +15% of target) L/min;
Resp.rate ≥ 12 bpm
Resp. rate Adult: Off, 2 to 50 bpm 1 bpm ±2%
Paed: Off, 5 to 80 bpm
Target patient rate Adult: 8 to 30 bpm 1 bpm ±2%
Ti (Inspiration time) (A)CV, V-SIMV (mandatory Adult: 0.1 sec ±(20 ms + 5% of setting)
breaths): Paed: 0.05 sec
0.3 to 3.0 sec
P(A)CV, P-SIMV (mandatory
breaths):
0.2 to 5.0 sec
P(A)C:
0.3 to 4.0 sec
PIF (Peak Inspiratory Flow) (A)CV, V-SIMV (mandatory 1 L/min N/A
breaths):
Adult: 10 to 120 L/min
Paed: 5 to 60 L/min
Cycle (expiratory trigger) 5 to 90%, Auto 5% N/A
English 145
Accuracy of controls
146
Accuracy of controls
WARNING
ResMed does not recommend 500 mL as the upper limit for the pediatric tidal volume use;
however, clinicians may choose this upper limit based on their clinical determination.
Notes:
• All flows and volumes are measured at BTPS conditions.
• To achieve specified accuracies, a successful Learn Circuit must be performed prior to measurement testing.
• Delivered pressure is regulated at the patient port.
English 147
Measurement System and Accuracy
In accordance with ISO 80601-2-72 2015 the tolerance of monitoring parameters inclusive of
measurement uncertainty is:
Time parameters ± (30ms + 5%)
Flow parameters ± 12%
Volume parameters ± 12 ml or 12% whichever is greater
Pressure parameters ± (0.5 hPa + 5%)
Ventilator performance comparisons should be made on the basis of wholly including or excluding
measurement uncertainty.
Functional variants
Functional variations between Astral 100 and Astral 150.
Feature Astral 100 Astral 150
Single limb with expiratory valve Y Y
Single limb intentional leak Y Y
Circuit
Mouthpiece (tube only) Y Y
Double limb N Y
Preset programs 2 4
Leak therapy modes Y Y
Valve therapy modes Y Y
Apnoea ventilation Y Y
Therapy
Manual breath button N Y
Sigh (pre-programmed) N Y
SpO2 monitoring Y Y
FiO2 monitoring Optional extra Y*
O2 Oxygen inlet Low flow Low flow
*Not applicable on all device variants.
148
Guidance and Manufacturer’s Declaration Electromagnetic Emissions & Immunity
WARNING
• The device should not be used adjacent to or stacked with other equipment. If adjacent or
stacked use is necessary, the device should be observed to verify normal operation in the
configuration in which it will be used.
• The use of accessories other than those specified for the device is not recommended. They
may result in increased emissions or decreased immunity of the device.
• Additional equipment connected to medical electrical equipment must comply with the
respective IEC or ISO standards (eg, IEC 60950 for data processing equipment). Furthermore all
configurations shall comply with the requirements for medical electrical systems (see
IEC 60601-1-1 or clause 16 of the 3Ed. of IEC 60601-1, respectively). Anybody connecting
additional equipment to medical electrical equipment configures a medical system and is
therefore responsible that the system complies with the requirements for medical electrical
systems. Attention is drawn to the fact that local laws take priority over the above mentioned
requirements. If in doubt, consult your local representative or the technical service department.
• Portable RF communications equipment (including peripherals such as antenna cables and
external antennas) should be used no closer than 30 cm to any part of the device, including
cables specified by the manufacturer. Otherwise, degradation of the performance of this
equipment could result.
English 149
Guidance and Manufacturer’s Declaration Electromagnetic Emissions & Immunity
150
Guidance and Manufacturer’s Declaration Electromagnetic Emissions & Immunity
Notes:
• Ut is the AC mains voltage prior to application of the test level.
• At 80 MHz and 800 MHz, the higher frequency range applies.
• These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
• ISM bands: 6.765 to 6.795MHz, 13.533 to 13.567MHz, 26.957 to 27.283MHz and 40.66 to 40.70Mz
English 151
Guidance and Manufacturer’s Declaration Electromagnetic Emissions & Immunity
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in metres (m) can be determined using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer.
Notes:
• At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
• These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
152
Symbols
Symbols
The following symbols may appear on your product or packaging.
Power plug
Temperature Limitation
Oximeter connector
Keep upright
Ventilation indicator
Keep dry Alternating current
European Authorised Representative Connector for control line of external expiratory valve
CE Labelling in accordance with EC directive 93/42/EEC Connector for the breathing pressure measuring line
Environmental information
This device should be disposed of separately, not as unsorted municipal waste. To dispose of your
device, you should use appropriate collection, reuse and recycling systems available in your region. The
use of these collection, reuse and recycling systems is designed to reduce pressure on natural resources
and prevent hazardous substances from damaging the environment.
If you need information on these disposal systems, please contact your local waste administration. The
crossed-bin symbol invites you to use these disposal systems. If you require information on collection
and disposal of your ResMed device please contact your ResMed office, local distributor or go to
ResMed.com/environment.
English 153
Standards compliance
Standards compliance
The Astral meets the following standards:
• IEC 60601-1 Medical Electrical Equipment - General Requirements for Basic Safety and Essential
Performance
• IEC 60601-1-2 Medical electrical equipment - Part 1-2: General requirements for basic safety and
essential performance - Collateral standard: Electromagnetic compatibility -Requirements and tests
• IEC 60601-1-8 General requirements, test and guidance for alarm systems in medical electrical
equipment and medical electrical systems
• IEC 60601-1-11 Medical electrical equipment - Part 1-11: General requirements for basic safety and
essential performance - Collateral standard: Requirements for medical electrical equipment and
medical electrical systems used in the home healthcare environment
• ISO 80601-2-72 Medical electrical equipment - Part 2-72: Particular requirements for basic safety and
essential performance of home healthcare environment ventilators for ventilator-dependent patients
154
Troubleshooting
Troubleshooting
If there is a problem, try the following suggestions. If the problem cannot be solved, contact your care
provider or ResMed.
Alarm troubleshooting
The most common reason for an alarm to sound is because the system has not been properly
assembled or a Learn Circuit has not been correctly performed for each program.
Notes:
• The alarm actions listed below are based on having the appropriate alarm settings for the patient's therapy.
When an adjustable alarm is activated, re-confirm the alarm settings.
• The alarm log and alarm settings are maintained when the device is powered down and in the event of a power
loss.
• If an alarm activates repeatedly, discontinue use, switch to a backup ventilator and return the device for
servicing.
If the alarm log reaches its storage capacity, the oldest data will be discarded to allow new entries to be
written to the log.
English 155
Troubleshooting
156
Troubleshooting
English 157
Troubleshooting
158
Troubleshooting
English 159
Troubleshooting
160
Troubleshooting
If problem persists, this circuit may not be compatible with the Astral device. Consider
selecting another circuit configuration.
205 The measured circuit resistance exceeds safe operating limits for this device.
1. Check the circuit and attached accessories for blockages.
2. Ensure that there are no sharp bends or kinks in the circuit and the patient end is not
blocked.
3. If using humidification, ensure that the humidifier tub is not overfilled.
4. Follow the on-screen instructions carefully:
• circuit should not be blocked during step 2
• circuit should be completely blocked during step 3.
If problem persists, this circuit may not be compatible with the Astral device. Consider
selecting another circuit configuration.
206 Hardware fault detected. Contact an authorised Service Centre.
303 Unable to calibrate oxygen sensor.
1. Check that supplemental oxygen is not added during the Learn Circuit.
2. Repeat Learn Circuit. If the problem persists, replace the oxygen sensor as described
in Replacing the Oxygen sensor.
404, 405, 406 Hardware fault detected. Contact an authorised Service Centre.
409 Learn Circuit was unable to complete due to excessive leak from the circuit.
1. Check that the circuit is completely blocked during the third step of the Learn Circuit.
2. Check that the circuit is assembled correctly and there are no leaks in the circuit.
3. Check that the circuit adapter is firmly inserted.
4. This circuit may not be compatible with the Astral device. Try another circuit.
Repeat Learn Circuit. If the problem persists, contact an authorised Service Centre.
415 Hardware fault detected. Contact an authorised Service Centre.
420 The measured circuit compliance exceeds safe operating limits for this device.
Check that the circuit is assembled correctly and completely blocked during the third step
of the Learn Circuit.
If problem persists, this circuit may not be compatible with the Astral device. Consider
selecting another circuit configuration.
426 Hardware fault detected. Contact an authorised Service Centre.
English 161
Troubleshooting
If problem persists, this circuit may not be compatible with the Astral device. Consider
selecting another circuit configuration.
505 The measured circuit resistance exceeds safe operating limits for this device.
1. Check the circuit and attached accessories for blockages.
2. Ensure that there are no sharp bends or kinks in the circuit and the patient end is not
blocked.
3. If using humidification, ensure that the humidifier tub is not overfilled.
4. Follow the on-screen instructions carefully:
• circuit should not be blocked during step 2
• circuit should be completely blocked during step 3.
If problem persists, this circuit may not be compatible with the Astral device. Consider
selecting another circuit configuration.
506, 512 Hardware fault detected. Contact an authorised Service Centre.
600 Unable to calibrate Expiratory Flow Sensor.
1. Check the circuit adapter is clean, dry and firmly inserted.
• If the adapter is wet, then removing the adapter and vigorously shaking to clear
water can be effective. Re-insert adapter firmly and repeat Learn Circuit.
• If the adapter is not clean, then it will need to be replaced.
2. If using a small diameter paediatric breathing circuit, consider using an anti-bacterial
filter or a 22-mm adapter on the expiratory adapter port.
3. Repeat Learn Circuit and ensure that the circuit is not moved until completion of the
test.
If problem persists, then the expiratory flow sensor may need to be replaced.
162
Troubleshooting
General troubleshooting
Issue Action
Condensation forming in circuit Condensation may form due to high humidity settings and low ambient temperatures.
Adjust humidifier settings in accordance with manufacturer's instructions.
Touch screen damaged or If you are unable to power off the Astral device normally, use the following forced
non-responsive shutdown procedure:
1. Disconnect any external power source (eg, AC mains or external battery).
2. Press and hold the green on/off button and the alarm mute/reset button for at least
10 seconds. After 10 seconds the alarm bar will flash yellow.
3. Release both buttons. Astral will then power off.
4. The Astral device can be powered back on by pressing the on/off button and used as
intended.
Unable to save data from Astral to 1. Remove and reinsert the USB stick.
USB or USB is not detected by 2. Use a new USB stick.
device. 3. Remove the AC or external DC power supply, the restart the Astral by switching it off
then on.
4. Reformat your USB stick. Note that any data currently saved on the USB will be lost.
Learn Circuit failed If the Learn Circuit fails and a warning message appears on the top of the Learn Circuit
results page, try the following:
1. Check the circuit for Leak.
2. Check the module, the blue membrane and sensor are pressed all the way in and sit
flush with the enclosure.
3. Select the paediatric patient type. This will allow higher resistance circuits to pass.
(10mm and 15mm corrugated circuits will not pass Learn Circuit with adult patient
type selected).
4. Hold the circuit straight to reduce resistance.
5. Ensure that the ventilation and alarms are effective before proceeding further.
Note: It is acceptable to use a circuit that gives a caution message as the Astral device
will compensate for circuit resistance and compliance.
Flow Sensor failed If the Flow Sensor fails and a message appears on the bottom of the Learn Circuit results
(Astral 150 only) page, try the following:
1. Check the circuit for Leak.
2. Check the expiratory valve, the blue membrane and sensor are pressed all the way in
and sit flush with the enclosure.
3. If required, replace the flow sensor by following the instructions in 'Replacing
components" in the Cleaning and maintenance section.
English 163
Recommendations for suctioning
and/or PEEP alarms may be triggered. Press to pre-silence alarms for two minutes.
CAUTION
Following open or closed suctioning, restore the patient circuit and check ventilation is correctly
restored.
To assess the patient condition and the effectiveness of the secretions removal, peak inspiratory flow
(during pressure-controlled ventilation), peak inspiratory pressure (during volume-controlled ventilation) or
tidal volume during pressure-controlled ventilation, as well as SpO2 can be monitored.
Astral places no restrictions on which ventilation mode is used while suctioning. Astral may respond
differently depending on the mode and therapy settings. It is recommended that the 'expected response'
is clearly documented in the patient care plan.
164
Limited warranty
Limited warranty
ResMed Ltd (hereafter 'ResMed') warrants that your ResMed product shall be free from defects in
material and workmanship from the date of purchase for the period specified below.
Product Warranty period
• Mask systems (including mask frame, cushion, headgear and tubing)—excluding 90 days
single-use devices
• Accessories—excluding single-use devices
• Flex-type finger pulse sensors
• Humidifier water tubs
• Batteries for use in ResMed internal and external battery systems 6 months
• Clip-type finger pulse sensors 1 year
• CPAP and bilevel device data modules
• Oximeters and CPAP and bilevel device oximeter adapters
• Humidifiers and humidifier cleanable water tubs
• Titration control devices
• CPAP, bilevel and ventilation devices (including external power supply units) 2 years
• Battery accessories
• Portable diagnostic/screening devices
English 165
Appendix A: Definitions
Appendix A: Definitions
Ventilation settings definitions
The available settings will vary with the selection of the ventilation mode. Each mode details the settings
available.
Setting Definition
Apnoea Definition Apnoea Definition sets the type of breath which must be delayed for an apnoea to be
detected.
Apnoea Interval Apnoea Interval (T apnoea) sets the period without breath or spontaneous breath
(T apnoea) required for an apnoea to be detected.
Apnoea Response Apnoea Response sets the behaviour of the ventilator when an apnoea is detected.
Circuit Type Circuit Type sets whether a Double limb circuit, Single limb circuit with expiratory valve
or Single limb circuit with intentional leak is in use.
CPAP Continuous Positive Airway Pressure (CPAP) sets the pressure maintained throughout a
spontaneous breath.
Cycle Cycle (also known as Expiratory Trigger) sets the threshold where start of expiration
within a breath is detected.
EPAP Expiratory Positive Airway Pressure (EPAP) sets the pressure to be delivered to the
patient during expiration.
Flow shape Sets the target flow waveform for the delivery of mandatory controlled volume breaths.
Inspiratory duration option Inspiratory duration option (Insp Duration Option) sets whether Inspiration Time (Ti) or
(Insp Duration Option) Peak Inspiratory Flow (PIF) is used to configure volume controlled breaths.
Interface type Invasive, mask, or mouthpiece
Interval Sigh interval sets the period between sigh breaths.
IPAP Inspiratory Positive Airway Pressure (IPAP) sets the pressure to be delivered to the
patient during inspiration.
Magnitude Magnitude sets the size of the manual or sigh breath delivered relative to the size of the
normal ventilation breath. Separate magnitude settings are available for configuration of
manual or sigh breaths.
Manual Breath Manual Breath sets whether a manual breath is available for delivery.
Mask Type Mask Type sets the type of mask or in-line vent in use when the circuit type is single
with leak.
Max EPAP Maximum Expiratory Positive Airway Pressure (Max EPAP) sets the maximum pressure to
be delivered to the patient during expiration to maintain upper airway patency.
Max PS Maximum Pressure Support (Max PS) sets the maximum pressure support above EPAP
allowed to achieve the Target Va.
Min EPAP Minimum Expiratory Positive Pressure (Min EPAP) sets the minimum pressure allowed to
be delivered to the patient during expiration to maintain upper airway patency.
The Min EPAP should be set to treat any lower airway condition.
Min PS Minimum Pressure Support (Min PS) sets the minimum pressure support above EPAP
allowed to achieve the Target Va (iVAPS).
P control Pressure control (P control) sets the pressure support above PEEP to be delivered during
inspiration for pressure assisted breaths.
166
Appendix A: Definitions
Setting Definition
P control max Maximum allowed pressure control (P control max) sets the maximum pressure control
above PEEP allowed to achieve the target safety volume.
Patient type Select from Adult or Paediatric. This setting configures the default values and ranges
available for ventilation settings and determines circuit resistance acceptance criteria
applied in the Learn Circuit.
PEEP Positive End Expiratory Pressure (PEEP) sets the pressure maintained during exhalation.
PIF Peak Inspiratory Flow (PIF) sets the maximum delivered flow for volume controlled
breaths.
PS Sets the pressure support above PEEP to be delivered during inspiration for pressure
supported breaths (spontaneous breaths).
PS Max Maximum allowed Pressure Support (PS Max) sets the maximum pressure support above
PEEP allowed to achieve the target safety tidal volume.
Pt Height Patient Height (Pt Height) is used to estimate the patient's anatomical deadspace and
Ideal Body Weight (IBW).
Resp. rate Respiratory rate (Resp. rate) sets the breaths per minute (bpm) to be delivered by the
ventilator to the patient. The measured respiratory rate may be greater due to patient
triggered breaths.
Rise Time Rise time sets the time taken for the ventilator to reach inspiratory pressure for pressure
controlled breaths.
Safety Vt Safety tidal volume sets the target minimum tidal volume (Vt) for each ventilator
delivered breath.
Sigh Alert Sigh alert sets whether the ventilator gives a single beep just prior to delivery of a sigh
breath.
Sigh Breath Sigh Breath sets whether a magnified breath (a sigh breath) will be delivered at the sigh
interval.
Target Pt Rate Target Patient Rate (Target Pt Rate) sets the upper boundary for the iVAPS intelligent
Backup Rate (iBR).
Target Va Target Alveolar Minute Ventilation (Target Va) sets the servo-ventilation target for
iVAPS.
Ti Inspiration time (Ti) sets the duration of the inspiratory phase of a breath.
Ti Max Maximum inspiratory time (Ti Max) sets the maximum duration of the inspiratory phase
of a breath.
Ti Min Minimum Inspiratory Time (Ti Min) set the minimum duration of the inspiratory phase of
a breath.
Trigger Sets the trigger threshold above which the ventilator triggers a new breath.
The trigger is blocked for the first 300 ms following the start of exhalation.
Trigger type Trigger type sets whether a pressure based trigger threshold or flow based trigger
threshold is used when a Double circuit is selected.
Vt The Tidal Volume (Vt) sets the volume of gas, measured in mL, to be delivered to the
patient in a mandatory controlled volume breath.
English 167
Appendix A: Definitions
Parameter Definition
FiO2 Average of percentage of Oxygen delivered to circuit.
I:E I:E is the ratio of the inspiratory period to the expiratory period.
The measured I:E ratio is displayed as a monitored parameter during ventilation.
The expected I:E ratio is calculated and displayed on the settings screens if the Resp.
rate setting is not set to Off.
Leak Leak is the average unintentional leak. It is reported as a percentage for Double limb
circuits and as a flow for Single limb circuits with intentional leak.
The measured Leak is displayed as a monitored parameter during ventilation.
MV Minute Ventilation (MV) is the product of the Target Patient Rate (Target Pt Rate) and
expired tidal volume averaged over the last eight breaths.
The MV is displayed as a calculated parameter during iVAPS configuration.
MVe Expiratory Minute Volume (MVe) is the product of the respiratory rate and expired tidal
volume averaged over the last eight breaths.
The measured MVe is displayed as a monitored parameter during ventilation.
MVi Inspiratory Minute Volume (MVi) is the product of the respiratory rate and inspired tidal
volume averaged over the last eight breaths.
The measured MVi is displayed as a monitored parameter during ventilation.
Pressure Pressure is the current airway pressure of the patient as measured at the patient port.
The measured Pressure is displayed as a monitored parameter during ventilation.
PEEP End expiratory pressure (PEEP) is the airway pressure measured 50 ms prior to the end of
the last expiration.
The measured PEEP is displayed as a monitored parameter during ventilation.
Pmean Mean airway pressure of the patient over the last breath.
% Spont cycle % Spont cycle is the percentage of breaths that are spontaneously cycled over the past
20 breaths.
% Spont trig % Spont trig is the percentage of breaths that are spontaneously triggered over the last
20 breaths.
The measured %Spont Trig is displayed as a monitored parameter during ventilation.
PIF Peak Inspiratory Flow (PIF) is the maximum flow reached during the last inspiration.
The measured PIF is displayed as a monitored parameter during ventilation.
The expected PIF is calculated and displayed for volume controlled breaths on the
settings screens when the Inspiratory Phase Duration Option is set to Ti.
PIP Peak Inspiratory Pressure (PIP) is the maximum airway pressure reached during the last
inspiration.
The measured PIP is displayed as a monitored parameter during ventilation.
Pulse rate The measured Pulse rate (pulse) is displayed as a monitored parameter when a pulse
oximeter is used.
168
Appendix A: Definitions
Parameter Definition
Resp. rate Respiratory rate (Resp. rate) is the number of breaths per minute averaged over the last
eight breaths.
The measured Resp. rate is displayed as a monitored parameter during ventilation.
RSBI Rapid Shallow Breathing Index (RSBI) is calculated by dividing the breath rate by Tidal
Volume.
The measured RSBI is displayed as a monitored parameter during ventilation.
SpO2 The measured functional Oxygen Saturation (SpO2) is displayed as a monitored
parameter when a pulse oximeter is used.
Te Expiratory time Te is the period in seconds of the last expiratory phase.
Ti Inspiratory time Ti is the period in seconds of the last inspiratory phase.
The measured Ti is displayed as a monitored parameter during ventilation.
The expected Ti is calculated and displayed for volume controlled breaths on the settings
screens when the Inspiratory Phase Duration Option is set to PIF.
Va Alveolar Minute Ventilation (Va) is calculated by (Tidal Volume - Deadspace) x Resp.
Rate.
The measured Va is shown as a monitored parameter during ventilation.
Vte Expiratory Tidal Volume (Vte) is the volume expired during the last breath.
The measured Vte is displayed as a monitored parameter during ventilation.
Vti Inspiratory Tidal Volume (VTi) is the volume inspired during the last breath.
The measured VTi is displayed as a monitored parameter during ventilation.
Average Vt Average Tidal Volume (Average Vt) is the average volume expired during the last five
minutes of ventilation.
The Average Vt is displayed as a calculation parameter during iVAPS configuration.
Average Vt/kg Average Tidal Volume per kg (Average Vt/kg) is the Average Vt divided by Ideal Body
Weight (IBW).
The Average Vt is displayed as a calculation parameter during iVAPS configuration.
English 169
Appendix B: Ventilation parameters
V-SIMV
P-SIMV
P(A)CV
iVAPS
CPAP
CPAP
P(A)C
ACV
(S)T
Respiratory rate PS
(A)CV, P(A)CV, P(A)C
• Adult: Off, 2 to 50 [15]
• Paed: Off, 5 to 80 [15]
Mandatory breaths: V-SIMV,
P-SIMV
• Adult: 2 to 50 [15],
• Paed: 5 to 80 [15]
PS, (S)T
• Adult: Off, 2 to 50 [15]
• Paed: Off, 5 to 80 [15]
EPAP 2 to 25 [5]
(cmH2O)
170
Appendix B: Ventilation parameters
V-SIMV
P-SIMV
P(A)CV
iVAPS
CPAP
CPAP
P(A)C
ACV
(S)T
PS
PS PS
(cmH2O) • Adult: 2 to 50 [7]
• Paed: 2 to 50 [7]
Spontaneous breaths: V-SIMV,
P-SIMV
• Adult: 2 to 50 [7]
• Paed: 2 to 50 [7]
Min PS 0 to 50 [2]
(cmH2O)
English 171
Appendix B: Ventilation parameters
V-SIMV
P-SIMV
P(A)CV
iVAPS
CPAP
CPAP
P(A)C
ACV
(S)T
PS
Cycle (%) P-SIMV, V-SIMV (spontaneous
breaths), PS
5 to 90 [25%] (leak)
5 to 90 [Auto] (valve)
Cycle is fixed at 15% for valve,
10% for leak.
Trigger type Flow/Pressure
(double limb circuit only)
172
Appendix B: Ventilation parameters
V-SIMV
P-SIMV
P(A)CV
iVAPS
CPAP
CPAP
P(A)C
ACV
(S)T
PS
Ti Min PS
(sec)
0.2 to 4.0 [0.2]
(S)T, iVAPS
0.1 to 4.0 [0.2]
Ti Min is fixed at:
• 0.2 for P-SIMV, V-SIMV,
and CPAP (valved)
• 0.1 for CPAP (leak)
Ti Max (Maximum Adult: 0.3 to 4.0 [1.5]
Inspiration Time)
Paed: 0.3 to 4.0 [0.8]
(sec)
Ti Max is fixed at:
• The lesser of
( (2/3) x (60/f) ) or 4 for
P-SIMV and V-SIMV
• 4 for CPAP (leak)
• 3 for CPAP (valved) Adult
• 1.5 for CPAP (valved)
Paediatric
Pt Height Adult:
cm: 110 to 250 [175]
inches: 44 to 100 [70]
Target Pt 8 to 30 [15]
Rate
Target Va 1 to 30 [5.2]
(L/min)
*The International ventilator standard indicates that Paediatric patient type is intended to be used for a patient receiving less
than 300 mL, however Astral permits adjustment of ‘Vt’ setting parameter up to 500 mL for cases where ‘Vt’ is set such that it
compensates for leak in the breathing circuit.
WARNING
ResMed does not recommend 500 mL as the upper limit for the pediatric tidal volume use;
however, clinicians may choose this upper limit based on their clinical determination.
English 173
Appendix B: Ventilation parameters
V-SIMV
P-SIMV
P(A)CV
iVAPS
CPAP
CPAP
P(A)C
ACV
(S)T
PS
I:E Ratio Displayed if Respiratory Rate
not set to Off
Supplementary Features
The following table indicates the supplementary features applicable to each ventilation mode. The
adjustable parameters, the available setting range and default are displayed.
iVAPS
CPAP
CPAP
P(A)C
ACV
(S)T
PS
174
Appendix B: Ventilation parameters
V-SIMV
P-SIMV
P(A)CV
iVAPS
CPAP
CPAP
P(A)C
ACV
(S)T
PS
Sigh Breath Off / On [Off]
English 175
Appendix B: Ventilation parameters
V-SIMV
P-SIMV
P(A)CV
iVAPS
CPAP
CPAP
P(A)C
ACV
(S)T
PS
Apnoea Vt If Apnoea Response is (A)CV +
(mL) Alarm:
Adult: 100 to 2500 [500]
Paed: 50 to 500 [100]
Apnoea Flow Shape Constant
(%)
*Adult Tapnoea can be extended to 15min when the mouthpiece interface is selected.
176
Appendix B: Ventilation parameters
V-SIMV
P-SIMV
P(A)CV
iVAPS
CPAP
CPAP
P(A)C
ACV
(S)T
PS
Manual Breath Ti
(sec)
Manual Breath Vt
(mL)
Manual Breath PS
(cmH2O)
Sigh Ti
(sec)
Sigh PIF
(L/min)
Sigh Vt
(mL)
Sigh P Control
Apnoea I:E
English 177
Appendix B: Ventilation parameters
V-SIMV
P-SIMV
P(A)CV
iVAPS
CPAP
CPAP
P(A)C
ACV
(S)T
PS
Apnoea PIF Display
(L/min)
178
Appendix C: Alarm parameters
V-SIMV
P-SIMV
P(A)CV
iVAPS
CPAP
CPAP
P(A)C
ACV
(S)T
PS
Vti - Low Adult: Off, 50 to 2990 [100]
(Tidal Volume)
Paed: Off, 10 to 995 [25]
(mL)
English 179
Appendix C: Alarm parameters
V-SIMV
P-SIMV
P(A)CV
iVAPS
CPAP
CPAP
P(A)C
ACV
(S)T
PS
Pressure - Low Off, 2 to 79 [5]
(Peak Inspiratory Pressure)
(cmH2O)
180
Appendix C: Alarm parameters
V-SIMV
P-SIMV
P(A)CV
iVAPS
CPAP
CPAP
P(A)C
ACV
(S)T
PS
FiO2 - Low (%) Off, 18 to 99 [Off]
English 181
ResMed Ltd
1 Elizabeth Macarthur Drive, Bella Vista NSW 2153 Australia
See ResMed.com for other ResMed locations worldwide. Astral, AirView and ResScan are trademarks and/or registered trademarks of the ResMed family of companies.
For patent and other intellectual property information, see ResMed.com/ip. Actichlor is a trademark of Ecolab US. Inc. Aerogen and Aeroneb are registered trademarks of Aerogen,
Inc. CaviCide is a trademark of Metrex. Mikrozid is a trademark of Schülke & Mayr. NONIN is a trademark of Nonin Medical, Inc. Velcro is a trademark of Velcro Industries B.V.
© 2018 ResMed Ltd. 278579/1 2018-05
ResMed.com