IFU Bahasa Inggris DFI DUS R-300 Urine Chemistry Analyzer
IFU Bahasa Inggris DFI DUS R-300 Urine Chemistry Analyzer
DU(V3). 2009.11
Contents
1. Introduction
2. System Description
2.1 Principles of Measurement
2.2 Components and Functions
2.3 Technical Information
3. Instrument Installation
3.1 Contents of package
3.2 Installation
4. Main Process
5. How to operate instrument
5.1 Calibration
5.2 General Mode
5.3 One by One Mode
5.4 Quick Mode
5.5 How to input ID
6. System Setup
6.1 System Clock Set
6.2 System Config
6.3 Registration
6.4 Data Base
6.5 DB Block Control
7. How to Clean the Strip Loader(Plate) after Use
8. Troubleshooting & Maintenance
2
1. Introduction
Urine test strips simplify laboratory diagnosis through ease of use, high sensitivity and
specificity. These benefits allow you to identify pathological changes in the urine quick
and reliably. The addition of the Urine Analyzer to the urinalysis process allows for
standardization and efficiency of urine test by eliminating potential sources of error
associate with the visual reading of test strips, such as improper lighting at the workplace,
or different color discrimination by the user or different timing when the values are read.
The urine test strips that shall be used with the instrument are multi parameter strips for
the determination of specific gravity, pH, leukocytes, nitrite, protein, glucose, ketones,
urobilinogen, bilirubin and blood in urine.
2. System Description
The Urine Analyzer is a urine strip reader and analyzer device for laboratories. The reader
is semiautomatic; the forwarding, reading and evaluating are automated. The only thing
the operator has to do is to dip the strip into the sample and place it on the strip loading
plate.
The Urine Analyzer is a reflectance photometer. The strip is illuminated by white light,
and the reflected light from the strip is detected by the Sensor. The RGB signals are
digitized, and this digitized image is interpreted by the processor. The intelligent image
analyzer SW locates the strip and the pads, and based on these color data the parameter
values are determined. The results including the date and time of the measurement,
sequence number and the ID are stored and printed out by the internal printer.
3
2.2 Components and Functions
3
8
2 7
5
4 6
Component Function
4
2.3 Technical Information
3. Instrument Installation
Optional accessories
Barcode Reader (Scanner)
External Keyboard
5
3.2 Installation
1) Carefully unpack the Urine Analyzer and place it on a firm and level surface. Do not
place the analyzer under direct sunlight or any direct light source.
2) Insert the plate: Grasp the plate end and insert it along the visible sliding rail until it
gets between motor gear and bearings sufficiently.
4) Unpack the adapter and connect it to the socket in the backside of Urine Analyzer. Plug
the adapter into the wall socket. Use only the provided adapter. Use of other type adapter
can cause defective measurement and general failure.
5) In case of viewing the test results in your computer connect the serial cable to a serial
port COM1 of Urine Analyzer and a port for COM1 in the back of computer. Open the
PC management program in the connected computer. Then you can see the test results
automatically in your computer after testing
6) To use a Barcode reader/scanner to identify samples etc., plug it into the Serial Port
(COM2) on the backside of the Urine Analyzer.
7) To use keyboard to identify samples etc., plug it into the socket on the backside of the
Urine Analyzer and input the patient‟s ID.
6
4. Main Process
Step 1. Put the plate into the middle of the instrument and connect AC adapter to the port
on the backside of instrument and power on.
Step 2. Automatically the Instrument executes System Check and then the plate moves to
the middle of the reader.
Step 3. Calibration (Refer to 5.1 Calibration)
Step 4. Measurement MODE
On condition the plate is stopped, place each strip dipped in the urine one by one.
At this time, there are three selective modes for detecting the test strips.
General Mode: Calculate placing time of „whole strips‟. [press start(▶) key „after‟
putting 1st test strip on the groove located in the nearest to center of instrument, place next
test strip (max. 10 test strips) and then press End (ENT) key after placing the last test
strip] (Refer to 5.2 General Mode)
One by One Mode: Recognize placing time of „each strip‟. [press start(▶) key „after
putting 1st test strip on plate like general mode and press another next(▶) key whenever
you place each additional strip(max. 10 test strips) and then press End (ENT) key after
placing the last test strip] (Refer to 5.3 One by One Mode)
Quick Mode: Place all test strips and then press End (ENT) key. (Refer to 5.4 Quick
Mode)
Step 5. Waiting
Standby Urine Analyzer until the color incubation of 1 st strip is completed.
Step 6. Reading
After the 1st strip finish color incubation, the instrument starts to measure.
Step 7. Remove Strips
As soon as Urine Analyzer finishes the measurement, it will be changed to waiting mode.
At this time, operator can check the result by watching display or prints out with the
naked eye before he/she throw away these already checked out strips.
7
5. How to operate instrument
Turn ON the Power switch in the rear of instrument. Then the start page will be shown as
[Fig.5] and the instrument checks the system automatically.
General Mode
Quick Mode
For more detailed working procedures under each mode, please refer to 5.2, 5.3 and
5.4.
8
5.1 Calibration
The Urine Analyzer is calibrated before shipment by the manufacturer. The Calibration of
instrument should be done prior to first use and then the same process from the 2 nd time
can be recommended to do in every 4 weeks with Calibration Strip in the package.
Calibration Strip is used for checking aging processes of the optical system and variation
of other internal conditions of instrument. When strong variation is detected which may
be caused, for instance, by contamination of the reference pad or low light intensity of a
defective light source, an error message will be printed.
The Calibration Strips are plastic white standard strips with defined and constant
reflectance characteristics. The Calibration Strip should remain in the vial until use and
should be used only once. Do not touch in the middle areas of the strip. The plate strip
tray should be clean and dry before a calibration is performed.
Step2. Move to Calibration by using ▲▼ (up, down) keys and select YES by using
◄► (left, right) keys.
Step3. Press ESC key twice
Step4. Put the „Calibration Strip(CAL)‟ into the 1st position of Strip Loading Plate and
then press < ENT >.
The calibration checking procedure can be stopped by pressing <ESC>.
Step5. After calibration finished, the result of calibration print out.
If the calibrations values are between 333+/-10(323 ~ 343), it is OK.
9
◄ 2)
ESC
ESC ENT
ESC
Error Success
10
Press ESC key to cancel placing strip and evaluation.
3) Waiting for incubation of 1st strip
2) After the 1st strip dipped in the urine is placed on the plate, press Start key (▶).
Button Place another key(▶) whenever each of next strip is placed on the plate one by
one from the center of instrument after dipping in the collected urine and placing on the
plate. (The Progress Bar display for100 sec incubating time) By pressing Place another
key(▶) per measurement of each strip, the exact measuring/starting time is recognized.
After the last reagent strip of a batch is placed, press End (ENT) button.
11
Press ESC key to cancel placing strip and evaluation.
3) Waiting for incubation of 1st strip
4) After incubation time of the 1st test strip, it start to read test result of the strips on the
plate one by one.
5) Test result will be printed out, shown on the LCD and automatically saved in the
memory.
6) Press ESC key and you will return to One by One mode.
2) Test strips are temporarily incubated outside the instrument and inserted at the correct
time. After the last reagent strip of a batch is placed, press End (ENT) button.
3) Test result will be printed out, shown on the LCD and automatically saved in the
memory.
4) Press ESC key and you will return to Quick mode.
12
5.5 How to input ID
There are 2 ways to input ID into Urine Analyzer.
1st Method
The user can input patient/sample‟s ID using barcode Reader (scanner) while
measurement.
2rd Method (Multi ID input)
In System Control mode press ③key and the display [Fig.6.3] will be shown as below.
Use ▲▼ key to move to each item. Press ESC key and you will return to System Setup
mode.
ⓐ ⓑ
[Fig.6.3] Registration Mode
ID
1) Enter patient/sample‟s ID using Keyboard or Barcode Reader. Press ENT key or ►
(Right direction) key. Then enter next patient/sample‟s ID. If you want to change the ID
number, you can go to the ID using ◄► (Right direction, Left direction) key and enter
ID number again.
2) Part ⓐ is serial number of urine samples to be tested. As you register patient/sample‟s
ID numbers, the number in entry ⓐ increases in order.
3) Part ⓑ is data location number; this will be stored in data base memory after
measurement.
4) If patient/sample‟s IDs are as same as serial numbers, you may register only the first
patient/sample‟s ID. Then the others will be numbered in order automatically. In that case,
next ID number will be increased one by one from the first number. This number will
increase even when restarting the system after turning off the power.
5) If patient/sample‟s ID is not registered before a test, ID columns will be blank.
6) To stop printing patient/sample‟s ID before the test, make it blank in ID (just input
only”0” key, this will erase ID area)
13
6. System Setup
In Standby Mode of [Fig.5] press ESC key twice to display System Setup [Fig.6].
14
[Fig.6.2] System Configuration Mode
Beep Enable
If you select YES, beep will sound while operating instrument.
Use ◄► keys to select YES or NO.
Set Bps(PC)
This is to select speed rate while transferring data with each other between the instrument
and PC. Under general conditions 115.2 should be selected.
Disp Unit
Select the unit of test results: mg or mol.
Strip Type
This is to select the Strip type for test.
The Strip type should be matched with the Strip type being used.
Name 1 2 3 4 5 6 7 8 9 10 11
Urine 4(=DUS4) G P H O
Urine 5(=DUS5) B K G P H
Urine 6(=DUS6) P H O K B G
Urine 7(=DUS7) U P H O K B G
Urine 8(=DUS8) N U P H O K B G
Urine 9(=DUS9) N U P H O S K B G
Urine 10(=DUS10) L N U P H O S K B G
Urine 11(=DUS11) L N U P H O S K B G A
Print Enable
Select printer on or off by selecting YES or NO. If you select YES, test results will be
shown both on LCD and printing paper.
When you finish selecting all conditions, press ESC key.
If change from present conditions is not required, just return to System Setup mode. But
if any change is needful, the display will ask you to confirm saving the change.
If you decide to change anything (YES or NO) and press ENT key, you will go back to
System Setup mode.
6.3 Registration
In System Setup mode press 3) key and the display [Fig.6.3] will be shown as below.
Input data on each item and use ▲▼ key to move to next item. Press ESC key and you
will return to System Setup mode.
ⓐ ⓑ
[Fig.6.3] Registration Mode
Oper , Lot
Enter operator‟s ID and strip lot number by using numeric keys. If you don‟t need to
record the test operator‟s ID you may skip and leave it blank.
16
ID
Enter patient/sample‟s ID using numeric keys, external keyboard or barcode reader
(scanner). Press ENT key or ►(Right direction) key. Then enter next patient/sample‟s ID.
If you want change ID number, you can go to the ID using ◄► key and enter ID number
again.
(Refer to 5.5 How to input ID)
Color
Color changes in order of YELLOW, DK YELLOW, STRAW, AMBER, RED, ORANGE,
GREEN and OTHER, by pressing ◄► keys.
Turb
Turbidity changes in order of CLEAR, SL CLOUDY, CLOUDY, TURBID and OTHERS,
by pressing ◄► keys.
1) Select patient/sample‟s Color/Turb using ◄► keys. Press ENT key. And then enter
next patient/sample‟s Color/Turb.
2) Part ⓐ is serial number of urine samples to be tested. As you register patient/sample‟s
Color/Turb one by one, the number in entry ⓐ increases in order.
3) Part ⓑ is data location number; this will be stored in data base memory after
measurement.
4) If patient/sample‟s Color/Turb is not registered before a test, Color/turb columns will
be blank.
5) To stop printing patient/sample‟s Color/Turb before the test, make it blank in
Color/Turb.
17
6.4 Data Base
In System Setup mode press 4) key and the Searching by will be shown as [Fig.6.4].
The data saved in the memory of instrument can be searched by using ▲▼ keys.
Go End: Press ENT key and then latest result will be shown as [Fig.6.4.1].
Go Mem : Input SEQ Number by Keyboard and then the result will be shown
as [Fig.6.4.1].
Go ID: Input patient/sample‟s ID by Keyboard and then the result will be
shown as [Fig.6.4.1].
ⓐ ⓑ
- ⓐ shows the time of measurement.
- ⓑ shows strip type.
- ⓒ is data‟s number saved in data base. ⓒ
[Fig.6.4.1] Data Base
1) You can also go to the data by using ◀▶ keys.
2) You can print out the result on the display by pressing ENT key.
Data base can save up to 2000 test results in its memory. When memory is full, the data
can be transferred to PC by RS232c interface cable and sorted by its tested Date in PC.
Otherwise the oldest data of the instrument will be erased on the „first in first out‟ basis.
3) To erase all data in Data Base, type in “9999” and press “ENT” key at areaⓒ by using
keyboard . After clearing this number restarts at 0001.
Do this with care as stored data cannot be recovered.
18
6.5 DB Block Control
In System Setup mode press 5)key and the display will be shown as [Fig.6.5].
Print
To print out the results data.
Note: Printing Area is limited to max. 50 test results at one time.
19
7. How to Clean the Strip Loader Plate after Use
Caution: It is better for the strip loader plate to be cleaned everyday because it
can be stained by the tested strip and remaining urine smeared into the loader.
Step1. Turn the instrument power OFF in advance and then separate the plate from
the body of the instrument.
Step2. Completely clean all the grooves of the detached strip loader plate and clean the
contaminated area using a neutral detergent or a water-wet towel. Do not use volatile
material like thinner, gasoline, and benzene for cleaning the body.
Step 3. Polish the plate with a soft cloth and make sure to dry it completely
20
8. Troubleshooting & Maintenance
Symptoms of Cause/Symptom
Possible Cause or Related Symptoms
Error No.& Remedy
A-1 The instrument will 1. The connection part for power source may be loose. 1. Turn the instrument
not work even though 2. Defective system electronics power OFF. Check that
power is on the power cord is
firmly plugged into the
instrument and into a
live AC electrical
outlet then turn the
power ON
(AC adapter : 100-
240V/12VDC 3A)
2. Contact your
Customer Service
Office
21
1. Examine state of used strip and check out if any belong 1.2 : Careful handling
B-1 Always shows to a case described below; of the strips by
„false positive (+) test a) Order (sequence) of test parameter on printed results referring to instruction
result‟ differ from it of the label on strip's bottle. (Tested strip of use for strip and
model is not fit for this mode analyzer) operating manual for
B-2 Result by visual test b) Reacted color(s) of reagent pad on the strip is (are) the analyzer
with color chart shows much different from those of the color chart label on the
negative results strip bottle. 3.4: The Calibration
although instrument c) Detection was done with dry strip(before dipping in the should be done with
result is positive urine) or any other test strip instead of the reagent strip for new calibration strip.
this model analyzer
B-3 Other reference d) Tested without placing the dipped strip on the plate
tests shows Negative e) Strip position was turned upside-down or wrongly
results although positioned
instrument result is f) Reagent pad(s) were not fully wetted by the urine
positive g) Reagent strip was not placed on the tray 'immediately'
after dipping in the urine and removal of excessive urine
f) Shape of used strip was bent(curved) strip
2. Strip type is wrongly set up
(Location for strip set-up: System setup /2.System Config /
strip type)
3.There is a lot of dirty urine(contamination) on the
calibration strip
4. The Calibration Value not included 333(+/-10) (How to
see RGB calibration value can be referred to page 9 on
Operator Manual)
C-1 Does not print 1. Printer mode is set as "NO" 1. Reset “Yes”
(Location for changing printer mode:
System Setup /2.System Config /Print enable) 2. Replace „Printer Module‟
2. Test results can be displayed through LCD with a new one
monitor
C-2 Printer can't push out Can hear a sound that the motor inside of the Take out printer paper and
paper normally printer makes re-install it into the analyzer.
If you have the same error
C-3 Very small characters even after re-install of the
are printed on the paper paper correctly, Contact
and there is strange sound your Customer Service
Office
22
C-4 Printed characters are The printer paper is not properly installed. If you have the same error
not clear or not readable (Results should be printed on shinny side of even after re-install of the
the paper.) paper correctly, Contact
C-5 Characters are printed your Customer Service
partially Office
C-6 The strange characters The strange characters are also found on result Push „Power button & ESC
are printed on the paper through LCD monitor & ◀ key‟ together at a
time (reset of the machine)
and restart. If you have
the same error, Contact
your Customer Service
Office
D-1 Display a message The strip for test is wrongly positioned on the Misuse by the user
"Misplaced strip" strip plate
E-1 Strip Plate does not 1. Not used power supply of adapter 1. Misuse by the user
move into(out from) the ( 12V, 3A) 2. Replace „Plate‟ with a
machine 2. Plate malfunction problem new one.
E-2 Strip Plate always pull 3. Plate Sensor Problem 3,4 Contact your Customer
out 4. Motor Problem Service Office
E-3 A Strip Plate stops
unexpectedly during
operation
F-1 Key does not work 1. A function key is out of order and Other 1. Replace LCD Module
function keys (strip detection, printing etc.)
works normally 2. Contact your Customer
Service Office
2. Other function keys (strip detection,
23
printing etc.) also works incorrectly
G-1 LCD is black (shows 1. The error(s) is(are) found as soon as Power- 1. Replace LCD Module
nothing) or wrong On and Other functions (strip detection, with a new one
printing etc.) works normally
2. Contact your Customer
2.Other functions (strip detection, printing Service Office
etc.) also works incorrectly
H-1 Can not make 1. Check out settings are as followings; 1. It the set up condition is
"HexTrans" program Baud rate : 115200, Com Port: COM1 different from the left
update conditions, make it as
H-2 Can not connect to 2. Serial cable between analyzer and PC is same like left side
LIS incorrectly connected
H-3 LIS results include 2. Correctly connected.
strange letters 3. Serial port in the computer is USB Port But if it is not working,
H-4 When connect to replacement RS232C
PC, the items can not be Cable
shown correctly on the
results 3. Use „USB To RS232
Converter Cable‟ instead
of USB
Registration sheet
Please complete the following Registration Sheet, and fax back within two weeks to your
local distributor.
These data are recorded in the Service Registration File, providing assistance for our
Support Service.
Date of delivery:
Date of Installation:
24
The Institution / Laboratory
address:
telephone:
Your distributor:
25