GLUCOSE
GLUCOSE OXIDASE/PEROXIDASE
The application parameters comprised here constitute a guide to facilitate the validation of our reagents by
the instrument. It is advisable to validate the use when there is any change in software or reagent versions.
INSTRUMENT: BIOBASE BK200 MINI
REAGENT PREPARATION
Reagents are provided ready to use.
ITEM PARAMETERS
Item Nº * Name GLUCOSE
Item Name GLU Sample volum (uL) 3
Min Absorbancy 0 R1 volume (uL) 300
Max Absorbancy 3 R2 Volume (uL)
Normal Low value 60 Sample type SER
Normal High Value 100 Units mg/dL
Lower Value of Test 3
Highest Value of
tests 500 Prozone check *
Sample Blank yes
Pri Wavelenght 510
Sub Wavelenght 700 Pri-Point
Start Point 34
Decimals Places 2 End Point 35
One Point
Analysis Method End Sub-Point
Start Point -
End point -
Linearity
K 1
b 0
GLUCOSE
GLUCOSE OXIDASE/PEROXIDASE
CALIBRATION
Item Name GLU Repeat Counts 3
Two point
Calibration Type linear Sample Type SER
Standard Counts 2 Factor
Standard Absorbance Concentration Cup nº
0 * 0.0 *
1 * * *
* User defined
Version 03/21