This document provides parameters for validating and calibrating a glucose reagent on a Biobase BK200 Mini instrument. It specifies the item name as GLU, a sample volume of 3uL, reagent volumes of 300uL for R1 and 300uL for R2, and a normal range of 60-100 mg/dL. Calibration is done with a two-point linear method using 2 standards of unknown concentration in serum/plasma samples.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
335 views2 pages

Glucose Bk200 Biobase

This document provides parameters for validating and calibrating a glucose reagent on a Biobase BK200 Mini instrument. It specifies the item name as GLU, a sample volume of 3uL, reagent volumes of 300uL for R1 and 300uL for R2, and a normal range of 60-100 mg/dL. Calibration is done with a two-point linear method using 2 standards of unknown concentration in serum/plasma samples.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 2

GLUCOSE

GLUCOSE OXIDASE/PEROXIDASE

The application parameters comprised here constitute a guide to facilitate the validation of our reagents by
the instrument. It is advisable to validate the use when there is any change in software or reagent versions.

INSTRUMENT: BIOBASE BK200 MINI

REAGENT PREPARATION
Reagents are provided ready to use.

ITEM PARAMETERS

Item Nº * Name GLUCOSE


Item Name GLU Sample volum (uL) 3
Min Absorbancy 0 R1 volume (uL) 300
Max Absorbancy 3 R2 Volume (uL)
Normal Low value 60 Sample type SER
Normal High Value 100 Units mg/dL
Lower Value of Test 3
Highest Value of
tests 500 Prozone check *
Sample Blank yes
Pri Wavelenght 510
Sub Wavelenght 700 Pri-Point
Start Point 34
Decimals Places 2 End Point 35

One Point
Analysis Method End Sub-Point
Start Point -
End point -

Linearity
K 1
b 0
GLUCOSE
GLUCOSE OXIDASE/PEROXIDASE

CALIBRATION

Item Name GLU Repeat Counts 3


Two point
Calibration Type linear Sample Type SER
Standard Counts 2 Factor

Standard Absorbance Concentration Cup nº


0 * 0.0 *
1 * * *

* User defined

Version 03/21