TCC 151
TCC 151
TCC-151W
CAT.NO.E20132
A thermal cycle chamber that achieves a
temperature change rate of 23℃/min. and
specimen temperature ramp control of 15℃/min.
Used for everything from JEDEC standard testing to screening,
this rapid-rate thermal cycle chamber rapidly changes air temperature while controlling
specimen temperature.
It features a built-in web application that can be used to operate the chamber from a PC or tablet.
This web application allows you to check the status of the chamber
remotely using a web browser.
25
HRG-357HS-25 HRG-600HS-25 HRG-800HS-25 HRG-1800HS-25
Temperature pull-down rate (K / min.)
TCC-151W
Heat-up HRG-357HS-20 HRG-600HS-20 HRG-800HS-20 HRG-1800HS-20
20 23℃/min.
ARGF-0250-15
TCC-151W
Cool-down
18℃/min.
15
ARGF-0400-15 ARGF-0800-15
10
1
Rapid-Rate Thermal Cycle Chamber TCC-151W
ARGF
HRG
2
Performance
Specimen temperature ramp control (example) Fatigue life depends on the rate of strain
■ Specimen
and the strain waveform. In the case of
temperature control data
thermal cycle chambers, the strain rate
140
125 Setting f l u c tu ate s b a s e d o n th e te m p e r atu re
temperature change rate, while the strain waveform is
100
influenced by the symmetry of the specimen
Specimen
te mpe rature change wavefor m dur ing
Temperature (℃)
60 temperature
temperature increase and decrease.
20
60
40 carried out with ease and accuracy.
20 In addition, this chamber is also designed
0 to execute tests at a temperature change
−20 rate of 10 to 15℃ per minute as stated
−40 in IEC60749-25 and at 15℃ per minute
−60
0 10 20 30 40 50
mentioned in JESD22-A104E.
Time (min.) T h is cha mbe r is ideal ly su it e d t o
■ Temperature rate of change at twelve measuring points (Average) automotive test requi rements, life
* Conforms to IEC60068-3-5: 2001 assessments for solder joi nts, and
Temperature rate of change (℃/min.)
25
reliability assessments for semiconductor
For temperature heat up: 13.9 to 15.1℃/min.
20 For temperature pull down: 14.9 to 15.4℃/min. devices and packages.
15
10
Maximized temperature
5 uniformity for equal thermal
load to the specimen
0
1 2 3 4 5 6 7 8 9 10 11 12
Measuring point
Test conditions
High temp. soak : +125℃ Specimen : Printed Circuit Board,
Low temp. soak : −40℃ 145 × 130 mm, ⑥ ⑤ ④
Ramp rate : 15℃/min. 90 pcs.
Control point : Air outlet sensor ③ ② ①
⑫ ⑪ ⑩
Measurement method
As shown on the right, thermocouples are attached to ⑨ ⑧ ⑦
the specimens at twelve measuring points.
3
Performance
Test area
Specimen can be inserted or
removed during testing
Testing can be paused upon completion
of any given cycle. Specimen can be
inserted or removed during testing,
enabling joint testing to reduce overall
test time.
Conductor resistance
evaluations
The TCC can be used in conjunction
with the Espec Resistance Evaluation
System (AMR) (sold separately) used
for continuos measurement of micro
resistance of conductor components,
such as solder joints under temperature
cycle conditions, allowing for real-time
detection of micro-crack formation. Example of connection between Conductor Resistance Evaluation System (AMR) and TCC
I n a d d it ion , ef fe c t ive s che d u l i ng
management has been facilitated by the
integration of automatic measurement
and data logging systems.
4
Controller
N-instrumentation P-310
TCC-151W TCC-151W
Test profile copying without a PC
小型環境試験器
Edit
The chamber comes with(Wi-Fia USB機能搭載)
port
※Some items may not be copied between different models chambers with different options. that can be used with USB memory
devices (not included) to share test
profiles with other chambers.
PC
Ethernet Ethernet
スマートフ
小型環境試験器
Smartphone Tablet controller
TCC-151W 内蔵ワイヤレスダイレクトによ
5 ワイヤレスダイレクト
(オプション)
PC
Ethernet Intranet
(company LAN)
Networks TCC-151W
Mail server
Mobile phone
* An Intranet environment is required. Supported browser:Internet Explorer 11
Edit
「安心」
「
・ 安全」
につながるアフターサービス
OnlineCore Network
camera
6
SPECIFICATIONS
Model TCC−151W
System Balanced Temperature Control system (BTC system)
Temperature range −70 to +180℃ (−94 to +356℉)
±0.5℃
Temperature fluctuation
−70 to +180℃, after temperature stabilization
−45 → +155℃ *2 +155 → −45℃ *2 −23.5 → +108.5℃ +108.5 → −23.5℃ +108.5 ⇔ −23.5℃
Temperature range Target temp.: Target temp.: Target temp.: Target temp.: Target temp.:
Performance *1
−70 → +180℃ +180 → −70℃ −40 → +125℃ +125 → −40℃ +125 ⇔ −40℃
Power supply
380V AC 3φ 50Hz 61A
400V AC 3φ 50Hz 60A
Cooling water supply pressure *5 0.2 to 0.5 Mpa (2 to 5 kg/cm2G)
Cooling water supply rate *6 4100L/h (at reference water temp. +25℃)
Piping connection size Carbon steel pipe, ID 32 mm (drain and supply)
Operating cooling water temp. range +5 to +32℃ (+41 to +89.6℉)
Noise level *7 Max. 65 dB
*1 The performance values are based on IEC60068-3-5:2006, and JTM K07:2007, under the conditions of a +23℃ ambient temperature,
cooling water temperature +25℃, rated voltage, and no specimen.
*2 Refer to Fig on page 8.
*3 Specimen: (glass epoxy PCB) 5kg + Jig: 4kg (ESPEC standard jig)
*4 Excluding protrusions.
*5 Rate depends on the cleanliness of the heat exchanger
*6 A pressure regulator valve is required if the pressure exceeds 0.5MPa (5kg/ cm2G)
*7 Noise level was measured in an anechoic room at a height of 1.2 m from the floor and a distance of 1 m from the chamber front panel
(JIS-Z-8731:1999 A-weighted sound pressure level).
7
TEMPERATURE CHANGE GRAPH
200
180
Temperature range: −45 to +155℃ (setting: −70, +180℃)
160 +155 to −45℃ (setting: +180, −70℃)
140 Specimen load: None
120 Temperature control: Chamber temperature
Temperature (℃)
DIMENSIONS/FITTINGS LOCATION
Unit: mm
Ethernet port (LAN port)
Specimen power supply
control terminal
Time signal × 2
Specimen temperature
input terminal
Cable port
Power supply φ25×100㎜,
cable port right & left side
160
RS-485
ETHERNET
試料電源制御出力
POWER SUPPLY
CONTROL
FOR SPECIMEN
No.1
No.2
タイムシグナル出力
記録計用出力
RECORDER OUTPUT
空気温度 (JIS T )
CHAMBER (TYPE T)
TEMPERATURE
試料温度 (JIS T )
SPECIMEN (TYPE T)
TEMPERATURE
試料温度入力
(JIS T )
SPECIMEN
TEMPERATUREINPUT
(TYPE T)
500
状態出力
STATUS OUTPUT
COM
NO
NC
設定温度到達出力
TEMP. ATTAINMENT
OUTPUT
扉開閉時出力
DOOR OPENING
SIGNAL OUTPUT
トプロテクタ
CIRCUIT PROTECTOR
試料制御用
FOR SPECIMEN
TOTAL 100V AC 3A
計測器用
FOR INSTRUMENT
Drain pipe
400 1000
1915
8
SAFETY DEVICES ACCESSORIES
• Leakage breaker for 200, 220, 380V AC supply • Flat cable port rubber plug (Silicone sponge rubber) 2
• Circuit breaker for 400V AC supply • Specimen basket 2
• Mechanical compartment cover and (18-8 Cr-Ni stainless steel: 5 mesh per inch)
electrical compartment door switch W700×H40×D346 mm/ load capacity 5kg
• Thermal fuse for control circuit short-circuit protection
• System error
• Motor reverse prevention relay
• Thermal fuse
• Air circulator short-circuit protection
• Air circulator Thermal switch
• Overheat protector
• Temp upper limit deviation alarm
• Temp upper / lower limit absolute alarm
• Chamber door switch
• Room temperature compensation burn-out detection circuit • Shelf brackets (7 positions available, pitch 60mm) 4
• Dry bulb temperature burn-out detection circuit • Cartridge fuse
• Product temperature burn-out detection circuit 200V AC
(only when product temperature control) Type B, 250V 6A 1
• Specimen power supply control terminal 250V 4A 1
• Cooling water pressure switch 220V AC, 380V AC, 400V AC
• Heater overcurrent protection Type B, 250V 4A 1
• Refrigeration circuit temperature burn-out detection circuit 250V 5A 1
• Refrigerator short-circuit protection 250V 6A 1
• Refrigerator overcurrent protection • Specimen temperature measuring thermocouple 1
• Refrigerator high/low pressure switch • Specimen temperature input connector 1
• Cooling tower interlock terminal • Strainer R11/4 in. (32A) 1
• Nipple R11/4 in. (32A) 1
• Strainer element (#30 mesh) 1
• Operation manual 1 set
• Warranty card 1
*Power cable is optional, not equipped as standard fitting.
Safety precautions
● Do not use specimens which are explosive or inflammable, or which contain such
substances. To do so could be hazardous, as this may lead to fire or explosion.
● Do not place corrosive materials in the chamber. If corrosive substances or liquid
is used, the life of the unit may be significantly shortened specifically because of
the corrosion of stainless steel, resin and silicone materials.
● Do not place life forms or substances that exceed allowable heat generation.
● Be sure to read the operation manual before operation.
9
TEST STANDARD (TCC-151W COMPATIBILTY)
+175 ±2
1± 0.2K / min.
+155 ±2 3 hours,
− 65 ±3 3± 0.6K / min.
+125 ±2 2 hours,
−55 ±3 5±1K / min.
+100 ±2 1 hour,
IEC 60068-2-14 Nb −40 ±3 10± 2K / min.
+85 ±2 30 min., 2
(JIS C 60068-2-14 Nb) −25 ±3 15± 2K / min.
+70 ±2 10 min.
−5 ±3 (AVG)
+55 ±2 3 hours if not specified in
+5 ±3 Average for
+40 ±2 product specifications
up to five minutes
+30 ±2
+100 ±2
+95 ±2
+90 ±2 −50 ±3
+85 ±2 −45 ±3
+80 ±2 −40 ±3 3 hours,
1± 0.2℃ / min.
+75 ±2 −35 ±3 2 hours,
3± 0.6℃ / min.
+70 ±2 −30 ±3 1 hour,
IEC-61747-5 5±1.0℃ / min.
+65 ±2 −25 ±3 30 min., 2
(EIAJ ED-2531B) (AVG)
+60 ±2 −20 ±3 10 min.
Average for
+55 ±2 −15 ±3 3 hours if not specified in
up to five minutes
+50 ±2 −10 ±3 product specifications
+45 ±2 −5 ±3
+40 ±2 0 ±3
+35 ±2
+30 ±2
10
Unit: mm
Ethernet port (LAN port)
Specimen power supply
OPTIONS control terminal
Time signal × 2
Specimen temperature
Power cable input terminal Additional cable port Paperless recorder
160
RS-485
ETHERNET
試料電源制御出力
POWER SUPPLY
CONTROL
FOR SPECIMEN
No.1
No.2
タイムシグナル出力
記録計用出力
試料温度 (JIS T )
SPECIMEN (TYPE T)
TEMPERATURE
試料温度入力
(JIS T )
SPECIMEN
TEMPERATUREINPUT
(TYPE T)
500
状態出力
STATUS OUTPUT
COM
NO
NC
1808
設定温度到達出力
TEMP. ATTAINMENT
cooling water
port.
100V コンセント
POWER OUTLET
サーキッ
トプロテクタ
CIRCUIT PROTECTOR
試料制御用
FOR SPECIMEN
TOTAL 100V AC 3A
計測器用
FOR INSTRUMENT
Drain pipe
Chart recorder
Interface 400 1000
1915 −100 to +220℃ /100 mm
・ RS-485 ・RK-63: 3 pens
・ RS-232C ・RK-64: 6 dots
・ GPIB
Location: Terminal panel
Recorder wiring
Recorder terminal
RS-485
ETHERNET
試料電源制御出力
Additional cable port
POWER SUPPLY
CONTROL
FOR SPECIMEN
No.1
No.2
タイムシグナル出力
TIME SIGNALS
記録計用出力
RECORDER OUTPUT
空気温度 (JIS T )
CHAMBER (TYPE T)
TEMPERATURE
試料温度 (JIS T )
SPECIMEN (TYPE T)
TEMPERATURE
試料温度入力
(JIS T )
SPECIMEN
TEMPERATUREINPUT
(TYPE T)
Cable port
Terminal panel Emergency
状態出力
STATUS OUTPUT
COM
NO
NC
設定温度到達出力
TEMP. ATTAINMENT
OUTPUT
扉開閉時出力
DOOR OPENING
SIGNAL OUTPUT
stop pushbutton
100V コンセント
POWER OUTLET
サーキッ
トプロテクタ
CIRCUIT PROTECTOR
試料制御用
FOR SPECIMEN
TOTAL 100V AC 3A
計測器用
FOR INSTRUMENT
Recorder
11
OPTIONS
Attached to specimens to measure When the chamber is setting operation Used to bolt the chamber to the floor.
specimen temperature. such as “Error”, interlock with
・Thermocouple type T without ball connecting devices.
(Copper/ Copper-Nickel) Chamber dew tray
* Same as accessory items
Operation:
When connecting with N.O. contact Prevents water leaks from the chamber
(normally open contact), output onto the floor.
Temperature attainment output “close” contact.
When connecting with N.C. contact
When the temperature in the chamber (normally close contact), output
reaches the set values, the chamber “open” contact.
sends out a contact signal. Power supply capacity: 250 V AC, 3 A
Accessory: Plug
Location: Terminal panel
Additional overheat protector Right side or within the control
board (retrofit is not available)
Additional preventive measures can be Image
*The circuit shall be connected by customer.
taken for excessive temperature rise in
t h e c h a m b e r, i n a d d it io n t o t h e *To prevent damage in the event of water
standard equipped overheat protector. Status indicator light leakage, other preventive measures are
also available.
Select light color, lighting, and
blinking or buzzer sound.
Overcool protector
• 1 level, light: 1 color, height: 534 mm
If the temperature inside the chamber • 2 levels, light: 2 colors, height: 574 mm Casters
• 3 levels, light: 3 colors, height: 614 mm
decreases excessively, the chamber
stops operating to prevent the speci- • 4 levels, light: 4 colors, height: 654 mm Installed for mobility.
mens from being damaged. Pole length: 290 mm Casters: 4
* T he pole can be shortened in units of Levelling-feet: 4
10 mm to a minimum height of 50 mm.
13
HRG
Image
14
https://www.espec.co.jp/english
Head Office
3-5-6, Tenjinbashi, Kita-ku, Osaka 530-8550, Japan
Tel : 81-6-6358-4741 Fax : 81-6-6358-5500