Safety of IPL SHR
Any people who are eligible in using and maintaining can safely use this machine .
Any operators and his/her assistants must know well the following safe notices and
make sure both the operator and the customers are safe during the treatments .
a) The machine will start inspection automatically after it is on.
b) Adopt the close filter /crystal of transferring light, the IPL SHR can only emit
from the front face
c) Independent electrical circuit, it will automatically stop working if the energy is
too high.
d) The emergency switch is convenient when you are in emergency
e) The key switch can protect it from stolen because they can't start the machine
without certain key
.
Any intensive light without correct using will hurt people. It can work out high
voltage when working;we suggest anybody who operates this machine should read
this safety notices before using it.
For Operator: it’s possible to touch the intensive light. So make sure you wear the 3
grade glasses.
For customer: it is better if the operator is trained eligible and arrange a proper
treatment room for the customer. It is even better if the customer knows the treatment
theory and cooperates with the operator during the treatment. He/she must make sure
wear the professional protective eye patch which protects his/her eyes from burning.
About the treatment room: there must be clear mark which shows “using intensive
light NOW!”
【MAIN WARNING】
Please make sure you will carefully read the following warning tips:
a) We need to warn you that it has dangerous high voltage .Anybody except
technicians that authorized by our company can repair this machine.
Especially the inside of device e.g.: electrical source,cooling system,
optic organs and treatment head. etc.
b) Make sure the voltage of this machine is same with yours, (110V or
220V)
c) It will hurt both you and the machine if you repair it when it is still working, we
strongly warn you that you must turn off the machine and pull out the electricity line
and then start repairing it.
d) Keep the probe clean. MAKE SURE there is no cooling gel inside the probe.
e) Stop using the machine when the handpiece is leaking water, if it is already on,
please close it immediately.
1
d) This machine is about 40kg in net weight .it might hurt people if you move it in not proper
ways. So pls make sure you move it slowly and carefully
a) The temperature of objects will rise after absorbing light energy, so please take actions to
reduce the flanmable risk on treatment areas and surrounding areas.
b) Do not use alcohol, acetone or other flanmable things on skin before treatment. You can just
wash the skin with water or sometimes a little soap.
c) If you use alcohol to disinfect the treatment head or any other parts, please wait till dry and
then start the machine, otherwise it is dangerous.
【AUTOMATICAL INSPECTION】
It will start inspecting itself automatically when you start the machine. You need
to turn on and turn off the machine repeatedly for 3 times and then start using it.
【LEVEL OF EQUIPMENT】
a) Level of motor protection: 1 CLASS, BF CLASS
b) Anti‐corrosive liquid: general
c) Not allowed to use this machine in room that exists flanmable anesthetic and nitrous oxide.
d) Operation manner: continuous operation
Theory
OPT(SHR) is with new AFT technology (Advanced Fluorescence Technology) uses
low and equal energy. The special filter cut s 950-1200nm wavelength, which is
useless in treatment and absorbs water to make patient feel painless
SHR stands for Super Hair Removal, a technology for permanent hair removal which
is having a sweeping success. The system combines laser technology and the benefits
of the pulsating light method achieving practically painless results. Even hairs which
until now have been difficult or even impossible to be removed, can now be treated.
“In Motion” represents a breakthrough in permanent hair removal with light
technology. The treatment is more pleasant than with the conventional systems and
your skin is better protected.
2
Compared to normal IPL machines, SHR(OPT)( is much
more professional:
3
Stationary
Work Stationary technology “Moving IPL” can use less time to do
technology technology & In-motion treatment on back, leg and arm.
technology
Filters
480nm,530nm,640nm.
Installation
◆ Assembly parts
Power line
Water pipe
Water funnel
Professional protective glasses
Professional protective eye patch
Pictures
4
About switch:
Emergency switch: keep it bounce in working;
when press it downwards, the machine will be
power off immediately; turn right to recover)
d. Input water:
(1)Water source: pure water or distilled water;
(2)Screw off the caps of water inlet and water spilt at the same time;
(3)Take advantage of water pipe and water funnel to input water
5
Ventilation fan
Attention: Must be
fulled of water before
start the machine
After add full of water , just need to close the cap of “Water Outlet” ,
place the other two to a clear and safe area during treatment . Close all of the
three caps after turned off machine .
e.Handle installation
The handle socket
6
Nomally,the white button is protuberant.when
connect the handpiece socket,press the white
button downwards firstly.
【Function test】
a. Turn on machine, when show the first LCD display, please do not do any other
operation.
b. Turn off machine
PS: This is to make sure good water circulation before normal operation. It’s allowed
to operate only based on good water circulation.
7
If the water doesn’t cycle well or the probe does not cold, there might be
some air in the pump. We suggest do as following:
a. Pull the handle ,Turn on machine, when show the first LCD display, please do not
do any other operation.
b.Let the handle put the earth Lower than the machine . Check the probe whether
cooling well . If no, write the next step .
c. Button the water injection , until it spurt the water ,then turn off machine .
d. Connect the probe well . Turn on the machine .
8
And then the system turn to
9
Now take Hair Remove mode for example , show as follows :
Current mode
Parameter
record
IPL switch
and energy
Parameter
RF switch
and energy
Filter
Cooling System
Return selection
Pulse Interval
Pulse Number 10
Cycle
Ⅰ【Clinic operation】
A. Before treatment
To the machine:
b). Input enough water and make good water circulation before actual operation
c). Water source: distilled water or pure water
d). Before the operation, make sure probe cold. Then it’s able to start the operation. (if
take operation under the state of no water circulation or hot probe, the handpiece is
easy to damaged.)
e). Inflammable and explosive things should be kept away from operating room, such
as stupefacient,liquid and gas.
To the patients:
11
a). Classification of skin: it's safe and effective to the white, brown or yellow skin.
b). Clean skin to eliminate the perfume and cosmetics.
c). Use digital camera for memory, take photo from different orientations, if necessary
d). If necessary, use external anesthetic one hour before the treatment; Please shave
hair firstly under hair removal treatment.
f). Spread the cooling gel on skin, about 2 mm thickness
g). Wear eyepatch before actual option
Note: can not look at the intensive pulsed light directly, or alarm to people’s eyes, such
as keratitis, cornea brushing off, or even lose sight.
h). forbidden to wear watch and necklace etc. during the treatment.
To the operators:
a). Adjust the right parameters
b). Touch the " " on the LCD to start
c). Test the device working, please shot the 3 first pulse to ground at beginning,
make sure even energy output in treatment.
d). Wear glasses before actual option
e). forbidden to wear watch and necklace etc. during the treatment.
f). can not look at the intensive pulsed light directly, or alarm to people’s eyes, such as
keratitis,cornea brushing off, or even lose sight.
g). Please do not optionally open the panel of the machine even power stopped, in
case high voltage danger.
h). It must tighten up the screw caps of machine before movement.
i). vertically slide the probe on skin, and keep close to skin
J). Keep the machine rest every 2‐3 hours continuously working
k). The handle is fragile, please keep special care when hanging to the holder or in
handle
l). Please do treatment from the lower energy at the beginning, and adjust the energy
according to actual conditions.
If the probe is not cold or it even becomes hot, please stop using this machine
and turn to us for help. Otherwise it will hurt people and damage the machine.
Please do not press the button on handle one press by one press or step the foot
pedal step by step. That will make skin gets different energies which will sure hurt the
customer.
Others:
System alarm of water temperature:
The lowest system settings of water temperature is 25 degree the highest system
setting of water temperature is 60 degree. During 25---60 degree, the machine
works well. If lower 25 degree or higher 60C,now pulse output from the probe.
Please do not press the button on handle one press by one press
or step the foot pedal step by step. That will make skin gets different
energies which will sure hurt the customer.
f. When ending the treatment or try to change another treatment part or parameter
adjustment,firstly release the button on the handpiece or pedal firstly, avoiding
electric shock by chance.
g. It’s can be adjusted the energy, according to the actual conditions of the patients.
C. Special Notices
a. Avoid sunlight and burn before and after the treatment, could use the suntan lotion
(cream),(SPF>30) to avoid direct sunshine.
b. Avoid chemical or mechanical stimulation before and after 1 week of the treatment
c. Do not take filled things or injection treatment before and after 2 weeks of the
treatment
d. Avoid direct or indirect heat stimulation to the treatment area after 3 days of the
treatment,especially to those of high pigment deposit.
e. If taking laser skin care treatment before, please wait for 3 weeks at least and then
do this treatment.
f. After the treatment, patients are not allowed to eat the foods of dark color or great
pungent within two weeks, such as the sauce, coke, seafood, capsicum etc.
g. It’s not allowed insolation in sun before and after the treatment
h. It’s not allowed that use any cosmetics after the treatment, suggested that use the
natural moisturize products
i. Please cooling the bad reaction area after the treatment, like tumescence and
13
causalgiase.
j. After the operation, make sure clean the probe.
k. It’s not allowed to operate the machine without professional operators.
14
Comparative chart after
Thread Vein treatment
Comparative chart
after Face Reparing
treatment
15
Comparative
chart after
Freckle
removal & skin
rejuvenation
treatment
9. PHYSIOLOGY of HAIR
16
years, it will grow up 10cm every year, but its length will not over 1meter usually.
2. Transition
After period of hair growup, it will enter transition. Transition normally will keep
1-2weeks. During transition, hair follicle will be shrinkened to be one sixth of normal
hair length.
3.Dormancy
Dormancy is after Transition, hair will not grow up, but it also connected with hair
follicle , dermis nipple is in dormacy and will keep 5-6weeks. Normally, there is
10-15% hair in dormacy. During course of laser hair removal, it is important to choose
proper hair grow-up. During growup, mother cell of hair rapidly split, melanin
quantity is most, melania develop in period of hair growup, and will become a part of
hair follicle, it is extremely sensitive on laser. We only can use this equipment to do
hair removal during this time.
Part of body Proportion of Dormancy Density Deepness
developing period /cm2 /mm
Mandible 70% 10 week 500 2-4
Muffle 65% 6 week 500 1-2.5
Oxter 30% 12week 65 3.5-4.5
Bikini area 30% 12week 70 3.5-5
Leg 20% 24week 60 2.5-4
17
Arround eyes, wrinkle obvious and eyelid Upper eyelid and eyebrow lift obvious,
drooping before treatment wrinkle improvement after treatment
18
Eyebag obvious and tumidness After one year therapy, the before
before treatment eyebag improvement and smooth.
Lower half-face
19
Before therapy, Jaw and
After one month therapy,
neck, both have obvious
turkey neck disappear
slack, form turkey neck.
Firm the whole face,
Firm the whole face, After
Before
20
Obvious scar, concave holes After 4 months therapy, they are
before therapy improved obviously.
21
COMPARISON of HAIR REMOVAL
22
Taboos(who are not suitable to use this machine?)
a) Pregnant
g) Herpes or trauma
h) Pimple record
i) People who accept Hirudin etc. treatment (should stop using it before 2 weeks).
23
Technical Parameter & Packing List
★ Technical Parameter
Wavelength 480-950nm or 530-950nm or 640-950nm
Counter Handle Yes.
Spot size 15*50mm2
Energy 1-36J/cm2
Continually shooting timer 1-30s
Pulse repetition rate 1-10 Hz
Pulse Single
Pulse on 2-15ms
Water temperature tank 38 degree
Display system 8.4" colorful touch screen
Drive Power 2,000W
Power supply 220V, 50Hz or 110V, 60Hz
★ Packing List
Main device 1pcs
OPT(SHR) Handle 1pcs
Filters 3 pcs
Water Funnel 1pcs
Protective Eyepatch 1pcs
Protective Glasses 1pcs
Holder 1pcs
Fuse 2pcs
Power cord 1pcs
User manual 1book
24
Question & Answer
Which advantages does this have for you?
An optimal intensity can be achieved given that the same area is treated
multiple times with low energy impulses. The treatment is more comfortable
and the skin is better protected than with conventional methods.
How are different hair and skin colours affected?
Light skin with dark hair responds most favourably to the treatments. In
addition to targeting the melanin, SHR-technology affects the chromphore, a
protein located in the follicle, enabling the successful treatment of dark skin
removing achromous, fine hairs.
What does the client feel during treatment?
Treatment with SHR is practically painless as opposed to treatment with the
conventional IPL devices. On areas with dense hair growth, warmth from
the light impulses is evoked and the light impulses may be felt to a much
lesser extent than with laser or IPL.
Which areas can be treated?
Body hair can be removed on any part of the body. It is recommended a test
be performed on a small patch before the first treatment.
How many treatments are necessary?
Until now hair had to be treated during the growing phase (20%-30%). In
the telogen phase stem cells are in a resting phase but can still successfully
be treated with the SHR system. Since growth cycles vary, approx. 8
treatments are usually necessary.
How long should the intervals between treatments be?
The hair growth cycle and the interval between treatments should
correspond requiring spacing of approximately 6-8 weeks.
How much time is necessary for a treatment?
Depending on the size of the area to be treated, the time varies from 15 to 30
minutes for a facial treatment and about 3 hours for both legs.
Will there be any side-effects?
Skin is more sensitive in some areas of the body than in others and redness
can occur as a side effect. It should disappear in a matter of hours but could
potentially be as intense as a sunburn lasting a few days. Contrary to
conventional methods, a light crusting which will vanish after 2-3 weeks, or
an alteration of the sin’s pigmentation which will disappear after 4-6
25
months, are very rare.
Precautions to be taken before the treatment?
At least 1 week before the first treatment, intense sun exposure or solarium
visits should be avoided. Hairs should not be waxed or plucked less than 6
weeks before treatment. Shaving is permitted any time. Areas to be treated
should be shaved one day before the scheduled appointment.
Precautions to be taken after the treatment?
The skin’s natural protection is highly weakened after this intense light
treatment. Therefore a high level of UV protection (30-50) in the form of
suntan lotion should be applied for about 1 week.
After 1-5 weeks hair will fall out. Shaving is possible any time but hair
should not be plucked.
Maintenance of Machine
26
Ⅰ【precautions】
1) No stress on the touch‐LCD, prevent it from breaking up;
2) Keep the handle clean. It’s not allowed to clean it with ethanol;
3) The handle is fragile, please do be careful when you pick up the headpiece;
4) Change water 10-15 days. make sure water circulate normally before you start the treatment.
1. No power supplied
Please don’t worry. You can DO AS following the sequence below to find out
what the problem is.
① Check up whether there is power supplied from electrical source.
② Check up whether the power line is connected well.
③ Check up if the emergency switch is on
④ Check up if the inside electricity line is connected well
⑤ Check up the fuse, which could be pulled out at the bottom of hole for power plug.
4. Only LCD screen displays, but can’t enter into the treatment menu.
Touch LCD screen broken down
27
After-sale Service And Maintenance
The term of maintaining guarantee is main device 1 year and spare part 3 months
from the date of purchasing according to the receipt. (Expendable is not included in it.)
Others not included in maintaining guarantee or over the time, we also supply to
maintain and renew but need charges. Please contact with the dealer if need change of
purchasing the parts and easily expendable products.
A. We will guarantee the quality of our equipment, normally there is no quality
problem
1. If there is a deficiency, please email or call us on the reasons, it would be best to
dealt with by email or telephone. Because some of the problems are caused by
using the machine in a wrong way.
2. We suggest the client to take photos of the area where the problem occurs, we
will pass them to our technician , our technician will give a satisfactory answer
within 1-7 working days
3. If the problem comes from the spare part within the warranty, we could consider
in changing the spare part to our client, which could be sent with his/her next
order. If the next order is uncertain, we could pay freight cost to send the spare
parts.
4. If the problem comes from the main device within the warranty, we could advice
our client to post the device back to our company for repair by ordinary post, or
we could re-send a new one to our client. If the cost of posting the problematic
device back to BEST is more than the value of a new device, BEST would take it
into consideration, according to the actual status; apply to re-send a completely
new device back to our client.
B. We won’t supply maintaining guarantee if troubles as below:
a) Disassembling the machine by self;
b) Fall it without caution;
c) Lack of reasonable keeping and protect;
d) Not operate according to the user manual.
e) Expendable reasons
If changes of the design and specification of the machine, we will not notice you in
addition. If any queries, please feel free to contact with the local marketing company.
All in one, we will try our best to serve you and also need your fully cooperation.
Thanks.
28
Product name : Purchase date: deadline:
Maintain list
29