Liquid Nitrogen Plant Specification Sheet
LNP40
Liquid nitrogen production rate* ≥ 40 liter/day (1.7 liter/hour); 10.6 gal/day (0.44 gal/hour)
*reduced production rate on 50Hz
Dewar assembly
Capacity 160 liters (42 gal)
Empty Weight 260 lb (118 kg)
Full Weight 545 lb (247 kg)
Dimensions (OD x H) 24 x 60 in (61 x 152 cm)
Cold head model AL200, mounted on dewar
Helium compressor model CP2850, available as water or air cooled
Water cooled:
Weight 243 lb (110 kg)
Dimensions - L x W x H 19 x 18 x 24.5 in (48 x 46 x 62 cm)
Electrical rating 200/230 or 440/480VAC, 3Ph, 60Hz // 200 or 380/415VAC, 3Ph, 50Hz
Power consumption @ steady state 5.1 kW // 4.2 kW
Cooling water flow rate Minimum flow 2.3 GPM (9 LPM) @ 80°F (27°C) maximum temperature
Air cooled:
Weight 384 lb (174 kg)
Dimensions - L x W x H 23.5 x 21 x 43 in (60 x 54 x 109 cm)
Electrical rating 200/230 or 440/480VAC, 3Ph, 60Hz // 200 or 380/415VAC, 3Ph, 50Hz
Power consumption @ steady state 5.7 kW // 4.8 kW
Helium flexible lines
Standard length 10 ft (3 m)
Weight per pair 9.2 lb (4.2 kg)
Nitrogen generator assembly
Nitrogen purity 98%
Weight 66 lb (30 kg)
Dimensions (L x W x H) 42 x 7 x 29 in (107 x 18 x 74 cm)
Nitrogen Gas Flow Rate 0.76 SCFM at 7 PSIG (21.6 LPM at 0.5 bar)
System parameters
Helium pressure 210 ± 5 PSIG (14.5 ± .34 bar) @ 60 Hz
210 ± 5 PSIG (14.5 ± .34 bar) @ 50 Hz
Ambient temperature range 45°F to 100°F (7°C to 38°C)
Maximum altitude (standard LNP) 10,000 ft (3000 m)
Specifications subject to change without notice. Page 1 of 2 Revised 27Nov13
Liquid Nitrogen Plant Specification Sheet
LNP40
Compressed air requirement
Dew point 35°F (2°C)
Flow rate 5.1 ACFM at 85 PSIG (144 ALPM at 6 bar) FAD**
**FAD (free air delivery) measures the compressor’s output while operating at a specific
pressure. The result, expressed in actual cubic feet per minute (ACFM), rates the outlet
flow based on the inlet conditions.
Maximum sound level
Water cooled 70 dBA @ 1 meter
Air cooled 74 dBA @ 1 meter
Minimum footprint requirement
Water cooled helium compressor 4.5 x 3.5 ft (1.4 m x 1.1 m)
Air cooled helium compressor 6 x 6 ft (1.8 m x 1.8 m)
Shipping weight and dimensions The standard LNP is packaged in one wooden crate
Crate 1(Cold Head, Dewar and
Helium Compressor):
Water Cooled- Weight 831 lb (377 kg)
Air Cooled- Weight 972 lb (441 kg)
Dimensions (L x W x H) 62 x 38 x 70 in (157 x 97 x 178 cm)
Crate 2(Optional Air Compressor):
Weight 975 lb (442 kg)
Dimensions (L x W x H) 60 x 40 x 54 in (152 x 102 x 137 cm)
Optional Equipment
Air compressor
Flow rate 6.3 ACFM at 175 PSIG (178 ALPM at 12.1 bar) FAD
Weight 500 lb (227 kg)
Dimensions (L x W x H) 53 x 28 x 48 in (135 x 71 x 122 cm)
Duty cycle 1/3
Horsepower 3
Electrical rating 208/230; 460VAC, 3Ph, 60Hz // 220; 380VAC, 3Ph, 50Hz
Sound level 74 dBA @ 1meter
Refrigerated air dryer
Nominal capacity 15 SCFM (425 SLPM)
Weight 44 lb (20 kg)
Dimensions (L x W x H) 18 x 8 x 18 in (46 x 20 x 46 cm)
Power consumption 0.23 kW
Electrical rating 115VAC, 1Ph, 60Hz; 230VAC, 1Ph, 50Hz
Specifications subject to change without notice. Page 2 of 2 Revised 27Nov13
2 1
LNP40 SYSTEM ALL INFORMATION CONTAINED IN THIS DOCUMENT SHOULD BE
REV C 3/4"
USED FOR REFERENCE PURPOSES ONLY & ARE SUBJECT TO
AEROQUIP
CHANGE WITHOUT PRIOR NOTIFICATION. ALSO, THIS DOCUMENT
3/4" IS CONSIDERED PROPRIETARY & THE SOLE PROPERTY OF
AEROQUIP CRYOMECH, INC. & SHOULD NOT BE REPRODUCED IN ANY
MANNER WITHOUT PRIOR WRITTEN APPROVAL.
COLD HEAD
MOTOR CORD
(SAME LENGTH 1/2"
AS FLEX LINES) AEROQUIP
(2) HELIUM FLEX LINES (1) 1/2"ID 3/4"
B (1) 3/4"ID AEROQUIP B
STANDARD LENGTH: 10ft
LONGER LENGTHS AVAILABLE
NITROGEN FLEX LINE 1/4" ID
STANDARD LENGTH: 10ft
LONGER LENGTHS AVAILABLE
AL200
COLD HEAD
POWER CORD
CP2850 COMPRESSOR PACKAGE
(WATER COOLED)
NITROGEN GENERATOR ASSEMBLY
(STAND OPTIONAL)
AIR COMPRESSOR
160 LITER LIQUID
COLD HEAD (OPTIONAL)
MOTOR CORD NITROGEN DEWAR
(SAME LENGTH ASSEMBLY
AS FLEX LINES)
THIRD ANGLE PROJECTION
A CRYOMECH, INC A
113 FALSO DRIVE SYRACUSE, N.Y. 13211
Tel: (315)455-2555 Fax:(315)455-2544
ALL DIMENSIONS ARE IN INCHES
TOLERANCES: FRACTIONS ±1/32 DECIMALS XXX=±.005 XX=±.01 X=±.1
ANGLES ±.5° ALLOVER FINISH 25 Ra
CONCENTRICITY .005 TOTAL INDICATOR RUN OUT
PERPENDICULAR ±.002 (UNLESS OTHERWISE SPECIFIED)
1 YEAR
NAME: LNP40 LIQUID NITROGEN PLANT
MANUAL TOOL KIT SPARE PARTS
DWG #: LNP40 SYSTEM SHT 1 OF 1
POWER CORD 3072 - C ADDED CP2800 COMPRESSOR PACKAGE 15MAY13 MATERIAL: AS STATED
SCALE: TO DWG
2682 - B UPDATED COMPRESSOR SIDE PANELS 23FEB12 DWN BY: BL DATE: 18AUG04 FIT SIZE: A
CP2850 COMPRESSOR PACKAGE 1419 - A SEE ECN 1419 03JUL07 1ST CHK: RED DATE: 26AUG04 2ND CHK: NLO DATE: 26AUG04
(AIR COOLED) ECN ZONE REV DESCRIPTION DATE CUST: BTA:
2 1