TLD PDF
TLD PDF
net/publication/301780379
Hand book on the use of TLD badge based on CaSO4:Dy Teflon TLD discs for
individual monitoring
CITATIONS READS
3 12,703
1 author:
SEE PROFILE
Some of the authors of this publication are also working on these related projects:
All content following this page was uploaded by Ambika Sahai Pradhan on 03 May 2016.
The scope of this handbook is limited to the use of the prevalent official TLD badge system
in India. The present TLD badge is designed to asses the doses incurred by workers in the radiation
fields of X rays, high & low energy gamma rays, beta rays, mixed fields of X-rays and gamma rays
and mixed fields of beta rays & high energy gamma rays. This badge is not intended for mixed
fields of beta rays and low energy X-rays and to measure neutron doses. However its response is
adequate to measure gamma ray doses in the mixed fields of gamma ray and neutrons. Other
personnel dosimeters such as pocket dosimeters, film badges, neutron dosimeters, criticality
accident dosimeters and customized extremity dosimeters etc. are beyond the scope of this
handbook.
The purpose of this hand book is to provide the details of the infrastructure requirements
and the procedures to be followed by the processing laboratories / TLD units for carrying out the
individual monitoring using TLD system based on CaSO4:Dy Teflon TLD discs (Appendix-A).The
specimen formats provided in the appendices are for general guidelines and the exact format /
specification would vary depending upon the local requirement of a TLD unit. The handbook is
aimed at providing a ready reference material to the new entrants / forthcoming units and existing
laboratories / units.
3.2. Space: Adequate space to be provided for following activities of the laboratory / unit and
basic amenities for the staff.
3.2.1. TLD Processing Room: Two air-conditioned rooms each of total area about 30 sq. m. are
required for the processing, handling, receipt & dispatch of the TLD cards. For processing of
additional TLD cards, floor area of air conditioned rooms at a rate of 10 sq. m. per 1000 radiation
workers should be added.
3.2.2. Oven Room: An oven room of area 10 sq. m for annealing of the TLD card is required. The
room should have a platform to accommodate at least two electric ovens for annealing of TLD
cards and washbasin for cleaning of trays, cards etc.
3.2.3. Calibration/Source Room: A source cum calibration room of size 5 m x 4 m x 3 m is
required to carry out exposure of TLD cards for calibration purpose and storage of radioactive
source. The source room should have appropriate design and wall thickness in order to provide
adequate radiation shielding. Facility for storage of source in an appropriate pit (in a lead container
of appropriate wall thickness) with lock and key should be provided at one corner of the room.
3.3. Equipment and Materials: Following equipments are required at each unit.
3.3.1. TLD Cards: (Appendix –B) Sufficient number of TLD cards of the order of three to four
times of radiation workers to be monitored, should be available depending upon whether the
service is conducted in-house as at power stations or service is conducted through post.
3.3.2. TLD Cassettes: (Appendix-C) Number of TLD cassettes available should be at least one
and half times the number of radiation workers; one cassette for each card in use and remaining
to provide the replacement for the broken / damaged / lost cassettes.
3.3.3. TLD Readers: Three manual TLD badge readers (BR-3A/ BR-3B type) or two
semiautomatic N2 gas readers (BR-7B type) with an additional standby manual TLD reader are
required for the above workload. For monitoring of additional 2000 workers in a month, two
more manual readers or a semiautomatic reader will be required. Specifications of manual TLD
badge reader BR-3A/BR-3B and semiautomatic TLD badge reader BR-7B type are given in
Appendix-D and Appendix-E respectively.
3.3.4. UPS/ Voltage Stabiliser: An UPS (uninterrupted power supply)/Voltage Stabiliser of
about 3-5 KW capacity is necessary to take care of fluctuations in electric supply and power
failure.
3.3.5. Check Light Source: 63Ni beta ray source incorporated with plastic scintillator in the card
form is required to check sensitivity of PMT of TLD reader BR-3A /BR-3B type.
3.3.6. Wrappers & Polythene Pouches: (Appendix-F) A minimum requirement for one year in
advance of paper wrappers and polythene pouches should be made available. Paper wrappers
of different colours (dark) should be used.
3.3.7. Air Circulating Ovens: (Appendix-G) Minimum two air circulating ovens having
maximum temperature up to 450 °C with an accuracy of ± 2 °C in the range from 200 °C to 400
°C are required for annealing of TLD cards.
3.3.8. Radiation Source: A standard radioactive source Cs-137 (∼10-37 GBq) is required for
calibration purpose. Calibration must be traceable to national standards.
3.3.9. Contamination Monitor: (Appendix-H) One Contamination Monitor is required to check
the radioactive contamination of TLD cards received from the user institutions.
3.3.10. Gamma Zone Monitor/Area Monitor: (Appendix-I) One Area Monitor / Gamma Zone
Monitor, capable of measuring low level (<10mR/hr) radiation field is required to check the
radiation level at the calibration / source room entrance.
3.3.11. Calibration Ring/Table: (Appendix-J) A Perspex calibration ring/wooden table is
required for exposure of TLD badges / TLD cards.
3.3.12. Computer: A personal computer along with the other related accessories is required for
data entry, dose analysis and for dose record maintenance. Additional computers may be required
depending upon the type of TLD reader to be used and workload of radiation workers with the
laboratory. Computer should have facility to connect reader through interface for direct TLD
reading transfer while using manual TLD reader (BR -3A or BR- 3B).
3.3.13. Glow Curve Recorder: Normally, the PC based TLD Readers have the facility to record
the glow curve. In case of manual TLD reader without any PC interface, a two-pen glow curve
recorder, having scale in the range 1 mV to 10 V and speed of 2.5 - 25 cm per min is
recommended for recording of glow curves.
3.3.14. Annealing / Drying Trays: Sufficient number of steel trays (~20), preferably with cover,
of size 25 cm length x 4 cm height x 3.5 cm width and thickness less than 1 mm are required for
annealing of TLD cards. Minimum two trays of size 30x30x5 cm3 are required for drying the TLD
cards after cleaning with acetone.
3.3.15. Forms: Sufficient stock of TLD-5, TLD-4 and TLD-2 forms should be ensured for at least
one year. Specimen format for TLD-5, TLD-4 and TLD-2 forms are given in Appendix - K,
Appendix – L and Appendix – M, respectively.
3.3.16. Number Punching Machine: Number Punching Machine is required for engraving the
individual card number on the TLD cards.
3.3.17. Sealing Machine: A heat-sealing machine is required for sealing TLD card in polythene
pouch before despatch.
3.3.18. Thermometer & Thermocouple: An additional thermometer with thermocouple is
required to verify the temperature of the oven.
3.3.19. Gadget: A suitable electrical gadget (could be made locally, also available from the
supplier of Manual TLD Reader, Appendix-U, Kaustabh Industrial Engineers) having indicator
light facility to detect the fault with the TLD card is preferred to check the physical dimension of
the card and the appropriateness of clipping and centering of TLD discs on the card. This gadget is
more relevant for the TLD cards to be used in manual reader.
3.3.20. Vernier Calliper: A Vernier calliper is required for measuring the dimension of TLD
cards, filters in the cassette etc.
3.3.21. View Box: Commercially available view box is required to check spots and presence of
voids and foreign particles in the TLD discs.
3.3.22. Stop Watch: A stop watch is required to set exact time in terms of hour, min and sec with
alarm for irradiation of TLD cards.
3.3.23. Air Blower: An air blower is required for drying / evaporating acetone from the TLD card
after cleaning.
3.3.24. Nitrogen Cylinder and accessories: Sufficient number of nitrogen cylinders is required
for auto TLD reader along with accessories such as gas regulator, rubber tube, spanner, key and
trolley. Safe storage and proper immobilization of nitrogen cylinders during use should be ensured.
3.3.25. Long forceps: One or two stainless steel forceps of length 45-60 cm are required to handle
radiation source.
3.3.26. Acetone: Sufficient stock of analytical grade (AR) acetone is required for cleaning the
TLD cards on receipt from the supplier and during the field use.
3.4. Utility Services and other Materials
3.4.1. Boxes: Sufficient number of cardboard and plastic boxes of various sizes for despatch and
storage of TLD badges / cards should be stocked to last for at least six months.
3.4.2. Stationery: Sufficient stationery items such as computer printing paper, note books /
registers for log books for keeping and maintaining various records should be made available.
3.4.3. Furniture: Adequate furniture for the staff as well as for the equipment should be provided.
3.4.4. Communication facilities: Laboratory should have an arrangement for the required
communication such as telephone, fax, email, internet etc.
3.4.5. Fire Protection: Provision should be made for protection of laboratory from hazards arising
out of fire, electricity, fumes etc.
3.4.6. Water and Electricity: Provision should be made for adequate number of power points for
TLD readers, ovens & other equipment and tap water facility in laboratory.
4. TLD BADGE
TLD badge is based on CaSO4:Dy thermoluminescent phosphor as a radiation detector.
The concentration of Dy dopant to the extent of 0.05 mol.% in CaSO4 makes it highly sensitive
thermoluminescent (TL) phosphor (about 40 times that of the most popular LiF TLD-100). The
response to neutrons remains negligible (only 3.8 mGy equivalent Co-60 gamma ray response per
1010 thermal neutron cm-2). The phosphor exhibits a main TL glow peak at about 220°C (TL peak
of CaSO4: Dy Teflon Disc occurs at about 240 °C) with very small satellite glow peaks on higher
and lower temperature sides of the main peak. The TL response is linear and glow curve structure
remains stable for both low and high LET radiation in the dose range of interest in radiation
protection. The response and the chemical form are highly stable to the climatic variations.
Though the TL signal fades faster when exposed to bright sunlight, exposure to normal room light
does not affect the response. However, the response of TLDs contained in paper wrapper and
polythene pouches, is not affected by exposure to sunlight. In order to obtain large number of
integrated dosimeters having the same TL sensitivity, CaSO4:Dy phosphor was embedded in
Teflon (polytetrafluoroethylene PTFE) to make TLD discs. Teflon is a plastic which can
withstand temperatures necessary for TL readout and annealing process for repeated use and thus
provides rugged and handy dosimeter suitable for field use. Some of the characteristics of the
CaSO4:Dy Teflon embedded TLD discs are given in Appendix -A.
The TLD badge comprises of a TLD card and plastic cassette / holder. TLD card has three
Teflon TLD discs (13.3 mm dia. and 0.8 mm thick) that are mechanically clipped onto circular
holes (12.0 mm diameter) punched in an aluminium plate of size 52.5 x 30 x 1 mm3. An
asymmetric “V” cut is provided in the card to ensure its orientation. Specification of TLD card is
given in Appendix-B. The card wrapped in a thin paper wrapper (thickness 10 mg/cm2) containing
printed information regarding the user, service period and type of radiation etc. is put in a thin
plastic pouch (thickness 3-4 mg/cm2) before loading it in the plastic cassette. The paper wrapper
protects the TLD discs from exposure to light whereas plastic pouch prevents contamination of the
TLD card. Specifications of TLD cassette are given in Appendix-C. Specifications of paper
wrapper and plastic pouch are given in Appendix-F. There are three well-defined regions in the
plastic cassette / holder corresponding to three TLD discs of the TLD card. The asymmetric “V”
cut of the card permits its loading in the plastic cassette in only one orientation and ensures proper
positioning of three discs as follows: (i) Disc D1- sandwiched between a pair of filter combination
of 1.0 mm thick Al and 1.0 mm thick Cu (Copper filter nearer to the disc), (ii) Disc D2-
sandwiched between a pair of 1.6 mm thick (180 mg/cm2) plastic filters and (iii) Disc D3- under a
circular open window. For identification purposes, photograph of the user could also be
permanently fixed on the central transparent region of the badge.
There are three types of TLD badges / cassettes in use namely, 1- Chest Badge, 2- Wrist
Badge and 3- Head Badge. Though the dosimeter and design of all the three types of TLD badges
are same, they have different attachment (clip/strap) for wearing purpose depending on their use.
5.2. Semiautomatic N2 Gas Heating TLD Reader (Model BR-7B): Recently a semiautomatic
TLD reader has become very popular. The TLD badge reader (model BR-7B) is capable of
automatically reading 50 TLD cards in 100 min. The reader specifications are given in Appendix-
E. The main features of this reader are - measurement of doses from few µSv to 1 Sv without any
range switching, hot gas (air or N2) heating of TLD discs, auto dark current subtraction, PC based
operation (AT 386 or advance versions) and built-in data base management software. The TLD
badge reader consists of microprocessor based electronic control circuits, the PMT housing, card
transport system for moving the TLD card to the reading position, gas heater and temperature
control unit, cooling fans and a solenoid for switching the gas flow on/off. Entire operation of the
reader is PC controlled. Once the cards are loaded in the magazine and inserted into the reader,
with all corresponding badge numbers and other relevant data entered manually in the same
sequence in which the cards are loaded, the reader automatically reads the 50 cards loaded in the
magazine. It prints the TL readings as and when each card is read in order to provide a hard copy
and stores the glow curve & the TL readings in a floppy/hard disk. Elaborate self-diagnostic
software keeps an eye on the vital circuits of the reader such as EHT applied to the PMT,
temperature, gas flow etc. In case of fault detection, the readout process is terminated
automatically with a message displayed on the PC monitor to alert the user.
7.2. Setting of TLD Reader :The TLD readers should be kept ‘ON’ for at least half an hour for
warm-up before starting the actual measurements on it. The EHT, count suppression (only in case
of manual TLD reader) and initial temperature should be noted.
In case of manual TLD reader, dark current of the reader should be noted by taking the
readout in the ‘EXP’ mode with heater in ‘OFF’ position and without inserting any TLD card.
Then light source reading should be taken to check the performance of the PMT of the reader.
For this, minimum three readings of in-built LED in ‘CAL’/’LS’ mode as well as that of an
external 63Ni light source in ‘EXP’/ ‘TEST’ mode should be noted for manual reader (TLD BR-
3A/BR-3B). In case of semiautomatic reader, at least three reading of in-built 63Ni light source
should be taken in the ‘Reader Test’ mode. All the data should be recorded in the log book with
date, model, serial number of TLD reader etc.
After LS measurements, readout of 2-3 dummy TLD cards should be taken to check the
temperature and heating profile. Then minimum three exposed calibration cards should
be read to confirm the calibration of the reader and if required either Reader Calibration Factor
(RCF) should be changed or EHT to the PM tube should be adjusted to indicate counts in desired
proportion to dose (10 µSv - 1 count). If EHT is adjusted, calibration should be verified from
readout of more number of exposed calibration cards. The relative ratio of readings from three
discs of each exposed card should also be recorded and maintained in the logbook.
7.3. Intermittent Check of TLD Reader Sensitivity: The reading of 63Ni light source and
exposed calibration card should be taken intermittently in a day (at least once in the after-noon), in
between the readout of routine service cards to confirm the stability of PM tube and calibration of
the reader. In case of semiautomatic TLD reader, exposed calibration cards (at least two) are kept
at fixed positions in the magazine to verify the sequence of cards. This will ensure the calibration
of reader also.
9.2. Preparation of Cards for Annealing: TLD cards should be cleaned with acetone after the
receipt from the supplier. The cards should be checked for cleanliness before each annealing
during the field use also and should be cleaned appropriately as and when necessary. The trays
should also be cleaned with acetone prior to their use for annealing and stored properly in clean
and dust free condition. After the acetone cleaning, the cards should be dried at room temperature
for 12 –16 hrs.
9.3. Temperature and Duration of Annealing: Annealing trays containing TLD cards should
be put into the oven and the temperature should be raised from ambient to 230°C. This
temperature of 230°C should be maintained for 4 hours. The oven temperature should be allowed
to come down and trays with TLD cards should be removed only when the temperature is below
80°C.
9.4. Verification of Proper Annealing: A minimum of 5 TLD cards from each tray should be
taken and read on a calibrated reader. The TL readout of these cards should be within the limits
as given in section 8.1. The record of these readouts should be maintained in a separate logbook.
10.1. List of Radiation Workers: List of institutions and all the radiation workers (Master List)
in each institutions provided by the Concerned Authority should be updated before the dispatch of
TLD cards in each monitoring cycle. A register/log book entitled “ISSUE and DISPATCH”
should be prepared for each institution having columns to indicate details like date of despatch and
receipt, number of the cards issued for each service period.
10.2. Personnel History Form: In the case of new radiation workers duly filled in personnel data/
history form (Appendix-L), of each worker should be obtained. It should be ensured that
personnel data/history form (TLD-4) of each radiation worker is available and a personnel number
is allotted to each radiation worker by the concerned authority, before issuing any TLD badge to a
radiation worker.
10.3. Master List and TLD-5 Form: The names of radiation workers with personnel numbers
should be entered into the computer in the appropriate format which can be used for dose data
entry as well as for the preparation of the dose report. A hard copy of names and personnel
numbers should also be made available for ready reference and for further addition or deletion
of names of radiation workers. TLD-5 form should be prepared as per the updated ‘Master List’
for use in the subsequent monitoring period through the computer programme developed for this
purpose.
10.4. Wrappers for TLD Cards: The name and personnel number of each radiation worker
should be printed on the paper wrapper as per the master list. Type of radiation and service period
should also be printed on the wrapper for each radiation worker. Two or three set of paper
wrappers (depending upon in-house monitoring or postal services, respectively) may be made with
different colour codes to be used in different service periods for ease of identification of TLD
cards of different service periods.
10.5. Recording of TLD Card Numbers: The TLD card number could be noted against the
personal number and name as per the master list. Legibility of the number on the cards should be
checked before and after annealing. The cards should be physically verified for any damage like
discoloration of the disc, dislocation of disc from clips etc. and if necessary, should be replaced.
Paper wrapper should also be checked for any damage and should be replaced if required.
10.6. Dispatch of TLD Cards/Badges: TLD card identified with a number should be put in a
paper wrapper and sealed in polythene pouch of matching size. For the new institutions and new
radiation workers, the cards should be loaded in TLD cassettes/holders before despatch. In the
subsequent service period, only TLD cards wrapped in paper wrapper and sealed in polythene
pouch should be sent to the institution. The cards/badges should be dispatched to the respective
institution by registered post/courier in case of postal service or by hand delivery if arranged by the
user institution. Additional TLD cassettes are provided as per the requirement. All the instructions
and user’s manual should be sent for the new institutions.
10.7. Addition and Deletion of Names: The changes should be made in the master list for any
addition or deletion of name of radiation workers after each service period. The number of the
TLD cards and/or cassettes should be sent as per the changed requirement.
11.1. Tally of Dispatch and Receipt: The number of TLD cards/badges received from the user
institutions should be tallied and compared with the number of badges / cards dispatched as per the
entry in the log book entitled as “ISSUE and RECEIPT”. Discrepancy if any should be noted.
11.2. Contamination Check of TLD Cards on Receipt: The TLD cards received from institution
after the use should be first checked for contamination and if a card is found to be contaminated,
procedure described in section15 should be followed.
11.3. Arranging Cards for Readout: The number of cards received should be checked and if any
card is found missing or spoiled, it should be noted on the TLD-5 form. Polythene pouches
should be cut and kept open so that the cards can be easily removed at the time of readout. TLD
cards along with paper wrapper & polythene pouch may be kept in trays in sequence.
11.4. Intimation of Missing/Spoiled/Contaminated Cards: Report about the missing / spoiled /
contaminated cards should be sent to the concerned person (health physicist/radiation safety
officer/head of the institution) as soon as it is noticed. A record of missing TLD card number /
badge and the corresponding personnel number should be maintained in the log book and should
be communicated to concerned parties and for the purpose of calculating the charges.
12.1. Recording of Reader Details and Personnel Number: The TLD badge reader should be
calibrated as described in Section 7. Before taking actual readout of the cards, all the details like
Reader no., reader calibration factor (RCF), EHT value, date of measurement etc. should be noted
on each TLD-5 form and it should be signed by operator (person taking measurement). TLD card
number and personal number of the worker should be checked on the TLD-5 form before heating
the cards in the reader. In case of semiautomatic gas TLD reader, exposed calibration cards (at
least two) should be kept at fixed positions in the magazine to the verify sequence of cards which
will, in turn ensure the calibration of TLD reader also.
12.2. Recording of Readout of TLD Cards: Irrespective of availability of computer interface
facilities or TLD readers having readout recording facility, the TL readings should be recorded on
the TLD-5 form (Appendix-K) in the respective columns against their personnel number by the
operator.
12.3. Dose Evaluation: After completion of the readouts of TLD cards, the doses should be
evaluated and entered in the respective column on TLD-5 forms by using appropriate algorithm
(Appendix-O). In case of computerised dose evaluation, it should be verified manually for the
proper entry of data and functioning of the algorithm.
12.4. High Reading: In case of the manual TLD reader, if the reading of Disc 1(D1) in any of the
TLD card is recorded abnormally high (more than 10 mSv), before taking further readings,
the performance of the TLD reader should be checked by taking calibration card reading. In such
cases, procedure given in Section 16 has to be followed.
12.5. Anomalous Reading: If any abnormal pattern of readings is observed, the same should be
investigated as described in Section 18.3.
12.6. Dose Reports: In case of manual reader the dose report could be obtained through computer
by feeding the required dose data in the required format, in case of semiautomatic reader, dose
report could be prepared using the software provided by the supplier or developed by the
processing laboratory/unit. The specimen format for dose report is given in Appendix-N.
Receipt
Date of Number of TLD Received Received Remarks
Receipt Cards Received By From
13.2. Accounting of Cards: Record of changes in the TLD-5 form, i.e. addition or deletion of
any personnel number, any non-returned/missing TLD card or receipt of any old card should be
kept separately in the logbook entitled `ACCOUNTING’.
13.3
Sr. No. Inst. No. Changes in each Total No. of TLD cards Remark .
Ma
Service Month
ster
Addition Deletion Issued Not returned File
:
‘Ma
ster list’ containing personnel number and names of workers should be maintained for future
reference.
13.4. Dose Records: All the dose data should be kept in the appropriate format in the computer
by each processing laboratory/unit in addition to the data maintained by the Central Dose Records
at BARC.
13.5. Status of TLD Cards, TLD Readers and Other Equipment: Record of stock and
performance of readers, cards & other equipments and also the details regarding repair if any,
should be maintained in the respective log books.
13.6. Record of QA performance: Records of internal and external QA checks should be
maintained properly and separately.
13.7. Status of TLD Cards: Each laboratory/unit should maintain a record of data on selection of
TLD cards, sensitivity check and quality assurance programme, experimental card reading,
blind test readings, annealed card reading etc. in the separate log book meant for each purpose.
13.8. Other records: Records of overexposure cases, TLD-DRD discrepancy cases (in
case of units at power stations) and abnormal pattern of TL readings should be maintained in
proper logbooks.
In certain situations, TLD badges of radiation workers while handling open sources or
working in environment of radioactive materials, may get contaminated. For handling such cases,
the following procedures are to be followed before processing of the contaminated TLD badges by
the laboratory/unit.
15.1. Checking for Contamination :The TLD cards received after use should be checked for
contamination. For this purpose, the whole lot of cards is to be checked first and in case of any
doubt, each card from the lot should be checked. If required polythene pouch, paper wrapper
and card should also be checked individually. If the card, paper wrapper or polythene pouch is
found contaminated, it should be removed and stored separately. The problem may be indicated
on the TLD-5 form against the particular number. The same should also be reported to the
concerned agencies. Contaminated TLD cards should not be read in the TLD reader. In case,
the contaminated card has been read in the TLD reader, a thorough check of the reader
magazine/card holder should be carried out and if necessary decontamination may be done before
taking any further readout or introducing next magazine in this TLD reader. Readout of
contaminated card should not be considered for reporting the dose.
15.2. Contamination Check of Cards with Very High TL Reading: Apart from recording the
glow curve, the cards which gives abnormally high TL reading should also be checked for
contamination of the cassette / storage place etc. to rule out the possibility of contribution to the
reading due to radioactive contamination. In case of detection of any contamination, suitable action
should be initiated.
15.3. Decontamination: If the card is found to have radioactive contamination or higher reading is
confirmed due to contamination, the card must be removed from the lot of field cards and
discarded. TLD badge reader and work place should be decontaminated if necessary. Further
readout on this TLD reader should be carried out only after ensuring that there is no contamination
and re-confirming the calibration of the reader. Contaminated cards / wrappers should be disposed
off as per the procedure of disposal of radioactive waste which depends on the level of
contamination. Investigation should be initiated to know the cause of contamination.
15.4. Reconciliation of Dose: In case of contaminated card, personal doses are usually not
estimated from the readout of the card. However, for radiation protection requirements, the doses
could be estimated from the work details, doses received by co-workers and / or dose measured by
other dosimeters such as use of DRD (pocket dosimeter) etc. if any. If the source of
contamination is confirmed, the dose can also be reconciled by taking into account the exposure
to the disc due to contamination. This can be achieved by collecting the work details of the worker
and also by calculation and/ or experimental verification.
16. HANDLING OF OVER EXPOSURE CASES
During routine processing of the TLD cards, if dose of any radiation worker found to be
more than 10 mSv in a monitoring period (monthly / quarterly), this should be considered as a case
for overexposure (OE). Special procedure should be followed for the overexposed (OE) cards as
given below:-
1. After recording a high reading of disc D1 in the manual reader, TLD card, wrapper, polythene
pouch & if available, cassette should be checked for any contamination and senior person
should be informed about OE case. The remaining readings in the manual reader should be
taken in the presence of another concerned expert / senior person. In case of semiautomatic gas
reader, after the readout of the OE TLD cards in the magazine the reading of the other cards
should be terminated if the readout of the OE cards approaches counts equivalent to 1Sv of 137
Cs gamma rays. For lower readouts, all the remaining cards in the magazine could be read
before starting the special procedure of the OE card.
2. Three annealed TLD cards (as test cards) should be taken from the same lot of field card and
exposed to a known dose of 137Cs gamma ray source.
3. In case of semiautomatic reader, readout of freshly exposed test cards should be taken on the
same reader on which over exposed card was read. Second readout of the OE card and freshly
exposed test cards should also be taken on the same reader and reading of second readout
should be compared. In case of manual reader, readout of disc D2 of OE card, 2nd readout of
OE card and freshly exposed test cards should be taken in presence of Senior Scientific Staff
(for accredited laboratory) or Health Physicist (for TLD units at power station). All the glow
curves of the OE card and test card of first and second readout should be recorded.
4. In case of very high exposure (≥500 mSv), special precaution should be taken to avoid PMT
fatigue while taking the readout and the TLD reader should be checked by using calibration
card for any change / fatigue in the PMT tube response before taking further readouts of the
TLD card.
5. All the OE and test TLD cards should be annealed and should be exposed to about 5 mSv of
137
Cs gamma rays after taking the background reading.
6. Readout of all TLD cards should be taken on same TLD reader on which OE card was read.
7. If all steps go in convincing direction, report the over exposure. In case of any doubts, final
discussion could be arrived by the mutual consent of the concerned persons.
9. Once a TLD card records an over exposure, it has to be removed from the service.
18. MISCELLANEOUS
18.1. Calibration of Radioactive Source: The Radioactive source used for exposure of TLD
cards should be calibrated by the Radiation Standards Section (RSS) of Radiation Safety System
Division, BARC. The air-kerma rate should be measured by RSS at the processing laboratory/unit
in the actual geometry to be used for the exposure of calibration/experimental cards and the source
calibration should be re-confirmed after every 5 years.
18.2. Swipe Test of Radioactive Source: Swipe sample of radioactive source should be collected
occasionally (once in a years) and sample should be checked using contamination monitor for any
radioactive contamination or damage to the source. If significant counts (> equivalent of activity
185 Bq) are observed, the source should be sealed in the polythene bag and kept in container.
Matter should be reported to Head of the institution; Head, RP&AD, BARC & AERB and source
should be disposed off as per the procedure approved by regulatory body AERB. Arrangements
should be made to procure a new source.
18.3. Investigation of Abnormal Pattern of TL Readings: During the routine readout of cards, if
the TL readout pattern of the discs in any card is noticed to be abnormal then special measures
should be taken. The card, paper wrapper and polythene pouch should be first checked for any
contamination (radioactive/chemical). If necessary, the history of that TLD card should be traced
out to check the initial performance of that card before introduction into service. The matter
should be immediately reported to the Health Physicist/Concerned Authority or Agency, who
would in turn inform the concerned institution to check the physical integrity/ contamination of
the particular cassette and obtain the details regarding the contamination/ radiation level in the
area where the individual has worked and where cards were stored while not in use for that
service period. Area monitoring should be advised in such cases and nature of radiation should be
traced out to ensure remedial steps. If the nature of radiation is not confirmed and no reason is
assigned for abnormal pattern, dose should be reconciled either by calculation method or from
dose/s received by co-worker/s . In case of Nuclear Power Stations, the dose can be assigned on
the basis of the readings recorded by alternate dosimeter for the said period such as DRD (pocket
dosimeter) reading.
If the TLD card is found to be contaminated then the procedure as described in section 15 should
be followed and dose should be assigned accordingly. In case, there is no radioactive
contamination, the card should be checked for contamination of luminescent material by taking the
readout of cards without heating. If the readout without heating is high, the card should be
discarded and dose should be reconciled as above. If the readout without heating is close to the
expected value, the second readout of each of the discs should be taken and compared. The second
readout should be about 10% of the first readout. The card should then be cleaned in the acetone,
annealed and read in the TLD reader to confirm the TL readout to be close to the background
signal. This card should then be exposed to a known dose of 137Cs gamma rays and readout to
confirm the behaviour of the card by recording the glow curve. If found to follow the normal
pattern, investigation for conditions of exposure may be carried out subject to the discretion of
scientific officer /in-charge of the laboratory/unit.
REFERENCES
1. ANSI (1983), American National Standard for Dosimetry - Personnel Dosimetry Radiation
Performance - Criteria for Testing, American National Standards Institute (ANSI) Inc., New
York, HPSN 1311 – 1983.
2. ANSI (1993), American National Standard for Dosimetry - Personnel Dosimetry Radiation
Performance - Criteria for Testing, American National Standards Institute (ANSI) Inc., New
York, HPSN 1311 – 1993
3. ANSI (2001), American National Standard for Dosimetry - Personnel Dosimetry Radiation
Performance - Criteria for Testing, American National Standards Institute (ANSI) Inc., New
York, HPSN 1311 - 2001, 11/2000 Version, 2001.
4. Bartlett, D.T. and Lberts W.G., (1994), Type Testing and Calibration of Personal Dosemeters,
Radiat. Prot. Dosim. 54, 259-265.
5. Bhatt B.C., Srivastava J.K., Sanaye S.S., Shinde S.S., Patil A.S. and Supe S.J. (1994), Personal
Dosimeter Intercomparisons as Indices of Performance - A Quality Assurance Procedure,
Radiat. Prot. Dosim., 54, 343-346.
6. Bohm, J., Lebedev, V.N. and McDonalds (1994), Performance Testing of Dosimetry Services
and its Regulatory Aspects, Radiat. Prot. Dosim., 54, 311-319.
7. BSS ( 1996), International Basic Safety Standards for Protection against Ionizing Radiation
and for the Safety of Radiation Sources, IAEA Safety Series No.115, International Atomic
Energy Agency, Vienna.
8. Dere. A.V., Popli K.L., Nagpal, J.S. Kapoor D.K. and Patel P.H. (1993), Dose Computation
Algorithms for Individual Thermoluminescent Dosimetry, Bull. Radiat. Prot. (IARP, India),
16, 54-55.
9. DOE (1986), Department of Energy Standard for the Performance Testing of Personnel
Dosimetry Systems, DOE Laboratory Accreditation Program for Personnel Dosimetry System,
U.S. Department of Energy, DOE/EH/0027UC-41), December 1985.
10. Gangadharan P., Sankaran A. and Kannan S., (1975), A Semi-Automatic TLD Personnel
Monitoring Badge Processor for Routine Use, Proc. Nat. Symp. On Thermoluminescence and
its Application, Kalpakkam, 563-579.
15. ICRU (1985), International Commission on Radiation Units and Measurements, Determination
of Dose Equivalents Resulting from External Radiation Sources ICRU Report 39, Bethesda,
MD.
16. ICRU (1988), International Commission on Radiation Units and Measurements, Determination
of Dose Equivalents from External Radiation Sources - Part 2 ICRU Report 43.
17. ICRU (1992), International Commission on Radiation Units and Measurements, Determination
of Dose Equivalents from External Photon and Electron Radiations ICRU Report 46.
18. ICRU (1998), International Commission on Radiation Units and Measurements, Conversion
Coefficients for use in Radiological Protection Against External Radiation, ICRU Report 57.
20. Intercomparison for individual monitoring of external exposure from photon radiation, IAEA-
TECDOC – 1126, December 1999.
21. ISO/IEC (1997), General Requirements for the Competence of Testing and Calibration
Laboratories, Draft, Ref No. :ISO/IEC Guide25.
22. Julius, H.W., Marshall, T.O., Christensen, P and Von Dijk, W.E. (1997) - Type Testing of
Personal Dosemeters of Photon Energy and Angular Response, Radiat, Prot. Dosim. 54, 273-
276.
23. Kannan., S. Kulkarni M.S., Ratna P and Lalsare M.D (1997) - An Automated TLD Badge
Reader For Large Scale Personnel Monitoring, Proc. 23rd IARP conference of Recent Advance
in Radiation Measurements and Radiation Protection held at Guru Nanak Dev University
Amritsar, Feb. 19-21,
24. Kher, R.K., Adtani, M. M., Supe S.J. and Vohra K.G. (1983), Experiences in the Use of TLD
Badge for Personal Monitoring, Bull. Radiat. Prot., 6, 83-88.
25. Kramer, H.M., Bohn, J., Iles, W.J. and Thompson, I.M.G (1994), On the Current Status of an
ISO working Document on the Calibration and Type Testing of Radiation Protection
Dosimeters for Photons, Radiat. Prot. Dosim. 54, 267-272.
26. Lakshmanan A.R., Popli K.L.. and Kher R.K. (1989a), Photon Energy Dependence of
CaSO4:Dy TLD under Differenct Metal Filters in Terms of the New ICRU Quantitites,
Radiat. Prot. Dosim., 28, 273-275.
27. Lakshmanan, A.R., Gopalkrishnan A.K. and Kher R.K. (1989b), Photon Energy Dependence
of CaSO4:Dy TLD Thermoluminescence Dosimeter Badge to Diagnostic X-rays and Dose
Evaluation Procedures, Radiat. Prot. Dosim., 28, 263-272.
28. Lakshmanna., A.R. Bhuwan Chandra, Pradhan A.S. and Supe S.J. (1986) - Application of
Thermoluminescence Dosemeters for Personnel Monitoring in India, Radiat. Prot. Dosim., 17,
49-52.
29. NAVLAP (1995), National Voluntary Laboratory Accreditation Program Bulletin, Dosimetry,
Voll-II, 1, NIST/NAVLAP, Gaithersburg, MD 20899.
30. Nagpal., J.S, Udaykumar., J. Page, A.G. and Venkatraman G. (1995) - Studies on 63Ni Based
Check Light Source For TL Readers - Radiat. Prot. Dosim, 60, 181 - 184.
31. Nagpal J.S. (1993) - TL Monitoring System, Dose Evaluation Procedure and Quality
Assurance Workshop on Individual Monitoring, July 19-23, BARC, Bombay.
33. Pradhan A.S., Dere A. and Popli, K.L (1979) –Annealing and Repeated Readout of TLD Cards
based on CaSO4:Dy Teflon Discs, Int. Journal Appl. Radiat. Isotope, 30, 317-319.
34. Pradhan A.S. Bhuwan Chandra and Ayyanger K., (1975), Development of CaSO4:Dy Teflon
Discs for Thermoluminescence Dosimetry, Proc. Nat. Symp. On Thermoluminescence and its
Application, Kalpakkam 409-413.
35. Pradhan A.S. (1981), Thermoluminescence Dosimetery and its Applications, Radiat. Prot.
Dosim. 1, 153-166.
36. Pradhan A.S. and Bakshi A.K. (1997), Role of Post-Irradiation Annealing for Instant
Evaluation of Personal Monitoring Badges based on CaSO4:Dy Teflon TLD Discs, Proc. Nat.
Symp. On Thermoluminescence and its Application, Raipur, 196-199.
37. Vohra K.G., Bhatt R.C., Bhuwan Chandra, Pradhan A.S., Lakshmanan A.R. and Shastry S.S
(1980) - A Personnel Dosimetry TLD Badge based on CaSO4:Dy Teflon TLD Discs, Health
Physics, 38, 193-196.
38. Pradhan A.S. and Bakshi A.K. (2002), Calibration of TLD Badges for Photons of Energy
above 6 MeV and Dosimetric Intricacies in High Energy Gamma Ray Fields Encountered in
Nuclear Power Plants, Radiat. Prot. Dosim., 98, 283-290.
APPENDIX - A
Characteristics of CaSO4:Dy Teflon TLD disc
Ratio of CaSO4: Dy and Teflon : 1: 3
Effective Atomic Number (Z) : 15.1
Density of the TLD Disc : 2.52 g/cm3
Softening Point of Teflon : 330°C
Main Glow Peak Temperature : 230°C (as shown in following figure)
Sensitivity of TLD Disc : About 30-40 times more than LiF TLD-100
Fading : 2 -3% in six months.
Climactic Effect (Environmental) : Negligible
Effect of Sunlight : Negligible when covered by paper wrapper &
polythene pouch and loaded in the badge
TL (arbitrary units)
Reusability : 20 cycles
Beta Response : 60% of 60Co gammas for Nat.U (effective energy 0.8 MeV)
Thermal Neutron Response: 60 Co gamma ray equivalent to 2.4 mGy per 1010 n/cm2
Energy Response: Following figure shows the photon energy dependence of CaSO4 : Dy Teflon
TLD disc from 30 keV to 1.25 MeV. The TLD disc under the metal filter shows a modified
response, with a 65 % over response at 80 keV and 25% under response at 30 keV.
10
8
Relative TL Response
b
6
2 a
10 2 5 102 2 5 103 2
12.0 13.3 mm
mm dia dia disc
Fig. 3a TLD Card Without Disc Fig. 3b TLD Card With Discs
APPENDIX- C
Specifications of TLD Cassette
TLD chest badge cassette consists of three regions corresponding to three TLD discs of the
card. Rectangular copper filter is fixed with adhesive or embedded by molding in main body and
circular copper filter is embedded by molding in slider. Aluminium emblem is fixed by adhesive
(Araldite) in the main body and on the slider. Transparent windows is embedded by double
moulding within main body and slider. TLD cassette has a S.S Crocodile clip fixed with silver
plated eyelets to the slider part for affixing to the user’s clothing. Quality of filter and plastic
material of the cassette should be as per the specifications given below.
Dimensions
(a) Main body: (b) Slider part:
Cu filter : 30 mm x 16 mm x 1 mm Cu filter : Dia – 16 mm, thickness- 1 mm
Al filter : Dia- 15.6 mm, thickness- 1.0 mm Al filter : Dia –15.6 mm, thickness-1 mm
Plastic filter : 30.5 mm x 21 x 1.6 mm Plastic filter : Dia –25 mm, thickness-1.6 mm
Open window : Dia - 14.5 mm Open window: Dia – 13.5 mm
Material
Eyelet : Silver eyelet 2030
Transparent window : General purpose polystyrene 666 clear
Copper Filter : Copper 99.9 % purity
Alluminium Filter : Aluminium (Emblem)
Slider : High impact polystyrene -975 ( Ultra white)
Main Body : High impact Polystyrene -975 (Ultra white)
Fig. 4
APPENDIX- C (contd.)
Specifications of TLD Cassette of Altered Filter Dimension
In this design of the TLD cassette, dimension of some of the filters was altered and crocodile
clip was replaced by a smaller size clip. The cassette was made of ABS plastic (white) and filters
were embedded into the plastic body. Dimensions of the filters are as given below.
Dimensions:
Main body:
Cu filter (rectangular) : 32 mm x 16 mm x 1 mm
Al filter (circular) : 13.6 mm & thickness- 0.6 mm
Plastic filter (rectangular) : 30.5 mm x 21 mm x 1.6 mm
Open window : Dia- 14.5 mm
Slider part:
Cu filter (circular) : Dia - 15.6 mm, thickness- 1 mm
Al filter (circular) : Dia -12.6 mm, thickness – 0.6 mm
Plastic filter (circular) : Dia - 25 mm, thickness- 1.5 mm
Open window : Dia - 13.5 mm
APPENDIX - D
Specifications of Manual TLD Reader (Model BR-3A)
Heating method : Electrical
Light detector : Photo multiplier tube (PMT) selected for dark current
< 1 nA at 1000V
Weight : 14 Kg (approx.)
APPENDIX-E
Specifications of Semiautomatic TLD Reader (Model BR-7B)
Dosimeter :Three-element BARC CaSO4:Dy Teflon disc dosimeter
Dark current : Dark current is 1µSv (TL output from CaSO4:Dy) equivalent
with software based sampling & subtraction
Heating cycle : The temperature is raised to 300 °C in 8 –10 sec and clamped at
300 °C
Facilities available
*Entry of badge ID numbers, calibration factor, etc.
*Storage of dose and glow curve data of badges in floppy/hard disk,
* Motorized driver assembly for automatic feeding of 50 dosimeter cards loaded in a magazine.
Software
IBM PC compatible menu driven software for transfer of data to a user defined file, storage and
display of glow curves, computation of dose and generation of dose reports.
Temperature monitoring
Chromel - Alumel thermocouple in hot gas stream. External temperature monitoring facility is
provided at the back panel
Calibration
Coarse adjustment by varying the EHT through a potentiometer in the EHT circuit; Fine through
software.
Power
Power supply: 230V, 50 Hz: Power requirements: 500 VA (including PC)
PC requirement
IBM PC 486 or upward compatible with FDD & HDD, SVGA monitor, Serial & Parallel ports,
Dos 6.20 and MS Windows 95/98/NT.
APPENDIX-E (contd.)
Safeguards: A constant vigil should be kept during the entire readout procedure to avoid loss of
any TL reading due to reader malfunction/failure. Special attention should be given to the
following.
(1) Heater/Gas flow Failure: The heater and gas flow should be checked for failure in every
dosimeter readout cycle. In the event of failure of heater or gas flow, the readout should be
terminated and a message indicating heater/gas flow failure is flashed on the PC monitor.
(2) Mechanical failure: Any mechanical failure during readout cycle is sensed using a time-out
watch dog programme and the cycle is terminated with an option for the user to restart the cycle.
(3) EHT/Input circuit failure: Failure of EHT or the input circuit (I-F converter) is sensed and
reading cycle is terminated in case of failure.
APPENDIX-F
2) Polythene pouch
Thickness : 3 - 4 mg /cm2
Size : 32 mm width x 60 mm length
Paper Wrapper
APPENDIX-G
B) Heater:
1) Material - Nichrome
2) Type - Tubular
3) Rating - 3.0 kW.
4) Heating Rate - Three different heating rates.
C) Temperature:
D) Air Circulation:
E) Electrical Supply:
Indication : Visual - Four and half digit seven segment LCD display to
indicate contamination levels in counts per second or mR/h
Climatic conditions
Temperature range : 10 °C to 50 °C
Relative Humidity : 0 to 99%
Alarms : Loud, continuous pulsed tone, large area red alarm light with
remote indication on a I/O connector (optional)
Wooden Table
APPENDIX-K
Specimen Format* of TLD-5 Form – Recording of TL Readout
Name & Address of the Laboratory / Unit
INSTITUTION NUMBER:
No. of cards sent: Reader No. :
No. of cards received: EHT:
TLD cards read on: RCF:
*Format / specification may vary depending upon the requirements of the TLD unit
APPENDIX-L
Specimen Format* of TLD-4 Form - Personal Data of Radiation Worker
Name & Address of the Laboratory / Unit
(Instruction: Use block letters, leave one blank square between each word)
6. Name of father/ :
Husband
14. a) Have you availed the personnel monitoring service in the previous institution Yes No
b) If the answer is “Yes” , please give the following details:
(In case of more than one institution, please furnish the information on a separate sheet)
19.TLD badge required for CHEST only/ both CHEST & WRIST
20. Details of radiation sources to be handled:
Date Date
*Format / specification may vary depending upon the requirements of the TLD unit
APPENDIX-M
#
Specimen Format of TLD-2 Form - Application for Personnel Monitoring
Services
Name & Address of the Laboratory / Unit
Note: (i) Please read TLD Personnel Monitoring Service Manual before filling the form.
(ii) Separate sheet may be attached if space is not sufficient.
1. Name of the Institute: ___________________________________________________
2. Postal Address for Correspondence: ________________________________________
(with phone no, fax and email address): ________________________________________
________________________________________
3. Details of persons who will be using the TLD badges:
Sr. Name in Full Sex Date of C/W* Remarks**
No. Birth
(c) Medical X-ray units (viz. Radiography, Fluoroscopy, Dental, Therapeautic, CT Scan, Cardiac
Catheterisation, Angiography, etc.)
Sr. No. Make, Operating Current Fixed or Year of Remarks
Model and Voltage (mA) Mobile Installation
Type of unit (kVp)
(d) Telegamma Units
Sr. No. Details of Name of Activity Date of Remarks
Make and the Isotope Curie On Date Installation
Type of unit or GBq
(e) Sealed Sources such as Ra-226, Co-60, Ta-182, Sr-90 Au-198 Cs-137, Ir-192 etc.
Sr. No. Source & Activity Type of Use Remarks
its Physical form Curie or On
GBq Date
(f) Unsealed Sources such as Tc-99m, Na-24, P-32, K-42, Cr-51, Fe-59, Y-90, I-131, Au-198 etc.
in liquid and powder form
Activity in Quantity Physical Chemical Activity to be Remarks
Sr. Source Stock handled at a form form procured
No. mCi On time (mCi/ Month Year
(MBq) date MBq)
Note: Please give similar details on a separate sheet if any other radiation generating machine
or sources which have not been mentioned above are available with you
DECLARATION
I HAVE READ THE TLD BADGE USER’S INSTRUCTION MANUAL AND THE
ATTACHED TERMS AND CONDITIONS AND I UNDERTAKE TO
1. Comply with all instructions stated in the manual and any other instruction to be given
time to time by the competent authority.
2. Comply, agree and follow all terms & conditions
3. Investigate the cases of over exposures or contamination of TLD cards and send the
reports promptly whenever called for
Date Signature of the Head of the Institution with Stamp and Seal
Place Name of the Signatory
Designation
# Format / specification may vary depending upon the requirements of the TLD unit
APPENDIX-N
Specimen Format# of Dose Report
Name & Address of the Laboratory / Unit
Per Name Worn Whole Body Dose (mSv) Beta Current Wrist Current
No. at Jan- Apr- Jul - Oct- Current Dose Year Dose Year
Mar Jun Sept Dec Year (mSv) Beta (mSv) Wrist
Total Total Total
(mSv) (mSv)
# Format / specification may vary depending upon the requirements of the TLD unit
Notes:
1. All doses are reported in millisievert (1mSv=100 mrem) 2. No Dose Reported unless PDF (TLD-4 ) form
received.
3. When more than one cards are used, dose is added. 4. Dose below measurable dose is shown zero.(low
energy X= 0.05 mSv, β= 0.5 mSv or γ= 0.2 mSv).
5. Annual effective dose in any calendar year shall 6. Cumulative effective dose constraint for the 5 year
not exceed 30 mSv, block from 1.1.1999 to 31.12. 2003 is 100 mSv.
7. ‘_’ means in any service period dosemeter not used. 8. ‘* ’ means Total dose> 20 mSv. Inform details of the
radiation work/ practice to BARC/AERB
9. ‘* * ’ means letter sent to the institution. 10. ‘* * * ’ means under any service period dose
>100 mSv.
APPENDIX-N (Contd.)
Specimen Format# of Dose Report
Name & Address of the Laboratory / Unit
# Format / specification may vary depending upon the requirements of the TLD unit
Notes:
1. All doses are reported in millisievert (1mSv=100 mrem) 2. No Dose Reported unless PDF (TLD-4 ) form received.
3. When more than one cards are used dose is added. 4. Dose below measurable dose is shown zero (low energy
X= 0.05 mSv, β= 0.5 mSv or γ= 0.2 mSv).
5. Annual effective dose in any calendar year shall 6. Cumulative effective dose constraint for the 5 year
not exceed 30 mSv. block from 1.1.1999 to 31.12.2003 is 100 mSv.
7. ‘_’ means in any service period dosemeter not used. 8. ‘* ’ means Total dose> 20 mSv. Inform details of the
radiation work/ practice to BARC/AERB
9. ‘* * ’ means letter sent to the institution. 10. ‘* * * ’ means under any service period dose >100 mSv.
APPENDIX -O
Algorithm for Dose Evaluation
Key to abbreviations:
D1 : reading of dosimeter under Cu-Al filter
D2 : reading of dosimeter under plastic window
D3 : reading of dosimeter under open window
RV : reporting value.
D1’, D2’, D3’ are D1, D2, D3 minus the control card reading.
1. Gamma Ray Dose: If the ratio of D3’ to D1’ is less than 1.3, only gamma ray dose should be
evaluated using the value of D1’ i.e.
Gamma Dose = D1’ (equated to zero if less than RV)
2. Beta Dose: If the ratio of D3’ to D1’ is more than 1.3, also the ratio of D3’ to D2’ is more
than 1.3 and D1’ is very close to background reading, then beta dose should be evaluated as
follows.
3. Beta and Gamma Ray Doses: If the ratio of D3’ to D1’ and D2 to D1 are more than 1.3 also
ratio of D3’ to D2’ is more than 1.2 and D1 is significantly higher than background, both beta
and gamma doses should be evaluated.
Beta dose should be calculated as per the calculation shown above taking appropriate value of
BMF from the ratio of D2 & D3
APPENDIX –O (contd.)
Case –II
If the ratio of D2’ to D1’ is more than “1.3” but less than or equal to “5” (i.e. 1.3<D2’/
D1’≤ 5) which indicates that the TLD badge is exposed to X rays of energy between 40 keV to
200 keV or TLD badge is worn under lead apron and exposed to diagnostic X-rays. In this case
X-ray dose should be evaluated as follows
Case -III
If the ratio of D2’ to D1’ and D3’ to D1’ are more than “5” and D2~D3 (within 20%),
which means the TLD badge is exposed to X rays of energy less than 40 keV or the badge is
worn over the lead apron. Dose should be evaluated as follows:
Dose = (D2’ + D3’) / 20
Note:
1. In case there is any ambiguity in the ratio of D2’/D1’, D3’/D1’ etc., information on the
type of radiation source handled by the worker should be obtained and accordingly the dose
should be evaluated.
2. Beta & Gamma doses are evaluated when it is ensured that photons of effective energy less
than 60 keV are not encountered significantly.
3. X ray doses are evaluated when it is ensured that beta radiation is not encountered.
4. For the algorithm example of Case-II, other tested methods developed by individual TLD
units could also be used. It may be noted that Dose = D1 (Case-I) remains valid even for
this category (Case-II) if over estimation up to 50 % could be tolerated.
5. Evaluated doses in all the above cases should be rounded off to the nearest multiple of 0.05
in the dose report.
APPENDIX – P
ANSI Criteria for Testing the Performance of Personnel Dosemeters
American National Standard Institute (ANSI), provides a procedure for testing the
performance of dosimetry system. These procedures are periodically reviewed(1,2) by ANSI and the
latest revision has been carried out in 2001(3). According to ANSI performance criteria, the
definition of some parameters which need to be evaluated are given below
Performance quotient is the relative difference of the personal dose equivalent reported by
the TLD unit from the delivered personal dose equivalent which for the ith dosimeter is
where, Hi is the personal dose equivalent assigned by the testing / irradiating laboratory to
irradiated dosimeter and Hi′ is the corresponding personal dose equivalent reported by the test
participant / TLD unit.
The mean value of performance quotient Pi, of a set of dosimeters (P) or Bias (B) is
n
B ≡ P = 1 / n∑ Pi
i =1
Where sum is extended over all n values of Pi. The large magnitude of B indicates the bias
in the data. A positive value of B means the reported dose is higher than the delivered dose, and a
negative means that the reported values are less. IfB is systematically too high or too low for all
the test categories, it may indicate calibration or algorithm error.
The standard deviation S (σp ) of the values of performance quotient Pi, is
1/ 2
n
(2
)
S ≡ σ p = ∑ Pi − P / (n − 1)
i =1
Where the sum is extended over all n values of Pi for a particular test in a given radiation
category or test category. If the values of S (σp) are large, it indicates a lack of precision or large
random scatter.
Performance in a given radiation categoy is considered to be acceptable if the specified
Tolerance Level (L) is not exceeded by the sum of the absolute value bias B and the standard
deviation S i. e.
| B| + S ≤ L
Tolerence level L, is a quantity which collectively reflects the overall uncertainty in the
calibration, measurement and dose evaluation procedure as well as it takes into account the
recommendations of Competent Authorities in the field of Radiation Protection.
According to the latest criteria, the tolerance level (L) value has been brought down from
0.5 to 0.4. Also, no separate limit has been kept for the bias B or standard deviation S. A
performance quotient limit PQL (commonly called 10% rule) which is consistent with the
current practices of International Organization for Standardization (ISO) has been adopted by
ANSI. This means that regardless the value of the tolerance level L, not more than 10% of the
tested dosimeters for any test category should have absolute value of performance quotient |Pi|
greater than the value of permitted tolerance level (0.4). With fifteen dosimeters tested in any
category, if two or more have Pigreater than 0.4, the results of the test should be considered
as fail i.e. unacceptable and unsatisfactory.
The revision by ANSI became inevitable because it was noted that as per the previous test
criteria, any test category could be passed even if 25% of the test dosimeters (3 or 4 of 15)
exhibited Pi values more than the permitted value. The adoption of trumpet curve criteria by
us in addition to the ANSI criteria was however prevented such situation in our QA programme.
(a) (b)
Fig. 5 Acceptable upper and lower limits for the ratio of measured dose / conventional
true dose as a function of dose: (a) for Hp (10); and (b) for Hp (0.07). (Broken lines:
monthly monitoring periods; solid lines bi-monthly monitoring periods.)
The upper limit of the permissible accuracy limit ( i.e. upper part of the trumpet curve) is
given by
Hul = 1.5 [1+ H0 /(2H0+H1)]
Where H1 is conventional true dose (delivered dose) and H0 is the lowest measurable dose (i.e.
recording level). Recording level H0 can be calculated from the formula 1 mSv x Frequency of
service/ 12. For monitoring period of one month the recording level H0 is not lower than 1 mSv/12
= 0.085 mSv according to ICRP 60/75
APPENDIX - R
Procedure for External Quality Assurance Check
Dose evaluation by
Despatch of performance the TLD unit and
status report & submission of
recommendations to all results to BARC
the TLD units and
concerned authorities
Analysis of the
results at BARC
APPENDIX – S
Procedure for Blind Testing
Assignment of fake
names of users of TLD
badge by the HPU
Corrective measures,
if any
APPENDIX - T
Procedure for Spot Testing of Accredited Laboratory
TLD CARDS:
3. ANI Associates
302, Daffodil, Tata Colony
Sainath Gardens, Mulund (East)
Mumbai- 400 081
Phone: 022-25607484
TLD CASSETTE:
ANNEALING OVEN: