Del Medical EV800 Table
Del Medical EV800 Table
Contents
Safety
EV-800 Introduction
Installation, Operation,
& Service Manual Operation
Periodic
Maintenance
Maintenance
Component
Replacement
Trouble-
Shooting
Electrical
Schematics
Illustrated
Parts List
P/N 8000-EV800-D
Notes
Revision: A, March 6, 2012
Copyright 2012, Del Medical Inc. All rights reserved.
This document is the property of Del Medical Inc. and contains confidential and proprietary
information owned by Del Medical Inc. Any unauthorized copying, use or disclosure of it without the
prior written permission of Del Medical Inc. is strictly prohibited.
Record of Revisions
Revision History . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ii–1
List of Affected Pages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ii–1
Chapter 1 — Introduction
1.1 Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–1
1.2 Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–2
1.3 Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–3
1.4 Minimum Space Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–5
1.5 Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–6
1.6 Abbreviations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–7
1.7 Optional - Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–8
Chapter 2 — Installation
2.1 Installation Instructions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–1
PBL Option Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–7
Ion Chamber Option Installation . . . . . . . . . . . . . . . . . . . . . . 2–11
Grid Installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–16
Testing Collision Avoidance Switches . . . . . . . . . . . . . . . . . . 2–29
Cassette Tray Insertion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–31
i
EV800 Elevating Table Installation, Operation, and Service Manual
Chapter 3 — Operation
3.1 Safety Precautions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–1
3.2 Operation Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–4
3.3 Controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–5
Down Pedal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–5
Up Pedal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–6
Table Top Lock Release Pedals . . . . . . . . . . . . . . . . . . . . . . . . . 3–6
Bucky Carriage Lock Release . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–6
Power Indicator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–6
3.4 Cassette Tray Removal. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–7
3.5 Emergency Stop Button (For Service Technicians Only) . . . . . . . 3–8
Chapter 5 — Maintenance
5.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5–1
5.2 Removing Front Panels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5–2
5.3 Replacing Fuses F1, FL1, & FL2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5–4
5.4 Drive Motor Override . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5–5
5.5 Adjusting Table Elevation Mid-Point Height . . . . . . . . . . . . . . . . 5–10
5.6 Testing & Adjusting Collision Avoidance Switches . . . . . . . . . . 5–12
Testing Switches. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5–12
Adjusting Switches . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5–13
5.7 Changing Elevation Restart Mode . . . . . . . . . . . . . . . . . . . . . . . . . 5–17
ii
EV800 Elevating Table Installation, Operation, and Service Manual
Chapter 7 — Troubleshooting
7.1 Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–1
7.2 Troubleshooting Index. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–2
7.3 Troubleshooting Charts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–3
7.4 Switch Location. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–10
7.4.1 Treadle Switch Location/Layout . . . . . . . . . . . . . . . . . . . . . 7–10
7.4.2 Limit Switch Location/ Layout . . . . . . . . . . . . . . . . . . . . . . 7–11
7.4.3 Bucky Lock Release Switch Location . . . . . . . . . . . . . . . . . . 7–12
7.4.4 Various Switch and Lock Locations . . . . . . . . . . . . . . . . . . . 7–13
7.4.5 Bucky Coil Locks. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–14
iii
EV800 Elevating Table Installation, Operation, and Service Manual
Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .N-1
iv
Safety Information
i
Introduction
The policy of Del Medical, Inc. is to manufacture X-ray equipment that
meet high standards of performance and reliability. We enforce strict
quality control techniques to eliminate the potential for defects and
hazards in our products.
The intended use of this equipment is to provide an adjustable elevating
platform for positioning of a patient with respect to an X-ray source for the
purpose of acquiring X-ray images of the desired parts of a patient’s
anatomy. Use of this equipment in any other fashion may lead to serious
personal injury.
The safety guidelines provided in this section of the manual are intended
to educate the operator on all safety issues in order to operate and
maintain the table in a safe manner.
Introduction i-1
EV800 Elevating Table Installation, Operation, & Service Manual
Statement of Liability
To prevent excess radiation exposure to the patient and the operator from
either primary or secondary radiation, this table must be operated and
serviced by trained personnel who are familiar with the safety precautions
required. While this table has been designed for safe operation, improper
operation or carelessness may result in serious injury or damage to
equipment. The manufacturer or its agents and representatives assume no
responsibility for the following:
1 Injury or danger to any person from X-ray exposure.
2 Overexposure due to poor technique selection.
3 Injury or danger from improper use of the elevator function.
4 Problems or hazards resulting from failure to maintain the equipment
as specified in the maintenance chapters.
5 Equipment which has been tampered with or modified. Del Medical,
Inc. is not liable for any damage or injury arising from failure to
follow the instructions and procedures provided within the manuals
or associated informational material, or from user failure to use
caution when installing, operating, adjusting, or servicing this
equipment. Del Medical, Inc. is not liable for damage or injury arising
from the use of this product for any other use than that intended by
the manufacturer.
Definitions
The table below defines the meaning of various symbols found on device
labels on the machine.
Definitions i-3
EV800 Elevating Table Installation, Operation, & Service Manual
Warning Statements:
• Are used to indicate hazards or unsafe practices which COULD result
in severe personal injury or death.
• Appear in bold type.
• Have a triangular symbol with an exclamation point above the text.
• Are preceded by the word Warning.
• Are always found before the step or piece of information to which
they refer.
• Look like the following example:
Warning
This text will describe special safety precautions to follow in order
to avoid unsafe practices that COULD result in severe personal
injury or death.
Caution Statements:
• Are used to indicate hazards or unsafe practices which could result in
minor personal injury or product/property damage.
• Appear in bold type.
• Have a triangular symbol with an exclamation point above the text.
• Are preceded by the word Caution.
• Are always found before the step or piece of information to which
they refer.
• Look like the following example:
Caution
This text will describe special safety precautions to follow in order
to avoid unsafe practices that could result in personal injury or
product/property damage.
Warning
Turn off electrical power to the table at the power source before
servicing the table. Also, make sure that the power source is
locked out and tagged “Table Being Serviced” before servicing the
table; you could get seriously shocked or burned if you do not.
Warning
Do not place any objects within the vertical motion area above or
below table. Objects in such areas can present crush hazards for
the patient or operator and can also damage the table. See shaded
areas in figure below.
Warning
Make sure the operator or the patient does not touch the table’s
unlock pedals when the patient is getting on or off table. This may
cause the patient to fall off the table and cause serious injury. See
figure below.
Warning
Do not operate the table in an explosive atmosphere (such as
anesthetic gas). Doing so can cause an explosion or a fire hazard
which could cause serious injury.
Warning
All of the movable assemblies and parts of this equipment should
be operated with care and routinely inspected in accordance with
the manufacturer’s recommendations contained in this manual.
Warning
This table is intended to be used as part of a system for the
intended generation of X-rays for medical diagnosis.
Warning
The Emergency Stop button DOES NOT turn off all power to the
table. There may still be lethal voltages present on the main
chassis and throughout the table when this button is turned off.
These voltages come from sources outside the table (i.e. generator,
wallstand, etc.) that are connected to the table.
Warning
Do not work on the table before the capacitor’s LED is off; you can
be severely shocked if you do.
Identification Labels
The EV800 components have manufacturing and certification information
affixed. The manufacturing label contains:
• The full name and address of the manufacturer of the component
• The place, month, and year of manufacture
• The model number and serial number of the component
The certification label also states that the component complies with either
“21CFR, Subchapter J”or the applicable DHHS standards under the
Radiation Control for Health and Safety Act of 1968 (or its equivalent).
A label may combine both manufacturing and certification information.
The underside of the removable float-top also has a label with matching
serial number and date of manufacturing.
EV800 Label
The location of the EV800 identification label is shown in Figure i-1.
Product Life
The table has an estimated life of 10 years from the point of purchase. This
may vary depending on the (1) product use, (2) product maintenance, and
(3) environmental conditions. When it has been determined that the table
has completed its useful service life, local regulations must be complied
with in regards to the proper disposal of the possible hazardous materials
used in the manufacturing of the table.
The table contains electronic components, such as: soldered (may contain
lead) circuit boards, tinned copper wire, transistors, diodes, resistors,
capacitors, etc. The cabinet is made from steel and/or aluminum.
NOTE
DO NOT INCINERATE THIS PRODUCT. DO NOT PUT IN TRASH
THAT IS DISPOSED OF IN LANDFILLS. DISPOSAL OF THIS
PRODUCT MUST ADHERE TO LOCAL ORDANANCES AND
REGULATIONS. LOCAL AUTORITIES MUST BE CONTACTED FOR
PROPER DISPOSAL AND/OR RECYCLING OPTIONS FOR ALL OF
THE COMPONENTS LISTED ABOVE.
A 1-2-2012 Original
All A
Introduction 1-1
EV800 Elevating Table Installation, Operation, & Service Manual
1.2 Description
1-2 Description
EV800 Elevating Table Installation, Operation, & Service Manual
1.3 Dimensions
Dimensions 1-3
EV800 Elevating Table Installation, Operation, & Service Manual
Caution
Do not drill equipment mounting holes in pre-tensioned or post
tensioned concrete floors before determining the location of the
tensioned wire ropes. Consult with customer or responsible
project engineer to locate and avoid drilling through wire rope.
Cutting through tensioning cables can cause severe structural
damage.
1-4 Dimensions
EV800 Elevating Table Installation, Operation, & Service Manual
6” (15 cm)
9” (23 cm)
Front
Figure 1-3. Plan View (Top View) of Minimum Space Requirements
1.5 Specifications
Specifications
Compatibility The EV800 Elevating Table is compatible with a wide variety of
generators, wallstands, and tubestands. It is intended to be used
in a stationary diagnostic X-ray configuration.
Voltage 120VAC or 230/240 VAC
Current 10 Amps, Momentary 15 sec ON, 85 sec OFF
Frequency 50/60 Hz Single Phase
Incoming Power Line 3 #16 AWG (1.5mm2) for 50ft (15.25M) or Less
External Heat Generation 300 Btu (88 watts)
Fuse Type Line (FL1) & (FL2) 10 Amp
Motor (F1) 10 A
Classification Class 1 Type B
1-6 Specifications
EV800 Elevating Table Installation, Operation, & Service Manual
1.6 Abbreviations
% Percent
AWG American Wire Gauge
Btu British Thermal Unit
°C Degree Celsius
CE Communautés Européennes
cm Centimeter
C.R.S. Cold Rolled Steel
°F Degree Fahrenheit
ga Gauge
hPa Hecto Pascal
inHg Inches Mercury
Kg Kilogram
Lb Pound
M Meter
max. Maximum
min. Minimum
mm Millimeter
PBL Positive Beam Limitation
Sq/Ft Square Foot
Sq/M Square Meter
UL Underwriters Laboratories
Abbreviations 1-7
EV800 Elevating Table Installation, Operation, & Service Manual
1.7 Optional-Accessories
See page 9-36 for Part ID for accessories.
1-8 Optional-Accessories
Installation
2
2.1 Installation Instructions
Tools Required:
• Diagonal cutters (side cutters)
• Medium phillips screwdriver
• Pallet hoist
• Power drill and masonry bit (size determined by installer)
• Scale for measuring downward force of up to 120 lbs (53 Kg.)
• Set of hex wrenches
• Set of nut drivers
• Set of open-end wrenches
• Small flat-tip screwdriver
• Utility knife
Your shipment will arrive in two boxes as shown below. One box will
contain the table top and the other box will contain the base. Save the
boxes until the table is inspected for shipping damage and is up and
operating.
1 Move the base pallet to the approximate position where the base will
be installed.
2 Use diagonal cutters to cut the steel straps (1 in Figure 2-2). Remove
the straps and the top cover (2).
3 Remove the front cover (1 in Figure 2-3) from the box and unwrap it.
4 Lift off outer box shell (2).
5 Remove the shrink wrapping (3) from around the base.
3 1 2
10 Press and hold the Up pedal (1 in Figure 2-5) until the base rises to its
mid-level position where it will stop. Press and hold the Up pedal
again until the base rises to its highest position where it will stop.
Up Pedal
Warning
Unplug the base’s power cord before performing the following
steps. You could get seriously injured if you do not.
12 Unscrew the two lower panel clamp screws (1 in Figure 2-6), then
remove the lower panhead screws (2).
13 Lift the lower panel (3) up slightly and then pull it out. The bottom of
the lower panel is connected to the table by slots (4) shown below.
1
2
3
Caution
Use extreme care when moving the base off the pallet and not to
damage the foot pedals. Do not pry on the pedals or let the pedals
rest on the edge of the pallet when moving the base.
14 Lift the base off the pallet and place the base in its final resting
position.
Warning
You must use four 1/2’-13 x 3” (M12 X 75mm) bolts and appropriate
anchors and washers to mount table base to floor. Use of any size
smaller may cause the table base to break free from the floor under
extreme load and leverage conditions. This condition could cause
severe injury to the patient and/or user.
Caution
It is up to the installer and the customer to determine the best
method for mounting the base to the floor. Before mounting the
base, consult with the building’s maintenance supervisor about
drilling holes in the floor. Make sure that there are no hazards
under the floor such as pipes, conduits, or structural cables that
can be damaged by drilling holes in the floor.
Caution
Do not mount the table base to the floor until the tubestand or
overhead tube crane that accompanies it has been mounted and
aligned correctly; you may have to reposition the table if you do.
15 Mount the base to floor using holes through bottom of the base. There
are several sets of holes that allow some adjustability in positioning
the base. Choose the set of holes that best suits your needs, and then
make sure that base is level after mounting. If necessary, use shims
from accessory kit to aid leveling.
1 2
Jumpers
20 Install the PBL plate assembly (1 in Figure 2-10) on the bucky and
secure it in place with four mounting screws (2).
21 Unscrew the two acorn nuts (3) and remove the cover (4).
2 (x4)
4
1 3
Figure 2-10. PBL Plate Assembly
25 Secure the internal cable with p-clamp & screw (1 & 2 in Figure 2-12).
26 Reattach the cover (3) with two nuts (4).
1&2 4 3
Figure 2-12. Clamp & Cover
29 Unscrew the four cover screws (2 on each side) (1 in Figure 2-14) and
remove the top cover (2).
1 3
3
2 (x2)
32 Unscrew the four cover screws (1 in Figure 2-16) and remove the
cover (2).
1 (x4) 2
Figure 2-16. Ion Chamber Cover
3(x2)
4
2 2
1
35 Connect the internal ion chamber cable (1 in Figure 2-19) to the ion
chamber and secure with the two mounting screws (2).
36 Install the clamp (3) over the cable and secure with the two nuts (4).
37 Insert the grommet (5) over the cable.
5 4 3 2 1
38 Reinstall the cover (1 in Figure 2-20) and secure with the four screws
(2). Make sure that the chamber mates with the grommet installed in
the previous step when reattaching the cover.
2 (x4) 1
Grid Installation
39 If not already done so, unscrew the four top cover screws (2 on each
side)(1 in Figure 2-21) then remove the top cover (2).
40 Unscrew the four clamp screws (1 in Figure 2-22) and remove the
clamps (2).
2 (x4)
1 (x4)
Caution
Be careful not to over-tighten the grid clamp screws in the
following step or you may crack the grid.
42 Install the grid as shown below and secure in place with clamps and
screws.
43 Reattach the bucky top cover (1 in Figure 2-25) and secure in place
with four screws (2).
1
Figure 2-26. Bottom Cable Port
Warning
Make sure power to the table and all peripheral equipment
(generator, tubestand, etc.) is turned off before performing the
following steps; you can be seriously injured if you do not.
46 Disconnect the power cord from the main chassis shown below and
remove the power cord.
Wiring Safety Information
• X-ray apparatus shall be grounded and powered in accordance
with the National and Local Electrical Codes.
• During installation, make sure that all protective ground wire
connections are completed before the equipment is turned on.
• It is required that a separate 10-amp (min.) line disconnect and
current limiting device be provided for table power.
• The protective ground wires between system components are
connected as shown in the wiring diagram in service manual.
• Regulations of professional associations concerning safety and
accident prevention must be observed.
• When installation and maintenance is required, use only
personnel trained in the mechanical and electrical operation of
this equipment.
Caution
If you are connecting 120 nominal input voltage to this table, then
a 120 VAC needs to be connected to the line input. JP1, JP2, and
JP3 on the main board must have the jumper plugs installed
between pins 2-3.
Caution
If you are connecting 240 nominal input voltage to this table, then
a 240 VAC needs to be connected to the line input. JP1, JP2, and
JP3 on the main board must have the jumper plugs installed
between pins 1-2.
Warning
Make sure power to the table is turned off at disconnect switch
before performing the following steps.
Also, make sure power to the bucky is off by turning off the
generator before performing the following steps.
49 Pull the port cover (1 in Figure 2-29) out, flip it over, and reinsert it.
3
2
1
Figure 2-31. Upper Cover Mounting Screws
53 Hook the top cover (1 in Figure 2-32 below) onto the upper rail (3 in
Figure 2-31 above) and secure in place with the two screws (2 in
Figure 2-32 below).
2
Figure 2-32. Upper and Lower Panel Screws
Note: Steps 58-68 are for rooms that are too small to allow the table top to
be slid on from the end.
56 Remove guide rollers (1 in Figure 2-34) and bearings (2) from front of
table.
2 1 2 1 2
Figure 2-34. Guide Roller and Bearing Removal
57 Slide the table to the right and reinstall the left guide roller (1 in
Figure 2-35) and the left bearing assembly (2).
2 1
1
2
60 While holding the eccentric nut (1 in Figure 2-37) still, tighten the
bearing nut (2).
61 Repeat steps 60-63 for right side guide roller and bearing.
62 Have your partner slide the table half way off and hold it there as
shown below.
63 Install the middle bearing (1 in Figure 2-38) and adjust it as you did
for the others.
64 Slide the table top back and forth to ensure smoothness of motion.
Readjust bearing gap if required.
65 Center the table top on the table base and go to the next step.
Warning
The bumpers must be installed. If they are not, the table top may
slide off of its base during use and severely injure the patient or
operator.
67 Mount the bumper (1 in Figure 2-40) with two screws (2) in all four
corners of table.
1
2
Figure 2-40. Bumper Mounting
Caution
Do not exceed 110 lbs (49 Kg) of force in the following step or the
table may get damaged. If the force appears that it will exceed 110
lbs (49 Kg), release the down pedal immediately.
73 Press the Down pedal and lower the table until the table top contacts
the scale and the table stops by itself. The force on the scale when the
table stops should be between 90 to 110 lbs (40 to 49 Kg.). If it is not or
if the force exceeds 110lbs (49 Kg), adjust the collision avoidance
switch according to Section 5.6 “Testing & Adjusting Collision
Avoidance Switches” on page 5-12. If it is, go to the next step.
74 Raise the table back up.
75 Move the table top to its rightmost position.
Scale
78 Remove the shipping tape (1 in Figure 2-43) from the cassette tray.
79 Pull the pin (2) back and remove the cassette blocker (3).
80 Place the guide in the desired slots (4).
2
1
4 (Typical)
Figure 2-43. Cassette Tray Preparation
81 Insert the cassette tray into the bucky as shown in the figure below.
Installation is complete.
Warning
Do not place any objects within the vertical motion area above or
below table. Objects in such areas can present crush hazards (1 in
Figure 3-1) for the patient or operator, and can also damage the
table. See areas shown shaded in Figure below.
Warning
Make sure that the operator or the patient does not touch the table
top lock release pedals when the patient is getting on or off table.
This may cause the patient to fall off the table and cause serious
injury. See figure below.
Warning
No foreign objects which can attenuate or scatter the X-ray beam
are allowed between the X-ray tube and the table top during
exposure. Failure to follow this may result in serious injury.
Warning
This table is intended to be used as part of a system for the
intended generation of X-rays for medical diagnosis.
3.3 Controls
This section describes the controls of the table. Figure 3-3 below shows the
controls of the table.
Table Top
Bucky
Carriage
Power
Indicator
Bucky
Carriage
Lock Release
Up Pedal
Down Pedal
Table Top Lock Release Pedal
Down Pedal
Press the Down pedal to lower the table. The table can be lowered to
within 22” (56cm) of the floor.
The table moves downward at a constant speed when you press and hold
the Down pedal. When the table reaches its mid-point level, it will
automatically stop. Restarting the downward motion of the table depends
on how your table was set up.
The default setup is automatic restart. In this mode, if you do not release
the pedal after the table stops, it will automatically restart after a few
seconds. The table will then continue downward until it reaches its lowest
position and automatically stops.
The optional setup is manual restart. In this mode, you release the pedal
and then press and hold the pedal again to restart motion. The table will
then continue downward until it reaches its lowest position and
automatically stops.
The procedure for changing from one mode to another is covered in
Section 5.7 and should only be performed by a service technician.
Up Pedal
Press the Up pedal to raise the table. The table can be raised to 34” (86 cm)
above the floor.
The table moves up at a constant speed when you press and hold the Up
pedal. When the table reaches its mid-point level, it will automatically
stop. Restarting the upward motion of the table depends on how your
table was set up.
The default setup is automatic restart. In this mode, if you do not release
the pedal after the table stops, it will automatically restart after a few
seconds. The table will then continue upward until it reaches its highest
position and automatically stops.
The optional setup is manual restart. In this mode, you release the pedal
and then press and hold the pedal again to restart motion. The table will
then continue upward until it reaches its highest position and
automatically stops.
The procedure for changing from one mode to another is covered in
Section 5.7 and should only be performed by a service technician.
Power Indicator
Indicates that power to the table is turned on and that the table is ready to
operate.
Warning
The Emergency Stop button DOES NOT turn off all power to the
table. There may still be lethal voltages present on the main
chassis and throughout the table when this button is turned off.
These voltages come from sources outside the table (i.e. generator,
wallstand, etc.) that are connected to the table.
Clean External Surfaces Every Week or as Required “4.2 Cleaning External Sur-
faces” on page 4-2
Warning
Turn off all electrical power to the table and all its peripheral
equipment (generator, tubestand, etc.) at the power sources before
servicing the table. Also, make sure that the power sources are
locked out and tagged “Equipment Being Serviced” before
servicing the table. Some components inside the table have power
sources other than that of the table, which is why all peripheral
equipment must be turned off; you could get seriously injured if
you not.
Warning
This equipment is NOT classified as anaesthetic-proof and may
ignite inflammable anesthetics. Flammable agents used for skin
cleaning or disinfecting may also produce an explosion hazard.
• Ensure the power has been disconnected and that the emergency
switches have been activated before starting any cleaning operation.
• Ensure no liquid gets into the unit.
• Do not immerse the equipment, including any components or
accessories, in liquid.
• Do not autoclave the equipment, including any component or
accessories.
• Do not use water. Water can short-circuit the electrical installation and
cause corrosion to mechanical parts.
• Do not use acid or abrasive products.
• Use only a dry cloth to clean chrome-plated parts.
Warning
Turn off all electrical power to the table and all its peripheral
equipment (generator, tubestand, etc.) at the power sources before
servicing the table. Also, make sure that the power sources are
locked out and tagged “Equipment Being Serviced” before
servicing the table. Some components inside the table have power
sources other than that of the table, which is why all peripheral
equipment must be turned off; you could get seriously injured if
you not.
1
Figure 4-1. Upper Panel Screws
4 Unscrew the two lower panel clamps screws (1 in Figure 4-2), then
remove the upward panhead screws (2) .
5 Lift the lower panel (3) up slightly and then pull it out. The bottom of
the lower panel is connected to the table by slots (4) shown below.
1
1
1
Tools Required:
• Set of hex wrenches
• Scale (ruler) at least 6” (150mm) long
• Two open-end 9/16” wrenches
1 Run table to its uppermost position
Warning
Turn off all electrical power to the table and all its peripheral
equipment (generator, tubestand, etc.) at the power sources before
servicing the table. Also, make sure that the power sources are
locked out and tagged “Equipment Being Serviced” before
servicing the table. Some components inside the table have power
sources other than that of the table, which is why all peripheral
equipment must be turned off; you could get seriously injured if
you not.
1
Figure 4-4. Upper Panel Screws
5 Unscrew the two lower panel clamps screws (1 in Figure 4-2), then
remove the lower panhead screws (2).
6 Lift the lower panel (3) up slightly and then pull it out. The bottom of
the lower panel is connected to the table by slots (4) shown below.
1
7 With the chain now tight between the three sprockets, remove the
remaining slack by tightening the chain tension bolt. When properly
tensioned, the chain should deflect approximately 3/16” at point
shown with moderate finger pressure. If tension is OK, reassemble
table. If tension is more than 3/16”, tighten chain according to the
steps that follow.
1
Figure 4-6. Check Chain Tension
1
Step 13 - tension added here
9 With the chain now tight between the three sprockets, remove the
remaining slack by tightening the chain tension bolt. When properly
tensioned, the chain should deflect approximately 3/16” at the point
shown with moderate finger pressure.
1
Figure 4-8. Braces in place under the upper frame gear racks
10 Reassemble table.
Warning
Turn off all electrical power to the table and all its peripheral
equipment (generator, tubestand, etc.) at the power sources before
servicing the table. Also, make sure that the power sources are
locked out and tagged “Equipment Being Serviced” before
servicing the table. Some components inside the table have power
sources other than that of the table, which is why all peripheral
equipment must be turned off; you could get seriously injured if
you not.
1
Figure 5-1. Upper Panel Screws
4 Unscrew the two lower panel clamps screws (1 in Figure 5-2), then
remove the lower panhead screws (2).
5 Lift the lower panel (3) up slightly and then pull it out. The bottom of
the lower panel is connected to the table by slots (4) shown below.
1
Warning
Turn off all electrical power to the table and all its peripheral
equipment (generator, tubestand, etc.) at the power sources before
servicing the table. Also, make sure that the power sources are
locked out and tagged “Equipment Being Serviced” before
servicing the table. Some components inside the table have power
sources other than that of the table, which is why all peripheral
equipment must be turned off; you could get seriously injured if
you not. Wait five minutes after turning power off to work on the
table.
Caution
Replace fuses only with same type and rating of fuse or machine
may get damaged.
2 Access to the fuses is on the back to the table’s base. Unscrew the fuse
cap of the faulty fuse and replace the fuse. See figure below for
location and fuse rating.
3 Reverse steps to reassemble.
Warning
Turn off all electrical power to the table and all its peripheral
equipment (generator, tubestand, etc.) at the power sources before
servicing the table. Also, make sure that the power sources are
locked out and tagged “Equipment Being Serviced” before
servicing the table. Some components inside the table have power
sources other than that of the table, which is why all peripheral
equipment must be turned off; you could get seriously injured if
you not. Wait five minutes after turning power off to work on
table.
Caution
Replace the fuses only with the same type and rating of the fuse or
the machine may get damaged.
2 If room dimensions allow, remove the stop bumpers from the end of
table top assembly and carefully remove the table top from the table
frame.
3 If unable to remove the table top it will be necessary to remove the
stop bumpers and slide the top to the head of the table.
6 It is necessary to remove the grease cap to access the rear hub of the
input shaft. Using a small, flat screwdriver and hammer, carefully pry
off the grease cap (1 in Figure 5-6) and set it aside.
7 Carefully wipe off excess grease from the end of the input shaft (1 in
Figure 5-7).
9 Using pliers or vise grips, hold the overrride coupling while securely
tightening the black clamping screw (1 in Fgure 5-9).
Note: A light tap with a hammer can be used to seat the coupling flat on the input
shaft.
10 Using a 9/16” socket and ratchet (1 in Figure 5-10), the table can now
by manually raised or lowered for service.
Warning
Override coupling must be removed from the gearbox prior to any
attempt to return power and operate the table.
Warning
Turn off all electrical power to the table and all its peripheral
equipment (generator, tubestand, etc.) at the power sources before
servicing the table. Also, make sure that the power sources are
locked out and tagged “Equipment Being Serviced” before
servicing the table. Some components inside the table have power
sources other than that of the table, which is why all peripheral
equipment must be turned off; you could get seriously injured if
you not. Wait five minutes after turning power off to work on
table.
2 Turn off all power to the table and wait five minutes for table’s
capacitors to discharge.
3 Remove front panels according to Section 5.2 Removing “Front
Panels” on page 5-2.
4 Loosen the two clamp screws (1 in Figure 5-12), move the plate (2) to
the desired position, and retighten the screws.
5 Test the new mid-table pause position. Readjust as necessary.
6 Reassemble table.
Testing Switches
1 Move table top to its leftmost position.
2 Turn on power to the table.
3 Run table to its highest position.
4 Place a stool or other sturdy support under the left end of the table top
as shown in Figure 5-13 on the next page.
5 Place a weight scale on the stool.
Caution
Do not exceed 110 lbs (49 Kg) of force in the following step or the
table may get damaged. If the force appears that it will exceed 110
lbs (49 Kg), release the down pedal immediately.
6 Press the down pedal and lower the table until the table top contacts
the scale and the table stops by itself. The force on the scale when the
table stops should be between 90 to 110 lbs (40 to 49 Kg.). If it is not or
if the force exceeds 110lbs (49 Kg), adjust the collision avoidance
switch according to step 10. If it is, go to the next step.
7 Raise table back up.
8 Move table top to its rightmost position.
9 Repeat steps 4-6 for right end of table top.
Scale
Adjusting Switches
10 Turn off all power to the table and wait five minutes for the table’s
capacitors to discharge.
Warning
Turn off all electrical power to the table and all its peripheral
equipment (generator, tubestand, etc.) at the power sources before
servicing the table. Also, make sure that the power sources are
locked out and tagged “Equipment Being Serviced” before
servicing the table. Some components inside the table have power
sources other than that of the table, which is why all peripheral
equipment must be turned off; you could get seriously injured if
you not. Wait five minutes after turning power off to work on
table.
12 Slide the table top (1 in Figure 5-15) and remove cover (2) to expose
switch assembly.
Caution
Be careful not to adjust the pressure so much that the switch’s arm
(3) rests on the switch (4) or the switch may get damaged.
15 When proper pressure is achieved, tighten the jam nut and repeat the
procedure for the switch on other side of the table.
Access
1 2 Hole 3 4
Warning
Turn off all electrical power to the table and all its peripheral
equipment (generator, tubestand, etc.) at the power sources before
servicing the table. Also, make sure that the power sources are
locked out and tagged “Equipment Being Serviced” before
servicing the table. Some components inside the table have power
sources other than that of the table, which is why all peripheral
equipment must be turned off; you could get seriously injured if
you not. Wait five minutes after turning power off to work on
table.
4 Locate Restart Mode Pot and Jumpers on the motor control board
using figure below. Then go to the next step.
Main Chassis
Restart Mode
Pot & Jumpers
DETENT
delay of JP2A
restart P2A
= Jumper
JP1A
DETENT
Turning pot in JP2A
manual mode P2A
will have no
effect.
Manual Restart Mode
Jumper Positions
Figure 5-19. Elevation Adjustment Pots
Warning
Turn off all electrical power to the table and all its peripheral
equipment (generator, tubestand, etc.) at the power sources before
servicing the table. Also, make sure that the power sources are
locked out and tagged “Equipment Being Serviced” before
servicing the table. Some components inside the table have power
sources other than that of the table, which is why all peripheral
equipment must be turned off; you could get seriously injured if
you do not. Wait five minutes after turning power off to work on
table.
1
Figure 6-1. Upper Panel Screws
4 Unscrew the two lower panel clamps screws (1 in Figure 6-2), then
remove the lower panhead screws (2).
5 Lift the lower panel (3) up slightly and then pull it out. The bottom of
the lower panel is connected to the table by slots (4) shown below.
1
Warning
Turn off all electrical power to the table and all its peripheral
equipment (generator, tubestand, etc.) at the power sources before
servicing the table. Also, make sure that the power sources are
locked out and tagged “Equipment Being Serviced” before
servicing the table. Some components inside the table have power
sources other than that of the table, which is why all peripheral
equipment must be turned off; you could get seriously injured if
you not. Wait five minutes after turning power off to work on
table.
2 Turn off all power to the table and wait five minutes for the table’s
capacitors to discharge.
3 Unplug the connectors (1 in Figure 6-3).
4 Unscrew all mounting screws (2) and remove the board.
5 Reverse steps to reassemble.
2 (Typ) 1 (Typ)
3
Figure 6-3. Front Chassis Screws
Warning
Turn off all electrical power to the table and all its peripheral
equipment (generator, tubestand, etc.) at the power sources before
servicing the table. Also, make sure that the power sources are
locked out and tagged “Equipment Being Serviced” before
servicing the table. Some components inside the table have power
sources other than that of the table, which is why all peripheral
equipment must be turned off; you could get seriously injured if
you not. Wait five minutes after turning power off to work on
table.
2 Turn off all power to the table and wait five minutes for the table’s
capacitors to discharge.
3 Unplug the connector (1 in Figure 6-4).
4 Cut cable ties (2).
5 Unscrew the ground nut (3).
2 1 3
Figure 6-4. Foot Treadle P-Clamps & Connector
6 Unscrew the two foot treadle mounting screws, located inside the
table as shown (1 in Figure 6-5), and remove the treadle assembly (2).
7 Reverse steps to install the treadle.
1 1
2
Note: Do not use an off-the-shelf replacement motor. The motor in your table has
been specially modified at the Del factory for this application. You must use part
#112-5387G1 as a replacement motor.
Warning
Turn off all electrical power to the table and all its peripheral
equipment (generator, tubestand, etc.) at the power sources before
servicing the table. Also, make sure that the power sources are
locked out and tagged “Equipment Being Serviced” before
servicing the table. Some components inside the table have power
sources other than that of the table, which is why all peripheral
equipment must be turned off; you could get seriously injured if
you not. Wait five minutes after turning power off to work on
table.
2 Turn off all power to the table and wait five minutes for the table’s
capacitors to discharge.
3 Unplug the connector (1 in Figure 6-6).
1
Figure 6-6. Motor Connector
1 (x4)
2 3
8 Install the coupling to the new motor assembly. Be sure to space the
coupling according to the mount distance measured in step #6.
9 Reverse the steps to install the motor. Use care when realigning the
motor/gearbox coupling.
Warning
Turn off all electrical power to the table and all its peripheral
equipment (generator, tubestand, etc.) at the power sources before
servicing the table. Also, make sure that the power sources are
locked out and tagged “Equipment Being Serviced” before
servicing the table. Some components inside the table have power
sources other than that of the table, which is why all peripheral
equipment must be turned off; you could get seriously injured if
you not. Wait five minutes after turning power off to work on
table.
Opposite 3
Side of Switch 2 1
Figure 6-10. Treadle Switch (Typical)
5 Reverse the steps to install switch. Adjust the switch so that its arm (1
in Figure 6-11) engages the switch, but still has a little play left in it so
that it’s not forced against the switch.
Caution
Carefully test the treadle in the circuit. The table may not operate
or may drive uncommanded if not adjusted properly.
Warning
Turn off all electrical power to the table and all its peripheral
equipment (generator, tubestand, etc.) at the power sources before
servicing the table. Also, make sure that the power sources are
locked out and tagged “Equipment Being Serviced” before
servicing the table. Some components inside the table have power
sources other than that of the table, which is why all peripheral
equipment must be turned off; you could get seriously injured if
you not. Wait five minutes after turning power off to work on
table.
3 Carefully slide the table top off of the base far enough to expose the
brakes.
4 Loosen the terminal block screws (1 in Figure 6-13) and slide the lock
wires out of the terminal block.
5 Unscrew the lock screw (2) and remove the lock (3).
6 Reverse the steps to install brake.
Caution
Use care not to damage the locks when sliding the table top back
onto the base.
1 2 3
Height Limit
Switches
Warning
Turn off all electrical power to the table and all its peripheral
equipment (generator, tubestand, etc.) at the power sources before
servicing the table. Also, make sure that the power sources are
locked out and tagged “Equipment Being Serviced” before
servicing the table. Some components inside the table have power
sources other than that of the table, which is why all peripheral
equipment must be turned off; you could get seriously injured if
you not. Wait five minutes after turning power off to work on
table.
2 3
1
Figure 6-15. Limit Switch (Typical)
Warning
Turn off all electrical power to the table and all its peripheral
equipment (generator, tubestand, etc.) at the power sources before
servicing the table. Also, make sure that the power sources are
locked out and tagged “Equipment Being Serviced” before
servicing the table. Some components inside the table have power
sources other than that of the table, which is why all peripheral
equipment must be turned off; you could get seriously injured if
you not. Wait five minutes after turning power off to work on
table.
1
Figure 6-16. Bumper Removal
4 Carefully slide the table top off of the base far enough to expose the
bucky.
6 Loosen the terminal block screws (1 in Figure Figure 6-19) and slide
the lock wires out of the terminal block. The terminal block is located
under the bucky and you can access it through the front of the
machine.
Bucky
Locks
1
Figure 6-19. Terminal Block
7 Unscrew the lock screws (1 in Figure 6-20) and remove the locks. You
will have to lift the bucky (2) up to access the screws.
8 Reverse the steps to install lock.
Caution
Use care not to damage the table locks when sliding the table top
back onto the base.
Warning
Turn off all electrical power to the table and all its peripheral
equipment (generator, tubestand, etc.) at the power sources before
servicing the table. Also, make sure that the power sources are
locked out and tagged “Equipment Being Serviced” before
servicing the table. Some components inside the table have power
sources other than that of the table, which is why all peripheral
equipment must be turned off; you could get seriously injured if
you not. Wait five minutes after turning power off to work on
table.
3 Slide table top (1 in Figure 6-22) remove the cover (2) to expose switch
assembly.
2 5 3 2
4 1
Warning
Turn off all electrical power to the table and all its peripheral
equipment (generator, tubestand, etc.) at the power sources before
servicing the table. Also, make sure that the power sources are
locked out and tagged “Equipment Being Serviced” before
servicing the table. Some components inside the table have power
sources other than that of the table, which is why all peripheral
equipment must be turned off; you could get seriously injured if
you not. Wait five minutes after turning power off to work on
table.
3 Carefully slide the table top off the base far enough to expose the
locks.
4 Unscrew the cover screws and remove the lock cover (1 in Figure 6-26)
from the side of the table where the suspected faulty lock is located
(there are two sets of locks - on left and right side of table).
1
Figure 6-26. Lock Cover (Right Side Shown)
Caution
Use care not to damage the left-right locks when sliding the table
top back onto the base.
2 5 4 3 1 2
Warning
Use extreme caution when operating the table without the
protective covers in place. Objects within the table can present
crush hazards.
Disassembly
Figure 6-28. Braces in place under the upper frame gear racks
5 Reset the emergency stop switch, then slowly lower the table until the
gear racks engage the tops of the braces.
Caution
Use short taps of the foot switch to engage the braces. The braces
will unload the gearbox and drive the chain assembly to allow for
safe removal.
6 Engage the emergency stop switch inside the table and disconnect the
power to the table.
7 Remove tension from the chain by loosening the two socket head cap
screws (1 in Figure 6-29) and backing down the single tension bolt (2).
Carefully lay the slack chain down to the floor pan of table. Be sure to
maintain chain/tooth indexing on the two top sprockets of the drive
shafts.
2
Figure 6-29. Loosen screws to remove chain tension
8 Disconnect the power cable from the motor and table chassis.
Disconnect the ground wire. Remove the four bolts (1 in Figure 6-30)
securing the motor to the gearbox. Remove the motor from table.
3
2
1 (x4)
9 Remove the four bolts (2 in Figure 6-30) securing the gearbox to the
motor drive plate (3). Remove gearbox from table.
Re-assembly
1
Figure 6-31. New motor/gearbox assembly
11 Install the new gearbox to the table and secure the four bolts.
Carefully loop the chain around the drive sprocket (reference
Figure 6-29 for correct path). Do not adjust the chain tensioner at this
time.
12 Verify that the chain is completely engaged on the drive sprocket,
tensioner, and the two drive shaft sprockets. With hand force only,
turn the input shaft of the gearbox clockwise to tighten the chain on
the small drive sprocket and the two drive shaft sprockets. The small
drive sprocket should turn in a counter-clockwise direction when
observed from the chain side of the aluminum motor/gearbox
mounting plate.
Step 12 - tension added here
1
Step 13 - tension added here
13 With the chain now tight between the three sprockets, remove the
remaining slack by tightening the chain tension bolt. When properly
tensioned, the chain should deflect approximately 3/16” at point
shown with moderate finger pressure.
1
Figure 6-32. Check Chain Tension
14 Lock the tensioner in place with the two socket head cap screws and
lock nut on the chain tension bolt.
15 Install the motor onto the gearbox. Position coupling, urethane insert,
and D-hole coupling onto the motor. Visually align the flats of D-hole
coupling with the mating flats on the gearbox shaft and join the motor
to the gearbox. Secure with the four mounting bolts.
16 Reconnect the power cable to the motor and the main chassis, and
then secure grounds.
17 Reconnect the power to table and reset the emergency stop switch.
18 Slowly raise the table to upper stop. Remove 2” x 4” supports. Cycle
the table up and down several times to check for proper operation.
S11
S12
S13
S14
Figure 7-1. Treadle Location
S101
S103
S102
S301
Figure 7-3. Bucky Lock Release Switch Location
Transverse Transverse
Locks Locks
Left Right
L301 L302
To Order by Fax
Fax your order to Del Medical Inc. at 1-800-288-7011. Fax orders can be
sent 24 hours a day, 7 days a week.
If you need additional assistance, please call Del Medical Inc. at 1-800-800-
6006 and speak to one our Customer Service Representatives. Telephone
hours are 8:00 a.m. to 5:00 p.m., Monday through Friday (Central
Standard or Daylight Time).
Up Pedal 202-5160P1
112-5538G2 36”
Table Top
6 785-11-16000011 WASHER, #8 6
1 230-5013P1 BLOCK-STOPPER/BUMPER 4
2 401-0005P1 BUMPER 4
22 764-20-16203111 SCREW 1
9-12
EV800 Elevating Table Installation, Operation, & Service Manual
2 400-5005P1 BRG-RADIAL 2
12MMx28MMXx12MM
2 400-5005P1 BRG-RADIAL 2
12MMx28MMXx12MM
19 423-0007P1 NUT,SELF-LOCKING,10-32 1
9 232-0016P1 STUD-ECCENTRIC 6
3
4,5
9 14
6,7,8
11
2
12,13
3 201-5107P1 Bracket 1
0920-0022B Patient Head Clamps (Must be ordered with part #0920- Pair
0040)
103101M18 10:1 103 Line Grid Replaces 8:1 103 Line Grid 1
103121M18 12:1 103 Line Grid Replaces 8:1 103 Line Grid 1
Notes
n-1
EV800 Elevating Table Installation, Operation, & Service Manual
Notes
n-2
EV800 Elevating Table Installation, Operation, & Service Manual
Notes
n-3
EV800 Elevating Table Installation, Operation, & Service Manual
Notes
n-4