0% found this document useful (0 votes)
365 views2 pages

VS600 Specifications

The document provides specifications for the VS-600 vital signs monitor. It can measure blood oxygen saturation, blood pressure, pulse rate, and temperature. It is portable, has rechargeable batteries, and complies with various medical device standards.

Uploaded by

huntrex
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
365 views2 pages

VS600 Specifications

The document provides specifications for the VS-600 vital signs monitor. It can measure blood oxygen saturation, blood pressure, pulse rate, and temperature. It is portable, has rechargeable batteries, and complies with various medical device standards.

Uploaded by

huntrex
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 2

VS-600

Vital Signs Monitor

Physical Specifications
Weight ≤1.9 kg (with SpO2, NIBP and a battery ) Nellcor SpO2
≤1.7 kg (with NIBP and a battery ) Meet standards of ISO 80601-2-61.
Size 134 mm × 120 mm × 243 mm SpO2 range 0 to 100 %
Screen type Segment display SpO2 resolution 1%
Screen size 90mm × 90mm SpO2 accuracy 70 to 100%: ±2% (adult/pediatric)
70 to 100%: ±3% (neonate)
EMC Compliance Meets the requirements of IEC 60601-1-2. 0% to 69%: Not specified.
Perfusion indicator (PI) No
Classifications
Defibrillation proof (protection against electrical shock) PR from Mindray SpO2
Type CF: SpO2, NIBP and Temp PR range 20 to 254 bpm
Water proof IPX1 PR accuracy ± 3 bpm (without motion)
Degree of mobility Portable
PR from Masimo SpO2
Patient Type Adult, Pediatric, Neonatal PR range 25 to 240 bpm
PR accuracy ±3 bpm (without motion)
NIBP ±5 bpm (with motion)
Meet standards of IEC 80601-2-30, ISO 81060-2.
Technique Oscillometry PR from Nellcor SpO2
Operation mode Manual PR range 20 to 300 bpm
Parameters Systolic, diastolic, mean, Heart Rate PR accuracy ±3 bpm (20 to 250 bpm)
Max Measure time Adult/Pediatric: 180 s, Neonate: 90 s Not specified ( 251 to 300 bpm)
Systolic range Adult: 40 to 270 mmHg
Pediatric: 40 to 200 mmHg PR from NIBP
Neonate: 40 to 135 mmHg PR range 40 to 240 bpm
Diastolic range Adult: 10 to 210 mmHg PR accuracy ±3 bpm or ±3%, whichever is greater
Pediatric: 10 to 150 mmHg
Neonate: 10 to 100 mmHg SmarTemp™ Temperature
Mean range Adult: 20 to 230 mmHg Meet standard of ISO 80601-2-56.
Pediatric: 20 to 165 mmHg Technique Thermal resistance
Neonate: 20 to 110 mmHg Operating mode Adjusted mode (predictive mode)
NIBP accuracy Max mean error: ± 5 mmHg Direct mode (monitor mode)
Max standard deviation: 8 mmHg Measurement range Monitor mode: 25 °C to 44 °C
NIBP resolution 1 mmHg (77 °F to 111.2 °F)
Predictive mode: 35 °C to 43 °C
Mindray SpO2 (95 °F to 109.4 °F)
Meet standards of ISO 80601-2-61. Accuracy (Monitor mode)
SpO2 range 0 to 100 % 25 °C to 32 °C (not include 32 °C): ±0.2 °C
SpO2 resolution 1% 32 °C to 44 °C (include 32 °C): ±0.1 °C
SpO2 accuracy 70 to 100%: ±2% or
(without motion in adult /pediatric mode) 77°F to 89.6°F (not include 89.6°F): ±0.4 °F
70 to 100%: ±3% 89.6°F to 111.2°F (include 89.6 °F): ±0.2 °F
(without motion in neonate mode) Resolution ±0.1 °C or ± 0.2 °F
0% to 69%: Not specified.
Perfusion indicator (PI) Yes, for Mindray SpO2 Indicator
LED indicator Power on LED (1), two color: yellow/green
Masimo SpO2 AC power LED (1), green
Meet standards of ISO 80601-2-61. Battery LED (1), green
SpO2 range 1 to 100 % Audible indicator Give pulse tone, power-on self check tone
SpO2 resolution 1% Other Technical alert on screen
SpO2 accuracy 70 to 100%: ±2%
(without motion in adult /pediatric mode) Interfacing
70 to 100%: ±3% Main unit AC power connector (1)
(without motion in neonate mode) RS 232 connector (1), for DIAP protocol
70 to 100%: ±3%
(measured with motion) Power
1% to 69%: Not specified. AC line voltage 100 to 240 VAC (±10 %)
Perfusion indicator (PI) Yes, for Masimo SpO2 AC frequency 50/60 Hz (±3 Hz)
AC current 0.9 to 0.5 A Relative humidity Operating: 15 to 95 % (non condensing)
Battery type Rechargeable lithium-ion battery, Storage: 10 to 95 % (non condensing)
Battery run time At least 11 hours (2600mAh battery) Barometric Operating: 57.0 to 107.4 kPa
At least 22 hours (4500mAh battery) Storage: 16.0 to 107.4 kPa)
--------------------------------------------------------------
Environmental Requirements Some of functions marked with an asterisk may not be available.
Temperature Operating: 0 to 40 °C (without Temp) Please contact your local Mindray sales representative for the most
5 to 40 °C (with Temp) current information.
Storage: -20 to 60 °C

www.mindray.com