VS600 Specifications
VS600 Specifications
Physical Specifications
Weight ≤1.9 kg (with SpO2, NIBP and a battery ) Nellcor SpO2
≤1.7 kg (with NIBP and a battery ) Meet standards of ISO 80601-2-61.
Size 134 mm × 120 mm × 243 mm SpO2 range 0 to 100 %
Screen type Segment display SpO2 resolution 1%
Screen size 90mm × 90mm SpO2 accuracy 70 to 100%: ±2% (adult/pediatric)
70 to 100%: ±3% (neonate)
EMC Compliance Meets the requirements of IEC 60601-1-2. 0% to 69%: Not specified.
Perfusion indicator (PI) No
Classifications
Defibrillation proof (protection against electrical shock) PR from Mindray SpO2
Type CF: SpO2, NIBP and Temp PR range 20 to 254 bpm
Water proof IPX1 PR accuracy ± 3 bpm (without motion)
Degree of mobility Portable
PR from Masimo SpO2
Patient Type Adult, Pediatric, Neonatal PR range 25 to 240 bpm
PR accuracy ±3 bpm (without motion)
NIBP ±5 bpm (with motion)
Meet standards of IEC 80601-2-30, ISO 81060-2.
Technique Oscillometry PR from Nellcor SpO2
Operation mode Manual PR range 20 to 300 bpm
Parameters Systolic, diastolic, mean, Heart Rate PR accuracy ±3 bpm (20 to 250 bpm)
Max Measure time Adult/Pediatric: 180 s, Neonate: 90 s Not specified ( 251 to 300 bpm)
Systolic range Adult: 40 to 270 mmHg
Pediatric: 40 to 200 mmHg PR from NIBP
Neonate: 40 to 135 mmHg PR range 40 to 240 bpm
Diastolic range Adult: 10 to 210 mmHg PR accuracy ±3 bpm or ±3%, whichever is greater
Pediatric: 10 to 150 mmHg
Neonate: 10 to 100 mmHg SmarTemp™ Temperature
Mean range Adult: 20 to 230 mmHg Meet standard of ISO 80601-2-56.
Pediatric: 20 to 165 mmHg Technique Thermal resistance
Neonate: 20 to 110 mmHg Operating mode Adjusted mode (predictive mode)
NIBP accuracy Max mean error: ± 5 mmHg Direct mode (monitor mode)
Max standard deviation: 8 mmHg Measurement range Monitor mode: 25 °C to 44 °C
NIBP resolution 1 mmHg (77 °F to 111.2 °F)
Predictive mode: 35 °C to 43 °C
Mindray SpO2 (95 °F to 109.4 °F)
Meet standards of ISO 80601-2-61. Accuracy (Monitor mode)
SpO2 range 0 to 100 % 25 °C to 32 °C (not include 32 °C): ±0.2 °C
SpO2 resolution 1% 32 °C to 44 °C (include 32 °C): ±0.1 °C
SpO2 accuracy 70 to 100%: ±2% or
(without motion in adult /pediatric mode) 77°F to 89.6°F (not include 89.6°F): ±0.4 °F
70 to 100%: ±3% 89.6°F to 111.2°F (include 89.6 °F): ±0.2 °F
(without motion in neonate mode) Resolution ±0.1 °C or ± 0.2 °F
0% to 69%: Not specified.
Perfusion indicator (PI) Yes, for Mindray SpO2 Indicator
LED indicator Power on LED (1), two color: yellow/green
Masimo SpO2 AC power LED (1), green
Meet standards of ISO 80601-2-61. Battery LED (1), green
SpO2 range 1 to 100 % Audible indicator Give pulse tone, power-on self check tone
SpO2 resolution 1% Other Technical alert on screen
SpO2 accuracy 70 to 100%: ±2%
(without motion in adult /pediatric mode) Interfacing
70 to 100%: ±3% Main unit AC power connector (1)
(without motion in neonate mode) RS 232 connector (1), for DIAP protocol
70 to 100%: ±3%
(measured with motion) Power
1% to 69%: Not specified. AC line voltage 100 to 240 VAC (±10 %)
Perfusion indicator (PI) Yes, for Masimo SpO2 AC frequency 50/60 Hz (±3 Hz)
AC current 0.9 to 0.5 A Relative humidity Operating: 15 to 95 % (non condensing)
Battery type Rechargeable lithium-ion battery, Storage: 10 to 95 % (non condensing)
Battery run time At least 11 hours (2600mAh battery) Barometric Operating: 57.0 to 107.4 kPa
At least 22 hours (4500mAh battery) Storage: 16.0 to 107.4 kPa)
--------------------------------------------------------------
Environmental Requirements Some of functions marked with an asterisk may not be available.
Temperature Operating: 0 to 40 °C (without Temp) Please contact your local Mindray sales representative for the most
5 to 40 °C (with Temp) current information.
Storage: -20 to 60 °C
www.mindray.com
P/N:ENG-VS-600 Datasheet -210285X2P-20190529
©2019 Shenzhen Mindray Bio-Medical Electronics Co.,Ltd. All rights reserved.