2022-10-28 BA S3C ES Komm en
2022-10-28 BA S3C ES Komm en
SensoStar C
DE-18-MI004-PTB037 (MID heat)
DE-18-M-PTB-0049 (national German cooling)
CH-T2-18769-00 (national Swiss cooling)
3 General Information
- Valid standards for the application of calculators for heat metering: EN 1434, parts 1 – 6; the Measuring
Instruments Directive 2014/32/EU, Annexes I and MI-004; and the relevant national verification
regulations.
- For the selection, installation, commissioning, monitoring and maintenance of the instrument observe the
standard EN 1434 part 6, as well as the verification regulations PTB TR K8 + K9 for Germany (and any
relevant national verification regulations in other countries).
- National regulations for the consumption measurement of cooling must be observed.
- The technical regulations for electrical installations must be observed.
- This product fulfils the requirements of the European Council Directive on Electromagnetic Compatibility
(EMC Directive) 2014/30/EU.
- The identification plate of the instrument and the seals must not be removed or damaged – otherwise the
guarantee and the approved application of the instrument are no longer valid!
- The calculator left the factory in conformance with all applicable safety regulations. All maintenance and
repair work is to be carried out only by qualified and authorized technical personnel.
- Instruments with activated radio function are not allowed on air freight.
- The correct installation point in the system must be chosen: inlet or outlet flow (see item 3.1 ’Pictograms
installation point’).
- To clean the calculator (only if necessary) use a slightly moist cloth.
- To protect against damage and dirt the calculator should only be removed from the packaging directly
before installation.
- All specifications and instructions listed on the data sheet and in the Application Notes must be adhered to.
Further information can be obtained at www.engelmann.de.
- The heat meter has a lithium-metal-battery. Do not open the batteries, do not bring the batteries into
contact with water or expose them to temperatures above 80 °C. Do not charge them or short-circuit them.
- Instruments which have been replaced or exchanged must be disposed of according to relevant
environmental regulations.
- The display is deactivated and can be activated for two minutes by pushing the button.
- Unit of energy and installation point (outlet flow / inlet flow) can be set on location, only once, before
start of operation by pushing the button or alternatively using the „Device Monitor“ software.
- Type and concentration of glycol in the medium of those calculator types designed to be used with glycol
can be set on location at any time using the „Device Monitor“ software (see item 8.1 for details).
Page 1 of 23 Article no.: 1080622201 2022_10_28 Subject to technical change! Errors excepted.
3.1 Pictograms installation point of calculator (in the information loops)
On the right in the calculator display in all information loops you will find one of the following two pictograms. The
pictogram indicates in which pipe the calculator is to be mounted.
Installation in outlet flow
At delivery, the display shows „H 05“ until temperature sensors have been attached. This message disappears as
soon as temperature sensors have been connected and the first temperature measurement is carried out (every 15
minutes without flow).
The calculator connections have been designed to meet the valid standard EN1434-2. All terminal strips have been
labelled according to this standard.
The terminal strips are located under the cover of the calculator housing.
Page 2 of 23 Article no.: 1080622201 2022_10_28 Subject to technical change! Errors excepted.
- Sensor cables that are too long should not be rolled up tightly into an ‘air-core coil’. The cables should
either be laid out disordered, or rolled up loosely into a wide coil which can be turned and tied into an ‘8’.
Mounting
- Loosen the two cable glands without blind plugs and glide them over the sensor cables.
- Feed the temperature sensors through the appropriate openings of the cable glands into the terminal box.
- Clamp the wires (see identification label temperature sensors) as shown in the illustration:
The inlet flow temperature sensor must always be connected to clamp 5 and 6 (inlet).
The outlet flow temperature sensor must always be connected to clamp 7 and 8 (outlet).
The color of the wires does not matter.
Mounting
- Loosen the middle cable gland and glide it over the flow meter cable.
- Remove the blind plug in the cable gland opening. Feed the pulse cable of the flow meter through the
opening into the terminal box.
- Clamp 10 and 11 are used for the connection to the flow meter.
- Clamp on the wires as shown in the illustration.
Note: For flow meters with open collector connections (electronic outputs) make sure the polarity is correct.
Please check the unused cable glands to make sure that the necessary blind plugs are inserted and then tighten
the cable glands by hand.
Page 3 of 23 Article no.: 1080622201 2022_10_28 Subject to technical change! Errors excepted.
5.2.1 Special features when connecting to a Diehl VMT „Sharky FS 473“
For the correct connection of the cables, please note the polarity described in the following drawing:
white Clamp 10
Useful pulse
brown
Clamp 11
Important note: The colored marking of the cables corresponds to the scheme shown above.
The white cable (+) is connected to clamp 10, the brown cable (-) to clamp 11.
Important note: The pulse value will be set permanently after the first input pulses and cannot be changed
afterwards. Pay special attention that the flow meter does not register a flow before the correct pulse value has
been chosen (factory setting 1 l/pulse).
Page 4 of 23 Article no.: 1080622201 2022_10_28 Subject to technical change! Errors excepted.
25 - 0,00 MWh 0,00 GJ 0,00 m³ 0,00 m³/h 0,00 kW
100 - 0,0 MWh 0,0 GJ 0,0 m³ 0,0 m³/h 0,0 kW
250 - 0,0 MWh 0,0 GJ 0,0 m³ 0,0 m³/h 0,0 kW
1000 - 0 MWh 0 GJ 0 m³ 0 m³/h 0 kW
2500 - 0 MWh 0 GJ 0 m³ 0 m³/h 0 kW
6 Start of Operation
- Slowly open the shut-off valves.
Check the following points:
- Is the flow meter of the right size?
- Does the directional arrow on the flow meter match the actual direction of flow?
- Check that there are no leaks.
- Is a flow volume displayed?
- Are all shut-off valves open?
- Is the heating (heating/cooling) system clear (dirt filters not clogged)?
- Is a plausible temperature difference displayed?
When the components are functioning properly, attach the seals to the temperature sensors and the flow meter.
Protect the calculator against unauthorized opening using the numbered adhesive seal enclosed. The added bar
code label can be used for the purpose of documentation.
During installation of the retrofitting module observe the ESD requirements according to EN 61340-5-1.
This means that on location an antistatic wrist strap with an integrated 1 MΩ resistor has to be used which must be
connected to a proper spot: This is either a grounded pipe or – only with an appropriate adapter! – a Schuko plug
grounding socket. The antistatic wrist strap must be worn tightly on the skin of the wrist.
Open the calculator by pulling the clamping bracket at the upper housing´s leading edge up.
Plug the interface module on the right side of the PC board. If there are module cables you have to loosen the
needed number of cable glands and to glide them over the cables. Remove the blind plugs from the cable gland
openings and feed the cables into the calculator.
Protect the calculator against unauthorized opening using one of the numbered adhesive seals enclosed to the
modules. The added bar code label can be used for the purpose of documentation.
Page 5 of 23 Article no.: 1080622201 2022_10_28 Subject to technical change! Errors excepted.
7.4 Mounting a power pack
If an external power supply is needed, only the power pack designed for our calculator may be used.
To connect it to the calculator, please open the calculator. First remove the battery from the calculator and plug it
into the battery connector in the power pack. Protect the power pack against unauthorized opening using one of
the numbered adhesive seals enclosed. The added bar code label can be used for the purpose of documentation.
(The battery is a backup in case of a power outage.)
Remove the left blind grommet in the calculator housing and feed the power pack cable (A) through the cable
feedthrough. Press the cable grommet into the cable feedthrough. Connect the black service plug with the contact
fingers on the left hand side of the calculator’s PC board under the display. Plug the white connector into the
battery connector on the PC board.
Only authorized technical personnel may connect the power pack to 230 V / 24 V and check it.
When the calculator detects external power supply the pictogram of a mains plug appears in the display.
Protect the calculator against unauthorized opening using one of the numbered adhesive seals enclosed to the
power pack. The added bar code label can be used for the purpose of documentation.
A
Page 6 of 23 Article no.: 1080622201 2022_10_28 Subject to technical change! Errors excepted.
8 Display
The calculator has a liquid crystal display with 8 digits and special characters. The values that can be shown are
divided into five display loops. All data is retrieved using the push-button next to the display.
At the start you are automatically in the main loop (1st level). By pressing the push-button longer than 4 seconds
you change to the next display loop. Keep the push-button pressed until you reach the desired information loop. By
pressing the push-button briefly each time you can scan all the information within a loop.
After 2 minutes of non-use of the push-button, the display will automatically be deactivated.
Page 7 of 23 Article no.: 1080622201 2022_10_28 Subject to technical change! Errors excepted.
1)
Up to the end of the month / the 15th of the month (for the semimonthly values) the consumption and date will be
shown as 0.
2)
Three pulse inputs are an option. They can be set using the software „Device Monitor“.
1) Current power in kW 2) Current flow in m³/h 3) Inlet flow temperature 4) Outlet flow
in °C temperature in °C
Page 8 of 23 Article no.: 1080622201 2022_10_28 Subject to technical change! Errors excepted.
5) Maximum temperature
difference
alternating with date
and time
1) Parametrizing 2) Parametrizing
„unit of energy“ „installation point“
These parametrizing options are only available when the amount of energy is still <= 10 kWh. Make sure that
these characteristics are set as needed before starting up the system.
Setup by pushing the button: In order to start the editing mode for parametrizing you must select the respective
item in the parametrizing loop and then push the button once again for 2-3 seconds. As an aid, after 2 seconds the
„editing pen“ will be displayed bottom left in the LCD (see below picture). As soon as it appears you have to let go
of the button. Then the current display will start blinking.
By pressing the push-button briefly you can switch to the next option. By pressing the push-button longer the
currently displayed option will be set. If no option is chosen there will be no change and as soon as the LCD goes
out the edit mode will end automatically.
b) The following characteristic of those calculator types designed to be used with glycol can be set on location at
any time using the „Device Monitor“ software:
- type and concentration of glycol in the medium (propylene glycol; ethylene glycol; 20 %; 30 %; 40 %; 50%).
Page 9 of 23 Article no.: 1080622201 2022_10_28 Subject to technical change! Errors excepted.
9 Application Conditions
Calculator
Temperature range medium °C 0 – 150
heat
Temperature range medium °C 0 – 50
cooling
Mechanical class M2
Electromagnetic class E2
Protection class IP54
Ambient temperature in the °C 5 – 55 at 95 % relative humidity
field
Transport temperature °C -25 – 70 (for maximal 168 h)
Storage temperature °C -25 – 55
Pulse input interface microcontroller CMOS input class IB according to EN 1434-2: 2015
Pulse values standard see type identification label
TX versions adjustable pulse values will be shown in the display: 1; 2,5; 10; 25;
100; 250; 1000; 2500
10 Information Messages
When the instrument has detected an information message, the message symbol is displayed:
The specific message can be found at menu item 6 ‘Information message’ in level 1 / main loop (see section 8,
Display). The message code is displayed alternately in binary and hexadecimal form.
The instrument recognizes seven message causes, which can also occur in combination with each other.
Page 10 of 23 Article no.: 1080622201 2022_10_28 Subject to technical change! Errors excepted.
Example: Temperature sensor inlet cable break
Message
Temperature sensor
Temperature sensor
Temperature sensor
Temperature sensor
Low battery
Bit 7 Reset
6 5 4 3 2 1 0
Alternating hexadecimal
Display location 1 2 3 4 5 6 7 8 message displayed (LCD)
Alternating binary
message
displayed (LCD)
When a message appears in the standard display (total heat energy), with the exception of the messages
- Low battery (H 80)
- Reset (H 40),
the instrument must be exchanged and sent to the supplier for examination.
11 Manufacturer
Engelmann Sensor GmbH
Rudolf-Diesel-Str. 24-28
69168 Wiesloch-Baiertal
Germany
Page 11 of 23 Article no.: 1080622201 2022_10_28 Subject to technical change! Errors excepted.
Operating Instructions
The factory-setting of the clock in the meter is standard (winter) Central European Time (GMT +1). There is no
automatic changeover to daylight savings (summer) time.
The radio function is deactivated upon delivery (factory-setting). See chapter 1.4.1.3 & 1.4.2.3 regarding the
activation of the radio interface.
Page 13 of 23 Article no.: 1080622201 2022_10_28 Subject to technical change! Errors excepted.
1.4.1 Radio interface wireless M-Bus EN 13757-3, -4 (optional)
The radio interface transmits meter data (absolute values).
b) The radio function can also be activated using the software “Device Monitor”.
Page 14 of 23 Article no.: 1080622201 2022_10_28 Subject to technical change! Errors excepted.
This software can be ordered separately as an option.
The radio function can only be deactivated using the software “Device Monitor”.
After activation of the radio function or modification of the radio parameters the meter remains in installation
mode for 60 minutes. During this time he sends telegrams in a 36-seconds-interval.
If using the compact mode, after activation the meter transmits during installation mode format telegrams and
compact telegrams alternately.
During installation mode at least one meter of the version being installed (inlet or outlet flow, heat or heat/cooling,
pulse inputs, display units) must be read out with the Engelmann “Read-out Software walk-by”. The format of the
telegram will be stored locally in the PC in an .xml file.
After completion of the installation mode only compact telegrams will be transmitted.
b) The encryption can also be activated using the software “Device Monitor”. This software can be ordered
separately as an option.
The encryption can only be deactivated using the software “Device Monitor”.
Radio characteristics
Frequency 868 MHz Frequency 868 MHz
Output power 14 dBm Output power 14 dBm
Receiver sensitivity -135 dBm Receiver sensitivity -135 dBm
LoRaWAN characteristics
Device class Class A, Bi-directional
LoRa version 1.0.2 Rev B
Activation OTAA* or ABP*
Data rate DR0-DR5 (250 bit/s-5470 bit/s)
*OTAA = Over-the-air activation
** ABP = Activation by personalization
Page 15 of 23 Article no.: 1080622201 2022_10_28 Subject to technical change! Errors excepted.
1.4.2.2 LoRa module configuration
The module can be configurated by using the Device Monitors or the OTC App.
Parameter Possible settings Factory setting
Power mode Active Inactive
Inactive
Configuration Lock Open Open
Locked
Synchronize meter time on off
off
Activation type OTAA OTAA
ABP
EcoMode off EcoMode 10 years
6 years
10 years
Transmit interval *[Min.] 5 … 1440 60
Message format** Standard Standard
Engelmann
Compact
JSON
Scheduled - daily redundant
Scheduled - extended
Combined heat/cooling
Pulse input selection Choice between 0-3 pulse inputs 0
* The actual transmission interval depends on the type of telegram and the current data rate. The transmission
interval is adjusted accordingly in order to guarantee the set battery life (Eco Mode 10 or 6 years). You can find
more information in the „Manual LoRa Module“.
** For pulse input option, the telegram type „Engelmann“ must be selected.
a) NTC interface - via Elvaco OTC App; further information is available in the operating instructions of the Elvaco
OTC App: https://www.elvaco.se/Image/GetDocument/en/269/elvaco-otc-app-manual-english.pdf
b) Optical interface - via the Engelmann configuration software Device Monitor – from version 2.22; further
information is available in the operating instructions of the Device Monitor. The software can be ordered
separately.
c) Via the heat meter menu;
Press and hold the button to switch to display loop "6" (= module loop; see Chapter
3 Display in the Module Loop (Optional)). Then switch to the second loop with a
short press of the button – 6-02 – „EnA oFF“ (see image).
In order to start the edit mode, you must then press the key
press once more for 2-3 seconds. As an aid, an “editing pen” symbol appears in the
lower left corner of the LCD after 2 seconds. As soon as it can be seen, you must
release the button. The display now shows „EnA on“ and a black triangle in all
display loops (see image).
Page 16 of 23 Article no.: 1080622201 2022_10_28 Subject to technical change! Errors excepted.
As soon as the meter has connected to the LoRaWAN network, “LorA Con” appears in
the LCD.
1.5 Three additional pulse inputs (optional; only in conjunction with M-Bus or radio)
With this option, additional instruments with pulse outputs can be read out via optical interface, M-Bus or radio.
General information about pulse inputs:
It is important to note that the acknowledged state of the art technology rules and the relevant legal restraints
(international and local; see “Relevant Norms / Standards / Literature Pulse Inputs”) are to be observed.
The installation has to be performed by authorized, skilled persons.
If the regulations and the information in the installation and operating instruction manuals are not strictly followed,
or if the installation is shown to be faulty, any resulting expenses will be charged to the company responsible for
the installation.
1.5.3 Possible combinations of the different input (class IB) and output (class OA) devices
Class IA Class IB Class IC Class ID Class IE
Class OA yes yes no yes no
Class OB yes no no yes yes
Class OC no yes yes no no
Class OD no no yes no no
Class OE no no no no yes
Page 17 of 23 Article no.: 1080622201 2022_10_28 Subject to technical change! Errors excepted.
1.5.5 Set-up possibilities
Pulse value Units
1 liters / kWh / pulse without unit
2,5 liters / kWh / pulse without unit
10 liters / kWh / pulse without unit
25 liters / kWh / pulse without unit
100 liters / kWh / pulse without unit
250 liters / kWh / pulse without unit
1000 liters / kWh / pulse without unit
For transmitting the values of the pulse inputs via radio, transmission must be set using the “Device Monitor”
software, if the meters were not already ordered with the transmission of these values set.
Select the menu item „Parameterization of the module“ in the software. Set the radio transmission in the menu
item „Transmission of values for the pulse inputs“. (The transmission of these values via M-Bus is always set in the
factory setting.)
From firmware version 1.03 (S3) and 1.00 (S3C) on the meter recognizes nominal size and unit of energy and
autonomously sets the pulse values for energy and volume according to the following notes.
Page 18 of 23 Article no.: 1080622201 2022_10_28 Subject to technical change! Errors excepted.
S3 - pulse output for energy:
Display shows Display shows Display shows Display shows
kWh / MWh Gcal GJ MMBTU
qp 0,6 m3/h 1 kWh/pulse 1 Mcal/pulse 10 MJ/pulse 10 MMBTU/pulse
qp 1,5 m3/h 1 kWh/pulse 1 Mcal/pulse 10 MJ/pulse 10 MMBTU/pulse
qp 2,5 m3/h 1 kWh/pulse 1 Mcal/pulse 10 MJ/pulse 10 MMBTU/pulse
qp 3,5 m3/h 10 kWh/pulse 10 Mcal/pulse 10 MJ/pulse 10 MMBTU/pulse
qp 6 m3/h 10 kWh/pulse 10 Mcal/pulse 10 MJ/pulse 10 MMBTU/pulse
1.6.2 Technical data for one pulse output and two pulse outputs
Pulse outputs class OA (electronic switch) according to EN 1434-2:2016
Connecting cable length 1m
Switching voltage, maximum 30 V
Switching current, maximum 27 mA
Contact resistance (on) max. 74 Ω
Contact resistance (off) min. 6 MΩ
Closure time 100 ms
Interval between pulses 100 ms
Page 19 of 23 Article no.: 1080622201 2022_10_28 Subject to technical change! Errors excepted.
Pulse outputs for energy:
The pulse value for energy is always determined by the last place of the energy display.
Examples:
Display: 0 kWh -> pulse value: 1 kWh/pulse
Display: 0,000 MWh -> pulse value: 0,001 MWh/pulse
Display: 0,000 GJ -> pulse value: 0,001 GJ/pulse
To retrofit such a meter with an additional communication interface the calculator´s adhesive seal has to be
destroyed and the device´s calculator to be opened. Use a screwdriver with a wide tip (4 - 5 mm) and carefully
press in the two round predetermined breaking points above the cable feedthroughs (see picture 1).
Then insert the screwdriver into one of the two openings at an angle of approx. 45° and carefully lift the handle up
to approx. 90° (see picture 2). The upper housing piece of the calculator is then no longer latched on this side.
Repeat this with the other opening. Now the upper housing piece can be taken off.
Page 20 of 23 Article no.: 1080622201 2022_10_28 Subject to technical change! Errors excepted.
Plug the interface module on the right side of the PC board (see picture 3). The cables are to be fed through the
rightmost cable feedthroughs into the calculator after removing the blind grommets. Close the calculator.
Protect the calculator against unauthorized opening using one of the numbered adhesive seals enclosed to the
modules (stick it upon the seal destroyed). The added bar code label can be used for the purpose of
documentation.
To remove a module the upper housing piece must be carefully pressed against the rear panel of the lower housing
piece during opening the calculator. Thus the two rear snap-fits of the upper housing piece lever out the module of
the PC board (see picture 4).
Page 21 of 23 Article no.: 1080622201 2022_10_28 Subject to technical change! Errors excepted.
2.2 Retrofitting S3C with an additional communication interface
To our calculator further communication interfaces can be added later.
Open the calculator by pulling the clamping bracket at the upper housing´s leading edge up.
Plug the interface module on the right side of the PC board. If there are module cables you have to loosen the
needed number of cable glands and to glide them over the cables. Remove the blind plugs from the cable gland
openings and feed the cables into the calculator.
Protect the calculator against unauthorized opening using one of the numbered adhesive seals enclosed to the
modules (stick it upon the seal destroyed). The added bar code label can be used for the purpose of
documentation.
Page 22 of 23 Article no.: 1080622201 2022_10_28 Subject to technical change! Errors excepted.
3 Display in the Module Loop (Optional)
or: or:
4 Imprint
Engelmann Sensor GmbH
Rudolf-Diesel-Str. 24-28
69168 Wiesloch-Baiertal
Germany
E-Mail: [email protected]
www.engelmann.de
Page 23 of 23 Article no.: 1080622201 2022_10_28 Subject to technical change! Errors excepted.