Thermo Accela 600 Pump Service Manual - Rev 2 - 122209a
Thermo Accela 600 Pump Service Manual - Rev 2 - 122209a
Service Manual
Accela™, Chromquest™, and Xcalibur™ are trademarks of Thermo Fisher Scientific. Freon™ and Flourinert™ are
registered trademarks of DuPont. Fluorinert™ is a registered trademark of 3M. All other trademarks are the
property of Thermo Fisher Scientific and its subsidiaries.
Thermo Fisher Scientific Inc. provides this document to its customers with a product purchase to use in the
product operation. This document is copyright protected and any reproduction of the whole or any part of this
document is strictly prohibited, except with the written authorization of Thermo Fisher Scientific Inc.
The contents of this document are subject to change without notice. All technical information in this
document is for reference purposes only. System configurations and specifications in this document supersede
all previous information received by the purchaser.
Thermo Fisher Scientific Inc. makes no representations that this document is complete, accurate or error-
free and assumes no responsibility and will not be liable for any errors, omissions, damage or loss that might
result from any use of this document, even if the information in the document is followed properly.
This document is not part of any sales contract between Thermo Fisher Scientific Inc. and a purchaser. This
document shall in no way govern or modify any Terms and Conditions of Sale, which Terms and Conditions
of Sale shall govern all conflicting information between the two documents.
For Research Use Only. Not regulated for medical or veterinary diagnostic use by U.S. Federal Drug
Administration or other competent authorities.
C
Contents
Preface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .v
Related Documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . v
Safety and Special Notices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .vi
Safety Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .vi
Environmental Conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . vii
Good Laboratory Practices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .viii
Keeping Good Records . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .viii
Chapter 1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
Functional Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Hardware Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Status LEDs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Chapter 2 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
Installation Checklist. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Unpacking and Inspecting the Instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Making the Initial Instrument Preparations. . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Connecting the Power Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Connecting the Pump to the Computer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Connecting the Pump to Other System Modules . . . . . . . . . . . . . . . . . . . . . . . 14
Connecting the Solvent Lines . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Installing the Optional Dynamic Mixer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Installing the Optional Seal Wash Pump . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Setting up Continuous Flushing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Connecting the Pump to the Autosampler. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Configuring the Pump to the Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Powering On the Pump. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .111
Preface
This manual provides you with information on how to install, maintain, and troubleshoot the
Thermo Scientific 600 Pump. The Introduction, Installation, Maintenance, Troubleshooting,
and Replacement Parts chapters are also in the customer’s 600 Pump Hardware Manual. The
Health Check and Calibration Procedures and Replacement Parts for Service chapters contain
information intended only for individuals that have been trained on servicing the 600 Pump.
You operate the 600 Pump from a computer that runs Aria™ operating system (OS) (version
1.6.3 or higher), Xcalibur™, or ChromQuest™ data system.
Several Thermo Scientific chromatography systems use the 600 Pump. Each of these systems
has specific pump installation requirements that supersede any installation procedures
described in this manual. Also, procedures for controlling the pump and performing
diagnostic procedures vary depending on the system that controls the pump. Refer to your
system manual for information on using the system features to control the pump.
Related Documentation
In addition to this guide, you will find information related to 600 Pump in the
following documents:
• Aria OS User Guide—Provides information on controlling the 600 Pump using Aria OS
when the pump is installed in an Aria or Transcend™ system.
• Accela 600 Pump Hardware Manual —Provides installation and spare parts information
when the 600 Pump is installed in an Accela™ system.
• Accela Getting Started with ChromQuest Guide—See this guide for information on
controlling the 600 Pump using Chromquest software when the pump is installed in an
Accela system.
• Accela Getting Started with Xcalibur Guide—Provides information on controlling the 600
Pump using Xcalibur software when the pump is installed in an Accela system.
CAUTION Highlights electric shock related hazards to humans. Each electric shock notice
is accompanied by the international high voltage symbol.
Safety Information
Observe the precautions listed in this section to ensure the safe operation and longevity of the
instrument.
CAUTION When you use the 600 Pump, follow the generally accepted procedures for
quality control and method development. If you use the 600 Pump in the field of
chromatographic analysis and you observe a change in retention time of a particular
compound, in the resolution between two compounds, or in peak shape, immediately
determine the reason for the changes. Until you determine the cause of the change, do not
rely on separation results.
CAUTION Use equipment only in the manner specified by its manufacturer, or risk
impairing the protection provided by the equipment.
CAUTION Do not service any part when operating the instrument. Do not remove the
front panel when the instrument is running. Do not run the instrument without the front
panel.
CAUTION The instrument contains voltage lines. Switch off the power and disconnect
the power cable prior to servicing the instrument. There is no need to open the pump
housing as all user serviceable components are outside of the instrument.
CAUTION Never try to repair or replace any components of the instrument not described
in this manual without the assistance of a Thermo Fisher Scientific representative. There
are no operator serviceable parts inside the power supply of the 600 Pump.
CAUTION To prevent personal injury, observe good laboratory practice when handling
solvents, changing tube lines. or both. Consult the pertinent material safety data sheets
(MSDSs) for the solvents used for HPLC analysis.
Environmental Conditions
The installation category (over voltage category) for this instrument is Category 2, which
pertains to equipment that receives its electrical power from the local level, such as an
electrical wall outlet.
The pollution degree for this instrument is Pollution Category 2 (Laboratory): Normally only
nonconductive pollution occurs. Expect temporary conductivity caused by condensation.
CAUTION Only use the pump according to the conditions stated, or risk personal injury
or damage to the pump.
Chemical Toxicity
Although the large volume of toxic and flammable solvents used and stored in laboratories can
be quite dangerous, do not ignore the potential hazards posed by your samples. Take special
care to read and follow all precautions that ensure proper ventilation, storage, handling, and
disposal of both solvents and samples. Become familiar with the toxicity data and potential
hazards associated with all chemicals by referring to the manufacturers’ MSDSs.
Solvent Requirements
Use HPLC-grade solvents that are free of particulates. Choose a mobile phase that is
compatible with the sample and column you have selected for your separation. Be aware that
some solvents are corrosive to stainless steel.
CAUTION Do not use solvents containing Freon™ and perfluorinated solvents, such as
Fluorinert™, FC-75, FC-40, and Fomblin perfluoro polyether solvents from Ausimont.
They adversely affect the Teflon™ AF degassing membrane.
Solvent Disposal
Make sure you have a solvent waste container or other kind of drain system available at or
below the benchtop level. Most solvents have special disposal requirements prohibiting
disposal directly down a drain. Follow all governmental regulations when disposing of any
chemical.
Contacting Us
There are several ways to contact Thermo Fisher Scientific for the information you need.
Introduction
This chapter provides an introduction to the 600 Pump.
Contents
• Functional Description
• Hardware Components
• Status LEDs
• Specifications
Figure 2. Image of the 600 Pump in a Transcend pump modular housing with covers open
Functional Description
The 600 Pump is a quaternary, low-pressure mixing pump with a built-in solvent degassing
system and an automatic calibration feature. The pumping system provides optimum
performance in the flow rate range of 1 μL/min to 5 mL/min and a maximum operating
pressure of 600 bar. For ultra high performance liquid chomatography systems and
TurboFlow technology applications, the 600 Pump is an excellent choice.
The 600 Pump uses a force sensor feedback controller, which continuously calibrates valve
timing and pumping efficiency based on the measured compressibility of the solvent. This
patent-pending feature enables the 600 Pump to form accurate gradients virtually
pulsation-free; no pulse damping device is required. The force sensor technology measures
system pressure without an additional pressure transducer in the flow path, thereby omitting a
large delay volume that is cumbersome to flush.
The integral vacuum degasser offers superb solvent degassing efficiency using a built-in
vacuum membrane degasser that removes dissolved gasses from the mobile phase.
The 600 Pump, a bench-top unit, is included in several Thermo Scientific systems including
Accela and TLX systems. It uses either a modular pump housing or a system housing. The
pump is remotely controlled with a USB communication link from a computer using Aria
OS, Xcalibur, or ChromQuest software.
Hardware Components
The 600 Pump has these major components: the vacuum degasser assembly, the liquid
displacement assembly (LDA), a 65 μL static mixer, and an optional dynamic mixer. The
pump may also contain status LEDs and an optional seal wash pump for continuously
flushing the piston guide bushings.
Purge valve
The following topics describe the main components of the 600 Pump:
• Vacuum Degasser Assembly
• Liquid Displacement Assembly (LDA)
• Purge Valve
• Dynamic Mixer
• Seal Wash Pump
As the lower piston withdraws, mobile phase from the proportioning valve assembly passes
through the inlet check valve and fills the lower piston chamber. As the lower piston returns,
the inlet check valve closes and the mobile phase is pushed through the outlet check valve and
into the upper piston chamber. Both check valves are low volume and incorporate a dual
ball-and-seat design, which ensures a positive seal with minimal delay volume.
Next, the upper piston moves into the upper piston chamber, closing the outlet check valve
and forcing the mobile phase out the front of the outlet module.
Purge Valve
The purge valve is located to the left of the LDA. See Figure 3. Turning the wingnut
counterclockwise opens the purge valve.
When the purge valve is open and the solvent flow is experiencing back pressure from either
an LC column or a back pressure regulator, mobile phase exits the LDA from the left side of
the outlet module.
When the purge valve is closed, mobile phase flows out the front of the LDA assembly and
into the high-pressure tubing that connects the LDA to the mixer or outlet adapter and the
mixer to the autosampler.
Dynamic Mixer
Adding the optional dynamic mixer to the 600 Pump ensures mixing of complex or relatively
incompatible solvents and reduces the chromatographic baseline noise for applications using
proportioned mobile phases. The integrated dynamic mixer also increases the accuracy and
reproducibility of complex ternary and quaternary gradients. The dynamic mixer adds 35 μL
of gradient delay volume to the system. You can plumb the system to bypass the dynamic
mixer for mobile phases that do not require additional mixing. 600 Pumps that do not
contain the dynamic mixer have an outlet adapter installed in its place.
As the pistons move back and forth within the displacement chambers, a small quantity of
mobile phase leaks behind the primary piston seals. With buffered mobile phases, a precipitate
(also known as creep or salt build-up) forms as the liquid fraction evaporates. The abrasion
caused by this precipitate can scratch the piston and shorten the life of the piston seals. With
the seal wash pump option, a large reservoir above the pump continuously feeds the pump
and flushes the piston guide bushings to prevent the precipitation of salt crystals behind the
primary piston seals.
Status LEDs
If the 600 Pump is housed in a modular housing, four status LEDs provide information about
the pump. Table 2 describes the four status LEDs.
Table 1. Status LEDs for the 600 Pump
LED Status Description
Power Green The pump is switched on and is receiving power.
Comm Amber The pump is not communicating with the data system PC.
Green Communication to the data system PC has been established.
Flashing green A pump program is downloading from the data system PC.
Specifications
The tables in this section list the pump’s physical dimensions, pump and degasser operating
specifications, and electronic specifications.
Table 2. Physical dimensions
Specification Description
Height (with tray) 178 mm
Width 355 mm
Depth 457 mm
Weight 13.8 kg
Installation
This chapter provides instructions for the initial installation of your 600 Pump.
Note Installation procedures vary depending on the instrument control or data system
that controls the 600 Pump. Refer to system installation documentation when applicable.
Contents
• Installation Checklist
• Unpacking and Inspecting the Instrument
• Making the Initial Instrument Preparations
• Connecting the Power Cable
• Connecting the Pump to the Computer
• Connecting the Pump to Other System Modules
• Connecting the Solvent Lines
• Installing the Optional Dynamic Mixer
• Installing the Optional Seal Wash Pump
• Setting up Continuous Flushing
• Connecting the Pump to the Autosampler
• Configuring the Pump to the Software
• Powering On the Pump
Installation Checklist
This section provides a brief summary of the steps that must be completed for the proper
installation of your 600 Pump:
❒ Unpack and inspect your instrument (see “Unpacking and Inspecting the Instrument”
on page 10).
❒ Read safety precautions in “Safety Information” on page vi.
❒ Position the pump appropriately (see “Making the Initial Instrument Preparations” on
page 11).
❒ Make back panel connections (see system installation procedures or “Connecting the
Pump to the Computer” on page 12 and “Connecting the Pump to Other System
Modules” on page 14).
❒ Connect solvent lines (see “Connecting the Solvent Lines” on page 15).
❒ Connect the power cord and turn on the instrument (see “Connecting the Power
Cable” on page 11).
❒ Configure the pump to the instrument control or data system (contact Thermo Fisher
Scientific technical support).
(Name) (Date)
The shipping container contains the items listed in Table 7. Depending on the instrument in
which the pump is used, an accessory kit that includes additional tools and tubing might be
included in the system packaging. Ensure that you have received all the items listed on the
packing list. If any items are missing, contact your Thermo Fisher Scientific field service
engineer immediately.
CAUTION Never connect the pump to an electrical line source that has power drops or
fluctuations of 10% below the nominal rated line voltage.
2. Connect the series B plug of the USB cable to the USB port on the back of the 600
Pump.
Figure 6. 600 Pump back panel
RS232
connector
USB series B
socket
See Figure 7 for a description of the back panel connections. Table 8 and Table 9 describe the
Accela system interconnect cable socket and D-sub 25-pin connector.
Figure 7. Back panel connections
Accela System USB series B
interconnect cable socket
socket
4. Insert the solvent reservoir filter and inlet line into one of the solvent reservoir bottles,
and screw the cap onto the solvent reservoir bottle until it is secure.
Note The cap is a two-piece assembly. The upper section (Figure 8) snaps onto a
threaded section. You can screw the threaded section onto the bottle and snap on the
upper section after installing the tube, or, if you are replacing existing tubing, you can
unscrew the entire cap from the bottle.
5. Position the bottle in the solvent platform, allowing the solvent inlet line to hang down
along the left side of the system.
6. Repeat step 1 through step 5 for solvents B, C, and D if applicable.
CAUTION Do not use solvents containing Freon and perfluorinated solvents, such as
Fluorinert, FC-75, FC-40, and Fomblin perfluoro polyether solvents from Ausimont.
They adversely affect the Teflon AF degassing membrane.
CAUTION To prevent personal injury, observe good laboratory practice when handling
solvents, changing tube lines, or both. Consult the pertinent material safety data sheets for
the solvents used for HPLC analysis.
2. Push a length of tubing through a PEEK 1/8 inch externally threaded ¼-28 fitting and
slide a ferrule over the tubing end. Cut the Teflon tubing so that the end is flat. See
Figure 10.
Figure 10. Tubing with nut and ferrule installed
3. Screw the ¼-28 fitting onto outlet port A on the vacuum degasser. Connect the other end
of the tubing to inlet valve A on the LDA. See Figure 11.
Note Use the supplied extender tool to tighten the plastic connectors (see Figure 12).
Be careful not to over tighten them, or risk damaging the threads.
Extender tool
4. Repeat step 1 through step 3 to connect additional solvent lines. Use ports B, C, and D.
5. Using the supplied plugs, plug all unused ports. Press the ports in by hand.
6. Continue with “Priming the Degassing Membranes” on page 19.
Tool Use
1/4 in. open-end wrench Disconnect the high-pressure lines
#2 Phillips head screwdriver Disconnect the chassis from the housing
2.5 mm ball driver wrench Disconnect the screws that secure the blind outlet module
or the dynamic mixer to the front panel of the pump
CAUTION To prevent electrical shock, turn off the pump and disconnect it from line
power before you install the dynamic mixer.
3. Using a 1/4 in. open-end wrench, disconnect the static mixer from the blind outlet
module (see Figure 15 and Figure 16).
4. Disconnect the tubing between the pump and the autosampler, using a 1/4 in. open-end
wrench, disconnect the high-pressure solvent line from the blind outlet module.
Figure 15. View of the static mixer connection to the blind outlet module
Static mixer connection to Blind outlet
the blind outlet module module
Figure 16. View of the static mixer disconnected from the blind outlet module
Static mixer
5. Using a #2 Phillips head screwdriver, remove the screw that secures the metal plate
covering the LED board, and then remove the metal plate (see Figure 17).
6. Disconnect the LED cable from the left door of the pump.
7. Using a #2 Phillips head screwdriver, remove the four screws that secure the pump chassis
to the housing, and then pull the chassis out of the housing (see Figure 17).
8. Using a 2.5 mm hex ball driver, remove the two screws (with washers) that secure the
blind outlet module to the front panel of the pump (see Figure 18). Set the screws and
washers aside. You use them in step 12 to secure the dynamic mixer to the pump.
Figure 18. View of the back side of the front panel
9. Remove the blind outlet module from the front panel of the pump (see Figure 19).
Figure 19. View of the blind outlet module being removed from the front panel
Blind outlet module
10. Using a 1/4 in. open-end wrench, disconnect the static mixer from the LDA
(see Figure 20).
Figure 20. view of the connection between the LDA and the static mixer
Connection between LDA and
static mixer
11. Insert the dynamic mixer into the port vacated by the blind outlet module
(see Figure 21).
Figure 21. View of the dynamic mixer inserted into the port vacated by the blind outlet
module
Dynamic mixer
12. From the back side of the front panel (see Figure 22), secure the dynamic mixer with the
screws that you removed in step 8 on page 22.
Figure 22. View of the dynamic mixer from inside the pump chassis
Screws that
secure the
dynamic mixer
to the pump
13. Using a 2.5 mm hex ball driver, tighten the screws that secure the dynamic mixer to the
pump (see Figure 22).
14. Push the pump chassis back into the housing, and then secure the chassis to the housing
with the four screws that you removed in step 7 on page 22.
15. Reconnect the LED cable to the LED board.
16. Reconnect the metal plate that covers the LED board. Using a #2 Phillips head
screwdriver, secure the metal plate to the left door of the pump.
17. Reinstall the static mixer as follows:
a. Using a 1/4 in. open-end wrench, reconnect the static mixer to the LDA.
b. Using a 1/4 in. open-end wrench, connect the static mixer to the inlet of the dynamic
mixer (see Figure 23).
Figure 23. View of the static mixer connected to the LDA and the dynamic mixer
LDA outlet Static mixer Dynamic mixer inlet
18. Connect the outlet of the dynamic mixer to the inlet of the autosampler’s heat exchanger
(see “Connecting the Pump to the Autosampler” on page 31).
To install the optional seal wash pump, you must have these tools:
Tool Use
#2 Phillips head screwdriver Disconnect the LED cover plate from the inside of the
pump’s left door and the chassis from the housing
2.5 mm hex ball driver Disconnect the screws that secure the cover plate to the
front panel and secure the seal wash pump to the front
panel
4. With the cable connector facing up (see Figure 25), slide the seal wash pump into the
front panel of the pump (see Figure 26).
Figure 25. Seal wash pump
Cable connector
Figure 26. View of the seal wash pump placed into the front panel of the pump
5. Using the screws that you removed in step 3 on page 26, secure the seal wash pump to the
front panel of the pump (see Figure 27).
Figure 27. View of the seal wash pump secured to the front panel of the pump
Screws that secure the seal wash pump
CAUTION To prevent electrical shock, turn off the pump and disconnect it from line
power before you install the seal wash pump.
7. Using a #2 Phillips head screwdriver, remove the screw that secures the metal plate
covering the LED board, and then remove the metal plate (see Figure 17 on page 22).
8. Disconnect the LED cable from the left door of the pump.
9. Using a #2 Phillips head screwdriver, remove the four screws that secure the pump chassis
to the housing, and then pull the chassis out of the housing (see Figure 17).
10. Using the cable supplied with the seal wash pump, connect the seal wash pump to
Valve E on the Controller PCB (see Figure 28).
Figure 28. View of the connection between the seal wash pump and the Controller PCB
Cable connection to
seal wash pump
Cable connection to
Controller board
11. Push the pump chassis back into the housing, and then secure the chassis to the housing
with the four screws that you removed in step 9 on page 28.
12. Reconnect the LED cable to the LED board.
13. Reconnect the metal plate that covers the LED board. Using a #2 Phillips head
screwdriver, secure the metal plate to the left door of the pump.
14. Remove the inner wash line from the guide bushings (see Figure 29).
Figure 29. View of the guide bushings with the inner wash line removed
Inner wash line removed
15. To set up the seal wash pump for continuous flushing, go to the next procedure.
2. Connect the other end of the tubing to the appropriate port on the autosampler injection
valve. Contact technical support for information on the autosampler injection valve
connections.
Maintenance
To perform preventive and corrective maintenance procedures for the 600 Pump, you must
have access its components. This chapter also describes how to remove and dismantle the
LDA and basic maintenance procedures.
CAUTION Read all safety precautions before performing any maintenance procedures on
the 600 Pump. See “Safety Information” on page vi.
CAUTION The instrument contains voltage lines. Switch off the power and disconnect
the power cable prior to servicing the instrument. There is no need to open the pump
housing, however, as all user serviceable components are outside of the instrument.
Contents
• Maintenance Schedule
• Accessing the Liquid Displacement Assembly Components
• Replacing the Primary Piston Seals
• Replacing the Secondary Piston Seals
• Cleaning the Check Valves
• Cleaning the Pistons
• Cleaning the Instrument Casing
• Flushing the Pump
• Flushing the Piston Guide Bushings
• Purging Air from the Piston Chambers
• Cleaning the Displacement Chambers and the Inlet Module
• Replacing the Fuses
Maintenance Schedule
To ensure the 600 Pump functions properly, perform the preventive maintenance procedures
at regularly scheduled intervals. Perform the corrective maintenance procedures when you
observe poor performance or when you have been instructed to do so by a Thermo Fisher
Scientific service representative.
Table 10 lists the minimum recommended frequencies for performing the maintenance
procedures described in this manual. However, the volume and type of mobile phases that
flow through the pump affect the how often you should perform these procedures. If
warranted, increase the frequency of these procedures.
Table 10. Maintenance frequency
Procedure Frequency
Replacing the Primary Piston Seals Every 6 to 12 months depending on usage
Replacing the Secondary Piston Seals Every 6 to 12 months depending on usage
Cleaning the Check Valves Every 6 to 12 months if you use buffered mobile
phases
Cleaning the Pistons Every 6 to 12 months depending on usage
Cleaning the Instrument Casing As needed
Flushing the Pump Once per day if you use buffered mobile phases
and you do not have the seal wash pump installed
Flushing the Piston Guide Bushings Once per day if you use buffered mobile phases
and you do not have the seal wash pump installed
Purging Air from the Piston Chambers As needed
Cleaning the Displacement Chambers As needed
and the Inlet Module
Replacing the Fuses As needed
Once you have accessed the liquid displacement assembly (LDA) components, perform all the
necessary maintenance procedures that require access to components before reassembling the
LDA. The following procedure require access to the LDA components:
• “Replacing the Secondary Piston Seals” on page 53.
• “Cleaning the Check Valves” on page 55.
• “Cleaning the Pistons” on page 56.
• “Cleaning the Displacement Chambers and the Inlet Module” on page 58.
Inlet valves
Solvent lines
Figure 34. View of the pump with the solvent lines disconnected
4. Using a wrench, unscrew the static mixer at the outlet of the LDA and at the inlet of the
outlet module.
Figure 35. View of the pump with the static mixer
5. Using a 2.5 mm hex tool, loosen the two 2.5 M screws on the Z-arm cover and remove
the cover.
Figure 36. View of the pump with the Z-arm cover in place
6. Using a 4mm hex tool, loosen the two clamping screws by three or four turns.
Figure 37. View of the pump showing the two clamping screws
8. Gently push the piston heads to the left so that they do not contact the Z-arms.
Figure 39. View of the pump with the Z-arms and pistons not touching
Figure 40. View of the pump showing the location of the LDA attachment screws.
CAUTION When loosening the screw near the inlet valves, do not touch or apply
force on inlet valve D to prevent breakage.
10. Gently pull the LDA off the two guide pins. Do not let the pistons touch the Z-arms.
Figure 41. Removing the LDA from the pump
Piston back
flush tubes
2. Put the LDA onto a level surface. Using the 4 mm hex tool, loosen the fastening screws
that secure the holding hooks until you release the holding hooks.
Figure 43. View of the LDA showing the location of the holding hooks
fastening screws
3. Carefully remove the pistons, guide bushings, and holding hooks from the LDA body.
Figure 44. LDA with the pistons, guide bushings, and holding hooks removed
Holding hook
Piston
Guide
bushings
4. Using the 4 mm hex tool, loosen the compression lid set screw. Remove the compression
lid by sliding it out (see Figure 45).
5. Slide out the outlet module, check valves, and displacement chambers.
Figure 46. Sliding out the check valves, displacement chambers, and outlet modules
Displacement Displacement
chamber chamber
4. Verify that the remaining check valve is facing the appropriate direction and insert it into
the main body of the LDA.
5. Insert the upper displacement chamber into the main body of the LDA.
6. Insert the outlet module into the main body of the LDA.
Figure 47. Inserting the displacement chambers, check valves, and outlet module into the LDA
Displacement Displacement
chamber chamber
7. Carefully slide the compression lid into the LDA main body (see Figure 49). If the
compression lid does not fit, verify that the parts are installed correctly.
8. Using the 4 mm hex tool, firmly tighten the compression lid screw.
Figure 50. Tightening the compression lid screw
9. If you have not already done so, insert the pistons into the holding hook bores and
bushing guides. Align the piston wash tubes so that they protrude away from the holding
hook. See Figure 70.
10. Carefully insert the piston into the bore of the displacement chamber and seal. At the
same time, tighten the holding hook using the 4 mm hex key. Repeat the same procedure
for the second piston unit.
Figure 51. Inserting the piston/guide bushing/holding hook assembly
2. Align the pistons in such a way that they do not touch the Z-arms while you insert them.
Do not insert the piston springs at this time.
3. Reinstall the piston back flush tubes.
Figure 54. LDA with the piston back flush tubes reinstalled
Piston back
flush tubes
4. Place the LDA onto the drive unit using the two alignment pins as a guide (see
Figure 55).
Alignment pins
CAUTION When tightening the screw near the inlet valve, do not touch or apply
force on inlet valve D.
Figure 56. Tightening the LDA attachment screw near the inlet valve
6. Insert the piston springs. The lip of the spring faces toward the right.
Figure 57. View of the pump with the piston springs installed
8. Replace the Z-arm cover and tighten the screws using a 3 mm hex key (see Figure 59).
Figure 59. View of the pump with the Z-arm cover replaced
M3 2.5 mm
screws
Static mixer
Figure 61. View of the pump with the solvent lines replaced
Solvent lines
Inlet valves
Piston
Guide bushings
Holding hook
4. Insert the piston into the primary seal no more than 3 mm. The primary seal is located
inside the displacement chamber.
Figure 64. Displacement chamber with piston inserted
Displacement chamber
Piston
5. Gently remove the seal by pulling on the piston while applying side force. Do not allow
the piston to touch any metal parts.
6. Remove the primary piston seal from the piston and discard the primary piston seal.
7. Insert the piston into the guide bushing and slide the replacement seal over the piston.
Ensure that the replacement seal is flush with the guide bushing and the piston protrudes
out of the seal by 6 mm.
Figure 65. View of piston, guide bushing, and replacement piston seal
piston seal
6 mm
8. Insert the piston into the bore of the displacement chamber. Do not apply any side load
onto the seal during this process.
Figure 66. Using the guide bushing and piston to insert the piston seal
9. Hold down the guide bushing and slide the piston up until it is out of the seal.
Figure 67. Sliding the piston out of the seal
10. Remove the guide bushing/piston assembly from the displacement chamber.
Figure 68. Removing the bushing/piston assembly from the displacement chamber
11. Repeat step 1 through step 8 to replace the primary piston seal of the second
displacement chamber with a new piston seal.
12. Continue with one of the following:
• Perform other procedures that require access to the LDA. See “Accessing the Liquid
Displacement Assembly Components” on page 35.
• Rebuild the LDA and replace it onto the pump. See “Reassembling the Liquid
Displacement Assembly” on page 42 and “Reinstalling the Liquid Displacement
Assembly onto the Pump” on page 46.
5. Remove the secondary piston spring from the piston guide bushing.
6. Carefully slide the piston guide bushing over the piston and align the piston wash tubes
so that they face away from the holding hook.
Figure 70. Proper alignment of the piston wash tubes
Piston wash
tubes
7. Gently push the new secondary piston seal into its groove by pushing the guide bushing
up toward the end of the holding hook.
3 Maintenance
Cleaning the Check Valves
8. Repeat step 4 through step 6 for the remaining secondary piston seal.
9. Continue with one of the following:
• Perform other procedures that require access to the LDA. See “Accessing the Liquid
Displacement Assembly Components” on page 35.
• Rebuild the LDA and replace it onto the pump. See “Reassembling the Liquid
Displacement Assembly” on page 42 and “Reinstalling the Liquid Displacement
Assembly onto the Pump” on page 46.
CAUTION Do not use an ultrasonic bath to clean the check valves. Sonicating the check
valves can damage their small ruby balls.
If you use mobile phases that contain a high concentration of salts and you do not have the
seal wash pump, flush the piston seals at least once a day. The seal wash pump is an optional
pump that periodically flushes the piston guide bushings.
2. Dismantle the LDA. See “Dismantling the Liquid Displacement Assembly” on page 40.
3. Pry the piston seals out of the bores of the displacement chambers:
a. Carefully insert a piston into the hole in the center of the seal without pushing the
piston further into the bore of the displacement chamber.
b. Gently pry the seal out of the bore.
4. Using a needle, graphite pencil, or equivalent tool, carefully remove the PEEK film seals.
Be careful not to scratch the displacement chambers as you remove the film seals, as this
might cause leaking.
Note Both ends of the lower displacement chamber, the lower end of the upper
displacement chamber, and the bottom of the outlet module contain a PEEK film seal
for a total of four film seals.
5. Place the displacement chambers and the inlet module in a container of water and place
the container in an ultrasonic bath for 10 minutes.
6. Place the displacement chamber and the inlet module in a container of acetone and place
the container in an ultrasonic bath for 10 minutes.
7. Install the PEEK film seals if they are not damaged; otherwise replace them with
new ones.
8. Replace the primary and secondary piston seals. See “Replacing the Primary Piston Seals”
on page 51 and “Replacing the Secondary Piston Seals” on page 53.
9. Continue with one of the following:
• Perform other procedures that require access to the LDA. See “Accessing the Liquid
Displacement Assembly Components” on page 35.
• Reassemble the LDA and place it onto the pump. See “Reassembling the Liquid
Displacement Assembly” on page 42 and “Reinstalling the Liquid Displacement
Assembly onto the Pump” on page 46.
Fuse holder
3. Remove the fuses and replace with 5×20mm, 2A, 250V, TL, slo-blo fuses.
4. Plug the pump power cord into the power outlet.
Troubleshooting
This chapter provides procedures for several diagnostic tests and troubleshooting guides for
the pump and degasser.
Contents
• Diagnostic Tests
• Error Messages
• Troubleshooting Guides
Diagnostic Tests
Run the tests in this section if you want to validate the pump’s performance or if you suspect
the pump is not functioning properly.
2. Fill a mobile phase reservoir with HPLC-grade methanol and connect it to inlet
channel A.
3. Use the instrument control system to set channel A to 100%.
4. Set the flow rate to 1mL/min.
5. Do one of the following:
• Connect the other end of the restrictor cartridge to an instrument that measures flow
rate.
• Connect the other end of the restrictor cartridge to a length of tubing and place the
tubing into an empty, 10 ml graduated cylinder.
6. Start the pump.
7. Wait 2 minutes.
8. Stop the pump.
9. If you do not have a flow rate meter, measure the fluid in the graduated cylinder. The
actual flow rate equals the measured volume divided by 2.
10. Compare the actual flow rate with the set flow rate. Under these conditions, the accuracy
of the flow should be ±0.5%.
2. Fill a mobile phase reservoir with HPLC-grade methanol and connect it to inlet
channel A.
3. Open the purge valve.
4. Set the pressure unit to bar in the instrument control software (See the instrument
control software manual).
5. Verify that the pressure indicated by the instrument control software is 0.
3. Fill a mobile phase reservoir bottle with HPLC-grade methanol and connect it to inlet
channels A and B.
4. Fill a mobile phase reservoir bottle with HPLC-grade methanol spiked with 0.5% acetone
and connect it to inlet channels C and D.
5. Set each channel to 25%.
19. Compare your results for each step gradient with the pass criteria in Table 14.
Table 14. Pass criteria for gradient test.
Gradient forming performance Pass criteria
20% Step Height: _______% 19.5–20.5%
40% Step Height: _______% 39.5–40.5%
60% Step Height: _______% 59.5–60.5%
80% Step Height: _______% 79.5–80.5%
20% Step Height: _______% 19.5–20.5%
40% Step Height: _______% 39.5–41.5%
60% Step Height: _______% 59.5–60.5%
80% Step Height: _______% 79.5–80.5%
20. If any value falls outside the range, call technical support.
Error Messages
Table 15 describes the pump-related error messages that might appear in the system control
software event log or status area.
Table 15. Error messages
Error message Description
Low Pressure Error The system pressure has fallen below the set minimum pressure.
High Pressure Error The system pressure has risen above the set maximum pressure.
Index Error The system expected an index signal that never arrived. Physical
problem with the hardware or connectors: Contact an
authorized dealer.
Motor Error The motor is malfunctioning. Indicates that too much torque is
required to drive the pump accurately.
Position Error The system was unable to maintain accurate positioning of the
motor. Restart the pump.
Pressure Sensor Error The pressure transducer is malfunctioning.
Solenoid Error A proportioning valve’s solenoid has been disconnected or is
defective. Physical problem with the hardware or connectors:
Contact an authorized dealer.
Troubleshooting Guides
This section provides troubleshooting guides for the pump and degasser.
Calibration Tools
Use the tools described in the following table when performing the procedures described in
this chapter.
Table 18.
Description Used for... Picture Part number Quantity
Force sensor calibration wizard Force sensor
downloaded from calibration
http://support.flux.ch/milestones/200
9/04/index.html
Set manometer reference Force sensor Flux part 1
calibration number 50098
includes the following items:
• 50090 Manometer Adapter,
• 50092 Manometer Reference
• 50093 Cable Converter RS485
• 50099 Certificate Of Compliance
• 50164 Assembly Manometer
Adapter
Hypersil Gold 30x2.1 5μm Force sensor 1
calibration
Table 18.
Description Used for... Picture Part number Quantity
GJC HPLC liquid flowmeter Transfer volume Flux part 1
calibration number 2000,
model number.
5025000
❖ To perform the Thermo Rheos Series Pump Force Sensor Calibration wizard
1. Connect the Accela 600 Pump outlet to the inlet of the Keller LEO 1 Manometer.
Figure 79. Manometer attached to LDA outlet
8. At the wizard home page, click Next. The following window appears.
9. Click on the Pump ID drop down menu, select the pump you want to calibrate, and
click Next. The following window appears.
10. Click on the Manometer ID drop down menu, select the manometer you want to use,
and click Next. The following window appears.
13. Unscrew the two M3x40 mm locking screws and remove the Z-arm cover.
Purge valve
Clamping screws
Piston springs
16. Remove the two piston springs and click Next. The following window appears.
22. In the Operator name field, type your name and click Next. The following
window appears.
2. Set the GC Flow meter to Terminal mode by doing the following, or see the GJC
flowmeter manual page 6-9.
a. Disconnect the flow meter power.
b. Unscrew the flow meter cover.
c. Reset Dil Switch 7 to off.
9. Start the Thermo Rheos Series Transfer Volume Calibration Wizard. The wizard home
page appears.
11. Click on the Pump ID drop down menu, select the pump you want to calibrate, and
click Next. The following window appears.
12. Click on the Flow Meter ID drop down menu, select the flow meter you want to use, and
click Next. The following window appears.
Table 22.
Field Entry
Pressure SD allowed 5
Flow Rate Precision 0.5%
15. Click Next. When the calibration is complete, the “Calibration successfully completed”
message appears.
17. In the Operator Name field, type your name and click Next. The following
window appears.
1. Connect the Accela 600 Pump Outlet to the back pressure restrictor. Use restrictor type
1000 PSI Upchurch P-455.
11. In the Pump ID field, select the pump you want to validate and click Next. The following
window appears.
12. In the Detector field, select Micro UV Detector K-200 and click Next. The following
window apepars.
Field Description
Intensity SD, mAU Enter the intensity standard deviation
limit in mAU.
Step height deviation, % Enter the percent step height deviation
limit.
Intensity drift, mAU Enter the intensity drift limit in mAU.
Correlation Enter the correlation value. The expected
minimum is 0.999.
17. In the Operator name field, type your name and click Next. The following window
appears.
Software Compatibility
Table 23.
Pump Type Software Compatibility Firmware
TMO 600 Pump Aria OS 2332e
Accela 600 Xcalibur
Table 23.
Pump Type Software Compatibility Firmware
Transcend 600 Chromquest
Janeiro II 3.0
Firmware Manager Tool
Pump Identification Wizard
Force Sensor Calibration
Wizard
Transfer Volume Calibration
Wizard
Gradient Validation Wizard
Replacement Parts
This chapter contains a list of replaceable parts and their corresponding part numbers. To
place an order, contact your local Thermo Fisher Scientific support representative or contact
technical support.
Main Body
The replaceable parts listed in Table 25 correspond to the items listed in Figure 81.
Table 25. Main assembly parts
Item Quantity Description Part number
1 1 Degassing bracket assembly Accela 50038
2 1 Screw ISO 4762 M3×8 50026
3 1 Cover purge pump Accela 600 50025
4 1 LDA Accela 600 50007
5 1 Mixer static Accela 600 50009
6 1 Placeholder Accela 600 50010
7 1 Piston spring stainless steel 20008
8 1 Z-Arm cover Accela 600 50087
9 1 Zyl I-6kt Sch A2/BN 00611 M3×40 90072
10 1 Front panel Accela 600 5022
11 1 Drip tray Accela 600 50027
12 1 Cable valve D Accela 600 50014
13 1 Inlet Module Accela 600 50006
14 1 Cable valve C Accela 600 50013
15 1 Cable valve B Accela 600 50012
16 1 Cable valve A Accela 600 50011
(1) Degassing bracket (2) Screw ISO (3) Cover (4) LDA (5) Mixer static
4762 M3×8 purge pump
(6) Placeholder
(3) Displacement
chamber upper
(4) Check valve gamma
titan
(5) Displacement
chamber lower
Table 28.
Description Picture Part number Quantity
Connector AC with Cable Accela 600 50029 1
Table 28.
Description Picture Part number Quantity
Cable Interface Board Degasser Accela 50054 1
600
Pump Components
The components in the following table refer to the items in Figure 84. Other pump
components appear in “Miscellaneous Parts” on page 98 and Chapter 6, “Replacement
Parts.”
Table 29. Pump components
Item number Quantity Description Part number
1 1 Degassing bracket assembly Accela 600 50038
2 1 Tray complete Accela 600 50039
3 1 Drive unit Accela 600 50001
4 1 Controller Accela 600 50002
5 1 UHP LC interface board holder Accela 600 50036
6 1 UHP LC interface board holder Accela 600 50035
7 1 Power supply Accela 600 50045
Miscellaneous Parts
The following table describes components not pictured in Figure 84.
Table 30. Miscellaneous parts
Description Picture Part number Quantity
Static mixer assembly 5009 1
Figure 85.
Index
A inlet module 58
instrument casing 56
Accela system interconnect cable 14
pistons 56
Accuracy, testing 73
computer, connecting 12
autosampler, connecting 31
connections
low-pressure solvent lines 15
B USB cable 12
back panel Connector AC with Cable Accela 600 96
clearance requirements 11 continuous flushing 5, 30
computer connections 12
connecting the power cables 11 D
connecting to other modules 14
D-sub 25-pin connector 14 degasser
connecting solvents 16
description 2, 4
C priming membranes 16
Cable DC Power Accela 600 96 Troubleshooting Guide 69
Cable DMS Accela 600 97 detecting leaks 79
Cable D-Sub25 96 D-sub 25-pin connector 14
Cable Interface Board Acela 600 96 dynamic mixer 20
Cable Interface Board Degasser Accela 600 97 description 5
Cable RS232 Accela 600 96 installing 20
Cable USB Accela 600 96
Cable Valve A Acceka 600 95 E
Cable Valve B Accela 600 95 electronic specifications 7
Cable Valve C Accela 600 95 environmental conditions vii
Cable Valve D Accela 600 95
Cable Valve E Accela 600 97
cables
F
Accela system interconnect 14 film seals, removing 59
packing location 12, 14 flow rate
USB 12 testing 61
Calibration wizard 74 flow rate, accuracy 79
check valves Force Sensor technology 2
accessing 35 fuses, replacing 58
cleaning 55
checklist, installation 10 H
cleaning
high pressure error 67
check valves 55
displacement chambers 58 Hypersil Gold 30x2.1 5μm 71
S V
safety information vi vacuum degasser
seal wash pump 5 description 2
figure 27 vacuum degasser, description 4
installing 26 valve, purge 4
secondary piston seals, replacing 53
shipping container, contents 10 W
solenoid error 67
Witch with Bowden Cable Accela 600 97
solvent
platform 16
reservoir filter 16
solvent lines
connecting 15
solvent requirements ix
solvent reservoir bottles, assembling 15
specifications
degasser operating 7
electronic 7
physical dimensions 6
pump operating 6
system housing 2
T
Thermo Rheos Series Pump Force Sensor Calibration wizard
74
Thermo Rheos Series Transfer Volume Calibration Wizard
79
Tools, calibration 71
transfer volume calibration wizard 79
troubleshooting guide
degasser 69
pump 68
U
unpacking your pump 10
USB cable 12, 13
USB port 13