Quality Assurance In Diagnostic
Radiology
Robert G. Gould, Sc.D.
Professor and Vice Chair
Department of Radiology
University of California
San Francisco, California
Why Do Quality Control?
• Improve clinical results
• Preempt quality or safety problems
• Maintain standard of care
• Minimize patient radiation dose
• Satisfy government regulations
1
QC Testing
• Acceptance testing
– Upon installation prior to patient use
– Medical physicist
• Annual inspection
– Medical physicist
– Equipment vendor/service provider
• Daily and weekly tests
– QC technologist
Quality Control (QC)
• Team approach
– Radiologists, Medical Physicists,
Technologists
• Use eyes and experience
• Don’t “work around” problems
• Try to be preemptive
2
Mechanical Integrity
• Fix problems as soon as possible
– They only get worse
• If things become loose, tighten them!
Regulations
• Are more better?
• Are all of equal value?
• Do they cover all aspects of IQ and
safety?
– Should I stop when all the regulatory tests
are complete?
3
Quality Control
• Emphasize those tests that are
important to IQ and/or safety
– Concentrate on those functions that effect
quality and safety
– Minimize time on activities done primarily
to meet regulations
Digital Projection Imaging: QC
X-ray tube!
Collimator!
X-rays!
Patient!
Grid!
Detector!
4
X-Ray Tube Concerns
• Focal spot size
– Component in spatial resolution
• Worn anode
– Variation in intensity across field
– Increase in HVL due to metal coating on
inside of glass
• Instabilities, arching
Focal Spot Measurement
• When?
– Acceptance
– Annually
– Tube replacement
• How?
– Star pattern-measure spatial
resolution
– Pinhole camera
5
Spatial Resolution Measurement
• Image a lead bar
test pattern
• Assess using vendor
QC software to
determine contrast
of specific line pairs
– MTF can be obtained
• Determine along
both axis or at an
angle of 45°
Pinhole Camera
• The best Focal spot, F
– Shows emission distribution L1
– Difficult and time
consuming Pinhole
– Not possible for some tubes
• Use CR and large L2
magnification factor (≥ 5x) Image, S
• Careful alignment Detector
F = S (L1/L2)
= S [1/(M-1)]
6
Generator QC
• Consistent x-ray output for same
technical factors (KVp, mA, exposure
time)
• mA and time settings
– Should be linear
– Should be consistent
– mR/mAs should be a constant
• KVp calibration
Collimators
• Restrict primary X-ray beam to
detector size or a smaller
anatomic region-of-interest Tube housing
• Restriction and alignment X-ray tube
of X-ray beam to detector Collimator
Added
• Major component in radiation filtration
protection Light
Mirror
• Reduces scatter
X-rays
• Lower personnel exposure
Patient
• Improved image quality
• Reduction in patient radiation Grid
burden Detector
• Component in beam filtration
7
Collimator
• What?
– X-ray field - detector alignment
– X-ray - light field alignment
• When?
– Acceptance
– Annually
– Tube replacement
X-ray - Light Field Alignment
• Sum on opposite
sides < 2% of the
source-image
distance (SID) Mirror
Light
D
source
D
8
Filtration
• Minimum amount set by
regulation
• Determined by measuring
the HVL
Intensity
Peak tube
– Thickness required to Potential
reduce X-ray intensity to
half its initial value
– Measured in mm of Al
– Measure of X-ray beam 0 20
40 60 80 100 120
penetrance (hardness) Energy, KeV
Filtration/Beam Quality
Tube Minimum
• Indicated by
measuring half Potential, HVL,
KVp Mm Al
value layer (HVL)
• Need to measure at 50 1.5
only a single KVp 71 2.1
– Tube potential
indication should be 80 2.3
calibrated 100 2.7
120 3.2
9
Filtration
• Current digital R/F and angio systems
have variable filtration
– Combinations of Al and Cu of various
thicknesses
– Anatomic protocols automatically change
• Measure HVL at minimum filtration
Basic Imaging Geometry: Detector
Tube housing
• Converts X-ray intensity to X-ray tube
electrical signal Collimator
Added
• Major component of spatial filtration
resolution Light
• Major determinate of patient Mirror
dose X-rays
• Component of automatic Patient
exposure control system
Grid
Detector
10
Digital Detectors: Radiography
• What?
– Uniformity
– Artifacts
– AEC
• When?
– Acceptance
– Annually
– Component replacement
– Manufacturers recomendation
Digital Detectors: Flat Field Uniformity
• Digital detectors do not
respond uniformly across field
– Produces density variations
within the image
– ‘Structured’ noise
• Assessed by uniformity of pixel
values (eg. Mean and
standard deviation)
• Most systems have software Mean,
that automates testing standard deviation
11
Digital Detector Structured Noise
Normalized “flat-field”
correction matrix
“Flat field”
algorithm
Flat-field “corrected” image
Raw image
• Periodically generate new correction
matrix
– Follow manufacturers recommendation
– Often done by technologist
Digital Detectors: Artifacts
• Non-uniformities
• Dropouts and dead pixels
• Determined by imaging uniform plastic
block
– View with narrow W/L
12
Automatic Exposure Control (AEC)
• Should be able to maintain a pixel mean
value within ~15%
– Track with changes in KVp
• Clinically used range (~ 50 - 120 KVp)
– Track with changes in patient thickness
• 5 - 35 cm of water equivalent
Annual Testing - Key Measurements
• Mechanical integrity
• Linearity of mAs
• Half value layer
• X-ray field - detector size
• Light - x-ray field alignment
• Spatial resolution
• Artifacts/uniformity
• AEC consistency
13
Fluoroscopy QC
• What?
– Table-top exposure rate
– Automatic brightness control
• When?
– Installation
– Annually
– Major component changes
– Manufacturer’s recommendation
Typical Regulations
Fluoroscopic Equipment
• Table-top exposure rate cannot exceed 10
R/min
• During routine fluoroscopy the table-top
(patient entrance) exposure rate shall not
exceed 5 R/min for a typical patient
• Determined by use of a phantom equivalent to
8” of water
14
Automatic ExposureControl (AEC)
• Feedback mechanism that attempts to maintain a
constant brightness level from the center portion of the
output screen
– center weighted exposure meter
– adjusts the X-ray technique factors (mA and/or KV)
• Determinate in patient dose
Fluctuating patient entrance dose! Constant input exposure!
Constant
light level!
X-ray tube!
+ KV! Patient! Image intensifier!
+ mA!
Feedback loop!
Automatic Exposure Control
• For a given object size, should require
same kVp and mA
• California:
– 8 inch plastic (lucite) phantom
– 12 inch table-top to entrance surface
distance
– 6.25 x 6.25 inch field at table-top
– Record kVp and mA weekly
15
Fluoroscopy - Image Quality
• Image resolution pattern
– Bar pattern (line pairs/ mm)
• Contrast sensitivity
– Low contrast phantom
Problem: very subjective
Computed Tomography QC
• What?
– Dose
– Slice thickness/sensitivity profile
– Table incrementation accuracy
– Image quality factors
• When?
– Installation
– Annually
– Major component changes
– Manufacturer’s recommendation
16
Image Quality CT
• Uniformity
• Artifacts
• Linearity
• Noise
• Spatial resolution
• Contrast sensitivity AAPM phantom
CT Dose Measurements
• CTDI
• In air at isocenter (mR/mAs)
7.8
7.8
4.3
7.8
120
KVp
100
mAs
Cylindrical
PMMA
360°
rota<on
7.8
64
slice
CT
Body
32
cm
diameter
17
QC Challenges
• Man-machine interfaces
– What goes on in the software black box?
– How to test?
Cedars-Sinai CT Overexposures
• What happened?
• Brain perfusion procedures
– Used in stroke assessment
• Over-rode ‘default’ protocol
settings
– Protocols come with the
machine
– Changed technique factors that
effect dose
• Eight times the protocol dose
18
Cedars-Sinai CT Overexposure
• Went on for 18 months because no one
made the association of hair loss and
skin reddening with CT procedure
– 2-3 weeks after exposure before onset of
hair loss
Cedars-Sinai CT Overexposure
• Errors at multiple levels
– Originally caused by changing default
protocol
– Dose indicators appear at time of scan:
should have been recognized at time of
scan
– Radiologist should have realized overdose
from the images
• Not found during any QC testing
19
Conclusions
• QC is a necessary and valuable aspect of
x-ray imaging
• QC should be a meaningful endeavor not
just going through the motions
– React to problems before they interfere with
patient images
• Not all QC tasks are of equal value
– Concentrate on the important ones (those
that effect patient safety and image quality)
20