IV Pump Flo-Gard 6301
IV Pump Flo-Gard 6301
Table of Contents
Introduction.............................................................................................. 1 Physical Description ............................................................................... 3 Operating Instructions .......................................................................... 10 Precautions ...................................................................................... 10 Loading the Pumps ........................................................................... 11 Starting a Primary Infusion ............................................................... 15 Starting a Secondary Infusion........................................................... 18 Using Volume-Time Programming.................................................... 19 Changing the Flow Rate While the Pump Is Running....................... 21 The OPTIONS Key ........................................................................... 22 Using the Programmed Delivery Profile Feature........................ 22 Computer Control ....................................................................... 25 Computer Monitoring.................................................................. 26 Locking the Front Panel .................................................................... 26 Battery Powered Operation .............................................................. 27 FLOW CHECK Information Display .................................................. 27 Cleaning and Storage ........................................................................... 28 Alerts and Alarms ................................................................................. 29 Technical Specifications ...................................................................... 34 Configurable Options............................................................................ 36 Warranty and Service Information ....................................................... 41 Warranty ........................................................................................... 41 Service Information .......................................................................... 42
7-19-B1-564
Change record 7-19-01-564 Original issue: 7/92 Rev. A: 9/92 Rev. B: 05/93 Page Front cover i ii 14 5 68 9 1014 15 1617 18-30 3132 33 34 3538 3942 Back cover Revision 0 B B 0 A 0 A B 0 B 0 B 0 A 0 B B
ii
7-19-B1-564
Introduction
The Flo-Gard 6301 Dual Channel Volumetric Infusion Pump from Baxter can deliver a wide variety of fluids over a broad range of infusion rates. The device's features include: Uses only standard Baxter solution administration sets. Two separate pump channels allow it to do the work of two conventional pumps, resulting in space savings. Configurable settings allow the device to be customtailored to best suit the hospital's needs. Computer control capability allows remote control or monitoring of the device via a standard RS-232C interface. Infuses a wide variety of fluids, including blood and fat emulsions. Suitable for use in epidural administration. Safety clamp automatically occludes the tubing when the pump door is opened. Occlusion sensors detect both upstream and downstream restrictions. Sensitivity of the downstream occlusion sensor is selectable through the device's configuration. Flow check display shows resistance to flow. Ultrasonic air-in-line detectors. Sensitivity of the air detectors is selectable through the device's configuration. Each pump channel features a Programmed Delivery Profile that enables programming of up to 10 sequential infusion programs for situations where ramping of medications are indicated.
7-19-B1-564
Slide clamp loading option provides flow shut-off when I.V. set is removed from the pump. Locking control panel helps prevent patient tampering. Each pump channel features an independent secondary medication program. The channel automatically switches over to the primary program upon completion. Volume-Time programming automatically calculates flow rate. Automatic self-test routine checks for proper function before use. Five-hour memory retains infusion data after power-off. Easily replaceable fuse, battery, and power cord. Access to test points and internal program simplifies servicing and software upgrades. Device can be configured to display a hospital area designator, such as ONCOLOGY or CARDIAC ICU, upon power-up.
7-19-B1-564
Physical Description
Items 1 through 9 are associated with Pump 1 and are described below. The controls associated with Pump 2 are identical and function in exactly the same manner. Items 10 through 28 are common to the operation of both pumps. All front panel items are shown in Figure 1.
Front Panel
ITEM 1. Pump 1 ONOFF/CHARGE key FUNCTION Turns Pump 1 on and off. The internal battery charger remains on regardless of the ON-OFF/CHARGE key as long as the device is plugged in. Opens and closes pump door for I.V. set loading and removal.
Stops Pump 1 until further instructions are given. The message STOPPED appears when the key is pressed. An alert sounds if Pump 1 is stopped for more than two minutes. Clears all programming alerts while pump is running. Allows keypad and other controls common to both pumps to accept data for Pump 1 programming. Yellow LED lights to indicate that Pump 1 is selected. Shows rate, volume to be infused (VTBI), and total volume infused for Pump 1 primary and secondary infusion programs. The decimal point is displayed as an underscore (example: 99_9). Shows all Pump 1 messages and messages that apply to the device as a whole. Red LED that blinks on and off during a Pump 1 or general alarm, accompanied by a message display and a repeated sequence of three tones. An alarm indicates that Pump 1 or the device as a whole requires immediate attention. Green LED which is constantly lit while Pump 1 is pumping.
Yellow LED which lights during Pump 1 and general alerts, accompanied by a message display and a repeated single tone. An alert indicates that Pump 1 or the device as a whole needs timely attention.
7-19-B1-564
Front Panel
ITEM 10. BACK LIGHT key FUNCTION Backlights the displays when pressed. Pressing the key again turns the backlight off. If the device is operating on battery power, the backlight remains on for 60 seconds when the BACK LIGHT key is pressed. Temporarily silences an audible alarm or alert for two minutes, unless another alarm or alert occurs on either pump within the two-minute silence period. All messages remain displayed. Lights when theTOT VOL STATUS key is functioning as the NEXT key. Displays total volume delivered for each pump. The key is also used to advance to the next step when programming a delivery profile. (See NEXT legend, item 12 above.) Resets the total volume delivered by the selected pump to zero if pressed when the pump is stopped. Enters the time over which an infusion is to take place during Volume-Time programming. Allows programming of the primary infusion rate for the selected pump. Allows programming of the primary VTBI for the selected pump. Starts the primary infusion for the selected pump. Always lit while the device is plugged in and the battery is charging. Lights when the device is being queried by the computer during computer monitoring. Lights when the device is operating on battery power. Lights when the device is being controlled by a remote computer. Clears any programming values currently being entered. Starts the delivery of the secondary infusion for the selected pump.
21. Battery icon 22. COMPUTER CONTROL legend 23. CLR key 24. SEC START key
7-19-B1-564
7 6 5
9 10
11 12 13
14
15 16
BACK LIGHT
ALARM
PUMPING
ALERT
SILENCE
NEXT
ALERT
PUMPING
ALARM
4 3 2
P U S H
PU MP
17 18
TIME
PUMP
PRI RATE
PRI VTBI
PRI START
2
P U S H
STOP
7 4
8 5 2
0
SEC VTBI
9 6 3
CLR
STOP
ON OFF CHARGE
1
.
SEC RATE
ON OFF CHARGE
OPTIONS
19 20 22 21
28 27
Flo-Gard 6301
26
25
24
23
Front Panel
ITEM 25. SEC VTBI key FUNCTION Allows programming of the secondary VTBI for the selected pump. Allows programming of the secondary infusion rate for the selected pump. Programming values and decimal point are entered with these keys. Allows the device to enter an enabled optional mode. These may include the Programmed Delivery Profile (PDP) or Computer Control modes. If Programmed Delivery Profile and/or Computer control is available, the OPTIONS key is lit.
7-19-B1-564 Rev. A
Figure 2 shows Pump 2 only. Pump 1 has identical features, which function in the same manner.
2. Pumping fingers
4. Air sensor
5. Safety clamp
7-19-B1-564
5 6 7
7-19-B1-564
The following items are located on the rear of the device and are shown in Figure 3.
5. Fuse
6. Power cord
8. COMMUNICATIONS PORT
9. VOLUME knob
7-19-B1-564
2 1 3
9
VOLUME
COMMUNICATIONS PORT
5 6 4
7-19-B1-564 Rev. A
Operating Instructions
Precautions
General
DANGERPossible explosion hazard if used in the presence of flammable anesthetics. Operate device from 115 V, 60 Hz, hospital-grade earth-grounded outlet only. Though the factory-supplied configuration settings are suitable for most therapies, the operator and hospital professionals should verify that the device's settings are appropriate for their clinical applications. Use only with standard Baxter soft tubing administration sets with an s designation. Read and understand this manual before using this device. When infusing fluid through a central line catheter, Baxter recommends that sets with a Luer lock adapter be used. If using a filter set, use only filters which are suitable for use with infusion pumps. Read and follow instructions of the filter to be used. To use the automatic piggyback function, use only Continu-Flo sets from Baxter as the primary line with a compatible secondary set for the secondary line. As with all medical electronic equipment, care must be exercised to avoid exposing this device to powerful sources of electromagnetic interference. This device design has been tested to the requirements of
10
7-19-B1-564
MDS201004, and applicable portions of MILSTD461C, which are voluntary test guidelines for electromagnetic susceptibility and emissions. This device was not found to be adversely affected by the susceptibility tests in these specifications, and will perform safely. The device's emissions were also found to be acceptable.
Epidural Administration Epidural administration of anesthetics is limited to short-term infusion (not to exceed 96 hours) with indwelling catheters specifically indicated for shortterm anesthetic epidural drug delivery.
Epidural administration of analgesics is limited to use with indwelling catheters specifically indicated for either short-term or long-term analgesic epidural drug delivery. To prevent infusion of drugs that are not indicated for epidural use, do not use administration sets that incorporate injection sites during epidural delivery. It is strongly recommended that pumps used for epidural drug delivery be clearly differentiated from pumps used for other routes of administration. WARNINGEpidural administration of drugs other than those indicated for epidural use could result in serious injury to the patient.
7-19-B1-564
11
Note:
The procedure for preparing and loading the administration set (steps 2 through 9) is the same for Pump 1 and Pump 2. Prepare fluid container and administration set according to the directions accompanying the products. Prime the set. Ensure all air is expelled from the set.
2. Prepare solution.
3. Close set regulating clamp. 4. Open pump door. Raise pump door latch to horizontal position and pull door open. a. Press SAFETY CLAMP to open position. b. If your hospital has enabled the slide clamp loading feature, insert the slide clamp on the I.V. set fully into the slot, so that it is flush with the pump housing. (See Figure 4.) If the slide clamp loading feature is not enabled, a black spring retainer occupies the slide clamp slot and the slide clamp cannot be inserted. Proceed to step c.
Note: During an emergency, in the alert mode, the pump can be operated without inserting the slide clamp. An alert tone sounds until the slide clamp is inserted or SILENCE is pressed. In the alarm mode (software versions later than 1.08), the pump will NOT start and an alarm tone will sound if the slide clamp is not loaded.
12
7-19-B1-564
c. Load the tubing through the guide channel from bottom to top as shown in Figure 4. d. Ensure that the I.V. tubing is loaded straight through the pump mechanism tubing guides and safety clamp. e. Ensure that the tubing is touching the pumping fingers before closing the pump door. 6. Close pump door. If resistance is felt when closing the door, check for a misloaded I.V. set. a. Verify that no drops are falling in drip chamber. If flow is observed, close regulating clamp, recheck I.V. set loading, and verify that the proper administration set is being used. b. If flow is again observed, do not use the pump. Have it inspected by service personnel.
Note:
Always close the administration set regulating clamp(s) before opening pump door and removing set.
8. Attach set to I.V. access site. 9. Turn pump(s) on by pressing appropriate ON/OFF CHARGE key(s). Verify that the following self-test occurs when you turn on the first pump: a. All segments of the Pump 1 and Pump 2 message displays appear momentarily. All segments of the main display of the powered-on pump appear momentarily. b. Three separate audible tones sound.
7-19-B1-564
13
c. If a Hospital Area Designator (HAD) has been programmed into the device, it is displayed for three seconds in the Pump 1 message display. d. The occlusion detection level is momentarily displayed in the Pump 1 message display (LEVEL 1, 2, or 3), followed by AUDIBLE SWITCHOVER if the Audible Switchover option is enabled. e. If Auto Restart and Flow Check are both enabled, AUTO RESTART appears for one second in the Pump 1 message display. f. If the device is plugged into an AC outlet, the plug icon is lit. If the device is operating on battery power, the battery icon is lit.
g. If the message INSERT SLIDE CLAMP appears when you close the door, the slide clamp loading option is enabled. Close the I.V. set's regulating clamp, open the door, and insert the I.V. set's compatible slide clamp fully into the slide clamp slot located below the safety clamp. Close the door and open the regulating clamp. 10. Set VOLUME knob on the rear of the device to the desired level.
14
7-19-B1-564
7-19-B1-564
15
5. If necessary, reset the volume previously infused on the selected pump to zero.
Note:
Always verify programmed information prior to starting the selected pump. a. Press PRI START. The green PUMPING LED lights and a moving bar appears next to the appropriate flow rate setting. b. If the alarm (software versions later than 1.08) or the alert tone sounds with the message INSERT SLIDE CLAMP when you press PRI START, the slide clamp loading option is enabled and the slide clamp has not been loaded into the slide clamp slot. Close the I.V. set's regulating clamp, open the door, and insert the I.V. set's slide clamp into the slide clamp slot located below the safety clamp. Close the door, then open the regulating clamp. c. Confirm flow by checking for drops in the I.V. set drip chamber. d. If Auto Restart is enabled and Flow Check is not enabled, AUTO RESTART is displayed in the Pump 1 message display. e. If you wish to program the other pump, press its ON-OFF/CHARGE key and repeat steps 1 through 6.
16
7-19-B1-564
b. The selected pump's PUMPING LED goes out and STOPPED appears in the message display. c. If the pump is not restarted or powered off within two minutes, an audible alert sounds. 8. To restart an infusion: a. Press the PUMP 1 or PUMP 2 key as appropriate. b. Press the appropriate START key. 9. When the infusion completes: When a pump has delivered the selected volume, it sounds an alert tone and switches to a KVO (Keep Vein Open) rate of 5 mL/hr or the current rate setting, whichever is lower. a. Press the appropriate STOP key. b. Close administration set regulating clamp. c. Open appropriate pump door. d. If the slide clamp loading option is enabled, press the slide clamp fully into its slot. e. Press the position. f.
SAFETY CLAMP
to the open
Check that no fluid is flowing in set, then remove set from pump.
7-19-B1-564
17
18
7-19-B1-564
b. Enter the desired fluid amount (in mL) on the keypad. Set VTBI equal to the volume of fluid in the container. c. To correct a mistake, press CLR or SEC VTBI again and re-enter the correct value. 6. Open secondary set regulating clamp. 7. Press SEC START to begin the infusion. Check also that the primary set regulating clamp is open. a. Verify that drops begin falling in the secondary set drip chamber only. b. Verify that the secondary program is displayed. c. When the VTBI of the secondary program reaches zero, the pump will revert to the primary program. When this happens, if the primary flow rate is set above 999 mL/hr, be sure to close the secondary set regulating clamp.
Note:
The Volume-Time Programming procedures are identical for Pump 1 and Pump 2.
7-19-B1-564
19
a. Press the appropriate ON-OFF/CHARGE key. b. Press PUMP 1 or PUMP 2 as appropriate to select the desired pump.
2. Enter a VTBI.
a. Press the PRI VTBI key (or SEC VTBI, if you are programming a secondary infusion). b. Program the desired VTBI on the keypad. c. To correct a mistake, press the CLR key and re-enter the correct VTBI.
3. Enter a time.
a. Press the TIME key. Enter the amount of time (up to 99 hours and 99 minutes) over which you wish the infusion to take place. b. To correct a mistake, press the CLR key or TIME key again and re-enter the correct time.
a. The device calculates the flow rate required to deliver the desired VTBI in the specified time period. b. If the calculated flow rate is higher than the device's capabilities, the message Hi is displayed. To correct a Hi message, repeat the procedure from step 2 and enter a longer time period if appropriate. c. If the calculated rate is too low, the message Lo is displayed. To correct a Lo message, repeat the procedure from step 2 and enter a shorter time period if appropriate.
20
7-19-B1-564
Verify that the calculated rate is acceptable before pressing PRI START (or SEC START) to begin the infusion.
2. Press PRI RATE (or SEC RATE) while pump is running. 3. Enter the new flow rate on the keypad.
NEW RATE appears in the selected pump's message display and an alert tone sounds periodically.
a. If the primary rate is changed above 999 mL/hr, ensure that the secondary set regulating clamp is closed. b. If a new rate higher than the allowable maximum is entered, Hi appears in the appropriate rate display.
The pump begins delivering fluid at the new rate, the alert tone stops, and the NEW RATE message disappears.
7-19-B1-564
21
NOTE:
If the word OPTIONS is not visible when a pump is powered on, you cannot use computer control or PDP. Consult qualified hospital biomedical personnel regarding the appropriateness of the device's settings.
Using the Programmed Delivery Profile Feature Each pump channel has a Programmed Delivery Profile (PDP) that allows you to program an infusion which changes over time. This feature, which can be made available through the device's configuration, lets you program an infusion of up to 10 different steps. The pump automatically changes to the next programmed rate once the VTBI of a step has been delivered. To use a PDP: 1. Verify that the OPTIONS key is lit. If OPTIONS is not lit, your device is not configured for PDP. Consult with authorized hospital personnel if you have questions regarding the appropriateness of your device's configuration settings. Press PUMP 1 or PUMP 2 as appropriate. The yellow LED on the PUMP key lights to indicate that the pump is selected.
2. Select a pump.
22
7-19-B1-564
3. Place the pump into PDP mode by pressing the OPTIONS key.
The message PGM DELIV appears and the NEXT legend over the TOT VOL/STATUS key lights. The pump is now in PDP mode. a. If a previously set PDP is stored in the pump's memory, STEP 1 appears in the message display and the rate and VTBI of the first programmed step appear in the main display. b. If you wish to clear the stored PDP, press the CLR key. The message CLEAR ALL? appears. Press the CLR key again to actually clear the PDP. The CLEAR ALL? message disappears. c. If you wish to use the stored PDP, continue pressing the NEXT key to scroll through all steps of the PDP until the first step is again displayed. Review each step to ensure that the PDP is appropriate. d. To begin the PDP, press PRI START. The message REVIEW PDP appears if you have not completed reviewing the stored PDP. After you have reviewed all steps and pressed PRI START, the pump begins delivering fluid.
Note:
If you enter PDP mode and then try to start the pump without first entering a PDP, the message ENTER PGM appears on the second line of the message display. The message disappears once you start to enter a PDP. a. After completing steps 1 through 3 above, press the NEXT key to reach the PDP step you wish to modify.
7-19-B1-564
23
b. Press the key for the infusion parameter you wish to modify (PRI VTBI or PRI RATE). c. Enter the new value on the keypad. d. Press the value.
NEXT
e. Modify additional PDP steps if desired. After reviewing or changing the last step, press NEXT again, until the first step is displayed. f. To begin the PDP, press PRI START. The pump begins delivering fluid.
a. After completing steps 1 through 3 above, enter the desired values using one of the following methods: Entering rate and VTBIUse PRI RATE, PRI VTBI and the numeric keypad exactly as if programming a primary infusion. You may enter rate and volume in either order. Entering VTBI and timeFirst press PRI VTBI and enter a value. Then press TIME and enter a time span for this step (up to 99 hours and 99 minutes). Finally, press PRI RATE. The device calculates and displays the resulting flow rate. If the display reads Hi or Lo, the calculated rate is outside the device's capabilities. Re-enter the VTBI and the time and press PRI RATE again. Entering rate and timeFirst press PRI RATE and enter a value.
24
7-19-B1-564
Then press TIME and enter a time span for this step (up to 99 hours and 99 minutes). Finally, press PRI VTBI. The device calculates and displays the resulting volume. If the display reads Hi or Lo, the calculated VTBI is outside the device's capabilities. Re-enter the rate and the time and press PRI VTBI again. b. Press the NEXT key to lock in the programmed values and advance to the next step. c. Program all remaining steps (up to 10) in the same manner. After entering settings for the last step, press NEXT again, until the first step is displayed. d. To begin the PDP, press PRI START. The pump begins delivering fluid. To pause during a PDP, press the pump's STOP key. To resume from the point at which you left off, select the pump (with PUMP 1 or PUMP2) and press PRI START. To drop out of PDP, select the pump and then press TIONS, PRI RATE, SEC RATE, or SEC VTBI.
OP-
In PDP mode, the pump switches to the KVO rate when it reaches a step that has a rate and VTBI of zero or when it finishes the tenth step. Computer Control The Computer Control option allows the device to be controlled by a remote computer. Computer Control is an option made available through the configuration. When the device is under computer control, the COMPUTER CONTROL legend on the front panel is lit.
7-19-B1-564
25
Pressing any enabled key (other than SILENCE, TOT VOL/STATUS, BACK LIGHT, PANEL LOCK, or the key combination for the Flow Check display or the Alarm Log display) causes the device to drop out of computer control, allowing you to clear alarms and/or change infusion data. Computer Monitoring When the device is set up for computer monitoring, information is gathered by a remote computer via the COMMUNICATIONS PORT. The computer connected to the device does not control it; it simply monitors infusion data. The MONITOR legend on the front panel lights whenever the computer queries the device for information. The device is operated in exactly the same manner as if it were not connected to the computer.
26
7-19-B1-564
7-19-B1-564
27
A solution of 10% bleach and water Soapy water Isopropyl alcohol up to 95%
Note:
Do not clean, disinfect, or sterilize any part of the device by autoclaving or with ethylene oxide gas. Doing so may damage the device and void the warranty. Do not use the following chemicals on the device, as they will damage the front panel: acetone, ammonia, benzene, hydroxytoluene, methylene chloride, n-alkyl dimethyl ethylbenzyl ammonium chloride, and ozone. It is recommended that the pump be plugged in during storage to help maintain the batteries at full charge. Do not store the device with either ON-OFF/CHARGE key ON. The battery may discharge completely.
28
7-19-B1-564
Alert Messages
STOPPED KVO PRI VTBI = 0
No flow KVO Pump has been in STOPPED mode for two minutes. Primary VTBI has been delivered or the PDP has finished. The pump has switched to a KVO rate of 5 mL/hr or programmed rate, whichever is less. This alert also occurs if you start the pump without first entering a non-zero value for the primary VTBI. The slide clamp loading option is enabled and the slide clamp is not loaded into the slide clamp slot. Close the regulating clamp, open the pump door, and insert the slide clamp fully into the slot, so that it is flush with the pump housing. Close the door, open the regulating clamp, and restart the pump. Flow rate is being changed while pump is running. Pump remains in NEW RATE alert condition until the appropriate START key is pressed. Pressing STOP or TOT VOL/STATUS displays the current flow rate.
No effect
NEW RATE
PRI RATE = 0
A primary flow rate of zero has been entered. The pump remains in this alert condition until a non-zero primary flow rate is entered and the appropriate START key is pressed. Battery needs recharging. Pump will stop operating in approximately fifteen minutes unless it is plugged into an AC outlet.
No change
No change
The SEC COMPLETE alert option is enabled and the pump has completed its secondary program and has switched to primary. Press any enabled key to exit from this alert.
7-19-B1-564
29
Message
Condition
SEC PROGRAM
Secondary program data is being entered while pump is running. Pump remains in alert condition until you press SEC START or STOP. An attempt has been made to start the secondary program with a secondary flow rate of zero. The pump remains in this alert condition until you enter a non-zero secondary flow rate and press SEC START, or start the primary program. A secondary VTBI of zero has been entered. The pump remains in this alert condition until you enter a non-zero secondary VTBI and press SEC START, or start the primary program. During Volume-Time Programming, if a flow rate outside the pump's capabilities is calculated, the message Hi or Lo appears in the rate display. If an attempt is made to start the pump with either of these messages displayed, a FLOW RATE alert is triggered. To exit this alert condition, reprogram the pump for a rate within the range set in the configuration and press the appropriate START key. A VTBI outside the acceptable range has been entered. To exit this alert condition, enter a VTBI within the range set in the configuration and press the appropriate START key. The pump is waiting for a Programmed Delivery Profile. Either enter the profile (as described on p. 24), or press OPTIONS to leave PDP mode. An attempt has been made to start a Programmed Delivery Profile before reviewing all the steps. Either press NEXT until the first step is again displayed, or press OPTIONS to leave PDP mode. The pump is asking for confirmation that all steps of the existing Programmed Delivery Profile should be erased. Do one of the following: press CLR to erase the profile; press NEXT to begin reviewing the rest of the profile; press PRI RATE or PRI VTBI to modify this step; or press OPTIONS to leave PDP mode.
SEC RATE = 0
No flow
SEC VTBI = 0
No flow
FLOW RATE
No flow
CHECK VTBI
No flow
No flow
No flow
No flow
30
7-19-B1-564
Message
Condition
COM TIMEOUT
There has been no communication between the device and the computer for the specified time period. Check for a disconnected cable or computer problem. To clear this alert, press the OPTIONS key to return the device to computer control, or press SILENCE or PUMP 1 or PUMP 2 to use the pump(s) without the computer. The controlling or monitoring computer is sending multiple questions or instructions to the device without waiting for the device's replies. The condition has been caused by the computer, not the device. To clear the alert, press OPTIONS to return the device to computer control (if appropriate), or press SILENCE or PUMP 1 or PUMP 2 to control the pump(s) manually. Notify the technical personnel responsible for the computer. If the alert recurs, disconnect the cable from the communications port. Reprogram the pump(s) for manual operation.
No change
Alarm Messages
CHECK SET LOADING
No flow I.V. set is not loaded properly. Close the regulating clamp, open the pump door, and verify that the set is loaded properly against the pumping fingers and through the upstream and downstream occlusion sensors. Close the door, open the regulating clamp, and restart the pump. a. Air bubble at detector b. Empty fluid container c. No I.V. set in pump Closed distal clamp, stopcock, clogged filter or other blockage downstream of the pump. Note: The pump can be set up to restart itself automatically when the occlusion is removed. If you want to prevent the pump from restarting automatically, press any enabled key (such as SILENCE) during the alarm. The slide clamp loading option is enabled and the slide clamp is not loaded in the slide clamp slot. To clear this alarm, open the pump door and insert the slide clamp into the slot. Close the door.
AIR
No flow
OCCLUSION
No flow
INSERT SLIDE CLAMP Note: This alarm option is available only on pumps running software versions later than 1.08.
No flow
7-19-B1-564
31
Message
Condition
No flow
Pump door not fully closed. The door latch must be fully lowered to a vertical position. Battery power has been exhausted. Plug device into AC outlet to restore operation and recharge battery.
BATTERY LOW with rapid three-tone alarm FAILURE in message display with code number in main display COMMON FAILURE in left message display and code number in right SEC VTBI display COM TIMEOUT
No flow
No flow
Pump-specific failure. Press appropriate ONOFF/CHARGE key twice to reset. If FAILURE does not clear, record the failure code number, remove the device from use and have it serviced.
No flow
General failure. Press each ON-OFF/CHARGE key in turn until both pumps are powered off. Power one or both pumps on again. If COMMON FAILURE does not clear, record the failure code number, remove the device from use, and have it serviced.
No flow
There has been no communication between the device and the computer for the specified time period. Check for a disconnected cable or computer problem. To clear this alarm, press the OPTIONS key to return the device to computer control, or press SILENCE or PUMP 1 or PUMP 2 to use the pump(s) without the computer. The controlling computer is sending multiple questions or instructions to the device without waiting for the device's replies. The condition has been caused by the computer, not the device. To clear the alarm, press OPTIONS to return the device to computer control, or press SILENCE or PUMP 1 or PUMP 2 to control the pump(s) manually. Notify the technical personnel responsible for the computer. If the alarm recurs, disconnect the cable from the communications port. Reprogram the pump(s) for manual operation.
No flow
32
7-19-B1-564
Note:
The last ten alarm codes are stored in memory for each pump. You can review these codes when the pump is powered on and stopped. To recall the stored alarm codes, press SILENCE and TOT VOL/STATUS simultaneously. The alarm code that occurred most recently for each powered-on and stopped pump appears in that pump's message display. To scroll back through the other stored alarm codes, press CLEAR TOT VOL within one second. Alarm Recall is exited automatically if the CLEAR TOT VOL key is not pressed within one second. The alarm codes are date and time stamped. They are detailed in the Service Manual.
7-19-B1-564
33
Technical Specifications
ITEM Catalog Code Number Description Administration Set Keep Vein Open (KVO) Rate Nurse Call Battery Battery Life 2M8064 Dual Channel Linear Peristaltic Volumetric Infusion Pump Baxter s designated sets 5 mL/hr or programmed rate, whichever is less SPECIFICATION
Available through the COMMUNICATIONS PORT 12 volt, 3.2 ampere-hour sealed lead-acid -Approximately 6 hours with one pump running at a rate from 1 to 1400 mL/hr -Approximately 4 hours with both pumps running at rates from 1 to 1400 mL/hr 8 hours for more than 80% recharge 110/120 V, 60 Hz 2.9 m (9.5 ft) long, with Hospital Grade plug 0.8 A, 250 V, SB, 5.2 mm (13 64 in.) 20 mm (25 32 in.) Less than 50 microamps (using UL-544 specified test methods) 8.1 kg (17.9 lbs) 33 cm W 13 cm D 29 cm H (13" W 5.1" D 11.4" H) Primary program for each pump: 1.099.9 mL/hr in 0.1 mL increments or 11999 mL/hr in 1 mL increments. Upper limit can be reduced by authorized service personnel. Secondary program: 1.099.9 mL/hr in 0.1 mL increments or 1999 mL/hr in 1 mL increments, subject to the upper limit that also applies to the primary program. 1.099.9 mL in 0.1 mL increments or 19999 mL for both primary and secondary of each pump. Upper limit can be reduced by authorized service personnel.
VTBI Range
34
7-19-B1-564 Rev. A
SPECIFICATION Factory set to NORM, which causes the device to alarm on air bubbles approximately 75 L or larger. The MIN setting causes the device to alarm on air bubbles approximately 50 L or larger.
7-19-B1-564
35
Configurable Options
The device has several features which can be selected to best suit the needs of the hospital. The settings for these features are referred to as the configuration. The configuration applies to the device as a whole unit, and thus to both pumping channels. To see your device's configuration settings: a. Both pumps must be stopped. Press TIME and TOT VOL/STATUS simultaneously, and hold for one second. The message REVIEW CONFIG appears in the left message display. The first option and its current setting appear in the right message display. b. To view the next option, press the NEXT or SEC START key. c. To stop viewing the configuration, again press TIME and TOT VOL/STATUS simultaneously.
Note:
If there is any question regarding the device's current configuration or applicability for a particular clinical application, the operator and hospital professionals should verify that the settings are appropriate. The configuration can be changed only by designated personnel. The following table shows all of the configurable items, their available settings, and the way they are set when the device leaves the factory.
36
7-19-B1-564
Configurable Options
Item Descriptor Options Factory Setting LEVEL 1
Occlusion Alarm Level (Pressure increments which trigger an OCCLUSION alarm. This setting is displayed for one second after the selftest when the pump is turned on.) Audible Switchover from secondary to primary program. When ON, AUDIBLE SWITCHOVER is displayed for 1 second after the self-test when the pump is turned on.
OCCLUSION LEVEL n
AUDIB SWI
1 = OFF 2 = ON
OFF
Number of Automatic Re- AUTO RES starts The number of times the pump will restart automatically after a downstream occlusion occurs and is then relieved. When ON, AUTO RESTART is displayed while the pump is running. If Flow Check is also enabled, AUTO RESTART is displayed for one second after the self-test when the pump is turned on. Door Open Required DOOR OPEN
0 to 9
1 = OFF 2 = ON (door must be opened after a downstream occlusion alarm.) 1 = MIN (2585 L) 2 = NORM (50110 L) 17 seconds 17 seconds
OFF
AIR SIZE
NORM
1 sec. 7 sec.
7-19-B1-564
37
Configurable Options
Item Descriptor Options Factory Setting 1999 mL/hr 9999 mL OFF
MAX RATE
11999 mL/hr
19999 mL 1 = OFF (The decimal point & TOT VOL/STATUS keys must be pressed to view the Flow Check display) 2 = ON (Flow Check display visible whenever pump is running) 1 = 300 2 = 1200 3 = 2400 4 = 4800 5 = 9600
BAUD RATE Baud Rate (The rate of information exchange between the device and the computer when in computer control or computer monitoring) Computer Control (Selects the type of computer control available at power-up) COMP CONT
9600
1 = DISABLED (computer control not available) 2 = OFF W ALRM (device drops out of computer control when an alarm occurs) 3 = ON W ALRM (device remains in computer control during an alarm) 0 = No message 1 = NICU 2 = PICU 3 = MED/SURGICAL 4 = TRAUMA/BURN UNIT 5 = OPER ROOM 6 = CARDIAC/ICU 7 = SURGICAL/ICU 8 = ICU 9 = ONCOLOGY
DISABLED
Hospital Area Designator HAD (displayed for 3 seconds after self-test. Custom designators can be created using a computer.)
No message
38
7-19-B1-564
Configurable Options
Item Descriptor Options Factory Setting ON
Close Clamp Message (CLOSECLAMP appears with DOOR OPEN message to remind the user to close the I.V. set regulating clamp when the pump door is open.)
CLOSE CLMP
1 = OFF 2 = ON
Slide Clamp Loading Op- INS CLAMP tion Note: The alarm option is available only on pumps running software versions later than 1.08.
1 = OFF 2 = ON for software versions earlier than 1.09, or ALERT for software versions later than 1.08 (An audible alert occurs and the message INSERT SLIDE CLAMP is displayed if the pump is used without first loading the set's slide clamp into the slide clamp slot.) 3=ALARM (An audible alarm occurs and the message INSERT SLIDE CLAMP is displayed. The pump will not operate.) 1 = DISABLED (No PDP allowed) 2 = FIVE HOUR (profile can be entered from front panel and is saved in 5-hour memory) 3 = SEMI PERM (profile can be entered from front panel; pump remembers the most recent profile) 4 = PERM (Profile is saved in permanent memory. If a new PDP is entered, the pump will revert to the PDP in permanent memory after the current run.) Real time entered in hours and minutes
OFF
PDP
DISABLED
Time Setting
TIME SET
7-19-B1-564
39
Configurable Options
Item Descriptor Options Factory Setting Current date for Central Standard Time
Date Setting
DATE SET
Month/Day/Year
40
7-19-B1-564
Warranty
Baxter Healthcare Corporation (Baxter) warrants that the equipment shall be free from defects in material and workmanship when delivered to the original purchaser. Baxter's sole obligation shall be limited to repair or replacement, at Baxter's option and expense, of the defective part or unit, excluding batteries, for a period of one year following the date of initial delivery. The warranty period for batteries is limited to a period of six months following the date of initial delivery. The warranty extends only to the original purchaser and is not assignable or transferable, and shall not apply to auxiliary equipment or disposable accessories. THERE ARE NO OTHER WARRANTIES INCLUDING ANY IMPLIED WARRANTY AND ANY WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE WHICH EXTEND BEYOND THE DESCRIPTION OF THE PRODUCT AND THOSE EXPRESSLY SET FORTH IN ITS LABELING. UNLESS USED ACCORDING TO THE DIRECTIONS ACCOMPANYING THE PRODUCT, ALL WARRANTIES ARE SPECIFICALLY EXCLUDED. In no event shall Baxter Healthcare Corporation be responsible for incidental, consequential or exemplary damages. Modification, alteration, recalibration or abuse, and service by other than a Baxter Healthcare Corporation authorized representative may void the warranty.
7-19-B1-564
41
Service Information
While under Baxter Healthcare Corporation Warranty, Service Agreement (optional), or lease agreement, the instrument must not be opened by unauthorized personnel. To contact Baxter Healthcare Corporation Customer Service Division for service and repair information for all instruments, call 1-(800) THE PUMP. Shipping costs for all units returned to Baxter Healthcare Corporation shall be paid by the customer. The unit must be packed in its original container or in another Baxter approved container that will provide adequate protection during shipment. To ensure prompt return, a Baxter Product Service representative must be notified before shipping any unit for repair. When calling Baxter Product Service, please be prepared to provide code number and serial number of the unit. A service request number will be issued and should accompany all communications. A brief written description of the problem should be attached to the instrument when it is returned for service. Baxter Healthcare Corporation will not be responsible for unauthorized returns or for units damaged in shipment due to improper packing.
42
7-19-B1-564