AD629 Technical Specificatons Page 1
License Overview
International configuration
AD629
Basic license Extended license
Licenses Audiometry
Basic Audiometry x x
Békésy x x
SISI x x
Langenbeck x x
Stenger x x
Modified Hughson-Westlake x x
Weber x x
ABLB x x
Speech testing with CD/Mic x x
Build-in wave files x x
Binaural Speech x
Hearing Loss Simulator (HLS) x
MHA x
QuickSIN optional optional
TEN test optional optional
High Frequency (HF) optional optional
Freefield Lineout x x
Hyrbid mode - PC controlled mode optional optional
Sync mode - Audiogram transfer optional optional
Fuctions available only in Diagnostics suite
MaskingHelper x x
AD629 Technical Specificatons Page 2
Languages supported in IMP and AUD
Portuguese
Norwegian
Japanese
Chinese
Russian
Spanish
German
English
Turkish
Finnish
Korean
French
Polish
Czech
Italian
Greek
IMP
MT10 X X X
Existing AT235 X X
New AT235 X X X X X X X X X X X X X X X X
Existing AA222 X X
New AA222 X X X X X X X X X X X X X X X X
Titan X X X X X X X X X X X
AUD
AS608 X X X X
AD226 X X X X X X X X X X
AD629 / AD229 X* X* X X* X X X* X X* X* X* X X* X* X X
AC40 X* X* X X* X X X* X X* X* X* X X* X* X X
Suites
Titan suite X X X X X X X X X X X X
DS X X X X X X X X X X X X X X
AD629 Technical Specificatons Page 3
Included and Optional Parts
Included parts DD45 Audiometric headset
B71 Bone conductor
APS3 Patient response button
Goose neck microphone
Power cable
Operation manual CD
Multilingual instructions for use
Optional parts Diagnostic Suite software
OtoAccess database
21925 Amplivox audiocups, noise reducing headset
Carrying case (Standard or Trolly Style)
EARTone3A/5A Audiometric insert phones
IP30 Audiometric insert phones
HDA300 Audiometric headset with double mono 6.3mm jack
TDH39 Audiometric headset
CIR33 Insert earphone set for masking or monitoring
DD450 Audiometric headset
Talk back microphone
Sound field speakers SP90 (with external power amp)
AP12 Power Amplifier 2x12 Watt
AP70 Power Amplifier 2x70 Watt
AD629 Technical Specificatons Page 4
General Technical Specifications
AD629 Technical Specification
Safety Standards IEC 60601-1, ES60601-1, CAN/CSA-C22.2 No.60601-1
Class I, Applied parts type B, Continuous operation
EMC Standard IEC 60601-1-2:2001 + A1:2004
Audiometer Standards Tone Audiometer: IEC 60645 -1, ANSI S3.6 -2010, Type 2, HF IEC 60645-4.
Speech Audiometer: IEC 60645-2/ANSI S3.6 type B or B-E.
Auto threshold tests: ISO 8253-1
Calibration Calibration information and instructions is located in the AD629 Service manual
Air Conduction DD45: PTB/DTU report 2009
TDH39: ISO 389-1 1998, ANSI S3.6-2010
HDA300: PTB report PTB 1.61 – 4064893/13
HDA280 PTB report 2004
E.A.R Tone 3A/5A: ISO 389-2 1994, ANSI S3.6-2010
IP 30: ISO 389-2 1994, ANSI S3.6-2010 DES-2361 CIR 33
ISO 389-2
Bone Conduction B71: ISO 389-3 1994, ANSI S3.6-2010
Placement: Mastoid
Free Field ISO 389-7 2005, ANSI S3.6-2010
High Frequency ISO 389-5 2004, ANSI S3.6-2010
Effective masking ISO 389-4 1994, ANSI S3.6-2010
Transducers DD45 Headband Static Force 4.5N ±0.5N
TDH39 Headband Static Force 4.5N ±0.5N
HDA300 Headband Static Force 4.5N ±0.5N
HDA280 Headband Static Force 5N ±0.5N
DD450 Headband Static Force 10N ±0.5N
B71 Bone Headband Static Force 5.4N ±0.5N
E.A.R Tone 3A/5A:
IP30
CIR 33
Patient Response switch One hand held push button.
Patient communication Talk Forward (TF) and Talk Back (TB).
Monitor Output through built-in speaker or through external earphone or speaker.
Special tests/test battery SISI. ABLB. Stenger. Stenger Speech. Langenbeck (tone in noise). Békésy Test. Weber,
2 channel speech, 2 channels Master Hearing Aid, Auto threshold.
Auto threshold tests:
Available time for patient to respond: Same as tone presentation
Increment of hearing level: 5dB.
Auto threshold test (Békésy):
Mode of operation: Békésy
Rate of level change: 2.5 dB/s ±20%
Smallest increment of level: 0.5 dB
Stimuli
Tone 125-20000Hz separated in two ranges 125-8000Hz and 8000-20000Hz.
Resolution 1/2-1/24 octave.
Warble Tone 1-10 Hz sine +/- 5% modulation
AD629 Technical Specificatons Page 5
Wave file 44100Hz sampling, 16 bits, 2 channels
Masking Automatic selection of narrow band noise (or white noise) for tone presentation and
speech noise for speech presentation.
Narrow band noise:
IEC 60645-1:2001, 5/12 Octave filter with the same centre frequency resolution as pure
Tone.
White noise:
80-20000Hz measured with constant bandwidth
Speech Noise.
IEC 60645-2:1993 125-6000Hz falling 12dB/octave above 1KHz +/-5dB
Presentation Manual or Reverse. Single or multiple pulses.
Intensity Check the accompanying Appendix
Available Intensity Steps is 1, 2 or 5dB
Extended range function: If not activated, the Air Conduction output will be limited to 20 dB
below maximum output.
Frequency range 125Hz to 8kHz (Optional High Frequency: 8 kHz to 20 kHz)
125Hz, 250Hz, 750Hz, 1500Hz and 8kHz may freely be deselected
Speech Frequency Response:
(Typical) Frequecy Linear (dB) FFequv (dB
(Hz) Ext sign1 Int. Sign2 Ext sign1 Int. Sign2
TDH39 125-250 +0/-2 +0/-2 +0/-8 +0/-8
(IEC 60318-3 Coupler) 250-4000 +2/-2 +2/-1 +2/-2 +2/-2
4000-6300 +1/-0 +1/-0 +1/-0 +1/-0
DD45 125-250 +0/-2 +1/-0 +0/- +0/-7
(IEC 60318-3 Coupler) 250-4000 +1/-1 +1/-1 +2/-2 +2/-3
4000-6300 +0/-2 +0/-2 +1/-1 +1/-1
E.A.R Tone 3A 250-4000 +2/-3 +4/-1 (Non
(IEC 60318-5 Coupler) linear)
IP 30 250-4000 +2/-3 +4/-1 (Non linear)
(IEC 60318-5 Coupler)
B71 Bone Conductor 250-4000 +12/-12 +12/-12 (Non linear)
(IEC 60318-6 Coupler)
2% THD at 1000 Hz max output
+9 dB (increasing at lower
frequency)
Level range: -10 to 50 dB HL
1. Ext. sign: CD input 2. Int. sign: Wave files
External signal Speech replaying equipment connected to the CD input must have a signal-to-noise ratio
of 45 dB or higher.
The speech material used must include a calibration signal suitable for adjusting the input
to 0 dBVU.
Free Field Power amplifier and loudspeakers
With an input of 7 Vrms - Amplifier and loudspeakers must be able to create a Sound
Pressure Level of 100 dB in a distance of 1 meter - and meet the following requirements:
Frequency Response Total Harmonic Distortion
125-250 Hz +0/-10 dB 80 dB SPL < 3%
250-4000 Hz ±3 dB 100 dB SPL < 10%
4000-6300 Hz ±5 dB
Internal storage 250 patients and unlimited sessions/measurements/audiograms
AD629 Technical Specificatons Page 6
Signal Indicator(VU) Time weighting: 300mS
Dynamic range: 23dB
Rectifier characteristics: RMS
Selectable inputs are provide with an attenuator by which the level can be adjusted to the
indicator reference position(0dB)
Data Connections (sockets) 4 x USB A (compatible with USB 1.1 and later)
1 x USB B (compatible with USB 1.1 and later)
1 x LAN Ethernet
External devices (USB) Standard PC mouse and keyboard (for data entry)
Supported printers: Standard PCL3 printers (HP, Epson, Canon)
Input Specifications TB 100uVrms at max. gain for 0dB reading
Input impedance : 3.2KOhm
Mic.2 100uVrms at max. gain for 0dB reading
Input impedance : 3.2KOhm
CD 7mVrms at max. gain for 0dB reading
Input impedance : 47KOhm
TF (side panel) 100uVrms at max. gain for 0dB reading
Input impedance : 3.2KOhm
TF (front panel) 100uVrms at max. gain for 0dB reading
Input impedance : 3.2KOhm
Wave files Plays wave file from hard disk drive
Output Specifications FF1 & 2 7Vrms at min. 2KOhm load
60-20000Hz -3dB
Left & Right 7Vrms at 10 Ohms load
60-20000Hz -3dB
Ins. Left & Right 7Vrms at 10 Ohms load
60-20000Hz -3dB
Bone 7Vrms at 10 Ohms load
60-10000Hz -3dB
Ins. Mask 7Vrms at 10 Ohms load
60-20000Hz -3dB
Monitor(side panel) 2x 3Vrms at 32 Ohms / 1.5Vrms at 8 Ohms load
60-20000Hz -3dB
Display 5,7 inch high resolution color display 640x480 pixels
Compatible software Diagnostic Suite - Noah, OtoAccess and XML compatible
Dimensions (LxWxH) 36.5 x 29.5 x 6.5 cm / 14.4 x 11.6 x 2.6 inches
Weight 3.3kg/6.3lb
Power supply 100-240 V~, 50-60Hz max 0.5A
Operation environment Temperature: 15-35°C
Re. Humidity: 30-90% Non condensing
Transport and storage Transport temperature: -20-50°C
Storage temperature: 0-50°C
Re. Humidity: 10-95% Non condensing