U600 ULTRASONIC SCALER - Woodpecker Medical

The product is mainly used for teeth cleaning and also an indisensable equipment for teeth disease prevention and treatment.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
108 views18 pages

U600 ULTRASONIC SCALER - Woodpecker Medical

The product is mainly used for teeth cleaning and also an indisensable equipment for teeth disease prevention and treatment.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 18

U600 ULTRASONIC SCALER

INSTRUCTION MANUAL
Contents

1. The installation and components of equipment.................................... 1


2. Installation and adjustment................................................................... 3
3. Maintenance and sterilization............................................................... 5
4. Precaution............................................................................................. 8
5. After service.......................................................................................... 9
6. Symbol instruction................................................................................ 9
7. Environmental protection.................................................................... 1 0
8. Manufacturer’s right........................................................................... 1 0
9. European authorized representative.................................................... 1 1
10. EMC - Declaration of conformity..................................................... 1 1
11. Statement........................................................................................... 1 4
TABLE OF OPERATING POWER OF THE TIPS................................ 1 5
1. The installation and components of equipment

1.1 Instruction
Guilin Woodpecker Medical Instrument Co., Ltd. is a professional
manufacturer in researching, developing and producing ultrasonic scalers. The
product is mainly used for teeth cleaning and also an indisensable equipment for
teeth disease prevention and treatment.
The ultrasonic scaler U600 has scaling, perio, endo and auto-water supply
functions with the following features:
1.1.1 The handpiece is detachable and can be autoclaved to high temperature
134°C and high pressure 0.22Mpa.
1.1.2 Automatic frequency tracking ensures that the machine always works on
the best frequency and more steadily.
1.1.3 Digital control, easy operation and more efficient for scaling.
1.2 Components
1.2.1 The components of machine are listed in the packing list.
1.2.2 Product performance and structural composition
Ultrasonic scaler U600 is composed of electrocircuit, water way and
ultrasonic transducer.
1.2.3 Scope of application
Ultrasonic scaler U600 is used for the dental calculus elimination and root
canal treatment.
1.3 The main technical specifications
1.3.1 Technical specifications of ultrasonic scaler
a) Main unit input: 220-240V~ 50Hz/60Hz 150mA
b) Output primary tip vibration excursion: ≤90μm
c) Output half-excursion force: <2N
d) Output tip vibration frequency: 28kHz±3kHz
e) Output power: 3W to 20W
f) Main unit fuse: T0.5AL 250V
g) Water pressure: 0.01MPa to 0.5MPa
h) Weight of main unit: 1.8kg
i) Operating mode: Continuous operation
j) Type of protection against electric shock: class II equipment
k) Degree of protection against electric shock: Type BF applied part
l) Applied part of the equipment: handpiece and tip
m) Degree of protection against harmful ingress of water: Ordinary
equipment, the foot switch is drip-proof equipment (IPX1)
n) Degree of safety of application in the presence of a Flammable Anesthetic
Mixture with air, Oxygen or Nitrous Oxide: Equipment not suitable for being used
in the presence of a flammable anesthetic mixture with air, oxygen or nitrous oxide.

1
1.4 Instruction of the main components
Instruction and component sketch map
1.4.1 Front of the main unit sketch map

Picture 1
1.4.2 Back of the main unit sketch map

Picture 2
1.4.3 Instruction of using wrench

2
Picture 3

Picture 4

2. Installation and adjustment

2.1 Operation
2.1.1 Open the packing box, make sure that all the parts and accessories are
complete according to the packing list, take the main unit out of the box, and put

3
it on the the stable plane facing to the operator.
2.1.2 Turn the water control knob to the max according to the picture 1
direction, Do not screw it over tight in case of damage. [ note 1]
2.1.3 Insert the plug of the foot switch to its socket. (see picture 2)
2.1.4 Connect one end of the water pipe to the water entrance, and the other
end to the clean water sourse. (see picture 2)
2.1.5 Choose the scaling tip according to the requirement, and fix the scaling
tip with the wrench. ( see picture 4)
2.1.6 Turn on the power switch, the power indicator lighted and the machine
is ready for work. Touch panel is applied to this machine, water supply mode or
power can be adjusted by directly touching the water supply mode identification
or power indicator on the touch panel.
2.1.7 Under normal working condition, the frequency of the tips is very high,
light touch and a certain to-and-fro motion will eliminate the tartar without
obvious heating, overexetion and overatay are forbidden.
2.1.8 Vibrating intensity: Adjust the vibrating intensity according to your
need, usually adjust to the middle grade, and adjust the vibrating during the
clinical treatment according to the patient's sensitivity and the rigidity of the
tartar.
2.1.9 Water volume adjustment: Step on the foot switch, and the tip begins
to vibrate, then turn the water control switch to fine spray to cool down the
handpiece and clean the teeth.
2.1.10 The handpiece can be handled in the same gesture as a pen in hand.
2.1.11 Be sure not to make the end of the tip touch the teeth vertically, and not
use too much pressure when the tip touch the surface of the teeth, in order not to
hurt the teeth and the tip.
2.1.12 After finishing operation, keep the machine working for 30 seconds
with the water supply to clean the handpiece and the tip.
2.1.13 Unscrew the scaling tip and sterilize it.
Note: Don't screw the scaling tips when stepping on the foot switch, and the
machine is working.
2.2 Cleaning mode
It is suggested to wash and disinfect the water pipe after scaling everyday.
The “Cleaning mode” can wash and disinfect the water pipe, reducing the
crystal substance and bacterias in water pipe.
Operation:
1.Put distilled water or mineral substance into water tank.
2.Press the automatically water supply button and outside water button at the
same time (1s) to start the “Cleaning mode” after buzzer beeps. The button of
automatic water supply mode would flashes and others buttons will go out.
3.Link the connector and the handpiece with the drainage device.
4.Step the pedal, the device would start self-cleaning. After that the pedal

4
could be loosen.
5.After cleaning for 30s, the device would stop self-cleaning. Or u can stop by
step the pedal again or press the automatic water supply button.
6.After cleaning, press the automatically water supply button and outside
water button at the same time (1s) to exit the “Cleaning mode” after buzzer
beeps.

3. Maintenance and sterilization

3.1 All the scaling tips ,Handpiece,wrench can be autoclaved.


3.1.1Autoclaved under high temperature, pressure, time:
134°C, 2.0bar~2.3bar (0.20MPa~0.23MPa), 4min.
3.2 The tip and wrench can be cleaned by ultrasonic cleaner.
3.3 Troubles shotting
Fault Possible cause Solutions
The power plug is in loose Make the plug insert to the
The scaling tip doesn't contact. socket well.
vibrate and there is no
The foot switch is in loose con- Insert the foot switch to its
water flowing out when
tact. socket tightly.
stepping on the foot
switch. The fuse in the main unit is
Contact our dealers or us.
broken.
Screw the tip on the hand-
The tip is in loose contact.
piece tightly (See Picture 6).
The connect plug between the
The scaling tip doesn't handpiece and the circuit board Contact our dealers or us.
vibrate but there is water is in loose contact.
flowing out when stepping
on the switch. Something wrong with the Send the handpiece to our
handpiece. company to repair.
Something wrong with the
Contact our dealers or us.
cable.

The scaling tip vibrates


but there is no spray The water control knob is not Turn on the water contrl
when stepping on the foot on. knob [note 1].
switch.

5
Fault Possible cause Solutions

The tip hasn't been screwed on Screw the tip on the hand-
to the handpiece tightly. piece tightly (See Picture 4).

The vibration of the tip


becomes weak. The tip is loose because of Screw on the tip tightly (See
vibration. Picture 4).
The coupling between the hand-
Dry it by the hot air.
piece and the cable isn't dry.
The tip is damaged [note 2]. Change a new one.
There is water seeping
The waterproof "O"ring is dam- Change a new waterproof
from the coupling between
aged. "O"ring.
the handpiece and cable.
There is water flow out There is impurity in the solenoid Contact with the local dis-
when turn off the power. valve. tributor or manufacturer.
The amount of spouting water is Turn the water control switch
The handpiece generates too little. to a higher grade [note 1].
heat.
The potentiometer is broken. Change a new one.
The water control knob is a low Turn the knob to a high
grade. grade [note 1].
The water pressure is not
The amount of spouting Enhance the water pressure.
enough.
water is too little.
Clean water pipe with
The water pipe is jammed. multi-function syringe
[note2].

The screw is loose. Tighten it.


The u-file doesn't vibrate.
Endochuck is damaged. Change a new one.
There is noise coming
The screw is loose. Tighten it.
from the endochuck.

There is no water coming


out from the handpiece Turn the water control to the
There is air in the water pipe.
(automatic water supply Max, reinsert the bottle.
mode).

If the problem still can't be solved, please contact local dealer or manufacturer.
3.4 Notice
[Note 1] The water control knob can adjust the water volume according to the
symbol.
[Note 2] To clean the water pipe with the multi-function syringe of the dental
unit ( see picture 5):

6
picture 5
a) Snip the water pipe at a distance of 10cm to 20cm from the water entrance.
b) Turn on the power switch, get through to the power.
c) Connect the multi-function syringe of the dental unit to the water pipe.
d) Screw off the scaling tip or pull out the handpiece.
e) Step on the foot switch.
f) Turn on the switch of the multi-function syringe, press the air or water into
the water pipe to clean and eliminate the impurity.
[Note 3 ] If the scaling tip has been screwed on tightly and there is fine spray
too, the following phenomena show that the scaling tip is damaged:
a) The vibrating intensity and the pulverization degree become weak
obviously.
b) During operating, there is some buzz when the scaling tip is working.
3.4Cleaning mode
It is suggested to wash and disinfect the water pipe after scaling everyday.
The “Cleaning mode” can wash and disinfect the water pipe, reducing the
crystal substance and bacterias in water pipe.
Operation
1.Put distilled water or mineral substance into water tank.
2.Press the automatically water supply button and outside water button at the
same time (1s) to start the “Cleaning mode” after buzzer beeps. The button of
automatic water supply mode would flashes and others buttons will go out.
3.Link the connector and the handpiece with the drainage device.
4.Step the pedal, the device would start self-cleaning. After that the pedal
could be loosen.
5.After cleaning for 30s, the device would stop self-cleaning. Or u can stop by
step the pedal again or press the automatic water supply button.
6.After cleaning, press the automatically water supply button and outside
water button at the same time (1s) to exit the “Cleaning mode” after buzzer

7
beeps.
4. Precaution

4.1 usage notice


4.1.1 Keep the scaler clean before and after operation.
4.1.2 The scaling tip, wrench and handpiece must be sterilized before each
treatment.
4.1.3 Don't screw the scaling tip when stepping on the foot switch.
4.1.4 The scaling tip must be fastened. There must be fine spray coming out
from the tip when operating.
4.1.5 Change a new one when the tip is damaged or worn excessively.
4.1.6 Don't twist or rub the tip.
4.1.7 While scaler working, the heat of scaling tip will become higher if there
is no water flowing out, please keep the water flow smoothly.
4.1.8 Don't use impure water source, and be sure not to use normal brine
instead of pure water source.
4.1.9 If use the water source without hydraulic pressure, the water surface
should be one meter higher than the head of the patient.
4.1.10 Don't knock or rub the handpiece.
4.1.11 After each clinical operation with clinical liquid, change a bottle with
purified water, turn the water supply to max, make the machine work with auto-
water supply for 30 seconds in order to keep the water way and spare parts clean
and durable.
4.1.12 Please put the power plug into the socket easy to pull out, to make sure
it can be pull out in emergency.
4.1.13 When using the equipment, please keep the water get through
smoothly, otherwise patient's tooth surface would be injured by overheat in the
handpiece.
4.1.14 After operating, turn off power source supply, and then pull out the
plug.
4.1.15 As a professional manufacturer of medical instruments, we are only
responsible for the safety on the following conditions:
Ⅰ.The maintenance, repair and modification are made by the manufacturer
or the authorized dealer.
Ⅱ. The changed components are original of "WOODPECKER" and operated
correctly according to instruction manual.
4.1.16 The screw thread of the scaling tips produced by other manufacturers
maybe coarse, rusty and collapsed, which will damage the screw thread of the
handpiece irretrievably. Please use "WOODPECKER" brand scaling tip.
4.1.17 Please select a suitable power when using different type of tips (refer to
“TABLE OF OPERATING POWER OF THE TIPS”).

8
4.2 Contraindication
4.2.1 The patient who has hemophilia is not allowed to use this equipment.
4.2.2 The patient or doctor who equips with heart pacemaker is forbidden to
use this equipment.
4.2.3 The heart disease patient, pregnant woman and children should be
cautious to use the equipment.
4.3 Storage and maintenance
4.3.1 The equipment should be handled carefully and lightly. Be sure that it is
far from the vibration, and is installed or kept in a cool, dry and ventilated place.
4.3.2 Don't store the machine together with the articles that are combustible
poisonous, caustic, or explosive.
4.3.3 This equipment should be stored in a room where the relative humidity
is 10% ~ 93%, atmospheric pressure is 70kPa to106kPa, and the temperature is
-20°C ~ +55°C.
4.3.4 Please turn off the power switch and pull out the power plug when the
equipment is not used. If the machine is not used for a long time, please make it
get through to the power and water once per month for five minutes.
4.4 Transportation
4.4.1 Excessive impact and shake should be prevented in transportation. Lay
it carefully and lightly and don't invert it.
4.4.2 Don't put it together with dangerous goods during transportation.
4.4.3 Avoid solarization and getting wet in rain or snow during transportation.
4.5 Working condition
4.5.1 Environment temperature: +5°C to +40°C
4.5.2 Relative humidity: 30% ~ 75%
4.5.3 Atmosphere pressure: 70kPa to 106kPa
4.5.4 A temperature of the water at the inlet:not higher than +25°C

5. After service
We offer one year free repair to the equipment according to the warranty card.
The repair of the equipment should be carried out by professional technician.
We are not responsible for any irretrievable damage caused by the not
professional person

6. Symbol instruction

Trademark Ordinary equipment

9
Alternating current Drip-proof
Date of manufacture Manufacturer
Class II equipment Type BF applied part

Foot switch interface Used indoor only


Adjustment for the water Sterilizable up to the
flow temperature specified
Mode of outside-water
Mode of auto-water system
system
CE marked product Water entrance pressure

Earth (ground) Follow Instructions for Use

Adjustment for the water flow,in Mode of auto-water system

Appliance compliance WEEE directive

Atmospheric pressure for storage

Temperature limitation for storage

Humidity limitation for storage

Authorised Representative in the EUROPEAN COMMUNITY

7. Environmental protection
Please dispose according to the local laws.

8. Manufacturer’s right
We reserve the rights to change the design of the equipment, the technique,
fittings, the instruction manual and the content of the original packing list at any
time without notice. If there are some differences between blueprint and real

10
equipment, take the real equipment as the norm.
9. European authorized representative

10. EMC - Declaration of conformity

11
12
13
The device has been tested and homologated in accordance with EN 60601-
1-2 for EMC. This does not guarantee in any way that this device will not
be effected by electromagnetic interference Avoid using the device in high
electromagnetic environment.

11. Statement
All rights of modifying the product are reserved to the manufacturer without
further notice. The pictures are only for reference. The final interpretation rights
belong to GUILIN WOODPECKER MEDICAL INSTRUMENT CO., LTD.
The industrial design, inner structure, etc, have claimed for several patents by
WOODPECKER, any copy or fake product must take legal responsibilities.

14
TABLE OF OPERATING POWER OF THE TIPS

Scaling Endodontics
Tip Model Power Tip Model Power
G3 1-10(G) E3 1-6(E)
G4 1-10(G) E3D 1-3(E)
G5 1-10(G) E4 1-6(E)
G6 1-10(G) E4D 1-3(E)
G7 1-10(G) E5 1-6(E)
G8 1-10(G) E5D 1-3(E)
G9 1-10(G) E8 1-10(E)
G 10 1-10(G) E9 1-10(E)
G 11 1-10(G) E10 1-6(E)
E10D 1-6(E)
Periodontics E11 1-6(E)
Tip Model Power E11D 1-6(E)
P1 1-10(P) E14 1-3(E)
P2L 1-3(P) E15 1-3(E)
P2LD 1-2(P)
P2R 1-3(P) Cavity Preparation
P2RD 1-2(P) Tip Model Power
P3 1-6(P) SB1 1-10(P)
P3D 1-6(P) SB2 1-10(P)
P4 1-6(P) SB3 1-10(P)
P4D 1-6(P) SBL 1-10(P)
SBR 1-10(P)

15
ZMN/WI-09-241 V1.6- 20191225