0% found this document useful (0 votes)
51 views2 pages

Specific Proteins Calibrator-150721009-BS600

This document provides calibration values and uncertainties for several specific protein calibrators, including C3, C4, CRP II, IgA II, IgG, and IgM. The calibrators were measured using a turbidimetry method and referenced to certified reference materials, with values reported in multiple units including g/L and μmol/L. Ranges of calibration values are given at varying concentration levels for each protein, along with expanded measurement uncertainties calculated at a coverage factor of k=2.

Uploaded by

S6b2
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
51 views2 pages

Specific Proteins Calibrator-150721009-BS600

This document provides calibration values and uncertainties for several specific protein calibrators, including C3, C4, CRP II, IgA II, IgG, and IgM. The calibrators were measured using a turbidimetry method and referenced to certified reference materials, with values reported in multiple units including g/L and μmol/L. Ranges of calibration values are given at varying concentration levels for each protein, along with expanded measurement uncertainties calculated at a coverage factor of k=2.

Uploaded by

S6b2
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 2

TRACEABILITY and UNCERTAINTY

BS600-Specific Proteins Calibrator


LOT:150721009

Abbreviated Calibrator Uncertainty Calibrator Uncertainty


Method Reference Measurement Procedure Reference Material Unit 1 Unit 2
name Value (k=2) Value (k=2)

0.190 0.01

0.540 0.03

C3 Turbidimetry Method / ERM-DA470K g/L 1.06 0.06 / / /

1.59 0.09

2.77 0.16

0.120 0.01 0.60 0.03

0.238 0.01 1.19 0.07

C4 Turbidimetry Method / ERM-DA470K g/L 0.482 0.03 μmol/L 2.41 0.13

0.551 0.03 2.76 0.15

1.060 0.06 5.30 0.29

10.20 0.84 97.10 7.98

33.8 2.78 321.78 26.45


CRP II
Turbidimetry Method / ERM-DA472 mg/L 90.7 7.46 nmol/L 863.46 70.98

150 12.33 1428.00 117.38

289 23.76 2751.28 226.16

page 1 of 2
TRACEABILITY and UNCERTAINTY
BS600-Specific Proteins Calibrator
LOT:150721009

Abbreviated Calibrator Uncertainty Calibrator Uncertainty


Method Reference Measurement Procedure Reference Material Unit 1 Unit 2
name Value (k=2) Value (k=2)

0.420 0.03 2.63 0.21

1.02 0.08 6.38 0.50

IgA II Turbidimetry Method / ERM-DA470K g/L 1.90 0.15 μmol/L 11.88 0.93

4.13 0.32 25.81 2.02

8.35 0.65 52.19 4.09

1.52 0.09 10.14 0.57

4.08 0.23 27.21 1.53

IgG Turbidimetry Method / ERM-DA470K g/L 8.54 0.48 μmol/L 56.96 3.20

17.7 0.99 118.06 6.62

37.5 2.10 250.13 14.03

0.270 0.02 0.28 0.02

0.630 0.04 0.65 0.05


IgM
Turbidimetry Method / ERM-DA470K g/L 0.880 0.06 μmol/L 0.91 0.06

1.26 0.09 1.30 0.09

4.90 0.35 5.05 0.36