Specifications:
Perkin Elmer 5100 AA w\ Data System
Spectrometer
Dimensions 120W x 62H x 63D (cm)
Weight Approx. 175kg (385 lbs.)
Power 115 or 230 volts±10%, 50 or 60Hz, 250 watts
Requirements (300 watts with optional background corrector)
Burner System
Dual-option, premix design using a high-
strength, Ryton® mixing chamber, internally
coated with polypropylene to resist chemical
attack.
Angled to help drainage.
Uses impact bead (for optimum sensitivity and
detection limits for most elements determined
Burner Chamber
using an air-acetyleneflame) or a flow spoiler
(for optimum performance for most elements
determined using a nitrous oxide-acetylene
flame and for solutions having a high dissolved
solids content).
Separate fuel and oxidant lines.
Separate auxiliary oxidant inlet.
Adjustable, stainless steel neulixer.
Nebulizer
Optional corrosion-resistant nebulizers.
10cm(4 inch), titanium, single-slot burner head
for use with air-acetylene flame.
Burner Head Optional 5cm(2inch), titanium, single-slot
burner head for use with a nitrous oxide-
acetylene flame.
Provides horizontal, vertical and rotational
Burner Mount
adjustment.
HGA-600 Graphite Furnace
Power 206, 220 or 240 VAC±10%
Requirements 50 or 60Hz
Power Approx. 3.6 kVA for a furnace temperature of
Consumption 2700°C
Dimensions 42W x 32H x 60D (cm)
(power unit)
Weight approx. 65 kg (143 lb)
Temperature Programmable from ambient (20°C) to 3000°C in
Range increments of 10°C
Argon or Nitrogen
Min. pressure 300kPa
Required Inert
Max. pressure 450kPa
Gas
Max. consumption: 1500 mL/min
Automatic gas shut-down in standby mode
Mains water or filtered industrial water
Water Coolant
max consumption 2.5 L/min
Isothermal
200-3000°C in increments of 10°C
Temperature
0-99s in increments of 1s.
Ramp Time
Zero ramp time selects maximum power heating
Hold Time 0-99s in increments of 1s
Internal Gas 0-300 mL/min in increments of 10 mL/min
Stream Changeover to an alternate gas is possible
Max 11Voff (continuous control of the effective
Output Voltage
voltage)
Automatic protection against: data transfer errors,
operation with open furnace, malfunctioning of the
teperature sensor, program errors, electrical
Safety Features
supply failure, tripped power circuit breader, low
gas inlet pressure, overheated furnace or
transformer, graphite tube breakage.
From the 5100 PC software via an IEEE-488
Operation
interface
Microprocessor
Motorola 6809 with 24Kb BPROM and 2 Kb RAM
System
Complies with the requirements for technical
Technical
instruments stipulated in IEC 348, VDE 0411, and
Standard
CSA File 45 111
Radio Interface Complies with legal requirements of the PRG (VDE
Suppression 0871, grade B)
Ambient
10 to 40°C
Operating Range