0% found this document useful (0 votes)
350 views67 pages

Rayomatic 20-100 Manual

Manual

Uploaded by

Paul Chiriacescu
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
350 views67 pages

Rayomatic 20-100 Manual

Manual

Uploaded by

Paul Chiriacescu
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 67

IRtec Rayomatic 20/40

IRtec Rayomatic 60/100


Process Infrared thermometer
Instruction Manual MM850341 ed. 06f
Instruction Manual MM850341 ed.06f

INTRODUCTORY NOTE

ATTENTION: THIS MANUAL IS VALID FOR IRTEC RAYOMATIC 20/40/60/100.

This publication contains operating instructions, as well as a description of the principles of operation, of
IRtec Rayomatic 20/40/60/100 IR thermometers.
This information covers all models of the instrument, including the basic equipment and its options and
accessories. This manual is a complete “USER GUIDE”, providing step-by-step instructions to operate
the instrument in each of its designed functions.

Eurotron has used the best care and efforts in preparing this book and believes the information in this
publication are accurate. The Eurotron products are subjected to continuous improvement, in order to
pursue the technological leadership; these improvements could require changes to the information of this
book.
Eurotron reserves the right to change such information without notice.

No part of this document may be stored in a retrieval system, or transmitted in any form, electronic or
mechanical, without prior written permission of Eurotron.

IRtec IR thermometers uses sophisticated analogic and digital technologies. Any maintenance operation
must be carried out by qualified personnel ONLY. We recommend to contact our technicians for any
support requirements.

IRtec is fully tested in conformity with the directive n°89/336/CEE Electromagnetic Compatibility.
Eurotron shall not be liable in any event, technical and publishing error or omissions, for any incidental
and consequential damages, in connection with, or arising out of the use of this book.

All right reserved


Copyright © 1998, 2009
EUROPEAN Headquarters USA Headquarters
Eurotron Instruments SpA E-Instruments Group LLC
Viale F.lli Casiraghi 409/413 172 Middletown Blvd – Suite B201
20099 Sesto S. Giovanni (MI) Langhorne, PA 19047
Tel. : +39-02 24 88 201 Tel.: 215 750 1212
FAX: +39-02 24 40 286 FAX: 215 750 1399
Mail: [email protected] Mail: [email protected]

2
Instruction Manual MM850341 ed.06f

TABLE OF CONTENTS
1 GENERAL DESCRIPTION 5
1.1 Specifications ............................................................................................................................7
1.1.1 IRtec Rayomatic 20 ...................................................................................................................7
1.1.2 IRtec Rayomatic 40 ...................................................................................................................9
1.1.3 IRtec Rayomatic 60 ................................................................................................................. 11
1.1.4 IRtec Rayomatic 100 ............................................................................................................... 13
1.2 Ordering code.......................................................................................................................... 15
1.2.1 IRtec Rayomatic 20 ................................................................................................................. 15
1.2.2 IRtec Rayomatic 40 ................................................................................................................. 16
1.2.3 IRtec Rayomatic 60 ................................................................................................................. 17
1.2.4 IRtec Rayomatic 100 ............................................................................................................... 18
2 PHYSICAL DESCRIPTION 19
2.1 IRtec Rayomatic 20/40 ............................................................................................................ 19
2.2 IRtec Rayomatic 60 ................................................................................................................. 20
2.3 IRtec Rayomatic 100 ............................................................................................................... 21
3 PRINCIPLE OF OPERATION 22
3.1 Modular systems ..................................................................................................................... 22
3.2 Basic elements ........................................................................................................................ 22
3.2.1 Monochromatic thermometers ................................................................................................. 22
3.2.2 Two-Color thermometers.........................................................................................................23
3.2.3 Optical system ......................................................................................................................... 23
3.2.4 Detectors ................................................................................................................................. 23
3.2.5 Input stage............................................................................................................................... 24
3.2.6 Microcontroller ......................................................................................................................... 24
3.2.7 Serial communication .............................................................................................................. 24
3.2.8 Output stage ............................................................................................................................ 24
3.2.9 Temperature compensation..................................................................................................... 25
3.2.10 Power supply ...................................................................................................................... 25
3.2.11 Laser pinpointing ................................................................................................................ 25
4 SIGNAL PROCESSING 26
4.1 Average ................................................................................................................................... 26
4.2 Peak ........................................................................................................................................ 26
4.3 Valley....................................................................................................................................... 26
4.4 Peak-Picker ............................................................................................................................. 27
4.4.1 Peak-Picker with reset............................................................................................................. 27
4.4.2 Peak-Picker delayed................................................................................................................ 27
4.5 Valley-Picker ........................................................................................................................... 28
4.5.1 Valley-Picker with reset ........................................................................................................... 28
4.6 Track and Hold ........................................................................................................................ 28
5 UNPACKING 30
6 INSTALLATION 31
6.1 Important notes........................................................................................................................ 31
6.2 Electrical connection................................................................................................................ 32
6.2.1 DigiMax II connections and programming ............................................................................... 34
6.2.2 DigiMax III connections and programming .............................................................................. 34
6.2.2 Laser Pinpointing systems....................................................................................................... 35
6.2.3 Reset input .............................................................................................................................. 36
6.3 Alarms ..................................................................................................................................... 36
6.3.1 Alarm output ............................................................................................................................ 37
6.4 Emissivity adjustment .............................................................................................................. 38
6.5 Positioning............................................................................................................................... 38
6.6 Mounting and Alignment.......................................................................................................... 38
7 MAINTENANCE 40
7.1 Rayomatic 100 fiber optic ........................................................................................................ 40
7.2 Rayomatic 60 fiber optic .......................................................................................................... 40
7.3 Purge Air Supply...................................................................................................................... 40
7.4 Water Supply ........................................................................................................................... 41
7.5 Optic cleaning.......................................................................................................................... 41
7.6 Mounting Device...................................................................................................................... 41

3
Instruction Manual MM850341 ed.06f

7.7 Interconnection Cable ............................................................................................................. 41


7.8 Storage ................................................................................................................................... 41
8 SUPPORT AND ACCESSORIES 42
8.1 Rayomatic 20/40 ..................................................................................................................... 43
8.1.1 Positioning plates .................................................................................................................... 44
8.1.2 Water cooling jacket................................................................................................................ 44
8.1.3 Air purge devices .................................................................................................................... 46
8.1.4 Flanges ................................................................................................................................... 47
8.1.5 Sighting tubes ......................................................................................................................... 49
8.1.6 Other accessories ................................................................................................................... 50
8.2 Rayomatic 60/100 ................................................................................................................... 51
8.2.1 Positioning plates .................................................................................................................... 52
8.2.2 Air purge devices .................................................................................................................... 52
8.2.3 Flanges ................................................................................................................................... 53
8.2.4 Sighting tubes ......................................................................................................................... 55
8.2.5 Other accessories ................................................................................................................... 55
8.3 Advice ..................................................................................................................................... 55
8.4 Air treatment equipment .......................................................................................................... 56
9 SERIAL COMMUNICATION 57
9.1 Smart/RS232 adapter ............................................................................................................. 57
9.1.1 Connections ............................................................................................................................ 57
10 CERTIFICATES 60
10.1 Warranty Terms ...................................................................................................................... 60
10.2 Letter of Conformity................................................................................................................. 60
APPENDIX 61
A1 EMC Conformity........................................................................................................................... 61
A2 How to determine an object emissivity......................................................................................... 64
A2.1 Typical Emissivity Values ........................................................................................................ 64
A2.2 Metals - Typical Emissivity Values .......................................................................................... 65
A2.3 Non-Metals - Typical Emissivity Values ................................................................................. 67

4
Instruction Manual MM850341 ed.06f

1 GENERAL DESCRIPTION
The temperature measurement of a liquid or gaseous compound has been successfully made with thermoelectric or
expansion thermometers thanks to the good thermal exchange between the fluid and the sensor.
With solid bodies a good thermal exchange is difficult to be obtained and an additional measuring error should be
considered.
Direct contact measurement is impractical when the object being measured is moving, or electrically hazardous, or
for any other reasons cannot be touched with a thermocouple. A non-contact IR temperature measurement is the
only solution to the above problems. Other applications benefit because non-contact thermometers do not add or
remove heat or disturb the process in any way.

Eurotron has applied the most recent development in IR sensor technology and, based on an extensive application
experience, has designed the new line of non-contact thermometers series IRtec Rayomatic applying a total
modularity concept. Each IRtec Rayomatic thermometer consists of a cylindrical chassis able to contain: lens, optical
filters, IR sensor and electronic. IRtec Rayomatic 100 has a remote optical head that allow a good location of the
instrument and the best accessibility of measuring head .

Eurotron IRtec infrared thermometers have been designed to be used on a wide range of industrial and research
applications.
These thermometers include the emissivity adjusment and a 4/20 mA linear output in order to work using the
DigiMax display units. DigiMax is a microprocessor unit and it is able to provide the necessary 24Vdc current loop
power supply. You can order the indicator complete of one or two programmable alarm relè.
Any other acquisition unit with 4/20 mA two wires current loop input can be used with the thermometers.

IRtec Rayomatic 20 IRtec Rayomatic 40

IRtec Rayomatic 20
IRtec Rayomatic 20 thermometers are easy to use and reliable pyrometers that can be manually set acting on
trimmers situated on the back side of the instrument.

IRtec Rayomatic 40
IRtec Rayomatic 40 are sophisticated thermometers that implement a digital communication protocol superimposed
to the standard 4-20mA current loop. This allows the user to set the parameters by means of a PC, a Bell202/RS232
adapter and the dedicated software as well as continuously monitoring the measurements.

IRtec Rayomatic 60
Fibre optic IRtec Rayomatic 60 infrared thermometers are designed for non contact temperature measurements.
Two series are available with different spectral band: series 100 in 0.9 µm and series 160 in 1.6 µm.

5
Instruction Manual MM850341 ed.06f

IRtec Rayomatic 60

IRtec Rayomatic 100


IRtec Rayomatic 100 thermometers represent the new frontier in non-contact temperature measurement. Eurotron
applied the most recent development in IR sensor technology and, based on an extensive application experience,
has designed the new line of PC reconfigurable and programmable IR thermometers.
The 2-colors principle measurement is performed at the same time by two indipendent detectors with different but
adjacent narrow band IR filters. By rationing the output of these 2 sensors, the temperature measurement becomes
indipendent from a number of factors that can affect the desired accuracy.
An internal modem allows using the same two wires 4/20mA current loop for smart communication. A smart/RS232
adapter is available for PC bi-directional communication. You can read the temperature measurement, set all
parameters (temperature range, averaging, peak-picking, alarms, emissivity, etc.).

IRtec Rayomatic 100

Rayomatic 20 Rayomatic 40 Rayomatic 60 Rayomatic 100


2-wire 4/20 mA output STD STD STD STD
Serial comm. Bell202 NO STD STD STD
ALARM NO STD STD NO
RESET NO STD STD NO
LASER OPT OPT OPT NO
Emissivity Adj. Rotative switches PC Rotative switches + PC PC
Configuration Factory PC PC PC

6
Instruction Manual MM850341 ed.06f

1.1 Specifications

1.1.1 IRtec Rayomatic 20


• Measuring ranges:
see par 1.2.1 Table A
• Accuracy: @ 23°C ±5°C and ε=1.0
IRtec Rayomatic 20-100 and 160: ±0.5% of rdg.
IRtec Rayomatic 20-510: ±1% of rdg.
IRtec Rayomatic 20-814 : ±1% of rdg. or ±1 °C whichever is greater
• Repeatability:
IRtec Rayomatic 20-100 and 160: ± 0.25% of the reading
IRtec Rayomatic 20-510 : ± 0.5% of the reading
IRtec Rayomatic 20-814 : ± 0.5% of the reading
• Response time:
IRtec Rayomatic 20-100 and 160: 28 ms (t95)
IRtec Rayomatic 20-510 : 100 ms (t95)
IRtec Rayomatic 20-814 : 100 ms (t95)
• Type of detector:
IRtec Rayomatic 20-100: Si
IRtec Rayomatic 20-160: InGaAs
IRtec Rayomatic 20-510 : Thermopile
IRtec Rayomatic 20-814 : Thermopile
• Spectral band:
see par 1.2.1 Table A
• Emissivity:
adjustable with 2 rotative switch from 0.30 to 1.00
• Working temperature:
from -20 to +60 °C (without cooling)
from 0 to +50 °C (for laser operation)
• Storage temperature:
from -30 to +70 °C
from -10 to +50 °C (for laser operation)
• Output signal:
4/20 mA 2 wire current loop – max load 700 Ω
• Temperature stability:
±0.1 °C/°C for the band exceeding +18 to +28°C
• Target pinpointing:
optional on-board laser pinpointing system with remote command
• Environmental rating:
IP65 (NEMA-4)
• Power supply:
4-20mA loop: from 12 to 32 Vdc
Laser pinpointing: 12 to 32 Vdc max. 100mA
• Dimension / Weight:
ø 45 mm x 200 mm - 0.5 Kg net weight

7
Instruction Manual MM850341 ed.06f

Note:
• Nominal target @95% energy
When the laser pinpointing system is installed, multiply the actual target diameter by 1.2

8
Instruction Manual MM850341 ed.06f

1.1.2 IRtec Rayomatic 40


• Accuracy: @ 23°C ±5°C and ε=1.0
IRtec Rayomatic 40-100: ±0.5% of rdg.
IRtec Rayomatic 40-160: ±0.5% of rdg.
IRtec Rayomatic 40-814: ±1% of rdg. or ±1°C whichever is greater
IRtec Rayomatic 40-390: ±1% of rdg.
IRtec Rayomatic 40-460: ±1% of rdg.
IRtec Rayomatic 40-510: ±1% of rdg.
IRtec Rayomatic 40-343: ±1% of rdg. or ±3°C whichever is greater
IRtec Rayomatic 40-790: ±1% of rdg. or ±2°C whichever is greater
• Repeatability:
IRtec Rayomatic 40-100: ±0.25% of rdg.
IRtec Rayomatic 40-160: ±0.25% of rdg.
IRtec Rayomatic 40-814: ±0.5% of rdg. or ±0.5°C whichever is greater
IRtec Rayomatic 40-390: ±0.5% of rdg.
IRtec Rayomatic 40-460: ±0.5% of rdg.
IRtec Rayomatic 40-510: ±0.5% of rdg.
IRtec Rayomatic 40-343: ±0.5% of rdg. or ±1.5°C whichever is greater
IRtec Rayomatic 40-790: ±0.5% of rdg. or ±1.5°C whichever is greater
• Response time:
IRtec Rayomatic 40-100: 28ms (t95)
IRtec Rayomatic 40-160: 28ms (t95)
IRtec Rayomatic 40-814: 100ms (t95)
IRtec Rayomatic 40-390: 100ms (t95)
IRtec Rayomatic 40-460: 100ms (t95)
IRtec Rayomatic 40-510: 100ms (t95)
IRtec Rayomatic 40-343: 1000ms (t95)
IRtec Rayomatic 40-790: 1000ms (t95)
• Type of detector:
IRtec Rayomatic 40-100: Si
IRtec Rayomatic 40-160: InGaAs
IRtec Rayomatic 40-814 : Thermopile
• Emissivity:
adjustable by PC from 0.30 to 1.00
• Temperature drift:
±0.1 °C/°C for the band exceeding +18 to +28°C
• Working temperature:
from -20 to +60 °C (without cooling)
from 0 to +50 °C (for laser operation)
• Storage temperature:
from -30 to +70 °C Si and InGaAs detectors
from +10 to +60 °C Thermopile
• Digital communication:
Bell 202 superimposed on 2-wire current loop (4/20 mA)
RS232 with optional adapter
• Output signal:
4/20 mA 2 wire current loop – max load 700 Ω
• Target pinpointing:
optional on-board laser pinpointing system with remote command
• Environmental rating: IP65 (NEMA-4)
• Power supply:
Current Loop: from 12 to 32 Vdc
Laser: from 12 to 32 Vdc
• Dimension / Weight: ø 45 mm x 200 mm - 0.5 Kg

9
Instruction Manual MM850341 ed.06f

Note:
• Nominal target @95% energy
• When the laser pinpointing system is installed, multipy the actual target diameter by 1.2

10
Instruction Manual MM850341 ed.06f

1.1.3 IRtec Rayomatic 60


Accuracy: ±(0.5% of rdg. + 1°C)
Repeatability: ±0.25% of rdg.
Spectral Band:
Series 100: 0.9 mm (Silicon detector)
Series 160: 1.6 mm (InGaAs detector)
Response Time:
adjustable by PC from 50ms to 10s
Reference temperature:
from +15 to +35°C
Emissivity: adjustable from 0.30 to 1.00 in 0.01 steps with selector swithces or from PC
Digital communication:
Bell 202 superimposed on 2-wire current loop
RS232 with optional adapter
Output signal: 4/20 mA 2-wire current loop - max load 700 Ω
Environmental rating: IP65 (NEMA-4)
Power supply:
Current loop: from 12 to 32 Vdc
Laser (optional): from 12 to 32 Vdc
Green Laser (optional): 12Vdc±2Vdc
Temperature stability : <0.05% of rdg./°C for temperature exceeding the reference band
Working temperature:
Optical head: 200°C max
Optic fiber: 200°C max
Electronic module: from -20 to +60°C
Green Laser (optional): from +20°C to +30°C
Storage temperature: from -30 to +70°C
Dimensions / Weight :
Electronic module: ø 45x200 mm / 0.8 kg
Optical head: ø 16x52 mm - Fast Lock
ø 30x94 mm – Threaded M1

11
Instruction Manual MM850341 ed.06f

Note:
• Nominal target @95% energy

12
Instruction Manual MM850341 ed.06f

1.1.4 IRtec Rayomatic 100


• Accuracy: @ 23°C ±5°C and ε=1.0
2-color: ±0.75% FS (with 95% attenuation)
monocromatic: ±0.5% FS
• Repeatability:
2-color: ±0.25% FS
monocromatic: ±0.25% FS
• Response time: 38ms (t95)
• Emissivity:
adjustable by PC from 0.30 to 1.00 in 0.01 steps
E-Slope adjustable by PC from 0.800 to 1.200 in 0.001 steps
• Temperature drift:
<±0.05 % rdg./°C for the band exceeding +18 to +28°C
• Working temperature:
from -20 to +60 °C (without cooling)
• Storage temperature:
from -30 to +70 °C
• Digital communication:
Bell 202 superimposed on 2-wire current loop
RS232 with optional adapter
• Output signal:
4/20 mA 2 wire current loop – max load 700 Ω
• Environmental rating: IP65 (NEMA-4)
• Power supply:
Loop: from 12 to 32 Vdc
• Operating temperature:
optical head: 200 °C max
fiber optic: 200 °C max
electronic module: from -20 to +60 °C
• Dimension / Weight:
Elecronic module ø 45 mm x 200 mm - 0.5 Kg
fiber optic ø 8 mm
optic head ø 16 mm x 52 mm

13
Instruction Manual MM850341 ed.06f

Note:
• Nominal target @95% energy

14
Instruction Manual MM850341 ed.06f

1.2 Ordering code

1.2.1 IRtec Rayomatic 20

cat. 1130 - A - B - C – D – E - F

Table A CWL φ target/distance Range


100-1 0.9µm 6mm @ 500mm 600 to 1600°C (1100 to 2900°F)
100-2 0.9µm 16mm @ 1000mm 600 to 1600°C (1100 to 2900°F)
160-1 1.6µm 6mm @ 500mm 300 to 1300°C (570 to 2350°F)
160-2 1.6µm 3mm @ 300mm 300 to 1300°C (570 to 2350°F)
160-3 1.6µm 2mm @ 100mm 300 to 1300°C (570 to 2350°F)
510-1 5.1µm 10.5mm @ 300mm 150 to 1300°C (300 to 2350°F)
510-2 5.1µm 17mm @ 500mm 800 to 2000°C (1470 to 3630°F)
814-1 8-14µm 21mm @ 600mm -25 to 1000°C (-15 to 1800°F)
814-5 8-14µm 80mm @ 1000mm 0 to 800°C (32 to 1450°F)
814-6 8-14µm 10mm @ 65mm 0 to 400°C (32 to 750°F)

Table B Signal output


2 4÷20 mA 2-wire current loop

Table C Pinpointing system


0 None
1 built-in pinpointing laser

Table D Electrical connection cable


1 2m long - 2 wires shielded
2 8m long - 2 wires shielded
9 Length on request

Table E Report of Calibration


0 None
1 Eurotron traceable SIT

15
Instruction Manual MM850341 ed.06f

1.2.2 IRtec Rayomatic 40

cat. 1140 - A - B - C – D – E – F

Table A - B CWL φ target/distance Range


100-1 0.9µm 6mm @ 500mm 600 to 1600°C (1100 to 2900°F)
100-2 0.9µm 16mm @ 1000mm 600 to 1600°C (1100 to 2900°F)
160-1 1.6µm 6mm @ 500mm 300 to 1300°C (570 to 2350°F)
160-2 1.6µm 3mm @ 300mm 300 to 1300°C (570 to 2350°F)
160-3 1.6µm 2mm @ 100mm 300 to 1300°C (570 to 2350°F)
814-1 8-14µm 21mm @ 600mm -25 to 1000°C (-15 to 1800°F)
814-2 8-14µm 9mm @ 300mm 0 to 1000°C (32 to 1800°F)
814-3 8-14µm 3.5mm @ 140mm 0 to 1000°C (32 to 1800°F)
814-5 8-14µm 80mm @ 1000mm 0 to 800°C (32 to 1450°F)
814-6 8-14µm 10mm @ 65mm 0 to 400°C (32 to 750°F)
343-1 3.43µm 26mm @ 500mm 100 to 400°C (210 to 750°F)
390-1 3.9µm 17mm @ 550mm 600 to 1300°C (1100 to 2350°F)
460-1 4.6µm 17mm @ 550mm 400 to 1600°C (750 to 2900°F)
510-1 5.1µm 17mm @ 550mm 150 to 1300°C (300 to 2350°F)
510-2 5.1µm 8.5mm @ 300mm 800 to 2000°C (1470 to 3630°F)
790-1 7.9µm 17mm @ 550mm 40 to 600°C (104 to 1100°F)

Table C Signal output


2 4÷20 mA 2-wire current loop with superimposed serial communication

Table D Pinpointing system


0 None
1 built-in laser pinpointing

Table E Electrical connection cable


1 2 m long - 6 wires shielded
2 8 m long - 6 wires shielded
9 Special length

Table F Report of Calibration


0 None
1 Eurotron NIST or EA traceable certificate with data

16
Instruction Manual MM850341 ed.06f

1.2.3 IRtec Rayomatic 60

cat. 1110 – A - B - C – D – E – F – G
Table A Range
100-1 from +600 to +1600°C
160-1 from +300 to +1300°C

Table B Signal output


2 4÷20 mA linear

Table C Optic fiber length


2 3.5 m
3 6m
4 8m
9 Special lenght

Table D Target / Distance


1 2mm OD @ 75mm – fast lock head
2 4mm OD @ 150mm – fast lock head
3 8mm OD @ 300mm – fast lock head
4 5mm OD @ 500mm – threaded M1
5 12mm OD @ 900mm – threaded M1
9 special

Table E Options
0 None
1 Red laser pinpointing (650nm)
2 Green laser pinpointing (532nm)

Table F Electrical connection


1 2m long - shielded cable
2 8m long - shielded cable
4 2m long – high temperature (200°C) shielded cable

Table G Calibration certificate


1 Eurotron NIST or EA traceable certificate with data

17
Instruction Manual MM850341 ed.06f

1.2.4 IRtec Rayomatic 100

cat. 1148 – 100 - A - B - C – D – E - F

Table A Range
1 600 to 1600°C (1100 to 2900°F)
2 800 to 2200°C (1470 to 3630°F)
3 1000 to 2200°C (1830 to 4000°F)
4 1500 to 2700°C (2700 to 4900°F)

Table B Signal output


2 4÷20 mA 2-wire current loop with superimposed serial communication

Table C Optic fiber length


1 1m
2 3.5 m
3 6m
4 8m
9 Special lenght

Table D Target / Distance


1 20mm OD @ 1000mm (50:1)
2 10mm OD @ 650mm (65:1)
3 5mm OD @ 350mm (70:1)
4 2mm OD @ 150mm (75:1)
5 1mm OD @ 100mm (100:1)

Table E Electrical connection


1 2m long - 6 wires standard shielded cable
2 8m long - 6 wires standard shielded cable
4 2m long - 6 wires high temperature (200°C) shielded cable
9 Special length shielded cable

Table F Calibration certificate


1 Eurotron NIST or EA traceable certificate with data

18
Instruction Manual MM850341 ed.06f

2 PHYSICAL DESCRIPTION

2.1 IRtec Rayomatic 20/40


IRtec Rayomatic 20/40 radiation thermometers are designed by using the most recent sensor technology. The
external case is made of extruded aluminium with M44x1.5 threading compatible with a wide range of accessory. The
case provides the room for the optic system, the pointing system, the optic filter, the sensor, the electronics for signal
conditioning, and the laser pinpointing system. The instruments are IP65 protected and the electrical connections are
very sturdy.
All thermometers could install an integrated laser pinpointing system.

There are two different series of IRtec Rayomatic thermometers available in order to satisfy all applications and
using conditions:

IRtec Rayomatic 20
Three different models are available:

MODEL SENSOR WAVELENGTH


IRtec Rayomatic 20 – 100 Silicon 0.9 µm
IRtec Rayomatic 20 – 160 InGaAs 1.6 µm
IRtec Rayomatic 20 – 814 Thermopile 8-14 µm

Removing the rear flange, the binding posts for 4/20 mA signal output and laser pinpointing system are accessible.
Two rotative switches allows the operator to set the correct emissivity value.
Response time, range, and special mathematical output functions can be programmed by factory on request.

Optional laser 4/20 mA


pinpointing P.S. signal output

Emissivity
Adj. switches

IRtec Rayomatic 40
IRtec Rayomatic 40 thermometers represent the new frontier in non-contact temperature measurement. Eurotron
applied the most recent development in IR sensor technology and, based on an extensive application experience,
has designed the new line of PC reconfigurable and programmable IR thermometers. An internal modem allows to

19
Instruction Manual MM850341 ed.06f

use the same two wire 4/20mA current loop for smart communication. A smart/RS232 adapter is available for PC bi-
directional communication. You can read the temperature measurement, set all parameters (range, averaging, peak-
picking, alarms, emissivity, etc.). The communication data is fixed; 1200baud, 8bit ,no parity, 1 stop bit.

Two auxiliary lines can be programmed and used for function and alarm reset and to drive alarm relè.

Four different models are available:

MODEL WAVELENGTH
IRtec Rayomatic 40 – 100 0.9 µm
IRtec Rayomatic 40 – 160 1.6 µm
IRtec Rayomatic 40 – 814 8-14 µm

Optional laser 4/20 mA


pinpointing P.S. signal output

Alarm output HW Reset

2.2 IRtec Rayomatic 60


Fibre optic IRtec Rayomatic 60 infrared thermometers are designed for non contact temperature measurements.
Two series are available with different spectral band: series 100 in 0.9 mm and series 160 in 1.6 mm.

Measuring System
Each IRtec 60 thermometer consists of three separate and interchangeable parts:
• Single lens optical adapter (head) to collect the IR radiation;
• Flexible fiber optic cable steel coated to transmit the IR radiation from the “head” to the electronic module;
• Microprocessor electronic module to convert the IR radiation in electric signal. The thermometer can be
ordered for different measuring ranges as specified in ordering instruction table A.

The Optic fiber IR thermometers are the ideal solution when:


• The ambient temperature is very high (up to 200 °C);
• The target is not easily accessible.

Optical head
Five different interchangeable optical adapters are available for differents target/distance vision cones.

Flexible fiber optic


You can choose between different fiber optic cable lengths. The fiber is protected by a flexible steel band for
environment applications up to 200°C.

WARNING
DO NOT CURVE THE FIBER OPTIC WITH RADIUS SMALLER THAN 200 MM !
SMALLER RADIUS CAN CAUSE A BREAKING OF THE FIBER.

Fast lock connectors


The optic head can be connected and disconnected from the process using a special fast connection system. It
allows the quick installing and maintenance of the measuring system.

Electronic module
The electronic module contains a detector, processing electronics, and Bell202 serial comunication interface. The
Bell202 interface consist of a modem to modulate the digital communication over the 4-20 mA 2-wire current loop.
The communication data is fixed; 1200baud, 8bit ,no parity, 1 stop bit.

20
Instruction Manual MM850341 ed.06f

This feature allows you to change the thermometer setting connecting a PC on the signal current loop using a simple
RS232 converter. E-slope, response time, and range can be set from the PC with the optional Configuration
Software. Also advanced function as Peak, Valley, Peak-Picker, Average, Track&Hold, etc. can be set on the IRtec
Rayomatic 60.

Removing the rear flange, the binding posts for 4/20 mA signal output and laser pinpointing system are accessible.
Two rotative switches allows the operator to set the correct emissivity value.
Response time, range, and special mathematical output functions can be programmed by PC.

Optional laser 4/20 mA


pinpointing P.S. signal output

Emissivity
Adj. switches

Two different models are available:

MODEL SENSOR WAVELENGTH


IRtec Rayomatic 60 – 100 Silicon 0.9 µm
IRtec Rayomatic 60 – 160 InGaAs 1.6 µm

2.3 IRtec Rayomatic 100


IRtec Rayomatic 100 radiation thermometers are designed by using the most recent sensor technology. The
external case is made of extruded aluminium with M44x1.5 threading compatible with a wide range of accessory. The
case provides the room for the optic system, the optic filter, the sensor, the electronics for signal conditioning, and the
laser pinpointing system. The optical head is fiber made of different length (1 mt., 3mt., 8 mt.) and grant an easy
installation with fast lock attack. The instruments are IP65 protected and the electrical connections are very sturdy.

WARNING
DO NOT CURVE THE FIBER OPTIC WITH RADIUS SMALLER THAN 200 MM !
SMALLER RADIUS CAN CAUSE A BREAKING OF THE FIBER.

The infrared radiation emitted from the target is focalized from a beamsplitter on 2 sensors inside the instrument. The
electronic module treats the termal signal and makes a report between the 2 different received signals, filtered on
different bands. As a function of this report, the memory assigns a measuring temperature. This system grant a right
temperature also if the target is obscured

Removing the rear flange, the binding posts interconnections are accessible.
On the instrument are situated 2 binding posts for 4/20 mA signal output system.

4/20 mA
signal output

Software for IRtec Rayomatic 40/60/100


An optional RS232 PC adapter and a configuration software allow the changes of the most important instrument
parameters: emissivity, response time, measuring span, average, peak-picker, decay, etc.
The LogMan data logger software is developed by Eurotron for displaying on a Personal Computer the trend of
temperature or help you to calculate the correct programmable parameters.
TM
The acquired data can also be stored in your hard-disk with Excel file format.

21
Instruction Manual MM850341 ed.06f

3 PRINCIPLE OF OPERATION

3.1 Modular systems


IRtec radiation thermometers uses a modular structure and component interchangeability to ensure an instrument to
be exactly suited for each particular application. The modular construction allows an easy add and substitution of any
option directly on the field. It also simplifies field troubleshooting, repair and calibration.
Different lenses, filters, IR sensors, electronic module, target sighting systems and accessories provide over 100
possible standard instruments to exactly fulfil your application requirement.

3.2 Basic elements

3.2.1 Monochromatic thermometers


IRtec radiation thermometers contains the following basic elements: optical system, detector, and signal-processing
electronic. The general features of these elements are first examined in the following sections.

IRtec Rayomatic 20

Pt100
Ambient Output
temperature D/A converter
compensation

(mV or µA) RISC


Σ-∆ converter Microcontrollor

12-32 Vcc
Laser Power Supply Loop
Power Supply

Data Acquisition
System

Bell 202/RS232 Personal Computer


converter
IRtec Rayomatic 40

Pt100
Ambient Modem
+
temperature BELL 202 D/A converter
compensation

Data Acquisition
(mV or µA) RISC
Max 450 ohm System
Σ-∆ converter Microcontrollor

Laser Power Supply

Allarm Reset - +

12-32 Vdc
Power Supply

Relè

22
Instruction Manual MM850341 ed.06f

3.2.2 Two-Color thermometers


The functioning system is the following.
The visual cone is determined by the optical head, using the achromatic lens and the fiber, that act as limiting range
(f.i. 20mm@1000mm). The fiber collect the energy and carry it in an optical device that divide the signal with a proper
mirror, filter it on 2 bands (0.88µm; 0.98µm) and focalize it on 2 sensors.
The infrared radiation received form the sensor is converted in electric signal that is not linear function of the
temperature of the target. The electric signal is treated from an electronic circuit inside the microprocessor that
generate an output signal in linear current 4-20mA.

Bell 202/RS232 Personal Computer


converter
IRtec Rayomatic 100

Modem
+
BELL 202 D/A converter

Data Acquisition
(mV or µA) RISC System
Σ-∆ converter Microcontrollor

Power Supply

Allarm Reset - +

12-32 Vdc
Power Supply

Relè

3.2.3 Optical system


The detector is positioned at a focus distance of a high quality lens. The lens forms an image of the source in the
plane of the detector. The size of the cone of radiation accepted from an axial point of the target and, the irradiance
at the detector, is determined by the clear aperture of the lens, while the active area of the detector determines the
field of view. The lens and detector then act as the aperture and field stop of the system respectively.

3.2.4 Detectors
Thermopile
A thermopile contain a large number of thermocouples in a very small area. The thermocouples are connected
together and the output is the difference between the ambient temperature (cold junction) and the temperature
collected by the optical system (hot junction).
Inspite of large number of thermocouples, the signal output is very small: few µV/°C. The response time of the
detector is very long because the detector must be warming up (or cooled) by the collected energy.
The waveband is determined by an additional optical filter chose by Eurotron for the specific application. This solution
permits to reduce disturbs due by atmosphere absorption and to maximise the output signal.

Photodiode
Both Silicium and InGaAs photodiode principle is completely different from thermopile: the collected energy is
transformed in a electron flow. The result is a current proportional with the incident energy.
This transformation is very fast (only few µS). This time the response time of the measuring system is limited by
electronics: high resolution and low energy consumption, make it a little bit slower.
The waveband is determined by the physical characteristic of the sensor.

23
Instruction Manual MM850341 ed.06f

Ratio detectors
Special IR applications could not be solved using the standard monochromatic thermometers. A ratio detector could
be the solution.
Typical problems which will cause an incorrect reading include:
• small objects e.g. to small to fill the target area
• smoke, dust or steam which obscure the line of sight
• dirty windows in the process
• emissivity of the product changes (due to changes in alloy or surface condition).

A ratio detector is a single chip twin detector with different working wavelengths. The ratio of the two simultaneous
measurements is independent from emissivity and the result is the correct target temperature.
Every ratio thermometer has a limit as to how much signal can be lost. This is referred as the reduction ratio that can
vary up to 90% without effecting the measurement accuracy. To be also kept in consideration that the loss in signal
can come from 3 sources:
• low emissivity of the target
• object too small to fill the cone of vision
• obstruction caused by steam, smoke, dirt and dirty windows

In all the above condition a ratio thermometer will operate with better accuracy than any monochromatic
thermometer. The IRtec “Rayomatic” thermometers was designed to operate with spectral bands with limited
temperature coefficient.

3.2.5 Input stage


Is the most important part of the thermometer. It should be able to interface the detector to the micro-controller. IRtec
Rayomatic series uses different kind of detectors (thermopile, photodiode, and ratio detectors) with different operating
specifications.
• Thermopile models needs electronics with high impedance input and very low drift.
• Photodiode models needs fast electronics.
• Ratio models needs electronics able to drive a double detector.
For the best modularity of the measuring system, Eurotron has developed special configurable electronics to meet all
requirements.

3.2.6 Microcontroller
Microcontroller has to manage all thermometer functionality. It should be fast, with low absorption, with integrate
memory, serial interface, etc.
Eurotron’s IRtec Rayomatic use a RISC processor with internal re-writable FLASH memory.

3.2.7 Serial communication


This module is included in IRtec Rayomatic 40/60 only. This module consist of a special modem to implement serial
communication on the thermometer.
The smart IRtec Rayomatic line uses a digital communication method superimposed on the 2 wire 4-20 mA loop to
communicate with PC, lap-top and/or handheld communicators.
The communication capability also means that a PC can be used both for remote calibration and transmitter operative
mode configuration.
The digital protocol operates using the frequency shift keying (FSK) principle, which is based on the Bell 202
communication standard where sine waves are superimposed on the dc analog signal to gives simultaneous analog
and digital communication
Temperature unit, peak hold, emissivity, average can be programmed through the digital port using either a PC or a
dedicated communicator.

3.2.8 Output stage


This module is based by a D/A converter to transform the temperature measurements from the microcontroller digital
format to the current signal. The signal output from the thermometer is a 4/20 mA on 2-wire current loop linear with
the temperature range.

24
Instruction Manual MM850341 ed.06f

3.2.9 Temperature compensation


All IR detector are ambient temperature sensible. Detectors measure the difference between detector and target
energy. To obtain the correct value of target temperature, it is necessary to measure the detector temperature and
add it to the detector measurement. A Pt100 is mounted on the detector and microcontroller reads the temperature
and makes the computation.

3.2.10 Power supply


A power supply stage is built-in the thermometer for adapting the current loop supply (from 12 to 32 Vdc) to all digital
and analogue circuits.

3.2.11 Laser pinpointing


IRtec Rayomatic 20/40/60 thermometers can be optionally equipped with a laser pinpointing system to simplify
alignement operation.
When target has small and critical dimensions the laser aiming system can be activated by operator to check and
adjust appropriate alignment of thermometer.
A separate (12-32 Vdc max. 100mA and 12Vdc±2Vdc for the green laser) power supply input allows operator to use it
only when necessary.

NOTE: THE WORKING TEMPERATURE FOR THE GREEN LASER IS FROM +20°C TO +30°C

WARNING
USE A SEPARATE POWER SUPPLY OF 12VDC±2VDC 100M A ONLY WHEN USING THE GREEN LASER

25
Instruction Manual MM850341 ed.06f

4 SIGNAL PROCESSING
When a radiation thermometer is employed in an on-line system, its signal may be subjected to large fluctuations.
Time function facilities are useful to retain the signal levels that are more likely to represent the true target
temperature.
Eurotron IR thermometers can be factory pre-set with one of the signal processing facilities: averaging, Track & Hold,
Peak, Valley, Peak-Picking, Valley-Picking, etc.

IRtec Rayomatic 40/60/100 functions can be programmed from user by using a Personal Computer, a
Bell202/RS232 adapter and a configuration software.

IRtec Rayomatic 20 can be programmed by factory only.

4.1 Average
Rapid temperature fluctuations about a true mean value can make the thermometer output unsuitable for recording
and control. In these cases, the average function can be used to provide a smoothed signal. The output signal Avu is
given by:

Avu =Avp + (M – Avp) * 2 ∆T / D

where:
M = Actual temperature
Avu = Output average
Avp = Previous average
∆T = sampling time = 20ms
D = Average time (sec) (programmable)

Temperature

Temperature with constant D>0


Range 0..99

Actual temperature
(D=0)

Time

4.2 Peak
The Peak function holds a maximum temperature and retain its until a higher value appear or an external reset signal
appear to decay the output to the current temperature measurement.

4.3 Valley
The Valley function holds the minimum temperature and retain its until a lower value appear or an external reset
signal appear to increase the output to the current temperature measurement.

26
Instruction Manual MM850341 ed.06f

4.4 Peak-Picker
The peak-picker function holds a peak signal and allows it to decay slowly until the arrival of the next peak. It is used
in such applications as rolling mills where the signal is occasionally lowered by steam, smoke, water droplets, metal
scale, etc. When Peak-Picker function is programmed, the 4/20 mA output will follows the mathematical expression
below:

U = Peak – (PickDec * 2 ∆T)

dove :
U = output signal
Peak = peak signal
PickDec = Decay rate °C/sec (programmable)
∆T = sampling time

Temperature
Decay rate (PickDec)
Programmable (0.0 to 1999.9 °C/S)

Time

4.4.1 Peak-Picker with reset


When the Peak-Picker with reset function is selected, the external reset input is used to force the signal output to the
current temperature measurement.

reset

4.4.2 Peak-Picker delayed


When the Peak-Picker delayed function is selected, you have to set the following parameters:
PickDec = Decay rate
Threshold = Temperature threshold level
StartPeak = delay for starting peak-picking
EndPeak = delay for ending peak-picking
Delayed Mode 2 = when selected, enable mode 2 output sustain

27
Instruction Manual MM850341 ed.06f

When the signal exceed the programmed trigger level (Threshold), the thermometer will wait the “StartPeak” delay
before starting with the standard Peak-Picker function. When the input signal will decrease below the trigger level
(Threshold), the thermometer will wait the “EndPeak” delay before reset the output signal.
The output signal will be reduced to low scale if Mode 1 is selected or to the last calculated value if Mode 2 is
selected.

Temperature

Decay rate (0.0 to 1999.9 °C/sec)

Mode 2

Trigger level EndPeak


StartPeak

Mode 1

Time

4.5 Valley-Picker
Reverse the Peak-picker function principle to describe the Valley Picker.

U = Valley + (PickDec * 2 ∆T)

where :
U = output signal
Valley = Valley signal
PickDec = Increasing rate °C/sec (programmable)
∆T = sampling time

4.5.1 Valley-Picker with reset


Reverse the Peak-Picker with reset function principle to describe the Valley-Picker with reset.

4.6 Track and Hold


The track (sample) and Hold function allows smoothed signals to be obtained from intermittent events. These may
originate from objects on a conveyor belt or targets where the view is periodically obscured by rotating machinery.
The output sampling is activated by a command signal received by an external switch that can be actuated by the
belt or rotating machinery itself. The output is held when the switch is operated until the next command is received.

28
Instruction Manual MM850341 ed.06f

Temp.

Time - tempo
Reset
Track & Hold

29
Instruction Manual MM850341 ed.06f

5 UNPACKING
Remove the instrument from its packing case and remove any shipping ties, clamps or packing materials.
Carefully follow any instruction given by any attached tags.
Inspect the instrument from scratches, dents, damages to case corners etc. which may have occurred during
shipment.
If any mechanical damage is noted, report the damage to the shipping carrier and then notify Eurotron directly or its
nearest agent, and save the damaged packaging for inspection.
A label, on the instrument, shows the serial number and model of the unit. Refer to these numbers for any inquiry for
service, spare parts supply or application and technical support requirements.

30
Instruction Manual MM850341 ed.06f

6 INSTALLATION
IRtec Rayomatic installing is very easy. The cylindrical aluminium chassis is externally threaded (M44x1.5) and
allows quick connecting with all available accessories.

The following procedure could help you for installing your IRtec Rayomatic 20/40/60 (monochromatic)
thermometer.
• Configure the thermometer. Set the emissivity on the IRtec Rayomatic 20 moving the 2 rotative switches.
Connect the IRtec Rayomatic 40 to the PC and set the mathematical function, the emissivity and the
temperature range by software. For Rayomatic 60 you can adjust the emissivity either by PC or by the
selector switches on the back panel.
• Connect, if necessary, all accessories (air purge, support, sighting tube, etc.). Fix the IRtec Rayomatic to the
process.
• Connect the signal cable from your data acquisition system to the instrument terminals. Pay attention to the
polarity.
• Aim accurately the thermometer to your target. Refers to Target vs. distance table on section 1. Be sure that
your target is bigger than the measuring area. The optional laser pinpointing can help you for best
installation.

The following procedure could help you for installing your IRtec Rayomatic 100 (2-Colors) thermometer.
• Install the thermometer on your plant.
• Supply the instrument and connect them to the PC. Use the Smart/RS232 converter.
• Run the configuring software.
• Select the right IRtec model
• Select the “Monocromatic” mode and start the PC temperature monitoring.
• Move the IRtec Rayomatic until the measured temperature correspond to maximum (the optical cone of
thermometer is centered to the target).
• Select the two-color mode on software.
• Adjust the “E-Slope” (non-gray) function until the measured temperature correspond to the real target
temperature.

6.1 Important notes

ATTENTION
ARE TO BE AVOID BACKSCATTERING, SUNLIGHT OR OTHER LIGHTING FONTS. TO CHECK THIS CONDITIONS: POINT THE
INSTRUMENT, LOWER THE TEMPERATURE OF THE PROCESS UNDER 600°C: DISTURB RADIATIONS ARE PRESENT ONLY IF THE
INSTRUMENT READ OVER 600°C. POINTING ON SUNLIGHT YOU CAN READ ALSO 400°C BUT THIS IS FOR THE SPECTRUM
COMPONENTS OF PLANK EMISSION AT 600°C. A CLOSED MEASURING AREA WILL BE THE BEST SOLUTION.

IMPORTANT
DO NOT PULL OUT THE ELECTRONIC BOARD FROM THE CYLINDER. IT COULD BE DIFFICULT TO REASSEMBLE THE THERMOMETER.

WARNING
DO NOT INSTALLING THE THERMOMETER IN ROOM WITH TEMPERATURE HIGHER THAN +60°C WITHOUT AN ADEQUATE COOLING
SYSTEM.

WARNING
DO NOT CURVE THE FIBER OPTIC WITH RADIUS SMALLER THAN 200 MM !
SMALLER RADIUS CAN CAUSE A BREAKING OF THE FIBER.

WARNING
IRTEC RAYOMATIC 100 OPTICAL FIBER ASSEMBLY
REMOVE THE CENTRAL SCREW (SEE FIGURE BELOW) BEFORE TO REMOVE THE FIBER OPTIC FROM THE ELECTRONIC MODULE.

31
Instruction Manual MM850341 ed.06f

Remove

IMPORTANT
IRTEC RAYOMATIC 60 OPTICAL FIBER ASSEMBLY
IN ORDER TO ACHIEVE THE BEST RESULTS, BEFORE SCREWING THE OPTICAL FIBER TO THE MAIN INSTRUMENT, SWITCH ON THE
LASER PINPOINTING SYSTEM AND PROPERLY TURN THE FIBER WITH THE AIM OF OBTAINING THE MAXIMUM BEAM INTENSITY EXITING
FROM THE OPTICAL HEAD.

6.2 Electrical connection

IMPORTANT
TO SHEILD THE CONNECTION CABLES, EXTEND THE SHIELD WIRE UNTILL THE DESTINATION POINT (DISPLAY, CONTROLLER…) AND
CONNECT IT TO GROUND.

The temperature signal is converted, amplified and linearized by the internal electronics that makes available a 4/20
mA linear output. Eurotron supply a connector for 4/20 mA active current loop connection.
The unit is supplied complete with a shielded multi-wires electrical cable.

The following procedure could help you for electrical cable substitution on your IRtec Rayomatic thermometer:
1. Remove the 3 screw on the flange on the rear of the instrument.
2. Open the cable gland and insert the shielded signal cable (φ 3.9 – 6.5 mm) through the flange.
3. Unsheathe the cable and prepare the wire for connection.
4. Unscrew partially the cable clamp. Insert correctly the wires. Pay attention to the polarity. Screw the cable clamp.
5. Put near the flange to the cylinder and screw firmly.
6. Close the cable gland.
7. Insert the 3 screw to the flange holes and accompain them manually. Use an adequate screw driver. Look up (1
or 2 rotation) alternatively the 3 screw.

IRtec Rayomatic 20 connections:

100.0 -
IN 4/20mA Laser Power
Supply
+ Output
4/20mA

Indicator with 24Vdc internal power supply

32
Instruction Manual MM850341 ed.06f

12-32 Vdc
Power Supply

IN 4/20mA
+ -
100.0
Laser Power
- Supply

Indicator with 24Vdc external power supply

IRtec Rayomatic 40/60 connections:

-
100.0 IN 4/20mA
Laser Power
+ Output Supply
4/20mA
Indicator with internal power supply

Reset Alarm
12-32 Vdc
Power Supply

IN 4/20mA
+ -
100.0
Laser Power
- Supply

Indicator with external power supply

WARNING
DO NOT GO THROUGH THE SIGNAL CABLE NEAR ELECTROMAGNETIC SOURCE
(EG. ELECTRICAL MOTORS, HIGH POWER CABLE, WELDINGS, ETC). IF NECESSARY USE METAL CONDUIT.

IRtec Rayomatic 100 connections:

-
100.0 I N 4/ 20mA
+ Output
4/20mA
Indicator with internal power supply

33
Instruction Manual MM850341 ed.06f

12-32 Vdc
Power Supply

IN 4/20mA
+ -
100.0
-

Indicator with external power supply

6.2.1 DigiMax II connections and programming


NOTE: REFER TO DIGIMAX II MANAUL FOR PROGRAMMING DETAILS.
The outoput of the pyrometer is a current signal. In any case you must power the termometer: for this aim, use the
following configuration where the pyrometer is considered as a transmitter and is powered by the Digimax itself.
6 7
DigiMax II

IRtec Rayomatic
-
+

• Connect the loop+ and loop- wires to clamps 6 and 7.

DigiMax II
Connectors on the back side:
use the Rayomatic as a
transmitter.

In order to set the Digimax indicator, follow the procedure beginning from “Normal operation” and access to the
subsequent pages to modify the parameters:

• Press z , the indicator shows “Par”


• Press z , the indicator shows “Conf”
• Press twice z to enter the configuration, the indicator shows “PASS”
• Insert the configuration password: “3333” and press z
• Insert the configuration code “9000”, it means a 4-20mA input and press z
• The indicator shows “Scdd”, set the number of decimal digits (0 to 3) and press z
• The indicator shows “ScLo”, set the beginning value of the scale (e.g. 0°C) and press z
• The indicator shows “Schi” , set the end of scale value (e.g. 600°C) and press z
• Set to “0” the “sqrt” parameter (square root) and press z

6.2.2 DigiMax III connections and programming


NOTE: REFER TO DIGIMAX III MANAUL FOR PROGRAMMING DETAILS.

34
Instruction Manual MM850341 ed.06f

The outoput of the pyrometer is a current signal, hence you need to insert the proper connector (a 2.5ohm shunt
resistor) to adapt the voltage input of the DigiMax. Anyway, you must power the termometer: for this aim, use the
following configuration where the pyrometer is considered as a transmitter and is powered by the Digimax itself.
• Connect the loop+ and loop- wires to clamps 6 and 9.

9
DigiMax III
DigiMax III Connectors on the back side:
notice the external shunt
5 6
1 2 resistance between 5 and 6

IRtec Rayomatic
-
220 Vac Rs
+

NOTE: THE SHUNT RESISTOR IS SUPPLIED WITH THE DIGIMAX

EXAMPLE: how to read continuously a temperature by means of Digimax III in a range of 0-600°C and without any
programming alarm. In order to set the Digimax indicator, follow the procedure beginning from “Operator mode” and
access to the subsequent pages to modify the parameters:

• Press a number of times enough to show the indication “PASS”


• Insert the password code “33” with keys and press
• Press , the indicator shows “conf” and press
• The indicator shows “Conf”
• Insert the configuration code “8000” with keys (it means a 10-50mV input) and press
• Press , the indicator shows “Unit”
• Select °C or °F with keys
• Press , the indicator shows “Scdd”
• set the number of decimal digits (0 to 3) with keys and press
• Press , the indicator shows “Sc.Lo”
• set the beginning value of the scale (e.g. 0°C) with keys and press
• Press , the indicator shows “Sc.Hi”
• set the end of scale value (e.g. 600°C) with keys and press
• press to return to the actual reading

6.2.2 Laser Pinpointing systems


IRtec Rayomatic 20/40 thermometers can be optionally ordered complete of an integrated laser pinpointing system.
The advantages of an integrate laser system with an independent power supply are:
• More accuracy in target aiming
• More stability with vibration
• Aiming system always available for maintenance
• Remote laser switching on/off

NOTE: Laser pinpointing is NOT available on Ratio thermometers

Remove the flange on the back of the instrument to access to the power supply terminals.
Connect the power supply (12…32 Vdc max. 100mA) to the instrument. Before switch the laser on verify the correct
polarity.

WARNING
USE A SEPARATE POWER SUPPLY OF 12VDC±2VDC 100M A ONLY WHEN USING THE GREEN LASER

35
Instruction Manual MM850341 ed.06f

Laser power supply


12..32 Vdc

IRtec Rayomatic 20

Laser power supply


12..32 Vdc

IRtec Rayomatic 40/60

WARNING: LASER EMISSIONS

DO NOT STARE INTO BEAM OR VIEW DIRECTLY WITH OPTICAL INSTRUMENTS.


CLASS 2 LASER PRODUCT CONFORMS TO IEC825-1:1993 AND 21CFR CAP.1 PAR.J
MAX. OUTPUT: <1mW – WAVELENGTH: 630-670 nm

6.2.3 Reset input


IRtec Rayomatic 40 thermometers only, have an external input terminal that allows resetting of special functions
(e.g. Alarms, Peak-Picker, Peak, etc.).

ATTENTION
IF THE HARDWARE RESET IS ENABLED FOR A MATHEMATICAL FUNCTION, IT IS USED FOR ALARM RESETTING TOO.

Laser Power Supply 4/20 mA


(optional) output

Alarm output Reset input


Reset connections

6.3 Alarms
IRtec Rayomatic 40 thermometers only, have two programmable alarms (Low and High) with hysteresis.
• System alarm: always enabled with output directed to the hardware alarm output. The thermometer activate the
output when the detector temperature is out of the operative range (e.g. When the water cooling system is
malfunctioning). It is factory set and you cannot disable it.
• Low alarm: you can set the minimum temperature of your process. When the temperature goes below the
programmed value, the alarm output became active.

36
Instruction Manual MM850341 ed.06f

• High alarm: you can set the maximum temperature of your process. When the temperature goes above the
programmed value, the alarm output became active.
• Hysteresis: you can program the hysteresis band for re-entry from the alarm status.
• Alarm Output: Two output modes are available. When alarms are programmed, the instrument uses the current
loop output to signal the alarm status; the high alarm status change the current loop to 21 mA, the low alarm
status change the output to 3.75 mA. You can also enable an hardware output line (20 mA sink current) to drive
an external relais.
• Alarm mode: if the external alarm output is selected, you can choose between normally high or normally low
output.
• Alarm reset: you can enable the reset input to reset (quit) the alarm status.

NOTE:
• All alarm set point are referred to the effective measured temperature.
• The alarm output and reset input terminals are available on model IRtec Rayomatic 40 only.
• Anyway, the alarm output is ever enabled to signal the system alarm event.

Temperature

Max setpoint

Hysteresis

Min setpoint

Time - tempo

Reset

Alarm output

6.3.1 Alarm output


IRtec Rayomatic 40 thermometers only, have a programmable alarms output terminal to allows the connection with
an external relais or a data acquisition system (max. 20mA). Anyway, this output is used by the thermometer to signal
if the operative temperature of the electronic module is out of range.

37
Instruction Manual MM850341 ed.06f

4/20 mA
output +
1300 °C
24 Vdc
Power suppy

Relè
Reset
Alarm output
Alarm output and reset connections

6.4 Emissivity adjustment


IRtec Rayomatic 20/60 has two rotative switches to set the emissivity of your target. You can change it in the range
from 0.30 to 1.00.
• Remove the rubber plug on the rear of the instrument.
• Use a small, plastic screw driver to set the emissivity to the correct value. Thermometer uses the “00” value to
set the emissivity to 1.00. Values lower than “30” are considered by the thermometer equal to 0.30.

Decimal Centesimal

Emissivity Adjustment
Es. 85 = 0.85
IRtec Rayomatic 20

IRtec Rayomatic 40 emissivity can be changed only by using a PC connected to the current loop.

IRtec Rayomatic 100 E-Slope can be changed only by using a PC connected to the current loop.

6.5 Positioning
The thermometer has to be installed in an accessible place for any further maintenance operation, and it hasn't to be
exposed to excessive heat, smoke and steam.
The optical path between the lens and the target, should be as much free as possible from smoke and steam.
The pointing axis should be placed with a 90° angle against the target surface; an angle smaller than 45° could be
acceptable for most materials.

NOTE: A LARGE NUMBER OF MATERIALS HAVE THE EMISSIVITY COEFFICIENT DEPENDENT ON THE VIEWING ANGLE. LARGE
ANGLES (MORE THAN 45°) CAN INDUCE A LARGE ERROR IN MEASUREMENT.

Before the installation, it iss useful to look at the tables supplied together with the instrument, in order to determine
the correct distance and visual field.

NOTE: USING THE MONOCHROMATIC THERMOMETERS THE TARGET AREA HAVE TO BE LARGER THAN THE CALCULATED
MEASURING AREA.

6.6 Mounting and Alignment

38
Instruction Manual MM850341 ed.06f

The Raryomatic IR beam profile is a cone described by the optical charts. The IR monochromatic thermometer must
be installed so that the target is bigger than circle described by optics and the optical path is free from obstruction
(steam, dust, etc.).
Install the mounting plate in a suitable place in the most favourable position from a thermal, and mechanical point of
view. Mount on the plate, if it is necessary, the cooling device and/or the purge air ejector. If a cooling jacket has to
be used, use the water outlet on the upper part of the support in order not to create air bubbles in the jacket.

WARNING
IT ADVISABLE TO USE A PROTECTIVE SCREEN DURING THE THERMOMETER ALIGNMENT OPERATIONS IF THE TARGET HAS A STRONG
RADIATION.

In order to align the support with the target, look through the support hole (without the thermometer) and find the best
position. It's important for the optic path to be free from any obstacle. In particular, when the surface is targeted
through a hole, the hole diameter has to be big enough according to the distance from the instrument. If optional laser
pinpointing system is installed, switch it on for spot centre viewing.

39
Instruction Manual MM850341 ed.06f

7 MAINTENANCE
Each IRtec Rayomatic is factory calibrated and certified against Eurotron Standards and shipped with a Report of
calibration stating the nominal and actual values and the deviation error. The instrument calibration has to be
periodically verified. IRtec uses a sophisticated analog and digital technology. All the maintenance operations must
be carried on by qualified personnel. Eurotron supplies instructions and operative modalities for any maintenance
operation. Please contact Eurotron for any support.

For a correct working of the instrument, the optic system must be kept clean and it mustn't reach temperatures higher
than the specified ones. The maintenance department should ensure these working conditions with a periodical
check of the cooling system and cleaning the lens.

7.1 Rayomatic 100 fiber optic


IRtec Rayomatic 100 allow to change the fiber optic without the instrument calibration.

IMPORTANT
REMOVE THE CENTRAL SCREW (SEE FIGURE BELOW) BEFORE TO REMOVE THE FIBER OPTIC FROM THE ELECTRONIC MODULE.

Remove

WARNING
DO NOT CURVE THE FIBER OPTIC WITH RADIUS SMALLER THAN 200 MM !
SMALLER RADIUS CAN CAUSE A BREAKING OF THE FIBER.

7.2 Rayomatic 60 fiber optic


IRtec Rayomatic 60 allow to change the fiber optic without the instrument calibration.

IMPORTANT
IRTEC RAYOMATIC 60 OPTICAL FIBER ASSEMBLY
IN ORDER TO ACHIEVE THE BEST RESULTS, BEFORE SCREWING THE OPTICAL FIBER TO THE MAIN INSTRUMENT, SWITCH ON THE
LASER PINPOINTING SYSTEM AND PROPERLY TURN THE FIBER WITH THE AIM OF OBTAINING THE MAXIMUM BEAM INTENSITY EXITING
FROM THE OPTICAL HEAD.

WARNING
DO NOT CURVE THE FIBER OPTIC WITH RADIUS SMALLER THAN 200 MM !
SMALLER RADIUS CAN CAUSE A BREAKING OF THE FIBER.

7.3 Purge Air Supply


The air filters cleanliness must be checked at regular intervals.
Our suggestion is to check it every day, then, according to your experience, find a correct time interval.
If the optic system reaches temperatures higher than the working one, it has to be recalibrate in Eurotron
laboratories.

40
Instruction Manual MM850341 ed.06f

The purge air device is to be accurately checked, as the diffuser may become obstructed by non-clean air. When this
happens, the air flow from the diffuser is not uniform, and dust particles appear on the lens.
In this case, the diffuser should be drowned in a detergent solution and blown with compressed air, then dried. A
good air filtering can solve this problems.

7.4 Water Supply


Verify the water flow according to your experience: daily first, and then when the system is running well, weekly.
Check the thermometer temperature: it has to be high enough to prevent any condense formation. Once the water
continuity is defined, it's enough to verify the support temperature, that has to be slightly warm.
If the thermometer reaches too high temperatures, due to water absence or to a flow partial interruption, it has to be
verified and calibrated by Eurotron.

7.5 Optic cleaning


Remove the connection wires from the terminal board.
Remove the thermometer from its support. Verify the cleanliness of the internal part of the cylinder removing any dirt
particles or oil.
If necessary, clean the lens with a very soft cloth and then reinstall the thermometer.
Verify the alignment and reconnect everything.

Use an air purge system to keep the lens clean.

7.6 Mounting Device


Verify at regular intervals that these devices are in good conditions and that no damage has occurred.

7.7 Interconnection Cable


Verify at regular intervals that it is in good conditions and that no damage has occurred. Verify also the good
connections with the indicator or the acquisition system.

7.8 Storage
Store the instrument in the original package, at temperature from –30°C to +70°C (from +10 to +70°C for thermopile
sensors) with non condensing R.H. less than 90%.

41
Instruction Manual MM850341 ed.06f

8 SUPPORT AND ACCESSORIES


This chapter shows the full range of available accessories that can be supplied with the measuring systems. Special
configurations or mounting systems can be realised by Eurotron to satisfy every application necessity. Eurotron
R&D department will be happy to help you solving the problem with the best choice.

42
Instruction Manual MM850341 ed.06f

8.1 Rayomatic 20/40

43
Instruction Manual MM850341 ed.06f

8.1.1 Positioning plates


When the thermometer is installed with standard environmental conditions you do not need to use any special cooling
or lens cleaner systems. The thermometer could be installed directly on your process.
To make the positioning more easy use the Eurotron supports.

2-axis orientable plate cod. EE290072

1-axis orientable plate cod. EE290074

8.1.2 Water cooling jacket


When the thermometer is installed in a very hot environment it is useful to cool its electronics using a special cooling
jacket. Use clean water in order to create no obstruction and cold enough to ease the thermometer working.

44
Instruction Manual MM850341 ed.06f

Water cooling jacket cod. EE290075

IMPORTANT

DO NOT USE TOO COOL WATER AS IT COULD CREATE CONDENSE ON THE SUPPORT OR ON THE LENS.
IT IS ADVISABLE TO HAVE A LOW WATER FLOW TO KEEP THE TEMPERATURE OF THE SUPPORT HIGHER THAN THE DEW-POINT OF A
HOT AND HUMID DAY.
THE FLOW SHOULD NOT BE HIGHER THAN 0.7 BAR AND THE AMBIENT TEMPERATURE SHOULD BE LOWER THAN 200°C.

The cooling water line has to be made using the following precautions:
ƒ The outlet of the cooling jacket must be kept free in order to control the flow and water temperature;
ƒ It's useful to use a thermometer to control the water temperature.

NOTE: IN CLOSE CIRCUIT SYSTEMS, ALWAYS USE A FLOW WATER WITH A LOW LEVEL ALARM.

Before the installation, it is useful to look at the tables supplied together with the instrument, in order to determine the
correct distance and visual field.

To properly install the water cooling jacket, follow this procedure:

Spacer

Jacket end

Top flange
Rayomatic 20/40 Cooling jacket Purge air adapter
with or without lens

Mounting sequence of cooling system EE290075 + Rayomatic

ƒ Insert the Rayomatic inside the jacket and fix it.


ƒ Insert the cable through the spacer and insert it inside the jacket in order to fasten the Rayomatic

45
Instruction Manual MM850341 ed.06f

Cooling jacket mounting: back side

ƒ Fix the flange and the purge air adapter (eventually with a proper lens) with 4 screws:

Water cooling jacket mounting: flange and purge air adapter M44x1.5 (EE280329 or EE280315)

ƒ Insert the cable through the jacket end and fix it at the end of the system

Water cooling jacket: final aspect

8.1.3 Air purge devices


To eliminate fumes or vapours from the front of the optic head, and to keep the lens clean, an air flux is used. This air
flux can also be used to cool, when necessary the thermometer.

46
Instruction Manual MM850341 ed.06f

Air purge system with M44x1.5 M adapter cat. EE290076

Air purge system with 1” ½ BSPP F adapter cat. EE290079

Air purge system cat. EE290077

8.1.4 Flanges
Code From ∅ To ∅ Fori Material
EE280113 1 ½ “ BSPP male ¾ “ BSPP female 4 x ∅7.2 mm AISI 304
EE280223 1 ½ “ BSPP male ¾ “ BSPP female 4 x ∅15.9 mm AISI 304
EE280292 1 ½ “ BSPP male ¼ “ BSPP female 4 x ∅6 mm AISI 304
EE280115 Through hole 1 ½” BSPP AISI 304
EE280282 Through hole 1 ½” BSPP 4 x ∅6 mm AISI 304

47
Instruction Manual MM850341 ed.06f

EE280113
22
10
Ø7.2
Ø127
Ø101.6

1" 1/2 BSPP


3/4" BSPP
EE280115
6
Ø157

1" 1/2 BSPP

EE280223
22
10
Ø15.9
Ø127
Ø101.6
1" 1/2 BSPP
3/4" BSPP

48
Instruction Manual MM850341 ed.06f

EE280292
18
Ø6 6
Ø157
Ø101

1" 1/4 BSPP

1" 1/2 BSPP


EE280282
Ø6 6
Ø157
Ø101

1" 1/2 BSPP

8.1.5 Sighting tubes


The sighting tubes must be used when the target is covered by fumes or flames. They shold be used together with an
air purge system and help you to have a clean viewing path from target to the thermometer.

Code Material Tube length Instrument connection


EE280059 AISI 310 300 mm 1 ½ “ BSPP Male (1)
EE280278 AISI 310 600 mm 1 ½ “ BSPP Male (1)
EE280212 INCONEL 600 300 mm 1 ½ “ BSPP Male (1)
EE280279 INCONEL 600 600 mm 1 ½ “ BSPP Male (1)
EE280215 ALLUMINA 600 mm 1 ½ “ BSPP Male (1)
EE280216 ALLUMINA 600 mm 1 ½ “ BSPP Male (1)
EE280218 Silicon Carbide 450 mm 1 ½ “ BSPP Male (1)
EE280286 Aluminium 1000 mm M44 x 1.5 Female (2)
EE280293 Aluminium 200 mm M44 x 1.5 Female (2)
EE280294 Aluminium 300 mm M44 x 1.5 Female (2)
EE280114 INCONEL 600 150 mm ¾ “ BSPP Male (1)
EE280222 INCONEL 600 300 mm ¾ “ BSPP Male (1)
EE280281 AISI 310 150 mm ¾ “ BSPP Male (1)

49
Instruction Manual MM850341 ed.06f

EE280280 AISI 310 300 mm ¾ “ BSPP Male (1)


EE280137 INCONEL 600 400 mm 1 ¼” BSPP Male (1)
(1) compatible with cat. EE290079 air purge system
(2) compatible with cat. EE290076 air purge system

8.1.6 Other accessories


CODE DESCRIPTION
BB530200 Bell202/RS232 adapter (IRtec Rayomatic 40/100)
BB530018 RS232 adapter (IRtec Rayomatic 20)
BB260195 Setup Software (IRtec Rayomatic 40/100)
BB260196 Setup & LogMan Graphic Software (IRtec Rayomatic 40/100 only)

50
Instruction Manual MM850341 ed.06f

8.2 Rayomatic 60/100

51
Instruction Manual MM850341 ed.06f

8.2.1 Positioning plates


When the thermometer is installed with standard environmental conditions you do not need to use any special cooling
or lens cleaner systems. The thermometer could be installed directly on your process.
To make the positioning more easy use the Eurotron supports.

cod. EE290074 1-axis orientable plate for electronic module

8.2.2 Air purge devices

EE290060

99
55

Ø32

Ø30

70

84

Compact Air purge code EE290060 complete of fast lock adapter and positioning plate (EE290063)

52
Instruction Manual MM850341 ed.06f

EE290046

1" 1/2 BSPP

88
Ø30
Ø69

Ø63
68

Air purge code EE290046 complete of fast lock adapter

8.2.3 Flanges
Code From ∅ To ∅ Fori Material
EE280113 1 ½ “ BSPP ¾ “ BSPP 4 x ∅7.2 mm AISI 304
EE280223 1 ½ “ BSPP ¾ “ BSPP 4 x ∅15.9 mm AISI 304
EE280292 1 ½ “ BSPP ¼ “ BSPP 4 x ∅6 mm AISI 304
EE280115 Through hole 1 ½” BSPP AISI 304
EE280282 Through hole 1 ½” BSPP 4 x ∅6 mm AISI 304

EE280113
22
10
Ø7.2
Ø127
Ø101.6
1" 1/2 BSPP
3/4" BSPP

53
Instruction Manual MM850341 ed.06f

EE280115
6
Ø157

1" 1/2 BSPP


EE280223
22
10
Ø15.9
Ø127
Ø101.6

1" 1/2 BSPP


3/4" BSPP

EE280292
18
Ø6 6
Ø157
Ø101
1" 1/4 BSPP

1" 1/2 BSPP

54
Instruction Manual MM850341 ed.06f

EE280282
Ø6 6
Ø157
Ø101

1" 1/2 BSPP


8.2.4 Sighting tubes
The sighting tubes must be used when the target is covered by fumes or flames. They shold be used together with an
air purge system and help you to have a clean viewing path from target to the thermometer.

Code Material Tube length Process connection


EE280059 AISI 310 300 mm 1 ½ “ BSPP Male
EE280278 AISI 310 600 mm 1 ½ “ BSPP Male
EE280212 INCONEL 600 300 mm 1 ½ “ BSPP Male
EE280215 ALLUMINA 600 mm 1 ½ “ BSPP Male
EE280278 AISI 314/310 600 mm 1 ½ “ BSPP Male
EE280279 INCONEL 600 600 mm 1 ½ “ BSPP Male
EE280216 ALLUMINA 600 mm 1 ½ “ BSPP Male
EE280217 Silicon Carbide 450 mm 1 ½ “ BSPP Male
EE280114 INCONEL 600 150 mm ¾ “ BSPP Male
EE280222 INCONEL 600 300 mm ¾ “ BSPP Male
EE280281 AISI 310 150 mm ¾ “ BSPP Male
EE280280 AISI 310 300 mm ¾ “ BSPP Male
EE280137 INCONEL 600 400 mm 1 ¼” BSPP Male

8.2.5 Other accessories


CODE DESCRIPTION
BB530200 Bell202/RS232 adapter (IRtec Rayomatic 40/100)
BB530018 RS232 adapter (IRtec Rayomatic 20)
BB260195 Setup Software (IRtec Rayomatic 40/100)
BB260196 Setup & LogMan Graphic Software (IRtec Rayomatic 40/100 only)

8.3 Advice
It's important to use very clean air to prevent dirt from depositing on the main lens. Every particle in the air could
obstruct part of the diffuser, and modify the aerodynamic characteristics in the air flux. So, the flux would be modified,
and dirt air could reach the lens. The air system must be periodically checked. To add a filtering system is strongly
recommended.

Eurotron can supply an air filtering system (cod. EE290015) to be used on the already present air network. This
system needs dry and clean air at a pressure up to 10 bar. If a compressed air network isn't available, a fan
generating system can be used.

According to the application, it could be necessary to cool the thermometer using an air flow. The maximum pressure
for an air purge system is 0.7 bar. Just for a normal air purge, only the tenth part of this flow is required.

55
Instruction Manual MM850341 ed.06f

Remember that the air tube must have a diameter bigger enough not to have flow losses.

IMPORTANT NOTE:

WHEN A PURGE AIR IS USED, IT MUST BE CLEAN AND DRY. THE DIFFUSOR PRODUCES A LAMINAR FLOW THAT KEEPS THE LENS
CLEAN. IF THE AIR ISN'T PERFECTLY CLEAN, TRACKS OF OIL AND DIRT MAY DEPOSIT ON THE LENS.

IMPORTANT NOTE:

WHEN MEASURING TEMPERATURES INSIDE OVENS, DO NOT USE ANY AIR COMING FROM THE FAN OF THE OVEN. THE AIR FROM THE
OVEN, IN FACT, MAY VARY, AND THE PRESSURE ON THE THERMOMETER MAY BE NOT SUFFICIENT TO GRANT ITS CLEANING AND
COOLING.

8.4 Air treatment equipment


The reduction and air filtering group (code EE290015) is composed by a modular system including:
• An interception valve (1/8")
• A pressure reducer with a filtering element
• An anti-oil filter
• A pressure gauge (0 ÷ 10 bar)

Two different air flow controllers are available:


1) For low flows: when only the purge air is required.
2) For high flows: when an air cooling system is required instead of a water one.

We suggest to minimise the distance between the controller group and the thermometer:

56
Instruction Manual MM850341 ed.06f

9 SERIAL COMMUNICATION
IRtec Rayomatic 40 is a programmable infrared thermometer using the same current loop wires to transmit and
receive digital informations from a Personal Computer. To do this, Eurotron has included inside the unit, a modem
with Bell202 communication protocol.

9.1 Smart/RS232 adapter


For connecting a PC to the IRtec Rayomatic 40/100, it is necessary to use the optional SMART/RS232 adapter (cat.
BB530200). The SMART/RS232 adapter could be used for the following operations:
• Powering and Programming the unit (fig. 1). Switch to LAB Setup position.
• Programming the unit connected to an acquisition loop (fig. 2, 3 and 4). You can have different hardware
configurations depending on the indicator internal resistance. Switch to PROCESS Setup position.
• Powering the laser pinpointing (fig.5). You have to switch to 24Vdc Supply position.

Two WindowsTM 98/2000/NT/XP compatible software packages are available to set the unit and for data acquisition.

Adapter Selector

RS232
connection
towards PC

Connections to
Rayomatic

SMART/RS232 adapter (BB530200)

9.1.1 Connections

Converter
SMART/RS232

fig. 1 – Direct PC connection

57
Instruction Manual MM850341 ed.06f

Converter
SMART/RS232

_
Indicator
4-20 mA

_ + Internal resistance:
Ri = 220..270Ω
Power Supply
12-32 Vdc

fig. 2 – Connection on a powered current loop

Add R0 in order to have:


220Ω<R0+Ri<270Ω

Converter
SMART/RS232

_
R0
Indicator
4-20 mA

_ + Internal resistance:
Ri<220Ω
Power supply
12-32 Vdc

fig. 3 – Connection on a powered current loop

Use: 220Ω<R0<270Ω

Converter
SMART/RS232

_
R0
Indicator
4-20 mA

_ + Internal resistance:
Ri>270Ω
Power supply
12-32 Vdc

fig. 4 – Connection on a powered current loop

58
Instruction Manual MM850341 ed.06f

Converter
SMART/RS232

_
Indicator
4-20 mA

_ +

Power supply
12-32 Vdc
(12Vdc±2Vdc for
the green laser)

fig 5 – Laser pinpointing supply

WARNING
USE A SEPARATE POWER SUPPLY OF 12VDC±2VDC 100M A ONLY WHEN USING THE GREEN LASER

59
Instruction Manual MM850341 ed.06f

10 CERTIFICATES

10.1 Warranty Terms


Each instrument is shipped with a Warranty Certificate that indicates the validity conditions of the warranty itself.
Eurotron warrants its products against defects in materials and workmanship.
If the unit should malfunction, it must be returned during the warranty period, transportation prepaid, to Eurotron for
evaluation. Upon examination, if the unit is found to be defective it will be repaired or replaced at no charge.
Eurotron‘s WARRANTY does not apply to defects resulting from any action of the purchaser, including but not
limited to mishandling, improper interfacing, operation outside of design limits, improper repair, or unauthorized
modification. This WARRANTY is VOID if the unit shows evidence of having been tampered with or shows evidence
of being damaged as a result of excessive corrosion; or current, heat, moisture or vibration; improper specification;
misapplication; misuse or other operating conditions outside of Eurotron‘s control.
This warranty applies to the original purchaser only. Please include a copy of the original invoice or a small service
charge may be applied.

Direct all warranty and repair requests/inquiries to the Eurotron Customer Service Department. BEFORE
RETURNING ANY PRODUCT(S) TO EUROTRON, PURCHASER MUST OBTAIN AN AUTHORIZED RETURN (AR)
NUMBER FROM EUROTRON‘S CUSTOMER SERVICE DEPARTMENT (IN ORDER TO AVOID PROCESSING
DELAYS).
The purchaser is responsible for shipping charges, freight, insurance and proper packaging to prevent breakage in
transit.

10.2 Letter of Conformity


Each instrument is shipped with a Letter of Conformity, to grant that the characteristics of the instrument correspond
to the required ones, and that the instrument calibration is traceable to the National and International Standards.

60
Instruction Manual MM850341 ed.06f

APPENDIX

A1 EMC Conformity
The instrument is designed to fulfil the prevision of the directive 89/336/CEE Electromagnetic Compatibility.
In the following page you will find the EMC declaration of conformity

61
Instruction Manual MM850341 ed.06f

Declaration of Conformity

We : Eurotron Instruments S.p.A.


(Supplier's name)

Viale F.lli Casiraghi, 409/413 20099 Sesto S. Giovanni (MI) - Italy


(Address)

declare under our sole responsibility that the product :


IR thermometers series IRtec Rayomatic 20 & 40
(Name and type)

Cat. 1130 & cat. 1140


(Model)

to which this declaration relates is in conformity with the following


normative documents :
EN 50082-2 (3/95)
IEC 1000-4-2 / IEC 1000-4-4 / IEC 1000-4-11
ENV 50140 - ENV 50141 - ENV 50204

EN 55011
(Title, number and date of issue of normative documents)

following the prevision of directive :


89/336/CEE Electromagnetic Compatibility (EMC)

Sesto S. Giovanni, July 20th, 2000


(Place and date of issue) (Signature of authorised person)

eurotron

62
Instruction Manual MM850341 ed.06f

Declaration of Conformity

We : Eurotron Instruments S.p.A.


(Supplier's name)

Viale F.lli Casiraghi, 409/413 20099 Sesto S. Giovanni (MI) - Italy


(Address)

declare under our sole responsibility that the product :


IR thermometers series IRtec Rayomatic 100
(Name and type)

Cat. 1148
(Model)

to which this declaration relates is in conformity with the following


normative documents :
EN 50082-2 (3/95)
IEC 1000-4-2 / IEC 1000-4-4 / IEC 1000-4-11
ENV 50140 - ENV 50141 - ENV 50204

EN 55011
(Title, number and date of issue of normative documents)

following the prevision of directive :


89/336/CEE Electromagnetic Compatibility (EMC)

Sesto S. Giovanni, July 20th, 2000


(Place and date of issue) (Signature of authorised person)

eurotron

63
Instruction Manual MM850341 ed.06f

A2 How to determine an object emissivity


Emissivity is the measure of an object ability to absorb, transmit, and emit infrared energy. It can have a value from 0
(shiny mirror) to 1.0 (blackbody). If a value of emissivity higher than the actual one is set, the output will read low,
provided that the target temperature is above the ambient one. For example, if 0.95 is set in and the actual emissivity
is 0.9, the reading will be lower than the true temperature when the target temperature is above the ambient one.

The emissivity can be determined by one of the following methods, in order of preference:

1. Determine the actual temperature of the material using a sensor such as a RTD, thermocouple or another
suitable method. Next, measure the object temperature and adjust the emissivity setting until the correct value
is reached. This is the correct emissivity for the measured material.

2. For relatively low temperature objects (up to 260°C or 500°F, place a piece of tape, such as a masking, on the
object. Make sure the tape is large enough to cover the field of view. Next, measure the tape temperature
using an emissivity setting of 0.95. Finally, measure an adjacent area on the object and adjust the emissivity
setting until the same temperature is reached. This is the correct emissivity for the measured material.

3. If a portion of the surface of the object can be coated, use a flat black paint, which will have an emissivity of
about 0.98. Next, measure the painted area using an emissivity setting of 0.98. Finally, measure an adjacent
area on the object and adjust the emissivity setting until the same temperature is reached. This is the correct
emissivity for the measured material.

A2.1 Typical Emissivity Values


The following table provides a brief reference guide to determine emissivity and can be used when one of the above
methods is not practical. Emissivity values shown in the table below are only approximate, since several parameters
may effect the emissivity of an object. These include the following ones:

1. Temperature
2. Angle of measurement
3. Geometry (plane, concave, convex, etc.)
4. Thickness
5. Surface quality (polished, rough, oxidized, sandblasted)
6. Spectral region of measurement
7. Transmissivity (i.e., thin film plastics)

64
Instruction Manual MM850341 ed.06f

A2.2 Metals - Typical Emissivity Values

1.0 µm 1.6 µm 5.1 µm 8-14 µm


Aluminium
Non-Oxidized 0.1-0.2 0.02-0.2 0.02-0.2 0.02-0.1
Oxidized 0.4 0.4 0.2-0.4 0.2-0.4
Alloy A 3003
Oxidized — 0.4 0.4 0.3
Roughened 0.2-0.8 0.2-0.6 0.1-0.4 0.1-0.3
Polished 0.1-0.2 0.02-0.1 0.02-0.1 0.02-0.1
Brass
Polished 0.8-0.95 0.01-0.05 0.01-0.05 0.01-0.05
Burnished — — 0.3 0.3
Oxidized 0.6 0.6 0.5 0.5
Carbon
Non-oxidized 0.8-0.95 0.8-0.9 0.8-0.9 0.8-0.9
Graphite 0.8-0.9 0.8-0.9 0.7-0.9 0.7-0.8
Chromium 0.4 0.4 0.03-0.3 0.02-0.2
Copper
Polished 0.05 0.03 0.03 0-03
Roughened 0.05-0.2 0.05-0.2 0.05-0.15 0.05-0.1
Oxidized 0.2-0.8 0.2-0.9 0.5-0.8 0.4-0.8
Gold 0.3 0.01-0.1 0.01-0.1 0.01-0.1
Haynes Alloy 0.5-0.9 0.6-0.9 0.3-0.8 0.3-0.8
Inconel
Oxidized 0.4-0.9 0.6-0.9 0.6-0.9 0.7-0.95
Sandblasted 0.3-0.4 0.3-0.6 0.3-0.6 0.3-0.6
Electropolished 0.2-0.5 0.25 0.15 0.15
Iron
Oxidized 0.4-0.8 0.5-0.9 0.6-0.9 0.5-0.9
Non-oxidized 0.35 0.1-0.3 0.05-0.25 0.05-0.2
Rusted — 0.6-0.9 0.5-0.8 0.5-0.7
Molten 0.35 0.4-0.6 — —
Iron Cast
Oxidized 0.7-0.9 0.7-0.9 0.65-0.95 0.6-0.95
Non-oxidized 0.35 0.3 0.25 0.2
Molten 0.35 0.3-0.4 0.2-0.3 0.2-0.3
Iron Wrought
Dull 0.9 0.9 0.9 0.9
Lead
Polished 0.35 0.05-0.2 0.05-0.2 0.05-0.1
Rough 0 65 0.6 0.4 0-4
Oxidized — 0.3-0.7 0.2-0.6 0.2-0.6
Magnesium 0.3-0.8 0.05-0.3 0.03-0.15 0.02-0.1
Mercury — 0.05-0.15 0.05-0.15 0.05-0.15
Molybdenum
Oxidized 0.5-0.9 0.4-0.9 0.3-0.7 0.2-0.6
Non-oxidized 0.25-0.35 0.1-0.3 0.1-0.15 0.1

65
Instruction Manual MM850341 ed.06f

1.0 µm 1.6 µm 5.1 µm 8-14 µm

Monel (Ni-Cu) 0.3 0.2-0.6 0.1-0.5 0.1-0.14


Nickel
Oxidized 0.8-0.9 0.4-0.7 0.3-0.6 0.2-0.5
Electrolytic 0.2-0.4 0.1-0.3 0.1-0.15 0.05-0.15
Platinum
Black — 0.95 0.9 0.9
Silver 0.04 0.02 0.02 0.02
Steel
Cold-Rolled 0.8-0.9 0.8-0.9 0.8-0.9 0.7-0.9
Ground Sheet — — 0.5-0.7 0.4-0.6
Polished Sheet 0.35 0.25 0.15 0.1
Molten 0.35 0.25-0.4 0.1-0.2 —
Oxidized 0.8-0.9 0.8-0.9 0.7-0.9 0.7-0.9
Stainless 0.35 0.2-0.9 0.15-0.8 0.1-0.8
Tin (Non-oxidized) 0.25 0.1-0.3 0.05 0.05
Titanium
Polished 0.5-0.75 0.3-0.5 0.1-0.3 0.05-0.2
Oxidized — 0.6-0.8 0.5-0.7 0.5-0.6
Tungsten 0.1-0.6 0.05-0.5 0.03
Polished 0.35-0.4 0.1-0.3 0.05-0.25 0.03-0.1
Zinc
Oxidized 0.6 0.15 0.1 0.1
Polished 0.5 0.05 0.03 0.02

66
Instruction Manual MM850341 ed.06f

A2.3 Non-Metals - Typical Emissivity Values


1.0 µm 2.2 µm 5.1 µm 8-14 µm

Asbestos 0.9 0.8 0.9 0.95


Asphalt — — 0.95 0.95
Basalt — — 0.7 0.7
Carborundum — 0.95 0.9 0.9
Ceramic 0.4 0.8-0.95 0.85-0.95 0.95
Clay — 0.8-0.95 0.85-0.95 0.95
Concrete 0.65 0.9 0.9 0.95
Cloth — — 0 95 0.95
Glass
Plate — 0.2 0.98 0.85
"Gob" — 0.4-0.9 0.9
Gravel — — 0.95 0.95
Gypsum — — 0.4-0.97 0.8-0.95
Ice — — — 0.98
Limestone — — 0.4-0.98 0.98
Paint 0.9-0.95
Paper (any colour) — — 0.95 0.9S
Plastic (opaque, — — 0.95 0.95
over 20 mils)
Rubber — — 0.9 0 95
Sand — — 0.9 0.9
Snow — — 0.9
Soil — — — 0.9-0.98
Water — — — 0.93
Wood, Natural — — 0.9-0.95 0.9-0.95

To optimize surface temperature measurements consider the following guidelines:


1. Determine the object emissivity using the instrument to be used for the measurement.
2. Avoid reflections by shielding the object from surrounding high temperature sources.
3. For Higher temperature objects use shorter wavelength instruments, whenever any overlap occurs.
4. For semi-transparent materials such as plastic films and glasses, assure that the background is uniform and lower in temperature than the
object.
5. Mount the sensor perpendicular to the surface whenever the emissivity is less than 0.9. In any case, do not exceed angles more than 30
degrees from incidence.

67

You might also like