Rayomatic 20-100 Manual
Rayomatic 20-100 Manual
INTRODUCTORY NOTE
This publication contains operating instructions, as well as a description of the principles of operation, of
IRtec Rayomatic 20/40/60/100 IR thermometers.
This information covers all models of the instrument, including the basic equipment and its options and
accessories. This manual is a complete “USER GUIDE”, providing step-by-step instructions to operate
the instrument in each of its designed functions.
Eurotron has used the best care and efforts in preparing this book and believes the information in this
publication are accurate. The Eurotron products are subjected to continuous improvement, in order to
pursue the technological leadership; these improvements could require changes to the information of this
book.
Eurotron reserves the right to change such information without notice.
No part of this document may be stored in a retrieval system, or transmitted in any form, electronic or
mechanical, without prior written permission of Eurotron.
IRtec IR thermometers uses sophisticated analogic and digital technologies. Any maintenance operation
must be carried out by qualified personnel ONLY. We recommend to contact our technicians for any
support requirements.
IRtec is fully tested in conformity with the directive n°89/336/CEE Electromagnetic Compatibility.
Eurotron shall not be liable in any event, technical and publishing error or omissions, for any incidental
and consequential damages, in connection with, or arising out of the use of this book.
2
Instruction Manual MM850341 ed.06f
TABLE OF CONTENTS
1 GENERAL DESCRIPTION 5
1.1 Specifications ............................................................................................................................7
1.1.1 IRtec Rayomatic 20 ...................................................................................................................7
1.1.2 IRtec Rayomatic 40 ...................................................................................................................9
1.1.3 IRtec Rayomatic 60 ................................................................................................................. 11
1.1.4 IRtec Rayomatic 100 ............................................................................................................... 13
1.2 Ordering code.......................................................................................................................... 15
1.2.1 IRtec Rayomatic 20 ................................................................................................................. 15
1.2.2 IRtec Rayomatic 40 ................................................................................................................. 16
1.2.3 IRtec Rayomatic 60 ................................................................................................................. 17
1.2.4 IRtec Rayomatic 100 ............................................................................................................... 18
2 PHYSICAL DESCRIPTION 19
2.1 IRtec Rayomatic 20/40 ............................................................................................................ 19
2.2 IRtec Rayomatic 60 ................................................................................................................. 20
2.3 IRtec Rayomatic 100 ............................................................................................................... 21
3 PRINCIPLE OF OPERATION 22
3.1 Modular systems ..................................................................................................................... 22
3.2 Basic elements ........................................................................................................................ 22
3.2.1 Monochromatic thermometers ................................................................................................. 22
3.2.2 Two-Color thermometers.........................................................................................................23
3.2.3 Optical system ......................................................................................................................... 23
3.2.4 Detectors ................................................................................................................................. 23
3.2.5 Input stage............................................................................................................................... 24
3.2.6 Microcontroller ......................................................................................................................... 24
3.2.7 Serial communication .............................................................................................................. 24
3.2.8 Output stage ............................................................................................................................ 24
3.2.9 Temperature compensation..................................................................................................... 25
3.2.10 Power supply ...................................................................................................................... 25
3.2.11 Laser pinpointing ................................................................................................................ 25
4 SIGNAL PROCESSING 26
4.1 Average ................................................................................................................................... 26
4.2 Peak ........................................................................................................................................ 26
4.3 Valley....................................................................................................................................... 26
4.4 Peak-Picker ............................................................................................................................. 27
4.4.1 Peak-Picker with reset............................................................................................................. 27
4.4.2 Peak-Picker delayed................................................................................................................ 27
4.5 Valley-Picker ........................................................................................................................... 28
4.5.1 Valley-Picker with reset ........................................................................................................... 28
4.6 Track and Hold ........................................................................................................................ 28
5 UNPACKING 30
6 INSTALLATION 31
6.1 Important notes........................................................................................................................ 31
6.2 Electrical connection................................................................................................................ 32
6.2.1 DigiMax II connections and programming ............................................................................... 34
6.2.2 DigiMax III connections and programming .............................................................................. 34
6.2.2 Laser Pinpointing systems....................................................................................................... 35
6.2.3 Reset input .............................................................................................................................. 36
6.3 Alarms ..................................................................................................................................... 36
6.3.1 Alarm output ............................................................................................................................ 37
6.4 Emissivity adjustment .............................................................................................................. 38
6.5 Positioning............................................................................................................................... 38
6.6 Mounting and Alignment.......................................................................................................... 38
7 MAINTENANCE 40
7.1 Rayomatic 100 fiber optic ........................................................................................................ 40
7.2 Rayomatic 60 fiber optic .......................................................................................................... 40
7.3 Purge Air Supply...................................................................................................................... 40
7.4 Water Supply ........................................................................................................................... 41
7.5 Optic cleaning.......................................................................................................................... 41
7.6 Mounting Device...................................................................................................................... 41
3
Instruction Manual MM850341 ed.06f
4
Instruction Manual MM850341 ed.06f
1 GENERAL DESCRIPTION
The temperature measurement of a liquid or gaseous compound has been successfully made with thermoelectric or
expansion thermometers thanks to the good thermal exchange between the fluid and the sensor.
With solid bodies a good thermal exchange is difficult to be obtained and an additional measuring error should be
considered.
Direct contact measurement is impractical when the object being measured is moving, or electrically hazardous, or
for any other reasons cannot be touched with a thermocouple. A non-contact IR temperature measurement is the
only solution to the above problems. Other applications benefit because non-contact thermometers do not add or
remove heat or disturb the process in any way.
Eurotron has applied the most recent development in IR sensor technology and, based on an extensive application
experience, has designed the new line of non-contact thermometers series IRtec Rayomatic applying a total
modularity concept. Each IRtec Rayomatic thermometer consists of a cylindrical chassis able to contain: lens, optical
filters, IR sensor and electronic. IRtec Rayomatic 100 has a remote optical head that allow a good location of the
instrument and the best accessibility of measuring head .
Eurotron IRtec infrared thermometers have been designed to be used on a wide range of industrial and research
applications.
These thermometers include the emissivity adjusment and a 4/20 mA linear output in order to work using the
DigiMax display units. DigiMax is a microprocessor unit and it is able to provide the necessary 24Vdc current loop
power supply. You can order the indicator complete of one or two programmable alarm relè.
Any other acquisition unit with 4/20 mA two wires current loop input can be used with the thermometers.
IRtec Rayomatic 20
IRtec Rayomatic 20 thermometers are easy to use and reliable pyrometers that can be manually set acting on
trimmers situated on the back side of the instrument.
IRtec Rayomatic 40
IRtec Rayomatic 40 are sophisticated thermometers that implement a digital communication protocol superimposed
to the standard 4-20mA current loop. This allows the user to set the parameters by means of a PC, a Bell202/RS232
adapter and the dedicated software as well as continuously monitoring the measurements.
IRtec Rayomatic 60
Fibre optic IRtec Rayomatic 60 infrared thermometers are designed for non contact temperature measurements.
Two series are available with different spectral band: series 100 in 0.9 µm and series 160 in 1.6 µm.
5
Instruction Manual MM850341 ed.06f
IRtec Rayomatic 60
6
Instruction Manual MM850341 ed.06f
1.1 Specifications
7
Instruction Manual MM850341 ed.06f
Note:
• Nominal target @95% energy
When the laser pinpointing system is installed, multiply the actual target diameter by 1.2
8
Instruction Manual MM850341 ed.06f
9
Instruction Manual MM850341 ed.06f
Note:
• Nominal target @95% energy
• When the laser pinpointing system is installed, multipy the actual target diameter by 1.2
10
Instruction Manual MM850341 ed.06f
11
Instruction Manual MM850341 ed.06f
Note:
• Nominal target @95% energy
12
Instruction Manual MM850341 ed.06f
13
Instruction Manual MM850341 ed.06f
Note:
• Nominal target @95% energy
14
Instruction Manual MM850341 ed.06f
cat. 1130 - A - B - C – D – E - F
15
Instruction Manual MM850341 ed.06f
cat. 1140 - A - B - C – D – E – F
16
Instruction Manual MM850341 ed.06f
cat. 1110 – A - B - C – D – E – F – G
Table A Range
100-1 from +600 to +1600°C
160-1 from +300 to +1300°C
Table E Options
0 None
1 Red laser pinpointing (650nm)
2 Green laser pinpointing (532nm)
17
Instruction Manual MM850341 ed.06f
Table A Range
1 600 to 1600°C (1100 to 2900°F)
2 800 to 2200°C (1470 to 3630°F)
3 1000 to 2200°C (1830 to 4000°F)
4 1500 to 2700°C (2700 to 4900°F)
18
Instruction Manual MM850341 ed.06f
2 PHYSICAL DESCRIPTION
There are two different series of IRtec Rayomatic thermometers available in order to satisfy all applications and
using conditions:
IRtec Rayomatic 20
Three different models are available:
Removing the rear flange, the binding posts for 4/20 mA signal output and laser pinpointing system are accessible.
Two rotative switches allows the operator to set the correct emissivity value.
Response time, range, and special mathematical output functions can be programmed by factory on request.
Emissivity
Adj. switches
IRtec Rayomatic 40
IRtec Rayomatic 40 thermometers represent the new frontier in non-contact temperature measurement. Eurotron
applied the most recent development in IR sensor technology and, based on an extensive application experience,
has designed the new line of PC reconfigurable and programmable IR thermometers. An internal modem allows to
19
Instruction Manual MM850341 ed.06f
use the same two wire 4/20mA current loop for smart communication. A smart/RS232 adapter is available for PC bi-
directional communication. You can read the temperature measurement, set all parameters (range, averaging, peak-
picking, alarms, emissivity, etc.). The communication data is fixed; 1200baud, 8bit ,no parity, 1 stop bit.
Two auxiliary lines can be programmed and used for function and alarm reset and to drive alarm relè.
MODEL WAVELENGTH
IRtec Rayomatic 40 – 100 0.9 µm
IRtec Rayomatic 40 – 160 1.6 µm
IRtec Rayomatic 40 – 814 8-14 µm
Measuring System
Each IRtec 60 thermometer consists of three separate and interchangeable parts:
• Single lens optical adapter (head) to collect the IR radiation;
• Flexible fiber optic cable steel coated to transmit the IR radiation from the “head” to the electronic module;
• Microprocessor electronic module to convert the IR radiation in electric signal. The thermometer can be
ordered for different measuring ranges as specified in ordering instruction table A.
Optical head
Five different interchangeable optical adapters are available for differents target/distance vision cones.
WARNING
DO NOT CURVE THE FIBER OPTIC WITH RADIUS SMALLER THAN 200 MM !
SMALLER RADIUS CAN CAUSE A BREAKING OF THE FIBER.
Electronic module
The electronic module contains a detector, processing electronics, and Bell202 serial comunication interface. The
Bell202 interface consist of a modem to modulate the digital communication over the 4-20 mA 2-wire current loop.
The communication data is fixed; 1200baud, 8bit ,no parity, 1 stop bit.
20
Instruction Manual MM850341 ed.06f
This feature allows you to change the thermometer setting connecting a PC on the signal current loop using a simple
RS232 converter. E-slope, response time, and range can be set from the PC with the optional Configuration
Software. Also advanced function as Peak, Valley, Peak-Picker, Average, Track&Hold, etc. can be set on the IRtec
Rayomatic 60.
Removing the rear flange, the binding posts for 4/20 mA signal output and laser pinpointing system are accessible.
Two rotative switches allows the operator to set the correct emissivity value.
Response time, range, and special mathematical output functions can be programmed by PC.
Emissivity
Adj. switches
WARNING
DO NOT CURVE THE FIBER OPTIC WITH RADIUS SMALLER THAN 200 MM !
SMALLER RADIUS CAN CAUSE A BREAKING OF THE FIBER.
The infrared radiation emitted from the target is focalized from a beamsplitter on 2 sensors inside the instrument. The
electronic module treats the termal signal and makes a report between the 2 different received signals, filtered on
different bands. As a function of this report, the memory assigns a measuring temperature. This system grant a right
temperature also if the target is obscured
Removing the rear flange, the binding posts interconnections are accessible.
On the instrument are situated 2 binding posts for 4/20 mA signal output system.
4/20 mA
signal output
21
Instruction Manual MM850341 ed.06f
3 PRINCIPLE OF OPERATION
IRtec Rayomatic 20
Pt100
Ambient Output
temperature D/A converter
compensation
12-32 Vcc
Laser Power Supply Loop
Power Supply
Data Acquisition
System
Pt100
Ambient Modem
+
temperature BELL 202 D/A converter
compensation
Data Acquisition
(mV or µA) RISC
Max 450 ohm System
Σ-∆ converter Microcontrollor
Allarm Reset - +
12-32 Vdc
Power Supply
Relè
22
Instruction Manual MM850341 ed.06f
Modem
+
BELL 202 D/A converter
Data Acquisition
(mV or µA) RISC System
Σ-∆ converter Microcontrollor
Power Supply
Allarm Reset - +
12-32 Vdc
Power Supply
Relè
3.2.4 Detectors
Thermopile
A thermopile contain a large number of thermocouples in a very small area. The thermocouples are connected
together and the output is the difference between the ambient temperature (cold junction) and the temperature
collected by the optical system (hot junction).
Inspite of large number of thermocouples, the signal output is very small: few µV/°C. The response time of the
detector is very long because the detector must be warming up (or cooled) by the collected energy.
The waveband is determined by an additional optical filter chose by Eurotron for the specific application. This solution
permits to reduce disturbs due by atmosphere absorption and to maximise the output signal.
Photodiode
Both Silicium and InGaAs photodiode principle is completely different from thermopile: the collected energy is
transformed in a electron flow. The result is a current proportional with the incident energy.
This transformation is very fast (only few µS). This time the response time of the measuring system is limited by
electronics: high resolution and low energy consumption, make it a little bit slower.
The waveband is determined by the physical characteristic of the sensor.
23
Instruction Manual MM850341 ed.06f
Ratio detectors
Special IR applications could not be solved using the standard monochromatic thermometers. A ratio detector could
be the solution.
Typical problems which will cause an incorrect reading include:
• small objects e.g. to small to fill the target area
• smoke, dust or steam which obscure the line of sight
• dirty windows in the process
• emissivity of the product changes (due to changes in alloy or surface condition).
A ratio detector is a single chip twin detector with different working wavelengths. The ratio of the two simultaneous
measurements is independent from emissivity and the result is the correct target temperature.
Every ratio thermometer has a limit as to how much signal can be lost. This is referred as the reduction ratio that can
vary up to 90% without effecting the measurement accuracy. To be also kept in consideration that the loss in signal
can come from 3 sources:
• low emissivity of the target
• object too small to fill the cone of vision
• obstruction caused by steam, smoke, dirt and dirty windows
In all the above condition a ratio thermometer will operate with better accuracy than any monochromatic
thermometer. The IRtec “Rayomatic” thermometers was designed to operate with spectral bands with limited
temperature coefficient.
3.2.6 Microcontroller
Microcontroller has to manage all thermometer functionality. It should be fast, with low absorption, with integrate
memory, serial interface, etc.
Eurotron’s IRtec Rayomatic use a RISC processor with internal re-writable FLASH memory.
24
Instruction Manual MM850341 ed.06f
NOTE: THE WORKING TEMPERATURE FOR THE GREEN LASER IS FROM +20°C TO +30°C
WARNING
USE A SEPARATE POWER SUPPLY OF 12VDC±2VDC 100M A ONLY WHEN USING THE GREEN LASER
25
Instruction Manual MM850341 ed.06f
4 SIGNAL PROCESSING
When a radiation thermometer is employed in an on-line system, its signal may be subjected to large fluctuations.
Time function facilities are useful to retain the signal levels that are more likely to represent the true target
temperature.
Eurotron IR thermometers can be factory pre-set with one of the signal processing facilities: averaging, Track & Hold,
Peak, Valley, Peak-Picking, Valley-Picking, etc.
IRtec Rayomatic 40/60/100 functions can be programmed from user by using a Personal Computer, a
Bell202/RS232 adapter and a configuration software.
4.1 Average
Rapid temperature fluctuations about a true mean value can make the thermometer output unsuitable for recording
and control. In these cases, the average function can be used to provide a smoothed signal. The output signal Avu is
given by:
where:
M = Actual temperature
Avu = Output average
Avp = Previous average
∆T = sampling time = 20ms
D = Average time (sec) (programmable)
Temperature
Actual temperature
(D=0)
Time
4.2 Peak
The Peak function holds a maximum temperature and retain its until a higher value appear or an external reset signal
appear to decay the output to the current temperature measurement.
4.3 Valley
The Valley function holds the minimum temperature and retain its until a lower value appear or an external reset
signal appear to increase the output to the current temperature measurement.
26
Instruction Manual MM850341 ed.06f
4.4 Peak-Picker
The peak-picker function holds a peak signal and allows it to decay slowly until the arrival of the next peak. It is used
in such applications as rolling mills where the signal is occasionally lowered by steam, smoke, water droplets, metal
scale, etc. When Peak-Picker function is programmed, the 4/20 mA output will follows the mathematical expression
below:
dove :
U = output signal
Peak = peak signal
PickDec = Decay rate °C/sec (programmable)
∆T = sampling time
Temperature
Decay rate (PickDec)
Programmable (0.0 to 1999.9 °C/S)
Time
reset
27
Instruction Manual MM850341 ed.06f
When the signal exceed the programmed trigger level (Threshold), the thermometer will wait the “StartPeak” delay
before starting with the standard Peak-Picker function. When the input signal will decrease below the trigger level
(Threshold), the thermometer will wait the “EndPeak” delay before reset the output signal.
The output signal will be reduced to low scale if Mode 1 is selected or to the last calculated value if Mode 2 is
selected.
Temperature
Mode 2
Mode 1
Time
4.5 Valley-Picker
Reverse the Peak-picker function principle to describe the Valley Picker.
where :
U = output signal
Valley = Valley signal
PickDec = Increasing rate °C/sec (programmable)
∆T = sampling time
28
Instruction Manual MM850341 ed.06f
Temp.
Time - tempo
Reset
Track & Hold
29
Instruction Manual MM850341 ed.06f
5 UNPACKING
Remove the instrument from its packing case and remove any shipping ties, clamps or packing materials.
Carefully follow any instruction given by any attached tags.
Inspect the instrument from scratches, dents, damages to case corners etc. which may have occurred during
shipment.
If any mechanical damage is noted, report the damage to the shipping carrier and then notify Eurotron directly or its
nearest agent, and save the damaged packaging for inspection.
A label, on the instrument, shows the serial number and model of the unit. Refer to these numbers for any inquiry for
service, spare parts supply or application and technical support requirements.
30
Instruction Manual MM850341 ed.06f
6 INSTALLATION
IRtec Rayomatic installing is very easy. The cylindrical aluminium chassis is externally threaded (M44x1.5) and
allows quick connecting with all available accessories.
The following procedure could help you for installing your IRtec Rayomatic 20/40/60 (monochromatic)
thermometer.
• Configure the thermometer. Set the emissivity on the IRtec Rayomatic 20 moving the 2 rotative switches.
Connect the IRtec Rayomatic 40 to the PC and set the mathematical function, the emissivity and the
temperature range by software. For Rayomatic 60 you can adjust the emissivity either by PC or by the
selector switches on the back panel.
• Connect, if necessary, all accessories (air purge, support, sighting tube, etc.). Fix the IRtec Rayomatic to the
process.
• Connect the signal cable from your data acquisition system to the instrument terminals. Pay attention to the
polarity.
• Aim accurately the thermometer to your target. Refers to Target vs. distance table on section 1. Be sure that
your target is bigger than the measuring area. The optional laser pinpointing can help you for best
installation.
The following procedure could help you for installing your IRtec Rayomatic 100 (2-Colors) thermometer.
• Install the thermometer on your plant.
• Supply the instrument and connect them to the PC. Use the Smart/RS232 converter.
• Run the configuring software.
• Select the right IRtec model
• Select the “Monocromatic” mode and start the PC temperature monitoring.
• Move the IRtec Rayomatic until the measured temperature correspond to maximum (the optical cone of
thermometer is centered to the target).
• Select the two-color mode on software.
• Adjust the “E-Slope” (non-gray) function until the measured temperature correspond to the real target
temperature.
ATTENTION
ARE TO BE AVOID BACKSCATTERING, SUNLIGHT OR OTHER LIGHTING FONTS. TO CHECK THIS CONDITIONS: POINT THE
INSTRUMENT, LOWER THE TEMPERATURE OF THE PROCESS UNDER 600°C: DISTURB RADIATIONS ARE PRESENT ONLY IF THE
INSTRUMENT READ OVER 600°C. POINTING ON SUNLIGHT YOU CAN READ ALSO 400°C BUT THIS IS FOR THE SPECTRUM
COMPONENTS OF PLANK EMISSION AT 600°C. A CLOSED MEASURING AREA WILL BE THE BEST SOLUTION.
IMPORTANT
DO NOT PULL OUT THE ELECTRONIC BOARD FROM THE CYLINDER. IT COULD BE DIFFICULT TO REASSEMBLE THE THERMOMETER.
WARNING
DO NOT INSTALLING THE THERMOMETER IN ROOM WITH TEMPERATURE HIGHER THAN +60°C WITHOUT AN ADEQUATE COOLING
SYSTEM.
WARNING
DO NOT CURVE THE FIBER OPTIC WITH RADIUS SMALLER THAN 200 MM !
SMALLER RADIUS CAN CAUSE A BREAKING OF THE FIBER.
WARNING
IRTEC RAYOMATIC 100 OPTICAL FIBER ASSEMBLY
REMOVE THE CENTRAL SCREW (SEE FIGURE BELOW) BEFORE TO REMOVE THE FIBER OPTIC FROM THE ELECTRONIC MODULE.
31
Instruction Manual MM850341 ed.06f
Remove
IMPORTANT
IRTEC RAYOMATIC 60 OPTICAL FIBER ASSEMBLY
IN ORDER TO ACHIEVE THE BEST RESULTS, BEFORE SCREWING THE OPTICAL FIBER TO THE MAIN INSTRUMENT, SWITCH ON THE
LASER PINPOINTING SYSTEM AND PROPERLY TURN THE FIBER WITH THE AIM OF OBTAINING THE MAXIMUM BEAM INTENSITY EXITING
FROM THE OPTICAL HEAD.
IMPORTANT
TO SHEILD THE CONNECTION CABLES, EXTEND THE SHIELD WIRE UNTILL THE DESTINATION POINT (DISPLAY, CONTROLLER…) AND
CONNECT IT TO GROUND.
The temperature signal is converted, amplified and linearized by the internal electronics that makes available a 4/20
mA linear output. Eurotron supply a connector for 4/20 mA active current loop connection.
The unit is supplied complete with a shielded multi-wires electrical cable.
The following procedure could help you for electrical cable substitution on your IRtec Rayomatic thermometer:
1. Remove the 3 screw on the flange on the rear of the instrument.
2. Open the cable gland and insert the shielded signal cable (φ 3.9 – 6.5 mm) through the flange.
3. Unsheathe the cable and prepare the wire for connection.
4. Unscrew partially the cable clamp. Insert correctly the wires. Pay attention to the polarity. Screw the cable clamp.
5. Put near the flange to the cylinder and screw firmly.
6. Close the cable gland.
7. Insert the 3 screw to the flange holes and accompain them manually. Use an adequate screw driver. Look up (1
or 2 rotation) alternatively the 3 screw.
100.0 -
IN 4/20mA Laser Power
Supply
+ Output
4/20mA
32
Instruction Manual MM850341 ed.06f
12-32 Vdc
Power Supply
IN 4/20mA
+ -
100.0
Laser Power
- Supply
-
100.0 IN 4/20mA
Laser Power
+ Output Supply
4/20mA
Indicator with internal power supply
Reset Alarm
12-32 Vdc
Power Supply
IN 4/20mA
+ -
100.0
Laser Power
- Supply
WARNING
DO NOT GO THROUGH THE SIGNAL CABLE NEAR ELECTROMAGNETIC SOURCE
(EG. ELECTRICAL MOTORS, HIGH POWER CABLE, WELDINGS, ETC). IF NECESSARY USE METAL CONDUIT.
-
100.0 I N 4/ 20mA
+ Output
4/20mA
Indicator with internal power supply
33
Instruction Manual MM850341 ed.06f
12-32 Vdc
Power Supply
IN 4/20mA
+ -
100.0
-
IRtec Rayomatic
-
+
DigiMax II
Connectors on the back side:
use the Rayomatic as a
transmitter.
In order to set the Digimax indicator, follow the procedure beginning from “Normal operation” and access to the
subsequent pages to modify the parameters:
34
Instruction Manual MM850341 ed.06f
The outoput of the pyrometer is a current signal, hence you need to insert the proper connector (a 2.5ohm shunt
resistor) to adapt the voltage input of the DigiMax. Anyway, you must power the termometer: for this aim, use the
following configuration where the pyrometer is considered as a transmitter and is powered by the Digimax itself.
• Connect the loop+ and loop- wires to clamps 6 and 9.
9
DigiMax III
DigiMax III Connectors on the back side:
notice the external shunt
5 6
1 2 resistance between 5 and 6
IRtec Rayomatic
-
220 Vac Rs
+
EXAMPLE: how to read continuously a temperature by means of Digimax III in a range of 0-600°C and without any
programming alarm. In order to set the Digimax indicator, follow the procedure beginning from “Operator mode” and
access to the subsequent pages to modify the parameters:
Remove the flange on the back of the instrument to access to the power supply terminals.
Connect the power supply (12…32 Vdc max. 100mA) to the instrument. Before switch the laser on verify the correct
polarity.
WARNING
USE A SEPARATE POWER SUPPLY OF 12VDC±2VDC 100M A ONLY WHEN USING THE GREEN LASER
35
Instruction Manual MM850341 ed.06f
IRtec Rayomatic 20
ATTENTION
IF THE HARDWARE RESET IS ENABLED FOR A MATHEMATICAL FUNCTION, IT IS USED FOR ALARM RESETTING TOO.
6.3 Alarms
IRtec Rayomatic 40 thermometers only, have two programmable alarms (Low and High) with hysteresis.
• System alarm: always enabled with output directed to the hardware alarm output. The thermometer activate the
output when the detector temperature is out of the operative range (e.g. When the water cooling system is
malfunctioning). It is factory set and you cannot disable it.
• Low alarm: you can set the minimum temperature of your process. When the temperature goes below the
programmed value, the alarm output became active.
36
Instruction Manual MM850341 ed.06f
• High alarm: you can set the maximum temperature of your process. When the temperature goes above the
programmed value, the alarm output became active.
• Hysteresis: you can program the hysteresis band for re-entry from the alarm status.
• Alarm Output: Two output modes are available. When alarms are programmed, the instrument uses the current
loop output to signal the alarm status; the high alarm status change the current loop to 21 mA, the low alarm
status change the output to 3.75 mA. You can also enable an hardware output line (20 mA sink current) to drive
an external relais.
• Alarm mode: if the external alarm output is selected, you can choose between normally high or normally low
output.
• Alarm reset: you can enable the reset input to reset (quit) the alarm status.
NOTE:
• All alarm set point are referred to the effective measured temperature.
• The alarm output and reset input terminals are available on model IRtec Rayomatic 40 only.
• Anyway, the alarm output is ever enabled to signal the system alarm event.
Temperature
Max setpoint
Hysteresis
Min setpoint
Time - tempo
Reset
Alarm output
37
Instruction Manual MM850341 ed.06f
4/20 mA
output +
1300 °C
24 Vdc
Power suppy
Relè
Reset
Alarm output
Alarm output and reset connections
Decimal Centesimal
Emissivity Adjustment
Es. 85 = 0.85
IRtec Rayomatic 20
IRtec Rayomatic 40 emissivity can be changed only by using a PC connected to the current loop.
IRtec Rayomatic 100 E-Slope can be changed only by using a PC connected to the current loop.
6.5 Positioning
The thermometer has to be installed in an accessible place for any further maintenance operation, and it hasn't to be
exposed to excessive heat, smoke and steam.
The optical path between the lens and the target, should be as much free as possible from smoke and steam.
The pointing axis should be placed with a 90° angle against the target surface; an angle smaller than 45° could be
acceptable for most materials.
NOTE: A LARGE NUMBER OF MATERIALS HAVE THE EMISSIVITY COEFFICIENT DEPENDENT ON THE VIEWING ANGLE. LARGE
ANGLES (MORE THAN 45°) CAN INDUCE A LARGE ERROR IN MEASUREMENT.
Before the installation, it iss useful to look at the tables supplied together with the instrument, in order to determine
the correct distance and visual field.
NOTE: USING THE MONOCHROMATIC THERMOMETERS THE TARGET AREA HAVE TO BE LARGER THAN THE CALCULATED
MEASURING AREA.
38
Instruction Manual MM850341 ed.06f
The Raryomatic IR beam profile is a cone described by the optical charts. The IR monochromatic thermometer must
be installed so that the target is bigger than circle described by optics and the optical path is free from obstruction
(steam, dust, etc.).
Install the mounting plate in a suitable place in the most favourable position from a thermal, and mechanical point of
view. Mount on the plate, if it is necessary, the cooling device and/or the purge air ejector. If a cooling jacket has to
be used, use the water outlet on the upper part of the support in order not to create air bubbles in the jacket.
WARNING
IT ADVISABLE TO USE A PROTECTIVE SCREEN DURING THE THERMOMETER ALIGNMENT OPERATIONS IF THE TARGET HAS A STRONG
RADIATION.
In order to align the support with the target, look through the support hole (without the thermometer) and find the best
position. It's important for the optic path to be free from any obstacle. In particular, when the surface is targeted
through a hole, the hole diameter has to be big enough according to the distance from the instrument. If optional laser
pinpointing system is installed, switch it on for spot centre viewing.
39
Instruction Manual MM850341 ed.06f
7 MAINTENANCE
Each IRtec Rayomatic is factory calibrated and certified against Eurotron Standards and shipped with a Report of
calibration stating the nominal and actual values and the deviation error. The instrument calibration has to be
periodically verified. IRtec uses a sophisticated analog and digital technology. All the maintenance operations must
be carried on by qualified personnel. Eurotron supplies instructions and operative modalities for any maintenance
operation. Please contact Eurotron for any support.
For a correct working of the instrument, the optic system must be kept clean and it mustn't reach temperatures higher
than the specified ones. The maintenance department should ensure these working conditions with a periodical
check of the cooling system and cleaning the lens.
IMPORTANT
REMOVE THE CENTRAL SCREW (SEE FIGURE BELOW) BEFORE TO REMOVE THE FIBER OPTIC FROM THE ELECTRONIC MODULE.
Remove
WARNING
DO NOT CURVE THE FIBER OPTIC WITH RADIUS SMALLER THAN 200 MM !
SMALLER RADIUS CAN CAUSE A BREAKING OF THE FIBER.
IMPORTANT
IRTEC RAYOMATIC 60 OPTICAL FIBER ASSEMBLY
IN ORDER TO ACHIEVE THE BEST RESULTS, BEFORE SCREWING THE OPTICAL FIBER TO THE MAIN INSTRUMENT, SWITCH ON THE
LASER PINPOINTING SYSTEM AND PROPERLY TURN THE FIBER WITH THE AIM OF OBTAINING THE MAXIMUM BEAM INTENSITY EXITING
FROM THE OPTICAL HEAD.
WARNING
DO NOT CURVE THE FIBER OPTIC WITH RADIUS SMALLER THAN 200 MM !
SMALLER RADIUS CAN CAUSE A BREAKING OF THE FIBER.
40
Instruction Manual MM850341 ed.06f
The purge air device is to be accurately checked, as the diffuser may become obstructed by non-clean air. When this
happens, the air flow from the diffuser is not uniform, and dust particles appear on the lens.
In this case, the diffuser should be drowned in a detergent solution and blown with compressed air, then dried. A
good air filtering can solve this problems.
7.8 Storage
Store the instrument in the original package, at temperature from –30°C to +70°C (from +10 to +70°C for thermopile
sensors) with non condensing R.H. less than 90%.
41
Instruction Manual MM850341 ed.06f
42
Instruction Manual MM850341 ed.06f
43
Instruction Manual MM850341 ed.06f
44
Instruction Manual MM850341 ed.06f
IMPORTANT
DO NOT USE TOO COOL WATER AS IT COULD CREATE CONDENSE ON THE SUPPORT OR ON THE LENS.
IT IS ADVISABLE TO HAVE A LOW WATER FLOW TO KEEP THE TEMPERATURE OF THE SUPPORT HIGHER THAN THE DEW-POINT OF A
HOT AND HUMID DAY.
THE FLOW SHOULD NOT BE HIGHER THAN 0.7 BAR AND THE AMBIENT TEMPERATURE SHOULD BE LOWER THAN 200°C.
The cooling water line has to be made using the following precautions:
The outlet of the cooling jacket must be kept free in order to control the flow and water temperature;
It's useful to use a thermometer to control the water temperature.
NOTE: IN CLOSE CIRCUIT SYSTEMS, ALWAYS USE A FLOW WATER WITH A LOW LEVEL ALARM.
Before the installation, it is useful to look at the tables supplied together with the instrument, in order to determine the
correct distance and visual field.
Spacer
Jacket end
Top flange
Rayomatic 20/40 Cooling jacket Purge air adapter
with or without lens
45
Instruction Manual MM850341 ed.06f
Fix the flange and the purge air adapter (eventually with a proper lens) with 4 screws:
Water cooling jacket mounting: flange and purge air adapter M44x1.5 (EE280329 or EE280315)
Insert the cable through the jacket end and fix it at the end of the system
46
Instruction Manual MM850341 ed.06f
8.1.4 Flanges
Code From ∅ To ∅ Fori Material
EE280113 1 ½ “ BSPP male ¾ “ BSPP female 4 x ∅7.2 mm AISI 304
EE280223 1 ½ “ BSPP male ¾ “ BSPP female 4 x ∅15.9 mm AISI 304
EE280292 1 ½ “ BSPP male ¼ “ BSPP female 4 x ∅6 mm AISI 304
EE280115 Through hole 1 ½” BSPP AISI 304
EE280282 Through hole 1 ½” BSPP 4 x ∅6 mm AISI 304
47
Instruction Manual MM850341 ed.06f
EE280113
22
10
Ø7.2
Ø127
Ø101.6
EE280223
22
10
Ø15.9
Ø127
Ø101.6
1" 1/2 BSPP
3/4" BSPP
48
Instruction Manual MM850341 ed.06f
EE280292
18
Ø6 6
Ø157
Ø101
49
Instruction Manual MM850341 ed.06f
50
Instruction Manual MM850341 ed.06f
51
Instruction Manual MM850341 ed.06f
EE290060
99
55
Ø32
Ø30
70
84
Compact Air purge code EE290060 complete of fast lock adapter and positioning plate (EE290063)
52
Instruction Manual MM850341 ed.06f
EE290046
88
Ø30
Ø69
Ø63
68
8.2.3 Flanges
Code From ∅ To ∅ Fori Material
EE280113 1 ½ “ BSPP ¾ “ BSPP 4 x ∅7.2 mm AISI 304
EE280223 1 ½ “ BSPP ¾ “ BSPP 4 x ∅15.9 mm AISI 304
EE280292 1 ½ “ BSPP ¼ “ BSPP 4 x ∅6 mm AISI 304
EE280115 Through hole 1 ½” BSPP AISI 304
EE280282 Through hole 1 ½” BSPP 4 x ∅6 mm AISI 304
EE280113
22
10
Ø7.2
Ø127
Ø101.6
1" 1/2 BSPP
3/4" BSPP
53
Instruction Manual MM850341 ed.06f
EE280115
6
Ø157
EE280292
18
Ø6 6
Ø157
Ø101
1" 1/4 BSPP
54
Instruction Manual MM850341 ed.06f
EE280282
Ø6 6
Ø157
Ø101
8.3 Advice
It's important to use very clean air to prevent dirt from depositing on the main lens. Every particle in the air could
obstruct part of the diffuser, and modify the aerodynamic characteristics in the air flux. So, the flux would be modified,
and dirt air could reach the lens. The air system must be periodically checked. To add a filtering system is strongly
recommended.
Eurotron can supply an air filtering system (cod. EE290015) to be used on the already present air network. This
system needs dry and clean air at a pressure up to 10 bar. If a compressed air network isn't available, a fan
generating system can be used.
According to the application, it could be necessary to cool the thermometer using an air flow. The maximum pressure
for an air purge system is 0.7 bar. Just for a normal air purge, only the tenth part of this flow is required.
55
Instruction Manual MM850341 ed.06f
Remember that the air tube must have a diameter bigger enough not to have flow losses.
IMPORTANT NOTE:
WHEN A PURGE AIR IS USED, IT MUST BE CLEAN AND DRY. THE DIFFUSOR PRODUCES A LAMINAR FLOW THAT KEEPS THE LENS
CLEAN. IF THE AIR ISN'T PERFECTLY CLEAN, TRACKS OF OIL AND DIRT MAY DEPOSIT ON THE LENS.
IMPORTANT NOTE:
WHEN MEASURING TEMPERATURES INSIDE OVENS, DO NOT USE ANY AIR COMING FROM THE FAN OF THE OVEN. THE AIR FROM THE
OVEN, IN FACT, MAY VARY, AND THE PRESSURE ON THE THERMOMETER MAY BE NOT SUFFICIENT TO GRANT ITS CLEANING AND
COOLING.
We suggest to minimise the distance between the controller group and the thermometer:
56
Instruction Manual MM850341 ed.06f
9 SERIAL COMMUNICATION
IRtec Rayomatic 40 is a programmable infrared thermometer using the same current loop wires to transmit and
receive digital informations from a Personal Computer. To do this, Eurotron has included inside the unit, a modem
with Bell202 communication protocol.
Two WindowsTM 98/2000/NT/XP compatible software packages are available to set the unit and for data acquisition.
Adapter Selector
RS232
connection
towards PC
Connections to
Rayomatic
9.1.1 Connections
Converter
SMART/RS232
57
Instruction Manual MM850341 ed.06f
Converter
SMART/RS232
_
Indicator
4-20 mA
_ + Internal resistance:
Ri = 220..270Ω
Power Supply
12-32 Vdc
Converter
SMART/RS232
_
R0
Indicator
4-20 mA
_ + Internal resistance:
Ri<220Ω
Power supply
12-32 Vdc
Use: 220Ω<R0<270Ω
Converter
SMART/RS232
_
R0
Indicator
4-20 mA
_ + Internal resistance:
Ri>270Ω
Power supply
12-32 Vdc
58
Instruction Manual MM850341 ed.06f
Converter
SMART/RS232
_
Indicator
4-20 mA
_ +
Power supply
12-32 Vdc
(12Vdc±2Vdc for
the green laser)
WARNING
USE A SEPARATE POWER SUPPLY OF 12VDC±2VDC 100M A ONLY WHEN USING THE GREEN LASER
59
Instruction Manual MM850341 ed.06f
10 CERTIFICATES
Direct all warranty and repair requests/inquiries to the Eurotron Customer Service Department. BEFORE
RETURNING ANY PRODUCT(S) TO EUROTRON, PURCHASER MUST OBTAIN AN AUTHORIZED RETURN (AR)
NUMBER FROM EUROTRON‘S CUSTOMER SERVICE DEPARTMENT (IN ORDER TO AVOID PROCESSING
DELAYS).
The purchaser is responsible for shipping charges, freight, insurance and proper packaging to prevent breakage in
transit.
60
Instruction Manual MM850341 ed.06f
APPENDIX
A1 EMC Conformity
The instrument is designed to fulfil the prevision of the directive 89/336/CEE Electromagnetic Compatibility.
In the following page you will find the EMC declaration of conformity
61
Instruction Manual MM850341 ed.06f
Declaration of Conformity
EN 55011
(Title, number and date of issue of normative documents)
eurotron
62
Instruction Manual MM850341 ed.06f
Declaration of Conformity
Cat. 1148
(Model)
EN 55011
(Title, number and date of issue of normative documents)
eurotron
63
Instruction Manual MM850341 ed.06f
The emissivity can be determined by one of the following methods, in order of preference:
1. Determine the actual temperature of the material using a sensor such as a RTD, thermocouple or another
suitable method. Next, measure the object temperature and adjust the emissivity setting until the correct value
is reached. This is the correct emissivity for the measured material.
2. For relatively low temperature objects (up to 260°C or 500°F, place a piece of tape, such as a masking, on the
object. Make sure the tape is large enough to cover the field of view. Next, measure the tape temperature
using an emissivity setting of 0.95. Finally, measure an adjacent area on the object and adjust the emissivity
setting until the same temperature is reached. This is the correct emissivity for the measured material.
3. If a portion of the surface of the object can be coated, use a flat black paint, which will have an emissivity of
about 0.98. Next, measure the painted area using an emissivity setting of 0.98. Finally, measure an adjacent
area on the object and adjust the emissivity setting until the same temperature is reached. This is the correct
emissivity for the measured material.
1. Temperature
2. Angle of measurement
3. Geometry (plane, concave, convex, etc.)
4. Thickness
5. Surface quality (polished, rough, oxidized, sandblasted)
6. Spectral region of measurement
7. Transmissivity (i.e., thin film plastics)
64
Instruction Manual MM850341 ed.06f
65
Instruction Manual MM850341 ed.06f
66
Instruction Manual MM850341 ed.06f
67