Laboratory Birla Institute of Scientific Research, 27, Malviya Industrial Area,
Jaipur, Rajasthan
Accreditation Standard ISO/IEC 17025: 2005
Certificate Number CC-2841 (In lieu of C-0118) Page 1 of 3
Validity 20.10.2018 to 19.10.2020 Last Amended on -
Quantity Measured / Range/Frequency *Calibration Measurement Remarks
Instrument Capability (±)
ELECTRO TECHNICAL CALIBRATION
I. SOURCE
$
1. DC Voltage 330 mV to 1000 V 0.012% Using Fluke Multi
Product Calibrator
5500A by Direct Method
$
2. DC Current 3.3. mA to 100 mA 0.07 % to 0.025% Using Fluke Multi
100 mA to 1.0 A 0.025% to 0.076% Product Calibrator
1A to 10 A 0.076% to 0.11% 5500A by Direct Method
$
3. AC Voltage 50Hz to1 kHz Using Fluke Multi
3.3. mA to 100 mA 0.22 % to 0.14 % Product Calibrator
100 mA to 10 A 0.14% to 0.13% 5500A by Direct Method
$
4. AC Current 50Hz to1 kHz Using Fluke Multi
3.3. mA to 100 mA 0.22 % to 0.14% Product Calibrator
100 mA to 10 A 0.14% to 0.13% 5500A by Direct Method
$
5. Frequency 10 Hz to 1 MHz 0.05 % to 0.007% Using Fluke Multi
Product Calibrator
5500A by Direct Method
0.001 Ω
$
6. Resistance 0.05 % Using Standard
(4 Wire) 0.01 Ω 0.05 % Resistance (Discrete
0.1 Ω 0.05 % Values) by VSI Method
1.0 Ω 0.05 %
10 Ω 0.05 %
7. DC Variable 10 Ω to 100 Ω 0.2 % to 0.027 Using Fluke Multi
100 Ω to 10 kΩ
$
Resistance 0.027 % to 0.017 Product Calibrator
(2 Wire) 10 kΩ to 100 kΩ 0.017 % to 0.02 % 5500A by Direct Method
100 kΩ to 1MΩ 0.02 % to 0.025 %
1 MΩ to 10 MΩ 0.025 % to 0.13 %
10 MΩ to 100 MΩ 0.13 % to 0.16 %
Dheeraj Chawla Srikanth R
Convenor Program Manager
Laboratory Birla Institute of Scientific Research, 27, Malviya Industrial Area,
Jaipur, Rajasthan
Accreditation Standard ISO/IEC 17025: 2005
Certificate Number CC-2841 (In lieu of C-0118) Page 2 of 3
Validity 20.10.2018 to 19.10.2020 Last Amended on -
Quantity Measured / Range/Frequency *Calibration Measurement Remarks
Instrument Capability (±)
$
8. High Resistance 2.0Ω 2.28 Using Variable
20.0Ω 2.28 Resistance by Direct
200.0Ω 2.28 Method
1000 MΩ 2.28
2000 MΩ 2.28
$
9. AC Power 50 Hz UPF 0.17 % Using Fluke Multi
40 to 600 V Product Calibrator
0.01 A/10 A 5500A by Direct Method
$
10. AC Power 240 V/0.1 A to 10A 0.55 % Using Fluke Multi
50 Hz PF=0.5 Product Calibrator
5500A by Direct Method
$
11. DC Power 10 to 600 V/ 100 mA to 0.16 % Using Fluke Multi
10 A Product Calibrator
5500A by Direct Method
II. MEASURE
$
1. DC Voltage 1 mV to 100 mV 0.5 % to 0.01 % Using 6½ Digit DMM
100 mV to 1 V 0.01 % to 0.005 % Agilent 34401 A by
1 V to 1000 V 0.005 % to 0.006 % Direct Method
$
2. DC Current 10 mA to 100 mA 0.08 % to 0.062 % Using 6½ Digit DMM
100 mA to 1 A 0.062 % to 0.13 % Agilent 34401 A by
1 A to 3 A 0.13 % to 0.17 % Direct Method
$
3. AC Voltage 50 Hz Using 6½ Digit DMM
10 mV to 100 mA 1.0 % to 0.55 % Agilent 34401 A by
100 mA to 1 V 0.55 % to 0.11 % Direct Method
1 V to 750 V 0.11 % to 0.005 %
$
4. AC Current 50 Hz Using 6½ Digit DMM
100 mV to 1A 0.56 % to 0.17 Agilent 34401 A by
1 A to 3 A 0.17% to 0.25 % Direct Method
Dheeraj Chawla Srikanth R
Convenor Program Manager
Laboratory Birla Institute of Scientific Research, 27, Malviya Industrial Area,
Jaipur, Rajasthan
Accreditation Standard ISO/IEC 17025: 2005
Certificate Number CC-2841 (In lieu of C-0118) Page 3 of 3
Validity 20.10.2018 to 19.10.2020 Last Amended on -
Quantity Measured / Range/Frequency *Calibration Measurement Remarks
Instrument Capability (±)
$
5. Frequency 10 Hz to 100 Hz 0.1 % to 0.057 % Using 6½ Digit DMM
100 Hz to 100 kHz 0.057 % to 0.012% Agilent 34401 A by
Direct Method
10Ω to 100 Ω
$
6. DC Resistance 0.05 % to 0.016% Using 6½ Digit DMM
(2 Wire) 100 Ω to 1 kΩ 0.016 % to 0.017 % Agilent 34401 A by
1 kΩ to 10 MΩ 0.017 % to 0.05 % Direct Method
10 MΩ to 100 MΩ 0.05 % to 0.93 %
$
7. CT Using CT Bridge and
Ratio Error 5-10-25-50 0.07 % Standard CT by
Phase Error 100A/5 A 3.7 min Comparison Method
* Measurement Capability is expressed as an uncertainty () at a confidence probability of 95.45%
$
Only in Permanent Laboratory
Dheeraj Chawla Srikanth R
Convenor Program Manager