The document provides service information for the NGL XCF 3000 Blood Components Separator, including descriptions of its components, principles of operation, and troubleshooting guidelines. It details each major system such as the computer board, display, centrifuge driver, pumps, valves and more. The document also includes factory and service settings for parameters of the centrifuge driver.

Uploaded by

Suarez O.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as DOC, PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
319 views29 pages

Service Manual-NGL XCF 3000

The document provides service information for the NGL XCF 3000 Blood Components Separator, including descriptions of its components, principles of operation, and troubleshooting guidelines. It details each major system such as the computer board, display, centrifuge driver, pumps, valves and more. The document also includes factory and service settings for parameters of the centrifuge driver.

Uploaded by

Suarez O.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as DOC, PDF, TXT or read online on Scribd
You are on page 1/ 29

BLOOD COMPONENTS SEPARATOR

NGL XCF 3000

Service Manual

Sichuan Nigale Biomedical Co., Ltd.

Chengdu China
Sichuan Nigale Biomedical Co.,Ltd. Blood Component Separator Service Manual

Table of Contents

1.0 INTRODUCTION………………….…………………………………………….…………….…...
…2
1.1 Warranty……………………………………………………………………..
……………………….…2
1.2 Service Information……………………………………………………………………………….……2
2.0 PRINCIPLES OF OPERATION ……………….………………………………………………...….3
2.1 Components Description…………………………………………………………………………...…..3
2.2 Power Entry………………………………………………………………………………..……….…. 5
2.3 Power Supply 3 &Computer Power Supply…………………………………...…………………….…5
2.4 Computer Board……………………………………………………………………………….…….….6
2.5 Keyboard Board……………………………………………………...………………………… .….….7
2.6 Display………………………………………………………………………………………….………8
2.7 Centrifuge Power Supply………………………………………………………………………...…..…9
2.8 Centrifuge Driver………………………………………………………………………………....….…9
2.9 Centrifuge………………………………………………………………………………….………….14
2.10 DC Power Supply Board and Power Supply 1…………………………………………………….….15
2.11 Pump Controlled Board……………………………………………………………………….………16
2.12 Blood Pump & Anticoagulant Pump Driver…...………………………………………….……...…..17
2.13 Blood Pump & Anticoagulant Pump……………………………………………………..…….….….19
2.14 Valve Controlled Board………………………………………………………………………….……20
2.15 Valve…………………………………………………………………………………………………..21
2.16 Pressure Monitor (DPM&SPM)……………………………………………………………..………..21
2.17 Weigher……………………………………………………………………….…………….…………22
2.18 Cuff Board………………………………………………………………………………………….…22
2.19 Air Pump…………………………………………………………………………….…………….…..23
2.20 Air Detector………………………………………………………………………………………..….23
2.21 Line Sensor………………………………………………………………………………………..…..24
2.22 Bowl Optical Detector…………………………………………………………………….…..……....25
2.22 Cooling Fan……………………………………………………………………………………..….....25
3.0 TROUBLESHOOTING ……………………………………………………..…………………….....26
3.1 Preliminary Action…………………………………………………………………………….……...26
3.2 Weigher Channel………………………………………………………………………………….…..26
3.3 System control & Display Channel…………………………………………………………………...26
3.4 Air Detector Channel……………………………………………………………………………….…26
3.5 Pump Drive Channel………………………………………………………………………….…..…..26
3.6 Centrifuge Drive Channel………………………………………………………………………..…...27
3.7 Cuff Working Channel……………………………………………………………………………...…28
3.8 Line Sensor Monitor Channel………………………………………………………….………..….…28
3.9 Valve Working Channel…………………………………………………………………….…..……..28
3.10 DPM & SPM Channel……………………………………………………………………………..….28
3.11 Bowl Optical Detector Channel…………………………………….……………………….……...…28
4.0 SYSTEM BLOCK DIAGRAM

1
Sichuan Nigale Biomedical Co.,Ltd. Blood Component Separator Service Manual

1.0 INTRODUCTION
This service manual contains information required to calibrate, troubleshooting, and maintain the XCF3000
Blood Component Separator.

1.1 WARRANTY
The warranty extends only to the original purchase and is not assignable or transferable, and shall not apply
to auxiliary equipment, disposable accessories.

1.2 SERVICE INFORMATION


While under Nigale Warranty, the machine must not be opened by unauthorized personnel.
All repairs and adjustments must do by qualified personnel only, and must follow procedures given in this
Service Manual.
Nigale will not be responsible for unauthorized repairs and adjustments.

2.0 PRINCIPLES OF OPERATION

2
Sichuan Nigale Biomedical Co.,Ltd. Blood Component Separator Service Manual

2.1 Components Description (front view)

Components Description (back view)

3
Sichuan Nigale Biomedical Co.,Ltd. Blood Component Separator Service Manual

2.2 Power Entry

4
Sichuan Nigale Biomedical Co.,Ltd. Blood Component Separator Service Manual

The power entry is located on the left side of the machine.


The Power Socket is a filter type socket, Fuse type is 4A,Φ5×20.
2.3 Power Supply 3 & Computer Power Supply

The Power Supply 3 is an isolated transformer, which decrease the disturbance between computer board and
outer, Computer Power Supply is a tri-output power supply, it’s input voltage is 220VAC from power supply
3 , output +5V DC,±12VDC. It provides all power, which Computer Board needed.

2.4 Computer Board

5
Sichuan Nigale Biomedical Co.,Ltd. Blood Component Separator Service Manual

The computer Board contains the microcomputer that controls all equipment operations. The program has
been programmed in the microcomputer and extended in the ROM, according the program, the computer
outputs various control signals, monitors the keyboard, and monitors instrument status signals to effect. The
computer board also receives all switch and transducer signals, and commands all indicators and components
loads in the instrument.

6
Sichuan Nigale Biomedical Co.,Ltd. Blood Component Separator Service Manual

The following describe main components of Computer Board.


2.4.1 3D1 N87C196KC20: The CPU is a high performance microcontroller single chip, the CPU-chip on the
board is 16 bits wide and connects to the interrupt controller and the memory controller by a 16_bit bus. In
addition, there is an 8-bits bus, which transfers instruction bytes from the memory controller to the CPU. An
extension of the 16-bit bus connects the CPU to the peripheral devices. The CPU also has high-speed input
& output ports, 10 bits precise A/D converter, serial port, timers etc.
2.4.1 3D4 BS64-70: contain memory area for temporarily storing operational information (parameters, etc),
and for storing various software variables. To read data from or write data into RAM.
2.4.2 3D5 DS12C887: Real Time Clock, Multiplexed Bi-directional Address/Data Bus, nonvolatile with over
10 years of operation in the absence of power, self-contained subsystem includes lithium, quartz, and support
circuitry, counts seconds, minutes, hours, days, day of the week, month, and year.
2.4.3 3D3 27C512 EPROM (erase programmable read-only memory): which contains 64K×8 bits. The
program has been embedded.
2.4.4 3 D14 3D15 82C55: Programmable peripheral interface, this chip is a high-performance, CHMOS
version of industry standard 8255A general-purpose programmable I/O device. The functional configuration
of this chip is programmed by the system software so the normally no external logic is necessary to interface
peripheral devices or structures.
2.4.5 Base voltage: GND VERF, 5V-0.005V; 3RP1 is adjustable.
Interrupt voltage: GND PFI 1.5V±0.05V, 3RP2 is adjustable.

2.5 Keyboard Board

7
Sichuan Nigale Biomedical Co.,Ltd. Blood Component Separator Service Manual

2.5.1 2D1 8279: Programmable keyboard and display I/O interface device, it communicate with computer to
drive display and send keyboard control signal to computer.
2.5.2 2D2 ULN2003A is an Darlington Transistor Array to drive lights in the Keyboard.
2.5.3 The potentiometer 2RP1 is used to adjust brightness of LCD. The voltage of V0-and V0+ should be
12.6V±0.05V
2.6 Display

This is a LCD device, no part needed adjust or repair.


2.7 Centrifuge Power Supply

8
Sichuan Nigale Biomedical Co.,Ltd. Blood Component Separator Service Manual

This is an isolation transformer, the purpose of this device is to decrease distinctly disturbance between
outer, at same time decrease centrifugal motor to disturb other devices.
2.8 Centrifuge Driver

9
Sichuan Nigale Biomedical Co.,Ltd. Blood Component Separator Service Manual

2.8.1The centrifuge driver is a AC motor driver which incorporating the microprocessor technology, the
220VAC power is applied to the driver from Power Supply 2. It receives the control signal from computer
board through the channel of Pump Controlled Board. If the turning is not clockwise, exchange random two
lines of U/T1、V/T2、W/T3。
2.8.2 All parameters have been programmed in the factory before you receive this machine.

10
Sichuan Nigale Biomedical Co.,Ltd. Blood Component Separator Service Manual

In the following form, the Factory Setting is a default, Nigale Setting means the different from the default. If
the parameters are mixed up, Please reference 0-02,Reset all parameters to Factory Setting ,then set
parameters to Nigale Setting.

Group 0: User Parameters.

Factory Nigale
Pr. Functions Settings
Setting Setting

0-00 Identity code of 1:40W 4:400W


drive (Read only) 2:100W 5:750W
3: 200W 6:1.5KW

0-01 Rated current 40W: 0.4A 400W: 2.5A


display (Read only) 100W: 0.8A 750W:4 .2A
200W: 1.6A 1.5KW:7.0A

0-02 Parameter reset 10: Reset Parameters to Factory


0
Setting
0:F(Frequency command)
0-03 Start-up display of
AC drive 1:H(output frequency)

0
2:U(user-defined unit)

3: A (output current)
0: Display User-Defined Unit (u)
0-04 User-defined Unit
1: Display Counter Value (C)

2: Display Process Operation (1=tt)


0

3: Display DC-BUS voltage (U)

4: Display output voltage (E)

0-05 User-defined 0.1~160


1.0
coefficient K

0-06 Software version Read only #,#

0-07 Password input 0 ~999 0

11
Sichuan Nigale Biomedical Co.,Ltd. Blood Component Separator Service Manual

0-08
Password 0 ~999 0
configuration

Group 1:Basic Parameters

1-00 Maximum operation Freq. 50.0 ~400Hz 60.0 82

1-01 Maximum setting Freq. 10.0 ~400Hz 60.0 82

1-02 Maximum output voltage 2.0 ~255V 220 95

1-03 Mid-point freq. 1.0 ~400Hz 1.0

1-04 Mid-point voltage 2.0 ~255V 12.0

1-05 Minimum output freq. 1.0 ~60.0Hz 1.0

1-06 Minimum output voltage 2.0 ~255V 12.0

1-07 Upper bound of freq. 1~110% 100

1-08 Lower bound of freq. 0~100% 0.0

1-09 Accel time 1 (Tacc1) 0.1~600Sec 10.0 30

1-10 Decel time 1 (Tdec1) 0.1~600Sec 10.0 40

1-11 Accel time 2 0.1~600Sec 10.0

1-12 Decel time 2 0.1~600Sec 10.0

1-13 JOG Accel time 0.1~600Sec 10.0

12
Sichuan Nigale Biomedical Co.,Ltd. Blood Component Separator Service Manual

1-14 JOG Decel time 0.0~600Sec 10.0

1-15 JOG frequency 1.0Hz~400Hz 6.0

1-16 Auto-accel/decel 0:Linear Accel/Decel

1:Auto accel, linear decel

2:Linear accel,auto decel

3:Auto accel/Decel

4:Linear accel.Auto decel,stall

prevention during deceleration

5:Auto accel. Auto decel,stall

prevention during deceleration

1-17 S-curve setting in 0~7


0
acceleration

1-18 S-curve setting in 0~7


0
deceleration

Group 2: Operation Method Parameters

Factory Nigale
Pr. Functions Settings
Setting Setting

2-00 Source of frequency 0: Digital keypad 0

1: 0 ~10V from AVI

2: 4~20mA from AVI

3: Controlled by V.R on drive

4: RS-485 communication

interface

2-01 Source of operation Command 0:By digital keypad 0 2

13
Sichuan Nigale Biomedical Co.,Ltd. Blood Component Separator Service Manual

1:By external terminals, keypad

STOP enable

2: By external terminals,

keypad STOP disable

3: By RS-485 communication

interface, keypad STOP enable

4: By RS-485 communication

interface, keypad STOP

disable

2-02 Stop method 0: Ramp stop 1: Coast stop 0 0

2-03 Carrier freq. 3 ~10K Hz 10

2-04 Reverse operation inhibit 0: Enable reverse 0

1: Disable reverse

2: Disable forward

2-05 ACI (4 ~ 20mA) input loss 0: Decel to 0Hz 0

detection 1: Stop immediately, display EF

2: Run with the last freq.

2-06 Line Start Lockout 0:Enable 1:Disable 0

2.9 Centrifuge

14
Sichuan Nigale Biomedical Co.,Ltd. Blood Component Separator Service Manual

The centrifuge consists of an AC motor, tachometer, cover switch, fluid sensor, centrifuge turntable which
gets six claws to lock the centrifuge bowl, when the centrifuge is running..

15
Sichuan Nigale Biomedical Co.,Ltd. Blood Component Separator Service Manual

2.9.1 Motor is an 80W AC motor, running at 4800 rpm clockwise.


2.9.2 Tachometer is a Hall effect magnetic devices. The magnetic encoder has a solid rotating disk rather
than the optical encoder.
2.9.3 Cover Switch: two micro-switches in series to represent the status of the cover position.
2.9.4 Fluid Sensor: there are two sensors, it is a humidity sensitive device, when there is no fluid on the
surface of the sensor, the value of resistance is very high; when there is fluid on the surface of the sensor, the
value of resistance is very low. Thus, the computer through this device to detect the fluid flow.
2.10 DC Power Supply Board and Power Supply 1

The DC Power Supply Board provides the power to valve controlled board, pump drivers, cooling fans.
2.10.1 The 0V1 0V2 0V3 0V4 and 0V5 are Silicon bridge rectifiers, the 0IC1 0IC2 and 0IC3 are 3-terminal
voltage regulators, the 0F21 0F2 0F3 0F4 and 0F5 are self-recoverable fuse, this board also including line
regulator which consists of 0V10 0V11 and 0V17 etc.
2.10.2 0×1 connects the Power Supply 1 which input 12VAC 9VAC 24VAC 24VAC, the terminal output
0×2 is connected valve controlled board, 0×3 is connected cooling fan, 0×4 is connected pump controlled
board, to provide power they needed.
2.10.3 Output Voltage test:
0×2 pin1 pin2 24V±0.2V, potentiometer 0RP1 adjust if necessary
pin3 pin4 5V±0.5V
0×3 pin1 pin2 24V±0.5V
0×4 pin1 pin2 12V±0.5V
pin3 pin4 31V±5V
pin5 pin6 31V±5V

16
Sichuan Nigale Biomedical Co.,Ltd. Blood Component Separator Service Manual

2.10.3 Power Supply 1 is a round transformer which be applied 220VAC, and output AC to the DC Power
Supply Board.
2.11 Pump Controlled Board

2.11.1 Safety Control Circuit: consist of 11V7 11V8 11V5 11R24 11R25 and a relay 11K1, as the relay
energizes or not, the power of pump driver connected or disconnected to power according to the control
signal from the computer board.
2.11.2 Centrifuge control signal channel
2.11.3 Blood pump & AC pump control signal channel
2.11.4 Centrifuge speed signal feeding channel

17
Sichuan Nigale Biomedical Co.,Ltd. Blood Component Separator Service Manual

2.11.5 Blood pump & AC pump feeding channel


2.11.6 Blood pump driver & AC pump driver
2.12 Blood Pump & Anticoagulant Pump Driver

This device is a stepping motor driver with over current, over voltage and short circuit protection. The blood
pump and AC pump use same type of driver, but different resolution setting.
2.12.1 Specifications of the driver
Pin Description

Pin
Signal Description
No.

Pulse signal: this input represents pulse signal, effective for


1 PUL
each upward rising edge

2 DIR Direction signal: high/low voltage levels signal

P1 3 COM Opto-Isolated Power

Enable signal: this signal is used for enable/disable, high level


4 ENA for enabling driver and low level for disabling deriver. No
connection when staying

5 RST Reset signal;

6 NC
P2 1、2 DC-/ DC power supply
DC+

18
Sichuan Nigale Biomedical Co.,Ltd. Blood Component Separator Service Manual

3、4 A+/A- Motor coil A (leads A+ and A-)

5、6 B+/B- Motor coil B (leads B+ and B-)

CURRENT OUTPUT :
VALUE SW1 SW2 SW3
1.3A on on on
1.6A off on on
1.9A on off on
2.2A off off on
2.5A on on off
2.9A off on off
3.2A on off off
3.5A off off off

RESOLUTION :

Resolution SW5 SW6 SW7 SW8

2 on on on on

4 on off on on

8 on on off on

16 on off off on

32 on on on off

64 on off on off

128 on on off off

256 on off off off

5 off on on on

10 off off on on

25 off on off on

50 off off off on

19
Sichuan Nigale Biomedical Co.,Ltd. Blood Component Separator Service Manual

125 off on on off

250 off off on off


2.12.2 Setting:
2.12.2.1 Resolution of Blood Pump Driver

On 1 2 3 4 5 6 7 8
2.12.2.2 Resolution of AC Pump Driver

On 1 2 3 4 5 6 7 8
2.13 Blood Pump & Anticoagulant Pump

2.13.1 Blood Pump & AC Pump is same structure; it equipped stepping motor that is controlled by pump
driver. To operate the pump, a control signal is received from CPU board to pump driver, pump driver
control voltage applied to the motor, causing the pump to rotate.
2.13.2 Tachometer is located in the middle of the pump. The tachometer is optical pickup devices that send

20
Sichuan Nigale Biomedical Co.,Ltd. Blood Component Separator Service Manual

pulses to circuits on the pump controlled board for processing. The frequency of these pulses represents the
rotating speed of the corresponding pump assembly.
2.14 Valve Controlled Board

2.14.1 Valve control channel: the board has four electric magnetic valves control circuit.
2.14.2 The board including buzzer drive circuit, consist of 6VT23 etc.
2.14.3 Air detector signal channel: blood line air detector and anticoagulant line air detector be connected to
6×10, 6×9 respectively.
2.14.4 Centrifuge Cover Switch & Fluid Sensor signal channel: connector 6×8
2.14.5 Donor Pressure Monitor signal channel: connector 6×11
2.14.6 Power Indicator 6VD26: 24VDC is applied to this board , the light is on.

2.15 Valve

21
Sichuan Nigale Biomedical Co.,Ltd. Blood Component Separator Service Manual

2.15.1 The valve is a electric magnetic valve, the central pole moved down when 24VDC is applied to the
winding, thus open the valve.
2.15.2 The position tester: at the bottom there is a photoelectric cell to detect the central pole position, and
send the signal to computer board.
2.16 Pressure Monitor (DPM & SPM )

This is a pressure component, DPM & SPM are same device, DPM connected to valve controlled board 6×4.
SPM Connected to Computer Board 3×11.

22
Sichuan Nigale Biomedical Co.,Ltd. Blood Component Separator Service Manual

The input is +12VDC, and the output range is 0~5VDC according the pressure changed in the pressure
monitor filter line.
2.17 Weigher

2.17.1 Voltage Converted Circuit: consist of 7N2 7RP2 7R3 7C4 7C3, +12VDC is converted to +10VDC,
7RP2 is a potentiometer for adjust base voltage..
2.17.2 Amplifier Circuit: consist of 7N1 7RP1 7R1 etc, 7N1( AD620) is a instrumentation amplifier, 7RP1 is
a potentiometer which change the gain, use for calibration..
2.17.3 Off-set Adjust: 7R4 7RP3 be used to adjust zero float.
2.17.4 Weight Sensor: it is a weight transducer bridge, which output voltage, when weight change.
2.17.5 Software Calibration: according to the Operating Manual.
2.18 Cuff Board

2.18.1 Cuff pressure measure: 8RP1 8R6 8N1 8R5 made a regulating circuit, output 5VDC to pressure
transducer.
2.18.2 Staleness Valve Drive circuit: 8V2 8R3 8R4 etc.
2.18.3 Air Pump Drive circuit: 8V1 8R1 8R2, Connector 8×3 is linked with air pump..

23
Sichuan Nigale Biomedical Co.,Ltd. Blood Component Separator Service Manual

2.18.4 Base Voltage Calibration: measure the 5V test-point, 5V±0.05V, if necessary, adjust 8RP1
potentiometer.
2.18.5 Cooling Fans is connected of cuff board via 8×5 8×4 8×3 8×2.
2.19 Air Pump

The air pump produce the pressure of cuff, the working voltage is 5VDC. During draw , 5VDC is applied to
the air pump, to provide the air pressure; during return , the pump receives 0VDC, not running. The air pump
power via connector 8A1 of Cuff Board.
2.20 Air detector

2.20.1 The Air Detector is a ultrasonic component. There are four Air Detectors in this machine to monitor
the liquid in the line, it is same physical structure and same circuit, so it can exchange freely. To do different
task, it gets different name, Blood Line Air Detector, Donor Line Air Detector 1, Donor Line Air Detector 2,
and Anticoagulant Line Air Detector.
This detector contains circuits that transmit an ultrasonic signal through the line. As long as the fluid level in
the line is at a level high enough to remain in the ultrasonic frequency path, the fluid acts as a good medium
and enables the signal to be received by receiver transducer in the detector, the receiving transducer outputs
the received signal to the channel of computer.
When the light on, liquid is in the line; when the light off, air is found in the line.
The input working voltage is ±12VDC, and the output signal high-level (light on) means liquid in the line,
low-level(light off) means air is found in the line.
2.21 Line Sensor

24
Sichuan Nigale Biomedical Co.,Ltd. Blood Component Separator Service Manual

2.21.1 It is a liquid density detector, consists of a light transmitter, receiver, and reflector. Light is pass
through and across the line to the receiver. According the changes of the light, the receiver output a
corresponding voltage to computer board.
2.21.2 Power on, at main menu press ‘ENTER’, then press ‘5’, a parameter is displayed.
XXXXNNNN
‘NNNN’ is represent of line sensor A/D convert value, when there is no line in the sensor, the range is 700~
750, the best value is near 700; where there is an empty line in the sensor, the range is 100~550.
2.21.3 As the illustration, a potentiometer is for calibration if need.
2.21 Bowl Optical Detector

2.21.1This device transmit a laser to the bowl and receive the reflection, according to the receiving different
voltage, the computer send information about separation status on display.
2.21.2 Power on,at main menu press ‘ENTER’, then press ‘5’, a parameter is displayed.
NNNNXXXX
‘NNNN’ is represent of bowl optical voltage A/D value, when there is no bowl in the centrifuge, the range is
50~200; where there is an empty bowl in the centrifuge, the range is 700~900, best near 700.
2.21.3 Water Test value adjustment: loose four screws in the bowl optical, up will decrease the value, down
will increase the value. Normal, the range is from 45 to 50, but best value is 47.
2.22 Cooling Fan

Cooling fan is a DC fan that cools the electronics, the fan receives 24VDC power via connector 0×3 in the

25
Sichuan Nigale Biomedical Co.,Ltd. Blood Component Separator Service Manual

DC Power Supply, the fan cools the electronics in the console of the machine.
3.0 TROUBLESHOOTING
Troubleshooting the NGL XCF 3000 consists of using standard troubleshooting techniques such as board or
component replacement and signal tracing to correct an instrument malfunction. Before troubleshooting the
instrument, the service engineer should become thoroughly familiar with Chapter2, Principles Of Operation,
in order to understand microcomputer control of instrument operation.
WARNING: DANGEROUS VOLTAGE ARE PRESENT WITHIN THE INSTRUMENT. EXERCISE
EXTREME CAUTION IF POWER IS APPLIED WHEN TROUBLESHOOTING WITH REAR
PANEL OR FRONT PANEL REMOVED.
3.1 Preliminary Action
The first step in troubleshooting the instrument is obtaining from the instrument user. With the complexity of
the instrument, no failure analysis should commence until it is first verified that the instrument will not
perform according to the Operating Instruction.
3.2 Weigher Channel
Power on, at the main menu, press ‘Enter’ key, then press ‘8’, gets the procedure of weigher calibration
procedure.
3.2.1 Tolerance over range: adjust the potentiometer 7RP1.
3.2.2 “000” or “999” in the display, weigher may be permanently damaged, it must be changed.
3.3.3 Channel failure, check computer board, try to exchange another computer board.
3.3 System control & Display Channel
3.3.1 No character in the display, check the cable from keyboard to computer.
3.3.2 Check computer working voltage supply.
3.3.3 Try to exchange another computer board. Change it if needed.
3.3.4 Auto quit: program interfered by outer, check ground device or power system. Another possibility is
interfere from the centrifuge motor, if persist, replace centrifuge.
3.4 Air Detectors Channel
3.4.1 High sensitivity: no fluid, just line inserted, the light is on. Action: replace it.
3.4.2 Low sensitivity: abnormally select special position, then the light is on. Action: replace it.
3.4.3 Always light on or always light off: the device is damaged. Action: replace it.
3.5 Pumps Drive Channel
Including blood pump and AC pump. Enter the test procedure, at the main menu, press ‘ENTER’ and ‘1’,
then press ‘5’, ‘9’, and press ‘start/stop’ key, blood pump and AC pump are running, the pump speed control
keys can increase or decrease the speed, and press ‘start/stop’ key again, will pause the running of two
pumps, illustration:
Time count
fse1 fse2 rate
blood pump set speed XX AA
blood pump actual speed AC pump actual speed YY
counts of blood pump counts of AC pump speed of centrifuge
Note:The AC pump set speed is equal to XX/100;the blood pump set speed is equal to AA;/AC pump
set speed/the actual rate is equal to YY.
Press “ESC” to quit the procedure.

26
Sichuan Nigale Biomedical Co.,Ltd. Blood Component Separator Service Manual

3.5.1 Not running: check the DC power supply, pump driver, if failure, replace it.
3.5.1 Over speed: pump driver failure, replace it.
3.5.3 Low speed: clear tachometer, if persist, change tachometer.
3.5.4 No speed: the pump is running, but ‘No speed’ alarm, replace the tachometer.
3.5.5 Noise: clear pump head or replace with new one.
3.6 Centrifuge Drive Channel
3.6.1 Centrifuge Driver fault description and corrective action:

Fault
Fault Name Corrective Actions
Descriptions
1. Check whether the motors horsepower
oc The AC drive corresponds to the AC drive output power.
detects an
abnormal increase 2. Check the wiring connections between the AC
in current. drive and motor for possible short circuits.

3. Increase the Acceleration time (Pr.1-09,


Pr.1-11).

4. Check for possible excessive loading


conditions at the motor.

5. If there are any abnormal conditions when


operating the AC drive after the short-
circuit is removed, the drive should be sent
back to manufacture.
1. Check whether the input voltage falls within
ou The AC drive the rated AC drive’s input voltage.
detects that the
DC bus voltage 2. Check for possible voltage transients.
has exceeded its
maximum allowable 3. Bus over-voltage may also be caused by motor
value. regeneration. Increase the decel time.
1. Ensure that the ambient temperature falls
oH The AC drive within the specified temperature range.
temperature
sensor detects 2. Make sure that the ventilation holes are not
excessive heat. obstructed.

3. Remove any foreign objects on the heat sink


and check for possible dirty heat-sink fins.

4. Provide enough spacing for adequate


ventilation.

27
Sichuan Nigale Biomedical Co.,Ltd. Blood Component Separator Service Manual

3.6.2 Not running: check centrifuge driver, watch the alarm code, take an action; if motor is error, and
change it.
3.6.1 Over speed: check centrifuge driver, watch the alarm code, and take an action;
3.6.3 Lowe speed: check centrifuge driver, watch the alarm code, take an action; if motor is error, and
change it.
3.6.4 Centrifuge No Speed: Centrifuge is running, but the alarm is on. Enter the test procedure, at the main
menu, press ‘ENTER’ and ‘1’ key, then press ‘5’, ‘9’, then press ‘MODIFY, to start centrifuge, if the
centrifuge is running, loot at the value of speed of centrifuge, if the value=0, the Hall effect magnetic
devices is failure, replace it, illustration:
Time count
fse1 fse2 rate
blood pump set speed XX AA
blood pump actual speed AC pump actual speed YY
counts of blood pump counts of AC pump speed of centrifuge
Note:The AC pump set speed is equal to XX/100;the blood pump set speed is equal to AA;/AC
pump set speed/the actual rate is equal to YY.
Press “ESC” to quit the procedure.
3.6.5 Big Noise: centrifuge motor is failure, replace it with new one.
3.7 Cuff Working Channel
3.7.1 No air pressure: the air pump is failure, change it. If persist, check cuff board, cable to computer
board. Replace it if need.
3.7.2 No deflate: the Staleness Valve is failure, change the cuff board or deflate valve.
3.8 Line Sensor Monitor Channel
3.8.1 Over range: adjust potentiometer in it, if necessary, if failure, replace it.
3.8.2 Light off: replace it..
3.9 Valve Working Channel
3.9.1 Valve position error: look at the pole move or not, check the position tester, replace it if needed.
3,9.2 Valve close is not tight: means liquid is still flow slowly when the valve is closed, replace it.
3.9.3 Valve pole not move: check valve itself, then look at valve-controlled board, cable etc. replace failure
parts.
3.10 DPM&SPM Channel
3.10.1 Self test failure when no pressure applied to the transducer, replace it.
3.10.2 Position Error: if the DPM filter is inserted into the adapter of DPM, but the alarm is still on, replace
it.
3.11 Bowl Optical Detector Channel
3.11.1 light off: replace it.
3.11.2 Over range: adjust potentiometer. If failure, replace it.

28