This is to attest that
UNIVERSAL INSPECTION CO.LTD.
BLDG.NO.176, BLOCK NO.248, AZAIBA NORTH
MUSCAT 130, SULTANATE OF OMAN
Calibration Laboratory CL-221
has met the requirements of AC204, IAS Accreditation Criteria for Calibration Laboratories, and has demonstrated
compliance with ISO/IEC Standard 17025:2017, General requirements for the competence of testing and calibration
laboratories. This organization is accredited to provide the services specified in the scope of accreditation.
Effective Date February 10, 2023
Expiration Date May 1, 2024
President
UNIVERSAL TESTING CO.LTD.
www.ui.com.sa
Contact Name Dinesh Kumar Kesavan Contact Phone +966-508836773
Accredited to ISO/IEC 17025:2017 Effective Date February 10, 2023
CALIBRATION AND MEASUREMENT CAPABILITY (CMC)*
MEASURED RANGE UNCERTAINTY1,2 CALIBRATION METHOD OR
QUANTITY or DEVICE (±) PROCEDURE, STANDARD
TYPE CALIBRATED EQUIPMENT (OPTIONAL)
Dimensional
Caliper 0 mm to 300 mm 17 µm Caliper Checker
(Digital, Dial, Vernier)
Height Gauge 0 mm to 300 mm 13 µm Caliper Checker
(Digital, Dial, Vernier)
Micrometer 2.5 mm to 25 mm 1.3 µm Gauge Block Set
Dial Gauge 0 mm to 25 mm 6.0 µm Dial Gauge Calibrator
Mechanical
Pressure Gauge Up to 1000 bar 1.5 bar Dead Weight tester
High Pressure Comparison
Pump and Digital Pressure
Gauge
Pressure Gauge Pneumatic 0 bar to 100 bar 0.02 bar High Pressure Pneumatic
Hand Pump and Digital
Pressure Gauge
Vacuum Gauge -0.85 bar to 0 bar 0.02 bar High Pressure Pneumatic
Hand Pump and Digital
Vacuum Gauge
Weighing Balance 1 mg to 500 mg 0.14 mg E2 Class Weights
1 g to 200 g 160 mg
201 g to 5000 g 170 mg
Tachometer 12 rpm to 12000 rpm 1.0 rpm Tachometer Calibrator
(Contact Type)
Thermal
RTD / -10 °C to 1200 °C 0.3 °C Dry Bath Calibrator
Thermocouple
Temperature Bath 50 °C to 600 °C 0.6 °C SSPRT PT-100(4W) with
Fluke DMM 8846A
* If information in this CMC is presented in non-SI units, the conversion factors stated in NIST Special Publication
811 “Guide for the Use of the International System of Units (SI)” apply.
•
CL-221 Effective Date February 10, 2023
Universal Testing Co. Ltd. Page 2 of 4
IAS/CL/100-1 IAS/CL/100-3
MEASURED RANGE UNCERTAINTY1,2 CALIBRATION METHOD OR
QUANTITY or DEVICE (±) PROCEDURE, STANDARD
TYPE CALIBRATED EQUIPMENT (OPTIONAL)
Electrical – DC/LF
DC Voltage – Source3 1 mV to 20 mV 0.08 mV CLARK HESS Multifunction
20 mV to 200 mV 0.60 mV Electrical Calibrator
200 mV to 2 V 0.01 V
2 V to 20 V 0.06 V
20 V to 240 V 0.72 V
240 V to 1000 V 3.0 V
AC Voltage – Source3 1 mV to 20 mV 0.06 mV CLARK HESS Multifunction
@50 Hz 20 mV to 200 mV 0.60 mV Electrical Calibrator
200 mV to 2 V 0.01 V
2 V to 20 V 0.06 V
20 V to 240 V 0.72 V
240 V to 1000 V 3.0 V
DC Current - Source3 1 µA to 200 µA 0.60 µA CLARK HESS Multifunction
200 µA to 2 mA 0.01 mA Electrical Calibrator
2 mA to 20 mA 0.06 mA
20 mA to 200 mA 0.60 mA
200 mA to 2 A 0.01 A
2 A to 20 A 0.06 A
AC Current – Source3 1 µA to 200 µA 0.62 µA CLARK HESS Multifunction
@50 Hz 200 µA to 2 mA 0.01 mA Electrical Calibrator
2 mA to 20 mA 0.06 mA
20 mA to 200 mA 0.60 mA
200 mA to 2 A 0.01 A
2 A to 20 A 0.06 A
Resistance - Source3 1 kΩ to 10 kΩ 0.03 kΩ ZEAL ZMDRB Decade
10 kΩ to 100 kΩ 0.30 kΩ Resistance Box
100 kΩ to 1 MΩ 0.003 MΩ
1 MΩ to 10 MΩ 0.03 MΩ
10 MΩ to 100 MΩ 1.0 MΩ
100 MΩ to 1000 MΩ 23 MΩ
DC Voltage -Measure4 0 mV to 100 mV 0.30 mV Fluke 8846A Precision
100 mV to 1 V 0.003 V Multimeter
1 V to 10 V 0.03 V
10 V to 100 V 0.30 V
100 V to 1000 V 3.0 V
AC Voltage- Measure4 0 mV to 100 mV 0.29 mV Fluke 8846A Precision
@50 Hz 100 mV to 1 V 0.003 V Multimeter
1 V to 10 V 0.03 V
10 V to 100 V 0.29 V
100 V to 1000 V 3.0 V
CL-221 Effective Date February 10, 2023
Universal Testing Co. Ltd. Page 3 of 4
IAS/CL/100-1 IAS/CL/100-3
MEASURED RANGE UNCERTAINTY1,2 CALIBRATION METHOD OR
QUANTITY or DEVICE (±) PROCEDURE, STANDARD
TYPE CALIBRATED EQUIPMENT (OPTIONAL)
DC Current – Measure4 0 μA to 100 μA 0.30 µA Fluke 8846A Precision
100 uA to 1 mA 0.003 mA Multimeter
1 mA to 10 mA 0.03 mA
10 mA to 100 mA 0.30 mA
100 mA to 400 mA 1.2 mA
400 mA to 1 A 0.003 A
1 A to 3 A 0.009 A
3 A to 10 A 0.03 A
AC Current - Measure4 0 μA to 100 μA 0.30 µA Fluke 8846A Precision
@50Hz 100 uA to 1 mA 0.003 mA Multimeter
1 mA to 10 mA 0.03 mA
10 mA to 100 mA 0.30 mA
100 mA to 400 mA 1.2 mA
400 mA to 1 A 0.002 A
1 A to 3 A 0.01 A
3 A to 10 A 0.03 A
DC Resistance – Measure4 1 kΩ to 10 kΩ 0.03 kΩ Fluke 8846A Precision
10 kΩ to 100 kΩ 0.30 kΩ Multimeter
100 kΩ to 1 MΩ 0.003 MΩ
1 MΩ to 10 MΩ 0.004 MΩ
10 MΩ to 100 MΩ 0.30 MΩ
100 MΩ to 1000 MΩ 3.0 MΩ
Electric & Diesel Welding Up to 600 A 0.84 A ESAB Check Master 9000 as
Machine per BS EN 50504:2008
1The uncertainty covered by the Calibration and Measurement Capability (CMC) is expressed as the expanded
uncertainty having a coverage probability of approximately 95 %. It is the smallest measurement uncertainty that a
laboratory can achieve within its scope of accreditation when performing calibrations of a best existing device. The
measurement uncertainty reported on a calibration certificate may be greater than that provided in the CMC due to the
behavior of the calibration item and other factors that may contribute to the uncertainty of a specific calibration.
2When uncertainty is stated in relative terms (such as percent, a multiplier expressed as a decimal fraction or in scientific
notation), it is in relation to instrument reading or instrument output, as appropriate, unless otherwise indicated.
3Capability is suitable for the calibration of measuring devices in the stated ranges.
4Capability is suitable for the calibration of devices intended to generate the indicated quantity in the stated ranges.
CL-221 Effective Date February 10, 2023
Universal Testing Co. Ltd. Page 4 of 4
IAS/CL/100-1 IAS/CL/100-3