BMS Medical Gas System T&C

This document outlines test procedures for medical compressed air and dental compressed air equipment. It includes 8 tests to verify the monitoring and alarm functions of various system components like air compressors, air receivers, and dryers. Each test step lists the expected result to check, such as confirming alarm statuses change from "Normal" to "Alarm" as simulated fault conditions are triggered. Personnel initial and date the results column after completing each step.

Uploaded by

Ranjith Kumar
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as DOCX, PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
99 views23 pages

BMS Medical Gas System T&C

This document outlines test procedures for medical compressed air and dental compressed air equipment. It includes 8 tests to verify the monitoring and alarm functions of various system components like air compressors, air receivers, and dryers. Each test step lists the expected result to check, such as confirming alarm statuses change from "Normal" to "Alarm" as simulated fault conditions are triggered. Personnel initial and date the results column after completing each step.

Uploaded by

Ranjith Kumar
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as DOCX, PDF, TXT or read online on Scribd
You are on page 1/ 23

MEDICAL COMPRESSED AIR / DENTAL COMPRESSED AIR Test Form

SOFTWARE FUNCTIONAL TEST FORM -1

Form No: ECC-NSCKWE-BMS-


Project: PROJECT THE ECC & NDCS @ SGH TESTFROM-0001

Equipment ID: GAS-XX-XX Date: Revision: 0

Process Description: Testing and Commissioning of GAS-XX-XX

Area/Location: Panel ID: GAS-XX-XX

EXPECTED RESULT

RESUL (INITIAL & REMARK


SR.NO. TESTING ACTIVITY
T DATE) S
BMS – SOFTWARE ONLINE &
GRAPHIC DISPLAY

LMCP/MG/CA/B2/01
TEST –
LMCP/MG/CA/B2/01
1
Monitoring Points

GAS-XX-XX_EqAlmSts=
Alarm Status is “Normal”
“NORMAL”
1.
Simulate the condition is
GAS-XX-XX_EqAlmSts =
“Alarm”
“ALARM”

Monitor the Mode to “Local” GAS-XX-XX_Mode = “LOCAL”


2.
Switch the Mode to “BMS” GAS-XX-XX_Mode = “BMS”
EXPECTED RESULT

RESUL (INITIAL & REMARK


SR.NO. TESTING ACTIVITY
T DATE) S
BMS – SOFTWARE ONLINE &
GRAPHIC DISPLAY

Monitor Close Status is


GAS -XX-XX_EqSts =
“NORMAL”
3. “NORMAL”

Monitor Close Status is


GAS -XX-XX_EqSts = “CLOSE”
“CLOSE”

Fault Status is “Normal” GAS-XX-XX_AlmSts=


4. “NORMAL”
Simulate the condition is
“Alarm” GAS-XX-XX_AlmSts = “FAULT”

Monitor the Incoming Power GAS-W-XX-XX_PwrSts=


Status is “Normal” “NORMAL”
5.
Monitor the Incoming Power GAS-W-XX-XX_PwrSts=
Status is “ALARM” “ALARM”

Air Compressors
TEST –
Air Compressors
2
Monitoring Points

Monitor the Status is “ON” GAS -XX-XX_EqSts = “ON


1.
Monitor the Status is “OFF” GAS -XX-XX_EqSts = “OFF”

GAS-XX-XX_EqAlmSts=
Alarm Status is “Normal”
“NORMAL”
2.
Simulate the condition is
GAS-XX-XX_EqAlmSts =
“Alarm”
“ALARM”

Fault Status is “Normal” GAS-XX-XX_AlmSts=


3. “NORMAL”
Simulate the condition is
“Alarm” GAS-XX-XX_AlmSts = “ALARM”

Monitor the Pump Status is


“ON” GAS -XX-XX_EqSts = “ON
4.
Monitor the Pump Status is GAS -XX-XX_EqSts = “OFF”
“OFF”
EXPECTED RESULT

RESUL (INITIAL & REMARK


SR.NO. TESTING ACTIVITY
T DATE) S
BMS – SOFTWARE ONLINE &
GRAPHIC DISPLAY

Pump Status is “Normal” GAS-XX-XX_AlmSts=


“NORMAL”
5
Simulate the condition is
“Alarm” GAS-XX-XX_AlmSts = “FAIL”

TEST – Analog input Monitoring


3 Points

Check the Medical gas Medical gas pressure Sensor


1
pressure Sensor. readings: ______ Pa

Check the Delivery Pressure Delivery Pressure Sensor


2
Sensor. readings ______ Pa

Air Compressor After Coolers


TEST –
Air Compressor After Coolers
4
Monitoring Points

GAS-XX-XX_EqAlmSts=
Alarm Status is “Normal”
“NORMAL”
1
Simulate the condition is
GAS-XX-XX_EqAlmSts =
“Alarm”
“ALARM”

Monitor the Mode to “Local” GAS-XX-XX_Mode = “LOCAL”


2
Switch the Mode to “BMS” GAS-XX-XX_Mode = “BMS”

Fault Status is “Normal” GAS-XX-XX_AlmSts=


“NORMAL”
3
Simulate the condition is
“Alarm” GAS-XX-XX_AlmSts = “FAULT”

TEST – Analog input Monitoring


5 Points

Check the Cooler Cooler Temperature Sensor


1
Temperature Sensor. readings: ______ ºC
EXPECTED RESULT

RESUL (INITIAL & REMARK


SR.NO. TESTING ACTIVITY
T DATE) S
BMS – SOFTWARE ONLINE &
GRAPHIC DISPLAY

Air Receivers
TEST –
Air Receivers
6
Monitoring Points

GAS-XX-XX_EqAlmSts=
Alarm Status is “Normal”
“NORMAL”
1
Simulate the condition is
GAS-XX-XX_EqAlmSts =
“Alarm”
“ALARM”

Monitor the Mode to “Local” GAS-XX-XX_Mode = “LOCAL”


2
Switch the Mode to “BMS” GAS-XX-XX_Mode = “BMS”

Fault Status is “Normal” GAS-XX-XX_AlmSts=


“NORMAL”
3
Simulate the condition is
“Alarm” GAS-XX-XX_AlmSts = “FAULT”

Pressure Fault Status is


GAS-XX-XX_AlmSts=
“Normal”
“NORMAL”
4
Simulate the condition is
GAS-XX-XX_AlmSts = “FAULT”
“Alarm”

GAS-XX-XX_EqAlmSts=
Pressure Status is “Normal”
“NORMAL”
5
Simulate the condition is
GAS-XX-XX_EqAlmSts =
“Alarm”
“ALARM”

TEST – Analog input Monitoring


7 Points

Check the Medical gas HL/LL Medical gas pressure HL/LL


1
pressure Sensor. Sensor readings: ______ Pa

Check the Air Receiver Air Receiver Pressure Sensor


2
Pressure Sensor. readings ______ Pa ºC
EXPECTED RESULT

RESUL (INITIAL & REMARK


SR.NO. TESTING ACTIVITY
T DATE) S
BMS – SOFTWARE ONLINE &
GRAPHIC DISPLAY

Duplex Column Dryers


TEST –
Duplex Column Dryers
8
Monitoring Points

GAS-XX-XX_EqAlmSts=
Alarm Status is “Normal”
“NORMAL”
1
Simulate the condition is
GAS-XX-XX_EqAlmSts =
“Alarm”
“ALARM”

Monitor the Mode to “Local” GAS-XX-XX_Mode = “LOCAL”


2
Switch the Mode to “BMS” GAS-XX-XX_Mode = “BMS”

Fault Status is “Normal” GAS-XX-XX_AlmSts=


“NORMAL”
3
Simulate the condition is
“Alarm” GAS-XX-XX_AlmSts = “FAULT”

Pressure Fault Status is


GAS-XX-XX_AlmSts=
“Normal”
“NORMAL”
4
Simulate the condition is
GAS-XX-XX_AlmSts = “FAULT”
“Alarm”

GAS-XX-XX_EqAlmSts=
Pressure Status is “Normal”
“NORMAL”
5
Simulate the condition is
GAS-XX-XX_EqAlmSts =
“Alarm”
“ALARM”

Space Status is “Normal” GAS-XX-XX_AlmSts=


“NORMAL”
6
Simulate the condition is
“Alarm” GAS-XX-XX_AlmSts = “FAULT”

Dewpoint the Status is “ON” GAS -XX-XX_EqSts = “ON


7
Monitor the Status is “OFF” GAS -XX-XX_EqSts = “OFF”
EXPECTED RESULT

RESUL (INITIAL & REMARK


SR.NO. TESTING ACTIVITY
T DATE) S
BMS – SOFTWARE ONLINE &
GRAPHIC DISPLAY

Pressure Regulators (4 bar)


TEST –
Pressure Regulators (4 bar)
9
Analog Input Points

Check the Medical gas HL/LL Medical gas pressure HL/LL


1
pressure Sensor. Sensor readings: ______ Pa

Check the Delivery Pressure Delivery Pressure Sensor


2
Sensor. readings ______ Pa

Pressure Regulators (7 bar)


TEST –
Pressure Regulators (7 bar)
10
Analog Input Points

Check the Medical gas HL/LL Medical gas pressure HL/LL


1
pressure Sensor. Sensor readings: ______ Pa

Check the Delivery Pressure Delivery Pressure Sensor


2
Sensor. readings ______ Pa

Header Pipe (4 bar)


TEST –
Header Pipe (4 bar)
11
Monitoring Points

Pressure Fault Status is


GAS-XX-XX_AlmSts=
“Normal”
“NORMAL”
1
Simulate the condition is
GAS-XX-XX_AlmSts = “FAULT”
“Alarm”

Header Pipe (4 bar)


TEST –
Header Pipe (4 bar)
12
Analog Input Points

Check the Medical gas HL/LL Medical gas pressure HL/LL


1
pressure Sensor. Sensor readings: ______ Pa
EXPECTED RESULT

RESUL (INITIAL & REMARK


SR.NO. TESTING ACTIVITY
T DATE) S
BMS – SOFTWARE ONLINE &
GRAPHIC DISPLAY

Header Pipe (7 bar)


TEST –
Header Pipe (7 bar)
13
Monitoring Points

Pressure Fault Status is


GAS-XX-XX_AlmSts=
“Normal”
“NORMAL”
1
Simulate the condition is
GAS-XX-XX_AlmSts = “FAULT”
“Alarm”

Header Pipe (7 bar)


TEST –
Header Pipe (7 bar)
14
Analog Input Points

Check the Medical gas HL/LL Medical gas pressure HL/LL


1
pressure Sensor. Sensor readings: ______ Pa

MEDICAL COMPRESSED
TEST – AIR / DENTAL
15 COMPRESSED AIR High
Level Integration.

Test the MEDICAL


All the parameters are displayed
COMPRESSED AIR /
in BMS as per BMS interface
DENTAL COMPRESSED AIR
document.
for HLI.

Tested By: Acknowledged


Honeywell Pte Ltd
By: Kilowatts

NAME: NAME:

DATE: DATE:
Submitted and Penta-Ocean Construction Co, Witnessed and
RE/RTO
Checked By: Ltd. Checked BY:

NAME: NAME:

DATE: DATE:
MEDICAL VACUUM / DENTAL VACUUM Test Form

SOFTWARE FUNCTIONAL TEST FORM -2

Form No: ECC-NSCKWE-BMS-


Project: PROJECT THE ECC & NDCS @ SGH TESTFROM-0002

Equipment ID: GAS-XX-XX Date: Revision: 0

Process Description: Testing and Commissioning of GAS-XX-XX

Area/Location: Panel ID: GAS-XX-XX

EXPECTED RESULT

RESUL (INITIAL & REMARK


SR.NO. TESTING ACTIVITY
T DATE) S
BMS – SOFTWARE ONLINE &
GRAPHIC DISPLAY

TEST – LMCP/MG/VAC/B2/01
LMCP/MG/VAC/B2/01
1 Monitoring Points

GAS-XX-XX_EqAlmSts=
Alarm Status is “Normal”
“NORMAL”
1
Simulate the condition is
GAS-XX-XX_EqAlmSts =
“Alarm”
“ALARM”

Monitor the Mode to “Local” GAS-XX-XX_Mode = “LOCAL”


2
Switch the Mode to “BMS” GAS-XX-XX_Mode = “BMS”
EXPECTED RESULT

RESUL (INITIAL & REMARK


SR.NO. TESTING ACTIVITY
T DATE) S
BMS – SOFTWARE ONLINE &
GRAPHIC DISPLAY

Monitor Close Status is


GAS -XX-XX_EqSts =
“NORMAL”
“NORMAL”
3
Monitor Close Status is
GAS -XX-XX_EqSts = “CLOSE”
“CLOSE”

Fault Status is “Normal” GAS-XX-XX_AlmSts=


“NORMAL”
4
Simulate the condition is
“Alarm” GAS-XX-XX_AlmSts = “FAULT”

Monitor the Incoming Power GAS-W-XX-XX_PwrSts=


Status is “Normal” “NORMAL”
5
Monitor the Incoming Power GAS-W-XX-XX_PwrSts=
Status is “ALARM” “ALARM”

TEST – Vacuum Pumps


Vacuum Pumps
2 Monitoring Points

Monitor the Status is “ON” GAS -XX-XX_EqSts = “ON


1
Monitor the Status is “OFF” GAS -XX-XX_EqSts = “OFF”

GAS-XX-XX_EqAlmSts=
Alarm Status is “Normal”
“NORMAL”
2
Simulate the condition is
GAS-XX-XX_EqAlmSts =
“Alarm”
“ALARM”

Fault Status is “Normal” GAS-XX-XX_AlmSts=


“NORMAL”
3
Simulate the condition is
“Alarm” GAS-XX-XX_AlmSts = “ALARM”

Monitor the Pump Status is


“ON” GAS -XX-XX_EqSts = “ON
4
Monitor the Pump Status is GAS -XX-XX_EqSts = “OFF”
“OFF”
EXPECTED RESULT

RESUL (INITIAL & REMARK


SR.NO. TESTING ACTIVITY
T DATE) S
BMS – SOFTWARE ONLINE &
GRAPHIC DISPLAY

Pump Status is “Normal” GAS-XX-XX_AlmSts=


“NORMAL”
5
Simulate the condition is
“Alarm” GAS-XX-XX_AlmSts = “FAIL”

TEST – Analog input Monitoring


3 Points

Check the VACUUM ACTUAL VACUUM ACTUAL PRESSURE


1
PRESSURE Sensor. readings: ______ Pa

TEST – Vacuum Reserviors


Vacuum Reserviors
4 Monitoring Points

GAS-XX-XX_EqAlmSts=
Alarm Status is “Normal”
“NORMAL”
1
Simulate the condition is
GAS-XX-XX_EqAlmSts =
“Alarm”
“ALARM”

Monitor the Mode to “Local” GAS-XX-XX_Mode = “LOCAL”


2
Switch the Mode to “BMS” GAS-XX-XX_Mode = “BMS”

Fault Status is “Normal” GAS-XX-XX_AlmSts=


“NORMAL”
3
Simulate the condition is
“Alarm” GAS-XX-XX_AlmSts = “FAULT”

TEST – Analog input Monitoring


5 Points

Check the VACUUM ACTUAL VACUUM ACTUAL PRESSURE


1
PRESSURE Sensor. Sensor readings: ______ ºC

Filters
TEST –
Filters
6
Monitoring Points
EXPECTED RESULT

RESUL (INITIAL & REMARK


SR.NO. TESTING ACTIVITY
T DATE) S
BMS – SOFTWARE ONLINE &
GRAPHIC DISPLAY

GAS-XX-XX_EqAlmSts=
Alarm Status is “Normal”
“NORMAL”
1
Simulate the condition is
GAS-XX-XX_EqAlmSts =
“Alarm”
“ALARM”

Monitor the Mode to “Local” GAS-XX-XX_Mode = “LOCAL”


2
Switch the Mode to “BMS” GAS-XX-XX_Mode = “BMS”

Fault Status is “Normal” GAS-XX-XX_AlmSts=


“NORMAL”
3
Simulate the condition is
“Alarm” GAS-XX-XX_AlmSts = “FAULT”

Pressure Fault Status is


GAS-XX-XX_AlmSts=
“Normal”
“NORMAL”
4
Simulate the condition is
GAS-XX-XX_AlmSts = “FAULT”
“Alarm”

GAS-XX-XX_EqAlmSts=
Pressure Status is “Normal”
“NORMAL”
5
Simulate the condition is
GAS-XX-XX_EqAlmSts =
“Alarm”
“ALARM”

TEST – Analog input Monitoring


7 Points

VACUUM ACTUAL VACUUM ACTUAL PRESSURE


1
PRESSURE Sensor. Sensor readings: ______ Pa

Header Pipe (4 bar)


TEST –
Header Pipe (4 bar)
11
Monitoring Points
EXPECTED RESULT

RESUL (INITIAL & REMARK


SR.NO. TESTING ACTIVITY
T DATE) S
BMS – SOFTWARE ONLINE &
GRAPHIC DISPLAY

Pressure Fault Status is


GAS-XX-XX_AlmSts=
“Normal”
“NORMAL”
1
Simulate the condition is
GAS-XX-XX_AlmSts = “FAULT”
“Alarm”

MEDICAL COMPRESSED
TEST – AIR / DENTAL
15 COMPRESSED AIR High
Level Integration.

Test the MEDICAL


All the parameters are displayed
COMPRESSED AIR /
1 in BMS as per BMS interface
DENTAL COMPRESSED AIR
document.
for HLI.
Tested By: Acknowledged
Honeywell Pte Ltd
By: Kilowatts

NAME: NAME:

DATE: DATE:

Submitted and Penta-Ocean Construction Co, Witnessed and


RE/RTO
Checked By: Ltd. Checked BY:

NAME: NAME:

DATE: DATE:
AGSS Plant Test Form

SOFTWARE FUNCTIONAL TEST FORM -3

Form No: ECC-NSCKWE-BMS-


Project: PROJECT THE ECC & NDCS @ SGH TESTFROM-0003

Equipment ID: GAS-XX-XX Date: Revision: 0

Process Description: Testing and Commissioning of GAS-XX-XX

Area/Location: Panel ID: GAS-XX-XX

EXPECTED RESULT

RESUL (INITIAL & REMARK


SR.NO. TESTING ACTIVITY
T DATE) S
BMS – SOFTWARE ONLINE &
GRAPHIC DISPLAY

TEST – AGSS Pumps (Duplex)


AGSS Pumps (Duplex)
1 Monitoring Points

GAS-XX-XX_EqAlmSts=
Alarm Status is “Normal”
“NORMAL”
1
Simulate the condition is
GAS-XX-XX_EqAlmSts =
“Alarm”
“ALARM”

Monitor the Mode to “Local” GAS-XX-XX_Mode = “LOCAL”


2
Switch the Mode to “BMS” GAS-XX-XX_Mode = “BMS”
EXPECTED RESULT

RESUL (INITIAL & REMARK


SR.NO. TESTING ACTIVITY
T DATE) S
BMS – SOFTWARE ONLINE &
GRAPHIC DISPLAY

Monitor Close Status is


GAS -XX-XX_EqSts =
“NORMAL”
“NORMAL”
3
Monitor Close Status is
GAS -XX-XX_EqSts = “CLOSE”
“CLOSE”

Fault Status is “Normal” GAS-XX-XX_AlmSts=


“NORMAL”
4
Simulate the condition is
“Alarm” GAS-XX-XX_AlmSts = “FAULT”

Monitor the Incoming Power GAS-W-XX-XX_PwrSts=


Status is “Normal” “NORMAL”
5
Monitor the Incoming Power GAS-W-XX-XX_PwrSts=
Status is “ALARM” “ALARM”

Pump Status is “Normal” GAS-XX-XX_AlmSts=


“NORMAL”
6
Simulate the condition is
“Alarm” GAS-XX-XX_AlmSts = “FAIL”

TEST – Analog input Monitoring


2 Points

Check the VACUUM ACTUAL VACUUM ACTUAL PRESSURE


1
PRESSURE Sensor. readings: ______ Pa

Check the AGSS VACUUM AGSS VACUUM PRESSURE


1
PRESSURE Sensor. readings: ______ Pa

TEST – AGSS Plant High Level


3 Integration.

All the parameters are displayed


Test the AGSS Plant for HLI. in BMS as per BMS interface
document.
Tested By: Acknowledged
Honeywell Pte Ltd
By: Kilowatts

NAME: NAME:

DATE: DATE:

Submitted and Penta-Ocean Construction Co, Witnessed and


RE/RTO
Checked By: Ltd. Checked BY:

NAME: NAME:

DATE: DATE:
GAS CYLINDER MANIFOLD / O2 PRESSURE REGULATOR Test Form

SOFTWARE FUNCTIONAL TEST FORM -4

Form No: ECC-NSCKWE-BMS-


Project: PROJECT THE ECC & NDCS @ SGH TESTFROM-0004

Equipment ID: GAS-XX-XX Date: Revision: 0

Process Description: Testing and Commissioning of GAS-XX-XX

Area/Location: Panel ID: GAS-XX-XX

EXPECTED RESULT

RESUL (INITIAL & REMARK


SR.NO. TESTING ACTIVITY
T DATE) S
BMS – SOFTWARE ONLINE &
GRAPHIC DISPLAY

TEST – AGSS Pumps (Duplex)


AGSS Pumps (Duplex)
1 Monitoring Points

Monitor the Status is “ON” GAS -XX-XX_EqSts = “ON


1
Monitor the Status is “OFF” GAS -XX-XX_EqSts = “OFF”

Monitor the Incoming Power GAS-W-XX-XX_PwrSts=


Status is “Normal” “NORMAL”
2
Monitor the Incoming Power GAS-W-XX-XX_PwrSts=
Status is “ALARM” “ALARM”
EXPECTED RESULT

RESUL (INITIAL & REMARK


SR.NO. TESTING ACTIVITY
T DATE) S
BMS – SOFTWARE ONLINE &
GRAPHIC DISPLAY

Monitor the Duty bank


running Status is “ON” GAS -XX-XX_EqSts = “ON
3
Monitor the Duty bank GAS -XX-XX_EqSts = “OFF”
running Status is “OFF”

Monitor the Standby bank


running Status is “ON” GAS -XX-XX_EqSts = “ON
4
Monitor the Standby bank GAS -XX-XX_EqSts = “OFF”
running Status is “OFF”

Monitor the Emergency bank


running Status is “ON” GAS -XX-XX_EqSts = “ON
5
Monitor the Emergency bank GAS -XX-XX_EqSts = “OFF”
running Status is “OFF”

Pump Status is “Normal” GAS-XX-XX_AlmSts=


“NORMAL”
6
Simulate the condition is
“Alarm” GAS-XX-XX_AlmSts = “FAIL”

TEST – Analog input Monitoring


2 Points

Check the MEDICAL GAS MEDICAL GAS PRESSURE


1
PRESSURE Sensor. readings: ______ Pa

Check the VACUUM ACTUAL VACUUM ACTUAL PRESSURE


2
PRESSURE Sensor. readings: ______ Pa

Check the LEFT DUTY BANK LEFT DUTY BANK PRESSURE


3
PRESSURE Sensor. readings: ______ Pa

RIGHT DUTY BANK


Check the RIGHT DUTY
4 PRESSURE readings: ______
BANK PRESSURE Sensor.
Pa
EXPECTED RESULT

RESUL (INITIAL & REMARK


SR.NO. TESTING ACTIVITY
T DATE) S
BMS – SOFTWARE ONLINE &
GRAPHIC DISPLAY

Check the Standby Standby PRESSURE readings:


5
PRESSURE Sensor. ______ Pa

GAS CYLINDER
TEST – MANIFOLD / O2 PRESSURE
3 REGULATOR High Level
Integration.

Test the GAS CYLINDER All the parameters are displayed


MANIFOLD / O2 PRESSURE in BMS as per BMS interface
REGULATOR for HLI. document.

Tested By: Acknowledged


Honeywell Pte Ltd
By: Kilowatts

NAME: NAME:

DATE: DATE:

Submitted and Penta-Ocean Construction Co, Witnessed and


RE/RTO
Checked By: Ltd. Checked BY:

NAME: NAME:

DATE: DATE:
MEDICAL GAS WARNING AND ALARM SYSTEM (AVSU/MGAP) Test Form

SOFTWARE FUNCTIONAL TEST FORM -5

Form No: ECC-NSCKWE-BMS-


Project: PROJECT THE ECC & NDCS @ SGH TESTFROM-0005

Equipment ID: GAS-XX-XX Date: Revision: 0

Process Description: Testing and Commissioning of GAS-XX-XX

Area/Location: Panel ID: GAS-XX-XX

EXPECTED RESULT

(INITIAL &
SR.NO. TESTING ACTIVITY RESULT REMARKS
DATE)
BMS – SOFTWARE ONLINE &
GRAPHIC DISPLAY

AGSS Pumps (Duplex)


TEST – 1 AGSS Pumps (Duplex)
Monitoring Points
EXPECTED RESULT

(INITIAL &
SR.NO. TESTING ACTIVITY RESULT REMARKS
DATE)
BMS – SOFTWARE ONLINE &
GRAPHIC DISPLAY

Monitor the Medical Gas Status


GAS-W-XX-XX_PwrSts=
is “Normal”
“NORMAL”
1
Monitor the Medical Gas Status
GAS-W-XX-XX_PwrSts= “ALARM”
is “ALARM”

Monitor the Incoming Power


GAS-W-XX-XX_PwrSts=
Status is “Normal”
“NORMAL”
2
Monitor the Incoming Power
GAS-W-XX-XX_PwrSts= “ALARM”
Status is “ALARM”

Monitor the SPARE AUXILIARY


GAS-W-XX-XX_PwrSts=
Status is “Normal”
“NORMAL”
3
Monitor the SPARE AUXILIARY
GAS-W-XX-XX_PwrSts= “ALARM”
Status is “ALARM”

Tested By: Acknowledged


Honeywell Pte Ltd
By: Kilowatts

NAME: NAME:

DATE: DATE:

Submitted and Penta-Ocean Construction Co, Witnessed and


RE/RTO
Checked By: Ltd. Checked BY:

NAME: NAME:

DATE: DATE: