0% found this document useful (0 votes)
91 views107 pages

Tonometer Edited

The document discusses intraocular pressure (IOP) and various methods used to measure it. IOP is the fluid pressure inside the eye and is an important factor in evaluating glaucoma risk. The Schiotz tonometer is described as an indentation tonometer that measures IOP based on the indentation of a weighted plunger into the cornea. A mathematical formula developed by Friedenwald is used to convert the tonometer readings into estimated IOP measurements based on the weighted plunger and scale reading.

Uploaded by

ANUSAYA SAHU
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
91 views107 pages

Tonometer Edited

The document discusses intraocular pressure (IOP) and various methods used to measure it. IOP is the fluid pressure inside the eye and is an important factor in evaluating glaucoma risk. The Schiotz tonometer is described as an indentation tonometer that measures IOP based on the indentation of a weighted plunger into the cornea. A mathematical formula developed by Friedenwald is used to convert the tonometer readings into estimated IOP measurements based on the weighted plunger and scale reading.

Uploaded by

ANUSAYA SAHU
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 107

TONOMETER

What is IOP ?

increased risk to develop glaucoma: >21mm Hg


Intraocular pressure (IOP)

 It is the fluid pressure inside the eye.


 IOP is an important aspect in the evaluation of
patients at risk from glaucoma.
 Eye pressure is measured in millimeters of mercury
(mm Hg).
 Normal eye pressure ranges from 12-22 mm Hg, and
eye pressure of greater than 22 mm Hg is considered
higher than normal.
 When the IOP is higher than normal but the person
does not show signs of glaucoma, this is referred to
as ocular hypertension.
 High eye pressure alone does not cause glaucoma.
However, it is a significant risk factor.
 Individuals diagnosed with high eye pressure should
have regular comprehensive eye examinations by an
eyecare professional to check for signs of the onset of
glaucoma.
HISTORY
1826-William Bowman Digital tonometry(as routine
examination)
1863-Albrecht von Grafe Designed first instrument to attempt to
measure IOP
1865- Further instruments followed
Donders Indentation type(noanaesthetic was
1880-Preistly used until 1884)
1885-Malkalov 1stApplanation tonometer
1905-Hjalmar Schiotz Indentation tonometry
1948,1955-Friedenwald Coefficient of ocular rigidity
1954-Goldmann Prototype Applanation
tonometer(constant area)
1972-Grolmann NCT
Grant Electronic indentation tonometer
Halberg Hand held tonometer
TONOMETER

 Definition: A tool used to measure the pressure of the eye.

Application :
 Used to help detect glaucoma.
 Measures the production of aqueous humor, the liquid found
inside the eye, and the rate at which it drains in to the tissue
surrounding the cornea
 Procedure performed to determine the intraocular pressure (IOP)
 Most tonometers are calibrated to measure pressure in
millimeters of mercury (mmHg).
Albrecht von Grafe tonometer
Donders tonometer
IDEAL TONOMETER

 Should give accurate and reasonable IOP


measurement
 Convenient to use
 Simple to calibrate
 Stable from day to day
 Easier to standardise
 Free of maintenance problems
FACTORS MODIFYING IOP

1. Physiological variations : the IOP normally


fluctuates 2-5mmHg throughtout the day :
• with respiration and heart beat
• with time of the day
• with the venous pressure
• with the arterial pressure
• with the osmotic pressure of blood
Continue

2. Local mechanical factors :


• dilatation of the pupil
• changes in the solid content of the eye
• pressure from outside

3. Pharmacological factors:
The ciliary muscle is inserted into the trabeculum,
so the contraction of the ciliary muscle makes the
trabecular meshwork more porous -> increases
the facility of outflow -> reduces IOP
• outflow facility
• reduction of aqueous production
• atropine
Measurement of IOP

 Manometry : it is measured by inserting a cannula,


directly into the anterior chamber which is
connected with a manometer.
Continue

 Digital tonometry
Tonometry Direct

Indirect Manometry

TYPES OF TONOMETRY
INDENTATION APPLANATION
Schiotz Variable force
Examples- GAT, Perkins, MMT,
Tonopen, Pneum. Tonometer
Mercurial Constant force
Examples- Maklakov, Glucotest,
Applanometer
Electronic
Scleral Tonometry
 Vogelsang 1927 – Ballistic Tonometer .The rebound of
a small metal ball from the eye is measured and this
depends to a large extent on the physical properties of
the coats of the eye.

 Roth & Blake 1963- Vibration tonometer cause


minimal deformation by oscillating force by a probe,
which also functions as a sensor and measures the
resonant frequency of the eye.
 Newer tonometers
1. Trans –palpebral
Tonometer
2. Disposable tonometer
 Tonosafe – acrylic biprism
 Tonoshield- silicone
shield
3. Dynamic contour
tonometer
4. Ocular Response Analyzer
A.HIGH DISPLACEMENT B.LOW DISPLACEMENT
TONOMETERS TONOMETERS
Tonometers that displace a large Tonometers in which the IOP is
volume of fluid and consequently negligibly raised during tonometry
raise IOP significantly. (less than 5%).
Examples Examples
1. Schiotz. 1. GoldmannApplanation
2. Maklakov. Tonometer.
2. Mackay-Marg tonometer.

The Goldmann tonometer displaces


only 0.5 μl of aqueous humor and
raises IOP by only 3%.
Less accurate More accurate
Needle inserted into the AC through a self-sealing, beveled
corneal puncture

Needle is connected to a fluid-filled tubing

Height of the fluid in the tubing corresponds to IOP.

Tubing can also be connected to a fluid-filled reservoir that


has a pressure-sensitive membrane.

Movement of membrane, recorded optically or


electronically, is a measure of IOP.
 USES:

1. It is used for continuous


measurements of IOP.

2. Used in experiment, research


work on animal eyes.

 DISADVANTAGES:

1. Not practical method for


human beings.
2. Needs general anesthesia.

3. Introduction of needle
produces breakdown of blood
aqueous barrier and release of
prostaglandins which alter
IOP.
 It is an indirect method of measuring the IOP.

 It includes digital tonometry.

 Three basic types of Tonometers:


o Indentation
o Applanation(flattening)
o Non-Contact
 Intraocular pressure (IOP) is estimated by response of
eye to pressure applied by finger pulp.

 PROCEDURE: Patient looks down

Index finger of both hands used

One finger is kept stationary which feels the fluctuation


produced by the indentation of globe by the other finger.
If IOP is raisedfluctuation produced is feeble or absent and
the eyeball feels firm to hard.
When the IOP is very loweye feels soft like a partially filled
balloon.
`

ADVANTAGES DISADVANTAGES
Easiest to perform Reading not proper
No equipment Only depends on examiner
No anaesthesia Over-estimation or under-
estimation
No staining
Estimation of IOP with irregular
corneas,where applanation
tonometry not possible.
 All clinical tonometers measure the IOP by relating a
deformation of the globe to the force responsible for
the deformation.

 In indentation tonometry, a known weight is placed on


the cornea, and the IOP is estimated by measuring the
deformation or indentation of the globe.

 Eg-Schiotz tonometer
Tonometer weight = 11g
scale

needle

Additional weights
7.5,10,15g
lever
Weight 5.5g
3mm diameter plunger
holder

ROC 15mm Foot plate


 The extent to which cornea is indented by plunger is
measured as the distance from the foot plate curve to
the plunger base and a lever system moves a needle on
calibrated scale.

 The indicated scale reading and the plunger weight are


converted to an IOP measurement.

 More the plunger indents the cornea, higher the


scale reading and lower the IOP

 Each scale unit represents 0.05 mm protrusion of


the plunger.
 When the plunger indents the cornea, the baseline
or resting pressure(Po) is artificially raised to a new
value(Pt).
 Change in pressure from Po to Pt  Expression of
resistance an eye offers to displacement of a volume of
fluid (Vc).

Pt=P0 + E(Scleral Rigidity)


 Because the tonometer actually measures Pt , it is
necessary to estimate Po for each scale reading &
weight.
 Friedenwald’s gave a mathematical formula.

 The formula has a single numerical constant, the coefficient


of ocular rigidity (K), which is roughly an expression of the
distensibility of the eye.Its average value is 0.025.

 He developed a nomogram for estimating K on the basis of


two tonometric readings with different weights

Log Pt2/Pt1 = K (V2-V1)

in which Pt1 , and V1 , represent the tonometric pressure and


the volume of the indentation caused by the bar in the
determination made with the first weight, whereas Pt2 and V2
represent the respective values as obtained with the second
weight.
 On the basis of this formula, he developed a set of
conversion tables for IOP.
Plunger Load
Scale Reading 5.5 g 7.5 g 10 g 15g
3.0 24.4 35.8 50.6 81.8
3.5 22.4 33.0 46.9 76.2
4.0 20.6 30.4 43.4 71.0
4.5 18.9 28.0 40.2 66.2
5.0 17.3 25.8 37.2 61.8
5.5 15.9 23.8 34.4 57.6
6.0 14.6 21.9 31.8 53.6
6.5 13.4 20.1 29.4 49.9
7.0 12.2 18.5 27.2 46.5
7.5 11.2 17.0 25.1 43.2
8.0 10.2 15.6 23.1 40.2
8.5 9.4 14.3 21.3 38.1
9.0 8.5 13.1 19.6 34.6
9.5 7.8 12.0 18.0 32.0
10.0 7.1 10.9 16.5 29.6
 Patient should be anasthetised with 4% lignocaine or 0.5% proparacaine

 With the patient in supine position, looking up at a fixation target while


examiner separates the lids and lowers the tonometer plate to rest on the
anesthetized cornea so that plunger is free to move vertically .

 Scale reading is measured.

 The 5.5 gm weight is initially used.

 If scale reading is 4 or less, additional weight is added to plunger.

 IOP measurement is repeated until 3 consecutive readings agree within


0.5 scale units.

 Conversion table is used to derive IOP in mm Hg from scale reading and


plunger weight.
OCULARRIGIDITY
 Measure of distensibility or resistance to deformation of ocular
coats.
 Important in indentation tonometer
 Increase in ocular rigidity  increase IOP
 Long standing glaucoma
 ARMD
 High Hyperopia
 Vasoconstrictors
 Decrease in ocular rigidity  decrease in IOP
 Increasing age
 Strong Miotic therapy
 Vasodilator therapy
 Post operative after RD surgery(Vitrectomy,
cryopexy,scleral band)
 High Myopia
 Compressible gas
ADVANTAGES DISADVANTAGES LIMITATIONS
1.Portable 1. Falsely high/low IOP 1. Instrumentalerrors
Ocular rigidity
2. Sturdy 2. Cannot be used in 2. Muscular contractions
traumatic cases/early  of extraocular muscles
post op cases/corneal raises IOP
diseases.  Accomodatio
n decreases
IOP
3. Relatively inexpensive 3. Variations in volumeof
globe
 Micropthalmos
 High myopia
 Buphthalmos
4. Easy to operate 4. Recorded in supine
position only.
5. Easy to clean &maintain 5. Reading also influenced
by the size of the footplate
hole and the thickness and
curvature of the cornea.
6. Does notrequire slit
lamp/power supply.
 A calibration check should be done at the start of
every day. Place the footplate of the instrument on
the rounded test block (the dummy cornea)
provided with the tonometer’s storage case.

 With the footplate resting on the test block, a


correctly calibrated instrument will have a scale
reading of zero.

 If not, you can calibrate it to zero.

 If the needle is to the left of zero, rotate the


footplate in a clockwise direction and check agai n.

 If the needle is to the right of the zero position,


rotate the footplate in an anti-clockwise direction .
 The tonometer is disassembled between each use and the
barrel is cleaned with 2 pipe cleaners, the first soaked in
isopropyl alcohol 70 % or methylated spirit and the
second dry.
 The foot plate is cleaned with alcohol swab.
 All surfaces must be dried before reassembling.
 The instrument can be sterilized with ultraviolet radiation,
steam, ethylene oxide.
 In between patients, the Schiotz tonometer should be
disinfected by soaking it in sodium hypochlorite.
 As with other tonometer tips, the Schiotz can be damaged by
some disinfecting solutions such as hydrogen peroxide and
bleach.
 It is done to get rid from ocular rigidity.
 A reading is taken with one weight on the Plunger and thena
second reading' is taken with a different weight.
 Making a diagnosis of glaucoma in a pt. with myopia presents
unusual difficulties. The low ocular rigidity in these eyes resultin
Schiotz readings within normal limits.

5.5g 10g Ocular rigidity IOP


18 mm Hg 15mm Hg lower >18
18 mm Hg 21mm Hg higher <18

18 mm Hg 18 mm Hg equal 18
 Has a continuous recording of IOP that is used for
tonography.

 ADVANTAGE:

 Scale is magnified, which makes it easier to detect


small changes in IOP.
 New and updated version of an indentation
tonometer.
PROCEDURE ADVANTAGES DISADVANTAGES
Sterile probe is propelled forward 1. Very light 1. Tend to read
into the cornea by a solenoid; the 2. Disposable slightly higher
time taken for the probe to 3. Probe is extremely than the
return to its resting position and light and itscontact Goldmann.
the characteristics of the rebound with the cornea is 2. Accuracy falls off
motion are indicative of the IOP very short , used in scarred
(and also the biomechanical without first corneas.
properties of the cornea) anesthetizing the
eye.
Time taken for the probe to 4.Comparable to the
return to its resting position is Goldmann in both
longer in eyes with lower IOP and normal and post-
faster in eyes with higher IOP. keratoplasty human
eyes.

Rebound tonometer does 5.Used in screening


correlate, like the Goldmann, with situations when
central corneal thickness. patients are unable to
be seated or measured
at the slitlamp.
(Haag Streit, Koeniz, Switzerland)
PARTS OF GOLDMANN APPLANATION
TONOMETER

Connects to the slit


lamp
Biprism
(measuring prism)

Feeder arm

Control weight insert


Housing
Adjusting knob
PRINCIPLE

 The concept was introduced by Goldmann is 1954.


 It is based on IMBERT FICKS LAW.
 It states that the pressure inside an ideal sphere (P) is equal
to force (F) necessary to flatten its surface divided by the
area of the flattening (A).

P=F/A
 P can be determined if
Force F is fixed or
Area A is fixed
 The ideal sphere is dry, thin-walled and flexible.

 The cornea is not ideal sphere.

 Two extra forces acting on cornea -


 Capillary attraction of tear meniscus (T), tends to pull
tonometer towards cornea
 Corneal rigidity (C) resists flattening

 Thus, Modified Imbert Ficks Law


F = PA , becomes

F + T = PA + C , or

P =( F + T - C) /A
 These two forces cancel each other when flattened area has
diameter of 3.06 mm.

 Applanation tonometry displaces only about 0.5 microlitre of


aqueous humor, which raises IOP by about 3%. Because the
volume displaced is so small, ocular rigidity, or the
‘stretchability’ of the globe, has little effect on the pressure
readings.
PROCEDURE

Patient is asked not to drink alcoholic beverages or large amounts of


fluid (e.g., 500 ml or more) for 2 hours before the test, as the former
will lower IOP and the latter may raise it.

Patient is told the purpose of the test and is reassured that the
measurement is not painful. The patient is instructed to relax,
maintain position, and hold the eyes open wide.

One drop of a topical anesthetic, such as 0.5% proparacaine, is


placed in each eye, and the tip of a moistened fluorescein strip is
touched to the tear layer on the inner surface of each lower lid.
Tonometer tip is cleaned with a sterilizing solution, and the tip and
prism are set in correct position on the slit lamp.

Tension knob is set at 1g. If the knob is set at 0, the prism head may vibrate
when it touches the eye and damage the corneal epithelium. The 1 g position
is used before each measurement. As a rule, it is more accurate to measure
IOP by increasing rather than decreasing the force ofapplanation.

0 graduation mark of the prism is set at the white line on the prism
holder.

Cobalt blue filter is used with the slit beam opened maximally. The
angle between the illumination and the microscope should be
approximately 60°.The room illumination is reduced.
Heights of the slit lamp, chair, and chin rest are adjusted until the
patient is comfortable and in the correct position for the measurement.

Palpebral fissure is a little wider if the patient looks up. However, the gaze
should be no more than 15° above the horizontal to prevent an elevationof IOP
that is especially marked in the presence of restrictive neuromuscular disease
such as dysthyroid ophthalmopathy.

Operator sits opposite the patient, the assembly is advanced towards the
patient with the tester observing from the side until the limbal zone has
a bluish hue.

If the tonometer tip touches the lids, the fluorescein rings will
thicken, which may cause an overestimation of IOP.
Clinician observes the applanation through the biprism at low power.
A monocular view is obtained of the central applanated zone and the
surrounding fluorescein-stained tear film.

Using the control stick, the observer raises, lowers, and centers the
assembly until two equal semicircles are seen in the center of the
field of view.

If the two semicircles are not equal in size, IOP is overestimated.The


clinician turns the tension knob in both directions to ensure that the
instrument is in good position.

If the semicircles cannot be made ‘too small,’ the instrument is too


far forward. If the semicircles cannot be made ‘too large,’ the
instrument is too far from the eye.
Fluorescein rings should be approximately 0.25–0.3mm in
thickness – or about one-tenth the diameter of the flattened area. If
the rings are too narrow, the patient should blink two or three times
to replenish the fluorescein.

If the fluorescein rings are too narrow, IOP is underestimated.


The tension knob is rotated until the inner borders of the fluorescein
rings touch each other at the midpoint of their pulsations.

Intraocular pressure is measured in the right eye until three


successive readings are within 1 mmHg. Intraocular pressure is then
measured in the left eye.

Reading obtained in grams is multiplied by 10 to give the IOP in


millimeters of mercury.
cont….

The fluorescein rings should be The fluorescent semicircles are viewed


approximately 0.25–0.3 mm in through the biprism and the force against
thickness – or about one-tenth the the cornea is adjusted until the inner
diameter of the flattened area. edges overlap.
FALSELY LOW IOP FALSELY HIGH IOP
Too little fluoroscein Too much fluoroscein
Thin cornea Thick cornea
Corneal edema Steep cornea
With the rule astigmatism Against the rule astigmatism
 1mm Hg per 4D  1 m Hg per 3D
Prolonged Cataract Wider meniscus
Repeated tonometry Widening the lid fissure
excessively
Elevating the eyes more than
150
 Wider meniscus or improper vertical alignment gives
higher IOP readings
 If the two semicircles are not equal in size, IOP is
overestimated.
 For every 3D increase in corneal curvature, IOP raises
about 1 mm Hg as more fluid is displaced under steeper
corneas causing increase in ocular rigidity
 More than 6 D astigmatism produces an elliptical area
on applanation that gives erroneous IOP. 4D with-the-
rule astigmatism underestimate IOP and 4D against-the-
rule astigmatism overestimate IOP.
 Mires may be distorted on applanating on irregular
corneas .
 Elevating the eyes more than 15° above the
horizontal causes an overestimation of IOP.

 Widening the lid fissure excessively causes an


overestimation of IOP

 Repeated tonometry reduces IOP, causing an


underestimation of the true level.This effect is greatest
between the first and second readings, but the trend
continues through a number of repetitions.

 A natural bias for even numbers may cause slight errors


in readings.
 THINNER cornea  less force to applanate 
Underestimation

 THICKER cornea  more force to applanate 


Overestimation

 Goldmann applanation tonometer was designed to give


accurate readings when the CCT was 550 μm.

The deviation of CCT from 550 μm yields a change in


applanation readings of 0.7 mm Hg per 10 μm.
 IOP measurements are also modified after PRK and
LASIK.

 Thinning of the central cornea is gives lower readings


on applanation.
Potential Sources of Error – During Measurement
Potential Sources of Error – During Measurement
Potential Sources of Error – During Measurement
Potential Sources of Error – During Measurement
Potential Sources of Error – During Measurement
Potential Sources of Error – During Measurement
Potential Sources of Error – During Measurement
Potential Sources of Error – During Measurement
Potential Sources of Error – During Measurement
Potential Sources of Error – During Measurement
ADVANTAGES DISADVANTAGES

Most accurate Need slit lamp

No indentation, so not much


Dark room
force is applied on cornea

Does not get affected by Need staining &cobalt


corneo-scleral rigidity light

Readings are directly from


knob

Can be done on post op


cases/injury cases
CALIBRATION

 Goldmann tonometer should be calibrated at least once


a month.

 If the Goldmann tonometer is not within 0.1g


(+1mmHg) of the correct calibration, the instrument
should be repaired; however, calibration errors of up to
+2.5 mmHg may still be tolerated clinically.
STERILIZATION

 Biprism should be rinsed and dried immediately after


use.

 Between uses, the prism head should be soaked in a


solution such as diluted bleach or 3% hydrogen
peroxide.

 70% ethanol and 70% isopropanol are effective as


sterilizing solutions but were shown in one study to
cause mild damage to the tonometer tip after one month
of immersion.
 Other methods of sterilization include: 10 min of
rinsing in running tap water, wash with soap and water,
cover the tip with a disposable film, and exposure to
UV light.

 It is possible to transfer bacteria, viruses, and other


infectious agents with the tonometer head, including
such potentially serious infections as epidemic
keratoconjunctivitis, hepatitis B, Jacob-Kreutzfeld and,
theoretically, acquired immunodeficiency syndrome.
 Care must be taken to be sure any sterilizing solution
has been completely rinsed off the tonometer tip, as
some of these solutions may be toxic to the corneal
epithelium, especially after LASIK or other corneal
procedures.

 If the tonometer tip is not mechanically wiped after


each use, epithelial cells may stick to the tip with the
small but serious risk of transmitting Jacob-Kreutzfeld
virus.

 Disposable tonometer tips may be an acceptable


alternative to soaking in, and wiping with, antiseptic
solutions.
PERKIN’STONOMETER

 Similar to the Goldmann tonometer.

 ADVANTAGES OVER
GOLDMANN TONOMETER:

 Portable and counterbalanced, so it


can be used in any position.

 Therefore,useful in a number of
situations, including in the operating
room, at the bedside, and with
patients who are obese or for other
reasons cannot be examined at the
slit lamp.
 Similar to the Goldmann and
Perkins tonometers.

 Except :

 Uses a different biprism.

 Force for applanation is supplied


by an electric motor.
MacKay-MargTonometer

 1.5 mm diameter
plunger

 Rigid spring

 Rubber sleeve.

 Movement of plunger
is electronically
monitored by a
transducer and
recorded on a moving
paper strip.
 USEFUL FOR MEASURING IOP IN EYES
WITH:

1. Scarred, irregular, or edematous corneas

2. Accurate when used over therapeutic soft contact


lenses.
TONOPEN

 Same principle as that of Mackay-Marg tonometer.


 Battery operated, portable.
 USES:
1. Community health fairs
2. Ward rounds
3. Children
4. Irregular surfaces
5. Measuring through an amniotic membrane patch graft.
 A disposable latex cover which is discarded after each use
provides infection control.
TONOPEN

• For pressures from 6 to 24 mmHg, it measured an average of


1.7 mm higher than the Goldmann tonometer.

• Above 24 mmHg, the readings were similar.


DISADVANTAGES Overestimate the IOP in infants so its
usefulness in congenital glaucoma
screening and monitoring is somewhat
limited.

In band keratopathy where the surface


of the pathology is harder than normal
cornea, the Tono-Pen tends to
overestimate the IOP.
PNEUMATICTONOMETER
 Principle is similar to the MacKay-Marg tonometer.

 Cornea is applanated by touching apex by silastic


diaphragm covering sensing nozzle.

 It is connected to central chamber containing


pressurized air.

 There is pneumatic to electronic transducer.

 It converts the air pressure to recording on paper strip


and IOP is read.
MAKLAKOV T ONOMETER

 Differs from the other


applanation instruments
 A known force is applied to
the eye, and the area of
applanation is measured – a
technique known as constant-
force rather than constant-area
applanation.
 The instrument consists of a
wire holder into which a flat-
bottom weight, ranging from
5 to 15g, is inserted.
 Surface of the weight is painted with a dye, such as
mild silver protein (Argyrol) mixed with glycerin, and
then the weight is lowered onto the cornea.

 During the procedure the patient is supine, and the


cornea is anesthetized.

 Weight is lifted from the cornea, and the area of


applanation is taken to be the area of missing dye,
which is measured either directly or indirectly from an
imprint on test paper.
 Intraocular pressure is inferred from the weight (W)
and the diameter of the area of applanation (d) by using
the following formula:

Pt= W /π(d/2)2
 Intraocular pressure is measured in grams per square
centimeter and is converted to millimeters of mercury
by dividing by 1.36.

DISADVANTAGE:
 Displaces a greater volume of aqueous humor than the other
applanation devices (but less than a schiøtz tonometer),
which means that the IOP readings are more influenced by
ocular rigidity.
OCUTONTONOMETER

 Hand-held tonometer.

 Works on the applanation principle.

 Probe is so light that it is barely felt and, therefore,


needs no anesthetic in most patients.

 Device is comparable to Goldmann tonometry but


tends to read higher than the Goldmann tonometer
when the cornea is thicker, and its accuracy may be
compromised by diurnal changes in corneal thickness.
NON-CONTACTTONOMETER
 Introduced by Grolman.
 Original NCT has 3 subsystems:
 1. Alignment system: It aligns patient’s eye in 3 dimensions.
 2. Optoelectronic applanation monitoring system:
 It comprises transmitter, receiver and detector, and timer.
 a. Transmitter directs a collimated beam of light at corneal
apex.
 b. Receiver and detector accept only parallel coaxial rays of
light reflected from cornea.
 c. Timer measures from an internal reference to the point of
peak light intensity.
3. Pneumatic system: It generates a puff of room air
directed against cornea.
PRINCIPLE
 A puff of room air creates a constant force that momentarily
flattens the cornea. The corneal apex is deformed by a jet of
air.

 The force of air jet which is generated by a solenoid


activated piston increases linearly over time.

 When the reflected light is at peak intensity, the cornea is


presumed to be flattened.

 The time elapsed is directly related to the force of jet


necessary to flatten the cornea and correspondingly to IOP.

 The time from an internal reference point to the moment of


flattening is measured and converted to IOP.
 A puff of air of known area is generated against cornea
(B).

 At the moment of corneal applanation,a light (T),


which is usually reflected from the normal cornea into
space, suddenly is reflected (R) into an optical sensor
(A).

 When the sensor is activated by the reflected light, the


air generator is switched off. The level of force at
which the generator stops is recorded, and a computer
calculates and displays the intraocular pressure.
ADVANTAGES LIMITATIONS
 Screening procedure  IOP is near normal,accuracy
decreases with increase in IOP &
in eyes with abnormal
cornea/poor fixation.
 Can be operated by non-medical
porsennel
 No anaesthesia required
 No direct contact between
instrument & eye
 New NCT, Pulsair is a portable hand held tonometer.
I-CARETONOMETER

 Handheld,portable & rebound tonometer.

 Turn on the unit by pressing the measurement button. It


will beep and then display LOAD.

 Place the single use probe into the collar and push the
measurement button again.

 When the I-care tonometer is ready to use, it is brought


to the patient’s eye with the central grove in the
horizontal position.
 The distance of the eye to
the tip of the probe is from
4-8 millimeters.

 Press the measurement


button and take no less
than six measurements.

 After the six


measurements the IOP
will be displayed.
TRANS-PALPEBRAL TONOMETER
PRINCIPLE USES LIMITATIONS
Pressure on the eyelid Young children Scleral rigidity
in most eyes
produces retinal
phosphenes.
Pressure on the eyelid Demented patients Thickness of the
required to induce and severely eyelids
these phosphenes is developmentally-
proportional to the challenged patients.
intraocular pressure.

Portable patients can Orbicularis muscle


measure their own tone
IOP at home

Potential intra
palpebral scarring
DYNAMIC CONTOUR TONOMETRY

Introduced by Kanngiesser
 Based on Pascal’s Law of Pressure, which states that
pressure applied to a confined fluid is transmitted
undiminished throughout the confining vessel of the
system.

 Concave shape of the tip generates minimal corneal


distortion and theoretically eliminates errors in
measuring IOP induced by ocular rigidity when the
cornea is applanated with a flat tipped tonometer.

 Operated in a fashion similar to a goldmann


applanation tonometer.
Pascal Dynamic Contour Tonometer (A) utilizes the
principle that when the contours of the cornea and
tonometer match, then the pressure measured at the
surface of the eye equals the pressure inside the eye (B).
 Microchip-enabled, solid-state sensor embedded within
the tip records 100 IOP measurements per second and
averages them over fluctuations in ocular pulse
amplitude.

 Digital display shows the final averaged IOP as well as


a Q-value that can be used objectively to judge the
quality of the final measurement.
OCULAR RESPONSEANALYZER

 Ocular response analyser (ORA) is a non-contact (air puff)


tonometer that does not require topical anaesthesia and provides
additional information on the biomechanical properties of the
cornea.

 Uses an air pulse to deform the cornea into a slight concavity.

 Measures not one but two pressures at which applanation


occurs:

1. When the air jet flattens the cornea as the cornea is bent
inward .

2. As the air jet lessens in force and the cornea recovers.


 The difference between
the pressures at which
the cornea flattens
inward and outward is
measured by the machine
and termed corneal
hysteresis (CH).

 The machine uses this


value to correct for the
effects of the cornea on
measurement.
 IOP correlate well with Goldmann tonometry but, on
average, measure a few millimeters higher.

 Congdon et al found that a ‘low’ hysteresis reading


with the ORA correlates with progression of
glaucoma, whereas thin central corneal thickness
correlates with glaucoma damage.
Tonometry for special clinicalcircumstances

• Pneumatic tonometer-Preferred
Irregular • Goldmann,NCT,Tonopen-limited accuracy
cornea

Soft contact
• Pneumo tonometry,Tonopen
lenses

Gas filled
• Pneumatic tonometry,Tonopen
eyes
• Goldmann,pneumotonometer,tonopen do
not correlate well with manometrically
FlatAC determined pressures.

• Tactile assessment
Keratoprostheses
Non-contact tonometer

Air-puff tonometer

continue

 non-contact
 only screening purpose
 It applanates the cornea by means of a jet of air.
 Once the instrument is properly aligned with the
patient's eye, a fixed distance separates the cornea
from the instrument.
 An optical system measures the time that it takes for
the air puff to flatten the cornea.
 This can be correlated with the IOP.
Continue

 Mean IOP readings compare favorably with


Goldmann tonometry

Advantage :The instrument is beneficial in mass


glaucoma screenings because it does not require
topical anesthetic and, with proper use, there is no
risk of injuring the cornea.
REFERENCES

 BECKER-SHAFFERS GLAUCOMA 8th EDITION

 SHIELDS GLAUCOMA 6th EDITION

 ANATOMY & PHYSIOLOGY OF EYE


A.K.KHURANA

 DIAGNOSTIC PROCEDURES IN
OPHTHALMOLOGY -BY HV NEMA, NITIN NEMA

You might also like