INSTALLATION/SERVICE MANUAL
Did you find this document useful?
Our EU Authorised Representative:-
Medical Systems Ltd. JMC Medical Sprl,
S.C.O. 212-214, SEC. 34-A, 58, Avenue du Parc de Woluwe
CHANDIGARH (INDIA) B-1160 Bruxelles
TEL.: +91-172-3012287, 3012285, Belgium
FAX: +91-172-2621912 Tel: +32.2.672.56.44
E-mail:
[email protected] Fax: +32.2.662.29.92
Website: www.allengers.com E-mail:
[email protected]Is this content inappropriate? Report this Document
We Aim For Your Delight
PREFACE
This Installation and Service Manual of Allengers 325 X-Ray Machine
covers all aspects of Installation and Servicing of the Machine.
The Manual covers the Installation Procedure & checkpoints, Circuit
Diagram, its Explanation, Calibration Procedure, Trouble Shooting and
major parts list.
Although all efforts have been made to keep this Manual simple to
understand, in case of any clarification, Allengers International Customer
Care Cell can be contacted for more detailed information on any aspect.
It is assumed that anyone referring to this manual and attempting to service
the X-Ray Machine has relevant Technical background and Training to
service this Equipment.
This Manual is company’s confidential document and is meant to be
circulated only to company’s Service Engineers and authorized Service
Dealers.
Photocopying and free circulation of this document is prohibited.
Contents
CONTENTS
1. System Overview………………………………………………….. 1
1.1 Introduction…………………………………………………… 1
1.2 Technical Specifications……………………………………… 4
1.3 Description of Control………………………………………... 5
2. Packing List………………………………………………………... 9
3. Unpacking Instructions…………………………………………… 10
4. Infrastructural Requirements for Installation………………….. 11
4.1 Supply and Wiring Requirements……………………………... 11
4.2 Earthing Requirement…………………………………………. 11
4.3 Suggested Room Plan…………………………………………. 12
4.4 Air Conditioning………………………………………………. 14
4.5 Environmental Conditions…………………………………….. 15
4.6 Room Layout Approval from BARC/AERB………………….. 15
5. Installation Procedure…………………………………………….. 16
5.1 Inter Connections of Various Assemblies……………………... 16
5.2 Assembly Procedure…………………………………………... 18
5.3 Installation Checkpoints………………………………………. 20
6. Circuit Diagram & Circuit Explanation…………………………. 23
7. Calibration & Testing…..…………………………………………. 31
7.1 Energizing the Equipment…………………………………….. 31
7.2 Carrying out Preliminary Tests………………………………... 31
7.3 Calibrating the Equipment…………………………………….. 32
7.4 Procedure for carrying out QA Tests as per Regulatory Board 43
8. Troubleshooting…………………………………………………… 44
9. Post Installation Instructions……………………………………... 52
Allengers 325 Installation/Service Manual
Rev. ‘0’ Eff. Date: 12.02.2011
Contents
10. Dark Room Procedures & X-Ray Film Processing……………. 53
11. Radiation Monitoring & Protection…………….………………. 54
12. Preventive Maintenance Procedure…………………………….. 55
13. Preventive Maintenance Check Points………………………….. 56
Checked By__________________________
Allengers 325 Installation/Service Manual
Rev. ‘0’ Eff. Date: 12.02.2011
Approvals
This machine is type approved for Mechanical, Electrical & Radiation safety standards by following
regulatory boards:
PRODUCT APPROVALS
Approved for meeting Radiation Safety
Standards by Atomic Energy Regulatory
Board (AERB)
Approved for meeting Mechanical &
Electrical Safety Standards by Bureau of
Indian Standards (BIS)
CE Certified; Approved for meeting
International Safety Standards
SYSTEM APPROVALS
ISO 9001:2008 & ISO 13485:2003 Approved
for meeting Quality Management System
Requirements
_________________________________________________________________________________
Allengers 325 Installation/Service Manual
Rev. ‘0’ Eff. Date: 12.02.11
System Overview
1. SYSTEM OVERVIEW
1.1 INTRODUCTION
Allengers 325 is an all-purpose high power Diagnostic X-Ray Generator, which caters to all
the requirements of generalized and specialized Radiography when used with Manual or
motorized table. This generator is also available in various options like Fluoro/Spot Filming
capabilities. This can also be made available for R+F Tables having I.I. T.V. System.
This unit incorporates State - of – the- Art technology and is designed keeping in mind the
diverse requirements of the Radiologists. It is a Solid State Silicon Rectified Unit (Two
Pulse) and consists of Control Panel, High Voltage Tank, High Voltage Cables & X-Ray
Tube Head with a collimator.
This unit can be supplied in different combinations of Tube Stands, Tables & Vertical Bucky
Stand as per individual customer requirement.
This machine is Type Approved for Electrical and Mechanical Safety Standards and for the
Radiation Safety (vide IS: 7620) from the Atomic Energy Regulatory Board (AERB), Govt.
of India, thus making it entirely safe from the Radiation Hazards.
Allengers 325 Installation/Service Manual Page 1
Rev. ‘0’ Eff. Date: 12.02.11
Trusted by over 1 million members
Try Scribd FREE for 30 days to access over 125
million titles without ads or interruptions!
Start Free Trial
Cancel Anytime.
System Overview
The unit consists of following modules:
1. CONTROL
2. H.V. TANK*
Allengers 325 Installation/Service Manual Page 2
Rev. ‘0’ Eff. Date: 12.02.11
System Overview
3. X-RAY TUBE
4. H.V. Tank is also available for two Tubes operation.*
5. IMAGE INTENSIFIER*
6. TROLLEY WITH MONITORS*
7. TUBE STAND*
8. TABLE*
9. VERTICAL BUCKY STAND*
* OPTIONALLY AVAILABLE AS PER CUSTOMER REQUIREMENT (REFER INSTALLATION
MANUAL OF RESPECTIVE SYSTEM)
Allengers 325 Installation/Service Manual Page 3
Rev. ‘0’ Eff. Date: 12.02.11
System Overview
1.2 TECHNICAL SPECIFICATIONS
Parameters Specifications
Output (As per IS: 7620) 30 KW
Max. mA output 300 mA
Radiographic KVP 35–125 KVP
(Selectable in Steps by Major and Minor KVP
Selector Switch)
Fluoroscopy KVP* 45-100 KVP
(Selectable in steps by Selector Switch)
Radiographic Timer 0.01 to 5 Seconds (in 24 Steps)
Fluoroscopy Timer* Commulative timer of 5 Minutes
Radiographic mA 50 mA, 100 mA at Small Focus
200 mA, 300 mA at Large Focus
Fluoroscopic mA* Up to 3 mA maximum
Electronic Overload Simultaneous protection from excess selection of
mA, Time, KVP and input voltage.
Voltage Compensator Switch For input voltage variations
Tube Head with Collimator Indian/Imported Rotating Anode, Dual Focus 20/40
KW X-Ray Tube Unit with Light Beam Device
(Manual or Motorized)
Small Focus: 1.0 mm Square
Large Focus: 2.0 mm Square
H.V. Cables One pair of High Voltage Cables of length 6 meters
(Diff. Lengths available on request)
Power Supply 440 Volts, AC, 50/60 Hz. 3 Phase Max. Allowable
Line Regulation ± 10 %
Digital Displays KVP, FmA & mAS
Visual Indicators X-Ray ON, Live ON, Overload
*OPTIONALLY AVAILABLE AS PER CUSTOMER REQUIREMENT
Allengers 325 Installation/Service Manual Page 4
Rev. ‘0’ Eff. Date: 12.02.11
System Overview
1.3 DESCRIPTION OF CONTROL
16 17 18 19 20
14 15
1 13
2 12
3 11
4 5 6 7 8 9 10
Allengers 325 Installation/Service Manual Page 5
Rev. ‘0’ Eff. Date: 12.02.11
System Overview
Description of Control Functions
1. Fluoroscopic KVP Selector:
Used to select Fluoroscopic KVP from 45 to 100 KVP in steps of 5 KVP. This
function works only in machines provided with Fluoroscopy mode.
2. Voltage Compensator Switch:
Used to adjust Voltage on Voltmeter in fine steps.
3. Fluoroscopy/Radiography Mode Selection Switch:
Used to select either Radiographic or Fluoroscopic / Spot Filming Mode
4. Ready Switch:
Used to start Rotor & Filament Boosting.
5. X-Ray Switch:
X-Ray switch when pressed keeping the Ready switch pressed starts exposure.
Exposure terminates after preset time.
6. Horizontal/Vertical Bucky Selection Switch:
When we switched ON the machine “HORIZONTAL” Bucky will energize by
default. If we press “VER” switch then “VERTICAL” Bucky will select.
7. ON Switch/OFF Switch:
These Switches are used to energize/de-energize the equipment. By pressing ON
Switch equipment get energizes and by pressing OFF Switch it de-
energizes.
Allengers 325 Installation/Service Manual Page 6
Rev. ‘0’ Eff. Date: 12.02.11
System Overview
8. Fluoro mA Increase/Decrease Switch (Only in Machines with
Fluoroscopy option)
When (+) switch is pressed, Fluoro mA increases & when (-) switch is pressed,
Fluoro mA decreases.
9. Fluoro Timer Reset Switch:
This switch is used to reset the Fluoro timer after completing 300 seconds of Fluoro
exposure cycle.
10. X-Ray Indicator:
This indicator glows when X-Ray exposure is taking place.
11. Overload Indicator:
Glows when selected parameters viz. KVP, mA, Time or input voltage exceed Tube
Rating Chart.
12. Radiographic Minor KVP Selection Switch:
Used to adjust Radiographic KVP in fine steps.
13. Radiographic Timer Switch:
Used to select radiographic time from 0.01 to 5 sec. in 24 steps.
14. Technic Selector Switch:
Used to select Radiographic mA from 50, 100, 200 or 300 mA.
Allengers 325 Installation/Service Manual Page 7
Rev. ‘0’ Eff. Date: 12.02.11
System Overview
15. Radiographic Major KVP Selection Switch:
Used to adjust Radiographic KVP in coarse steps.
16. Voltmeter: 230 V.A.C.:
Indicates Input Voltage to the machine.
17. Fluoro mA Digital Display:
Displays Fluoroscopic mA. This function works only in machines provided with
Fluoroscopy. In Radiography machines, display remains blank.
18. mAs Digital Display:
Displays mAs as per mA selected from Technic selector and Time selected from
Timer switch.
19. Radiographic KVP Display:
Displays active Radiographic KVP.
20. mA Meter:
Indicates Tube Current during exposure.
Exposure
Ready
A. HAND SWITCH
This is used to press Ready switch and Exposure switch using retractable type cord