MI 3155 - EurotestXD ANG Ver 1.12.16 20752716
MI 3155 - EurotestXD ANG Ver 1.12.16 20752716
MI 3155
Instruction manual
Version 1.12.16, Code no. 20 752 716
Distributor:
Manufacturer:
Metrel d.d.
Ljubljanska cesta 77
SI-1354 Horjul
Slovenia
https://www.metrel.si
[email protected]
It is the user’s responsibility to ensure the integrity and security of the data carrier and to
regularly backup and validate the integrity of backups of the data. METREL HAS NO
OBLIGATION OR LIABILITY FOR ANY LOSS, ALTERATION, DESTRUCTION, DAMAGE,
CORRUPTION OR RECOVERY OF USER DATA, REGARDLESS OF WHERE THE DATA IS
STORED.
Hereby, Metrel d.d. declares that the MI 3155 is in compliance with Directive
2014/53/EU (RED) and all other subjected EU directive. The full text of the EU
declaration of conformity is available at the following internet address
https://www.metrel.si/DoC.
© 2022 METREL
The trade names Metrel®, Smartec®, Eurotest®, Auto Sequence® are trademarks registered in Europe and other
countries.
No part of this publication may be reproduced or utilized in any form or by any means without
permission in writing from METREL.
2
i. About the instruction manual
This Instruction manual contains detailed information on the EurotestXD, its key
features, functionalities and use.
It is intended for technically qualified personnel responsible for the product and its use.
Please note that LCD screenshots in this document may differ from the actual instrument
screens in details due to firmware variations and modifications.
Metrel reserve the right to make technical modifications without notice as part of the
further development of the product.
3
MI 3155 EurotestXD Table of Contents
Table of Contents
1 General description .......................................................................................................... 9
1.1 Warnings and notes .................................................................................................... 9
1.1.1 Safety warnings ....................................................................................................... 9
1.1.2 Markings on the instrument ................................................................................... 10
1.1.3 Warnings related to safety of batteries .................................................................. 10
1.1.4 Warnings related to safety of measurement functions ........................................... 10
1.1.5 Notes related to measurement functions ............................................................... 11
1.2 Testing potential on PE terminal ................................................................................ 14
1.3 Battery and charging of Li-ion battery pack ............................................................... 16
1.3.1 Precharge.............................................................................................................. 17
1.3.2 Li – ion battery pack guidelines ............................................................................. 19
1.4 Standards applied ..................................................................................................... 20
2 Instrument set and accessories..................................................................................... 22
2.1 Standard set MI 3155 EurotestXD ............................................................................. 22
2.1.1 Optional accessories ............................................................................................. 22
3 Instrument description ................................................................................................... 23
3.1 Front panel ................................................................................................................ 23
3.2 Connector panel ........................................................................................................ 24
3.3 Back side .................................................................................................................. 25
3.4 Carrying the instrument ............................................................................................. 27
3.4.1 Secure attachment of the strap.............................................................................. 27
4 Instrument operation ...................................................................................................... 29
4.1 General meaning of keys .......................................................................................... 29
4.2 General meaning of touch gestures........................................................................... 30
4.3 Virtual keyboard ........................................................................................................ 31
4.4 Display and sound ..................................................................................................... 32
4.4.1 Terminal voltage monitor ....................................................................................... 32
4.4.2 Battery indication ................................................................................................... 33
4.4.3 Bluetooth ............................................................................................................... 33
4.4.4 Measurement actions and messages .................................................................... 33
4.4.5 Result indication .................................................................................................... 35
4.4.6 Auto Sequence® result indication .......................................................................... 35
4.5 Instruments main menu ............................................................................................. 37
4.6 General Settings ....................................................................................................... 38
4.6.1 Language .............................................................................................................. 39
4.6.2 Power Save ........................................................................................................... 39
4.6.3 Date and time ........................................................................................................ 40
4.6.4 Workspace manager ............................................................................................. 40
4.6.5 Auto Sequences groups ..................................................................................... 40
4.6.6 User accounts ....................................................................................................... 40
4.6.7 Profiles .................................................................................................................. 45
4.6.8 Settings ................................................................................................................. 45
4.6.9 Devices ................................................................................................................. 50
4.6.10 Bluetooth initialization ........................................................................................ 50
4.6.11 Initial Settings .................................................................................................... 50
4.6.12 About ................................................................................................................. 51
4.7 Instrument profiles ..................................................................................................... 52
4
MI 3155 EurotestXD Table of Contents
5
MI 3155 EurotestXD Table of Contents
7.16 Z line 4W – Line impedance and prospective short-circuit current ........................... 145
7.17 Z line m – High precision line impedance and prospective short-circuit current .... 147
7.18 High Current (MI 3143 and MI 3144) ....................................................................... 150
7.19 Voltage Drop ........................................................................................................... 152
7.20 U touch – Touch voltage (MI 3143 and MI 3144) ..................................................... 155
7.21 Z auto - Auto test sequence for fast line and loop testing ........................................ 157
7.22 R line m – DC resistance measurement (MI 3144) ............................................... 159
7.23 ELR Current Injection Test (MI 3144) ...................................................................... 161
7.24 ELR Combination Time Test (MI 3144) ................................................................... 163
7.25 EVSE Diagnostic Test (A 1632) .............................................................................. 165
7.26 Earth – Earth resistance (3-wire test) ...................................................................... 168
7.27 Earth 2 clamp – Contactless earthing resistance measurement (with two current
clamps) ............................................................................................................................... 169
7.28 Ro – Specific earth resistance (A 1199) .................................................................. 170
7.29 Power ...................................................................................................................... 172
7.30 Harmonics ............................................................................................................... 174
7.31 Currents .................................................................................................................. 176
7.32 Current Clamp Meter (MI 3144) ............................................................................... 178
7.33 ISFL – First fault leakage current............................................................................. 180
7.34 IMD – Testing of insulation monitoring devices........................................................ 182
7.35 Rpe – PE conductor resistance ............................................................................... 186
7.36 Illumination .............................................................................................................. 188
7.37 Discharging time ..................................................................................................... 189
7.38 AUTO TT – Auto test sequence for TT earthing system .......................................... 192
7.39 AUTO TN (RCD) – Auto test sequence for TN earthing system with RCD............... 194
7.40 AUTO TN – Auto test sequence for TN earthing system without RCD ..................... 196
7.41 AUTO IT – Auto test sequence for IT earthing system............................................. 198
7.42 Locator .................................................................................................................... 200
7.43 Visual and Functional inspections ........................................................................... 202
7.44 Measurements using adapter MD 9273 ................................................................... 203
7.44.1 Power CLAMP ................................................................................................. 203
7.44.2 Voltage CLAMP ............................................................................................... 205
7.44.3 Current CLAMP ............................................................................................... 206
7.44.4 Inrush CLAMP ................................................................................................. 207
7.44.5 Harmonics U CLAMP ...................................................................................... 209
7.44.6 Harmonics I CLAMP ........................................................................................ 211
8 Auto Sequences® ......................................................................................................... 213
8.1 Selection of Auto Sequences® ................................................................................ 213
8.1.1 Selecting an active Auto Sequence® group in Auto Sequences® menu.............. 213
8.1.2 Searching in Auto Sequences® menu ................................................................. 214
8.1.3 Organization of Auto Sequences® in Auto Sequences® menu ........................... 216
8.2 Organization of an Auto Sequence® ....................................................................... 216
8.2.1 Auto Sequence® view menu ............................................................................... 217
8.2.2 Step by step execution of Auto Sequences® ....................................................... 219
8.2.3 Auto Sequence® result screen ............................................................................ 221
8.2.4 Auto Sequence® memory screen ........................................................................ 222
9 Communication ............................................................................................................. 224
9.1 USB and RS232 communication ............................................................................. 224
9.2 Bluetooth communication with Android devices ....................................................... 224
9.3 Communication with adapters ................................................................................. 225
9.4 Bluetooth and RS-232 communication with scanners .............................................. 226
10 Upgrading the instrument ............................................................................................ 227
6
MI 3155 EurotestXD Table of Contents
7
MI 3155 EurotestXD Table of Contents
8
MI 3155 EurotestXD General description
1 General description
1.1 Warnings and notes
9
MI 3155 EurotestXD General description
Hazardous voltages exist inside the instrument. Disconnect all test leads, remove
the power supply cable and switch off the instrument before removing battery
/fuse compartment cover.
Do not connect any voltage source on C1 inputs. It is intended only for connection
of current clamps. Maximal input voltage is 3 V!
Read the Instruction manual with special care to safety operation«. The
symbol requires an action!
10
MI 3155 EurotestXD General description
Diagnostic test
If any insulation resistance values (RISO(15 s) or RISO(60 s )) are over-ranged the DAR
factor is not calculated. The result field is blank: DAR:_____!
If any insulation resistance values (RISO(60 s) or RISO(10 min)) are over-ranged the PI
factor is not calculated. The result field is blank: PI :_____!
11
MI 3155 EurotestXD General description
conductor – this can cause misunderstanding, because PRCD reacts at half the tripping
fault current. In order to prevent this, use parameter Sensitivity set to 'Ipe monitoring',
where test current will be a half of selected nominal tripping current.
If Sensitivity parameter is set to 'standard', a test current with nominal tripping current
is used.
The a.c. part of MI and EV RCDs is tested as general (non-delayed) RCDs.
The d.c. part of MI and EV RCDs is tested with a d. c. test current The Pass limit is
between 0.5 and 1.0 IdNDC.
The Zs rcd function takes longer to complete but offers much better accuracy of fault loop
resistance (in comparison to the RL sub-result in Contact voltage function).
Auto test is finished without x5 tests in case of testing the RCD types A, F, B and B+ with
rated residual currents of IdN = 300 mA, 500 mA, and 1000 mA or testing the RCD type
AC with rated residual current of IdN = 1000 mA . In this case Auto test result passes if all
other results pass, and indications for x5 are omitted.
Auto test is finished without x1 tests in case of testing the RCD types B and B+ with rated
residual currents of IdN = 1000 mA. In this case Auto test result passes if all other results
pass, and indications for x1 are omitted.
Tests for sensitivity Idn(+) and Idn(-) are omitted for selective type RCD.
Trip out time measurement for B and B+ type RCDs in AUTO function is made with sine-
wave test current, while trip-out current measurement is made with DC test current.
Illumination
LUXmeter type B (A 1172) and LUXmeter type C (A 1173) probes are supported by the
instrument.
Artificial light sources reach full power of operation after a period of time (see technical
data for light sources) and should be therefore switched on for this period of time before
the measurements are taken.
12
MI 3155 EurotestXD General description
For accurate measurement make sure that the milk glass bulb is lit without any shadows
cast by hand, body or other unwanted objects.
Refer to the Illuminance handbook for more information.
Rpe
The specified accuracy of tested parameters is valid only if the mains voltage is stable
during the measurement.
Measurement will trip an RCD if the parameter RCD is set to ‘No’.
The measurement does not normally trip an RCD if the parameter RCD is set to ‘Yes’.
However, the RCD can trip if a leakage current from L to PE already flows.
IMD
It is recommended to disconnect all appliances from the tested supply to receive regular
test results. Any connected appliance will influence the insulation resistance threshold
test.
Discharging time
Interpretation of the ‘Repeat’ message:
It is not possible to differentiate between a disconnection moment at very low voltage and
a machine with a very low discharging time. In both cases the reading will be 0.0 s
together with the “Repeat” warning. If after few repetitions the result is always 0.0 s with
the “Repeat” message it can be considered as a valid 0.0 s result.
A 0.0 s reading without the “Repeat” message is a valid result.
Residual voltage is not displayed for DUTs with very low discharging time, or if detected
disconnection voltage is lower than 20 V.
Auto Sequences®
Metrel Auto Sequences® are designed as guidance to tests in order to significantly
reduce testing time, improve work scope and increase traceability of the tests performed.
METREL assumes no responsibility for any Auto Sequence® by any means. It is the
user’s responsibility, to check adequacy for the purpose of use of the selected Auto
Sequence®. This includes type and number of tests, sequence flow, test parameters and
limits.
Auto Sequences® mode allows building custom-defined test sequences.
See notes related to single tests in the selected Auto Sequence®.
Compensate test leads resistance before entering Auto Sequences®.
Zref value for Voltage drop test (ΔU) implemented in any Auto Sequence® should be set
in single test function.
13
MI 3155 EurotestXD General description
check the installation against this fault the key should be used as an indicator prior to
performing live tests.
Warning!
14
MI 3155 EurotestXD General description
Test procedure
Notes:
PE test terminal is active in the Voltage, Rpe, RCD tests, Z loop, Zs rcd, Z auto, Z line,
ΔU, AUTO TT, AUTO TN, AUTO TN (rcd), AUTO IT tests and Auto Sequences only!
In case of detection of phase voltage on PE terminal in IT earthing system, the tests can
be enabled/disabled according to setting of parameter ‘Ignore PE probe warning (IT)’.
For correct testing of PE terminal, the key has to be touched for at least 1 second.
Make sure that the TEST key is grounded through human body resistance without any
insulated material between (gloves, shoes, insulated floors, pens, ...). PE test could
otherwise be impaired and results of a single test or Auto Sequence® can mislead. Even
detected dangerous voltage on PE test terminal cannot prevent running of a single test or
Auto Sequence®. All such behaviour is regarded as misuse. Operator of the instrument
must stop the activity immediately and eliminate the fault/connection problem before
proceeding with any activity!
15
MI 3155 EurotestXD General description
Symbol:
Indication of low battery.
The battery is charged whenever the power supply is connected to the instrument. The charger
socket polarity is shown in Figure 1.4. Internal circuit controls (CC, CV) charging and assures
maximum battery lifetime. Nominal operating time is declared for battery with nominal capacity of
4.4 Ah.
The instrument automatically recognizes the connected power supply and begins charging.
Symbol:
Indication of battery charging
Figure 1.5: Charging indication
(animation)
16
MI 3155 EurotestXD General description
Typical charging profile which is also used in this instrument is shown in Figure 1.6.
ICH
Cha
C u r rg e
re n
t
VLOWV
ICH/8
Precharge
Fastcharge Safety Time
Time
where:
VREG ................................... Battery charging voltage
VLOWV.................................. Precharge threshold voltage
ICH ...................................... Battery charging current
ICH/8 .................................... 1/8 of the charging current
1.3.1 Precharge
On power up, if the battery voltage is below the VLOWV threshold, the charger applies 1/8 of the
charging current to the battery. The precharge feature is intended to revive deeply discharged
battery. If the VLOWV threshold is not reached within 30 minutes of initiating precharge, the
charger turns off and a FAULT is indicated.
Figure 1.7: Battery fault indication Figure 1.8: Battery full indication
(charging suspended, timer fault, battery (charging completed)
absent)
Note:
As a safety backup, the charger also provides an internal 5-hour charge timer for fast
charge.
17
MI 3155 EurotestXD General description
Typical charging time is 3 hours (Battery type: 18650T22A2S2P) in the temperature range of
5°C to 60°C.
Current
Charge Charge
Charge
Suspended Suspended
Ich
Ich/8
Temperature
TWARM
TCOOL
THTF
TLTF
where:
TLTF ........... Cold temperature threshold (typ. -15°C)
TCOOL ......... Cool temperature threshold (typ. 0°C)
TWARM ........ Warm temperature threshold (typ. +60°C)
THTF ........... Hot temperature threshold (typ. +75°C)
The charger continuously monitors battery temperature. To initiate a charge cycle, the battery
temperature must be within the TLTF to THTF thresholds. If battery temperature is outside of this
range, the controller suspends charge and waits until the battery temperature is within the TLTF to
THTF range.
If the battery temperature is between the TLTF and TCOOL thresholds or between the TWARM and
THTW thresholds, charge is automatically reduced to ICH/8 (1/8 of the charging current).
18
MI 3155 EurotestXD General description
Storage
Charge or discharge the instruments battery pack to approximately 50 % of capacity
before storage.
Charge the instrument battery pack to approximately 50 % of capacity at least once
every 6 months.
Transportation
Always check all applicable local, national, and international regulations before
transporting a Li – ion battery pack.
Handling Warnings
19
MI 3155 EurotestXD General description
20
MI 3155 EurotestXD General description
21
MI 3155 EurotestXD Instrument set and accessories
22
MI 3155 EurotestXD Instrument description
3 Instrument description
3.1 Front panel
23
MI 3155 EurotestXD Instrument description
Test connector
L/L1 pin – In 4-wire measurements used as a current probe C1.
1 N/L2 pin – In 4-wire measurements used as a current probe C2.
PE/L3 pin – In 4-wire measurements used as a voltage probe P2.
S pin – In 4-wire measurements used as a voltage probe P1.
2 Protection cover
3 Protection cover – PS/2 communication port
Charger socket
4
Warnings!
Maximum allowed voltage between any test terminal and ground is 550 V!
Maximum allowed voltage between test terminals on test connector is 550 V!
Maximum allowed voltage on test terminal C1 is 3 V!
Maximum short-term voltage of external power supply adapter is 14 V!
24
MI 3155 EurotestXD Instrument description
Type: 18650T22A2S2P
1 Li-ion battery pack
Type: 18650T22A2S4P (optional)
2 Fuse F1 M 315 mA / 250 V
3 Fuses F2 and F3 F 5 A / 500 V (breaking capacity 50 kA)
25
MI 3155 EurotestXD Instrument description
4 SD card slot
26
MI 3155 EurotestXD Instrument description
27
MI 3155 EurotestXD Instrument description
28
MI 3155 EurotestXD Instrument operation
4 Instrument operation
The EurotestXD instrument can be manipulated via a keypad or touch screen.
29
MI 3155 EurotestXD Instrument operation
30
MI 3155 EurotestXD Instrument operation
Backspace
Clears last character or all characters if selected.
(If held for 2 s, all characters are selected).
31
MI 3155 EurotestXD Instrument operation
4-wire test terminal indication with online voltage between P1 and P2 probes.
32
MI 3155 EurotestXD Instrument operation
Charging finished.
4.4.3 Bluetooth
33
MI 3155 EurotestXD Instrument operation
Use MI 3143 Euro Z 440 V, MI 3144 Euro Z 800 V or A 1143 Euro Z 290 A
adapter for this test.
Use A 1172 or A 1173 Illumination sensor for this test.
34
MI 3155 EurotestXD Instrument operation
High resistance to earth of potential and current test probes. Results may be
impaired.
Too small current for declared accuracy. Results may be impaired. Check in
Current Clamp Settings if sensitivity of current clamp can be increased.
In Earth 2 Clamp measurement results are very accurate for resistances
below 10 Ω. At higher values (several 10 Ω) the test current drops to few mA.
The measuring accuracy for small currents and immunity against noise
currents must be considered!
Measured signal is out of range (clipped). Results are impaired.
Fuse F1 is broken.
35
MI 3155 EurotestXD Instrument operation
36
MI 3155 EurotestXD Instrument operation
Options
Single Tests
Menu with single tests, see chapter 6 Single tests.
Auto Sequences
Menu with customized test sequences, see chapter 8 Auto
Sequences®.
Memory Organizer
Menu for working with and documentation of test data, see chapter 5
Memory Organizer.
General Settings
Menu for setup of the instrument, see chapter 4.6 General Settings.
37
MI 3155 EurotestXD Instrument operation
Options
Language
Instrument language selection.
Power Save
Brightness of LCD, enabling/disabling Bluetooth communication.
Date /Time
Instruments Date and time.
Workspace Manager
Manipulation with project files. Refer to chapter 4.8 Workspace Manager
for more information.
User accounts
User accounts settings. Refer to chapter 4.6.6 User accounts for more
information.
Instrument Profile
Selection of available instrument profiles. Refer to chapter 4.7 Instrument
profiles for more information.
Settings
Refer to chapter 4.6.8 Settings for more information.
Devices
Selection of external devices. Refer to chapter 4.6.9 Devices for more
information.
38
MI 3155 EurotestXD Instrument operation
Bluetooth initialization
Reset Bluetooth settings.
Initial Settings
Factory settings.
About
Instrument info.
4.6.1 Language
In this menu the language of the instrument can be set.
39
MI 3155 EurotestXD Instrument operation
Note:
If the batteries are removed the set date and time will be lost.
If a user account is set and the user is signed in the user's name will be stored in memory for
each measurement.
Individual users can change their passwords.
40
MI 3155 EurotestXD Instrument operation
4.6.6.1 Signing in
If signing in is demanded the user must enter the password in order to work with the instrument.
Options
User signing in
41
MI 3155 EurotestXD Instrument operation
Administrator signing in
Options
Signs out the set user.
42
MI 3155 EurotestXD Instrument operation
Options
The Account manager menu is accessed by selecting
Account manager in Sign in menu or User profile menu.
43
MI 3155 EurotestXD Instrument operation
Options
Opens the window for adding a new user.
44
MI 3155 EurotestXD Instrument operation
Blackbox password can be set by administrator from the Account manager menu. Set Blackbox
password is valid for all users. Default Blackbox password is empty (disabled).
Options
4.6.7 Profiles
Refer to chapter 4.7 Instrument profiles for more information.
4.6.8 Settings
In this menu different general parameters can be set.
45
MI 3155 EurotestXD Instrument operation
46
MI 3155 EurotestXD Instrument operation
line measurement.
Load pretest [yes, no] [yes]: Load pretest is enabled in Riso and
Riso-all measurements.
[no]: Load pretest is disabled.
External Device [None, Commander, [None]: This option is intended to disable the
A 1507] commander’s remote keys. In case of high EM
interfering noise the operation of the
commander can be irregular.
[Commander]: Working with commander
enabled.
[A 1507]: Working with 3-phase active switch
enabled.
Limit Uc [Custom, 12 V, 25 V, Contact voltage limit.
50 V]
4.6.8.1 Adapters
Settings menu provide selection and configuration of Metrel measuring adapters to perform
supported tests and measurements. See Appendix G for details on available Metrel adapters
and supported tests.
Maximum RCD disconnection times differ in various standards. The trip-out times defined in
individual standards are listed below.
47
MI 3155 EurotestXD Instrument operation
Notes:
Trip-out limit times for PRCD, PRCD-K and PRCD-S are equal to General (non-delayed)
RCDs.
The trip-out times of VDE 0664 are equal to trip-out times of EN 61008 / EN 61009.
The trip-out times of VDE 0100-410 TN/IT are equal to trip-out times of IEC 60364-4-41
TN/IT.
The trip-out times of VDE 0100-410 TT are equal to trip-out times of IEC 60364-4-41 TT.
48
MI 3155 EurotestXD Instrument operation
Supported standards:
IEC 62752:2016 In-cable control and protection device for mode 2 charging of electric road
vehicles (IC-CPD)
IEC 62955:2018 Residual direct current detecting device (RDC-DD) to be used for mode 3
charging of electric vehicles
6 mA 60 mA 200 mA 300 mA
Standard
(1 IN) (10 IN) (33 IN) (50 IN)
IEC 629551) < 10.0 s < 300 ms < 100 ms
IEC 627522) < 10.0 s < 300 ms < 40 ms
Table 4.7: Break times for d.c. residual currents
1) IEC 62955: Table 2 – Maximum values of break times for residual direct currents
2) IEC 62752: Table 3 – Limit values of break time for smooth d.c. residual currents
30 mA 60 mA 150 mA
Standard
(1 IN) (2 IN) (5 IN)
IEC 627523) < 300 ms < 150 ms < 40 ms
Table 4.8: Break times for a.c. residual currents
3) IEC 62752: Table 2 – Limit values of break time for a.c. residual currents at rated
frequency
Up to 30 mA 60 mA 150 mA
Standard
(1 IN) (2 IN) (5 IN)
IEC 629554) No tripping > 300 ms > 80 ms
Table 4.9: Non-operating times for a.c. residual currents
4) IEC 62955: Table 3 – Minimum values of non-operating time for alternating residual
currents (RMS values)
49
MI 3155 EurotestXD Instrument operation
4.6.9 Devices
In this menu operation with external devices is configured.
Reading devices
Type Sets appropriate reading device (QR or barcode scanner, RFID
reader, via aMESM application).
Port Sets communication port of selected reading device.
Bluetooth device Goes to menu for pairing with selected Bluetooth device.
name
50
MI 3155 EurotestXD Instrument operation
Warning!
Following customized settings will be lost when setting the instruments to initial settings:
measurement limits and parameters,
global parameters, system settings, and Devices in General settings menu,
opened Workspace and Auto Sequences group will be deselected,
user will be signed out.
Note:
Following customized settings will stay:
profile settings,
data in memory (Data in memory organizer, Workspaces, Auto Sequences groups,
Auto Sequences) and
user accounts.
4.6.12 About
In this menu instrument data (name, serial number, firmware (FW) and hardware (HW) version,
FW profile, hardware documentation (HD) version, fuse version and date of calibration) can be
viewed.
Note:
Adapter info is also displayed, if connected.
51
MI 3155 EurotestXD Instrument operation
The instrument uses different specific system and measuring settings in regard to the scope of
work or country it is used. These specific settings are stored in instrument profiles.
By default, each instrument has at least one profile activated. Proper licence keys must be
obtained to add more profiles to the instruments.
Options
52
MI 3155 EurotestXD Instrument operation
Options
List of Workspaces.
List of Exports.
53
MI 3155 EurotestXD Instrument operation
Options
Marks the opened Workspace in Memory Organizer.
Opens the selected Workspace in Memory Organizer.
Refer to chapter 4.8.6 Opening a Workspace for more information.
54
MI 3155 EurotestXD Instrument operation
Options
55
MI 3155 EurotestXD Instrument operation
56
MI 3155 EurotestXD Instrument operation
57
MI 3155 EurotestXD Instrument operation
58
MI 3155 EurotestXD Instrument operation
Folders with lists of Auto Sequences® are stored in Root\__MOS__\AT on the SD card.
Opens the selected list of Auto Sequences. Previously selected list of Auto
Sequences will be closed automatically.
Refer to chapter 4.9.1.1 Selecting a list of Auto Sequences for more information.
59
MI 3155 EurotestXD Instrument operation
Procedure
Note:
Previously selected list of Auto
Sequences is closed automatically.
Procedure
60
MI 3155 EurotestXD Memory Organizer
5 Memory Organizer
Memory Organizer is a tool for storing and working with test data.
A measurement can be a Single test or an Auto Sequence. For more information refer to
chapters 7 Tests and measurements and 8 Auto Sequences®.
61
MI 3155 EurotestXD Memory Organizer
Overall status of measurements under each structure element /sub-element can be seen
without spreading tree menu. This feature is useful for quick evaluation of test status and as
guidance for measurements.
62
MI 3155 EurotestXD Memory Organizer
Options
Note:
There is no status indication if all measurement results under each structure element /
sub-element have passed or if there is an empty structure element / sub-element
(without measurements).
63
MI 3155 EurotestXD Memory Organizer
Procedure
64
MI 3155 EurotestXD Memory Organizer
Procedure
Note:
To change the name of a Node, see chapter
5.1.5.15 Rename a Structure object
65
MI 3155 EurotestXD Memory Organizer
Options
Saves a measurement.
Saving of measurement on a position after the selected (empty or finished)
measurement.
66
MI 3155 EurotestXD Memory Organizer
Deletes a measurement.
Selected Measurement can be deleted. User is asked for confirmation before the
deleting. Refer to chapter 5.1.5.14 Delete a measurement for more information.
Options
67
MI 3155 EurotestXD Memory Organizer
68
MI 3155 EurotestXD Memory Organizer
The parameters and their content are displayed in this menu. To edit the selected parameter,
Procedure
Note:
Some structure parameters include
custom lists. See chapter 5.1.5.4 Custom
lists of Structure parameter values for
details.
69
MI 3155 EurotestXD Memory Organizer
Attachments
a
The name of attachment can be seen.
Operation with attachments is not
supported in the instrument.
Some structure parameters include option to create custom lists of values for particular
structure parameter. These custom values can be re-used easily, by selecting it from the
custom lists, and re-typing is therefore not required.
Procedure
70
MI 3155 EurotestXD Memory Organizer
71
MI 3155 EurotestXD Memory Organizer
This menu is intended to add new structure objects in the tree menu. A new structure object can
be selected and then added in the tree menu.
Procedure
72
MI 3155 EurotestXD Memory Organizer
73
MI 3155 EurotestXD Memory Organizer
In this menu new empty measurements can be set and then added in the structure tree. The
type of measurement, measurement function and its parameters are first selected and then
added under the selected Structure object.
Procedure
c
74
MI 3155 EurotestXD Memory Organizer
75
MI 3155 EurotestXD Memory Organizer
In this menu selected structure object can be copied (cloned) to same level in the structure tree.
Cloned structure object has the same name as the original.
Procedure
76
MI 3155 EurotestXD Memory Organizer
By using this function, a selected empty or finished measurement can be copied (cloned) as an
empty measurement to the same level in the structure tree. Parameters and limits of new
measurement are the same as they are set in original measurement. Its parameters/limits can
be changed when measurement is started.
Procedure
77
MI 3155 EurotestXD Memory Organizer
In this menu selected Structure object can be copied and pasted to any allowed location in the
structure tree.
Procedure
78
MI 3155 EurotestXD Memory Organizer
Note
The Paste command can be executed one
or more times.
When structure object is selected to be cloned, or copied & pasted, additional selection of its
sub-elements is needed. The following options are available:
Options
Parameters of selected structure object will be cloned /
pasted too.
Attachments of selected structure object will be cloned
/ pasted too.
Structure objects in sub-levels of selected structure
object will be cloned / pasted too.
Measurements in selected structure object and sub-
levels will be cloned / pasted too.
79
MI 3155 EurotestXD Memory Organizer
In this menu selected measurement can be copied as an empty measurement to any allowed
location in the structure tree. Selected measurement can be copied multiple times to different
locations in the structure tree. Parameters and limits of new measurement are the same as they
are set in original measurement. Its parameters/limits can be changed when measurement is
started.
Procedure
Note
The Paste command can be executed
one or more times.
80
MI 3155 EurotestXD Memory Organizer
In this menu selected Structure object with sub-items (sub-structures and measurements) can
be cut and pasted (moved) to any allowed location in the structure tree.
Procedure
81
MI 3155 EurotestXD Memory Organizer
Procedure
82
MI 3155 EurotestXD Memory Organizer
In this menu selected measurement can be deleted from selected location in the tree structure.
Procedure
83
MI 3155 EurotestXD Memory Organizer
Procedure
84
MI 3155 EurotestXD Memory Organizer
Procedure
Measurement is recalled.
85
MI 3155 EurotestXD Memory Organizer
86
MI 3155 EurotestXD Memory Organizer
Procedure
87
MI 3155 EurotestXD Memory Organizer
Figure 5.8: Search results screen (left), structure object selected (right)
Options
Next page (if available).
Attachments.
Name and link of attachment is displayed.
88
MI 3155 EurotestXD Memory Organizer
Views comment.
The instrument displays comment attached to the selected Structure object.
Note:
Search result page consist of up to 50 results.
89
MI 3155 EurotestXD Single tests
6 Single tests
Single tests can be selected in the main Single tests menu or in Memory organizer main
menu and sub-menus.
Options
Area Group
Groups
90
MI 3155 EurotestXD Single tests
Cross selector
Last used
91
MI 3155 EurotestXD Single tests
Header line:
ESC touch key
function name
battery status
real time clock
Control panel (available options)
Result field:
main result(s)
sub-result(s)
PASS / FAIL indication
Voltage monitor with info and warning symbols
92
MI 3155 EurotestXD Single tests
on
93
MI 3155 EurotestXD Single tests
on
Note:
Set parameters, set limits and entered user comments are saved to the memory. When
same Single test is used next time, settings and comments will remain the same.
Figure 6.2: Single test start screen, example of insulation resistance measurement
Options (before test, screen was opened in Memory organizer or Single test main menu):
Starts the measurement.
long
Opens help screens.
94
MI 3155 EurotestXD Single tests
long on
Expands column in control panel.
Previous value.
Next value.
95
MI 3155 EurotestXD Single tests
Figure 6.4: Single test results screen, example of insulation resistance measurement
results
long
Saves the result.
96
MI 3155 EurotestXD Single tests
on
Adds comment to the measurement. The
instrument opens keypad for entering a comment.
Opens help screens.
97
MI 3155 EurotestXD Single tests
Plot edit
Opens control panel for editing graphs.
98
MI 3155 EurotestXD Single tests
Legend
1 Overall status of the inspection
2 Selected inspection
3 Item
4 Child items
5 Status fields (for items and child items)
6 Control panel (available options)
Options (inspection screen was opened in Memory organizer or from Single test main
menu)
99
MI 3155 EurotestXD Single tests
Opens help screens. Refer to chapter 6.1.8 Help screens for more
information.
100
MI 3155 EurotestXD Single tests
The child item(s) will automatically get a status on base of status in the parent item.
All child items will get the same status as applied to the parent item.
Notes:
Inspections and even inspection items inside one inspection can have can have different
status types. For example, some basic inspections don’t have the ‘checked’ status.
Only inspections with overall statuses can be saved.
101
MI 3155 EurotestXD Single tests
Adds comment to the measurement. The instrument opens keypad for entering a
comment.
Opens help screens. Refer to chapter 6.1.8 Help screens for more information.
Options
Retest
Enters screen with ‘empty’ measurement.
102
MI 3155 EurotestXD Single tests
Options
Opens help screen.
on
103
MI 3155 EurotestXD Single tests
Options
Retest
Enters starting screen for a new measurement.
Refer to chapter 6.1.3 Single test start screen for
more information.
Opens menu for viewing parameters and limits.
Refer to chapter 6.1.2 Setting parameters, limits
and comments of single tests for more information.
on
Expands column in control panel.
104
MI 3155 EurotestXD Tests and measurements
Measurement parameters
105
MI 3155 EurotestXD Test and measurements
6)
In case of 1-phase voltage system and limit type se to %.
7)
In case of 3-phase voltage system and limit type set to voltage.
8)
In case of 3-phase voltage system and limit type set to %.
Connection diagrams
Figure 7.2: Connection of 3-wire test lead and optional adapter in three-phase system
Figure 7.3: Connection of Plug commander and 3-wire test lead in single-phase system
106
MI 3155 EurotestXD Test and measurements
Measurement procedure
107
MI 3155 EurotestXD Test and measurements
Uiso Nominal test voltage [50 V, 100 V, 250 V, 500 V, 1000 V, 2500 V]
Type of test [-, L/PE, L/N, N/PE, L/L, L1/L2, L1/L3, L2/L3, L1/N, L2/N, L3/N,
Type Riso1)
L1/PE, L2/PE, L3/PE]
Limit(Riso) Min. insulation resistance [Off, Custom, 0.01 M ... 100 M]
1)
Insulation measurement depends on Type Riso parameter setting, see table below.
Type Riso 3-wire test lead and Tip commander 2.5 kV test lead
parameter measuring terminals measuring terminals
(UN ≤ 1 kV) (UN =2.5 kV)
-
L-N
Lx-N L and N
L/L
HV+ and HV-
Lx-Ly
L-PE
L and PE
Lx-PE
N-PE N and PE
Table 7.1: Insulation resistance measuring terminals and Type Riso parameter
dependency
108
MI 3155 EurotestXD Test and measurements
Connection diagrams
Figure 7.7: Connections of 3-wire test lead and Tip commander (UN ≤ 1 kV)
109
MI 3155 EurotestXD Test and measurements
Measurement procedure
Start the measurement. A longer press on key or a longer press on ‘Start test’
option on touch screen starts a continuous measurement.
Stop the measurement. Wait until object under test is fully discharged.
Save results (optional).
110
MI 3155 EurotestXD Test and measurements
111
MI 3155 EurotestXD Test and measurements
Connection diagrams
Measurement procedure
112
MI 3155 EurotestXD Test and measurements
𝑅𝐼𝑆𝑂 (1 min)
𝐷𝐴𝑅 =
𝑅𝐼𝑆𝑂 (15 s)
PI (Polarization Index) is the ratio of insulation resistance values measured after 1 minute and
after 10 minutes. The DC test voltage is present during the whole period of the measurement
For additional information regarding PI and DAR diagnostic, please refer to Metrel’s handbook
Modern insulation testing.
Connection diagrams
Figure 7.15: Connection of 3-wire test lead and Tip commander (UN ≤ 1 kV)
113
MI 3155 EurotestXD Test and measurements
Measurement procedure
114
MI 3155 EurotestXD Test and measurements
A voltage ramp starts from 50 V and rises with a slope of 100 V/s (Range parameter set to 1000
V) or 350 V/s (Range parameter set to 2500 V). The measurement ends when the defined end
voltage is reached or if the test current exceeds the value of 1 mA.
Figure 7.19: Connection of 3-wire test lead and Tip commander (Range: 1000 V)
115
MI 3155 EurotestXD Test and measurements
Measurement procedure
116
MI 3155 EurotestXD Test and measurements
𝑈𝑑𝑐
𝑈𝑎𝑐 ≈
1.15 × √2
Uac voltage may be directly compared with the voltage declared on tested protection
device.
Connection diagram
Figure 7.23: Connection of 3-wire test lead plus optional Extension lead
117
MI 3155 EurotestXD Test and measurements
Measurement procedure
R Resistance
R+ Result at positive test polarity
R- Result at negative test polarity
7.7 R low 4W
118
MI 3155 EurotestXD Test and measurements
Connection diagram
Figure 7.26: Connection of 4-wire test lead plus optional extension leads
Measurement procedure
R Resistance
R+ Result at positive test polarity
R- Result at negative test polarity
119
MI 3155 EurotestXD Test and measurements
Connection diagrams
120
MI 3155 EurotestXD Test and measurements
Measurement procedure
R Resistance
121
MI 3155 EurotestXD Test and measurements
122
MI 3155 EurotestXD Test and measurements
Contact voltage,
Trip-out time,
Trip-out current and
RCD Auto test.
123
MI 3155 EurotestXD Test and measurements
information.
Earthing system Refer to chapter 4.6.8 Settings for more information.
1)
Parameter is available only when parameter Use is set to other (for Electrical Vehicle
(EV) RCDs/RCMs and Mobile installations (MI) RCDs).
2)
Parameter is available only when RCD I or RCD t test is selected and parameter Use is
set to ‘other’.
3)
Parameter is available only when parameter ‘Use’ is set to PRCD, PRCD-3p, PRCD-S+
or PRCD-K.
Connection diagrams
Figure 7.34: Connecting the Plug commander and the 3-wire test lead
Contact voltage measurement is performed before trip-out time/current tests. If limit voltage
(e.g. 50 V) is reached during this pre-test, trip-out test is aborted for safety reason.
Contact voltage measurement can also be performed using external test probe. Place external
probe to reference earth potential.
For connection see Figure 7.35 Connection for Uc(P) measurement.
124
MI 3155 EurotestXD Test and measurements
Before the contact voltage measurement make sure that Uc(P) parameter is set to On.
Test procedure
The contact voltage results Uc and Uc(P) relate to the rated nominal residual current of the
RCD and are multiplied by an appropriate factor (depending on RCD type and type of test
current). The 1.05 factor is applied to avoid negative tolerance of result. See Table 7.3 for
detailed contact voltage calculation factors.
Fault Loop resistance is indicative and calculated from Uc result (without additional proportional
𝑈
factors) according to: 𝑅𝐿 = 𝐼 𝐶 .
∆𝑁
Uc Contact voltage
Uc(P) – if selected Contact voltage, external probe
Rl Fault loop resistance
125
MI 3155 EurotestXD Test and measurements
t ∆N Trip-out time
Uc Contact voltage for rated IN
Slope range
RCD type Waveform
Start value End value
AC 0.2IN 1.1IN Sine
IEC 62752:
0.2IN 1.0IN Sine
EV RCD, EV RCM, MI RCD (a.c. part)
IEC 62955:
0.2IN 1.0IN Sine
EV RCD, EV RCM, MI RCD (a.c. part)
A, F (IN 30 mA) 0.2IN 1.5IN
Pulsed
A, F (IN = 10 mA) 0.2IN 2.2IN
B, B+ 0.2IN 2.2IN DC
IEC 62752:
1.2 mA 6.0 mA DC
EV RCD, EV RCM, MI RCD (d.c. part)
IEC 62955:
1.2 mA 6.0 mA DC
EV RCD, EV RCM, MI RCD (d.c. part)
Table 7.4: Relation between RCD type, slope range and test current
Maximum test current is I (trip-out current) or end value in case the RCD didn’t trip-out.
126
MI 3155 EurotestXD Test and measurements
Test procedure
I∆ Trip-out current
Uc Contact voltage
Uc I∆ Contact voltage at trip-out current I∆ or no value if the RCD didn’t trip
t I∆ Trip-out time at trip-out current I∆
127
MI 3155 EurotestXD Test and measurements
RCD’, trip-out times or non-operating times for a.c. residual current are measured according to IEC
62752 or IEC 62955.
128
MI 3155 EurotestXD Test and measurements
Step 1 Step 2
Step 3 Step 4
Step 5 Step 6
Step 7 Step 8
129
MI 3155 EurotestXD Test and measurements
Step 9 Step 10
Step 11 Step 12
Figure 7.39: Individual steps in RCD Auto test, example on testing EV RCD
130
MI 3155 EurotestXD Test and measurements
Connection diagrams
131
MI 3155 EurotestXD Test and measurements
Measurement procedure
Z Loop impedance
Ipsc Prospective fault current
Ulpe Voltage L-PE
R Resistance of loop impedance
XL Reactance of loop impedance
Uc (P) Contact voltage at prospective fault current (external probe)
𝑈n ∙ 𝑘sc
𝐼PSC =
𝑍
132
MI 3155 EurotestXD Test and measurements
where:
𝑈n ......... Nominal UL-PE voltage (see table below),
𝑘sc ........ Correction factor (Isc factor) for 𝐼PSC . Refer to chapter 4.6.8 Settings for more
information.
Uc(P) .... Voltage between external earthed point and main earthing point (P/S and PE terminals),
see calculation below
Uc(P) calculation
𝑈C (P) = 𝑍PE−P/S × 𝐼PSC
133
MI 3155 EurotestXD Test and measurements
Connection diagram
Measurement procedure
134
MI 3155 EurotestXD Test and measurements
Z Loop impedance
Ipsc Prospective fault current
Ulpe Voltage L-PE
R Resistance of loop impedance
XL Reactance of loop impedance
𝑈n ∙ 𝑘sc
𝐼PSC =
𝑍
where:
𝑈n ......... Nominal UL-PE voltage (see table below),
𝑘sc ........ Correction factor (Isc factor) for 𝐼PSC . Refer to chapter 4.6.8 Settings for more
information.
135
MI 3155 EurotestXD Test and measurements
136
MI 3155 EurotestXD Test and measurements
Connection diagrams
Measurement procedure
137
MI 3155 EurotestXD Test and measurements
Z Loop impedance
Ipsc Prospective fault current
Ulpe Voltage L-PE
Uc 1) Contact voltage at nominal residual current
Uc (P) Contact voltage at prospective fault current (external probe) 2)
Contact voltage at nominal residual current (external probe) 3)
R Resistance of loop impedance
XL Reactance of loop impedance
1)
Result is presented only if Protection is set to TTrcd.
2)
Protection type parameter set to TN.
3)
Protection type parameter set to TTrcd.
𝑈n ∙ 𝑘sc
𝐼PSC =
𝑍
where:
𝑈n ......... Nominal UL-PE voltage (see table below),
𝑘sc ........ Correction factor (Isc factor) for IPSC. Refer to chapter 4.6.8 Settings for more
information.
Uc(P) .... Voltage between external earthed point and main earthing point (P and PE terminals),
see calculation below
Uc(P) calculation
138
MI 3155 EurotestXD Test and measurements
Connection diagram
139
MI 3155 EurotestXD Test and measurements
Measurement procedure
Z Loop impedance
Ipsc Standard prospective fault current
Imax Maximal prospective fault current
Imin Minimal prospective fault current
Ub Contact voltage at maximal prospective fault current (contact voltage measured against
Probe S if used)
R Resistance of loop impedance
XL Reactance of loop impedance
140
MI 3155 EurotestXD Test and measurements
Refer to A 1143 – Euro Z 290 A, MI 3143 – Euro Z 440 V and MI 3144 – Euro Z 800 V
Instruction manuals for detailed information.
141
MI 3155 EurotestXD Test and measurements
Connection diagram
142
MI 3155 EurotestXD Test and measurements
Measurement procedure
Z Line impedance
Ipsc Prospective short-circuit current
Uln Voltage measured between L and N test terminals
R Resistance of line impedance
XL Reactance of line impedance
Imax3p Maximal three-phases prospective short-circuit current
Imin3p Minimal three-phases prospective short-circuit current
Imax2p Maximal two-phases prospective short-circuit current
Imin2p Minimal two-phases prospective short-circuit current
Imax Maximal single-phase prospective short-circuit current
Imin Minimal single-phase prospective short-circuit current
𝑈n ∙ 𝑘sc
𝐼PSC =
𝑍
where:
𝑈n ......... Nominal UL-N or UL-L voltage (see table below),
𝑘sc ........ Correction factor (Isc factor) for 𝐼PSC . Refer to chapter 4.6.8 Settings for more
information.
143
MI 3155 EurotestXD Test and measurements
The prospective short-circuit currents IMin, IMin2p, IMin3p and IMax, IMax2p, IMax3p are calculated as
follows:
2
𝐶min 𝑈n(L−N) 𝑍(L−N)hot = √(1.5 × 𝑅(L−N) )2 + 𝑋(L−N)
𝐼min = where
𝑍(L−N)hot 0.95; 𝑈n(L−N) = 230 V ± 10 %
𝐶min = {
1.00; otherwise
2 2
𝐶max 𝑈n(L−N) 𝑍(L−N) = √𝑅(L−N) + 𝑋(L−N)
𝐼max = where
𝑍(L−N) 1.05; 𝑈n(L−N) = 230 V ± 10 %
𝐶max = {
1.10; otherwise
2
𝐶min 𝑈n(L−L) 𝑍(L−L)hot = √(1.5 × 𝑅(L−L) )2 + 𝑋(L−L)
𝐼min2p = where
𝑍(L−L)hot 0.95; 𝑈n(L−L) = 400 V ± 10 %
𝐶min = {
1.00; otherwise
2 2
𝐶max 𝑈n(L−L) 𝑍(L−L) = √𝑅(L−L) + 𝑋(L−L)
𝐼max2p = where
𝑍(L−L) 1.05; 𝑈n(L−L) = 400 V ± 10 %
𝐶max = {
1.10; otherwise
2
𝐶min × 𝑈n(L−L) 2 𝑍(L−L)hot = √(1.5 × 𝑅(L−L) )2 + 𝑋(L−L)
𝐼min3p = where
√3 𝑍(L−L)hot 0.95; 𝑈n(L−L) = 400 V ± 10 %
𝐶min = {
1.00; otherwise
2 2
𝐶max × 𝑈n(L−L) 2 𝑍(L−L) = √𝑅(L−L) + 𝑋(L−L)
𝐼max3p = where
√3 𝑍(L−L) 1.05; 𝑈n(L−L) = 400 V ± 10 %
𝐶max = {
1.10; otherwise
144
MI 3155 EurotestXD Test and measurements
Connection diagram
145
MI 3155 EurotestXD Test and measurements
Measurement procedure
Z Line impedance
Ipsc Prospective short-circuit current
Uln Voltage measured between C1 and C2 terminals
R Resistance of line impedance
XL Reactance of line impedance
𝑈n ∙ 𝑘sc
𝐼PSC =
𝑍
where:
𝑈n ......... Nominal L-N or L-L voltage (see table below),
𝑘sc ........ Correction factor for Isc. Refer to chapter 4.6.8 Settings for more information.
146
MI 3155 EurotestXD Test and measurements
147
MI 3155 EurotestXD Test and measurements
Connection diagram
Measurement procedure
148
MI 3155 EurotestXD Test and measurements
Z Line impedance
Ipsc Standard prospective short-circuit current
Imax Maximal prospective short-circuit current
Imin Minimal prospective short-circuit current
Imax2p Maximal two-phases prospective short-circuit current
Imin2p Minimal two-phases prospective short-circuit current
Imax3p Maximal three-phases prospective short-circuit current
Imin3p Minimal three-phases prospective short-circuit current
R Resistance of line impedance
XL Reactance of line impedance
Refer to A 1143 – Euro Z 290 A, MI 3143 – Euro Z 440 V and MI 3144 – Euro Z 800 V
Instruction manuals for detailed information.
149
MI 3155 EurotestXD Test and measurements
Connection diagram
Measurement procedure
150
MI 3155 EurotestXD Test and measurements
R Resistance
Rsel1) Resistance (calculated from Clamp current)
Itest Test current
Ic1) Clamp current
U Voltage
f Frequency
U Voltage dip
U% Voltage dip in percentage [ΔU (%) = (ΔU / U_unloaded) x 100 %]
1)
Measurement with current clamps is supported by MI 3144 – Euro Z 800 V instrument only.
Voltage monitor:
Up1p2 Voltage P1-P2
Uc1c2 Voltage C1-C2
Freq Frequency
Refer to MI 3143 – Euro Z 440 V and MI 3144 – Euro Z 800 V Instruction manuals for
detailed information.
151
MI 3155 EurotestXD Test and measurements
Connection diagram
Figure 7.68: Voltage drop measurement – connection of Plug commander and 3-wire test
lead
152
MI 3155 EurotestXD Test and measurements
Measurement procedure
153
MI 3155 EurotestXD Test and measurements
U Voltage drop
Ipsc Prospective short-circuit current
Uln Voltage L-N
Zref Reference line impedance
Z Line impedance
(𝑍 − 𝑍𝑅𝐸𝐹 ) ⋅ 𝐼𝑁
𝛥𝑈[%] = ⋅ 100
𝑈𝑁
where:
154
MI 3155 EurotestXD Test and measurements
Connection diagram
Refer to MI 3143 – Euro Z 440 V and MI 3144 – Euro Z 800 V Instruction manual for detailed
information.
Measurement procedure
Connect MI 3155 instrument with MI 3143 or MI 3144 Euro Z instrument via serial
RS232 or pair them using Bluetooth communication. See chapter 4.6.8.1 Adapters.
Enter the U touch function.
Set test parameters / limits.
Check Bluetooth communication active sign if MI 3143 or MI 3144 Euro Z instrument
is connected to MI 3155 instrument via Bluetooth communication.
Connect test leads and A 1597 adapter to MI 3143 or MI 3144 Euro Z instrument.
Connect test leads to the object under test.
Refer to MI 3143 Euro Z 440 V or MI 3144 Euro Z 800 V Instruction manual for
detailed information.
155
MI 3155 EurotestXD Test and measurements
Voltage monitor:
Up1p2 Voltage P1-P2
Uc1c2 Voltage C1-C2
Freq Frequency
Refer to MI 3143 Euro Z 440 V and MI 3144 Euro Z 800 V Instruction manuals for detailed
information.
156
MI 3155 EurotestXD Test and measurements
7.21 Z auto - Auto test sequence for fast line and loop
testing
Tests / measurements implemented in Z auto test sequence
Voltage
Z line
Voltage Drop
Zs rcd
Uc
157
MI 3155 EurotestXD Test and measurements
Connection diagram
Measurement procedure
158
MI 3155 EurotestXD Test and measurements
Connection diagram
Measurement procedure
Connect MI 3155 instrument with MI 3144 Euro Z 800 V instrument via serial RS232
or pair them using Bluetooth communication. See chapter 4.6.8.1 Adapters.
Enter the R line m function.
Set test parameters / limits.
Check Bluetooth communication active sign if MI 3144 Euro Z 800 V instrument is
connected to MI 3155 instrument via Bluetooth communication.
Connect test leads to MI 3144 Euro Z 800 V instrument.
Connect test leads to the object under test.
Refer to MI 3144 Euro Z 800 V Instruction manual for detailed information.
159
MI 3155 EurotestXD Test and measurements
R Line resistance
Itest Test current
Udc Voltage
U Voltage drop
U% Voltage drop in percentage
Voltage monitor:
Up1p2 Voltage P1-P2
Uc1c2 Voltage C1-C2
Freq Frequency
160
MI 3155 EurotestXD Test and measurements
Connection diagram
Figure 7.81: ELR Current Injection Test / Combination Time Test connection
Measurement procedure
Connect MI 3155 instrument with MI 3144 Euro Z 800 V instrument via serial
RS232 or pair them using Bluetooth communication. See chapter 4.6.8.1
Adapters.
Enter the ELR Current Injection Test function.
Set test parameters / limits.
Check Bluetooth communication active sign if MI 3144 Euro Z 800 V instrument is
connected to MI 3155 instrument via Bluetooth communication.
Connect test leads to MI 3144 Euro Z 800 V instrument.
Connect test leads to the object under test. See Figure 7.81.
161
MI 3155 EurotestXD Test and measurements
Measurement result
I Current
Voltage monitor:
Up1p2 Voltage P1-P2
Uc1c2 Voltage C1-C2
Freq Frequency
162
MI 3155 EurotestXD Test and measurements
Connection diagram
Measurement procedure
Connect MI 3155 instrument with MI 3144 Euro Z 800 V instrument via serial RS232
or pair them using Bluetooth communication. See chapter 4.6.8.1 Adapters.
Enter the ELR Combination Time Test function.
Set test parameters / limits.
Check Bluetooth communication active sign if MI 3144 Euro Z 800 V instrument is
connected to MI 3155 instrument via Bluetooth communication.
Connect test leads to MI 3144 Euro Z 800 V instrument.
Connect test leads to the object under test. See Figure 7.81.
Refer to MI 3144 Euro Z 800 V Instruction manual for detailed information.
163
MI 3155 EurotestXD Test and measurements
Measurement result
t Time
Voltage monitor:
Up1p2 Voltage P1-P2
Uc1c2 Voltage C1-C2
Freq Frequency
164
MI 3155 EurotestXD Test and measurements
Figure 7.85: Diagnostic Test (EVSE) start screens – EV simulator, Errors and Monitor
Connection diagrams
Figure 7.86: Diagnostic test - EV simulator and Errors sub-tests - connection to EVSE
Figure 7.87: Diagnostic test - EV simulator and Errors sub-tests - connection to Mode 2
charging cable powered from Analyser
165
MI 3155 EurotestXD Test and measurements
Figure 7.88: Diagnostic test (EVSE) - Monitor sub-test - connection to EVSE or charging
cable
Pair and connect MI 3155 with A 1632 eMobility Analyzer instrument via Bluetooth
communication. See chapter 4.6.8.1 Adapters.
Enter the Diagnostic test (EVSE) function.
Set test parameters / limits.
Check Bluetooth communication active sign if A 1632 eMobility Analyzer is
connected to MI 3155 instrument via Bluetooth communication.
Connect the charging cable / station to A 1632 eMobility Analyzer adapter. See
Figure 7.86, Figure 7.87 and Figure 7.88.
Refer to A 1632 eMobility Analyzer Instruction manual for detailed information.
166
MI 3155 EurotestXD Test and measurements
167
MI 3155 EurotestXD Test and measurements
Connection diagrams
Figure 7.91: Resistance to earth, measurement of main installation earthing and lighting
protection system
Measurement procedure
Enter the Earth function.
Set test parameters / limits.
Connect test cable to the instrument.
Connect 3-wire test lead to the object under test, see Figure 7.91.
Start the measurement.
Save results (optional).
168
MI 3155 EurotestXD Test and measurements
Re Earth resistance
Rc Resistance of H (current) probe
Rp Resistance of S (potential) probe
Connection diagram
Measurement procedure
169
MI 3155 EurotestXD Test and measurements
Re Earth resistance
Connection diagram
170
MI 3155 EurotestXD Test and measurements
Measurement procedure
171
MI 3155 EurotestXD Test and measurements
7.29 Power
Connection diagram
Measurement procedure
172
MI 3155 EurotestXD Test and measurements
P Active power
S Apparent power
Q Reactive power (capacitive or inductive)
PF Power factor (capacitive or inductive)
THDu Voltage total harmonic distortion
173
MI 3155 EurotestXD Test and measurements
7.30 Harmonics
Connection diagram
Measurement procedure
174
MI 3155 EurotestXD Test and measurements
175
MI 3155 EurotestXD Test and measurements
7.31 Currents
Connection diagram
Measurement procedure
176
MI 3155 EurotestXD Test and measurements
177
MI 3155 EurotestXD Test and measurements
Connection diagram
Measurement procedure
Connect MI 3155 instrument with MI 3144 Euro Z instrument via serial RS232 or pair
them using Bluetooth communication. See chapter 4.6.8.1 Adapters.
Enter the Current Clamp Meter function.
Set test parameters / limits.
Check Bluetooth communication active sign if MI 3144 Euro Z 800 V instrument is
connected to MI 3155 instrument via Bluetooth communication.
Connect current clamp to MI 3144 Euro Z 800 V instrument.
Wrap the object under test with the measuring clamp. See Figure 7.109.
Refer to MI 3144 Euro Z 800 V Instruction manual for detailed information.
178
MI 3155 EurotestXD Test and measurements
I Current
f Frequency
179
MI 3155 EurotestXD Test and measurements
Imax(Isc1, Isc2) Maximum first fault leakage current [Off, Custom, 3.0 mA ... 19.5 mA]
Connection diagrams
Figure 7.112: Measurement of highest First fault leakage current with 3-wire test lead
Figure 7.113: Measurement of First fault leakage current for RCD protected circuit with 3-
wire test lead
180
MI 3155 EurotestXD Test and measurements
Measurement procedure
181
MI 3155 EurotestXD Test and measurements
Connection diagram
182
MI 3155 EurotestXD Test and measurements
183
MI 3155 EurotestXD Test and measurements
terminal selection to L2 and proceeds with the test when supply voltage is
detected.)
Insulation resistance between L2-PE is decreased automatically according to limit
value1) every time interval selected with timer. To speed up the test press the
184
MI 3155 EurotestXD Test and measurements
𝑈𝐿1−𝐿2
Calculated first fault leakage current at threshold insulation resistance is given as 𝐼1(2) = 𝑅1(2)
,
where UL1-L2 is line-line voltage. The calculated first fault current is the maximum current that
would flow when insulation resistance decreases to the same value as the applied test
resistance, and a first fault is assumed between opposite line and PE.
If any of the activation / disconnection time result (t1, t2) is out of set limit, overall status of the
test is “failed” and cannot be modified manually. Otherwise, overall status can be user defined.
If activation of IMD device is visual indication and/or audio alert, without voltage disconnection,
Time limit (t1,t2) parameter should be set to “Off” to disable timing limitation.
185
MI 3155 EurotestXD Test and measurements
Connection diagram
186
MI 3155 EurotestXD Test and measurements
Measurement procedure
187
MI 3155 EurotestXD Test and measurements
7.36 Illumination
Probe positioning
Measurement procedure
188
MI 3155 EurotestXD Test and measurements
E Illumination
Measuring principle
189
MI 3155 EurotestXD Test and measurements
Connection diagram
Measurement procedure
190
MI 3155 EurotestXD Test and measurements
191
MI 3155 EurotestXD Test and measurements
Voltage
Z line
Voltage Drop
Zs rcd
RCD Uc
I ΔN Rated RCD residual current sensitivity [10 mA, 15 mA, 30 mA, 100 mA,
300 mA, 500 mA, 1000 mA]
RCD type RCD type [AC, A, F, B, B+]
Selectivity Characteristic [G, S]
Fuse type Selection of fuse type [Off, Custom, gG, NV, B, C, D, K, Z, L, U]
Fuse I Rated current of selected fuse
Fuse t Maximum breaking time of selected fuse
I (U) 1) Rated current for U measurement (custom value)
Isc factor Isc factor [Custom, 0.20 … 3.00]
I test Test current [Standard, Low]
Limit(ΔU) Maximum voltage drop [Off, Custom, 3.0 % ... 9.0 %]
Limit Uc Conventional touch voltage limit [Custom, 12 V, 25 V, 50 V]
Ia(Ipsc (LN)) Minimum short circuit current for selected fuse or custom value
1)
Applicable if Fuse type is set to Off or Custom.
192
MI 3155 EurotestXD Test and measurements
Connection diagram
Measurement procedure
193
MI 3155 EurotestXD Test and measurements
Voltage
Z line
Voltage Drop
Zs rcd
Rpe rcd
Connection diagram
194
MI 3155 EurotestXD Test and measurements
Measurement procedure
195
MI 3155 EurotestXD Test and measurements
Voltage
Z line
Voltage Drop
Z loop
Rpe
Connection diagram
196
MI 3155 EurotestXD Test and measurements
Measurement procedure
197
MI 3155 EurotestXD Test and measurements
Voltage
Z line
Voltage Drop
ISFL
IMD
Connection diagram
198
MI 3155 EurotestXD Test and measurements
Measurement procedure
199
MI 3155 EurotestXD Test and measurements
7.42 Locator
This function is intended for tracing mains installation, like:
Tracing lines,
Finding shorts, breaks in lines,
Detecting fuses.
The instrument generates test signals that can be traced with the handheld tracer receiver
R10K. See Appendix C – Locator receiver R10K for additional information.
200
MI 3155 EurotestXD Test and measurements
201
MI 3155 EurotestXD Test and measurements
Inspection
202
MI 3155 EurotestXD Test and measurements
Required test is selected from the CLAMP section of the Single Tests menu, see Figure 7.147
below. Menu is available only when Adapter MD 9273 is set, see chapters 4.6.8 Settings and
9.3 Communication with adapters for details.
Selected test is configured from EurotestXD. Adapter MD 9273 acquires test signals, process
measurements and send results to the EurotestXD. Results are presented on the instrument
screen and can be saved to the Workspace memory for later use.
Measurement parameters
There are no parameters to be set.
Connection diagram
203
MI 3155 EurotestXD Test and measurements
Measurement procedure
Connect MD 9273 to the item to be tested and set Bluetooth® mode.
Enter the Power CLAMP function and wait for active Bluetooth® communication sign.
Start the continuous measurement.
Stop the measurement.
Save results (optional).
Note:
Voltage test terminals connection and current flow toward load should be taken into account;
the red voltage terminal should be connected to the Line terminal and the jaw should be
correctly oriented, to obtain positive sign of Power test result. If Power test result has negative
sign, connection of voltage terminal or jaw orientation are opposite and the result of phase
displacement angle has opposite sign too. Consequently, load character determination
(capacitive or inductive) is mismatched.
204
MI 3155 EurotestXD Test and measurements
Measurement parameters
h Harmonic setup [1 to 19, 1st is fundamental frequency]
Connection diagram
Measurement procedure
Connect MD 9273 to the item to be tested and set Bluetooth® mode.
Enter the Voltage CLAMP function and wait for active Bluetooth® communication sign.
Set test parameter.
Start the continuous measurement.
Stop the measurement.
Save results (optional).
205
MI 3155 EurotestXD Test and measurements
Measurement parameters
h Harmonic setup [1 to 19, 1st is fundamental frequency]
Connection diagram
Measurement procedure
Connect MD 9273 to the item to be tested and set Bluetooth® mode.
Enter the Current CLAMP function and wait for active Bluetooth® communication sign.
Set test parameter.
Start the continuous measurement.
Stop the measurement.
206
MI 3155 EurotestXD Test and measurements
207
MI 3155 EurotestXD Test and measurements
Figure 7.157: Inrush CLAMP menu – setup on the left, waiting for trigger on the right
Test parameters
Inrush threshold Inrush current threshold setting [Off, 5 mA … 90 A]
Voltage threshold Voltage dip threshold setting [Off, 50 V … 500 V]
Duration Recording duration [3 s, 10 s]
Connection diagram
Test procedure
Connect MD 9273 to the item to be tested and set Bluetooth® mode.
Enter the Inrush CLAMP function and wait for active Bluetooth® communication sign.
Set test parameters.
Set charts Y value range1) within expected values (optional; could be set later, after
the test)
Start the test.
Initiate set threshold event or manually trigger test recording.
Save results (optional) after test is finished and results and recorded charts are
presented on the screen.
1)
Chart range selection:
Voltage range [100 mV/div … 100 V/div]
Current range [10 mA/div … 200 A/div]
208
MI 3155 EurotestXD Test and measurements
Measurement parameters
Type [%, V]
% – harmonics and distortion are displayed as relative value
V – harmonics and distortion are displayed as absolute value
Connection diagram
209
MI 3155 EurotestXD Test and measurements
Measurement procedure
Connect MD 9273 to the item to be tested and set Bluetooth® mode.
Enter the Harmonics U CLAMP function and wait for active Bluetooth®
communication.
Set Type parameter.
Set charts Y value range3) within expected values (optional; could be set later, after
the test).
Start the continuous measurement.
Stop the measurement.
Save results (optional).
3)
Chart Voltage range selection: [100 mV/div … 100 V/div]
210
MI 3155 EurotestXD Test and measurements
Measurement parameters
Type [%, A]
% – harmonics and distortion are displayed as relative value
A – harmonics and distortion are displayed as absolute value
Connection diagram
Measurement procedure
Connect MD 9273 to the item to be tested and set Bluetooth® mode.
Enter the Harmonics I CLAMP function and wait for active Bluetooth® communication.
Set Type parameter.
Set charts Y value range within expected values (optional; could be set later, after the
test).
Start the continuous measurement.
Stop the measurement.
Save results (optional).
211
MI 3155 EurotestXD Test and measurements
212
MI 3155 EurotestXD Auto Sequences
8 Auto Sequences®
Pre-programmed sequences of measurements can be carried out in Auto Sequences® menu.
The sequence of measurements, their parameters and flow of the sequence can be
programmed. The results of an Auto sequence® can be stored in the memory together with all
related information.
Auto Sequences® can be pre-programmed on PC with the Metrel ES Manager software and
uploaded to the instrument. Refer to chapter Appendix F Programming of Auto Sequences
on Metrel ES Manager for detailed information on programming Auto Sequences.
On the instrument parameters and limits of individual single test in the Auto Sequence® can be
changed / set.
Procedure
213
MI 3155 EurotestXD Auto Sequences
Procedure
214
MI 3155 EurotestXD Auto Sequences
Figure 8.1: Search results screen (left), Auto Sequence selected (right)
Options
Next page.
Previous page.
Note:
Search result page consist of up to 50 results.
215
MI 3155 EurotestXD Auto Sequences
Figure 8.2: Examples of organized Auto Sequences® in main Auto Sequences® menu
Options
216
MI 3155 EurotestXD Auto Sequences
During the execution phase of an Auto Sequence®, pre-programmed single tests are
carried out. The sequence of single tests is controlled by pre-programmed flow
commands.
After the test sequence is finished the Auto Sequence® result menu is shown. Details of
individual tests can be viewed and the results can be saved to Memory organizer.
Note:
Once fuse and RCD parameters are changed in active Auto Sequence®, the new
settings are distributed through all single tests within the active Auto Sequence® and
stored for next use of the same Auto Sequence®.
Figure 8.3: Screen organization in Auto Sequence® view menu – Header selected
Legend
1 Short code
2 Auto Sequence® name
3 Header
4 Single tests
5 Description
6 Control panel (available options)
Option
217
MI 3155 EurotestXD Auto Sequences
Figure 8.4: Screen organization in Auto Sequence® view menu – measurement selected
Legend
1 Auto Sequence® name
2 Header
3 Single tests
4 Parameters / limits of selected single test
5 Control panel (available options)
Options
Selects single test.
on
218
MI 3155 EurotestXD Auto Sequences
The attached ‘x3’ at the end of a single test name indicates that a loop of single tests is
programmed. This means that the marked single test will be carried out as many times as the
number behind the ‘x’ indicates. It is possible to exit the loop before, at the end of each
individual measurement.
219
MI 3155 EurotestXD Auto Sequences
Figure 8.6: Auto Sequence® – Example of a finished measurement with options for
proceeding
Ends the Auto Sequence® and goes to Auto Sequence® result screen.
Refer to chapter 8.2.3 Auto Sequence® result screen for more information.
Exits the loop of single tests and proceeds to the next step in the Auto Sequence®.
The offered options in the control panel depend on the selected single test, its result and the
programmed test flow.
Notes:
During Auto Sequences® the popup Warning messages are displayed only before the
single test inside one Auto Sequence®. This default setting can be changed with
appropriate flow command. For more information about programming Auto Sequences®
refer to chapter Appendix F – Programming of Auto Sequences on Metrel ES
Manager.
220
MI 3155 EurotestXD Auto Sequences
Legend
1 Overall PASS / FAIL status
2 Short code
3 Auto Sequence® name
4 Single tests with individual PASS / FAIL statuses
5 Description
6 Control panel (available options)
Options
221
MI 3155 EurotestXD Auto Sequences
A new Auto Sequence® was started from the Auto Sequence® main
menu:
▪ Saving under the last selected Structure object will be offered by
default. The user can select another Structure object or create a
Options (menu for viewing details of Auto Sequence® and single test results)
Adds comment to the selected single test in Auto Sequence®. The instrument opens
keypad for entering a comment.
Figure 8.8: Details of menu for viewing details of Auto Sequence® results
222
MI 3155 EurotestXD Auto Sequences
Options
Enters menu for viewing details of the Auto Sequence®. Refer to chapter 8.2.3 Auto
Sequence® result screen for more information.
223
MI 3155 EurotestXD Communication
9 Communication
The instrument can communicate with the Metrel ES Manager PC software and aMESM
android application. The following action is supported:
Saved results and Tree structure from Memory organizer can be downloaded and stored
to a PC or android device.
Tree structure from Metrel ES Manager PC software or aMESM android aplication can
be uploaded to the instrument.
Custom Auto Sequences® can be uploaded to the instrument or downloaded and stored
to a PC.
Metrel ES Manager is a PC software running on Windows 8.1, Windows 10 and Windows 11.
There are three communication interfaces available on the instrument:
RS-232
USB
Bluetooth.
Instrument can also communicate to various external devices (test adapters, scanners …).
Figure 9.1: Interface connection for data transfer over PC COM port
224
MI 3155 EurotestXD Communication
If this option is not supported by the selected Android application, then configure a
Bluetooth link via Android device’s Bluetooth configuration tool. Usually, no code for
pairing the devices is needed.
The instrument and Android device are ready to communicate.
Notes
Sometimes there will be a demand from the Android device to enter the code. Enter
code ‘1234’ to correctly configure the Bluetooth link.
The name of correctly configured Bluetooth device must consist of the instrument type
plus serial number, e.g. MI 3155-12240429I. If the Bluetooth module got another name,
the configuration must be repeated.
In case of serious troubles with the Bluetooth communication it is possible to reinitialize
the internal Bluetooth module. The initialization is carried out during the Initial settings
procedure. In case of a successful initialization “INITIALIZING… OK!” is displayed at the
end of the procedure. See chapter 4.6.11 Initial Settings.
Metrel android application aMESM is available for download from Google play store:
When adapter is selected from the list, supported communication port is automatically offered.
To establish Bluetooth communication, adapter must be paired with EurotectXD.
Procedure:
1. Adapter: switch it ON and select BT mode, if not already automatically selected.
2. EurotestXD: Open General Settings / Settings menu and navigate to the Adapters
section.
3. Adapter type: select adapter by using left / right arrows or tap on the field and select it
225
MI 3155 EurotestXD Communication
Note:
Pairing between same Metrel adapter and same EurotestXD is necessary only when adapter is
first time used. If communication is not established when supported test is selected, adapter is
probably switched OFF or Bluetooth link is out of range.
226
MI 3155 EurotestXD Upgrading the instrument
227
MI 3155 EurotestXD Maintenance
11 Maintenance
Unauthorized persons are not allowed to open the EurotestXD instrument. There are no user
replaceable components inside the instrument, except the battery and fuses under back cover.
228
MI 3155 EurotestXD Maintenance
Warnings:
Disconnect all measuring accessory and switch off the instrument before opening battery
/ fuse compartment cover, hazardous voltage inside!
Replace blown fuse with original type only, otherwise the instrument or accessory may
be damaged and / or operator’s safety impaired!
229
MI 3155 EurotestXD Maintenance
Warnings:
Disconnect all measuring accessory and switch off the instrument before opening battery
/ fuse compartment cover, hazardous voltage inside!
Replace battery pack with original type only, otherwise the instrument may be damaged
and / or operator’s safety impaired!
Ensure batteries are used and disposed of in accordance with Manufacturers guidelines
and in accordance with Local and National Authority guidelines.
11.3 Cleaning
No special maintenance is required for the housing. To clean the surface of the instrument or
accessory use a soft cloth slightly moistened with soapy water or alcohol. Then leave the
instrument or accessory to dry totally before use.
Warnings:
Do not use liquids based on petrol or hydrocarbons!
Do not spill cleaning liquid over the instrument!
11.5 Service
For repairs under warranty, or at any other time, please contact your distributor.
230
MI 3155 EurotestXD Technical specifications
12 Technical specifications
12.1 R iso, R iso all – Insulation resistance
Uiso: 50 V, 100 V and 250 V (R iso, R iso all)
Riso – Insulation resistance (R iso)
R L-N, R L-PE, R N-PE – Insulation resistance (R iso all)
Measuring range according to EN 61557 is 0.15 M … 199.9 M.
Measuring range (M) Resolution (M) Accuracy
0.00 ... 19.99 0.01 (5 % of reading + 3 digits)
20.0 ... 99.9 (10 % of reading)
0.1
100.0 ... 199.9 (20 % of reading)
Nominal voltages Uiso ............................50 VDC, 100 VDC, 250 VDC, 500 VDC, 1000 VDC,
2500 VDC
Open circuit voltage ................................-0 % / +20 % of nominal voltage
Measuring current...................................min. 1 mA at RN = UN 1 k/V
Short circuit current ............................... max. 3 mA
Load pretest voltage .............................. < 20 VAC, 125 Hz
Load pretest warning ............................. < 50 k
The number of possible tests
231
MI 3155 EurotestXD Technical specifications
(R iso, Riso all) ...................................... up to 1000, with a fully charged battery (type:
18650T22A2S2P)
up to 2000, with a fully charged battery (type:
18650T22A2S4P)
Specified accuracy is valid if 3-wire test lead is used while it is valid up to 100 M if Tip
commander is used.
Specified accuracy is valid up to 100 M if relative humidity is > 85 %.
In case the instrument gets moistened, the results could be impaired. In such case, it is
recommended to dry the instrument and accessories for at least 24 hours.
The error in operating conditions could be at most the error for reference conditions (specified in
the manual for each function) 5 % of measured value.
PI – Polarization index
Measuring range Resolution Accuracy
0.01 ... 9.99 0.01 (5 % of reading + 2 digits)
10.0 ... 100.0 0.1 (5 % of reading)
For Riso, R60 and Um sub-results technical specifications defined in chapter 12.1 R iso, R iso
all – Insulation resistance apply.
R – Resistance
Measuring range () Resolution () Accuracy
0.00 ... 19.99 0.01 (3 % of reading + 3 digits)
20.0 ... 199.9 0.1
(5 % of reading)
200 ... 1999 1
R+, R – Resistance
Measuring range () Resolution () Accuracy
0.0 ... 199.9 0.1
(5 % of reading + 5 digits)
200 ... 1999 1
232
MI 3155 EurotestXD Technical specifications
(Rlow (Current = standard), Rlow 4W) ....up to 1700, with a fully charged battery (type:
18650T22A2S2P)
up to 3400, with a fully charged battery (type:
18650T22A2S4P)
Rlow (Current = ramp) ............................up to 1000, with a fully charged battery (type:
18650T22A2S2P)
up to 2000, with a fully charged battery (type:
18650T22A2S4P)
R+, R – Resistance
Measuring range () Resolution () Accuracy
0.00 ... 19.99 0.01
20.0 ... 199.9 0.1 (5 % of reading + 5 digits)
200 ... 1999 1
233
MI 3155 EurotestXD Technical specifications
Test current shape..................................Sine-wave (AC), pulsed (A, F), smooth DC (B, B+)
DC offset for pulsed test current .............< 2 mA (typical)
RCD type ................................................(non-delayed), S (time-delayed), PRCD, PRCD-2p,
PRCD-3p, PRCD-S, PRCD-S+, PRCD-K, EV RCD,
EV RCM, MI RCD
Test current starting polarity ...................0º or 180º
Voltage range .........................................93 V ... 134 V (16 Hz … 400 Hz)
185 V ... 266 V (16 Hz … 400 Hz)
RCD test current in relation to RCD type, nominal RCD current and multiplication factor
IN × 1/2 IN × 1 IN × 2 IN × 5 RCD I
(mA) (mA) (mA) (mA)
IN (mA) AC A, F B, B+ AC A, F B, B+ AC A, F B, B+ AC A, F B, B+ AC A, F B, B+
10 5 3.5 5 10 20 20 20 40 40 50 100 100 ✓ ✓ ✓
15 7.5 5.3 7.5 15 30 30 30 60 60 75 150 150 ✓ ✓ ✓
30 15 10.5 15 30 42 60 60 84 120 150 212 300 ✓ ✓ ✓
100 50 35 50 100 141 200 200 282 400 500 707 1000 ✓ ✓ ✓
300 150 105 150 300 424 600 600 848 × 1500 × × ✓ ✓ ✓
500 250 175 250 500 707 1000 1000 1410 × 2500 × × ✓ ✓ ✓
1000 500 350 500 1000 1410 × 2000 × × × × × ✓ ✓ ×
×. ............................................................not applicable
✓.............................................................applicable
AC type...................................................sine wave test current
A, F types…… ........................................pulsed current
B, B+ types .............................................smooth DC current
234
MI 3155 EurotestXD Technical specifications
30 15 30 60 150 × × × ✓ ×
a.c.
6 3 6 × × 60 200 300 × ✓
d.c.
×. ............................................................not applicable
✓.............................................................applicable
MI / EV type (a.c. part) ............................sine wave test current
MI / EV type (d.c. part) ............................smooth DC current
t N –Trip-out time
Measuring range (ms) Resolution (ms) Accuracy
0.0 ... 40.0 0.1 1 ms
0.0 ... max. time* 0.1 3 ms
* For max. time see normative references in chapter 4.6.8.2 RCD standard. This specification
applies to max. time >40 ms.
I – Trip-out current
Measuring range Resolution I Accuracy
0.2IN ... 1.1IN
0.05IN 0.1IN
(AC type)
0.2IN ... 1.0IN 0.05IN 0.1IN
235
MI 3155 EurotestXD Technical specifications
t I – Trip out-time
Measuring range (ms) Resolution (ms) Accuracy
0 ... 300 1 3 ms
The accuracy is valid if mains voltage is stabile during the measurement and PE terminal is free
of interfering voltages. Specified accuracy is valid for complete operating range.
Trip-out measurement is not available for IN=1000 mA (RCD types B, B+).
236
MI 3155 EurotestXD Technical specifications
Ulpe – Voltage
Measuring range (V) Resolution (V) Accuracy
0 ... 550 1 (2 % of reading + 2 digits)
237
MI 3155 EurotestXD Technical specifications
Ulpe – Voltage
Measuring range (V) Resolution (V) Accuracy
0 ... 550 1 (2 % of reading + 2 digits)
Uc – Contact voltage
Refer to chapter 12.6.2 RCD Uc – Contact voltage for detailed technical specification.
238
MI 3155 EurotestXD Technical specifications
Uln – Voltage
Measuring range (V) Resolution (V) Accuracy
0 ... 550 1 (2 % of reading + 2 digits)
12.12Voltage Drop
∆U – Voltage drop
Measuring range (%) Resolution (%) Accuracy
Consider accuracy of line
0.0 ... 99.9 0.1
impedance measurement(s)*
239
MI 3155 EurotestXD Technical specifications
12.7 Z loop, Z loop 4W – Fault loop impedance and prospective fault current,
12.8 Zs rcd – Fault loop impedance and prospective fault current in system with RCD,
12.27 IMD.
240
MI 3155 EurotestXD Technical specifications
241
MI 3155 EurotestXD Technical specifications
Principle:
= 2··d·Re,
where Re is a measured resistance in 4-wire method and d is distance between the probes.
Accuracy note:
Accuracy of the specific earth resistance result depends on measured earth resistance Re as
follows:
Re – Earth resistance
Measuring range () Accuracy
1.00 ... 1999 5 % of measured value
2000 ... 19.99 k 10 % of measured value
>20 k 20 % of measured value
Additional error:
See Earth resistance three-wire method.
242
MI 3155 EurotestXD Technical specifications
12.20.3 Frequency
Measuring range (Hz) Resolution (Hz) Accuracy
0.00 ... 9.99 0.01
(0.2 % of reading + 1 digit)
10.0 ... 499.9 0.1
Uac – AC voltage
Measuring range (V) Resolution (V) Accuracy
0 … 1562 1 Consider accuracy of DC
voltage
243
MI 3155 EurotestXD Technical specifications
12.22 Currents
Maximum voltage on C1 measuring input ........ 3 V
Nominal frequency ................................. ......... 0 Hz, 40 Hz ... 500 Hz
244
MI 3155 EurotestXD Technical specifications
12.24 Power
Measurement characteristics
Function symbols Class according to Measuring range
IEC 61557-12
P – Active power 2.5 5 % ... 100 % INom*)
S – Apparent power 2.5 5 % ... 100 % INom*)
Q – Reactive power 2.5 5 % ... 100 % INom*)
PF – Power factor 1 - 1 ... 1
THDu 2.5 0 % ... 20 % UNom
*)
INom depends on selected current clamp type and selected range as follows:
A 1018:[20 A]
A 1391: [40 A, 300 A]
Error of external voltage and current transducers is not considered in this specification.
12.25 Harmonics
Measurement characteristics
Function symbols Class according to Measuring range
IEC 61557-12
Uh 2.5 0 % ... 20 % UNom
THDu 2.5 0 % ... 20 % UNom
Ih 2.5 0 % ... 100 % INom*)
THDi 2.5 0 % ... 100 % INom*)
*)
INom depends on selected current clamp type and selected range as follows:
A 1018: [20 A]
A 1391: [40 A, 300 A]
Error of external voltage and current transducers is not considered in this specification.
245
MI 3155 EurotestXD Technical specifications
12.27 IMD
R1, R2 – Threshold insulation resistance
R (kΩ) Resolution (kΩ) Note
5 ... 640 5 up to 128 steps
12.28 Illumination
E – Illumination (A 1172)
Specified accuracy is valid for complete operating range.
Measuring range (lux) Resolution (lux) Accuracy
0.01 ... 19.99 0.01 (5 % of reading + 2 digits)
20.0 ... 199.9 0.1
200 ... 1999 1 (5 % of reading)
2.00 k ... 19.99 k 10
E – Illumination (A 1173)
Specified accuracy is valid for complete operating range.
Measuring range (lux) Resolution (lux) Accuracy
0.01 ... 19.99 0.01 (10 % of reading + 3 digits)
20.0 ... 199.9 0.1
(10 % of reading)
200 ... 1999 1
246
MI 3155 EurotestXD Technical specifications
Up – Peak voltage
Range (V) Resolution (V) Accuracy
0 ... 372 1 (5 % of reading + 3 digits)
U – RMS voltage
Range (V) Resolution (V) Accuracy
70.0 ... 99.9 0.1
(2 % of reading + 2 digits)
100 … 263 1
Freq – Frequency
Range (Hz) Resolution (Hz) Accuracy
45.0 ... 65.0 0.1 (0.2 % of reading + 1 digits)
247
MI 3155 EurotestXD Technical specifications
248
MI 3155 EurotestXD Technical specifications
12.41General data
Power supply ......................................... Li-Ion battery pack 7.2 V,
4400 mAh (type: 18650T22A2S2P)
8800 mAh (type: 18650T22A2S4P) optional
Operation............................................... Typical 16 h (type: 18650T22A2S2P)
Typical 32 h (type: 18650T22A2S4P)
Charger socket input voltage ................. 12 V 10 %
Charger socket input current.................. 3000 mA max.
Battery charging current ........................ up to 2200 mA (Battery type: 18650T22A2S2P)
up to 3000 mA (Battery type: 18650T22A2S4P)
Measuring category ............................... 600 V CAT III
300 V CAT IV
Protection classification ......................... double insulation
Pollution degree..................................... 2
Protection degree .................................. 56 (with protective covers on USB, Charger and
PS/2 terminals)
Altitude .................................................. ≤ 2000 m
Display .................................................. 4.3 inch (10.9 cm) 480x272 pixels TFT colour display
with touch screen
Dimensions (w h d) .......................... 252 mm 111 mm 165 mm
Weight ................................................... 1.78 kg, with battery pack (type: 18650T22A2S2P)
Reference conditions
Reference temperature range ................ 10 C ... 30 C
Reference humidity range ...................... 40 %RH ... 70 %RH
Operation conditions
Working temperature range ................... 0 C ... 40 C
Maximum relative humidity .................... 95 %RH (0 C ... 40 C), non-condensing
Operation............................................... Indoor use
Storage conditions
Temperature range ................................ -10 C ... +70 C
Maximum relative humidity .................... 90 %RH (-10 C ... +40 C)
80 %RH (40 C ... 60 C)
Locator
Locator .................................................. supports inductive mode
Maximum operation voltage ................... 440 V a.c.
Communication ports, memory
RS 232 .................................................. 115200 bits/s, 8N1 serial protocol
USB ....................................................... USB 2.0 Hi speed interface
with USB type B receptacle connector
Data storage capacity ............................ 8 GB SD memory card
Bluetooth module................................... Class 2
EMC
Emission ................................................ Class B
Immunity ................................................ Basic electromagnetic environment
(Portable test and measurement equipment)
The error in operating conditions could be at most the error for reference conditions (specified in
the manual for each function) +1 % of measured value + 1 digit, unless otherwise specified in
the manual for particular function.
249
MI 3155 EurotestXD Appendix A – Profile notes
Special delayed G type RCD selection added in the Selectivity parameter in Test Parameters /
Limits section as follows:
Time limits are the same as for general type RCD and contact voltage is calculated the same as
for general type RCD.
Selective (time delayed) RCDs and RCDs with (G) - time delayed characteristic demonstrate
delayed response characteristics. They contain residual current integrating mechanism for
generation of delayed trip out. However, contact voltage pre-test in the measuring procedure
also influences the RCD and it takes a period to recover into idle state. Time delay of 30 s is
inserted before performing trip-out test to recover S type RCD after pre-tests and time delay of 5
s is inserted for the same purpose for G type RCD.
Table 7.3: Relation between Uc, Uc(P) and IN changed as follows:
250
MI 3155 EurotestXD Appendix A – Profile notes
Measurement procedure
251
MI 3155 EurotestXD Appendix A – Profile notes
Figure A.3: Example of individual steps in RCD Auto test – Inserted 2 new steps
252
MI 3155 EurotestXD Appendix A – Profile notes
I N Rated RCD residual current sensitivity [10 mA, 15 mA, 30 mA, 100 mA, 300
mA, 500 mA, 650 mA, 1000 mA]
Nominal residual current (A,AC) .............10 mA, 15 mA, 30 mA, 100 mA, 300 mA, 500 mA,
650 mA, 1000 mA
RCD test current in relation to RCD type, nominal RCD current and multiplication factor
IN × 1/2 IN × 1 IN × 2 IN × 5 RCD I
(mA) (mA) (mA) (mA)
IN AC A, F B, AC A, FB, AC A, F B, AC A, B, AC A, B,
(mA) B+ B+ B+ F B+ F B+
10 5 3.5 5 10 20 20 20 40 40 50 100 100 ✓ ✓ ✓
15 7.5 5.3 7.5 15 30 30 30 60 60 75 150 150 ✓ ✓ ✓
30 15 10.5 15 30 42 60 60 84 120 150 212 300 ✓ ✓ ✓
100 50 35 50 100 141 200 200 282 400 500 707 1000 ✓ ✓ ✓
300 150 105 150 300 424 600 600 848 × 1500 × × ✓ ✓ ✓
500 250 175 250 500 707 1000 1000 1410 × 2500 × × ✓ ✓ ✓
650 325 227.5 250 650 916.5 1300 1300 × × × × × ✓ ✓ ✓
1000 500 350 500 1000 1410 × 2000 × × × × × ✓ ✓ ×
×. ............................................................not applicable
✓.............................................................applicable
AC type...................................................sine wave test current
A, F types…… ........................................pulsed current
B, B+ types .............................................smooth DC current
I N Rated RCD residual current sensitivity [10 mA, 15 mA, 30 mA, 100 mA, 300
mA, 500 mA, 650 mA, 1000 mA]
Nominal residual current (A,AC) .............10 mA, 15 mA, 30 mA, 100 mA, 300 mA, 500 mA,
650 mA, 1000 mA
253
MI 3155 EurotestXD Appendix A – Profile notes
RCD test current in relation to RCD type, nominal RCD current and multiplication factor
IN × 1/2 IN × 1 IN × 2 IN × 5 RCD I
(mA) (mA) (mA) (mA)
IN AC A, F B, AC A, FB, AC A, F B, AC A, B, AC A, B,
(mA) B+ B+ B+ F B+ F B+
10 5 3.5 5 10 20 20 20 40 40 50 100 100 ✓ ✓ ✓
15 7.5 5.3 7.5 15 30 30 30 60 60 75 150 150 ✓ ✓ ✓
30 15 10.5 15 30 42 60 60 84 120 150 212 300 ✓ ✓ ✓
100 50 35 50 100 141 200 200 282 400 500 707 1000 ✓ ✓ ✓
300 150 105 150 300 424 600 600 848 × 1500 × × ✓ ✓ ✓
500 250 175 250 500 707 1000 1000 1410 × 2500 × × ✓ ✓ ✓
650 325 227.5 250 650 916.5 1300 1300 × × × × × ✓ ✓ ✓
1000 500 350 500 1000 1410 × 2000 × × × × × ✓ ✓ ×
×. ............................................................not applicable
✓.............................................................applicable
AC type...................................................sine wave test current
A, F types…… ........................................pulsed current
B, B+ types .............................................smooth DC current
Online voltages are displayed together with test terminal indication. All three
test terminals are used for selected measurement.
254
MI 3155 EurotestXD Appendix B – Commanders
B.2 Battery
The commander uses two AAA size alkaline or rechargeable Ni-MH battery cells.
Nominal operating time is at least 40 h and is declared for cells with nominal capacity of 850
mAh.
Notes:
If the commander is not used for a long period of time, remove all batteries from the
battery compartment.
Alkaline or rechargeable Ni-MH batteries (size AAA) can be used. Metrel recommends
only using rechargeable batteries with a capacity of 800 mAh or above.
Ensure that the battery cells are inserted correctly otherwise the commander will not
operate and the batteries could be discharged.
255
MI 3155 EurotestXD Appendix B – Commanders
256
MI 3155 EurotestXD Appendix C – Locator receiver R10K
Detectors Operation
In built inductive sensor (IND) Tracing hidden wires.
Current clamp (optional) Connected through the rear connector.
Locating wires.
Selective probe Connected through the rear connector.
Locating fuses in fuse cabinets.
The user can choose between three sensitivity levels (low, middle and high). An extra
potentiometer is added for fine sensitivity adjustment. A buzzer sound and 10-level LED bar
graph indicator indicates the strength of the magnetic field e.g. proximity of the traced object.
Note:
The field strength can vary during tracing. The sensitivity should always be adjusted
to optimum for each individual tracing.
257
MI 3155 EurotestXD Appendix D – Structure objects
Object Object
Circuit Circuit
Connection Connection
Socket Socket
Light Light
RCD RCD
MPE MPE
258
MI 3155 EurotestXD Appendix D – Structure objects
Data proc. Inst. Protective conductor for Lift Data processing installation
Inverter Inverter
Module Module
259
MI 3155 EurotestXD Appendix D – Structure objects
Level 2 Level 2
Level 3 Level 3
Varistor Varistor
LS connection LS connection
Machine Machine
260
MI 3155 EurotestXD Appendix E – Default list of AutoSequences
261
MI 3155 EurotestXD Appendix F – Programming of AutoSequences
on Metrel ES Manager
the left side ,structure of selected group of Auto Sequence® is displayed. In the middle
part of the workspace , the elements of the selected Auto Sequence® are shown. On the
right side, list of available Single tests (measurements and inspections) and list of Flow
262
MI 3155 EurotestXD Appendix F – Programming of AutoSequences
on Metrel ES Manager
An Auto Sequence® begins with Name, Description and Image, followed by the first step
(Header), one or more measuring steps and ends with the last step (Result). By inserting
appropriate Single tests (measurements and inspections) and Flow commands and
setting their parameters, arbitrary Auto Sequences® can be created.
263
MI 3155 EurotestXD Appendix F – Programming of AutoSequences
on Metrel ES Manager
Operation options on Group of Auto Sequences® are available from menu bar at the top of
Auto Sequence® Editor workspace.
File operation options:
Toggle between Search by name within Auto Sequence® group and normal
view. See chapter F.4 Search within selected Auto Sequence® group for
details.
Group of Auto Sequences® operation options (also available by right clicking on Folder or
Auto Sequence®):
Deletes:
- the selected Auto Sequence®
- the selected folder with all subfolders and Auto Sequences®
Right click on the selected Auto Sequence® or Folder opens menu with additional
possibilities:
Auto Sequence®: Edit Name, Description and Image (see Figure F.6.)
Folder: Edit folder name
264
MI 3155 EurotestXD Appendix F – Programming of AutoSequences
on Metrel ES Manager
DOUBLE CLICK
Folder name: Edit folder name
Drag and drop of the selected Auto Sequence® or Folder / Subfolder moves it to a new
location:
265
MI 3155 EurotestXD Appendix F – Programming of AutoSequences
on Metrel ES Manager
Figure F.6: Editing the Auto Sequence® Name, Description and Image
266
MI 3155 EurotestXD Appendix F – Programming of AutoSequences
on Metrel ES Manager
When function is selected, Search menu as presented on Figure F.7 appear on the
screen. By entering the text into search box, found results are automatically highlighted with
yellow background. Search functionality is implemented in Folders, Subfolders and Auto
Sequences® of selected Auto Sequence® Group. Search text can be cleared by selecting
the Clear button.
267
MI 3155 EurotestXD Appendix F – Programming of AutoSequences
on Metrel ES Manager
Header
The Header step is empty by default.
Flow commands can be added to the Header step.
Measurement step
The Measurement step contains a Single test and the Operation after end of test flow
command by default. Other Flow commands can also be added to the Measurement step.
Result
The Result step contains the Result screen flow command by default. Other Flow commands
can also be added to the Result step.
268
MI 3155 EurotestXD Appendix F – Programming of AutoSequences
on Metrel ES Manager
Options
Adding flow commands Selected flow command can be dragged from the list
of Flow Commands and dropped on the appropriate
place in any Auto Sequence® step.
Changing position of flow command By a click on an element and use of and keys.
within measurement step
Delete
Deletes the selected measurement step / flow
command.
269
MI 3155 EurotestXD Appendix F – Programming of AutoSequences
on Metrel ES Manager
Pause
A Pause command with text message or picture can be inserted anywhere in the measuring
steps. Warning icon can be set alone or added to text message. Arbitrary text message can
be entered in prepared field Text of menu window.
Parameters
Pause type Show text and/or warning ( check to show warning icon).
Show picture ( browse for image path).
Buzzer mode
Passed or failed measurement is indicated with beeps.
Pass – double beep after the test
Fail – long beep after the test
Beep happens right after single test measurement.
Parameters
270
MI 3155 EurotestXD Appendix F – Programming of AutoSequences
on Metrel ES Manager
Parameters
Operation after end of test The operation can be individually set for the case the
measurement passed, failed or ended without a status.
pass
fail Manual: The test sequence stops and waits for
appropriate command (RUN key, external
no status
command …) to proceed.
Result screen
This flow command controls the proceeding after the Auto Sequence® has ended.
Parameters
Auto save Auto Sequence® results are stored in the momentary workspace.
A new Node with the current month and year will be created. Under the
Node Auto Sequence® results will be stored.
Up to 100 Auto Sequence® results can be automatically stored under the
same node. If more results are available, they are split to multiple nodes.
Auto Save Flow setting is disabled by default.
Note
This flow command is active only if Auto Sequence® is started from
the Auto Sequence® Main menu (not from the Memory organizer).
271
MI 3155 EurotestXD Appendix F – Programming of AutoSequences
on Metrel ES Manager
Name of Custom Inspection Item tasks and Type of Item Pass / Fail checkbox marking
272
MI 3155 EurotestXD Appendix F – Programming of AutoSequences
on Metrel ES Manager
Browse to the location, confirm overwriting, if file already exists or edit file name to
save it as a new Custom Inspection Data file.
273
MI 3155 EurotestXD Appendix F – Programming of AutoSequences
on Metrel ES Manager
Item tasks containing Child Item tasks are marked with triangle in
front of their Name.
Click on triangle mark:
274
MI 3155 EurotestXD Appendix F – Programming of AutoSequences
on Metrel ES Manager
275
MI 3155 EurotestXD Appendix G – Tests and Measurements with adapters
276
MI 3155 EurotestXD Appendix G – Tests and Measurements with adapters
277