RPQP05 Calibration Procedure
RPQP05 Calibration Procedure
MASTER COPY
AMENDMENT HISTORY
Consider all the instruments and devices that measure something in your facility. These
measurements may include temperature, weight, pressure, humidity, pH, salinity, and the list
continues. Each device has a purpose, and the resulting measurements are needed for monitoring
and managing the equipment or product. But what happens when these devices become inaccurate
and fail to provide precise results? Does your facility have an instrument calibration or accuracy check
program with procedures to verify that the devices are precise and accurate? Management should be
confident that their measuring devices are reading correctly. The following steps will help you
establish an instrument calibration procedure, or tune up your existing one.
If the instruments do not have a manufacturer identification number, clearly mark each
device with a unique number by attaching an asset number tag, etching the device, or using another
labeling method. Identification by location (i.e., table scale) may be appropriate if there is only one of
the devices; however, a numbering system is recommended, especially if any of the following
circumstances apply:
The device is the standard used to calibrate other devices (such as a NIST-certified
thermometer).
The device is used to monitor a critical control point (CCP).
Your facility has more than one of the devices. For example, table scales can be moved
easily and the same one may not necessarily be returned to the same location.
Assign a procedure number for the method of certification, calibration, and accuracy checks used for
each device. Documented procedures should include:
Certification of Accuracy: The use of a recognized standard, such as calibration against the
National Institute for Standards and Technology (NIST), is a regulatory requirement for some
segments of the food industry. The use of a NIST-certified device or certified-reference standards is
typical for most measuring devices, based on the critical nature of the measuring device accuracy,
Good Laboratory Practices, and other standards of the industry.
PRO NO RPQP05
PROCEDURE FOR REV 01
CALIBRATION DATE 01-JUL-2023
At a minimum, the traceable reference standard should be recertified annually. Obtain and maintain
the documentation provided by the certification company. Note: for digital thermometers with
interchangeable probes, both the digital device and affiliated probe are to be recertified.
If an outside company is used to calibrate the devices, the resulting documentation should include
the NIST-certified standard’s identifying number.
Calibration: Calibration involves testing the device with two different measurements or standards,
typically just above and below the range of actual use. Examples include:
Two different buffer solutions would be used to calibrate a pH meter (such as 4.0 and 7.0 if
the products being tested are at a range of 4.2 to 5.0).
A thermometer would be checked against the NIST-certified thermometer at two
temperatures, such as ice chip (for cold) or oil tank (for hot) AND at room temperature.
Ensure the instrument is capable of handling the calibration standards. An extreme example
would be testing a sensitive table scale with a 100-pound weight. Similarly, a bi-metal or dial-type
thermometer can be damaged if exposed to rapid temperature extremes (such as plunging the probe
into ice water directly after being in a hot oil bath)
Frequency Determination: The frequency of conducting each type of accuracy check depends on
multiple factors, including:
Regulatory requirements.
Manufacturer’s recommendations.
Industry standards/Good Laboratory Practices.
Level of risk to product safety if the device is inaccurate.
Inherent reliability. (For example, the least reliable types of thermometers are dial and bi-
metal probe thermometers, whereas most digital thermometers are fairly reliable.)
Environment of use. Is the measuring device moved (e.g., table-top scale moved during
daily sanitation)?
Potential abuse, such as receiving thermometers that may be put into pockets, sat on, etc.
Historic information about accuracy/past performance.
Methodology
Each method of determining the accuracy of a measuring device is to have clearly outlined
procedures that an individual can follow with training. A key resource for accuracy `checks or
calibration procedures is the equipment manufacturer. These procedures do not necessarily need to
be elaborate
PRO NO RPQP05
PROCEDURE FOR REV 01
CALIBRATION DATE 01-JUL-2023
Verification:
Remember the adage, “If it is not documented, it was not done.” A number of documents
have been identified, including the procedures for conducting the accuracy check/calibration, the
National Institute for Standards and Technology certification, the accuracy check/calibration findings
and potential corrective action, and record verification. The record retention schedule for these
documents should match any federal, state, local, or company requirements for the documents that
can recreate the history of the product.
Conclusion:
A Calibration Program is necessary to ensure that the instruments and devices used to
measure key processing or product parameters are accurate. These suggested steps will help any
food plant set up a program so they can rest easy knowing that their measuring devices are reading
correctly.