System Control Cabinet Layout
System Control Cabinet Layout
Legend:
S W = Water Lines
= Signal Cables
Shield 1
VCIM console CJ712 27676A S or = Optional Selectable in GTC
12 2 X Ray On Light TB2
X-Ray ON
13 3
18 Shield 4 X-Ray on Light CJ2 50001 S
X-Ray 5
Power ON RT Bus CJ301
Lamp 6
19 Shield 7
16 8 24 V
X-Ray 9
Without Fluoro UPS 53030
Lamp
Door 100 wires signal table data #1 BJ400 27672A S CONTROL BOOTH
17 Switch Dose Monitor Control
100 wires signal table data #2 BJ201 27673A S 29183 DISPLAY
J3
RT Bus (From RTAC Gitane board) J22 25 wires UIF table power data BJ403 27674A P S 27126A ION CHAMBER A
F10
BJ5 Injector 24 wires Motor power AJ200 27669A P S 27126B ION CHAMBER B
Ethernet (Fluoro UPS J48) Ethernet RJ POS J5 27662A S ensure that the ECG
power cable is
RJ45 for FE Laptop J8 27505A S connected to a line that Control Booth Live Monitor
is protected against
short circuit hazards,
V 34004 VIDEO IN
Keyboard, VGA, Mouse extender J6 27493A S according to local
P
regulations and that
the ECG monitor is
powered according to
Video Splitter IEC60601-1
Splitter 1 Out2 34004 V requirements (leakage
Live Control Booth Simul Monitor
current and
grounding).
Splitter 1 Out1 34007/34070 V C9 V 34006 VIDEO IN
Live
Splitter 2 Out2
Review 34005 V C13
Hospital Main Power Control Booth Review Monitor
480 VAC
Splitter 3 Out2 34006 V C13 AW Video Splitter
Simul
34009/34073
C Splitter 3 Out1
Simul 34072 V V
C
AW Workstation Central Unit
With keyboard and mouse
P
POS-TAB-SYS BULKHEAD AW Power Strip
To 230Vac
4 OUTLETS AW Workstation Monitor #2
P G DL
DL Infra Red Receiver J15 28195 S P
Power
Distribution DL Keypad J14 27494A S AW Workstation Monitor #1
Box
(Wall PDB) V
Image box Control Booth Power Strip P
System ON/OFF Loop 27682A S
S 27523A Detector Chiller Control J17 Fiber Channel J7 27512A S P 29012A
P 27472A 5 OUTLETS
Provided by installator P 34075
JEDI POWER 3 Phases P
2
P 29010A
System Control Cabinet/PDU Provided by installator Detector power CAN Bus J11 29172A S
Power 3 Phases P G
P 29011A
Secured Power
Strip 3 RED OUTLETS
Provided by installator
Injector P 29024A
Main Ground Wire G EXAM ROOM FLOOR G
Power Terminal
To PDB P S
Emergency UPS Stop 27561A S P JEDI Power Input 3 phase
D P
29285A P Live Monitor D
POS & PDU Power Input 3 phase
Provided by installator P G
Chiller 3 Phase
AWG6
29060
Jedi
AWG6
STRIP UPS J2 VCIM On-Off Console with
27675A AC IN / TB1 (1-2)
FUSE BOX Handswitch
PDB
AWG10 AWG12
53022
LC Gantry
PDU & BUS BAR 28254A S SRB POWER SUPPLY
DL Keypad
J5 J6 J8 J4B
29172A S P 27586A J1 Detector PS Power & CAN Signals
20A Non UPS Not Used UPS Not Used J3 Detector Power Supply Out S 27494A Serial Link with Power
C1-
C1-CABINET
PDU Power 2 phase SUSPENSION 28189 G
Positioner DETECTOR L BUS BAR
P Detector Supply
27470A G
Not Used J2B
C2-
C2- Optic fiber L BUS BAR 27471A FE Laptop
S 27512A w/ bulkhead
G BUS BAR from PDB Not Used J3A
W 27511A Detector water S 27505A NO BULKHEAD (Free)
pipe in flow
SERVICE PWS J15 BUS BAR 27468A G Detector water L BUS BAR 27030 G
27510A
pipe out flow
System Control Cabinet See Connecting Details *
BUS BAR 27473A
[
TUBE
J4A/J2A Strip AC BUS BAR
* Connecting Details 27472A P G X-Ray tube water pipe
(For US only) Control Booth Power Strip W 27666A
out flow GND 27030 G
S 27561A Emergency UPS Stop J17
[
X-Ray tube water Remote Box (Optional)
J9 Strip AC 27667A pipe in flow
BUS BAR 28186 P G P S 27452A
Ceiling Power H 27589A ANODE +
TB02
P/N 2364241
27663A X-Ray tube ingrid 3 mon 28045
SUSPENSION
CONNECTION BOX
J100 Power COLOR 1
Not used
16A
25A
P 27669A V VGA
F AWG10 AWG12 motor
F
2772513
7
Detector Conditioner
2358 531
2315
C1 AWG10 AWG12
S 27670A J206 Collimator
PDB
AWG10
COLOR 2
WATER OUTLET 27511A S 27126A ION CHAMBER A
C1 - CABINET
S 27523A Chiller control RS232
B13
S 27126B ION CHAMBER B
20A
X-Ray Tube Water Chiller RS232 28254A S
53022
OR
G G
3 LCD Ergo Monitor Suspension 4 LCD Monitor Dual Arm Suspension
SUSPENSION CONNECTION BOX SUSPENSION CONNECTION BOX SUSPENSION FRAMEHOLDER
SUSPENSION FRAMEHOLDER
OR 36 m WITH MONITORS
WITH MONITORS
HOSPITAL UPS LINE 28121 P ECG POWER P 28122 P
P-U ECG VIDEO POWER - AC IN
P 28188 Power 28125/6/7 P 3x MONITORS POWER – AC IN P 28188 Power 28128>31 P 4x MONITORS POWER – AC IN
28186 P G MONITORS POWER 28186 P MONITORS POWER strip
P-NV strip P-NV
28154>57 G 4xMONITORS GROUND
Extension
G 28192 Column 28603
UPPER SPRING ARM 28605 G LOWER BEAM GROUND
28189 G GROUND G 28192 COLUMN 28191 ARM 28191 28189 G GROUND
28601 Spring upper joint 28609 28604 G UPPER BEAM GROUND
34070 V V V
IN V
34003 V 28045 V V
34009 V VIDEO x-CAT7-x V VIDEO VIDEO COLOR AW
V V
IN 28048 V
28056 V
28045 V VIDEO/VGA V VIDEO
CAUTION: For proper operation and regulatory compliance, it is mandatory to follow the guidelines described
in direction 46-019668, System Facility Power and Ground Requirements
REVISIONS
APPROVALS YY/MM/DD TITLE
K NOTE: All sheets of this schematic will be at the same revision of this sheet. MADE
REV
K
REV MADE BY DATE PCN NO. DESCRIPTION -
The revision of this sheet will be the revision of the entire document. - Optima IGS 330/320/Plus System MISMAPS
1 Cheng Guoxing 2020-Oct-26 - Initial Release
CHECK
- 4
- FIRST MADE FOR:
APPR
-
5849801-1EN
OTHER GENERAL ELECTRIC
- - SHEET 1 OF 1
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16