System Control Cabinet Layout

The document provides a legend for various cable types used in a technical room, including signal cables (S), water lines (W), video cables (V), internal cables (I), high voltage cables (H), power cables (P), ground wires (G), and field supplied cables (FS). It also shows the connection details between a power distribution box (PDB), positioner bulkhead, system control cabinet, and various cable runs and MIS numbers.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
382 views1 page

System Control Cabinet Layout

The document provides a legend for various cable types used in a technical room, including signal cables (S), water lines (W), video cables (V), internal cables (I), high voltage cables (H), power cables (P), ground wires (G), and field supplied cables (FS). It also shows the connection details between a power distribution box (PDB), positioner bulkhead, system control cabinet, and various cable runs and MIS numbers.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 1

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16

Legend:

S W = Water Lines
= Signal Cables

V = Video Cables I = Internal Cable


H = High Voltage Cables
34 = Cable Run Numbers
P = Power Cables

G = Ground Wires FS = Field Supplied Cables


** Connecting Details TECHNICAL ROOM (Independent of
Power Cables) 11198A = MIS Numbers
PDB Positioner Bulkhead
20 System Control Cabinet
Room POS-TAB-SYS BULKHEAD
Light
A Control
21
TB3 27682A S PDB On-Off CJ 512/1 w/Bolus chasing AJ128 27083A P S A
Ground Room or
14 Light = Not included in GTC
Switch Room Light TB3
(TSSC)
15
TB2 Door Interlock TB2

Shield 1
VCIM console CJ712 27676A S or = Optional Selectable in GTC
12 2 X Ray On Light TB2
X-Ray ON
13 3
18 Shield 4 X-Ray on Light CJ2 50001 S
X-Ray 5
Power ON RT Bus CJ301
Lamp 6
19 Shield 7
16 8 24 V
X-Ray 9
Without Fluoro UPS 53030
Lamp
Door 100 wires signal table data #1 BJ400 27672A S CONTROL BOOTH
17 Switch Dose Monitor Control
100 wires signal table data #2 BJ201 27673A S 29183 DISPLAY
J3
RT Bus (From RTAC Gitane board) J22 25 wires UIF table power data BJ403 27674A P S 27126A ION CHAMBER A
F10
BJ5 Injector 24 wires Motor power AJ200 27669A P S 27126B ION CHAMBER B

15 wires signal cable AJ300 27670A S

100 wires signal cable data AJ 202 27671A


Bolus Wallbox
Buzzer kit PWR J49 Buzzer kit
29087 P
B in exam room
WARNING: Prior to
P S 27083A W/BOLUS CHASING
B
DICOM ETH J9 S Hospital
Network each installation, S 27041A BOLUS CHASING

Ethernet (Fluoro UPS J48) Ethernet RJ POS J5 27662A S ensure that the ECG
power cable is
RJ45 for FE Laptop J8 27505A S connected to a line that Control Booth Live Monitor
is protected against
short circuit hazards,
V 34004 VIDEO IN
Keyboard, VGA, Mouse extender J6 27493A S according to local
P
regulations and that
the ECG monitor is
powered according to
Video Splitter IEC60601-1
Splitter 1 Out2 34004 V requirements (leakage
Live Control Booth Simul Monitor
current and
grounding).
Splitter 1 Out1 34007/34070 V C9 V 34006 VIDEO IN
Live

Splitter 2 Out2
Review 34005 V C13
Hospital Main Power Control Booth Review Monitor

3 x 380 to Splitter 2 Out1


Review
34008/34071 V C9 V 34005 VIDEO IN

480 VAC
Splitter 3 Out2 34006 V C13 AW Video Splitter
Simul
34009/34073
C Splitter 3 Out1
Simul 34072 V V
C
AW Workstation Central Unit
With keyboard and mouse
P
POS-TAB-SYS BULKHEAD AW Power Strip
To 230Vac
4 OUTLETS AW Workstation Monitor #2
P G DL
DL Infra Red Receiver J15 28195 S P
Power
Distribution DL Keypad J14 27494A S AW Workstation Monitor #1
Box
(Wall PDB) V
Image box Control Booth Power Strip P
System ON/OFF Loop 27682A S
S 27523A Detector Chiller Control J17 Fiber Channel J7 27512A S P 29012A
P 27472A 5 OUTLETS
Provided by installator P 34075
JEDI POWER 3 Phases P
2
P 29010A

System Control Cabinet/PDU Provided by installator Detector power CAN Bus J11 29172A S
Power 3 Phases P G
P 29011A
Secured Power
Strip 3 RED OUTLETS
Provided by installator
Injector P 29024A
Main Ground Wire G EXAM ROOM FLOOR G
Power Terminal
To PDB P S
Emergency UPS Stop 27561A S P JEDI Power Input 3 phase
D P
29285A P Live Monitor D
POS & PDU Power Input 3 phase
Provided by installator P G
Chiller 3 Phase

To PDU Power 2 phase KVM - R Mouse


Injector P
Omega Table S 27493A
DL Keyboard
JEDI X-Ray tube stator cde DJ2 27664A S S 27662A RJ45 Ethernet
Tube SW &Oil Pump Power V
Door Light X-Ray tube security DJ3 27665A S Injector (J2) S
BUS BAR
System ON G 27468A
From Injector X-Ray tube Ingrid DJ1 27663A L BUS BAR G DL Flat Panel
1 27473A BUS BAR
* Connecting Details S 27677A X-Ray Tube Chiller Control DJ4 Anode + 27589A H
(Not for US) w/ Bolus chasing only 27041A P S 29041A V P
Cathode - 27590A S 27672A J600 Table Data #1
TB02 RS232 J3 (Generic IF Board) 28245
ITU Interface Board
3x 160A
27673A J601 Table Data #2
Relay command
AWG6
J3 ITU
2347573

P 27674A J605 Table Pwr Data


27711/
29059/

AWG6
29060

Jedi
AWG6
STRIP UPS J2 VCIM On-Off Console with
27675A AC IN / TB1 (1-2)
FUSE BOX Handswitch
PDB

P/N 2364241 L BUS BAR


27676A 27676A or
L BUS BAR 27470A G S 29044A
Signals and 24Vdc
16A
25A
J47 System Bulkhead
AWG10 AWG12 STRIP NON UPS L BUS BAR 27471A
2315513
27727
2358531

AWG10 AWG12
53022

PDU P/N 2305956-


5491772
E E
AWG10
AWG10
AWG10

LC Gantry
PDU & BUS BAR 28254A S SRB POWER SUPPLY
DL Keypad
J5 J6 J8 J4B
29172A S P 27586A J1 Detector PS Power & CAN Signals
20A Non UPS Not Used UPS Not Used J3 Detector Power Supply Out S 27494A Serial Link with Power
C1-
C1-CABINET
PDU Power 2 phase SUSPENSION 28189 G
Positioner DETECTOR L BUS BAR
P Detector Supply
27470A G
Not Used J2B
C2-
C2- Optic fiber L BUS BAR 27471A FE Laptop
S 27512A w/ bulkhead
G BUS BAR from PDB Not Used J3A
W 27511A Detector water S 27505A NO BULKHEAD (Free)
pipe in flow
SERVICE PWS J15 BUS BAR 27468A G Detector water L BUS BAR 27030 G
27510A
pipe out flow
System Control Cabinet See Connecting Details *
BUS BAR 27473A

[
TUBE
J4A/J2A Strip AC BUS BAR
* Connecting Details 27472A P G X-Ray tube water pipe
(For US only) Control Booth Power Strip W 27666A
out flow GND 27030 G
S 27561A Emergency UPS Stop J17

[
X-Ray tube water Remote Box (Optional)
J9 Strip AC 27667A pipe in flow
BUS BAR 28186 P G P S 27452A
Ceiling Power H 27589A ANODE +
TB02

3x 160A 27590A CATHODE - G 27508A ECG Control Unit (Optional)


2 wires AC power cable CB600 P
AWG6
P 27522A Detector Conditioner Power J3B Omega Table Power 27675A
S S 27560A* Data Reserved
2347 573

AWG6 27664A X-ray tube stator cde J1


AWG6
Jedi
27665A X-ray tube temp status 36 m * NOTE:
FUSE BOX
Cable 27560A for ECG Control Unit is not used.
PDB

P/N 2364241
27663A X-Ray tube ingrid 3 mon 28045
SUSPENSION
CONNECTION BOX
J100 Power COLOR 1
Not used
16A
25A
P 27669A V VGA
F AWG10 AWG12 motor
F
2772513
7

Detector Conditioner
2358 531

2315

C1 AWG10 AWG12
S 27670A J206 Collimator
PDB

PDU P/N 5491772


2305956 - 2
WATER INLET 27510A W 27671A J205 Data LC SUSPENSION 4 mon 28048
P 27522A Chiller AC power 3 wires CONNECTION BOX
AWG10

AWG10

COLOR 2
WATER OUTLET 27511A S 27126A ION CHAMBER A
C1 - CABINET
S 27523A Chiller control RS232
B13
S 27126B ION CHAMBER B

20A
X-Ray Tube Water Chiller RS232 28254A S
53022

C2 Positioner S 27677A Remote Control WATER OUTLET 27666A W


C2 - CABINET
System Control Cabinet G P 1TB - AC POWER & GROUND WATER INLET 27667A

OR

G G
3 LCD Ergo Monitor Suspension 4 LCD Monitor Dual Arm Suspension
SUSPENSION CONNECTION BOX SUSPENSION CONNECTION BOX SUSPENSION FRAMEHOLDER
SUSPENSION FRAMEHOLDER
OR 36 m WITH MONITORS
WITH MONITORS
HOSPITAL UPS LINE 28121 P ECG POWER P 28122 P
P-U ECG VIDEO POWER - AC IN

36m FRAME FRAME

P 28188 Power 28125/6/7 P 3x MONITORS POWER – AC IN P 28188 Power 28128>31 P 4x MONITORS POWER – AC IN
28186 P G MONITORS POWER 28186 P MONITORS POWER strip
P-NV strip P-NV
28154>57 G 4xMONITORS GROUND
Extension
G 28192 Column 28603
UPPER SPRING ARM 28605 G LOWER BEAM GROUND
28189 G GROUND G 28192 COLUMN 28191 ARM 28191 28189 G GROUND
28601 Spring upper joint 28609 28604 G UPPER BEAM GROUND

LOWER Spring 28602 Spring lower joint


28216 FRAME 28151/2/3 G 28607>08 G TWO BEAM REAR COVER GROUND
COLUMN 3x MONITORS GROUND
50001 S X-Ray on Light
28606

34070 V V V

34007 V V 34001 V IN 34071 V V V CAT7-1 CAT7-2


VIDEO VIDEO
H 34008 V
x- CAT7-x
V 34002 V VIDEO
34072 V V V H
In Room 34073 V V V
2nd 2nd 4-4
2nd shielding shielding CAT7-3 CAT7-2 CAT7-1 2nd 2nd
shielding
shielding 2nd shielding
/VGA shielding CAT7-3 CAT7-4
Review /VGA /VGA
AW Live

IN V
34003 V 28045 V V
34009 V VIDEO x-CAT7-x V VIDEO VIDEO COLOR AW
V V
IN 28048 V
28056 V
28045 V VIDEO/VGA V VIDEO

INFRARED REMOTE RECIEVER INFRARED REMOTE RECIEVER


I/F RECEIVER s s I/F RECEIVER s 28197 s
28195 s F-1
28197 s SIGNAL 28195 SIGNAL

DOSE MONITOR DISPLAY 1


DOSIMETER DISPLAY 1 s
J B13 29183 s F-3 s 28200 DOSE
J

CAUTION: For proper operation and regulatory compliance, it is mandatory to follow the guidelines described
in direction 46-019668, System Facility Power and Ground Requirements

Optima IGS 320, Optima IGS 330, Optima IGS Plus


MISMAPS

REVISIONS
APPROVALS YY/MM/DD TITLE

K NOTE: All sheets of this schematic will be at the same revision of this sheet. MADE
REV
K
REV MADE BY DATE PCN NO. DESCRIPTION -
The revision of this sheet will be the revision of the entire document. - Optima IGS 330/320/Plus System MISMAPS
1 Cheng Guoxing 2020-Oct-26 - Initial Release
CHECK
- 4
- FIRST MADE FOR:

APPR
-
5849801-1EN
OTHER GENERAL ELECTRIC
- - SHEET 1 OF 1