Amplified Pressure Sensors
-10 cm H2O to120 cm H2O Pressure Sensor
Features
• 4 Volt ratiometric output
• Temperature Compensated
all sensors
• Calibrated Zero and Span
Applications
• Medical Instrumentation
• Respiratory Breathing
e www.allsensors.com
General Description
This Amplified Output pressure sensor is based upon a proprietary technology to compensate all errors. This model
provides a calibrated amplified output with superior output characteristics. Output characteristics are tested at
pressure and temperature and then digitally compensated using a proprietary 3D mapping scheme. In addition the
sensor utilizes a silicon, micromachined, stress concentration enhanced structure to provide a very linear output to
measured pressure.
f 408 225 2079
These calibrated and temperature compensated sensors give an accurate and stable output over a wide temperature
range. This series is intended for use with non-corrosive, non-ionic working fluids such as air, dry gases and the
like.
p 408 225 4314
The output of the device is ratiometric to the supply voltage over a supply voltage range of 4.5 to 5.5 volts.
Physical Dimensions Equivalent Circuit
0.25 MIN Pin Size 0.10 Typical Vs
(6.4) 0.010 x 0.020 (2.54)
(0.25 x 0.50)
Pin 1 2 3 4
a 6296 San Ignacio, Suite E & F, San Jose, CA 95119
0.174 to 0.190
(4.4 to 4.8)
1
0.21 - Vout
(5.3)
+
A 0.42 Gnd
1.08 (10.5)
(27.4)
0.66
(16.0) B
pin 1: Vs
0.48 0.55 0.80
(12.2) (14.0) (20.3) pin 2: Gnd
NOTES: 0.85 pin 3: Vout
1) Dimensions in inches (millimeters) (21.6)
1.10 pin 4: do not connect
(27.9)
Approvals
MKT DATE MFG DATE ENG DATE QA DATE
As Is With Change As Is With Change As Is With Change As Is With Change
All Sensors DS-0027 Rev B
Pressure Sensor Characteristics Maximum Ratings Environmental Specifications
Supply Supply Voltage VS +4.5 to +5.5 Vdc Temperature Ranges
Common-mode pressure 10 psig Compensated see specification
Lead Temperature 250°C Operating -25 to 85° C
(soldering 2-4 sec.)
Storage -40 to 125° C
Humidity Limits 0 to 95% RH
(non condensing)
Standard Pressure Ranges
Part Number Operating Pressure Compensated Range Proof Pressure Burst Pressure
120 CMH2O-D-4V -10 to 120 cm H2O 5 to 50° C 300 cmH2O 600 cmH2O
120 CMH2O-D-4V-PRIME -10 to 120 cm H2O -25 to 85° C 300 cmH2O 600 cmH2O
Performance Characteristics for 120 CMH2O-D-4V
Parameter, note 1 Minimum Nominal Maximum Units
Operating Range, pressure -10 -- 120 cmH2O
Output Voltage, @120 cmH2O 4.40 4.5 4.6 volt
Output Voltage @ zero pressure 0.30 0.35 0.40 volt
Output Voltage @-10 cmH2O 0.23 0.28 0.33 volt
Offset Temperature Shift (5°C-50°C), note 2 -- -- ±1.0 %span
Linearity, hysteresis error, note 4 -- 0.05 ±0.25 %fs
Span Shift (5°C-50°C), note 2 -- -- ±1.0 %span
Performance Characteristics for 120 CMH2O-D-4V-PRIME
Parameter, note 1 Minimum Nominal Maximum Units
Operating Range, pressure -10 -- 120 cmH2O
Output Voltage, @120 cmH2O 4.40 4.5 4.6 volt
Output Voltage @ zero pressure 0.30 0.35 0.40 volt
Output Voltage @-10 cmH2O 0.23 0.28 0.33 volt
Offset Temperature Shift (-25 to 85°C), note 2 -- -- ±1.0 %span
Linearity, hysteresis error, note 4 -- 0.05 ±0.25 %fs
Span Shift (-25 to 85°C), note 2 -- -- ±1.0 %span
Specification Notes
note 1: All parameters are measured at 5.0 volt excitation, for the nominal full scale pressure and room temperature unless
otherwise specified. Pressure measurements are with positive pressure applied to PORT B.
note 2: Shift is relative to 25°C.
note 3: Shift is within the first hour of excitation applied to the device.
note 4: Measured at one-half full scale rated pressure using best straight line curve fit.
note 5: The voltage added to the offset voltage at full scale pressure. Nominally the output voltage range is 0.35V to 4.5V
volts for minus to plus full scale pressure.
All Sensors reserves the right to make changes to any products herein. All Sensors does not assume any liability arising out of the application or use of
any product or circuit described herein, neither does it convey any license under its patent rights nor the rights of others.
Amplified Pressure Sensors