Philips Optimus Rad RF Release 3.3 Xgrscope 3.14 Help Manual
Philips Optimus Rad RF Release 3.3 Xgrscope 3.14 Help Manual
24 22335 Hamburg
Service Support & Field Assistance Axel Duve [email protected] Box141
March 28, 2001 OM_33_rev2.doc PC version page 1-52
XRGSCOPE 3.14 can be used for Optimus generators release 3.3 and 2.3 with one exception for tube
programming: Release 3 generators must use the TUBE_R3.TDL file, release 2 generators TUBE_R2.TDL.
Previous Optimus releases ( 0.2 / 0.3 / 1.1 / 2.1 / 2.2 / 3.1 / 3.2 ) are no longer supported.
Text explanations in ‘italic letters and quotation marks’ refer to bold and underlined screen or item names.
Some bold and underlined screen names have red FILE NAMES added. These files have been saved with
<F3> online with a generator. To open the files call XRGSCOPE, push <ESC> when it says "Attempting link
to device ...", then use "File" and select the file name mentioned. It allows to program the screen within its
limits and selectable items as if one is online with the generator.
Note:
Whenever using XRGSCOPE, start from DOS if possible. Using WIN95/98 might cause some unexpected
problems (WIN3.x also) depending on the number of multitasking background programs.
XRGSCOPE runs best under pure DOS. If it runs in a DOS-box under WIN 3.x, the file dosprmpt.pif (or
possibly a special self-made xrgscope.pif) should be set to exclusive and full-screen-mode, which can be
done with the PIF-EDITOR.
XRGSCOPE is not tested with WIN95/98. In case of problems they may be solved if WIN95/98 is started in
MSDOS-MODE.
WIN NT requires special drivers to get access to the PC hardkey, available on the customer support CD-
ROM 2422 210 4220x.
To ensure safe loading procedures from and to the generator screensavers and any kind of WIN and PC
power management must be switched off. These should never be active when using XRGSCOPE.
Do not start XRGSCOPE or FLASH1/2 with NORTON, some loading process might take hours !
There are two disks with the documentation. Each disk has a self-extracting OMA0220x.EXE and
OMB0230x.EXE file. The contents of the packed.EXE files are listed in the OMA/OMBxxxx.TXT files
attached. A free disk space of 5 MB is required.
Create a directory on the PC hard disk (e.g. C:\OPT_R33) and load all files of both installation disks into the
new DIR (copy A:\*.* C:\OPT_R33). Run both .EXE files to inflate the programs.
Note! Since OMA 02204.EXE a new tube firmware TUBE_R3.TDL exists to drive the new version of rotor
control 4512 104 71421 or 4512 104 71461 with firmware 4512 113 22322. It also runs on all predecessor
rotor controls 4512 104 71401...6 with firmware 4512 113 22312.
Generators (preinstalled or as replacement) and spare CU-PCBs are always delivered with Release 3.3.
Just in case release 3.3 is not loaded in the flash-PROMs it has to be loaded first. Afterwards the generator
can be programmed:
FLASH1.BAT (COM1) or
FLASH2.BAT (COM2).
Philips Medical Systems DMC Röntgenstr. 24 22335 Hamburg
Service Support & Field Assistance Axel Duve [email protected] Box141
March 28, 2001 OM_33_rev2.doc PC version page 2-52
Now the firmware will be loaded automatically to the Flash-PROMs on CU EZ139. It takes 15...50 minutes
depending on the processor type of the PC.
- !! After the end of the loading procedure (100%) one must wait until the message
Flash Loaded
<OK>
comes up.
Restart normal.
The normal communication (e.g. to run the setting to work) to the generator can be started now.
If the time between ‘Flash Loaded <OK>‘ and pushing of <ENTER> on the PC takes too long <ESC> has to
be pushed once or twice on the PC.
If an original NVRAM still exists, the control desk will come up as usual with the parameters and settings
when the generator has been switched off.
If nothing has been programmed or if e.g. the mA_control PCB has been exchanged and no tube has yet
been programmed, the control desk will only display PHILIPS OPTIMUS and dots will fill all APR segments.
Then the generator has to be programmed according to the local configuration.
Starting of XRGSCOPE
If the command XRGSCOPE is used in a DOS environment the program will start asking for the hardkey
password.
Without the hardkey connected the access is restricted.
In a WIN 95/98/NT environment the command line has to be set according to the PC configuration.
File
Whenever XRGSCOPE is started, the program will first of all display the screen CUSTOMER.TDL (with what
is actually stored in CUSTOMER.TDL, details see ‘Customer Data’). The default screen is empty.
Note: During every log-in an entry 00S* will appear in the error log index to indicate service access to the
generator (including APRMANager). To prevent from getting too many 00S* entries one can leave
XRGSCOPE active on the PC and turn the generator off. After turn-on the communication generator
starts at the same screen which is open on the PC (even if the communication cable has been
removed when the generator was off).
Allows to open data files of the current DIR from which XRGSCOPE has been started. Files can be
opened and the content can be displayed. Some files might be too big to be displayed, then a screen
table too big will come up. It depends on the free DOS memory and the file size.
To enter in the file list push the <TAB> button and select with the cursor buttons or type in the file name.
If xxx.TDL files shall be displayed from a floppy disk, just type A: in the command line and the file list will
be displayed. ‘Open Data File’ can also be used offline.
Whenever data screens like ‘error log index’ are saved to a xxx.TDL file (function Save: <F3> appears in
the bottom line), ‘customer data’ being saved in the CUSTOMER.TDL file are attached to the saved data
screen. It helps to separate saved screen files of different sites, customers or rooms in the same hospital.
Site data must be stored in CUSTOMER.TDL file, only data of this file are attached to the saved screens.
One can save site specific customer data in self-made files using <F3>. To recall site data use the Load:
<F4> function. <F4> does not appear in the bottom command line, but it is possible to use it.
Procedure: Either the CUSTOMER.TDL screen is open or open the ‘Customer Data’ screen. Push <F4>
and select a site data file. The old data screen comes up. Now save this screen with <F3> typing in
CUSTOMER as file name.
Notes:
Several screens, once transmitted with <F2>, are followed by a screen
It is recommended to reset the generator whenever the screen says so. Then one can be sure to get the full
functionality of the generator afterwards.
If a generator has to be programmed from scratch one can program as many steps in a row as necessary
and make one reset at the end of the programming procedure. It saves a lot of time.
If only a change of a programming screen takes place (e.g. ‘RGDV x Data Set A’) it would be sufficient to
push the RGDV button at the control desk after Transmit <F2> to validate the modified data.
A down-pointing arrow at the end of a programming line allows a selection of values or options after pushing
<ENTER>. Select items with the cursor and push <ENTER> or <F2>.
[Square bracket ] fields allow to enter values within a given range. If the entered value is out of range the
limits will automatically be displayed. Value fields without brackets can not be changed, for reading only.
- Mains data - U_mains_nominal [V] available values = 380 / 400 / 440 / 460 / 480 V
MAINS default value = 400V
(for 415V select 400V)
!! It is important to select (U) and type in [Ri] the true values. The calculation of the
!! kV_control for the duty cycle to drive the IGBT's is based on that.
The max kW limit and with it the max mA is programmed in the function key EZ139 D38
(depending on the customer order, see STAMMKARTE on frontal kV power unit).
Philips Medical Systems DMC Röntgenstr. 24 22335 Hamburg
Service Support & Field Assistance Axel Duve [email protected] Box141
March 28, 2001 OM_33_rev2.doc PC version page 5-52
- Tubes
- Tube 1...3
Note! Since OMA 02204.EXE a new tube firmware TUBE_R3.TDL exists to drive the new
version of rotor control 4512 104 71421 or 4512 104 71461 with rotor control firmware
4512 113 22322.
It also runs on all predecessor rotor controls 4512 104 71401...6 with rotor control firmware
4512 113 22312.
After 35 sec of Reading (spinning bar) the screen Load Data from Disk comes up, offering
the default TUBE_R3.TDL. file
With <OK> a screen Available Tubes (Rel.3 Format) offers a selection of all available
tubes in non-cooled housings ROT350 (250W) or water cooled housings ROT351 (500W).
After tube selection a screen Analyzing will come up, followed by Transmitting and then
Percent transmitted (how much of file has been loaded to the generator). Wait until the PC
comes up with the message to reset the generator:
D O N O T R E S E T B E F O R E !!
!! In case of a single focus tube like RO30: APR's loaded to a RGDV with this tube
!! connected should only be programmed on the large focus (use APRMANager).
The speed for the tube type and technique is automatically taken from the tube data file.
Fluoroscopy without rotation is not possible.
Tube: 1 ... 3
The value must be programmed according to the max kV of the tube (see tube label). If a
tube arcs during adaptation the max kV value can be reduced (or repeat the break in
procedure, see doc INSTALLATION). After adaptation the programmed value is the max
value which can be selected at the control desk (the adaptation table has no values beyond
the limit kV value), even if a higher value is programmed thereafter in this field. The kV value
with which the tube has successfully been adapted appears in 'Adapted To [kV]:'.
If a tube has been adapted with a higher kV than it should be used for (e.g. max 100kV for
veterinary surgeons) the limit must be programmed in this data field afterwards.
The middle focus would be a third physical filament in a tube (not yet available), it is not the
VARIO focus.
Philips Medical Systems DMC Röntgenstr. 24 22335 Hamburg
Service Support & Field Assistance Axel Duve [email protected] Box141
March 28, 2001 OM_33_rev2.doc PC version page 6-52
All standard diagnostic tubes of the TUBE_R3.TDL tube file have a default min kV value of
40 kV, it can only be modified in a range > 40kV.
Adapted To [kV]: gives the max available kV value at the control desk (after adaptation)
If this value shall be increased, first the 'Max Tube Voltage Limit [kV]:' must be increased.
Then the tube has to be readapted afterwards to have the higher kV available for application.
0.1 mA is the lowest current of the Optimus (fluoroscopy and TDC only).
The lowest mA value one can select at the control desk for exposures is 1 mA.
The kV dependent max mA value is automatically limited by the tube type and the generator
version:
OM 50 kW max = 650 mA OM 65 kW max = 900 mA OM 80 kW max = 1100 mA
The max tube mA value is initially set by the tube data set of the tube file and can not be
increased. It will be adapted to the individual focus limits during adaptation (within the total
tube limits), the max mA table can be seen in 'Select Unit' >> 'FU_mA' >>'Programming'.
!! If there is any application reason to decrease the max mA value it is possible, but it will
!! influence all registration devices attached to this tube and might lead to problems, if
!! additional reductions of the emission current and tube power are programmed in the
- Capacitance Tube Connection - formula and table = see doc INSTALLATION CABLECAP
Recommended for children hospitals and casualty rooms (long preparation times).
- rotation prolongation after prep: !! Only possible with HIGH SPEED rotor control units !!
disable = (default)
Anode stops immediately after exposure end or let go of the PREP hand-switch
if PREP has been started from standstill.
Philips Medical Systems DMC Röntgenstr. 24 22335 Hamburg
Service Support & Field Assistance Axel Duve [email protected] Box141
March 28, 2001 OM_33_rev2.doc PC version page 7-52
enable = Once started with PREP from standstill the tube keeps rotating for 30 seconds
after the last let go of PREP as long as no exposure has been switched.
The tube finally stops if an exposure has been switched (or after 30 sec).
Recommended for children hospitals and casualty rooms.
- For exposure series, if ‘Exposure series / Tomo movement’ = yes has been
programmed at RGDV ‘Data Set A’.
Exposure series have to be started with a fluoro command, afterwards the tube
keeps 9000rpm exposure speed for 30 seconds after the last prep or fluoro.
- Disable Tube
- If a tube (e.g. 2nd tube) shall be removed from a system one should use this function. If a tube
shall be removed temporarily one should do it at the registration device ‘RGDV1...8 Data Set A’.
!! In combination with ‘Bucky Controller 1’: DO NOT use this function before all RGDV associated
!! to ‘Bucky Controller 1’ have been changed to ‘none’ instead of ‘Bucky Controller 1’.
- This function must be used if a double focus tube shall be replaced by a single focus tube. The
non-used small filament circuit of FU_mA must be switched off.
- If a double focus tube shall be replaced by any other type, it is not required to disable the tube.
Just load the other tube type according to ‘Tube 1...3 Data Set’.
- Registration Devices
- RGDV 1 ... 8
- Room: Room 1...3 for room door contact and radiation warning lamp wiring see
doc INSTALLATION, WG unit required if more than 1 room
- Tube: None RGDV not available, ‘None’ can be used to disable a RGDV selection
without erasing the tube data in mA_control and CU
- one out of four release circuits of a 'Release circuit adaptation unit' 1WA, 2WA, 1WB, (2WB
not yet available)
(e.g. Adapter for 4 aux. units WA: 1 = EWAX1 2 = EWAX2 3 = EWAX3 4 = EWAX4)
- ignore in case of a 'Bucky controller 1' or 'Thorav. / Dig. Bucky'
- ignore, if 'none' is programmed for a 'Release circuit adaptation unit' (e.g. free
cassette RGDV without cassette present interlock)
RGDV number and 'Release circuit number' do not have to be the same, if required all 8
RGDV's can be programmed to one release circuit only.
yes = release with the desk hand-switch and via release decade of WA or WB possible
double step = individual PREPARATION and EXPOSURE request with the desk hand-
switch or via the release decades of WA/WB
single step = instant EXPOSURE request with PREPARATION only, released from
!! If 'intermediate boost' mode has been 'enable'd to activate double boost mode, it will
automatically be off if 'single step' is programmed.
- Bucky format density correction (6% steps): range = - 8 ... + 8 steps see [1] page 25
- correction during collimation, input at WAX11/12 pin 1-2 or WBX11 pin 1-2,
side-fields active when contact closed (>24x24cm), center field only when open
- with GALILEO collimator Bucky TH, TH2, Thoravision or Digital Diagnost, ask application
- Cone density correction (6% steps): range = - 8 ... + 8 steps see [1] page 25
- correction when cone in use, input at WB X13 pin 4-5, active when closed, Optimus R/F only
- Dose measurement input:
- EZX61 in use for dose rate inputs, in this case the exposure
on signal X_ACT/ has to be connected to the central part
via signal bus
EZX21 = chamber 1\
EZX22 = chamber 2 \ to be used for
EZX31 = chamber 3 >backpanel input plugs Amplimat and/or
EZX32 = chamber 4 / R/F photo sensor input
EZX41 = chamber 5 / (R/F (e.g.DSI) EZX41 preferred)
Photo sensor input via EZX61, if ‘none’ has been programmed at ‘Dose
measurement input’, the exposure on signal X_ACT/ to the central
part has to be connected via signal bus EZX23
Photo sensor/ampl. inp. photo sensor input only via EZX41 to get the REL signal (as
EXON signal) for AEC and all free techniques, e.g. kV-mAs
enable Must be set for AEC / AECF / TDC / PHOTO TIMER techniques to switch on
kV not before the dose integrator has been reset and the offset of the chamber
signal line has been measured by Dose Rate Control.
Must also be enabled for mobile Amplimat stands without grid sync connected.
- Automatically ‘disable’d if any non-AEC technique is selected.
Bucky controller 1 = RGDV 1...4 BuckyTH and BuckyTH2 via system CAN
RGDV 1...4 Digital Diagnost via system CAN
RGDV 5...8 Bucky unit at easyDiagnost via system CAN
- Release circuit adaptation unit: (Adapter decade cable for 4 aux units RAD - R/F)
none = to be programmed in case of free cassette, mobile cassette stands without grid
sync contact and if no cassette present interlock is required
If an exposure end beeper is required in the examination room (even for free cassette) the
RGDV must be assigned to the ‘Release circuit adaptation unit’ to activate the EXON
optocoupler at WA X13/14. R/F systems provide the beeper driven from WBX13.
- Mounted tomo extension: only, if at least one 'Release circuit adaptation unit' 1WA and/or
2WA option is present
none = no tomo auxilary
Philips Medical Systems DMC Röntgenstr. 24 22335 Hamburg
Service Support & Field Assistance Axel Duve [email protected] Box141
March 28, 2001 OM_33_rev2.doc PC version page 11-52
1WA to activate tomo functions, see also ‘RGDV1...4 Data Set B’ 'Used for tomo'
2WA and RGDV Interface Assignments' 'Decade Bucky 1/2' + 'Tomo Time'
- Medium II format kV corr. (dose equiv. steps): range = 0...8 dose equivalent
+ kV correction steps
(see ‘kV steps’ page 12)
- Medium II format density corr. (-6% steps): range = 0 ... 8 steps of -6% density [1]
- Medium II format mAs corr. (-6% steps): range = 0 ... 8 steps of -6% mAs [1]
- Small II format kV corr. (dose equiv. steps): range = 0...8 dose equivalent
+ kV correction steps
(see ‘kV steps’ page 12)
- Small II format density corr. (-6% steps): range = 0 ... 8 -6% density steps [1]
- Small II format mAs corr. (-6% steps): range = 0 ... 8 -6% mAs steps [1]
yes = - must be set if any 'Mounted tomo extension' is programmed at 'RGDV Data Set A'
(tomo time input active only in combination with option Automatic input of tomo
times 9890 000 0222x)
- must be programmed in connection with a Bucky TH, TH2, Digital Diagnost with
TOMO, which gives access to the 16 tomo times in the System (tomo time input
only in combination with option Automatic input of tomo times 9890 000 0222x)
- if 'yes' is set exposure time +/- corrections are not possible at the control desk
no = no fluoro possible
yes = - only possible in combination with ‘Release decade adaptation unit’ WB and if
programmed in generator function key EZ139 D38, generator base must be R/F
- if fluoro is set to ‘yes’ the menu/APR field at the control desk line 1 column 2 is
reserved to display fluoro parameters (kV, mA, time) and can not be used for menu
or APR names anymore
- CT add on: for therapy simulator purpose, allows run times >= 16 seconds in continuous
fluoroscopy mode
- kV steps:
values: 40 - 41 - 42 - 44 - 46 - 48 - 50 - 52 - 55 - 57 - 60 - 63 - 66 - 70 - 73 - 77 -
81 - 85 - 90 - 102 - 109 - 117 - 125 - 133 -141 - 150 kV
- Density steps: 25 % all RGDV should get the same density step factor
12 % (default) (and should also be equal to generators next door)
6% (SCP and MedioCP steps are always at 12% )
!! The density correction displayed at the desk only shows e.g. -1 or +2. This -1 or +2 can
!! stand for (a) step(s) of 6%, 12% or 25%.
!! All RGDV's (and all systems in the same hospital) should have the same correction factor to
!! get the same result when used. More explanation in 'Change APR Data Set'.
- Density correction (6% steps): range = - 8 ... + 8 steps (see [1] page 25)
yes The incorrect exposure indicator is blinking at the control desk to indicate that an
exposure has not been terminated by the generator (FU_mA controls all non-AEC
techniques kV-mAs, kV-mA-ms, kV-mAs-ms),
e.g. if
- the hand-switch has been released prior to the set exposure time
- the sweep time switched by the tomo table is shorter than the programmed
exposure time
- the exposure switch is bouncing.
on (default), the tube overload protection is on, tube load reductions are active
Philips Medical Systems DMC Röntgenstr. 24 22335 Hamburg
Service Support & Field Assistance Axel Duve [email protected] Box141
March 28, 2001 OM_33_rev2.doc PC version page 13-52
As long as PREP has been started from yellow-red status EXPOSURE is possible
with 100% load, even with red light on. PREP is not possible in red, ready disappears.
- Decade Bucky 1 (X11) e.g. WAX11 and/or Decade Bucky 2 (X12) WAX12
!! Once a RGDV is programmed all functions of WAX11 or WAX12 are active, functions
!! not provided by the system must be "simulated" with a short-link.
For a stand with only one remote switchover contact using the same set of tomo trajectory /
times for cassette and DSI tomo (e.g. D96) it is sufficient to use one remote switch-over
contact only.
none = default
RGDV 1...8
- side field selection for automatic collimators with Bucky chambers, pin 1-2
(closed if > 24x24 cm), density format size correction when programmed
!! if no automatic collimator present insert a link at pin 1-2 to allow the side field
selection
- defines the Bucky RDGV for Bucky/Tomo remote switchover, if 'Tomo mode
switch' is enabled, pin 3-4, (Bucky tomo [DSI tomo] = closed, Bucky [DSI] = open)
- activates the recognition of the Bucky Ready contact pin 9-10 (ready = closed)
none = default
Philips Medical Systems DMC Röntgenstr. 24 22335 Hamburg
Service Support & Field Assistance Axel Duve [email protected] Box141
March 28, 2001 OM_33_rev2.doc PC version page 14-52
RGDV 1...8
- side field selection for automatic collimators with Bucky chambers, pin 1-2
(closed if > 24x24 cm), density format size correction when programmed
!! if no automatic collimator present insert a link at pin 1-2 to allow the side field
selection
- defines the Bucky RDGV for Bucky/Tomo remote switchover, if 'Tomo mode
switch' is enabled, pin 3-4, (Bucky tomo [DSI tomo] = closed, Bucky [DSI] = open)
- activates the recognition of the Bucky Ready contact pin 9-10 (ready = closed)
none = default
RGDV 1...8
- side field selection for automatic collimators with Bucky chambers, pin 1-2
- defines the Tomo RDGV for Bucky/Tomo remote switchover, if 'Tomo mode switch' is
enabled, pin 3-4, (Bucky tomo [DSI tomo] = closed, Bucky [DSI] = open)
- must be set to use option automatic input of tomo times,
- activates recognition of the Tomo Ready contact, pin 5-6, (ready = closed)
Tomo numbers 1...8 can be programmed in the 'APR Data Set's. Programming the times
makes only sense if option automatic input of tomo times 9890 000 0222x exists
(programmed in the CU-function key D38, see STAMMKARTE).
Tomo numbers 1...8 are directly linked to the tomo trajectories WAX22 pins 1...8, while
tomo times 1...8 are linked to the tomo time input WAX21 pins 1...8.
Every tomo time has to be programmed with a time in-between 0.1 ... 6000 ms.
*** If tomo APRT's are programmed using option Automatic input of tomo times: Select every
APRT once. For a very short time the default time of an APRT can be seen at the desk, then
the tomo time contact might override the default time. This time must be saved using the
desk save function: push and hold the RESET labeled button + active APRT. Now the
programmed tomo time 1...8 is copied and saved at the APRT data set. If it is not saved then
the default APRT time will come up again if the APRT is pushed a second time or if an APRT
with the same tomo time is selected.
The tomo time input only reacts to a high >> low transition of the time contact input.
- If tomo is programmed at 'RGDV Data Set A + B' it is not possible to change the exposure
time at the desk. If APRT with the same tomo trajectory have to be corrected time dependent
then an APRT with another tomo trajectory has to be selected. Afterwards select the APRT
which has to be modified to get the high >> low time input response.
- It might happen that three tomo times can be seen at the desk after pushing an APRT
button: At first the APR time is displayed. The second time might be tomo time number 1, as
long as all time input contacts of the tomo stand are open. If then any tomo time contact is
closed, the status change will cause the display of the programmed tomo time.
- Bucky RGDV:
Eight RGDV’s can be programmed at the same system ready contact WBX11 pin 9-10
(R/F controller, ready = closed).
Activates also the format size correction contact (closed if > 24x24 cm) WB X11 pin 1-2
(Bucky chambers only).
!! if no automatic collimator exists, insert a link at pin 1-2 to get access to the side field
selection (Bucky chambers only)
- if no Bucky Ready contact exists, but the format size correction has to be active: insert
a link at WBX11 pin 9-10 (Bucky chambers only)
Philips Medical Systems DMC Röntgenstr. 24 22335 Hamburg
Service Support & Field Assistance Axel Duve [email protected] Box141
March 28, 2001 OM_33_rev2.doc PC version page 15-52
- AMPLIMAT
- Sensitivity: AEC_SENS
low (default)
!! Must be low in case of PCB EZ150 versions up to and including 4512 108 05963
and versions >= 4512 108 05964 if jumper W4 is in position 3 = x1
'low' must be the recommended range if at least one film screen combination of
the system is < 200 (only >= 4512 108 05964 with jumper W4 in position 3 = x1).
The chamber signal = density voltage "U off" to the dose rate control input of CU
(see 'DRC Logging' - 'AEC' - 'AEC Calculation') should not increase 10VDC.
For ‘Photo sensor/ampl. inp.’ at ‘RGDV Data Set A’ it does not matter if high or low
is programmed. The density voltage 'U_off' will always be 1V.
- Chamber 1 ... 5
Note: For detailed information a booklet "Radiographic screens and films" , also
including imaging plates, is available: 4512 980 50592.
(1) If the film type is not in the default data list, then select one of the film files:
FILM_BL.TDL (blue), FILM_GR.TDL (green), FILM_UV.TDL (ultra violet).
Select the one which matches the sensitivity factor S and RLF compensation.
S is a multiplication factor for the speed type of the screen:
If the screen = 400 and S_film = 0.5, then the total system is 200.
(2) If the screen type is not in the default data list file LUMAT.TDL can be selected:
It contains the luminous groups LG with different colors and different speeds.
(3) The following list gives the PEI-No's of the chambers which can be selected from
the data file CHAMBER.TDL and the typical dose request values.
(6) The default value is 1 representing a density of 1.0. If no PC-hardkey exists, the
density adjustment must be carried out by typing in the appropriate factor.
desired density
= new [‘CORRECTION FACTOR’]
measured density
The name of the FSC is the abbreviation of the resulting speed and the color of
the film screen combination. There might be names like B400 for a blue 400
speed combination or G200 for a green 200 speed combination. The default
abbreviation name is taken from the speed class of the screen data set.
The name can be changed (only with the PC-hardkey), see 'Start Automatic
DRC Processing’ = ‘<CANCEL>' or ‘<ESC>‘.
The name must be changed if different film screen systems with the same
resulting name exist, otherwise the change from e.g. one G200 to a second one
can not be recognized at the control desk.
Combinations of systems with different colors can not be programmed the
automatic way (even though they are used), it requires a special procedure.
If film screen combinations are used at different chambers a data set can easily
be copied to other chambers/data sets once the first has been programmed
(only with the PC-hardkey):
Use the '<CANCEL>' function, the data screen will be displayed. With the Save
function <F3> the data screen can be loaded to the disk/hard disk with any file
name (e.g. Ch1D1). If a data set is opened at another chamber the saved file
can be Loaded with <F4> + file name to the open data set. Transmit the data
set with <F2> thereafter. The density alignment must be carried out for each
chamber + data set.
Only two values of the 'Data Set's of the 'Chamber's should be changed:
'Abbreviation' and 'Dose of FSC'.
All others must not be changed (e.g. all kV dependent characteristics).
1000µGy 1000µGy
------------------- = Ks example: -------------- = 5µGy
speed system 200
Abbreviation:
The name of the film screen combination can be changed into every 6 digit
string.
An FSC can be erased by filling the name field with blanks,
but: !!
There should never be a gap in-between the FSC data sets of a chamber. If
only one FSC is used, it must be at data set 1, the second must be on 2 etc. If
e.g. 4 data sets are active and the 1st will be erased by blanks (spaces), all
FSC's disappear and AEC is no more available for this RGDV.
If any FSC will be removed from the last position all APR linked to this FSC
have to be programmed to any other still existing FSC, otherwise no FSC will
be displayed at the desk. When these APR are changed to another FSC the
background mAs values have to be adapted to the FSC sensitivity.
Since Release 3 predefined APR data sets can be programmed with the FSC
name = 'Abbreviation'. If the abbreviation string and the name in the APR data
set are identical, the FSC will be associated to the APR data set.
If no name is given in the default APR data sets, APR will be assigned to the
FSC which has the closest 'Dose of FSC[µGy]:' value to the Recommended
Dose in the default APR data sets. The value can only be seen with the help of
APRMANager if the APR data screen is open or with the print preview.
The initial value 'Dose of FSC [µGy]:' of the ‘DRC Handling / Start Automatic
DRC Processing’ (calculated by XRGSCOPE in the service PC) is set to
achieve a density of 1.0.
It can be changed according to the customers density "taste", e.g. 1.5:
desired density
* default [Dose of FSC] = new [Dose of FSC]
measured density
If default APR of the installation disk are selected with AEC technique the
exposure data at the desk will display the abbreviation of the FSC and a default
zero with a triangle for the density. This zero should be the basic value for the
desired density. It will not change whatever value for the ‘[Dose of FSC]’ is
programmed.
The II lead time, density voltage correction (default 1V), and the dynamic factors can be
modified if necessary.
off = No 4% dose supervision at 10% of the backup exposure time, automatically off
after APR overriding
on = (default) At 10 % of the backup exposure time (APR mAs value) DRC detects if at
least 4% of the expected dose (signal must be > 20mV) have been measured. If
not, the exposure will be switched off by CU.
off = No 4% dose supervision at 10% of the backup exposure time, automatically off
after APR overriding
on = (default) At 10 % of the backup exposure time (APR mAs value, not before 250ms,
tomo time > 1000ms) DRC detects if at least 4% of the expected dose (signal must
be > 20mV) have been measured. If not, the exposure will be switched off by CU.
20 Fluoroscopy curve memories can be loaded with one of the 3 predefined fluoro curves of the
installation disk file FLUOCURV.TDL:
One out of the programmed curves (1...20) has to be attached to the APR data set,
see ‘Change APR’.
If other curves for special applications are required, send your request to Helpdesk Hamburg.
Philips Medical Systems DMC Röntgenstr. 24 22335 Hamburg
Service Support & Field Assistance Axel Duve [email protected] Box141
March 28, 2001 OM_33_rev2.doc PC version page 20-52
- Application Limits
Limits can be modified within the range of the value fields. The displayed default limit values
are higher than the values which can be selected at the control desk.
Values should not increase the local regulation limits.
A short exposure time in combination with a low emission current can cause an error 00L$ if
the resulting mAs value is below 0.5 mAs (only if sent online with XRGSCOPE).
The max exposure time which can be selected at the control desk is always 16000 ms.
If tomo is programmed the max exposure time is 6000 ms.
60000 ms is only available if ‘CT add on’ has been selected at ‘RGDVx Data Set B’ and if
option is present.
!!! The max time limit for AEC must not be shorter than 4000 ms, 6000 ms for TDC !!!
Otherwise error 00L3 comes up and the generator might be dead afterwards.
0.5 mAs is the lowest value one can select at the control desk (any non-AEC technique
RUIT = kV-mA-ms, RUQ = kV-mAs, RUQT = kV-mAs-ms).
A short exposure time in combination with a low emission current can cause an error 00L$ if
the resulting mAs value is below 0.5 mAs (only if sent online with XRGSCOPE).
AEC = AEC falling load and TDC = Tomo Density Control techniques have a default max
mAs limit of 580 mAs programmed to meet the HHS limit of 600mAs (AEC and AECF use
the same limit). AECF = AEC fixed current is covered by the AEC limit.
Even if the non-AEC techniques RUIT, RUQ, RUQT show 1000mAs as default limit, the
absolute limit which can be selected at the control desk is 850mAs.
It can also be selected in HHS countries which are limited at 600mAs for AEC exposures.
Table of default max mAs values (kV dependent) to prevent an over-saturation of the
Thoravision detector drum (memory effect).
Max. current during overl. for conti. fluoro [mA]: range = 1 ... 3 mA
- Human Interface
Philips Medical Systems DMC Röntgenstr. 24 22335 Hamburg
Service Support & Field Assistance Axel Duve [email protected] Box141
March 28, 2001 OM_33_rev2.doc PC version page 21-52
Select any APR at the desk, the name will be highlighted. After <ENTER> at the PC the
APR-# will be displayed on the PC screen. This # can be used for 'Inquire APR
Assignments' or to move APR.
Once this 'Select APR Data Set' window is transmitted with <F2>, the following
'Change APR Data Set' command allows to look into the complete APR data set.
(It is not possible with the default Test APR.)
Any APR # in-between 1 and 1024 can be typed in, no matter if this APR # is assigned to a
RGDV or not.
This function opens the window to modify the entire APR data set. Data marked with a
[P] can only be changed in this window via service PC or with the APRMANager.
All other non-marked data can also be changed at the desk and saved thereafter via the
button combination RESET (hold it) + the highlighted APR,
but: see ‘APR modifiable by User’.
('Change APR Data Set' is not possible with the default Test APR.)
Note: If only a few APR data sets have to be maintained one can use these functions. If
more APR data sets shall be modified the use of APRMAN.EXE (part of the installation disk)
is much easier, more convenient and a lot faster.
This number can be used to copy a data set into another data set. To do so the destination
APR number ('Select APR Data Set') has to be filled in the 'APR number' field, afterwards
transmit with <F2>. The source data set does not change.
To move an APR data set copy a data set to the desired destination and delete the source
APR data set afterwards with ‘APR deassign'.
There has to be an active destination APR. Error 00BR will come up if a data set is loaded/
copied to an APR number which is not assigned to any RGDV.
If APR backups are reloaded the APR numbers of the backup file will be the same after
reloading except
- if the APR numbers are already occupied
- if a default data set from the installation disk is used
- if an individual RGDV backup file will be loaded to another RGDV of the same generator.
Note: 'APR number's should not be changed if the generator is in combination with a
BuckyTH or TH2 system via CAN. Otherwise the assignment of APR#s for the control
handle will be lost.
- [P] APR name: 16 characters are possible
If special letters are needed (language dependent) see doc INSTALLATION character
codes.
Note: Since Bucky controller release 5.x for BuckyTH2 an "=" character must be at the
end of the APR name string to get a ready display for grid used RGDV / APR.
APR for RGDVs 1+2+3 with Inalfa trays must always have the "=".
APR for ACL4 trays with removable grid must have the "=" when the grid shall be
used
and no "=" when the grid shall not be used (removed from the tray to get a
system ready) for application reasons.
Philips Medical Systems DMC Röntgenstr. 24 22335 Hamburg
Service Support & Field Assistance Axel Duve [email protected] Box141
March 28, 2001 OM_33_rev2.doc PC version page 22-52
Digital Diagnost systems also require the "=" character for a system ready when
the grid shall be used and no "=" without grid.
If APR data are changed at the desk the two last digits of the 16 character field will be used
to indicate '_*' = 'overriding'.
- [P] Vario focus ratio: range: 20%, 35%, 50% (default) , 65%, 80% of small focus
Sets the ratio of small and large focus, for tubes with superimposed focal spots.
If possible indicate the ratio in the 'APR name' field.
Example: 'Thorax_ap____50%'
To get the best application exposure parameters use APRMANager, it takes the tube current
limits into account and calculates the optimum.
At least one of the measuring fields must be 'on' to have access to AEC, AECF or TDC no
matter if the 'Preferred technique' is set to Automatic' or 'Non automatic'.
A chamber must be programmed at 'RGDVx Data Set A' + 'Dose measurement input'.
The default APR data set [### APR name ###] has all 3 fields 'on'.
Some customers like ho have technique APR's only. If so switch 'off' all fields for the non-
AEC techniques so that there is no way to change to AEC technique if AEC is possible at the
same RGDV / menu / page.
If a Mobile Cassette Stand MCS is used at RGDV4 together with free cassette of a
BuckyTH, TH2 system the APR/menu structure should offer free cassette APR with all fields
off and
MCS APR with the required field(s) on.
Non automatic
One of the three 'No AEC technique’s will be displayed at the major exposure data
display.
With the mA-s labeled button at the desk one can switch over from kV-mA-ms (LED on)
to either kV-mAs or kV-mAs-ms technique (LED off), depending on what has been
programmed.
If a 'Dose measurement input' and/or ‘Dose measurement sensor type’ is programmed at
'RGDVx Data Set A' and at least one 'dose measurement field' is 'on' one can switch over
to AEC technique, which will cause the overriding sign ' _*' coming up.
Automatic
AEC / AECF / TDC technique will appear when this APR is selected (AEC technique
button LED and 1...3 measuring field LED's on).
Philips Medical Systems DMC Röntgenstr. 24 22335 Hamburg
Service Support & Field Assistance Axel Duve [email protected] Box141
March 28, 2001 OM_33_rev2.doc PC version page 23-52
*** Most of the predefined APR data sets from the installation disk are programmed for a
400 speed system (Recommended Dose: 2.5 µGy). If these are taken all organ
dependent mAs values have to be adapted to the local film speed systems (doubled
for 200 speed systems, divided by 2 for 800 speed systems).
AEC falling load kV the kV dependent mA start value of the falling load exposure
will appear in the 'Exposure data I [mA]' field
AEC fixed current kV-mA the fixed mA value has to programmed in the
'Exposure data I [mA]' field, mA and exposure time
values have to match the required organ dependent mAs
TDC (Tomo Density Control) (option 9890 000 0223x, see STAMMKARTE)
Two out of the three 'No AEC technique's are available for every APR.
If either kV-mAs or kV-mAs-ms technique is selected one can switch over to kV-mA-ms
technique pushing the "mA-s" labeled button at the desk.
If kV-mA-ms (LED on) is programmed as preferred non-AEC technique and if it will be
deselected by pushing the "mA-s" labeled button (LED off) it changes to kV-mAs
technique only.
The preferred 'No AEC technique' after AEC-deselect can be changed at the control desk
with the overriding-save function.
Must be selected in case of 'AEC fixed current kV-mA' as preferred AEC technique
and can be selected for 'TDC (Tomo Density Control)' techniques.
Should be selected as the best non-AEC technique for TDC technique APR.
'RGDVx Data Set B'. It must be programmed in the APR or the time selection is
controlled by automatic input of tomo times (decade WAX21 or Bucky controller).
Note: Depending on which of the three non-AEC techniques is selected only the relevant
parameters in the following 'Exposure data nn [xx]' have to be modified. All other
parameters will be calculated by the generator itself after transmission with <F2>. If the
same APR data set screen will be called again, the calculated "slave values" appear.
*** Most of the predefined APR data sets from the installation disk are programmed for a
400 speed system (Recommended Dose: 2.5 µGy). If these are taken all organ
dependent mAs values have to be adapted to the local film speed systems (doubled
for 200 speed systems, divided by 2 for 800 speed systems).
- [P] Tube current max. factor [%]: range = 1...100% (default 100 %)
The max emission current can be modified within the above range.
The reduced current (see 'Exposure data I [mA]') will be the start mA value of an AEC
falling load exposure. It can also be the max mA value e.g. for an AEC fixed current
exposure. It will be a constant value during the exposure until it comes to a point where
the mA have to be reduced (falling load) within 4000ms. The mA value as the result of the
% reduction is different for each kV value.
The mA value can be seen if the APR data set screen has been transmitted with <F2>
once. Call the same ‘Change APR’ data set again and the calculated mA value appears.
Another way to see the mA is at the desk: Push the APR button, switch off AEC. The
background technique will come up. In case of kV-mAs or kV-mAs-s push the "mA-s"
labeled button to switch over to kV-mA-ms technique. The displayed mA is the value
which is one step below the max value in the row of the programmed mA steps (see
'RGDVx Data Set B'). Now push the +mA button to get the max mA value. This is
identical to the data set displayed on the PC screen.
If an APR with small focus will be changed by pushing the large focus button the +mA
has to be pushed until the max value appears. A change from large to small will
immediately display the max mA.
!! The modification of the max factor will lead error 00L$ if the resulting mAs value is
below 0.5 mAs.
!! Take care that there are not too many additional reductions programmed !!
- [P] PSC U thin (dose equivalent steps): range = 0...-5 dose equiv. kV steps
- [P] PSC U thick (dose equivalent steps): range = 0...+5 dose equiv. kV steps
- [P] PSC Q thin (6% steps): range = 0...-10 mAs steps see [1] page 25
- [P] PSC Q thick (6% steps): range = 0...+10 mAs steps see [1] page 25
- [P] PSC density thin (6% steps): range = 0...-10 dens. steps see [1] page 25
- [P] PSC density thick (6% steps): range = 0...+10 dens. steps see [1] page 25
[1] Depending on the step rate of mAs- and density-correction programmed at 'RGDVx Data
Set B' these correction values must have an even factor with the basic step rate:
For one 25 % mAs-correction step (programmed at 'RGDVx Data Set B') four 6%-steps have
to be programmed.
To get a +1 density step displayed two 6% steps have to be programmed if 12% 'Density
steps' are programmed at 'RGDVx Data Set B'.
Depending on the generator version, tube type, focal spot, selected technique and all
reduction factors: This value can be a master or slave value.
The result of a [mA]*[ms] multiplication must always be >= 0.5 mAs, otherwise error 00L$
appears.
If full mA range is not available, check ‘Tube limit’ settings or all other reduction tables.
- Exposure data Q [mAs]: range = 0.5...1000 mAs (generally limited at 850 mAs)
Depending on the tube type, focal spot, selected technique and all reduction factors: This
value can be a master or slave value.
If full mAs range is not available, check ‘Tube limit’ settings or all other reduction tables.
Depending on the tube type, focal spot, selected technique and all reduction factors: This
value can be a master or slave value.
The result of a [mA]*[ms] multiplication must always be >= 0.5 mAs, otherwise error 00L$
appears.
If full range is not available, check ‘Tube limit’ settings or all other reduction tables.
Depending on which step rate has been programmed at 'RGDV Data Set B',
explanation see [1] page 25.
*** If APR with AEC are associated to a certain Film-Screen-Combination the basic density
value of the FSC appears as a zero. This zero - value is the base density value which
has been adjusted to the customers taste.
There are organs which work fine with the basic zero density, but others require e.g. a
zero-2 or zero+3 correction. SCP and MedioCP are based on 12% steps for every +/-1
correction, so should the Optimus be programmed at 'RGDVx Data Set B'. For every +/-1
12% step two +/-6% steps have to be programmed.
- Film screen comb.: Offers all available FSC names programmed to the 'Data Set's of a
chamber assigned to the RGDV of the selected APR.
Since Release 3 predefined APR data sets can be programmed with the FSC name =
‘Abbreviation’. If the ‘Abbreviation’ string and the name in the APR data set is identical, the
FSC will be associated to the APR data set.
If no name is given in the default APR data set the Recommended Dose µGy value of the
individual APR data set will determine the assignment of the APR to the FSC Data Set which
has the closest ‘Dose of FSC [µGy/V]’ value.
Philips Medical Systems DMC Röntgenstr. 24 22335 Hamburg
Service Support & Field Assistance Axel Duve [email protected] Box141
March 28, 2001 OM_33_rev2.doc PC version page 26-52
!! The Recommended Dose µGy value can not be seen in this screen via XRGSCOPE !!
!! It is only possible with the APRMANager.
*** Most of the predefined APR data sets from the installation disk are programmed for a 400
speed system (Recommended Dose: 2.5 µGy). If these are taken all organ dependent
mAs values have to be adapted to the local film speed systems (doubled for 200 speed
systems, divided by 2 for 800 speed systems).
Option Automatic Input of Tomo Times 9890 000 0222x (see STAMMKARTE) must be
present. If no tomo unit is present: ignore.
Decade systems:
Active if programmed at 'RGDVx Data Set A + B', 'RGDV Interface Assignments' and
'Tomo Time' for the programmed 'Release circuit adaptation unit'. Then 8 tomo numbers
are available for non-CAN systems.
*** If tomo APRT's are programmed using option Automatic input of tomo times: Select
every APRT once. For a very short time the default time of an APRT can be seen at the
desk, then the tomo time contact might override the default time. This time must be saved
using the desk save function: push and hold the RESET labeled button + active APRT.
Now the programmed tomo time 1...8 is copied and saved at the APRT data set. If it is
not saved then the default APRT time will come up again if the APRT is pushed a second
time or if an APRT with the same tomo time is selected.
The tomo time input only reacts to a high >> low transition of the time contact input.
- If tomo is programmed at 'RGDV Data Set A + B' it is not possible to change the
exposure time at the desk. If APRT with the same tomo trajectory have to be corrected
time dependent then an APRT with another tomo trajectory has to be selected.
Afterwards select the APRT which has to be modified to get the high >> low time input
response.
- It might happen that three tomo times can be seen at the desk after pushing an APRT
button: At first the APR time is displayed. The second time might be tomo time number 1,
as long as all time input contacts of the tomo stand are open. If then any tomo time
contact is closed, the status change will cause the display of the programmed tomo time
CAN systems:
Active if programmed at 'RGDVx Data Set A + B' for CAN systems:
Bucky TH, TH2, Digital Diagnost ("<" marked not for DiDi) systems with TOMO option
provide 16 tomo sweeps:
tomo # Tomo #
sweep sweep
Bucky Optimus Bucky Optimus
0 1 8º / 0.8s 8 9 30º / 1s
1 2 8º / 1s 9 10 30º / 2s
2 3 8º / 2s 10 11 30º / 3s
3 4 20º / 0.8s 11< 12< 30º / 4s
4 5 20º / 1s 12 13 40º / 1.2s
5 6 20º / 2s 13 14 40º / 2s
6 7 20º / 3s 14 15 40º / 3s
7 8 30º / 0.8s 15< 16< 40º / 4s
Gives a selection of the three filter types of the GALILEO collimator BuckyTH, TH2,
Digital Diagnost, Thoravision:
no filter
2 mm AL
0.1 mm CU + 1 mm Al
0.2 mm CU + 1 mm Al
One out of the 20 ‘Fluoroscopy curves’ (the ones which have been loaded with default
data sets can) be selected.
OPTIMUS provides memory for 1024 APR. Once activated an APR data set gets an APR-#.
These #s are given without any special system (it is more ore less random). If a certain APR-
# is known, typed in and transmitted with <F2>, the window coming up gives the APR name,
the RGDV and menu to which it is assigned to (blank in case of none), the page, line and
column of its position.
If an APR-# is typed in which is not assigned to any RGDV the location fields (RGDV, menu,
submenu, page etc.) are empty, the name might be [### APR name ###] or any old name of
an APR which has already been deassigned.
To get the APR-#, just select any APR button at the desk. Then use 'Select APR Data Set'.
(It is not possible with the default Test APR.)
- RGDV 1 ... 8
- Predefined Assignments
The Load Data from Disk window offers all predefined APR data sets of the installation disk
or other predefined files (or use APRMAN.EXE to create customized APR data files).
...DE(U) German
...EN(G) English
...ES(P) Spanish
...FR(A) French
The files may contain e.g. data sets in a window Available Examination Unit Type :
Philips Medical Systems DMC Röntgenstr. 24 22335 Hamburg
Service Support & Field Assistance Axel Duve [email protected] Box141
March 28, 2001 OM_33_rev2.doc PC version page 28-52
After the end of the transmission from the PC to the generator "waiting" will appear on the
PC screen. Now the generator is calculating data to the individual installation parameters
(generator power, tube type etc.). The generator must not be reset before "waiting"
disappears, otherwise APR which have not yet been calculated at that moment are lost.
Individual RGDV related backup files (APR_BAKx.TDL, x = 1...8, see 'Accept' + 'Backup' )
can also be loaded to the generator with the same 'Predefined Assignment' function.
It is possible to load all RGDV with predefined APR data sets and make one generator reset
at the end of all loading procedures. After every reset all newly loaded APR and menus
appear at the desk.
It is not possible to load a second data set on top of an existing data set. The old RGDV
menu/APR structure has to be erased before with 'Delete Menu'.
With the XRGSCOPE - ‘File’ option one can use the editor to modify the predefined APR
data sets or backup data sets (e.g. to change all APR to large filament in case of a single
filament tube).
This is easier and faster as if every APR will be selected at the desk with 'Select APR'
followed by the 'Change APR'. One advantage is that this can be done at the PC offline the
generator.
The handling is easy, is almost similar to the 'Change APR' screen online with the generator.
After modification the data set has to be saved with <F3>, the filename could be the same or
any other.
Disadvantage: Only data sets with up to 50 APR can be modified. There is no way to add or
erase APR.
- Manual Assignment
- APR Assignment
The menu name field is as long as the longest menu name. If the field is only two
digits long and only one field is displayed after <ENTER> then there is no menu.
If no menu structure exists, proceed directly to 'Assign APR'.
If APR's shall be in a menu structure, assign this/these before in 'Menu Assignment'.
If an APR shall be added to a menu it has to be selected first. Transmit this window
with <F2> after selection.
An APR position has to be opened first, after a cold/warm reset the new open position
can be selected at the control desk to be modified thereafter (see 'APR Data Set').
APR:
Philips Medical Systems DMC Röntgenstr. 24 22335 Hamburg
Service Support & Field Assistance Axel Duve [email protected] Box141
March 28, 2001 OM_33_rev2.doc PC version page 29-52
The window will offer either ### APR name ###, which is the default data set or, if
there have already been other APR, some previous names of deassigned APR's (see
'APR Deassign').
Page No.:
Line No.:
Column No.:
It indicates (automatically) the next position where it will be loaded to:
page, line and column.
Any position can be typed in. If the position is already occupied, an error will come up.
If only one APR at position column 2 line 3 exists at a menu structure, the next APR
position will be on the next page, even if all other positions on this page are free. The
automatically given position can be changed to fill up empty positions.
Do not assign an APR at position column 2 line 4 at a menu structure or to column 2
line 1 at a fluoro RGDV, it will be transmitted to nowhere.
- APR Deassign
- Select Menu
The menu name field is as long as the longest menu name. If the field is only two
digits long and only one field is displayed after <ENTER> there is no menu.
If an APR shall be removed from a menu structure this menu has to be selected first.
If no menu structure exists, proceed directly to 'Deassign APR'.
- Menu Assignment
There are structures possible with a menu layer first and a second (up to 10)
submenu layer.
Creating submenu structures: First assign the main menus at 'Assign Menu'. Then
'Select Menu' and program the submenu names at the main menu with 'Assign Menu'.
Philips Medical Systems DMC Röntgenstr. 24 22335 Hamburg
Service Support & Field Assistance Axel Duve [email protected] Box141
March 28, 2001 OM_33_rev2.doc PC version page 30-52
!!! At least one APR must be programmed to one of the lowest menu or submenu
layer to have the structure fixed.
If once an APR is programmed to a menu there is no further submenu layer possible.
The 'Select Menu' name field is as long as the longest available menu name. If no
menu is available there will be the message The requested table is empty.
This function can be used to assign menus to an empty RGDV. It can also be used to
add a menu to an existing menu structure (in the same layer) or to create / add
submenus.
Menus can not be assigned if a paging structure with APR only is programmed or
within an APR layer.
Submenus can not be assigned at least one APR exists at a menu. In this case no
menus are offered at 'Select Menu'.
Menu Name:
Note: Since Bucky controller release 5.x for BuckyTH2 and also for Digital Diagnost a "="
character must be at the end of the APR name string to get a system ready for grid
used RGDV / APR. Menus with such APR should also have this "=" character to
indicate the use of the grid (especially if menus without grid also exist).
APR for RGDV 1+2+3 with Inalfa trays must always have the "=".
APR for ACL4 trays with removable grid must have the "=" when the grid shall be
used
and no "=" when the grid shall not be used (removed from the tray to get ready) for
application reasons.
Page No.:
Line No.:
Column No.:
It indicates (automatically) the position where it will be loaded to:
page, line and column
!! The automatic positioning unfortunately also takes the position which is occupied
!! by the fluoro kV/mA and time label.
Any position can be typed in. If the position is already occupied, an error will come up.
If only one menu at position column 2 line 3 exists at a menu structure, the next menu
position will be on the next page, even if all other positions on this page are free. The
automatically given position can be changed to fill up empty positions.
Do not assign a menu at position column 2 line 4 at a menu structure or to column 2
line 1at a fluoro RGDV, it will be transmitted to nowhere.
Philips Medical Systems DMC Röntgenstr. 24 22335 Hamburg
Service Support & Field Assistance Axel Duve [email protected] Box141
March 28, 2001 OM_33_rev2.doc PC version page 31-52
This function allows to delete the selected menu including the whole structure under it.
It can be an APR layer only or with submenus including all APR's.
The function offers all menus of a RGDV including an empty field on top. If the empty
field will be selected and transmitted with <F2> the complete menu/submenu/APR
structure will be erased. After generator reset "test APR" will appear at the desk. This
function should be used to clear a RGDV before new predefined APR data sets are
loaded or the 'Accept' >> 'Restore' (to reload backup files) function is used.
Menus or submenus can be moved within its layer to a free position (column and line),
which has to be defined before. Moving menus to other RGDV's is not possible. Use
APRMANager for a more convenient, faster and easier maintenance.
The menu name can be changed into everything desired (characters see doc
INSTALLATION).
The 'RGDV key 1/2' positions can be programmed with any of the active RGDV's. Six
APR of every active RGDV can be assigned to the six buttons 3...8. To get the ‘APR
key’ numbers use 'Select APR'.
The APR numbers must be entered at the RGDV # where they are assigned to. The
two programmed RGDV keys appear at all RGDV screens.
If an assigned RGDV and/or an assigned APR is selected at the control desk, the
RGDV/APR button will light up at the external APR module and vice versa.
APR key 1...8 (0 no APR): positions 1..8 = keys 1..8 (see page 32)
8 APR of each active RGDV can be assigned to the 8 APR keys ( >> max 64 !!).
It is recommended to use one RGDV only.
If an assigned APR is selected at the control desk, the APR button will light up at the
external APR module and vice versa.
With the RGDV key 1 and RGDV key 2 assignment the RGDV remote switch over
inputs WB X21 pin 1 and 2 will be activated. The selected RGDV must get a low
signal from the R/F controller, while the other RGDV signal must go high.
8 APR of each activated RGDV can be assigned to the 8 APR keys (e.g. SCOPO)
(max 64 !!).
It is recommended to use one RGDV only.
If an assigned APR is selected at the control desk, the APR button will light up in the
external APR module and vice versa.
Yes = All data which can be overridden at the desk can be saved:
- Push and hold the "RESET" labeled button
- push the highlighted (overridden) APR button
- the asterisk indicating overriding disappears.
- Adjust
- Tube Adaptation
The tube should be properly conditioned before starting the adaptation procedure.
Break-in procedure see doc INSTALLATION chapter Tube conditioning.
The generator should be in ready condition. Select a RGDV with free cassette programming (no grid
sync contact).
Free cassette
RGDV 4 at BuckyTH2 and Digital Diagnost
or
RGDV 8 Bucky unit easyDiagnost
'Data Set A' has to be modified as follows (only these settings have to be changed):
- Syncmaster present = No
- Mounted radiographical controller = none
Transmit the 'Data Set A' screen with <F2> and push the RGDV button, then the modified data
are valid (a generator warm start is not required).
After selection of 'Tube Adaptation' with <ENTER> a screen asks to wait 20 seconds after the screen
coming up thereafter has been sent with <TRANSMIT> or <F2>.
Focus: small
medium (a tube with a (third) medium filament does not exist yet, it is not VARIOfocus)
large
After <TRANSMIT> the green ready disappears from the desk for 20 seconds. The "Test" display on
the right side of the major exposure parameter indicating a non-adapted focus will change to "Adap".
With adapted tubes parameters like "film-screen-combinations" or "xxx ms" will disappear and will also
be replaced by a "Adap" during re-adaptation.
Ready returns with a first data set 40kV 0.00mAs (to measure the mA measuring circuit offset).
- Start the adaptation procedure by pressing the hand-switch in PREP and EXP position continuously.
- The anode will be accelerated and braked several times during the entire adaptation process.
Philips Medical Systems DMC Röntgenstr. 24 22335 Hamburg
Service Support & Field Assistance Axel Duve [email protected] Box141
March 28, 2001 OM_33_rev2.doc PC version page 34-52
After about 30 sec after the fist 40kV 0.00mAs exposure 40kV 1.5mAs will be switched for 6 seconds
to measure the individual standby filament current.
After another 20..30 seconds exposures with different kV stages are switched (about 120 / filament).
4 min after start of adaptation the boost adaptation takes place. One exposure is for the positive boost
to measure the boost up time, a second one (which is the last exposure of adap) for the negative
boost to measure the cooling time. The tube will brake the last time, a beep at the PC and a window
on the screen will ask to reset the generator.
There is no exposure end beep during the entire adaptation procedure.
During the adaptation it might happen, that the procedure stops for a while. The red light of the
temperature traffic light might turn on and warning entries 00BV are logged. If the TTS (Tube
Temperature Supervision) detects a temperature in a critical range it will stop adaptation for a while to
keep adaptation always at a 100% load range.
After adaptation followed by a generator reset all techniques are available and "Test" disappears.
AEC, AECF, TDC, kV-mAs, kV-mA-ms, kV-mAs--ms techniques and Fluoro (small focus only) are
not available as long as the filaments are not adapted.
VARIOfocus as a "calculated" virtual focus requires that small and large focus are adapted. APR
using VARIOfocus can not be selected as long as both, small and large, are not adapted.
<OK> gives the command for CU to scan on the internal generator CAN bus and update the
generator CAN member list (assign / de-assign units on the list).
Minimum configuration of an OPTIMUS (see also ‘Faultfind - Power ON Results - Internal
CAN Configuration’:
FU_mA (EZ119), FU_kV (EZ130), FU_CU (EZ139), FU_CIE (EZ150).
If now an optional Function Unit shall be added (e.g. FU_ADAP A = 1WA = PCB
EWA102) this command must be given to register this board on the CU CAN list.
If a generator is getting a new (empty memory) PCB CU or the NVRAM has to be erased
for any reason CU updates the member list during the very first turn-on itself.
- Area Exposure Product If the area dose shall be displayed at the control desk the option must
exist (9890 000 0256x, see STAMMKARTE) in the function key EZ139
D38 (see ‘Faultfind - Power ON Results - Options’). The option requires
the GALILEO collimator via ‘Bucky controller 1’ or ‘Thorav. / Dig. Bucky’.
Thoravision and Digital Bucky: The dose value is not displayed at the
control desk, as the system controller takes the value and resets the
value afterwards with the command next patient. The dose value will be
displayed on the images
Procedure to measure and modify the specific tube yield see doc ADJUSTMENTS. The default
yield data table is on the installation disk: RE_YIEL.TDL.
- Add Filter Correction Table to adjust deviations of the default filter correction tables of the
(additional) homogenous GALILEO collimator filters
FIL_CORR
- 2 mm AL The default data table is on the installation disk: RE_2AL.TDL.
- Wedge Filter Correction Tables to adjust deviations of the default filter correction tables of
the GALILEO wedge filters (not yet available)
WED_FILT
- Wedge 1 The default data table is on the installation disk: RE_WED1.TDL.
- Finger Wedge The default data table is on the installation disk: REF_FING.TDL.
LOCK IN: LOCK = for dose adjustment procedures, this function replaces
the kv and mA potmeters
automatically disabled after reset
!!! If no link is inserted at 1WB X12:07 (DR_FL_LO_1) to
:10 (GND2) the screen will be in 'UNLOCK' condition
when the screen is called again.
- Backup >hardkey<
RGDV 1...8
APR Assignments After Reading (spinning bar) the PC offers the default backup name
Backup File Name: APR_BAK1...8.TDL.
The default name can be changed into any other (up to 8 digit) file name.
The path from which XRGSCOPE has been started on the hard disk will
automatically be taken into account.
It is also possible to type A:\filename <ENTER> to load the backup file
directly to a floppy disk.
- CU Complete
The default backup name Backup File Name: CUBACKUP.TDL (or A:\filename <ENTER>)
can be changed into any other (up to 8 digit) file name (e.g. CU97xxxx.TDL representing the
generator serial number). A disk space of 700 kByte is required. It takes about 8...15 minutes to
save the data to the floppy / hard disk.
The path (from which XRGSCOPE has been started on the hard disk) will automatically be taken
into account.
It is also possible to type A:\filename <ENTER> to load the backup file directly to a floppy disk.
- Restore >hardkey<
RGDV1...8
APR Assignments The path (from which XRGSCOPE has been started on the hard disk) will
automatically be taken into account
or type in either A:\filename <ENTER> or just A: <ENTER>, <TAB> and
select a file with the cursor up/down buttons.
Load Data from Disk: will offer the default file name APR_BAK1...8.
- CU Complete
Load Data from Disk offers the default File Name: CUBACKUP.TDL
(taking the path from which XRGSCOPE has been started automatically into account)
or type in either A:\filename <ENTER> or just A: <ENTER>, <TAB> and select a file with the
cursor up/down buttons.
It takes about 15-30 minutes to restore the data (PC type dependent). If the spinning bar during
the loading procedure seem to turn very slowly then stop the procedure. Type <ALT> x to leave
the XRGSCOPE program and type in the following command in DOS or WIN: XRGSCOPE D=5 ,
the smallest value should be 1 for a PC processor with a low speed.
!!! Date and Time must be set after restore is finished with a new PCB CU or after the NVRAM
has been erased by pulling the battery jumper. !!!
- Inspect
- Tube Statistic
Philips Medical Systems DMC Röntgenstr. 24 22335 Hamburg
Service Support & Field Assistance Axel Duve [email protected] Box141
March 28, 2001 OM_33_rev2.doc PC version page 37-52
Note: No reset function available for the counters, the actual status has to recorded in the
system logbook in case of a tube exchange.
- Generator Statistic
for future use [s]: to count the switch on time, not yet implemented
Tube name + housing type as programmed from the tube data file.
Empty field: no tube programmed.
Philips Medical Systems DMC Röntgenstr. 24 22335 Hamburg
Service Support & Field Assistance Axel Duve [email protected] Box141
March 28, 2001 OM_33_rev2.doc PC version page 38-52
- Power ON Results
- Options OPTIONS
This screen displays all options programmed in the function key EZ139 D38 (compare with
STAMMKARTE at the cover of the frontal kV converter and add-on STAMMKARTE updates).
This screen displays the actual status of the assigned members on the internal generator CAN.
The CU as main controller of the generator is not part of this list.
This screen displays the actual firmware versions Release, Version (don’t care) and Level.
For PCB Central Unit it also displays the hardware version (must be 2 for Release 3.x).
- Logging Table
- Error Log Index gives an overview of the events logged in the CU NVRAM
Depending on the free memory of the PC (no matter if it runs with DOS or any WINxx) the
error log index might appear as a table in which one can go up/down with the cursor buttons
or
with just one single error, the frame indicates "Error Log Index 1/50". With the <TAB> button
one can jump to the right end of the individual screen, with the <page up/down> keys one
can step through the error index list, the actual index number will also change to e.g. 12/50.
Index: [ ]
Max 50 events can be logged. The last event is always in the highest index number. If all 50
lines are filled, line 50 will log the last event and the oldest event (line 1) disappears.
The last entry will always be "00S*" if the first action after log-in is checking the error log
index. "00S*" indicates that a PC log-in has taken place, no matter if XRGSCOPE or
APRMAN was used.
Code: [ ]
Date of Error: [ ]
Error Explanation: [ ]
Index: [ ]
After looking through the ‘Error Log Index’ list enter the index number of an event to get the
details. (The last index number is automatically in the input field). Transmit with <F2>.
Not every function unit provides detailed event logs. If event source is not CU, kV (see ‘FU-
kV’) or mA (see ‘FU-mA’), "FU not supported" will be displayed.
- Error Detail of CU
Philips Medical Systems DMC Röntgenstr. 24 22335 Hamburg
Service Support & Field Assistance Axel Duve [email protected] Box141
March 28, 2001 OM_33_rev2.doc PC version page 40-52
Start function: Clear Error Log Once sent with <OK> all entries are erased.
- X-Ray Log
The tube temperatures are supervised by the TTS calculation model. Every second the
temperatures are recalculated and updated in this table. Overload flags (see part: 23456)
are tube and housing type dependent. Tables disappear with switch off, but actual
temperatures will be calculated and updated during switch on or warm start.
All temperatures should be at 20ºC (basic value) after a long switch-off period.
0 0 1 0 0 0 0 0
yellow red T(2) T(3) T(4) T(5) T(6)
traffic light condition: 1=
green =00 standby 0 = temperature value within the max limit
green + yellow = 0 0
yellow =10 0=
yellow + red =11 exposure 1 = temperature value increased max limit
red =11
This Tube Load Table displays all loads to the tube and housing. The table disappears after
switch off or warm start.
time indicates at which second the event has been logged after turn-on
tube tube number (1...3)
tube disk energy [Ws] parameters
peak load [W] max peak load in
rotor energy [Ws] acceleration and brake energy in one package
focus filament used for the exposure
VARIO focus is displayed as large filament at a ratio of 20%, 35% and
50% of small focus, as small with 65% and 80% of small focus
to be continued
- Dose Rate Control Logging
Philips Medical Systems DMC Röntgenstr. 24 22335 Hamburg
Service Support & Field Assistance Axel Duve [email protected] Box141
March 28, 2001 OM_33_rev2.doc PC version page 41-52
selected mode:
time measured [ms]:
last I [mA]:
last U [kV]:
dose relative [%]:
dose offset:
last dose measured:
LOCK IN state:
DRL (dose rate limitation) state:
U nominal [kV]:
I nominal [mA]:
t nominal [ms]:
Always 2000ms if Test is displayed at the desk indicating a non-adapted focus.
C eff ht [nF]:
t corrected [ms]:
- AEC
U nominal [kV]:
I nominal [mA]:
t backup [ms]:
C eff ht [nF]:
selected sensor:
dose measurement input:
film screen comb.:
dose nominal [OD]:
dose calculated:
kV factor:
U off [V]:
t corrected [ms]:
t lead AEC [ms]:
- AEC Trace
t sample [ms]
dose nominal
dose actual
dose rate
t dr [ms]
t esti [ms]
- TDC
U nominal [kV] :
I start [mA] :
t backup [ms] :
C eff ht [nF] :
dose measurement input :
film screen comb. :
dose nominal [OD] :
I max [mA] :
Philips Medical Systems DMC Röntgenstr. 24 22335 Hamburg
Service Support & Field Assistance Axel Duve [email protected] Box141
March 28, 2001 OM_33_rev2.doc PC version page 42-52
I min [mA] :
dose calculated [ms] :
kV factor :
U off [V] :
t corrected [ms] :
number of steps :
time / sample [ms] :
- TDC Trace
idx
dose nominal
dose actual
dose rate
reg in
reg out [%]
new I [mA]
- CONT
U start [kV] :
I start [mA] :
C eff ht [nF] :
measure_int [ms] :
max dU [kV] :
U [kV] :
I [mA] :
dr_abs :
- CONT Trace
act_idx :
idx :
delta_dose_rate :
PID_out factor :
dr_abs :
new I [mA] :
new U [kV] :
- CU Trace CU_TRACE
Rel Time
FG
Trace Info
Philips Medical Systems DMC Röntgenstr. 24 22335 Hamburg
Service Support & Field Assistance Axel Duve [email protected] Box141
March 28, 2001 OM_33_rev2.doc PC version page 43-52
- FU - mA
- Program
Window asks for Focus number. Type in focus number FU-mA (1...6):
deinstall :
Uiso [kV]: focus specific value, default value will be overridden after adaptation
(generator version = max available emission current dependent)
Uiso is the kV value, from which on the filament current can be decreased to
drive always the max kV dependent emission current (with rising kV)
Pmax [W]: maximum focus output power, default value in tube data file
value might be different (lower) after tube adaptation
(e.g.SRO33/100 with Optimus 50: Pmax after adap =
50kW instead of 80kW)
IFmax [mA]: max filament current limit, default value in tube data file
value might be different (lower) after tube adaptation
e.g. Pmax tube > Pmax generator
IFregelmax [mA]: = IFmax + analog offset value for regulation purposes during boost up,
typically IFmax + 200mA
Philips tube :
FILsel :
FILcir: physical filament circuit, see table ‘filament converter PCB FU-mA EZ119’
above
I0 [ A]: I0 = max emission current at Umin (typically 40kV = default)
up to I110 = e.g. max emission current at Umax (typically 150kV = default)
I120 [ A]: (I0 equals 40kV) + (I110 equals 110kV) = 150 kV
Philips Medical Systems DMC Röntgenstr. 24 22335 Hamburg
Service Support & Field Assistance Axel Duve [email protected] Box141
March 28, 2001 OM_33_rev2.doc PC version page 44-52
all values in-between are (default) specified tube data coming from the
installation disk data file, values (might) change when modified in the ‘Tube
Limits’ screen and after adaptation
CU physical fil: CU can handle 9 filaments at max 3 tubes, see table ‘focus number CU
EZ139 (0..8 = max 9)’ page 44
CU checksum fil:
- Faultfind
- Logging Tables
Gives the actual status of the filament circuit. Several entries are possible in the three fields.
Displays the last 10 activities with FU_mA. All fields have several entries possible.
details to be continued
- Error: Info
Displays an error code and sometimes the reason for the error entry.
- fatal: Fatal errors might be displayed at the control desk, but not
in all cases. If there is a communication breakdown a fatal
error might not be possible on the CAN bus.
As all error entries are logged in CU the information has to
get to CU.
None of the units has a memory to keep the fatal condition
when the generator has been turned off. So the information
is lost. The only way to log it is by pushing the on button to
warm reset the generator. Then the fatal error condition
might be transmitted to CU.
task name: is the task name of the function unit process during the error
ignore for troubleshooting
procedure name: is the procedure name of the FU process during the error
ignore for troubleshooting
Displays actual status during error: see ‘Read Actual Status’ page 44.
Displays status trace during error, only 3 entries: see ‘Read Status Trace’ page 44.
- error condition
- single focus tube (using large only)
EN_STOP_X : on =
off =
EN_X_ACT : on =
off =
GRID control : on =
off =
GRID mode : on =
off =
IF actual (filament circuit 1) [mA]: filament current actual value small focus
measuring point EZ119 X5 - (X8 gnd) 2.5A / V
IF actual (filament circuit 2) [mA]: filament current actual value large focus
measuring point EZ119 X7 - (X8 gnd) 2.5A / V
intermed circuit voltage (fil. cir. 1) [V]: intermediate DC supply of filament converter 1
small focus (typically 325VDC)
intermed circuit voltage (fil. cir. 2) [V]: intermediate DC supply of filament converter 2
large focus (typically 325VDC)
- Functional Tests
- test watchdog
activates watchdog of FU_mA, red LED turns on and remains on, activate warm reset to
restart generator
no PC access & communication during this (FATAL ERROR) status
- Monitoring
type in filament circuit number, see table ‘filament converter on FU-mA EZ119’ page 44
type in filament circuit number, see table 'filament converter on FU-mA EZ119’ page 44
type in filament circuit number, see table ‘filament converter on FU-mA EZ119’ page 44
A ‘table’ of 100 individual screen comes up, screen 1/100 is the last logged value, 100/100
the oldest.
The mA value is measured every 2 ms. With low mA values (e.g. during fluoro or low-mAs
tomo) there might be 0mA entries.
- read memory
ofs [uA] default mA offset current value of mA measuring circuit G100 = 4000uA
corrected value appears after adaptation
- Adaptation Results
focus: type in filament number, see table ‘focus number FU-mA’ page 44
U [kV]: type in a kV value between 40...150kV (or whatever the limit is)
Displays tables of different sizes (kV stage dependent) after the filament has successfully
been adapted.
Screen (or value line) 1/xx always displays the individual standby filament current at 100uA.
Screen xx/xx displays a filament current which is just one step above the limit filament
current.
boost time [ms] : how long it takes to boost up from a specific emission current
(filament temperature) to a higher emission current (higher
temperature) with the boost current = max filament current + 2A
(positive boost)
blank time [ms] : how long it takes for the filament to ‘cool down’ from a specific
emission current to a lower emission current (negative boost)
typically the negative boost time is 2-3 times positive boost time
Philips Medical Systems DMC Röntgenstr. 24 22335 Hamburg
Service Support & Field Assistance Axel Duve [email protected] Box141
March 28, 2001 OM_33_rev2.doc PC version page 49-52
- FU - kV
- Adjust
- Faultfind
- Power On Results
- Logging Tables
- Read Error
- Error: Info
class :
error code :
error explanation :
task name :
procedure name :
previous warning :
Philips Medical Systems DMC Röntgenstr. 24 22335 Hamburg
Service Support & Field Assistance Axel Duve [email protected] Box141
March 28, 2001 OM_33_rev2.doc PC version page 50-52
previous explanation :
parameter :
- Error: Trace
tube : 1 of 1...3
EN_X_C : on / off
act: kV anode during divider test [kV] : actual value of what appears in the text
min: kV anode during divider test [kV] : min deviation value of setpoint value
max: kV anode during divider test [kV] : max deviation value of setpoint value
- Functional Tests
- Test Watchdog
activates watchdog of FU_kV, red LED turns on and remains on, activate warm reset to
restart generator
no PC access & communication during this (FATAL ERROR) status
Start function: execution <OK> starts the digital - analog - digital test of the kV
setpoint and the duty cycle channel
- Monitoring
tube temp switch : short circuit (against ground) = red traffic light on at control desk
closed = normal condition, green traffic light on at control desk
open = open cable connection / tube too hot, red traffic light
at control desk
- kV Measurements KV_MEAS
to be continued
kV anode [kV] : kV anode actual value, measurement only during exp. possible
- Read Memory