Nanosurf EasyScan 2 AFM Operating Instructions
Nanosurf EasyScan 2 AFM Operating Instructions
easyScan 2 AFM
Operating Instructions
for
4
4.4: Approaching the sample ......................................................................... 44
4.4.1: Manual coarse approach................................................................................45
4.4.2: Manual approach using the motorized approach stage....................46
4.4.3: Automatic final approach ..............................................................................47
4.5: Starting a measurement .......................................................................... 49
4.6: Selecting a measurement area................................................................ 50
4.7: Storing the measurement........................................................................ 52
4.8: Creating a basic report............................................................................. 53
4.9: Further options......................................................................................... 53
CHAPTER 8: Maintenance 85
8.1: Introduction.............................................................................................. 86
8.2: The easyScan 2 AFM Scan Head .............................................................. 86
8.3: The easyScan 2 controller........................................................................ 86
5
9.2: Software and driver problems ................................................................ 88
9.2.1: No connection to microscope......................................................................88
9.2.2: USB Port error.....................................................................................................88
9.2.3: Driver problems.................................................................................................89
9.3: AFM measurement problems .................................................................. 91
9.3.1: Probe Status light blinks red .........................................................................91
9.3.2: Automatic final approach fails .....................................................................92
9.3.3: Image quality suddenly deteriorates.........................................................92
9.4: Nanosurf support ..................................................................................... 93
9.4.1: Self help................................................................................................................93
9.4.2: Assistance ............................................................................................................94
6
CHAPTER 13: Imaging 121
13.1: Introduction............................................................................................. 122
13.2: Imaging panel.......................................................................................... 123
13.3: The Imaging toolbar................................................................................ 127
13.4: Acquisition tab ........................................................................................ 128
13.4.1: Preparation group ...........................................................................................129
13.4.2: Approach group ...............................................................................................130
13.4.3: Imaging group ..................................................................................................131
13.4.4: Scripting group.................................................................................................132
13.5: Stage panel .............................................................................................. 133
13.5.1: Move Stage To dialog .....................................................................................134
13.6: Video panel .............................................................................................. 135
13.6.1: Analog video camera display ......................................................................136
13.6.2: Digital Video Camera display.......................................................................137
13.6.3: Illumination section ........................................................................................140
13.6.4: Digital Video Properties dialog ...................................................................140
13.7: Online panel............................................................................................. 142
13.7.1: Scan Position section......................................................................................142
13.7.2: Master Image section .....................................................................................143
13.7.3: Illumination section ........................................................................................144
7
CHAPTER 16: Working with documents 169
16.1: Introduction............................................................................................. 170
16.2: Data Info panel ........................................................................................ 170
16.2.1: Data Info toolbar ..............................................................................................171
16.3: Charts ....................................................................................................... 171
16.3.1: Working with multiple charts......................................................................173
16.3.2: Chart Properties dialog..................................................................................174
16.4: Gallery panel............................................................................................ 181
16.4.1: History File mask ..............................................................................................182
16.4.2: Image list.............................................................................................................182
16.4.3: Gallery toolbar ..................................................................................................182
16.4.4: Mask Editor dialog ...........................................................................................183
16.4.5: File Rename dialog ..........................................................................................185
16.5: Analysis tab.............................................................................................. 186
16.5.1: Measure group..................................................................................................187
16.5.2: Correction group..............................................................................................189
16.5.3: Roughness group.............................................................................................191
16.5.4: Filter group.........................................................................................................193
16.5.5: Tools group ........................................................................................................195
16.5.6: Report Group.....................................................................................................196
16.5.7: Scripting group.................................................................................................198
16.6: Tool panel................................................................................................. 198
8
17.9: SPM Parameters dialog........................................................................... 221
17.9.1: Imaging ...............................................................................................................222
17.9.2: Spectroscopy.....................................................................................................225
17.9.3: Lithography........................................................................................................228
17.9.4: Operating Mode...............................................................................................229
17.9.5: Approach ............................................................................................................231
17.9.6: Z-Controller........................................................................................................233
17.9.7: Signal Access .....................................................................................................235
17.10: User Signal Editor dialog........................................................................ 238
17.11: Vibration Frequency Search dialog ....................................................... 239
17.11.1: General concept ...............................................................................................239
17.11.2: Automated vibration frequency search...................................................240
17.11.3: Manual sweep controls..................................................................................241
17.11.4: Auto Frequency Config dialog ....................................................................243
17.12: Laser Alignment dialog .......................................................................... 244
17.13: Cantilever Browser dialog ...................................................................... 245
17.13.1: Cantilever Editor dialog .................................................................................247
17.14: ATS Stage and TSC 3000 driver configuration ...................................... 248
17.14.1: Stage Configuration dialog ..........................................................................250
17.14.2: The COM Port Configuration dialog..........................................................252
9
About this Manual
This manual is divided into two parts: The first part provides instructions on how to set up
and use your Nanosurf easyScan 2 AFM system. The second part is a reference for the
software that comes with the easyScan 2 AFM system. It applies to Nanosurf SPM Control
Software version 3.0. If you are using newer software versions, download the latest manual
from the Nanosurf support pages, or refer to the “What’s new in this version.pdf” file that is
installed in the Manuals subdirectory of the directory where the SPM Control Software is
installed.
The first part of the manual starts with Chapter 1: Introduction (page 12), which provides an
introduction to the easyScan 2 AFM system, and with Chapter 2: Installing the easyScan 2
AFM (page 19), which should be read when installing your system. Chapter 3: Preparing for
measurement (page 27) and Chapter 4: A first measurement (page 41) should be read by all
users, because they contain useful instructions for everyday measurements. The other
chapters provide more information for advanced or interested users.
The second part of the manual can be used as a reference for the SPM Control Software that
controls the AFM. It starts with Chapter 12: The SPM Control Software user interface (page
109) and ends with Chapter 17: Advanced settings (page 201). This part describes the
functions of all buttons, inputs, dialogs, and control panels of the SPM Control Software.
The final chapter of this manual, Quick reference (page 254), contains an index to the
software reference part of the manual for quick retrieval of the relevant information
locations.
For more information on the scripting interface of the software packages, refer to the on-
line help file easyScan 2 Script Programmers Manual that is installed together with the SPM
Control Software.
For more information on the optional Nanosurf Report software, refer to the on-line help
included with the Nanosurf Report software.
CHAPTER 1:
1.1: Introduction
The Nanosurf easyScan 2 AFM system is an atomic force microscope that can measure the
topography and several other properties of a sample with nanometer resolution. These
measurements are performed, displayed, and evaluated using the SPM Control Software.
The easyScan 2 AFM system is a modular scanning probe system that can be upgraded to
obtain more measurement capabilities. The main parts of the basic system are the
easyScan 2 AFM Scan Head, the AFM Sample stage, the easyScan 2 Controller with AFM
Basic module, and the SPM Control Software.
The content of the system and the function of its major components are described in this
chapter. Detailed technical specifications and system features can be found in Chapter 11:
Technical data (page 99).
Several other Nanosurf products can be used in conjunction with the easyScan 2 AFM:
• AFM Dynamic Module: adds dynamic mode measurement capabilities for measuring
delicate samples.
• AFM Mode Extension Module: adds phase contrast, force modulation and current
measurement capabilities.
• AFM Video Module: allows observation of the approach on the computer screen. This is
useful when observation using the lenses is impractical.
• Signal Modules: allow monitoring signals (Module S) and creating custom operating
modes (Module A). Refer to Section 11.4.2: The Signal Modules (page 105) for more details.
• Nanosurf Micrometer Translation Stage: allows locating of a position on a sample with
micrometer accuracy. Should be used together with the Nanosurf easyScan 2 Sample
Stage.
• Nanosurf Report: software for simple automatic evaluation and report generation of
SPM measurements.
• Nanosurf Analysis: software for detailed analysis of SPM measurements.
• Scripting Interface: software for automating measurements. Refer to Section 13.4.4:
Scripting group (page 132) and the Programmer’s Manual for more details.
• Lithography Option: software for professional lithography applications. Refer to Chapter
15: Lithography (page 151) for more information.
• The Nanosurf isoStage: a highly compact active vibration isolation table, equipped with
a special easyScan 2 AFM Sample Stage (The isoStage Adapter Plate), and dedicated for
use with Nanosurf AFM scan heads.
• The Halcyonics_i4 Active Vibration Isolation Table: a larger and heavier active vibration
isolation solution, which features load adjustment.
12
COMPONENTS OF THE SYSTEM
1 2 3
4 5 6 7
18
8 9 17
19
1. easyScan 2 Controller with a built-in AFM Basic Module, and optionally with built-in
AFM Dynamic Module, AFM Mode Extension Module, Video Module electronics and
Signal Module A or S electronics.
2. easyScan 2 AFM Scan Head(s) with AFM Video Camera (comes with AFM Video
Module).
3. Scan Head Case.
4. USB cable.
5. Mains cable.
6. Scan Head cable, connects the Scan Head to the easyScan 2 controller.
7. Video Camera cable (see item 2; comes with AFM Video Module).
13
CHAPTER 1: THE EASYSCAN 2 AFM
14
CONNECTORS, INDICATORS AND CONTROLS
1 2 3 8
4 7
5 6
9 10 11 12
15
CHAPTER 1: THE EASYSCAN 2 AFM
Camera
connector
Figure 1-3: Parts of the Scan Head. Scan Head with video Camera.
Leveling screws
For coarse approach of the sample (Section 4.4.1: Manual coarse approach (page 45)), and
for aligning the plane of the scanner with the plane of the sample (Section 5.2: Adjusting the
measurement plane (page 57)).
Scan Head cable connector
For connecting the Scan Head cable that also connects to the easyScan 2 Controller.
Ground connector
For connecting a cable that puts the sample or the Sample Holder at the same ground
potential as the scan head.
Alignment chip and hole for cantilever insertion tool
Used for mounting the cantilever on the scan head (Section 3.3: Installing the cantilever
(page 29)).
Scan Head serial number
Shows what serial number and version of the scan head you have.
16
CONNECTORS, INDICATORS AND CONTROLS
Module lights
Scan Head lights
Probe Status light
Video Out
connector Video In
(optional) connector
(optional)
Signal Out
connector Signal In Scan head
(optional) connector cable connector
(optional)
Controller
Power Serial number
S/N: 23-05-001
switch
USB outputs
(to dongle)
USB power
light
Mains power USB input USB active
connector (from PC) light
17
CHAPTER 1: THE EASYSCAN 2 AFM
– Blinking red
There is no laser signal with which to do feedback: see Section 9.3.1: Probe Status light
blinks red (page 91) for possible causes and for more information on how to resolve this
situation).
The Scan Head lights
Indicate the Scan Head type that is connected to the instrument. The Scan Head lights blink
when no Scan Head can be detected, or when the controller has not been initialized yet.
The Module lights
Indicate the modules that are built in into the controller. The module lights blink when the
controller has not been initialized yet. During initialization, the module lights are turned on
one after the other.
The Video Out connector
An RCA/cinch connector that outputs a PAL video signal that can be connected to a video
monitor.
18
CHAPTER 2:
IMPORTANT
• Make sure that the mains power connection is protected against excess voltage
surges.
• Place the instrument on a stable support in a location that has a low level of building
vibrations, acoustic noise, electrical fields, and air currents.
IMPORTANT
• Never touch the cantilever tips, the cantilevers (Figure 1-2: Contents of the Tool set
(page 15), item 10 and item 11), or any part of the Cantilever deflection detection
system (Figure 3-2: Cantilever deflection detection system (page 30)).
• Ensure that the surface to be measured is free of dust and possible residues.
• Always put the Scan Head it in the Scan Head Case during transport and storage (see
Figure 7-2: Scan Head storage).
IMPORTANT
If you inadvertently connected the USB Cable to both the easyScan 2 Controller and a
running computer, Windows will attempt to install drivers for the newly found
hardware. When this happens, do the following:
• Do NOT break off the installation!
• Insert the Software Installation CD (if the Software Installation program should start,
choose “Exit” first) and follow the steps described for the USB Video Adapter in Section
2.2.4: Manual installation of the USB Video Adapter driver (page 25) to let Windows
search for the necessary drivers on this CD.
• When this process has finished, disconnect the USB cable from the computer, finish
the remaining steps below, and then go through the Software Installation procedure
as described in Section 2.2: Installing the SPM Control Software (page 23).
20
INSTALLING THE HARDWARE
2 Connect the Scan Head Cable (item 6) to the easyScan 2 controller but not to the Scan
Head (item 2) yet.
3 Connect the easyScan 2 Controller to the mains power using the Mains Cable (item 5),
but do not turn on the controller yet.
Figure 2-1: Measurement setup. Complete easyScan 2 AFM system with Sample Stage, Micrometer
Translation Stage, AFM Video Module and Signal Module A.
21
CHAPTER 2: INSTALLING THE EASYSCAN 2 AFM
> Connect the Break-out cable (Figure 1-1: Components (page 13), item 17) to the Signal
Out connector on the Controller (Figure 1-4: The easyScan 2 controller (page 17)).
In case of an upgrade, the Controller must be sent in to your local Nanosurf distributor for
installing the Signal Module S electronics inside the Controller.
! WARNING
LASER RADIATION (650nm)
DO NOT STARE INTO THE BEAM
OR VIEW DIRECTLY WITH OPTICAL
INSTRUMENTS (MAGNIFIERS)
CLASS 2M LASER PRODUCT
Older Scan heads may contain lasers with 850 nm wavelength infrared light, and lower
optical power. These lasers have class 1 rating, which does not require special protection,
even when viewing the laser radiation directly with optical instruments. The wavelength
and laser class are indicated next to the serial number on the scan head (see Figure 1-3: Parts
of the Scan Head (page 16)).
22
INSTALLING THE SPM CONTROL SOFTWARE
IMPORTANT
Do not run any other programs while installing the easyScan 2 software.
23
CHAPTER 2: INSTALLING THE EASYSCAN 2 AFM
IMPORTANT
The easyScan 2 Installation CD contains calibration information (.hed files) specific to
your instrument! Therefore, always store (a backup copy of ) the CD delivered with the
instrument in a safe place.
24
INSTALLING THE SPM CONTROL SOFTWARE
25
CHAPTER 2: INSTALLING THE EASYSCAN 2 AFM
Windows Vista/7
To manually install the USB Video Adapter driver:
1 In the “Found New Hardware” dialog for an “Unknown Device”, click the “Locate and
install driver software (recommended)” button.
The User Account Control (UAC) dialog may pop up after pressing this button,
displaying the text “Windows needs your permission to continue” for a “Device driver
software installation”. If this is the case:
> Click the “Continue” button.
2 In the next dialog, which states “Windows couldn’t find driver software for your
device”, click the “Browse my computer for driver software (advanced)” button.
3 In the next dialog, “Browse” to your CD drive (usually D:) or manually type “D:\” (or the
corresponding drive letter) into the Path field. Make sure “include subfolders” is
checked. Then click the “Next” button.
Windows begins searching the specified path, and — since an unsigned driver is found
— a Windows Security window opens, stating that “Windows can’t verify the publisher
of this driver software”
4 In the Windows Security window, click the “Install this driver software anyway” button.
The Found New Hardware window now displays a “USB Video Adapter” and driver
installation will take place.
5 When Windows has finished installing the Video driver software, click “Close”.
Windows 2000/XP
To manually install the USB Video Adapter driver:
1 When the “Found New Hardware” dialog displays the text “Can Windows connect to
Windows Update to search for software?”, select “No, not this time” and click “Next”.
2 In the next dialog, select “Install automatically (recommended)” and click “Next”.
Windows now begins searching for the appropriate driver, and — since an unsigned
driver is found — a Warning window opens, stating that “The software that you are
trying to install for this hardware: USB Video Adapter has not passed Windows Logo
testing for compatibility with Windows XP”.
3 In the Warning window, click the “Continue anyway” button.
Under some circumstances a “Files Needed” window may pop up. If this is the case:
> “Browse” to the Installation CD’s “DriverVideo” folder and click “OK”.
4 When Windows has finished installing the Video driver, click “Finish”.
26
CHAPTER 3:
Preparing for
measurement
0
0
CHAPTER 3: PREPARING FOR MEASUREMENT
3.1: Introduction
Once the system has been set up (see Chapter 2: Installing the easyScan 2 AFM (page 19)),
the instrument and the sample have to be prepared for measurement. The preparation
consists of three steps: Initializing the easyScan 2 Controller, Installing the cantilever, and
Installing the sample.
28
INSTALLING THE CANTILEVER
> Select the desired operating mode from the Operating mode drop-down menu
by clicking the currently selected operating mode:
29
CHAPTER 3: PREPARING FOR MEASUREMENT
Figure 3-1: Cantilever. (Left) Alignment system. (Center) Cantilever chip viewed from the top. (Right)
Cantilever, 450 μm long, 50 μm wide with integrated tip.
Photodetector Laser
Alignment chip
Cantilever
holder spring
Hole for cantilever
insertion tool
30
INSTALLING THE CANTILEVER
! WARNING
LASER RADIATION (650nm)
DO NOT STARE INTO THE BEAM
OR VIEW DIRECTLY WITH OPTICAL
INSTRUMENTS (MAGNIFIERS)
CLASS 2M LASER PRODUCT
Older Scan heads may contain lasers with 850 nm wavelength infrared light, and lower
optical power. These lasers have class 1 rating, which does not require special protection,
even when viewing the laser radiation directly with optical instruments. The wavelength
and laser class are indicated next to the serial number on the scan head (see Figure 1-3: Parts
of the Scan Head (page 16)).
CAUTION
• Nothing should ever touch the cantilever.
• The Cantilever Holder Spring is very delicate. NEVER touch or pull on it! It will become
bent and unusable otherwise!
• Always close the DropStop before handling the cantilever. If you fail to do so, the
cantilever can fall into the Scan Head, causing malfunction of the microscope,
particularly of the scanner.
• If a cantilever has dropped into the Scan Head, and the microscope is malfunctioning,
contact your local support. Never open the Scan Head, because this will cause
damage to the scanner and the laser beam deflection detection system.
31
CHAPTER 3: PREPARING FOR MEASUREMENT
CAUTION
Always store and ship the Scan Head with a cantilever installed. Otherwise, the
cantilever holder spring may damage the alignment chip.
32
INSTALLING THE CANTILEVER
Figure 3-4: Mounting the cantilever. (top left) inserting the cantilever insertion tool, (top right)
inserting/removing the cantilever, (bottom right) correctly inserted cantilever.
Figure 3-5: Cantilever Alignment. (Left) correct: the mirrored environment shows a reflection that is
continuous over the cantilever and the alignment chip, and small triangular gaps can be seen between
the edges of the alignment chip and the corners of the cantilever chip, (center & right) incorrect: the
mirrored environment shows a reflection that is different on the cantilever and on the alignment chip,
and no nice triangular gaps can be discerned.
33
CHAPTER 3: PREPARING FOR MEASUREMENT
Grid: 10 μm / 100 nm
The Grid: 10 μm / 100 nm can be used for testing the XY-calibration of the 70 μm and
100 μm scanners, and for testing the Z-calibration. It is manufactured using standard silicon
34
INSTALLING THE SAMPLE
100 nm
10 µm
Calibrated values of period and height (with 3% accuracy) are printed on the package.
Certified Calibration grids are available as an option.
Grid: 660 nm
The Grid: 660 nm can be used to test the XY-calibration of the 10 μm scanner. Depending
on the grid version, it consists either of:
• Silicon oxide hills on a silicon substrate with a period of 660 nm and an unspecified
height. The approximate height is 149 nm,
• Holes in a silicon oxide layer with a period of 660 nm and an unspecified depth. At the
time of writing, the approximate depth is 60 nm.
35
CHAPTER 3: PREPARING FOR MEASUREMENT
Sample specifications:
Size: 5 mm × 5 mm
Material: Silicon oxide on silicon
Structure: Square array of holes or hills in the silicon oxide layer
660 nm. Calibrated value of period (with 3% accuracy) is printed on the
Period:
package. Certified Calibration grids are available as an option.
60 nm
660 nm
Flatness sample
The Flatness sample is a polished silicon sample. It can be used for testing the Flatness of
the scanned plane.
Sample specifications:
Size: 5 mm × 5 mm
Material: Silicon
Thickness: Approx. 320 μm
CD-ROM piece
Sample for demonstrating the AFM imaging. The CD sample is a piece from a CD, without
any coating applied to it.
Sample specifications:
Material: Polycarbonate
Structure: 100 nm deep pits arranged in tracks that are spaced 1.6 μm apart.
Microstructure
Sample for demonstrating AFM imaging (no longer available). The microstructure is
approximately the negative of the Grid: 10μm / 100nm. It consists of holes in a silicon oxide
36
INSTALLING THE SAMPLE
layer with an unspecified period and depth. The approximate period is 10 μm, the
approximate depth is 100 nm.
The surface of the graphite sample can be cleaned when it is very dirty or uneven. Due to
the layered structure of graphite this can easily be done using a piece of adhesive tape
(Figure 3-8: Cleaving graphite):
1 Put the sample on the table using a pair of tweezers.
2 Stick a piece of adhesive tape gently to the graphite and then pull it off again.
The topmost layer of the sample should stick to the tape.
3 Remove any loose flakes with the tweezers.
The graphite sample is now ready for use and should not be touched anymore.
37
CHAPTER 3: PREPARING FOR MEASUREMENT
For stand-alone measurements, put the Scan head directly on top of the sample. Protection
feet to be placed under the three alignment screws are provided to protect delicate
samples from being scratched (Figure 1-2: Contents of the Tool set (page 15), item 2).
Figure 3-9: The easyScan 2 AFM Scan Head on its Sample Stage
38
INSTALLING THE SAMPLE
4 Put the sample on the sticky side of the Post-it® note, and press on it lightly.
The result should resemble Figure 3-10: Sample mounted onto the Sample Holder.
5 Remove the scan head and place the Sample Holder on top of the Sample plate.
It is recommended to always connect the Sample Holder to the ground connector on
the Scan Head using the ground cable.
6 Place the scan head back onto the Sample Stage, but be sure to keep a safe distance
between the cantilever and the sample surface. If necessary, adjust the tip–sample
distance via the leveling screws before placing back the scan head.
39
CHAPTER 3: PREPARING FOR MEASUREMENT
40
CHAPTER 4:
A first measurement
0
0
CHAPTER 4: A FIRST MEASUREMENT
4.1: Introduction
In this chapter, step-by-step instructions are given to operate the microscope and to
perform a simple measurement. More detailed explanations of the software and of the
system can be found elsewhere in this manual.
42
PREPARING THE INSTRUMENT
IMPORTANT
• Never touch the cantilever or the surface of the sample! Good results rely heavily on a
correct treatment of the tip and the sample.
• Avoid exposing the system to direct light while measuring. This could influence the
cantilever deflection detection system and reduce the quality of the measurement.
Prepare the instrument as follows (see Chapter 3: Preparing for measurement (page 27) for
more detailed instructions):
1 If you have the AFM Dynamic Module, install an NCLR type cantilever, otherwise install
a CONTR type cantilever.
2 Install one of the samples from the Nanosurf AFM Basic Sample Kit or Calibration
Sample Kit. Preferably, install the 10 μm Calibration grid when using a 70-μm or 110-
μm scan head, and the 660 nm Calibration grid when using a 10-μm scan head.
The measurement examples shown here were made with the 10 μm Calibration grid.
To make sure that the configuration is correct, do the following:
1 Open the menu item “File” >> “Parameters” >> “Load...”, and load the file “Default_EZ2-
AFM.par” from the directory that holds the default easyScan 2 configurations.
Usually this is “C:\Program Files\Nanosurf\Nansurf easyScan 2\Config”.
2 Open the menu item “File” >> “Chart Arrangement” >> “Load...”, and load the file
“Default_EZ2-AFM.chart” from the directory that holds the default easyScan 2
configurations.
43
CHAPTER 4: A FIRST MEASUREMENT
2 In case of a drop-down menu selection list, change the selection using the mouse or
the up and down arrows on the keyboard. In case of a numerical value, use one of the
following methods:
• Use the up and down arrow keys on the keyboard to increase or decrease its value.
The new value is automatically used after one second.
• Click the arrow buttons next to the parameter value with the mouse pointer.
Normally, the parameter value is changed by a small amount (usually in the range of
1–10%). Some edit boxes are doubling or dividing the parameter value by two (e.g.
the “points/line” parameter). The new value is automatically used after one second.
• Enter the new value using the keyboard. The entered value is applied upon pressing
the “Enter” or “Return” key, or by activating another input. The entered value is
discarded upon pressing the “Esc” key. The unit prefix can be changed by typing one
of the following keyboard keys:
f = femto space bar = no prefix
p = pico k = kilo
n = nano M (shift-m) = mega
u = micro G (shift-g) = giga
m = milli T (shift-t) = tera
Examples: if the basic unit is Volts, type “m” to change to millivolts, type the space
bar for volts, type “u” for microvolts.
Sometimes the program will change an entered value to a slightly different value. This
happens when the desired value is outside the digitization range of the easyScan 2
controller, for example due to resolution or timing limits. In such cases, the desired
value is automatically changed to the nearest possible value.
44
APPROACHING THE SAMPLE
– Green
The approach has finished successfully: the Z-scanner is within the measuring range.
To prepare for the approach process:
> Select the Acquisition tab
The controls for positioning the cantilever with respect to the sample are located in the
Approach group.
During the approach steps described in the following sections, use the side view of the
cantilever to judge the distance between tip and sample surface:
> If the AFM Video Camera is installed, select “Side view” in the Video panel located in
the Info pane (see Section 13.6: Video panel (page 135)).
If the Video Camera is not installed, use the side view lens of the scan head to observe
the sample instead.
45
CHAPTER 4: A FIRST MEASUREMENT
Figure 4-1: Side view of the cantilever after manual coarse approach.
1 Observe the distance between tip and sample in the side view of the integrated optics.
Figure 4-2: View of sample and cantilever after manual approach using the motorized approach
stage. (Left) side view, (Right) top view.
46
APPROACHING THE SAMPLE
2 While observing the tip–sample distance, click and hold the “Advance” button in the
Approach group of the Acquisition tab until the tip is close enough to the sample:
The tip should not come closer to the sample than a few times the cantilever width
(Figure 4-2: View of sample and cantilever after manual approach using the motorized
approach stage, left).
Now that the sample is in focus, the top view image can be used to find a suitable location
to measure on. In top view, the sample is seen from a direction perpendicular to its surface
(see Figure 4-2: View of sample and cantilever after manual approach using the motorized
approach stage, right).
To use the top view:
1 If the AFM Video Camera is installed, select Top view in the Video panel (see Section
13.6: Video panel (page 135)). If the Video Camera is not installed, use the top view lens
to observe the sample instead.
2 If necessary, move the Sample Holder to find a suitable location that is free of dust
particles.
47
CHAPTER 4: A FIRST MEASUREMENT
frequency. It is instructive to see both frequency sweep measurements in all detail at least
once. To do this, it is possible to manually perform the frequency sweeps:
1 In the Preparation group of the Acquisition tab, click the “Freq. Sweep” button:
The Vibration Frequency Search dialog now opens (see Section 17.11: Vibration
Frequency Search dialog (page 239) for details).
2 Click the “Auto frequency set” button.
The SPM Control Software now sets appropriate values for the coarse and fine sweeps
and performs these sweeps. The fine sweep will overwrite the data of the coarse
sweep in the charts displayed in the “Vibration frequency search” dialog. To see the
results of the individual sweeps:
> Press the “Coarse sweep” and “Fine sweep” buttons sequentially.
Before final approach of the sample, it is necessary to set the scanning and feedback
parameters of the control software to suitable initial values. The easiest way to do this is to
use the “Auto Set” wizard:
1 In the Preparation group of the Acquisition tab, click the “Auto Set” button:
A dialog will pop up, which will ask you some basic questions about your sample and
your measurement needs.
2 Answer the questions of the wizard to the best of your knowledge.
For descriptions of the features of standard Nanosurf samples refer to Section 3.4.2:
Nanosurf samples (page 34).
Now that the initial software settings have been given suitable values, you need to name
the measurement series (see Section 16.4.1: History File mask (page 182)). Each completed
measurement (scan/image) will be temporarily saved (automatically) in the History folder
under this name, with index numbers (or, optionally, date and time attributes) added to
identify the individual measurements. It is best to enter the measurement series’ name
now, since the control software will (by default) start measuring as soon as the final
approach is done. It is also strongly recommended to move all relevant measurements to a
48
STARTING A MEASUREMENT
new folder when you are finished, since the files in the History folder will be overwritten
over time (see Max History Files (page 211)).
To set the measurement series name:
1 Activate the Gallery panel (see Section 16.4: Gallery panel (page 181)) in the Info pane.
2 Click the History tab at the top of the Gallery panel with the mouse.
3 In the entry box at the top of the panel, enter a name by hand or use the Mask Editor
dialog (see Section 16.4.4: Mask Editor dialog (page 183)) to create the name mask.
If no [INDEX] attribute is explicitly added to the name mask, it will be implicitly applied
to the end of the file name so that individual measurements can be stored and
distinguished.
The automated final approach can now be started. To do this:
1 In the Approach group of the Acquisition tab, click the “Approach” button:
The cantilever is moved towards the sample via the approach stage, with the Z-
Controller turned on. This movement continues until the Z-Controller error becomes
zero. From this point onward, the distance between sample and tip is maintained
automatically by the electronics. The probe status light changes to a constant green,
and a message “Approach done” appears:
49
CHAPTER 4: A FIRST MEASUREMENT
> Start measurements manually by clicking the “Start” button in the Imaging group of
the Acquisition tab:
Two representations of the ongoing measurement are drawn in the Imaging panel. One
representation is a color coded height image (Topography) called a Color map. The other is
a plot of height as a function of X* position called a Line graph. With the current settings,
the software automatically adjusts the contrast of the Color map, and height range of the
Line graph to the data that have been measured.
To judge the imaging quality, watch the displays until at least one fourth of the
measurement has been completed.
IMPORTANT
Measurements on the micrometer/nanometer scale are very sensitive to environment
influences. Direct light or fast movements — causing air flow and temperature
variations near the Scan Head — can influence and disturb the measurement.
When a measurement contains large disturbances, or no two scan lines are similar, stop
measuring and reduce or eliminate the disturbances. To do this:
> Click the “Stop” button in the Imaging group of the Acquisition tab and follow the
instructions in Chapter 9: Problems and solutions (page 87):
50
SELECTING A MEASUREMENT AREA
The mouse pointer becomes pen-shaped when moving over the color map.
3 Click on one corner of the region to be selected using the left mouse button, and keep
the button pressed.
4 Drag the mouse to the other corner of the region.
The size and the position of the square are shown in the Tool results panel of the Info
pane.
5 Release the mouse button when the size of the square covers approximately one
period of the grid.
6 Confirm the selection by double clicking the color map graph using the left mouse
button. Now the selection is enlarged to the whole display size. You can abort the
zoom function by clicking the “Zoom” button again.
The microscope will now start measuring a single grid period. Once the measurement is
completed, you should get an image such as the one in Figure 4-4: Zoomed measurement,
left. The depression in the sample surface to the sides of the structure is due to the LineFit
data filter used. To remove it, click on “LineFit” and select the data filter “Raw Data” from the
drop-down menu. The chart will now look like Figure 4-4: Zoomed measurement, right.
51
CHAPTER 4: A FIRST MEASUREMENT
Figure 4-4: Zoomed measurement. (Left) with line subtraction, (Right) without line subtraction.
The current measurement is immediately stored and will show up in the History page
of the Gallery panel, together with all other finished/stored measurements (see Section
16.4: Gallery panel (page 181) for details). In addition, the captured document will
remain open in the Document space of the SPM Control Software.
Measurement documents in the temporary History folder should always be moved to a
new location for permanent storage when you are done measuring. For details on how to
do this, see Save as (page 182). Measurement documents thus permanently stored can
always be loaded with the SPM Control Software or the optional Nanosurf Report or
Nanosurf Analysis software packages for later viewing, analysis, and printing. A brief
introduction on how to create a basic report using the Report software is given in the next
section. For more detailed information on starting and using the Report software, see
Section 22.2: Creating a report (page 452), or refer to the Nanosurf Report online help.
52
CREATING A BASIC REPORT
The Nanosurf Report software will now launch, import the currently open
measurement, and evaluate the data using the default template.
IMPORTANT
After a fresh installation of the Report software, the Report software has to be run at least
once before it can be automatically started from within the SPM Control Software. To run
the Report software for the first time, select it from the Microsoft Windows “Start” menu.
53
CHAPTER 4: A FIRST MEASUREMENT
54
CHAPTER 5:
Improving measurement
quality
0
0
CHAPTER 5: IMPROVING MEASUREMENT QUALITY
56
ADJUSTING THE MEASUREMENT PLANE
3 Remove interfering electromagnetic field sources, such cathode ray tube displays,
loudspeakers, …
57
CHAPTER 5: IMPROVING MEASUREMENT QUALITY
'Rotation'
angle
Measurement plane
'Y-Slope'
angle Y* X*
Image area
'X-Slope' angle
Y
X Scanner XY-plane
Z, Z*
Figure 5-2: Sample and measurement orientation before slope adjustment
58
ADJUSTING THE MEASUREMENT PLANE
The SPM Parameter dialog will now open (see Section 17.9: SPM Parameters dialog
(page 221) for details). This dialog contains all possible parameters and settings that
influence the behavior of your Nanosurf easyScan 2 AFM system.
59
CHAPTER 5: IMPROVING MEASUREMENT QUALITY
IMPORTANT
If the sample surface contains large jumps or steps in height, the line fitting procedure
used to determine the slope in X- and Y-direction may not deliver the best possible
results. In such cases it is recommended to perform a manual slope adjustment as
described below.
60
JUDGING TIP QUALITY
When the image quality deteriorates dramatically during a previously good measurement,
the tip has most probably picked up some material from the sample. As a result, the image
in the color map charts consist of uncorrelated lines or the image appear blurred (see
Chapter 5: Measurement containing tip artifacts). In such cases:
> Follow the suggestions in Section 9.3.3: Image quality suddenly deteriorates (page 92).
If these do not help, the cantilever should be replaced.
When all peaks in the image have the same, usually triangular shape, the sharp end of the
tip may have broken off. The measured structure reflects the shape of the tip rather than
that of the sample and is called a tip artefact. In such cases:
> Replace the cantilever.
61
CHAPTER 5: IMPROVING MEASUREMENT QUALITY
62
CHAPTER 6:
Operating modes
CHAPTER 6: OPERATING MODES
6.1: Introduction
This chapter instructs you on how to use the Static Force mode, Dynamic Force mode, Phase
Contrast mode, Force Modulation mode, and Spreading Resistance mode. If you are unfamiliar
with Atomic Force Microscopy techniques, it is recommended to first read Chapter 10: AFM
theory (page 95).The amount of available modes depends on the Scan head and the
available modules built into the easyScan 2 controller. The modules required to be able to
use a certain operating mode are listed in Table 6-1: Operating modes and required modules.
If additional hardware is required, it will be discussed in the respective section explaining
the measurement mode.
64
DYNAMIC FORCE MODE
65
CHAPTER 6: OPERATING MODES
Phase measurement
In Phase contrast mode, the phase of the measured cantilever vibration is compared to the
phase of a reference sine wave. The phase comparison is performed by multiplying the
measured vibration signal with the reference. The measured phase, φmeasured , is the output
of this multiplication, and is related to the actual phase shift between the measured
vibration and the excitation sine wave, φactual , according to the following equation:
Figure 6-5: Phase shift folded back. Frequency spectrum with phase shift folded back over the +90°
limit.
66
FORCE MODULATION MODE
> Select the Phase Contrast mode in the Preparation group of the Acquisition tab.
A new Color map and line chart for the phase data will automatically be added to the
chart area. If necessary, you can increase the size of the measurement pane to make
room for the new charts.
After approach, the reference phase is usually set automatically in such a way that the
average phase shift lies in the center of the measurement range. To re-adjust the phase
shift:
Click the “Freq. Sweep” button in the Preparation group of the Acquisition tab.
If you do not want the reference phase to be set automatically:
1 Open the SPM Parameter dialog by clicking the “More” button in the Mode Properties
section of the Imaging window.
2 Uncheck the reference phase “Auto set” checkbox.
3 Enter the reference phase manually using the “Reference phase” input box.
67
CHAPTER 6: OPERATING MODES
5 In the Preparation group of the Acquisition tab, click the “Freq. seep” button.
6 Click the “Auto frequency set” button to measure the frequency characteristic of the
cantilever.
7 Make a note of the cantilever resonance frequency from the sweep chart and set
“Vibration frequency” to this value.
8 Set “Vibration amplitude” to the desired modulation amplitude.
9 Increase the vibration amplitude if no proper Force Modulation signal can be
measured.
IMPORTANT
• Standard CONTR cantilevers cannot be used due to the surface oxide on the tip, and
the semiconducting contact between the tip and the sample.
68
KELVIN PROBE FORCE MICROSCOPY
IMPORTANT
• EFM cantilevers cannot be used for static force mode measurements, due to their
insufficient width.
3 Install the cantilever as described in Section 3.3: Installing the cantilever (page 29).
4 In the Preparation group of the Acquisition tab, select the Spreading Resistance
operating mode.
5 In the Z-Controller section of the Imaging window, verify the Setpoint.
6 In the Mode Properties section, set the Tip voltage to your requirements.
6.7.1: Introduction
Kelvin Probe Force Microscopy (KPFM) is an extension of AFM. The technique was first
published in 1991 by Nonnenmacher and co-workers. Using KPFM, images can be recorded
that contain information on the local work function or local contact potential difference
between tip and sample.
69
CHAPTER 6: OPERATING MODES
KPFM, the FlexAFM has demonstrated best KPFM performance and is therefore the
instrument of choice for this type of measurement.
Hardware setup
The KPFM setup for Nanosurf AFMs will schematically look as depicted in Figure 6-6: KPFM
connection schematics. The User Controller 1 of the SPM Control Software is used as PID
controller to regulate the applied DC voltage. Signal Module A is used for communication
between the AFM and the Lock-in amplifier.
Three cables are required to connect the Signal Module A (featuring BNC connectors) to the
Lock-in amplifier (depending on the model used, most likely also equipped with BNC
connectors).
The reference output of the Lock-in amplifier is used to generate the AC voltage. This wire
is connected to the cantilever via the “Tip Voltage” connector on the input side of Signal
Module A.
70
KELVIN PROBE FORCE MICROSCOPY
Detector
Lock In
Cantilever
Alternative setup
Instead of applying the alternating voltage to the cantilever tip, it can also be applied to the
sample support. In this case the ground connection has to be removed from the sample
support and needs to be replaced by a direct connection to the “Reference out” connector
71
CHAPTER 6: OPERATING MODES
of the Lock-in amplifier, which in turn should no longer be connected to “Tip Voltage” of
Signal Module A.
6.7.4: Procedures
Lock-in settings
There are several approaches for selecting the reference frequency of the Lock-in:
– AC frequency below the resonance frequency of the cantilever
At these low frequencies, the cantilever is in a forced oscillation and the mechanical
response of the cantilever and the excitation are almost at the same phase or close to the
opposite phase (depending on whether the forces between tip and surface are attractive
or repulsive).
– AC frequency at a higher oscillation mode of the cantilever
In these situations, the cantilever oscillation is self-amplifying.
Note that in both cases, the frequency that can be applied to the tip is limited by a filter
present in the “Tip Voltage” channel of Signal Module A. The maximum frequency that can
be applied depends on the version number of your Signal Module A.
To identify the installed Signal Module A version number:
1 In the upper right corner of the control software window, just beneath the Window’s
Close button, click the “About” button:
72
KELVIN PROBE FORCE MICROSCOPY
Software settings
Before KPFM imaging can start, the control software must be adjusted to allow application
of a voltage to the tip, and the phase settings of the Lock-in amplifier must be optimized.
There is a Wizard that can automatically prepare the software and some of the settings for
you (see More (page 129)), but it is suggested to go through the manual procedure at least
once so that you know what to adjust and optimize when you really do use the wizard.
To allow tip voltages to be set:
1 In the “Hardware” group of the “Settings” tab, select the “Signal Access” button:
2 In the SP Parameter dialog, which now opens with the Signal Access page, select
“Voltage Source Output” as Tip Signal mode in the Signal Module Config section (see
Figure 6-7: Activating Voltage Source Output).
73
CHAPTER 6: OPERATING MODES
Further optimization can optionally be carried out as outlined below (see also Figure 6-11:
Tip Voltage Spectroscopy). This procedure will additionally give an indication of the signal
variations that are to be expected in the imaging experiment.
To further optimize the phase settings of the Lock-in amplifier, perform the following steps
while approached:
1 In the Signal Access page of the SPM Parameter dialog, select “Enable Input 1”.
2 Configure User Input 1 and User Output 1 (see Figure 6-9: Configuration of User Input 1
and User Output 1) by clicking the respective “Config” buttons.
For correct voltage display, you will need to apply the conversion that the Lock-in
amplifier produces (see Example below).
Example
• If the X-channels output has a range equal to ±10 V, the sensitivity range of the lock-
in amplifier can be inserted directly into the Calibration section of the User Signal
Editor for User Input 1 (see Figure 6-9: Configuration of User Input 1 and User Output 1,
left, which shows the calibration settings for a ±10 mV sensitivity range).
• If the X-channel output has a range different from ±10 V, the sensitivity range that
would be equivalent to ±10 V should be entered. Thus, a ±10 mV sensitivity range on
a ±2.5 V output range would correspond to a ±40 mV sensitivity range on a ±10 V
output range.
74
KELVIN PROBE FORCE MICROSCOPY
Figure 6-9: Configuration of User Input 1 and User Output 1. (Left) Suggested name and
calibration settings for User Input 1; see example above. (Right) Suggested name and calibration
settings for User Output 1.
3 Make sure that “Enable Feedback Loop” in the User Controller section of the SPM
Parameter dialog is unchecked.
4 Open the Spectroscopy window (see Figure 6-11: Tip Voltage Spectroscopy) by clicking
the Spectroscopy tab at the bottom of the Measurement pane.
5 In the Parameters section of the Spectroscopy window, click the “More” button.
The SPM Parameter dialog will open on the Spectroscopy page.
6 Set all parameters and settings to the ones shown in Figure 6-10: Spectroscopy
parameters
7 Select User Input 1 (or the name assigned to it) to be displayed in the Line graph chart
(see Figure 6-10: Spectroscopy parameters).
8 Adjust the phase on the Lock-in amplifier until the maximum slope is observed.
KPFM Imaging
After proper adjustment of the phase, imaging can start (see Figure 6-12: KPFM imaging). On
Nanosurf AFMs, the KPFM signal and sample topography can be recorded simultaneously
in a single scan. To operate the FlexAFM in simultaneous KPFM mode:
1 In the Mode Properties section of the Imaging window, make sure that “Tip Voltage” is
set to 0 V.
2 Set all Signal Access parameters and settings to the ones shown in Figure 6-7:
Activating Voltage Source Output (page 73).
75
CHAPTER 6: OPERATING MODES
3 Configure User Output 1 (see Figure 6-9: Configuration of User Input 1 and User Output
1, right).
4 In the User Controller section of Signal Access page of the SPM Parameter dialog, check
or uncheck the “Inverted Feedback” option (see also Figure 6-7: Activating Voltage
Source Output (page 73)).
The correct setting depends on the settings and type of Lock-in amplifier. If set
incorrectly, the feedback responds in the wrong direction and the output voltage will
go to its maximum or minimum value. When this happens:
> Change the status of the “Inverted Feedback” check box.
76
KELVIN PROBE FORCE MICROSCOPY
Figure 6-11: Tip Voltage Spectroscopy. The phase optimization procedure. The “Tip voltage” setting
(Mode Properties section of the Imaging window) is ignored during running of the voltage ramps. In
the example shown here, a sequence of 8 curves was recorded to allow an easy comparison between
the individual curves.
77
CHAPTER 6: OPERATING MODES
Tip:
If signals are not as expected, the applied voltage can be checked with a multimeter. The
tip voltage specified in the software should be directly measurable on the cantilever
chip or spring. One test lead of the multimeter is pressed gently onto the cantilever chip,
while the other is connected to the ground.
CAUTION
The test lead on the cantilever chip must be put on with great care, as excessive force on
the chip will cause damage to the cantilever holder, optical unit or approach
mechanism!
Tip
Upon request, a parameter file for KPFM measurements can be obtained from
Nanosurf’s Support department.
78
KELVIN PROBE FORCE MICROSCOPY
Figure 6-13: KPFM measurement. Topography (left) and KPFM signal (right) of local charges that
were placed on an insulating (oxide) surface layer in a “Swiss cross” pattern. Image courtesy: Marcin
Kisiel, Thilo Glatzel and students of the Nanocurriculum of the University of Basel.
KPFM Wizard
As mentioned earlier, a KPFM wizard exists, which can set many of the parameters and
settings for you, although the optimization processes described above will remain to be
performed. To start the wizard:
1 In the Preparation group of the Acquisition tab, Select “KPFM” from the drop-down
menu of the “More” button:
79
CHAPTER 6: OPERATING MODES
1. The same scan area must be scanned twice, thereby consuming more time.
2. A lateral mismatch between the topography and KPFM images may occur as a result
of drift or vertical movement.
These disadvantages do not occur during simultaneous collection of the topography and
surface potential data. In addition, the lateral resolution of the KPFM signal is improved
when recording all data in parallel, because the tip and sample can be kept at a smaller
distance throughout the experiment. Nanosurf therefore recommends the simultaneous
measurement procedure for KPFM.
80
CHAPTER 7:
Finishing measurements
0
0
CHAPTER 7: FINISHING MEASUREMENTS
Figure 7-1: Side view of the cantilever after retracting. Minimal distance required for safe removal
of the scan head and sample.
82
STORING THE INSTRUMENT
IMPORTANT
Always store the scan head with a(n old) cantilever installed. This will prevent dust from
gathering on the alignment chip, and will protect the alignment chip against damage.
Figure 7-2: Scan Head storage. The easyScan 2 AFM Scan Head stored in its Scan Head case.
4 Pack all components in the original Nanosurf packaging material or instrument case,
as shown in Figure 7-3: Packing.
83
CHAPTER 7: FINISHING MEASUREMENTS
IMPORTANT
Before transport, always put the Scan Head in the Scan Head case. Make sure the Scan
Head is locked in tight. Put the instrument in the original Nanosurf packaging material
or instrument case.
Figure 7-3: Packing. The easyScan 2 AFM system packed in the Instrument Case.
84
CHAPTER 8:
Maintenance
0
0
CHAPTER 8: MAINTENANCE
8.1: Introduction
To ensure fault-free operation of the microscope, the maintenance instructions below have
to be observed.
CAUTION
When using a (compressed gas) air duster to clean the alignment chip on the easyScan
2 AFM Scan Head, be sure to place the DropStop and to only gently blow over the
alignment chip rather than fully onto it. A failure to do so will results in strong air
currents inside the scan head that could damage sensitive wires belonging to the scan
head electronics!
86
CHAPTER 9:
9.1: Introduction
The problems described here can occur during normal operation of the microscope. If the
suggested course of action does not solve the problem, or the problem is not described
here, refer to Section 9.4: Nanosurf support (page 93).
This error message appears when the easyScan 2 software is waiting for an answer from the
controller. Most likely, the easyScan 2 is not connected to the mains power, or it is not
turned on. In this case the status lights on the top of the controller are off. To fix this
problem:
> Check the connections and the power switch.
The USB serial converter is not available. The USB cable is not properly connected. In this
case the USB power light on the easyScan 2 controller rear panel) does not light up (Figure
1-4: The easyScan 2 controller (page 17)). To fix this problem:
1 Check if the a second copy of the easyScan 2 is already running and occupying the USB
port.
2 Check that the USB cable is properly connected.
88
SOFTWARE AND DRIVER PROBLEMS
If this does not solve the problem, check if there is a driver problem with the USB Serial
port/USB Serial converter drivers, as described in the next section.
89
CHAPTER 9: PROBLEMS AND SOLUTIONS
USB Hub
Figure 9-1: Device manager. The drivers that may be installed on your system when your controller
is connected to the computer.
90
AFM MEASUREMENT PROBLEMS
91
CHAPTER 9: PROBLEMS AND SOLUTIONS
92
NANOSURF SUPPORT
Setpoint drift
When part of the scan line in the Line graph is drawn red, the tip has moved to its maximum
or minimum Z-position. This should also be visible in the range display on the right hand
side of the graph. The tip will move to its maximum Z-position when it has lost contact with
the sample. To correct this:
1 In the Static Force based modes, increase the force Setpoint, in the Dynamic Force
based modes, reduce the amplitude Setpoint.
2 If this does not help, retract and re-approach the tip.
93
CHAPTER 9: PROBLEMS AND SOLUTIONS
9.4.2: Assistance
If the standard solutions are not sufficient, contact your local distributor for help. In order
to resolve the problem as fast as possible, please provide as much information as possible,
such as:
• A detailed description of what happened before the problem occurred. For example:
“When I click the ‘Approach’ button, then quickly click abort, the controller will not react
to anything I do anymore. This only happens when measuring in Dynamic Force Mode.”
• If an error message was displayed: The exact text of the message, or at least the start of
the message.
• The serial number of your scan head and/or controller.
• A description of the computer hardware and software on which the control software is
running: computer brand, type (laptop or desktop), operating system, software version
etc.
• Original Nanosurf image data (.nid) files that show the problem, rather than bitmap
screen shots, because these files contain all the settings that were used to make them.
• Parameter (.par) files with the instrument settings that were used when the problem
occurred.
• Script files, if the problem occurs during the operation of a script.
IMPORANT
Sending “.vbs” scripts by e-mail often does not work, because these files are usually
blocked as a security measure. To successfully e-mail a script, you may either:
• Add the script text to the body of the e-mail.
• Change the extension of the script file to “.txt” and attach it to the e-mail.
• Compress the script file to a “.zip” archive and attach it to the e-mail.
94
CHAPTER 10:
AFM theory
0
0
CHAPTER 10: AFM THEORY
Figure 10-1: Cantilever. 228 μm long micro-fabricated silicon cantilever with integrated tip.
The measurement of the cantilever deflection can be used to control the tip–surface
distance on an atomic scale. Thus, enormous resolution can be achieved, so that even the
atomic arrangement of surfaces can be probed. This measurement is a so-called static
operating mode, in which the static deflection of the cantilever is used. Generally, the
forces acting on the tip will cause it to snap onto the sample, which result in an effective,
nanometer-range flattening of the tip, and friction and stiction between the tip and the
sample.
To circumvent the aforementioned problems, the so-called Dynamic force microscopy
modes can be used, as was pointed out by the AFM inventors. In these modes, the
cantilever vibrates during the operation. In the dynamic modes, changes in the free
96
THE EASYSCAN 2 AFM
resonance frequency and the damping of the cantilever vibration caused by the forces
between the tip and the cantilever can be measured and used to regulate the tip–sample
distance.
To achieve atomic resolution, ultra-clean and flat surfaces prepared in highly sophisticated
vacuum systems are needed. Nevertheless, measurements in air can give useful results for
many technically relevant surfaces. In this manual, the use of the dynamic modes in air on
such surfaces is described.
Control
electronics
Figure 10-2: SPM system. Cantilever with deflection detection system scanning the sample. The
sample is visualized on a computer with installed scan software, which also directs the scan itself.
In dynamic operating modes, the cantilever is excited using a piezo element. This piezo is
oscillated with a fixed amplitude at an operating frequency close to the free resonance
frequency of the cantilever. The repulsive force acting on the tip will increase the resonance
frequency of the cantilever. This will cause the vibration amplitude of the cantilever to
decrease. The vibration of the cantilever is also detected using the cantilever deflection
detection system.
97
CHAPTER 10: AFM THEORY
The measured laser beam deflection or cantilever vibration amplitude can now be used as
an input for a feedback loop that keeps the tip–sample interaction constant by changing
the tip height. The output of this feedback loop thus corresponds to the local sample
height.
An image of the surface is made by scanning over the sample surface in the X and Y
direction. The sample structure image is now obtained by recording the output of the
height control loop as a function of the tip position. The direction of the X- and Y-axes of
the scanner is shown in Figure 10-3: Scanner coordinate system. The scanner axes may not
be the same as the measurement axes, when the measurement is rotated, or the
measurement plane is tilted. Therefore, the image X- and Y-axes are denoted by an asterisk
to avoid confusion (i.e. X*, Y*).
98
CHAPTER 11:
Technical data
0
0
0
CHAPTER 11: TECHNICAL DATA
11.1: Introduction
The specifications given in this chapter represent typical values of the Nanosurf easyScan
2 AFM system. The exact specifications of all components belonging to your system are
given on the calibration certificate delivered with the instrument.
100
THE EASYSCAN 2 AFM SCAN HEADS
Compatible cantilevers
Cantilevers used in conjunction with the easyScan 2 AFM should have the following
properties:
• Grooves that are compatible with the alignment chip used by Applied Nanostructures,
BudgetSensors, NanoSensors, NanoWorld, and VISTAprobes.
• A nominal length of 225 μm or more, and a width of 40 μm or more.
• A coating on the backside of the cantilever that reflects (infra)red light.
11.2.2: Dimensions
10.5
2.1
44
24
24.5
9.5
2.6
43.4
37.6
85.5
18.8
37.1
42.1
10
32.6 32.6
75.1
Figure 11-1: The Nanosurf easyScan 2 AFM Scan Head. All dimensions in mm.
101
CHAPTER 11: TECHNICAL DATA
102
THE EASYSCAN 2 CONTROLLER
Computer requirements
For a detailed description of computer hardware requirements and supported operating
systems, see Nanosurf technical note “TN00477 — Computer requirements for Nanosurf
Products”. The technical note can be downloaded from the Support section of the Nanosurf
website by all registered users.
103
CHAPTER 11: TECHNICAL DATA
104
HARDWARE MODULES AND OPTIONS
Signal Module S
Signal Module S can be used for monitoring signals from the AFM with external
instruments, such as Volt-meters, oscilloscopes or spectrum analyzers. The signals are
available on nine BNC connectors that are labelled with their signal names. The signal
names and their function are listed in the table Table 11-5: Monitor signals (page 106).
The calibration of the monitor signals can be found by looking up the signal calibration in
the Scan Head Calibration Dialog, reached via the menu “Options” >> “Config Scan Head...”.
The magnitude of the physical signal can be calculated from the Monitor Signal voltage
using the formula:
MonitorSignalVoltage
PhysicalSignal [ Unit ] = ----------------------------------------------------------- ⋅ Maximum + Offset
10 [ V ]
In addition to the monitor voltages, +15V and –15V voltage sources are available for driving
small home-made electronics. The voltages are available on a three-pole mini-DIN
connector. The pin-out of this connector is shown in Figure 11-2: Voltage source connector
(page 108).
Signal Module A
Signal Module A can be used to add functionality (custom operating modes) to the
easyScan 2 system, in addition to the applications of Signal Module S. The signals are
available on twenty-one BNC connectors that are labelled with their signal names. The
signal names and their function are listed in Table 11-5: Monitor signals and Table 11-6:
Custom operating mode signals.
105
CHAPTER 11: TECHNICAL DATA
106
HARDWARE MODULES AND OPTIONS
107
CHAPTER 11: TECHNICAL DATA
1: +15V 2: GND
3: -15V
Figure 11-2: Voltage source connector. Connector as seen from outside.
Various other User Input and Output settings can be configured in the Signal Access page
of the SPM Parameter dialog. Refer to Section 17.9.7: Signal Access (page 235) for details on
how to change them.
108
CHAPTER 12:
1 2 3
The main SPM Control Software window (also referred to as workspace) consists of five
major areas:
1. The Measurement pane on the left. This area contains the so-called Operating
windows, which are used to acquire and display ongoing measurement data.
2. The Document space in the middle. This area is used for displaying and analyzing
previously stored measurement documents.
3. The Info pane on the right. This area contains several stacked Panels and is used to
group a diverse array of functionality and information.
4. The Ribbon at the top. This area is used to access all action functions.
5. The Status bar at the bottom. This area is used to display additional information.
110
THE WORKSPACE
In “Normal” mode, the emphasis lies on the Measurement and Info panes. The inside
border of the two panes can be dragged by the mouse to adjust their individual widths to
your needs. Document space on the other hand is rather limited (see Figure 12-1: The main
window in “Normal” workspace mode). This mode is most suited for measurements.
In “Document” mode, the Document space is maximized while the Measurement and Info
panes are minimized to the left and right side of the main window, respectively (see Figure
12-2: The main window in “Document” workspace mode). The various window and panel
titles are shown in tabs so that you can still open them when needed. A click or a mouse-
over on one of these tabs will cause the respective window or panel to slide out
111
CHAPTER 12: THE SPM CONTROL SOFTWARE USER INTERFACE
automatically, so that you can work on it. It will automatically minimize again when you are
done. This mode is most suited for analyzing stored measurement data.
Figure 12-3: Elements of Operating windows. Shown here is the Imaging window, with the
Parameter area (Imaging panel) on the left, the Chart area on the right, and the Chart bar on top.
Clicking the tabs at the bottom of the Measurement pane switches between the Operating windows.
112
DOCUMENT SPACE
113
CHAPTER 12: THE SPM CONTROL SOFTWARE USER INTERFACE
By default, all measurements are temporarily stored (automatically) during imaging and
spectroscopy. They can be opened at all times from the Gallery panel (see Section 16.4:
Gallery panel (page 181)), but should be moved to a new folder for permanent storage as
soon as you have finished measuring (see Save as (page 182)).
Everything related to documents is described in more detail in Chapter 16: Working with
documents (page 169).
12.5: Panels
In the panels of the Info pane, the control software provides additional information that
can be useful to the user. These panels are normally docked to the Info pane and are
stacked to save space. The panels have several features, however, that allow you to arrange
them in a way that is most efficient for your application (see Figure 12-5: Arranging panels).
Figure 12-5: Arranging panels. (Left) Stacked panels. (Center) Separated panel that is docked to a
stack. (Right) Panels that are minimized in “Document” mode.
To separate a panel and dock it individually to the side of/below another panel that is
already docked to this window, drag its title bar to the desired position using the mouse
cursor.
To add a control panel to a stack, drag either its title bar or its label to either the title bar or
labels of the stack. To remove a panel from a stack, drag its label away from the stack.
When panels are stacked, their title labels are displayed on the bottom of the Info pane. To
move a control panel to the top of the stack, click its tab.
With the “pin” button ( ) in the tile bar of the individual panel or the Info pane, the auto
hide feature is controlled. If “unpinned”, the panel or the Info pane minimizes to the right
114
PANELS
border of the main window and only the panel titles are visible (similar to the “Documents”
workspace mode, but now only for the panel/Info pane and not for the Measurement area).
A mouse hover over (or click on) a title tab will slide this panel into view.
It is possible to scroll the content of a control panel up and down, when it is too small to
display all the parameters it contains. To do this, move the mouse cursor over an area where
it changes to a four pointed arrow. Then, drag the content up and down with the mouse.
Tool
The Tool panel contains the results of the various analysis tools available to you during and
after measurement, displays the current mouse position during selections, and displays the
size of those selections (e.g. during zooming). The Tool panel is described in more detail in
Section 16.6: Tool panel (page 198).
Gallery
The Gallery panel displays a list of stored measurements for quick opening (viewing and
analysis). A File Browser is also integrated for general file management tasks. The Gallery
panel is described in more detail in Section 16.4: Gallery panel (page 181).
Help
The Help Panel provides quick access to PDF versions of the user manuals belonging to
your system, to relevant application notes and technical notes, and to online sources of
information (direct links to the Nanosurf website).
115
CHAPTER 12: THE SPM CONTROL SOFTWARE USER INTERFACE
Online
The Online Panel provides you with an overview of the current scan range within the
maximum range the scanner is capable of (Scan Position section), a “Master image” that
can be used as a reference for multiple zoomed scans on different points of interest (Master
Image section) and a simple illumination control (Illumination section). The Online panel is
described in more detail in Section 13.7: Online panel (page 142).
Video
The Video panel with its top and side view of cantilever and sample is particularly useful
during sample positioning and approach (see also Section 4.4: Approaching the sample
(page 44)). The elements and usage of the Video panel is described in Section 13.6: Video
panel (page 135).
Stage
The Stage panel can be used to control the automated translation stages (ATS A100 or ATS
C301), but is only available when the Translation Stage software is installed. The Stage
panel is explained in Section 13.5: Stage panel (page 133).
12.6: Ribbon
The Ribbon provides access to all major actions and commands by grouping them
according to their usage.
116
STATUS BAR
tab displays different buttons for each of the Measurement windows (Imaging,
Spectroscopy and Lithography) and is therefore described in their respective Chapters.
1 2 3 4 5 6 7
Figure 12-7: The Status bar. Numbers in this figure correspond to those in the list below.
The status bar displays relevant microscope information and the loaded settings (see Figure
12-7: The Status bar). It contains the following elements:
1. Help text information about the current menu button or latest error messages
2. Status signal of the Z-feedback controller (red: in upper limit positions; orange: in
lower limit position; green: tip in normal feedback contact with sample).
3. Software status: “Online” or “Simulation” (depends on the presence or absence of a
scan head, or on user choice).
4. Currently loaded file (“.chart”) used for chart settings.
5. Currently loaded file (“.par”) used for parameter settings.
6. Currently loaded scan head calibration file (“.hed”).
7. Buttons to access the Workspace “Normal” and “Document” view.
117
CHAPTER 12: THE SPM CONTROL SOFTWARE USER INTERFACE
The workspace group in the view tab gives you the ability to switch between the two
workspace modes “Normal” and “Documents”. To switch, click on either of the buttons, or
on the corresponding smaller buttons ( ) on the right-hand side of the status bar (see
also Section 12.7: Status bar). With these you don't even need to switch to the view tab while
measuring.
Normal
The optimal workspace choice during measurements.
Documents
The best choice for viewing or analysis of stored documents (see Chapter 16: Working with
documents (page 169)).
The buttons of the Panels group have the same function as the tabs at the bottom of the
Info pane (i.e., to bring the respective panel to the top of Info pane). If the panel was
undocked from the Info pane (see Section 12.5: Panels) and subsequently closed (i.e., no
longer visible as panel or tab), it will re-appear by pressing its button in the Panels group.
118
VIEW TAB
IMPORTANT
The “Video” and “Stage” buttons are only available when the appropriate hardware is
present.
The window group provides you with tools to arrange open measurement documents in
different ways or quickly close them all.
Cascade
Open document windows are stacked on top of each other and slightly offset with respect
to each other so that individual windows can be easily accessed. Width and position of the
windows is optimized by the control software.
Tile H.
Tiles the open document windows horizontally, so that individual measurements can be
easily compared. Width of the document windows is maximized. Height is evenly
distributed over the available document space.
Tile V.
Tiles the open document windows vertically, so that individual measurements can be easily
compared. Height of the document windows is maximized. Width is evenly distributed
over the available document space.
Close all
Closes all open document windows. If unsaved data exists, you will be asked to save it.
119
CHAPTER 12: THE SPM CONTROL SOFTWARE USER INTERFACE
120
CHAPTER 13:
Imaging
0
0
CHAPTER 13: IMAGING
13.1: Introduction
Imaging measurements of the sample are controlled using the Imaging window. This
chapter describes all elements of the Imaging window in detail. For procedures describing
a basic measurements, refer to Chapter 4: A first measurement (page 41). For details on how
to use the charts see Section 16.3: Charts (page 171). For advanced imaging settings see
(Section 17.9.1: Imaging (page 222).
The Imaging window is contained within the Measurement pane and can be opened by
clicking the Imaging tab. The largest part of the Imaging window consists of a number of
charts that display the data from the ongoing measurement: the Chart area. The imaging
window can display as many charts as required. Scroll bars will appear automatically as
soon as the content is larger than the window can accommodate. By default, two charts
groups are displayed: 2 color maps of the sample and their corresponding line graphs.
Usually, these show Topography on the left and another measurement signal on the right
(e.g. Deflection), depending on the current operating mode. For more information on
adding and changing charts, which basically works the same for charts in Operating
windows and in stored measurements documents, see Chapter 16: Working with documents
(page 169).
The Parameter area on the left side of the Imaging window, the so-called Imaging panel, is
organized in 3 sections: the “Parameters”, “Z-Controller” and “Mode Properties” section.
These sections are an integral part of the Imaging window and represent the most
commonly used parameters for the currently selected operating mode. Possible
parameters in these sections are described in Section 13.2: Imaging panel. Advanced
parameters can be accessed via the “More” button in each section, which will open the SPM
Parameter dialog on the respective page (see Section 17.9: SPM Parameters dialog (page
221)).
At the top, the Imaging window contains a toolbar with commands to control the imaging
process: the Imaging toolbar. It is described in Section 13.3: The Imaging toolbar.
Apart from the necessary settings, several actions have to be performed before being able
to image a sample. These are accessed via the Acquisition tab of the Ribbon, the elements
of which are described in detail in Section 13.4: Acquisition tab.
Before and during imaging, several panels of the Info pane provide additional information
for your reference. The relevant panels are explained in Section 13.6: Video panel and Section
13.7: Online panel. A description of the other panels is found elsewhere in this manual.
Please refer to the Quick reference (page 300) to locate them.
122
IMAGING PANEL
123
CHAPTER 13: IMAGING
'Rotation'
angle
Measurement plane
'Y-Slope'
angle Y* X*
Image area
'X-Slope' angle
Y
X Scanner XY-plane
Z, Z*
Figure 13-2: Coordinate systems
Image size
Defines the image size in both the X* and Y* direction. The size is doubled or halved when
the arrows next to the edit box are used.
Time / Line
The time needed to acquire a single data line. The time needed for the entire image is
displayed in the status bar.
Points / Line
The number of measured data points per line. It also sets the number of lines to the same
value.
Rotation
The angle between the X-direction of the scanner and the X* direction of the measurement
(Figure 15-2: Coordinate systems).
“More” button
Opens up the SPM Parameter dialog on the “Imaging” page for more advanced parameters
(see Section 17.9.1: Imaging (page 222)).
124
IMAGING PANEL
Z-Controller section
During imaging, the tip–sample interaction is kept constant through the Z-Controller. The
Z-Controller is a standard PID controller as is shown in Figure 13-3: Z-Controller.
P-Gain
I-Gain
Error
Control Signal Error
Range +
(Tip Current, +
Signal +
Deflection, dt Topography
Amplitude, ...) - 1..16x
+
D-Gain
Set point d
dt
Setpoint
The working point for the Z-Controller. Depending on the operating mode, this is the
tunneling current (STM mode), cantilever deflection (static force mode) or relative
cantilever vibration amplitude (dynamic force mode). In the later case, the set amplitude is
relative to the operating amplitude, set in the Mode Properties Section. For example, when
a Setpoint of 70% is used, the Z-Controller will move the tip closer to the sample until the
vibration amplitude has decreased to 70% of the vibration amplitude far away from the
sample.
P-Gain
The strength of the Z-Controller reaction that is proportional to the error signal. Increasing
the P-Gain decreases the error signal.
125
CHAPTER 13: IMAGING
I-Gain
The strength of the Z-Controller reaction that is proportional to the integral of the error
signal. Increasing the I-Gain decreases the error signal over time. It is the least sensitive to
noise, and usually the dominant contributor to the topography measurement.
D-Gain
The strength of the Z-Controller reaction that is proportional to the derivative of the error
signal. Increasing the D-Gain decreases fast changes in the error signal, but also amplifies
high frequency noise.
“More” button
Opens the SPM Parameter dialog on the “Z-controller” page for more advanced parameters
(see Section 17.9.6: Z-Controller (page 233)).
Tip Voltage
This parameter defines the potential to be applied to the tip. The voltage that can be used
lies between -10V and +10V.
IMPORTANT
With the STM scan head the sample is automatically connected to the ground of the
instrument. With AFM scan heads the sample has to be electrically connected to the
instrument’s chassis ground for accurate measurements.
126
THE IMAGING TOOLBAR
127
CHAPTER 13: IMAGING
Figure 13-4: Zooming. (Left) The Zoom tool area marker. (Right) The Zoom tool information in the
Tool panel of the Info pane.
The change in position is indicated by an arrow. The start of the arrow is defined by the
mouse cursor position where the left mouse button is pressed; the end of the arrow by the
position where the button is released. With a single click of the left mouse button an arrow
ending in the center of the measurement is drawn. The direction of the arrow can be
adjusted by dragging its end markers. It can be moved by dragging the center marker.
The image is moved by double clicking, or clicking the “Move” button in the Tool Results
panel. To abort the Move function, click the Move-Button again or click the right mouse
button and select “Abort” in the context menu.
Full
The Full button returns the parameters Scan range to the largest possible values, and “X-
Offset” and “Y-Offset” to zero (see Section 17.9.1: Imaging (page 222)).
Capture
With the “Capture” button, you can immediately copy the current measurement to the
History page of the Gallery panel without waiting for the scan to be completed (see also
Section 16.4: Gallery panel (page 181)). It is stored as a new document and remains open in
the Document space of the SPM Control Software.
128
ACQUISITION TAB
Auto set
The “Auto set” button opens the Parameter preparation wizard. This wizard sets the
measurement parameters to reasonable values, based on the sample feature size and
desired imaging quality that you enter:
More
Additional preparation wizards to help setting up certain operating modes are found here.
Currently, only the KPFM Wizard is available:
129
CHAPTER 13: IMAGING
Environment
This drop-down box allows the selection of either “Air” or “Liquid” as measurement
environment. Depending on mounted scan head, certain options may or may not be
available.
Operating Mode
This drop-down box allows the selection of the Operating mode used for measuring (see
Chapter 6: Operating modes (page 63)). Depending on the installed controller modules and
the mounted scan head, certain options may or may not be available.
Cantilever Selector
This drop-down box allows the selection of the mounted cantilever. For more details see
Section 3.3.1: Selecting a cantilever (page 30). Depending on your scan head, certain options
may or may not be available. The entry “Cantilever Browser…” opens the Cantilever
Browser dialog, which allows you to edit existing or define new cantilever types (see Section
17.13: Cantilever Browser dialog (page 245))
Freq. Sweep
This button opens the Vibration Frequency Search dialog. In this dialog, amplitude versus
frequency plots can be measured and the operating frequency for dynamic modes can be
set manually. In “Air” environment, setting of the vibration frequency is normally
performed fully automatically, without user intervention. In “Liquid” environment, the
Vibration Frequency Search dialog opens after the control software determined a suitable
vibration frequency. You need to fine adjust and confirm the selected resonance frequency
prior to approach. For more details about this dialog see Section 17.11: Vibration Frequency
Search dialog (page 239).
Laser Align
This button opens the Laser Align dialog. In this dialog, laser power and the current
position of the laser beam on the detector are monitored. It is used to check for good
detector and laser alignment. With FlexAFM scan heads it guides the user to manually
adjust these positions if required or desired (see FlexAFM Laser and detector alignment).
Launcher icon
More advanced settings are available through the “Dialog Launcher” icon ( at the bottom
right corner of the Preparation group), which opens up the SPM Parameter dialog on the
Operating Mode page (see Section 17.9.4: Operating Mode (page 229)).
130
ACQUISITION TAB
Home
Increases the tip–sample distance to its maximum value to ensure that the maximum
motorized approach range is available during final automatic approach.
Retract
Increases the tip–sample distance at maximum speed until the button is released.
Advance
Decreases the tip–sample distance at maximum speed until the button is released.
Approach
Starts the automatic approach. During automatic approach, the tip–sample distance is
decreased until the Setpoint (set in the Z-Controller section) is reached, or until the
maximum number of approach steps is reached (see Section 17.9.5: Approach (page 231)).
Withdraw
Increases the tip–sample distance with approach speed settings.
Launcher icon
More advanced settings are available through the “Dialog Launcher” icon ( at the bottom
right corner of the Preparation group), which opens up the SPM Parameter dialog on the
Approach page (see Section 17.9.5: Approach (page 231)).
Start
Clicking “Start” starts a measurement and changes the button to “Stop”. Clicking “Stop”
aborts the measurement as soon as the current scan line is finished.
Finish
Selecting “Finish” will set the “Finish” flag, which will not abort the measurement directly,
but will do so when the measurement is finished. Deselecting (i.e., clicking it again) will
disable the “Finish” flag so that the measurement will no longer stop automatically when it
is finished. The “Finish” button is highlighted when it is flagged.
Up / Down
Starts a single measurement or restarts an ongoing measurement from the selected
scanning direction. With the “Up” button the image is scanned from bottom to top. With
the “Down” button it is scanned from top to bottom.
131
CHAPTER 13: IMAGING
Launcher icon
More advanced settings are available through the “Dialog Launcher” icon ( at the bottom
right corner of the Preparation group), which opens up the SPM Parameter dialog on the
Imaging page (see Section 17.9.1: Imaging (page 222)).
The Scripting Interface is an optional component for creating user defined scripts (software
components) to add new features or automating tasks. For details, see “Help Panel” >>
“Manuals” >> “Script Programmer’s Manual”.
Important:
The Scripting Interface is a purchase option and has to be activated using the Access
Codes page (see Access Code (page 212) of the Options dialog).
132
STAGE PANEL
This panel is only available to Nanosurf AFM systems when the Translation Stage Software
is installed on the computer that contains the SPM Control Software (usually with Nanite
systems). The Stage panel allows moving of the translation stage in the same way as the
corresponding panel in the Batch manager is operated.
The upper part of the Stage Position section displays the current position of the translation
stage in coordinates relative to the zero point of the translation stage, or in coordinates
relative to a previously set zero point. On start-up, the coordinates are shown relative to the
start up position, until the reference position of the stage has been found (e.g. after clicking
the “Search” button).
133
CHAPTER 13: IMAGING
Absolute
Clicking the “Absolute” button ( ) (re)sets the zero point of this axis’ coordinate display to
the absolute zero point of the translation stage. The Stage Position coordinates (X,Y,Z) now
display the absolute position of the stage.
Relative
Clicking the “Relative” button ( ) sets the zero point of this axis’ coordinate display to the
current position. The Stage Position coordinates (X,Y,Z) now display the position of the
stage relative to the position where the “Relative” button was last clicked.
Search
Clicking the “Search” button ( ) starts a search for the reference position of the
translation stage. The reference position can be different from the zero point of the stage.
Move To
Clicking the “Move To” button ( ) opens the Move Stage To dialog (see Section
13.5.1: Move Stage To dialog).
Arrow buttons
Move the translation stage in the direction of the arrow for as long as the button is clicked
and held.
Speed
The speed with which the translation stage moves.
CAUTION
It is recommended to always keep the approach stage in the upper limit position. When
repeatedly moving to the same position, it is possible that the tip will crash into the
sample if the approach stage is in a lower position than it was before.
134
VIDEO PANEL
This dialog is only available to Nanosurf Nanite AFM systems with an automated translation
stage attached and the corresponding software installed. It is used to move the translation
stage to a specific destination position. It is opened by clicking the “Move To” button in the
Manual Move section of the Stage panel (see Manual Move section (page 134)).
X/Y/Z
The (X, Y, Z) coordinate of the destination position.
Move relative to current position
When checked, the (X,Y,Z) coordinates are interpreted relative to the current position.
When not checked, the coordinates are interpreted as a position relative to the absolute
zero point of the translation stage. This checkbox is only enabled when the reference
position has been found (e.g. after clicking the “Search” button in the Stage Position
section of the Stage panel).
Move
Clicking the “Move” button ( ) moves the translation stage to the destination
position entered by the user (see above).
INFO
If your system is not equipped with a Video Camera or the SPM Controller is not
equipped with a Video Module, this Panel will not be available.
135
CHAPTER 13: IMAGING
Figure 13-5: Analog video signal in the Video panel. (Left) Top view. (Right) Side view.
Show Window
The “Show Window” button ( ) allows you to switch between display of the video signal
in a section of the Video panel and display of the video signal in a separate window (see
Figure 13-6: Free floating video window).
136
VIDEO PANEL
Switch view
The “Switch View” button ( ) allows you to switch between the two analog video
cameras. Either the Top view signal or the Side view signal is displayed.
Brightness
The “Brightness” slider ( ) regulates the brightness of the video display.
Contrast
The “Contrast” slider ( ) regulates the contrast of the video display.
Save As...
The “Save As...” button ( ) allows you to save the current video image as a JPG file.
Copy
The “Copy” button ( ) allows you to copies the current video image to the Windows
Clipboard for “pasting” into other applications.
Figure 13-7: Digital video signal in the Video panel. (Left) Top view. (Right) Side view.
137
CHAPTER 13: IMAGING
In addition to the options available for analog cameras (see Analog Video Camera toolbar),
the following options exist:
Zoom + / Zoom –
The “Zoom” buttons ( ) allow you to zoom in or out digitally (by binning of pixels). The
zoom area is always in the center of the image. The number of pixels displayed is kept
constant. Therefore, the video rate is independent of the zoom factor. The current zoom
factor is displayed in an overlay at the bottom of the video image.
Focus up / Focus down
The “Focus” buttons ( ) adjust the focus of the current camera view. Motors in the
respective camera physically adjust lenses to change the focus plane.
Gain slider
The “Gain” slider ( ) regulates the amount of gain applied to the video signal.
Anti-Moiré Button
The “Anti-Moiré” button ensures that the ratio between the number of camera pixels and
video display pixels is always a whole number. This prevents moiré patterns from occurring
in the video image. The moiré effect is mostly visible on samples with regular structures like
grids or lines if this option is not enabled.
When the Anti-Moiré mode is active, the button is highlighted. Clicking the highlighted
button will deactivate the Anti-Moiré mode and will cancel the button’s highlighting.
138
VIDEO PANEL
The “High-Resolution Mode” button If this mode is activated, the video display shows all
camera pixels on the display one-to-one. No Zoom is available in this mode. If the video
display’s pixel size is lower than the camera’s, scroll bars are automatically shown on the
borders of the video display. This mode is only useful in a large separate window (e.g. on a
second LCD monitor) when you need to see a sample in full detail.
When the High Resolution mode is active, the button is highlighted. Clicking the
highlighted button will deactivate the High-Resolution mode and will cancel the button’s
highlighting.
IMPORTANT
In high-resolution mode the amount of (video pixel) data that needs to be transferred
over the USB connection is huge. As a result, the video frame rate typically drops below
5 Hz and a high-performance PC is required to keep the system stable.
Pause Button
If the “Pause” button ( ) is activated the video image update is stopped and the frozen
image (last frame received) is shown continuously. A second click on this button restarts
the real time video display.
Video Properties Button
A click on this button opens the Digital Video Properties Dialog. See the next section
(Section 13.6.4: Digital Video Properties dialog) for details.
139
CHAPTER 13: IMAGING
The Digital Video Properties dialog can be accessed via the Video Display toolbar or the
video display’s context menu. The properties accessible through this dialog directly
influence the respective camera's behavior at the camera driver level.
140
VIDEO PANEL
Image
Exposure time
Adjusts the exposure time (time to record one image frame) for the respective video
camera. Changing this setting allows the camera to cope with very bright or very dark
sample illumination conditions, or to fine-tune the predefined illumination condition
ranges (see Range below).
Range
Allows the selection of different illumination ranges (levels) for quick adjustment of the
video camera to the current sample illumination conditions. Lower level numbers are
suitable for lower illumination conditions.
Gain
Identical to the “Gain” slider in Digital Video Camera toolbar (page 137), which adjusts the
amount of gain applied to the video signal.
Auto Levels
When checked, this option automatically adjusts the camera data to fill the dynamic range
of the Video display (similar to the “Auto set” option for the Chart data range (see Section
16.3.2: Chart Properties dialog (page 174)). Using this option will automatically give you a
good quality image. You will find that changing the other image parameters while this
option is checked (within limits) hardly has an effect on the video display anymore.
Color adjustment
Red ratio
Adjusts the relative amount of red color information in the RGB color mix for each pixel. Not
available for the (monochrome) side view.
Green ratio
Adjusts the relative amount of green color information in the RGB color mix for each pixel.
Not available for the (monochrome) side view.
Blue ratio
Adjusts the relative amount of blue color information in the RGB color mix for each pixel.
Not available for the (monochrome) side view.
Gamma
Adjusts the midpoint of the video image’s dynamic range. Adjusting this setting can be
beneficial for images that show too low or too high contrast. In such cases, increase or
decrease the gamma setting, respectively.
141
CHAPTER 13: IMAGING
The Scan Position section displays information about the maximum scan head scan area,
the current measurement area and the tip position.
The left side of the section shows the maximum scan range of the scan head (outer square
and numbers) and the currently set image size (inner square). The dotted bars to the left
142
ONLINE PANEL
and right of the inner square represent the “Overscan” area (see Overscan in Section 17.9.1:
Imaging (page 222)).
The red line in the inner square represents the currently measured scan line during
imaging. The two arrows represent the orientation of the axes of the image coordinate
system. The small dot in the center of the image square represents the Offset and Rotation
point (see Image offset X/Y and Rotation in Section 17.9.1: Imaging (page 222)).
The dotted bar on the right side of the Scan Position section represents the maximum Z-
range (bar and numbers), the used range for the current measurement (gray box) and the
actual Z-position (average of the current scan line) of the tip (red line).
The Master Image section displays a topography measurement, which can be used as a
reference for comparison with other measurements, or as an overview image for multiple
(zoomed) measurements on several points of interest. The Master Image section starts out
blank. A reference image has to be loaded manually. When this is done, a box with a black
outline will show the current measurement area inside the reference/overview image.
Load
The Load button captures the actual topography image of the active Measurement Tab
into the Master Image Section.
Zoom
A click to this tool activates the Zoom Tool in the Master Image chart area to select a new
scan area size and position (within the loaded image). After selecting the Zoom button the
user draws a zoom frame in the chart area. To select the new scan area, double-click within
the selection area. A right mouse click aborts the zoom operation.
143
CHAPTER 13: IMAGING
With the Illumination slider the intensity of the sample illumination LED is controlled. This
slider is only present in the Online panel when no video camera is present in your system.
When a camera is present, this slider automatically moves to the Video panel (see Section
13.6.3: Illumination section).
144
CHAPTER 14:
Spectroscopy
0
0
CHAPTER 14: SPECTROSCOPY
14.1: Introduction
146
INTRODUCTION
IMPORTANT
AFM users must electrically connect the sample to the ground connector on the scan
head to apply a tip–sample voltage difference.
147
CHAPTER 14: SPECTROSCOPY
Modulated output
This parameter defines the signal used to drive the spectroscopy (X-Axis). All possible
signals are recorded while this modulation output signal is changing its value from “Start
value” to “End value” (Y-Axis). The number of available modulated outputs depends on the
scan head and the number of installed modules. Possible values are: “Z-Axis”, “Tip Potential”
and the names of the User Outputs.
Start value / End value
The range over which the Modulated output is changed. The “Spec Forward” data is
measured from the Start to the End value, the “Spec backward” data is measured in the
opposite direction. The “Spec forward” data is always measured before the “Spec
backward” data. For spectroscopy as a function of distance (Z-axis modulation), more
negative values are further away from the sample whereas more positive values go towards
(or even into) the sample.
Modulation time
The time used to change the Modulated output from the Start to the End value.
Data points
The number of data points measured while the Modulation output changes from Start to
End value. The data points are equally distributed over the modulation range.
Averages
The number of times the modulation is repeated to obtain an averaged spectroscopic
measurement. The measurement results of aborted modulations are discarded during
averaging.
“More” button
Opens up the SPM Parameter dialog on the “Spectroscopy” page for more advanced
parameter settings (see Section 17.9.2: Spectroscopy (page 225)).
148
SPECTROSCOPY TOOLBAR
Position section
Point
Activates the single point spectroscopy mode. It defines the position of the spectroscopy
measurement by mouse. Click in the topography Color Map chart at the position where the
spectroscopy measurement should take place. A small black dot appears at this position.
The positions coordinate is transferred to the “X/Y-Pos From” parameters in the Position
section. The value in “Sequence” defines the number of spectroscopy measurements
measured at these point.
Line
Activates the line spectroscopy mode. Defines the start and end position of the
spectroscopy measurement by mouse. Click and hold the left mouse button in the
topography Color Map chart at the position where the spectroscopy measurement should
start. Move the mouse to the end position and release the left mouse button. A line is
149
CHAPTER 14: SPECTROSCOPY
drawn between these two points. The positions coordinate is transferred to the “X/Y-Pos
From” and X/Y-Pos To” parameters in the Position section. The value in “Sequence” defines
the number of spectroscopy measurements measured along these line.
Load
Fills the Topography Color Map chart in the Spectroscopy window with the current
measurement of the Imaging window. Selection of “point” or “line” measurement position
can be done in this chart.
Capture
A click on “Capture” saves the current spectroscopy measurement data to the History page
of the Gallery panel, even when the measurement(s) have not been completed yet. The
spectroscopy data are stored as a new document and remain open in the Document space
of the SPM software.
Start
Clicking “Start” starts a spectroscopy measurement sequence and changes the button to
“Stop” until the measurement sequence is finished. Clicking “Stop” aborts the
measurement sequence as soon as the current modulation period is finished.
Launcher icon
More advanced settings are available through the “Dialog Launcher” icon ( at the bottom
right corner of the Spectroscopy group), which opens up the SPM Parameter dialog on the
Spectroscopy page (see Section 17.9.2: Spectroscopy (page 225).
150
CHAPTER 15:
Lithography
CHAPTER 15: LITHOGRAPHY
15.1: Introduction
152
PERFORMING LITHOGRAPHY
By default, the Lithography window also contains the Lithography Preview display (see
Section 15.6: Lithography preview (page 168)) and a Color map chart of the current
Topography measurement. The Lithography window can however display more charts,
should this be desirable. For more information on adding and changing charts, see Section
16.3.1: Working with multiple charts (page 173).
IMPORTANT
Lithography of objects drawn by hand and direct manipulation of the tip is available as
standard. Issuing of lithography commands through the Scripting Interface requires a
license for the Scripting Interface. Import of vector or pixel graphics files to be used as
patterns in the lithography process requires the licensed Lithography Option. For
information on how to activate the scripting interface or the Lithography Option, refer
to Access Code (page 212).
153
CHAPTER 15: LITHOGRAPHY
vector or pixel graphics file, the pattern is referred to as a “Lithography object” in the
Lithography window.
IMPORTANT
• In case of vector-based objects, multiple lithography objects may be present (e.g.
through sequential import) and used for lithography. In case of pixel-based objects,
only one pixel-based object can be present at any given time (other objects will be
deleted upon import).
• A separate CAD program called “LayoutEditor” is included on the installation CD to
create suitable GDS vector graphics (Newest version of LayoutEditor; http://
layout.sourceforge.net). Pixel graphics files can be created or edited in any pixel-
based image editor like the Windows Paint.
• As an alternative to designing the lithography pattern in a vector or pixel graphics file
and then importing it into the lithography software, a freehand drawing mode is
available in the Lithography window.
4. The imported object is positioned and scaled to fit the target area.
5. The Lithography sequence is executed.
6. The sample surface is re-imaged to view the Lithography results.
Tip
As alternatives to step 3–5, you can also use the Direct Tip Manipulation or Free Drawing
modes.
154
LITHOGRAPHY PANEL
– Dynamic Force
– STM (STM only)
Inactive pen mode
Action to be performed when the tip is moving from one end point to a new start point, in
case the end point and start point are not the same. The following options are available:
– Lift up tip
Only lift the tip (upper position of the Z-actuator of the scan head). No feedback will be
performed by the Z-Controller during travel to the new start point.
– Standard operating mode
Switch the Z-Controller operating mode back to the one selected in the “Operating
Mode Panel” during imaging. All values such as Tip speed, Tip voltage, Setpoint etc. will
temporarily chance back to the values used for imaging. The Z-Controller will be active
during travel to the new start point.
“More” button
Opens up the SPM Parameters dialog on the “Lithography” page (see 1Section 17.9.3:
Lithography (page 228)).
Layer list
Lists all layers that are present in the objects shown in the Lithography Objects list. Layer 0
is always present, even if no lithography objects exists, and may be used to set the default
Lithography parameter values (see Parameters (page 166) in Section 15.5.2: Pixel Graphic
Import dialog for details).
Edit
Edit will open the Layer Editor dialog to edit the selected layer.
Copy
Copy will open the Layer Editor dialog (see Section 15.3.1: Layer Editor dialog) to edit the
selected layer before copying it. When changes have been made (if any) and the “OK”
button is clicked, a new layer is generated.
155
CHAPTER 15: LITHOGRAPHY
IMPORTANT
Upon creation of a new layer, the layer number will be incremented to the next available
layer number. If a total of 256 layers is reached, no more layers can be added.
Delete
Used to delete a layer. Delete will open a warning dialog to confirm the deletion of the
selected layer.
IMPORTANT
Only layers currently not assigned to any object can be deleted.
Load
Load a predefined layer list “.lld”. Layers that are needed to display the current objects that
are not part of the loaded list will be created.
Save
Save all the layers to a layer list file “.lld”.
Contains a list of all available Lithography objects. Objects may be selected or deselected
by checking or unchecking the checkbox. If the object is unchecked it will not be used for
a lithography session.
Edit
Edit will open the Object Editor dialog to edit the selected object (see Section 15.3.2: Object
Editor dialog (page 159)).
Copy
Copy will open the Object Editor dialog to edit the selected object before copying it.
Delete
Delete will delete the selected object. A warning dialog will appear for confirmation of this
action.
156
LITHOGRAPHY PANEL
The Layer Editor sets the controller parameter values to be used during lithography.
Layer
Layer
Displays the selected layer’s number.
Name
Displays and allows editing of the selected layer’s name.
Color
Allows selection of the layer color for display of the layer elements in the Topography chart
of the Lithography window.
Common parameters
Tip speed
Determines the drawing speed during lithography,
157
CHAPTER 15: LITHOGRAPHY
Tip voltage
Determines the voltage set to the tip during oxidative Lithography.
STM parameters
Setpoint
Used to set the tunnelling current Setpoint of the Z-Controller during STM Lithography.
158
LITHOGRAPHY PANEL
Preview
Graphical area that provides a preview of the selected Lithography object. The red cross (if
visible) indicates the origin position of the object.
Object
Name
The name that is used to describe the object. Default names are generated during import,
based on the GDS II object names, or on the pixel graphic filename, but may be edited here
afterwards.
159
CHAPTER 15: LITHOGRAPHY
Size
Width/Height
Displays the width and the height of the object.
Scale factor
The factor by which the object can be scaled. If the scale factor is changed, the width and
the height will be automatically recalculated. Scale factor 1 represents the original size.
Position
X-Pos/Y-Pos
The X-Pos and the Y-Pos may be used to move the object within the space of the
topography map.
Move to Center
Moves the origin on the selected object back to the center of the topography map.
Start
Starts the lithography sequence and changes to “Stop” until the lithography sequence is
finished. Clicking “Stop” aborts the sequence.
Rescan
Starts a single image measurement and changes to “Stop” until a full image has been
scanned. The image is scanned from the bottom to top. Clicking “Stop” aborts the
measurement.
IMPORTANT
When performing imaging from within the Lithography window, be sure to set valid
imaging parameters in the Imaging and Z-Controller sections of the Imaging panel.
160
LITHOGRAPHY TOOLBAR
Launcher icon
The Lithography parameters can also be accessed through the “Dialog Launcher” icon (
at the bottom right corner of the Lithography group), which opens up the SPM Parameter
dialog on the Lithography page (see Section 17.9.2: Spectroscopy (page 225).
Manipulate
Starts the direct tip manipulation mode. It is now possible to control the movement of the
tip by moving the mouse around the topography color map chart. When the left mouse
button is held down, Lithography will be performed with the lithography operating mode
set in Lithography panel, and with the parameters set in Layer 0 (the Tip speed setting is
ignored). When the left mouse button is released, the tip will go to the inactive pen mode
set in the Lithography panel, and will not move until the left mouse button is pressed again.
Dragging the mouse slowly will produce smoother lines than dragging it fast.
Draw
Starts the free hand drawing mode. A shape can now be drawn in the topography color
map chart by clicking and holding the left mouse button. A shape can only consist of a
single line. Repeating the above will erase the previous drawing. Double clicking the
drawing will save it to the Lithography Object list. The drawn shape can be executed by a
click on “Start” in the Acquisition tab.
Import vector
Opens an “Open File” dialog to import a GDS II vector graphic file (extension “.gds”). Other
formats (DXF, OAS, OASIS, CIF) can be converted to GDS II using the external program
LayoutEditor (provided on installation CD).
IMPORTANT
• Since the Lithography software only supports a subset of the GDS II file format, an
error message will appear when a file containing non-supported elements is loaded.
• To avoid most load error messages, the vector graphics project should be fully
flattened before saving it as a GDS II file. LayoutEditor and most other CAD programs
provide some form of flattening functionality. Refer to the manual or (online) help of
your CAD program for details.
For more information on the available import options after selecting a valid GDS II file, refer
to Section 15.5.1: Vector Graphic Import dialog.
161
CHAPTER 15: LITHOGRAPHY
Import bitmap
Opens a “Open File” dialog for importing a BMP, DIB, GIF, TIFF, PNG, or JPEG pixel graphics
file.
IMPORTANT
The Lithography software supports files with 256 pixels or less in width and height.
For more information on the available import options after selecting a valid pixel graphics
file, refer to Section 15.5.2: Pixel Graphic Import dialog.
Load
Loads the Topography image from the Imaging window into the Lithography Topography
chart.
Capture
This button captures the measurement currently displayed in the “Lithography window” to
the History page of the Gallery panel. If clicked during a measurement, a copy is generated
as soon as the measurement in progress is finished. The capture process is cancelled by
clicking second time. The captured measurement is stored as a new document and remains
open in the Document space of the SPM Control Software.
162
LITHOGRAPHY TOOLBAR
The Vector Graphic Import dialog appears after clicking the “Import Vector” button and
selecting a valid GDS II file. It can be used to select the object (cell) of the GDS II file to
import. Size and origin of the resulting lithography object can be set during import using
the Size and Origin fields (see description below), or after import using the Object Editor
(see Section 17.3.2: The Object Editor dialog (page 191) for details).
Preview
A graphical area that displays the selected object in the available objects list (see above).
The red cross (if visible) indicates the position of the object's origin.
163
CHAPTER 15: LITHOGRAPHY
Size
Width / Height
Displays width and height of the selected object (cell).
Scale factor
The factor by which the selected object (cell) will be scaled. A scale factor of 1 corresponds
to the original object size. If the scale factor is changed manually, the object's width and
height will be recalculated and displayed automatically.
Origin
X-Offset / Y-Offset
The X-Offset and the Y-Offset of the origin of the selected object (cell).
Set origin to center
When enabled, the origin of the object (cell) will be set to the center of the rectangle that
encloses the object. When disabled, the origin will remain at the position that is defined in
the object.
164
LITHOGRAPHY TOOLBAR
The Pixel Graphic Import dialog appears after clicking and selecting a supported pixel
graphics file, and can be used to specify how such a file is converted to a lithography object.
All images are first converted to an 8-bit grayscale (256 levels). Each set of pixels with the
same grayscale value will correspond to a separate layer in the resulting lithography object.
Layers will only be generated for those grayscale values that are actually occupied. In
addition, the number of layers can be reduced upon import (see Simplify to (page 186)).
For each layer, individual lithography parameters can be set. One of these lithography
parameters can be automatically varied upon import, by using the grayscale values of the
imported pixel graphics file to define the selected parameter's range (see Parameters
below). All parameters can of course always be modified manually after import (see Section
17.3.1: The Layer Editor dialog (page 188)).
165
CHAPTER 15: LITHOGRAPHY
Preview
Graphical area that displays the selected pixel graphics file and information about the file.
Size
Width / Height
The width and the height of the lithography object resulting from the Pixel Graphic import.
The default settings for width and height are taken from the dimensions of the current
color topography map of the Lithography window. The pixel graphic is automatically
resized to fit into the area defined by these dimensions while maintain its aspect ratio. It is
at this point however possible to change the automatically calculated size manually. If the
width is changed manually, the height is recalculated to keep the aspect ratio. If the height
is changed manually, the width is not recalculated.
Origin
X-Offset / Y-Offset
By default the origin is in the center of the pixel graphic. By manually changing X-Offset and
Y-Offset, the origin may be moved to a different position.
Parameters
This area allows selection of the lithography parameter that will be automatically varied,
based on the different grayscale values of the imported pixel graphics file. The parameter
values for the black and white pixels of the imported pixel graphic can be set, after which
the parameter values corresponding to any in-between grayscale values are interpolated.
IMPORTANT
Since it is only possible to select one automatically adjusted lithography parameter per
import, the other parameters must be set elsewhere. This is achieved by setting the
Layer 0 parameters before import. The values entered here will be used as default values
for all imported layers, except for the parameter that was explicitly selected in the Pixel
Graphics Import dialog.
All values and settings in the parameters section are stored when the dialog is closed. They
will be automatically used the next time the dialog is opened.
From the drop-down list box, one of the following parameters may be selected for
automatic calculation:
– Tip speed
– Tip voltage
– STM Setpoint
166
LITHOGRAPHY TOOLBAR
Example
Setting the automatically adjusted Lithography parameter to “Static Force Setpoint”,
and Black (layer 0) and White (layer 3) to 25 μN and 10 μN, respectively, will result in:
• Layer 0 (black pixel layer) having a Static Force Setpoint of 25 μN
• Layer 1 (gray pixel layer 1) having a Static Force Setpoint of 20 μN
• Layer 2 (gray pixel layer 2) having a Static Force Setpoint of 15 μN
• Layer 3 (white pixel layer) having a Static Force Setpoint of 10 μN.
Simplify to
Select the number of layers the imported pixel graphics file should be simplified to.
Selecting the number of layers to be identical to the number of grayscale values in the pixel
graphics file will results in no simplification taking place. In all other cases, simplifications
are performed through binning of layers.
167
CHAPTER 15: LITHOGRAPHY
Figure 15-3: Lithography preview. (Left) Before the Lithography sequence has been started. (Right)
While the Lithography sequence is running.
The Lithography Topography chart is the only chart present in the Chart area of the
Lithography window (Figure 15-3: Lithography preview, left). It displays the topography
image of the sample surface to be used for lithography (after the Topography information
has been loaded via the “Load” button of the Lithography toolbar (see Load (page 156)) and
the superimposed preview images of the Lithography objects (when these have been
loaded and selected; see Import vector or Import bitmap (page 162)).
Before running a lithography sequence, a box with the size and content of the selected
lithography object is superimposed on the surface map. When selecting the center of the
box with the mouse, the corresponding object can be moved around the scan area to
reposition it. The new object location (X-Pos and Y-Pos) is however only transferred to the
object's properties (as displayed in the Lithography Objects list, and graphically shown on
the Topography chart) when the selection box is double-clicked after repositioning. If it is
not, any changes made are not implemented. Position of an object (and in addition its size)
can also be modified by editing the respective parameters for the selected object inside the
Lithography Object Editor dialog (see Section 15.3.2: Object Editor dialog (page 159)).
When a lithography sequence is started, the selection box will disappear, while a red circle
and a darker red trace line will be drawn on the Lithography object preview to provide a
live progress report on the Lithography drawing process (see Figure 15-3: Lithography
preview (page 168), right). After a lithography sequence has been completed, the red line
and circle will remain visible until a new lithography sequence is started.
168
CHAPTER 16:
16.1: Introduction
When working with the Nanosurf SPM Control Software, all finished measurements (a full
image was recorded) will be temporarily stored according to the file name mask you
specified in the Gallery panel (see Section 16.4: Gallery panel). These measurement
documents can be opened and displayed in the document area of the SPM Control
Software’s workspace (see Section 12.1: General concept and layout (page 110) and Section
12.4: Document space (page 113)). It is strongly recommended to permanently store
relevant documents to a new folder (see Save as (page 182) in Section 16.4.3: Gallery
toolbar).
Figure 16-1: Measurement document. Typical measurement window with the Data Info panel
expanded (see Figure 12-4: Example of a measurement document window (page 113) for a window with
a minimized Data Info panel).
Charts and the Data Info panel together display all available measurement information.
170
CHARTS
Just like the panels of the Info pane, the Data Info panel can be “pinned” and “unpinned”
to disable or enable the Auto-hide function of the panel (see Section 12.5: Panels (page
114)). It cannot be undocked from the document window, however.
The Data Info panel also contains a small toolbar, which allows you to customize the
presentation of the data and offers ways to export it.
Categorized
The “Categorized” button ( ) will group the data entries for the measurement document
by category. This is the default display method of the Data Info panel.
Alphabetical
The “Alphabetical” button ( ) will sort the data entries for the measurement document
alphabetically.
Save
The “Save” button ( ) will save the information in the Data Info panel to file. Possible
formats are text (.TXT) and comma separated values (.CSV) files.
Copy to Clipboard
The “Copy to Clipboard” button ( ) will copy the entries of the Data Info panel to the
Windows clipboard for easy pasting into other applications.
16.3: Charts
Charts provide a graphical display of the measured data. Charts occur in Measurement
document windows, in Operating windows, and in various other windows and dialogs. You
can adjust them to your needs and liking. How to do this is explained in this section. This
information is valid for charts in stored measurement documents as well as for ongoing
measurements in one of the Operating windows (Imaging, Spectroscopy, and Lithography).
A Chart consists of a graphical representation of the measured data itself and elements that
provide additional information. There are three basic chart types: Line graph, color map
and 3D view (see Figure 16-2: Elements of a chart, items 1–3).
Chart titles
The title elements of each chart display the signal name and the background line filtering
type that are used. A click on each of these titles opens a drop-down menu with other
possible signals or filters:
171
CHAPTER 16: WORKING WITH DOCUMENTS
1 2 3
5 4
6
7
Figure 16-2: Elements of a chart. (1) Line graph. (2) Color map. (3) 3D view. (4) Data Info panel (see
Section 16.2: Data Info panel). (5) Color scale for data Z-range. (6) Data range indicator, with scan head
Z-range as dotted box, data Z-range as solid gray box and current scan line height as red line. (7) Line
selection arrow. (8) “Chart Properties” button.
Color scales
The Color scale (Figure 16-2: Elements of a chart, item 5) shows which measured signal level
is mapped to which color. The color mapping can be changed using the Color Palette
dialog (see Section Color Palette (page 208)) .
172
CHARTS
This histogram displays the current height distribution of the measurement data and the
used color range. With the top and bottom range selectors, the color bar range can be
adjusted to the actual height distribution of the measurement data. Changing these range
settings changes the “Center” and “Span” parameters of the Chart Properties dialog (see
Section 16.3.2: Chart Properties dialog) and immediately updates the color display of the
data in the chart.
Chart properties
The “Chart Properties” button ( ; Figure 16-2: Elements of a chart, item 8) opens the Chart
properties dialog. This dialog is the center of all chart parameters and is described in more
detail in section Section 16.3.2: Chart Properties dialog.
Most chart settings can also be accessed from a context menu, which is opened by right-
clicking a chart.
173
CHAPTER 16: WORKING WITH DOCUMENTS
chart is indicated by a thin blue line around the chart area. A chart is activated by clicking
on it with the mouse cursor anywhere in the chart area.
Arrangement of the charts is performed automatically by the control software based on
the size of the window and based on the order in which the charts were generated/added.
If the window is to small to display all charts, scrollbars are displayed at the border of the
window.
Short cuts to add and remove charts are found in the chart context menu. Select “Create
new chart” or “Delete current chart” from the menu list. It is also possible to use the “Insert”
or “Delete” key of your computer’s keyboard for this task.
The Charts Properties dialog is used to set all chart properties that influence data display
by the respective chart. It may be kept open at all times if many parameters have to be set
for different charts.
Some parameters are chart type specific. They are therefore displayed at the bottom of the
Chart Properties dialog in a separate group.
174
CHARTS
Add chart
The “Add Chart” button ( ) creates a copy of the currently selected or active chart and
adds it to the active window in last position.
Remove chart
The “Remove Chart” button ( ) removes the currently active chart.
Previous chart
The “Previous Chart” button ( ) activates the previous chart and updates the
parameters displayed in the Chart Properties dialog to those of that chart.
Next chart
The “Next Chart” button ( ) activates the next chart and updates the parameters
displayed in the Chart Properties dialog to those of that chart.
Chart parameters
Type
Selects the chart type to be used for display of the measurement data:
– Line graph
Data is displayed as a line plot. Points outside the range of the scanner are displayed in
red. The line being displayed is selected by dragging the Line selection arrow in a Color
map (see Line selection arrow (page 173)). In ongoing measurements (e.g. during
imaging) the position of the Line selection arrow is updated automatically and
corresponds to the last measured scan line (but even here it is possible to select a
different line for view in a line graph by dragging/holding the Line selection arrow in a
different location).
– Color map
Z-height data is encoded using a color scale and displayed 2-dimensionally.
175
CHAPTER 16: WORKING WITH DOCUMENTS
– 3D view
Data is shown in a 3-dimensional representation in parallel perspective. Color
information (such as implemented in the Color map) is maintained.
Signal
Selects the input channel (signal data) to be used for the chart. The available signals
depend on the operating mode (selected or used) and the status of the User inputs.
Filter
Selects the line filter method. The control software applies this filter to the measured data
before displaying it (see Figure 16-3: Data filter types). No modification of the original
measurement data occurs (selecting another filter is always possible). Available data filters
are:
– Raw data
No data processing.
– Mean fit
Calculates the mean value of each line of data points and subtracts this number from the
raw measurement data for each data point of that line.
– Line fit
Calculates the first order least squares fit (mean value and slope) for each line of data
points and subtracts the fitted values from the raw measurement data for each data
point of that line.
– Derived data
Calculates the difference between two consecutive data points (derivative) and displays
this instead of the raw image data.
– Parabola fit
Calculates the second order least squares fit for each line of data points and subtracts the
fitted values from the raw measurement data for each data point of that line.
– Polynomial fit
Calculates the fourth order least squares fit for each line of data points and subtracts the
fitted values from the raw measurement data for each data point of that line.
Display size
The size of the chart in pixels.
Show Axis
When checked (default), the axis labels, color and range scales, and titles are displayed
alongside the graph. When unchecked, they are hidden.
Keep Aspect ratio
When checked, the axis in the color map are drawn in their correct size-relation (according
to their value and unit). When unchecked (default), the size of the display is always a square
and data pixels are stretched if necessary.
176
CHARTS
Figure 16-3: Data filter types. The same measurement data displayed using the available filters. A
defective area on a calibration grid is shown here to illustrate the effect of the filters.
Span
The span that corresponds to the chart’s displayed Z-range. Increasing Span decreases
feature contrast and vice-versa. The current span is displayed next to the color scale in color
maps, or can be inferred from the Z-axis labels in Line graphs and 3D views.
Center
The signal value that corresponds to the center of the “Span” parameter.
Auto set
When checked, the chart’s Z-range is automatically set to optimally match the
measurement data. During measurements, the Span and Center parameters will be
updated continuously (i.e., the chart adapts to the available data).
177
CHAPTER 16: WORKING WITH DOCUMENTS
Set button
Clicking this button starts the optimization of the Z-range manually. Mostly used when
“Auto set” is off.
Figure 16-4: Show reverse line option. (Left) Reverse line disabled. (Right) Reverse line enabled.
Shading
When checked, the color map creates the impression of a 3-dimensional surface which is
lighted from the left. This is achieved by combining the topography with its derivative. The
number of pixels in the edit box defines the amplification of the derivative add on.
Smooth pixel
When checked, the screen edge rendering of individual data pixels is smoothed with their
neighboring pixels. Alternative data pixels are drawn as individual squares. This smoothing
shows the most effect when the display size is larger that the number of measured of data
points (e.g. a 256×256 measurement displayed at 512×512 pixels).
178
CHARTS
Figure 16-5: Shading option. (Left) Shading disabled. (Right) Five-pixel shading enabled.
3D View options
Pos X, Pos Y
Defines the center position of the 3D plot inside the chart area.
Rotation
Defines the z-axis rotation of the 3D plot relative to the view point.
Tilt
Defines the off-plane angle of the 3D plot.
Z-Scale
Defines a Z-axis ‘stretch’ factor. Use this e.g. to enlarge surface details.
Zoom
Defines the magnification of the 3D plot.
179
CHAPTER 16: WORKING WITH DOCUMENTS
Light Rot
Defines the rotation angle of the light source relative to the Z-axis (360°)
Light Tilt
Defines the off-plane angle of the light source. The lowest value (0°) corresponds to
“sunset” lighting, the highest value (90°) corresponds to mid-day lighting at the equator.
Default
The “Default” button resets all 3D parameters to their default values.
Keyboard and Mouse short cuts
Always click and hold the left mouse button on the 3D view chart while moving around the
mouse to change the 3D view. The surface is reduced in feature complexity once the left
mouse button is pressed to speed up redrawing on the screen. The surface will return to full
detail once the mouse button is released.
Press the following additional keys/buttons to determine which chart property is changed:
– Surface rotation
Mouse left/right
– Surface tilt
Mouse up/down.
– Size displayed surface
“Ctrl”- key + mouse up/down
– Surface position
“Shift”-key + mouse up/down/left/right
– Z-scale magnification
Left mouse button + right mouse button + mouse up/down
– Light source direction
“Shift” + ”Ctrl”-key + mouse left/right
– Light source height
“Shift” + ”Ctrl”-key + mouse up/down
180
GALLERY PANEL
The Gallery Panel displays lists of thumbnails representing previous measurements for
quick access to those documents. It contains two pages: the “History” page, with the
temporarily (automatically) stored measurements (for details on how to change the default
History folder and the maximum number of files to be stored see Gallery Settings (page
211)) and the “File Browser” page with measurements from a user-selected directory (e.g.
containing older measurements that were previously moved there). The various elements
of the Gallery panel are described in the next sections.
181
CHAPTER 16: WORKING WITH DOCUMENTS
The temporary (automatic) storage of new measurements uses a File mask to create new
file names each time a new measurement has to be saved. This mask can contain normal
text, but also special variables like index number or date and time stamps. You may enter a
new mask directly into the edit field in the History page, select an old mask from the drop-
down menu, or define a new mask with the help of the Mask Editor dialog (see Section
16.4.4: Mask Editor dialog), which is opened by clicking the “Mask Editor” button next to the
edit box.
The Gallery toolbar is present in both the “History” and “File Browser” pages. It performs
similar functions in both pages.
Save as
Selected measurements can be saved to a new location with this button.
If only a single measurement is selected a standard Windows “Save” Dialog is shown. Here
you select the new location and the new file name.
If multiple measurements are selected a “Folder” dialog is shown, which allows you to
select (or create) the folder that all selected files are to be copied to. It is strongly
182
GALLERY PANEL
recommended to do this for files in temporary storage of the History folder that you want
to keep, because they may be overwritten as soon as the maximum number of files in the
History folder is reached (see Section Gallery Settings (page 211)
Rename
Single or multiple measurements can be renamed by clicking the “Rename” button. It will
open the File Rename dialog (see Section 16.4.5: File Rename dialog), which allows you to
specify a mask for the new file names.
Delete
Single or multiple measurements can be deleted by clicking this button.
The Mask Editor dialog assists you in the creation of file name masks.
History mask
Filename
A file name mask is the template that is used to generate the file names for documents that
are temporarily stored (automatically) during measurement. A file mask consists of
standard text entered by the user and of variables for software-generated text or numbers
(either specified by the user or added automatically).
Mask variables
Mask variables can be entered as specific words surrounded by square brackets. The
following variables are defined in the SPM Control Software:
– [INDEX]
This variable represents a number that is automatically incremented each time a new
filename is created. The index number is 5 digits in length and filled with zeros for
missing digits. The next index number that will be used is shown in the “Next Index” field.
183
CHAPTER 16: WORKING WITH DOCUMENTS
– [TIME]
This variable represents the actual time of file name creation. It is formatted with two
digit numbers for the hours, minutes and seconds (HHMMSS). This time format is used
regardless of the Regional Settings of the Windows operating system.
– [DATE]
This variable represents the date of the day the file was created (i.e., the day the
measurement was performed). It contains four digit numbers for the year and two digit
numbers for month and day (YYYYMMDD). This date format is used regardless of the
Regional Settings of the Windows operating system.
To quickly insert a Mask variable at the current cursor position in the mask edit box, click
the corresponding button.
Next Index
This entry field defines the next index number to be used. By default this value is identical
to the highest number present in the History files, increased by one.
IMPORTANT
If no Mask variable is used, an index is automatically added to the text string defined in
the filename mask.
Preview
The filename that will be used for the next measurement document to be saved (as
specified by your entries) is shown here.
184
GALLERY PANEL
The File Rename dialog is used to rename (or move) multiple files using a Rename mask
(see below). The dialog is opened by clicking the “Rename” button in the Gallery panel. To
rename a file or multiple files, the Rename mask can be defined here following the same
principles as for the History mask. A preview of the new filenames is shown in the preview
section.
Rename mask
See History mask (page 183).
Preview
Source Filename
The left column of the Preview section shows the original filename(s).
Target Filename
The right column of the preview section shows what the filename(s) will be after pressing
the “Rename” button.
185
CHAPTER 16: WORKING WITH DOCUMENTS
Refresh
This button updates the preview list.
Rename
This button renames the selected files.
IMPORTANT
If no unique filename(s) would result from the specified Rename mask, an index is
automatically added to the files. If this still does not result in unique filenames, the text
“Copy of” is added to each filename as often as is required to make it unique.
Cancel
This button closes the dialog without renaming the files.
All of these tools can also be used while a measurement is still being acquired.
To use a quick evaluation tool:
1 Click on the chart that you want to evaluate to activate it.
2 Select the desired tool using one of the following approaches:
• Click on one of the tool buttons in the Analysis tab.
• Select the tool from the Chart’s context menu (right-click on the chart).
3 Define the evaluation. The procedure to define the evaluation is different for each tool.
Details can be found in the tool-specific instructions below.
When a tool has been selected, the Tool panel (see Section 16.6: Tool panel) moves to the
top of the panel stack of the Info pane.
IMPORTANT
Depending on the selected chart type, some tools may be unavailable.
186
ANALYSIS TAB
Tip
For more elaborate evaluations, the optional Nanosurf Analysis or Nanosurf Report
software can be used.
Measure Length
Calculates the distance and signal difference between two points. Graphically, a line with
arrowheads on each end represents the selection marker. The line is defined by drawing a
line on the measurement chart. The first point is positioned by moving the mouse cursor
to the desired location and clicking and holding the left mouse button. The second point
is positioned when the mouse button is released. When the mouse is not moved between
clicking and releasing, an line parallel to the X*-axis is drawn.
The direction and length of the selection marker can be adjusted by dragging the end
markers. The line can be moved as a whole by dragging the center marker.
The Tool status section of the Tool panel displays the calculated “Length”, “DeltaZ”, “Width”
and “Height”. This data will also be stored in the “Tool” data category of the Data Info panel
187
CHAPTER 16: WORKING WITH DOCUMENTS
(see Section 16.2: Data Info panel) for the respective measurement document as long as the
tool is active when the document is stored. For more information on the data in the Tool
status section (see Tool status section (page 199)).
Measure Distance
Calculates the distance between two parallel lines. The parallel lines are defined by drawing
them in the chart. The first point of the first line is defined by the mouse cursor position
where the left mouse button is clicked, the second point by the position where the button
is released. When the mouse is not moved between clicking and releasing, a line parallel to
the X*axis is drawn. After releasing the mouse button, a second parallel line sticks to the
mouse cursor, that is released by clicking its desired position. The direction of the parallel
lines can be adjusted by dragging their end markers; they can be moved by dragging the
center marker.
The Tool status section of the Tool panel displays the calculated distance. The distance
value only depends on the cursor positions, it does not on depend the displayed data
values. The distance data will also be stored in the “Tool” data category of the Data Info
panel (see Section 16.2: Data Info panel) for the respective measurement document as long
as the tool is active when the document is stored. For more information on the data in the
Tool status section (see Tool status section (page 199)).
188
ANALYSIS TAB
Measure Angle
Calculates the angle between two lines. In Line graph-type displays, this tool can only be
used when the chart displays data that has the unit “meters”.
The two lines are defined by drawing them in the chart. The first point of the first line is
defined by the mouse cursor position where the left mouse button is clicked, the second
point by the position where the button is released. When the mouse is not moved between
clicking and releasing, a line parallel to the X*-axis is drawn. After releasing the mouse
button, the end of the second line sticks to the mouse pointer. The end is released by
clicking its desired position. The angle can be changed by dragging the line end point
markers or the corner mark; it can be moved by dragging the line center markers.
The Tool status section of the Tool panel displays the calculated angle. This data will also be
stored in the “Tool” data category of the Data Info panel (see Section 16.2: Data Info panel)
for the respective measurement document as long as the tool is active when the document
is stored. For more information on the data in the Tool status section (see Tool status section
(page 199)).
In contrast to the evaluation tools of the Measure and Roughness groups, the tools of the
Correction group (and also those of the Filter group (see Filter group (page 193)) actually
change measurement data. This is done in a copy of the original measurement document,
though, so you won’t lose any data and will always be able to access the original
measurement data in addition to the corrected or filtered data.
189
CHAPTER 16: WORKING WITH DOCUMENTS
Correct Background
Removes the effect of an ill-aligned scan plane when the line filter options (see Filter (page
176)) do not give satisfactory results. This may be the case when the scan lines in different
parts of the measurement have a different average height. An example of such a
measurement is shown in Figure 16-6: Correct Background.
Figure 16-6: Correct Background. (Left) Uncorrected image; the end points of the selection marker
have been moved to points that should have the same height. (Right) Corrected image.
To use the tool, select three points that should be on the same height. This is done in the
same way as with the angle tool (see Measure Angle (page 339)). The selected points
become the end points of the selection marker.
After clicking the “Execute” button in the Tool status section of the Tool panel, a copy of the
original measurement document is made and the plane that is defined by the selection
maker is subtracted from the measurement data in the newly created document. To get
useful results, the Data filter option for the corrected image in the new document will be
automatically set to “Raw data”.
Correct scan line levels
Removes the effect of drift when the line filter options (see Filter (page 176)) do not give
satisfactory results. This may occur when the scan lines in different parts of the
measurement have a different average height. An example of such a measurement is
shown in Figure 16-7: Correct scan line levels.
To use the tool, draw a line through points that should have the same height in the same
way as with the Measure Length tool.
After clicking the “Execute” button in the Tool status section of the Tool panel, a copy of the
original measurement document is made and the average level of each scan line in the
newly created document is adjusted so that all points along the drawn line have the same
190
ANALYSIS TAB
Figure 16-7: Correct scan line levels. (Left) Uncorrected image with a selection marker through
points that should be at the same height. (Right) Corrected image
height. To get useful results, the Data filter option for the corrected image in the new
document will be automatically set to “Raw data”.
191
CHAPTER 16: WORKING WITH DOCUMENTS
The Tool status section of the Tool panel displays the calculated “Length” and “DeltaZ” of
the selected area.
The Tool result section displays the roughness values that are calculated from the data
according to the following formulas:
The Roughness Average, Sa The Mean Value, Sm The Root Mean Square, Sq
N-1 N-1 N-1
Sa = 1 | z( x l ) | Sm =
1
z( x l ) Sq =
1
( z( x l )) 2
N l =0 N l =0 N l =0
The roughness values depend on the line filter option (see Filter (page 176)) that is applied
to the chart, because they are calculated from the filtered data.
Clicking the “Store” button in the Tool result section stores the roughness values in the
“Roughness” data category of the Data Info panel for the active measurement document.
Calculate Area Roughness
Calculates several roughness parameters from the data points in a selected area.
The area is selected in the same way as with the Cut out Area tool.
The Tool status section of the Tool Results panel displays the calculated Size or Width and
Height of the selected area. For more information on the Tool status section, see The Tool
status section (page 207).
The Tool result section displays the roughness values that are calculated from the data
according to the following formulas:
192
ANALYSIS TAB
The Roughness Average, Sa The Mean Value, Sm The Root Mean Square, Sq
M-1 N-1 M-1 N-1
Sa = 1 1 1 M-1 N-1
| z( x k , y l ) | Sm = z( x k , y l ) Sq = ( z( x k , y l )) 2
MN k = 0 l =0 MN k = 0 l =0 MN k =0 l =0
The roughness values depend on the Data filter that is applied to the chart, because they
values are calculated from the filtered data. More information on data filters is provided in
Section 18.2: The Chart bar under Data filter (page 199).
Clicking the “Store” button in the Tool result section stores the roughness values in the
“Roughness” data category of the Data Info panel for the active measurement document.
Tip
The Area Roughness tool can be used to determine the mean height difference between
two plateaus with more accuracy than with the “Measure Distance” tool. To determine
the mean height difference, select an area on each plateau, and calculate the difference
between their Sm-values.
Glitch Filter
The Glitch Filter removes the effect of small defects in the image such as single short
glitches in the scan. Compared to the Noise Filter (see below), it has the advantage of not
reducing resolution on step edges. The glitch filter is implemented as a Median filter on a
3×3 pixel matrix.
To apply the filter, activate the color map chart that is to be filtered, then click the “Glitch
Filter” button. A new Measurement document with the filtered data is created.
Noise Filter
The Noise filter removes high frequency noise from the image, but applying the filter will
also decrease the resolution of the image. The Noise Filter is implemented as a convolution
with a 3×3 pixel Gaussian kernel function.
193
CHAPTER 16: WORKING WITH DOCUMENTS
Figure 16-8: Glitch Filter. (Left) Unfiltered image with some glitches where the tip lost contact with
the sample. (Right) Corrected image.
Figure 16-9: Noise Filter. (Left) Noisy (unfiltered) image of an AFM measurement on HOPG. (Right)
Filtered image
To apply the filter, activate the color map chart that is to be filtered, then click the “Noise
Filter” button in the Tools bar. A new measurement document with the filtered data is
created.
Tip
• Filters are especially useful for improving the appearance of 3D views.
• Applying filters may changes the result of the other tools. This may result in incorrect
results, e.g. when evaluating sample roughness.
194
ANALYSIS TAB
Create Cross-Section
Creates a new measurement document containing a line cross-section of a Color map or
Line View display.
The line is defined by drawing a selection arrow. The arrow points toward the forward
direction of the line. The start of the arrow is defined by the mouse cursor position where
the left mouse button is clicked, the end of the arrow by the position where the button is
released. When the mouse is not moved between clicking and releasing, an arrow ending
in the center of the measurement is drawn. The direction of the arrow can be adjusted by
dragging its end markers; it can be moved by dragging the center marker.
Double-clicking the graph, or clicking the “Cut out line”-button in the Tool status section of
the Tool panel creates a new document that contains the line section.
The Tool chart section of the Tool Results panel displays a preview chart of the selected line.
The Tool status section of the Tool Results panel displays the calculated “Length” and
“DeltaZ” of the selected line. For more information on the data in the Tool status section
(see Tool status section (page 199)).
Cut Out Area
Creates a new measurement document containing a subsection of an existing
measurement.
195
CHAPTER 16: WORKING WITH DOCUMENTS
One corner of the area is defined by the mouse cursor position where the left mouse button
is clicked, the opposite corner by the position where the button is released. When the
mouse is not moved between clicking and releasing, an area is defined that has a size of
33% of the current measurement, and is centered on the clicked location.
Once an area is defined, it can be resized by dragging one of its corners, and moved as a
whole by dragging its center point.
Pressing the “Shift” key while dragging a corner defines a non-square (i.e. rectangular) area.
Double-clicking the graph, or clicking the “Cut out area” button in the Tool Results panel
creates a new measurement document that contains the selected area.
The Tool status section of the Tool Results panel displays the calculated “Size” or “Width”
and “Height” of the selected area. For more information on the data in the Tool status
section (see Tool status section (page 199)).
The Nanosurf Report software offers a powerful and extensive set of analysis functions.
Complex analyses can be created interactively, and then displayed and printed in visually
appealing reports. These reports can then be used as templates to consistently apply the
same analysis to other measurements.
196
ANALYSIS TAB
Report
The “Report” button starts the Report software (if installed) from within the SPM Control
Software:
When a measurement is opened by the Report software, it will import all measurement
channels that are displayed in the current measurement document. By default, a Basic
Report is generated. Other Report styles can be chosen from the “Report” button’s drop-
down-menu:
This drop-down menu lists templates stored in a template folder. The Menu Item's name is
equal to the template name without the extension (*.mnt) and is sorted alphabetically. This
standard directory is configured by the Report Template File path (see Reporting (page
210)).
IMPORTANT
After a fresh installation of the Report software, the Report software has to be run at
least one time before you can automatically start it from the SPM Control Software. To
run the Report software for the first time, select it from the Windows “Start” menu.
New Report
An empty report is opened.
Add Measurement
The currently active measurement is added to the currently opened report.
Apply Template
Opens a dialog that allows you to select a template that is applied to the currently active
measurement. If selected, a menu item the template is applied to the current selected
measurement Document.
IMPORTANT
If you do not save the measurement in the control software, but only save the report, the
data in measurement channels that were not displayed is lost.
197
CHAPTER 16: WORKING WITH DOCUMENTS
IMPORTANT
A measurement document should only display those channels that are used in a
template. When a template is applied to a measurement document that displays
different, or a different number of measurement channels than the template uses, the
results may not be correct.
The Tool panel of the Info pane displays varying information, which depends on the tool
currently selected in the Analysis tab.
198
TOOL PANEL
2 2
Length = Width + Height
th
ng
Le
Height
Width
In a Color map chart, length is calculated in the XY-Plane. In a Line graph chart, length is
calculated in the XZ-Plane.
“Length” is not displayed when “Width” and “Height” are of different physical units (e.g. in
Amplitude Spectroscopy, where the X-Axis is given in [m] and the Z-Axis in [V]).
Width, Height
The “Width” and “Height” of the measurement tool in the chart, calculated in the chart
plane.
199
CHAPTER 16: WORKING WITH DOCUMENTS
DeltaZ
The difference between the “Z-Pos” values at both ends of the selection marker.
In a Color map chart, “DeltaZ” is the difference in the (filtered) sample height between the
start and the end point.
IMPORTANT
The calculated values of “Length”, “Width” and “Height” only depend on the cursor
positions, they do not depend on the displayed data values.
200
CHAPTER 17:
Advanced settings
0
0
CHAPTER 17: ADVANCED SETTINGS
The About dialog displays information that may be useful for diagnostics when you have
problems with your instrument. The About dialog is opened by clicking the “Information”
button on the upper right corner of the program window, just below the “close window”
button:
202
FILE MENU
The File menu is accessed by clicking the blue File tab at the left side of the Ribbon. It
provides access to mostly software related settings and options, but also to functionality
such as opening, storing and printing of measurements. The latter function can also be
performed using the Quick Access toolbar, in which these commands are present by
default:
Open
Launches a system “File open” dialog for opening Nanosurf “.nid” or “.ezd” (easyScan 1) files.
It is possible to select more than one file at the same time by using the “Shift” and/or “Ctrl”
keys.
Selected files will open in measurement windows, which contains a chart area and a data
info panel. There is no Imaging toolbar like there is for ongoing measurements in the
Imaging window. You can however still customize the charts through a context menu,
which opens through a right-click. The data info Panel displays all significant parameters
were used for the measurement. For more information on Measurement document
functionality, see Chapter 16: Working with documents (page 169).
203
CHAPTER 17: ADVANCED SETTINGS
204
FILE MENU
The data lines are stored from the bottom to the top. A small header at the beginning of the
first data line contains the names of the channel and frame, as well as X-, Y-, and Z-ranges
with their physical units.
Comma separated z values (.csv)
This format stores all the measured data in a chart, as a matrix of floating point numbers in
ASCII format separated by a “comma” and “SPACE” character. This enables easy data
exchange with commonly used spread sheet and database applications.
(X, Y, Z)-Points (.csv)
This format stores the coordinates of all measured points in a chart as a list of floating point
number pairs. For Line graphs, only X and Z points are exported.
Parameters
All measurement parameters are stored in a configuration file with the extension “.par”.
When the SPM Control Software is started, default values are loaded from a file that is
selected in the Controller Configuration Dialog (Section 17.8: Controller Configuration
dialog). Functions for storing and retrieving parameters are accessed via the application
button.
Save
Saves the parameters to the currently selected parameter file. The name of this file is
indicated in the status bar at the bottom of the main window.
Save as...
Saves the parameters under a new file name.
Load
Loads a previously saved parameter file.
Chart Arrangement
The chart arrangement of the Imaging and Spectroscopy windows is stored in a
configuration file with the extension “.chart”. When the SPM Control Software is started, a
default arrangement is loaded from a file that is selected in the Controller Configuration
dialog (Section 17.8: Controller Configuration dialog). Functions for storing and retrieving
the chart arrangement are accessed via the application button.
Save
Saves the chart arrangement to the currently selected chart file. The name of this file is
indicated in the status bar at the bottom of the main window.
205
CHAPTER 17: ADVANCED SETTINGS
Save as...
Saves the chart arrangement under a new file name.
Load
Loads a previously saved chart file.
Options
The “Options” button opens the Options dialog (see Section 17.2.1: Options dialog), which
configures various general control software settings.
Exit
With exit you can close the SPM Control Software. If you exit the program while still having
unsaved data, you will be asked to save it.
Allows changes to the content of the Quick Access toolbar (see File menu (page 203)). This
process is very similar to that in the Microsoft Office applications and is therefore not
explained in this manual. You are free to try it out and can always use the “Reset” button to
reload standard settings.
206
FILE MENU
User Interface
Language
Specifies the SPM Control Software language.
Save workspace on exit
When checked, the workspace settings are automatically saved to the system registry
when the control software is closed (see also Parameters and Chart Arrangement (page
205)).
207
CHAPTER 17: ADVANCED SETTINGS
Color Palette
The color palette dialog is reached via the menu item “Options” >> “Config Color palette...”.
The color palette is used to map the display range of the measured values to a color. Three
different palette types are available:
– Black&White
The color map is a linear gray scale.
– Color
The color selection uses the HSB-color model where the color (H) is set in ° value. The
color is selected by entering a number or by clicking a color in the color bar.
– Look Up Table
A user definable palette (with max 256 color entries) can be selected. This palette is
stored in a “.lut” file that contains an ASCII table with RGB color values. A different look up
table can be selected by clicking the “Browse...” button.
208
FILE MENU
Scripting
Allows you to set the search paths for the Acquisition and Analysis scripts that are displayed
in the Scripting group of the Acquisition tab and Analysis tab, respectively.
Scripts can be organized in subdirectories inside each of the “Script” directories, which are
displayed as submenus in the control software. These submenus are displayed before
individual scripts in the Script drop-down menu.
209
CHAPTER 17: ADVANCED SETTINGS
Reporting
Used to configure the behavior of the “Report” button in the Report group of the Analysis
tab.
Apply the following template
When enabled, automatically applies the template specified in the box below. You can
search for a template using the “Browse” button.
Display templates in the following directory
As with the Stridulating paths (see Scripting (page 209)), the content of the specified
directory is displayed as choices in the Report drop-down menu.
210
FILE MENU
Gallery Settings
History files
Sets the directory where the temporarily (automatically) stored measurements (which are
listed in the Gallery panel of the Info pane) are stored.
Max History Files
Sets the maximum number of files to keep in the above directory. When the maximum is
reached, the oldest measurement is deleted from disk to allow the latest measurement to
be saved.
211
CHAPTER 17: ADVANCED SETTINGS
Access Code
Used to enter the access code for software modules, such as the Scripting Interface and the
Lithography Option...”
IMPORTANT
If you receive the warning “To change access codes you need Windows administrator
rights”, please restart the control software with the “Run as Administrator...” option from
the Windows Explorer context menu.
212
SETTINGS TAB
Calibration
This button opens the Scan Head Selector dialog to load, save or edit a scan head
calibration file. For more details, see Section 17.4: Scan Head Selector dialog (page 214).
Diagnosis
This button opens the Scan Head Diagnosis dialog where the actual health state of the scan
head can be seen. For more details, see Section 17.7: Scan Head Diagnosis dialog (page 218).
Controller
Opens the Controller Configuration dialog where different hardware related settings can
be defined. It defines communications port, video driver settings, start up parameter and
others. For more details, see Section 17.8: Controller Configuration dialog (page 220).
Signal Access
Opens the SPM Parameter dialog on the Signal Access page (see Section 17.9.7: Signal
Access (page 235)).
Simulation
Check or uncheck the Simulation button to enter or exit the control software’s microscope
simulation mode. Once the simulation mode is active, the status bar of the control software
displays the text “Simulation”. Otherwise, this field displays the text “Online”.
In microscope simulation mode, many functions of the microscope are performed on a
mathematically generated surface. Thus, software functionality and acquisition
procedures can be practised without danger of harming the instrument.
213
CHAPTER 17: ADVANCED SETTINGS
The Scan Head Selector dialog is used to load, save or edit scan head calibration files. These
files store all calibration values specific to a certain scan head. The Scan Head selector
dialog is opened via the “Calibration” button in the Scan Head group of the Settings tab.
The configuration of each scan head is stored in a file with a filename that corresponds to
the serial-number of that particular scan head and with the extension “.hed” (e.g. “10-11-
584.hed” for an easyScan 2 AFM scan head). The currently loaded scan head calibration file
is displayed in the status bar.
Tip:
The specific scan head calibration file(s) for each customer is automatically copied and
selected as default during the installation of the control software from the installation
CD. It can be found in the “Calibrations” sub directory of your installation path.
IMPORTANT:
When you change a scan head, you have to load the correct configuration file too. If you
do not, scan ranges and other important calibration settings are incorrect and the scan
head may not operate properly.
Load...
Loads a different scan head calibration file.
Save as...
Saves the current scan head calibration file with a different name.
Edit...
Edit the currently loaded scan head calibration file using the Scan Head Calibration Editor
dialog (see Section 17.5: Scan Head Calibration Editor dialog). Always save a backup copy of
the original scan head calibration files by clicking 'Save As...' first.
214
SCAN HEAD CALIBRATION EDITOR DIALOG
CAUTION!
Changes to these settings should be performed with great care. False settings can lead
to false interpretation of the data and incorrect operation of the controller.
215
CHAPTER 17: ADVANCED SETTINGS
Axis Orthogonality
The X- and Y-Axes of the scanner are generally not perfectly orthogonal, and their
orientation with respect to the AFM housing may vary. The controller corrects these errors
by adding/subtracting some of the X scanner command signal to the Y scanner command
signal and vice versa.
X/Y Angle
The angle between then the X- and Y-axis of the scanner hardware. The control software
uses this value to correct the scan command signals such that the scan axes are orthogonal.
Rotation
The angle between the X-axis of the scanner and the X-axis of the microscope body (see
Figure 10-3: Scanner coordinate system (page 98)). The control software uses this value to
correct the scan command signals in such a way that the scan axis is parallel to the X-axis
of the microscope body.
Tip
With this value, the alignment of the scanner's 0° Rotation and another system’s
coordinate system may be calibrated (e.g. images scanned with the LensAFM or Nanite
AFM and optical images by external video cameras of optical profilometers or Nano-
Indenters).
216
SCAN HEAD CALIBRATION EDITOR DIALOG
IMPORTANT
This value has been pre-configured by Nanosurf for Static Force Mode operating mode
with CONTR Cantilever and a Laser Spot at 225 μm. If other Cantilevers are used, or if the
laser spot has been adjusted manually in case of a FlexAFM scan head, a recalibration of
this value has to be performed. If not, the Set Point in [N] may be incorrect.
217
CHAPTER 17: ADVANCED SETTINGS
This dialog can be used to correct the scan range by entering a correction factor based on
a measured distance and a known real distance.
This correction factor could for example be determined by evaluating the height
information in a measurement of a calibration grid with known properties.
Tip
The Scan Head Diagnostics dialog cannot be accessed once the tip has been
approached to the sample. In this case, retract the tip first.
218
SCAN HEAD DIAGNOSIS DIALOG
Laser Status
Status about the laser / detector system is shown here.
If the Status light on the SPM controller is blinking red, more detailed information about the
failure is displayed here (see also Section 9.3.1: Probe Status light blinks red (page 91)).
Information about the preamplifier that is present in the STM scan head is displayed here.
The offset current is a leakage current that is measured by the preamplifier when it is not in
contact with the sample.
219
CHAPTER 17: ADVANCED SETTINGS
Note
The offset current is not only a measured value, but is also used as a compensating value
during measurement. Therefore, even high values are not problematic for measuring at
low tunnelling current.
With this dialog some controller hardware related settings can be configured. On a
correctly installed system, it should not be necessary to change these settings manually.
The Controller configuration dialog is opened via Controller” button in the Hardware group
of the Settings tab.
Start configuration
The parameter and chart arrangement files that are loaded when the SPM Control Software
starts. Each Windows user has his/her own set of these two files a personal “Local Settings”
directory. Therefore, each windows user can configure the control software to his/her own
personal preferences without any consequences for other users.
220
SPM PARAMETERS DIALOG
USB Connection
The SPM controller uses a virtual serial port that is connected to the USB port. The number
of this virtual serial port should be the same as the one shown in your the windows device
manager dialog. Activate “Auto detect” to let the control software search for the right COM
port at each program start. This is highly recommended, because Windows assigns
individual COM port numbers to different USB connectors. With auto detect, you will be
able to plug in the USB cable to different ports.
Tip
If the port number is set to “No Controller (Simulation only)” and “Auto Detect” is
switched off, the control software will always start in Simulation mode. This could be
useful if the software will be used mainly for analysis and is installed on a PC without
microscope hardware.
Video Signal
Allows the selection of the video device driver used to handle the video camera of the SPM
controller or scan head. Activate “Auto detect” to let the control software search for the
right device driver. The control software then automatically selects the correct device
driver for different scan heads if available. If a video device is found, the Video Panel will
automatically be present in the Info pane.
Select “No video” in the list if you wish to completely suppress the video display.
Microscope Firmware
Here the currently used firmware version of the controller is shown.
In case of a standard software update (downloaded from the Nanosurf homepage) manual
update of new firmware is normally not necessary since it is performed automatically at the
start of the updated SPM Control Software. Automatic firmware update is always
performed each time a different (older or newer!) software version is started since last time.
Click the “Update” button to install individual firmware updates you received from
Nanosurf support.
221
CHAPTER 17: ADVANCED SETTINGS
• Spectroscopy
• Lithography
• Operating Mode
• Approach
• Z-Controller
• Signal Access
These pages are described in more detail in the next sections.
17.9.1: Imaging
Image parameter
Image size
The image size in X*-direction and the image size in Y*-direction. When the Check-box is
active, the image Height is always identical to the Image width.
Measurements
The number of measured data points and data lines in an image. When the Check-box is
active, the number of Lines is always equal to the number of Points / Line.
222
SPM PARAMETERS DIALOG
Time / Line
The time needed to acquire a data line. The time needed for the entire image is displayed
in the status bar.
Rotation
The angle between the X-direction of the scanner and the X* direction of the measurement
(Figure 15-2: Coordinate systems).
Imaging options
Overscan
The “Overscan” determines how much the effective scan range is increased relative to the
image width. This will eliminate edge effects caused by the reversal of the scanning motion
by not recording or displaying them in the measurement image. Disadvantages of using
Overscan are that the maximum scan range is reduced, the tip moves slightly faster over
the sample with the same “Time/Line” setting, and the tip may hit large features outside the
measured image.
Ref. Z-Plane
The height of the reference plane. This height reference is used when the Z-Controller
output is cleared, and when the Z-position is not modulated relative to the current surface
position during spectroscopy measurements.
The reference plane for the image can be aligned to the surface of the sample using the
slope parameters (see Figure 5-2: Sample and measurement orientation before slope
adjustment (page 58) or Figure 13-2: Coordinate systems (page 124)).
Image offset X/Y
The center position of the measured area.
Slope X
A positive value rotates the image plane around the Y-axis counterclockwise.
Slope Y
A positive value rotates the image plane around the X-axis counterclockwise.
The center position of the measured area can be changed by typing its position as well as
by using the Move tool in the Imaging toolbar. The zero position corresponds to the center
position of the scanner.
Adjust slope
The “Adjust slope” button will cause the control software to set appropriate values for X-
and Y-slope by performing two single line scans (one in X- and one in Y-direction) and
determining the respective slopes via line fitting, thus electronically compensating for
these measurement plane slopes (see Section 5.2: Adjusting the measurement plane (page
57) for details).
223
CHAPTER 17: ADVANCED SETTINGS
Auto slope
Automatically performs the same action as the “Adjust slope” button does. It adjusts the
slopes with each new “Start” of imaging.
Imaging modes
Scan mode
This parameter defines how the images are acquired and displayed:
– Continuous
The acquisition direction is reversed after each scan: from bottom to top and vice versa.
– Cont.Up
The acquisition direction is always from bottom to top.
– Cont.Down
The acquisition direction is always from top to bottom.
Measurement mode
This parameter defines how each imaging line is acquired and stored:
– Forward
During forward scan only (left to right in the image).
– Backward
During backward scan only (right to left in the image).
– Forw.&Backw.
During both forward and backward scan.
224
SPM PARAMETERS DIALOG
Rel. Tip-Pos
This parameter defines the distance by which the Tip is moved towards the sample from
the position that corresponds to the Setpoint. A negative setting will move the tip away
from the sample.
Chart automation
Auto. clear old chart data
Automatically clear the chart data from a measurement when a measurement is restarted
(either when a scan is restarted manually, or when a previous scan has finished and
measurement recommences as determined by the scan mode (see Scan mode (page 224).
Auto chart settings
Automatically adjusts the Z-Range of all Charts to encompass the measurement data. The
Z-Range will also be automatically updated during measurements.
17.9.2: Spectroscopy
Modulated parameter
Modulated output
This parameter defines the signal used to drive the spectroscopy (X-Axis). All possible
signals are recorded while this modulation output signal is changing its value from “Start
225
CHAPTER 17: ADVANCED SETTINGS
value” to “End value” (Y-Axis). The number of available modulated outputs depends on the
scan head and the number of installed controller modules. Possible values are: “Z-Axis”, “Tip
Potential” and the names of the User Outputs.
Start value / End value
The range over which the Modulated output is changed.
The “Spec Forward” data is measured from the Start to the End value, the “Spec backward”
data is measured in the opposite direction. The “Spec forward” data is always measured
before the “Spec backward” data. For spectroscopy as a function of distance (Z-axis
modulation), more negative values are further away from the sample whereas more
positive values go towards (or even into) the sample.
Relative to current value
When checked, the Start and the End values are added to the value the modulated output
had before starting the modulation:
• When the Tip Potential is modulated, the current value is the Tip voltage set in the Z-
Controller panel.
• When the Z-Axis is modulated, the current value is the sample surface height, as
measured using the Z-Controller output. Otherwise, the measurement Z-position is
given by the value of the Ref. Z-Plane in the Imaging Panel.
Modulation time
The time used to change the Modulated output from the Start to the End value.
Sequence Points
The number of Spectroscopy Measurements to be made in the sequence. If Sequence
points is set to one (1), a single point spectroscopy is performed. If Sequence value is two
or more, a line spectroscopy is performed. The spectroscopic measurements positions that
are equally distributed over the line defined by “X/Y-Pos from” and “X/Y-Pos to”.
X-Pos from / Y-Pos from
The XY-coordinates of the measured point in a spectroscopy measurement. Sets the XY-
coordinates of the starting point of the line in a spectroscopy measurement sequence.
X-Pos to / Y-Pos to
The XY-coordinates of the end-point of the line in a spectroscopy measurement sequence.
The from and to coordinates are more conveniently chosen using the point and line tools
in the Spectroscopy toolbar.
Measurement parameter
Data points
The number of data points measured while the Modulation output changes from Start to
End value. The data points are equally distributed over the modulation range.
226
SPM PARAMETERS DIALOG
Averages
The number of times the modulation is repeated to obtain an averaged spectroscopic
measurement. The measurement results of aborted modulations are discarded during
averaging.
Keep Z-Controller active
When checked, the Z-Controller will continue to change the Z-position to keep the tip–
sample interaction constant. This option is not available when the Modulated output is set
to the Z-Axis. This setting can for example be used to measure tip current as a function of
applied voltage while keeping the tip–sample force constant.
227
CHAPTER 17: ADVANCED SETTINGS
17.9.3: Lithography
Lithography modes
Lithography operating mode
Used to select the operating mode during lithography operation. The following options are
available:
– Static Force
– Dynamic Force
– STM (STM only)
Inactive pen mode
Action to be performed when the tip is moving from one end point to a new start point, in
case the end point and start point are not the same. The following options are available:
– Lift up tip
Only lift the tip (upper position of the Z-actuator of the scan head). No feedback will be
performed by the Z-Controller during travel to the new start point.
– Standard operating mode
Switch the Z-Controller operating mode back to the one selected in the “Operating
Mode Panel” during imaging. All values such as Tip speed, Tip voltage, Setpoint etc. will
228
SPM PARAMETERS DIALOG
temporarily chance back to the values used for imaging. The Z-Controller will be active
during travel to the new start point.
229
CHAPTER 17: ADVANCED SETTINGS
the phase signal is zero. When “Auto set” is enabled the phase reference is automatically set
after finishing the approach.
Tip Parameter
Tip Voltage
This parameter defines the potential to be applied to the tip. The voltage that can be used
lies between -10V and +10V.
Info:
With the STM scan head the sample is automatically connected to the ground of the
instrument. With AFM scan heads the sample has to be electrically connected to the
instrument chassis ground for accurate measurements.
230
SPM PARAMETERS DIALOG
17.9.5: Approach
Approach parameter
Approach Mode
For STM scan heads no selection is possible (approach is always performed step-by-step
through the stick-slip motion of the STM approach stage).
For AFM scan heads, two approach options are available for AFM:
– Continuous approach
Approach with continuous slow motorized stage movement until surface contact point
is reached. Z-Axis stays at Tip-Position during approach. This is the default approach
method, and was the only method available in the Nanosurf SPM Control Software
before version 3.
– Step-By-Step approach
Approach is performed by moving the motorized stage quickly over a distance that is
less than the scanner’s Z-range. During the movement of the motorized stage, the tip is
fully retracted. When the motorized movement is finished, the scanner extends the tip
along the Z-axis to probe for the sample surface. The approach is considered done when
the Setpoint (defined in the Z-Controller page of the SPM Parameter dialog, or in the Z-
Controller section of the Imaging panel of the Imaging window) has been reached. If it is
231
CHAPTER 17: ADVANCED SETTINGS
not reached within the Z-range of the scanner, the tip is again fully retracted and the next
motorized step is performed. This process of ‘step-and-probe’ is repeated until the
Setpoint has been reached (and approach is done). This approach method is considered
to be more ‘gentle’ to tip and sample and should be considered for very sharp tips and/
or very soft samples. In general, it does however take more time than Continuous
approach.
Max. Slope
This parameter defines the speed of extending the z-axis. This parameter is only available
in Step-By-Step approach mode. Slower speeds help to preserve sharp tips.
Max. Steps
This parameter defines the maximal duration of an automatic approach:
• In Continuous approach mode it defines the maximum time.
• In Step-By-Step approach mode it defines the maximum number of cycles.
Move Speed
This parameter defines the move speed during automatic approach and withdraw:
• In Continuous Approach mode this value should be small. If the approach is too fast, the
tip or the sample surface can be damaged. On the other hand, the motor will not move
when the move speed is too small.
• In Step-By-Step Approach mode should be around full speed. Lower values help to stop
in the z-range of the scanner. On the other hand the approach time increases.
Tip-Position
This value determines the Z-Position of the scanner when the approach motor stops. When
the Tip-Position is changed when the tip is already approached to the sample, the motor
will move the approach stage so, that the Z-Position of the tip becomes equal to the set Z-
Position. When a high resolution scanner is used, the Tip-Position before approach is set to
approximately +500 nm (advanced) by default. This compensates for the residual motion
of the approach stage that occurs after the approach motor has stopped.
232
SPM PARAMETERS DIALOG
17.9.6: Z-Controller
233
CHAPTER 17: ADVANCED SETTINGS
D-Gain
The strength of the Z-Controller reaction that is proportional to the derivative of the error
signal. Increasing the D-Gain decreases fast changes in the error signal, but also amplifies
high frequency noise.
Z-Controller options
Z-Feedback Mode
The following modes are available:
– Free Running
The Z-Controller is active.
– Freeze Position
The Z-Controller is not active, the scanner remains in its current Z-position.
– Stop and Clear
The Z-Controller is not active, the scanner is moved to the “Ref. Z Plane”, set in the
Imaging page of the SPM Parameter dialog. The Probe Status Light will blink green as
long as the Z-Controller is deactivated.
CAUTION
The tip may be damaged when the Z-Controller is not active during scanning. This will
happen when Ref. Z Plane is much lower than the current position of the tip, or when
the scan range contains large height differences.
Z-Feedback algorithm
The following algorithms are available:
– Standard PID
A standard PID controller is used for Z-Feedback.
– Adaptive PID
A standard PID controller is used for Z-feedback. In addition, the bandwidth of the
Topography measurement is adapted to the number of measured points per second. The
adaptive PID controller thus reduces noise in the measurement. However, topography
changes that happen faster than the time between two measured points are also lost.
This makes it more difficult to detect vibrations due to instability of the feedback loop.
These vibrations remain visible in the Current, Amplitude, or Deflection signal, however,
so always monitor these signals when optimizing Z-Controller settings, especially when
using the Adaptive PID.
Error Range
The range of the error signal used to control the Z-Position. The error signal is the difference
between the signal used for topography feedback and the current Setpoint. When the
value of “Error Range” is reduced, the resolution of the error signal is increased.
234
SPM PARAMETERS DIALOG
When the Signal Module A is installed, the Signal Access Page allows several parameters to
be configured.
235
CHAPTER 17: ADVANCED SETTINGS
Config
The “Config” button opens the User Signal Editor dialog (see Section 17.10: User Signal Editor
dialog (page 238)) to set the signal's name, range and unit.
User Controller
Enable Feedback loop
Activates an auxiliary feedback loop controller. This feedback loop controller uses the User
Input 1 as its signal input and drives the User Output 1 with it control value. It can be used
for custom measurements in all modes; Static as well as Dynamic mode (See “Help panel”
>> “AppNotes” >> “AN00031 — KPFM Operating Mode”).
Setpoint
Defines the Setpoint of the feedback controller. The controller keeps the User Input Signal
1 at this value.
I-Gain
Defines the I-Gain of the feedback controller.
Add output to Tip voltage
When checked, adds the feedback controller's output voltage to the Tip voltage output.
Inverted Feedback
Inverts the direction of the feedback response. Depending on the experiment to be
controlled, the feedback may have to increase or decrease its output value to force the User
Input signal to reach the Set Point.
Tip
If the overall gain of the experimental setup from its input to its output is positive,
feedback has to be non-inverted. If it has a negative gain, feedback has to be inverted.
236
SPM PARAMETERS DIALOG
3 Cable Cantilever
2
1
Figure 17-1: Tip Signal Mode schematic. Describes the electronics behind the three Tip Signal
modes.
237
CHAPTER 17: ADVANCED SETTINGS
The User Signal Editor dialog is used for editing the calibration of the User Input/Output
signals. The settings made in this dialog are stored in the active Scan Head calibration file.
Name
The name of the user signal. The entered name is used throughout the control software.
Unit
The base unit of the physical signal, without prefix (i.e. ‘m’, not ‘nm’ or ‘μm’).
Calibration
The physical signal values that correspond to the maximum and minimum signal voltages
should be entered here. Prefixes can be used.
238
VIBRATION FREQUENCY SEARCH DIALOG
239
CHAPTER 17: ADVANCED SETTINGS
automatically detect the cantilever resonance peak and adjust the operating frequency
accordingly. In some cases (mostly in Liquid environment) it is necessary for the user to
adjust the operating frequency manually (based on the Bode-plot result), because the
automatic algorithm is not able to find the right frequency peak.
For recording the Bode-plot, the SPM Control Software tunes the cantilever excitation from
a start frequency slowly up to a higher frequency. During this tuning (or sweeping) the
cantilevers amplitude and phase response is measured and saved. The result is plotted in
an Amplitude and a Phase versus Frequency chart. The Excitation amplitude is held
constant during the sweep. Therefore, any change in detector amplitude and phase signal
is a result of cantilever response.
Resonance frequencies vary greatly between cantilevers, depending on the cantilever’s
respective physical properties. Cantilever manufacturers provide information on the
resonance frequency range of each cantilever type. The SPM Control Software therefore
has a built-in database of commonly used cantilevers. Other cantilever can be manually
added by the user (see Section 17.13: Cantilever Browser dialog (page 245)). Based on this
database information, the control software selects the frequency sweep range.
After a coarse and a fine sweep, the optimal operating frequency is chosen close to the
resonance frequency of the cantilever.
IMPORTANT
The Phase measurement and chart plot is performed only if a Mode Extension Module
is installed in the SPM controller. The Phase is always measured, independent on the
selected Operating Mode.
240
VIBRATION FREQUENCY SEARCH DIALOG
selected and the excitation amplitude tuned to get the desired free vibration
amplitude at this frequency.
4. After the amplitude tuning, a final fine sweep is performed with the new excitation
amplitude.
At the end of the procedure, the new operating frequency is shown by the marker (both
graphically and in numbers). Additionally, the “excitation amplitude” parameter shows the
amplitude required to obtain the desired free vibration amplitude.
The user may now just leave the dialog and accept the new frequency using the “OK”
button, reset to previous values with the “Cancel” button, do manual adjustment with the
marker, or change amplitude settings and/or customize frequency sweep ranges through
use of the edit boxes and action buttons.
Capture
Stores a copy of the current Bode-plot in the History page of the Gallery panel. The data is
stored as a new measurement and remains open in the Document space of the SPM
Control Software.
Config...
Selected from the “Auto frequency set” button’s drop-down menu, this menu option opens
the Auto frequency config dialog (see Section 17.11.4: Auto Frequency Config dialog (page
243)).
241
CHAPTER 17: ADVANCED SETTINGS
The value of this frequency can be changed manually by typing in the desired value, by
using the arrow keys beside the parameter field, or by dragging the frequency marker line
in the charts as described above.
Free vibration amplitude
This parameter shows the actual amplitude at the current marker frequency. If the user
enters a new value, the excitation amplitude is adjusted to obtain a free cantilever
amplitude with this new value. A new frequency sweep is immediately performed and its
results displayed. If the desired amplitude could not be set, an error dialog is shown.
Zoom
A click on this button starts a new frequency sweep with a smaller frequency range than
currently displayed in the charts. The center of the sweep is at the markers position. The
new sweep range and center position is shown in the Fine sweep parameter section.
Coarse sweep
Start / End Frequency
These parameters defines the sweep range of a coarse frequency sweep. This range is
normally set large enough to search for the initial resonance peak. “Auto frequency set”
sets these parameters to the range found in the Cantilever database for the currently
selected cantilever.
Step Frequency
This parameter defines the frequency steps used during the sweep. Therefore, the number
of data points of the sweep is defined as:
242
VIBRATION FREQUENCY SEARCH DIALOG
243
CHAPTER 17: ADVANCED SETTINGS
The Scan Head Laser Alignment dialog displays the current position of the AFM laser spot
on the detector and the used laser power. It is only available when a Nanosurf FlexAFM scan
head is attached. The dialog is opened by clicking the “Laser align” button in the
Preparation group of the Acquisition tab.
Laser working point
This graph shows the efficiency of the laser detection system. A small white diamond-
shaped marker should be found anywhere within the green area, meaning that a sufficient
amount of the laser signal is being reflected by the cantilever and correctly makes it back
to the photodiode detector.
244
CANTILEVER BROWSER DIALOG
If the marker resides in the red area to the right, it will turn red too, signaling that too little
laser light is being detected. This can be caused by a misaligned cantilever, or by something
blocking the optical path of the laser.
If the marker resides in the red area to the left, too much light is being detected by the
photodiode detector, usually caused by strong environmental light. This condition is very
rare, however.
Laser position
This graphical area shows where the deflected laser beam hits the photodiode detector. A
green spot anywhere within the area enclosed by the dotted square means that the
cantilever deflection detection system (consisting of laser, cantilever, and detector) is
properly aligned. If the laser spot falls outside this area, it will become red, meaning that
the alignment does not allow proper measurements to take place. Usually this is caused by
a misaligned cantilever, which can be easily corrected. The position of the laser spot is also
given as a percentage of the maximum deflection the detector can identify.
The Cantilever Browser dialog displays the list of stored cantilever types in the database of
the SPM Control Software. In this database, many cantilevers of different manufactures are
saved with their typical physical properties. The Cantilever Browser dialog is opened via the
drop-down menu entry “Cantilever browser” of the “cantilever” button in the Preparation
group of the Acquisition tab.
245
CHAPTER 17: ADVANCED SETTINGS
New...
Opens the Cantilever Editor dialog for a new cantilever type. You can create new cantilever
types that are not defined in the default configuration. See section x.1 The Cantilever Editor
Edit...
Opens the “Cantilever Editor” dialog to modify the currently selected cantilever type. See
section x.1 The Cantilever Editor
Delete
Deletes the currently selected cantilever type.
Default
Sets all known cantilevers types back to their default factory settings. It also restores
deleted known cantilevers. After a software update, new known cantilevers are added. User
defined cantilevers are always kept as defined.
Attention:
Please note that scan heads with cantilever holders based on the Alignment Chip
technology can only be used with cantilevers possessing the following properties:
• The sensor chip must have grooves that fit onto the alignment chip.
• The cantilever should have a nominal length of 225 μm or more, and a top width of 40
μm or more (please distinguish between mean width specified by most
manufacturers and the actual top width of trapezoid-shaped cantilevers).
• The back of the cantilever must have a coating that reflects infrared light. Uncoated
cantilevers transmit much of the infrared light of the cantilever deflection detection
system.
246
CANTILEVER BROWSER DIALOG
Chart
Sets the deflection signal unit as displayed in the Chart area.
Setpoint
Sets the deflection signal unit to be used for entering the PID controller’s Setpoint.
The Cantilever Editor dialog allows and editing of existing cantilever types and the creation
of new cantilever types that are not defined in the default configuration.
Cantilever
Name
Name of the cantilever type. This name appears in the Cantilever type drop-down menu in
the Preparation group of the Acquisition tab. The list is sorted by the manufacturer's name.
Manufacturer
Name of the cantilever manufacturer.
247
CHAPTER 17: ADVANCED SETTINGS
Properties
Spring constant
This parameter defines the nominal spring constant of the cantilever type. This value is
used to calculate the correct force used for Z-Controller Setpoint and chart signals in all
operating modes that use the Static Force for Z-control.
IMPORTANT
The deflection calibration value of the current scan head calibration file is used to
calculate the force:
Force [N] = Spring Constant [N/m] * Deflection [m]
Cantilever length
The nominal length of this cantilever type (for information only, currently not used by the
control software).
Cantilever width
The mean width of this cantilever type (for information only, currently not used by the
control software).
Resonance frequency air / Resonance frequency liquid
The nominal resonance frequency of the cantilever type in air or in liquid measurement
environment. This frequency is used for calculation of the automatic coarse frequency
sweep search range (see Section 17.11: Vibration Frequency Search dialog (page 239)).
Q-factor air / Q-factor liquid
The apparent quality factor of the cantilever in air or in liquid measurement environment.
The higher the number, the sharper the peak. By default, this number is 500 in air and 5 in
liquid. The quality factor is used for calculation of the automatic fine frequency sweep
search range (see Section 17.11: Vibration Frequency Search dialog (page 239)).
248
ATS STAGE AND TSC 3000 DRIVER CONFIGURATION
CAUTION
Using an incorrect stage configuration will cause the stage to behave incorrectly. This
may cause damage to the system. Only change settings when instructed to do so by
Nanosurf support personnel.
To change the TSC 3000 driver configuration and to access further driver functionality, the
driver menu has to be opened. To do this:
> Right-click the icon in the Windows Notification Area.
From the opened TSC 3000 driver menu, the following options are
available:
Load configuration...
Allows a predefined or custom-saved configuration (“.stage” file) to be loaded into the
driver software.
Once you have changed the configuration, you must search for the reference position of
the stage (see Stage Position section (page 133) in Section 13.5: Stage panel).
249
CHAPTER 17: ADVANCED SETTINGS
CAUTION
Using an incorrect stage configuration will cause the stage to behave incorrectly. This
may cause damage to the system. Only change the settings when instructed to do so by
Nanosurf support personnel.
Axis calibration
The axis calibration determines the relationship between the number of steps that the
motor that drives a certain axis has made and the physical position coordinates of that axis.
The motor controller converts the Stepcount to the position according to the formula:
Enable
Allows individual axis (X, Y, Z, or phi) to be enabled or disabled. Depending on the type of
translation stage used and on the axes selected, the rotation axis “phi” may become
available.
250
ATS STAGE AND TSC 3000 DRIVER CONFIGURATION
Fullstep size
The amount of distance that the stage travels when the stepper motor makes one step.
Max Movement
The maximum distance that the stage moves.
Ministeps/Fullstep
The number of steps in which a full stepper motor step can be sub-divided by the
controller.
Ref Switch Pos.
“At the end” must be checked when the limit switch is located at the end of the translation
stage’s movement range (i.e., the position most distant to the motor). When checked, the
controller will move the stage towards the end of the range until it finds the position of the
reference switch of the stage.
Position value
“Inverted” must be checked when the axis coordinate should decrease (rather than
increase) when the stage moves towards the end.
Speed Settings
Maximal move speed:
In (Ministep)counts/second. If the Maximal move speed is set too high, the stepper motor
in the stage may fail to finish a step before the TSC 3000 controller commands it to make
the next one. As a consequence the stage will not be in the correct position after moving.
251
CHAPTER 17: ADVANCED SETTINGS
The TSC 3000 COM Port Configuration dialog sets the Serial Interface/COM port that the
TSC 3000 Controller is connected to. The dialog is opened via the TSC 3000 Driver menu
option “Set COM Port” (see Section 17.14: ATS Stage and TSC 3000 driver configuration (page
248)).
252
Quick reference
Quick reference
Dialogs Ribbon tabs/groups
About ............................................................... 202 Acquisition .....................................................128
Auto Frequency Config ............................. 243 Approach ...............................................130
Cantilever Browser ...................................... 245 Imaging ..................................................131
Cantilever Editor .......................................... 247 Lithography ..........................................160
Chart Properties ........................................... 174 Preparation ...........................................129
COM Port Configuration ........................... 252 Scripting .................................................132
Controller Configuration .......................... 220 Analysis ............................................................186
Digital Video Properties ............................ 140 Correction ..............................................189
File Rename ................................................... 185 Filter .........................................................193
Laser Alignment ........................................... 244 Measure ..................................................187
Layer Editor .................................................... 157 Report .....................................................196
Mask Editor .................................................... 183 Roughness .............................................191
Move Stage To .............................................. 134 Tools ........................................................195
Object Editor ................................................. 159 File .....................................................................203
Options ............................................................ 206 Chart arrangement ............................205
Pixel Graphic Import ................................... 165 Exit ...........................................................206
Scan Axis Correction ................................... 218 Export chart/document ....................204
Scan Head Calibration ............................... 215 Open ........................................................203
Scan Head Diagnosis .................................. 218 Options ...................................................206
Scan Head Selector ..................................... 214 Parameters ............................................205
SPM Parameters ........................................... 221 Print/Print preview .............................205
Stage Configuration ................................... 250 Save/Save as .........................................204
User Signal Editor ........................................ 238 Settings ...........................................................212
Vector Graphic Import ............................... 163 Hardware ...............................................213
Vibration Frequency Search .................... 239 Scan Head ..............................................213
View ..................................................................118
Operating windows
Panels ......................................................118
Imaging ........................................................... 121
Window ..................................................119
Lithography ................................................... 151
Workspace .............................................118
Spectroscopy ................................................ 145
Toolbars
Panels
Analog Video Camera ................................136
Data Info ......................................................... 170
Data Info ..........................................................171
Gallery .............................................................. 181
Digital Video Camera ..................................137
Imaging ........................................................... 123
Gallery ..............................................................182
Lithography ................................................... 154
Imaging ...........................................................127
Online .............................................................. 142
Lithography ...................................................161
Spectroscopy ................................................ 148
Quick Access ..................................................203
Stage ................................................................ 133
Spectroscopy .................................................149
Video ................................................................ 135
254