Bestx Series

This document provides installation and user instructions for the BEST-X-DC X-ray unit. It includes sections on general aspects, installation and use, operation, plates positioning, servicing, and safety. The manual aims to help users safely and efficiently obtain dental radiographs using the X-ray diagnostic equipment.

Uploaded by

sampaulsinghbme
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
643 views76 pages

Bestx Series

This document provides installation and user instructions for the BEST-X-DC X-ray unit. It includes sections on general aspects, installation and use, operation, plates positioning, servicing, and safety. The manual aims to help users safely and efficiently obtain dental radiographs using the X-ray diagnostic equipment.

Uploaded by

sampaulsinghbme
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 76

Manuale Doc.

HBI070-2

Manuale di Installazione e uso EVOSTYLE N.G. Mod. MQI006-0

BEST-X LINE INSTALLATION AND USER MANUAL


Mod. MQI006-0 Doc. HBE220-1
MANUAL Doc. HBE220-1

Installation and User Manual BEST-X-DC WITH Mod. MQI006-0


TIM-X

INDEX

GENERAL ASPECTS ..........................................................................................................................1

Introduction ....................................................................................................................................... 1

Descriptions ....................................................................................................................................... 1

Technical data ....................................................................................................................................... 5

Other Data ....................................................................................................................................... 6

INSTALLATION AND USE...................................................................................................................6

Use Conditions ....................................................................................................................................... 6

General symbols ....................................................................................................................................... 7

Symbols for transportation and storage................................................................................................... 9

Information for installation ....................................................................................................................... 10

Utensils and tools required for installation (not supplied) ........................................................10

Electrical indications .....................................................................................................................10

Wiring and connection conductors sections ............................................................................10

Installation ..................................................................................................................................... 12

Instructions for installation in wall mode.....................................................................................12

Settings ...........................................................................................................................................24

Springs adjustment of the double pantograph arm ............................................................................ 25

Instructions for installation in column mode .......................................................................................... 26

Palmtop operation .......................................................................................................................32

Operation ..................................................................................................................................... 36

Matching Screen ..................................................................................................................................... 42

Display Options Screen ............................................................................................................................ 43

Calibration Screen .................................................................................................................................... 44

Functioning e time scale: ......................................................................................................................... 45


MANUAL Doc. HBE220-1

Installation and User Manual BEST-X-DC WITH Mod. MQI006-0


TIM-X

The Monoblock Does Not Emit X-Rays .......................................................................................51

Procedures for the filament current measures:.........................................................................54

Measurement of the filament current ........................................................................................54

Plates positioning ..........................................................................................................................57

Servicing of the x-ray unit.............................................................................................................62

Extraordinary maintenance.........................................................................................................63

Thermal Curves ..................................................................................................................................... 69

Emission and filament characteristics .................................................................................................... 69

Safety aspects ..................................................................................................................................... 70

Possible drawbacks found in the radiographs results .......................................................................... 71

Clear image ..................................................................................................................................... 71

Recommendations .......................................................................................................................72
Manual Doc. HBE220-1

Installation and User Manual BEST-X-DC WITH Mod. MQI006-0


TIM-X

GENERAL ASPECTS

Introduction
Dear Customer,

Thank you for the preference granted to our product. We invite you to attentively read the
present instructions that will help you to get the maximum of your X-ray diagnostic
information with minimal x-ray use.

This manual has the purpose to provide the User with instructions for proper, safe and
efficient operation.

The equipment must be used in accordance with the procedures contained in the manual
and never for purposes other than those specified herein.

The User is responsible for what concern the fulfilments in legal matters facing installation
and equipment functioning.

The plant can only be used by medical personnel in possession of the related licenses
enablers and aware of the risks associated with the use of ionizing radiation sources. The use
of X-ray sources for purpose of medical diagnostics is subject to specific authorizations
and/or communications to the Authorities responsible for vigilance. The User is responsible
for the use unauthorized of the plant. The User of the X-ray plant for dental complementary
radiology is also required, without exception, to observe the regulations governing the
safety of exposure to ionized radiations sources for workers, for member of the public, for
population and patients.

If the equipment is not operated correctly or it is not made proper maintenance, the
manufacturer can not be held responsible for any breakages, lesions and mal functioning.

Descriptions
“CAUTION: Do not modify this equipment without the manufacturer’s permission.”

BEST-X-DC radiological unit is an equipment designed to obtain intraoral dental radiographs


arranged to be used with acquisition systems (conventional films, phosphor plates and
videoradiografie).

According to Directive 2007/47/CE (Legislative Decree 37/2010) it is classified in CLASS IIb.

NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO)
Italy phone and fax + 39.011. 781994 [email protected] 1
www.newliferadiology.it
Manual Doc. HBE220-1

Installation and User Manual BEST-X-DC WITH Mod. MQI006-0


TIM-X

The unit is manufactured in accordance with the actual international standards on the
protection of ionizing radiation, electrical safety, mechanical safety and electromagnetic
compatibility for electromedical equipment.

The equipment consists of the following parts:

Monoblock

BEST-X-DC Monobloc require the use of X-


ray tubes CEI OX/70-3

BEST-X-DC complies with EU directives on


electromagnetic compatibility.

However, it may be appropriate to avoid


installing the equipment in the immediate
vicinity of other electrical equipment with
which they could generate
electromagnetic fields of mutual
disturbance. It is also important to avoid
using electrical appliances (eg.
electrosurgery, cell phones, etc…) in the
immediate vicinity of the apparatus during
its use.

Articulated support double pantograph with wall support or mobile

The monoblock is directed within the


installation area by an articulated arm
double pantograph; this arm, coupled to the
wall support through an extension cable of
variable size (mm 400-800-1100), has a
maximum extension which varies between
1730 mm and 2430, depending on the
extension cord used.

NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO)
Italy phone and fax + 39.011. 781994 [email protected] 2
www.newliferadiology.it
Manual Doc. HBE220-1

Installation and User Manual BEST-X-DC WITH Mod. MQI006-0


TIM-X

Palmtop

Palmtop of the BEST-X-DC allows


managing the exposure times and
adjustment of kV(60/70) and mA (4/7) and
at the same time it guarantees maximum
safety in the use of X-ray tubes for
electromedical purposes for intraoral
diagnostic;

The management of the palmtop occurs


through the use of the RF radio control;

The control panel of the palmtop is


provided with touchscreen display so as to
facilitate its use;

The functioning of the palmtop is


managed by RF radio control to vigilant
device better known as dead man’s
device, in such a way as to guarantee
maximum safety for both, the operator
and the patient;

The palmtop does not permit use in the X-


ray scopy;

The time scale set by the values of the


factory complies with scale R10 relative to
the standard EN60601-2-7 (IEC 60601-2-7);

The exposure times shown on the display of


the timer are expressed in ms.

The maximum time that can be set is 1 s


(the display will read 1000) while the
minimum time is 20 ms. (the display will
read 0020);

NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO)
Italy phone and fax + 39.011. 781994 [email protected] 3
www.newliferadiology.it
Manual Doc. HBE220-1

Installation and User Manual BEST-X-DC WITH Mod. MQI006-0


TIM-X

BEST-X-DC has a unique safety system against short circuits or malfunction of the unit, this
system is the automatic fuse THERMOSWITCH.

This device intervenes in case of conduction for a time longer than 6 sec. In this situation
it blocks the continuity between the control unit and the monoblock, thus preventing the
emission of anomalous X-ray.

In case of intervention of the THERMOSWITCH, this will be replaced, this operation requires
a service call to technical assistance.

Drawings, schematics, components lists, instructions for repairs:


New Life Radiology is committed to providing, upon request, drawings, circuit diagrams,
component parts lists, instructions, or other information that can serve for qualified technical
personnel, to perform the repair of any parts that can be repaired.

The Manufacturer reserves the right to make changes at any time without notice.

NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO)
Italy phone and fax + 39.011. 781994 [email protected] 4
www.newliferadiology.it
Manual Doc. HBE220-1

Installation and User Manual BEST-X-DC WITH Mod. MQI006-0


TIM-X

Technical data
Classification: Electro medical equipment of Class I

With part applied type B

Head: Monoblock with potential constant 100KHz


model CEI OX/70-3

Type: 60-70 kVp selectable (±5%)

Feeding voltage: 230 V~ (50 Hz) monophase

Maximun Power absorbed from the net: 0,800 KW

Cone diameter: 60 mm

Max symmetrical rays field: Ø60 mm to SSD 200 mm

Anodic current: 4 mA – 6.5 mA selectable

Anodic voltage: 60, 70 kV selectable

Emission time: from 20 mS to 1000 mS scale R10

Exposition time: predefined

Max anodic voltage: 70 kV

Reference current time: 7mAs to 70 kV, 7mA, 1 s

Max supplied power: 0,49 kW a 70 kV, 7mA

Maximum apparent resistance of the power supply: 0.44 

Weight:

Wall model 27 kg

Mobile model 40 kg

Accuracy Load Factors


Voltage accuracy ±5%

Current accuracy ±3%

Time accuracy ±5%

Dose accuracy ±5%

NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO)
Italy phone and fax + 39.011. 781994 [email protected] 5
www.newliferadiology.it
Manual Doc. HBE220-1

Installation and User Manual BEST-X-DC WITH Mod. MQI006-0


TIM-X

Other Data
Maximum current automatic switches for the feeding Magnetic thermal switch from 10 A (CEI
net 23-3)

High voltage measure method: Not invasive method

Current measure method in the radiogenic pipe: See page 54

Charge application time determination method: Not invasive method

Interposition aluminium filter between rays window and Al 1 mm (AIP99,9 UNI3567)


cone collimator:

INSTALLATION AND USE


Use Conditions

The equipment is designed for continuous operation with intermittent charge.

The operating times are with intermittent charge with a duty cycle of 1.30 (for each
working period will correspond to 30 pause)

Classification according to Directive 2007/47 / EC Class IIb


(Legislative Decree no. 37/2010)

Protection against electrical dangers: Class I

Protection degree against the direct and indirect Equipment with parts applied type B
contacts:

Protection degree against water penetration: Common equipment IPX0

Use safety degree in presence of inflammable Equipment not suitable to an use in


anaesthetic mixture: presence of an inflammable anaesthetic
mixture with air or with oxygen or with
nitrogen protoxide

Use conditions: Equipment for continuous operation with


intermittent charge

Installation: Permanent e mobile

NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO)
Italy phone and fax + 39.011. 781994 [email protected] 6
www.newliferadiology.it
Manual Doc. HBE220-1

Installation and User Manual BEST-X-DC WITH Mod. MQI006-0


TIM-X

General symbols

Caution: consult annexed documentation

Ionizing radiation

Apparatus with Type B part applied

Earth protection

O Switch open(disconnected from mains supply)

I Switch close(connected to mains supply)

Red point, placed on the cover of the unit, indicate focal spot

Alternating current

Radiation emission symbol

Follow annex instructions

Symbol in conformity with European Directive 2002/96/Ec (Weee).

NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO)
Italy phone and fax + 39.011. 781994 [email protected] 7
www.newliferadiology.it
Manual Doc. HBE220-1

Installation and User Manual BEST-X-DC WITH Mod. MQI006-0


TIM-X

Manufacturer

Symbol in conformity with Community Legislation. The symbol is followed by a


number which identifies the notified organism which certifies and monitors
such compliance

REF Product Code

SN Equipment Serial Number. To be used for all communication with


manufacturer / service technician

TUBE X-ray Tube Serial number

NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO)
Italy phone and fax + 39.011. 781994 [email protected] 8
www.newliferadiology.it
Manual Doc. HBE220-1

Installation and User Manual BEST-X-DC WITH Mod. MQI006-0


TIM-X

Symbols for transportation and storage

27 kg

NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO)
Italy phone and fax + 39.011. 781994 [email protected] 9
www.newliferadiology.it
Manual Doc. HBE220-1

Installation and User Manual BEST-X-DC WITH Mod. MQI006-0


TIM-X

Information for installation


PLASTIC MATERIAL PLUGS NOT ALLOWED

Utensils and tools required for installation (not supplied)


 1 multi meter

 1 meter

 1 fixed key from 13 mm

 1 nandle spanner from 5,5 mm

 1 set of Allen spanner

 1 spirit level

 1 plastic mallet

 1 percussion drill with points from Ø 3 to Ø 13 mm

 1 thin screwdriver for electrical connection

 1 medium screwdriver

 1 net feeding cable with three wires (2 conductors + 1 ground) of section 1,5 mmq for
a reduced length of not over 40 m (for superior lengths use the section cable
corresponding to that indicated from the Country actual standards).

Electrical indications
All works for the feeding electric plant must be performed in conformance with the actual
reference standards for plants and locals for medical use. (CEI 64-8-710).

A feeding of 230 V~ to 50 Hz. is required. For the phase conductors, neutral and ground, the
minimum section must be of 1,5 mmq of copper. It recalls the necessity to perform the
ground connection as required by current legislation.

The plant must be performed and tested by qualified personnel.

The warranty excludes damages caused by erroneous connection.

Wiring and connection conductors sections


A three conductors cable (3x1,5 mmq) is expected between the net switch and the
equipment. For the assembled versions and in all cases of separate installation from the
control unit, a second cable, always 3 x 1.5 sq mm, shall be provided between the control
unit and the support. Please note that the radiological groups are supplied without a plug

NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO)
Italy phone and fax + 39.011. 781994 [email protected] 10
www.newliferadiology.it
Manual Doc. HBE220-1

Installation and User Manual BEST-X-DC WITH Mod. MQI006-0


TIM-X

for leaving the buyer free to connect to equipment already on site.

Efficient grounding is a first necessary condition for the proper functioning of the system,
respecting the symbols indicated for connection:

F – Line (brown) N – Neutral (blue) T – Ground (yellow-green)

Point A grounded power cord; Point B output external lamp

Check carefully all connection cables, plugs and line contacts.

Remember that the suppliers must support a current of at least 10A.

The connection to the external lamps must be done by connecting the output LAMP
indicated with the letter A in the figure above. The lamp must be of value 30W 230-240 Vac.

NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO)
Italy phone and fax + 39.011. 781994 [email protected] 11
www.newliferadiology.it
Manual Doc. HBE220-1

Installation and User Manual BEST-X-DC WITH Mod. MQI006-0


TIM-X

according to the actual reference standard for plants and locals for medical use (IEC 64-8-
710).

Installation
“CAUTION: To avoid the risk of any electric shock, this device must be connected only to
power mains gifted with ground protection.”

The installation is provided in both wall and mobile version on column.

Instructions for installation in wall mode


The system in wall version consists of the components shown in the following legend:

A wall plate
B extension
C double pantograph arm
D palmtop
E monoblock

NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO)
Italy phone and fax + 39.011. 781994 [email protected] 12
www.newliferadiology.it
Manual Doc. HBE220-1

Installation and User Manual BEST-X-DC WITH Mod. MQI006-0


TIM-X

Quotas with relative useful space – EXTENSION FROM mm 400

NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO)
Italy phone and fax + 39.011. 781994 [email protected] 13
www.newliferadiology.it
Manual Doc. HBE220-1

Installation and User Manual BEST-X-DC WITH Mod. MQI006-0


TIM-X

Quotas with relative useful space – EXTENSION FROM mm 800

NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO)
Italy phone and fax + 39.011. 781994 [email protected] 14
www.newliferadiology.it
Manual Doc. HBE220-1

Installation and User Manual BEST-X-DC WITH Mod. MQI006-0


TIM-X

Quotas with relative useful space – EXTENSION FROM mm 1100

NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO)
Italy phone and fax + 39.011. 781994 [email protected] 15
www.newliferadiology.it
Manual Doc. HBE220-1

Installation and User Manual BEST-X-DC WITH Mod. MQI006-0


TIM-X

Wall plate installation


 This equipment is supplied with a plate for mural fixing (A), closed by a plastic cover
(fig.1). Remove the plastic cover.

fig. 1 fig. 2

 Define the position on the wall of the WALL PLATE (A) with respect to the amplitude
of the working field chosen and to the size of the structure that the same can assume
right or left of the axis of the wall plate when it is inactive.

 Plot on the wall the position of the six holes making sure to check the perpendicularity
with a plumb line. If the electric plant is recessed, track also the corresponding hole
(Fig. 2)

 Drill six holes on the wall starting with the tip of  7, enlarging gradually. This to not
demolish the stability and maintain under control the interaxes. For walls of full or
holed or cement bricks use metallic plugs preferable of  12 equipped with
unmissable female grain and separated screw  6 with hexagonal head and washer.

 For types of wall of insufficient reliability it is necessary to resort to the construction of a


reinforcement to be defined case by case.

 Apply the wall plate and bring it closer parallel to the wall by tightening the 6 screws
alternately, if the wall is not perfectly flat put adequate thicknesses so as not to
deform the wall plate.

NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO)
Italy phone and fax + 39.011. 781994 [email protected] 16
www.newliferadiology.it
Manual Doc. HBE220-1

Installation and User Manual BEST-X-DC WITH Mod. MQI006-0


TIM-X

Extension installation
 Extract the SEGER from the extension (B), and enter it in the wall plate (A), as shown
in fig. 3

fig. 3

NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO)
Italy phone and fax + 39.011. 781994 [email protected] 17
www.newliferadiology.it
Manual Doc. HBE220-1

Installation and User Manual BEST-X-DC WITH Mod. MQI006-0


TIM-X

 Replace the SEGER, unhook the front cover and remove the lower strip cover strips
(fig. 4)

fig. 4

NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO)
Italy phone and fax + 39.011. 781994 [email protected] 18
www.newliferadiology.it
Manual Doc. HBE220-1

Installation and User Manual BEST-X-DC WITH Mod. MQI006-0


TIM-X

Double pantograph arm installation


 Enter the double pantograph arm (C) in the extension (B) and route the cable as
shown in Figure 5. Close the extension repositioning the lower strip cover strips.
Connect the cable from the double pantograph arm to output X-RAY TUBE and the
ground wire IN in its slot.

fig. 5

NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO)
Italy phone and fax + 39.011. 781994 [email protected] 19
www.newliferadiology.it
Manual Doc. HBE220-1

Installation and User Manual BEST-X-DC WITH Mod. MQI006-0


TIM-X

Monoblock installation

 Before inserting the monoblock (E) it is essential to work in safety by


opening the double pantograph arm (C) as shown in fig. 6, to avoid that it
snaps abruptly since it has the springs charged and calibrated to support
the weight of the monoblock.

fig. 6

NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO)
Italy phone and fax + 39.011. 781994 [email protected] 20
www.newliferadiology.it
Manual Doc. HBE220-1

Installation and User Manual BEST-X-DC WITH Mod. MQI006-0


TIM-X

 Unscrew the screw from the front tubular carrier of the double arm pantograph (C)
by lifting the cylinder and remove the half-moon(fig. 7)

fig. 7

NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO)
Italy phone and fax + 39.011. 781994 [email protected] 21
www.newliferadiology.it
Manual Doc. HBE220-1

Installation and User Manual BEST-X-DC WITH Mod. MQI006-0


TIM-X

 Insert Monoblock (E) in the double pantograph arm (C) as shown in Fig. 8

fig. 8

Reinsert the half-moon in its slot. Lower the cylinder, securing the double arm
pantograph with the screw.

NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO)
Italy phone and fax + 39.011. 781994 [email protected] 22
www.newliferadiology.it
Manual Doc. HBE220-1

Installation and User Manual BEST-X-DC WITH Mod. MQI006-0


TIM-X

fig. 9

NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO)
Italy phone and fax + 39.011. 781994 [email protected] 23
www.newliferadiology.it
Manual Doc. HBE220-1

Installation and User Manual BEST-X-DC WITH Mod. MQI006-0


TIM-X

Settings
After installing the equipment, make a dynamic test to ensure that the movements of the
assembly are properly calibrated.

Clutch screws adjustment


 If necessary adjust the clutch screws of the wall plate (A) and the extension (B) as
shown in fig.10.

fig.10

After making the adjustments necessary to close the wall plate with the respective cover,
making sure to replace the lower back lining strip and the front cover of the extension.

NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO)
Italy phone and fax + 39.011. 781994 [email protected] 24
www.newliferadiology.it
Manual Doc. HBE220-1

Installation and User Manual BEST-X-DC WITH Mod. MQI006-0


TIM-X

Springs adjustment of the double pantograph arm


Do not intervene personally in the arm’s settings, have this operation made exclusively by
specialized personnel only.

 In the case in which the double pantograph arm does not remain stationary in all
positions wanted by the user, it means that requires some adjustments. The
calibration must be performed as follows: with reference to fig. 11, if it was found that
the section A is spread when the double pantograph tends to go back, it means
that the spring is too tight and in this case we must unscrew slightly the push-spring
nut through the socket wrench supplied. To do this, tilt the section A of about 15 °
and insert the socket wrench into the hole 1 and unscrew slightly.

 If on the contrary it is found that stretching the double pantograph, the stretch A
tends to fall forward, it means that the spring must be slightly loaded, then as before,
tilt the stretch A of 15 ° and insert the socket wrench into the hole 1 and screw slightly.

 This procedure is identical for the section B in which the adjusting key must be
inserted in the hole 2.

fig.11

NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO)
Italy phone and fax + 39.011. 781994 [email protected] 25
www.newliferadiology.it
Manual Doc. HBE220-1

Installation and User Manual BEST-X-DC WITH Mod. MQI006-0


TIM-X

Instructions for installation in column mode


This device is not intended for mobile use, the wheels are intended for a better positioning of
the main appliance above the patient.

The system in column mode consists of the components shown in the following legend:

Detail E

A cart
B vertical support
C monoblock
D double pantograph arm
E fuse box
F palmtop

NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO)
Italy phone and fax + 39.011. 781994 [email protected] 26
www.newliferadiology.it
Manual Doc. HBE220-1

Installation and User Manual BEST-X-DC WITH Mod. MQI006-0


TIM-X

Quotas with appropriate useful space

Fig. 12

NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO)
Italy phone and fax + 39.011. 781994 [email protected] 27
www.newliferadiology.it
Manual Doc. HBE220-1

Installation and User Manual BEST-X-DC WITH Mod. MQI006-0


TIM-X

Column assembly
 The first procedure to be carried out is to assemble the carriage (A), this is done by
mounting the two tubular "1" and "2" on the base "3" using four screws 8x20, as shown
in fig. 13

Fig. 13

 Once assembled the cart (A) is possible to fix on it the vertical support (B) using 4
screws 6x20 as shown in fig. 14

fig. 14

NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO)
Italy phone and fax + 39.011. 781994 [email protected] 28
www.newliferadiology.it
Manual Doc. HBE220-1

Installation and User Manual BEST-X-DC WITH Mod. MQI006-0


TIM-X

 Insert the two handles provided into the hole as in fig. 15

fig. 15

Installing double pantograph arm


 Remove the screws holding the card fuse holder (E) to the vertical support (B) as
shown in Fig. 16

fig. 16

NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO)
Italy phone and fax + 39.011. 781994 [email protected] 29
www.newliferadiology.it
Manual Doc. HBE220-1

Installation and User Manual BEST-X-DC WITH Mod. MQI006-0


TIM-X

 Insert the double pantograph arm (D) in the vertical support (B) as shown in fig. 17

fig. 17

NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO)
Italy phone and fax + 39.011. 781994 [email protected] 30
www.newliferadiology.it
Manual Doc. HBE220-1

Installation and User Manual BEST-X-DC WITH Mod. MQI006-0


TIM-X

 Connect the cable coming from the double pantograph arm (D) to the switch
board fuse holder (E) as shown in figure 18

fig. 18

NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO)
Italy phone and fax + 39.011. 781994 [email protected] 31
www.newliferadiology.it
Manual Doc. HBE220-1

Installation and User Manual BEST-X-DC WITH Mod. MQI006-0


TIM-X

Monoblock installation

 To install the monoblock refer to the procedure described previously for the wall
version from page 20.

 To adjust the springs of the double pantograph arm refer the procedures for the wall
version on page 25.

Palmtop operation

 The palmtop of the device allows the management of the exposure times of the
monoblock and the regulation of kV (60/70) and mA (4/7), at the same time
guarantees maximum safety in the use of X-ray tubes for electromedical purposes for
radiological diagnosis;

 Palmtop is provided with touchscreen display so as to facilitate its use;

 The palmtop functioning is managed by radio-frequency remote control to dead


man’s device in such a way as to ensure maximum safety for both the operator and
patient;

 The palmtop does not allow the use of X-ray in scopy;

 The time range set by the factory values complies with R10 range relative to
standard EN60601-2-7 (IEC 60601-2-7);

 The exposure time displayed on the timer are expressed in ms

 The maximum time that can be set is 1 s (the display will read 1000) while the
minimum time is 20 ms. (the display will read 0020);

NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO)
Italy phone and fax + 39.011. 781994 [email protected] 32
www.newliferadiology.it
Manual Doc. HBE220-1

Installation and User Manual BEST-X-DC WITH Mod. MQI006-0


TIM-X

Introduction to the palmtop:


Thru the palmtop it is possible to handle the connected endoral monoblock in the best way
possible.

Front view
Manual time increase
Time schedule
setting
Beam emission Battery level
time selection
Type of tooth selection emssone ragg x
Manual time decrease

Menu

Pre-Start

mA Selection

kV Selection

Type of film selection

Type of patient selection

NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO)
Italy phone and fax + 39.011. 781994 [email protected] 33
www.newliferadiology.it
Manual Doc. HBE220-1

Installation and User Manual BEST-X-DC WITH Mod. MQI006-0


TIM-X

Menu It allows to browse thru the various menus of the system


indicated as follows:

It allows to go back to the initial


configuration screen;

It allows to modify the exposure


parameters;

It allows to associate the hand-held


to a receiver;

It allows to modify some functions of


the display and to speed up the
self-calibration;

Type of tooth selection It allows to select the type of tooth;

Type of patient selection It allows to select the type of patient;

X-ray emission selection It allows to set the exposure time;

Type of film selection It allows to select the type of film or RVG;

kV selection It allows to select kV (60 o 70);

mA selection It allows to select mA (4 o 7);

Pre-Start It allows to implement the monoblock for the x-ray emission;

Manual time increase It allows the manual increase of the exposure time;

Manual time decrease It allows the manual decrease of the exposure time;

Time schedule It allows customizing the time schedule;

Battery level It shows the battery charge level;

NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO)
Italy phone and fax + 39.011. 781994 [email protected] 34
www.newliferadiology.it
Manual Doc. HBE220-1

Installation and User Manual BEST-X-DC WITH Mod. MQI006-0


TIM-X

Rear view

USB

On/Off button
Reset
button

On/Off button It allows turning on/turning off the hand-held or put it on


standby;

Reset button It allows resetting the hand-held;

USB Micro USB slot for recharging;

NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO)
Italy phone and fax + 39.011. 781994 [email protected] 35
www.newliferadiology.it
Manual Doc. HBE220-1

Installation and User Manual BEST-X-DC WITH Mod. MQI006-0


TIM-X

Operation

Environmental working conditions from 0 to 40 C°.

After possibly plugging in the battery, push and release the on/off button, the display will
light up and show the “New Life Radiology” logo, the software version and, ultimately, the
exposure parameters screen.

As of this moment, after approximately a ten second inner processing time, it is possible to
put the system on standby pushing and releasing the button. To exit the standby mode and
to get back to the latest screen, repeat the same operation.

The system will enter the saving mode also after a period of time customizable through a
further screen and, next time, it will turn off automatically.

To turn off the display, keep the on/off button pressed for as long as 6 seconds: differently
from the standby mode, where the display turns off gradually, in this case, the display will be
suddenly turning off.

The touch-screen display will allow the user to interact with the interface through virtual
buttons and sliders as well as to read all the information regarding configurations, feedback
and mistakes.

Below, all the screens and all of the actually implemented functions will be explained in
detail.

NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO)
Italy phone and fax + 39.011. 781994 [email protected] 36
www.newliferadiology.it
Manual Doc. HBE220-1

Installation and User Manual BEST-X-DC WITH Mod. MQI006-0


TIM-X

X-Ray Configuration Screen

In the above screen, it is possible to set all of the parameters for the next exposition, in
particular the exposure time, in line with the tooth, animal and film, the kV and the mA.

Pressing on the tooth, on the animal, the user selection will be “surrounded” by a green line,
whereas, to change the film (or RGV digital sensor), the kV and mA, it will be sufficient to
press the desired button just one time.

In case you suppose that the times are wrong, it is also possible to change them according
to the current configuration with the + and - buttons and save the new time pushing SET: the
new time is saved permanently in the memory until a new parameter reset occurs.

The charge level of the battery always shows.

In order to access another function push the Menu button where as to start the exposition
just select the X-ray icon (Pre-Start).

NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO)
Italy phone and fax + 39.011. 781994 [email protected] 37
www.newliferadiology.it
Manual Doc. HBE220-1

Installation and User Manual BEST-X-DC WITH Mod. MQI006-0


TIM-X

X-Ray Pre-Start

In this screen the hand-held seeks the receiver and, once found it, programs it making
“Configuration done” appear.

At this point it is possible to do the x-ray pressing the X-Ray button and keeping it pressed.

If the button is pressed prior to the configuration the x-ray does not occur and the button
releasing is requested, if instead, the receiver is not found, after a certain period of time it is
necessary to exit such screen.

To exit, in any case, it is sufficient to press the Cancel button.

To carry out an X-ray, you need to keep the X-Ray button pressed until the cooling process
screen appears.

NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO)
Italy phone and fax + 39.011. 781994 [email protected] 38
www.newliferadiology.it
Manual Doc. HBE220-1

Installation and User Manual BEST-X-DC WITH Mod. MQI006-0


TIM-X

Cooling Screen And Report

In the above screen, the bar indicates a percentage waiting time, in order to cool down
the head of the instrument and it is not possible to exit until the 0% is reached. Under the
please wait request, a general exposition report can be seen, that is to say whether the
exposition was successful or it was interrupted. Once the count is through, it is possible to
exit and to go back to the parameter configuration screen or to have a visual of the
feedback graph of the electrical current and voltage as they are read by the receiver on
the inverter generator.

NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO)
Italy phone and fax + 39.011. 781994 [email protected] 39
www.newliferadiology.it
Manual Doc. HBE220-1

Installation and User Manual BEST-X-DC WITH Mod. MQI006-0


TIM-X

Graphical Report Screen

In the above screen you can press the KV button and have a visual of the voltage (yellow
line) and/or the mA button just to have a visual of the electrical current (blue line).

NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO)
Italy phone and fax + 39.011. 781994 [email protected] 40
www.newliferadiology.it
Manual Doc. HBE220-1

Installation and User Manual BEST-X-DC WITH Mod. MQI006-0


TIM-X

Menu Screen

From here you can go back to the configuration screen pushing Radiographic or you can
access various functions:

- Parameters: The screen parameter is accessible by New Life Radiology personnel only

- Matching: to connect the hand-held to a receiver

- Disp.Options: to modify some functions of the display and to access the self-calibration
mode.

NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO)
Italy phone and fax + 39.011. 781994 [email protected] 41
www.newliferadiology.it
Manual Doc. HBE220-1

Installation and User Manual BEST-X-DC WITH Mod. MQI006-0


TIM-X

Matching Screen

In this screen you can connect the hand-held to a new receiver or reset the inner
parameters, going back to the default parameters, pressing Reset Param.
As it is indicated on the display, to start a matching process, you need beginning with the
radiographic device on off and then pressing the START button, turning on the device itself.
After a few seconds you will have a visual of the successful connection: it shows on the
display and the receiver will emit a long beep sound, then a short beep sound and again a
long one.
To go back to the menu, press the Menu button.

NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO)
Italy phone and fax + 39.011. 781994 [email protected] 42
www.newliferadiology.it
Manual Doc. HBE220-1

Installation and User Manual BEST-X-DC WITH Mod. MQI006-0


TIM-X

Display Options Screen

Here you can change some of the hand-held’s functions which are not connected with the
radiographic device, for instance: the display brightness (Brightness), the display sleep time
(energy saving mode) and the time after which the whole system turns off completely. To
modify them, you simply need to scroll the corresponding slider: if you exit (BACK) without
saving (Save) these settings will only be utilized until the device is turned on again, otherwise
they will be kept in its memory. From now on you can also access the self-calibration of the
radiographic device pressing the Calib. button.

NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO)
Italy phone and fax + 39.011. 781994 [email protected] 43
www.newliferadiology.it
Manual Doc. HBE220-1

Installation and User Manual BEST-X-DC WITH Mod. MQI006-0


TIM-X

Calibration Screen

When you have chosen the x-ray tube, you need to select, from the list on the right, what
type of exposition you wish to calibrate: after such choice, you have to press ButtonStart to
commence the calibration.
During the calibration, the radiographic will make up to a maximum of 20 expositions during
which it will modify the parameters to find the best ones. At the end of this cycle, if the
optimal point has been found, it will save the new parameters in its memory and the result
of the calibration will be sent to the hand-held. Such result will be shown in the text line
placed in the bottom.
To go back to the Menu, press Back.

NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO)
Italy phone and fax + 39.011. 781994 [email protected] 44
www.newliferadiology.it
Manual Doc. HBE220-1

Installation and User Manual BEST-X-DC WITH Mod. MQI006-0


TIM-X

Functioning e time scale:


On timer, by default, it is installed R10 time scale (in milliseconds [ms]), these values are
intended as recommended values to get with less exposure time maximum image quality,
they can be modified depending on the needs of the operator. In the tables below the
time scales shows programmed depending on the type of film or sensor used with the
relative dose value, expressed in milligray [mGy] which has to 20 cm from the fire tube (this
value is to be considered an estimate of the dose issued and should not be understood as a
measure of the same):

Film D
70 kV / 7 mA Tooth type

Patient type

320 [ms] 220 [ms] 160 [ms]


3.06 [mGy] 2.10 [mGy] 1.53 [mGy]

240 [ms] 180 [ms] 120 [ms]


2.29 [mGy] 1.72 [mGy] 1.14 [mGy]

Film D
70 kV / 4 mA Tooth type

Patient type

600 [ms] 440 [ms] 320 [ms]


3.26 [mGy] 2.39 [mGy] 1.74 [mGy]

480 [ms] 360 [ms] 240 [ms]


2.61 [mGy] 1.95 [mGy] 1.30 [mGy]

NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO)
Italy phone and fax + 39.011. 781994 [email protected] 45
www.newliferadiology.it
Manual Doc. HBE220-1

Installation and User Manual BEST-X-DC WITH Mod. MQI006-0


TIM-X

Film D
60 kV / 7 mA Tooth type

Patient type

340 [ms] 240 [ms] 180 [ms]


2.27 [mGy] 1.60 [mGy] 1.20 [mGy]

260 [ms] 200 [ms] 140 [ms]


1.74 [mGy] 1.34 [mGy] 0.93 [mGy]

Film D
60 kV / 4 mA Tooth type

Patient type

600 [ms] 480 [ms] 360 [ms]


2.28 [mGy] 1.82 [mGy] 1.36 [mGy]

500 [ms] 400 [ms] 280 [ms]


1.90 [mGy] 1.52 [mGy] 1.06 [mGy]

The timer is designed to work with systems of video radiography, by selecting the button

the times are reduced so as to be compatible with such systems. The following tables show

the scales used for digital film and phosphor:

Digital
70 kV / 7 mA Tooth type

Patient type

60 [ms] 40 [ms] 40 [ms]


0.57 [mGy] 0.38 [mGy] 0.38 [mGy]

60 [ms] 40 [ms] 40 [ms]


0.57 [mGy] 0.38 [mGy] 0.38 [mGy]

NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO)
Italy phone and fax + 39.011. 781994 [email protected] 46
www.newliferadiology.it
Manual Doc. HBE220-1

Installation and User Manual BEST-X-DC WITH Mod. MQI006-0


TIM-X

Digital
70 kV / 4 mA Tooth type

Patient type

100 [ms] 80 [ms] 80 [ms]


0.54 [mGy] 0.43 [mGy] 0.43 [mGy]

100 [ms] 80 [ms] 80 [ms]


0.54 [mGy] 0.43 [mGy] 0.43 [mGy]

Digital
60 kV / 7 mA Tooth type

Patient type

100 [ms] 100 [ms] 100 [ms]


0.67 [mGy] 0.67 [mGy] 0.67 [mGy]

100 [ms] 100 [ms] 100 [ms]


0.67 [mGy] 0.67 [mGy] 0.67 [mGy]

Digital
60 kV / 4 mA Tooth type

Patient type

180 [ms] 180 [ms] 180 [ms]


0.68 [mGy] 0.68 [mGy] 0.68 [mGy]

180 [ms] 180 [ms] 180 [ms]


0.68 [mGy] 0.68 [mGy] 0.68 [mGy]

NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO)
Italy phone and fax + 39.011. 781994 [email protected] 47
www.newliferadiology.it
Manual Doc. HBE220-1

Installation and User Manual BEST-X-DC WITH Mod. MQI006-0


TIM-X

Phosphorus
70 kV / 7 mA Tooth type

Patient type

200 [ms] 100 [ms] 100 [ms]


1.91 [mGy] 0.95 [mGy] 0.95 [mGy]

100 [ms] 80 [ms] 80 [ms]


0.95 [mGy] 0.76 [mGy] 0.76 [mGy]

Phosphorus
70 kV / 4 mA Tooth type

Patient type

300 [ms] 200 [ms] 100 [ms]


1.63 [mGy] 1.08 [mGy] 0.54 [mGy]

180 [ms] 160 [ms] 120 [ms]


0.97 [mGy] 0.87 [mGy] 0.65 [mGy]

Phosphorus
60 kV / 7 mA Tooth type

Patient type

200 [ms] 150 [ms] 120 [ms]


1.34 [mGy] 1.00 [mGy] 0.80 [mGy]

180 [ms] 100 [ms] 100 [ms]


1.20 [mGy] 0.67 [mGy] 0.67 [mGy]

NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO)
Italy phone and fax + 39.011. 781994 [email protected] 48
www.newliferadiology.it
Manual Doc. HBE220-1

Installation and User Manual BEST-X-DC WITH Mod. MQI006-0


TIM-X

Phosphorus
60 kV / 4 mA Tooth type

Patient type

400 [ms] 300 [ms] 200 [ms]


1.52 [mGy] 1.14 [mGy] 0.76 [mGy]

300 [ms] 200 [ms] 200 [ms]


1.14 [mGy] 0.76 [mGy] 0.76 [mGy]

To change the time, you just select the type of patient and the type of tooth,

The tooth and the patient selected will be highlighted by the blu LED in their

correspondence;

To change the type of tooth, just press the desired tooth icon.

If you have the need to manually increase or decrease the exposure time without following

the preset scale, this can be done by pressing the button to increase the time or by

pressing the button to decrease it; any increase/decrease of the time takes place in

steps of 20 ms.. Press button to save your edits of time.

When using the appliance we recommend the use of protective devices such as collars or

aprons.

NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO)
Italy phone and fax + 39.011. 781994 [email protected] 49
www.newliferadiology.it
Manual Doc. HBE220-1

Installation and User Manual BEST-X-DC WITH Mod. MQI006-0


TIM-X

Table Of Errors

When errors are visualized, the note ERROR appears, followed by an error code:

ERROR CODE CAUSE SOLUTION


Radiographic not found. The palmtop cannot connect Turn off and turn on the
Please, turn off the to the receiver radiological unit and repeat
radiographic and retry the matching procedure
After pressing the pre-start Press the button Cancel and
button, the palmtop is not from the Radiographic screen,
Searching for Radiographic… connected with the receiver repeat the procedure of the
Timeout prestart exposition configuration

The programmed exposure Press the Exit button and


time has not been successfully repeat the exposition
Exposition interrupted
accomplished procedure

NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO)
Italy phone and fax + 39.011. 781994 [email protected] 50
www.newliferadiology.it
Manual Doc. HBE220-1

Installation and User Manual BEST-X-DC WITH Mod. MQI006-0


TIM-X

The Monoblock Does Not Emit X-Rays


If the monoblock does not emit X-rays, as first check the continuity of the
THERMOSWITCH; if this test is negative, you must bypass it as shown in fig.19

fig. 19

NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO)
Italy phone and fax + 39.011. 781994 [email protected] 51
www.newliferadiology.it
Manual Doc. HBE220-1

Installation and User Manual BEST-X-DC WITH Mod. MQI006-0


TIM-X

If after this operation there is still no x-ray emission, the problem must be sought through a
tester in the continuity of the power cord that goes, if the wall version, from the monoblock
to the output X-RAY TUBE card fuse placed on the plate at points A and B (Fig. 20a); while
for the column version such continuity must be sought, always by means of a tester,
between points A and B indicated in fig 20b.

fig.20a (wall version)

NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO)
Italy phone and fax + 39.011. 781994 [email protected] 52
www.newliferadiology.it
Manual Doc. HBE220-1

Installation and User Manual BEST-X-DC WITH Mod. MQI006-0


TIM-X

fig.20b (column version)

NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO)
Italy phone and fax + 39.011. 781994 [email protected] 53
www.newliferadiology.it
Manual Doc. HBE220-1

Installation and User Manual BEST-X-DC WITH Mod. MQI006-0


TIM-X

Procedures for the filament current measures:


Measurement of the filament current
The measurement of the filament current can be carried out through the use of an
oscilloscope or by means of a digital multimeter. To perform this measurement, proceed as
follows:

 After removing the back cover, making reference to fig.21, connect oscilloscope / tester
terminals across the resistor to 1KΩ.

fig. 21

NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO)
Italy phone and fax + 39.011. 781994 [email protected] 54
www.newliferadiology.it
Manual Doc. HBE220-1

Installation and User Manual BEST-X-DC WITH Mod. MQI006-0


TIM-X

Measurement using the oscilloscope:


Measure across the resistor the voltage value. To make the measurement with an
oscilloscope, you need:

set the time base [SEC/DIV] and Volt time scale [VOLTS/DIV] of the oscilloscope in an
appropriate manner, so that the waveform of the measured voltage fills the screen of the
oscilloscope, as shown in fig.22 (the order of magnitude of exposure time to perform the
current measurement is of a few hundred ms)

fig. 22

By setting the timer on 7 mA, if the measurement is performed correctly it is expected to see
an average value of about 7 V (to each V corresponds 1 mA). As if 4 mA are set on timer
the average value that is expected to read is about 4 V.

NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO)
Italy phone and fax + 39.011. 781994 [email protected] 55
www.newliferadiology.it
Manual Doc. HBE220-1

Installation and User Manual BEST-X-DC WITH Mod. MQI006-0


TIM-X

Measurement using the digital multimeter:

It is necessary to set the full scale of the multimeter in an appropriate manner in order to
measure the voltage across the resistance; then after positioning the end sleeves of the
multimeter across the resistor (as shown previously in fig. 21) set an exposure time long
enough to allow the multimeter to read this voltage (the order of magnitude of the
exposure is approximately 1 sec). The value that is expected to read is like in the previous
case of about 7 V / 4 V depending on the current value set on the timer (to each V
corresponds 1 mA)

Procedure for dosimetric measurements:

During an inspection test are carried out dosimetric measurements to verify correct
operation. These measures are carried out by a special instrument cluster with the following
procedure:

 The instrument used for dosimetric measurements is the RTI ELECTRONICS PIRANHA
255;

 Is positioned the instrument at a distance of 50 cm from the fire hose and found the
dose values expressed in mGy;

 It checks the performance of the X-ray tube is in compliance with the project
specifications;

 The tests are stored in digital format and made available for those who request them;

NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO)
Italy phone and fax + 39.011. 781994 [email protected] 56
www.newliferadiology.it
Manual Doc. HBE220-1

Installation and User Manual BEST-X-DC WITH Mod. MQI006-0


TIM-X

Plates positioning

 Below, in fig.23 it shows the layout plates positioning on BEST-X-DC

fig.23

NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO)
Italy phone and fax + 39.011. 781994 [email protected] 57
www.newliferadiology.it
Manual Doc. HBE220-1

Installation and User Manual BEST-X-DC WITH Mod. MQI006-0


TIM-X

(“A” wall version)

(“A” column version)

NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO)
Italy phone and fax + 39.011. 781994 [email protected] 58
www.newliferadiology.it
Manual Doc. HBE220-1

Installation and User Manual BEST-X-DC WITH Mod. MQI006-0


TIM-X

1 2 3

Mod. T201

Description plate fields:

1. TUBE:

It indicates the model of the X-ray tube used;

2. MAT: XXXXXXXX

It indicates the serial number of the X-ray tube;

3. PALMTOP: XX. T XYZ

XX It indicates the year of manufacture;

T XYZ It indicates the number of the palmtop series;

4. S/N XX.XY.XYZ (It represents the device serial number)

XX It indicates the year;

XY It indicates the month;

XYZ indicates the serial sequential number;

NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO)
Italy phone and fax + 39.011. 781994 [email protected] 59
www.newliferadiology.it
Manual Doc. HBE220-1

Installation and User Manual BEST-X-DC WITH Mod. MQI006-0


TIM-X

(“B”)

(“C”)

NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO)
Italy phone and fax + 39.011. 781994 [email protected] 60
www.newliferadiology.it
Manual Doc. HBE220-1

Installation and User Manual BEST-X-DC WITH Mod. MQI006-0


TIM-X

(”D”)

1. S/N XX. T XYZ (It indicates the timer’s serial number)

XX It indicates the year of manufacture

T XYZ It is the palmtop’s progressive serial number

NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO)
Italy phone and fax + 39.011. 781994 [email protected] 61
www.newliferadiology.it
Manual Doc. HBE220-1

Installation and User Manual BEST-X-DC WITH Mod. MQI006-0


TIM-X

Servicing of the x-ray unit

CARRIED OUT BY THE DENTAL LABORATORY OPERATOR, WITH ANNUAL CADENCE

Informations for programming routine maintenance on the plant are provided to ensure the
best conditions of use by fully exploiting the diagnostic potential, without compromising any
safety and reliability aspect that characterize the Company.

The verifications and controls are divided into blocks and the operator before starting the
procedures of controls must ensure the availability of the instruction manual:

Radiogenic head monoblock:


plate integrity with the identification data;

Ionizing radiations warning signals integrity;

oil leaks absence verification;

complex integrity: covers shells, spacer cone attachment, rotation system;

monoblock’s correct anchoring verification at the pantograph;

verify the correct 360 ° rotation of the monoblock;

Wall and column support and pantograph:


verify the correct fixing of the wall support (only for wall version);

move the pantograph in all directions to verify the stability and bilance;

verify that the movement of the complex is fluid, flexible, without obstacles and that it is not
hardening;

check for drawings from fluidifying liquid;

NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO)
Italy phone and fax + 39.011. 781994 [email protected] 62
www.newliferadiology.it
Manual Doc. HBE220-1

Installation and User Manual BEST-X-DC WITH Mod. MQI006-0


TIM-X

Supply unit
plate integrity with the identification data;

verify the integrity of the snap command and its spiral cable;

integrity of the signal lights (yellow);

integrity of the acoustic signaling system;

If that anomalies are found on the plant during all of the above checks, these should be
reported to your distributor maintenance technician's New Life Radiology.

The extraordinary maintenance interventions are not carried out by the operator

Extraordinary maintenance
CARRIED OUT BY THE NEW LIFE RADIOLOGY SRL TECHNICAL ASSISTANCE IN CASE OF NEED
AND IN ANY CASE AFTER RELEVANT REPAIRS .

We provide a guideline on controls, on minimum checks that are to be performed in any


technical intervention on the plant:

monoblock, support and pantograph stability;

wear of the joints of rotation, of the plant balance and of the springs;

power supply's sliding contact of the head;

power cord of the head and of the control unit;

TESTING SERVICES

The frequency of checks carried out by a qualified expert on the performance of the X-ray
(high voltage values, the dose rate, etc ..) will allow to always obtain perfect images. The X-
ray, and in particular the single unit, does not contains parts subject to maintenance or
external assistance.

NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO)
Italy phone and fax + 39.011. 781994 [email protected] 63
www.newliferadiology.it
Manual Doc. HBE220-1

Installation and User Manual BEST-X-DC WITH Mod. MQI006-0


TIM-X

Manufacturer responsibility
 The manufacturer is responsible for the effects of safety, reliability and performance
of the equipment only if:

 assembly and any intervention on the equipment have been made by specialists;

 if the environment electrical system in which the equipment is installed is in


compliance with current standards in the field of plant safety;

 if the equipment is used in accordance with use

Parts of input and output signal


It is not provided for the equipment connections with other parts of entry and exit signal
outside.

Cleaning and disinfection of the unit and the parts in contact with
the patient

You should always disconnect the unit from the mains power supply before starting
cleaning activities and / or disinfection.

The method used for disinfection must meet the regulations and the recommendations
in force, including those regarding the prevention of risks of explosion.

Cleaning and disinfection of the parts in contact with the patient


 The parts in contact with the patient are represented by the cone collimator (cod.
CL01). This part should be carefully disinfected after the use through disposable
disinfectant wipes category "medical surgical".

Cleaning and disinfection of the unit


 For these operations can be used a cloth moistened with neutral water-based
detergent products. Make sure that no liquid seeps inside the equipment because it
may cause short circuits and corrosion. Do not use abrasive polishing.

 The accessories and connection cables must be disinfected only with a cloth soaked
in disinfectant solution. Do not use solvents or corrosive disinfectants.

NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO)
Italy phone and fax + 39.011. 781994 [email protected] 64
www.newliferadiology.it
Manual Doc. HBE220-1

Installation and User Manual BEST-X-DC WITH Mod. MQI006-0


TIM-X

 Spray disinfectants are not recommended because they may enter into the device
and cause short circuits and corrosion. If the spray use is essential, take the following
precautions:

 If the room where the appliance is installed is subjected to disinfection treatment, the
same must be carefully covered with a protective sheet, taking care to turn it off well
in advance so as to cool completely.

 After the dispersion of the vaporized disinfectant, remove the protective sheet and
disinfect the device as described above.

 Do not use the device in the presence of disinfectants which vaporize to form
explosive mixtures and wait until the vapors have dissipated before using it.

Environmental protection and dismantling


 The symbol indicates that this product is in conformity with European Directive
2002/96/Ec (Weee). It should not be treated as household waste but should be
handed in at the appropriate collection centre for the recycling of electrical and
electronic equipment. Discard it according to local regulations for waste disposal.

 The monoblock unit is made of lead parts and contains oil. The dismantling at end-of-
life of these parts must be done in a controlled way through authorized disposal
companies according to current regulations.

 In case of monoblock's damage for impact or crushing with a consequent oil


leaking, prevent dispersal into the environment, proceeding to decommissioning as
indicated above.

NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO)
Italy phone and fax + 39.011. 781994 [email protected] 65
www.newliferadiology.it
Manual Doc. HBE220-1

Installation and User Manual BEST-X-DC WITH Mod. MQI006-0


TIM-X

Features of the transmitter/receiver module


The device is powered by a 3.7 V / 1200 mA/h LI-POL battery. You can recharge the battery
thru the USB micro main.

Technical specifics

Features: Min Type Max Unit

Power Unit: 2.1 3.3 3.6 Vdc

Electrical
current Rx mode 3
absorbed mA
with Tx mode 25
modulation:

Transmission
863 870 MHz
frequency :

FSK: -107 at
25 kbps
Receiver
dBm
Sensitivity:
OOK: -113 at
2 kbps

Work
-40 +85 °C
temperature:

Battery charger features


Type of power Constant Voltage

Output voltage 5V DC

Max output current 1.2A

Type of output connector Micro USB

Power 6W

Power voltage 90÷264 Vac

NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO)
Italy phone and fax + 39.011. 781994 [email protected] 66
www.newliferadiology.it
Manual Doc. HBE220-1

Installation and User Manual BEST-X-DC WITH Mod. MQI006-0


TIM-X

External dimensions 64x40.5x25.5 [mm]

Protection Overcharge protection, overcurrent, anti-


short circuit

Work temperature 0÷40°C

Characteristics of the X-ray tube CEI OX/70-3

X-ray tube Technical description


Target material that characterizes the radiation spectrum : Tungsten
Reference axis for target angle and focal spot Orthogonal to the anode-cathode
characteristics: axis
Target angle compared to the reference axis: 13°
Focal spot value for the reference axis: 0,3 mm
Filtration: eq. 1 mm + 1 mm Al added,
not removable without tools
X-ray tube nominal voltage: 70kV
Current intensity and frequency: 4mA-7mA 60 Hz
Operating cycle: 1/30

X-ray tube- sheath assembly technical description


Reference axis for angle of the target and characteristics Orthogonal to the anode-cathode
of the focal spot: axis
Target angle compared to the reference axis: 13°
Focal spot value for the reference axis: 0,3 mm
Load factors values concerning radiation leakage: 1/30
Classification: Class IB (IEC 60601-1)
Data for high voltage connections: See detailed figures
High voltage connection polarity: Phase and Neutral (sinusoidal
alternating current)
Precautions to be observed to installation completed None
before the first load:

Beam limiting devices technical description


distance FFD 200 mm max  60 mm
Limiter beam (cone collimator) coated in lead:
mod. C.L.01

NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO)
Italy phone and fax + 39.011. 781994 [email protected] 67
www.newliferadiology.it
Manual Doc. HBE220-1

Installation and User Manual BEST-X-DC WITH Mod. MQI006-0


TIM-X

Technical description of the X-ray radiant complex for diagnostic


Reference axis at which the slope of the anode and the Orthogonal to the anode-cathode
characteristics of the focal spot refer: axis
Anode slope compared to the specific reference axis: 13°
Focal spot position on the reference axis: See detailed figures
Focal spot value for the reference axis: 0,3 mm

X-ray tube:

fig.24

NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO)
Italy phone and fax + 39.011. 781994 [email protected] 68
www.newliferadiology.it
Manual Doc. HBE220-1

Installation and User Manual BEST-X-DC WITH Mod. MQI006-0


TIM-X

Reference curves

Thermal Curves

Emission and filament characteristics

DC Rating Charts

fig.25

NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO)
Italy phone and fax + 39.011. 781994 [email protected] 69
www.newliferadiology.it
Manual Doc. HBE220-1

Installation and User Manual BEST-X-DC WITH Mod. MQI006-0


TIM-X

Safety aspects

 The equipment is not designed to be used in the presence of explosive gases or


vapours.

 It is forbidden to pour water or other liquids on the equipment so as not to cause short
circuits and corrosion

 Only service technicians are authorized to remove the monoblock from its support.

 The personnel authorized to the radiological examinations performance must observe


the protection rules against radiation.

 To protect patient from widespread radiation, it is recommended to wear protective


clothing for radiological use.

 All personnel present during an x-ray examination must comply with safety regulations
concerning protection against radiation. For their own safety, the operator must
always keep a distance of more than 2 meters and out of the path of the x-ray
beam, in order to avoid the exposition to the stray radiation.

NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO)
Italy phone and fax + 39.011. 781994 [email protected] 70
www.newliferadiology.it
Manual Doc. HBE220-1

Installation and User Manual BEST-X-DC WITH Mod. MQI006-0


TIM-X

Possible drawbacks found in the radiographs results


Clear image
Possible causes:

 Exhausted development liquid


 Excessive dilution of the developing liquid
 Too short exposure time to X-rays
 Insufficient development time
 Exposure liquid temperature below the recommended range
Dark image

Possible causes:

 Wrong dilution of developing liquid


 Too long exposure time to X-rays
 Excessive development time
 Exposure liquid temperature higher than the recommended

Not detailed image

Possible causes:

 Animal movement
 Monoblock movement

Radiography partially exposed

Possible causes:

 Error in the centering between rays’s beams and film


 Liquid development too low with a consequent film partial development
 Contact between two or more films during development

Veiled Image

Possible causes:

 Films that exceeded the expiration date


 Film accidental exposure to rays
 Film accidental exposure to heat sources
 Film accidental exposure to daylight or safety lamp darkroom no longer suitable

Black line occurrence on X-rays

Possible causes:

NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO)
Italy phone and fax + 39.011. 781994 [email protected] 71
www.newliferadiology.it
Manual Doc. HBE220-1

Installation and User Manual BEST-X-DC WITH Mod. MQI006-0


TIM-X

 A sharp film folding can be the cause of the appearance of a black line on the film
Recommendations
For maximum radiological image quality with minimal X-ray dose, we recommend the use
of films with high sensitivity and the respect of development time proposed by the
manufacturer of the films, shaking continuously films during the development itself. If the
image thus obtained is too dark, it is necessary to decrease the X-ray exposure time and not
the development duration.

In the case of manual development it is good to know that the developer liquid preserves
its efficiency in average for a week regardless of the number of films processed. Please note
that the treatment liquids are harmful to the environment and must be disposed of as
indicated by the manufacturer.

Film positioning
During normal operation, the film must be positioned at 90 ° with respect to the cone
collimator, as shown in fig.26

Upper jaw Lower jaw

Fig.26

NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO)
Italy phone and fax + 39.011. 781994 [email protected] 72
www.newliferadiology.it
Manual Doc. HBE220-1

Installation and User Manual BEST-X-DC WITH Mod. MQI006-0


TIM-X

According copyright laws, this


manual can not be copied, in
whole or in part, without the written
consent of the New Life Radiology.

Although every effort is made to


ensure the accuracy of the
information contained in this
manual, the New Life Radiology
does not assume responsibility for
any mistakes, omissions or
inaccuracies.

NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO)
Italy phone and fax + 39.011. 781994 [email protected] 73