HbA1c Testing Kit Guide
HbA1c Testing Kit Guide
A1C-3
Tina-quant Hemoglobin A1c Gen.3 - Whole blood and Hemolysate Application
Order information
Analyzer(s) on which kit(s) can be used
05336180 190 Tina-quant Hemoglobin A1c Gen.3 (2 × 100 tests) cobas c 111
Materials required (but not provided):
04528417 190 C.f.a.s. HbA1c (3 × 2 mL) Code 674
05479207 190 PreciControl HbA1c norm (4 × 1 mL) Code 208
05912504 190 PreciControl HbA1c path (4 × 1 mL) Code 209
05007232 190 Hemolyzing Reagent Gen.2 (8 × 6.3 mL) Code 952
11488457 122 HbA1c Hemolyzing Reagent for Tina-quant HbA1c (1000 mL)
11930630 001 Chimneys
A1C-3
Tina-quant Hemoglobin A1c Gen.3 - Whole blood and Hemolysate Application
This kit contains components classified as follows in accordance with the from various manufacturers may contain differing materials which could
Regulation (EC) No. 1272/2008: affect the test results in some cases. When processing samples in primary
tubes (sample collection systems), follow the instructions of the tube
manufacturer.
See the limitations and interferences section for details about possible
sample interferences.
A1C-3
Tina-quant Hemoglobin A1c Gen.3 - Whole blood and Hemolysate Application
A1C-3
Tina-quant Hemoglobin A1c Gen.3 - Whole blood and Hemolysate Application
3. As a matter of principle, care must be taken when interpreting any In rare cases of ">test rng" flags that might occur with the whole blood
HbA1c result from patients with Hb variants. Abnormal hemoglobins application remix the whole blood sample and repeat the analysis with the
might affect the half life of the red cells or the in vivo glycation rates. In same settings.
these cases even analytically correct results do not reflect the same Lower limits of measurement
level of glycemic control that would be expected in patients with normal
hemoglobin.21 Limit of Blank and Limit of Detection
Whenever it is suspected that the presence of an Hb variant (e.g. HbSS,
HbCC, or HbSC) affects the correlation between the HbA1c value and Hb: Limit of Blank = 0.31 mmol/L (0.50 g/dL)
glycemic control HbA1c must not be used for the diagnosis of diabetes Limit of Detection = 0.62 mmol/L (1.00 g/dL)
mellitus.
HbA1c: Limit of Blank = 0.12 mmol/L (0.19 g/dL)
4. Any cause of shortened erythrocyte survival will reduce exposure of
erythrocytes to glucose with a consequent decrease in mmol/mol HbA1c Limit of Detection = 0.18 mmol/L (0.29 g/dL)
values (IFCC) and % HbA1c values (DCCT/NGSP), even though the The Limit of Blank and Limit of Detection were determined in accordance
time-averaged blood glucose level may be elevated. Causes of with the CLSI (Clinical and Laboratory Standards Institute) EP17‑A
shortened erythrocyte lifetime might be hemolytic anemia or other requirements.
hemolytic diseases, homozygous sickle cell trait, pregnancy, recent
significant or chronic blood loss, etc. Caution should be used when The Limit of Blank is the 95th percentile value from n ≥ 60 measurements of
interpreting the HbA1c results from patients with these conditions. analyte‑free samples over several independent series. The Limit of Blank
HbA1c must not be used for the diagnosis of diabetes mellitus in the corresponds to the concentration below which analyte‑free samples are
presence of such conditions. found with a probability of 95 %.
5. Glycated HbF is not detected by the assay as it does not contain the The Limit of Detection is determined based on the Limit of Blank and the
glycated β-chain that characterizes HbA1c. However, HbF is measured standard deviation of low concentration samples.
in the total Hb assay and as a consequence, specimens containing high The Limit of Detection corresponds to the lowest analyte concentration
amounts of HbF (≥ 10 %) may result in lower than expected mmol/mol which can be detected (value above the Limit of Blank with a probability of
HbA1c values (IFCC) and % HbA1c values (DCCT/NGSP).13,23 95 %).
6. mmol/mol HbA1c values (IFCC) and % HbA1c values (DCCT/NGSP) Expected values
are not suitable for diagnosis of gestational diabetes.24 Protocol 1 (mmol/mol HbA1c acc. to IFCC): 29‑42 mmol/mol HbA1c27
7. In very rare cases of rapidly evolving type 1 diabetes the increase of Protocol 2 (% HbA1c acc. to DCCT/NGSP): 4.8‑5.9 % HbA1c27
HbA1c values might be delayed compared to the acute increase in
glucose concentrations. In these conditions diabetes mellitus must be This reference range was obtained by measuring 482 well-characterized
diagnosed based on plasma glucose concentrations and/or the typical healthy individuals without diabetes mellitus. HbA1c levels higher than the
clinical symptoms.24 upper end of this reference range are an indication of hyperglycemia during
the preceding 2 to 3 months or longer. According to the recommendations
Criterion: Recovery within ± 10 % of initial value. of the American Diabetes Association values above 48 mmol/mol HbA1c
Icterus: No significant interference up to a conjugated and unconjugated (IFCC) or 6.5 % HbA1c (DCCT/NGSP) are suitable for the diagnosis of
bilirubin concentration of 1026 μmol/L or 60 mg/dL. diabetes mellitus.24,28 Patients with HbA1c values in the range of
Lipemia: No significant interference up to an Intralipid concentration of 39‑46 mmol/mol HbA1c (IFCC) or 5.7‑6.4 % HbA1c (DCCT/NGSP) may be
600 mg/dL. There is poor correlation between the triglycerides at a risk of developing diabetes.24,28
concentration and turbidity. HbA1c levels may reach 195 mmol/mol HbA1c (IFCC) or 20 % HbA1c
Glycemia: No significant interference up to a glucose level of 55.5 mmol/L (DCCT/NGSP) and more in poorly controlled diabetes. Therapeutic action is
(1000 mg/dL). A fasting sample is not required. suggested at levels above 64 mmol/mol HbA1c (IFCC) or 8 % HbA1c
(DCCT/NGSP). Diabetes patients with HbA1c levels below 53 mmol/mol
Rheumatoid factors: No significant interference up to a rheumatoid factor HbA1c (IFCC) or 7 % HbA1c (DCCT/NGSP) meet the goal of the American
level of 750 IU/mL. Diabetes Association.18,19
Drugs: No interference was found at therapeutic concentrations using HbA1c levels below the established reference range may indicate recent
common drug panels.25,26 episodes of hypoglycemia, the presence of Hb variants, or shortened
Other: No cross reactions with HbA0, HbA1a, HbA1b, acetylated lifetime of erythrocytes.
hemoglobin, carbamylated hemoglobin, glycated albumin and labile HbA1c Each laboratory should investigate the transferability of the expected values
were found for the anti-HbA1c antibodies used in this kit. to its own patient population and if necessary determine its own reference
For diagnostic purposes, the results should always be assessed in ranges.
conjunction with the patient’s medical history, clinical examination and other
findings. Specific performance data
Representative performance data on the cobas c 111 analyzer are given
ACTION REQUIRED below. Results obtained in individual laboratories may differ.
Special Wash Programming: The use of special wash steps is mandatory
when certain test combinations are run together on the cobas c 111 Precision
analyzer. For information about test combinations requiring special wash Repeatability and intermediate precision were determined using human
steps, please refer to the latest version of the carry-over evasion list found samples and controls in accordance with the CLSI (Clinical and Laboratory
with the CLEAN Method Sheet and the operator’s manual for further Standards Institute) EP5 requirements (2 aliquots per run, 2 runs per day,
instructions. 21 days). The following results were obtained (data based on DCCT/NGSP
Where required, special wash/carry-over evasion programming must values):
be implemented prior to reporting results with this test.
Whole Blood application
Limits and ranges
Measuring range Repeatability Mean SD CV
Hb: 2.48‑24.8 mmol/L (4‑40 g/dL) % HbA1c % HbA1c %
HbA1c: 0.186‑1.61 mmol/L (0.3‑2.6 g/dL)* PreciControl HbA1c norm 5.63 0.04 0.7
*The measuring range for HbA1c lies between 0.186 mmol/L and the PreciControl HbA1c path 10.7 0.06 0.6
concentration of the highest standard. The test range stated above is based
on a typical calibrator value of 1.61 mmol/L. Human sample 1 5.21 0.04 0.7
This corresponds to a measuring range of 23‑196 mmol/mol HbA1c (IFCC) Human sample 2 6.22 0.04 0.7
and 4.2‑20.1 % HbA1c (DCCT/NGSP) at a typical hemoglobin
concentration of 8.2 mmol/L (13.2 g/dL). Human sample 3 7.34 0.05 0.7
A1C-3
Tina-quant Hemoglobin A1c Gen.3 - Whole blood and Hemolysate Application
A1C-3
Tina-quant Hemoglobin A1c Gen.3 - Whole blood and Hemolysate Application
10 Wolf HU, Lang W, Zander R. Alkaline haematin D-575, a new tool for Any serious incident that has occurred in relation to the device shall be
the determination of haemoglobin as an alternative to the reported to the manufacturer and the competent authority of the Member
cyanhaemiglobin method. II. Standardization of the method using pure State in which the user and/or the patient is established.
chlorohaemin. Clin Chim Acta 1984;136:95-104. The Summary of Safety & Performance Report can be found here:
11 Little RR, Wiedmeyer HM, England JD, et al. Interlaboratory https://ec.europa.eu/tools/eudamed
standardization of measurements of glycohemoglobins. Clin Chem Symbols
1992;38:2472-2478.
Roche Diagnostics uses the following symbols and signs in addition to
12 Frank EL, Moulton L, Little RR, et al. Effects of hemoglobin C and S those listed in the ISO 15223‑1 standard (for USA: see dialog.roche.com for
traits on seven glycated hemoglobin methods. Clin Chem definition of symbols used):
2000;46(6):864-867.
13 Chang J, Hoke C, Ettinger B, et al. Evaluation and Interference Study Contents of kit
of Hemoglobin A1c Measured by Turbidimetric Inhibition Immunoassay. Reagent
Am J Clin Pathol 1998;109(3):274-278.
Volume after reconstitution or mixing
14 Kobold U, Jeppsson JO, Duelffer T, et al. Candidate reference methods
for hemoglobin A1c based on peptide mapping. Clin Chem GTIN Global Trade Item Number
1997;43:1944-1951.
15 Jeppsson JO, Kobold U, Finke A, et al. Approved IFCC reference COBAS, COBAS C, COBAS INTEGRA, TINA-QUANT and PRECICONTROL are trademarks of Roche.
method for the measurement of HbA1c in human blood. Clin Chem Lab All other product names and trademarks are the property of their respective owners.
Med 2002;40:78-89. Additions, deletions or changes are indicated by a change bar in the margin.
© 2020, Roche Diagnostics
16 Martina WV, Martijn EG, van der Molen M, et al. β-N-terminal
glycohemoglobins in subjects with common hemoglobinopathies:
relation with fructosamine and mean erythrocyte age. Clin Chem
1993;39:2259-2265.
Roche Diagnostics GmbH, Sandhofer Strasse 116, D-68305 Mannheim
17 Weykamp CW, Penders TJ, Muskiet FAJ, et al. Influence of www.roche.com
hemoglobin variants and derivatives on glycohemoglobin
+800 5505 6606
determinations, as investigated by 102 laboratories using 16 methods.
Clin Chem 1993;39:1717-1723.
18 American Diabetes Association. Standards of Medical Care for patients
with diabetes mellitus. Diabetes Care [Suppl.] 1995;18(1):8-15.
19 Sacks BW, Bruns DE, Goldstein DE, et al. Guidelines and
recommendations for laboratory analysis in the diagnosis and
management of diabetes mellitus. Clin Chem 2002;48:436-472.
20 Glick MR, Ryder KW, Jackson SA. Graphical Comparisons of
Interferences in Clinical Chemistry Instrumentation.
Clin Chem 1986;32:470-475.
21 Miedema K. Influence of hemoglobin variants on the determination of
glycated hemoglobin. Klin Lab 1993;39:1029-1032.
22 Niederau C, Coe A, Katayama Y. Interference of Non-glucose Adducts
on the Determination of Glycated Hemoglobins. Klin Lab
1993;39:1015-1023.
23 Rohlfing C, Connolly J, England J, et al. Effect of Elevated Fetal
Hemoglobin on HbA1c Measurements: Four Common Assay Methods
compared to the IFCC Reference Method. Poster Abstract AACC
Annual Meeting 2006, Chicago. Clin Chem 2006;52(6) Suppl A 108.
24 International Expert Committee Report on the Role of the A1C Assay in
the Diagnosis of Diabetes. Diabetes Care 2009;32(7):1327-1334.
25 Breuer J. Report on the Symposium “Drug effects in Clinical Chemistry
Methods”. Eur J Clin Chem Clin Biochem 1996;34:385-386.
26 Sonntag O, Scholer A. Drug interference in clinical chemistry:
recommendation of drugs and their concentrations to be used in drug
interference studies. Ann Clin Biochem 2001;38:376-385.
27 Junge,W, Wilke B, Halabi A, et al. Determination of reference levels in
adults for hemoglobin A1c (HbA1c). Poster presentation
EUROMEDLAB, Barcelona 2003.
28 Diagnosis and Classification of Diabetes Mellitus. Diabetes Care
2010;33(1):62-69.
29 Consensus statement on the worldwide standardization of the
hemoglobin A1c measurement. American Diabetes Association,
European Association for the Study of Diabetes, International
Federation of Clinical Chemistry and Laboratory Medicine and
International Diabetes Federation Consensus Committee. Diabetes
Care 2007;30:2399-2400.
A point (period/stop) is always used in this Method Sheet as the decimal
separator to mark the border between the integral and the fractional parts of
a decimal numeral. Separators for thousands are not used.