Sme5 Eng+Front
Sme5 Eng+Front
O2 Analyser system
with SME 5 Electronics
version 8.129-116np
Please note: The photograph shows as an example the KES5001 probe with a
standard IP65 SME 5 electronics package. Your order may have a different style
of probe and electronics housing selected.
OXITEC® 5000 O2 Analyser System with SME 5 Electronics Version 8.129-116np Page 2
11.7.8 Measuring Range Shift ...................................................................................................................... 35
11.7.9 Alter code........................................................................................................................................... 35
11.7.10 Damping mA output ....................................................................................................................... 35
11.7.11 Call up max./min. values................................................................................................................ 35
11.7.12 Change language........................................................................................................................... 35
11.7.13 Service........................................................................................................................................... 35
11.7.14 Software ........................................................................................................................................ 35
11.8 STATUS MESSAGES ....................................................................................................................36
12 SYSTEM DESCRIPTION / ADJUSTMENTS .................................................................................38
12.1 IN CASE OF LOSS THE SYSTEM CODE ...........................................................................................38
12.2 BRIDGE ON BACK OF DISPLAY UNIT ..............................................................................................38
12.3 ADJUSTMENT DISPLAY CONTRAST ...............................................................................................39
12.4 RELAY OUTPUTS / FUNCTIONS ....................................................................................................40
12.5 DIGITAL INPUTS..........................................................................................................................40
12.5.1 Digital inputs at Option limit curves .................................................................................................... 40
12.6 AMOUNTS OF TEST AIR AND/OR REFERENCE AIR...........................................................................40
12.7 STABILITY CRITERIA FOR CALIBRATION.........................................................................................41
12.8 REACTION TIME OF THE MA OUTPUT............................................................................................41
12.9 EXTENSION MODULES ................................................................................................................41
12.9.1 RS 232 Interface (option)................................................................................................................... 41
13 TECHNICAL SPECIFICATION ......................................................................................................42
13.1 TECHNICAL SPECIFICATION ELECTRONIC UNIT ............................................................................42
13.2 TECHNICAL SPECIFICATION PROBE .............................................................................................43
14 SERVICE AND MAINTENANCE ...................................................................................................44
14.1 SWITCHING BETWEEN 230 V AND 115 V OPERATION ...................................................................44
14.2 REPLACING THE FUSES ..............................................................................................................44
14.3 ADJUSTING THE FLOW RATE OF REFERENCE AND TEST AIR ...........................................................45
14.4 REPLACING FILTER ....................................................................................................................46
14.5 EXCHANGE OF ZRO2-MEASURING CELL .......................................................................................46
14.6 EXCHANGE OF PROBE INNER PARTS ............................................................................................49
15 TROUBLE SHOOTING ..................................................................................................................49
15.1 O2 DISPLAY 0 %, ALTHOUGH THE OPERATION MODE INDICATES A HIGHER O2 VALUE ......................50
15.2 LOCAL DISPLAYS CORRECT, OUTPUT NOT CORRECT .....................................................................51
15.3 UNSTEADY, WIDELY VARYING MEASURING VALUE .........................................................................51
15.4 DISPLAY REMAINS AT THE END OF THE MEASURING RANGE OR IS HIGHER THAN EXPECTED .............52
16 DIMENSIONAL SHEETS OF THE EVALUATION ELECTRONICS .............................................53
19 INDEX.............................................................................................................................................62
OXITEC® 5000 O2 Analyser System with SME 5 Electronics Version 8.129-116np Page 3
Introduction
Dear customer,
Thank you for selecting our OXITEC® 5000 as your in-situ Flue Gas Oxygen measuring system.
For many years now, our OXITEC® oxygen analyser systems have been operating in numerous
applications with over 1000 units being produced and shipped around the world. ENOTEC are committed
to total quality and performance and we have continuously enhanced our products to integrate various
additional features and functions. In this package, the electronics uses the very latest Microprocessor
Technology, making the SME5 electronics one of the most advanced and up-to-date monitoring units
permitting you to reduce your maintenance & fuel costs to achieve an increased measuring accuracy with
more operational reliability due to new monitoring functions.
In our oxygen measuring probes you will find the Zirconium Oxide measuring cell is soldered in place
using a special process and technique developed by our company. This leads to a considerably
increased service life as compared to "glued or cemented" measuring cells, that have a tendency to leak
or crack during operation. The OXITEC cell is a proven gas-tight design providing greater measuring
accuracy, durability and longer working life.
All ENOTEC instruments are thoroughly tested in the factory and are subject to a strict ISO 9001 Quality
Assurance procedure. Therefore, with the correct installation, the operation of the OXITEC-5000 oxygen
analyser system is very easy and user friendly. It will provide you with many years of operation with
perfect measuring results.
Please read this manual carefully and follow the instructions as directed. If you have any questions or are
unclear about any aspect of the installation or operation, our Service Department, regional sales office or
approved distributor will be pleased to assist. The factory contact points are Tel: +49 2264-45780 or Fax:
+49 2264 457831 You may also visit us at our website http://www.enotec.de or http://www.enotec.com or
contact us by e-mail: [email protected] or [email protected].
OXITEC® 5000 O2 Analyser System with SME 5 Electronics Version 8.129-116np Page 4
ENOTEC GmbH
Höher Birken 6
51709 Marienheide/Rodt
Germany
CE Mark - Conformity
to EC Directive Electromagnetic Compatibility 89/336/EEC
We declare that the OXITEC® Oxygen Analyser System as detailed below due to its design and construction as well
as the type sold by us complies with the relevant basic safety and health requirements of the EC Directive. This
declaration becomes invalid if the appliance is modified without our prior consent.
Relevant
EC-directives: EC Low-voltage Directive (73/23/EEC)
EC Directive Electromagnetic Compatibility (89/336/EEC)
EC Directive Product Liability (85/374/EEC)
Applied
harmonised
generic standards: EN 50082-2 1995 Resistance to interference
.....................................................................................
EN 55011 Kl.B 1992 Emission
.....................................................................................
.....................................................................................
Applied
standards and technical
specification, in particular: EN 61000-4-2 ESD
....................................................................................
EN 61000-4-3 Irradiation E-field
....................................................................................
EN 61000-4-4 Burst
....................................................................................
EN 61000-4-6 Inflow
....................................................................................
EN 61000-3-2 Current Harmonics
....................................................................................
EN 61000-3-3 Flicker
....................................................................................
EN 61326-1 Immunity partly
....................................................................................
OXITEC® 5000 O2 Analyser System with SME 5 Electronics Version 8.129-116np Page 5
ENOTEC
2 YEAR WARRANTY
ENOTEC warrants that the equipment manufactured and sold by it will, upon shipment, be free of defects
in workmanship or material. Should any failure to conform to this warranty become apparent during a
period of two years after the date of shipment, ENOTEC shall upon prompt written notice from the
purchaser, correct such nonconformity by repair or replacement of the defective part or parts. The
purchaser is not entitled to claim any other legal remedies on the basis of this warranty. Please refer to
the complete ENOTEC warranty policy for details.
ENOTEC does not warrant equipment supplied by it against normal deterioration. Corrosive gases and
solid particles may cause damages and make a repair or a replacement necessary as a consequence of
normal wear and tear during the warranty period.
Note: When installing the equipment, the customer must ensure that all necessary supply lines are
connected and the operating temperature of the probe is reached. Experience has shown that products
installed but not taken into operation may be damaged by the process or by external influence. ENOTEC
will not accept any responsibility for such damages.
________________________________________________________________________________________
MANUFACTURER:
ENOTEC GmbH
Höher Birken 6
D-51709 Marienheide
Germany
TEL: +44(0)2264/45780
FAX: +44(0)2266/457831
Email: [email protected] and
[email protected]
OXITEC® 5000 O2 Analyser System with SME 5 Electronics Version 8.129-116np Page 6
1 Intended use
The OXITEC Oxygen Analyser System is a system for measuring the oxygen
concentration in flue gases and other non-combustible gases. For reasons of safety and
the possible occurrence of accidents, unauthorized conversions and modifications of the
system are prohibited.
OXITEC Oxygen Analyser Systems are available in an explosion-proof design for zone-1-installation. The
electronic system is certified EExd IIC T6 and the probe EExd IIC T3.
Please note the corresponding annexes.
The OXITEC Oxygen Analyser System may only be operated by authorized, trained skilled persons. For
reasons of immunity to interfering it must be ensured that the connection between measuring probe and
electronic system is only made with the special ENOTEC probe cable and pneumatic tubing.
The O2 measuring probes supplied by us may only be used with ENOTEC electronics or with compatible
electronics approved by ENOTEC. All probes operated at 840°C.
Exception: The explosion-proof probes are approved only for the operating temperature as per Ex-
certification. All OXITEC SME 5 electronics are factory set to 840°C operating temperature (with the
exception of explosion-proof electronics).
Note!
Under no circumstance may the measuring probe be directly connected to the 230
V mains power, as this will immediately destroy the probe heater element!
OXITEC® 5000 O2 Analyser System with SME 5 Electronics Version 8.129-116np Page 7
4 Safety hazards
During operation, the temperature of the probe filter head and of all parts
exposed to flue gas is 150°-840°C. Direct contact with the hot parts for
dismantling or maintenance will cause severe burns!
The probe may only be removed with heat-insulated gloves. Before removing the probe,
always switch off the supply voltage of the electronic system. After removal, store the
probe in a safe, protected place and wait until it has cooled down below 35°C.
The enclosed ferrite sleeves must be put over the terminal leads of
the cable.
(see Figure 11)
CE-conformity is void if these ferrite sleeves are not fitted!
The ground (Earth) connection must be installed thoroughly using the correct
cable size of at least 1,5mm2 minimum. It is very important that the equipotential
bonding is installed correctly.
Please note that some electrical devices (eg. frequency converter etc.) close to the
system can create considerable electrical disturbance and instability. So
for normal function of the OXITEC 5000 System it is important that these devices
be installed away from the system and in accordance with the manufacturers
installation notes and guidelines.
Please do not lay the probe cable near power supply cables, or near motors and
their cables, as they are controlled by frequency converters because this could
cause interference effects. If possible don’t cross the cables because this could
cause interference effects. However, if this is unavoidable please make sure that
this cables are only crossed at right angles.
OXITEC® 5000 O2 Analyser System with SME 5 Electronics Version 8.129-116np Page 8
6 GENERAL SYSTEM DESCRIPTION
The OXITEC 5000 measures the net concentration of oxygen in the flue gas of combustion processes
and other non-combustible gases.
The EMF (electrical motive force) produced by a temperature-stabilised Zirconia cell indicates the amount
of oxygen concentration.
Measurement is made direct and in-situ, i.e. the measuring cell is inside the flue gas duct at the end of a
probe in the middle third of the gas stream.
The water content existing in any flue gas and which has an influence on the actual O2 concentration, is
taken into account in this wet measuring process. The wet process provides a lower measuring value
compared to the dry process, because it refers to the total volume. The difference between both values is
in direct proportion to the water content in the flue gas.
The measuring cell consists of a small zirconium-oxide disc about the size of a penny, which is coated
with porous layers of platinum on both sides and soldered with a gas-tight seal into the end of a steel
tube cell holder.
The temperature of the measuring cell is stabilised by a built-in heater whose temperature is kept
constant by a temperature controller. At a constant temperature the mV-output of the cell is calculated as
follows:
The reference gas (Instrument Air) is clean, dry and oil-free air (20.95% by volume O2).
When different oxygen concentrations occur between measuring and reference side of the
electrochemical cell, there is a migration of oxygen ions from the higher to the lower partial pressure side.
The mV output signal of the cell is inversely logarithmically proportional to the oxygen content of the
process gas. With reducing oxygen content in the process gas the mV signal of the measuring cell
increases. Air on the process gas side, the ZrO2-cell generates approx. 0 mV ± 5 mV (cell constant) and
with 2.1% O2 approx. + 50 mV (rise of the characteristic line or decade voltage).
OXITEC® 5000 O2 Analyser System with SME 5 Electronics Version 8.129-116np Page 9
7 Design of the OXITEC 5000 Oxygen Analyser System
Every OXITEC 5000 Oxygen Analyser System consists of the measuring probe, which detects the
oxygen partial pressure direct in the flue gas duct, the electrical connection cable, the pneumatic
connection cable and the electronics. In the following paragraphs, the individual components will be
explained in detail.
ENOTEC
OXITEC 5000
OXITEC
Sauerstoffanalysensysteme
ENOTEC GmbH
51709 Marienheide / Germany
Modell
Zell-Nr.
Ser.- Nr.
Figure 1
electronic unit SME 5 OXITEC oxygen measuring probe with cooling protection tube
control cable FEP-0003 for solenoid valve (option)
ENOTEC
gas outlet
OXITEC
Sauerstoffanalysensysteme
ENOTEC GmbH
OXITEC 5000 51709 Mari enheide / Germany
Mod ell
gas entry
Zell- Nr.
Ser.- Nr.
succing/blow back
connection
cloosed gastight
insulation
flue gas duct
instrument
air in* pneumatic cable FEP-0002 length 150 m max.
Figure 2
OXITEC® 5000 O2 Analyser System with SME 5 Electronics Version 8.129-116np Page 10
8 DESCRIPTION OF THE SYSTEM COMPONENTS
The probe assembly consists of the connecting box (ASK-...), the inner probe assembly part (SIK-...) with
heater, thermocouple and signal wire, the measuring probe tube (MSR-...) with test gas line, the ZrO2
measuring cell, the filter head and the protection tube (optional for model KES/KIS200x). Refer to drawing
on page 57.
The connecting box contains the electrical connectors for the measuring cell (terminals {-1/+2}), the
thermocouple (terminals {+3/-4}), the heater (terminals 6-8) and additionally the pneumatic connector for
the reference air.
The probe tube with flange is itself the negative pole and serves as protection for the positive measuring
line and the reference gas line to the reference side of the measuring cell.
The filter head has a gas deflector plate referred to as V-shield.
The measuring cell itself consists of several layers: platinum, zirconium oxide and platinum with the
platinum being applied porously. The measuring cell with its 4-hole connection flange and gasket seal
can be field replaceable by the customer as required.
Figure 3
The test gas flows through to the filter and fills the flue gas side of the measuring cell. On the other side,
reference gas (clean, dry and oil-free instrument air (20.95 percent by volume O2)) continuously flows
through the inner part of the probe to the measuring cell. Because of the two different gases on each side
of the measuring cell, a logarithmic mV output occurs (high O2-value in the test gas = low voltage (e.g.
21% ≈ 0 mV); low O2-value in the test gas = high voltage (e.g. 2.1% ≈ 50 mV)). By means of the
electronics, this voltage (mV) is converted into a current 4-20 mA (or 0-20 mA) with a maximum load of
500 Ohm. The current output is then a direct relationship with the measured Oxygen value and available
as an electrically isolated output signal from the terminals (see terminal drawing, page 16 ).
OXITEC® 5000 O2 Analyser System with SME 5 Electronics Version 8.129-116np Page 11
The temperature of 840°C required for the measuring cell is generated by a heater element installed
inside the probe. To keep the temperature constant, it is measured by a Ni-Cr-Ni thermocouple and the
heater voltage controlled. The SME 5 electronics is equipped with a built-in temperature reference point
to compensate ambient temperature variations (Cold Junction Compensation).
Between probe and electronics, only the ENOTEC O2 special 7-core cable may be used,
which contains the conductors for the heater power, measuring signal, thermocouple,
ground/earth and compensation wire. This cable must not be longer than 150 m. Other
cable lengths must be technically approved with ENOTEC in advance.
The ENOTEC special cable must be treated as a measuring signal line.
ENOTEC offers a pneumatic cable with 2 tubes for the reference air supply to the probe (30 l/h
continuously) and the test gas supply (150 l/h - 200 l/h).
Both cables have a PUR insulation and can be used with ambient temperatures up to 90°C.
Please do not lay the probe cable near power supply cables, or
near motors and their cables, as they are controlled by frequency
converters because this could cause interference effects. If possible
don’t cross the cables because this could cause interference
effects. However, if this is unavoidable please make sure that this cables are only
crossed at right angles.
As a standard, the electronics are supplied in a housing for wall mounting, with pneumatic system.
Two different pneumatic systems are available:
• an instrument-air version requiring the supply of clean, dry, oil-free instrument air at the point of
installation.
• a pump version using the surrounding ambient air.
To recognize the type of pneumatic version look at the valve for adjust Testair. (see Figure on page 45)
To ensure the full functional range of the electronics, one of the above pneumatic
units is required.
The housing has bottom entry electrical cable glands, and compression fittings for 6mm or ¼” tubing, as
specified, for the pneumatic lines.
Optionally, the electronic unit SME 5 is also available in a 19” sub-rack, in a weather-resistant glass fibre
reinforced plastic housing or in an EExd housing for hazardous areas.
With the EExd housing, the pneumatic unit is mounted in a separate sheet steel field housing that does
not require certification since it only contains the pneumatic components.
OXITEC® 5000 O2 Analyser System with SME 5 Electronics Version 8.129-116np Page 12
9 INSTALLATION OF THE O2 ANALYSER SYSTEM
Only qualified and authorized staff may work with this appliance. This staff must
be familiar with all warnings, safety instructions and maintenance work in
accordance with these operating instructions. The faultless and safe operation of
this equipment requires proper transportation, professional storage, setup and
installation as well as careful operation and maintenance.
Flue gas temperature, pressure and all other process conditions must be in accordance with the
specification. Leave enough space for insertion/removal of the O2 probe and protection tube (if supplied)
and ensure access to the measuring probe and/or connecting box.
Before cutting the hole in the flue gas duct, make sure that the inside of the duct has enough space for
probe installation, and that no soot is blown out nearby, or any obstacles are in the way.
For probe lengths exceeding 2000 mm, a support must be mounted inside the duct (every 2 m) to prevent
the probe and mounting tube from flexing or bending.
The dimensions for the hole in the flue gas duct and the mounting of the adapter plate (mating flange) are
shown in the dimensional sheet, pages 57. The mating flange is fitted by the customer and when
mounting, ensure an air tight seal.
The part of the protection tube projecting from the duct wall must be insulated or heated, if
necessary, to prevent its temperature from dropping below the dew point.
Electrical heaters are available from ENOTEC as accessories. The bolts of the probe flange must remain
accessible.
The special cables for pneumatic and electronic systems must be left long enough (coiled up) so that the
probe can be drawn out of the protection tube without disconnecting it.
OXITEC® 5000 O2 Analyser System with SME 5 Electronics Version 8.129-116np Page 13
mounting O2-probes with adpater plate ADP-2000
gas kets
weld the adapterplate bolts
duct wall so that the probe M16 x 45
O2-probe get the declare
KES/KIS-200X flue gas duct incline.
ca. 1-2ø
min. opening
in the duct wall: í 80 mm
max. opening:
the protection tube í 150 mm
must be insulated in the adapterplate í76,1
this ara to avoid has to be welded
condensation. gastight to the outside diameter protection tube
flu gas duct.
mounting O2-probes with adapter plate ASR-1320
duct wall bolts
gas ket M8 x 29
flue gas duct
ca. 1-2ø
O2-probe
KES-132X min. opening in the
weld the adapterplateduct wall: í 72 mm
so that the probe í76,1
max. opening: í 76,1 mm
the adapterplate the adapterplate get the declare
must be insulated in has to be welded incline.
this ara to avoid gastight to the
condensation. flu gas duct.
Figure 4
The ground (Earth) connection must be installed thoroughly using the correct
cable size of at least 1,5mm2 minimum. It is very important that the equipotential
bonding is installed correctly.
Please note that some electrical devices (eg. frequency converter etc.) close to the
system can create considerable electrical disturbance and instability. So
for normal function of the OXITEC 5000 System it is important that these devices
be installed away from the system and in accordance with the manufacturers
installation notes and guidelines.
Please do not lay the probe cable near power supply cables, or near motors and
their cables, as they are controlled by frequency converters because this could
cause interference effects. If possible don’t cross the cables because this could
cause interference effects. However, if this is unavoidable please make sure that
this cables are only crossed at right angles.
OXITEC® 5000 O2 Analyser System with SME 5 Electronics Version 8.129-116np Page 14
9.3 Adjusting the filter head
Before installing the probe, the direction of flue gas flow must be determined and
the filter head assembly turned so that the flue gas flow hits the V - deflector. The
filter head can be freely turned a full 360° for this purpose, by loosening the holding
screws and lock nuts on the filter head, rotating the filter head/V-shield to the
required position, and then tightening both holding screws and lock nuts.
filterhead
filter
V-shield
flow direction
of flue gas
Figure 5
OXITEC® 5000 O2 Analyser System with SME 5 Electronics Version 8.129-116np Page 15
9.4 Electrical and pneumatic connections
The special ENOTEC probe cable (FEP-0001) and the pneumatic tubing (FEP-0002) must be connected
in accordance with Figure 1 and Figure 6. Both cables have a PUR sheathing and are designed for a
maximum permissible ambient temperature of +90°C.
test gas valve (option) test gas line measuring cell filter head
test gas supply
filter
OXITEC
Sauerstoffanalysensysteme
ENOTEC GmbH
51709 Marienheide / Germany
Modell
Zell-Nr.
V-shield
Ser.- Nr.
Figure 6
probe connection box
measuring cellsolenoid valve
wh-br/br gr-bl/gr
thermocouple 2 * 0.75 ý 2 * 0.75 ý
2 * 0.75 ý probe heating
SoNiCr/SoNiCr
wh/gr bla/blu/gr-ye
3 * 1.5 ý
Attention !
Do not connect shieldings at
the probe side and ensure,that
these do not have a conductive
1 2 3 4 6 7 8
connection with the cabinet or
wh/br brown greenwhite black blue gr.ye. the PG - screwing.
- mV+ +mV- L N
115 V ~
cell thermocouple probe heating
PUR-covering
í 11,7 mm
to electronic unit
Figure 7
OXITEC® 5000 O2 Analyser System with SME 5 Electronics Version 8.129-116np Page 16
9.5 Installing the SME5 electronics unit
When installing the electronic unit, please ensure sufficient air circulation around the
electronics housing.
The permissible ambient temperature range for the standard steel box field housing with
instrument air is –20°C to +55°C, and for the unit with built-in reference air and calibration gas pumps,
-20° to +50°. For the 19” rack mounting version, the permissible ambient temperature is 0°C to +60°C.
The location of the electronics unit must be chosen so that it is easily accessible ensuring that the door
can be fully opened after installation.
The place should be largely weather-protected so as to enable inspection even under the worst weather
conditions.
Before putting the system into operation, it is essential that all electrical connections, supply voltage and
pneumatic connections are checked, and, if applicable, the instrument air pressure and quality.
Special care must be taken not to mix up the connections for test gas and reference air.
The enclosed ferrite sleeves must be put over the leads of the stripped cable.
(see Figure 11)
CE-conformity is invalid if these ferrite sleeves are not fitted!
The ground (Earth) connection must be installed thoroughly using the correct
cable size of at least 1,5mm2 minimum. It is very important that the equipotential
bonding is installed correctly.
Please note that some electrical devices (eg. frequency converter etc.) close to the
system can create considerable electrical disturbance and instability. So
for normal function of the OXITEC 5000 System it is important that these devices
be installed away from the system and in accordance with the manufacturers
installation notes and guidelines.
Please do not lay the probe cable near power supply cables, or near motors and
their cables, as they are controlled by frequency converters because this could
cause interference effects. If possible don’t cross the cables because this could
cause interference effects. However, if this is unavoidable please make sure that
this cables are only crossed at right angles.
OXITEC® 5000 O2 Analyser System with SME 5 Electronics Version 8.129-116np Page 17
back view electronic unit with pneumatic unit (19" rack)
testgas testgas
output input X1 X2 X3
PE
0 1
cable glands
PG11 PG13,5 blind covers
pneumatic unit testgas testgas reference air reference air test air pneumatic unit testgas testgas instrument air reference air
reference air pumpinput output input output input instrument air input output input output
pneumatic unit testgas testgas instrument air reference air
instrument air input output input output
Attention:
plastic pneumatic tube should only be
screwed with support sleeves
nut support sleeve clamp ring
OXITEC® 5000 O2 Analyser System with SME 5 Electronics Version 8.129-116np Page 18
10 Description of the SME 5 microprocessor electronics
As a standard, the electronics are supplied in a housing for wall mounting, with pneumatic system.
Two different pneumatic systems are available:
• an instrument-air version requiring the supply of clean, dry, oil-free instrument air at the point of
installation.
• a pump version using the surrounding ambient air.
To ensure the full functional range of the electronics, both versions require a test
gas unit with pressure reducer and flow limiter.
Flow measurement of the test and reference gases is performed electronically in the SME5 unit, and no manual
control or adjustment is required.
The electronics unit has an LED graphic display with back lighting.
To operate the electronics, there are softkeys and a numeric keypad.
The entire menu structure of the electronics is laid out for a simple user-friendly interface.
The bottom of the housing has screw connectors for the cables and pneumatic lines.
Optionally, the SME 5 electronics are also available in a 19” rack or in an EExd housing.
The SME 5 electronics unit is designed for connection to an oxygen measuring probe. The mV signal of
the measuring cell is amplified and then converted into a digital signal in the electronics (14 Bit
converter). The µ-processor linearises and corrects the signal according to Nernst’s equation of the
measuring cell. The O2 value is then displayed as a percent value and the output has an isolated output,
either 0 - 20 mA or 4 - 20 mA - keypad selectable.
As a standard, the electronics has two measuring ranges. These measuring ranges can be freely defined
in their initial and/or final value (within permissible system-related limits).
The second measuring range can be switched on from outside (digital input see side 20 )
To the outside the second measuring range is signalled by a relay contact.
The electronics is completely self-controlled. Occurring system errors are displayed as clear text with on-
line help function information.
The ENOTEC user interface permits intuitive operation by means of a self-explanatory menu structure.
The system is multi-lingual, presently the languages German and English are available. Spanish, Italian
and French are presently in preparation.
Until the O2 measuring probe has reached its set temperature, the System Error contact is open and the
display shows that the probe is heating up. The progress of the heating phase is shown by a bar
indicating the current probe temperature.
During the heating phase, all system parameters, as well as actual values in the corresponding menus,
can be selected on demand from the key pad.
When the probe has reached its operating temperature and is stable within ± 0.5°C, actual
measurements are read, the System Error contact closes and the standard actual value display appears
on the screen.
OXITEC® 5000 O2 Analyser System with SME 5 Electronics Version 8.129-116np Page 19
10.3 Terminal Connections for the SME 5 Electronics
17 18 19 20 21 22 23 24 25 26 27 28 29 30
white/brown
brown
green
gn/ye
white
black
blue
L N P L N P L N P + - + -
E E E
Current output 0 (4) …
Calibration permission
Measurement range
Input 12-24V
Input 12-24V
Probe valve
supply.
contact
contact
20mA
contact
Figure 8 Terminal
Connection 17 18 19 20 21 22 23 24 25 26 27 28 29 30
AB AB AB AB AB AB AB AB AB AB AB AB AB AB
B - N - - - - -
A + L + + + + +
10.3.1 Terminal connection at installed options
Where options are selected, terminals 28 to 30 have different functions depending on the option chosen.
WARNING: The external current for the inputs (28 to 30) must be electrically Isolated
from Ground. Otherwise a malfunction or damage to the electronics is possible.
Do not exceed the max. current of 20mA.
OXITEC® 5000 O2 Analyser System with SME 5 Electronics Version 8.129-116np Page 20
X1 X2 X3
PE
0 1
X1 X2 status signals
1 - probe solenoid valve L 1 - maintenance
2 - probe solenoid valve N 2 -
3 - PE 3 - system fault
4 - probe heater L (black) 4 -
5 - probe heater N (blue) 5 - measure range
6 - PE 6 -
7 - cell + (brown) 7 - limit 1
8 - cell - (white) 8 -
9 - thermocouple + (green) 9 - limit 2
10 - thermocouple - (white) 10 -
11 - 11 - Dig.Input calibration release12-24V +
12 - output 0/4 - 20 mA + 12 - Dig.Input calibration release 12-24V -
13 - output 0/4 - 20 mA - 13 - Dig.Input measure range switch 12-24V+
14 - 14 - Dig.Input measure range switch 12-24V-
OXITEC® 5000 O2 Analyser System with SME 5 Electronics Version 8.129-116np Page 21
10.4 connection diagram
Terminal OXITEC 5000 - Elektronic
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30
white/brown
required.
Power Supply
brown
green
white
230V~ bzw.115V~
black
blue
10A
PE
N
L
+
-
Current output 0(4)..20mA
1 2 3 4 5 6 7 8
brown
green
black
white
blue
Figure 10
The colour code of the wires for the thermocouple has changed from the old
German standard : (+) red and (–) green to the European standard DIN / IEC 584:
(+) green
(-) white
The result is that there are two white wires in our special cable FEP-0001.
Don’t mix up the two white wires. For differentiation on of the white wire is marked with a brown line, this
wire belong to the brown wire.
OXITEC® 5000 O2 Analyser System with SME 5 Electronics Version 8.129-116np Page 22
Table 1 International colour codes for NiCrNi thermocouple wires
polarity of thermocouple DIN IEC 584 DIN 43710 (old german) ANSI MC 96.1(US)
+ green red yellow
- white green red
Figure 11
OXITEC® 5000 O2 Analyser System with SME 5 Electronics Version 8.129-116np Page 23
11 FUNCTION OF THE SME5 MICROPROCESSOR ELECTRONICS
Figure 12
Figure 13
Various options can be selected from a list (as in Figure 13) with the softkeys [Up] and [Down]. If there
more than 4 options, this is indicated by an arrow on the right hand side of the menu list. This arrow
indicates that there are more options to follow. To select one of the options, it is highlighted and the
selection confirmed with the softkey [Enter].
OXITEC® 5000 O2 Analyser System with SME 5 Electronics Version 8.129-116np Page 24
11.3 The heating phase
After switching on the electronics, the OXITEC logo will appear on the display for a short time. During this
time, all parameters relevant for a faultless functioning and start up are checked automatically.
If during this check any system errors are found, the electronic unit will show them in clear text. (see 11.8)
If the check shows no problems, the heater window appears. In this heating window the progress of the
heating phase is indicated by a bar (0°C to 900°C).
When the probe temperature is stable within ±0.5°C (after approx. 20 to 60 minutes, depending on the
process conditions), O2 measurement is released and the main windows appears on the display.
During the heating phase, all system parameters as well as actual values can be called up in the
corresponding menu windows.
Figure 15
In the main window (Figure 15) the currently measured oxygen value is displayed. A bar graph shows the
measuring range of the output signal. At a moisture of > 0% "oxygen dry" is displayed to indicate that the
O2 Value is an corrected Value.
• Status resp. MR1/2 : With this key in faultless operation the measure range can be selected.
Otherwise display possible errors or alarms.
• System : System menu, where system settings, calibrations, system tests etc. can be
made
• Actual values : Display of all actual values of the system
• Parameters : In this menu, parameters are shown and can be altered.
OXITEC® 5000 O2 Analyser System with SME 5 Electronics Version 8.129-116np Page 25
11.5 Actual value menu
Figure 16
In the actual value menu the following current actual values of the electronics unit can be called up:
• cell voltage [mV]
• cell temperature [°C]
• current output [mA]
• oxygen dry [%]
• oxygen wet [%]
• lambda
• reference point temperature [°C]
• reference gas flow [l/h]
• proces pressure [bar] (only at option process pressure correction)
• proces pressure [psi] (only at option process pressure correction)
• burner load [%] (only at option limit curves)
• actual limit1[%] (only at option limit curves)
• actual limit2 [%] (only at option limit curves)
• fuel (only at option limit curves)
If instead of a value only underscores are shown, this means that presently no valid value for this
parameter can be shown.
11.5.1 Lambda
The value of lambda (air ratio) is be calculated by an approximate equation and is only valid by
combustible free of own CO2 or N2 content.
21
λ≈
21 − O2
Approximate equation
OXITEC® 5000 O2 Analyser System with SME 5 Electronics Version 8.129-116np Page 26
11.6 Parameter menu
Figure 17
In the parameter menu the system parameters can be displayed and altered.
After having selected a parameter for the first time with the softkey [Enter] a window appears requesting
the entry of a code. After having entered the valid service code, there appears a window where the
corresponding value can be altered.
OXITEC® 5000 O2 Analyser System with SME 5 Electronics Version 8.129-116np Page 27
11.6.1 Altering parameters
By means of the parameter “Limit value 1” the entry of parameters will be explained.
Figure 18
In the window Parameter Entry, the parameter to be altered is displayed in clear text with the current
value. If special functions are assigned to this parameter (as in this example), these are displayed, too.
Further, the permissible entry limits of this parameter are indicated.
With the ten-key keypad the value can be altered. To correct, the cursor can be moved back with the [←]
key. The sign can be selected independently of the cursor position with the [±] key.
When pressing the [Back] key, the altered value is cancelled and the old value
maintained.
The measuring ranges of the electronics can be keypad entered. In the menu item “Measuring range
from” the initial value of the measuring range can be entered.
It should be noted that the initial value of the measuring range cannot be greater
than the final value. Also, the final value cannot be smaller than the initial value.
Therefore the electronics will not accept any such values.
Also, the second measuring range must always be smaller than the first one.
Therefore, the electronics will not accept such settings, either.
The final value of the measuring range can be entered in the menu item “Measuring range up to”.
If an moisture value greater than 0% entered, the electronic indicated in the main
window that this is an corrected value.
OXITEC® 5000 O2 Analyser System with SME 5 Electronics Version 8.129-116np Page 28
11.6.4 limits, limit curves (option)
The standard electronic unit is fitted with two adjustable limits.
The limits can be configured as min- or max-alarm. An exceed of a limit will be signalised by relay
(brake contact). This means that the contact is closed in "normal" use, if a limit exceed the contact
opened.
As an option, the electronic unit is available with limit adjustable curves for a maximum of different 4
fuels.
At digital input 1 and 2 the fuel should be selected. This selection has following coding:
Digital- Digital-
Input 2 Input 1
fuel 1 0 0
fuel 2 0 1
fuel 3 1 0
fuel 4 1 1
WARNING: The external current for the inputs must be electrically Isolated from
Ground. Otherwise a malfunction or damage to the electronics is possible.
Do not exceed the max. current of 20mA.
3,5
2,5
Vol % O2
2 limit curve 1
1,5 limit curve 2
0,5
0
20 40 60 80 100
burner load
The actual burner load and the corresponding limits could be shown in the actual value menu.
OXITEC® 5000 O2 Analyser System with SME 5 Electronics Version 8.129-116np Page 29
11.6.5 max range pressure (option)
In most applications the process pressure will be negative relative to ambient conditions. However, If the
process pressure is positive and has a greater than 50mbar differential to ambient, then a process partial
pressure correction will be necessary.
The electronics has an analogue input for the actual process pressure using a standard 4-20mA input
signal. The maximum range of this input signal, corresponding to 20mA, can be adjusted here. Note the
relative pressure differential. The starting value for the 4-20 mA input is always Zero (0) bar standard
gauge pressure.
WARNING: The external current for this input must be electrically Isolated from
ground. Otherwise a malfunction or damage to the electronics is possible.
Do not exceed the max. current of 20mA.
OXITEC® 5000 O2 Analyser System with SME 5 Electronics Version 8.129-116np Page 30
11.7 System menu
When selecting the system menu with the corresponding softkey, a code enquiry first appears, where the
system code (factory-set to 0000) must be entered and confirmed with [ENTER].
Figure 20
After having entered the system code, the system selection menu appears. Figure 21
Figure 21
In this menu, the following functions can be called up:
• Single point calibration
• Two point calibration
• Automatic calibration
• Measuring value memory
• System test with test gas
• Setting current output
• Measuring range shift
• Change code
• Damping mA output
• Call up max./min. values
• Change language
• Service
• Software version
OXITEC® 5000 O2 Analyser System with SME 5 Electronics Version 8.129-116np Page 31
11.7.1 Single point calibration
Figure 22
In this menu item, a single point calibration with air can be made. After selecting this menu item you are
requested to begin the calibration with the softkey [Start]. The single point calibration is fully automatic, no
manual intervention is required.* The progress of calibration is indicated by a moving time bar.
The amount of test air is monitored by the electronics, if it is insufficient, an error message appears.(only
by integrated pneumatic unit)
During calibration it is possible to accept the value of the cell voltage with [Accept] and to have the
calibration parameters calculated in advance.
It should however be noted that the advance acceptance of the cell voltage is not recommended,
because the electronic system then does not know if the output is stable, and therefore may not be
correct.
* If the electronics are not connected to a probe with solenoid valve, the test gas
valve must be opened prior to calibration.
Figure 23
In this menu item, a Two point calibration with air and with a suitable test gas concentration, can be
performed. It should be noted that the concentration of the test gas must correspond to the value
shown in the parameter menu.
After having selected this menu item, you are requested to begin the calibration with the softkey [Start].
Like the single point calibration, the Two point calibration is fully automatic.* The progress is indicated by
a moving time bar.
If the amount of test gas seems insufficient during calibration, the message “Test gas flow insufficient”
appears. In this window then the amount of test gas can be adjusted by means of a bar graph display.
After having adjusted the test gas flow, calibration is continued with [OK]..(only by integrated pneumatic
unit)
OXITEC® 5000 O2 Analyser System with SME 5 Electronics Version 8.129-116np Page 32
During the Two point calibration, the cell voltage can be accepted in advance with [Accept], but it should
be noted that for the Two point calibration two cell voltage values are required. It may therefore occur that
after the manual acceptance of a value, the calibration progress bar is reduced to half. This is not an
error, but means that the electronics are now determining the second value, which can be accepted
manually.
It should, however, be noted that an advance acceptance of the cell voltage is not recommended,
because the electronic system then does not know if the output is stable and may therefore not be
correct.
If the parameters determined after calibration are outside the plausibility limits, this is indicated by an
error message. In case of errors during calibration the electronics will continue to work with the old
parameters.
* If the electronics are not connected to a probe with solenoid valve, the test gas
valve must be opened prior to calibration.
OXITEC 5000 offers the possibility of automatically performing a calibration at regular intervals.
All settings that are required for an automatic calibration can be made in this menu.
These are:
• Type of calibration Single Point/Two Point
In this point the type of calibration can be selected. (Only in Systems with integrated Pneumatic)
• Automatic calibration On/Off
As a standard, ACAL is off. In this item, ACAL can be activated.
• Calibration every xx days (e.g. 90 days)
Indication of the calibration cycle time in days. At an value of 0 days the calibration starts immediately
upon release, this is important for an calibration from the outside.
• Release internal/external
If External is selected, a calibration is made only when the digital input Calibration release has a
“high” signal and the cycle time has elapsed. See Chapter 12.5
Internal means that independently of the release input status, calibration will be made when the cycle
time has elapsed. If the cycle time has elapsed, but no release has been given, calibration starts
immediately upon release.
• Remaining time xx days
The remaining time until the next calibration is indicated under this item.
For automatic calibration, it is essential to connect a probe with solenoid valve and
to ensure that the test gas “Air” is always available and under pressure (instrument
air version) or a pneumatic unit with pumps must be installed (for details see Annex
.... Gas plans).
WARNING: If the Calibration Time Interval in the parameter menu is set to zero Secs,
for example to make a remote calibration, it is very important that the Release Signal
has a minimum time setting of 15 seconds and a maximum time of 60 sec. A short
Release Signal (<15 Secs) the calibration will not start, and with too long Release
Signal (>60 Secs) it could be that the calibration function restarts for a second
time, or more.
NOTE: There is no "real time" clock in the electronics. It does not know time of day, month or
year. Only a clock counter in electronically used for automatic calibration purposes.
OXITEC® 5000 O2 Analyser System with SME 5 Electronics Version 8.129-116np Page 33
11.7.4 Measuring value memory
Under this menu item a measuring value memory can be switched on during calibration.
With the measuring value memory on, the last measured value prior to calibration is saved so that at the
current output the measuring value which changes because of the calibration, has no effect. The
currently measured value is released again after an adjustable delay, when the calibration is completed.
Figure 24
To perform a system test, in this menu the test gas on the probe can be entered.*
This function therefore permits a system test.
Furthermore, this menu permits to adjust the correct amount of test gas required for Two point calibration.
* If the electronics are not connected to a probe with solenoid valve, the test gas
valve must be opened prior to calibration.
11.7.6 System test with test air (Only in Systems with integrated pneumatic)
To perform a system test with test air, in this menu the test air on the probe can be entered.*
This function therefore permits a system test.
Furthermore, this menu permits to adjust the correct amount of test air required for calibration.
* If the electronics are not connected to a probe with solenoid valve, the test gas
valve must be opened prior to calibration.
Here, the initial value of the current output can be selected with the keys [Up] and [Down] (0 mA or 4 mA).
OXITEC® 5000 O2 Analyser System with SME 5 Electronics Version 8.129-116np Page 34
11.7.8 Measuring Range Shift
The electronics has two measuring ranges. The switching mode of the ranges can be made manually
from the instrument keypad, or by an external remote 12- 24 voltage input signal. This microprocessor
parameter location is used to determine which mode of range switching is required - Manual or Remote.
The factory preset default in the electronics is "Remote" range switching. This means that in the case of
an open digital input, the measuring range 1 is active. At the modus "local" the measuring range can
be changed in the main menu with the soft key MR1/2 by hand. In case of errors or alarms, the soft key
MR1/2 in the main menu is occupied by status information. The status function has the highest
priority, Therefore this means in case of errors the measuring range can not be changed by hand.
In case of errors or alarms, the soft key MR1/2 in the main menu is occupied by
status information. The status function has the highest priority, Therefore this
means in case of errors the measuring range can not be changed by hand.
The electronics continuously save the smallest and greatest measured values. These values can be
called up here. After quitting this item, the current min. and/or max. values can be deleted in order to
restart recording minima and maxima from this point.
11.7.13 Service
This menu item is reserved for Enotec service personnel only, it is factory-coded.
11.7.14 Software
OXITEC® 5000 O2 Analyser System with SME 5 Electronics Version 8.129-116np Page 35
11.8 Status messages
Figure 25
If errors or alarms occur during operation, these are displayed as clear text in the status menu.
All errors and alarms must be acknowledged in this menu.
A help function for the status messages is available in this menu.
If the electronics is put into operation without a probe connected, some system errors
will occur. E.g.. heating system don’t work, error cell signal out of range, error wire
breakage thermoelement. When the electronics is then connected to a probe, these
errors will be shown again, because the microprocessor stores the errors as they occur
in a permanent memory (EEPROM). After acknowledgement of each of these errors,
the system operates normally."
OXITEC® 5000 O2 Analyser System with SME 5 Electronics Version 8.129-116np Page 36
The temperature of the electronic is outside the allowed operating limits (-20°C to +70°C
within its housing)
9. probe temperature to high
The indicated temperature of the probe is over 890°C. Please test the temperature of the
flue gas (maximum 890°C, without cooling tube) Please contact ENOTEC.
10. probe temperature to low
The temperature of the probe was or is too low (<820°C) Please look at the heater fuse and
measure the actual power supply voltage to the electronics unit(230V/115V ±10%).
11. alarm limit1
Oxygen limit 1 out of range.
12. alarm limit2
Oxygen limit 2 out of range.
13. data fault EEProm
A data error has occurred in the microprocessor and the electronics continues with factory
default values. Please contact ENOTEC.
14. measuring value memory active
The measuring value memory is still active.
15. error by trying ACAL
Automatic Calibration has failed. Possible reasons: mV output value/signal from probe not
stable, cell constant or Mv/Dec slope outside of microprocessor tolerance/range. (Note: the
corresponding error is shown on the electronics display), no Calibration air (instrument air
version only) System descriptions / Settings
OXITEC® 5000 O2 Analyser System with SME 5 Electronics Version 8.129-116np Page 37
12 System description / Adjustments
On the back side of the display unit an bridge exist. This bridge is use for the first
run in factory. Don’t move or remove this bridge, otherwise an fault function of the
electronic is possible. Therefore contact to ENOTEC is required.
Postion bridge
for working
Figure 26
OXITEC® 5000 O2 Analyser System with SME 5 Electronics Version 8.129-116np Page 38
12.3 Adjustment display contrast
The contrast of the display can be adjusted if necessary. This could be necessary at higher ambient
temperatures. The trimming potentiometer is shown on Figure 27. To adjust the contrast it is necessary to
dismount the display board and power on the system. Now the contast can be adjusted by a small screw
driver (type 2.5)
Attention, working under tension is only allowed for authorized, trained skilled
persons in keep to the corresponding safety instructions. (eg VDE 0105)
trimming
potentiometer
Adjust contrast
Figure 27
OXITEC® 5000 O2 Analyser System with SME 5 Electronics Version 8.129-116np Page 39
12.4 Relay outputs / Functions
The relay contacts are designed for 230 V and 10A ~ , 5A =.
The digital inputs are designed for a direct voltage of 12 V-30 V for logical high.
Logical "Low" agree an voltage less than 1 V.
Where option limit curves selected, there are folowing additional digital inputs
WARNING: The external current for the inputs (28 to 30) must be electrically
Isolated from Ground. Otherwise a malfunction or damage to the electronics is
possible.
Do not exceed the max. current of 20mA.
The instrument air versions are designed for an admission pressure of 1-10 bar, with a higher admission
pressure as 6bar it is necessary to readjust the amount of reference air and/or test air.(see Chapter 14.3)
The amount of air should be in the following range:
Test air: 150l/h - 180l/h
Reference air: 30l/h - 40l/h
The amount of reference air can be checked in the Actual value menu.
To adjust the amount of calibration air, the test air line must be disconnected from the electronics housing
and a flow meter can be connected to the test air output.
OXITEC® 5000 O2 Analyser System with SME 5 Electronics Version 8.129-116np Page 40
12.7 Stability criteria for calibration
During calibration, the cell voltage is checked for stability. This check works according to the following
criteria.
Always the last measured value is temporarily saved, when the next value is outside the tolerance, the
internal timer is reset and the new value temporarily saved. This means that if the timer was not reset, the
value is stable. In this way, the last measured value after the timer has lapsed (2 min.) is used to
calculate the constant or slope.
The reaction time of the mA output to a change of the O2 cell input voltage is less than 175 ms.
The transfer protocol contains a timer index, this timer is increased every second after power on. It is
possible to count the time of one year.
The oxygen is shown by the protocol only in a ppm range.
......[STX]tim:+0000000.0secoxi:+0000000.0ppmcell+0000003.9mVtemp+0000006.9°Cout:+000
0000.0mA[ETX][chk][CR][LF][STX]tim:+0000001.0secoxi:+0014846.3ppmcell+0000012.0mVte
mp+0000014.3°Cout:+0000021.3mA[ETX][chk][CR][LF][STX]tim:+0000001.0secoxi:+0003061.8
ppmcell+0000009.1mVtemp+0000022.1°Cout:+0000021.3mA[ETX][chk][CR][LF]......
The character "STX" is the beginning of the following ACII string, the character "ETX" is the end of the
ACII string.
One data sequence beginns always with a short description (for example : oxi) by following the value (for
example: +0003061.8) the end of the data sequence is the measure unit (for example: ppm).
For example a complete sequence "oxi+0003061.8ppm".
OXITEC® 5000 O2 Analyser System with SME 5 Electronics Version 8.129-116np Page 41
13 Technical Specification
Reaction time of the mA output: With a change of 100 mV on the cell input < 175 ms
OXITEC® 5000 O2 Analyser System with SME 5 Electronics Version 8.129-116np Page 42
13.2 Technical Specification Probe
KES-2001: 495mm
KES-2002: 925mm
KES-2003: 1835mm
KES-2004: 2768mm
KES-2005: 3682mm
KES5001: 520mm
KES5002: 950mm
KES5003: 1865mm
KEX5001: 520mm
KEX5002: 950mm
Response time (lag time): 0,5s (flue gas flowing at > 10m/second)
OXITEC® 5000 O2 Analyser System with SME 5 Electronics Version 8.129-116np Page 43
14 Service and Maintenance
The working voltage can only be changed after removing the board cover.
The switch for changing the voltage (Figure 28) can be made with a screwdriver and rotating to the
voltage required.
selection of
power supply
F3
F1 F2 F4
Figure 28
It must be noted that the standard version with instrument air of the OXITEC 5000
is designed for 50/60 Hz.
For the pump version on 60 Hz mains, a special order is necessary. The mains
frequency cannot be changed by the customer.
The fuses on the circuit board below the cover protect the electronics and should
only be replaced by service personnel, because otherwise a damage of the
electronics cannot be excluded.
OXITEC® 5000 O2 Analyser System with SME 5 Electronics Version 8.129-116np Page 44
14.3 Adjusting the flow rate of reference and test air
At a System with a pneumatic unit it is possible to adjust the reference and test air.
In this point the pump and instrumental air versions distinguished.
• At the pump version only the flow rate of test air can adjusted. (see Figure 29)
• At the instrumental air version both flow rate (test air and reference air) can be adjusted. (see
Figure 30)
The flow rate of reference air can be read in the actual menu (see page 26)
The flow rate of calibration air can be read in the system menu under the menu point "system test with
test air" (see page 34)
OXITEC® 5000 O2 Analyser System with SME 5 Electronics Version 8.129-116np Page 45
14.4 Replacing Filter
Figure 31
2. Insert new filter with cement as follows:
Mix cement thoroughly and coat both the grooves in the seat and the filter around the seat at
the outside. Press the filter into the seat by turning it. Wipe the cement between filter and seat
and remove excess material inside as well.
EExd IIC3 probes have a sintered metal filter that is welded to the filter head
assembly.
Take the probe out of the protection tupe and wait until it has cooled down
If necessary, clean the probe from dirt and dust etc., after it is cooled down!
Pull out the probe inner parts (4-hole-ceramic-rod with signal -measuring wire, thermocouple element and
heater) by disconnection the wires in the connection head and by loosening the two screws which hold
the inner parts at the tube.
Pull off the thin transparent reference air pipe from the wall entrance screwing at the connection cabinet.
Carefully pull out probe internal part.
ATTENTION ! The measuring signal wire can be fixed solid with its
contact net to the inner part of the ZrO2-cell.
Pull off the orange isolation from the measuring signal wire and do not pull out the platinum wire with the
probe inner parts, if it is fixed solid to the internal cell contact side.
Loosen the 4 screws at the flange of the measuring cell and remove the measuring cell from the
measuring cell tube (with signal wire, if caked solid)
OXITEC® 5000 O2 Analyser System with SME 5 Electronics Version 8.129-116np Page 46
measuring cell holder tube
measuring cell
measuring signal wire
screwdriver blade
place twist
Anyhow, if the measuring cell is defect, the ZrO2-ceramic plate should be carefully broken in
order to loosen the platinum wire, if it is caked solid inside.
Lead the platinum wire into the 4-hole-rod and make sure, that the round platinum net lies in the
recess of the heating sleeve.
protection cap
4 screws M5x12
figure 33 mounting measure cell holder tube
Insert the probe inner parts and ensure, that the inner parts do not clamp in the probe tube.
Press the locking bolt against the spring, so that the inner parts are pressed against the
measuring cell with enough spring load.
OXITEC® 5000 O2 Analyser System with SME 5 Electronics Version 8.129-116np Page 47
Tighten locking bolt and connect wires as follows:
polarity of thermocouple DIN IEC 584 DIN 43710 (old german) ANSI MC 96.1(US)
+ green red yellow
- white green red
Now heat up probe outside the process to operating temperature and tighten the 4 screws of
the measuring cell flange-sealing.
ATTENTION!
Also at the outside the probe is hot (300°C).
Install the probe and take up a curve with test gases (see 11.7.2).
1 2 3 4 5 6 7 8 9 10 11 12
13
figure 34 construction of the measure cell holder tube
OXITEC® 5000 O2 Analyser System with SME 5 Electronics Version 8.129-116np Page 48
14.6 Exchange of probe inner parts
Proceed as described in paragraph 14.5, apart from exchange of thermo-couple element.
2. Exchange of heating
Disassemble probe inner part, loosen clamp, pull out platinum wire and pull off heating
with support from the 4-hole rod.
Assemble new heating, if necessary with new ceramic support and new 4-hole ceramic
rod.
Insert platinum wire and thermo-couple element, assemble clamps and connection
sleeves (for long probes).
ATTENTION:
The heating support has 3 drillings, which must correspond with 3
holes of the 4-hole ceramic rod. One hole is for the reference air and
the other for the platinum measuring wire, so that the platinum net
lies in the recess of the heating support.
At the end of the measuring cell the platinum measuring wire has a platinum net for
taking the mV-value of the reference electrode of the ZrO2-measuring cell. The platinum
net is pressed against the measuring cell by means of spring load via the 4-hole rod.
If there is not enough spring load or if the wire is broken, the mV-signal can no longer be
taken.
Depending on the length the probe has one or more 4-hole rods. In case of long probes
you have to ensure during assembling, that the holes fit to another, so that installation
can be carried out and the reference air can be lead to the inner reference air side of the
O2-measuring cell.
15 Trouble shooting
OXITEC® 5000 O2 Analyser System with SME 5 Electronics Version 8.129-116np Page 49
The following information is intended as a key element to assist trouble shooting. It is not a
comprehensive list of all possible system errors.
ENOTEC GmbH
Marienheide
Höher Birken 6
D-51709 Marienheide - Rodt
Germany
If the measured O2 value appears doubtful, first the ENOTEC-O2 analyser system should first be checked
with test gas, if all system parameters are o.k.. If the system works correctly with test gas and there are
no leakages in the area of the probe (flanges, calibration gas connection), then the measurement is
correct.
If for instance certain system parameters are wrong, then the measured O2 value cannot be correct and
with the following instructions it should be tried to localize or repair the disorder.
1. Check the measuring cell temperature (set value 840°C, or optionally 750 °C).
A cell temperature below 600°C may lead to a 0% display.
Causes:
a) Measuring probe heater defective (resistance must be approx. 35-42 Ohm, disconnect probe and
check). Note: De-energize probe first.
b) Thermocouple defective (check resistance, approx. 2-80 Ohm)
c) Fuse for heater voltage defective.
d) Transformer (230/115V) defective, check voltages.
e) Triggering for temperature control or electrical load relay defective.
Result 1:
The mV value is outside the usual range (see characteristic line) e.g. above 267 mV or below - 45 mV.
OXITEC® 5000 O2 Analyser System with SME 5 Electronics Version 8.129-116np Page 50
Result 1:
If on the probe a mV value is measured that according to the characteristic line and at the right measuring
temperature yields a probable O2 value, then the error is on the way to the electronics.
Possible causes:
a) Cable short circuit
b) Electronics input defective
c) Wire break
Remedies:
a) Check wiring
b) Measure probe cable
Result 2:
On the probe a mV value is measured which is outside the usual range from -45 mV to 267 mV.
Possible causes:
a) There is no mV tap in the probe (measuring signal wire) or it is interrupted.
b) Combustibles in the flue gas.
c) Measuring cell defective
Remedy: Check whether the probe reacts to test gas. If it does, there may be a high portion of
combustibles in the flue gas. In this case, there are reducing conditions on the probe cell
which reduce the oxygen content on the cell surface. Caution: Explosion hazard!
If the probe does not react, check the inner part.
1. Check measuring range. Check whether the current value is outside the measuring range
2. Measure the mA output on the strip terminal.
Result 2: no mA value
Possible causes:
Electronics defective.
Measures:
Measure the mV value on the terminals on the board.
Causes:
a) Intermittent contact caused by wire break.
Remedies:
a) Repair intermittent contact.
OXITEC® 5000 O2 Analyser System with SME 5 Electronics Version 8.129-116np Page 51
15.4 Display remains at the end of the measuring range or is higher than
expected
1. All system parameters are correct and the ENOTEC O2 analyser system shows the correct value with
test gas.
Cause: Leakages at the measuring probe or at the measuring cell flange seal.
Remedy: Exchange measuring cell or replace measuring cell flange seal. In case of a leakage in the
area of the zirconium oxide measuring cell, this must be exchanged.
OXITEC® 5000 O2 Analyser System with SME 5 Electronics Version 8.129-116np Page 52
16 Dimensional sheets of the evaluation electronics
300 í8,2
240
ENOTEC
400
420
OXITEC 5000
All dimensions in mm
260
Figure 35 Field housing IP65
177 (4HE)
483 360
510 310
330
150
M10
ENOT EC
450
OXITEC 5000
660
All dimansions in mm
OXITEC® 5000 O2 Analyser System with SME 5 Electronics Version 8.129-116np Page 53
17 Dimensional sheets for probes
test gas valve KES-1320010 filter head filter
A
OXITEC
Sauerstoffanalysensysteme
ENOTEC GmbH
51709 Marienheide / Germany
Modell
Zell-Nr.
Ser.- Nr.
V-shield
A (immersion depth)
A
View A - A
connection box * note: í 9
by mounting with the
148 adapter plate ASR-1320
the immersion depth
decrease around 100 mm
Adapter plate
for KES132.1-3 Mating flange mounting kit
with stand off tube; length 100 mm(4");diameter 76,1x2 mm
Article-No.: ASR-1320 45
adapter plate
í76,1*2
with 4 bolts
M8 * 29
ST 37-2
DN40/PN6
99 16
130 100
OXITEC® 5000 O2 Analyser System with SME 5 Electronics Version 8.129-116np Page 54
test gas valve (option)
A 120 B (immersion depth )
protection tube flange A
protection tube filter
í76,1
OXITEC
Sauerstoffanalysensysteme
ENOTEC GmbH
51709 Marienheide / Germany
Modell
Zell-Nr.
Ser.- Nr.
C V-shield
connection box D A
View A - A
18
dimensions in mm
type A B C D protection tube flange
4" ASA 150 LBS
KES/KIS 2001 135 495 380 615
KES/KIS 2002 150 925 800 1045
76,1
191
230
KES/KIS 2003 150 1835 1720 1955
KES/KIS 2004 150 2768 2648 2888
KES/KIS 2005 150 3682 3562 3802
KES/KIS 2006 150 4602 4482 4722
KES/KIS 2007 150 5522 5402 5642
flue gas direction all dimensions in mm
KES5001 520
KES5002 950 view A - A
90.0ø
150 (5002-5003)
135 (5001) 120
KES5003 1865 d2
test gas
all dimensions in mm
A
connection
B (immersion depth)
1/4"
í 57
OXITEC® 5000 O2 Analyser System with SME 5 Electronics Version 8.129-116np Page 55
adapter plate
mating flange mounting kit
Article-No.: ADP-2000
M16 18 45
adapter plate
with 8 bolts
M16 * 45
ST 37-2
4" ASA 150 lbs
OXITEC® 5000 O2 Analyser System with SME 5 Electronics Version 8.129-116np Page 56
17.1 Probe components
f h
3
measuring cell 7
g 4 6 d
2 flange gas ket 5
a b c e
measuring wire
Reference air tube article-no.: RLS-0001 article-no.: MSD-132X (for KES 1321-2
MSD-2001 (for KES 1323 a. 2001)
measuring wire MSD-200X (for KES 200X)
4-hole ceramic rod heater
article-no.: KER-132x (for KES 1321-1 article-no.: HEI-132x (for KES 1321-1
KER-2001 (for KES 1323 a. 2001) HEI-2001 (for KES 1323 a. 2001)
KER-200X (for KES 200X) HEI-200X (for KES 200X)
heater connection
heater support
article-no.: HZH-00001
spring reference
holding pins clamp spacer air
assy. thermocouple
article-no.: MT2-132X (for KES 1321-2
MT2-2001 (for KES 1323 a. 2001)
thermocouple connection screw kit for inner part MT2-200X (for KES 200X)
article-no.: RSI-1320 (for KES 1321-2)
RSI-2001 (for KES 2001)
RSI-200X (for KES 200X)
consisting of: holding pins with screws,spring
clamps,spacer assy.
OXITEC® 5000 O2 Analyser System with SME 5 Electronics Version 8.129-116np Page 57
18 pneumatic connection and mounting Overview
Flowmeter Flowmeter
2 2
solenoid 1 solenoid 1
valve valve
3 3
filter filter
pressure
regulator
reference air out
reference air in
test gas out
test gas in
max. 2 bar
test air in
max. 6 bar
OXITEC® 5000 O2 Analyser System with SME 5 Electronics Version 8.129-116np Page 58
mounting plate 1 (design with pumps)
0-A-000154
TRA-0017
0-P-000763
0-P-000757
PLU-0026
0-P-000751
0-P-000755
0-P-000777
Abstand zur MP : 15mm
0-A-000171
í 3,2 í 3,2
PLU-0025
í 3,2 í 3,2
DFM-00001
0-A-000158
0-P-000743
0-P-000758
0-P-000759
OXITEC® 5000 O2 Analyser System with SME 5 Electronics Version 8.129-116np Page 59
mounting plate 1 below (instrument air)
0-A-000154
TRA-0017
0-P-000763
0-P-000757
0-P-000751
0-P-000755
0-P-000745
0-P-000760
DFM-00001
0-P-000752
0-P-000753
0-A-000158
0-P-000758
0-P-000759
OXITEC® 5000 O2 Analyser System with SME 5 Electronics Version 8.129-116np Page 60
mounting plate 2 middle
0-L-000001
í 15
46.2 hoch
í10
0-P-000658
0-P-000658 test air
refrence air only at design
only at design with pumps
with pumps
0-A-000155
OXITEC® 5000 O2 Analyser System with SME 5 Electronics Version 8.129-116np Page 61
19 Index
µ-processor 19 mains voltage 42
19 19 maintenance 44
4-hole rod 49 measuring accurance 42
4-hole-rod 47, 49 measuring cell 46
actual values 25 measuring range 28, 42
adapter plate 13 measuring range shift 31, 35
altering parameters 28 measuring signal line 12
ambient temperature 17 measuring signal wire 46
amount of reference air 40 measuring value memory 31, 33
amounts of test air 40 oxygen 26
analog input 42 parameters 25
attenuation mA output 31 pneumatic cable 12
automatic calibration 31, 33 pneumatic tubing 16
calibration progress bar 33 power consumption 42
calibration release 40 probe inner part 49
call up max./min. values 31, 35 probe internal part 46
CE-certification 8 probe tube 47
CE-conformity 17, 23 probe valve 40
cell temperature 26 protection tube 11
cell voltage 26 pump version 12, 44, 45
change code 31 reaction time 41
change language 31, 35 recommended fuse 42
code 35 reference air 40
codes 48 reference point temperature 26
colour codes 23 relay outputs 40
conformity 5 replacing filter 46
connecting box 11 resistance to interference 42
connection diagram 22
resolution 42
current output 26, 34
service 31, 35, 44
digital inputs 40
Service 40
dimensions 42 setting current output 31
display 19 shield 16, 17
display unit 38
signal output 42
EExd housing 12
single point calibration 31, 32
electronics 19, 44
softkeys 19
electronics unit 17
software 35
extension modules 41
software version 31
external 33
stability criteria 41
factory-set 31
status 25
ferrite sleeves 8, 17, 23
status messages 36
field housing 17
switch for changing the voltage 44
filter head 11, 46
flow measurement 19 system 25
flue gas duct 13 system code 38
fuses 44 System error 40
heater 11 system menu 25
heating phase 25 system test 34
system test with test gas 31
heating support 49 technical specification 43
inner probe assembly part 11
instrument Air Version 44
temperature compensation 42
instrument-air 12 ten-key keypad 24
instrumental air version 45 terminal connections 20
internal 33 thermocouple 11, 23, 48
lambda 26 two point calibration 31, 32
limit value 1 40 underscores 26
limit value 2 40 valve 45
mains frequency 44 zirconium-oxide 9
OXITEC® 5000 O2 Analyser System with SME 5 Electronics Version 8.129-116np Page 62
Oxitec Distribution & Service :
ENOTEC GmbH
Höher Birken 6
D-51709 Marienheide
Germany
Enotec UK Limited
The Red House
84 High Street
Buntingford
Herts SG9 9AJ
England
ENOTEC Inc.
6206 Sandy Ridge Circle NW
North Canton
Ohio 44720
USA
Your Distribution:
ENOTEC ASIA PTE. Ltd.
Pasir Panjang Distripark
Block 1 Please contact:
#02-11 Pasir Panjang Road ENOTEC GmbH
Singapore 118480
OXITEC® 5000 O2 Analyser System with SME 5 Electronics Version 8.129-116np Page 63