0% found this document useful (0 votes)
123 views4 pages

Industrial Gas Analyzer Guide

The LGA-4000Z gas analyzer uses tunable diode laser absorption spectroscopy (TDLAS) to directly measure gas concentrations in pipelines without extracting samples. It shines a laser tuned to the target gas's absorption wavelength into the pipeline to detect the gas based on the Beer-Lambert law. The system consists of a transmitter, receiver, and connection box. It provides fast, sensitive measurements and has certifications allowing use in hazardous areas.

Uploaded by

01666754614
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
123 views4 pages

Industrial Gas Analyzer Guide

The LGA-4000Z gas analyzer uses tunable diode laser absorption spectroscopy (TDLAS) to directly measure gas concentrations in pipelines without extracting samples. It shines a laser tuned to the target gas's absorption wavelength into the pipeline to detect the gas based on the Beer-Lambert law. The system consists of a transmitter, receiver, and connection box. It provides fast, sensitive measurements and has certifications allowing use in hazardous areas.

Uploaded by

01666754614
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 4

MODEL LGA-4000Z

TUNABLE DIODE LASER GAS ANALYZER


Teledyne’s LGA-4000Z series leverages the latest in Tunable Diode Laser Absorption Spectroscopy (TDLAS) to achieve an
ultra-fast gas measurement in-line without resorting to extraction and sample conditioning. By tuning the laser to a specific
absorption wavelength, based on the gas of interest, the attenuation of the light through a sample cell and the detector can
be correlated to the concentration of each gas, per Beer-Lambert Law. The laser provides a narrow absorption band which
minimizes the effect of interfering compounds in the process

The System Wavelength Modulation Spectroscopy (WMS)


The LGA-4000Z consists of a transmitter, receiver and WMS is a method used to significantly increase the
connection box. The transmitter drives the tunable diode to sensitivity by reducing the low-frequency noise. The laser
emit a laser of a certain wavelength to pass through the light’s intensity is modulated at a high frequency to create a
environment being measured until it reaches the receiver. The periodic signal to the detector. The signal is then encoded and
receiver then performs a photo-electrical conversion, signal measured at a high frequency, eliminating the effect of low
processing and spectrum data analysis to obtain the frequency noise.
measuring result. The connection box, containing the I/O
Interface, makes the connection of the power cable and the
signal cable convenient.

* For extractive option only LGA-4000ZA

Features
• Flame proof design, IP-66 rated enclosure, ATEX and IECEx
• Zone 1 IIC T6 Certified
• Modular design simplifies service and field part process gas pipeline

replacements
• Integral automatic gain control (AGC) algorithm ensure Transmitting
instrument
flange
Welding
Flange roots of valve receiving
optimal signal integrity unit unit

• Excellent long-term drift characteristics, less than ±1% FS transmitting and receiving cables

over 6 months purge gas purge gas

• Local OLED display with magnetic buttons to maintain


area classification 8 Inlet Pipe

• Low power (rated 10W)


• 2 x 4-20mA outputs, 2 x alarm relays, RS-485 / RS-232/
GPRS communications

[email protected]
2 www.teledyne-ai.com
SPECIFICATIONS
Certifications ATEX and IECEx Zone 1 IIC T6

Path Length Typically 2-65 feet (0.6-20 meters), application-dependent

Repeatability ± 1% FS

Linearity ± 1% FS

Response Time Typically < 1 second (application-dependent)

Utility Gases Air or Nitrogen for purge (application-dependent), 30-115 psig (0.3-0.8 MPa)

Analog: 2 x 4-20mA (isolated), 750 ohm max load


Outputs Discrete: 2 x relay contact, Form-C type, rated 1A @ 24 VDC
Communications: RS-485 / RS-232

Inputs 2 x 4-20mA (for pressure and temperature compensation)

Power 24 VDC 10 W
Temperature -4° to 140°F (-20° to 60°C)

GAS DETECTION PARAMETERS


Gas Min Range Max Range Detection Limit
Acetylene (C2H2) 0-50 ppm 0-100% 0.5 ppm
Ammonia (NH3) 0-10 ppm 0-100% 0.5 ppm
Carbon Dioxide (CO2) 0-1000 ppm 0-100% 40 ppm

Carbon Monoxide (CO) 0-1000 ppm 0-100% 40 ppm

Ethylene (C2H4) 0-100 ppm 0-100% 1 ppm


Hydrochloric Acid (HCl) 0-100 ppm 0-50% 0.1 ppm

Hydrofluoric Acid (HF) 0-20 ppm 0-50% 0.05 ppm


Hydrogen Sulfide (H2S) 0-100 ppm 0-100% 5 ppm
Methane (CH4) 0-200 ppm 0-100% 10 ppm
Oxygen (O2) 0-1% 0-25% 0.01%
Water / Moisture (H2O) 0-100 ppm 0-100% 0.1 ppm

[email protected]
3 www.teledyne-ai.com
California Office colorado office Middle East Office Southeast Asia Office
City of industry, CA Englewood, co Sharjah petaling jaya
United states of america (USA) United states of america (USA) UNITED ARAB EMIRATES (UAE) MALAYSIA
TEL: +1 626.934.1500 TEL: +1 303.792.3300 TEL: +971.6557.9727 TEL: +603.7805.7712

[email protected] [email protected]
www.teledyne-ai.com www.teledyne-ml.com

Rev-A 07/30/2019

You might also like