KT800R Paperless Recorder Manual

This document is the user manual for the KT800R paperless recorder. It describes the product's functions, settings, wiring methods, operation methods, troubleshooting methods. The manual contains 16 chapters that cover an overview of functions, installation and wiring, operation instructions, configuration settings for the system, recording, inputs, outputs, alarms, displays, reports, printing, and communication. It also includes specifications for the recorder.

Uploaded by

Abdul Kurniadi
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
119 views141 pages

KT800R Paperless Recorder Manual

This document is the user manual for the KT800R paperless recorder. It describes the product's functions, settings, wiring methods, operation methods, troubleshooting methods. The manual contains 16 chapters that cover an overview of functions, installation and wiring, operation instructions, configuration settings for the system, recording, inputs, outputs, alarms, displays, reports, printing, and communication. It also includes specifications for the recorder.

Uploaded by

Abdul Kurniadi
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd

User’s

Manual

KT800R
USER’S MANUAL

IM01C82R01-01E
First Edition
FOREWORD
Thank you for purchasing our paperless recorder!
This manual is about the functions, settings, wiring methods, methods of
operation, failure of treatment methods of the paperless recorder. To ensure correct
use, please read this manual carefully and use properly before operation and keep this
manual in a safe place for quick reference.

Notice
 The contents of this manual are subject to change without prior notice as a result
of continuing upgrades to the instrument’s performance and functions.
 Every effort has been made in the preparation of this manual to ensure the
accuracy of its contents. However, if you have any questions or find any errors,
please feel free to contact us.
 Copying or reproducing all or any part of the contents of this manual without our
permission is strictly prohibited.

Revisions
February, 2012, First edition

I IM01C82R01-01E
SAFETY PRECAUTIONS
For your safe use of this instrument be sure to comply with the following safety
precautions. If the instrument is used in a manner not specified in this manual, the
protection provided by the instrument may be impaired. We assume no responsibility
for the customer’s failure to comply with these requirements.

WARNING
 Power Supply
Before connecting the power cord, ensure that the power supply voltage matches the
voltage rating for the instrument.
 Protective Grounding
Make sure to connect the protective grounding to prevent electric shock before
turning on the power.
 Necessity of Protective Grounding
Never cut off the internal or external protective grounding wire or disconnect the
wiring of the protective grounding terminal. Otherwise, it will pose a potential shock
hazard.
 Defect-Free Protection
Do not operate the instrument when the protective grounding might be defective. Also,
make sure to check them before operation.
 Do Not Operate in Explosive Atmosphere
Do not operate the instrument in the presence of flammable, explosive gas or vapors.
In such circumstances the use of this instrument is very dangerous.
 Do Not Remove Covers
Please do not tear down part of the front screws and open the front panel, because
some areas inside the instrument have high voltages .The cover should be removed by
our qualified personnel only.
 External Connection
Connect the protective grounding before connecting to the items under measurement
or external control circuit connections.
 Damage to the protection
Using the instrument in a manner not specified in this manual can damage the
instrument’s protection.

IM01C82R01-01E II
CHECKING THE CONTENTS OF THE PACKAGE
Unpack the wrapping box and check the contents before operating the instrument. If
some of the contents are not correct or missing or if there is any physical damage,
contact our company or the sales network from which you purchased it.

Accessories
The following accessories are supplied with the instrument. Make sure that all items
are present and undamaged.

Number Name Quantity Notes


Mounting
1 2 For panel mounting
bracket
2 User’s Manual 1 This manual
3 Certification 1 Manufacture Date
CD-ROM (Included only when the
Standard communication / U disk/ SD card
4 1
Software storage is ordered. For Windows
2000/XP/Windows Vista/Windows 7.
2m(Included only when the
Communication
5 1 Communication
Cord
is ordered.)
Included only when the external
6 U Disk 1 storage is ordered. Capacity and
model of the USB storage may vary.
Included only when the SD card
7 SD card 1 storage is ordered. Capacity and
model of the storage may vary.

III IM01C82R01-01E
Accessories (Sold Separately)
The following optional accessories are available for purchase separately. When
ordering, make sure whether there is a shortage or damage. For information about
ordering accessories, please contact us.

Number Name Quantity Notes


1 U Disk 1 1GB
2GB
4GB
2 SD Card 1 1GB
2GB
4GB

IM01C82R01-01E IV
HOW TO USE THIS MANUAL
This user's manual consists of the following 16 chapters:
Chapter Title and Contents
1 Overview of Functions
Describe the functions of the instrument
2 Before Using the Instrument
Describe the installation and wiring connection
3 Names of All Parts/ Mode of Usage/Common Operations
Describe the names of each part of the instrument, how to
use the storage medium drive, run mode, and common
key operations.
4 Switching Operation Screens
Describe how to use the operation screen such as the
trend display and digital display. Describes the
operations that can be performed using the arrow keys,
page key and the enter key on the front panel.
5 System Configuration and Its Function
Describe the related parameters of running system and its
functions
6 Record Configuration
Set the related parameters of the recording data
7 Input Configuration
Set the relevant parameters measured input channels
8 Output configuration
Set the relevant parameters of output transmitted
channel
9 Alarm Configuration
Set relevant parameters of alarms and relays.
10 Display configuration
Set related parameters of channel combinations and
display
11 Report Configuration
Set related parameters of channel accumulation and report
12 Print Configuration
Set related parameters of print function
13 Communication Configuration
Set parameters related to communications
14 Check System Information
Check system information, and AD board running
15 Import and Export Configuration File
Describe the import and export of configuration file
16 Specifications
List the specifications of the Paperless Recorder

V IM01C82R01-01E
Symbols Used in this Manual
Unit
K Denotes”1024”
k Denotes”1000”
M Denotes“1024K”
G Denotes“1024M”
B “Bytes”

Symbols (The following symbols are used in this manual)


WARNING: precautions that should be observed to prevent injury or death to
the user.

CAUTION: precautions that should be observed to prevent damage to the


instrument.

Note : Providing important information for the proper operation of the


instrument.

Operation of Marks
The following symbols are used during the explanation of operation
[ ] : represents key’s name. Example: [Page] [En]
『』: represents a reference item. Example: 『Input Channel』
【】 : indicates screen element .Example: 【Device Name】

IM01C82R01-01E VI
FOREWORD ..........................................................................................................I
SAFETY PRECAUTIONS ................................................................................... II
CHECKING THE CONTENTS OF THE PACKAGE ........................................ III
HOW TO USE THIS MANUAL .......................................................................... V
CHAPTER 1 OVERVIEW OF THE FUNCTIONS ...................................................... 1
1.1 OVERVIEW OF THE INSTRUMENT ........................................................... 1
1.2 FUNCTIONS OF THE INPUT SECTION ...................................................... 2
1.3 DISPLAY FUNCTION .................................................................................... 4
1.4 STORAGE FUNCTION. ................................................................................. 6
1.5ALARM FUNCTION ....................................................................................... 7
1.6 REPORT FUNCTIONS (ADDITONAL SPECIFICATIONS) ........................ 8
1.7 OUTPUT FUNCTIONS (ADDITIONAL SPECIFICATIONS) ...................... 9
1.8 COMMUNICATION FUNCTIONS (ADDITIONAL FUNCTIONS) .......... 10
1.9 OTHER FUNCTIONS ................................................................................... 11
CHAPTER 2 INSTALLATION AND WIRING .......................................................... 12
2.1 INSTRUMENT SITE .................................................................................... 12
2.2 INSTALLING THE INSTRUMENT ............................................................. 13
2.3 THE ARRANGEMENT OF VARIOUS TERMINALS ................................ 15
2.4 INPUT AND OUTPUT SIGNAL WIRING .................................................. 16
2.5 ALARM OUTPUT WIRING (OPTION) ...................................................... 19
2.6 CONNECT POWER SUPPLY ...................................................................... 21
2.7 CONNECT SWITCHING VALUE ............................................................... 22
2.8 24VDC TRANSMITTER POWER WIRING(OPTION) ......................... 23
2.9 RS232C/RS485 COMMUNICATION WIRING ........................................... 24
2.10 CONNECT SPECIAL SERIAL PRINTER ................................................. 25
CHAPTER 3 NAMES OF PARTS/RUN MODE/COMMON KEY OPERATIONS .. 26
3.1 NAMES OF PARTS AND FUNCTIONS ...................................................... 26
3.2 INSERTING/REMOVING THE EXTERNAL STORAGE MEDIUM ........ 27
3.3 RUN MODE .................................................................................................. 28
3.4 CONFIGURING THE FUNCTIONS ............................................................ 29
3.5 COMMON KEY OPERATIONS................................................................... 30
CHAPTER 4 DISPLAY SCREENS AND ITS FUNCTION ....................................... 37
4.1 SWITCHING OPERATION SCREENS. ...................................................... 37
4.2 EXPLANATION OF THE STATUS DISPLAY SECTION .......................... 38
4.3 DIGTAL DISPLAY SCREEN ....................................................................... 39
4.4 BAR GRAPH DISPLAY ............................................................................... 41
4.5 REAL-TIME TREND SCREEN ................................................................... 43
4.6 HISTORY TREND SCREEN ........................................................................ 45
4.7 Function screen .............................................................................................. 49
4.8 FILE LIST ...................................................................................................... 50
4.9 STORAGE OPERATIONS ............................................................................ 51
4.10 DATA BACKUP .......................................................................................... 52
4.11 DATA PRINTING ........................................................................................ 54
4.12 ALARM SUMMARY .................................................................................. 55
4.13 ACCUMULATIVE REPORT ...................................................................... 56
4.14 POWER-DOWN RECORD......................................................................... 57
4.15 SYSTEM LOG............................................................................................. 58
4.16 LOG CONFIGURATION ............................................................................ 59
CHAPTER 5 SYSTEM CONFIGURATIONS AND ITS FUNCTION ...................... 60
5.1 SETTING THE DATE AND TIME ............................................................... 60

VII IM01C82R01-01E
5.2 SETTING THE DEVICE NAME .................................................................. 61
5.3 SETTING THE SYSTEM PASSWORD ....................................................... 62
5.4 SET SYSTEM LANGUAGE......................................................................... 63
5.5 SET REFERENCE JUNCTION COMPENSATION .................................... 64
5.6 IMPLEMENTATION OF THE FACTORY SETTINGS ............................... 65
5.7 IMPLEMENTION OF PURGING DATA ..................................................... 68
CHAPTER 6 SET RECORDS OF THE CONFIGURATION ..................................... 69
6.1 SET THE RECORD INTERVAL .................................................................. 69
6.2 SET UP RECORD TYPE .............................................................................. 70
6.3 SET UP AUTOMATIC BACKUP ................................................................. 71
6.4 SET RECORD MODE................................................................................... 72
6.5 SET TRIGGER CONDITIONS ..................................................................... 73
6.6 PURGE HISTORY DATA ............................................................................. 74
CHAPTER 7 SET THE INPUT CONFIGURATION ................................................. 75
7.1 SET INPUT TYPE AND SCALING ............................................................. 80
7.2. SET CHANNEL TAG ................................................................................... 81
7.3 SET THE CHANNEL UNIT ......................................................................... 82
7.4 SET THE CHANNEL ADJUSTMENT ......................................................... 83
7.5 SET BURNOUT TREATMENT.................................................................... 84
7.6 SET SQUARE ROOT COMPUTATIONS .................................................... 85
7.7 SET THE PARAMETERS OF CHANNEL FILTER .................................... 86
7.8 SET THE REMOVAL OF SMALL SIGNAL................................................ 87
7.9 COPY CHANNEL CONFIGURATION ........................................................ 88
CHAPTER 8 SET OUTPUT CONFIGURATIONS .................................................... 89
8.1SET THE OUTPUT CHANNEL .................................................................... 90
8.2 SET THE OUTPUT ADJUSTMENT ............................................................ 91
CHAPTER 9 SET ALARM CONFIGURATION ........................................................ 92
9.1 SET THE ALARM AND RELAY CONTRACT ........................................... 94
9.2 SET THE DELAY PERIOD OF RELAY OUTPUT ...................................... 95
9.3 SET THE HYSTERESIS OF ALARM LIMIT .............................................. 96
9.4 SET THE VARIATION OF ALARM TIME .................................................. 97
CHAPTER 10 SET THE DISPLAY CONFIGURATION ........................................... 98
10.1 SET BACKGROUND COLOR ................................................................... 98
10.2 SET SCREEN BRIGHTNESS AND FIRST SCREEN ............................... 99
10.3 SET THE DIRECTION OF CURVES ON THE PAPER .......................... 100
10.4 SET THE COMBINATION AND COLOR OF THE CHANNELS .......... 101
CHAPTER 11 SET THE REPORT CONFIGURATION .......................................... 102
11.1 SET THE CHANNEL ACCUMULATION, THE INITIAL VALUE,
MAGNIFICATION, DECIMAL DIGITS AND SETTLING TIME ................. 102
11.2 SET THE NUMBER OF THE SHIFT REPORTS, TIME AND DISPLAY
NUMBER .......................................................................................................... 104
CHAPTER 12 SET PRINTING CONFIGURATION ............................................... 105
12.1 SET KEY TRIGGER PRINT ..................................................................... 105
12.2 SET TIMER PRINGTING AND THE TIME ............................................ 106
CHAPTER 13 SET COMMUNICATIONS CONFIGURATION ............................. 107
13.1 SET ADDRESS.......................................................................................... 107
13.2 SET COMMUNICATION BAUD RATE .................................................. 108
13.3 SET COMMUNICATION CHECK........................................................... 109
13.4 SET THE ORDER OF BYTE-SWAPPING .............................................. 110
CHAPTER 14 CHECK SYSTEM INFORMATION ................................................ 111

IM01C82R01-01E VIII
14.1 CHECK SYSTEM INFORMATION ......................................................... 111
14.2 CHECK AD BOARD ................................................................................ 112
CHAPTER 15 IMPORT AND EXPORT CONFIGURATION FILES ...................... 113
15.1 EXPORT CONFIGURATION FILE ......................................................... 113
15.2 IMPORT CONFIGURATION FILE .......................................................... 114
CHAPTER 16 SPECIFICATIONS ............................................................................ 115
16.1 SIGNAL INPUT AND ALARM ................................................................ 115
16.2 DISPLAY FUNCTION .............................................................................. 117
16.3 DATA STORAGE FUNCTION ................................................................. 119
16.4 OTHER STANDARD FEATURES ........................................................... 120
16.5 CHOOSE ACCESSORIES ........................................................................ 121
16.6 GENERAL SPECIFICATIONS ................................................................. 123
16.7 EXTERNAL DIMENSIONS ..................................................................... 127

IX IM01C82R01-01E
CHAPTER 1 OVERVIEW OF THE FUNCTIONS
1.1 OVERVIEW OF THE INSTRUMENT
Unlike conventional record that record data on charts, the Instrument displays the
measured data in a LCD. Moreover, the measured data can also be saved to external
storage media such as U Disk (additional specifications).
Measured/ operational data can be used as displayed data stored in the memory, or
you can also insert the external storage to save it in the external storage media through
automatic or manual backup.

Displayed data is shown as curves and other data on meter screen. It is the data filter
out from the samples by the instrument. A datum is equivalent to a point on the curve
of the screen. The time of a datum is determined by recorded meter intervals.
Historical data is equivalent to the original recording paper, which is very suitable for
long-time observation.

1 IM01C82R01-01E
1.2 FUNCTIONS OF THE INPUT SECTION
Number of Measurement Channels/Scan Interval
The number of measurement channels can be selected from 1~48.
The scan interval of measurement is 1 second.

Input Type and Computation


You can select the input type of a measurement channel as follows. You can also
perform computation for instance, the “square root,” on the measured data. For the
procedure related to setting the different modes, see section『System Configurations』
Input Mode Signal Type
DC current 0-20mA,4-20mA,0-10mA
DC voltage 0-5V,1-5V,0-10V,20mV,100mV
Resistance 400Ω,175Ω
frequency Fr
RTC PT100,PT100.,Cu50,Cu53,BA1,BA2
Thermocouple S,R,B,K,N,E,J,T,WRE5-26,WRE3-25,
F1,F2
General 20mA,5V,10V
vacuum
Section 20mA,5V,10V
vacuum

Input Range and Measurable Range


You can select the “Input range” that is appropriate for the input signal in “DC
current,” “DC voltage,” “Resistance,” “Thermocouple” “Thermal Resistance,”
“Frequency signal”.
For each “Input range,” a measurable range is defined.
Input
Resistance(℃) Thermocouple(℃) 0~20mA
type 20mV/100mV
WRE3- WRE5-
name Cu50 Cu53 Pt100 Pt100. BA1 BA2
25 26
F1 F2 S R B K N E J T
resistance:400Ω/175Ω
2500 frenquency:0 ~ 10KHz
2400 II type:0 ~ 10mA
2310 2310 0 ~ 5V
2300
III type:4 ~ 20mA
2200 1 ~ 5V
2100 2000 2000
2000
1820
1900
1768 1768
1800
1700
1600
1500
Input 1372
1400
range 1300
1300
1200
1200
1100 range:-
1000 9999~30000
1000
900
800
650 650 650
700
700 700
600
500
400
400
300
150 150 150
200
250
100
0
0 0
-100
-50 -50 -50 -50 -50
-200 -200 -200 -200 -200 -200 -200 -210 -200

IM01C82R01-01E 2
Burnout
When measuring temperature with a thermocouple / thermal resistence /4-20mA/1-5V
signal, the instrument provides check burnout function. You can specify the
measurement result to be set to maximum of measurable range, minimum of
measurable range, last value or Error Flag*1. For the setting procedure, see section
『Burnout』
*1 when the burnout setting is Error Flag, the measured value is indicated as “-.--”.

Reference Junction Compensation (RJC)


When measuring the temperature using a thermocouple, the reference junction
compensation can be used. Reference junction compensation current is installed. You
can adjust the temperature that measured by the instrument. For the setting procedure,
see section “reference junction compensation”.

Filter
The filter is used to suppress the disturbance that is curbing the signal. Filtering is
provided by the standard products. The filter can be set on each channel. For the
setting procedure, see section 『set filter parameter』.

3 IM01C82R01-01E
1.3 DISPLAY FUNCTION
Common items related to the display
TFT Color LCD and the Screen Configuration

The instrument has a 10.4 TFT color LCD (800-by-600 dot resolution).
The screen consists of the status display section and the data display section.

1. Status Display Section


Display the screen name, date and time, input acquisition board status, relay output
status, USB devices connection (additional specifications), cycle display signs,
searching status. For details, see section 『status display』.

2. Data Display Section


Display the operation screen such as the digital display, bar graph display and trend
display alarm, print, backup, etc.
When it has function configuration, display configuration mode.

Group Display
The data displayed on the curve displays are the data of measurement or computation
channels that are assigned to the group. Up to 8 channels can be assigned to a single
group. For the procedure used to assign channels to groups, see section『configuration
display 』.
Up to 8 sets of combinations can be used in the digital screen, bar graph screen,
real-time curve screen and historical curve screen effectively.

IM01C82R01-01E 4
Run Mode
This mode is entered when the power is turned on.
Display various operation screens such as signal’s digital, alarm status, trend,
historical trend, alarm summary, data print (option), and data backup display (option).

Running condition is under a number of remarkable figure images, digital display


screen, bar graph screen, real-time curve of the screen, and historical curve of the
screen and function screen. Function screen includes the alarm list screen, data print
screen (additional specifications), backup historical data (additional specifications),
the accumulative report (additional specifications), power-down recording and system
logs and other screens.

Configuration screen
This instrument is in Configuration mode to set various functions.
Under Configuration mode, this screen is used to set system configuration, recording
configuration, input configuration, output configuration, alarm configuration, display
configuration, report configuration, print the configuration and communication
configuration parameters.

5 IM01C82R01-01E
1.4 STORAGE FUNCTION.
The instrument can save measured data and computed data, which are first acquired to
the instrument’s internal memory as display data. Then, the data are saved to the
external storage medium on paper. For detail, see sections『data backup 』 and
『recording configuration』 .

External Storage Media


U disk, 1GB - 4GB (optional);
Normal SD card, 1GB - 4GB( optional)

Measurement period and recording intervals


AD sampling period is 120 milliseconds, which is an integer multiple of 50Hz power
period. This can effectively eliminate the interference of power frequency. The
measurement interval of signal is fixed at 1s. The measuring and computing are
completed in every scan interval. Display data is generated by these data.
Recording Interval is used by the instrument to record the time interval of data
which is saved to the internal memory.

Saving Data to Internal Memory


The display data are saved to internal memory continuously according to the
recording interval while the instrument is on.
If the instrument is unplugged, it will automatically generate a new history data file
when the power is turned ON.

Saving Data to External Storage Media


Automatic Backup: after inserting The SD card into the drive, the instrument
automatically back up the historical data to SD card every day from 0:00 ,even when
the instrument stops recording or it reaches the trigger condition that the users set. For
trigger condition setting and operation, please refer to 『Recording configuration
seeting』
Manual Backup: enter the screen “data backup”, and then you can back up internal
data to USB or SD cards.

IM01C82R01-01E 6
1.5ALARM FUNCTION
Real-time data alarm can be set or alarm relay output (additional specifications). For
details, see section 『Setting Alarms』.

Number of Alarms
You can set up to four alarms for each channel.
 High alarm (H)
An alarm occurs when the measured value is greater than or equal to the alarm value.
 Low alarm (L)
An alarm occurs when the Measured value is less than or equal to the alarm value.
 Incremental alarm (I)
An alarm occurs when increased measured value of the time point is greater than or
equal to the value of alarm value.
 Reduction alarm (D)
An alarm occurs when reduced measured value of the time point is greater than or
equal to alarm value.

Alarm Accessibility
Use the following auxiliary functions

Functional Content
Transfinite alarm Hysteresis Set any hysteresis within the range
Setting the length of incremental alarm
Incremental Alarm Time
time
Setting the length of reduced alarm
Reduced alarm time
time
Alarm relay output When alarm occurs, output contacts
(additional specifications) signal

Alarm display
Show alarm state in the digital display screen, bar screen and real-time curves of
screen, and display the icon of output relay in the System Column.

7 IM01C82R01-01E
1.6 REPORT FUNCTIONS (ADDITONAL
SPECIFICATIONS)
The Accumulation computation and reports can be performed, but the monthly
cumulative maximum value is no more than 999,999,999, otherwise the report is
invalid.

Accumulation Computation
The accumulation computation is performed every second.
The accumulation computation can be performed on each measurement channel.
For detail, see section『Report Setting』

Report function
Based on total accumulated data to show generated shift reports, hourly reports, daily
and monthly reports and to keep the data up to 410 days.
 Class report: According to the number of class and time settings to show the
cumulative amount of each shift.
 Horly report: Show the cumulative amount generated per hour.
 Daily report: Shows the cumulative amount generated every day.
 Monthly report: Shows the cumulative amount generated per month.

IM01C82R01-01E 8
1.7 OUTPUT FUNCTIONS (ADDITIONAL
SPECIFICATIONS)
Measured and Computed data can be transmitted and output as an analog signal.

Transmitter Output
The signal type of transmitter output is 4-20mA.
The instrument provides 8 channels transmitter output.
The maximum load of each channel is 750Ω.
For detail, see section『 Output Configuration Setting』

9 IM01C82R01-01E
1.8 COMMUNICATION FUNCTIONS (ADDITIONAL
FUNCTIONS)
The instrument can communicated with PC via serial cable(RS232C/RS485) . And
also can print data by micro printer.

RS232C/RS485 Communications
RS232C or RS485 can be select when the instrument communicates with PC.
By using the wired network (485 networks), radio, GPRS, Modem and other methods,
connect to the computer.
The Instrument adopts the international common standards: MODBUS-RTU
communication protocol.
The OPC software also provides for users, you can use it to connect the he
configuration software.

Connect to the panel printer


The instrument can be connected to the panel printer.
You can print history data and history curve manually.
Print real-time data can be automatic or manual.

IM01C82R01-01E 10
1.9 OTHER FUNCTIONS
24 VDC Power Supply for Transmitter (Option)
The instrument provides 6 channels power supply of 24 VDC. The maximum current
of each channel can be provided is 60mA.
This instrument supports 2-wire 4-20mA current input. For the wiring method, please
refer to 『24VDC Power Supply for Transmitter』

11 IM01C82R01-01E
CHAPTER 2 INSTALLATION AND WIRING
2.1 INSTRUMENT SITE
Installation location
Install the instrument in a location that meets the following conditions.
 Instrument panel
The instrument is designed for panel mounting.
 Location of installation
To prevent rain and direct heat, install the instrument indoors.
 Well-ventilated location
To prevent overheating, install the instrument in a well-ventilated location.
 Minimum mechanical vibrations
Choose an installation location with the minimum mechanical vibration.
 Horizontal
Install the instrument horizontally (However, the instrument can be inclined up to 30
degrees backwards for panel mounting).

Note
 Condensation may occur if the instrument is moved to another
place where both the ambient temperature and humidity are higher,
or if the temperature changes rapidly. In addition, measurement
errors will result when using thermocouples. In this case, let the
instrument adjust to the new environment for at least one hour
before using the instrument.
 The lifetime of the LCD may be shortened (lower picture quality,
etc.) if the instrument is used in a high-temperature environment
over a long period of time. When installing the instrument in a
high-temperature environment (greater than 40°C), please don not
do it.

Do not install the instrument in the following places:


 In direct sunlight or near heat sources
Install the instrument in a place with small temperature fluctuations near room
temperature (25.C). Placing the instrument in direct sunlight or near heat sources can
cause adverse effects on the internal circuitry.
 Where an excessive amount of soot, steam, moisture, dust, or corrosive gases are
present
Soot, steam, moisture, dust, and corrosive gases will adversely affect the instrument.
Avoid such locations.
 Near strong magnetic field sources
Do not bring magnets or instruments that produce electromagnetic fields close to the
instrument. Operating the instrument in strong magnetic fields can cause errors in the
measurements.
 Bad angle for viewing the screen
Because the instrument uses a TFT color LCD, it is difficult to view the display from
an extreme angle. Please install the instrument so that the monitor can be viewed from
the front.

IM01C82R01-01E 12
2.2 INSTALLING THE INSTRUMENT
Warning
 In order to prevent electric shock, be sure to turn off the power
before the installation.

Installation Procedure
The instrument should be mounted on a steel panel of thickness 1 mm to 10 mm.
1. Insert the instrument from the front side of the panel.
2. As shown in the figure below, mount the instrument to the panel using the mounting
brackets that came with the package.
●Use two brackets under the cover of the instrument
●The screws of instrument panel mounting bracket are the standard M4 screws.

Installation diagram
Positioning

13 IM01C82R01-01E
IM01C82R01-01E 14
2.3 THE ARRANGEMENT OF VARIOUS TERMINALS

Due to different configurations, the arrangement of every instrument for the wiring
terminal in various channels is different. Please refer to the terminal graph on the
terminal cover.
When relay loop number is more than 24, signal terminal 4 will function as the alarm
terminal.

15 IM01C82R01-01E
2.4 INPUT AND OUTPUT SIGNAL WIRING
Precautions to be taken while wiring
Take the following precautions when wring the input signal cables.
It is recommended that crimp-on lugs (designed for 4 mm screws) with insulation
sleeves be used on the lead wire ends.
crimp-on lugs (designed for 4 mm screws) with
insulation sleeves be used on the lead wire ends.

Take measures to prevent noise from entering the measurement circuit.


 Move the measurement circuit away from the power cable (power circuit) and
ground cable.
 It is desirable that the item being measured does not generate noise. However, if
this is unavoidable, isolate the measurement circuit from the item. Also, ground
the item being measured.
 Shielded wires should be used to minimize noise caused by electrostatic induction.
Connect the shield to the ground terminal of the instrument according to the need
(make sure you are not grounding at two points).
 For the interference generated by electromagnetic induction, if splicing the
equidistant dense circuit wiring, it will be more effective.
 The resistance of ground terminal is lower.

When using the thermocouple input, take measures to stabilize the temperature
at the input terminal.
 Always use the input terminal cover.
 Do not use thick wires which may cause large heat dissipation (cross sectional
area 0.5mm2 or less recommended).
 Make sure that the ambient temperature remains reasonably stable. Large
temperature fluctuations can occur if a nearby fan turns ON or OFF.

Connecting the input wires in parallel with other devices can cause signal
degradation, affecting all connected devices.
If you need to make a parallel connection, then
 Ground the instruments to the same point.
 Do not turn ON or OFF another instrument during operation. This can have
adverse effects on the other instruments.
 RTD cannot be wired in parallel.
 Current signal cannot be wired in parallel.

WARNING
To prevent electric shock, ensure the main power supply is turned OFF

IM01C82R01-01E 16
Wiring Method
1. Turn OFF the instrument and remove the input terminal cover.
2. Connect the input/output signal wires to the input/output terminals.
3. Fit with the rear terminal cover and supply electricity.

The arrangement of signal terminal for various channels


Take the 24-way universal input +4 way analog ouput as an example:

17 IM01C82R01-01E
Wiring diagram

IM01C82R01-01E 18
2.5 ALARM OUTPUT WIRING (OPTION)
WARNING
 To prevent electric shock, ensure the main power supply is turned
OFF.
 If a voltage of more than 30 VAC or 60 VDC is to be applied to the
alarm output terminal, use ring-tongue crimp-on lugs with
insulation sleeves on all terminals to prevent the wires from
slipping out when the screws become loose. Furthermore, use
double-insulated wires (dielectric strength of 2300 VAC or more)
for the signal wires on which a voltage of more than 30 VAC or 60
VDC is to be applied. For all other wires, use basic insulated wires
(dielectric strength of 1350 VAC). To prevent electric shock, attach
the terminal cover after wiring and make sure not to touch the
terminals.

Wiring Procedure
1. Turn OFF the instrument and remove the cover for the rear terminal.
2. Connect the alarm output cables to the terminal.
3. Replace the terminal cover and fasten it with screws.

CAUTION
Remove the rear terminal cover to wire easily. In order to prevent poor
contact, tighten the screws after wiring
.
Relay Specifications

Item Content
Output Relay transfer contact ( switchable)
Output capacity 250VAC/3A
30VDC/3A
Dielectric 500VAC/one minute

19 IM01C82R01-01E
The contact arrangement of alarm relay ( e.g. the maximum channel number
36-way)

* Note
When the relay way is more than 24, signal terminal 4 will function as alarm terminal.

Wiring diagram

IM01C82R01-01E 20
2.6 CONNECT POWER SUPPLY
Here explains power wiring method. Be sure to read this part before connecting
power.

Precautions to be taken when wiring the power supply


To prevent electric shock and damage to the instrument, observe the following
warnings.

WARNING
 To prevent electric shock, ensure the main power supply is turned
OFF.
 To prevent fire, use doubled-insulated wire
 Make sure to earth ground the protective earth terminal through a
grounding resistance less than 10Ω before turning ON the power.
 Use crimp-on lugs (designed for 4mm screws) for power and
ground wiring termination.
 Make sure to provide a power switch (double-pole type) on the
power supply line in order to separate the instrument from the main
power supply. Put an indication on this switch as the breaker on the
power supply line for the instrument.
Switch Specification Rated current: 3A or more
 Connect a fuse (between 2A and 15A) to the power line.

Wiring Procedure
1. Connect the power supply wires and the protective ground wire to the power
terminals.
2. Turn on the power.

Relay Specifications
Item AC supply
Input voltage 85VAC ~ 265VAC
Input frequency 50Hz

21 IM01C82R01-01E
2.7 CONNECT SWITCHING VALUE
Switching value (digital) input can be used as the trigger conditions of the batch
records. As for the related settings, please see "record configuration" section. Be sure
to read this section before connecting input signal line of switching value.
Wiring
1. Turn off the power and remove the rear terminal cover.
2. Connect digital input signal line and the switching value input terminal line.
3. Install the rear terminal cover, and turn on the power

The arrangement of terminals

Contact specification
Item Content
external Connect electrical level input terminal and
signal line termianl2 D+ terminal, and connect ground wire
and 2D- terminal。
Set the trigger of instrument record as “ on-off”.
When the electrical level input terminal is high
level ( more than 4 V and less than 24V), the
instrument stops recording; if it is low level
( less than 2 V), the instrument starts recording.

IM01C82R01-01E 22
2.8 24VDC TRANSMITTER POWER WIRING(OPTION)
Transmitter power output

CAUTION
●Never short-circuit the power supply terminals or apply an external
voltage, otherwise damage to the instrument may result.

Wiring Procedure
1. Turn OFF the instrument and remove the cover of the rear terminal.
2. Connect the power line of transmitter to the output terminal.
3. Fit with the terminal cover and turn on the power.

The arrangement of transmitter power terminal

Wiring diagram

2-wire transmitter wiring diagram

24VDC Transmitter Power Supply


Loop: 8
Output voltage: 24VDC±10%
Maximum output current : 65mADC(over-loading protect current: about
90mADC)

23 IM01C82R01-01E
2.9 RS232C/RS485 COMMUNICATION WIRING
This instrument provides a serial interface, and type ( RS 232C/ RS 485 ) can be
ordered according to the specification

RS232C interface

Pin
Signal Name Description
number
Receive data from the RS232 interface,
2 RXD
and the instrument input signal
Send data to the RS232 interface, and the
3 TXD
instrument output signal

5 GND Signal Ground

RS485 interface

Terminal name Description


A Differential signalA(+)
B Differential signalB(-)
SG Signal protective ground

Communication Specifications
Item Content
Speed(baud 1200/2400/4800/9600/19200/38400/57600/115200
rate)
Data format Eight data bits,One stop bit
Optional checkout code(None/odd/even)

IM01C82R01-01E 24
2.10 CONNECT SPECIAL SERIAL PRINTER
This section will explain how to connect serial printer. Be sure to read it before
connecting.

Wiring method
The terminal arrangement of serial printer interface

Explanation
The instrument and printer is connected through the printer special interface.

Instrument interface Micro-printer interface


RX RTS
TX RXD
GD GND

The instrument doesn’t need to set any parameter, so it can print when connected
to the printer.

25 IM01C82R01-01E
CHAPTER 3 NAMES OF PARTS/RUN
MODE/COMMON KEY OPERATIONS
3.1 NAMES OF PARTS AND FUNCTIONS
Front Panel

IM01C82R01-01E 26
3.2 INSERTING/REMOVING THE EXTERNAL
STORAGE MEDIUM
This section describes the procedures to insert and remove the external storage
medium.
USB disk and SD card can be used as external storage medium.
For the handling procedure of the storage medium, see section 『back up history
data』.

Procedure
Inserting the storage medium
1. Open the front cover. Insert the U disk and SD card into the USB port.
2. The instrument detects the U disk and SD card in the drive automatically.
3. Enter Backup screen, and prepare to back up data.

CAUTION
 Keep the front cover closed during operation except when
accessing the external storage medium. This will protect the storage
medium and the drive from particles such as dust.
 Keep the metal face upper when inserting the U disk, otherwise the
U disk can’t be inserted.
 If you are using a commercially available U disk or SD card, be
careful of static electricity.
 It is recommended using our company's products.

Remove the storage media


1. Make sure there is no data saved in storage media

CAUTION
 The access lamp is lit while the storage medium is being accessed.
 A message “Data are being saved to the medium.” is displayed
while data are being saved to the medium.

Description
Formatting external storage media
Use the external storage medium which has been formatted.
Please use the PC to carry on the formatting for the USB flash drives and choose
FAT 32 system, because this product doesn’t provide the formatted function.

Specification
Item Content
USB port Support the USB 2.0 protocol. The maximum
capacity is 4GB
SD card port Standard SD card, support SDHC [Link]
maximum capacity is 4GB

27 IM01C82R01-01E
3.3 RUN MODE
There are two run modes: operating and setting. This section describes the functions
and relationships of the two modes.

Mode Types

Mode Type Content Possible operations


Operation Data display and query mode of  Display measured
Mode operation. /operated data
This mode is entered when the  Data acquisition
power is turned ON.  Data printing
Use [Page Down] to switch the  Operation related to
screen, while pressing the [next the file on the
page] key and [confirmed] key external storage
to enter the configuration mode medium.
for more than 1 second.
Setting Operations such as  Set a variety of
Mode measurements, alarm detection, configuration
and data acquisition are parameters
continued.
This mode is entered by pressing
the [Page] and [Enter] key for
one second at the setting mode
menu.
Measured data can’t be
displayed in this mode.
Continue to do measurement,
alarm detection, data storage and
other operation.

IM01C82R01-01E 28
3.4 CONFIGURING THE FUNCTIONS
Setting Mode
The configuration mode is used to set the various parameters of the running
system.

Note
Change some system parameters will affect the data acquisition. We
Suggest backing up data before the change.

Setting Items
User can carry out the following content in configuration mode:
 Setting system parameters
 Setting display parameters
 Setting signal input parameters
 Setting analog output parameters
 Setting alarm parameters
 Setting accumulation and reports parameters
 Setting print parameters
 Setting communication parameters
 Clearing data in the instrument
 Checking instrument hardware information and firmware

29 IM01C82R01-01E
3.5 COMMON KEY OPERATIONS
This section describes common key operations which are often used.
[Left and right key] [Left key] [Right key]
[Up and down key] [Up key] [Down key]

Key operations in the running mode


Entering the running mode
1. Enter the running mode directly after power on
2. Select [Exit] to enter the running mode in the configuration mode.

Switching the screens under the running mode


Press [Page] to display each running screen in turn.

Key operations under the function screen


Enter the function screen

In the run mode, press the [PAGE] to switch to the function screen
Select Function screen
1. Use the[up and down ] button and [left and right] button to move the cursor
2. Use the [Enter] to enter the corresponding screen

IM01C82R01-01E 30
Function key operation
Click the corresponding function button below the screen to perform the
operation.

31 IM01C82R01-01E
Key operations in the configuration mode
Entering the configuration mode
1. Under the running mode, Press [Page] and [Enter] to enter the configuration
screen.
2. Move the cursor and press the [Enter] to enter in the function screen
3. Move the cursor and press the [Enter] to enter different configuration screens.

Selecting the parameter


Used [Left and right] button and [Up and down] button to move the cursor to the
target key

Confirming operation
Press [Enter] to enter the target operation screen

IM01C82R01-01E 32
Modifying the parameter
There are three types of the revisable parameters, namely, to adjust the input
parameters, edit the input parameters and select input parameters.

 Adjust the input parameters


Use [Up] and [down] to set the cursor to adjust the content. [Enter] has the same
function with [the up key] , except the function for pressing long.

 Edit the input parameters


If the editorial content is a value, use [up] and [down ] key to adjust subtlety the
parameters, and press [enter ] to modify in the pop-up input panel .When the
parameter can be edited, user can input Number, English (Caps), English, Symbol
and Chinese.

 Selecting input parameter


Press key [Up], [Down] [En] to pop up selection list, and use key [Up], [Down] to
move cursor. [En] is used to confirm selections; [Page ] means cancel.

Move the cursor to the parameters which need editing, then an input panel will
pop up after pressing [Enter], and the user can input.

Input Number/English (Caps)/English/Symbol


Used in inputting range, unit, tag etc.

33 IM01C82R01-01E
Procedure
[Left][Right][Up][Down] :Move the cursor to the soft key area (including
character soft key input and Pinyin Chinese character selection).

[Page]: Switch the cursor’s to【En】key if there is no [Link], switch the


cursor to the Chinese character, and switch between the Chinese character and
[En key]

[En]: Input the character to the input field, which under the place of cursor.

[Function Key]: the related operation in the corresponding function soft key:
[F1] : switch the input method( switching input method)
[F2]: 【Delete】(delete the last character in the input window)
[F3]和[F4]: no operation;
[F5]: 【Cancel】( canel all the editing operation and exit the inout panel);
[F6] : 【Finish】( finish the editing content and exit the input panel)

IM01C82R01-01E 34
Input Chinese characters
Used in inputting unit, tag, etc.

Operation
1. Use the [left] [right] to move the cursor to the alphabet ‘z’, use the [En],the
phonetic show bar display” z”, It is the same case with the operation of
alphabet 'i'.
2. Use the [Page Down] to jump the cursor to columns “selecting Chinese
characters” (also can use the [left and right key] and [up and down key])
3. Use the [up and down keys] to accented characters, which show icon
characters.
4. Use the [left and right button] to move the cursor to the character "Zi", use the
[En] to select.
5. Use the [Up] and [Down] to jump the cursor to the [En] (can also use the [left
and right button]), and use [En] to confirm exit.

35 IM01C82R01-01E
Examples
Revise unit to "wan m3 / h"

Remove the original information


[Enter]: Entering Input Panel.
【Delete】 soft key: Delete the original information in the input display
column for many times until the input display column is blank.

Input "Wan" character


Choose 【Pinyin input method】: [left ] to move the cursor, [enter ] to confirm
Input alphabet 'wan': [left ] to move the cursor, [enter ] to confirm
The cursor to select 【Chinese】area: [Page] to switch or [left and right
buttons] to move the cursor
Select "Wan" character: [up and down keys] to page down to accented
characters, [Enter] to select.

Input "m" word


Choose【lowercase input method】: Move the cursor with [Left and right]
button. [Enter] to select.
Choose the letter 'm': [left and right button] to move the cursor, [Enter] to
confirm.

Enter "3"
Select a “special Input Method”: Move the cursor with [Left and right] button,
[Enter] to select.
Select special character "3": [left and right button] to move the cursor, [Enter]
to select.

Enter "/" character


Choose【Symbol input method 】: Move the cursor with [Left and right]
button, [Enter] to select.
Select symbol '/': [left and right button] to move the cursor, [Enter] to select.

Enter "h" character


Choose 【lowercase input method】: Move the cursor with [Left and right]
button, [Enter] to select.
Choose the letter 'h': [left and right button] to move the cursor, [Enter] to
select.

Verify the editing


move the cursor to the 【Enter】 button: Move the cursor with [Left and right]
button, confirm editing, exit Input Panel: use [Enter] key。

IM01C82R01-01E 36
CHAPTER 4 DISPLAY SCREENS AND ITS
FUNCTION
4.1 SWITCHING OPERATION SCREENS.
This chapter describes the screen (operation screen) used to display the measured /
computed data.
The operation screens are composed of digital display, bar, real-time trend, history,
function screen.
Alarm summary, data print (additional function) and backup (additional function), the
cumulative report (additional specification) , power-down recording and system logs
and other screens enter through the data query screen.

Use the [Page] key to switch each screen.

digital display real-time curve


power ON bar screen
screen screen

data query history trend


screen screen

37 IM01C82R01-01E
4.2 EXPLANATION OF THE STATUS DISPLAY
SECTION
The following information is displayed in the status display section during the
operation mode and the configuration mode.
1 2 3 4 5 6 7 8

Alias 2010-09-20 [Link]

A C

B D

1. Screen name column


Data show the combination number of screen display, if not use the
combination number, then show 【device name】. Data query screen and
the configuration screen show the display name.
2. SD Card Connection logo
A: SD card connection is working.
B: Automatic backup of data is being written (A/B displays in turns.).
No show: No SD card connector or connection failure.
3. USB Device Identification
Display: There are USB devices and instrumentation connections.
No Show: No USB devices and instrumentation connections or
connection failure.
4. Circulation display flag
Circulation icon: Circulation Display each group
None: Fixed screen, there is no circulation.
5. Mode symbol of storage recording
C icon: the recording mode is circular, and the file is recording
D icon: the recording mode is circular, and the file stops recording
E icon: the record mode is not circular, and the file is recording
F icon: the record mode is not circular, and the file stops recording
6. Relay output flag
Display icon: There are alarms generated.
No Show: No alarm generated.
7. AD board working state flag
No Show: the instrument is working well.
Display icon: abnormal working state of AD board.
Note: if the instrument displays his flag, this means it doesn’t function
well, please contact the supplier.
8. Current data and time

IM01C82R01-01E 38
4.3 DIGTAL DISPLAY SCREEN
The digital displays measured / computed data. On the way of display, please refer
to the section 『switch to run the screen』.
Use [up and down keys] to switch displayed combinations .It can display up to 8
[Link] display of combinations don’t show configuration, skip this display
combinations.
The order to switch combination is shown in the following picture:

Use [Enter] to start or stop the automatic switching function of display combinations,
every 5 seconds automatically switch to the next combination curve, with a status bar
icons.
Screen dispays as follows:

1. Show combination numbers in the title bar, all channels show 【Device name】
2. Channel tag, can be set. With regard to setting method, See section『Channel Tag
and Unit Setting』
3. Measurement of channel /computed data. Blue is Normal, red is alarming
Note
 Measure the channel’s data display
When measuring data of measurement channel anomalies (see below), the
screen shows measurements for -.--.
 Data exception
4-20 mA : signal is bellow2 mA.
1-5V: measured signal is bellow 0.5V, and the [burnout] display an error flag
The thermocouple burnout, and the [burnout set] display an error flag.
The thermal resistence burnout, and the [burnout set] display an error flag.

39 IM01C82R01-01E
4. Channel accumulation value. It displays when the user selects [channel
accumulated] function.
5. Channel unit can be set. With regard to setting method, see section『Channel Tag
and Unit Setting』
6. Alarm symbol. 4-channel alarm can be set up, Green is normal, red is alarm.

IM01C82R01-01E 40
4.4 BAR GRAPH DISPLAY
Bar graph displays the measured / operated data. On the way to show, please refer to
『switch running screen』
Using [up and down keys] to switch combinations of display. It can display up to 8
[Link] display of combinations don’t show configuration, skip this display
combinations.
The order to switch combination is shown in the following picture:

Use [Enter] to start or stop the automatic switching function of display combinations,
every 5 seconds automatically switch to the next combination curve, with a status bar
icons.
Example of screen display is shown in the following picture:

1. Show combination numbers, all channels show 【Device name】


2. Channel tag, can be set. With regard to setting method, See section『Channel Tag
and Unit Setting』
3. Bar. Bar chart ruler has the length of 10 grids, and the length of filling color
represents proportion in percent. Green is normal, red is alarm.
4. Measurement of channel /computed data. Blue is normal, red is alarming
Note
 Measure the channel’s data display

41 IM01C82R01-01E
When measuring data of measurement channel anomalies (see below), the
screen shows measurements for -.--.
 Data exception
4-20 mA : signal is bellow2 mA.
1-5V: measured signal is bellow 0.5V, and the [burnout] display an error flag.
The thermocouple burnout, and the [burnout set] display an error flag.
The thermal resistence burnout, and the [burnout set] display an error flag.
5. Channel unit can be set. With regard to setting method, see section『Channel Tag
and Unit Setting』
6. Alarm symbol. 4-channel alarm can be set up, green is normal, red is alarm. The
place in which the warning signs are located in the bar is where the alarm sets
limits in the range of the position, which is only effective against H/L alarm.

IM01C82R01-01E 42
4.5 REAL-TIME TREND SCREEN
The waveform displays measured / operated data. On the way to show, please refer to
『switch running screen』
Using [up and down keys] to switch combinations of display. If display of
combinations don’t show configuration, skip this display combinations.. When all the
combinations show no configuration, each 8-channel has circular display according to
the order of combination of channels.
 Combined mode of curve cycle

 Uncombined model of curve cycle

Use [Enter] to start or stop the automatic switching function of display combinations,
every 5 seconds automatically switch to the next combination curve, with a status bar
icons.
An example of Real-time curve show (vertical)

43 IM01C82R01-01E
An example of Real-time curve show (horizontal)

1. Show combination numbers, all channels show 【Device name】


2. Trend cycle. Each grid means the length. The cycle is related with recording
interval, each grid time length = recording interval × 60.
3. Grid. Estimate the time and data value.
4. Data trend. Most display 8 trends at the same screen.
5. Time of the current grid.
6. Channel tag, which can be set. With regard to setting method, See section
『Channel Tag and Unit Setting』. Background color and this corresponding trend
color are identical.
7. Alarm symbol. 4-channel alarm can be set up, green is normal, red is alarm.
The place in which the warning signs are located is top-down arrangement, which
is corresponding to the configuration in the 【Alarm configuration】 screen.
8. Measurement of channel /computed data
Note
 Measure the channel’s data display
When measuring data of measurement channel anomalies (see below), the
screen shows measurements for#####。
 Data exception
4-20 mA: signal bellows 2 mA.
1-5V: measured signal is bellow 0.5V, and the [burnout] display an error
flag.
The thermocouple burnout, and the [burnout set] display an error flag.
The thermal resistence burnout, and the [burnout set] display an error flag.

9. Channel unit can be set. With regard to setting method, see section『Channel Tag
and Unit Setting』
10. Scale. Display the percentage of the trend.

Curve refresh: every pixel in the timeline of LCD screen timeline represents a record
of each interval. Each interval in the curve moves once.

IM01C82R01-01E 44
4.6 HISTORY TREND SCREEN
Display the measured / operated data stored in the memory with historical curve.
On the way to show, please refer to 『switch running screen』
There exist two methods to search historical data. One is continuous and the other is
fixed point. Historical Curves show no alarm.
Curve direction on paper is according to the configuration display, with horizontal or
vertical two options.

Horizontal historical Curve

1. Show combination numbers, all channels show 【Device name】


2. Trend cycle. Each grid means the length. The cycle is related with recording
interval, Time length = recording interval ×Zoom multiple× 60.
3. Grid. Estimate the time and data value.
4. Searching stick. Estimate the time and data value.
5. Data trend. Most display 8 trends at the same screen.
6. Searching time. Show the time where the cursor is located in the trend.
7. Channel tag can be set. With regard to setting method, See section『Channel Tag
and Unit Setting』.. Background color and this corresponding trend color are
identical.
8. Historical data of channels. Show the data of channels that the Meter recorded.
9. Channel unit can be set. With regard to setting method, see section『Channel Tag
and Unit Setting』
10. Trend display symbol. “√” display, “×” hide.
11. Scale. Display the percentage the trend.

45 IM01C82R01-01E
Procedure:
 Cursor searching state
Switching curve combination
Using [up and down keys] to switch combinations of display. If display of
combinations don’t show configuration, skip this display [Link] all
the combinations show no configuration, each 8-channel has circular display
according to the order of combination of channels.

Combined cycle mode of curve

Non-combined cycle mode of curve

Cursor searching mode operation


Use [left and right button] to move the cursor for time forward or backward
searching operation, and press [left and right button] to accelerate searching
Switch searching mode
Use [Enter key] to switch the searching mode.

 Searching mode
Searching mode operation
Use [left and left button] to move the cursor to the searching time, use the [up and
down keys] to set the date and time to the required searching time, then press
[Enter key], the trend will be automatically moved to the set point in time, and the
searching point is at the far right of the screen.

Scaling curve
Use [Left] [Right] to move the cursor to the curve of cycles, and use [Up] [Down]
to adjust zoom curve.

IM01C82R01-01E 46
Show or hide a curve
Use [Left] [Right] to move the cursor to the curve show / hide, press [Up] or
[down], you can hide or show the curve.

Switch the continuous state and the fixed point state


Press [En] to switch to the cursor state.

Longitudinal history curve

Procedure:
 Cursor state
Switching curve combination
Using [up and down keys] to switch combinations of display. If display of
combinations don’t show configuration, skip this display [Link] all
the combinations show no configuration, each 8-channel has circular display
according to the order of combination of channels.

Combined mode of curve cycle

47 IM01C82R01-01E
Non-combined cycle mode of curve

Cursor state
Uses [Up] and [Down] to move cursor so that forward or backward searching is
available. Long press [up and button] can accelerate continuous mode.

Switch the continuous state and the fixed point state


Press [En] to switch to the fixed point state.

 Searching mode state


Operation of searching mode state
Uses [Left] and [Right] to move cursor to searching time. Set the data time to a
desired point by using [Up] and [Down]. Press [En], the trend will be moved to
the fixed point automatically. The fixed point is on the right hand.

Scaling curve
Use [Left] [Right] to move the cursor to the curve of cycles, and use [Up] [Down]
to adjust zoom curve.

Show or hide a curve


Use [Left] [Right] to move the cursor to the curve show / hide, press [Up] or
[down], you can hide or show the curve.

Switch the continuous state and the fixed point state


Press [En] to switch to the fixed point state.

IM01C82R01-01E 48
4.7 FUNCTION SCREEN
This section describes the screen entrance for the instrument internal data (historical
data, alarm list, accumulation, accumulative statements, power-down recording and
system logs) to back up, print, view, etc.

Operation:
Use [left and right button] to move the cursor
Use [Enter key] to enter the corresponding screen

49 IM01C82R01-01E
4.8 FILE LIST
This section shows file storage and file storage status information.
On the display method, see the section『switch run screen』

1. Number: Records arranged by time, the nearer it occurs, the farther it is arranged.
2. Starting time: the starting time of data record in the file.
3. End time: the end time of data record in the file.
4. Status: Display the current state of file records, and the file state is as follows:
Being recorded: the file is recording data;
Manually stopped: because of changes in the configuration data or manually key
operation, file stops recording;
Power-down stop: due to power down, file stops recording;
Alarm stop: due to alarm cancellation, the file stops recording;
Switch stop: because the external switching device is disconnected , the file stops
recording;
Time stop: because the file reaches the set time for the end, it stops recording.
5. Key operation: Use the [up an down keys] to move the arrow to the file number
you want to view;
Press [F1] ( [trend] )to the historical trend screen for viewing the historical trend
data, and then use the [Page] to return to the file list screen;
press [F2 ] ([backup] )to the history data backup screen for a single file to back up
it . Then, use the [Page ] to return to the file list screen;
press [F3 ] ([print] )to data print screen . Then, use the [Page] to return to the file
list screen;
Press [Page] or [F6] ([Exit] key )to return to the current file list screen.

IM01C82R01-01E 50
4.9 STORAGE OPERATIONS
This section displays record configuration information and file storage, for the
recording configuration, please refer to the section『record configuration
setting』 .
On the display method, please refer to『switch run screen』

1. Instrument records the configuration information.


2. The starting time and end time of file storage in the instrument.
3. Menu: When the record configuration trigger is set to be manual: Press [Start key],
the instrument starts to recording; Press [Stop key] , the instrument stops
recording. Use the [Exit key] to return to the function screen.
Note: The Storage operation interface display varies according to the different
record configuration setting.

51 IM01C82R01-01E
4.10 DATA BACKUP
This section describes how to use the U disk and SD card to backup historical data.
On the way to show, please see the section and 【Switching the running screen 】

Procedures:
1. Properly connect U disk or SD card, then there is a corresponding connection
icon display in the system column.
2. Use [up and down keys] to select the backup contents, including all historical
data, historical data for single file, alarm recording, power-down record, the

IM01C82R01-01E 52
system log, accumulative shift report, the accumulative daily report,
accumulative monthly and accumulative annual report.
3. When backing up historical data for single files, use the [up and down keys]
to modify the file index, and make sure the backup start time and end time.
4. Modify the folder name or file name, use the [up and down keys] to adjust the
file number, use the [Enter key] to edit the file name.
5. Back up data, move the cursor to 【Backup (USB)】 or 【Backup (SD)】
button, press [Enter key] to back up data.
6. Back up all the historical data, the file storage path is / DATA /】 【Device
name】 / folder name; back up other data, the file storage path is / DATA /
【Device name】 / file name, for the【Device name】, Please check the system
configuration screen.
7. Use the [Page key] to return to data query screen.

53 IM01C82R01-01E
4.11 DATA PRINTING
This section describes how to use the mini-printer to print historical data.
On the way to display, please see the section 『switching the running screen』

Steps:
1. Connect mini-printer properly.
2. Set the start time and end time.
3. Set print channel (or combination) and print interval.
4. Printing curves or data.
5. Use [Page ] to return to data query screen.

Description
Connect the instrument and the printer through the special interface.

Instrument interface Micro-printer interface


RXD RTS
TXD RXD
GND GND

Instruments do not need to set the communication parameters.

IM01C82R01-01E 54
4.12 ALARM SUMMARY
Display channel alarm information and relay output state. It can save up to 256 pieces
alarm information.
On the way to display, please see the section and 『switching the running screen』

1. Channel sequence number:Records are arranged according to the stack, the closer
time of occurrence is, the more pre-arranged it is.
2. Status: Displays the current record status, alarm, or report cancel
3. Channel: Displays the channels of current record
4. Alarm type. Display the alarm type. Figures represent the 4-way alarm in the
number of (1-4), the letter on behalf of alarm type (H, L, I, D).
5. Alarm time. Displayed the alarm occur
6. Relay state. Display current relay output state.

Relay type Normal open relay Normal close relay


Green open close
Red close open

7. Menu: Use the [Enter key] to enter: Press [backup key] to the backup screen for
the alarm record backup; Press [Exit] to return to the current alarm record screen.
8. Use [left or right button] or [up and down keys] to inquire next page
9. Use [Page] to return to data query screen

55 IM01C82R01-01E
4.13 ACCUMULATIVE REPORT
According to the related setting of report configuration and current query range,
display accumulative report ( yearly report+ monthly report+ daily report and monthly
report+ shift report).
On the way to display, please see the section 『switching the running screen』

Operation instruction:
Channel: use the [Up and down keys] to modify the channel.
Type: Use the [Up and down keys] to select the display type of report
combinations, and use the [Enter] key to start the query.
Time: Use the [Up and down keys] to select the time for report query.
Function keys: Press the [last month], or [the next month] to switch the month;
press [ backup ] to skip to data backup screen for accumulative shift report
backup; press [Print], using the mini-printer to print the screen display content;
press [Exit] or [page] to return to the function screen.

IM01C82R01-01E 56
4.14 POWER-DOWN RECORD
Display time of power-down and power-on records, including the power-down and
power-on time, power-down frequency and the total time of power-down. It can save
up to 512 power-down record.
On the way to display, please see the section 『switching the running screen』

1. Channel sequence number:Records are arranged according to the stack, the closer
time of occurrence is, the more pre-arranged it is.
2. Power-down time: Power-down time of the Instrument.
3. Power-on time: power-on time of the Instrument.
4. Total power-down: the total number of power-down, accumulate each
power-down time for display.
5. Menu: Use the [Enter key] to enter: Press [backup key] to the backup screen, for
the power-down record backup; Press [Exit] to return to the current power-down
record screen.
6. Use [left or right button] or [up and down keys] to inquire on next page
7. Use [Page] to return to data query screen.

57 IM01C82R01-01E
4.15 SYSTEM LOG
Display related log of instrument operation.
On the way to display, please see the section 『switching the running screen』

Operation:
Use [left and right button] or [up and down keys] to inquire on next page
Use [Page] to return to data query screen.

1. Channel sequence number:Records are arranged according to the stack, the


closer time of occurrence is, the more pre-arranged it is.
2. Event: Operation event.
3. Time of occurrence: The incident occurred.
4. Menu: Use the [Enter key] to enter: Press [backup key] to the backup screen,
5. For the system log backup; Press [Exit] to return to the current system log
screen.

IM01C82R01-01E 58
4.16 LOG CONFIGURATION
This section describes how to log configuration.

Operation
Press the [next page] plus [En] to enter the login screen, enter the password, If the
password is correct then display configuration entry, select the appropriate
configuration entry to set the configuration; password error could not enter.

Press 【Quit】 button to exit configuration and enter the running screen.

59 IM01C82R01-01E
CHAPTER 5 SYSTEM CONFIGURATIONS AND
ITS FUNCTION
5.1 SETTING THE DATE AND TIME
Set the date and time of the instrument.

Procedure
Enter 【system configuration】screen, the method to enter please see “log
configuration”.

Date Time: This field is to set meter’s current date and time.

1. Use the [left and right button] to move the cursor to the【System time】
column.
2. Use the [up and down keys] to modify date and time.
3. Confirming operation
After the parameter setting, select 【Exit】there is a Frame.
Select 【Yes】 to save setting content, and exit [System] screen.
Select 【No】 to cancel setting content, and exit [System] screen.
Select 【Cance】 to continue setting parameters in [System] screen.

Caution
After setting system date / time, those historical data stored in
instrument is invalid.
New and effective data begins after setting system date / time.
Before setting system date / time, please back up history data.
After setting the system date / time, it is recommended that the
instrument should purge the data.

IM01C82R01-01E 60
5.2 SETTING THE DEVICE NAME
Set the device name; display it on the title bar.
Use that name to create the directory of data backup and automatic backup .

Operation
Enter 【System configuration screen】, the method to enter please see “log
configuration”.

Device Name: This field is to set the device name.


1. Use the [left and right button] to move the cursor to the【Alias】 column.
2. Use the [En] to input on the pop-up input panel, please refer to “common key
operation” about the input method.
3. Confirming operation
After the parameter setting, select 【Exit】 there is a Frame.
Select 【Yes】 to save setting content, and exit [System] screen.
Select 【No】to cancel setting content, and exit [System] screen.
Select 【Cancel】 to continue setting parameters in [System] screen.

61 IM01C82R01-01E
5.3 SETTING THE SYSTEM PASSWORD
System password is used to modify configuration parameters and prevent the system
parameters from being accidentally modified.
System password is six digits.

Operation
Enter 【System configuration screen】 , the method to enter please see “log
configuration”.

Password: This field is to set the system password.


1. Use the [left and right button] to move the cursor to the【Password】 column.
2. Use the [up and down keys] to change password.
3. Confirming operation
After the parameter setting, select 【Exit】there is a Frame.
Select 【Yes】 to save setting content, and exit [System] screen.
Select 【No】 to cancel setting content, and exit [System] screen.
Select 【Cancel】 to continue setting parameters in [System] screen.

NOTE
The system password is the only password which can enter the system
configuration to modify parameters. If it is lost, modification is
unavailable.
The initial password is 000000, after the purchase please change
password as soon as possible, and properly preserve it.

IM01C82R01-01E 62
5.4 SET SYSTEM LANGUAGE
Display language (Chinese and English).

Operation
Enter 【System configuration screen】 , the method to enter please see “log
configuration”.

System Language: The bar is to set system language.


1. Use the [left and right button] to move the cursor to the【Language】 column.
2. Use the [up and down keys] to select system language.
3. Confirming operation
After the parameter setting, select【Exit】 there is a Frame.
Select【Yes】to save setting content, and exit [System] screen.
Select【No】 to cancel setting content, and exit [System] screen.
Select【Cancel】 to continue setting parameters in [System] screen.

63 IM01C82R01-01E
5.5 SET REFERENCE JUNCTION COMPENSATION
Reference junction compensation can tune compensation temperature value of
thermocouple

Operation
Enter 【System configuration screen】 , the method to enter please see “log
configuration”.

Cold-side adjustment: the column is used to set the Cold-side adjustment.


1. Use the [left and right button] to move the cursor to the【RJC】 column.
2. Use the [up and down keys] to adjust the value of the cold side to the right
temperature.
3. Confirming operation
After the parameter setting, select【Exit】there is a Frame.
Select【Yes】to save setting content, and exit [System] screen.
Select【No】 to cancel setting content, and exit [System] screen.
Select【Cancel】 to continue setting parameters in [System] screen.

IM01C82R01-01E 64
5.6 IMPLEMENTATION OF THE FACTORY SETTINGS
This section describes the instrument functions of the factory settings and parameters
of reset value.
This will clear all stored data in memory of the instrument, including historical data,
the total cumulative amount, cumulative reports, alarm list, power-down records, but
the system log is not cleared.

Operation
Enter 【System configuration screen】 , the method to enter please see “log
configuration”.

1. Use the [ F 1] to move the cursor to the 【Factory set】 button.


2. Use the [En] to implement factory setting function.
3. Pop up a confirmation dialog box.
Select【Yes】, implement the factory settings.
Select【Cancel】or【No】, give up the factory settings.

65 IM01C82R01-01E
Reset instrument of configuration parameter values, please see the following table:

Configuration ## Parameter Reset value Note


The former in
Device name / Chinese, the
1 Device name
Alias latter in
English
2 Date and time Not affected No change
System
3 System language Not affected No change
Configuration
4 Password 000000
Displayed as
the cold-side
5 RJC 0.0
plus
adjustment
Background
6 white
color
Screen
7 10
brightness
Display
8 Trend direction Horizontal
configuration
9 First screen Overview screen
10 Channel None
According to
11 Color
trend number
The former in
Channel xx / Chinese, the
12 Display tag
Channelxx latter in
English
13 Channel unit %
14 Channel type Current
15 Channel type 4-20mA
Channel
16 2
precision
Input
17 Signal range 4.00 ~ 20.00
configuration
18 Display range 0.00 ~100.00
Adjusted value
19 1.00
K
20 Adjusted value B 0.00
Burnout
21 -.--
treatment
22 Filter 0.0s
Small signal
23 0.0%
cutting
24 Channel None
Output Adjusted value
25 1.00
configuration K
26 Adjusted value B 0.00
27 Record mode Circular
Record
28 Record type Real-time value
configuration
29 Record interval 2s

IM01C82R01-01E 66
Automatic
30 No
backup
Trigger
31 Power-down
condition
32 Relay delay 4s
33 Alarm hysteresis 0.00
34 I alarm time 1s
Alarm
35 R alarm time 1s
configuration
36 Alarm type None
37 Alarm value 0.00
38 relay None
39 Digit precision 2
40 Settle time 00:00
Number of shift
41 3
report
First shift
42 Display tag Second shifr
Third shift
Report
00:30~08:30
configuration
43 Starting time 08:30~16:30
16:30~00:30
44 Accumulation Close
Initial value of
45 0.00
accumulation
Accumulative
46 1.00
ratio
47 Key print No
Print
48 Time print No
configuration
49 Time interval 00:02(H:M)
Communication
50 001
Address
Communication
51 Bytes exchange 2-1 4-3
configuration
52 Baudrate 115200
53 Parity None

67 IM01C82R01-01E
5.7 IMPLEMENTION OF PURGING DATA
This section describes the instrument functions of purging data.
Clear all stored data in memory of the instrument, including historical data, the total
cumulative amount, cumulative reports, alarm list, power-down records, but the
system log is not cleared.

Operation
Enter 【System configuration screen】 , the method to enter please see “log
configuration”.

1. Use the [ F2] to move the cursor to the 【Purge data】 button.
2. Use the [En] to clear data function.
3. Pop up a confirmation dialog box.
Select【Yes】, implement data- clearing.
Select【Cancel】or【No】, give up clearing data.

IM01C82R01-01E 68
CHAPTER 6 SET RECORDS OF THE
CONFIGURATION
6.1 SET THE RECORD INTERVAL
This section describes the settings of historical data records interval. Optional
recording interval: 1 second, 2 seconds, 5 seconds, 10 seconds, 30 seconds, 1 minute,
2 minutes, 5 minutes, 10 minutes, 30 minutes.

Operation
Enter 【record configuration screen】 ,the method to enter please see “log
configuration”.

Record Interval: This field is used to set recorded interval


1. Use the [left and right button] to move the cursor to the【Interval】 column.
2. Use the [up and down keys] to select record interval.
3. Confirming operation
After the parameter setting, select【Exit】there is a Frame.
Select【Yes】to save setting content, and exit [System] screen.
Select【No】to cancel setting content, and exit [System] screen.
Select【Cancel】to continue setting parameters in [System] screen.
NOTE
After modifying the record interval, the record which is being stored in
the instrument is invalid.

69 IM01C82R01-01E
6.2 SET UP RECORD TYPE
This section describes setting and recording methods of historical data in the
instrument.
When the recording interval is more than 1 second, depending on the recording
method to select and calculate real-time value and record the calculated values.
Optional Recordings are: real-time values, average, maximum and minimum values.
Real-time value: Direct take the records of time corresponding to real-time value.
Average: take the average of real-time data before and after the recording points.
Maximum: Take the maximum real-time value before and after the recording data
points
Minimum: Take the minimum real-time value before and after the recording data
points

Operation:
Enter 【Record configuration screen】 , the method to enter please see “log
configuration”.

Recording type: This column is used to set recording method. ( It can choose
real-time value, maximum value, minimum value, and average value)

1. Use the [left and right button] to move the cursor to the 【Record type】
column.
2. Use the [up and down keys] to select recording method.
3. Confirming operation
After the parameter setting, select【Exit】there is a Frame.
Select【Yes】to save setting content, and exit [System] screen.
Select【No】to cancel setting content, and exit [System] screen.
Select【Cancel】to continue setting parameters in [System] screen.

IM01C82R01-01E 70
6.3 SET UP AUTOMATIC BACKUP
This section describes the principle and setting methods of historical data automatic
backup.

Automatic backup: every day at 0:00 the instrument will automatically back up the
history data to SD card; when stop recording or reach the trigger conditions set by the
user, the instrument will automatically back up historical data to the SD card.
Automatic backup file storage directory:
1. Every day at 0:00 backup file storage directory: DATA / 【Device Name】/
instrument date (year _month) / AUTO_ date (day). RDZ
2. Other automated file storage directory: DATA /【Device Name】/ instrument date
(month _year) / day / instrument time (hours _ minutes_ second). RDZ

Operation
Enter 【record configuration screen】 ,the method to enter please see “log
configuration”.

Automatic backup: The bar sets up automatic backups.

1. Use the [left and right button] to move the cursor to 【Auto backup】 column.
2. Use the [left and right keys] to open or turn off the automatic backup
function.
3. Confirming operation
After the parameter setting, select【Exit】there is a Frame.
Select【Yes】to save setting content, and exit [System] screen.
Select【No】to cancel setting content, and exit [System] screen.
Select【Cancel】to continue setting parameters in [System] screen.

71 IM01C82R01-01E
6.4 SET RECORD MODE
This section describes the method of instrument record mode setting.
There are 2 types of recording modes: circular and non-circular.
Circular: When the record in the last file of the instrument is full, it will continue to
record from the first file, the file will record historical data circularly according to the
way in which the new file replaces the old.
Non-circular: When the instrument records at the last file is full, and then it will stop
recording automatically

Operation
Enter 【Record configuration screen】 ,the method to enter please see “log
configuration”.

1. Use the [left and right button] to move the cursor to 【Record type】 column.
2. Use the [left and right keys] to select record type: circular or non-circular.
3. Confirming operation
After the parameter setting, select【Exit】there is a Frame.
Select【Yes】to save setting content, and exit [Configuration] screen.
Select【No】to cancel setting content, and exit [Configuration] screen.
Select【Cancel】to continue setting parameters in [Configuration] screen.

IM01C82R01-01E 72
6.5 SET TRIGGER CONDITIONS
This section describes the method of trigger conditions setting.
Trigger conditions include: manual trigger, alarm, relay, switch, timing, power.
Manual: enter the instrument storage interface, you can manually start or stop data
storage recording.
Alarm: can be set only when one channel occurs alarm, then start recording data
storage; when the alarm stops, then stop recording.
Relay: can be set only when one channel’s alarm occurs, then start recording data
storage; when the relay stops alarm, then stop recording.
Switch: Meter is connected to an external switch equipment, when the switch is on,
the began recording data storage, and when switch is off , then stop recording data
storage.
Timing: Timing cycles is fixed at 24 hours. Set start time and end time so that the
instrument only records the data storage in the set time every day.
Power: when the instrument is powered on, it will automatically create a new file and
start recording data.

Operation
Enter 【Record configuration screen】 ,the method to enter please see “log
configuration”.

1. Use the [left and right button] to move the cursor to 【Trigger condition】 column.
2. Use the [up and down keys] to select trigger condition.
3. Confirming operation
After the parameter setting, select [Exit] there is a Frame.
Select【Yes】to save setting content, and exit [Configuration] screen.
Select【No】to cancel setting content, and exit [Configuration] screen.
Select【Cancel】to continue setting parameters in [Configuration] screen.

73 IM01C82R01-01E
6.6 PURGE HISTORY DATA
This chapter describes the method for purging recording data.
It only purge the history data in the memory , the total accumulation, accumulative
reports, alarm list, power-down record,but the system log is not cleared.

Operation
Enter 【Record configuration screen】 ,the method to enter please see “log
configuration”.

1. Use the [ F1] to move the cursor to the【Purge History record】 button.
2. Use the [Enter key] to purge the data function
3. Confirming operation
Select【Yes】to purge data
Select【Cancel】or【No】to give up purging data

IM01C82R01-01E 74
CHAPTER 7 SET THE INPUT CONFIGURATION
This chapter describes the channel type, signal type, the relationship between the
signal range and range display, and the computing part of the channel involved

Channel types and signal types


Channel types Signal types Calculation
type
Current 0-20mA,4-20mA,0-10mA Linear
operation
Voltage 0-5V,1-5V,0-10V,20mV,100mV Linear
operation
Resistance 400Ω, 175Ω Linear
operation
Frequency Fr Linear
operation
RTD PT100,PT100.,Cu50,Cu53,BA1, No
BA2
Thermocouple S,R,B,K,N,E,J,T,WRE5-26, No
WRE3-25,F1,F2
Ordinary 4-20mA,0-5V,1-5V,0-10V Normal
vacuum vacuum
operation
Sub-vacuum 4-20mA,0-5V,1-5V,0-10V Sub-vacuum
operation

Note: The thermal resistance and thermocouple signal range can not be changed;
project quantity is the directly measured temperature.

 Linear operation
Semaphore transforms to project quantity according to the signal range and
practical range of linear

20mA 100℃ 20mA

100℃
10mA

4mA
0℃
0mA 0mA 0℃

75 IM01C82R01-01E
 Ordinary vacuum operation
Semaphores and vacuum degree of logarithm has a linear relationship, signal
range in the overall signal is free to set, the project range is expressed by 10 of the
index form, ranging from 10E-9.999 to 10E9.999.

 Sub-vacuum operation
According to the vacuum degree of order , divide it into a number of sections , the
corresponding signal of endpoint in every section and the vacuum degree has
a linear relationship, at the same time, corresponding signal in the same section
and the vacuum have a linear relationship too. Signal range in the overall signal
is free to set; the project range is expressed by 10 of the index form, ranging from
10E-9.999 to 10E9.999.

 Square root calculation


The percentage of the measured signal values need to have a square root
calculation

IM01C82R01-01E 76
20mA 100℃

50℃

8mA

4mA 0℃
(8 - 4)
50℃ = (100 - 0)* +0
(20 - 4)

 Filtering operation
Filter parameters range from 0.0 seconds to 9.9 seconds .Take an example when
the filter parameters is1.0 seconds

Previous collected data S0 = 10mA


current collected data S1 = 5mA

Filter parameter 1.0s

f * S0 + S1
Result
1+f
= 7.5mA

 Resection of the small-signal functions


less than the set value of the semaphore is zero, range: 0.0% ~ 9.9%.

 Line Regulation
Quantities according to the formula x = K * x + B linear adjustment.

 Burnout treatment,
The project quantity approach to deal with signal even break off, options are:
maximum, minimum, to maintain the data, error markers, error markers. When
choose error markers, collected data is the marked value.

 Reference Junction Compensation


reference junction compensation for thermocouple

77 IM01C82R01-01E
Channel types and the computation relationship
Reference Removal of Filter Square Specific Adjust treatment
junction small-signal root calculation linear of burnout
compensation
Current Linear
    
operation
voltage Linear
    
operation
Resistance Linear
  
operation
frequency Linear
  
operation
Thermal
  
Resistance
Thermocouple    
Ordinary General
vacuum  Vacuum

Sub vacuum  Sub-vacuum

 Tagging function is configurable function, and specific calculation can not be


changed.
For current,burnout treatment is only used for 4-20mA signal,For voltage, it is
only used for 1-5V signal.

IM01C82R01-01E 78
Signal Type Measurement Range
Type Range
0-20mA -20mA ~ 20mA
4-20mA 4mA ~ 20mA
0-10mA -10mA ~ 10mA
0-5V -5V ~ 5V
1-5V 1V ~ 5V
0-10V -10V ~ 10V
20mV -20mV ~ 20mV
100mV -100mV ~ 100mV
400Ω 0 ~ 400Ω
175Ω 0 ~ 175Ω
Fr 0 ~ 10KHz
Fr. 0.0~3000.0Hz
PT100 -200℃ ~ 650℃
PT100. --50℃ ~ 150℃
Cu50 -50℃ ~ 150℃
Cu53 -50℃ ~ 150℃
BA1 -200℃ ~ 650℃
BA2 -200℃ ~ 650℃
S -50℃ ~ 1768℃
R -50℃ ~ 1768℃
B 250℃ ~ 1817℃
K -200℃ ~ 1372℃
N -200℃ ~ 1300℃
E -200℃ ~ 1000℃
J -210℃ ~ 1200℃
T -200℃ ~ 385℃
WRE5-26 0℃ ~ 2310℃
WRE3-25 0℃ ~ 2310℃
F1 700℃ ~ 2000℃
F2 700℃ ~ 2000℃

Note:
 Channel tag: 15 characters in length mixed in English and Chinese
 Channel unit: Not involved in operations, only display, 7 characters in length,
support mixed English and Chinese symbols and special characters.
 Channel decimal point: The calculation and display precision, 0 to 3
decimals.
 Signal range: Within the entire range of signal, it is free to set signal
acquisition range.
 Project Range: According to the project need, it is to be set free, -9999 ~
30000, the decimal point 0 to 3.

79 IM01C82R01-01E
7.1 SET INPUT TYPE AND SCALING
This section describes the channel type and range setting method.

Operation:
Enter 【Input configuration screen】 ,the method to enter please see “log
configuration”.

1. Use the [left and right button] to move the cursor to the 【Channel close】, use the
[up and down keys] to open or close the channel.
2. Use the [left and right button] to move the cursor to the 【Channel type】, use the
[up and down keys] to select channel type.
3. Use the [left and right button] to move the cursor to the 【Signal type 】, use the
[up and down keys] to select signal types.
4. Use the [left and right button] to move the cursor to the 【Precision】, use the [up
and down keys] to adjust channel decimal point.
5. Use the [left button] to move the cursor to the 【Signal scale】, use the [En] to edit
data, use the [up and down keys] to tune the data.
6. Use the [left and right button] to move the cursor to the【Display scale】,
use the [En] to edit data, use the [up and down keys] to tune the data.
7. Confirming operation
After the parameter setting, select【Exit】there is a Frame.
Select【Yes】to save setting content, and exit [System] screen.
Select【No】to cancel setting content, and exit [System] screen.
Select【Cancel】to continue setting parameters in [System] screen.

Note: The thermal resistance and thermocouple signal range cannot be changed.

IM01C82R01-01E 80
7.2. SET CHANNEL TAG
This section describes how to set the channel tag.

Operation:
Enter 【Input configuration screen】 ,the method to enter please see “log
configuration”.

1. Use the [left and right button] to move the cursor to the【Tag 】
2. Use the [En] to edit bit number.
3. Confirming operation
After the parameter setting, select【Exit】there is a Frame.
Select【Yes】to save setting content, and exit [System] screen.
Select【No】to cancel setting content, and exit [System] screen.
Select【Cancel】to continue setting parameters in [System] screen.

Note: The channel tag, 15 characters in length, mixed in English and Chinese.

81 IM01C82R01-01E
7.3 SET THE CHANNEL UNIT
This section describes how to set the channel engineering units.

Operation:
Enter 【Input configuration screen】 ,the method to enter please see “log
configuration”.

1. Use the [left and right button] to move the cursor to the 【Unit】.
2. Use the [up and down keys] to select a common unit.
3. Use the [En] to edit a custom unit.
4. Confirming operation
After the parameter setting, select【Exit】there is a Frame.
Select【Yes】to save setting content, and exit [System] screen.
Select【No】to cancel setting content, and exit [System] screen.
Select【Cancel】to continue setting parameters in [System] screen.

Note: The channel unit is not involved in operations, only display, 7 characters in
length, supporting mixed English and Chinese symbols and special characters.

IM01C82R01-01E 82
7.4 SET THE CHANNEL ADJUSTMENT
This section describes the channel linear adjustment.
Formula: project quantities= k × Quantities + b.

Operation:
Enter 【Input configuration screen】 ,the method to enter please see “log
configuration”.

1. Use the [left and right button] to move the cursor to the linear【Adjust(kX+b)】.
2. Use the [up and down keys] to tune data; use the [En] to edit data.
3. Confirming operation
After the parameter setting, select【Exit】there is a Frame.
Select【Yes】to save setting content, and exit [System] screen.
Select【No】to cancel setting content, and exit [System] screen.
Select【Cancel】to continue setting parameters in [System] screen.

83 IM01C82R01-01E
7.5 SET BURNOUT TREATMENT
This section describes how to deal with burnout. Burnout treatment is effective on the
III type signal, thermal resistance and thermocouple signal.
Treatment methods: minimum, maximum, maintaining, -. -- (error flag).

Operation:
Enter 【Input configuration screen】 ,the method to enter please see “log
configuration”.

1. Use the [left and right button] to move the cursor to the 【Burnout 】
2. Use the [up and down keys] to choose approach.
3. Confirming operation
After the parameter setting, select【Exit】there is a Frame.
Select【Yes】to save setting content, and exit [System] screen.
Select【No】to cancel setting content, and exit [System] screen.
Select【Cancel】to continue setting parameters in [System] screen.

IM01C82R01-01E 84
7.6 SET SQUARE ROOT COMPUTATIONS
This section describes set the square root computations of channels. To operational
principle please see 『set the channel type and range』 .

Operation:
Enter 【Input configuration screen】 ,the method to enter please see “log
configuration”.

1. Use the [left and right button] to move the cursor to 【Channel type】
2. Use the [up and down keys] to choose the square root for the channel type.
3. Confirming operation
After the parameter setting, select【Exit】there is a Frame.
Select【Yes】to save setting content, and exit [System] screen.
Select【No】to cancel setting content, and exit [System] screen.
Select【Cancel】to continue setting parameters in [System] screen.

85 IM01C82R01-01E
7.7 SET THE PARAMETERS OF CHANNEL FILTER
This section describes the channel filtering settings. Setting theory of filtering
operation please see 『6.1』” channel types and range”.

Operation:
Enter 【Input configuration screen】 ,the method to enter please see “log
configuration”.

1. Use the [left and right button] to move the cursor to the 【Filter】.
2. Use the [up and down keys] to adjust the filter time parameter, ranging from 0.0
seconds ~ 9.9 seconds.
3. Confirming operation
After the parameter setting, select【Exit】there is a Frame.
Select【Yes】to save setting content, and exit [System] screen.
Select【No】to cancel setting content, and exit [System] screen.
Select【Cancel】to continue setting parameters in [System] screen.

IM01C82R01-01E 86
7.8 SET THE REMOVAL OF SMALL SIGNAL
This section describes the setting of channel small-signal removal function. Resection
of the small-signal operation principles please refer to 『6.1』”set the channel type
range”.

Operation:
Enter 【Input configuration screen】 ,the method to enter please see “log
configuration”.

1. Use the [left and right button] to move the cursor to the 【Cut】.
2. Use the [up and down keys] to adjust removal, general signal is 0.0% ~ 9.9%, the
frequency signal is frequency value: 0 ~ 99Hz.
3. Confirming operation
After the parameter setting, select【Exit】there is a Frame.
Select【Yes】to save setting content, and exit [System] screen.
Select【No】to cancel setting content, and exit [System] screen.
Select【Cancel】to continue setting parameters in [System] screen.

87 IM01C82R01-01E
7.9 COPY CHANNEL CONFIGURATION
This section describes the channel configuration replication.

Operation:
Enter 【Input configuration screen】 ,the method to enter please see “log
configuration”.

1. Use the [left and right button] to move the cursor to the 【Channel】.
2. Use the [up and down keys] to select the source- channel.
3. Use the [F1] to execute【Copy xx】key, and copy the configuration parameter in the
current channel.
4. Use the [up and down keys] to select the source channel.
5. Use the [F2] to execute 【Paste】key, and copy the source channel configuration to
the current channel configuration.

IM01C82R01-01E 88
CHAPTER 8 SET OUTPUT CONFIGURATIONS
This chapter describes the principle and setting method of output transmission.
According to the source-channel project quantity and project range, to output a linear
4-20mA transmission,
Support the output kX + b adjustment, the source input channel can be set freely. Up
to 8-channel can perform transmission output.

20mA 20mA
100℃

K*x+B

0℃
4mA 4mA

89 IM01C82R01-01E
8.1SET THE OUTPUT CHANNEL
This section describes the setting method of transmission output channel.

Operation:
Enter 【Input configuration screen】 ,the method to enter please see “log
configuration”.

1. Use the [left and right button] to move the cursor to the 【Channel】.
2. Use the [up and down keys] to select the output source channel or turn off the
output.
3. Confirming operation
After the parameter setting, select【Exit】there is a Frame.
Select【Yes】to save setting content, and exit [System] screen.
Select【No】to cancel setting content, and exit [System] screen.
Select【Cancel】to continue setting parameters in [System] screen.

IM01C82R01-01E 90
8.2 SET THE OUTPUT ADJUSTMENT
This section describes the setting methods of transmission output adjustment.
The formula: new output current = k × original output current+ b.
e.g
If the original channel scale is 0 ~ 100℃, and the current scale is 25℃, adjust K=2.00
and b = 3.00, then
Before adjusting, output = (25 / (100 - 0) * (20 – 4) + 4) mA = 8mA
After adjusting, output= = K × output + b = 2.00 * 8mA + 3.00 = 19mA

Operation:
Enter 【Input configuration screen】 ,the method to enter please see “log
configuration”.

1. Use the [left and right button] to move the cursor to the 【Channel】.
2. Use the [up and down keys] to tune data, and use [En] to edit data
3. Confirming operation
After the parameter setting, select【Exit】there is a Frame.
Select【Yes】to save setting content, and exit [System] screen.
Select【No】to cancel setting content, and exit [System] screen.
Select【Cancel】to continue setting parameters in [System] screen.

91 IM01C82R01-01E
CHAPTER 9 SET ALARM CONFIGURATION
This chapter describes the alarm principle, relay contacts and set methods.
The amount of channel project within a certain range of values performs alarm and
relay output. Optional alarm types are: H, L, I, D. 4-way alarm for each channel.

 H Alarm:
When the project is greater than or equal to alarm value which has been set, alarm
occurred. After alarm generated, the project is less than the amount or equal to (alarm
value - hysteresis), then eliminate alarm.
H alarm

hysteresis
alarm value

 L Alarm:
When the project is less than the amount equal to the alarm value, the alarm occurred.
After alarm generated, the project is greater than or equal to (alarm value + hysteresis),
then eliminate alarm.
L alarm

hysteresis

alarm value

 I alarm:
The project difference (V2 - V1) of a specific time interval point (t2 and t1) is greater
than or equal to alarm value which has been set, alarm occurred.
I alarm

project changed value


V2
set changed value
V1

t1 t2 t

IM01C82R01-01E 92
 D Alarm:
The project difference (V2 - V1) of a specific time interval point (t2 and t1) is greater
than or equal to alarm value which has been set, alarm occurred.
D alarm

V1
set changed value

V2 project changed value

t1 t2 t

 Relay Delay: Relay delay time after the occurrence of an alarm, 0-10 seconds
(optional). (Valid for all alarms)
 Hysteresis: hysteresis range of H/L‘s eliminating alarms (All channels is uniform)
 I Alarm Time: Alarm interval time of I. (All channels is uniform)
 D Alarm Time: Alarm interval time of D. (All channels is uniform)
 Relay Output: corresponding relay to alarm. Optional 1-12 or 1-8, when multiple
alarms use the same one relay output, Use the stack output.
 Each alarm action and eliminating actions will have reported alarm information to
form a warning list. The maximum of alarm list is 256.

93 IM01C82R01-01E
9.1 SET THE ALARM AND RELAY CONTRACT
This section describes the setting methods of alarm parameters and relay contacts.

Operation:
Enter 【Alarm configuration screen】 ,the method to enter please see “log
configuration”.

1. Use [left and right button] to move the cursor to the 【Channel】,use the [up and
down keys] to select channel configuration.
2. Use the [left and right button] to move the cursor to the 【Type】, use the [up and
down keys] to select alarm type and shut down alarm.
3. Use the [left and right button] to move the cursor to the 【Alarm value】, use the
[up and down keys] tune data; use the [En] to edit data.
4. Use the [left and right button] to move the cursor to the 【Relay】,use the [up and
down keys] to select relay contact number, OFF stands for closure; R1 ~ 12 stands
for the corresponding relay output.
5. Confirming operation
Press [F6] to execute【Exit】, and select [Exit] there is a Frame.
Select【Yes】to save setting content, and exit [System] screen.
Select【No】to cancel setting content, and exit [System] screen.
Select【Cancel】to continue setting parameters in [System] screen.

IM01C82R01-01E 94
9.2 SET THE DELAY PERIOD OF RELAY OUTPUT
This section describes the time delay settings method of relay.

Operation:
Enter 【Alarm configuration screen】 ,the method to enter please see “log
configuration”.

1. Use [left and right button] to move the cursor to the 【Relay delay】.
2. Use the [up and down keys] to adjust the output delay time, ranging from 0 to 10
seconds.
3. Confirming operation
After the parameter setting, select【Exit】there is a Frame.
Select【Yes】to save setting content, and exit [System] screen.
Select【No】to cancel setting content, and exit [System] screen.
Select【Cancel】to continue setting parameters in [System] screen.

95 IM01C82R01-01E
9.3 SET THE HYSTERESIS OF ALARM LIMIT
This section describes the setting method of hysteresis of alarm limit; hysteresis
parameters are only valid for HL alarm.

Operation:
Enter 【Alarm configuration screen】,the method to enter please see “log
configuration”.

1. Use the [left and right button] to move the cursor to the【Hysteresis】.
2. Use the [up and down keys] tune data; use the [En] to edit data.
3. Confirming operation
After the parameter setting, select【Exit】there is a Frame.
Select【Yes】to save setting content, and exit [System] screen.
Select【No】to cancel setting content, and exit [System] screen.
Select【Cancel】to continue setting parameters in [System] screen.

IM01C82R01-01E 96
9.4 SET THE VARIATION OF ALARM TIME
This section describes setting method of the changes in the amount of alarm hysteresis.
Incremental alarm periods is effective to I alarm, reduced alarm time is effective to D
alarm.

Operation:
Enter 【Alarm configuration screen】,the method to enter please see “log
configuration”.

1. Use the [left and right button] to move the cursor to the 【I Alarm Time】, use the
[up and down keys] to adjust the alarm time, the scope is 1 ~ 32 seconds.
2. Use the [left and right button] to move the cursor to the 【Reduced Alarm Time】,
use the [up and down keys] to adjust the alarm time, the scope is 1 ~ 32 seconds.
3. Confirming operation
After the parameter setting, select【Exit】there is a Frame.
Select【Yes】to save setting content, and exit [System] screen.
Select【No】to cancel setting content, and exit [System] screen.
Select【Cancel】to continue setting parameters in [System] screen.

97 IM01C82R01-01E
CHAPTER 10 SET THE DISPLAY
CONFIGURATION
This chapter describes the applications of channel combination.
According to the channel combinations, Digital display screen, Bar graph screen,
real-time curve screen and the historical curve screen have data or curves display;
default settings without channel combinations, digital screen and bar diagram show in
full-screen channel, real-time curve screen and the historical curve screen are in
accordance with 4-way channel to display as a set of cycles. Details please refer to the
screen description.

10.1 SET BACKGROUND COLOR


This section describes the background color setting method of digital display screen
and bar graph.

Operation
Enter the 【Display configuration】 screen , For the methods to enter, please refer
to the section『 log configuration』 .

1. Use the [left and right button] to move the cursor to the【Background color】.
2. Use the [up and down keys] to select direction: white or black;
3. Confirming operation
After the parameter setting, select【Exit】there is a Frame.
Select【Yes】to save setting content, and exit [Configuration] screen.
Select【No】to cancel setting content, and exit [Configuration] screen.
Select【Cancel】to continue setting parameters in [Configuration] screen.

IM01C82R01-01E 98
10.2 SET SCREEN BRIGHTNESS AND FIRST SCREEN
This section describes the setting method of the startup screen brightness and the first
screen display when the instrument starts.

Operation
Enter the 【Display configuration】 screen , For the methods to enter, please refer
to the section『 log configuration』 .

1. Use the [left and right button] to move the cursor to the【Screen brightness】, and
use the [up and down keys] to adjust the screen brightness, range from 0 to 10.
2. Use the [left and right button] to move the cursor to the 【The first screen】, and
use [the up and down keys] to select the first screen display when the meter starts.
3. Confirming operation
After the parameter setting, select【Exit】there is a Frame.
Select【Yes】to save setting content, and exit [Configuration] screen.
Select【No】to cancel setting content, and exit [Configuration] screen.
Select【Cancel】to continue setting parameters in [Configuration] screen.

99 IM01C82R01-01E
10.3 SET THE DIRECTION OF CURVES ON THE PAPER
This section describes the setting method of curve direction on the paper.

Operation
Enter the 【Display configuration】 screen , For the methods to enter, please refer
to the section『 log configuration』 .

1. Use the [left and right button] to move the cursor to the【Direction】.
2. Use the [up and down keys] to choose the direction, horizontal or vertical
(optional).
3. Confirming operation
After the parameter setting, select【Exit】there is a Frame.
Select【Yes】to save setting content, and exit [System] screen.
Select【No】to cancel setting content, and exit [System] screen.
Select【Cancel】to continue setting parameters in [System] screen.

IM01C82R01-01E 100
10.4 SET THE COMBINATION AND COLOR OF THE
CHANNELS
This section describes the setting methods of combination of channels.

Operation
Enter the 【Display configuration】 screen , For the methods to enter, please refer
to the section『 log configuration』 .

1. Use the [left and right button] to move the cursor to the 【Group】, use the [up and
down keys] to choose serial number.
2. Use the [left button] to move the cursor to the 【Channel】, use the [up and down
keys] to select open or close channel.
3. Use the [left and right button] to move the cursor to the 【Color】, optional colors:
red, green, blue, yellow, blue, [Link] the [up and down keys] to select channel
color.
4. Confirming operation
After the parameter setting, select【Exit】there is a Frame.
Select【Yes】to save setting content, and exit [System] screen.
Select【No】to cancel setting content, and exit [System] screen.
Select【Cancel】to continue setting parameters in [System] screen.

101 IM01C82R01-01E
CHAPTER 11 SET THE REPORT
CONFIGURATION
11.1 SET THE CHANNEL ACCUMULATION, THE
INITIAL VALUE, MAGNIFICATION, DECIMAL DIGITS
AND SETTLING TIME
This section describes the setting method of cumulative channel, the cumulative initial
value, cumulative magnification and settling time.
Accumulative Initial value: initial value of accumulation begins to accumulate.
Accumulative Rate: According to the normal rate of a cumulative accumulation to
multiply.
Purge accumulation: Clear the accumulated amount of the current channel.
Decimal digits: set the decimal digits for the accumulative value of all channels.
Settle time: from the setting settler time to the next one, which can be regarded as one
day accumulation when starting the accumulation function.

Operation:
Enter 【Report configuration screen】,the method to enter please see “log
configuration”.

1. Use the [left and right button] to move the cursor to the 【decimal digits】., use the
[up and down keys] to adjust, range from 0 to 2.
2. Use the [left and right button] to move the cursor to the【Settle time】 , use the [up
and down keys] to adjust settle time, range from 00 to 23.
3. Use the [left and right button] to move the cursor to the 【Channel 】. Use the [up

IM01C82R01-01E 102
and down keys] to select the channel.
4. Use the [left and left button] to move the cursor to 【Accumulation】. Use the [up
and down keys] to select to open or close the accumulative function.
5. Use the [left and right button] to move the cursor to the【Initial 】, use the [up and
down keys] to tune data, use the [En] to edit data, ranging from 0 to 999999.00.
6. Use the [left and right button] to move the cursor to the【Magnification】, use the
[up and down keys] to tune data, use the [En] to edit data, ranging from 0.00 ~
999999.00.
7. Confirming operation
After the parameter setting, select【Exit】there is a Frame.
Select【Yes】to save setting content, and exit [System] screen.
Select【No】to cancel setting content, and exit [System] screen.
Select【Cancel】to continue setting parameters in [System] screen.

103 IM01C82R01-01E
11.2 SET THE NUMBER OF THE SHIFT REPORTS,
TIME AND DISPLAY NUMBER
This section describes the number of shift reports and time and setting methods.
The number of shift report: the number of classes per day, range from 2 to 5.
Shift: start time and end time of every class, 24-hour clock.

Operation:
Enter 【Report configuration screen】 ,the method to enter please see “log
configuration”.

1. Use the [left and right button] to move the cursor to 【Shift number】, use the [up
and down keys] to choose the number of shift reports, five shifts are the largest.
2. Use the [left and right button] to move the cursor to the 【Display number】, and
use the Enter button to edit the shift report to show the display number.
3. Use the [left and right button] to move the cursor to the 【Start time】, use the [up
and down keys] to adjust the frequency of time.
4. Confirming operation
After the parameter setting, select【Exit】there is a Frame.
Select【Yes】to save setting content, and exit [System] screen.
Select【No】to cancel setting content, and exit [System] screen.
Select【Cancel】to continue setting parameters in [System] screen.

IM01C82R01-01E 104
CHAPTER 12 SET PRINTING CONFIGURATION
12.1 SET KEY TRIGGER PRINT
This section describes the key trigger printing functions and settings methods.
Key trigger print: In the digital display screen, Bar graph screen, real-time curve
screen, using the [Left] to trigger button printing, and then print the real-time data and
cumulative amount.

Operation:
Enter 【Print configuration screen】 ,the method to enter please see “log
configuration”.

1. Use the [left and right button] to move the cursor to the 【Key trigger print】, use
the [up and down keys] to open or close button printing functions.
2. Confirming operation
After the parameter setting, select【Exit】there is a Frame.
Select【Yes】to save setting content, and exit [System] screen.
Select【No】to cancel setting content, and exit [System] screen.
Select【Cancel】to continue setting parameters in [System] screen.

105 IM01C82R01-01E
12.2 SET TIMER PRINGTING AND THE TIME
This section describes the timer printing functions and settings methods.
Timer Print: each day, choose 00: 00:00 as the starting time and print real-time data
and accumulation according to time intervals.

Operation:
Enter 【Print configuration screen】 ,the method to enter please see “log
configuration”.

1. Use the [left and right button] to move the cursor to the【Timer Print】, use the [up
and down keys] to close to open timing printing functions.
2. Use the [left and right button] to move the cursor to the【Interval】, use the [up and
down keys] to adjust the print interval time in the format X/hour, X/minute.
3. Confirming operation
After the parameter setting, select【Exit】there is a Frame.
Select【Yes】to save setting content, and exit [System] screen.
Select【No】to cancel setting content, and exit [System] screen.
Select【Cancel】to continue setting parameters in [System] screen.

IM01C82R01-01E 106
CHAPTER 13 SET COMMUNICATIONS
CONFIGURATION
13.1 SET ADDRESS
This section describes the setting method of ModbusRTU mailing address.

Operation:
Enter 【Communication configuration screen】 ,the method to enter please see
“log configuration”.

1. Use the [left and right button] to move the cursor to the 【Address】.
2. Use the [up and down keys] to modify the mailing address, ranging from 0 to 247.
3. Confirming operation
After the parameter setting, select【Exit】there is a Frame.
Select【Yes】to save setting content, and exit [System] screen.
Select【No】to cancel setting content, and exit [System] screen.
Select【Cancel】to continue setting parameters in [System] screen.

107 IM01C82R01-01E
13.2 SET COMMUNICATION BAUD RATE
This section describes the setting method of serial communication baud rate.
Baud rate: 1200,2400,4800,9600,19200,38400,57600,115200. The default value is
115200.

Operation:
Enter 【Communication configuration screen】 ,the method to enter please see
“log configuration”.

1. Use the [left and right button] to move the cursor to the【Baud rate】.
2. Use the [up and down keys] to select communication baud rate.
3. Confirming operation
After the parameter setting, select【Exit】there is a Frame.
Select【Yes】to save setting content, and exit [System] screen.
Select【No】to cancel setting content, and exit [System] screen.
Select【Cancel】to continue setting parameters in [System] screen.

IM01C82R01-01E 108
13.3 SET COMMUNICATION CHECK
This section describes the setting method of serial communication check.
Parity: no parity, odd parity, even parity.
The default value is no check

Operation:
Enter 【Communication configuration screen】 ,the method to enter please see
“log configuration”.

1. Use the [left button] to move the cursor to the 【Parity】.


2. Use the [up and down keys] to Select check method.
3. Confirming operation
After the parameter setting, select【Exit】there is a Frame.
Select【Yes】to save setting content, and exit [System] screen.
Select【No】to cancel setting content, and exit [System] screen.
Select【Cancel】to continue setting parameters in [System] screen.

109 IM01C82R01-01E
13.4 SET THE ORDER OF BYTE-SWAPPING
This section describes the setting method of ModbusRTU 32-bit data byte-swapping.
In order to be suitable for ModbusRTU communication protocol, it provides
byte-swapping functions.
Byte-swapping: 1-2, 3-4, 2-1 4-3, 3-4 1-2, 4-3 2-1. The default value is 2-14-3.

Operation:
Enter 【Communication configuration screen】 ,the method to enter please see
“log configuration”.

1. Use the [left and right button] to move the cursor to the 【Byte swap】.
2. Use the [up and down keys] to choose byte-swapping method.
3. Confirming operation
After the parameter setting, select【Exit】there is a Frame.
Select【Yes】to save setting content, and exit [System] screen.
Select【No】to cancel setting content, and exit [System] screen.
Select【Cancel】to continue setting parameters in [System] screen.

IM01C82R01-01E 110
CHAPTER 14 CHECK SYSTEM INFORMATION
14.1 CHECK SYSTEM INFORMATION
This section describes how to check system information, and instrument setting.

Operation:
Enter 【System configuration screen】 ,the method to enter please see “log
configuration”.

111 IM01C82R01-01E
14.2 CHECK AD BOARD
This section describes the methods to check the status of AD board.

Operation:
Enter 【System configuration screen】 ,the method to enter please see “log
configuration”.

Blue Text (1: R7A8C: 8) is the inherent information in AD, black text (R7A8C: 8)
is the current AD board information, if there is any difference between them, this
shows that AD board failure, please contact the manufacturer for maintenance.

IM01C82R01-01E 112
CHAPTER 15 IMPORT AND EXPORT
CONFIGURATION FILES
15.1 EXPORT CONFIGURATION FILE
This section describes the instrument configuration file export functions and operation
methods.
Configuration file export, you can set up the information of the current instrument
channel, including input configuration, recording configuration, alarm configuration,
and export it to root directory of the external storage media .

Operation
Enter 【System configuration】 screen, for the method to enter, please refer to
the section『 log configuration 』.

1. Use the [left and right button] to move the cursor to 【Device 】, use the [up and
down keys] to select the connected backup storage media U disk / SD card; use
the [left and right button] to move the cursor to the 【File name】 ,and use the [up
and down keys] or [Enter button] to modify the file name; use the [left and right
button] to move the cursor to 【Export configuration】,use the [Enter key] to
export configuration.
2. Show the configuration file export situation.
3. Use the [left and right button] to move the cursor to the 【Exit】,and use [Enter key]
to exit the configuration file screen

113 IM01C82R01-01E
15.2 IMPORT CONFIGURATION FILE
This section describes the import meaning and operation method of configuration file.
Import the configuration file: To set the configuration of multiple models of the same
configuration, you can only configure one of the instrument configuration . You can
export the configuration file of this instrument, for the configuration file export ,
please refer to the section 『configuration file export 』,and for the remaining
instrument , they only need to import configuration file ,which makes multi-channel
configuration simplier.

Operation
Enter 【system configuration 】, the method to enter , please refer to the section
『 log configuration』.

1. Use the [left and right button] to move the cursor to the 【Device】, use the [up and
down keys] to select the connected backup storage media U disk / SD card; use
the [left and right button] to move the cursor to the 【File name】, and use the [up
and down keys] to confirm the imported file name; use the [left and right button]
to move the cursor to the 【Import configuration】, and use the [Enter key] to
import the configuration.
2. Show the configuration file import situation.
3. Use the [left and right button] to move the cursor to the 【Exit】,and use [Enter
key] to exit the configuration file screen.

IM01C82R01-01E 114
CHAPTER 16 SPECIFICATIONS
16.1 SIGNAL INPUT AND ALARM
Measurement input
Item Specifications
The number of simultaneously measuring up to 48 input channels
input channels
Measurement 1 second
period
Signal type DC current (I), DC voltage (V), thermocouple (TC), thermal
resistance (RTC), frequency (FR)
Input measurement range and the measurable range
Type Range Measurable range
Current 4-20mA 4.00mA ~ 20.00mA
0-20mA -20.00mA ~ 20.00mA
0-10mA -10.00mA ~ 10.00mA
Voltage 1-5V 1.000V ~ 5.000V
0-5V -5.000V ~ 5.000V
0-10V -10.00V ~ 10.00V
20mV -20.00mV ~ 20.00mV
100mV -100.00mV ~ 100.00mV
resistence 400Ω 0.0Ω ~ 400.0Ω
175Ω 0.0Ω ~ 175.0Ω
Digital signal FR 0Hz ~ 10000Hz
FR. 0.0Hz~3000.0Hz
RTD PT100 -200.0℃ ~ 650.0℃
PT100. -50℃ ~ 150.0℃
Cu50 -50.0℃ ~ 140.0℃
Cu53 -50.0℃ ~ 150.0℃
BA1 -100℃ ~ 600℃
BA2 -100℃ ~ 600℃
Thermocouple S -50℃ ~ 1768℃
R -50℃ ~ 1768℃
B 250℃ ~ 1820℃
K -50℃ ~ 1300℃
N -200℃ ~ 1300℃
E -100℃ ~ 1000℃
J -100℃ ~ 1000℃
T -200℃ ~ 400℃
WRE5-26 0℃ ~ 2310℃

115 IM01C82R01-01E
WRE3-25 0℃ ~ 2315℃
F1 600℃ ~ 2000℃
F2 600℃ ~ 2000℃
Measuring current: I = 0.25mA (resistance, heat resistance)
Thermocouple Choose the maximum, minimum, error flag or keeping the former
burnout displayed value
check current: About 2.5uA

Item Specifications
Thermocouple Using internal cold junction compensation, the compensation
cold junction accuracy :±0.5 ℃
compensation
Digital filter 0 ~ 9.9 seconds inertia filter
Vacuum
operation
General LOG curve is used in the whole range
algorithm
Sub-algorithm linear operation is used in every section

Alarm
Item Specifications
The number of Each channel has 4 alarm
alarms
alarm type High highalarm type , high alarm, I alarm , D alarm
Alarm delay 0-10s
time
Alarm output Alarm outputs to the internal relay
Internal relay number: 36 Or 24 ( If the channel number is more
than 36, it is 24, otherwise, it is 36)
Internal relay operations: OR operation
Display When alarm occurs, the corresponding screen displays the alarm
status; the status display section displays the alarm icon.
Alarm Display alarm log in the alarm list
information

IM01C82R01-01E 116
16.2 DISPLAY FUNCTION
Display
Item Specifications
Display * 10.4-inch TFT color LCD display (800× 600 points)
Display color 256 colors

Display Information
Item Specifications
Display channel Channels will be allocated to real-time curve and historical curve
group show and then display
group Number 8
the number of channels which an be allocated to each group
≤8
Real-time curve
display
Renewal rate 1 record interval
Waveform 1 pixel
width
Display method Display perpendicular axes with timeline (T) and the
measurement axis (Y)
Layout: horizontal, vertical
Div Curve cycle: recording interval × 60 div
Scale To display each channel (0% ~ 100%)
Digital Display Display measured values with digital
Update Rate 1 second
Bar graph Display measurement on the bar graph
display
Update Rate 1 second
ruler 0% ~ 100%
Historical curves Display stored data in the memory
show
Timeline can enlarge one times / 2 times / 4 times / 8 times / 16 times / 32
operation times
Alarm list Display up to 256 alarm records
display

117 IM01C82R01-01E
Other displayed information

Item Specifications
Channel-bit 7 Chinese characters or 15 letters(numbers)
Channel units Three Chinese characters 7 letters (numbers)
Status display Show the status of this instrument in the upper part of the display
section screen
Display content Searching status、 Display screen name, card status, alarm status,
SD card status, USB device status, circular display status, year,
month, day, hour, minute, seconds
Automatic At the specified time interval to switch display groups
switch of Show Time interval: 5 seconds
group
Language Simplified Chinese, English
System Show the number of input and output card channels, relay output
information channels, instrument additional features, firmware version,
display production date and released date

IM01C82R01-01E 118
16.3 DATA STORAGE FUNCTION
Setting
Item Specifications
External storage Back up the record data in the device
media
Media U Disk, SD card
Capacity Up to 4GB( U disk), 4 GB (SD card)
File format FAT32

File Name DAT+ "number" ( can edit )


Report Name CSV
Shift report: REP+ "number"( can edit)
Daily report: RPTD+“number”(can edit);
Monthly report: RPTM+“number”(can edit);
Annual report: RPTY+“number”(can edit)

Data type
Item Specifications
Data Size
Sampled data 2 bytes/ data values
Cumulative 8 bytes/ data values
amount
Data Format Binary Data
The sampling Equal to recording interval
period
Recording Always recording
method
Recording type Intermittent storage

119 IM01C82R01-01E
16.4 OTHER STANDARD FEATURES
Time-related functions
Item Specifications
Clock Can operate in 2000 ~ 2099
Clock Accuracy ±10ppm (0 ~ 50 ℃), does not include the delay caused by the
error when the power is on(less than 1 second )
Life of Clock About 10 years ( under room temperature)
battery

The type of characters which can be inputted


Item Specifications
Characters In Chinese, English (uppercase, lowercase), numbers, and
symbols (within limitations)

IM01C82R01-01E 120
16.5 CHOOSE ACCESSORIES
Alarm output relay
Item Specifications
Action Output relay contacts signal from the back panel of the terminal
when alarm occurs
Output points Normally open contact :
36 (input channel ≤ 36)
36 (input channels> 24)
Relay electric 250VAC(50/60Hz)/3A,250VDC/3A((load resistance))
shock rating
Output Type Normal open or closed
Relay Operation OR operation

RS232C/RS485
Item Specifications
Physic level RS-232、RS485(option)
Protocol MODBUS-RTU
Communication 1200/2400/4800/9600/19200/38400/57600 bps
rate
Byteswap 2-1 4-3、1-2 3-4、4-3 2-1、3-4 1-2
Data length 8bits
Stop bit 1bit
Parity bit 1bit
parity None/ odd/ even parity

24VDCPower Distribution
Item Specifications
Loop 6
Output voltage 24VDC±10%
Maximum output 60mADC(overload protection current: approximately 90mADC)
current
Allowed ≤ 750Ω
impedance

121 IM01C82R01-01E
USB Function
Item Specifications
USB port Compatible with USB2.0 protocol
Number of ports 1
Capacity Maximum: 4 GB
Power Supply 5V±10%,100mA
Devices can be U disk
connected

SD card function
Item Specifications
SD card port Compatible with SD protocol( NOT SDHC)
Number of ports 1
Capacity Maximum: 4 GB
Devices can be SD card
connected

Analog signal output


Item Specifications
Output Type 4-20mA
Number of Up to 8
output channels
Allowed ≤ 750Ω
impedance

IM01C82R01-01E 122
16.6 GENERAL SPECIFICATIONS
Frame
Item Specifications
Installation Embedded instrument panel installation (vertical plane)
Angle Allow for a maximum of 30 degrees after the tilt from horizontal
Plate thickness 2 ~ 10mm
of installation
Material Panels and Terminal Material: ABS plastic
Shell: Aluminum alloy
External 288(W)×288(H)×245(D)(D:the length from the installation panel
dimensions to terminal)
Weight About 9 Kg(cluding configuration)

Standard operating conditions


Item Specifications
Supply power 220VAC
Voltage
Power Frequency 50Hz
Ambient 0℃ ~ 50℃
temperature
Environment 20% ~ 85%RH( non-condensing)
Humidity
Warm-up Time 30 minutes after power is switched on
installation Indoor
location

Power
Item Specifications
Rated supply 220VAC
voltage
Allowed voltage 85VAC ~ 220VAC
range
Rated power 47-63Hz
frequency
Power ≤20W(including the matching function)
consumption

123 IM01C82R01-01E
Insulation
Item Specifications
Resistance Measurement input terminals:: 400V
voltage Between protective ground and measured terminal: 1000V
Grounding Grounding resistance ≤ 10Ω
resistance

Transportation and storage condition


Item Specifications
Temperature -10℃ ~ 60℃
Humidity 0% ~ 95%(non-condensing)
Shake-proof 10 ~ 55Hz,10m/S2,2 hours
Impact-proof Working:30/S2(3g),Transporting:100g

IM01C82R01-01E 124
Electrical characteristic

Signal type Range precision(25℃) Input impedance


4-20mA 4.00-20.00 ±0.2% ≤300
0-20mA 0.00-20.00 ±0.2% ≤300
0-10mA 0.00-10.00 ±0.2% ≤300
0-5V 0.000-5.000 ±0.2% 1M
1-5V 1.000-5.000 ±0.2% 1M
0-10V 0.000-10.000 ±0.2% 1M
0-20mV 0.00-20.00 ±0.2% 10M
0-100mV 0.00-100.00 ±0.2% 10M
400ohm 0.0-400.0 ±0.2% ---
175ohm 0.0-175.0 ±0.2% ---
Fr 0-10000 ±1Hz ---
Fr. 0.0-3000.0 ±0.1Hz ---
S -50-1768 ±2℃ 10M
R -50-1768 ±2℃ 10M
B 250-1820 ±2℃ 10M
K -200-1372 ±1℃ 10M
N -200-1300 ±1℃ 10M
E -200-1000 ±1℃ 10M
J 210-1200 ±1℃ 10M
T -200-400 ±1℃ 10M
WRE5-26 0-2310 ±2℃ 10M
WRE3-25 0-2310 ±2℃ 10M
F1 700-2000 ±2℃ 10M
F2 700-2000 ±2℃ 10M
Pt100 -200.0-650.0 ±0.4℃ ---
Pt100. -50.0-150.0 ±0.2℃ ---
Cu50 -50.0-150.0 ±0.4℃ ---
Cu53 -50.0-150.0 ±0.4℃ ---
BA1 -200.0-650.0 ±0.4℃ ---
BA2 -200.0-650.0 ±0.4℃ ---

125 IM01C82R01-01E
Frenquency input board
Item Specification
Channel number 8 channel/ per board
Low level Minimum:-5V,Maximun2V
High level Minimum:4V,Maximun26V
Duty ratio 10%-90%
Drive current Minimum:1.5mA

Anolog output board


Item Specification
Output signal 4-20mA
Resolution ratio 12 bit
load 750ohm
precison 0.2%F.S
Response time 1second
Temperature drift 1uA/℃
coefficient

Other standard
Item Specification
Data storage About 10 year
limit
Safty agency EN61010-1:2001
approval
Electromagnetic EN61326-1:2006
compatibility EN61000-3-2:2006
EN61000-3-3:2008
Protection grade Front panel IP50,complete machine:IP40

IM01C82R01-01E 126
16.7 EXTERNAL DIMENSIONS
Unit: mm

127 IM01C82R01-01E
Dimensions of Instrument installation
Unit: mm
Hole size of Instrument

Hole size of container

IM01C82R01-01E 128