Atellica VTLi Patient-side
Immunoassay Analyzer
A vital leap forward
in cardiac testing
Speed meets accuracy where it matters most―
the wait is over for hs-cTnI at the point of care.
siemens-healthineers.com/vtli
Atellica® VTLi Analyzer Specifications
System description Patient-side immunoassay analyzer
System menu Parameter High-sensitivity cardiac troponin I (hs-cTnI)
BNP*
Measurement range LoD–1250 ng/L†
LoD plasma 1.2 ng/L†
LoD whole blood 1.6 ng/L†
Sample types Whole blood Venous (lithium heparin†), capillary
Plasma Venous (lithium heparin†)
Sample size 30–100 µL
Time to result ~8 minutes
Calibration Reagent cartridges contain calibration parameters valid on RFiD tag until expiry.
Optical system calibration is automatically performed during every test
Quality control Sero PATHONORM™ Cardiac Acute Liq
Integrated barcode scanner For patient ID, operator ID, and bottled QC;1D (linear) and 2D bar-code symbologies
External interfaces POCT1A
Docking station Input 100–240 VAC, 50–60 Hz, 660 mA
power requirements Output 19 VDC, 1570 mA
Classification Class II
Atellica VTLi Analyzer Specifications (continued)
Operation on battery ~60 tests on a fully charged battery
Safety IEC 61010-1, IEC 61010-2-101
EMC IEC 61326-1, IEC 61326-2-6
Data security Patient data anonymization
TLS 1.2-encrypted end-to-end communication using client and server certificates
Configurable level of user authentication
No storage of passwords, no hard-coded passwords
No user access to operating system
Hardware Operating system Microsoft WEC2013
Processor NXP i.MX 6SoloX
Memory 512 MB RAM
WI-FI 802.11 b/g/n/ac
2.4 GHz
WI-FI security WPA/WPA2 Personal (pre-shared key)
Data capacity Patient/QA tests 100 tests
Operators Not limited
Reagent cartridge Storage 2–8ºC
Shelf life 7 months
Size of box (W x D x H) 152 x 95 x 167 mm
Cartridges per box 24
System dimensions Analyzer Length 25 cm
Height 5.2 cm
Width 8.5 cm
Weight (incl. battery) 780 g
Weight 580 g
Display 4.3-inch color LCD touchscreen with backlight
Docking station Length 29 cm
Height 6 cm
Width 10 cm
Weight 460 g
Environmental requirements Operating temperature 5–27°C
Humidity 20–80% (noncondensing)
Air pressure 70–110 kPa
Storage temperature –20 to 55ºC
Storage humidity 5–85% (noncondensing)
Atellica and all associated marks are trademarks of Siemens Healthcare
Diagnostics Inc., or its affiliates. All other trademarks and brands are the
property of their respective owners.
Not available for sale in the U.S. Product availability will vary from country to
country and is subject to varying regulatory requirements. Please contact
your local representative for availability.
Siemens Healthineers Headquarters Legal Manufacturer
Siemens Healthineers Headquarters Local Contact Information
Siemens Healthcare GmbH Diagnostics Inc.
Siemens Healthcare GmbH Siemens Healthcare Diagnostics Inc.
Henkestr. 127 Laboratory127
Henkestr. Diagnostics Point of Care Diagnostics
91052 Erlangen, Germany 511 Benedict
91052 Avenue
Erlangen, Germany 2 Edgewater Drive
Phone: +49 9131 84-0 Tarrytown,
Phone: +49NY 10591-5005
9131 84-0 Norwood, MA 02062-4637
siemens-healthineers.com USA
siemens-healthineers.com USA
Phone: +1 914-631-8000 Phone: +1 781-269-3000
Published by Siemens Healthcare Diagnostics Inc. · Order No. 40-21-DX-525-76 · 03-2021 · © Siemens Healthcare Diagnostics Inc., 2021