BWMS(PURIMAR) A/S TROUBLESHOOTING DOCUMENT
[TRO] – [Action] – [BWAST-TM-003]
# TRO Flow Rate Setting
※ In the cases below, please adjust the water flow rate again.
1) Air pressure for the sampling pump is adjusted.
2) The drain needle valve is adjusted.
3) The water regulator is cleaned, or replaced or its setting is changed.
1. Flow rate setting must be conducted while the ballast pump and the TRO sensor are
normally operating.
→
①Press [SERVICE] button. ②Wait until [HOLD] is displayed.
Then press [MODE] button.
→
③Press [▲] or [▼] button once. ④Check the counted number.
※Water sampling starts for flow check. HI: 1~5
Pulse count is displaying. GOOD: 6~12
LO: 13~
2. If the flow rate is [HI(1~5)] or [LO(13~ )], please adjust the regulator.
→
①Remove the rubber cap from the regulator. ②Unscrew the nut with wrench little bit.
Be careful not to loosen up the regulator.
1/2 Version. 0
BWMS(PURIMAR) A/S TROUBLESHOOTING DOCUMENT
[TRO] – [Action] – [BWAST-TM-003]
③If the message is HI, turn the regulator ③If the message is LO, turn the regulator
counterclockwise. (Turn quarter by quarter.) clockwise. (Turn quarter by quarter.)
→ →
④After adjusting the regulator, press [▲] or [▼] ⑤Tighten the locking nut.
button. ⑥Cover the rubber cap.
※Repeat the flow control steps until the flowrate
is 6~12. (8~10 is preferable.)
⑦Press [ ] button to return to normal operation.